Neometrix is proud to support our Nation for managing Medical Oxygen requirements during the Covid 19 pandemic.
We are pleased to announce that we are supplying PSA Medical Oxygen Generation plants to hospitals all over India through a running contract with ministry of health and family welfare, Government of India.
These NAMO™ series plants are for different hospitals of India with different capacities as per requirement of hospitals like 100 LPM, 200 LPM, 500 LPM, 1000 LPM, 1500 LPM, 2000 LPM, 2500LPM & 3000LPM.
Neometrix has enhanced its manufacturing capabilities to supply any of these plants within 4-6 weeks.
Neometrix is providing plants with purity as per Indian Medical standard i.e. 93±3.
Pressure swing adsorption processes utilize the fact that under high pressure, gases tend to be attracted to solid surfaces, or "adsorbed". The higher the pressure, the more gas is adsorbed. When the pressure is reduced, the gas is released, or desorbed.
PSA processes are used to separate gases in a mixture because different gases tend to be attracted to different solid surfaces more or less strongly.
PSA Medical Oxygen Plant MED Jaamu Kashmir-2400 LPM.pptx
1.
2.
3. Neometrix is proud to support our Nation for managing Medical Oxygen requirements
during the Covid 19 pandemic.
We are pleased to announce that we are supplying PSA Medical Oxygen Generation plants to
hospitals all over India through a running contract with ministry of health and family
welfare, Government of India.
These NAMO™ series plants are for different hospitals of India with different capacities as
per requirement of hospitals like 100 LPM, 200 LPM, 500 LPM, 1000 LPM, 1500 LPM, 2000
LPM, 2500LPM & 3000LPM.
Neometrix has enhanced its manufacturing capabilities to supply any of these plants within
4-6 weeks.
Neometrix is providing plants with purity as per Indian Medical standard i.e. 93±3.
Pressure swing adsorption processes utilize the fact that under high pressure, gases tend to
be attracted to solid surfaces, or "adsorbed". The higher the pressure, the more gas is
adsorbed. When the pressure is reduced, the gas is released, or desorbed.
PSA processes are used to separate gases in a mixture because different gases tend to be
attracted to different solid surfaces more or less strongly.
4.
5. The Installed Oxgene Generation Plant Is Psa Based Oxygen Generation System With Below
Major Specifications-
Plant Capacity : 2400LPM.
Oxygen Purity : 93±3% Conforming Ip 2010 Protocol
Oxygen Pressure : 4-4.5 Kg/Cm2
Oxygen Storage Capacity : 4000 Ltr
Autmatic Changeover Panel For Medical Cylinder And PSA Oxygen Lines
Iot For Remote Monitoring And Data Recording
Inbulit Safety Alarms And Visual Popup Indications
Dew Point: - 40 °C To - 73 °C
Operation: Plc Based Fully Automatic & Programmable Operation Suitable For
Unattended Operation With Auto Alarm Generation And Safety Interlocks
Outlet Connection: The Outlet Of Oxygen Generator Will Be Connected Directly To
Hospital Oxygen Pipeline For Consumption Of Patients Or To Cylinder Manifold To Fill
Cylinders If Applicable
Molecular Sieves Sodium/Silver Based
Low Pressure Interlock On Oxygen Storage Tank
Dust & Bacterial Filter At Outlet Of Oxygen Surge Vessel With Filtering Element
Minimum 3 Years Warranty Period
Power Backup For Plc Panel Provided For 30min Minimum.
6.
7. APPLICATION FOR MEDICAL OXYGEN SUPPLY FOR
COVID-19 PATIENT AND OTHER MEDICAL SUPPLIES
SUCH AS ICU AND GENERAL OXYGEN WARDS.
8.
9. Plant Capacity : 2400LPM
Oxygen Purity : 93±3% Conforming Ip 2010 Protocol
Oxygen Pressure : 4-4.5 Kg/Cm2
Oxygen Storage Capacity : 4000Ltr
Autmatic Changeover Panel For Medical Cylinder And PSA Oxygen Lines
Iot For Remote Monitoring And Data Recording
Inbulit Safety Alarms And Visual Popup Indications
Dew Point: - 40 °C To - 73 °C
Operation: Plc Based Fully Automatic & Programmable Operation Suitable
For Unattended Operation With Auto Alarm Generation And Safety
Interlocks
Outlet Connection: The Outlet Of Oxygen Generator Will Be Connected
Directly To Hospital Oxygen Pipeline For Consumption Of Patients Or To
Cylinder Manifold To Fill Cylinders If Applicable
Molecular Sieves Sodium/Silver Based.
Low Pressure Interlock On Oxygen Storage Tank
Dust & Bacterial Filter At Outlet Of Oxygen Surge Vessel With Filtering
Element
Power Backup For Plc Panel Provided For 30min Minimum