Cyber Physical Systems: Error Verification and Prediction in
Smart Medical Devices
Stephen O. Maitanmi1
*, Sunday A. Idowu2
, Olawale J. Omotosho3
,
Olusegun A. Ojesanmi4
123
Department of Computer Science, Babcock University, Ilisan Remo, Ogun State, Nigeria
4
Department of Computer Science, Federal University of Agriculture, Abeokuta, Ogun State, Nigeria
Abstract: Embedded software is a piece of software that is built in a system or hardware component to achieve
an objective. Cyber-Physical Systems (CPS) are integrations of computation with physical processes which are
monitored and controlled by the embedded systems. CPS has positively affected a number of application areas
which include communication, consumer energy, infrastructure, healthcare, manufacturing, military, robotics
and transportation. This research paper focuses on CPS with particular application interest on error prediction
in smart medical devices SMDs. SMDs are used for healthcare services by medical personnels in such a way
that they have to interact with the patients in one form or the other. A number of research works has focused on
errors arising from the use of SMDs. Errors arise from readings of the SMDs as a result of usability challenges
from embedded software leading to malfunctioning of the devices. This research paper is aimed at providing a
model for error verification and prediction in the use of SMDs using Euler’s method of representation with the
aims of establishing safety and reliability of the use of SMDs with possible reduction in risk of accidents. In
order to achieve the stated objective, a mathematical model using Euler’s method was adopted for its fast
convergence rate and simplicity for error prediction. Input data was provided through a critical incident
analysis of online database which provide readings from medical experts. These readings were compared to the
standard world benchmarks. The difference between the readings and the standard benchmark validates the
existence of errors. Due to the complexity of the model, an algorithm was developed to obtain an optimal
solution of P1-P5 within an acceptable threshold runtime. An implementation was carried out using Java
programming language because of its robustness and cross platform advantages which provides an efficient
error estimate and result analysis. The analysis showed that the optimum performances of the SMDs were
hindered by errors. The results generated from the use of thermometer for the diagnosis of Malaria predicted
98.1% accuracy in measurement, Upper respiratory tract infection predicted 99.3% accuracy in measurement,
Tonsilitis predicted 99.6% accuracy in measurement, Severe head injury predicted 99.9% accuracy in
measurement and Septicaemia predicted 99.9% accuracy.
These results show that Malaria generated the highest error in the use of thermometer. In conclusion,
CPS generally is an active area of research with a number of application domains in particular health
care systems. This particular work has provided a model for error verification in the use of SMDs.
Keywords: Cyber Physical Systems, Embedded system, SMDs, Euler’s method;
I. Introduction
The availability of the Internet alongside its advantages has led to the pervasive use of networking and
information technology (IT) across all sectors [1]. Interestingly, these technologies are combined with elements
of the physical world such as machines, devices and structures to create smart or intelligent systems that offer
increased usefulness, productivity, safety and speed which enabled previous challenging functions.
Embedded software is a piece of software that is hidden in a system or hardware components written
specifically in a coordinated fashion to achieve an objective. Embedded system is not only one of the most
important fields for current computer-based applications; it is also one of the most challenging fields of
computer science or more precisely, Software Engineering [2]. Embedded systems are used in many areas
ranging from vehicles and mobile phones to washing machines and printers to increase and enhance
productivity.
Cyber-Physical Systems (CPS) are integrations of computation with physical processes which are
monitored and controlled by the embedded systems usually with feedback loops where physical processes affect
computations [3]. All CPS have computational processes that interact with physical components. These
computational processes can either be simple or complex instructional processes. For example, simple processes
include water heater or cutting machine while the complex processes include vehicle, aircraft systems, oil
refineries among others. The computational and physical processes of such systems are highly interconnected
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 64 | Page
and coordinated to function together. The following diagram in Fig.1 sheds more light into the meaning of CPS
as regards integration of computation with physical processes.
Fig. 1. Cyber-Physical Systems Concept
CPS concepts as illustrated in Fig.1 shows that cyberspace is the Internet which is invisible in
information systems and it synergizes the physical operations that we carry out daily either in Computer
Science, Medicine or any other professional fields. The importance of the Internet cannot be overemphasized, it
harnesses the operations that were formerly carried out independently to a single, faster, robust and more
coordinated operation. To further illustrate the usefulness of CPS, a three months old baby can be monitored
from an independent room through a wearable body wire while a monitor is placed in the parent’s room with the
aid of sensors which is a device that detects a change in a physical stimulus and turns it into a signal that is
measurable. The movements within the defined locations are known with the aid of the sensors embedded as
part of the software.
Error verification is defined as investigating a device or software for the existence of error [4].
Meanwhile, the Institute of Medicine Patient Safety described patient safety as a freedom from accidental injury
due to medical care, software error or medical errors. It has increasingly been regarded as the key element of
quality in the last two decades [5]. The process began in 1999 with Harvard Medical Practice study that showed
adverse events in 3.7% of hospitalization and errors which could be related to 27.6% of adverse events [6]. In
2001, the Institute of Medicine published a study tagged ‘to err is human: building a safer health system’ [7].
This study estimated that preventable medical errors are responsible for between 44, 000 and 98, 000 deaths
annually in United States.
A more heart touching news from the Journal of Patient Safety reports that the numbers may be much
higher which was put between 210,000 and 440,000 annually for patients who visit the hospital. Such patients
are likely to suffer some type of preventable harm that contributes to their death [5]. This report would make
medical errors the third highest cause of death in America after heart disease and cancer.
Lack of patient safety measures make the care unsafe and produce opportunities for medical errors to
occur [7] while [8] stated that 2,000,000 defective devices were recalled in the last decades due to deficiencies
in outputs such examples include unmanned rocket launched in June 1998 called Ariane 5. Ariane 5 was a
project that lasted for ten years and cost $7 billion exploded after 37 seconds into the flight due to software
errors; code ported from Ariane 4 rocket did not work [9]. Similarly, on land transportation; a Honda CR-V
that cost $1.5M in US was recalled to update the software that controls its automatic transmissions. In similar
manner, a Toyota Hybrid was recalled because of software deficiency that made it enters fail-safe mode that
shuts down engines frequently [9].
Similarly, in Nigeria health services, [10] affirmed in a conference proceeding that seventy percent of
deaths in third world countries occur before age five and are caused by factors that are controllable.
The SMDs studied include thermometer, insulin pump, infusion pump and sphygmomanometer. They
are briefly defined as follows; thermometer measures the body temperature, insulin pump maintains the normal
body glucose level, infusion pump transfers fluids, medication or nutrients into a patient’s circulatory system
while sphygmomanometer measure the blood pressure of the body [11].
The objective of this paper was to design a model for error verification and prediction of SMDs in CPS.
In fulfilling this mandate, the research carried out a critical evaluation of some frequently used SMDs to verify
and predict the existence of errors.
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 65 | Page
The paper layout
The rest of the paper is organized as follows: apart from the introduction section, subsequent sections
handle the methodology of the research, results, discussions and conclusion.
II. Methodology
In order to achieve the stated objective which was to develop a model for error verification and
prediction of SMDs in CPS.; a critical incident analysis approach was adopted to look into the use of the SMDs
because of the sensitivity of the devices. A mathematical model using Euler’s method was employed for its fast
convergence rate and simplicity. Secondary input data were retrieved from an online database of Medline, Web
of Science, Health Technology Assessments and Health and Science Care Information Center which provides
database of readings from medical experts. These retrieved data were further compared with the standard world
benchmarks from literatures. The following sections consist of the breakdown of each SMD analysed alongside
its standard and practicable benchmarks. This input data was considered to achieve accuracy and correctness in
the research and experimentation process of the proposed model.
Due to the complexity of the model, an algorithm was developed to obtain an optimal solution within
an acceptable threshold runtime. Implementation was carried out using Java programming language because of
its robustness and cross platform advantages which provides an efficient error estimate and result analysis.
The SMDs analysed in this research paper include thermometer, sphygmomanometer, infusion pump
and insulin pump. These SMDs were chosen because of their relevance in the implementation of CPS as regards
technological advancement to health care systems.
Thermometer for both Benchmark and Readings
The following represent the analysis of digital thermometer:
Table 1. Temperature readings and standard benchmark (0
C)
Source : Appendix A-C
Where URTI= Upper respiratory tract infection.
Sphygmomanometer (mmHg) for both Benchmark and Readings
Symptoms Benchmark (P) Time(t) Readings P(t) Error(t)
Malaria 37.9 0 37.9 0
38.5 10 36.0 2.5
38.7 20 36.1 2.6
38.9 30 36.2 2.7
39.5 40 36.3 3.2
39.6 50 36.3 3.3
URTI 37.7 0 37.7 0
38.5 10 37.7 0.8
39.0 20 37.8 1.2
39.5 30 37.9 1.6
39.8 40 38.0 1.8
40.0 50 38.0 2.0
Tonsilitis 36.4 0 36.4 0
37.8 10 36.5 0.4
39.2 20 37.0 2.2
39.9 30 37.6 2.3
40.0 40 38.0 2.0
40.5 50 38.5 2.0
Severe Head
Injury
39.8 0 39.8 0
39.9 10 39.8 0.1
40.5 20 39.9 0.6
43.5 30 40.0 3.5
45.5 40 40.0 5.5
47.5 50 40.1 7.4
Septicaemia 39.9 0 39.9 0
40.0 10 39.0 1.0
45.3 20 39.5 5.8
47.3 30 39.9 7.4
47.8 40 40.0 7.8
48.0 50 40.0 8.0
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 66 | Page
Table 2. Sphygmomanometer readings and standard Benchmark (mmhg)
Source: Appendix D
Systolic Diastolic
BM/RD/error BM/RD/error BM/RD/error BM/RD/error BM/RD/error
≤84 85-89 90-99 100-109 110-119
≤129 2 1 1 3 2 1 4 3 1 5 4 1 6 5 1
<130-139 2 2 0 3 2 1 4 3 1 5 4 1 6 5 1
<140-159 3 3 0 3 3 0 4 3 1 5 4 1 6 5 1
<160-179 5 4 1 5 4 1 4 3 1 5 4 1 6 5 1
<180-209 5 5 0 5 5 0 5 4 1 6 5 1 6 5 1
≥210 7 6 1 6 6 0 5 4 1 7 6 1 7 6 1
Where N=161, BM= benchmark and RD= readings
Infusion Pump for both Benchmark and Readings
Table 3. Infusion Pump readings and standard Benchmark (Msl/hr.)
Source: [13]
Benchmark Readings Error Time (Mins)
Adult -100mls/hr 80mls/hr 10 60
500mls/hr 480mls/hr 20 60
500mls/hr 490mls/hr 10 60
4 litres/day 4litres/day 0 1,440
500mls/12rs 480mls/12rs 20 720
500mls/day 495mls/day 5 1,440
Children 300mls/day 280mls/day 20 1,440
10mls/hr 10mls/hr 0 60
1 liter/day 1 litre/day 0 1,440
2 litres/day 2 litres/day 0 1,440
Insulin Pump for both Benchmark and Readings
Table 4. Insulin Pump readings and standard Benchmark
1 unit of insulin for every 15g Carbohydrates taken. Insulin Pump Workbook.
Source: [14]
Time Benchmark
Blood Glucose
Amount of
Carbohydrates
Benchmark
Insulin intake
Readings
Insulin intake
Errors
Day 1
7:30am 5.7 60g 4 8 4
1:30pm 11.2 50g 3.3 6.6 3.3
6:00pm 7.2 72g 5 10 5
10:00pm 8
Day 2
7:30am 6.8 60g 4 8 4
1:30pm 12.3 50g 3.3 6.6 3.3
6:00pm 7.1 90g 6 12 6
10:00pm 11.3
2.1 Relationship between Model and Thermometer table
Performance at optimum level is represented in Table 1 as the Benchmark of temperature taken when
error is not involved (P) which is equivalent to 100%.
Performance with respect to time is represented in Table 1 taking note of the time in minutes when
temperatures were taken P(t). The recorded time were decimated to reduce the magnitudes of error such as 0.1,
0.2, 0.3 and so on for Malaria. These values are not constant for other ailments.
Error as shown in Table 1 gives the difference between benchmark and actual readings which was also
decimated to reduce error value ther eby realizing (E) of Malaria to be 0.25, 0.26, 0.27, 0.32 and 0.33
P1-P5 is the performance level of the SMDs and the model with respect to the introduced error.
2.2 Mathematical Modeling of Error Prediction in the use of Thermometer
P  Performance at optimum level
 P t  Performance with respect to t
 E t  Error introduced at any time t
    (1)P t P E t                              
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 67 | Page
(1) gives the difference between the benchmark and the actual reading which is a function of time.
   
1
( )P t E t P t
Where  
1
P t is the change in performance which is proportional to the product of performance and the error at
any time t
   
1
( )P t K E t P t 
   
1
( )P t E t P t 
substituting ( )P t in (1) gives
   
1
( )[ ] ( 2 )P t E t P E t                           
By Using Euler’s iteration method to project a possible outcome when variables such as error, and
time are used as input
1
( ) (3 )
1
P P h P
n n n
                            

Where n= 0-4 (five variables from temperature readings on malaria from Table 1)
While 0E = 0.25, 1E =0.26, 2E = 0.27, 3E =0.32 and 4E =0.33 decimated values from temperature readings in
Table 1
h = dt and dt =0.1 from model time used from Table 1
From (2) at optimum level when 0t  and 0E 
(0 ) 0P P from (2) = 1 or 100% when there is no error, meaning performance at best level
Starting Iteration
From (2)
1
( ) ( 0 .2 5*0 .1)[1 0 .2 5*0 .1] 0 .0 2 5*0 .9 7 5 ) 0 .0 2 40P       
1
( ) 1 ( 0 .1)( 0 .0 2 4 ) 1 0 .0 0 2 4 0 .9 9 7 61 0 0P P h P        or 99.8%
1
( ) ( 0 .2 6*0 .1)[1 0 .2 6*0 .1] 0 .0 2 6*0 .9 7 4 ) 0 .0 2 51P       
1
( ) 0 .9 9 8 ( 0 .1)( 0 .0 2 5 ) 0 .9 9 8 0 .0 0 2 5 0 .9 9 52 1 1P P h P        or 99.5%
1
( ) ( 0 .2 7 *0 .1)[1 0 .2 7 *0 .1] 0 .0 2 7 *0 .9 7 3 ) 0 .0 2 62P       
1
( ) 0 .9 9 5 ( 0 .1)( 0 .0 2 6 ) 0 .9 9 5 0 .0 0 2 6 0 .9 9 23 2 2P P h P        or 99.2%
1
( ) ( 0 .3 2*0 .1)[1 0 .3 2*0 .1] 0 .0 3 2*0 .9 6 8 ) 0 .0 3 13P       
1
( ) 0 .9 9 2 ( 0 .1)( 0 .0 3 1) 0 .9 9 2 0 .0 3 1 0 .9 8 94 3 3P P h P        or 98.9%
1
( ) (1 .3 3*0 .1)[1 0 .3 3*0 .1] 0 .0 3 3*0 .9 6 7 ) 0 .0 3 1 94P       
1
( ) 0 .9 8 9 ( 0 .1)( 0 .0 3 1 9 ) 0 .9 8 9 0 .0 0 3 1 9 0 .9 8 65 4 4P P h P        or 98.6%
2.3 Development of Algorithm
The algorithm is based on the following formulae:
   
1
( )[ ] (4)P t E t P E t                             
1
( ) (5 )
1
P P h P
n n n
                               

Declarations
Step 1: Initialize fractional variables (double) , ,h t E
Step 2: Initialize whole numbers (Int) ,P n
Step 3: let all values of n be between 0 to 4
Step 4: Let E0=0.25, E1=0.26, E2=0.27, E3=0.32 and E4=1.85
Step 5: Let 0.1 be stored in t as fractional number
Step 6: Let 1 be stored in P as whole number
Step 7: Let 0.1 be stored in h as fractional number
Starting the Euler’s iterations
Step 8: making reference to (4) above where 1
( ) ( )[ ( )]0 0 0P E t P E t  
Find
1
( )0P
to
1
( ) 4P
Step 9: starting iteration where the values of n= 0 to 4 starting with n=0 to begin
where K is assumed to be 1
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 68 | Page
1
( )1 0 0P P h P 
store the result in 1P
Step 10: Use the value obtained above in step 8 on
1
1
( P ) to find 1
( )2 1 1P P h P  and store the result in 2P
Step 11: Use the value obtained above in step 8 on
1
( ) 2P
to find
1
( )3 2 2P P h P 
and store the result in 3P
Step 12: Use the value obtained above in step 8 on
1
( )3P to find 1
( )4 3 3P P h P 
and store the result in 4P
Step 13: Use the value obtained above in step 8 on
1
( ) 4P
to find
1
( )5 4 4P P h P 
and store the
value in 5P
Step 14: Display the values of
, , , ,1 2 3 4 5P P P P P
III. Results & Discussions
Table 5 Error Prediction in the use of Thermometer for the diagnosis of Malaria
Table 5. shows the output of the error prediction that resulted in the use of thermometer for the diagnosis of
Table 5 shows the output of the error prediction that resulted in the use of thermometer for the
diagnosis of patient suspected to be having malaria fever. The variables displayed include E= error rate which
are values of 0.25, 0.26, 0.27, 0.32 and 0.33 taken from Table 1, it appears that the higher the values, the weaker
the functionality of the SMDs in terms of precision. P% refers to percentage analysis of the performance of the
predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1and constant as
discussed in the model development stages. P = performance, meaning maximum performance when there is no
error. The first output on column 1 therefore shows that the prediction generated 98.1% accuracy in
measurement. It appears also that the percentage error increases as the error in prediction increases.
Consequently, the rates of error increases down the column. The last column shows the highest error
rate with low performance. The performance value was 89.8%. From this value it is evidenced that performance
was affected.
Table 6. Error Prediction in the use of Thermometer for the diagnosis of Urinary Tract Infection (URTI)
Table 6 shows the output of the error prediction that resulted in the use of thermometer for the
diagnosis of patient suspected to be having URTI fever. The variables displayed include E= error rate which are
values of 0.08, 0.12, 0.16, 0.18 and 0.2 taken from Table 1, the higher the values, the weaker the functionality of
the SMDs in terms of correctness in measurement. P% refers to percentage analysis of the performance of the
predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1 and constant as
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 69 | Page
discussed in the model development stages. P = performance, meaning maximum performance when there is no
error. The first output on column 1 therefore shows that the prediction generated 99.3% accuracy in
measurement. It appears also that the percentage error increases as the error in prediction increases.
Therefore, the rates of error increases down the column. The last column shows the highest
involvement of errors. The performance value was 93.8%.From this value it is evidenced that performance was
affected.
Table 7. Error Prediction in the use of Thermometer for the diagnosis of Tonsilitis
Table 7 shows the output of the error prediction that resulted in the use of thermometer for the
diagnosis of patient suspected to be having Tonsilitis fever. The variables displayed include E= error rate which
are values of 0.04, 0.22, 0.23, 0.2 and 0.2 taken from Table 1, the higher the values, the weaker the functionality
of the SMDs in terms of correctness in measurement. P% refers to percentage analysis of the performance of the
predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1 and constant as
discussed in the model development stages. P = performance (benchmark), meaning maximum performance
when there is no error. The first output on column 1 shows that the prediction generated 99.6% accuracy in
measurement. It appears also that the percentage error increases with respect to increase in error.
Therefore, the rates of error increases down the column. The performance value was 92.9%. From this value it is
evidenced that performance was affected.
Table 8. Error Prediction in the use of Thermometer for the diagnosis of Severe Head Injury
Table 8 shows the output of the error prediction that resulted in the use of thermometer for the
diagnosis of patient suspected to be having severe head injury fever. The variables displayed include E= error
rate which are values of 0.01, 0.06, 0.35, 0.55 and 0.74 taken from Table 1, the higher the values, the weaker the
functionality of the SMDs in terms of correctness in measurement. P% refers to percentage analysis of the
performance of the predicted error, the higher the error, the lower the performance of the SMDs, t=time which is
0.1 and constant as discussed in the model development stages. P = performance (benchmark), meaning
maximum performance when there is no error. The first output on column 1 therefore shows that the prediction
generated 99.9% accuracy in measurement.
Therefore, the rates of error increases down the column. The last column shows the highest
involvement of errors. The performance value was 92.7%. It is evidenced from this value that performance was
affected.
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 70 | Page
Table 9. Error Prediction in the use of Thermometer for the diagnosis of Septicaemia
Table 9 shows the output of the error prediction that resulted in the use of thermometer for the
diagnosis of patient suspected to be having Septicaemia fever. The variables displayed include E= error rate
which are values of 0.01, 0.06, 0.35, 0.55 and 0.74 taken from Table 1, it appears that the higher the values, the
weaker the functionality of the SMDs in terms of precision. P% refers to percentage analysis of the performance
of the predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1 and
constant as discussed in the model development stages. P = performance (benchmark), meaning maximum
performance when there is no error. The first output on column 1 therefore shows that the prediction generated
99.9% accuracy in measurement. It appears also that the percentage error increases as the error in prediction
increases.
In addition, the rates of error increases down the column. The last column shows the highest error rate
with low performance. The performance value was 92.2%. It is therefore evidenced that performance was
affected.
IV. Conclusion
In conclusion, CPS generally is an active area of research with a number of evolving application
domains in particular, health care systems. The research paper investigated a number of SMDs and discovered
error in its applications. Based on these findings, a model was developed to guide medical personnel and
individual users in the appropriate use of SMDs.`
Acknowledgements
I will like to appreciate Mr. Semiu Ayinde of the department of Basic Sciences, Babcock University,
the Head of Department of Radiology, Babcock University Teaching Hospital (BUTH) Dr. Yusuf, Adebola and
Dr. Ajiboye, Bimpe who contributed immensely to the completion of this research paper.
References
[1]. Workshop Report."Foundations for Innovation in Cyber-Physical Systems Available:http://events.energetics.com /NIST-
CPSWorkshop/downloads.html,2013 [Nov. 20, 2014].
[2]. San Diego State University ."Introduction to computers". Information technology and Teachers Education International Conference.
Available: edweb.sdsu.edu/people/marshal/publication.html, 2009. [June 12, 2014]
[3]. E. A. Lee "Cyber physical systems: Design Challenges" Technical Report, EECS Department, University of California, Berkeley,
January (CPS Steering Group), 2008
[4]. D. R. Wallace, and R. U. Fujii. :Software verification and validation: an overview" Software, IEEE,Vol.6, no.3,pp.10- 17, 1989
[5]. T. J. John "A New Evidence-Based Estimate of Patient Harms Associated with Hospital Care" Journal of Patient Safety Volume 9,
Number 3. 2013
[6]. NAP Committee on Quality of Health Care in America IoM. ‘To err is human: building a Safer Health System’ Washington, D.C.:
National Academy Press., 2001
[7]. M. Soares, J. I. F. Salluh, and F. A. Bozza, Current definitions of patient safety. In: Chiche, J. D, Moreno R, Putensen C, Rhodes A,
editors. Patient Safety and Quality of Care in Intensive Care Medicine. Berlin:Medizinisch WissenschaftlicheVerlagsgesellschaft; p.
9-17., 2009
[8]. M. Paolo Formal verification of Medical user interface software in PVS, Queen Mary University of London, 2014.
[9]. K. Venkatasubramanian, "Challenges and Research directions in Medical Cyber Physical Systems" 2013.
[10]. D. C Stockwell, and A. D. Slonim. Quality and safety in the Intensive Care Unit. J Intensive Care Med 2006 Jul; 21(4):199-210.
2006
[11]. O. Akpabio, An address by the Executive Governor of Akwa Ibom state, his Excellency, Chief Godswill on the Occasion of the
National Health Conference 2009 (NHC 2009) Available:
http://www,unilorin.edu.ng/publications/sakamj/Technical%20Reports%20%20Saka%20MJ/NHC%202009 %20Proceedings.pdf
[May 28, 2014].
[12]. WHO Environmental Health Criteria 232, Static fields, 2006. Available www.icnirp.org/eb/publications/article [May 12, 2014]
[13]. FDA .Infusion Pump Improvement Initiative Center for Devices and Radiological Health U.S. Food and Drug Administration
2010.
[14]. E. Joan, and L. Helen "Insulin Pump Workbook"(n.d) Available: www.diabetes-
education.net/pdf/resources/insulin_pump_workbook.pdf [Jan. 12, 2015].
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 71 | Page
Appendix A Temperature Measurement
Fever in adults
If you or a family member has a fever, it means your body temperature is above normal.
Around 37ºC is normal
A digital thermometer is the best type to use to get an accurate temperature reading.
A fever is usually a normal response of your immune system to a virus or bacterial infection. Most healthy
adults can tolerate a fever well.
Fever ranges and symptoms
38–38.9°C – mild fever
With a mild fever you might have flushed cheeks, feel a little lethargic, and be warm to touch. You will
generally be able to carry out normal daily activities.
39–39.9°C – high fever
With a high fever you may not feel well enough to go to work, you may have aches and pains, and
you’ll feel hot to touch.
40°C or higher – very high fever
With a very high fever you will usually want to stay in bed or be inactive – you won’t feel well enough
to carry out normal activities. You may have lost your appetite. You’ll feel hot to touch.
When to see your doctor
Some mild diseases produce very high fevers – and severe illnesses can produce mild fever. Therefore,
when considering what medical attention you need, it’s important to look at other symptoms and how unwell
you feel.
You should see your doctor if you or a family member:
 has a very high fever (over 40ºC)
 is still feverish after three days of home treatment, or seems to be getting sicker
 is shivering or shaking uncontrollably, or has chattering teeth
 has a severe headache that doesn’t get better after taking painkillers
Appendix A Cont'd
 is having trouble breathing
 is getting confused or is unusually drowsy
 has recently travelled overseas.
When it’s urgent
See your doctor or go to the Emergency Department immediately if you notice the following symptoms
(along with a fever):
 Hallucinations
 Vomiting
 A stiff neck (they’re unable to put their chin on their chest or have pain when moving their neck forward)
 A skin rash
 A rapid heart rate.
Also get medical help if the person has a seizure (fit), or has signs of a seizure about to happen, such as
regular twitching or jerking.
Call Healthline 0800 611 116 if you are unsure what you should do.
Fever in pregnancy
If you’re pregnant and have a temperature of 38.5ºC – or any fever lasting for three days or more – you
must see your lead maternity carer. They’ll need to monitor the effects of the fever on your baby.
Self care
Most fevers last only three to four days – and a mild fever may not need any treatment at all.
Try these ideas if your fever is mild and you don’t have any other worrying symptoms:
 Drink plenty of fluids – water is best.
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 72 | Page
 Get plenty of rest.
 Wear light weight clothes and use lighter bedding. Keep the room temperature normal.
 Put cool cloths on your face, arms and neck to help you cool down. Don’t use any rapid cooling methods
that may make you shiver. (The muscle movement in shivering will actually raise your temperature and can
make your fever worse.)
Appendix B Temperature Measurement
Fever and Night Sweats
Fever is a common sign that on its own is usually little help in making a diagnosis. Persistent high
fever needs urgent treatment. Fever over 42.2°C (108°F) producesunconsciousness and leads to permanent brain
damage if sustained. Fever can be classified as:
 Low: 37.2-38°C (99°-100.4°F).
 Moderate: 38.1-40°C (100.5°-104°F).
 High: >40°C (104°F).
Fever may also be described as:
 Remitting - the most common type with daily temperatures fluctuating above the normal range.
 Intermittent - daily temperature drops into the normal range and then rises back above normal. If
temperature fluctuates widely causing chills and sweating, it is called a hectic fever.
 Sustained - persistent raised temperature with little fluctuation.
 Relapsing - alternating feverish and a febrile periods.
 Undulant - gradual increase in temperature, which stays high for a few days then gradually reduces.
Fever may also be described in terms of its duration; brief (<3 weeks), or prolonged. The term pyrexia of
unknown origin (PUO) is used to describe a condition where no underlying cause can be found.[1]
Night sweats are common and there is a long list of possible causes, mostly benign but important to
diagnose in order to manage effectively. Serious causes of night sweats can usually be excluded by a thorough
history, examination and simple investigations if required.[2]
Appendix C Body Temperature
What is body temperature?
Body temperature is a measure of the body's ability to generate and get rid of heat. The body is very
good at keeping its temperature within a narrow, safe range in spite of large variations in temperatures outside
the body.
When you are too hot, the blood vessels in your skin expand (dilate) to carry the excess heat to your
skin's surface. You may begin to sweat, and as the sweat evaporates, it helps cool your body. When you are too
cold, your blood vessels narrow (contract) so that blood flow to your skin is reduced to conserve body heat. You
may start shivering, which is an involuntary, rapid contraction of the muscles. This extra muscle activity helps
generate more heat. Under normal conditions, this keeps your body temperature within a narrow, safe range.
Where is body temperature measured?
Your body temperature can be measured in many locations on your body. The mouth, ear, armpit, and
rectum are the most commonly used places. Temperature can also be measured on your forehead.
What are Fahrenheit and Celsius?
Thermometers are calibrated in either degrees Fahrenheit (°F) or degrees Celsius (°C), depending on
the custom of the region. Temperatures in the United States are often measured in degrees Fahrenheit, but the
standard in most other countries is degrees Celsius.
What is normal body temperature?
Most people think of a "normal" body temperature as an oral temperature of 98.6°F (37°C). This is an
average of normal body temperatures. Your temperature may actually be 1°F (0.6°C) or more above or
below 98.6°F (37°C). Also, your normal body temperature changes by as much as 1°F (0.6°C) throughout the
day, depending on how active you are and the time of day. Body temperature is very sensitive to hormone levels
and may be higher or lower when a woman is ovulating or having her menstrual period.
Appendix D Blood pressure chart
Use the blood pressure chart below to see what your blood pressure means. The blood pressure chart is
suitable for adults of any age. (The level for high blood pressure does not change with age.)
Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices
DOI: 10.9790/0661-17546373 www.iosrjournals.org 73 | Page
Blood pressure readings have two numbers, for example 140/90mmHg.
The top number is your systolic blood pressure. (The highest pressure when your heart beats and
pushes the blood round your body.) The bottom one is your diastolicblood pressure. (The lowest pressure when
your heart relaxes between beats.)
The blood pressure chart below shows ranges of high, low and healthy blood pressure readings.
Blood pressure chart for adults
Using this blood pressure chart: To work out what your blood pressure readings mean, just find your top
number (systolic) on the left side of the blood pressure chart and read across, and your bottom number
(diastolic) on the bottom of the blood pressure chart. Where the two meet is your blood pressure.
Appendix D Cont'd
What blood pressure readings mean
As you can see from the blood pressure chart,only one of the numbers has to be higher or lower than it
should be to count as either high blood pressure or low blood pressure:
 90 over 60 (90/60) or less: You may have low blood pressure. More on low blood pressure.
 More than 90 over 60 (90/60) and less than 120 over 80 (120/80): Your blood pressure reading is ideal
and healthy. Follow a healthy lifestyle to keep it at this level.
 More than 120 over 80 and less than 140 over 90 (120/80-140/90): You have a normal blood pressure
reading but it is a little higher than it should be, and you should try to lower it. Make healthy changes to
your lifestyle.
 140 over 90 (140/90) or higher (over a number of weeks): You may have high blood pressure
(hypertension).Change your lifestyle - see your doctor or nurse and take any medicines they may give
you. More on high blood pressure
So:
 if your top number is 140 or more - then you may have high blood pressure, regardless of your bottom
number.
 if your bottom number is 90 or more - then you may have high blood pressure, regardless your top
number.
 if your top number is 90 or less - then you may have low blood pressure, regardless of your bottom
number.
 if your bottom number is 60 or less - then you may have low blood pressure, regardless of your top
number.

More Related Content

What's hot

A descriptive study to assess the need for the new Image based Electronic MLC...
A descriptive study to assess the need for the new Image based Electronic MLC...A descriptive study to assess the need for the new Image based Electronic MLC...
A descriptive study to assess the need for the new Image based Electronic MLC...All India Institute of Medical Sciences
 
Health Care Monitoring for the CVD Detection using Soft Computing Techniques
Health Care Monitoring for the CVD Detection using Soft Computing TechniquesHealth Care Monitoring for the CVD Detection using Soft Computing Techniques
Health Care Monitoring for the CVD Detection using Soft Computing Techniquesijfcstjournal
 
Mobile health applications risk management framework
Mobile health applications risk management frameworkMobile health applications risk management framework
Mobile health applications risk management frameworkKipkoech Benard
 
Iot based intelligent medicine box 1
Iot based intelligent medicine box 1Iot based intelligent medicine box 1
Iot based intelligent medicine box 1sidsingh680
 
Mhealth - mobile health
Mhealth - mobile healthMhealth - mobile health
Mhealth - mobile healthTushar Swami
 
Biomedical engineering and recent trends
Biomedical engineering and recent trendsBiomedical engineering and recent trends
Biomedical engineering and recent trendsHanzelah Khan
 
IRJET- Design and Implementation of Health Monitoring System
IRJET-  	  Design and Implementation of Health Monitoring SystemIRJET-  	  Design and Implementation of Health Monitoring System
IRJET- Design and Implementation of Health Monitoring SystemIRJET Journal
 
Use of mobile device in health care setting
Use of mobile device in health care settingUse of mobile device in health care setting
Use of mobile device in health care settingDr. Samir Sawli
 
Portable Real Time Cardiac Activity Monitoring System
Portable  Real Time Cardiac Activity Monitoring SystemPortable  Real Time Cardiac Activity Monitoring System
Portable Real Time Cardiac Activity Monitoring SystemIRJET Journal
 
IJSRED-V2I3P58
IJSRED-V2I3P58IJSRED-V2I3P58
IJSRED-V2I3P58IJSRED
 
IRJET- IoT based Smart Fitness Tracker for Gymnasiums
IRJET- IoT based Smart Fitness Tracker for GymnasiumsIRJET- IoT based Smart Fitness Tracker for Gymnasiums
IRJET- IoT based Smart Fitness Tracker for GymnasiumsIRJET Journal
 
03 cjece-2018-0001
03 cjece-2018-000103 cjece-2018-0001
03 cjece-2018-0001Stefan Oniga
 
Network Connected Medical Devices - A Case Study
Network Connected Medical Devices - A Case StudyNetwork Connected Medical Devices - A Case Study
Network Connected Medical Devices - A Case StudySophiaPalmira
 
Intelligent Healthcare Monitoring in IoT
Intelligent Healthcare Monitoring in IoTIntelligent Healthcare Monitoring in IoT
Intelligent Healthcare Monitoring in IoTIJAEMSJORNAL
 
Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...journalBEEI
 
IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...
IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...
IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...IRJET Journal
 
[Wroclaw #6] Medical device security
[Wroclaw #6] Medical device security[Wroclaw #6] Medical device security
[Wroclaw #6] Medical device securityOWASP
 

What's hot (20)

A descriptive study to assess the need for the new Image based Electronic MLC...
A descriptive study to assess the need for the new Image based Electronic MLC...A descriptive study to assess the need for the new Image based Electronic MLC...
A descriptive study to assess the need for the new Image based Electronic MLC...
 
Cobme
CobmeCobme
Cobme
 
Ijcet 06 06_004
Ijcet 06 06_004Ijcet 06 06_004
Ijcet 06 06_004
 
Health Care Monitoring for the CVD Detection using Soft Computing Techniques
Health Care Monitoring for the CVD Detection using Soft Computing TechniquesHealth Care Monitoring for the CVD Detection using Soft Computing Techniques
Health Care Monitoring for the CVD Detection using Soft Computing Techniques
 
Mobile health applications risk management framework
Mobile health applications risk management frameworkMobile health applications risk management framework
Mobile health applications risk management framework
 
Iot based intelligent medicine box 1
Iot based intelligent medicine box 1Iot based intelligent medicine box 1
Iot based intelligent medicine box 1
 
Mhealth - mobile health
Mhealth - mobile healthMhealth - mobile health
Mhealth - mobile health
 
Biomedical engineering and recent trends
Biomedical engineering and recent trendsBiomedical engineering and recent trends
Biomedical engineering and recent trends
 
IRJET- Design and Implementation of Health Monitoring System
IRJET-  	  Design and Implementation of Health Monitoring SystemIRJET-  	  Design and Implementation of Health Monitoring System
IRJET- Design and Implementation of Health Monitoring System
 
Use of mobile device in health care setting
Use of mobile device in health care settingUse of mobile device in health care setting
Use of mobile device in health care setting
 
Portable Real Time Cardiac Activity Monitoring System
Portable  Real Time Cardiac Activity Monitoring SystemPortable  Real Time Cardiac Activity Monitoring System
Portable Real Time Cardiac Activity Monitoring System
 
Biomedical application of wireless sensor 2
Biomedical application of wireless sensor 2 Biomedical application of wireless sensor 2
Biomedical application of wireless sensor 2
 
IJSRED-V2I3P58
IJSRED-V2I3P58IJSRED-V2I3P58
IJSRED-V2I3P58
 
IRJET- IoT based Smart Fitness Tracker for Gymnasiums
IRJET- IoT based Smart Fitness Tracker for GymnasiumsIRJET- IoT based Smart Fitness Tracker for Gymnasiums
IRJET- IoT based Smart Fitness Tracker for Gymnasiums
 
03 cjece-2018-0001
03 cjece-2018-000103 cjece-2018-0001
03 cjece-2018-0001
 
Network Connected Medical Devices - A Case Study
Network Connected Medical Devices - A Case StudyNetwork Connected Medical Devices - A Case Study
Network Connected Medical Devices - A Case Study
 
Intelligent Healthcare Monitoring in IoT
Intelligent Healthcare Monitoring in IoTIntelligent Healthcare Monitoring in IoT
Intelligent Healthcare Monitoring in IoT
 
Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...
 
IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...
IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...
IRJET- Applications of Wireless Sensor Networks in Healthcare: An Overview of...
 
[Wroclaw #6] Medical device security
[Wroclaw #6] Medical device security[Wroclaw #6] Medical device security
[Wroclaw #6] Medical device security
 

Viewers also liked

Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...
Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...
Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...IOSR Journals
 
The linkage between strategic human resource management, innovation and firm ...
The linkage between strategic human resource management, innovation and firm ...The linkage between strategic human resource management, innovation and firm ...
The linkage between strategic human resource management, innovation and firm ...IOSR Journals
 
Finite Elements Modeling and Analysis of Double Skin Composite Plates
Finite Elements Modeling and Analysis of Double Skin Composite PlatesFinite Elements Modeling and Analysis of Double Skin Composite Plates
Finite Elements Modeling and Analysis of Double Skin Composite PlatesIOSR Journals
 
Design of an effective automated machine for quality palm kernel production
Design of an effective automated machine for quality palm kernel  productionDesign of an effective automated machine for quality palm kernel  production
Design of an effective automated machine for quality palm kernel productionIOSR Journals
 
Explanation for the Increase in the Expansion of the Universe through Gravita...
Explanation for the Increase in the Expansion of the Universe through Gravita...Explanation for the Increase in the Expansion of the Universe through Gravita...
Explanation for the Increase in the Expansion of the Universe through Gravita...IOSR Journals
 
Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...
Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...
Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...IOSR Journals
 
Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...
Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...
Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...IOSR Journals
 
A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...
A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...
A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...IOSR Journals
 
Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...
Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...
Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...IOSR Journals
 
Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...
Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...
Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...IOSR Journals
 
Evaluation of Human Resource Management Practices on the Productivity and Per...
Evaluation of Human Resource Management Practices on the Productivity and Per...Evaluation of Human Resource Management Practices on the Productivity and Per...
Evaluation of Human Resource Management Practices on the Productivity and Per...IOSR Journals
 
Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...
Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...
Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...IOSR Journals
 
Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...
Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...
Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...IOSR Journals
 
A Review Study On Optimisation Of Process Of Wedm And Its Development
A Review Study On Optimisation Of Process Of Wedm And Its DevelopmentA Review Study On Optimisation Of Process Of Wedm And Its Development
A Review Study On Optimisation Of Process Of Wedm And Its DevelopmentIOSR Journals
 
A Review Paper on Cross Platform Mobile Application Development IDE
A Review Paper on Cross Platform Mobile Application Development IDEA Review Paper on Cross Platform Mobile Application Development IDE
A Review Paper on Cross Platform Mobile Application Development IDEIOSR Journals
 
Engine Emissions at Various Cetane Numbers with Exhaust Gas Recirculation
Engine Emissions at Various Cetane Numbers with Exhaust Gas RecirculationEngine Emissions at Various Cetane Numbers with Exhaust Gas Recirculation
Engine Emissions at Various Cetane Numbers with Exhaust Gas RecirculationIOSR Journals
 

Viewers also liked (20)

Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...
Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...
Estimation of Damping Derivative of a Delta Wing with Half Sine Wave Curved L...
 
The linkage between strategic human resource management, innovation and firm ...
The linkage between strategic human resource management, innovation and firm ...The linkage between strategic human resource management, innovation and firm ...
The linkage between strategic human resource management, innovation and firm ...
 
Finite Elements Modeling and Analysis of Double Skin Composite Plates
Finite Elements Modeling and Analysis of Double Skin Composite PlatesFinite Elements Modeling and Analysis of Double Skin Composite Plates
Finite Elements Modeling and Analysis of Double Skin Composite Plates
 
Design of an effective automated machine for quality palm kernel production
Design of an effective automated machine for quality palm kernel  productionDesign of an effective automated machine for quality palm kernel  production
Design of an effective automated machine for quality palm kernel production
 
I012415358
I012415358I012415358
I012415358
 
Explanation for the Increase in the Expansion of the Universe through Gravita...
Explanation for the Increase in the Expansion of the Universe through Gravita...Explanation for the Increase in the Expansion of the Universe through Gravita...
Explanation for the Increase in the Expansion of the Universe through Gravita...
 
B010310612
B010310612B010310612
B010310612
 
Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...
Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...
Comparative Analysis, Security Aspects & Optimization of Workload in Gfs Base...
 
Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...
Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...
Feasibility of Laying Fiber-Optic Cables underwater along River Nile Basin- S...
 
A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...
A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...
A Model for Optimum Scheduling Of Non-Resumable Jobs on Parallel Production M...
 
Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...
Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...
Effect of Post Annealing on Structural and Optical Propertie of Sno2 Thin Fil...
 
Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...
Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...
Quantization of the Orbital Motion of a Mass In The Presence Of Einstein’s Gr...
 
Evaluation of Human Resource Management Practices on the Productivity and Per...
Evaluation of Human Resource Management Practices on the Productivity and Per...Evaluation of Human Resource Management Practices on the Productivity and Per...
Evaluation of Human Resource Management Practices on the Productivity and Per...
 
Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...
Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...
Fatliquor preparation from Karanja seed oil (Pongamia pinnata L.) and its app...
 
A1303020104
A1303020104A1303020104
A1303020104
 
Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...
Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...
Synthesis, characterization and in vitro Antimicrobial activity of Cu (II) an...
 
G010245056
G010245056G010245056
G010245056
 
A Review Study On Optimisation Of Process Of Wedm And Its Development
A Review Study On Optimisation Of Process Of Wedm And Its DevelopmentA Review Study On Optimisation Of Process Of Wedm And Its Development
A Review Study On Optimisation Of Process Of Wedm And Its Development
 
A Review Paper on Cross Platform Mobile Application Development IDE
A Review Paper on Cross Platform Mobile Application Development IDEA Review Paper on Cross Platform Mobile Application Development IDE
A Review Paper on Cross Platform Mobile Application Development IDE
 
Engine Emissions at Various Cetane Numbers with Exhaust Gas Recirculation
Engine Emissions at Various Cetane Numbers with Exhaust Gas RecirculationEngine Emissions at Various Cetane Numbers with Exhaust Gas Recirculation
Engine Emissions at Various Cetane Numbers with Exhaust Gas Recirculation
 

Similar to Cyber Physical Systems Error Prediction in Medical Devices

“Detection of Diseases using Machine Learning”
“Detection of Diseases using Machine Learning”“Detection of Diseases using Machine Learning”
“Detection of Diseases using Machine Learning”IRJET Journal
 
Multiple disease prediction using Machine Learning Algorithms
Multiple disease prediction using Machine Learning AlgorithmsMultiple disease prediction using Machine Learning Algorithms
Multiple disease prediction using Machine Learning AlgorithmsIRJET Journal
 
IRJET - Review on Classi?cation and Prediction of Dengue and Malaria Dise...
IRJET -  	  Review on Classi?cation and Prediction of Dengue and Malaria Dise...IRJET -  	  Review on Classi?cation and Prediction of Dengue and Malaria Dise...
IRJET - Review on Classi?cation and Prediction of Dengue and Malaria Dise...IRJET Journal
 
New methodology to detect the effects of emotions on different biometrics in...
New methodology to detect the effects of emotions on different  biometrics in...New methodology to detect the effects of emotions on different  biometrics in...
New methodology to detect the effects of emotions on different biometrics in...IJECEIAES
 
COVID-19 knowledge-based system for diagnosis in Iraq using IoT environment
COVID-19 knowledge-based system for diagnosis in Iraq using IoT environmentCOVID-19 knowledge-based system for diagnosis in Iraq using IoT environment
COVID-19 knowledge-based system for diagnosis in Iraq using IoT environmentnooriasukmaningtyas
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESINTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESpijans
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESINTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESpijans
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESINTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESamsjournal
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES pijans
 
IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...
IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...
IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...IRJET Journal
 
Advancing the cybersecurity of the healthcare system with self- optimising an...
Advancing the cybersecurity of the healthcare system with self- optimising an...Advancing the cybersecurity of the healthcare system with self- optimising an...
Advancing the cybersecurity of the healthcare system with self- optimising an...Petar Radanliev
 
IRJET- Mobile Assisted Remote Healthcare Service
IRJET- Mobile Assisted Remote Healthcare ServiceIRJET- Mobile Assisted Remote Healthcare Service
IRJET- Mobile Assisted Remote Healthcare ServiceIRJET Journal
 
January_2024 Top 10 Read Articles in Computer Networks & Communications.pdf
January_2024 Top 10 Read Articles in Computer Networks & Communications.pdfJanuary_2024 Top 10 Read Articles in Computer Networks & Communications.pdf
January_2024 Top 10 Read Articles in Computer Networks & Communications.pdfIJCNCJournal
 
IRJET- Patient Health Monitoring using IoT with Machine Learning
IRJET- Patient Health Monitoring using IoT with Machine LearningIRJET- Patient Health Monitoring using IoT with Machine Learning
IRJET- Patient Health Monitoring using IoT with Machine LearningIRJET Journal
 
March 2024 - Top 10 Read Articles in Computer Networks & Communications
March 2024 - Top 10 Read Articles in Computer Networks & CommunicationsMarch 2024 - Top 10 Read Articles in Computer Networks & Communications
March 2024 - Top 10 Read Articles in Computer Networks & CommunicationsIJCNCJournal
 
Cloud Computing: A Key to Effective & Efficient Disease Surveillance System
Cloud Computing: A Key to Effective & Efficient Disease Surveillance SystemCloud Computing: A Key to Effective & Efficient Disease Surveillance System
Cloud Computing: A Key to Effective & Efficient Disease Surveillance Systemidescitation
 
Implementation Of Real Time IoT Based Health monitoring system
Implementation Of Real Time IoT Based Health monitoring systemImplementation Of Real Time IoT Based Health monitoring system
Implementation Of Real Time IoT Based Health monitoring systemkchakrireddy
 
A Novel Approach for Forecasting Disease Using Machine Learning
A Novel Approach for Forecasting Disease Using Machine LearningA Novel Approach for Forecasting Disease Using Machine Learning
A Novel Approach for Forecasting Disease Using Machine LearningIRJET Journal
 
IRJET- A Prediction Engine for Influenza Pandemic using Healthcare Analysis
IRJET- A Prediction Engine for Influenza  Pandemic using Healthcare AnalysisIRJET- A Prediction Engine for Influenza  Pandemic using Healthcare Analysis
IRJET- A Prediction Engine for Influenza Pandemic using Healthcare AnalysisIRJET Journal
 

Similar to Cyber Physical Systems Error Prediction in Medical Devices (20)

“Detection of Diseases using Machine Learning”
“Detection of Diseases using Machine Learning”“Detection of Diseases using Machine Learning”
“Detection of Diseases using Machine Learning”
 
Multiple disease prediction using Machine Learning Algorithms
Multiple disease prediction using Machine Learning AlgorithmsMultiple disease prediction using Machine Learning Algorithms
Multiple disease prediction using Machine Learning Algorithms
 
IRJET - Review on Classi?cation and Prediction of Dengue and Malaria Dise...
IRJET -  	  Review on Classi?cation and Prediction of Dengue and Malaria Dise...IRJET -  	  Review on Classi?cation and Prediction of Dengue and Malaria Dise...
IRJET - Review on Classi?cation and Prediction of Dengue and Malaria Dise...
 
New methodology to detect the effects of emotions on different biometrics in...
New methodology to detect the effects of emotions on different  biometrics in...New methodology to detect the effects of emotions on different  biometrics in...
New methodology to detect the effects of emotions on different biometrics in...
 
COVID-19 knowledge-based system for diagnosis in Iraq using IoT environment
COVID-19 knowledge-based system for diagnosis in Iraq using IoT environmentCOVID-19 knowledge-based system for diagnosis in Iraq using IoT environment
COVID-19 knowledge-based system for diagnosis in Iraq using IoT environment
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESINTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESINTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASESINTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
 
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
INTERNET OF THINGS BASED MODEL FOR IDENTIFYING PEDIATRIC EMERGENCY CASES
 
IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...
IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...
IRJET- Diabetes Mellitus Diagnosis using Artificial Intelligence Techniques C...
 
N018138696
N018138696N018138696
N018138696
 
Advancing the cybersecurity of the healthcare system with self- optimising an...
Advancing the cybersecurity of the healthcare system with self- optimising an...Advancing the cybersecurity of the healthcare system with self- optimising an...
Advancing the cybersecurity of the healthcare system with self- optimising an...
 
IRJET- Mobile Assisted Remote Healthcare Service
IRJET- Mobile Assisted Remote Healthcare ServiceIRJET- Mobile Assisted Remote Healthcare Service
IRJET- Mobile Assisted Remote Healthcare Service
 
January_2024 Top 10 Read Articles in Computer Networks & Communications.pdf
January_2024 Top 10 Read Articles in Computer Networks & Communications.pdfJanuary_2024 Top 10 Read Articles in Computer Networks & Communications.pdf
January_2024 Top 10 Read Articles in Computer Networks & Communications.pdf
 
IRJET- Patient Health Monitoring using IoT with Machine Learning
IRJET- Patient Health Monitoring using IoT with Machine LearningIRJET- Patient Health Monitoring using IoT with Machine Learning
IRJET- Patient Health Monitoring using IoT with Machine Learning
 
March 2024 - Top 10 Read Articles in Computer Networks & Communications
March 2024 - Top 10 Read Articles in Computer Networks & CommunicationsMarch 2024 - Top 10 Read Articles in Computer Networks & Communications
March 2024 - Top 10 Read Articles in Computer Networks & Communications
 
Cloud Computing: A Key to Effective & Efficient Disease Surveillance System
Cloud Computing: A Key to Effective & Efficient Disease Surveillance SystemCloud Computing: A Key to Effective & Efficient Disease Surveillance System
Cloud Computing: A Key to Effective & Efficient Disease Surveillance System
 
Implementation Of Real Time IoT Based Health monitoring system
Implementation Of Real Time IoT Based Health monitoring systemImplementation Of Real Time IoT Based Health monitoring system
Implementation Of Real Time IoT Based Health monitoring system
 
A Novel Approach for Forecasting Disease Using Machine Learning
A Novel Approach for Forecasting Disease Using Machine LearningA Novel Approach for Forecasting Disease Using Machine Learning
A Novel Approach for Forecasting Disease Using Machine Learning
 
IRJET- A Prediction Engine for Influenza Pandemic using Healthcare Analysis
IRJET- A Prediction Engine for Influenza  Pandemic using Healthcare AnalysisIRJET- A Prediction Engine for Influenza  Pandemic using Healthcare Analysis
IRJET- A Prediction Engine for Influenza Pandemic using Healthcare Analysis
 

More from IOSR Journals (20)

A011140104
A011140104A011140104
A011140104
 
M0111397100
M0111397100M0111397100
M0111397100
 
L011138596
L011138596L011138596
L011138596
 
K011138084
K011138084K011138084
K011138084
 
J011137479
J011137479J011137479
J011137479
 
I011136673
I011136673I011136673
I011136673
 
G011134454
G011134454G011134454
G011134454
 
H011135565
H011135565H011135565
H011135565
 
F011134043
F011134043F011134043
F011134043
 
E011133639
E011133639E011133639
E011133639
 
D011132635
D011132635D011132635
D011132635
 
C011131925
C011131925C011131925
C011131925
 
B011130918
B011130918B011130918
B011130918
 
A011130108
A011130108A011130108
A011130108
 
I011125160
I011125160I011125160
I011125160
 
H011124050
H011124050H011124050
H011124050
 
G011123539
G011123539G011123539
G011123539
 
F011123134
F011123134F011123134
F011123134
 
E011122530
E011122530E011122530
E011122530
 
D011121524
D011121524D011121524
D011121524
 

Recently uploaded

Vertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering TipsVertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering TipsMiki Katsuragi
 
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks..."LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...Fwdays
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brandgvaughan
 
CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):comworks
 
Pigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfAddepto
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
SAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxSAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxNavinnSomaal
 
Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Enterprise Knowledge
 
SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024Lorenzo Miniero
 
costume and set research powerpoint presentation
costume and set research powerpoint presentationcostume and set research powerpoint presentation
costume and set research powerpoint presentationphoebematthew05
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 3652toLead Limited
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfAlex Barbosa Coqueiro
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationSafe Software
 
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Mark Simos
 
Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Scott Keck-Warren
 
APIForce Zurich 5 April Automation LPDG
APIForce Zurich 5 April  Automation LPDGAPIForce Zurich 5 April  Automation LPDG
APIForce Zurich 5 April Automation LPDGMarianaLemus7
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupFlorian Wilhelm
 
My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024The Digital Insurer
 
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)Wonjun Hwang
 

Recently uploaded (20)

Vertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering TipsVertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering Tips
 
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks..."LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brand
 
CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):
 
Pigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food ManufacturingPigging Solutions in Pet Food Manufacturing
Pigging Solutions in Pet Food Manufacturing
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdf
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
SAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxSAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptx
 
Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024
 
SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024SIP trunking in Janus @ Kamailio World 2024
SIP trunking in Janus @ Kamailio World 2024
 
costume and set research powerpoint presentation
costume and set research powerpoint presentationcostume and set research powerpoint presentation
costume and set research powerpoint presentation
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdf
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
 
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
 
Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024
 
APIForce Zurich 5 April Automation LPDG
APIForce Zurich 5 April  Automation LPDGAPIForce Zurich 5 April  Automation LPDG
APIForce Zurich 5 April Automation LPDG
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project Setup
 
My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024
 
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
Bun (KitWorks Team Study 노별마루 발표 2024.4.22)
 

Cyber Physical Systems Error Prediction in Medical Devices

  • 1. IOSR Journal of Computer Engineering (IOSR-JCE) e-ISSN: 2278-0661,p-ISSN: 2278-8727, Volume 17, Issue 5, Ver. IV (Sep. – Oct. 2015), PP 63-73 www.iosrjournals.org DOI: 10.9790/0661-17546373 www.iosrjournals.org 63 | Page Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices Stephen O. Maitanmi1 *, Sunday A. Idowu2 , Olawale J. Omotosho3 , Olusegun A. Ojesanmi4 123 Department of Computer Science, Babcock University, Ilisan Remo, Ogun State, Nigeria 4 Department of Computer Science, Federal University of Agriculture, Abeokuta, Ogun State, Nigeria Abstract: Embedded software is a piece of software that is built in a system or hardware component to achieve an objective. Cyber-Physical Systems (CPS) are integrations of computation with physical processes which are monitored and controlled by the embedded systems. CPS has positively affected a number of application areas which include communication, consumer energy, infrastructure, healthcare, manufacturing, military, robotics and transportation. This research paper focuses on CPS with particular application interest on error prediction in smart medical devices SMDs. SMDs are used for healthcare services by medical personnels in such a way that they have to interact with the patients in one form or the other. A number of research works has focused on errors arising from the use of SMDs. Errors arise from readings of the SMDs as a result of usability challenges from embedded software leading to malfunctioning of the devices. This research paper is aimed at providing a model for error verification and prediction in the use of SMDs using Euler’s method of representation with the aims of establishing safety and reliability of the use of SMDs with possible reduction in risk of accidents. In order to achieve the stated objective, a mathematical model using Euler’s method was adopted for its fast convergence rate and simplicity for error prediction. Input data was provided through a critical incident analysis of online database which provide readings from medical experts. These readings were compared to the standard world benchmarks. The difference between the readings and the standard benchmark validates the existence of errors. Due to the complexity of the model, an algorithm was developed to obtain an optimal solution of P1-P5 within an acceptable threshold runtime. An implementation was carried out using Java programming language because of its robustness and cross platform advantages which provides an efficient error estimate and result analysis. The analysis showed that the optimum performances of the SMDs were hindered by errors. The results generated from the use of thermometer for the diagnosis of Malaria predicted 98.1% accuracy in measurement, Upper respiratory tract infection predicted 99.3% accuracy in measurement, Tonsilitis predicted 99.6% accuracy in measurement, Severe head injury predicted 99.9% accuracy in measurement and Septicaemia predicted 99.9% accuracy. These results show that Malaria generated the highest error in the use of thermometer. In conclusion, CPS generally is an active area of research with a number of application domains in particular health care systems. This particular work has provided a model for error verification in the use of SMDs. Keywords: Cyber Physical Systems, Embedded system, SMDs, Euler’s method; I. Introduction The availability of the Internet alongside its advantages has led to the pervasive use of networking and information technology (IT) across all sectors [1]. Interestingly, these technologies are combined with elements of the physical world such as machines, devices and structures to create smart or intelligent systems that offer increased usefulness, productivity, safety and speed which enabled previous challenging functions. Embedded software is a piece of software that is hidden in a system or hardware components written specifically in a coordinated fashion to achieve an objective. Embedded system is not only one of the most important fields for current computer-based applications; it is also one of the most challenging fields of computer science or more precisely, Software Engineering [2]. Embedded systems are used in many areas ranging from vehicles and mobile phones to washing machines and printers to increase and enhance productivity. Cyber-Physical Systems (CPS) are integrations of computation with physical processes which are monitored and controlled by the embedded systems usually with feedback loops where physical processes affect computations [3]. All CPS have computational processes that interact with physical components. These computational processes can either be simple or complex instructional processes. For example, simple processes include water heater or cutting machine while the complex processes include vehicle, aircraft systems, oil refineries among others. The computational and physical processes of such systems are highly interconnected
  • 2. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 64 | Page and coordinated to function together. The following diagram in Fig.1 sheds more light into the meaning of CPS as regards integration of computation with physical processes. Fig. 1. Cyber-Physical Systems Concept CPS concepts as illustrated in Fig.1 shows that cyberspace is the Internet which is invisible in information systems and it synergizes the physical operations that we carry out daily either in Computer Science, Medicine or any other professional fields. The importance of the Internet cannot be overemphasized, it harnesses the operations that were formerly carried out independently to a single, faster, robust and more coordinated operation. To further illustrate the usefulness of CPS, a three months old baby can be monitored from an independent room through a wearable body wire while a monitor is placed in the parent’s room with the aid of sensors which is a device that detects a change in a physical stimulus and turns it into a signal that is measurable. The movements within the defined locations are known with the aid of the sensors embedded as part of the software. Error verification is defined as investigating a device or software for the existence of error [4]. Meanwhile, the Institute of Medicine Patient Safety described patient safety as a freedom from accidental injury due to medical care, software error or medical errors. It has increasingly been regarded as the key element of quality in the last two decades [5]. The process began in 1999 with Harvard Medical Practice study that showed adverse events in 3.7% of hospitalization and errors which could be related to 27.6% of adverse events [6]. In 2001, the Institute of Medicine published a study tagged ‘to err is human: building a safer health system’ [7]. This study estimated that preventable medical errors are responsible for between 44, 000 and 98, 000 deaths annually in United States. A more heart touching news from the Journal of Patient Safety reports that the numbers may be much higher which was put between 210,000 and 440,000 annually for patients who visit the hospital. Such patients are likely to suffer some type of preventable harm that contributes to their death [5]. This report would make medical errors the third highest cause of death in America after heart disease and cancer. Lack of patient safety measures make the care unsafe and produce opportunities for medical errors to occur [7] while [8] stated that 2,000,000 defective devices were recalled in the last decades due to deficiencies in outputs such examples include unmanned rocket launched in June 1998 called Ariane 5. Ariane 5 was a project that lasted for ten years and cost $7 billion exploded after 37 seconds into the flight due to software errors; code ported from Ariane 4 rocket did not work [9]. Similarly, on land transportation; a Honda CR-V that cost $1.5M in US was recalled to update the software that controls its automatic transmissions. In similar manner, a Toyota Hybrid was recalled because of software deficiency that made it enters fail-safe mode that shuts down engines frequently [9]. Similarly, in Nigeria health services, [10] affirmed in a conference proceeding that seventy percent of deaths in third world countries occur before age five and are caused by factors that are controllable. The SMDs studied include thermometer, insulin pump, infusion pump and sphygmomanometer. They are briefly defined as follows; thermometer measures the body temperature, insulin pump maintains the normal body glucose level, infusion pump transfers fluids, medication or nutrients into a patient’s circulatory system while sphygmomanometer measure the blood pressure of the body [11]. The objective of this paper was to design a model for error verification and prediction of SMDs in CPS. In fulfilling this mandate, the research carried out a critical evaluation of some frequently used SMDs to verify and predict the existence of errors.
  • 3. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 65 | Page The paper layout The rest of the paper is organized as follows: apart from the introduction section, subsequent sections handle the methodology of the research, results, discussions and conclusion. II. Methodology In order to achieve the stated objective which was to develop a model for error verification and prediction of SMDs in CPS.; a critical incident analysis approach was adopted to look into the use of the SMDs because of the sensitivity of the devices. A mathematical model using Euler’s method was employed for its fast convergence rate and simplicity. Secondary input data were retrieved from an online database of Medline, Web of Science, Health Technology Assessments and Health and Science Care Information Center which provides database of readings from medical experts. These retrieved data were further compared with the standard world benchmarks from literatures. The following sections consist of the breakdown of each SMD analysed alongside its standard and practicable benchmarks. This input data was considered to achieve accuracy and correctness in the research and experimentation process of the proposed model. Due to the complexity of the model, an algorithm was developed to obtain an optimal solution within an acceptable threshold runtime. Implementation was carried out using Java programming language because of its robustness and cross platform advantages which provides an efficient error estimate and result analysis. The SMDs analysed in this research paper include thermometer, sphygmomanometer, infusion pump and insulin pump. These SMDs were chosen because of their relevance in the implementation of CPS as regards technological advancement to health care systems. Thermometer for both Benchmark and Readings The following represent the analysis of digital thermometer: Table 1. Temperature readings and standard benchmark (0 C) Source : Appendix A-C Where URTI= Upper respiratory tract infection. Sphygmomanometer (mmHg) for both Benchmark and Readings Symptoms Benchmark (P) Time(t) Readings P(t) Error(t) Malaria 37.9 0 37.9 0 38.5 10 36.0 2.5 38.7 20 36.1 2.6 38.9 30 36.2 2.7 39.5 40 36.3 3.2 39.6 50 36.3 3.3 URTI 37.7 0 37.7 0 38.5 10 37.7 0.8 39.0 20 37.8 1.2 39.5 30 37.9 1.6 39.8 40 38.0 1.8 40.0 50 38.0 2.0 Tonsilitis 36.4 0 36.4 0 37.8 10 36.5 0.4 39.2 20 37.0 2.2 39.9 30 37.6 2.3 40.0 40 38.0 2.0 40.5 50 38.5 2.0 Severe Head Injury 39.8 0 39.8 0 39.9 10 39.8 0.1 40.5 20 39.9 0.6 43.5 30 40.0 3.5 45.5 40 40.0 5.5 47.5 50 40.1 7.4 Septicaemia 39.9 0 39.9 0 40.0 10 39.0 1.0 45.3 20 39.5 5.8 47.3 30 39.9 7.4 47.8 40 40.0 7.8 48.0 50 40.0 8.0
  • 4. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 66 | Page Table 2. Sphygmomanometer readings and standard Benchmark (mmhg) Source: Appendix D Systolic Diastolic BM/RD/error BM/RD/error BM/RD/error BM/RD/error BM/RD/error ≤84 85-89 90-99 100-109 110-119 ≤129 2 1 1 3 2 1 4 3 1 5 4 1 6 5 1 <130-139 2 2 0 3 2 1 4 3 1 5 4 1 6 5 1 <140-159 3 3 0 3 3 0 4 3 1 5 4 1 6 5 1 <160-179 5 4 1 5 4 1 4 3 1 5 4 1 6 5 1 <180-209 5 5 0 5 5 0 5 4 1 6 5 1 6 5 1 ≥210 7 6 1 6 6 0 5 4 1 7 6 1 7 6 1 Where N=161, BM= benchmark and RD= readings Infusion Pump for both Benchmark and Readings Table 3. Infusion Pump readings and standard Benchmark (Msl/hr.) Source: [13] Benchmark Readings Error Time (Mins) Adult -100mls/hr 80mls/hr 10 60 500mls/hr 480mls/hr 20 60 500mls/hr 490mls/hr 10 60 4 litres/day 4litres/day 0 1,440 500mls/12rs 480mls/12rs 20 720 500mls/day 495mls/day 5 1,440 Children 300mls/day 280mls/day 20 1,440 10mls/hr 10mls/hr 0 60 1 liter/day 1 litre/day 0 1,440 2 litres/day 2 litres/day 0 1,440 Insulin Pump for both Benchmark and Readings Table 4. Insulin Pump readings and standard Benchmark 1 unit of insulin for every 15g Carbohydrates taken. Insulin Pump Workbook. Source: [14] Time Benchmark Blood Glucose Amount of Carbohydrates Benchmark Insulin intake Readings Insulin intake Errors Day 1 7:30am 5.7 60g 4 8 4 1:30pm 11.2 50g 3.3 6.6 3.3 6:00pm 7.2 72g 5 10 5 10:00pm 8 Day 2 7:30am 6.8 60g 4 8 4 1:30pm 12.3 50g 3.3 6.6 3.3 6:00pm 7.1 90g 6 12 6 10:00pm 11.3 2.1 Relationship between Model and Thermometer table Performance at optimum level is represented in Table 1 as the Benchmark of temperature taken when error is not involved (P) which is equivalent to 100%. Performance with respect to time is represented in Table 1 taking note of the time in minutes when temperatures were taken P(t). The recorded time were decimated to reduce the magnitudes of error such as 0.1, 0.2, 0.3 and so on for Malaria. These values are not constant for other ailments. Error as shown in Table 1 gives the difference between benchmark and actual readings which was also decimated to reduce error value ther eby realizing (E) of Malaria to be 0.25, 0.26, 0.27, 0.32 and 0.33 P1-P5 is the performance level of the SMDs and the model with respect to the introduced error. 2.2 Mathematical Modeling of Error Prediction in the use of Thermometer P  Performance at optimum level  P t  Performance with respect to t  E t  Error introduced at any time t     (1)P t P E t                              
  • 5. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 67 | Page (1) gives the difference between the benchmark and the actual reading which is a function of time.     1 ( )P t E t P t Where   1 P t is the change in performance which is proportional to the product of performance and the error at any time t     1 ( )P t K E t P t      1 ( )P t E t P t  substituting ( )P t in (1) gives     1 ( )[ ] ( 2 )P t E t P E t                            By Using Euler’s iteration method to project a possible outcome when variables such as error, and time are used as input 1 ( ) (3 ) 1 P P h P n n n                               Where n= 0-4 (five variables from temperature readings on malaria from Table 1) While 0E = 0.25, 1E =0.26, 2E = 0.27, 3E =0.32 and 4E =0.33 decimated values from temperature readings in Table 1 h = dt and dt =0.1 from model time used from Table 1 From (2) at optimum level when 0t  and 0E  (0 ) 0P P from (2) = 1 or 100% when there is no error, meaning performance at best level Starting Iteration From (2) 1 ( ) ( 0 .2 5*0 .1)[1 0 .2 5*0 .1] 0 .0 2 5*0 .9 7 5 ) 0 .0 2 40P        1 ( ) 1 ( 0 .1)( 0 .0 2 4 ) 1 0 .0 0 2 4 0 .9 9 7 61 0 0P P h P        or 99.8% 1 ( ) ( 0 .2 6*0 .1)[1 0 .2 6*0 .1] 0 .0 2 6*0 .9 7 4 ) 0 .0 2 51P        1 ( ) 0 .9 9 8 ( 0 .1)( 0 .0 2 5 ) 0 .9 9 8 0 .0 0 2 5 0 .9 9 52 1 1P P h P        or 99.5% 1 ( ) ( 0 .2 7 *0 .1)[1 0 .2 7 *0 .1] 0 .0 2 7 *0 .9 7 3 ) 0 .0 2 62P        1 ( ) 0 .9 9 5 ( 0 .1)( 0 .0 2 6 ) 0 .9 9 5 0 .0 0 2 6 0 .9 9 23 2 2P P h P        or 99.2% 1 ( ) ( 0 .3 2*0 .1)[1 0 .3 2*0 .1] 0 .0 3 2*0 .9 6 8 ) 0 .0 3 13P        1 ( ) 0 .9 9 2 ( 0 .1)( 0 .0 3 1) 0 .9 9 2 0 .0 3 1 0 .9 8 94 3 3P P h P        or 98.9% 1 ( ) (1 .3 3*0 .1)[1 0 .3 3*0 .1] 0 .0 3 3*0 .9 6 7 ) 0 .0 3 1 94P        1 ( ) 0 .9 8 9 ( 0 .1)( 0 .0 3 1 9 ) 0 .9 8 9 0 .0 0 3 1 9 0 .9 8 65 4 4P P h P        or 98.6% 2.3 Development of Algorithm The algorithm is based on the following formulae:     1 ( )[ ] (4)P t E t P E t                              1 ( ) (5 ) 1 P P h P n n n                                  Declarations Step 1: Initialize fractional variables (double) , ,h t E Step 2: Initialize whole numbers (Int) ,P n Step 3: let all values of n be between 0 to 4 Step 4: Let E0=0.25, E1=0.26, E2=0.27, E3=0.32 and E4=1.85 Step 5: Let 0.1 be stored in t as fractional number Step 6: Let 1 be stored in P as whole number Step 7: Let 0.1 be stored in h as fractional number Starting the Euler’s iterations Step 8: making reference to (4) above where 1 ( ) ( )[ ( )]0 0 0P E t P E t   Find 1 ( )0P to 1 ( ) 4P Step 9: starting iteration where the values of n= 0 to 4 starting with n=0 to begin where K is assumed to be 1
  • 6. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 68 | Page 1 ( )1 0 0P P h P  store the result in 1P Step 10: Use the value obtained above in step 8 on 1 1 ( P ) to find 1 ( )2 1 1P P h P  and store the result in 2P Step 11: Use the value obtained above in step 8 on 1 ( ) 2P to find 1 ( )3 2 2P P h P  and store the result in 3P Step 12: Use the value obtained above in step 8 on 1 ( )3P to find 1 ( )4 3 3P P h P  and store the result in 4P Step 13: Use the value obtained above in step 8 on 1 ( ) 4P to find 1 ( )5 4 4P P h P  and store the value in 5P Step 14: Display the values of , , , ,1 2 3 4 5P P P P P III. Results & Discussions Table 5 Error Prediction in the use of Thermometer for the diagnosis of Malaria Table 5. shows the output of the error prediction that resulted in the use of thermometer for the diagnosis of Table 5 shows the output of the error prediction that resulted in the use of thermometer for the diagnosis of patient suspected to be having malaria fever. The variables displayed include E= error rate which are values of 0.25, 0.26, 0.27, 0.32 and 0.33 taken from Table 1, it appears that the higher the values, the weaker the functionality of the SMDs in terms of precision. P% refers to percentage analysis of the performance of the predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1and constant as discussed in the model development stages. P = performance, meaning maximum performance when there is no error. The first output on column 1 therefore shows that the prediction generated 98.1% accuracy in measurement. It appears also that the percentage error increases as the error in prediction increases. Consequently, the rates of error increases down the column. The last column shows the highest error rate with low performance. The performance value was 89.8%. From this value it is evidenced that performance was affected. Table 6. Error Prediction in the use of Thermometer for the diagnosis of Urinary Tract Infection (URTI) Table 6 shows the output of the error prediction that resulted in the use of thermometer for the diagnosis of patient suspected to be having URTI fever. The variables displayed include E= error rate which are values of 0.08, 0.12, 0.16, 0.18 and 0.2 taken from Table 1, the higher the values, the weaker the functionality of the SMDs in terms of correctness in measurement. P% refers to percentage analysis of the performance of the predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1 and constant as
  • 7. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 69 | Page discussed in the model development stages. P = performance, meaning maximum performance when there is no error. The first output on column 1 therefore shows that the prediction generated 99.3% accuracy in measurement. It appears also that the percentage error increases as the error in prediction increases. Therefore, the rates of error increases down the column. The last column shows the highest involvement of errors. The performance value was 93.8%.From this value it is evidenced that performance was affected. Table 7. Error Prediction in the use of Thermometer for the diagnosis of Tonsilitis Table 7 shows the output of the error prediction that resulted in the use of thermometer for the diagnosis of patient suspected to be having Tonsilitis fever. The variables displayed include E= error rate which are values of 0.04, 0.22, 0.23, 0.2 and 0.2 taken from Table 1, the higher the values, the weaker the functionality of the SMDs in terms of correctness in measurement. P% refers to percentage analysis of the performance of the predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1 and constant as discussed in the model development stages. P = performance (benchmark), meaning maximum performance when there is no error. The first output on column 1 shows that the prediction generated 99.6% accuracy in measurement. It appears also that the percentage error increases with respect to increase in error. Therefore, the rates of error increases down the column. The performance value was 92.9%. From this value it is evidenced that performance was affected. Table 8. Error Prediction in the use of Thermometer for the diagnosis of Severe Head Injury Table 8 shows the output of the error prediction that resulted in the use of thermometer for the diagnosis of patient suspected to be having severe head injury fever. The variables displayed include E= error rate which are values of 0.01, 0.06, 0.35, 0.55 and 0.74 taken from Table 1, the higher the values, the weaker the functionality of the SMDs in terms of correctness in measurement. P% refers to percentage analysis of the performance of the predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1 and constant as discussed in the model development stages. P = performance (benchmark), meaning maximum performance when there is no error. The first output on column 1 therefore shows that the prediction generated 99.9% accuracy in measurement. Therefore, the rates of error increases down the column. The last column shows the highest involvement of errors. The performance value was 92.7%. It is evidenced from this value that performance was affected.
  • 8. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 70 | Page Table 9. Error Prediction in the use of Thermometer for the diagnosis of Septicaemia Table 9 shows the output of the error prediction that resulted in the use of thermometer for the diagnosis of patient suspected to be having Septicaemia fever. The variables displayed include E= error rate which are values of 0.01, 0.06, 0.35, 0.55 and 0.74 taken from Table 1, it appears that the higher the values, the weaker the functionality of the SMDs in terms of precision. P% refers to percentage analysis of the performance of the predicted error, the higher the error, the lower the performance of the SMDs, t=time which is 0.1 and constant as discussed in the model development stages. P = performance (benchmark), meaning maximum performance when there is no error. The first output on column 1 therefore shows that the prediction generated 99.9% accuracy in measurement. It appears also that the percentage error increases as the error in prediction increases. In addition, the rates of error increases down the column. The last column shows the highest error rate with low performance. The performance value was 92.2%. It is therefore evidenced that performance was affected. IV. Conclusion In conclusion, CPS generally is an active area of research with a number of evolving application domains in particular, health care systems. The research paper investigated a number of SMDs and discovered error in its applications. Based on these findings, a model was developed to guide medical personnel and individual users in the appropriate use of SMDs.` Acknowledgements I will like to appreciate Mr. Semiu Ayinde of the department of Basic Sciences, Babcock University, the Head of Department of Radiology, Babcock University Teaching Hospital (BUTH) Dr. Yusuf, Adebola and Dr. Ajiboye, Bimpe who contributed immensely to the completion of this research paper. References [1]. Workshop Report."Foundations for Innovation in Cyber-Physical Systems Available:http://events.energetics.com /NIST- CPSWorkshop/downloads.html,2013 [Nov. 20, 2014]. [2]. San Diego State University ."Introduction to computers". Information technology and Teachers Education International Conference. Available: edweb.sdsu.edu/people/marshal/publication.html, 2009. [June 12, 2014] [3]. E. A. Lee "Cyber physical systems: Design Challenges" Technical Report, EECS Department, University of California, Berkeley, January (CPS Steering Group), 2008 [4]. D. R. Wallace, and R. U. Fujii. :Software verification and validation: an overview" Software, IEEE,Vol.6, no.3,pp.10- 17, 1989 [5]. T. J. John "A New Evidence-Based Estimate of Patient Harms Associated with Hospital Care" Journal of Patient Safety Volume 9, Number 3. 2013 [6]. NAP Committee on Quality of Health Care in America IoM. ‘To err is human: building a Safer Health System’ Washington, D.C.: National Academy Press., 2001 [7]. M. Soares, J. I. F. Salluh, and F. A. Bozza, Current definitions of patient safety. In: Chiche, J. D, Moreno R, Putensen C, Rhodes A, editors. Patient Safety and Quality of Care in Intensive Care Medicine. Berlin:Medizinisch WissenschaftlicheVerlagsgesellschaft; p. 9-17., 2009 [8]. M. Paolo Formal verification of Medical user interface software in PVS, Queen Mary University of London, 2014. [9]. K. Venkatasubramanian, "Challenges and Research directions in Medical Cyber Physical Systems" 2013. [10]. D. C Stockwell, and A. D. Slonim. Quality and safety in the Intensive Care Unit. J Intensive Care Med 2006 Jul; 21(4):199-210. 2006 [11]. O. Akpabio, An address by the Executive Governor of Akwa Ibom state, his Excellency, Chief Godswill on the Occasion of the National Health Conference 2009 (NHC 2009) Available: http://www,unilorin.edu.ng/publications/sakamj/Technical%20Reports%20%20Saka%20MJ/NHC%202009 %20Proceedings.pdf [May 28, 2014]. [12]. WHO Environmental Health Criteria 232, Static fields, 2006. Available www.icnirp.org/eb/publications/article [May 12, 2014] [13]. FDA .Infusion Pump Improvement Initiative Center for Devices and Radiological Health U.S. Food and Drug Administration 2010. [14]. E. Joan, and L. Helen "Insulin Pump Workbook"(n.d) Available: www.diabetes- education.net/pdf/resources/insulin_pump_workbook.pdf [Jan. 12, 2015].
  • 9. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 71 | Page Appendix A Temperature Measurement Fever in adults If you or a family member has a fever, it means your body temperature is above normal. Around 37ºC is normal A digital thermometer is the best type to use to get an accurate temperature reading. A fever is usually a normal response of your immune system to a virus or bacterial infection. Most healthy adults can tolerate a fever well. Fever ranges and symptoms 38–38.9°C – mild fever With a mild fever you might have flushed cheeks, feel a little lethargic, and be warm to touch. You will generally be able to carry out normal daily activities. 39–39.9°C – high fever With a high fever you may not feel well enough to go to work, you may have aches and pains, and you’ll feel hot to touch. 40°C or higher – very high fever With a very high fever you will usually want to stay in bed or be inactive – you won’t feel well enough to carry out normal activities. You may have lost your appetite. You’ll feel hot to touch. When to see your doctor Some mild diseases produce very high fevers – and severe illnesses can produce mild fever. Therefore, when considering what medical attention you need, it’s important to look at other symptoms and how unwell you feel. You should see your doctor if you or a family member:  has a very high fever (over 40ºC)  is still feverish after three days of home treatment, or seems to be getting sicker  is shivering or shaking uncontrollably, or has chattering teeth  has a severe headache that doesn’t get better after taking painkillers Appendix A Cont'd  is having trouble breathing  is getting confused or is unusually drowsy  has recently travelled overseas. When it’s urgent See your doctor or go to the Emergency Department immediately if you notice the following symptoms (along with a fever):  Hallucinations  Vomiting  A stiff neck (they’re unable to put their chin on their chest or have pain when moving their neck forward)  A skin rash  A rapid heart rate. Also get medical help if the person has a seizure (fit), or has signs of a seizure about to happen, such as regular twitching or jerking. Call Healthline 0800 611 116 if you are unsure what you should do. Fever in pregnancy If you’re pregnant and have a temperature of 38.5ºC – or any fever lasting for three days or more – you must see your lead maternity carer. They’ll need to monitor the effects of the fever on your baby. Self care Most fevers last only three to four days – and a mild fever may not need any treatment at all. Try these ideas if your fever is mild and you don’t have any other worrying symptoms:  Drink plenty of fluids – water is best.
  • 10. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 72 | Page  Get plenty of rest.  Wear light weight clothes and use lighter bedding. Keep the room temperature normal.  Put cool cloths on your face, arms and neck to help you cool down. Don’t use any rapid cooling methods that may make you shiver. (The muscle movement in shivering will actually raise your temperature and can make your fever worse.) Appendix B Temperature Measurement Fever and Night Sweats Fever is a common sign that on its own is usually little help in making a diagnosis. Persistent high fever needs urgent treatment. Fever over 42.2°C (108°F) producesunconsciousness and leads to permanent brain damage if sustained. Fever can be classified as:  Low: 37.2-38°C (99°-100.4°F).  Moderate: 38.1-40°C (100.5°-104°F).  High: >40°C (104°F). Fever may also be described as:  Remitting - the most common type with daily temperatures fluctuating above the normal range.  Intermittent - daily temperature drops into the normal range and then rises back above normal. If temperature fluctuates widely causing chills and sweating, it is called a hectic fever.  Sustained - persistent raised temperature with little fluctuation.  Relapsing - alternating feverish and a febrile periods.  Undulant - gradual increase in temperature, which stays high for a few days then gradually reduces. Fever may also be described in terms of its duration; brief (<3 weeks), or prolonged. The term pyrexia of unknown origin (PUO) is used to describe a condition where no underlying cause can be found.[1] Night sweats are common and there is a long list of possible causes, mostly benign but important to diagnose in order to manage effectively. Serious causes of night sweats can usually be excluded by a thorough history, examination and simple investigations if required.[2] Appendix C Body Temperature What is body temperature? Body temperature is a measure of the body's ability to generate and get rid of heat. The body is very good at keeping its temperature within a narrow, safe range in spite of large variations in temperatures outside the body. When you are too hot, the blood vessels in your skin expand (dilate) to carry the excess heat to your skin's surface. You may begin to sweat, and as the sweat evaporates, it helps cool your body. When you are too cold, your blood vessels narrow (contract) so that blood flow to your skin is reduced to conserve body heat. You may start shivering, which is an involuntary, rapid contraction of the muscles. This extra muscle activity helps generate more heat. Under normal conditions, this keeps your body temperature within a narrow, safe range. Where is body temperature measured? Your body temperature can be measured in many locations on your body. The mouth, ear, armpit, and rectum are the most commonly used places. Temperature can also be measured on your forehead. What are Fahrenheit and Celsius? Thermometers are calibrated in either degrees Fahrenheit (°F) or degrees Celsius (°C), depending on the custom of the region. Temperatures in the United States are often measured in degrees Fahrenheit, but the standard in most other countries is degrees Celsius. What is normal body temperature? Most people think of a "normal" body temperature as an oral temperature of 98.6°F (37°C). This is an average of normal body temperatures. Your temperature may actually be 1°F (0.6°C) or more above or below 98.6°F (37°C). Also, your normal body temperature changes by as much as 1°F (0.6°C) throughout the day, depending on how active you are and the time of day. Body temperature is very sensitive to hormone levels and may be higher or lower when a woman is ovulating or having her menstrual period. Appendix D Blood pressure chart Use the blood pressure chart below to see what your blood pressure means. The blood pressure chart is suitable for adults of any age. (The level for high blood pressure does not change with age.)
  • 11. Cyber Physical Systems: Error Verification and Prediction in Smart Medical Devices DOI: 10.9790/0661-17546373 www.iosrjournals.org 73 | Page Blood pressure readings have two numbers, for example 140/90mmHg. The top number is your systolic blood pressure. (The highest pressure when your heart beats and pushes the blood round your body.) The bottom one is your diastolicblood pressure. (The lowest pressure when your heart relaxes between beats.) The blood pressure chart below shows ranges of high, low and healthy blood pressure readings. Blood pressure chart for adults Using this blood pressure chart: To work out what your blood pressure readings mean, just find your top number (systolic) on the left side of the blood pressure chart and read across, and your bottom number (diastolic) on the bottom of the blood pressure chart. Where the two meet is your blood pressure. Appendix D Cont'd What blood pressure readings mean As you can see from the blood pressure chart,only one of the numbers has to be higher or lower than it should be to count as either high blood pressure or low blood pressure:  90 over 60 (90/60) or less: You may have low blood pressure. More on low blood pressure.  More than 90 over 60 (90/60) and less than 120 over 80 (120/80): Your blood pressure reading is ideal and healthy. Follow a healthy lifestyle to keep it at this level.  More than 120 over 80 and less than 140 over 90 (120/80-140/90): You have a normal blood pressure reading but it is a little higher than it should be, and you should try to lower it. Make healthy changes to your lifestyle.  140 over 90 (140/90) or higher (over a number of weeks): You may have high blood pressure (hypertension).Change your lifestyle - see your doctor or nurse and take any medicines they may give you. More on high blood pressure So:  if your top number is 140 or more - then you may have high blood pressure, regardless of your bottom number.  if your bottom number is 90 or more - then you may have high blood pressure, regardless your top number.  if your top number is 90 or less - then you may have low blood pressure, regardless of your bottom number.  if your bottom number is 60 or less - then you may have low blood pressure, regardless of your top number.