4. Function of x-ray imaging system is to
provide a controlled flow of electrons intense
enough to produce an x-ray beam
appropriate for imaging.
4
Function of x-ray imaging system is to
provide a controlled flow of electrons intense
enough to produce an x-ray beam
appropriate for imaging.
5. Components of x-ray imaging
systems:
The x-ray tube
The operating console
High voltage generators
5
The x-ray tube
The operating console
High voltage generators
6. The operating console:
Controls the x-ray tube current.
Controls the x-ray tube voltage.
Control the exposure time.
Ensure x-ray beam produced is of proper
quantity and quality.
Quantity is the no. of x-rays or the intensity of
the beam (mR or mR / mAs).
Quality is the penetrability of x-ray beam
(kVp or HVL).6
Controls the x-ray tube current.
Controls the x-ray tube voltage.
Control the exposure time.
Ensure x-ray beam produced is of proper
quantity and quality.
Quantity is the no. of x-rays or the intensity of
the beam (mR or mR / mAs).
Quality is the penetrability of x-ray beam
(kVp or HVL).
7. High –voltage generators:
Contains the high-voltage step-up
transformer.
Contains the rectification circuit.
7
Contains the high-voltage step-up
transformer.
Contains the rectification circuit.
8. X-ray imaging systems; identification:
Depends on the energy of x-ray produced, or
Intended use, or purpose
Usually operates at 25-150 kV and 100 to
1200 mA tube current.
General purpose x-ray examination room will
have a radiographic as well as fluoroscopy
systems.
8
Depends on the energy of x-ray produced, or
Intended use, or purpose
Usually operates at 25-150 kV and 100 to
1200 mA tube current.
General purpose x-ray examination room will
have a radiographic as well as fluoroscopy
systems.
9. Examination table:
Uniform thickness.
Transparent to x-rays (carbon fiber).
Floating type.
Tilting, 90/30 degree of tilt.
9
Uniform thickness.
Transparent to x-rays (carbon fiber).
Floating type.
Tilting, 90/30 degree of tilt.