SlideShare a Scribd company logo
1 of 28
X-ray tube
 X-ray tube is an device for generating X-rays by
accelerating electrons to high energies and causing
them to strike a metal target from which X-rays are
emitted.
 High voltage is applied b/w cathode and anode.
 The electrons are emitted from the filament by thermionic emission and
accelerated towards the anode and strike the target.
Basic principle of x-ray tube:
- +
 An X-ray tube consists of an anode and cathode mounted inside an evacuated
glass tube.
 The cathode consists, of a small coil called filament , mounted in a focusing
cup.
 The anode target is mainly made of Tungsten , separated from filament by a
small gap.
 Electrons are boiled off the filament by applying an electric current so that it
becomes white hot- the process is called ‘Thermionic emission’ .
 A high voltage (HV) of 100 Kv or more applied and the electrons are attracted
across the gap to collide with the anode with the high energies, to produce X-
rays.
 Bremsstrahlung & Characteristic radiation process are involved here.
X-ray tube
Quality & Quantity:
 Quality refers to the overall energy of the beam
 As the X-ray beam is polyenergetic, any factors that
increase or decrease the average energy of photons in
the beam affect x-ray beam quality.
 Quality is directly affected by,
i. Changes in kVP
ii. Changes in the material(atomic number Z) of the
target material.
iii. Changes in the filtration.
iv. The type of waveform used (i.e., 1φ, 3φ, or high
frequency).
 Quantity refers to the number of X-ray photons in the
beam .
 As the number of photons increases, the beam
intensity increases & any factors that affect the
number of x-ray photons in the beam influence x-ray-
beam quantity.
 Quantity is affected by,
i. Changes in mA (tube current).
ii. Changes in the filtration.
iii. Changes in the material (Z number) of target.
iv. Changes in kVP
v. Changes in type of waveform used.
vi. Changes in distance from the tube (FFD).
Effects of changes in kVP:
Peak kilo voltage (kVp ) is the maximum voltage applied
across an X-ray tube, it determines the kinetic energy of
the electrons accelerated in the X-ray tube and the peak
energy of the X-ray spectrum.
•The changes in the kVp varies the energy of
the incoming electrons prior to maximum
energy of X-ray photons in Bremsstrahlung
spectrum.
•Hence as kVp increases more higher energy
photons are included in the beam , raising
the average energy of the beam.
•In addition, the increase in kVp increases
the speed with which incoming electrons
strike the target.
•Each time an incoming electron undergoes
a deceleration about a target atom nucleus,
Bremsstrahlung photons are produced,
resulting in overall Bremsstrahlung photons
 Exposure is approximately proportional to the square of kVp in the
diagnostic energy range
Exposure α (kVp) 2
Eg: the relative exposure of a beam generated with 80 kVp compared with that
of 60 kVp for the same tube current and exposure time is calculated as follows:
(80)2 ≈ 1.78
(60)2
Therefore, the output exposure increases by approximately 78%.
An increase in kVp increases the efficiency of x-ray production and the quantity
and quality of the x-ray beam.
Effect of changes in mA:
Tube current (mA) is the rate of electron flow from filament
to target (electron/sec) and measured in miliamperes.
•As the tube current (mA) increases, the
number of incoming electrons striking
target increases.
•In addition the number of X-ray photons
produced in both Bremsstrahlung and
characteristic x-ray portions of x-ray
spectrum is altered.
•The relationship between the mA and the
number of photons produced directly
affects the amplitude of the x-ray
emission spectrum.
•Eg: if we double the tube current (mA)
the number of the x-ray photons will be
doubled.
Number
of
X-rays
emitted
(per
Kev)
•The changes in the mA does not produces energy shifts in either portion of
spectrum .
Effect of changes in the target materials:
The target (anode) material affects the efficiency of Bremsstrahlung
radiation production, with output exposure roughly proportional to
atomic number.
•As the Z number of target increases the
amount of Bremsstrahlung radiation
produced also increases.
•Although the effect of using a higher Z
number target is more apparent on the
high-energy side of the Bremsstrahlung
peak, the spectra are same in many ways.
•The minimum & maximum energies are
the same.
•The peak of Bremsstrahlung spectrum
(greatest no. of X-rays production) occurs
at the same position on energy axis.
Effect of changes in the target materials:
•Hence in the Bremsstrahlung portion of x-ray spectrum there
is a change in the quantity not in quality of x-ray photons.
•As the Z number becomes larger the binding energy of shell
electrons also become larger. This increase in electronic
binding energies results in higher-energy characteristics x-rays.
•Low Z number target materials, such as molybdenum(Z=42) ,
tend to produce not only low energy characteristic x-rays but
also Bremsstrahlung x-rays.
•Low energy characteristic x-rays of molybdenum are useful in
soft tissue imaging such as mammography.
Effects of changes in mAs:
 Coulombs refers to charge carried by
individual electrons as they move towards
target plate.
 The exact number of electrons flowing
between the tube electrodes can be
determined by total charge in coulombs by
the charge on individual electron.
•Hence as mAs is increased the number of electrons striking the target
increases, the more electrons hit the target, the more x-ray are produced in
both complete & discrete portions of x-ray spectrum.
•Changes in mAs do not affect quality.
The exposure time is the duration of x-ray production. The quantity of x-rays is
directly proportional to the product of tube current and exposure time (mAs).
(mA in Ampere ) X (exposure in seconds)= coulombs X sec
sec
=coulombs
Effects of changes in added filtration:
 Filtration refers to the process by which photons in a
polyenergitic beam are selectively removed by
materials placed in the path of emerging x-ray
beam.
 Filtration materials may be part of the equipment such as,
1. Glass envelope of the tube
2. The insulating oil surrounding the tube
3. Tube housing window
Filtration by the parts of the equipment is called
Inherent filtration.
Effects of changes in added filtration
•Beryllium (Z=4) windows are used in X-
ray tubes when it is not desirable to
completely filter out low-energy photons
(mammography).
•Added filtration consists of additional
filters placed in the path of emerging
beam to absorb low-energy photons(as
these low energy photons contribute
radiation dose only to the patient).
Inherent filtration in most tubes is approximately the same
as 1-mm aluminium equivalent ( 1 mm of Al).
There are several effects of added filtration:
1. Increases the average energy of the x-ray beam::
 As low-energy photons are removed by the added filters.
 High energy photons constitutes a greater percentage of
the beam.
 This increases the average energy of the beam which is
referred as ‘Beam hardening’.
Beam hardening: It refers the process in which the quality, or
energy, of an x-ray beam is increased by removing lower-
energy photons with appropriate filtration.
Effect of changes in voltage waveform :
SINGLE PHASE (1φ) vs. THREE PHASE (3φ) :
The generator waveform affects the quality of the emitted
x-ray spectrum.
For the same kVp, a single-phase generator provides a lower
average potential difference than does a three-phase or
high-frequency generator.
 As the voltage across the tube electrodes fluctuate, the number of x-rays
produced also rises & falls.
 As voltage rises from zero , x-ray intensity gradually increases & intensity
reaches a peak as peak voltage (KVp) when voltage becomes even higher.
 As voltage increases there is an increased probability of multiple
interactions between incoming electrons and target atoms. each interaction
produces x-rays.
 During second half-cycle , X-ray intensity drops sharply until the voltage
reaches zero.
Single phase x-ray generators
produces a voltage waveform that
begins at zero, reaches maximum
and again drops back to zero
(cycle repeated throughout the
exposure).
Variation of x-ray production with tube voltage
SINGLE PHASE (1φ):
In three phase (3φ) generator the high voltage is applied
to the x-ray tube.
Three phase power line is supplied through three
separate wires and is stepped up by an transformer.
The voltage waveform in each wire is kept slightly out of
phase to each other, so that the voltage across tube is
always at maximum.
With three-phase power & full wave rectification, six
voltage pulses are applied to the x-ray tube during each
power cycle. This is known as three-phase six pulse
system.
The voltage ripple
[(Vmax –Vmin)/ Vmax] x 100
is 13% to 25% for 3φ six pulse system.
With voltage across tube remains significantly higher
throughout the exposure time, more electrons are
produced(affects quantity) & also average energy of beam
produced is also higher( affects quality)
Three phase generator (3φ):
SUMMARY
Ratings of X-ray tubes:
 Rating is used to describe the practical limits which
are inherent in any device.
 The rating of an X-ray unit is the combination of
exposure settings which the unit can withstand
without incurring unacceptable damage.
Selectable & non-selectable factors affecting ratings of a X-ray unit
Selectable factors kVp, mA ,exposure time ,focal spot.
Non-Selectable factors
(Rotating anode)
Rectification, thermal capacity of
anode & tube shield, diameter of
anode, rate of anode rotation, ratings
of high-tension cables & transformers.
X-RAY EXPOSURE RATING CHARTS :
 Rating charts are the charts which
determine operational limits of the x-ray tube
for single & multiple exposure & permissible
heat load of anode & tube housing.
 single exposure
 multiple rapid exposure (Angiographic capability)
Single-Exposure rating chart: It provides information on the allowed
combinations of kVp , mA & exposure time for a particular x-ray tube.
•For each tube, there are usually four individual charts with peak kilo-voltage
as the y-axis & exposure time as the x-axis are provided by manufacturer of x-
ray tube.
•These charts contains a series of curves, each for a particular mA value. Each
of the four charts is for a specific focal spot size and anode rotation speed.
•Each curve represents the maximal allowable tube current for a particular
kVp and exposure time.
X-RAY EXPOSURE RATING CHART:
In graph,
 Using a tube current (mA) of 300 mA & 110 kVp any exposure up to 0.036
sec can be used with single phase(1φ) & 0.2 sec with 3 phase (3φ) using high
speed rotor.
If a low speed rotator is employed, the maximum voltage permissible for the
above exposures at 300 mA are about are about 55kVp for single phase(1φ) &
69KvP for 3 phase (3φ).
Hence there is a great advantage of 3 phase power & improvement achieved
by the use of high speed rotor.
Multiple-exposure rating charts:
 Multiple-exposure rating charts are used in angiography and in
other imaging applications for which sequential exposures are
necessary.
 These charts can be helpful in the planning of an examination,
but they are used infrequently because of the variations in
imaging requirements.
 Most modern angiographic equipment includes a computerized
heat loading system to indicate the anode status and determine
the safety of a given exposure or series of exposures.
 Similarly, in CT, the computer system monitors the x-ray tube
exposure sequence.
 When the tube heat limit has been reached (for either the anode
loading or the housing loading), further image acquisition is
prevented until the anode (or housing) has cooled sufficiently to
ensure safety.
Heat unit:
 Heat unit(HU): In X-ray applications the heat is
measured in Heat units
HU = kVp x mA x s
-kVp is the kilo-voltage peak
-mA is the tube current
t-is the exposure time
 For Single phase : 1HU = 1kVp x 1mA x 1sec
 For three phase : HU=1.35 x kVp x mA x sec
Single phase Three phase
70 kVp
200 mA
0.25 second
HU= 70 X 200 X 0.25
HU = 3500 heat units
60 kVp
100 mA
0.1 second
HU= 60 X 100 X 0.1 X 1.35
HU= 810 heat units
Continuous Heating - Fluoroscopy
Fluoroscopy is an imaging
technique that uses X-rays to
obtain real time moving images of
interior of object.
•Fluoroscopy almost always single
phase
To Find appropriate curve
HU/sec = kVp X mA For right fluoro time, the graph
should be followed from current
heat to right.
Continuous Heating & cooling
Eg:
•For 100 kVp , 6mA & 600 HU , the exposure is started with 50000 KHU &
Exposure time is 3 minutes.
•After exposing for 3 minutes the heat will be around 1,10,000 HU.
•COOLING:
•If the cooling process is started at 1,10,000 HU .
•The time taken for the cooling process will be around 2 minutes.
•Then the Heat will reach around 40000HU. (point under the curve).
Factors affecting xray tube  pptx

More Related Content

Similar to Factors affecting xray tube pptx

Similar to Factors affecting xray tube pptx (20)

The x ray beam
The x ray beam The x ray beam
The x ray beam
 
xray production.pptx
xray production.pptxxray production.pptx
xray production.pptx
 
X ray production (Emission and Filtration)
X ray production (Emission and Filtration)X ray production (Emission and Filtration)
X ray production (Emission and Filtration)
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
 
x-ray_safety_presentation__2_.ppt
x-ray_safety_presentation__2_.pptx-ray_safety_presentation__2_.ppt
x-ray_safety_presentation__2_.ppt
 
721107.ppt
721107.ppt721107.ppt
721107.ppt
 
X-Ray Production.pptx
X-Ray Production.pptxX-Ray Production.pptx
X-Ray Production.pptx
 
Linear accelerator.pptx
Linear accelerator.pptxLinear accelerator.pptx
Linear accelerator.pptx
 
Linear Accelerators.pptx
Linear Accelerators.pptxLinear Accelerators.pptx
Linear Accelerators.pptx
 
CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS
 
Linear accelerator
Linear acceleratorLinear accelerator
Linear accelerator
 
1-Radiation physics had.pptx
1-Radiation physics had.pptx1-Radiation physics had.pptx
1-Radiation physics had.pptx
 
factors_affecting_xray_production.pdf
factors_affecting_xray_production.pdffactors_affecting_xray_production.pdf
factors_affecting_xray_production.pdf
 
x-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptxx-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptx
 
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTSUnit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
 
x-rays.pdf
x-rays.pdfx-rays.pdf
x-rays.pdf
 
GenerationofXRays.ppt
GenerationofXRays.pptGenerationofXRays.ppt
GenerationofXRays.ppt
 
Factors controlling the beam of x-ray
Factors controlling the beam of x-rayFactors controlling the beam of x-ray
Factors controlling the beam of x-ray
 
lecture 2_x ray part_1.pdf
lecture 2_x ray part_1.pdflecture 2_x ray part_1.pdf
lecture 2_x ray part_1.pdf
 
Chapter5 treatment machines for external beam
Chapter5 treatment machines for external beamChapter5 treatment machines for external beam
Chapter5 treatment machines for external beam
 

Recently uploaded

Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfSumit Tiwari
 
Gas measurement O2,Co2,& ph) 04/2024.pptx
Gas measurement O2,Co2,& ph) 04/2024.pptxGas measurement O2,Co2,& ph) 04/2024.pptx
Gas measurement O2,Co2,& ph) 04/2024.pptxDr.Ibrahim Hassaan
 
CELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxCELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxJiesonDelaCerna
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxRaymartEstabillo3
 
What is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERPWhat is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERPCeline George
 
Types of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptxTypes of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptxEyham Joco
 
Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Jisc
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaVirag Sontakke
 
AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.arsicmarija21
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxNirmalaLoungPoorunde1
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxSayali Powar
 
Pharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfPharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfMahmoud M. Sallam
 
Meghan Sutherland In Media Res Media Component
Meghan Sutherland In Media Res Media ComponentMeghan Sutherland In Media Res Media Component
Meghan Sutherland In Media Res Media ComponentInMediaRes1
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
 
How to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxHow to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxmanuelaromero2013
 
Full Stack Web Development Course for Beginners
Full Stack Web Development Course  for BeginnersFull Stack Web Development Course  for Beginners
Full Stack Web Development Course for BeginnersSabitha Banu
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfMr Bounab Samir
 

Recently uploaded (20)

Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
 
Gas measurement O2,Co2,& ph) 04/2024.pptx
Gas measurement O2,Co2,& ph) 04/2024.pptxGas measurement O2,Co2,& ph) 04/2024.pptx
Gas measurement O2,Co2,& ph) 04/2024.pptx
 
CELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptxCELL CYCLE Division Science 8 quarter IV.pptx
CELL CYCLE Division Science 8 quarter IV.pptx
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
 
What is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERPWhat is Model Inheritance in Odoo 17 ERP
What is Model Inheritance in Odoo 17 ERP
 
Types of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptxTypes of Journalistic Writing Grade 8.pptx
Types of Journalistic Writing Grade 8.pptx
 
Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...
 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of India
 
AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptx
 
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Bikash Puri  Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Bikash Puri Delhi reach out to us at 🔝9953056974🔝
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
 
Pharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfPharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdf
 
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Tilak Nagar Delhi reach out to us at 🔝9953056974🔝
 
Meghan Sutherland In Media Res Media Component
Meghan Sutherland In Media Res Media ComponentMeghan Sutherland In Media Res Media Component
Meghan Sutherland In Media Res Media Component
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
 
How to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxHow to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptx
 
Full Stack Web Development Course for Beginners
Full Stack Web Development Course  for BeginnersFull Stack Web Development Course  for Beginners
Full Stack Web Development Course for Beginners
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
 

Factors affecting xray tube pptx

  • 1.
  • 2. X-ray tube  X-ray tube is an device for generating X-rays by accelerating electrons to high energies and causing them to strike a metal target from which X-rays are emitted.
  • 3.  High voltage is applied b/w cathode and anode.  The electrons are emitted from the filament by thermionic emission and accelerated towards the anode and strike the target. Basic principle of x-ray tube: - +
  • 4.  An X-ray tube consists of an anode and cathode mounted inside an evacuated glass tube.  The cathode consists, of a small coil called filament , mounted in a focusing cup.  The anode target is mainly made of Tungsten , separated from filament by a small gap.  Electrons are boiled off the filament by applying an electric current so that it becomes white hot- the process is called ‘Thermionic emission’ .  A high voltage (HV) of 100 Kv or more applied and the electrons are attracted across the gap to collide with the anode with the high energies, to produce X- rays.  Bremsstrahlung & Characteristic radiation process are involved here. X-ray tube
  • 5. Quality & Quantity:  Quality refers to the overall energy of the beam  As the X-ray beam is polyenergetic, any factors that increase or decrease the average energy of photons in the beam affect x-ray beam quality.  Quality is directly affected by, i. Changes in kVP ii. Changes in the material(atomic number Z) of the target material. iii. Changes in the filtration. iv. The type of waveform used (i.e., 1φ, 3φ, or high frequency).
  • 6.  Quantity refers to the number of X-ray photons in the beam .  As the number of photons increases, the beam intensity increases & any factors that affect the number of x-ray photons in the beam influence x-ray- beam quantity.  Quantity is affected by, i. Changes in mA (tube current). ii. Changes in the filtration. iii. Changes in the material (Z number) of target. iv. Changes in kVP v. Changes in type of waveform used. vi. Changes in distance from the tube (FFD).
  • 7. Effects of changes in kVP: Peak kilo voltage (kVp ) is the maximum voltage applied across an X-ray tube, it determines the kinetic energy of the electrons accelerated in the X-ray tube and the peak energy of the X-ray spectrum. •The changes in the kVp varies the energy of the incoming electrons prior to maximum energy of X-ray photons in Bremsstrahlung spectrum. •Hence as kVp increases more higher energy photons are included in the beam , raising the average energy of the beam. •In addition, the increase in kVp increases the speed with which incoming electrons strike the target. •Each time an incoming electron undergoes a deceleration about a target atom nucleus, Bremsstrahlung photons are produced, resulting in overall Bremsstrahlung photons
  • 8.  Exposure is approximately proportional to the square of kVp in the diagnostic energy range Exposure α (kVp) 2 Eg: the relative exposure of a beam generated with 80 kVp compared with that of 60 kVp for the same tube current and exposure time is calculated as follows: (80)2 ≈ 1.78 (60)2 Therefore, the output exposure increases by approximately 78%. An increase in kVp increases the efficiency of x-ray production and the quantity and quality of the x-ray beam.
  • 9. Effect of changes in mA: Tube current (mA) is the rate of electron flow from filament to target (electron/sec) and measured in miliamperes. •As the tube current (mA) increases, the number of incoming electrons striking target increases. •In addition the number of X-ray photons produced in both Bremsstrahlung and characteristic x-ray portions of x-ray spectrum is altered. •The relationship between the mA and the number of photons produced directly affects the amplitude of the x-ray emission spectrum. •Eg: if we double the tube current (mA) the number of the x-ray photons will be doubled. Number of X-rays emitted (per Kev) •The changes in the mA does not produces energy shifts in either portion of spectrum .
  • 10. Effect of changes in the target materials: The target (anode) material affects the efficiency of Bremsstrahlung radiation production, with output exposure roughly proportional to atomic number. •As the Z number of target increases the amount of Bremsstrahlung radiation produced also increases. •Although the effect of using a higher Z number target is more apparent on the high-energy side of the Bremsstrahlung peak, the spectra are same in many ways. •The minimum & maximum energies are the same. •The peak of Bremsstrahlung spectrum (greatest no. of X-rays production) occurs at the same position on energy axis.
  • 11. Effect of changes in the target materials: •Hence in the Bremsstrahlung portion of x-ray spectrum there is a change in the quantity not in quality of x-ray photons. •As the Z number becomes larger the binding energy of shell electrons also become larger. This increase in electronic binding energies results in higher-energy characteristics x-rays. •Low Z number target materials, such as molybdenum(Z=42) , tend to produce not only low energy characteristic x-rays but also Bremsstrahlung x-rays. •Low energy characteristic x-rays of molybdenum are useful in soft tissue imaging such as mammography.
  • 12. Effects of changes in mAs:  Coulombs refers to charge carried by individual electrons as they move towards target plate.  The exact number of electrons flowing between the tube electrodes can be determined by total charge in coulombs by the charge on individual electron. •Hence as mAs is increased the number of electrons striking the target increases, the more electrons hit the target, the more x-ray are produced in both complete & discrete portions of x-ray spectrum. •Changes in mAs do not affect quality. The exposure time is the duration of x-ray production. The quantity of x-rays is directly proportional to the product of tube current and exposure time (mAs). (mA in Ampere ) X (exposure in seconds)= coulombs X sec sec =coulombs
  • 13. Effects of changes in added filtration:  Filtration refers to the process by which photons in a polyenergitic beam are selectively removed by materials placed in the path of emerging x-ray beam.  Filtration materials may be part of the equipment such as, 1. Glass envelope of the tube 2. The insulating oil surrounding the tube 3. Tube housing window Filtration by the parts of the equipment is called Inherent filtration.
  • 14. Effects of changes in added filtration •Beryllium (Z=4) windows are used in X- ray tubes when it is not desirable to completely filter out low-energy photons (mammography). •Added filtration consists of additional filters placed in the path of emerging beam to absorb low-energy photons(as these low energy photons contribute radiation dose only to the patient). Inherent filtration in most tubes is approximately the same as 1-mm aluminium equivalent ( 1 mm of Al).
  • 15. There are several effects of added filtration: 1. Increases the average energy of the x-ray beam::  As low-energy photons are removed by the added filters.  High energy photons constitutes a greater percentage of the beam.  This increases the average energy of the beam which is referred as ‘Beam hardening’. Beam hardening: It refers the process in which the quality, or energy, of an x-ray beam is increased by removing lower- energy photons with appropriate filtration.
  • 16. Effect of changes in voltage waveform : SINGLE PHASE (1φ) vs. THREE PHASE (3φ) : The generator waveform affects the quality of the emitted x-ray spectrum. For the same kVp, a single-phase generator provides a lower average potential difference than does a three-phase or high-frequency generator.
  • 17.  As the voltage across the tube electrodes fluctuate, the number of x-rays produced also rises & falls.  As voltage rises from zero , x-ray intensity gradually increases & intensity reaches a peak as peak voltage (KVp) when voltage becomes even higher.  As voltage increases there is an increased probability of multiple interactions between incoming electrons and target atoms. each interaction produces x-rays.  During second half-cycle , X-ray intensity drops sharply until the voltage reaches zero. Single phase x-ray generators produces a voltage waveform that begins at zero, reaches maximum and again drops back to zero (cycle repeated throughout the exposure). Variation of x-ray production with tube voltage SINGLE PHASE (1φ):
  • 18. In three phase (3φ) generator the high voltage is applied to the x-ray tube. Three phase power line is supplied through three separate wires and is stepped up by an transformer. The voltage waveform in each wire is kept slightly out of phase to each other, so that the voltage across tube is always at maximum. With three-phase power & full wave rectification, six voltage pulses are applied to the x-ray tube during each power cycle. This is known as three-phase six pulse system. The voltage ripple [(Vmax –Vmin)/ Vmax] x 100 is 13% to 25% for 3φ six pulse system. With voltage across tube remains significantly higher throughout the exposure time, more electrons are produced(affects quantity) & also average energy of beam produced is also higher( affects quality) Three phase generator (3φ):
  • 20. Ratings of X-ray tubes:  Rating is used to describe the practical limits which are inherent in any device.  The rating of an X-ray unit is the combination of exposure settings which the unit can withstand without incurring unacceptable damage. Selectable & non-selectable factors affecting ratings of a X-ray unit Selectable factors kVp, mA ,exposure time ,focal spot. Non-Selectable factors (Rotating anode) Rectification, thermal capacity of anode & tube shield, diameter of anode, rate of anode rotation, ratings of high-tension cables & transformers.
  • 21. X-RAY EXPOSURE RATING CHARTS :  Rating charts are the charts which determine operational limits of the x-ray tube for single & multiple exposure & permissible heat load of anode & tube housing.  single exposure  multiple rapid exposure (Angiographic capability)
  • 22. Single-Exposure rating chart: It provides information on the allowed combinations of kVp , mA & exposure time for a particular x-ray tube. •For each tube, there are usually four individual charts with peak kilo-voltage as the y-axis & exposure time as the x-axis are provided by manufacturer of x- ray tube. •These charts contains a series of curves, each for a particular mA value. Each of the four charts is for a specific focal spot size and anode rotation speed. •Each curve represents the maximal allowable tube current for a particular kVp and exposure time.
  • 23. X-RAY EXPOSURE RATING CHART: In graph,  Using a tube current (mA) of 300 mA & 110 kVp any exposure up to 0.036 sec can be used with single phase(1φ) & 0.2 sec with 3 phase (3φ) using high speed rotor. If a low speed rotator is employed, the maximum voltage permissible for the above exposures at 300 mA are about are about 55kVp for single phase(1φ) & 69KvP for 3 phase (3φ). Hence there is a great advantage of 3 phase power & improvement achieved by the use of high speed rotor.
  • 24. Multiple-exposure rating charts:  Multiple-exposure rating charts are used in angiography and in other imaging applications for which sequential exposures are necessary.  These charts can be helpful in the planning of an examination, but they are used infrequently because of the variations in imaging requirements.  Most modern angiographic equipment includes a computerized heat loading system to indicate the anode status and determine the safety of a given exposure or series of exposures.  Similarly, in CT, the computer system monitors the x-ray tube exposure sequence.  When the tube heat limit has been reached (for either the anode loading or the housing loading), further image acquisition is prevented until the anode (or housing) has cooled sufficiently to ensure safety.
  • 25. Heat unit:  Heat unit(HU): In X-ray applications the heat is measured in Heat units HU = kVp x mA x s -kVp is the kilo-voltage peak -mA is the tube current t-is the exposure time  For Single phase : 1HU = 1kVp x 1mA x 1sec  For three phase : HU=1.35 x kVp x mA x sec Single phase Three phase 70 kVp 200 mA 0.25 second HU= 70 X 200 X 0.25 HU = 3500 heat units 60 kVp 100 mA 0.1 second HU= 60 X 100 X 0.1 X 1.35 HU= 810 heat units
  • 26. Continuous Heating - Fluoroscopy Fluoroscopy is an imaging technique that uses X-rays to obtain real time moving images of interior of object. •Fluoroscopy almost always single phase To Find appropriate curve HU/sec = kVp X mA For right fluoro time, the graph should be followed from current heat to right.
  • 27. Continuous Heating & cooling Eg: •For 100 kVp , 6mA & 600 HU , the exposure is started with 50000 KHU & Exposure time is 3 minutes. •After exposing for 3 minutes the heat will be around 1,10,000 HU. •COOLING: •If the cooling process is started at 1,10,000 HU . •The time taken for the cooling process will be around 2 minutes. •Then the Heat will reach around 40000HU. (point under the curve).