4. INTRODUCTION
• Ultrasound biomicroscopy (UBM) provides high-resolution imaging
of ocular structures anterior to the pars plana region of the eye
• Developed by Pavlin, Sherar and Foster in Toronto in the late
1980s
• Provides exceptionally detailed two-dimensional gray-scale
images of the various anterior segment structures and evaluates
them both quantitatively and qualitatively
6. PRINCIPLE
• It acts on a principle similar to that of the B-scan (10Mhz)
• Frequency 50Mhz
• More frequency, less penetration(5mm) and more resolution
• Limited depth of penetration is also associated with a smaller
angular field (4x4 mm)
• addition to the tissues easily seen using conventional methods (ie,
slit lamp), such as the cornea, iris, and sclera, structures including
the ciliary body and zonules, hidden from clinical observation,
can be imaged and their morphology assessed
9. TECHNIQUE
There are three main components of the UBM machine.
1.Transducer
2. High-frequency signal processing.
3. Video monitor
10. Transducer 50 MHz
This radiofrequency travels the body tissue and is
reflected back to the transducer. The reflected
radio frequency is processed by the signal
processing unit.
Signal processing unit
The signal processing unit in UBM is specially
designed to handle high frequency signals.
Subtle movements.
Special motion control device.
Mounted on a pulley with the piezoelectric
crystal fixed on a large handle
11. Patient in supine position
Local anesthetic
Eye cup (plastic or silicone) which is
used to create a small water bath
Methyl cellulose or normal saline can
be used as coupling solution
The reflected signal is best detected
when the transducer is oriented so
that the ultrasound beam strikes the
targeted surface perpendicularly
The crystal of the transducer is placed
in saline approximately 2 mm from
the eye surface.
(This distance of 2 mm prevents injury
to the cornea and also helps as a fluid
standoff.)
17. The Scleral Spur Can Be Identified In The Region Where The Radiopaque Shadow Of The Sclera
Merges With The Relatively Radiolucent Shadow Of The Cornea And Is Often Visible As A Change In
The Configuration When The Inner Border As The Sclera Merges Into The Cornea.
24. GLAUCOMA
open-angle glaucoma
• can be used to measure the anterior
chamber angle in degrees
• assess the configuration of the
peripheral iris
• evaluate the iris insertion in relation to
the trabecular meshwork
• see if there is an anterior insertion of
the iris or an anteriorly displaced
ciliary body
Narrow angles
• UBM shows the extent of angle
closure
• depth of the anterior and posterior
chambers
• identifies pathologic processes
pushing the lens and iris forward.
determine the mechanism of elevated intraocular pressure
(angle-closure versus open-angle) by showing the
relationship between the peripheral iris and trabecular
meshwork.
25. PUPIL BLOCK GLAUCOMA
(A) The angle shows appositional closure owing to anterior
bowing of the iris.
(B) The angle is open with a flattened iris after laser peripheral
iridotomy.
26. PLATEAU IRIS
Plateau iris has been defined based on UBM by Kumar
et al if all criteria fulfilled in at least 2 quadrants:
1.The ciliary process was anteriorly directed, supporting
the peripheral iris so that it was parallel to the
trabecular meshwork.
2. The iris root had a steep rise from its point of
insertion, followed by a downward angulation from the
corneoscleral wall.
3. Presence of a central flat iris plane.
4. An absent ciliary sulcus.
5. Irido-angle contact (above the level of the scleral
spur) in the same quadrant.
Kumar, et al. Prevalence of plateau iris in primary angle closure
suspects, a UBM study. Ophthalmology 2008;115:430-34.
27. To Determine Occludability of the Angle
• perform dark room provocative testing with the UBM, to study the
spontaneous occlusion of the angle under conditions of decreased
illumination
• Helps to identify “at risk” population which can then be subjected to a
laser iridotomy
• better than dark room gonioscopy because the latter is time consuming
and standardization of slit-lamp illumination is difficult
28. Congenital glaucoma
Common features-
• thin stretched out ciliary body,
• abnormal tissue at the irido corneal
angle
• abnormal insertion of the ciliary body
31. Malignant glaucoma
• Extremely shallow anterior chamber,
• occluded angle,
• forward rotation of the ciliary body
• with or without fluid in the suprachoroidal
space.
32. Functional Status of a Filtering Surgery
• whether the sclerostomy aperture is patent
or blocked internally,
• whether the peripheral iridectomy is patent
• whether the filtering bleb is flat, shallow or
deep
36. TUMOURS
• Iris tumors, ciliary body tumors and anterior choroidal tumors can
be imaged
• Borders of the tumor are usually detectable by the change in
reflectivity from surrounding structures
37. CHOROIDAL TUMOUR Irido ciliary cyst
The most common
clinical presentation of
an irido-ciliary cyst is a
peripheral iris elevation -
the typical UBM finding of
a thin walled structure
with no internal
reflectivity is diagnostic.
39. Angle recession is imaged as a tear into the face of
the ciliary body. Ciliary body tissue is still imaged
attached to the scleral spur.
Zonular rupture
43. INTRAOCULAR LENSES
• Can be helpful in analyzing intraocular lens position and determining
the source of the problem if all does not go well
• Anterior chamber depth after surgery can be measured with a high
degree of accuracy.
• margins of the optic can be easily imaged and decentration analysed
• Haptic location in relationship to surrounding structures can be
determined and, in the case of posterior chamber lenses, one can
usually determine if the haptic is in the capsular bag
47. UBM AS-OCT
• contact technique • Non contact technique
• Coupling media used • Not used
• Sound energy used • Light energy used
• Scleral spur less distinct • More distinct
• Posterior lens capsule not seen • Seen
• patient supine, positioning theoretically causes the
iris diaphragm to fall back deepens the AC and opens
the angle.
• Patient sitting,so no alterations
• pressure on the eyecup used while scanning can
influence angle configuration
• Nil
• Only 1 quadrant can be imaged at a time • 4 quadrants can be scanned at once.
48. UBM AS-OCT
• there is no fixed reference point and the angle region
measured is located subjectively as nasal, temporal,
superior, inferior, and so forth, not in exact degrees of
an arc
• With AS-OCT, keeping the fixation angle (the angle
between the instrument’s optical axis and the eye’s
line of sight) at 0 degree, finding the exact location of
the measured angle in degrees of an arc is possible.
• UBM procedureis more time consuming and requires a
highly skilled operator to obtain high-quality precision
images.
• No skills required
Limitations
• risk for infection or corneal abrasion because of the
contact nature of the examination
• it is contraindicated in suspected open-globe injuries.
Limitations
• AS-OCT are that it cannot obtain clear images through
opaque media and is obstructed by the eyelids,
making imaging of the superior and inferior angles
difficult
• also provides limited visualization of the ciliary body.
49. LIMITATIONS
• The most important limitation of UBM is depth.UBM cannot
visualize structures deeper more that4 mm from the surface.
• The other limitation is that UBM cannot be performed in presence
of an open corneal or scleral wound.
In addition, imaging of the anterior segment structures is
possible even in eyes with corneal edema or corneal
opacification that precludes gonioscopic assessment.
PUPILLARY BLOCK AND FOLLOWING IRIDECTOMY. A space is present between the ciliary processes and posterior iris.
irido-zonular contact
The grading of the bleb is done according to intrableb
reflectivity, visibility of route under the scleral flap, formation
of cavernous fluid filled space, and bleb height. Following four
grades have been described:
• Low reflective L
• High reflective H
• Encapsulated E
• Flat F
Eyes with good IOP control mainly have ’L’ type blebs,
these have low to moderate intrableb reflectivity, visible
intrascleral route and higher intrableb height. Flat and
encapsulated blebs generally denote a surgical failure.