MANOJ ARYAL
B . Optometry
IOM,MMC
Slit lamp Biomicroscopy
(Instrumentation,Principle, Illumination
and Uses)
optometris
t
Introduction
 Biomicroscope derives its name from the
fact that it enables the practitioner to
observe the living tissue of eye under
magnification.
 It not only provides magnified view of every
part of eye but also allows quantitative
measurements and photography of every
part for documentation.
• The lamp facilitates an examination
which looks at anterior segment, or
frontal structures, of the human eye,
which includes the
–Eyelid
–Cornea
–Sclera
–Conjunctiva
–Iris
–Aqueous
–Natural crystalline lens and
–Anterior vitreous.
Important historical landmarks
 De Wecker 1863 devised a portable
ophthalmomicroscope .
 Albert and Greenough 1891,developed
a binocular microscope which provided
stereoscopic view.
 Gullstrand ,1911 introduced the
illumination system which had for the
first time a slit diapharm in it
 Therefore Gullstrand is credited with
the invention of slit lamp.
Large Gullstrand
Ophthalmomicroscop
e (1911)
TYPES
 There are 2 types of slit lamp
biomicroscope
1)Zeiss slit lamp biomicroscope
2)Haag streit slit lamp biomicroscope
 In Zeiss type light source is at the base of
the instrument while in Haag streit type it is
at the top of the instrument.
Zeiss slit lamp biomicroscope
Haag streit slit lamp biomicroscope
PRINCIPLE
 A "slit" beam of very bright light
produced by lamp. This beam is focused
on to the eye which is then viewed
under magnification with a microscope
Instrumentation
 Operational components of slit lamp
biomicroscope essentially consist of:
Illumination system
Observation system
Mechanical system
 Illumination system
 It consist of:
A bright ,focal source of light with a slit
mechanism
 Provides an illumination of 2*10^5 to 4*10^5
lux.
 The beam of light can be changed in
intensity,height,width,direction or angle and
color during the examination with the flick of
lever.
Condensing lens system:
Consist of a couple of planoconvex
lenses with their convex surface in
apposition.
Slit and other diapharm:
 Height and width of slit can be varied by
using knobs.
 Projection lens:
 Form an image of slit at eye.
 Advantages,
 1.keeps the aberration of lens down.
 2.increase the depth of focus of slit.
 Reflecting mirrors and prisms
 Filters
Yellow barrier filter
Red free filter
Neutral density filter
Cobalt blue filter
diffuser
Observation system(microscope)
 Observation system is essentially a compound
microscope composed of two optical elements
1.an objective ,2.an eyepiece
 It presents to the observer an enlarged image of
a near object.
 The objective lens consists of two planoconvex
lenses with their convexities put together
providing a composite power of +22D.
 Microscope is binocular i.e. it has two
eyepieces giving binocular observer a
 The eye piece has a lens of +10D.
 To overcome the problem of inverted
image produced by compound
microscope ,slit lamp microscope uses a
pair of prisms b/w the objective and
eyepiece to reinvert the image.
 Most slit lamp provide a range of
magnification from 6x to 40x
Mechanical system
 Joystick arrangement
 Movement of microscope and illumination
system towards and away from the eye
and from side and side is achieved via
joystick arrangement.
Up and down movement arrangement
Obtained via some sort or screw
devices.
Patient support arrangement
Vertically movable chin rest and the
Fixation target:
A movable fixation target greatly faciliates
the examination under some conditions.
Mechanical coupling :
Provides a coupling of microscope and
the illumination system along a common
axis of rotation that coincides their focal
planes.
 This ensures that light falls on the point
where the microscope is focused
 Has advantages when using the slit lamp
for routine examination of anterior
 Magnification control :
Including two or pair of readily
changeable objective lenses and two
sets of eyepieces.
 An on and off switch and illumination
control .
Topcon slit lamp model SL-3E
Light beam is controlled by
knobs
Joy stick arrangement
Chin rest
Reflecting mirror
biomicroscope
Illumination control
Magnification
may be changed
by
flipping a lever...
Changing filters. biomicroscope
Patient positioning
Alignmen
t mark
Microscope
and light
source rotate
indepedently
Filters used in slit lamp biomicroscopy
 Cobalt blue filter
 Used in conjunction with fluorescein stain
 Dye pods in area where the corneal epithelium
is broken or absent.
 The dye absorbs blue light and emits green.
 Uses:
 Ocular staining
 RGP lenses fitting
 Tear layer
 Red free(green)filter:
Obscure any thing that is red hence
the red free light , thus blood vessels
or haemorrhages appears black.
This increases contrast ,revealing the
path and pattern of inflammed blood
vessels.
Fleischer ring can also be viewed
satisfactorily with the red green filter.
Illumination techniques
 Includes
 Diffuse illumination
 Direct illumination
 Parallilepiped
 Optic section
 Conical(pinpoint)
 Tangential
 Specular reflection
 Indirect illumination
 Retro-illumination
 Sclerotic scatter
 Transillumination
 Proximal illumination
Diffuse illumination
 Angle between microscope and
illumination system should be 30-45
degree.
 Slit width should be widest.
 Filter to be used is diffusing filter.
 Magnification: low to medium
 Illumination: medium to high.
Applications:
General view of anterior of eye:
lids,lashes,sclera,cornea ,iris, pupil,
Gross pathology and media opacities
Contact lens fitting.
Assessment of lachrymal reflex.
Optics of diffuse illumination Diffuse illumination with slit beam and
background illumination
Direct illumination
 Involves placing the light source at an
angle of about 40-50 degree from
microscope.
 This arrangement permits both light beam
and microscope to be sharply focused on
the ocular tissue being observed.
 Wide beam direct illumination is commonly
used as a preliminary technique to
evaluate large area.
 it is particularly suitable for assessment
of cataracts,scars,nerves,vessels etc.
 It is also of great importance for the
determination of stabilization of axis of
toric contact lens.
 Parallelepiped:
Constructed by narrowing the beam to 1-
2mm in width to illuminate a rectangular
area of cornea.
Microscope is placed directly in front of
patients cornea.
Light source is approximately 45 degree
from straight ahead position.
Applications:
Used to detect and examine corneal
structures and defects.
Used to detect corneal striae that
develop when corneal edema occurs
with hydrogel lens wear and in
keratoconus.
Higher magnification than that used with
wide beam illumination is preferred to
evaluate both depth and extent of
corneal ,scarring or foreign bodies.
Conical beam(pinpoint)
 Produced by narrowing the vertical height of a
parallelepiped to produce a small circular or
square spot of light.
 Light source is 45-60 degree temporally and
directed into pupil.
 Biomicroscope: directly in front of eye.
 Magnification: high(16-25x)
 Intensity of light source to heighest setting.
 Focusing:
Beam is focused between cornea and
anterior lens surface and dark zone
between cornea and anterior lens
observed.
Principle is same as that of beam of sun
light streaming through a room
,illuminating airborne dust particles.
This occurance is called tyndall
phenomenon.
 Most useful when examining the
Tyndall phenomenon
 Cells, pigment or proteins in the
aqueous humour reflect the light like a
faint fog.
 To visualise this the slit illuminator is
adjusted to the smallest circular beam
and is projected through the anterior
chamber from a 42° to 90° angle.
 The strongest reflection is possible at
90°.
Optic section
 Optic section is a very thin parallelepiped and
optically cuts a very thin slice of the cornea.
 Axes of illuminating and viewing path intersect
in the area of anterior eye media to be
examined e.g. the individual corneal layers.
 Angle between illuminating and viewing path
is 45 degree.
 Slit length should be kept small to minimize
dazzling the patient.
 With narrow slit the depth and portion of
different objects(penetration depth of
foreign bodies, shape of lens etc) can be
resolved more easily.
 With wider slit their extension and shape
are visible more clearly.
 Magnification: maximum.
 Examination of AC depth is performed by
wider slit width .1-.3mm .
 Used to localize:
Nerve fibers
Blood vessels
Infiltrates
Cataracts
AC depth.
Optical section of lens
1.Corneal scar with wide beam illumination 2.optical section through scar
indicating scar is with in superficial layer of cornea.
Tangential illumination
 Requires that the illumination arm and the
viewing arm be separated by 90 degree.
 Medium –wide beam of moderate height is
used.
 Microscope is pointing straight ahead.
 Magnification of 10x,16x,or 25x are used.
 Observe:
Anterior and posterior cornea
Iris is best viewed without dilation by
this method.
Anterior lens (especially useful for
viewing pseudoexfolation).
Example of tangential illumination (iris).
Specular reflection
 Established by separating the microscope and
slit beam by equal angles from normal to cornea.
 Position of illuminator about 30 degree to one
side and the microscope 30 degree to otherside.
 Angle of illuminator to microscope must be equal
and opposite.
 Angle of light should be moved until a very bright
reflex obtained from corneal surface which is
called zone of specular reflection.
 Irregularities ,deposits ,or excavasation in these
smooth surface will fail to reflect light and these
appears darker than surrounding.
 Under specular reflection anterior corneal
surface appears as white uniform surface and
corneal endothelium takes on a mosaic pattern.
 Used to observe:
 Evaluate general appearance of corneal
endothelium
 Lens surfaces
 Corneal epithelium
Schematic of specularreflection.
Reflection from
front surface endothelium
Indirect illumination
 The beam is focused in an area adjacent to
ocular tissue to be observed.
 Main application:
 Examination of objects in direct vicinity of
corneal areas of reduced transparency e,g,
infiltrates,corneal scars,deposits,epithelial and
stromal defects
 Illumination:
 Narrow to medium slit beam
 Decentred beam
 Magnification: approx. m=12x (depending upon
object size)
Retroillumination
 Formed by reflecting light of slit beam from
a structure more posterior than the
structure under observation.
 A vertical slit beam 1-4mm wide can be
used.
 Purpose:
Place object of regard against a bright
background allowing object to appear
dark or black.
 Used most often in searching for keratic
precipitates and other debris on corneal
endothelium.
 The crystalline lens can also be
retroilluminated for viewing of water
clefts and vacuoles of anterior lens and
posterior subcapsular cataract
 Direct retroillumination from iris:
Used to view corneal pathology.
A moderately wide slit beam is aimed
towards the iris directly behind the
corneal anomaly.
Use magnification of 16x to 25x and
direct the light from 45 degree.
Microscope is directed straight ahead .
Schematic of
direct retroillumination from
the iris.
direct retroillumination from the iris.
 Indirect retroillumination from iris:
Performed as with direct
retroillumination but the beam is
directed to an area of the iris bordering
the portion of iris behind pathology.
It provides dark background allowing
corneal opacities to be viewed with
more contrast.
Observe:
Cornea, angles.
Retroillumination from fundus(red
reflex photography)
 The slit illuminator is positioned in an
almost coaxial position with the
biomicroscope.
 A wide slit beam is decentered and
adjusted to a half circle by using the slit
width and
 The decentred slit beam is projected near
the pupil margin through a dilated pupil.
Schematic of
retroillumination from the
retina.
Example of retroillumination from the reti
Sclerotic scatter
 It is formed by focusing a bright but narrow slit
beam on the limbus and using microscope on
low magnification.
 Such an illumination technique causes cornea
to take on total internal reflection.
 The slit beam should be placed approximately
40-60 degree from the microscope.
 When properly positioned this technique will
produce halo glow of light around the limbus
as the light is transmitted around the cornea.
 Corneal changes or abnormalities can be
visualized by reflecting the scattered light.
 Used to observe:
Central corneal epithelial edema
Corneal abrasions
Corneal nebulae and maculae.
Schematic of
sclerotic scatter. Example of
sclerotic scatter.
Proximal illumination
 This illumination technique is used to
observe internal detail, depth, and
density.
 Use a short,fairly narrow slit beam.
 Place the beam at the border of the
structure or pathology.
 The light will be scattered into the
surrounding tissue, creating a light
background that highlights the edges of
 Depending on the density of the
abnormality, the light from behind may
reflect through, allowing detailed
examination of the internal structure of
the pathology.
 Observe: corneal opacities (edema,
infiltrates, vessels, foreign bodies), lens,
iris
Transillumination
 In transillumination, a structure (in the
eye, the iris) is evaluated by how light
passes through it.
 Iris transillumination:
 This technique also takes advantage of
the red reflex.
 The pupil must be at mid mydriasis (3to
4 mm when light stimulated).
 Place the light source coaxial (directly
in line) with the microscope.
 Use a full circle beam of light equal to
the size of the pupil.
 Project the light through the pupil and
into the eye .
 Focus the microscope on the iris.
 Magnification of 10X to 16X is adequate
 Normally the iris pigment absorbs the
light, but pigmentation defects let the red
fundus light pass through..
 Observe: iris defects (they will glow with
the orange light reflected from the
fundus)
Basic slit lamp examination
 Patient positioning:
Head support unit
Adjust height of table or chair
Adjust height of chin rest such that
patients lateral canthus is aligned with
the mark.
Adjust ocular eyepieces.
Power up
Fixation
Magnification : begin with 6x -10x
magnification
Focusing
Special procedures
Protocol and documentation
Uses of slit lamp biomicroscopy
 Diagnostic:
OCT
FFA
Anterior segment and posterior
segment diseases
Dry eye
 Procedures:
 Applanation
 Tear evaluation
 Pachymetry
 Gonioscopy
 Contact lens fitting
 Therapeutic:
 Laser
 FB removal
 epilation
Anterior and posterior segment disease
evaluation
 Lids and lashes
 Conjunctiva and cornea
 Instillation of fluorescein and BUT
measurement
 Eversion of the lids
 Anterior chamber and angle measurement
 Iris
 Crystalline lens
 Anterior vitreous
Injected conjunctivaMeibomian gland openings
pinquecula,
INSTILLATION OF FLUORESCEIN PALPEBRAL CONJUNCTIVA
EXAMINATION
Evertion of lids
 This technique is used to examine the
inferior and superior palpebral
conjunctiva, particularly in contact lens
wear and when looking for allergic
conjunctival changes, papillae, and
foreign bodies.
 1. Ask the patient to look down and
grasp the superior eyelashes.
 2. Press gently on the superior margin of
the tarsal plate using a cotton swab (or
the index finger of the other hand), and
at the same time pull the eyelashes
upwards.
 3. To evert the lower eyelid, pull the
eyelid down and press under the eyelid
margin while moving finger upwards.
The eyelid will evert over finger.
Meibomian gland evaluation
 With the patient at the biomicroscope, use
white light and medium magnification to
inspect the lower eyelid margins.
 Look for capping of the meibomian gland
orifices (yellow mounds), notching of the
eyelid margins (indentations) and frothing of
the tears on the eyelid margins.
 Pull the lower eyelid down and look for
concretions in the palpebral conjunctiva.
 With mild pressure, press on the eyelid
margins near the eyelashes and watch
the meibomian gland orifices.
 Clear fluid should be expressed.
 Capping of the orifices, a cheesy
secretion on expression and frothing of
the eyelid margins indicates meibomian
gland dysfunction.
CENTRAL RETINA PHOTOGRAPHS
WITH A 90-DIOPTER LENS
 A moderate slit
beam in the
almost coaxial
position gives
the best results.
References
 Clinical procedure in optometry
 Primary care optometry
 Borishs clinical refraction
 Theory and practice of optics and refraction:AK
Khurana
 internet
Slit lump biomicroscope

More Related Content

What's hot

ultrasound biomicroscopy
ultrasound biomicroscopyultrasound biomicroscopy
ultrasound biomicroscopySSSIHMS-PG
 
Slitlamp bimicroscopy
Slitlamp bimicroscopy Slitlamp bimicroscopy
Slitlamp bimicroscopy Obehi Osoata
 
Cylinder prescription guidelines
Cylinder prescription guidelinesCylinder prescription guidelines
Cylinder prescription guidelinesPrashant Patel
 
Optics for ophthalmic residents
Optics for ophthalmic residentsOptics for ophthalmic residents
Optics for ophthalmic residentsLaurence Sullivan
 
Corneal topography wavefront analysis
Corneal topography wavefront analysisCorneal topography wavefront analysis
Corneal topography wavefront analysisikramdr01
 
Glaucoma drainage devices
Glaucoma drainage devicesGlaucoma drainage devices
Glaucoma drainage devicesDinesh Madduri
 
Ophthalmic lens material
Ophthalmic lens materialOphthalmic lens material
Ophthalmic lens materialManoj Mahat
 
Dissociated vertical deviation
Dissociated vertical deviationDissociated vertical deviation
Dissociated vertical deviationdoc_angie Shah
 
Slit Lamp Illumination Techniques
Slit Lamp Illumination TechniquesSlit Lamp Illumination Techniques
Slit Lamp Illumination TechniquesIrina Kezik
 
Keratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKeratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKamal Luitel
 
OPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSESOPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSESGREESHMA G
 
Optics of RGP contact lens
Optics of RGP contact lensOptics of RGP contact lens
Optics of RGP contact lensPabita Dhungel
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refractionsanju_95
 

What's hot (20)

ultrasound biomicroscopy
ultrasound biomicroscopyultrasound biomicroscopy
ultrasound biomicroscopy
 
Slitlamp bimicroscopy
Slitlamp bimicroscopy Slitlamp bimicroscopy
Slitlamp bimicroscopy
 
Cylinder prescription guidelines
Cylinder prescription guidelinesCylinder prescription guidelines
Cylinder prescription guidelines
 
Optics for ophthalmic residents
Optics for ophthalmic residentsOptics for ophthalmic residents
Optics for ophthalmic residents
 
Keratometer
KeratometerKeratometer
Keratometer
 
Bruckner test
Bruckner testBruckner test
Bruckner test
 
Corneal topography wavefront analysis
Corneal topography wavefront analysisCorneal topography wavefront analysis
Corneal topography wavefront analysis
 
Glaucoma drainage devices
Glaucoma drainage devicesGlaucoma drainage devices
Glaucoma drainage devices
 
Ophthalmic lens material
Ophthalmic lens materialOphthalmic lens material
Ophthalmic lens material
 
Dissociated vertical deviation
Dissociated vertical deviationDissociated vertical deviation
Dissociated vertical deviation
 
Slit Lamp Illumination Techniques
Slit Lamp Illumination TechniquesSlit Lamp Illumination Techniques
Slit Lamp Illumination Techniques
 
Absorptive lenses
Absorptive lenses Absorptive lenses
Absorptive lenses
 
Pentacam
PentacamPentacam
Pentacam
 
Keratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKeratometry and Dynamic Retinoscopy
Keratometry and Dynamic Retinoscopy
 
OPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSESOPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSES
 
Orbscan & topo
Orbscan & topoOrbscan & topo
Orbscan & topo
 
Optics of RGP contact lens
Optics of RGP contact lensOptics of RGP contact lens
Optics of RGP contact lens
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Introduction to cl fitting
Introduction to cl fittingIntroduction to cl fitting
Introduction to cl fitting
 
Rosek2 part1
Rosek2 part1Rosek2 part1
Rosek2 part1
 

Similar to Slit lump biomicroscope

Slit lamp biomicroscopy and operating microscope
Slit lamp biomicroscopy and operating microscopeSlit lamp biomicroscopy and operating microscope
Slit lamp biomicroscopy and operating microscopeSAMEEKSHA AGRAWAL
 
Slit lampbiomicroscopy-jinal & dhara
Slit lampbiomicroscopy-jinal & dharaSlit lampbiomicroscopy-jinal & dhara
Slit lampbiomicroscopy-jinal & dharadhara sorathiya
 
slit-lamp-160212142229.pdf
slit-lamp-160212142229.pdfslit-lamp-160212142229.pdf
slit-lamp-160212142229.pdfanju468752
 
Slit Lamp Biomicroscopy.
Slit Lamp Biomicroscopy.Slit Lamp Biomicroscopy.
Slit Lamp Biomicroscopy.Mystic Dider
 
what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...
what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...
what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...Naeem Ahmad
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopySSSIHMS-PG
 
slit lamp photography.pdf
slit lamp photography.pdfslit lamp photography.pdf
slit lamp photography.pdfMusabFathallah
 
Slit lamp presentation 2015
Slit lamp presentation 2015Slit lamp presentation 2015
Slit lamp presentation 2015amanimohsen1
 
Lenses in ophthalmology
Lenses in ophthalmologyLenses in ophthalmology
Lenses in ophthalmologyReshma Peter
 
lensesinophthalmology-171212163049 (1).pdf
lensesinophthalmology-171212163049 (1).pdflensesinophthalmology-171212163049 (1).pdf
lensesinophthalmology-171212163049 (1).pdfDrSunandaMajumder
 

Similar to Slit lump biomicroscope (20)

Slit lamp biomicroscopy and operating microscope
Slit lamp biomicroscopy and operating microscopeSlit lamp biomicroscopy and operating microscope
Slit lamp biomicroscopy and operating microscope
 
Slit lampbiomicroscopy-jinal & dhara
Slit lampbiomicroscopy-jinal & dharaSlit lampbiomicroscopy-jinal & dhara
Slit lampbiomicroscopy-jinal & dhara
 
slit-lamp-160212142229.pdf
slit-lamp-160212142229.pdfslit-lamp-160212142229.pdf
slit-lamp-160212142229.pdf
 
Slit lamp ..
Slit lamp ..Slit lamp ..
Slit lamp ..
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Slit lamp in Ophthalmology
Slit lamp in OphthalmologySlit lamp in Ophthalmology
Slit lamp in Ophthalmology
 
Slit lamp biomicroscope
Slit lamp biomicroscopeSlit lamp biomicroscope
Slit lamp biomicroscope
 
Slit Lamp Biomicroscopy.
Slit Lamp Biomicroscopy.Slit Lamp Biomicroscopy.
Slit Lamp Biomicroscopy.
 
SLIT LAMP
SLIT LAMPSLIT LAMP
SLIT LAMP
 
Slit lamp biomicroscope
Slit lamp biomicroscopeSlit lamp biomicroscope
Slit lamp biomicroscope
 
what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...
what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...
what is slit lamp Biomicroscopy | What is a slit lamp exam used for? |Differe...
 
SLIT LAMP BIOMICROSCOPY.pptx
SLIT LAMP BIOMICROSCOPY.pptxSLIT LAMP BIOMICROSCOPY.pptx
SLIT LAMP BIOMICROSCOPY.pptx
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Slit lamp examination
Slit lamp examinationSlit lamp examination
Slit lamp examination
 
slit lamp photography.pdf
slit lamp photography.pdfslit lamp photography.pdf
slit lamp photography.pdf
 
Slit lamp presentation 2015
Slit lamp presentation 2015Slit lamp presentation 2015
Slit lamp presentation 2015
 
Sl exam pt i
Sl exam pt iSl exam pt i
Sl exam pt i
 
basics of microscope - part 1
basics of microscope - part 1basics of microscope - part 1
basics of microscope - part 1
 
Lenses in ophthalmology
Lenses in ophthalmologyLenses in ophthalmology
Lenses in ophthalmology
 
lensesinophthalmology-171212163049 (1).pdf
lensesinophthalmology-171212163049 (1).pdflensesinophthalmology-171212163049 (1).pdf
lensesinophthalmology-171212163049 (1).pdf
 

More from C L GUPTA EYE INSTITUTE MORADABAD UTTER PRADESH

More from C L GUPTA EYE INSTITUTE MORADABAD UTTER PRADESH (20)

Astigmatism
Astigmatism Astigmatism
Astigmatism
 
Tonometry ppt
Tonometry pptTonometry ppt
Tonometry ppt
 
anatomy of sclera
anatomy of sclera anatomy of sclera
anatomy of sclera
 
Pupil ppt
Pupil  pptPupil  ppt
Pupil ppt
 
anatomy And Physiology of tear film
anatomy And Physiology of tear film anatomy And Physiology of tear film
anatomy And Physiology of tear film
 
cornea and layers
cornea and layers cornea and layers
cornea and layers
 
Limbus
LimbusLimbus
Limbus
 
anatomy And Physiology of lacrimal secretions
anatomy And Physiology of lacrimal secretions anatomy And Physiology of lacrimal secretions
anatomy And Physiology of lacrimal secretions
 
extra ocular muscles
extra ocular muscles extra ocular muscles
extra ocular muscles
 
Diseases of sclera
Diseases of sclera Diseases of sclera
Diseases of sclera
 
Corneal transparency
Corneal transparency Corneal transparency
Corneal transparency
 
Conjunctiva anatomy and physiology
Conjunctiva anatomy and physiology Conjunctiva anatomy and physiology
Conjunctiva anatomy and physiology
 
Color vision deficiency and ishiharas test
Color vision deficiency and ishiharas test Color vision deficiency and ishiharas test
Color vision deficiency and ishiharas test
 
Anatomy of eye
Anatomy of eye Anatomy of eye
Anatomy of eye
 
Anatomy and physiology of aqueous humor
Anatomy and physiology of aqueous humor Anatomy and physiology of aqueous humor
Anatomy and physiology of aqueous humor
 
Vitreous humour
Vitreous humourVitreous humour
Vitreous humour
 
Vitamins all
Vitamins allVitamins all
Vitamins all
 
Tear film
Tear filmTear film
Tear film
 
Tca cycle
Tca cycleTca cycle
Tca cycle
 
Synthesis
SynthesisSynthesis
Synthesis
 

Recently uploaded

College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Timedelhimodelshub1
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed RuleShelby Lewis
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowHyderabad Call Girls Services
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...soniya singh
 
Call Girls Madhapur 7001305949 all area service COD available Any Time
Call Girls Madhapur 7001305949 all area service COD available Any TimeCall Girls Madhapur 7001305949 all area service COD available Any Time
Call Girls Madhapur 7001305949 all area service COD available Any Timedelhimodelshub1
 
Call Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service GurgaonCall Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service GurgaonCall Girls Service Gurgaon
 
Call Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any TimeCall Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any Timedelhimodelshub1
 
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service HyderabadCall Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service GoaRussian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goanarwatsonia7
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknowgragteena
 
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...ggsonu500
 
Call Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any TimeCall Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any Timedelhimodelshub1
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliHigh Profile Call Girls Chandigarh Aarushi
 
Call Girls Dilsukhnagar 7001305949 all area service COD available Any Time
Call Girls Dilsukhnagar 7001305949 all area service COD available Any TimeCall Girls Dilsukhnagar 7001305949 all area service COD available Any Time
Call Girls Dilsukhnagar 7001305949 all area service COD available Any Timedelhimodelshub1
 

Recently uploaded (20)

College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Time
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
 
Call Girls Madhapur 7001305949 all area service COD available Any Time
Call Girls Madhapur 7001305949 all area service COD available Any TimeCall Girls Madhapur 7001305949 all area service COD available Any Time
Call Girls Madhapur 7001305949 all area service COD available Any Time
 
Call Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service GurgaonCall Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
Call Girls Gurgaon Parul 9711199012 Independent Escort Service Gurgaon
 
Call Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any TimeCall Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any Time
 
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service HyderabadCall Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
Call Girl Hyderabad Madhuri 9907093804 Independent Escort Service Hyderabad
 
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service GoaRussian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
 
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service GuwahatiCall Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
 
Russian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your bookingRussian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your booking
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
 
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
 
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 68 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
 
Call Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any TimeCall Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any Time
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
 
Call Girls Dilsukhnagar 7001305949 all area service COD available Any Time
Call Girls Dilsukhnagar 7001305949 all area service COD available Any TimeCall Girls Dilsukhnagar 7001305949 all area service COD available Any Time
Call Girls Dilsukhnagar 7001305949 all area service COD available Any Time
 

Slit lump biomicroscope

  • 1. MANOJ ARYAL B . Optometry IOM,MMC Slit lamp Biomicroscopy (Instrumentation,Principle, Illumination and Uses) optometris t
  • 2. Introduction  Biomicroscope derives its name from the fact that it enables the practitioner to observe the living tissue of eye under magnification.  It not only provides magnified view of every part of eye but also allows quantitative measurements and photography of every part for documentation.
  • 3. • The lamp facilitates an examination which looks at anterior segment, or frontal structures, of the human eye, which includes the –Eyelid –Cornea –Sclera –Conjunctiva –Iris –Aqueous –Natural crystalline lens and –Anterior vitreous.
  • 4. Important historical landmarks  De Wecker 1863 devised a portable ophthalmomicroscope .  Albert and Greenough 1891,developed a binocular microscope which provided stereoscopic view.  Gullstrand ,1911 introduced the illumination system which had for the first time a slit diapharm in it  Therefore Gullstrand is credited with the invention of slit lamp.
  • 6. TYPES  There are 2 types of slit lamp biomicroscope 1)Zeiss slit lamp biomicroscope 2)Haag streit slit lamp biomicroscope  In Zeiss type light source is at the base of the instrument while in Haag streit type it is at the top of the instrument.
  • 7. Zeiss slit lamp biomicroscope Haag streit slit lamp biomicroscope
  • 8. PRINCIPLE  A "slit" beam of very bright light produced by lamp. This beam is focused on to the eye which is then viewed under magnification with a microscope
  • 9.
  • 10. Instrumentation  Operational components of slit lamp biomicroscope essentially consist of: Illumination system Observation system Mechanical system
  • 11.  Illumination system  It consist of: A bright ,focal source of light with a slit mechanism  Provides an illumination of 2*10^5 to 4*10^5 lux.  The beam of light can be changed in intensity,height,width,direction or angle and color during the examination with the flick of lever.
  • 12. Condensing lens system: Consist of a couple of planoconvex lenses with their convex surface in apposition. Slit and other diapharm:  Height and width of slit can be varied by using knobs.
  • 13.  Projection lens:  Form an image of slit at eye.  Advantages,  1.keeps the aberration of lens down.  2.increase the depth of focus of slit.
  • 14.  Reflecting mirrors and prisms  Filters Yellow barrier filter Red free filter Neutral density filter Cobalt blue filter diffuser
  • 15. Observation system(microscope)  Observation system is essentially a compound microscope composed of two optical elements 1.an objective ,2.an eyepiece  It presents to the observer an enlarged image of a near object.  The objective lens consists of two planoconvex lenses with their convexities put together providing a composite power of +22D.  Microscope is binocular i.e. it has two eyepieces giving binocular observer a
  • 16.  The eye piece has a lens of +10D.  To overcome the problem of inverted image produced by compound microscope ,slit lamp microscope uses a pair of prisms b/w the objective and eyepiece to reinvert the image.  Most slit lamp provide a range of magnification from 6x to 40x
  • 17. Mechanical system  Joystick arrangement  Movement of microscope and illumination system towards and away from the eye and from side and side is achieved via joystick arrangement. Up and down movement arrangement Obtained via some sort or screw devices. Patient support arrangement Vertically movable chin rest and the
  • 18. Fixation target: A movable fixation target greatly faciliates the examination under some conditions. Mechanical coupling : Provides a coupling of microscope and the illumination system along a common axis of rotation that coincides their focal planes.  This ensures that light falls on the point where the microscope is focused  Has advantages when using the slit lamp for routine examination of anterior
  • 19.  Magnification control : Including two or pair of readily changeable objective lenses and two sets of eyepieces.  An on and off switch and illumination control .
  • 20. Topcon slit lamp model SL-3E Light beam is controlled by knobs Joy stick arrangement Chin rest Reflecting mirror biomicroscope Illumination control
  • 21. Magnification may be changed by flipping a lever... Changing filters. biomicroscope Patient positioning Alignmen t mark Microscope and light source rotate indepedently
  • 22. Filters used in slit lamp biomicroscopy  Cobalt blue filter  Used in conjunction with fluorescein stain  Dye pods in area where the corneal epithelium is broken or absent.  The dye absorbs blue light and emits green.  Uses:  Ocular staining  RGP lenses fitting  Tear layer
  • 23.  Red free(green)filter: Obscure any thing that is red hence the red free light , thus blood vessels or haemorrhages appears black. This increases contrast ,revealing the path and pattern of inflammed blood vessels. Fleischer ring can also be viewed satisfactorily with the red green filter.
  • 24.
  • 25. Illumination techniques  Includes  Diffuse illumination  Direct illumination  Parallilepiped  Optic section  Conical(pinpoint)  Tangential  Specular reflection  Indirect illumination  Retro-illumination  Sclerotic scatter  Transillumination  Proximal illumination
  • 26. Diffuse illumination  Angle between microscope and illumination system should be 30-45 degree.  Slit width should be widest.  Filter to be used is diffusing filter.  Magnification: low to medium  Illumination: medium to high.
  • 27. Applications: General view of anterior of eye: lids,lashes,sclera,cornea ,iris, pupil, Gross pathology and media opacities Contact lens fitting. Assessment of lachrymal reflex.
  • 28. Optics of diffuse illumination Diffuse illumination with slit beam and background illumination
  • 29. Direct illumination  Involves placing the light source at an angle of about 40-50 degree from microscope.  This arrangement permits both light beam and microscope to be sharply focused on the ocular tissue being observed.  Wide beam direct illumination is commonly used as a preliminary technique to evaluate large area.
  • 30.  it is particularly suitable for assessment of cataracts,scars,nerves,vessels etc.  It is also of great importance for the determination of stabilization of axis of toric contact lens.
  • 31.  Parallelepiped: Constructed by narrowing the beam to 1- 2mm in width to illuminate a rectangular area of cornea. Microscope is placed directly in front of patients cornea. Light source is approximately 45 degree from straight ahead position.
  • 32. Applications: Used to detect and examine corneal structures and defects. Used to detect corneal striae that develop when corneal edema occurs with hydrogel lens wear and in keratoconus. Higher magnification than that used with wide beam illumination is preferred to evaluate both depth and extent of corneal ,scarring or foreign bodies.
  • 33.
  • 34. Conical beam(pinpoint)  Produced by narrowing the vertical height of a parallelepiped to produce a small circular or square spot of light.  Light source is 45-60 degree temporally and directed into pupil.  Biomicroscope: directly in front of eye.  Magnification: high(16-25x)  Intensity of light source to heighest setting.
  • 35.  Focusing: Beam is focused between cornea and anterior lens surface and dark zone between cornea and anterior lens observed. Principle is same as that of beam of sun light streaming through a room ,illuminating airborne dust particles. This occurance is called tyndall phenomenon.  Most useful when examining the
  • 36. Tyndall phenomenon  Cells, pigment or proteins in the aqueous humour reflect the light like a faint fog.  To visualise this the slit illuminator is adjusted to the smallest circular beam and is projected through the anterior chamber from a 42° to 90° angle.  The strongest reflection is possible at 90°.
  • 37.
  • 38. Optic section  Optic section is a very thin parallelepiped and optically cuts a very thin slice of the cornea.  Axes of illuminating and viewing path intersect in the area of anterior eye media to be examined e.g. the individual corneal layers.  Angle between illuminating and viewing path is 45 degree.  Slit length should be kept small to minimize dazzling the patient.
  • 39.  With narrow slit the depth and portion of different objects(penetration depth of foreign bodies, shape of lens etc) can be resolved more easily.  With wider slit their extension and shape are visible more clearly.  Magnification: maximum.  Examination of AC depth is performed by wider slit width .1-.3mm .
  • 40.  Used to localize: Nerve fibers Blood vessels Infiltrates Cataracts AC depth.
  • 41. Optical section of lens 1.Corneal scar with wide beam illumination 2.optical section through scar indicating scar is with in superficial layer of cornea.
  • 42. Tangential illumination  Requires that the illumination arm and the viewing arm be separated by 90 degree.  Medium –wide beam of moderate height is used.  Microscope is pointing straight ahead.  Magnification of 10x,16x,or 25x are used.
  • 43.  Observe: Anterior and posterior cornea Iris is best viewed without dilation by this method. Anterior lens (especially useful for viewing pseudoexfolation).
  • 44. Example of tangential illumination (iris).
  • 45. Specular reflection  Established by separating the microscope and slit beam by equal angles from normal to cornea.  Position of illuminator about 30 degree to one side and the microscope 30 degree to otherside.  Angle of illuminator to microscope must be equal and opposite.  Angle of light should be moved until a very bright reflex obtained from corneal surface which is called zone of specular reflection.
  • 46.  Irregularities ,deposits ,or excavasation in these smooth surface will fail to reflect light and these appears darker than surrounding.  Under specular reflection anterior corneal surface appears as white uniform surface and corneal endothelium takes on a mosaic pattern.  Used to observe:  Evaluate general appearance of corneal endothelium  Lens surfaces  Corneal epithelium
  • 47. Schematic of specularreflection. Reflection from front surface endothelium
  • 48. Indirect illumination  The beam is focused in an area adjacent to ocular tissue to be observed.  Main application:  Examination of objects in direct vicinity of corneal areas of reduced transparency e,g, infiltrates,corneal scars,deposits,epithelial and stromal defects  Illumination:  Narrow to medium slit beam  Decentred beam  Magnification: approx. m=12x (depending upon object size)
  • 49. Retroillumination  Formed by reflecting light of slit beam from a structure more posterior than the structure under observation.  A vertical slit beam 1-4mm wide can be used.  Purpose: Place object of regard against a bright background allowing object to appear dark or black.
  • 50.  Used most often in searching for keratic precipitates and other debris on corneal endothelium.  The crystalline lens can also be retroilluminated for viewing of water clefts and vacuoles of anterior lens and posterior subcapsular cataract
  • 51.  Direct retroillumination from iris: Used to view corneal pathology. A moderately wide slit beam is aimed towards the iris directly behind the corneal anomaly. Use magnification of 16x to 25x and direct the light from 45 degree. Microscope is directed straight ahead .
  • 52. Schematic of direct retroillumination from the iris. direct retroillumination from the iris.
  • 53.  Indirect retroillumination from iris: Performed as with direct retroillumination but the beam is directed to an area of the iris bordering the portion of iris behind pathology. It provides dark background allowing corneal opacities to be viewed with more contrast. Observe: Cornea, angles.
  • 54.
  • 55. Retroillumination from fundus(red reflex photography)  The slit illuminator is positioned in an almost coaxial position with the biomicroscope.  A wide slit beam is decentered and adjusted to a half circle by using the slit width and  The decentred slit beam is projected near the pupil margin through a dilated pupil.
  • 56. Schematic of retroillumination from the retina. Example of retroillumination from the reti
  • 57. Sclerotic scatter  It is formed by focusing a bright but narrow slit beam on the limbus and using microscope on low magnification.  Such an illumination technique causes cornea to take on total internal reflection.  The slit beam should be placed approximately 40-60 degree from the microscope.  When properly positioned this technique will produce halo glow of light around the limbus as the light is transmitted around the cornea.  Corneal changes or abnormalities can be visualized by reflecting the scattered light.
  • 58.  Used to observe: Central corneal epithelial edema Corneal abrasions Corneal nebulae and maculae.
  • 59. Schematic of sclerotic scatter. Example of sclerotic scatter.
  • 60. Proximal illumination  This illumination technique is used to observe internal detail, depth, and density.  Use a short,fairly narrow slit beam.  Place the beam at the border of the structure or pathology.  The light will be scattered into the surrounding tissue, creating a light background that highlights the edges of
  • 61.  Depending on the density of the abnormality, the light from behind may reflect through, allowing detailed examination of the internal structure of the pathology.  Observe: corneal opacities (edema, infiltrates, vessels, foreign bodies), lens, iris
  • 62. Transillumination  In transillumination, a structure (in the eye, the iris) is evaluated by how light passes through it.  Iris transillumination:  This technique also takes advantage of the red reflex.  The pupil must be at mid mydriasis (3to 4 mm when light stimulated).  Place the light source coaxial (directly in line) with the microscope.
  • 63.  Use a full circle beam of light equal to the size of the pupil.  Project the light through the pupil and into the eye .  Focus the microscope on the iris.  Magnification of 10X to 16X is adequate  Normally the iris pigment absorbs the light, but pigmentation defects let the red fundus light pass through..  Observe: iris defects (they will glow with the orange light reflected from the fundus)
  • 64.
  • 65. Basic slit lamp examination  Patient positioning: Head support unit Adjust height of table or chair Adjust height of chin rest such that patients lateral canthus is aligned with the mark. Adjust ocular eyepieces.
  • 66. Power up Fixation Magnification : begin with 6x -10x magnification Focusing Special procedures Protocol and documentation
  • 67. Uses of slit lamp biomicroscopy  Diagnostic: OCT FFA Anterior segment and posterior segment diseases Dry eye
  • 68.  Procedures:  Applanation  Tear evaluation  Pachymetry  Gonioscopy  Contact lens fitting  Therapeutic:  Laser  FB removal  epilation
  • 69. Anterior and posterior segment disease evaluation  Lids and lashes  Conjunctiva and cornea  Instillation of fluorescein and BUT measurement  Eversion of the lids  Anterior chamber and angle measurement  Iris  Crystalline lens  Anterior vitreous
  • 70. Injected conjunctivaMeibomian gland openings pinquecula, INSTILLATION OF FLUORESCEIN PALPEBRAL CONJUNCTIVA EXAMINATION
  • 71. Evertion of lids  This technique is used to examine the inferior and superior palpebral conjunctiva, particularly in contact lens wear and when looking for allergic conjunctival changes, papillae, and foreign bodies.  1. Ask the patient to look down and grasp the superior eyelashes.
  • 72.  2. Press gently on the superior margin of the tarsal plate using a cotton swab (or the index finger of the other hand), and at the same time pull the eyelashes upwards.  3. To evert the lower eyelid, pull the eyelid down and press under the eyelid margin while moving finger upwards. The eyelid will evert over finger.
  • 73. Meibomian gland evaluation  With the patient at the biomicroscope, use white light and medium magnification to inspect the lower eyelid margins.  Look for capping of the meibomian gland orifices (yellow mounds), notching of the eyelid margins (indentations) and frothing of the tears on the eyelid margins.  Pull the lower eyelid down and look for concretions in the palpebral conjunctiva.
  • 74.  With mild pressure, press on the eyelid margins near the eyelashes and watch the meibomian gland orifices.  Clear fluid should be expressed.  Capping of the orifices, a cheesy secretion on expression and frothing of the eyelid margins indicates meibomian gland dysfunction.
  • 75. CENTRAL RETINA PHOTOGRAPHS WITH A 90-DIOPTER LENS  A moderate slit beam in the almost coaxial position gives the best results.
  • 76. References  Clinical procedure in optometry  Primary care optometry  Borishs clinical refraction  Theory and practice of optics and refraction:AK Khurana  internet