Temporary Pacemakers-
Objectives



Describe three common temporary pacemaker
settings and the situation or underlying rhythm
appropriate for each setting.
Discuss common complications of temporary
pacing and the assessments and
interventions needed to troubleshoot.
Describe general care and maintenance of a
temporary pacemaker.
Uses for Temporary Pacing


 To stimulate myocardium to contract in
absence of intrinsic rhythm.
To establish adequate cardiac output and
blood pressure to ensure tissue perfusion to
vital organs.
To reduce possibility of ventricular
dysrhythmias in presence of bradycardia.
Uses cont.


 To supplement inadequate rhythm in
presence of transient decrease in heart
rate.
To allow administration of
medications(beta-blockers) to treat
ischemia or tachyarrhythmias in presence
of conduction system dysfunction or
bradycardia.
AACN Procedure Manual for Critical Care, pg. 349.





Complete heart block
Symptomatic Second
degree heart block
Symptomatic bradycardia
or bradydysrhythmias
Dysrhythmias
complicating AMI
Alternating BBB


 New BBB with
transient CHB
Sinus node
dysfunction
Treatment of
bradycardia-
tachycardia syndrome
Rhythms Requiring Temporary Pacing
Rhythms Requiring Temp Pacing (cont.)




Ventricular standstill or
cardiac arrest
Long QT syndrome with
ventricular
dysrhythmias
Medication toxicity
Post-op cardiac surgery
 Prophylaxis with
cardiac diagnostic or
interventional
procedures
*AACN Procedure Manual for Critical Care, pg. 349.
3 Primary Methods @ Cone
 Transvenous-used mostly in CICU.
 Epicardial- used in SICU and on 2000.
 Transcutaneous- emergent.
Epicardial Pacing
 Method of stimulating the myocardium
through use of Teflon-coated, unipolar
stainless steel wires that are sutured loosely
to epicardium after surgery.
The NASPE/BPEG Generic (NBG) Code
Chamber(s)
Sensed
Response to
Sensing
Programmability
Rate Modulation
Antitachyarrhythmia
Function(s)
Position I
Category Chamber(s)
Paced
II III IV V
O = None
A = Atrium
V = Ventricle
D = Dual (A+V)
O = None
A = Atrium
V = Ventricle
D = Dual (A+V)
O = None
T = Triggered
I = Inhibited
D = Dual
(T+I)
O = None
P = Simple Programmable
M = Multiprogrammable
C = Communicating
R = Rate Modulation
O = None
P = Pacing
S = Shock
D = Dual
(P+S)
Manufacturer’sS = Single
Designation (A or V)
Only
S = Single
(A or V)
Note: Positions I through III are used exclusively for antibradyarrhythmia function
Common Settings
 AAI



Atrium is paced when necessary
Atrial activity is being sensed by pacer.
I = when atrial activity is sensed the pacer
does not fire.
 VVI



Ventricle is paced when necessary.
Ventricular activity is being sensed by pacer.
I= When ventricular activity is sensed, the
pacer does not fire.
 DDD


 Atrium and ventricle are paced when
necessary (not necessarily at the same time).
Atrial and ventricular activity is being sensed
by pacer.
D= Triggers pacer when atrial or ventricular
activity is not sensed. Inhibits pacer when
atrial or ventricular activity is sensed.
TEMPORARY PACING
DEFINITIONS
(from AACN procedure manual for Critical Care, 5th edition,
2005, pg.325)
Sensing

 The ability of the pacemaker to detect
intrinsic myocardial electrical activity.
The pacemaker either is inhibited from
delivering a stimulus or initiates an
electrical impulse.
Pulse Generation


 Occurs when the pacemaker produces a
programmed electrical current for a set
duration.
This energy travels through the epicardial
or transvenous wires to the myocardium.
This is known as pacemaker firing and
usually produces a line or spike on
the EKG recording.
Capture

 Refers to the successful stimulation of the
myocardium (by the pacemaker impulse)
that results in depolarization.
Evidenced on EKG by a pacemaker spike
followed either by an atrial or
ventricular complex, depending on
chamber being paced.
Synchronous Pacing



 Pacemaker coordinates with intrinsic
cardiac activity.
Able to sense the patient’s inherent cardiac
activity.
Will inhibit or trigger a stimulus as needed.
Examples: AAI, VVI, DDD
Asynchronous Pacing



Pacemaker functions at a fixed rate
regardless of patient’s underlying rhythm.
It is not able to sense the patient’s
underlying rhythm.
Examples: AOO, VOO, DOO
Failure to Pace
 Pacemaker does not discharge a pacing
stimulus at it’s programmed time to the
myocardium.
 Evidenced by absence of a pacing spike on
EKG where expected.
Sensing Failure
 Pacemaker either detected extraneous
signals that mimic intrinsic cardiac
activity(oversensing) or did not accurately
identify intrinsic activity(undersensing).
Oversensing
 Is recognized on EKG by pauses where
pacing beats were expected and
prolongation of the interval between paced
beats.
Oversensing
 Myopotential inhibition
 EMI
 T-waves outside of refractory period
 Dislodged/fractured lead
 Inappropriate sensitivity setting
Undersensing
 Recognized on EKG by inappropriate
pacemaker spikes relative to the intrinsic
electrical activity(pacemaker spikes
occurring within p wave, QRS complex, or
t wave) and shortened distances between
paced beats.
Undersensing
 Battery depletion
 Decreased QRS voltage
 Fusion beat
 Dislodged/fractured lead
 Inappropriate sensitivity setting
Failure to Capture


Pacemaker has delivered a pacing stimulus
that was unable to initiate depolarization
of the myocardium and subsequent
myocardial contraction.
Evidenced on EKG by pacemaker spikes
that are not followed by a p wave for atrial
pacing or a QRS complex for ventricular
pacing.
Loss of Capture
Feature of the Temporary DDD Pacemaker
2.5
1.25
0.5
Sensitivity
Medtronic 5388 Temporary
Pacemaker
*The following slides are courtesy of Medtronics*
Model 5388 Dual Chamber Temporary Pacemaker
1. Pace/Sense LEDs
2. Lock/Unlock Key
3. Lock Indicators
4. Rate Dial
5. Atrial Output Dial
6. Ventricular Output Dial
7. Menu Parameter Dial
8. Parameter Selection Key
9. Menu Selection Key
10. Pause Key
11. Power On Key
12. Power Off Key
13. Emergency/Asynchronous
Pacing Key
14. Lower Screen
15. Ventricular Output
Graphics
16. Atrial Output Graphics
17. Upper Screen
18. Rate Graphics
19. Setup Indicators
20. DDI Indicator
21. Low Battery Indicator
22. Setup Labels
Cable to Device Connections
Emergency Connections
Low Battery Indicator Battery Replacement
Off / On Keys Values at Power-On
• RATE
• A OUTPUT
• V OUTPUT
• A SENSITIVITY
80 ppm
10 mA
10 mA
0.5 mV
• V SENSITIVITY
• UPPER RATE
• AV INTERVAL
2.0 mV
110 ppm
170 ms
• PVARP 300 ms
Indicators
Pace/Sense Indicators
Low Battery
Indicator DDI Mode
Indicator
Setup
Indicators
Padlock
Indicator
Lock / Unlock
Padlock
Icon
Flashing Key
Icon
Lock/Unlock Key
Lock Indicator
Emergency Key Emergency Pacing Values
RATE Current setting
• A OUTPUT
(or 80 ppm if
device was off)
20 mA
• V OUTPUT
• A SENSITIVITY
• V SENSITIVITY
25 mA
ASYNC
ASYNC
• AV INTERVAL Current setting
(or 170 ms if
device was off)
Pause Key
Rate and Output Adjustments
Ventricular Output Dial
Atrial Output Dial
Rate Dial
Lower Screen Menus
Menu 1: Pacing
Parameters
Menu 2: Rate-Based
Pacing Parameters
Menu 3: Rapid
Atrial Pacing
Menu M:
Dial-A-Mode
Sensitivity Adjustments
Rate-Dependent Parameters
Rate Dependent Parameter Adjustments
PVARP Auto/Manual Setting
Upper Rate AV Interval
Rapid Atrial Pacing
Dial-A-Mode
Want to see:


 Ekg tracing shows paced rhythm consistent
with set parameters.
Patient exhibits hemodynamic stability.
Pacemaker leads/wires are isolated from
other electrical equipment by maintaining
secure connections into the pulse
generator.
Care and Maintenance


 Cleanse around insertion site with CHG or
Betadine.
Curl epicardial wires and use paper tape to
tape to abdomen in horizontal position.
Cleanse around temporary transvenous site
with CHG and place sterile dressing.
Complications




 Over/undersensing
causing competition
Failure to capture
Hiccups
Phlebitis, thrombosis,
embolism or
bacteremia
Ventricular
dysrhythmias





Pneumo/hemothorax
Myocardial
perforation & cardiac
tamponade
Air embolism
Lead dislodgement
Failure to pace
Troubleshooting:









Always check patient first!
Be sure pacer is on!!
Change battery.
Asynchronous pacing-only if no intrinsic rhythm
Be sure connections are secure.
Evaluate sensitivity threshold.
Increase MA.
Reverse polarity of wires/change cables or
generator.
Check rhythm. If patient goes into Afib, AAI will
not provide pacing. Go to VVI pacing.
Documentation




Patient & family education
Date/time pacing initiated
Description of events warranting
intervention
Vital signs & hemodynamic parameters
before & after pacing initiated (response to
pacing)
Documentation (cont.)




 EKG monitoring strip recording before &
after pacing initiated
Pacemaker settings: mode,
rate, output(MA), sensitivity
Medications administered and response
Adjustments to pacemaker
Date/time pacing discontinued
Resources:
•
•
•
AACN Procedure Manual for Critical Care,
Lynn-McHale Wiegand, Debra J. and Carlson,
Karen K, eds.Chapters 48-50, pp. 333-361. 2005.
“Pacemaker (Temporary) Insertion and Care.”
Nursing Clinical Policies and
Procedures/Cardiovascular Care, Moses Cone
Health System Homepage.
Wiederhold, Richard. Electrocardiography: The
Monitoring and Diagnostic Leads. pp. 77-83,
1999.

TPI TEMPORARY PACING INDICATIONS TECHNIQUE.pptx

  • 1.
  • 2.
    Objectives    Describe three commontemporary pacemaker settings and the situation or underlying rhythm appropriate for each setting. Discuss common complications of temporary pacing and the assessments and interventions needed to troubleshoot. Describe general care and maintenance of a temporary pacemaker.
  • 3.
    Uses for TemporaryPacing    To stimulate myocardium to contract in absence of intrinsic rhythm. To establish adequate cardiac output and blood pressure to ensure tissue perfusion to vital organs. To reduce possibility of ventricular dysrhythmias in presence of bradycardia.
  • 4.
    Uses cont.    Tosupplement inadequate rhythm in presence of transient decrease in heart rate. To allow administration of medications(beta-blockers) to treat ischemia or tachyarrhythmias in presence of conduction system dysfunction or bradycardia. AACN Procedure Manual for Critical Care, pg. 349.
  • 5.
         Complete heart block SymptomaticSecond degree heart block Symptomatic bradycardia or bradydysrhythmias Dysrhythmias complicating AMI Alternating BBB    New BBB with transient CHB Sinus node dysfunction Treatment of bradycardia- tachycardia syndrome Rhythms Requiring Temporary Pacing
  • 6.
    Rhythms Requiring TempPacing (cont.)     Ventricular standstill or cardiac arrest Long QT syndrome with ventricular dysrhythmias Medication toxicity Post-op cardiac surgery  Prophylaxis with cardiac diagnostic or interventional procedures *AACN Procedure Manual for Critical Care, pg. 349.
  • 7.
    3 Primary Methods@ Cone  Transvenous-used mostly in CICU.  Epicardial- used in SICU and on 2000.  Transcutaneous- emergent.
  • 8.
    Epicardial Pacing  Methodof stimulating the myocardium through use of Teflon-coated, unipolar stainless steel wires that are sutured loosely to epicardium after surgery.
  • 9.
    The NASPE/BPEG Generic(NBG) Code Chamber(s) Sensed Response to Sensing Programmability Rate Modulation Antitachyarrhythmia Function(s) Position I Category Chamber(s) Paced II III IV V O = None A = Atrium V = Ventricle D = Dual (A+V) O = None A = Atrium V = Ventricle D = Dual (A+V) O = None T = Triggered I = Inhibited D = Dual (T+I) O = None P = Simple Programmable M = Multiprogrammable C = Communicating R = Rate Modulation O = None P = Pacing S = Shock D = Dual (P+S) Manufacturer’sS = Single Designation (A or V) Only S = Single (A or V) Note: Positions I through III are used exclusively for antibradyarrhythmia function
  • 10.
    Common Settings  AAI    Atriumis paced when necessary Atrial activity is being sensed by pacer. I = when atrial activity is sensed the pacer does not fire.
  • 11.
     VVI    Ventricle ispaced when necessary. Ventricular activity is being sensed by pacer. I= When ventricular activity is sensed, the pacer does not fire.
  • 12.
     DDD    Atriumand ventricle are paced when necessary (not necessarily at the same time). Atrial and ventricular activity is being sensed by pacer. D= Triggers pacer when atrial or ventricular activity is not sensed. Inhibits pacer when atrial or ventricular activity is sensed.
  • 13.
    TEMPORARY PACING DEFINITIONS (from AACNprocedure manual for Critical Care, 5th edition, 2005, pg.325)
  • 14.
    Sensing   The abilityof the pacemaker to detect intrinsic myocardial electrical activity. The pacemaker either is inhibited from delivering a stimulus or initiates an electrical impulse.
  • 15.
    Pulse Generation    Occurswhen the pacemaker produces a programmed electrical current for a set duration. This energy travels through the epicardial or transvenous wires to the myocardium. This is known as pacemaker firing and usually produces a line or spike on the EKG recording.
  • 16.
    Capture   Refers tothe successful stimulation of the myocardium (by the pacemaker impulse) that results in depolarization. Evidenced on EKG by a pacemaker spike followed either by an atrial or ventricular complex, depending on chamber being paced.
  • 17.
    Synchronous Pacing     Pacemakercoordinates with intrinsic cardiac activity. Able to sense the patient’s inherent cardiac activity. Will inhibit or trigger a stimulus as needed. Examples: AAI, VVI, DDD
  • 18.
    Asynchronous Pacing    Pacemaker functionsat a fixed rate regardless of patient’s underlying rhythm. It is not able to sense the patient’s underlying rhythm. Examples: AOO, VOO, DOO
  • 19.
    Failure to Pace Pacemaker does not discharge a pacing stimulus at it’s programmed time to the myocardium.  Evidenced by absence of a pacing spike on EKG where expected.
  • 20.
    Sensing Failure  Pacemakereither detected extraneous signals that mimic intrinsic cardiac activity(oversensing) or did not accurately identify intrinsic activity(undersensing).
  • 21.
    Oversensing  Is recognizedon EKG by pauses where pacing beats were expected and prolongation of the interval between paced beats.
  • 22.
    Oversensing  Myopotential inhibition EMI  T-waves outside of refractory period  Dislodged/fractured lead  Inappropriate sensitivity setting
  • 23.
    Undersensing  Recognized onEKG by inappropriate pacemaker spikes relative to the intrinsic electrical activity(pacemaker spikes occurring within p wave, QRS complex, or t wave) and shortened distances between paced beats.
  • 24.
    Undersensing  Battery depletion Decreased QRS voltage  Fusion beat  Dislodged/fractured lead  Inappropriate sensitivity setting
  • 25.
    Failure to Capture   Pacemakerhas delivered a pacing stimulus that was unable to initiate depolarization of the myocardium and subsequent myocardial contraction. Evidenced on EKG by pacemaker spikes that are not followed by a p wave for atrial pacing or a QRS complex for ventricular pacing.
  • 26.
  • 27.
    Feature of theTemporary DDD Pacemaker 2.5 1.25 0.5 Sensitivity
  • 28.
    Medtronic 5388 Temporary Pacemaker *Thefollowing slides are courtesy of Medtronics*
  • 29.
    Model 5388 DualChamber Temporary Pacemaker 1. Pace/Sense LEDs 2. Lock/Unlock Key 3. Lock Indicators 4. Rate Dial 5. Atrial Output Dial 6. Ventricular Output Dial 7. Menu Parameter Dial 8. Parameter Selection Key 9. Menu Selection Key 10. Pause Key 11. Power On Key 12. Power Off Key 13. Emergency/Asynchronous Pacing Key 14. Lower Screen 15. Ventricular Output Graphics 16. Atrial Output Graphics 17. Upper Screen 18. Rate Graphics 19. Setup Indicators 20. DDI Indicator 21. Low Battery Indicator 22. Setup Labels
  • 30.
    Cable to DeviceConnections
  • 31.
  • 32.
    Low Battery IndicatorBattery Replacement
  • 33.
    Off / OnKeys Values at Power-On • RATE • A OUTPUT • V OUTPUT • A SENSITIVITY 80 ppm 10 mA 10 mA 0.5 mV • V SENSITIVITY • UPPER RATE • AV INTERVAL 2.0 mV 110 ppm 170 ms • PVARP 300 ms
  • 34.
    Indicators Pace/Sense Indicators Low Battery IndicatorDDI Mode Indicator Setup Indicators Padlock Indicator
  • 35.
    Lock / Unlock Padlock Icon FlashingKey Icon Lock/Unlock Key Lock Indicator
  • 36.
    Emergency Key EmergencyPacing Values RATE Current setting • A OUTPUT (or 80 ppm if device was off) 20 mA • V OUTPUT • A SENSITIVITY • V SENSITIVITY 25 mA ASYNC ASYNC • AV INTERVAL Current setting (or 170 ms if device was off)
  • 37.
  • 38.
    Rate and OutputAdjustments Ventricular Output Dial Atrial Output Dial Rate Dial
  • 39.
    Lower Screen Menus Menu1: Pacing Parameters Menu 2: Rate-Based Pacing Parameters Menu 3: Rapid Atrial Pacing Menu M: Dial-A-Mode
  • 40.
  • 41.
  • 42.
    Rate Dependent ParameterAdjustments PVARP Auto/Manual Setting Upper Rate AV Interval
  • 43.
  • 44.
  • 45.
    Want to see:   Ekg tracing shows paced rhythm consistent with set parameters. Patient exhibits hemodynamic stability. Pacemaker leads/wires are isolated from other electrical equipment by maintaining secure connections into the pulse generator.
  • 46.
    Care and Maintenance   Cleanse around insertion site with CHG or Betadine. Curl epicardial wires and use paper tape to tape to abdomen in horizontal position. Cleanse around temporary transvenous site with CHG and place sterile dressing.
  • 47.
    Complications      Over/undersensing causing competition Failureto capture Hiccups Phlebitis, thrombosis, embolism or bacteremia Ventricular dysrhythmias      Pneumo/hemothorax Myocardial perforation & cardiac tamponade Air embolism Lead dislodgement Failure to pace
  • 48.
    Troubleshooting:          Always check patientfirst! Be sure pacer is on!! Change battery. Asynchronous pacing-only if no intrinsic rhythm Be sure connections are secure. Evaluate sensitivity threshold. Increase MA. Reverse polarity of wires/change cables or generator. Check rhythm. If patient goes into Afib, AAI will not provide pacing. Go to VVI pacing.
  • 49.
    Documentation     Patient & familyeducation Date/time pacing initiated Description of events warranting intervention Vital signs & hemodynamic parameters before & after pacing initiated (response to pacing)
  • 50.
    Documentation (cont.)      EKGmonitoring strip recording before & after pacing initiated Pacemaker settings: mode, rate, output(MA), sensitivity Medications administered and response Adjustments to pacemaker Date/time pacing discontinued
  • 51.
    Resources: • • • AACN Procedure Manualfor Critical Care, Lynn-McHale Wiegand, Debra J. and Carlson, Karen K, eds.Chapters 48-50, pp. 333-361. 2005. “Pacemaker (Temporary) Insertion and Care.” Nursing Clinical Policies and Procedures/Cardiovascular Care, Moses Cone Health System Homepage. Wiederhold, Richard. Electrocardiography: The Monitoring and Diagnostic Leads. pp. 77-83, 1999.