SlideShare a Scribd company logo
1 of 14
Download to read offline
Final report, June 2017
TESTING AND COMPARISON OF THE FILTERSAFE STANDARD,
TURBO AND SUPERTURBO CONFIGURATIONS.
+ =
3 	 Executive Summary
4 	 Background and Objective
5 	 Experimental Setup
7 	 Test Procedure
8 	 Test Results/ Discussion
12 	 Backwash Frequency and Volume
13 	 Sampling and Analysis
14 	Conclusion
15 	 Quality Assurance/ Quality Control
CONTENT
InBallast DHI Lloyd'sCONTRIBUTORS:
EXECUTIVE
SUMMARY
Filtersafe Ltd. have 3 configurations of their filter product suite – Standard, Turbo
and Super Turbo. The three filter configuration’s design are identical except for
nozzle density, improving the filter’s capability to clean the filter screen. The filter
is commonly used as a pre-treatment step for several treatment systems that are
type approved by the USCG.
USCG’s Marine Safety Center has required vendors that want to use the Turbo
and Super Turbo configurations to validate that the configuration does not
compromise the separation efficiency. Lloyds Register is appointed an IL by the
USCG, qualifying them to oversee and evaluate testing related to ballast water type
approvals. The plan for testing as presented in email dated 27.04.2017 was agreed
with by the USCG in email of 01.05.2017. Together with Lloyds Register and DHI, it
has been defined that the acceptance criteria for the tests is that the removal rate
of the two “new” configurations should be as good as the standard version within
the standard deviation of the results.
A BS-101 Filtersafe filter with a corresponding rated capacity of 250 m3
/h
(Standard), 300 m3
/h (Turbo) and 350 m3/h (Super Turbo), has been tested at DHI’s
land based test facility in Hundested, Denmark under the supervision of Lloyds
Register. The tests were carried out in accordance to Test Plan which was approved
by Lloyds Register prior to the testing.
The intention of the test program has been to assess the BS-101 Filtersafe Turbo
and Super Turbo filter configurations against the standard version by addressing
their ability to produce comparable and reproducible test results and to determine
if the alternative configurations of the filter has any effect on the ability to remove
organisms ≥ 50 µm in minimum dimension.
All tests were carried out using water exceeding IMO/ USCG challenge water
requirements for the referred organism category and sampled and enumerated
according to IMO/ USCG requirements by DHI.
The results from the comparative testing of the Filtersafe – Standard, Turbo and
Super Turbo versions has not identified any deviations in their ability to remove
organisms ≥ 50 µm in minimum dimension. The increased rated challenging
conditions did effect the backwashing frequency and therefore the backwash
volume, however not to a level where the filter will clog or to the point where the
separation efficiency is compromised.
Standard Turbo SuperTurbo
BACKGROUND The Filtersafe filters have been applied successfully as an integral component of
different ballast water treatment systems in numerous formal type approval tests in
accordance with both the ETV protocol as well as with the IMO G8 Guidelines.
Filters are typically scaled linearly with reference to a specific flow per unit
filter screen area for a specified filtration size. This empirical approach is widely
recognized and used by most manufacturers. The Filtersafe standard product suite
has been further developed and two extensions from the base models are now
available: The Turbo and Super Turbo filters. These are identical to the standard
filter, however with a higher nozzle density (increased number of nozzles per
screen area unit), improving the filter’s capability to clean the filter screen. This
improvement can then be utilized by increasing filtrated water flow without
changing the physical properties of the filtration process.
The purpose of these tests has been to compare the separation and cleaning
capabilities of the Turbo and Super Turbo with the standard rated filter.
Filtersafe have undertaken a number of tests including that of a comparative
evaluation of multiple nozzle scanners using a BS-025 filter with 40µm screen at
their Tefen-based test laboratory (September 2016). The experimental results have
confirmed a relationship between increased nozzle density and increased capacity
whilst ensuring the filters ability to fully recover following a flushing sequence.
Tests have been undertaken triggering the flushing sequence at specified
differential pressure of 0.4 bar. The increased flow velocity and consequently the
more rapid build-up of filter-cake does not seem to impact the filters separation
capacity. It may be noted that the flow velocity inside a filter chamber is relatively
lower than that in the inlet and outlet pipe.
From this, it was concluded that an increase in cleaning capacity by adding
number of nozzles per screen area unit from the standard version, enables the
filter to sustain higher flowrates without jeopardizing its separation capacity. This
has opened for the opportunity of introducing higher capacity filters without
increasing their footprint/ volume. Thus, the Turbo and Super Turbo models has
been introduced.
In correspondence with the USCG, it has been made clear that the acceptance
of filter models based on a standard filter but with increased maximum flowrate
based on the above, such as the Turbo and Super Turbo models, will require an
independent assessment by a class society/ independent third party (such as an
Independent Laboratory, IL) to conclude that such models can be a replacement
for standard models.
The assessment including testing has been undertaken in compliance with the
Test Plan. This has been approved by Lloyds Register. Tests were undertaken in
accordance with this on 27.07.2017 and 28.07 2017.
OBJECTIVE
TEST SETUP The validation tests were undertaken by DHI Denmark. The testing was overseen
and evaluated by Lloyds register. The tests applied a challenge water with the
following characteristics containing an
average of 412 000 and 542 000 individual organisms of minimum diameter of ≥
50 μm.
•	 Most of the organisms (>90%) were natural and concentrated by filtration
•	 of ambient water near the test facility.
•	 In addition, some standard test organisms –STO’s were added (Artemia). The
intention of adding Artemia, was to verify the integrity of the filter mesh (e.g. in
light of the size of these, any presence of Artemia in the filtered sample water
would indicate a deficiency in the filter mesh.
•	 TSS was added to meet the ETV protocol/ IMO G8.
•	 The water was applied to the test filters from a single test water tank of 750 m3.
•	 A total of two batches was used allowing two tests of each filter configuration
per batch.
•	 Preparation, storage and mixing of the water in the tank followed the approved
QAPP to which the facility operates when undertaking compliance testing for
type approval under the USCG type approval regime1.
1
	 DHI. Quality Assurance Project Plan. Biological efficacy performance evaluation of ballast water management systems.
	 DHI Ballast Water Centre - Denmark.
2
	 Test Protocol for validating flow rates on Standard, Turbo and Super-Turbo scanners Doc. No TP01, Rev.2.2.2
Figure 1 -
P&ID of the Test
Setup Ref. Appendix
D in the Approved
Test Plan2 Test Tank
750m3
>100.000,
>50µm
Filter BS-101
Tank
1m3
Data Logger
Test Facility
V1
P1
S1 S2
F1 FT2
FT3
PT1 PT2
V2
V3
Over board
Over board
Data Logger
FS
Controller
FS
Filtersafe scope of delivery
Test facility scope of delivery
35 µm sieve
Tank
20L
The influent was sampled by taking a 20-liter time integrated sample for each test.
The effluent was sampled by use of a 1 m3
sample tank with a single integrated
(continuous) sample for each test run. Both effluent and backwash was discharged
to sea.
Inlet pressure, outlet pressure, inlet flow and backwash flow was monitored by the
Filtersafe (FS) logging system. The piping system’s Inlet flow was logged by the test
facility with their own instrumentation.
A BS-101 filter with a rated nominal flow of 250 m3
/h was used for the test
representing the reference base filter. By changing the scanner arm and associated
number of nozzles, this filter was converted into the Turbo version and the Super
Turbo version between tests. By these changes only, the filter meets the Turbo and
Super Turbo filter specification.
Testing was undertaken as with the following filters:
Table 1 filter capacity
The filter flow capacity settings defined (Turbo +20%, Super Turbo + 40%),
represent a flow increase for the entire product suite.
The filters backwash operation was automatically controlled by the standard
Filtersafe control unit that initiated an automatic backwash sequence when the
differential pressure (between the inlet and outlet of the filter) reached 0,4 bar.
The filter in each of the configurations was operated over a period of time
sufficient to undertake a representative sample for the ≥ 50 µm organism group
as defined by IMO and USCG. In order to ensure stable conditions and to allow
for sufficient time to undertake appropriate sampling, the test cycle had duration
of between 18 and 21 minutes. For each filter configuration, four test cycles with
water from two independent test tanks was conducted. The test plan matrix is
presented in Table 2 below.
A standard Filtersafe filter used for ballast water treatment has a 40 µm screen.
In the ballast water treatment context, the task of the filter is to reduce the
concentration of the ≥50 µm organism group defined by IMO and USCG.
Filter Rated max. capacity
BS-101 base filter 250 m3
/hr
BS-101 base filter converted by installing turbo scanner arm
(representing the Turbo version)
300 m3
/hr
BS-101 base filter converted by installing super turbo scanner arm
(representing the Super Turbo version)
350 m3
/hr
TEST
SAMPLING
Test
day
Test
Actual test flow
(rated flow)
(m3/h)
Filter
Actual
Accumulated
test volume
(m3
)
Sample
Inlet
≥50
Sample
Treated
≥50
1
1 265 (250) Filtersafe Base 81,2 1 1
2 264 (250) Filtersafe Base 81,2 1 1
3 315 (300) Filtersafe Turbo 107,6 1 1
4 315 (300) Filtersafe Turbo 115,2 1 1
5 364 (350) Filtersafe Super Turbo 117,8 1 1
6 358 (350) Filtersafe Super Turbo 105,1 1 1
2
7 359 (350) Filtersafe Super Turbo 116,3 1 1
8 360 (350) Filtersafe Super Turbo 129,2 1 1
9 317 (300) Filtersafe Turbo 96 1 1
10 317 (300) Filtersafe, Turbo 103,1 1 1
11 262 (250) Filtersafe, Base 88,9 1 1
12 263 (250) Filtersafe, Base 87,1 1 1
TEST
PROCEDURE
A verification checklist was prepared in collaboration with DHI and Lloyds Register.
The intention with this was to secure that the tests were run according to the
approved test plan and present in a more transparent way, what had been verified
on site.
The testing was run sequentially according to the following table.
Table 2 test run matrix
The inlet water delivered from the tank was assumed homogenous due to the
standard active mixing in operation in the tank.
Each test was run for 18-21 minutes at full flow specified by the rated capacity of
the filter in accordance with Table 2. The flow was adjusted by the pump of the test
facility. The influent and effluent was sampled for ≥50µm organisms by test facility
staff for later analysis.
After 2 sequential tests for each filter configurations, Filtersafe crew changed the
nozzle scanner from a standard to a Turbo, and from a Turbo to a Super Turbo, by
removing blinded nozzles. The design of the nozzle configuration is identical to
commercial product. The changeover procedure was overseen by the independent
test facility staff/ Lloyds Register to assure that the filter was changed to the
agreed configuration and to verify that other components of the filter were not
tampered with.
TEST RESULTS/
DISCUSSION
Inlet values:
The inlet values (average) of each tank was 4-5 times higher than that required
(> 100 000 required). Further, the concentration of natural organisms (excluding
STO’s) dominated the water with an average of > 90 % of the organisms.
In spite of the continuous mixing of the tank, the inlet values for each individual
test varied somewhat. However, DHI have confirmed these to reflect experienced
variations and that these are within acceptable limits.
Referring to the final results presented in DHI’s counting logs, there are no specific
trends related to the taxonomy of species. Additionally, the values show no defined
trend of layering of organisms in the water. This can be seen in Figure 2 below,
where the concentration peaks differ for the different fractions (tests) from each
tank. Due to the fact that all 3 filters were operated in different sequences over
the two test days, and that the test tank had mismatching concentration peaks of
species concentration, as illustrated in Figure 2, variations in inlet concentration
seems not to have had any influence on the final results.
The average inlet concentrations of organisms for the base filter was around
20% lower than for the “turbo” and super turbo” filter. For the base filter, the
average concentration was approximately 408 000 organisms/ m3 over the 4
tests. For the turbo and super turbo filter, the average inlet concentrations were
536 000 organisms/ m3
and 488 000 organisms/ m3
over the 4 tests, respectively.
Indeed, the Turbo and Super Turbo versions were challenged with a higher inlet
concentration and thus, the testing of these assumed conservative in comparison
with the base model. This in respect to both separation capability as well as to
their resistance against clogging and ability to fully recover following backwashing.
Removal Rates:
The removal rates for the different filter configurations showed to be similar
regardless of inlet concentration of organisms, inlet flow and respective nozzle
density. This is shown in Tables 3 and 4 as well as in Figure 3.
Table 3 Inlet and Discharge Concentrations of ≥50 µM Organisms
Day Test Filter Inlet (#/m3) Outlet (#/m3) Reduction (%)
1
1 Base 471 975 75 960 83,9
2 Base 406 433 65 766 83,8
3 Turbo 611 733 83 227 86,4
4 Turbo 720 976 74 556 89,7
5 Super 617 867 78 778 87,3
6 Super 427 542 89 397 79,1
2
7 Super 429 200 93 434 78,2
8 Super 476 867 88 264 81,5
9 Turbo 493 333 76 608 84,5
10 Turbo 317 675 80 505 74,7
11 Base 400 350 74 016 81,5
12 Base 355 200 64 918 81,7
Figure 2 -
Inlet Concentration
Challenge water Characteristics
(oeganisms/ m3) vs fraction of tank
Day 1	 Day 2
1,000,000
800,000
600,000
400,000
200,000
0
1# 2# 3# 4# 5# 6# AVG
Table 4 Removal Rates of ≥50 µM Organisms
Removal Rates (%)
Standard Turbo Super
Day Test Removal
(%)
AVG
(paired tests)
Test Removal
(%)
AVG
(paired tests)
Test Removal
(%)
AVG
(paired tests)
Day 1
Test
1
83,9
83,9
Test
3
86,4
88,0
Test
5
87,3
83,2
Day 1
Test
2
83,8
Test
4
89,7
Test
6
79,1
Day 2
Test
11
81,5
81,6
Test
9
84,5
79,6
Test
7
78,2
79,9
Day 2
Test
12
81,7
Test
10
74,7
Test
8
81,5
AVG (4
tests)
82,7 83,8 81,5
STD
DEV (4
tests)
1,3 6,5 4,1
Looking at the average removal rates achieved for the Base, Turbo and Super Tubo
filters at 82.7, 83.7 and 81.5, respectively, it may be concluded that perform very
similarely.
Following this, it has been shown that both the Turbo and the Super Turbo filter
versions meets the requirement of seperation as good as the standard filter within
its standard deviation.
Figure 3 -
Comparison of
Removal Rates of >50
µM Organisms
100
90
80
70
60
50
40
30
20
10
0
Base Turbo Super Turbo
1#	2#	3#	4#	AVG
%
BACKWASH
FREQUENCY
AND VOLUME
The tests show that that the FS filter will backwash every 3,5 - 4 minutes for the
predicted water quality (IMO challenge water) for the standard filter. For the
Turbo and Super Turbo model, the flushing increases a bit - every 3,3 -2,5 minutes
respectively. The backwash sequence time is fixed for all models, thus the outcome
of increasing the flow is increased the backwash frequency for the Turbo and
Super Turbo model to some extent, however, none of the filters were running in
continuous backwash. Figures 4-6 are representative charts, that illustrate that the
filters backwashing abilities is not effected by the higher flow. Table 5, describes
the flushing volume and flow rates for the tests.
Figure 4 -
Representative
Chart for a Standard
Filter Running on
Challenge Water
Test 1 - Standard-1, Day 1
Time line (hh:mm:ss)
Pressure(bar)
Flowrate(m3
/h)
FiltrationVelocity(m/h)
400
350
300
250
200
150
100
50
0
2.5
2
1.5
1
0.5
0
-0.5
0:00:00 0:02:53 0:05:46 0:08:38 0:11:31 0:14:24 0:17:17 0:20:10
Intel flowrate Flushing flowrate Inlet Pressure ∆P
Table 5 – Backwash characteristics
Scanner model No. of nozzles
Flush flow rate m3/hr @ 2 bar
(Pending on operating pressure)
Actual flush per cycle
(At 50 seconds) Liters
Base 4 12 200
6 18 300
8 23 350
Figure 5 -
Representative
Chart for a Turbo
Filter Running on
Challenge Water
Figure 6 -
Representative Chart
for a Super Turbo
Filter Running on
Challenge Water
Test 5 - Super Turbo-1, Day 1
Time line (hh:mm:ss)
Pressure(bar)
400
350
300
250
200
150
100
50
0
33
2.5
2
1.5
1
0.5
0
-0.5
0:00:00 0:02:53 0:05:46 0:08:38 0:11:31 0:14:24 0:17:17 0:20:10
Intel flowrate Flushing flowrate Inlet Pressure ∆P
Flowrate(m3
/h)
FiltrationVelocity(m/h)
Time line (hh:mm:ss)
Pressure(bar)
Flowrate(m3
/h)
FiltrationVelocity(m/h)
400
350
300
250
200
150
100
50
0
3.5
3
2.5
2
1.5
1
0.5
0
-0.5
0:00:00 0:02:53 0:05:46 0:08:38 0:11:31 0:14:24 0:17:17 0:20:10
Intel flowrate Flushing flowrate Inlet Pressure ∆P
Test 3 - Standard-1, Day 1
3
	 DHI. Quality Assurance Project Plan. Biological efficacy performance evaluation of ballast water management systems.
	 DHI Ballast Water Centre - Denmark.
SAMPLING
& ANALYSIS
Sampling was conducted using sampling ports according to the design specified
in ETV 5.3.2.5 and G2 of the IMO BWM convention and further positioned
according to the PID in appendix D of the approved test plan. A total of 20 L (inlet)
and 1m3
(discharge) was taken from the mainline upstream and downstream the
filter respectively in two separate tanks. This water was filtered through a net with
absolute mesh size of 50µm, and enumerated within 6 hours according to DHI’s
approved QAPP3
.
The rationale for using 20 l sampling volume for the determination of the inlet
concentration is based on a recommendation from the test facility (DHI, and in
agreement with Lloyds Register), that have experienced that sampling 1 m3
of the
inlet water will result in a too high density of organisms, making it challenging to
count and a giving high mortality, resulting in less accurate results.
QUALITY
ASSURANCE/
QUALITY
CONTROL
All testing was conducted by DHI, overseen by Lloyds Register without any
intervention by the manufacturer as verified by the signed checklist.
Data logged by the Filtersafe controller was as defined in the pre-approved Test
Plan handed over to DHI immediately after the test and was later forwarded by DHI
to the authors of this report.
Samples were conducted, handled, transported and enumerated according to the
pre-approved test plan.
REGIONAL OFFICE, MANUFACTURING
& ASSEMBLY
©CopyrightFiltersafe®2018DOC-00358
w w w . f i l t e r s a f e . n e t
Filtersafe reserves the right to change any of the information presented in this document without prior notice.
The data presented in this document is for general information only and cannot be used for quotation purposes.
Please contact Filtersafe for specific quote.
Filtersafe Offices
Filtersafe Manufacturing Facilities
Filtersafe Distributor's
ISRAEL	
A 4,000 sq/m premises comprising our regional sales offices,
R&D facilities and manufacturing/assembly plant. The facility is
capable of handling very large volumes of production in line
with market growth.
Contact Information:
Address: Ha’ella Street, P.O.B 23, Tefen Ind. Zone 2495900, Israel
Phone: +972 4 987 3620 ext. 105/106, Email: sales@filtersafe.net
HONG KONG
Our Hong Kong manufacturing/assembly plant supplies our entire
line of products to the region and is supported by a world-class
steel fabrication facility.
Contact Information:
Address: DD362  Lot 30  Tsing Chau Tsai Shipyard, 
Lantau Island N T, Hong Kong
Phone: +852 2454 4300, Email: fshongkong@filtersafe.net
BRAZIL, Rio de Janeiro
Branch Office
Contact Information:
Cell: +55 21 98899 6890
Direct: +55 21 3798 2307
Email: fsbrazil@filtersafe.net
CHINA, Beijing
Branch Office
Contact Information:
Address: Room 811, Building 1,
Xinyuanguoji Plaza, No 1,
Chengshousi Road, Fengtai
Dist., 100078 Beijing, PRC China
Email: fschina@filtersafe.net
GERMANY, Hamburg
Branch Office
Strategic Sales & Technical
Office serving all European
markets.
Contact Information:
Phone: +49 151 5111 9622
Email: fsgermany@filtersafe.net
SINGAPORE
Branch Office
Strategic Sales & Technical Office
Contact Information:
Phone: +65 9848 9856
Email: fssingapore@filtersafe.net

More Related Content

What's hot

Life Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for BiopharmaceuticalsLife Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for BiopharmaceuticalsWaters Corporation
 
IRJET- A Review on Utilization of Sequence Batch Reactor Technology (SBR)...
IRJET-  	  A Review on Utilization of Sequence Batch Reactor Technology (SBR)...IRJET-  	  A Review on Utilization of Sequence Batch Reactor Technology (SBR)...
IRJET- A Review on Utilization of Sequence Batch Reactor Technology (SBR)...IRJET Journal
 
A Comparison of Protein A Chromatographic Stationary Phases
A Comparison of Protein A Chromatographic Stationary PhasesA Comparison of Protein A Chromatographic Stationary Phases
A Comparison of Protein A Chromatographic Stationary PhasesKBI Biopharma
 
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa... Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...hydrologywebsite1
 
Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...
Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...
Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...MilliporeSigma
 
Inline Flocculation for Harvest and Perfusate Clarification
Inline Flocculation for Harvest and Perfusate ClarificationInline Flocculation for Harvest and Perfusate Clarification
Inline Flocculation for Harvest and Perfusate ClarificationMerck Life Sciences
 
Asahi Kasei Bioprocess Europe 2016 02 1
Asahi Kasei Bioprocess Europe 2016 02 1Asahi Kasei Bioprocess Europe 2016 02 1
Asahi Kasei Bioprocess Europe 2016 02 1GBX Events
 
Find your filter. What’s best for your process?
Find your filter. What’s best for your process? Find your filter. What’s best for your process?
Find your filter. What’s best for your process? Merck Life Sciences
 
Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)MilliporeSigma
 
Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)Merck Life Sciences
 
Membrane Chromatography Solutions for Single-Use, Intensified mAb Purification
Membrane Chromatography Solutions for Single-Use, Intensified mAb PurificationMembrane Chromatography Solutions for Single-Use, Intensified mAb Purification
Membrane Chromatography Solutions for Single-Use, Intensified mAb PurificationMilliporeSigma
 
Process Development for Continuous Flow-Through Polishing Purification for mA...
Process Development for Continuous Flow-Through Polishing Purification for mA...Process Development for Continuous Flow-Through Polishing Purification for mA...
Process Development for Continuous Flow-Through Polishing Purification for mA...Merck Life Sciences
 
Webinar: How to Develop a Regulatory-compliant Continued Process Verificatio...
Webinar: 	How to Develop a Regulatory-compliant Continued Process Verificatio...Webinar: 	How to Develop a Regulatory-compliant Continued Process Verificatio...
Webinar: How to Develop a Regulatory-compliant Continued Process Verificatio...MilliporeSigma
 
Potential Impact of Draft Annex 1 on Sterilizing Filtration
Potential Impact of Draft Annex 1 on Sterilizing FiltrationPotential Impact of Draft Annex 1 on Sterilizing Filtration
Potential Impact of Draft Annex 1 on Sterilizing FiltrationMilliporeSigma
 
An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...
An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...
An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...Merck Life Sciences
 

What's hot (16)

Life Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for BiopharmaceuticalsLife Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for Biopharmaceuticals
 
IRJET- A Review on Utilization of Sequence Batch Reactor Technology (SBR)...
IRJET-  	  A Review on Utilization of Sequence Batch Reactor Technology (SBR)...IRJET-  	  A Review on Utilization of Sequence Batch Reactor Technology (SBR)...
IRJET- A Review on Utilization of Sequence Batch Reactor Technology (SBR)...
 
A Comparison of Protein A Chromatographic Stationary Phases
A Comparison of Protein A Chromatographic Stationary PhasesA Comparison of Protein A Chromatographic Stationary Phases
A Comparison of Protein A Chromatographic Stationary Phases
 
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa... Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
Download-manuals-water quality-wq-manuals-gwvolume6fieldmanualwaterqualitysa...
 
Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...
Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...
Millipak® Final Fill Filters Reduce Contamination Risks and Simplify Filtrati...
 
Inline Flocculation for Harvest and Perfusate Clarification
Inline Flocculation for Harvest and Perfusate ClarificationInline Flocculation for Harvest and Perfusate Clarification
Inline Flocculation for Harvest and Perfusate Clarification
 
Asahi Kasei Bioprocess Europe 2016 02 1
Asahi Kasei Bioprocess Europe 2016 02 1Asahi Kasei Bioprocess Europe 2016 02 1
Asahi Kasei Bioprocess Europe 2016 02 1
 
Find your filter. What’s best for your process?
Find your filter. What’s best for your process? Find your filter. What’s best for your process?
Find your filter. What’s best for your process?
 
Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)
 
Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)Introduction to Tangential Flow Filtration (TFF)
Introduction to Tangential Flow Filtration (TFF)
 
Membrane Chromatography Solutions for Single-Use, Intensified mAb Purification
Membrane Chromatography Solutions for Single-Use, Intensified mAb PurificationMembrane Chromatography Solutions for Single-Use, Intensified mAb Purification
Membrane Chromatography Solutions for Single-Use, Intensified mAb Purification
 
Process Development for Continuous Flow-Through Polishing Purification for mA...
Process Development for Continuous Flow-Through Polishing Purification for mA...Process Development for Continuous Flow-Through Polishing Purification for mA...
Process Development for Continuous Flow-Through Polishing Purification for mA...
 
Innovation in size based polymer separations
Innovation in size based polymer separationsInnovation in size based polymer separations
Innovation in size based polymer separations
 
Webinar: How to Develop a Regulatory-compliant Continued Process Verificatio...
Webinar: 	How to Develop a Regulatory-compliant Continued Process Verificatio...Webinar: 	How to Develop a Regulatory-compliant Continued Process Verificatio...
Webinar: How to Develop a Regulatory-compliant Continued Process Verificatio...
 
Potential Impact of Draft Annex 1 on Sterilizing Filtration
Potential Impact of Draft Annex 1 on Sterilizing FiltrationPotential Impact of Draft Annex 1 on Sterilizing Filtration
Potential Impact of Draft Annex 1 on Sterilizing Filtration
 
An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...
An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...
An Efficient and cGMP-friendly Solution to Diafiltration for Intensified or C...
 

Similar to TESTING AND COMPARISON OF THE FILTERSAFE STANDARD, TURBO AND SUPERTURBO CONFIGURATIONS.

validation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfvalidation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfMd. Omar Faruq Khan
 
66 membrane filter integrity test
66 membrane filter integrity test66 membrane filter integrity test
66 membrane filter integrity testKethana Nellore
 
GFCM - SwRI report 7-21-11v4
GFCM - SwRI report 7-21-11v4GFCM - SwRI report 7-21-11v4
GFCM - SwRI report 7-21-11v4Wassim Shirry
 
Testing/Manufacture/flat sheet media
Testing/Manufacture/flat sheet mediaTesting/Manufacture/flat sheet media
Testing/Manufacture/flat sheet mediafec2020
 
Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet MediaTesting/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Mediafec2020
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
Testing/Manufacture/flat sheet medi 2
Testing/Manufacture/flat sheet medi   2Testing/Manufacture/flat sheet medi   2
Testing/Manufacture/flat sheet medi 2fec2020
 
Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Media  Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Media fec2020
 
1._BWMS_commissioning_post_webinar_doc.pdf
1._BWMS_commissioning_post_webinar_doc.pdf1._BWMS_commissioning_post_webinar_doc.pdf
1._BWMS_commissioning_post_webinar_doc.pdfrafael442748
 
Pharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingPharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingEric Ferguson
 
Key factors for developing a pharmaceutical cleaning strategy - Part 2
Key factors for developing a pharmaceutical cleaning strategy - Part 2Key factors for developing a pharmaceutical cleaning strategy - Part 2
Key factors for developing a pharmaceutical cleaning strategy - Part 2Fedegari Group
 
Testing/Manufacture/flat sheet media
Testing/Manufacture/flat sheet mediaTesting/Manufacture/flat sheet media
Testing/Manufacture/flat sheet mediafec2020
 
Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet MediaTesting/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Mediafec2020
 
Care and use manual waters x bridge columns
Care and use manual waters x bridge columnsCare and use manual waters x bridge columns
Care and use manual waters x bridge columnsJohn Omondi
 
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...IRJET Journal
 
Single-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and PracticeSingle-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and PracticeMerck Life Sciences
 
Single-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and PracticeSingle-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and PracticeMilliporeSigma
 
Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...
Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...
Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...MilliporeSigma
 
Parvovirus Filtration Best Practices - 25 Years of Hands-On Experience
Parvovirus Filtration Best Practices - 25 Years of Hands-On ExperienceParvovirus Filtration Best Practices - 25 Years of Hands-On Experience
Parvovirus Filtration Best Practices - 25 Years of Hands-On ExperienceMilliporeSigma
 

Similar to TESTING AND COMPARISON OF THE FILTERSAFE STANDARD, TURBO AND SUPERTURBO CONFIGURATIONS. (20)

validation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfvalidation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdf
 
66 membrane filter integrity test
66 membrane filter integrity test66 membrane filter integrity test
66 membrane filter integrity test
 
GFCM - SwRI report 7-21-11v4
GFCM - SwRI report 7-21-11v4GFCM - SwRI report 7-21-11v4
GFCM - SwRI report 7-21-11v4
 
Testing/Manufacture/flat sheet media
Testing/Manufacture/flat sheet mediaTesting/Manufacture/flat sheet media
Testing/Manufacture/flat sheet media
 
Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet MediaTesting/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Media
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Testing/Manufacture/flat sheet medi 2
Testing/Manufacture/flat sheet medi   2Testing/Manufacture/flat sheet medi   2
Testing/Manufacture/flat sheet medi 2
 
Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Media  Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Media
 
1._BWMS_commissioning_post_webinar_doc.pdf
1._BWMS_commissioning_post_webinar_doc.pdf1._BWMS_commissioning_post_webinar_doc.pdf
1._BWMS_commissioning_post_webinar_doc.pdf
 
Pharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingPharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth Testing
 
Key factors for developing a pharmaceutical cleaning strategy - Part 2
Key factors for developing a pharmaceutical cleaning strategy - Part 2Key factors for developing a pharmaceutical cleaning strategy - Part 2
Key factors for developing a pharmaceutical cleaning strategy - Part 2
 
Testing/Manufacture/flat sheet media
Testing/Manufacture/flat sheet mediaTesting/Manufacture/flat sheet media
Testing/Manufacture/flat sheet media
 
Testing/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet MediaTesting/Manufacture/Flat Sheet Media
Testing/Manufacture/Flat Sheet Media
 
Care and use manual waters x bridge columns
Care and use manual waters x bridge columnsCare and use manual waters x bridge columns
Care and use manual waters x bridge columns
 
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...
 
Single-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and PracticeSingle-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
 
Single-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and PracticeSingle-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
Single-Pass Tangential Flow Filtration (SPTFF) Theory and Practice
 
Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...
Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...
Redesigned Filter Capsule Improves Final Filtration Assembly Design and Opera...
 
Parvovirus Filtration Best Practices - 25 Years of Hands-On Experience
Parvovirus Filtration Best Practices - 25 Years of Hands-On ExperienceParvovirus Filtration Best Practices - 25 Years of Hands-On Experience
Parvovirus Filtration Best Practices - 25 Years of Hands-On Experience
 

Recently uploaded

(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130Suhani Kapoor
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingrakeshbaidya232001
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Christo Ananth
 
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSHARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSRajkumarAkumalla
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performancesivaprakash250
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlysanyuktamishra911
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSSIVASHANKAR N
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 

Recently uploaded (20)

(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
 
Porous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writingPorous Ceramics seminar and technical writing
Porous Ceramics seminar and technical writing
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
Call for Papers - African Journal of Biological Sciences, E-ISSN: 2663-2187, ...
 
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICSHARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
HARDNESS, FRACTURE TOUGHNESS AND STRENGTH OF CERAMICS
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performance
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 

TESTING AND COMPARISON OF THE FILTERSAFE STANDARD, TURBO AND SUPERTURBO CONFIGURATIONS.

  • 1. Final report, June 2017 TESTING AND COMPARISON OF THE FILTERSAFE STANDARD, TURBO AND SUPERTURBO CONFIGURATIONS. + =
  • 2. 3 Executive Summary 4 Background and Objective 5 Experimental Setup 7 Test Procedure 8 Test Results/ Discussion 12 Backwash Frequency and Volume 13 Sampling and Analysis 14 Conclusion 15 Quality Assurance/ Quality Control CONTENT InBallast DHI Lloyd'sCONTRIBUTORS:
  • 3. EXECUTIVE SUMMARY Filtersafe Ltd. have 3 configurations of their filter product suite – Standard, Turbo and Super Turbo. The three filter configuration’s design are identical except for nozzle density, improving the filter’s capability to clean the filter screen. The filter is commonly used as a pre-treatment step for several treatment systems that are type approved by the USCG. USCG’s Marine Safety Center has required vendors that want to use the Turbo and Super Turbo configurations to validate that the configuration does not compromise the separation efficiency. Lloyds Register is appointed an IL by the USCG, qualifying them to oversee and evaluate testing related to ballast water type approvals. The plan for testing as presented in email dated 27.04.2017 was agreed with by the USCG in email of 01.05.2017. Together with Lloyds Register and DHI, it has been defined that the acceptance criteria for the tests is that the removal rate of the two “new” configurations should be as good as the standard version within the standard deviation of the results. A BS-101 Filtersafe filter with a corresponding rated capacity of 250 m3 /h (Standard), 300 m3 /h (Turbo) and 350 m3/h (Super Turbo), has been tested at DHI’s land based test facility in Hundested, Denmark under the supervision of Lloyds Register. The tests were carried out in accordance to Test Plan which was approved by Lloyds Register prior to the testing. The intention of the test program has been to assess the BS-101 Filtersafe Turbo and Super Turbo filter configurations against the standard version by addressing their ability to produce comparable and reproducible test results and to determine if the alternative configurations of the filter has any effect on the ability to remove organisms ≥ 50 µm in minimum dimension. All tests were carried out using water exceeding IMO/ USCG challenge water requirements for the referred organism category and sampled and enumerated according to IMO/ USCG requirements by DHI. The results from the comparative testing of the Filtersafe – Standard, Turbo and Super Turbo versions has not identified any deviations in their ability to remove organisms ≥ 50 µm in minimum dimension. The increased rated challenging conditions did effect the backwashing frequency and therefore the backwash volume, however not to a level where the filter will clog or to the point where the separation efficiency is compromised. Standard Turbo SuperTurbo
  • 4. BACKGROUND The Filtersafe filters have been applied successfully as an integral component of different ballast water treatment systems in numerous formal type approval tests in accordance with both the ETV protocol as well as with the IMO G8 Guidelines. Filters are typically scaled linearly with reference to a specific flow per unit filter screen area for a specified filtration size. This empirical approach is widely recognized and used by most manufacturers. The Filtersafe standard product suite has been further developed and two extensions from the base models are now available: The Turbo and Super Turbo filters. These are identical to the standard filter, however with a higher nozzle density (increased number of nozzles per screen area unit), improving the filter’s capability to clean the filter screen. This improvement can then be utilized by increasing filtrated water flow without changing the physical properties of the filtration process. The purpose of these tests has been to compare the separation and cleaning capabilities of the Turbo and Super Turbo with the standard rated filter. Filtersafe have undertaken a number of tests including that of a comparative evaluation of multiple nozzle scanners using a BS-025 filter with 40µm screen at their Tefen-based test laboratory (September 2016). The experimental results have confirmed a relationship between increased nozzle density and increased capacity whilst ensuring the filters ability to fully recover following a flushing sequence. Tests have been undertaken triggering the flushing sequence at specified differential pressure of 0.4 bar. The increased flow velocity and consequently the more rapid build-up of filter-cake does not seem to impact the filters separation capacity. It may be noted that the flow velocity inside a filter chamber is relatively lower than that in the inlet and outlet pipe. From this, it was concluded that an increase in cleaning capacity by adding number of nozzles per screen area unit from the standard version, enables the filter to sustain higher flowrates without jeopardizing its separation capacity. This has opened for the opportunity of introducing higher capacity filters without increasing their footprint/ volume. Thus, the Turbo and Super Turbo models has been introduced. In correspondence with the USCG, it has been made clear that the acceptance of filter models based on a standard filter but with increased maximum flowrate based on the above, such as the Turbo and Super Turbo models, will require an independent assessment by a class society/ independent third party (such as an Independent Laboratory, IL) to conclude that such models can be a replacement for standard models. The assessment including testing has been undertaken in compliance with the Test Plan. This has been approved by Lloyds Register. Tests were undertaken in accordance with this on 27.07.2017 and 28.07 2017. OBJECTIVE
  • 5. TEST SETUP The validation tests were undertaken by DHI Denmark. The testing was overseen and evaluated by Lloyds register. The tests applied a challenge water with the following characteristics containing an average of 412 000 and 542 000 individual organisms of minimum diameter of ≥ 50 μm. • Most of the organisms (>90%) were natural and concentrated by filtration • of ambient water near the test facility. • In addition, some standard test organisms –STO’s were added (Artemia). The intention of adding Artemia, was to verify the integrity of the filter mesh (e.g. in light of the size of these, any presence of Artemia in the filtered sample water would indicate a deficiency in the filter mesh. • TSS was added to meet the ETV protocol/ IMO G8. • The water was applied to the test filters from a single test water tank of 750 m3. • A total of two batches was used allowing two tests of each filter configuration per batch. • Preparation, storage and mixing of the water in the tank followed the approved QAPP to which the facility operates when undertaking compliance testing for type approval under the USCG type approval regime1. 1 DHI. Quality Assurance Project Plan. Biological efficacy performance evaluation of ballast water management systems. DHI Ballast Water Centre - Denmark. 2 Test Protocol for validating flow rates on Standard, Turbo and Super-Turbo scanners Doc. No TP01, Rev.2.2.2 Figure 1 - P&ID of the Test Setup Ref. Appendix D in the Approved Test Plan2 Test Tank 750m3 >100.000, >50µm Filter BS-101 Tank 1m3 Data Logger Test Facility V1 P1 S1 S2 F1 FT2 FT3 PT1 PT2 V2 V3 Over board Over board Data Logger FS Controller FS Filtersafe scope of delivery Test facility scope of delivery 35 µm sieve Tank 20L
  • 6. The influent was sampled by taking a 20-liter time integrated sample for each test. The effluent was sampled by use of a 1 m3 sample tank with a single integrated (continuous) sample for each test run. Both effluent and backwash was discharged to sea. Inlet pressure, outlet pressure, inlet flow and backwash flow was monitored by the Filtersafe (FS) logging system. The piping system’s Inlet flow was logged by the test facility with their own instrumentation. A BS-101 filter with a rated nominal flow of 250 m3 /h was used for the test representing the reference base filter. By changing the scanner arm and associated number of nozzles, this filter was converted into the Turbo version and the Super Turbo version between tests. By these changes only, the filter meets the Turbo and Super Turbo filter specification. Testing was undertaken as with the following filters: Table 1 filter capacity The filter flow capacity settings defined (Turbo +20%, Super Turbo + 40%), represent a flow increase for the entire product suite. The filters backwash operation was automatically controlled by the standard Filtersafe control unit that initiated an automatic backwash sequence when the differential pressure (between the inlet and outlet of the filter) reached 0,4 bar. The filter in each of the configurations was operated over a period of time sufficient to undertake a representative sample for the ≥ 50 µm organism group as defined by IMO and USCG. In order to ensure stable conditions and to allow for sufficient time to undertake appropriate sampling, the test cycle had duration of between 18 and 21 minutes. For each filter configuration, four test cycles with water from two independent test tanks was conducted. The test plan matrix is presented in Table 2 below. A standard Filtersafe filter used for ballast water treatment has a 40 µm screen. In the ballast water treatment context, the task of the filter is to reduce the concentration of the ≥50 µm organism group defined by IMO and USCG. Filter Rated max. capacity BS-101 base filter 250 m3 /hr BS-101 base filter converted by installing turbo scanner arm (representing the Turbo version) 300 m3 /hr BS-101 base filter converted by installing super turbo scanner arm (representing the Super Turbo version) 350 m3 /hr TEST SAMPLING
  • 7. Test day Test Actual test flow (rated flow) (m3/h) Filter Actual Accumulated test volume (m3 ) Sample Inlet ≥50 Sample Treated ≥50 1 1 265 (250) Filtersafe Base 81,2 1 1 2 264 (250) Filtersafe Base 81,2 1 1 3 315 (300) Filtersafe Turbo 107,6 1 1 4 315 (300) Filtersafe Turbo 115,2 1 1 5 364 (350) Filtersafe Super Turbo 117,8 1 1 6 358 (350) Filtersafe Super Turbo 105,1 1 1 2 7 359 (350) Filtersafe Super Turbo 116,3 1 1 8 360 (350) Filtersafe Super Turbo 129,2 1 1 9 317 (300) Filtersafe Turbo 96 1 1 10 317 (300) Filtersafe, Turbo 103,1 1 1 11 262 (250) Filtersafe, Base 88,9 1 1 12 263 (250) Filtersafe, Base 87,1 1 1 TEST PROCEDURE A verification checklist was prepared in collaboration with DHI and Lloyds Register. The intention with this was to secure that the tests were run according to the approved test plan and present in a more transparent way, what had been verified on site. The testing was run sequentially according to the following table. Table 2 test run matrix The inlet water delivered from the tank was assumed homogenous due to the standard active mixing in operation in the tank. Each test was run for 18-21 minutes at full flow specified by the rated capacity of the filter in accordance with Table 2. The flow was adjusted by the pump of the test facility. The influent and effluent was sampled for ≥50µm organisms by test facility staff for later analysis. After 2 sequential tests for each filter configurations, Filtersafe crew changed the nozzle scanner from a standard to a Turbo, and from a Turbo to a Super Turbo, by removing blinded nozzles. The design of the nozzle configuration is identical to commercial product. The changeover procedure was overseen by the independent test facility staff/ Lloyds Register to assure that the filter was changed to the agreed configuration and to verify that other components of the filter were not tampered with.
  • 8. TEST RESULTS/ DISCUSSION Inlet values: The inlet values (average) of each tank was 4-5 times higher than that required (> 100 000 required). Further, the concentration of natural organisms (excluding STO’s) dominated the water with an average of > 90 % of the organisms. In spite of the continuous mixing of the tank, the inlet values for each individual test varied somewhat. However, DHI have confirmed these to reflect experienced variations and that these are within acceptable limits. Referring to the final results presented in DHI’s counting logs, there are no specific trends related to the taxonomy of species. Additionally, the values show no defined trend of layering of organisms in the water. This can be seen in Figure 2 below, where the concentration peaks differ for the different fractions (tests) from each tank. Due to the fact that all 3 filters were operated in different sequences over the two test days, and that the test tank had mismatching concentration peaks of species concentration, as illustrated in Figure 2, variations in inlet concentration seems not to have had any influence on the final results. The average inlet concentrations of organisms for the base filter was around 20% lower than for the “turbo” and super turbo” filter. For the base filter, the average concentration was approximately 408 000 organisms/ m3 over the 4 tests. For the turbo and super turbo filter, the average inlet concentrations were 536 000 organisms/ m3 and 488 000 organisms/ m3 over the 4 tests, respectively. Indeed, the Turbo and Super Turbo versions were challenged with a higher inlet concentration and thus, the testing of these assumed conservative in comparison with the base model. This in respect to both separation capability as well as to their resistance against clogging and ability to fully recover following backwashing.
  • 9. Removal Rates: The removal rates for the different filter configurations showed to be similar regardless of inlet concentration of organisms, inlet flow and respective nozzle density. This is shown in Tables 3 and 4 as well as in Figure 3. Table 3 Inlet and Discharge Concentrations of ≥50 µM Organisms Day Test Filter Inlet (#/m3) Outlet (#/m3) Reduction (%) 1 1 Base 471 975 75 960 83,9 2 Base 406 433 65 766 83,8 3 Turbo 611 733 83 227 86,4 4 Turbo 720 976 74 556 89,7 5 Super 617 867 78 778 87,3 6 Super 427 542 89 397 79,1 2 7 Super 429 200 93 434 78,2 8 Super 476 867 88 264 81,5 9 Turbo 493 333 76 608 84,5 10 Turbo 317 675 80 505 74,7 11 Base 400 350 74 016 81,5 12 Base 355 200 64 918 81,7 Figure 2 - Inlet Concentration Challenge water Characteristics (oeganisms/ m3) vs fraction of tank Day 1 Day 2 1,000,000 800,000 600,000 400,000 200,000 0 1# 2# 3# 4# 5# 6# AVG
  • 10. Table 4 Removal Rates of ≥50 µM Organisms Removal Rates (%) Standard Turbo Super Day Test Removal (%) AVG (paired tests) Test Removal (%) AVG (paired tests) Test Removal (%) AVG (paired tests) Day 1 Test 1 83,9 83,9 Test 3 86,4 88,0 Test 5 87,3 83,2 Day 1 Test 2 83,8 Test 4 89,7 Test 6 79,1 Day 2 Test 11 81,5 81,6 Test 9 84,5 79,6 Test 7 78,2 79,9 Day 2 Test 12 81,7 Test 10 74,7 Test 8 81,5 AVG (4 tests) 82,7 83,8 81,5 STD DEV (4 tests) 1,3 6,5 4,1 Looking at the average removal rates achieved for the Base, Turbo and Super Tubo filters at 82.7, 83.7 and 81.5, respectively, it may be concluded that perform very similarely. Following this, it has been shown that both the Turbo and the Super Turbo filter versions meets the requirement of seperation as good as the standard filter within its standard deviation. Figure 3 - Comparison of Removal Rates of >50 µM Organisms 100 90 80 70 60 50 40 30 20 10 0 Base Turbo Super Turbo 1# 2# 3# 4# AVG %
  • 11. BACKWASH FREQUENCY AND VOLUME The tests show that that the FS filter will backwash every 3,5 - 4 minutes for the predicted water quality (IMO challenge water) for the standard filter. For the Turbo and Super Turbo model, the flushing increases a bit - every 3,3 -2,5 minutes respectively. The backwash sequence time is fixed for all models, thus the outcome of increasing the flow is increased the backwash frequency for the Turbo and Super Turbo model to some extent, however, none of the filters were running in continuous backwash. Figures 4-6 are representative charts, that illustrate that the filters backwashing abilities is not effected by the higher flow. Table 5, describes the flushing volume and flow rates for the tests. Figure 4 - Representative Chart for a Standard Filter Running on Challenge Water Test 1 - Standard-1, Day 1 Time line (hh:mm:ss) Pressure(bar) Flowrate(m3 /h) FiltrationVelocity(m/h) 400 350 300 250 200 150 100 50 0 2.5 2 1.5 1 0.5 0 -0.5 0:00:00 0:02:53 0:05:46 0:08:38 0:11:31 0:14:24 0:17:17 0:20:10 Intel flowrate Flushing flowrate Inlet Pressure ∆P
  • 12. Table 5 – Backwash characteristics Scanner model No. of nozzles Flush flow rate m3/hr @ 2 bar (Pending on operating pressure) Actual flush per cycle (At 50 seconds) Liters Base 4 12 200 6 18 300 8 23 350 Figure 5 - Representative Chart for a Turbo Filter Running on Challenge Water Figure 6 - Representative Chart for a Super Turbo Filter Running on Challenge Water Test 5 - Super Turbo-1, Day 1 Time line (hh:mm:ss) Pressure(bar) 400 350 300 250 200 150 100 50 0 33 2.5 2 1.5 1 0.5 0 -0.5 0:00:00 0:02:53 0:05:46 0:08:38 0:11:31 0:14:24 0:17:17 0:20:10 Intel flowrate Flushing flowrate Inlet Pressure ∆P Flowrate(m3 /h) FiltrationVelocity(m/h) Time line (hh:mm:ss) Pressure(bar) Flowrate(m3 /h) FiltrationVelocity(m/h) 400 350 300 250 200 150 100 50 0 3.5 3 2.5 2 1.5 1 0.5 0 -0.5 0:00:00 0:02:53 0:05:46 0:08:38 0:11:31 0:14:24 0:17:17 0:20:10 Intel flowrate Flushing flowrate Inlet Pressure ∆P Test 3 - Standard-1, Day 1
  • 13. 3 DHI. Quality Assurance Project Plan. Biological efficacy performance evaluation of ballast water management systems. DHI Ballast Water Centre - Denmark. SAMPLING & ANALYSIS Sampling was conducted using sampling ports according to the design specified in ETV 5.3.2.5 and G2 of the IMO BWM convention and further positioned according to the PID in appendix D of the approved test plan. A total of 20 L (inlet) and 1m3 (discharge) was taken from the mainline upstream and downstream the filter respectively in two separate tanks. This water was filtered through a net with absolute mesh size of 50µm, and enumerated within 6 hours according to DHI’s approved QAPP3 . The rationale for using 20 l sampling volume for the determination of the inlet concentration is based on a recommendation from the test facility (DHI, and in agreement with Lloyds Register), that have experienced that sampling 1 m3 of the inlet water will result in a too high density of organisms, making it challenging to count and a giving high mortality, resulting in less accurate results. QUALITY ASSURANCE/ QUALITY CONTROL All testing was conducted by DHI, overseen by Lloyds Register without any intervention by the manufacturer as verified by the signed checklist. Data logged by the Filtersafe controller was as defined in the pre-approved Test Plan handed over to DHI immediately after the test and was later forwarded by DHI to the authors of this report. Samples were conducted, handled, transported and enumerated according to the pre-approved test plan.
  • 14. REGIONAL OFFICE, MANUFACTURING & ASSEMBLY ©CopyrightFiltersafe®2018DOC-00358 w w w . f i l t e r s a f e . n e t Filtersafe reserves the right to change any of the information presented in this document without prior notice. The data presented in this document is for general information only and cannot be used for quotation purposes. Please contact Filtersafe for specific quote. Filtersafe Offices Filtersafe Manufacturing Facilities Filtersafe Distributor's ISRAEL A 4,000 sq/m premises comprising our regional sales offices, R&D facilities and manufacturing/assembly plant. The facility is capable of handling very large volumes of production in line with market growth. Contact Information: Address: Ha’ella Street, P.O.B 23, Tefen Ind. Zone 2495900, Israel Phone: +972 4 987 3620 ext. 105/106, Email: sales@filtersafe.net HONG KONG Our Hong Kong manufacturing/assembly plant supplies our entire line of products to the region and is supported by a world-class steel fabrication facility. Contact Information: Address: DD362  Lot 30  Tsing Chau Tsai Shipyard,  Lantau Island N T, Hong Kong Phone: +852 2454 4300, Email: fshongkong@filtersafe.net BRAZIL, Rio de Janeiro Branch Office Contact Information: Cell: +55 21 98899 6890 Direct: +55 21 3798 2307 Email: fsbrazil@filtersafe.net CHINA, Beijing Branch Office Contact Information: Address: Room 811, Building 1, Xinyuanguoji Plaza, No 1, Chengshousi Road, Fengtai Dist., 100078 Beijing, PRC China Email: fschina@filtersafe.net GERMANY, Hamburg Branch Office Strategic Sales & Technical Office serving all European markets. Contact Information: Phone: +49 151 5111 9622 Email: fsgermany@filtersafe.net SINGAPORE Branch Office Strategic Sales & Technical Office Contact Information: Phone: +65 9848 9856 Email: fssingapore@filtersafe.net