This document provides information about the Neo Screw Remover Kit, including:
1) The kit contains various guides, drills, and removers for removing fractured screws from dental fixtures. The components are available in different sizes to accommodate different fixture and screw diameters.
2) Detailed instructions are provided on selecting the appropriate guide, drill, and remover based on the size of the fractured screw or fixture connection. Steps are outlined for installing guides and using drills and removers to remove fractured screws.
3) Safety tips are included, such as only using properly sterilized components and operating drills at the appropriate RPM to avoid damaging fixtures.
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety recommendations when working on starters. It then provides step-by-step instructions for dismantling and assembling starters for different models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, and reassembling with new parts as needed. Special tools are recommended for some procedures, and warnings are provided about spring tension and flying parts that could cause injury.
This surgical tool kit is used to remove fractured dental implant screws. It consists of various guides, drills, and removers sized to match common dental implant screws. The procedure involves using a guide to stabilize a reverse-rotating drill that creates a hole in the fractured screw. A sized screw remover is then inserted and reverse-rotated using low speed to extract the broken screw fragment. Care must be taken to properly clean and sterilize the tools before use and select the correctly sized tool to avoid damaging the implant fixture.
A 47-year-old male patient was referred for removal of a fractured implant screw. A guide was screwed into the implant to position a 1.0mm reverse drill directly over the fractured screw. The drill was used to core out approximately 1mm of the screw. Then, a matching 1.0mm screw remover was used to unscrew the fractured piece in a counterclockwise direction, successfully removing it.
This fixture remover kit consists of dental implant surgical tools for removing failed implants. The kit contains fixture removers, fixture remover screws, and a torque control device for safely removing implants. Proper preparation, use, cleaning, and sterilization are required for safe use. The document provides detailed instructions on how to use the kit to remove implants.
NeoBiotech Orders & Support - 415.496.9636 (415.49MyNeo)
Earn AID VIP Credits by submitting orders or inquires via
'1st Choice Advanced & Innovative Dentistry'
www.1stChoiceAID.com
* Save 10% Discount Code below for items ordering from NeoBiotech Website
1stChoiceNeo
- FIXTURE REMOVER KIT (Patent product) -
New and Innovative concept since 2009
Easy way to repair problematic implant
You can remove failed or failing implant without bone damage
Replace same diameter implant immediately
No, less patient trauma!
The unique product which Specialists' one of the most favorite product to become a top notch paleontologist.
P.T. International Corp. provides power transmission products including mounted bearings, drive products, and European linear components. They are headquartered in Charlotte, NC with an 80,000 square foot facility stocking over 25,000 SKUs. The document discusses their global manufacturing and sourcing, engineering capabilities, and ISO 9001:2008 certification. It also provides details on their product range, inspection processes, and installation guidelines for mounted spherical bearings.
The document describes a fixture remover tool that is used to remove failed dental implants or fixture components in 3 steps: 1) Assemble the remover tool and insert it into the failed fixture, 2) Gently rotate the tool to lock it to the fixture, 3) Continue rotating to detach the failed component without damaging surrounding bone. It also includes guides that attach to the fixture at different angles to help repair damaged internal threads.
Instruction Manual | EO Tech 2.5-10x FFP | Optics TradeOptics-Trade
This commodity is controlled under export regulations and requires an export license or exception to be exported abroad. The document provides an overview and instructions for the features, adjustments, mounting, and maintenance of the Vudu 2.5-10x riflescope. It describes the magnification, reticle focus, illumination, turret, battery, and eye relief adjustments as well as the zeroing procedure and warranty information.
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety recommendations when working on starters. It then provides step-by-step instructions for dismantling and assembling starters for different models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, and reassembling with new parts as needed. Special tools are recommended for some procedures, and warnings are provided about spring tension and flying parts that could cause injury.
This surgical tool kit is used to remove fractured dental implant screws. It consists of various guides, drills, and removers sized to match common dental implant screws. The procedure involves using a guide to stabilize a reverse-rotating drill that creates a hole in the fractured screw. A sized screw remover is then inserted and reverse-rotated using low speed to extract the broken screw fragment. Care must be taken to properly clean and sterilize the tools before use and select the correctly sized tool to avoid damaging the implant fixture.
A 47-year-old male patient was referred for removal of a fractured implant screw. A guide was screwed into the implant to position a 1.0mm reverse drill directly over the fractured screw. The drill was used to core out approximately 1mm of the screw. Then, a matching 1.0mm screw remover was used to unscrew the fractured piece in a counterclockwise direction, successfully removing it.
This fixture remover kit consists of dental implant surgical tools for removing failed implants. The kit contains fixture removers, fixture remover screws, and a torque control device for safely removing implants. Proper preparation, use, cleaning, and sterilization are required for safe use. The document provides detailed instructions on how to use the kit to remove implants.
NeoBiotech Orders & Support - 415.496.9636 (415.49MyNeo)
Earn AID VIP Credits by submitting orders or inquires via
'1st Choice Advanced & Innovative Dentistry'
www.1stChoiceAID.com
* Save 10% Discount Code below for items ordering from NeoBiotech Website
1stChoiceNeo
- FIXTURE REMOVER KIT (Patent product) -
New and Innovative concept since 2009
Easy way to repair problematic implant
You can remove failed or failing implant without bone damage
Replace same diameter implant immediately
No, less patient trauma!
The unique product which Specialists' one of the most favorite product to become a top notch paleontologist.
P.T. International Corp. provides power transmission products including mounted bearings, drive products, and European linear components. They are headquartered in Charlotte, NC with an 80,000 square foot facility stocking over 25,000 SKUs. The document discusses their global manufacturing and sourcing, engineering capabilities, and ISO 9001:2008 certification. It also provides details on their product range, inspection processes, and installation guidelines for mounted spherical bearings.
The document describes a fixture remover tool that is used to remove failed dental implants or fixture components in 3 steps: 1) Assemble the remover tool and insert it into the failed fixture, 2) Gently rotate the tool to lock it to the fixture, 3) Continue rotating to detach the failed component without damaging surrounding bone. It also includes guides that attach to the fixture at different angles to help repair damaged internal threads.
Instruction Manual | EO Tech 2.5-10x FFP | Optics TradeOptics-Trade
This commodity is controlled under export regulations and requires an export license or exception to be exported abroad. The document provides an overview and instructions for the features, adjustments, mounting, and maintenance of the Vudu 2.5-10x riflescope. It describes the magnification, reticle focus, illumination, turret, battery, and eye relief adjustments as well as the zeroing procedure and warranty information.
The work procedure outlines the steps for manufacturing a front standard part using various tools over 45 minutes. Key steps include loading the bottom assembly into a vice, drilling holes using a manual mill while changing drill bits as needed, loosening set screws to remove the top plate, and using a hand tap to thread holes before removing the finished part. Safety equipment like eye protection and gloves are required.
Husqvarna hedge trimmers service repair manualdikdkkdkdm
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general recommendations for working safely and properly with starters. It then provides step-by-step instructions for dismantling and assembling starters for specific models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, and replacing parts. Special tools are noted where needed and safety warnings are highlighted.
Husqvarna trimmers service repair manualdikdkkdkdm
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
Husqvarna ice drills service repair manualfjskemdme
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety recommendations when working on starters. It then provides step-by-step instructions for dismantling and assembling starters for different models, including the models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Replacement of parts like springs, pulleys, bearings and drive dogs are described.
Husqvarna pruning saws service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general recommendations for working safely and properly with starters. It then provides step-by-step instructions for dismantling and assembling starters for specific models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, and replacing parts. Special tools are noted where needed and safety warnings are highlighted.
Husqvarna ice drills service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
Husqvarna pruners service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general recommendations for working safely and properly with starters. It then provides step-by-step instructions for dismantling and assembling starters for specific models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, installing new springs and cords, and tensioning return springs. Special tools are noted where needed for certain procedures.
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety recommendations when working on starters. It then provides step-by-step instructions for dismantling and assembling starters for different models, including the models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Special tools required for some procedures are also noted.
Husqvarna blowers service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
Dr. Park is a clinical professor at UCLA School of Dentistry who teaches and conducts research in prosthodontics and implant dentistry. He has extensively researched medication-related osteonecrosis of the jaw and medication-assisted osseointegration using different implant surface topographies. Recently, he has been investigating the accuracy of measuring primary and secondary implant stability. Dr. Park received his dental degrees from Tufts University and completed advanced prosthodontics training and a fellowship at UCLA.
This document appears to contain a series of numbers with no other context. It lists numbers from 0405 to 1312 across multiple lines without any other text, headings or explanations. The essential information and purpose of the document cannot be determined from the limited numeric content provided.
This study evaluated the outcome of immediately loading 15 dental implants in 4 patients over a mean follow-up period of 4.8 months. The mean marginal bone loss from implant surgery to immediate loading was 0.03 ± 0.07 mm, and 0.16 ± 0.17 mm after 3 months of continued loading. No implants failed, resulting in a 100% survival rate. The results suggest that immediate loading of dental implants can achieve high success rates of up to 6 months. However, long-term clinical studies with larger sample sizes are still needed.
User guide universal prosthetic kit -151207Kei Lim
This Universal Prosthetic Kit contains dental implant procedure tools and drivers made of medical grade materials. It is used for implanting fixtures, connecting abutments, and detaching cover screws. The kit includes hex, SCS, and Unigrip drivers, a driver holder, and torque ratchet. It must be sterilized via autoclave before use according to the instructions, which include preparing components, washing after use, and storing at room temperature.
The document outlines an agenda for the 2016 Neobiotech World Symposium taking place June 25-26, 2016 in Los Angeles, California. The precongress event on June 25th will feature lectures and demonstrations on topics such as sinus surgery, implant placement, and guided bone regeneration. The main symposium on June 26th will include speakers presenting on innovative sinus surgery, implant loading protocols, prosthetic complications, peri-implant diseases, bone grafting techniques, and new technologies. Presenters include experts from the United States and South Korea specializing in oral surgery, periodontics, and prosthodontics.
The document introduces Neobiotech's AnyTime Loading concept for dental implants, which allows prosthesis loading within 1 to 4 weeks or at any time. This concept provides clinics with greater competitive advantages over other implant systems and more benefits to patients. Key advantages include reduced treatment time, immediate functional restoration, and increased patient satisfaction. Neobiotech's IS-II active implant is optimized for AnyTime Loading through its tapered design, reversed threading, specialized surface treatment and drilling protocol, which help achieve high primary stability and maintain stability longer to reduce the risk of early loading failure.
Prosthetic guide
www.neobiotechus.com
www.neobiotech.com
Download is available.
Click the content, you can see directly the content that you want to see.
Predictable and Accurate 3D Guided implant placement for a 3-D worldKei Lim
We live in a 3 dimensional world yet still place dental implants using 2D techniques involving guessing. New 3D guided implant placement allows for accurate and predictable implant positioning in all 3 planes using a surgical guide. This seminar will cover how to create a 3D surgical guide from scans or impressions, the equipment needed, and a hands-on demonstration of placing an implant using a 3D guide for improved accuracy over conventional techniques.
This document provides information on the Neobiotech CAD/CAM system, including specifications for its milling machines, furnaces, and materials. Key details include:
- The CM3i, CM4i, and CM5i milling machines with their axis configurations, spindle power, sizes, and material milling times.
- The module upgrade system that allows axis upgrading from 3 to 5 axes for increased accuracy and efficiency.
- Compatible materials that can be milled including zirconia, titanium, PMMA, and resin, as well as a sintering furnace for final processing.
- Associated tools, blocks, links, and scan bodies made from materials like titanium,
The document provides information on Neobiotech's Cameleon CAD/CAM system, an integrated digital solution for dentistry. It describes the TRIOS intraoral scanner, 3Shape model scanners including the D700, CAD and CAM software, milling machines, and dental materials compatible with the system. The Cameleon system aims to streamline workflows, improve communication, and provide accurate, esthetic restorations using innovative technology and materials.
The document is Neobiotech's 2015 product catalog. It includes:
- An introduction to Neobiotech's corporate philosophy and history of developing simpler, safer, and faster implant products.
- Details on Neobiotech's research and development facilities and production process.
- An overview of Neobiotech's implant systems including the IS-II active, IT-II active, EB-II active, and S-mini implants.
- Descriptions of the surgical kits, guides, prosthetic components, and digital solutions that accompany the implant systems.
The Neo Surgical Accessory Kit contains several tools to aid in implant surgery including a tissue punch for removing soft tissue, a dia marker for checking implant position before drilling, a bone planner for trimming irregular bone surfaces, and a bone profiler for trimming overgrown bone. It also includes QuickTight high speed hex drivers that automatically stop when torque exceeds 2Ncm.
The work procedure outlines the steps for manufacturing a front standard part using various tools over 45 minutes. Key steps include loading the bottom assembly into a vice, drilling holes using a manual mill while changing drill bits as needed, loosening set screws to remove the top plate, and using a hand tap to thread holes before removing the finished part. Safety equipment like eye protection and gloves are required.
Husqvarna hedge trimmers service repair manualdikdkkdkdm
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general recommendations for working safely and properly with starters. It then provides step-by-step instructions for dismantling and assembling starters for specific models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, and replacing parts. Special tools are noted where needed and safety warnings are highlighted.
Husqvarna trimmers service repair manualdikdkkdkdm
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
Husqvarna ice drills service repair manualfjskemdme
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety recommendations when working on starters. It then provides step-by-step instructions for dismantling and assembling starters for different models, including the models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Replacement of parts like springs, pulleys, bearings and drive dogs are described.
Husqvarna pruning saws service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general recommendations for working safely and properly with starters. It then provides step-by-step instructions for dismantling and assembling starters for specific models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, and replacing parts. Special tools are noted where needed and safety warnings are highlighted.
Husqvarna ice drills service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
Husqvarna pruners service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general recommendations for working safely and properly with starters. It then provides step-by-step instructions for dismantling and assembling starters for specific models, including releasing spring tension, removing starter pulleys and springs, checking components for wear, installing new springs and cords, and tensioning return springs. Special tools are noted where needed for certain procedures.
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety recommendations when working on starters. It then provides step-by-step instructions for dismantling and assembling starters for different models, including the models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Special tools required for some procedures are also noted.
Husqvarna blowers service repair manualdhjfjkmmmde
The document provides instructions for dismantling and assembling starters for various brush cutter and trimmer models. It begins with general safety warnings and recommendations. It then provides step-by-step instructions for dismantling and assembling starters for different models including the Models 265, 250/252, 240/245, 225, 232, 235, 240, 322, 325, 122, 32, Mondo, 18H, 140B and 141B. Instructions include releasing spring tension, removing starter pulleys and springs, checking parts for wear, fitting new springs and cords, and tensioning the return spring.
Dr. Park is a clinical professor at UCLA School of Dentistry who teaches and conducts research in prosthodontics and implant dentistry. He has extensively researched medication-related osteonecrosis of the jaw and medication-assisted osseointegration using different implant surface topographies. Recently, he has been investigating the accuracy of measuring primary and secondary implant stability. Dr. Park received his dental degrees from Tufts University and completed advanced prosthodontics training and a fellowship at UCLA.
This document appears to contain a series of numbers with no other context. It lists numbers from 0405 to 1312 across multiple lines without any other text, headings or explanations. The essential information and purpose of the document cannot be determined from the limited numeric content provided.
This study evaluated the outcome of immediately loading 15 dental implants in 4 patients over a mean follow-up period of 4.8 months. The mean marginal bone loss from implant surgery to immediate loading was 0.03 ± 0.07 mm, and 0.16 ± 0.17 mm after 3 months of continued loading. No implants failed, resulting in a 100% survival rate. The results suggest that immediate loading of dental implants can achieve high success rates of up to 6 months. However, long-term clinical studies with larger sample sizes are still needed.
User guide universal prosthetic kit -151207Kei Lim
This Universal Prosthetic Kit contains dental implant procedure tools and drivers made of medical grade materials. It is used for implanting fixtures, connecting abutments, and detaching cover screws. The kit includes hex, SCS, and Unigrip drivers, a driver holder, and torque ratchet. It must be sterilized via autoclave before use according to the instructions, which include preparing components, washing after use, and storing at room temperature.
The document outlines an agenda for the 2016 Neobiotech World Symposium taking place June 25-26, 2016 in Los Angeles, California. The precongress event on June 25th will feature lectures and demonstrations on topics such as sinus surgery, implant placement, and guided bone regeneration. The main symposium on June 26th will include speakers presenting on innovative sinus surgery, implant loading protocols, prosthetic complications, peri-implant diseases, bone grafting techniques, and new technologies. Presenters include experts from the United States and South Korea specializing in oral surgery, periodontics, and prosthodontics.
The document introduces Neobiotech's AnyTime Loading concept for dental implants, which allows prosthesis loading within 1 to 4 weeks or at any time. This concept provides clinics with greater competitive advantages over other implant systems and more benefits to patients. Key advantages include reduced treatment time, immediate functional restoration, and increased patient satisfaction. Neobiotech's IS-II active implant is optimized for AnyTime Loading through its tapered design, reversed threading, specialized surface treatment and drilling protocol, which help achieve high primary stability and maintain stability longer to reduce the risk of early loading failure.
Prosthetic guide
www.neobiotechus.com
www.neobiotech.com
Download is available.
Click the content, you can see directly the content that you want to see.
Predictable and Accurate 3D Guided implant placement for a 3-D worldKei Lim
We live in a 3 dimensional world yet still place dental implants using 2D techniques involving guessing. New 3D guided implant placement allows for accurate and predictable implant positioning in all 3 planes using a surgical guide. This seminar will cover how to create a 3D surgical guide from scans or impressions, the equipment needed, and a hands-on demonstration of placing an implant using a 3D guide for improved accuracy over conventional techniques.
This document provides information on the Neobiotech CAD/CAM system, including specifications for its milling machines, furnaces, and materials. Key details include:
- The CM3i, CM4i, and CM5i milling machines with their axis configurations, spindle power, sizes, and material milling times.
- The module upgrade system that allows axis upgrading from 3 to 5 axes for increased accuracy and efficiency.
- Compatible materials that can be milled including zirconia, titanium, PMMA, and resin, as well as a sintering furnace for final processing.
- Associated tools, blocks, links, and scan bodies made from materials like titanium,
The document provides information on Neobiotech's Cameleon CAD/CAM system, an integrated digital solution for dentistry. It describes the TRIOS intraoral scanner, 3Shape model scanners including the D700, CAD and CAM software, milling machines, and dental materials compatible with the system. The Cameleon system aims to streamline workflows, improve communication, and provide accurate, esthetic restorations using innovative technology and materials.
The document is Neobiotech's 2015 product catalog. It includes:
- An introduction to Neobiotech's corporate philosophy and history of developing simpler, safer, and faster implant products.
- Details on Neobiotech's research and development facilities and production process.
- An overview of Neobiotech's implant systems including the IS-II active, IT-II active, EB-II active, and S-mini implants.
- Descriptions of the surgical kits, guides, prosthetic components, and digital solutions that accompany the implant systems.
The Neo Surgical Accessory Kit contains several tools to aid in implant surgery including a tissue punch for removing soft tissue, a dia marker for checking implant position before drilling, a bone planner for trimming irregular bone surfaces, and a bone profiler for trimming overgrown bone. It also includes QuickTight high speed hex drivers that automatically stop when torque exceeds 2Ncm.
This document provides steps to treat peri-implantitis:
1. Clean the contaminated implant surface with a special brush to remove bacteria and promote reosseointegration after bone grafting.
2. For severe cases, use a brush to remove the rough implant surface and create a new machined surface for soft tissue attachment and bone preservation.
3. Replace a contaminated implant with a new one using an FR Kit to unscrew the old implant with minimal bone damage and allow for immediate replacement.
The document provides information about the ACMAuto Chip Maker instrument, which is used to collect autograft bone from mandibular molars or incisors for guided bone regeneration in defective areas. The instrument comes in various diameters and lengths. It must be cleaned, sterilized and stored properly after use. A clinical case and movie demonstrate how to use the instrument to drill and collect autograft bone chips within seconds for GBR procedures.
This document describes a new drill device called the NeoBiotech Auto Chip Maker (ACM) that can easily collect autogenous bone in substantial amounts from the posterior mandible. The drill is attached to a handpiece and used to collect bone from a donor site, such as the ramus, at 70 RPM without irrigation. The collected bone is trapped in the drill by a plastic stopper. It is then placed in an implant defect area and covered with a membrane for guided bone regeneration. At a follow-up surgery 3-4 months later, significant new bone formation was observed around the implant and previously deficient site. The collected bone has a coagulated and spongy consistency and the advantages of autogenous bone grafting while
This document provides information on the Neo SinusAll Kit, which combines several surgical kits into one for maxillary sinus implant procedures. It includes drills, surgical tools, and drivers made of medical grade materials like stainless steel. The kit allows sinus procedures to take place using a single kit and includes both crestal and lateral approach tool components, along with tools for implant placement. Instructions are provided on the use, sterilization, storage and maintenance of the kit components.
This document provides information about a basic implant training course offered by the Global Academy of Osseointegration. The 3-day course will teach doctors how to surgically place and restore dental implants, identify ideal implant candidates, and diagnose and treatment plan implant cases. The training includes hands-on exercises with typodont models, and each participant will have the opportunity to perform a live implant surgery under supervision. The course aims to provide dentists with the skills and knowledge to incorporate implant dentistry into their practices.
The CAMeleon CS is a 5-axis milling machine capable of high-precision machining of dental materials like zirconia, glass ceramics, composite resins, and PMMA. It features a 500W spindle that can reach speeds up to 100,000 RPM, allowing it to mill materials both dry and wet. The machine has a 6-tool automatic changer and accuracy of ±0.001mm via servo motors and harmonic drives installed on all axes. It is suitable for milling dental copings, crowns, and other small dental parts at a high capacity of 700-1000 parts per month.
The document describes the NeoGuideSystem for digital implant planning. It provides full digital workflow to quickly fabricate surgical guides for accurate, safe implant surgery. The system includes NeoGuide software for implant planning, surgical guides for guiding surgery, and a NeoGuide kit with drills and other tools. The full digital process takes only 5 days from scan to surgery. The system aims to provide faster, safer, and more accurate implant placement for patients, clinics, and dental labs.
Travel vaccination in Manchester offers comprehensive immunization services for individuals planning international trips. Expert healthcare providers administer vaccines tailored to your destination, ensuring you stay protected against various diseases. Conveniently located clinics and flexible appointment options make it easy to get the necessary shots before your journey. Stay healthy and travel with confidence by getting vaccinated in Manchester. Visit us: www.nxhealthcare.co.uk
“Psychiatry and the Humanities”: An Innovative Course at the University of Mo...Université de Montréal
“Psychiatry and the Humanities”: An Innovative Course at the University of Montreal Expanding the medical model to embrace the humanities. Link: https://www.psychiatrictimes.com/view/-psychiatry-and-the-humanities-an-innovative-course-at-the-university-of-montreal
Know the difference between Endodontics and Orthodontics.Gokuldas Hospital
Your smile is beautiful.
Let’s be honest. Maintaining that beautiful smile is not an easy task. It is more than brushing and flossing. Sometimes, you might encounter dental issues that need special dental care. These issues can range anywhere from misalignment of the jaw to pain in the root of teeth.
Computer in pharmaceutical research and development-Mpharm(Pharmaceutics)MuskanShingari
Statistics- Statistics is the science of collecting, organizing, presenting, analyzing and interpreting numerical data to assist in making more effective decisions.
A statistics is a measure which is used to estimate the population parameter
Parameters-It is used to describe the properties of an entire population.
Examples-Measures of central tendency Dispersion, Variance, Standard Deviation (SD), Absolute Error, Mean Absolute Error (MAE), Eigen Value
The biomechanics of running involves the study of the mechanical principles underlying running movements. It includes the analysis of the running gait cycle, which consists of the stance phase (foot contact to push-off) and the swing phase (foot lift-off to next contact). Key aspects include kinematics (joint angles and movements, stride length and frequency) and kinetics (forces involved in running, including ground reaction and muscle forces). Understanding these factors helps in improving running performance, optimizing technique, and preventing injuries.
How to Control Your Asthma Tips by gokuldas hospital.Gokuldas Hospital
Respiratory issues like asthma are the most sensitive issue that is affecting millions worldwide. It hampers the daily activities leaving the body tired and breathless.
The key to a good grip on asthma is proper knowledge and management strategies. Understanding the patient-specific symptoms and carving out an effective treatment likewise is the best way to keep asthma under control.
Nano-gold for Cancer Therapy chemistry investigatory projectSIVAVINAYAKPK
chemistry investigatory project
The development of nanogold-based cancer therapy could revolutionize oncology by providing a more targeted, less invasive treatment option. This project contributes to the growing body of research aimed at harnessing nanotechnology for medical applications, paving the way for future clinical trials and potential commercial applications.
Cancer remains one of the leading causes of death worldwide, prompting the need for innovative treatment methods. Nanotechnology offers promising new approaches, including the use of gold nanoparticles (nanogold) for targeted cancer therapy. Nanogold particles possess unique physical and chemical properties that make them suitable for drug delivery, imaging, and photothermal therapy.
2. Neo SR(Screw Remover) kit
140
NeoSRkit
This is kit consisting of the tools designed to remove the fractured screw fast and
safely in case that the screw inside the inserted fixture are fractured due to several reasons.
Introduction
Perfect Guide
Guide
Go to Index
4. NeoSRkit
142
Neo SR(Screw Remover) kit
Neo SR(Screw Remover) KIT Component
1. Perfect Guide & PG Driver
Perfect Guide plays a role of the guide of Reverse Drill and is used when the screw is fractured downward
female screw inside the fixture, and is used by attaching directly to the fixture female screw using a PG Driver.
- As there is a specially designed bump which enables to push the fractured screw together by turning it when attaching a Perfect
Guide, the guide can be attached perfect.
- Inside Perfect Guide, a drill stop is designed to avoid drilling more than 2mm.
- As this is attached directly to the Fixture for use, the tool to fix the guide is not needed when drilling.
Application fixture Application
Standard Product Name female screw Reverse Drill
ø0.8 / M1.6 PG 0816 M1.6 RCD 08
ø1.0 / M1.8 PG 1018 M1.8 RCD 10
ø1.2 / M2.0 PG 1220 M2.0 RCD 10, 12
ø1.4 / M2.5 PG 1425 M2.5 RCD 12, 14
Hex(mm) Product Name Application Product
2.5 PGHD 25S PG0816, PG1018, PG1220, PG1425
Be sure to use the same size of Perfect Guide as fixture female screw.
Perfect Guide
PG Driver
2. Conical, Internal & External Guide
Like Perfect Guide, these instruments play a role of the guide of Reverse Drill and are used when a Perfect
Guide can not be used, or when the screw is fractured in the deep area downward more than 2mm in the
beginning of female screw inside the fixture, and select the proper guide to use according to the internal
connection.
Size Product Name Application Fixture
11 /8 CG 00 Having 11 or 8 internal connection Fixture
Size Product Name Application Fixture
2.4 Hex IHG 24 Having 2.4 Hex internal connection Fixture
2.5 Hex IHG 25 Having 2.5 Hex internal connection Fixture
3.1 Octa IOG 31 Having 3.1 Octa internal connection Fixture
Size Product Name Application Fixture
2.4 Hex EHG 24 Having 2.4 Hex external connection Fixture
2.7 Hex EHG 27 Having 2.7 Hex external connection Fixture
3.4 Hex EHG 34 Having 3.4 Hex external connection Fixture
Conical Guide
Internal Guide
External Guide
Go to Index
5. NeoSRkit
143
Neo SR(Screw Remover) kit
4. Reverse Drill
- This is an instrument to form the hole for screw remover to insert on the
fractured screw before using a screw remover which has the function to
remove the screw. According to edge diameter, it is consisted of 4types
and selects a reverse drill suitable for the size of screw desired to remove.
- This enables to adjust the drilling the drilling depth 1mm~3mm as it is
marked.
1mm
1mm
1mm
Diameter( ) Product Name Fractured Screw Size
0.8 RCD 08 M1.6
1.0 RCD 10 M1.8 or M2.0
1.2 RCD 12 M2.0 or M2.5
1.4 RCD 14 M2.5
Diameter
Abutment Screw standard is different according to the manufacturer but generally, except for
special case, the screws of M1.6, M1.8, M2.0, M2.5 are used.
("M" means the meter screw and the number means the diameter of screw. For example,
“M1.6 screw” means the meter screw with screw diameter 1.6mm.)
Neo SR(Screw Remover) KIT Component
3. Guide Holder
This is an instrument to hold by hand and fix the Internal Guide, External Guide and Conical Guide which
play a role of the guide when using a Reverse Drill, after attaching them to this instrument.
Standard Product Name Application Product
Conical Guide : CG00
3.5 Double Hex GH 00 Internal Guide : IHG24, IHG25, IOG31
External Guide : EHG24, EHG27, EHG34
In case of impossible use of Perfect Guide situation, use Guide Holder with Conical, Internal, and
External Guide.
Go to Index
6. NeoSRkit
Neo SR(Screw Remover) kit
Neo SR(Screw Remover) KIT Component
144
5. Screw Remover
This is an instrument to remove the fractured screw by inserting the screw
remover suitable for the hole formed by a reverse drill and turning
counterclockwise at less than 80rpm. Screw Remover is designed with the
taper structure so that the more it turns the ore the friction force required to
loosen increases.
SR14 is use for broken hex of the screw.
Diameter( ) Product Name Application
0.8~1.1 SR 08 When the hole is formed by Reverse Drill ø0.8(RCD08)
1.0~1.3 SR 10 When the hole is formed by Reverse Drill ø1.0(RCD10)
1.2~1.5 SR 12 When the hole is formed by Reverse Drill ø1.2(RCD12)
1.4~1.7 SR 14 When the hole is formed by Reverse Drill ø1.4(RCD14)
6. Shank Driver
This is an instrument designed to use by hand or torque wrench or ratchet by
attaching the screw remover to this instrument.
Product Name Application Product
SHD 00 SR08, SR10, SR12, SR14
Go to Index
7. NeoSRkit
145
Neo SR(Screw Remover) kit
User Guide
1. Sterilize
Sterilize the desired components to use only or sterilize the whole kit that the components are
inserted.
2. Find standard size for Fixture & Screw
Check the size of Fixture connection or the desired fractured screw to remove or the female screw
inside the fixture.
If you don't know the size of Fixture connection or the fractured screw or the female screw of fixture, be sure
to ask the manufacturer or agent for information.
3. Guide for selection & connection
Select the suitable guide for Fixture connection and screw size and attach it to the fixture.
1. Usage of Perfect guide.
Use a PG Driver to attach the Perfect Guide which corresponds to the size of female screw of fixture, to the
female screw of the fixture.
Attach it with about 10Ncm which is a torque enough to tighten by the hand force.
2. Usage of External, conical, internal guide.
When there is a situation which cannot use Perfect Guide or abutment screw has been fractured more than
2mm, select one proper guide among External Guide, Conical Guide, and Internal Guide and install the
selected guide on the Guide Holder and then attach it to the fixture.
ex) External Regular ex) Internal Regular
Fractured screw Fractured screw
Fixture connection
External : 2.4Hex, 2.7Hex, 3.3Hex ,3.4Hex
Internal : 2.4Hex, 2.5Hex, 3.1 Octa
Screw standard
In general M1.6, M1.8, M2.0, M2.5 are mostly fitted
Go to Index
8. NeoSRkit
146
Neo SR(Screw Remover) kit
User Guide
4. Usage of Reverse Drill
Select suitable Reverse Drill to the size of screw desired to remove and attach it to the contra angle of
surgical engine and then insert it in the Guide installed in the fixture and turn it counterclockwise at
1,200rpm~1,300rpm to form 1~2mm hole on the fractured screw. As the heat may generate by drilling, the
sufficient irrigation is needed.
To reduce the mistake to attach the Perfect Guide which has different size with that of female screw of Fixture,
attach the Perfect Guide with a bigger diameter first and if not fitted, try to attach the smaller Perfect Guide.
5. Usage of Screw Remover
After removing the attached guide, select the Screw Remover suitable for the hole formed by a Reverse
Drill and attach it to the contra angle of surgical engine and then press the Screw Remover on the formed
hole with the proper force and turn it counterclockwise less than 80rpm to loosen the screw.
If a Screw Remover is attached to Shank Driver, it is available to loosen the fractured screw by hand or by
connecting a Ratchet or Torque Wrench.
Perfect Guide
As a drill stop is designed inside Perfect Guide, the drilling more than 2mm is not possible. However, for External
Guide, Conical Guide and Internal Guide, a drill stop is not designed other than Perfect Guide. Thus, when drilling,
be sure to check the strip marking which indicates the drilling depth of Reverse Drill.
Clock
wise
Counter
clockwise
Counter
clockwise
Insert to the hole
Connect PG Driver to Perfect Guide
before connecting Perfect Guide to
the fixture.
- In order to make 1~2mm hole of fractured
screw, it needs to insert Reverse drill to the
Guide and drill counterclockwise with irrigation.
- Do not connect the Guide while the drill is
rotating. You must drill when Reverse drill
toughs the surface of the screw.
Remove the Guide with the drill, the insert a suitable
Screw Remover to the hole and remove the fractured
screwbyrotatingcounterclockwise.
In case the screw is fractured downward
female screw inside the fixture. In order to use the
Perfect Guide, female
screw needs minimum
helix in the fixture.
In case the screw is fractured downward
female screw inside the fixture.
Go to Index
9. Neo SR(Screw Remover) kit
NeoSRkit
147
User Guide
External, Internal, Conical Guide
Install the selected guide on the Guide
Holder and then attach it to the fixture by
hand.
When there is over torque by combining the Fixture and the abutment, hex of the screw head can be distorted. On this case, using screw
remover kit, screw of abutment can be easily removed.
1. Insert 1.4 screw remover (SR14) counterclockwise by below 80 rpm to remove the screw. (Picture 1->2 )
2. In case the screw remover runs by itself not with the screw, then use 1.4 reverse drill (RCD14) to the hex and run 1,200~ 1,300rpm
counterclockwise and make 1~1.5mm hole and then use 1.4 screw remover(SR14) again. (Picture 1->2->3->1->2)
-In order to make 1~2mm hole of fractured screw, it
needs to insert Reverse drill to the Guide and drill
counterclockwisewithirrigation.
-Do not connect the Guide while the drill is rotating. You
must drill when Reverse drill toughs the surface of the
screw.
In case of broken hex of Abutment screw.
Remove the Guide with the drill, the insert a suitable
Screw Remover to the hole and remove the fractured
screwbyrotatingcounterclockwise.
1. Read the purpose of the product and how to use thoroughly before using.
2. If you select a Tool wrong, the fixture may be damaged. So, after checking the size of screw
desired to remove or fixture connection, select the proper Guide, Reverse Drill, and Reverse Tap
to use.
3. If Perfect Guide is not attached well, do not force to attach it but select the proper guide suitable
for the fixture connection among External Guide, Conical Guide, and Internal Guide. If you force to
attach the Perfect Guide, the fixture may be damaged.
4. Reverse Drill and Screw Remover should be used at optimal rpm.
5. If the used tool is left alone, it may cause the rust or corrosion. Be sure to dry thoroughly after
washing and store it in the dry place at ambient temperature.
Notices
Counter
clockwise
Counter
clockwise
Counter
clockwise
Counter
clockwise
Insert to the hole
Go to Index
10. Neo SR(Screw Remover) kit
148
NeoSRkit
Perfect Guide screw removal sequence.
When you use the Perfect Guide, in case of M2.0 and M2.5, you can remove the fractured screw by
sequence of method1 and 2.
In case of removing M2.0 and M2.5 size screw, if you fail to remove the screw by sequence of the method1,
you can have one more chance removing the screw by sequence of the method2.
Use a PG Driver to attach the Perfect
Guide which corresponds to the size of
female screw of fixture to the female
screw of the fixture by the hand force.
(Torque: About 10Ncm)
Connect the Reverse Drill with the contra
angle of surgical engine and then insert it
in the Perfect Guide installed in the fixture
and turn it at 1,200~1,300 rpm to from the
1~2mm hole on the fractured screw.
After removing the attached guide, select
the Screw Remover suitable for the hole
formed by a Reverse Drill and attach it to
the contra angle of surgical engine and
then press the Screw Remover on the
formed hole with the proper force and turn
it less than 80 rpm to loosen the screw.
Go to Index
11. Neo SR(Screw Remover) kit
NeoSRkit
149
External, Internal, Conical Guide screw
removal sequence
In case of a internal fixture, you can use the Internal Guide or Conical Guide
In case of removing M2.0 and M2.5 size screw, if you fail to remove the screw by sequence of the method1,
you can have one more chance removing the screw by sequence of the method2.
Use a PG Driver to attach the Perfect Guide
which corresponds to the size of female
screw of fixture to the female screw of the
fixture by the hand force. (Torque : About
10Ncm)
Connect the Reverse Drill with the contra
angle of surgical engine and then insert it in
the Perfect Guide installed in the fixture and
turn it at 1,200~1,300rpm to from the
1~2mm hole on the fractured screw.
After removing the attached guide, select the
Screw Remover suitable for the hole formed
by a Reverse Drill and attach it to the contra
angle of surgical engine and then press the
Screw Remover on the formed hole with the
proper force and turn it less than 80rpm to
loosen the screw.
According to Fixture connection, select one
paper guide among External Guide, Conical
Guide, and Internal Guide and install the
selected guide on the Guide Holder and
then attach it to the fixture.
Go to Index
13. Neo Titanium Mesh
166
NeoTitaniumMesh
Introduction
Type of Ti-mesh
Neo titanium mesh can be trimmed by customer’s need (Thickness 0.085mm hole
size 0.8)
20mm 25mm
35mm 50mm
20mm
12mm
35mm
25mm
Size Code Product Name
20 x 12 T1 TMN 122008
25 x 20 T2 TMN 202508
35 x 25 T3 TMN 253508
50 x 35 T4 TMN 355008
Go to Index