Presentation Topic:
Energy Dispersive Spectrometry
(EDS)
Energy Dispersive Spectrometry
➢Function
➢Introduction
➢working of EDS
➢Principle
➢Basic components
➢Results
Energy Dispersive Spectrometry:
Energy-dispersive X-ray spectroscopy (EDS, EDX, EDXS or XEDS), sometimes called
energy dispersive X-ray analysis (EDXA or EDAX) or energy dispersive X-ray microanalysis
(EDXMA)
Function:
EDS is a technique used for the elemental analysis or chemical characterization of a
sample. It relies on an interaction of some source of X-ray excitation and a sample.
EDS can be used to find the chemical composition of materials down to a spot size of
a few microns, and to create element composition maps over a much broader raster
area.
Energy Dispersive Spectrometry:
Introduction:
➢Energy Dispersive Spectrometry (EDS) was first introduced in the late
1960s.
➢ Before that time, the wavelength-dispersive spectrometer (WDS) was used
for x-ray characterization.
➢In the late 1960s, Fitzgerald, Keil, and Heinrich first used the solid state
detector as an electron beam micro analyzer.
Diagram:
The basic design of an energy dispersive spectrometer is shown in Figure.
Basic Principle:
Principle of EDS.
Basic Principle:
➢The incident beam may excite an electron in an inner shell, ejecting it from the shell
while creating an electron hole where the electron was.
➢An electron from an outer, higher-energy shell then fills the hole, and the difference
in energy between the higher-energy shell and the lower energy shell may be
released in the form of an X-ray.
➢The number and energy of the X-rays emitted from a specimen can be measured by
an energy-dispersive spectrometer. the energy of the X-rays are characteristic of the
difference in energy between the two shells, and of the atomic structure of the
element from which they were emitted, this allows the elemental composition of the
specimen to be measured.
Working:
➢The detector generates a charge pulse proportional to the X-ray energy
➢This pulse is first converted to a voltage.
➢ Then the signal is amplified through a field effect transistor (FET)
,isolated from other pulses , further amplified ,then identified electronically
as resulting from an X-ray of specific energy
➢Finally, a digitized signal is stored in a channel assigned to that energy in the
MCA (Multi Channel Analyzer).
Basic Components:
Detector
• Window
• Cryostat
Pulse
Processor
• Amplifier
Multi-
Chanel
Analyzer
• MCA
Detector:
Window:
➢The window is made up of Beryllium.
➢The window provides a barrier to maintain vacuum within the detector
whilst being as transparent as possible to low energy X-rays.
Cryostat:
➢- The charge signals generated by the detector are small and can only be
separated from the electronic noise of the detector if the noise is reduced
by cooling the crystal and FET.
➢The Field Effect Transistor is the first stage of the amplification process that
measures the charge liberated in the crystal by an incident X-ray and
converts it to a voltage output.
➢- The natural width (FWHM) of an X-ray peak is on the order of 2-10 eV .
Pulse Processor:
➢The signal (voltage step) from the preamplifier is transformed into a voltage pulse
that is suitable for the multi channel analyzer.
➢The time over which the waveform is averaged is called the process time (Tp).
➢Tp is under control of the operator. The longer the Tp, the lower the noise but
more time is spent measuring each X- ray, and the fewer events that can be
measured.
➢If noise is minimized, the resolution of the peak displayed in the spectrum is
improved, and it becomes easier to separate or resolve, from another peak that is
close in energy.
Graph:
Multi-Chanel Analyzer (MCA)
➢ The MCA takes the data from the pulse processor and displays it as a
histogram of intensity (number of counts) vs voltage.
➢ The voltage range (for ex., 20 keV) displayed on the x- axis is divided into a
number (1024, 2048 etc.) of channels each corresponding to a given energy
range (for example, 5,280 eV –5,300 eV).
➢ The MCA takes the peak height of each voltage pulse, converts it into a
digital value, and puts it into the appropriate channel.
➢Thus a count is registered at that energy level.
Output forms of EDS:
Spectrum
• Spectrum: a plot of the number of
X-rays detected versus their
energies.
• The Characteristic X-rays allow the
elements present in the sample to
be identified.
Conti…
Map:
Map: an image showing how the
concentration of one element varies
over an area of a sample. In this
example, red colors indicate higher
concentrations and blue colors reflect
lower concentrations.
Energy Dispersive Spectroscopy

Energy Dispersive Spectroscopy

  • 1.
  • 2.
    Energy Dispersive Spectrometry ➢Function ➢Introduction ➢workingof EDS ➢Principle ➢Basic components ➢Results
  • 3.
    Energy Dispersive Spectrometry: Energy-dispersiveX-ray spectroscopy (EDS, EDX, EDXS or XEDS), sometimes called energy dispersive X-ray analysis (EDXA or EDAX) or energy dispersive X-ray microanalysis (EDXMA) Function: EDS is a technique used for the elemental analysis or chemical characterization of a sample. It relies on an interaction of some source of X-ray excitation and a sample. EDS can be used to find the chemical composition of materials down to a spot size of a few microns, and to create element composition maps over a much broader raster area.
  • 4.
    Energy Dispersive Spectrometry: Introduction: ➢EnergyDispersive Spectrometry (EDS) was first introduced in the late 1960s. ➢ Before that time, the wavelength-dispersive spectrometer (WDS) was used for x-ray characterization. ➢In the late 1960s, Fitzgerald, Keil, and Heinrich first used the solid state detector as an electron beam micro analyzer.
  • 5.
    Diagram: The basic designof an energy dispersive spectrometer is shown in Figure.
  • 6.
  • 7.
    Basic Principle: ➢The incidentbeam may excite an electron in an inner shell, ejecting it from the shell while creating an electron hole where the electron was. ➢An electron from an outer, higher-energy shell then fills the hole, and the difference in energy between the higher-energy shell and the lower energy shell may be released in the form of an X-ray. ➢The number and energy of the X-rays emitted from a specimen can be measured by an energy-dispersive spectrometer. the energy of the X-rays are characteristic of the difference in energy between the two shells, and of the atomic structure of the element from which they were emitted, this allows the elemental composition of the specimen to be measured.
  • 8.
    Working: ➢The detector generatesa charge pulse proportional to the X-ray energy ➢This pulse is first converted to a voltage. ➢ Then the signal is amplified through a field effect transistor (FET) ,isolated from other pulses , further amplified ,then identified electronically as resulting from an X-ray of specific energy ➢Finally, a digitized signal is stored in a channel assigned to that energy in the MCA (Multi Channel Analyzer).
  • 9.
    Basic Components: Detector • Window •Cryostat Pulse Processor • Amplifier Multi- Chanel Analyzer • MCA
  • 10.
    Detector: Window: ➢The window ismade up of Beryllium. ➢The window provides a barrier to maintain vacuum within the detector whilst being as transparent as possible to low energy X-rays.
  • 11.
    Cryostat: ➢- The chargesignals generated by the detector are small and can only be separated from the electronic noise of the detector if the noise is reduced by cooling the crystal and FET. ➢The Field Effect Transistor is the first stage of the amplification process that measures the charge liberated in the crystal by an incident X-ray and converts it to a voltage output. ➢- The natural width (FWHM) of an X-ray peak is on the order of 2-10 eV .
  • 12.
    Pulse Processor: ➢The signal(voltage step) from the preamplifier is transformed into a voltage pulse that is suitable for the multi channel analyzer. ➢The time over which the waveform is averaged is called the process time (Tp). ➢Tp is under control of the operator. The longer the Tp, the lower the noise but more time is spent measuring each X- ray, and the fewer events that can be measured. ➢If noise is minimized, the resolution of the peak displayed in the spectrum is improved, and it becomes easier to separate or resolve, from another peak that is close in energy.
  • 13.
  • 14.
    Multi-Chanel Analyzer (MCA) ➢The MCA takes the data from the pulse processor and displays it as a histogram of intensity (number of counts) vs voltage. ➢ The voltage range (for ex., 20 keV) displayed on the x- axis is divided into a number (1024, 2048 etc.) of channels each corresponding to a given energy range (for example, 5,280 eV –5,300 eV). ➢ The MCA takes the peak height of each voltage pulse, converts it into a digital value, and puts it into the appropriate channel. ➢Thus a count is registered at that energy level.
  • 15.
    Output forms ofEDS: Spectrum • Spectrum: a plot of the number of X-rays detected versus their energies. • The Characteristic X-rays allow the elements present in the sample to be identified.
  • 16.
    Conti… Map: Map: an imageshowing how the concentration of one element varies over an area of a sample. In this example, red colors indicate higher concentrations and blue colors reflect lower concentrations.