Effects of Variation of tube
Voltage Current, Filtration
Presenter: Dheeraj Kumar
MRIT, Ph.D. (Radiology and Imaging)
Assistant Professor
Medical Radiology and Imaging Technology
School of Health Sciences, CSJM University, Kanpur
Introduction
• In the field of medical radiography, optimizing critical parameters
including tube voltage, current, and filtration is a crucial undertaking.
• This introduction seeks to underscore the paramount importance of
achieving a delicate equilibrium between these factors, emphasizing
their collective impact on diagnostic accuracy and radiation safety.
Significance of Optimization
• The essence of optimizing tube voltage, current, and filtration lies in
fine-tuning the delicate interplay between achieving superior image
quality and minimizing patient radiation exposure.
• Optimal parameter selection ensures that medical radiology
procedures not only meet diagnostic requirements but also adhere to
the fundamental principles of radiation protection and safety.
Balancing Image Quality and Patient Dose
• At the core of this optimization lies the intricate balance between image
quality and patient dose. Striking the right equilibrium is essential to yield
diagnostically valuable images while mitigating the potential risks
associated with unnecessary radiation exposure.
• Achieving this balance requires a nuanced understanding of the diverse
clinical scenarios encountered in medical radiology, demanding tailored
approaches for various patient populations and imaging protocols.
Overview of Impact on Diagnostic
Accuracy
• The optimization of tube voltage, current, and filtration directly
influences diagnostic accuracy, playing a pivotal role in the clarity,
contrast, and precision of medical images.
• Through careful parameter selection, radiographers and medical
physicists can enhance the visibility of anatomical structures,
facilitating accurate diagnoses and informed clinical decision-making.
Emphasis on Radiation Safety
• The overarching goal of optimization extends beyond diagnostic efficacy to
encompass radiation safety. By meticulously fine-tuning imaging
parameters, healthcare professionals contribute to minimizing unnecessary
radiation exposure, safeguarding both patients and healthcare workers.
• As custodians of radiation safety, the medical radiology community must
continuously strive for advancements in optimization practices, ensuring the
highest standard of care for patients while adhering to established safety
guidelines.
Collaborative Approach
• Recognizing the complexity of optimization, fostering a collaborative
approach among radiologists, technologists, and medical physicists is
essential.
• Through interdisciplinary collaboration, the medical radiology community
can harness collective expertise, share insights, and collectively advance the
field toward safer, more effective imaging practices.
Objectives
Objective 1: Understanding Tube Voltage (kVp)
• Tube voltage influences X-ray penetration.
• Examples: Higher kVp for thicker body parts, lower kVp for pediatric
imaging.
• The impact on image contrast with varying kVp settings
These images were generated at 60, 75, 120 kV, and required a relatively high
radiation intensity of 141 mAs.
Objective 2: Tube Voltage and Patient Dose
• Relationship between tube voltage and patient dose.
• Example: Higher kVp reduces patient dose due to increased X-ray
penetration.
• The importance of choosing optimal kVp for diagnostic needs.
Objective 3: Analyzing Tube Current (mA)
• Tube current and its role in controlling the amount of radiation
emitted.
• Examples: Increased mA for larger patients, decreased mA for
pediatric or sensitive areas.
• Showcase how tube current variations affect image brightness.
Tube Voltage and Image Quality
• Tube voltage, also known as kilovoltage (kV), plays a pivotal role in determining the energy level
of X-ray photons produced within the X-ray tube.
• Higher tube voltages result in X-ray photons with greater energy, enabling them to penetrate deeper
into the patient's body.
• This increased penetration is crucial for imaging thicker or denser anatomical structures, as it
ensures sufficient X-ray transmission to produce diagnostically useful images.
• Conversely, lower tube voltages produce X-ray photons with lower energy, which may be suitable
for imaging superficial structures but may result in inadequate penetration for thicker tissues.
Dark-field (top) and attenuation
radiograph (bottom) images
acquired with different tube
voltages. A distinct decrease in
dark-field signal strength can be
seen towards higher tube voltages.
At lower tube voltages, a higher
opacity of bone is present in
attenuation radiographs, impairing
the visualization of the "lung".
Relationship Between Tube Voltage and
Image Contrast
• Tube voltage significantly influences the contrast of the resulting radiographic image.
• Higher tube voltages tend to produce images with lower contrast, as the increased penetration leads
to more uniform attenuation of X-rays throughout the imaged region.
• Lower tube voltages, on the other hand, result in images with higher contrast, as the reduced
penetration causes greater attenuation of X-rays by varying tissue densities within the imaged area.
• The choice of tube voltage must be carefully balanced to achieve the desired diagnostic contrast
while ensuring adequate penetration for accurate anatomical visualization.
Examples of How Varying Tube Voltage
Influences Image Quality
1.Chest Radiography:
1. For chest radiographs, a moderate tube voltage
(e.g., 100-120 kV) is typically employed to ensure
adequate penetration through the thoracic cavity
while maintaining sufficient contrast to visualize
pulmonary structures.
2. Higher tube voltages may be used for larger
patients or to penetrate through bony structures,
such as in trauma imaging, where diagnostic
clarity is paramount.
Bone Radiography
• In bone radiography, where high contrast is
essential for visualizing fine skeletal detail,
lower tube voltages (e.g., 70-80 kV) are often
preferred to maximize the differential
attenuation between bone and surrounding
soft tissues.
• This results in images with enhanced bone
detail and sharp delineation of fractures or
pathologies.
Abdominal Imaging
• Abdominal radiography requires a balance
between penetration and contrast to visualize
both soft tissue structures and underlying organs.
• Depending on patient size and imaging
objectives, tube voltages may vary within a
moderate range (e.g., 80-100 kV) to optimize
diagnostic quality while minimizing radiation
dose to sensitive organs.
Pediatric Imaging
• In pediatric imaging, where minimizing
radiation dose is of utmost importance,
lower tube voltages are often used to reduce
scatter radiation and limit dose exposure to
developing tissues.
• Careful consideration of tube voltage
settings is necessary to achieve diagnostic
images with adequate contrast while
adhering to ALARA (As Low As
Reasonably Achievable) principles.

More Related Content

What's hot (20)

Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptx
 
Radiation and Spectra.pptx
Radiation and Spectra.pptxRadiation and Spectra.pptx
Radiation and Spectra.pptx
 
Computed Tomography Equipments.pptx
Computed Tomography Equipments.pptxComputed Tomography Equipments.pptx
Computed Tomography Equipments.pptx
 
Physical Principle of MRI.pptx
Physical Principle of MRI.pptxPhysical Principle of MRI.pptx
Physical Principle of MRI.pptx
 
Oral Cholecystography .pptx
Oral Cholecystography .pptxOral Cholecystography .pptx
Oral Cholecystography .pptx
 
X-Ray Properties.pptx
X-Ray Properties.pptxX-Ray Properties.pptx
X-Ray Properties.pptx
 
Clinical Applications and Procedure of MRI.pptx
Clinical Applications and Procedure of MRI.pptxClinical Applications and Procedure of MRI.pptx
Clinical Applications and Procedure of MRI.pptx
 
History of Ultrasound Imaging.pptx
History of Ultrasound Imaging.pptxHistory of Ultrasound Imaging.pptx
History of Ultrasound Imaging.pptx
 
Radiology Contrast Media.pptx
Radiology Contrast Media.pptxRadiology Contrast Media.pptx
Radiology Contrast Media.pptx
 
RGU.pptx
RGU.pptxRGU.pptx
RGU.pptx
 
MCU.pptx
MCU.pptxMCU.pptx
MCU.pptx
 
Emergency in Radiology Department.pptx
Emergency in Radiology Department.pptxEmergency in Radiology Department.pptx
Emergency in Radiology Department.pptx
 
Principle of Ultrasound.pptx
Principle of Ultrasound.pptxPrinciple of Ultrasound.pptx
Principle of Ultrasound.pptx
 
Tomography.pptx
Tomography.pptxTomography.pptx
Tomography.pptx
 
Nuclear medicine
Nuclear medicineNuclear medicine
Nuclear medicine
 
Guidelines for X-Ray Equipments.pptx
Guidelines for X-Ray Equipments.pptxGuidelines for X-Ray Equipments.pptx
Guidelines for X-Ray Equipments.pptx
 
PTC.pptx
PTC.pptxPTC.pptx
PTC.pptx
 
OSLD Prsentation
OSLD PrsentationOSLD Prsentation
OSLD Prsentation
 
physical interaction of x ray with matter
physical interaction of x ray with matter physical interaction of x ray with matter
physical interaction of x ray with matter
 
Radiographic Exposure.pptx
Radiographic Exposure.pptxRadiographic Exposure.pptx
Radiographic Exposure.pptx
 

Similar to Effects of variation of tube voltage current, filtration..pptx

J0362069073
J0362069073J0362069073
J0362069073theijes
 
Noise and artifacts in CT
Noise and artifacts in CTNoise and artifacts in CT
Noise and artifacts in CTAngelineSP
 
CT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protectionCT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protectionsanyengere
 
Principle of Ultrasound Imaging
Principle of Ultrasound ImagingPrinciple of Ultrasound Imaging
Principle of Ultrasound Imagingothman alameen
 
FINAL CTCA CAMRT presentation - Yili
FINAL CTCA CAMRT presentation - YiliFINAL CTCA CAMRT presentation - Yili
FINAL CTCA CAMRT presentation - YiliYili Wang
 
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdfNASERALHAQ
 
safe imaging.pptx
safe imaging.pptxsafe imaging.pptx
safe imaging.pptxAbinVl1
 
Radiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronicRadiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronicneestom1998
 
Dual energy CT in radiotherapy: Current applications and future outlook
Dual energy CT in radiotherapy: Current applications and future outlookDual energy CT in radiotherapy: Current applications and future outlook
Dual energy CT in radiotherapy: Current applications and future outlookWookjin Choi
 
Radiotherapy With Protons
Radiotherapy  With  ProtonsRadiotherapy  With  Protons
Radiotherapy With Protonsfondas vakalis
 
Radiotherapy With Protons
Radiotherapy  With  ProtonsRadiotherapy  With  Protons
Radiotherapy With Protonsfondas vakalis
 
Radiation dose to lens in CT
Radiation dose to lens in CTRadiation dose to lens in CT
Radiation dose to lens in CTKalpana Parajuli
 

Similar to Effects of variation of tube voltage current, filtration..pptx (20)

J0362069073
J0362069073J0362069073
J0362069073
 
Oncology imaging
Oncology imagingOncology imaging
Oncology imaging
 
CT dose reduction
CT dose reductionCT dose reduction
CT dose reduction
 
Noise and artifacts in CT
Noise and artifacts in CTNoise and artifacts in CT
Noise and artifacts in CT
 
Session 1 conventional x ray
Session 1  conventional x raySession 1  conventional x ray
Session 1 conventional x ray
 
Cone beam CT: concepts and applications.pptx
Cone beam CT: concepts and applications.pptxCone beam CT: concepts and applications.pptx
Cone beam CT: concepts and applications.pptx
 
Cone beam CT: Concepts and applicationss.pdf
Cone beam CT: Concepts and applicationss.pdfCone beam CT: Concepts and applicationss.pdf
Cone beam CT: Concepts and applicationss.pdf
 
CT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protectionCT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protection
 
Principle of Ultrasound Imaging
Principle of Ultrasound ImagingPrinciple of Ultrasound Imaging
Principle of Ultrasound Imaging
 
FINAL CTCA CAMRT presentation - Yili
FINAL CTCA CAMRT presentation - YiliFINAL CTCA CAMRT presentation - Yili
FINAL CTCA CAMRT presentation - Yili
 
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
7-RADIATION PROTECTION IN DENTISTRY===7=2023.pdf
 
safe imaging.pptx
safe imaging.pptxsafe imaging.pptx
safe imaging.pptx
 
Radiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronicRadiotherapy ppt. Types of radiation and chronic
Radiotherapy ppt. Types of radiation and chronic
 
Dual energy CT in radiotherapy: Current applications and future outlook
Dual energy CT in radiotherapy: Current applications and future outlookDual energy CT in radiotherapy: Current applications and future outlook
Dual energy CT in radiotherapy: Current applications and future outlook
 
Radiotherapy With Protons
Radiotherapy  With  ProtonsRadiotherapy  With  Protons
Radiotherapy With Protons
 
Radiotherapy With Protons
Radiotherapy  With  ProtonsRadiotherapy  With  Protons
Radiotherapy With Protons
 
CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION
 
Radiation dose to lens in CT
Radiation dose to lens in CTRadiation dose to lens in CT
Radiation dose to lens in CT
 
1.radiation of h&n tumors
1.radiation of h&n tumors1.radiation of h&n tumors
1.radiation of h&n tumors
 
2012_TMI
2012_TMI2012_TMI
2012_TMI
 

More from Dr. Dheeraj Kumar

Culture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxCulture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxDr. Dheeraj Kumar
 
X-ray Production A Journey Through History and the X-ray Tube.pptx
X-ray Production A Journey Through History and the X-ray Tube.pptxX-ray Production A Journey Through History and the X-ray Tube.pptx
X-ray Production A Journey Through History and the X-ray Tube.pptxDr. Dheeraj Kumar
 
Population of India and Family welfare programme.pptx
Population of India and Family welfare programme.pptxPopulation of India and Family welfare programme.pptx
Population of India and Family welfare programme.pptxDr. Dheeraj Kumar
 
FILM PROCESSING AND DEVELOPMENT.pptx
FILM PROCESSING AND DEVELOPMENT.pptxFILM PROCESSING AND DEVELOPMENT.pptx
FILM PROCESSING AND DEVELOPMENT.pptxDr. Dheeraj Kumar
 
Physical Principle of Ultrasound.pptx
Physical Principle of Ultrasound.pptxPhysical Principle of Ultrasound.pptx
Physical Principle of Ultrasound.pptxDr. Dheeraj Kumar
 
CLINICAL APPLICATIONS OF CT SCAN.pptx
CLINICAL APPLICATIONS OF CT SCAN.pptxCLINICAL APPLICATIONS OF CT SCAN.pptx
CLINICAL APPLICATIONS OF CT SCAN.pptxDr. Dheeraj Kumar
 
Radiographic Latent Image .pptx
Radiographic Latent Image .pptxRadiographic Latent Image .pptx
Radiographic Latent Image .pptxDr. Dheeraj Kumar
 

More from Dr. Dheeraj Kumar (10)

Culture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxCulture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptx
 
X-ray Production A Journey Through History and the X-ray Tube.pptx
X-ray Production A Journey Through History and the X-ray Tube.pptxX-ray Production A Journey Through History and the X-ray Tube.pptx
X-ray Production A Journey Through History and the X-ray Tube.pptx
 
Population of India and Family welfare programme.pptx
Population of India and Family welfare programme.pptxPopulation of India and Family welfare programme.pptx
Population of India and Family welfare programme.pptx
 
FILM PROCESSING AND DEVELOPMENT.pptx
FILM PROCESSING AND DEVELOPMENT.pptxFILM PROCESSING AND DEVELOPMENT.pptx
FILM PROCESSING AND DEVELOPMENT.pptx
 
Physical Principle of Ultrasound.pptx
Physical Principle of Ultrasound.pptxPhysical Principle of Ultrasound.pptx
Physical Principle of Ultrasound.pptx
 
Angiography.pptx
Angiography.pptxAngiography.pptx
Angiography.pptx
 
CLINICAL APPLICATIONS OF CT SCAN.pptx
CLINICAL APPLICATIONS OF CT SCAN.pptxCLINICAL APPLICATIONS OF CT SCAN.pptx
CLINICAL APPLICATIONS OF CT SCAN.pptx
 
Catheter and Guidewire.pptx
Catheter and Guidewire.pptxCatheter and Guidewire.pptx
Catheter and Guidewire.pptx
 
Radiographic Latent Image .pptx
Radiographic Latent Image .pptxRadiographic Latent Image .pptx
Radiographic Latent Image .pptx
 
Radiographic Grid.pptx
Radiographic Grid.pptxRadiographic Grid.pptx
Radiographic Grid.pptx
 

Recently uploaded

Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...chandars293
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsGfnyt
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 

Recently uploaded (20)

Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 

Effects of variation of tube voltage current, filtration..pptx

  • 1. Effects of Variation of tube Voltage Current, Filtration Presenter: Dheeraj Kumar MRIT, Ph.D. (Radiology and Imaging) Assistant Professor Medical Radiology and Imaging Technology School of Health Sciences, CSJM University, Kanpur
  • 2. Introduction • In the field of medical radiography, optimizing critical parameters including tube voltage, current, and filtration is a crucial undertaking. • This introduction seeks to underscore the paramount importance of achieving a delicate equilibrium between these factors, emphasizing their collective impact on diagnostic accuracy and radiation safety.
  • 3. Significance of Optimization • The essence of optimizing tube voltage, current, and filtration lies in fine-tuning the delicate interplay between achieving superior image quality and minimizing patient radiation exposure. • Optimal parameter selection ensures that medical radiology procedures not only meet diagnostic requirements but also adhere to the fundamental principles of radiation protection and safety.
  • 4. Balancing Image Quality and Patient Dose • At the core of this optimization lies the intricate balance between image quality and patient dose. Striking the right equilibrium is essential to yield diagnostically valuable images while mitigating the potential risks associated with unnecessary radiation exposure. • Achieving this balance requires a nuanced understanding of the diverse clinical scenarios encountered in medical radiology, demanding tailored approaches for various patient populations and imaging protocols.
  • 5. Overview of Impact on Diagnostic Accuracy • The optimization of tube voltage, current, and filtration directly influences diagnostic accuracy, playing a pivotal role in the clarity, contrast, and precision of medical images. • Through careful parameter selection, radiographers and medical physicists can enhance the visibility of anatomical structures, facilitating accurate diagnoses and informed clinical decision-making.
  • 6. Emphasis on Radiation Safety • The overarching goal of optimization extends beyond diagnostic efficacy to encompass radiation safety. By meticulously fine-tuning imaging parameters, healthcare professionals contribute to minimizing unnecessary radiation exposure, safeguarding both patients and healthcare workers. • As custodians of radiation safety, the medical radiology community must continuously strive for advancements in optimization practices, ensuring the highest standard of care for patients while adhering to established safety guidelines.
  • 7. Collaborative Approach • Recognizing the complexity of optimization, fostering a collaborative approach among radiologists, technologists, and medical physicists is essential. • Through interdisciplinary collaboration, the medical radiology community can harness collective expertise, share insights, and collectively advance the field toward safer, more effective imaging practices.
  • 8. Objectives Objective 1: Understanding Tube Voltage (kVp) • Tube voltage influences X-ray penetration. • Examples: Higher kVp for thicker body parts, lower kVp for pediatric imaging. • The impact on image contrast with varying kVp settings
  • 9. These images were generated at 60, 75, 120 kV, and required a relatively high radiation intensity of 141 mAs.
  • 10.
  • 11.
  • 12. Objective 2: Tube Voltage and Patient Dose • Relationship between tube voltage and patient dose. • Example: Higher kVp reduces patient dose due to increased X-ray penetration. • The importance of choosing optimal kVp for diagnostic needs.
  • 13.
  • 14.
  • 15.
  • 16. Objective 3: Analyzing Tube Current (mA) • Tube current and its role in controlling the amount of radiation emitted. • Examples: Increased mA for larger patients, decreased mA for pediatric or sensitive areas. • Showcase how tube current variations affect image brightness.
  • 17.
  • 18.
  • 19.
  • 20. Tube Voltage and Image Quality • Tube voltage, also known as kilovoltage (kV), plays a pivotal role in determining the energy level of X-ray photons produced within the X-ray tube. • Higher tube voltages result in X-ray photons with greater energy, enabling them to penetrate deeper into the patient's body. • This increased penetration is crucial for imaging thicker or denser anatomical structures, as it ensures sufficient X-ray transmission to produce diagnostically useful images. • Conversely, lower tube voltages produce X-ray photons with lower energy, which may be suitable for imaging superficial structures but may result in inadequate penetration for thicker tissues.
  • 21. Dark-field (top) and attenuation radiograph (bottom) images acquired with different tube voltages. A distinct decrease in dark-field signal strength can be seen towards higher tube voltages. At lower tube voltages, a higher opacity of bone is present in attenuation radiographs, impairing the visualization of the "lung".
  • 22. Relationship Between Tube Voltage and Image Contrast • Tube voltage significantly influences the contrast of the resulting radiographic image. • Higher tube voltages tend to produce images with lower contrast, as the increased penetration leads to more uniform attenuation of X-rays throughout the imaged region. • Lower tube voltages, on the other hand, result in images with higher contrast, as the reduced penetration causes greater attenuation of X-rays by varying tissue densities within the imaged area. • The choice of tube voltage must be carefully balanced to achieve the desired diagnostic contrast while ensuring adequate penetration for accurate anatomical visualization.
  • 23. Examples of How Varying Tube Voltage Influences Image Quality 1.Chest Radiography: 1. For chest radiographs, a moderate tube voltage (e.g., 100-120 kV) is typically employed to ensure adequate penetration through the thoracic cavity while maintaining sufficient contrast to visualize pulmonary structures. 2. Higher tube voltages may be used for larger patients or to penetrate through bony structures, such as in trauma imaging, where diagnostic clarity is paramount.
  • 24. Bone Radiography • In bone radiography, where high contrast is essential for visualizing fine skeletal detail, lower tube voltages (e.g., 70-80 kV) are often preferred to maximize the differential attenuation between bone and surrounding soft tissues. • This results in images with enhanced bone detail and sharp delineation of fractures or pathologies.
  • 25. Abdominal Imaging • Abdominal radiography requires a balance between penetration and contrast to visualize both soft tissue structures and underlying organs. • Depending on patient size and imaging objectives, tube voltages may vary within a moderate range (e.g., 80-100 kV) to optimize diagnostic quality while minimizing radiation dose to sensitive organs.
  • 26. Pediatric Imaging • In pediatric imaging, where minimizing radiation dose is of utmost importance, lower tube voltages are often used to reduce scatter radiation and limit dose exposure to developing tissues. • Careful consideration of tube voltage settings is necessary to achieve diagnostic images with adequate contrast while adhering to ALARA (As Low As Reasonably Achievable) principles.