SlideShare a Scribd company logo
1 of 14
Unit – 4
Artifacts and Noise in CT
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 1
Beam Hardening
• CT uses a poly energetic x-ray spectrum, with energies ranging from
about 25 to 120 keV
• Furthermore, x-ray attenuation coefficients are energy dependent:
• After passing through a given thickness of patient, lower-energy x-
rays are attenuated to a greater extent than higher-energy x-rays are.
• Therefore, as the x-ray beam propagates through a thickness of tissue
and bone, the shape of the spectrum becomes skewed toward the
higher energies (Fig. 13-38).
• Consequently, the average energy of the x-ray beam becomes greater
("harder") as it passes through tissue. Because the attenuation of
bone is greater than that of soft tissue, bone causes more beam
hardening than an equivalent thickness of soft tissue
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 2
• The beam-hardening phenomenon induces artifacts in CT because rays
from some projection angles are hardened to a differing extent than rays
from other angles, and this confuses the reconstruction algorithm.
• Most CT scanners include a simple beam-hardening correction algorithm
that corrects for beam hardening based on the relative attenuation of each
ray, and this helps to some extent.
• In a two pass algorithm, the image is reconstructed normally in the first
pass, the path length that each ray transits through bone and soft tissue is
determined from the first-pass image, and then each ray is compensated
for beam hardening based on the known bone and soft tissue thicknesses it
has traversed.
• A second-pass image reconstruction is performed using the corrected ray
values.
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 3
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 4
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 5
Motion artifacts
• Motion artifacts occur when the patient moves during the acquisition.
• Small motions cause image blurring, and larger physical displacements
during CT image acquisition produce artifacts that appear as double images
or image ghosting.
• If motion is suspected in a CT image, the adjacent CT scans in the study
may be evaluated to distinguish fact from artifact.
• In some cases, the patient needs to be rescanned.
• An example of a motion artifact is shown in Fig. 13-40B.
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 6
Partial Volume averaging
• The CT number in each pixel is proportional to the average µ in the
corresponding voxel.
• For voxels containing all one tissue type (e.g., all bone, all liver), µis
representative of that tissue.
• Some voxels in the image, however, contain a mixture of different tissue
types. When this occurs, for example with bone and soft tissue, the µ is not
representative of either tissue but instead is a weighted average of the two
different µ values.
• Partial volume averaging is most pronounced for softly rounded structures
that are almost parallel to the CT slice.
• The most evident example is near the top of the head, where the cranium
shares a substantial number of voxels with brain tissue, causing details of
the brain parenchyma to be lost because the large µ of bone dominates.
• This situation is easily recognizable and therefore seldom leads to
misdiagnosis. Partial volume artifacts can lead to misdiagnosis when the
presence of adjacent anatomic structures is not suspected (Fig. 13-41).
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 7
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 8
• There are several approaches to reducing partial volume artifacts, and
the obvious a priori approach is to use thinner CT slices.
• When a suspected partial volume artifact occurs with a helical study
and the raw scan data is still available, it is sometimes necessary to
use the raw data to reconstruct additional CT images at different
positions.
• Many institutions make use of interleaved reconstructions (e.g., 5-
mm slices every 2.5 mm) as a matter of routine protocol.
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 9
Noise in CT
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 10
• Compared with x-ray radiography, CT has significantly worse spatial
resolution and significantly better contrast resolution.
• Whereas the limiting spatial frequency for screen-film radiography is
about 7 line pairs (lp) per millimeter and for digital radiography it is 5
Ip/mm, the limiting spatial frequency for CT is approximately 1
Ip/mm.
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 11
• CT has the best contrast resolution of any clinical x-ray modaliry.
• Contrast resolution refers to the abilty of an imaging procedure to reliably depict
very subtle differences in contrast.
• It is generally accepted that the contrast resolution of screen-film radiography is
approximately 5%, whereas CT demonstrates contrast resolution of about 0.5%.
• A classic clinical example in which the contrast resolution capability of CT excels is
distinguishing subtle soft tissue tumors:
• The difference in CT number between the tumor and the surrounding tissue may
be small (e.g., 20 CT numbers), but because the noise in the CT numbers is
smaller (e.g., 3 CT numbers), the tumor is visible on the display to the trained
human observer.
• As is apparent from this example, contrast resolution is fundamentally tied to the
SNR.
• The SNR is also very much related to the number of x-ray quanta used per pixel in
the image. If one attempts to reduce the pixel size (and thereby increase spatial
resolution) and the dose levels are kept the same, the number of x-rays per pixel
is reduced. For example, for the same FOV and dose, changing to a 1,024 X 1,024
CT image from a 512 X 512 image would result in fewer x-ray photons passing
through each voxel, and therefore the SNR per pixel would drop.
• It should be clear from this example that there is a compromise between spatial
resolution and contrast resolution.
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 12
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 13
Thank You!
Dr. S. P. Angeline Kirubha, Biomedical Engineering Department,
SRMIST
12/5/2021 14

More Related Content

What's hot

Computed tomography by jay&jay
Computed tomography by jay&jayComputed tomography by jay&jay
Computed tomography by jay&jayPatel Jay
 
CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstructionGunjan Patel
 
Helical and Multislice CT
Helical and Multislice CTHelical and Multislice CT
Helical and Multislice CTManojzz Bhatta
 
CT Image reconstruction
CT Image reconstructionCT Image reconstruction
CT Image reconstructionSantosh Ojha
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)Upakar Paudel
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaAvinesh Shrestha
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptxdypradio
 
Gradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationGradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationJayanti Gyawali
 
6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptxRohit Bansal
 
Fat suppression techniques in MRI
Fat suppression techniques in MRIFat suppression techniques in MRI
Fat suppression techniques in MRISakshamKumar30
 
Portable ct mobile ct
Portable ct mobile ctPortable ct mobile ct
Portable ct mobile ctanilayyakutty
 
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONSMDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONShazem youssef
 

What's hot (20)

Ct
CtCt
Ct
 
Computed tomography by jay&jay
Computed tomography by jay&jayComputed tomography by jay&jay
Computed tomography by jay&jay
 
CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstruction
 
Helical and Multislice CT
Helical and Multislice CTHelical and Multislice CT
Helical and Multislice CT
 
MDCT (2)
MDCT (2)MDCT (2)
MDCT (2)
 
CT Image reconstruction
CT Image reconstructionCT Image reconstruction
CT Image reconstruction
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
 
Gradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationGradient echo pulse sequence and its application
Gradient echo pulse sequence and its application
 
6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx
 
CT artifact
CT artifact CT artifact
CT artifact
 
Computed tomography
Computed tomography Computed tomography
Computed tomography
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CT
 
Fat suppression techniques in MRI
Fat suppression techniques in MRIFat suppression techniques in MRI
Fat suppression techniques in MRI
 
Ct ppt 2
Ct ppt 2Ct ppt 2
Ct ppt 2
 
Portable ct mobile ct
Portable ct mobile ctPortable ct mobile ct
Portable ct mobile ct
 
Physics of ct mri
Physics of ct mriPhysics of ct mri
Physics of ct mri
 
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONSMDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
 

Similar to Noise and artifacts in CT

DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSDIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSMichaelAkortsu1
 
Comparison of ct and cbct
Comparison of ct and cbct Comparison of ct and cbct
Comparison of ct and cbct Amritha James
 
lecture continue.pdf
lecture continue.pdflecture continue.pdf
lecture continue.pdf8qq9rzc2j4
 
Specialised Techniques in Oral Radiology
Specialised Techniques in Oral RadiologySpecialised Techniques in Oral Radiology
Specialised Techniques in Oral RadiologyIAU Dent
 
European project PROMET - Deep Real-Time Optoacoustic Imaging of Breast Tumors
European project PROMET - Deep Real-Time Optoacoustic Imaging of Breast TumorsEuropean project PROMET - Deep Real-Time Optoacoustic Imaging of Breast Tumors
European project PROMET - Deep Real-Time Optoacoustic Imaging of Breast TumorsSébastien Sénégas
 
Effects of variation of tube voltage current, filtration..pptx
Effects of variation of tube voltage current, filtration..pptxEffects of variation of tube voltage current, filtration..pptx
Effects of variation of tube voltage current, filtration..pptxDr. Dheeraj Kumar
 
CBCT in Periodontology
CBCT in PeriodontologyCBCT in Periodontology
CBCT in PeriodontologyRinisha Sinha
 
IMAGE QUALITY AND IMAGE ARTIFACTS .pptx
IMAGE QUALITY AND IMAGE ARTIFACTS  .pptxIMAGE QUALITY AND IMAGE ARTIFACTS  .pptx
IMAGE QUALITY AND IMAGE ARTIFACTS .pptxlavanyarangaswamy31
 
CT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protectionCT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protectionsanyengere
 
Cbct in endodontics ppt
Cbct in endodontics pptCbct in endodontics ppt
Cbct in endodontics pptDr kausar banu
 
CBCT anatomical structures
CBCT anatomical structuresCBCT anatomical structures
CBCT anatomical structuressarahelgammal2
 
Diagnostic imaging in head and neck pathology
Diagnostic imaging in head and neck pathologyDiagnostic imaging in head and neck pathology
Diagnostic imaging in head and neck pathologyHayat Youssef
 
COMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCANCOMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCANShounak Nandi
 
Radiation dose to lens in CT
Radiation dose to lens in CTRadiation dose to lens in CT
Radiation dose to lens in CTKalpana Parajuli
 
Cone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptxCone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptxDrAnand22
 

Similar to Noise and artifacts in CT (20)

DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSDIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
 
Comparison of ct and cbct
Comparison of ct and cbct Comparison of ct and cbct
Comparison of ct and cbct
 
Cone beam CT: concepts and applications.pptx
Cone beam CT: concepts and applications.pptxCone beam CT: concepts and applications.pptx
Cone beam CT: concepts and applications.pptx
 
lecture continue.pdf
lecture continue.pdflecture continue.pdf
lecture continue.pdf
 
Specialised Techniques in Oral Radiology
Specialised Techniques in Oral RadiologySpecialised Techniques in Oral Radiology
Specialised Techniques in Oral Radiology
 
FUSION IMAGING
FUSION IMAGINGFUSION IMAGING
FUSION IMAGING
 
European project PROMET - Deep Real-Time Optoacoustic Imaging of Breast Tumors
European project PROMET - Deep Real-Time Optoacoustic Imaging of Breast TumorsEuropean project PROMET - Deep Real-Time Optoacoustic Imaging of Breast Tumors
European project PROMET - Deep Real-Time Optoacoustic Imaging of Breast Tumors
 
Cone beam CT: Concepts and applicationss.pdf
Cone beam CT: Concepts and applicationss.pdfCone beam CT: Concepts and applicationss.pdf
Cone beam CT: Concepts and applicationss.pdf
 
Oncology imaging
Oncology imagingOncology imaging
Oncology imaging
 
Effects of variation of tube voltage current, filtration..pptx
Effects of variation of tube voltage current, filtration..pptxEffects of variation of tube voltage current, filtration..pptx
Effects of variation of tube voltage current, filtration..pptx
 
CBCT in Periodontology
CBCT in PeriodontologyCBCT in Periodontology
CBCT in Periodontology
 
IMAGE QUALITY AND IMAGE ARTIFACTS .pptx
IMAGE QUALITY AND IMAGE ARTIFACTS  .pptxIMAGE QUALITY AND IMAGE ARTIFACTS  .pptx
IMAGE QUALITY AND IMAGE ARTIFACTS .pptx
 
CT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protectionCT Dose Issues.pptx on the factors to be considered on radiation protection
CT Dose Issues.pptx on the factors to be considered on radiation protection
 
Cbct in endodontics ppt
Cbct in endodontics pptCbct in endodontics ppt
Cbct in endodontics ppt
 
CBCT anatomical structures
CBCT anatomical structuresCBCT anatomical structures
CBCT anatomical structures
 
OCT
OCTOCT
OCT
 
Diagnostic imaging in head and neck pathology
Diagnostic imaging in head and neck pathologyDiagnostic imaging in head and neck pathology
Diagnostic imaging in head and neck pathology
 
COMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCANCOMPUTED TOMOGRAPHY SCAN
COMPUTED TOMOGRAPHY SCAN
 
Radiation dose to lens in CT
Radiation dose to lens in CTRadiation dose to lens in CT
Radiation dose to lens in CT
 
Cone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptxCone Beam Computed Tomography.pptx
Cone Beam Computed Tomography.pptx
 

Recently uploaded

Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591adityaroy0215
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...gragteena
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...indiancallgirl4rent
 
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7Miss joya
 
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130  Available With RoomVIP Kolkata Call Girl New Town 👉 8250192130  Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Roomdivyansh0kumar0
 
Call Girls In ludhiana For Fun 9053900678 By ludhiana Call Girls For Pick...
Call Girls In  ludhiana  For Fun 9053900678 By  ludhiana  Call Girls For Pick...Call Girls In  ludhiana  For Fun 9053900678 By  ludhiana  Call Girls For Pick...
Call Girls In ludhiana For Fun 9053900678 By ludhiana Call Girls For Pick...Russian Call Girls in Ludhiana
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...Gfnyt.com
 
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...Vip call girls In Chandigarh
 
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Call Girls Noida
 
Nepali Escort Girl * 9999965857 Naughty Call Girls Service in Faridabad
Nepali Escort Girl * 9999965857 Naughty Call Girls Service in FaridabadNepali Escort Girl * 9999965857 Naughty Call Girls Service in Faridabad
Nepali Escort Girl * 9999965857 Naughty Call Girls Service in Faridabadgragteena
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...Gfnyt.com
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknowgragteena
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Memriyagarg453
 
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service DehradunDehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service DehradunNiamh verma
 
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591adityaroy0215
 

Recently uploaded (20)

Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetChandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Chandigarh Call Girls 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
Call Girls Service Charbagh { Lucknow Call Girls Service 9548273370 } Book me...
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
 
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
 
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130  Available With RoomVIP Kolkata Call Girl New Town 👉 8250192130  Available With Room
VIP Kolkata Call Girl New Town 👉 8250192130 Available With Room
 
Call Girls In ludhiana For Fun 9053900678 By ludhiana Call Girls For Pick...
Call Girls In  ludhiana  For Fun 9053900678 By  ludhiana  Call Girls For Pick...Call Girls In  ludhiana  For Fun 9053900678 By  ludhiana  Call Girls For Pick...
Call Girls In ludhiana For Fun 9053900678 By ludhiana Call Girls For Pick...
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
 
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
 
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
 
Nepali Escort Girl * 9999965857 Naughty Call Girls Service in Faridabad
Nepali Escort Girl * 9999965857 Naughty Call Girls Service in FaridabadNepali Escort Girl * 9999965857 Naughty Call Girls Service in Faridabad
Nepali Escort Girl * 9999965857 Naughty Call Girls Service in Faridabad
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
 
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service DehradunDehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
 
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
 

Noise and artifacts in CT

  • 1. Unit – 4 Artifacts and Noise in CT Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 1
  • 2. Beam Hardening • CT uses a poly energetic x-ray spectrum, with energies ranging from about 25 to 120 keV • Furthermore, x-ray attenuation coefficients are energy dependent: • After passing through a given thickness of patient, lower-energy x- rays are attenuated to a greater extent than higher-energy x-rays are. • Therefore, as the x-ray beam propagates through a thickness of tissue and bone, the shape of the spectrum becomes skewed toward the higher energies (Fig. 13-38). • Consequently, the average energy of the x-ray beam becomes greater ("harder") as it passes through tissue. Because the attenuation of bone is greater than that of soft tissue, bone causes more beam hardening than an equivalent thickness of soft tissue Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 2
  • 3. • The beam-hardening phenomenon induces artifacts in CT because rays from some projection angles are hardened to a differing extent than rays from other angles, and this confuses the reconstruction algorithm. • Most CT scanners include a simple beam-hardening correction algorithm that corrects for beam hardening based on the relative attenuation of each ray, and this helps to some extent. • In a two pass algorithm, the image is reconstructed normally in the first pass, the path length that each ray transits through bone and soft tissue is determined from the first-pass image, and then each ray is compensated for beam hardening based on the known bone and soft tissue thicknesses it has traversed. • A second-pass image reconstruction is performed using the corrected ray values. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 3
  • 4. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 4
  • 5. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 5
  • 6. Motion artifacts • Motion artifacts occur when the patient moves during the acquisition. • Small motions cause image blurring, and larger physical displacements during CT image acquisition produce artifacts that appear as double images or image ghosting. • If motion is suspected in a CT image, the adjacent CT scans in the study may be evaluated to distinguish fact from artifact. • In some cases, the patient needs to be rescanned. • An example of a motion artifact is shown in Fig. 13-40B. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 6
  • 7. Partial Volume averaging • The CT number in each pixel is proportional to the average µ in the corresponding voxel. • For voxels containing all one tissue type (e.g., all bone, all liver), µis representative of that tissue. • Some voxels in the image, however, contain a mixture of different tissue types. When this occurs, for example with bone and soft tissue, the µ is not representative of either tissue but instead is a weighted average of the two different µ values. • Partial volume averaging is most pronounced for softly rounded structures that are almost parallel to the CT slice. • The most evident example is near the top of the head, where the cranium shares a substantial number of voxels with brain tissue, causing details of the brain parenchyma to be lost because the large µ of bone dominates. • This situation is easily recognizable and therefore seldom leads to misdiagnosis. Partial volume artifacts can lead to misdiagnosis when the presence of adjacent anatomic structures is not suspected (Fig. 13-41). Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 7
  • 8. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 8
  • 9. • There are several approaches to reducing partial volume artifacts, and the obvious a priori approach is to use thinner CT slices. • When a suspected partial volume artifact occurs with a helical study and the raw scan data is still available, it is sometimes necessary to use the raw data to reconstruct additional CT images at different positions. • Many institutions make use of interleaved reconstructions (e.g., 5- mm slices every 2.5 mm) as a matter of routine protocol. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 9
  • 10. Noise in CT Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 10
  • 11. • Compared with x-ray radiography, CT has significantly worse spatial resolution and significantly better contrast resolution. • Whereas the limiting spatial frequency for screen-film radiography is about 7 line pairs (lp) per millimeter and for digital radiography it is 5 Ip/mm, the limiting spatial frequency for CT is approximately 1 Ip/mm. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 11
  • 12. • CT has the best contrast resolution of any clinical x-ray modaliry. • Contrast resolution refers to the abilty of an imaging procedure to reliably depict very subtle differences in contrast. • It is generally accepted that the contrast resolution of screen-film radiography is approximately 5%, whereas CT demonstrates contrast resolution of about 0.5%. • A classic clinical example in which the contrast resolution capability of CT excels is distinguishing subtle soft tissue tumors: • The difference in CT number between the tumor and the surrounding tissue may be small (e.g., 20 CT numbers), but because the noise in the CT numbers is smaller (e.g., 3 CT numbers), the tumor is visible on the display to the trained human observer. • As is apparent from this example, contrast resolution is fundamentally tied to the SNR. • The SNR is also very much related to the number of x-ray quanta used per pixel in the image. If one attempts to reduce the pixel size (and thereby increase spatial resolution) and the dose levels are kept the same, the number of x-rays per pixel is reduced. For example, for the same FOV and dose, changing to a 1,024 X 1,024 CT image from a 512 X 512 image would result in fewer x-ray photons passing through each voxel, and therefore the SNR per pixel would drop. • It should be clear from this example that there is a compromise between spatial resolution and contrast resolution. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 12
  • 13. Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 13
  • 14. Thank You! Dr. S. P. Angeline Kirubha, Biomedical Engineering Department, SRMIST 12/5/2021 14