Lateral Flow
Achieving 5 x More Sensitivity
and Less Variation in Results than
Covalent Chemistry with Magnetic Particles
by Anteo Technologies
Introduction
Lateral flow immunoassays (LFIA) or strip
tests are conceptually simple immunoassays
well suited to Point of Care (PoC) applications
requiring little or no infrastructure support
(Wong and Tse 2009, Sharma et al 2015).
In its original conception, the test is easy-to-use,
relatively fast and low cost, giving good qualitative
information where a simple yes/no answer is sufficient.
However, this same test system has been evolving to
compete with more traditional laboratory based platforms.
To achieve the requirements of greater assay accuracy
and sensitivity without losing its basic advantages for
PoC conditions, the actual strip test is now at the core of
instruments such as hand held analysers comprising sample
handling, reading and analysis modules. In brief, LFIA is
no longer at the cheap and simple end of the spectrum in
terms of diagnostic technologies (Sharma et al, 2015).
The above trends have resulted in continuous improvements
over various materials and reagents used in LFIA of which
Anteo Technologies’ Lateral Flow Coupling Kit is one new
development. Historically, gold colloids have dominated in
LFIA with visualisation of conjugated gold colloids as they
were captured on pre-defined lines on the test strip.
Passive conjugation of antibodies on gold colloids is
conceptually simple but depending on the antibody,
can lead to difficulties such as poor Ab functionality
and loss of colloidal stability resulting in the formation
of aggregates (Wong and Tse 2009). These difficulties
change with scale, and achieving good uniformity
during scale-up can often be very challenging.
Summary
A new method of conjugating antibodies to particles was compared to covalent conjugation in a lateral flow
assay for human Chorionic Gonadotropin (hCG). Anteo Technologies will be launching a new Lateral Flow
Coupling Kit at AACC 2016 that contains Anteo’s patented Activation Reagent with all necessary buffers
for conjugating antibodies to a range of particles used for lateral flow applications. Here we demonstrate
the performance advantages of using activated magnetic particles, only requiring the addition of antibody
solution and blocking. Keeping all other variables constant, this kit was directly compared to covalently (EDAC
chemistry) conjugated magnetic particles of the same type using visual detection as the readout. In this hCG
study, this coupling kit achieves five times more sensitivity than the covalently conjugated magnetic particle
based assay using half the amount of antibody.
Comparison of Anteo’s Activated Magnetic Particles and Covalently Conjugated Magnetic
Particles in a Lateral Flow Assay for Human Chorionic Gonadotropin (hCG)
Find out more at anteotech.com
Using 200 nm Magnetic Particles
The use of 200 nm magnetic nanoparticles
activated with Anteo’s reagent for lateral flow as
an alternative to colloidal gold addresses a number
of potential problematic situations encountered
in the development of new lateral flow assays.
For example, preparation of labelled conjugates
is simplified using magnetic particles as they are
easy to separate during conjugation, washing
and blocking steps, with the additional benefit
of inclusion of a magnetic label for performing
magnetic immunoassays (MIA) (Shi et al 2015).
Alternatively, when used in conjunction with sufficiently
dark magnetic particles visual detection alone can
produce results rivalling those of other coloured
particle types. The kit contains Anteo’s Activation
Reagent that can be used to activate particle sizes
from 200 nm to 1 um that become immediately reactive
to antibodies once Anteo’s reagent is applied.
Conventionally, conjugation of antibodies to magnetic
particles is commonly done by covalent chemistry using
reagents such as 1-Ethyl-3-(3-dimethylaminopropyl)
carbodiimide (EDC) and N-hydroxysulfosuccinimide
(S-NHS). A reactive ester is formed on the magnetic
particles, which reacts with amino groups on the protein.
Chemically, the reaction is simple but fraught with multiple
variables to consider when trying to achieve optimum
results. Conditions that increase stability of the ester
during synthesis make it less reactive to amino groups
on the protein. Increasing ester reactivity leads to side
reactions. While different approaches can mitigate many
such problems, these methods require pre-activation of
particles which cannot be stored for any length of time.
The magnetic particles activated by Anteo’s Activation
Reagent are claimed to be stable in storage and highly
reactive to proteins with other advantages such as
ease of use, less antibodies used, faster conjugation
and reduced aggregation amongst its benefits.
Using a hCG lateral flow assay previously developed
using colloidal gold at Diagnostic Consulting Network
(DCN), the Anteo coupling kits were compared to
covalently conjugated magnetic particles using
the same assay materials (membrane, pads, and
antibodies) with the exception of the gold colloids.
Find out more at anteotech.com
Find out more at anteotech.com
Method
The hCG assay has previously been developed
using colloidal gold at Diagnostic Consulting
Network (DCN), Carlsbad, CA for internal use
and demonstration purposes. The same assay
materials (membrane, pads, and antibodies)
were used with Anteo’s technology and
covalently conjugated magnetic particles.
All materials and tests were performed in DCN’s laboratories
using magnetic particles conjugated with anti-hCG
antibody using Anteo’s protocol and DCN’s standard
covalent conjugation procedure. hCG concentrations
from 0 to 250 mIU/ml were prepared in 1%BSA, 0.1%Tw-
20 1xPBS and in negative pooled human urine. Tests
were performed using 100ul of sample and test strips
with the covalently conjugated and Anteo’s conjugated
particles dried onto a conjugate pad. Five replicates
for each concentration were tested and results were
recorded 15 minutes after the addition of the sample
using DCN’s visual grading scale and an ESE Quant
(Qiagen Lake Constance, Germany) reflectance reader.
The amount of antibody conjugated to the different particles
and the amount of particles used per test were optimised
to produce the highest sensitivity. Anteo’s conjugate
was tested with an antibody loading of 25ug of anti-hCG
antibody per mg of particles at 0.05% solids. The covalently
conjugated particle was tested with an antibody loading
of 50ug of anti-hCG per mg particles at 0.05% solids.
Results
The results for the visual grades and the values recorded
by the reflectance reader, including standard deviation
and %CV are shown in Table 1 for each of the two tests.
The limit of detection for the Anteo assay was ~10 mIU/
ml hCG in buffer and ~25 mIU/ml in urine for the visual
and reader based results. The limit of detection for the
covalently conjugated hCG assay was ~50 mIU/mL in
buffer and ~100 mIU/mL in urine. As a reference the limit
of the detection for the optimised colloidal gold assay is
at ~2.5 mIU/mL in buffer and ~25 mIU/mL in urine.
NOTE: A visual grade of 2 or higher at the
test line is considered positive, while the
cut-off for the reader is at 30mV.
Find out more at anteotech.com
Figure 1: Graph of the results from the Anteo and covalently conjugated magnetic particle hCG assays in buffer measured by a Qiagen reader.
Table 1: Visual and Reader results of the Anteo’s technology and covalently conjugated magnetic particle based hCG assays in buffer.
hCG
Concentration
(mIU/mL)
in Buffer
Anteo Mix&Go Magnetic Particle Covalently Conjugated Magnetic Particle
Visual
Grade
Average
TL (mv)
Stdev %CV
Visual
Grade
Average
TL (mv)
Stdev %CV
250 9.0 486.9 16.3 3% 5.4 163.5 25.8 16%
100 7.0 314.0 11.3 4% 4.0 108.0 12.3 11%
50 6.0 174.8 8.2 5% 3.0 62.8 6.0 10%
25 4.0 90.9 7.7 8% 2.6 59.5 12.8 21%
10 2.0 36.8 5.1 14% 2.2 44.7 6.2 14%
5 0.0 0.0 0.0 NA 1.8 34.3 11.4 33%
0 0.0 0.0 0.0 NA 1.0 6.5 12.4 189%
0.0
200.0
100.0
300.0
400.0
500.0
600.0
TestLineIntensity(mV)
hCG Concentration in Buffer (mIU/mL)
0 100 200 300
Anteo Mix&Go
Magnetic
Particle
Covalently
Conjugated
Magnetic
Particle
Find out more at anteotech.com
hCG
Concentration
(mIU/mL)
in Urine
Anteo Mix&Go Magnetic Particle Covalently Conjugated Magnetic Particle
Visual
Grade
Reader
Average
TL (mv)
Stdev %CV
Visual
Grade
Reader
Average
TL (mv)
Stdev %CV
250 8.0 378.6 32.0 8% 4.2 86.9 35.2 40%
100 5.8 174.9 26.6 15% 3.2 43.7 29.4 67%
50 4.0 79.8 26.2 33% 1.6 9.0 20.1 224%
25 2.0 10.0 16.0 160% 0.6 7.2 16.0 224%
10 0.2 6.4 14.4 224% 0.2 0.0 0.0 NA
5 0.0 0.0 0.0 NA 0.2 7.8 17.4 224%
0 0.0 0.0 0.0 NA 0.0 0.0 0.0 NA
Table 2: Visual and reader results of the Anteo and covalently conjugated magnetic particle hCG assays with urine.
Figures 3 and 4: Pictures of lateral flow strips run with the Anteo and covalently conjugated magnetic particles
with hCG diluted in buffer.
Figure 3: Anteo magnetic particle test strips run with hCG diluted in buffer
Figure 4: Covalently conjugated magnetic particle test strips run with hCG diluted in buffer.
250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL
250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL
Find out more at anteotech.com
Magnetic Particle Conjugation Evaluation
hCG Spiked in Urine Dose Response Curve
Visual Grades (Average of N=5; <25mIU/mL)
VisualGrades(0-10)
hCG Spiked in Urine (mIU/mL)
Anteo Mix&Go
Magnetic
Particle
Covalently
Conjugated
Magnetic
Particle0.0
1.0
0.5
1.5
2.0
2.5
0 10 20 30
Figures 5 and 6: Graphs of the reader results from the Anteo and covalently conjugated magnetic particle hCG assays in urine.
Magnetic Particle Conjugation Evaluation
hCG Spiked in Urine Dose Response Curve
Visual Grades (Average of N=5)
0.0
2.0
4.0
6.0
8.0
10.0
VisualGrades(0-10)
hCG Spiked in Urine (mIU/mL)
0 100 200 300
Anteo Mix&Go
Magnetic
Particle
Covalently
Conjugated
Magnetic
Particle
Conclusions
The direct comparison of the two tests show:
The Anteo technology based hCG assay is 5 times
more sensitive than the covalently conjugated magnetic
particle based test using the same critical reagents. The
estimated limit of detection for the Anteo assay is ~10 mIU/
mL of hCG in buffer and ~25 mIU/mL of hCG in urine
while the covalently conjugated hCG assay has limits
of detection of ~50 and 100 mIU/ml of hCG in buffer
and urine respectively. The hCG assay using the Anteo
particles used half the amount of antibody to achieve
better sensitivity than the covalently conjugated assay.
The variability of the Anteo technology based hCG assay is
lower than the covalently conjugated hCG assay. There is
also less non-specific binding with the Anteo particles.
The Anteo technology based magnetic particle hCG lateral
flow assay produced a comparable limit of detection to the
optimised colloidal gold reference assay of ~25 mIU/mL in
urine. This confirms that particles conjugated with the Anteo
Activation Reagent can rival traditionally used gold colloids in
lateral flow applications and bring with it benefits associated
with cost, handling and manufacturability improvements.
Find out more at anteotech.com
250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL
250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL
Figures 7 and 8: Pictures of the Anteo and covalently conjugated magnetic particle hCG assays run with hCG diluted in urine.
Figure 7: Anteo magnetic particle test strips run with hCG diluted with urine.
Figure 8: Covalently conjugated magnetic particle test strips run with hCG diluted with urine.
Anteo Products Lateral Flow
Anteo’s Lateral Flow Coupling Kit makes conducting
lateral flow tests with magnetic beads easier and
more flexible. Magnetic separation removes the need
to perform centrifugation and filtration concentration.
Advantages
•	 An alternative to EDC chemistry the
Anteo Kit is easy to use
•	 Less hazardous chemicals to work with,
Anteo’s Activation Reagent is water based
•	 Flexibility: you can bind any antibody
subclass to your surface
•	 Promotes monolayer binding which
increases antibody functionality
•	 Reduces aggregation
•	 Low background noise
The Anteo Lateral Flow Coupling Kit is
launching at AACC 2016. For more information
on this kit email sales@anteotech.com
About Anteo Technologies
Anteo Technologies is transforming the way
scientists work by providing new, powerful
tools developed with their patented nanoscale
engineered Activation Reagent. This Reagent
significantly advances traditional protein coupling
methods, leading to increased assay sensitivity.
Anteo’s Activation Reagent is designed by using cross-
linking metal ions to enable gentle yet secure multi-
point binding of biomolecules to synthetic materials.
The Reagent also delivers solutions to the challenges
of establishing highly functional biological interfaces
between fragile biomolecules, and synthetic and often
incompatible hydrophilic/hydrophobic materials such as
silica, polystyrene, COC/COP, metals and ceramics.
Water based, Anteo’s Activation Reagent is easy
to use. Simply pipette the reagent onto your
surface to activate. Hands on time for surface
activation and coupling is 60 minutes.
Find out more at anteotech.com
Find out more at anteotech.com
References
Sharma S, Zapatero-Rodríguez J, Estrela P, O’Kennedy R. Point-of-Care
Diagnostics in Low Resource Settings: Present Status and Future Role
of Microfluidics. Biosensors (Basel). 2015 Aug 13;5(3):577-601.
Shi L, Wu F, Wen Y, Zhao F, Xiang J, Ma L. A novel method to
detect Listeria monocytogenes via superparamagnetic lateral flow
immunoassay. Anal Bioanal Chem. 2015 Jan;407(2):529-35.
Wong RC and Tse HY. Lateral flow immunoassays, Humana
press, DOI 10.1007/978-1-59745-240-3, 2009

More Related Content

Viewers also liked

ASX Australian Biotech Market cap list
ASX Australian Biotech Market cap listASX Australian Biotech Market cap list
ASX Australian Biotech Market cap listMatt Sanderson
 
Anteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow testAnteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow testMatt Sanderson
 
Anteo Technologies Investor Presentation
Anteo Technologies Investor PresentationAnteo Technologies Investor Presentation
Anteo Technologies Investor PresentationMatt Sanderson
 
Anteo Technologies Lateral Flow
Anteo Technologies Lateral FlowAnteo Technologies Lateral Flow
Anteo Technologies Lateral FlowMatt Sanderson
 
Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource Matt Sanderson
 
Novel methods in Bioseparations
Novel methods in BioseparationsNovel methods in Bioseparations
Novel methods in BioseparationsMatt Sanderson
 
Gluten Test Kit Methods Usda
Gluten Test Kit Methods UsdaGluten Test Kit Methods Usda
Gluten Test Kit Methods UsdaHomer Hartage
 
Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...
Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...
Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...David Dazhia Lazarus
 
Multiplex energy metabolism
Multiplex energy metabolismMultiplex energy metabolism
Multiplex energy metabolismMatt Sanderson
 
FMDv NSP Ab Rapid test kit
FMDv NSP Ab Rapid test kitFMDv NSP Ab Rapid test kit
FMDv NSP Ab Rapid test kit小咪 翁
 
Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015Matt Sanderson
 
Top-Selling ELISA Kits 2016 - Elabscience
Top-Selling ELISA Kits 2016 - ElabscienceTop-Selling ELISA Kits 2016 - Elabscience
Top-Selling ELISA Kits 2016 - ElabscienceElabscience
 
Silicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium IonSilicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium IonMatt Sanderson
 
Multiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow AssaysMultiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow AssaysBrendan O'Farrell
 
A guide to lateral flow immunoassay development
A guide to lateral flow immunoassay developmentA guide to lateral flow immunoassay development
A guide to lateral flow immunoassay developmentExpedeon
 
immunoassay techniques
immunoassay techniquesimmunoassay techniques
immunoassay techniquessandeep kumar
 

Viewers also liked (20)

ASX Australian Biotech Market cap list
ASX Australian Biotech Market cap listASX Australian Biotech Market cap list
ASX Australian Biotech Market cap list
 
Anteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow testAnteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow test
 
Anteo Technologies Investor Presentation
Anteo Technologies Investor PresentationAnteo Technologies Investor Presentation
Anteo Technologies Investor Presentation
 
Anteo Technologies Lateral Flow
Anteo Technologies Lateral FlowAnteo Technologies Lateral Flow
Anteo Technologies Lateral Flow
 
Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource
 
Novel methods in Bioseparations
Novel methods in BioseparationsNovel methods in Bioseparations
Novel methods in Bioseparations
 
Gluten Test Kit Methods Usda
Gluten Test Kit Methods UsdaGluten Test Kit Methods Usda
Gluten Test Kit Methods Usda
 
Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...
Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...
Rapid detection of Foot and Mouth Disease Non-Structural Proteins using later...
 
Multiplex energy metabolism
Multiplex energy metabolismMultiplex energy metabolism
Multiplex energy metabolism
 
FMDv NSP Ab Rapid test kit
FMDv NSP Ab Rapid test kitFMDv NSP Ab Rapid test kit
FMDv NSP Ab Rapid test kit
 
Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015
 
Top-Selling ELISA Kits 2016 - Elabscience
Top-Selling ELISA Kits 2016 - ElabscienceTop-Selling ELISA Kits 2016 - Elabscience
Top-Selling ELISA Kits 2016 - Elabscience
 
Immunoassay: The uses of ELISA
Immunoassay: The uses of ELISAImmunoassay: The uses of ELISA
Immunoassay: The uses of ELISA
 
DCN Company Profile 2015
DCN Company Profile 2015DCN Company Profile 2015
DCN Company Profile 2015
 
Silicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium IonSilicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium Ion
 
Lateral Flow Assay
Lateral Flow AssayLateral Flow Assay
Lateral Flow Assay
 
Multiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow AssaysMultiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow Assays
 
A guide to lateral flow immunoassay development
A guide to lateral flow immunoassay developmentA guide to lateral flow immunoassay development
A guide to lateral flow immunoassay development
 
Immunoassay
ImmunoassayImmunoassay
Immunoassay
 
immunoassay techniques
immunoassay techniquesimmunoassay techniques
immunoassay techniques
 

Similar to Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Chemistry With Magnetic Particles

Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...Anteo Technologies
 
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...Anteo Technologies
 
ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS Lavaraj Devkota
 
Chromosomal damage risk assessment to benzene exposure
Chromosomal damage risk assessment to benzene exposureChromosomal damage risk assessment to benzene exposure
Chromosomal damage risk assessment to benzene exposureAlexander Decker
 
11.chromosomal damage risk assessment to benzene exposure
11.chromosomal damage risk assessment to benzene exposure11.chromosomal damage risk assessment to benzene exposure
11.chromosomal damage risk assessment to benzene exposureAlexander Decker
 
Novel composite electrodes:Preparation and application to the electroanalytic...
Novel composite electrodes:Preparation and application to the electroanalytic...Novel composite electrodes:Preparation and application to the electroanalytic...
Novel composite electrodes:Preparation and application to the electroanalytic...Université de Dschang
 
Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Arun kumar
 
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...Wael Ebied
 
Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...KBI Biopharma
 
biotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.pptbiotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.pptmisgana18
 
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...KBI Biopharma
 
Enzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide DetectionEnzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide DetectionSubhasis Sarkar
 
Quantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GCQuantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GCDr. Nimra Khan
 
CNV and aneuploidy detection by Ion semiconductor sequencing
CNV and aneuploidy detection by Ion semiconductor sequencingCNV and aneuploidy detection by Ion semiconductor sequencing
CNV and aneuploidy detection by Ion semiconductor sequencingThermo Fisher Scientific
 
Case Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical InstrumentsCase Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical InstrumentsJordi Labs
 
Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...
Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...
Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...CrimsonPublishers-SBB
 
Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...
Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...
Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...Erhunmwunsee Osazee
 

Similar to Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Chemistry With Magnetic Particles (20)

Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
 
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
 
ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS
 
Chromosomal damage risk assessment to benzene exposure
Chromosomal damage risk assessment to benzene exposureChromosomal damage risk assessment to benzene exposure
Chromosomal damage risk assessment to benzene exposure
 
11.chromosomal damage risk assessment to benzene exposure
11.chromosomal damage risk assessment to benzene exposure11.chromosomal damage risk assessment to benzene exposure
11.chromosomal damage risk assessment to benzene exposure
 
Novel composite electrodes:Preparation and application to the electroanalytic...
Novel composite electrodes:Preparation and application to the electroanalytic...Novel composite electrodes:Preparation and application to the electroanalytic...
Novel composite electrodes:Preparation and application to the electroanalytic...
 
Chap3 1
Chap3 1Chap3 1
Chap3 1
 
Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...
 
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
 
Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...
 
biotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.pptbiotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.ppt
 
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
 
Enzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide DetectionEnzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide Detection
 
Quantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GCQuantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GC
 
CNV and aneuploidy detection by Ion semiconductor sequencing
CNV and aneuploidy detection by Ion semiconductor sequencingCNV and aneuploidy detection by Ion semiconductor sequencing
CNV and aneuploidy detection by Ion semiconductor sequencing
 
Case Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical InstrumentsCase Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical Instruments
 
Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...
Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...
Crimson Publishers -A Sensor Multiplatform for Non Invasive Diagnosis of Pros...
 
1._1-8
1._1-81._1-8
1._1-8
 
Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...
Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...
Comparative Haematology and Urinary Analysis of Passive Inhalers of Petrol Fu...
 
Plegable
PlegablePlegable
Plegable
 

Recently uploaded

Botany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdfBotany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdfSumit Kumar yadav
 
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCRStunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCRDelhi Call girls
 
G9 Science Q4- Week 1-2 Projectile Motion.ppt
G9 Science Q4- Week 1-2 Projectile Motion.pptG9 Science Q4- Week 1-2 Projectile Motion.ppt
G9 Science Q4- Week 1-2 Projectile Motion.pptMAESTRELLAMesa2
 
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsHubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsSérgio Sacani
 
Grafana in space: Monitoring Japan's SLIM moon lander in real time
Grafana in space: Monitoring Japan's SLIM moon lander  in real timeGrafana in space: Monitoring Japan's SLIM moon lander  in real time
Grafana in space: Monitoring Japan's SLIM moon lander in real timeSatoshi NAKAHIRA
 
Botany 4th semester file By Sumit Kumar yadav.pdf
Botany 4th semester file By Sumit Kumar yadav.pdfBotany 4th semester file By Sumit Kumar yadav.pdf
Botany 4th semester file By Sumit Kumar yadav.pdfSumit Kumar yadav
 
STERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCE
STERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCESTERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCE
STERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCEPRINCE C P
 
SOLUBLE PATTERN RECOGNITION RECEPTORS.pptx
SOLUBLE PATTERN RECOGNITION RECEPTORS.pptxSOLUBLE PATTERN RECOGNITION RECEPTORS.pptx
SOLUBLE PATTERN RECOGNITION RECEPTORS.pptxkessiyaTpeter
 
Formation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disksFormation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disksSérgio Sacani
 
Cultivation of KODO MILLET . made by Ghanshyam pptx
Cultivation of KODO MILLET . made by Ghanshyam pptxCultivation of KODO MILLET . made by Ghanshyam pptx
Cultivation of KODO MILLET . made by Ghanshyam pptxpradhanghanshyam7136
 
Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...
Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...
Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...jana861314
 
A relative description on Sonoporation.pdf
A relative description on Sonoporation.pdfA relative description on Sonoporation.pdf
A relative description on Sonoporation.pdfnehabiju2046
 
Nanoparticles synthesis and characterization​ ​
Nanoparticles synthesis and characterization​  ​Nanoparticles synthesis and characterization​  ​
Nanoparticles synthesis and characterization​ ​kaibalyasahoo82800
 
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...Lokesh Kothari
 
Disentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOSTDisentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOSTSérgio Sacani
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPirithiRaju
 

Recently uploaded (20)

CELL -Structural and Functional unit of life.pdf
CELL -Structural and Functional unit of life.pdfCELL -Structural and Functional unit of life.pdf
CELL -Structural and Functional unit of life.pdf
 
Botany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdfBotany 4th semester series (krishna).pdf
Botany 4th semester series (krishna).pdf
 
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCRStunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
 
G9 Science Q4- Week 1-2 Projectile Motion.ppt
G9 Science Q4- Week 1-2 Projectile Motion.pptG9 Science Q4- Week 1-2 Projectile Motion.ppt
G9 Science Q4- Week 1-2 Projectile Motion.ppt
 
9953056974 Young Call Girls In Mahavir enclave Indian Quality Escort service
9953056974 Young Call Girls In Mahavir enclave Indian Quality Escort service9953056974 Young Call Girls In Mahavir enclave Indian Quality Escort service
9953056974 Young Call Girls In Mahavir enclave Indian Quality Escort service
 
Engler and Prantl system of classification in plant taxonomy
Engler and Prantl system of classification in plant taxonomyEngler and Prantl system of classification in plant taxonomy
Engler and Prantl system of classification in plant taxonomy
 
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroidsHubble Asteroid Hunter III. Physical properties of newly found asteroids
Hubble Asteroid Hunter III. Physical properties of newly found asteroids
 
Grafana in space: Monitoring Japan's SLIM moon lander in real time
Grafana in space: Monitoring Japan's SLIM moon lander  in real timeGrafana in space: Monitoring Japan's SLIM moon lander  in real time
Grafana in space: Monitoring Japan's SLIM moon lander in real time
 
Botany 4th semester file By Sumit Kumar yadav.pdf
Botany 4th semester file By Sumit Kumar yadav.pdfBotany 4th semester file By Sumit Kumar yadav.pdf
Botany 4th semester file By Sumit Kumar yadav.pdf
 
STERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCE
STERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCESTERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCE
STERILITY TESTING OF PHARMACEUTICALS ppt by DR.C.P.PRINCE
 
SOLUBLE PATTERN RECOGNITION RECEPTORS.pptx
SOLUBLE PATTERN RECOGNITION RECEPTORS.pptxSOLUBLE PATTERN RECOGNITION RECEPTORS.pptx
SOLUBLE PATTERN RECOGNITION RECEPTORS.pptx
 
Formation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disksFormation of low mass protostars and their circumstellar disks
Formation of low mass protostars and their circumstellar disks
 
Cultivation of KODO MILLET . made by Ghanshyam pptx
Cultivation of KODO MILLET . made by Ghanshyam pptxCultivation of KODO MILLET . made by Ghanshyam pptx
Cultivation of KODO MILLET . made by Ghanshyam pptx
 
Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...
Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...
Traditional Agroforestry System in India- Shifting Cultivation, Taungya, Home...
 
A relative description on Sonoporation.pdf
A relative description on Sonoporation.pdfA relative description on Sonoporation.pdf
A relative description on Sonoporation.pdf
 
Nanoparticles synthesis and characterization​ ​
Nanoparticles synthesis and characterization​  ​Nanoparticles synthesis and characterization​  ​
Nanoparticles synthesis and characterization​ ​
 
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
Labelling Requirements and Label Claims for Dietary Supplements and Recommend...
 
The Philosophy of Science
The Philosophy of ScienceThe Philosophy of Science
The Philosophy of Science
 
Disentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOSTDisentangling the origin of chemical differences using GHOST
Disentangling the origin of chemical differences using GHOST
 
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdfPests of cotton_Sucking_Pests_Dr.UPR.pdf
Pests of cotton_Sucking_Pests_Dr.UPR.pdf
 

Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Chemistry With Magnetic Particles

  • 1. Find out more at anteotech.com Lateral Flow Achieving 5 x More Sensitivity and Less Variation in Results than Covalent Chemistry with Magnetic Particles by Anteo Technologies
  • 2. Introduction Lateral flow immunoassays (LFIA) or strip tests are conceptually simple immunoassays well suited to Point of Care (PoC) applications requiring little or no infrastructure support (Wong and Tse 2009, Sharma et al 2015). In its original conception, the test is easy-to-use, relatively fast and low cost, giving good qualitative information where a simple yes/no answer is sufficient. However, this same test system has been evolving to compete with more traditional laboratory based platforms. To achieve the requirements of greater assay accuracy and sensitivity without losing its basic advantages for PoC conditions, the actual strip test is now at the core of instruments such as hand held analysers comprising sample handling, reading and analysis modules. In brief, LFIA is no longer at the cheap and simple end of the spectrum in terms of diagnostic technologies (Sharma et al, 2015). The above trends have resulted in continuous improvements over various materials and reagents used in LFIA of which Anteo Technologies’ Lateral Flow Coupling Kit is one new development. Historically, gold colloids have dominated in LFIA with visualisation of conjugated gold colloids as they were captured on pre-defined lines on the test strip. Passive conjugation of antibodies on gold colloids is conceptually simple but depending on the antibody, can lead to difficulties such as poor Ab functionality and loss of colloidal stability resulting in the formation of aggregates (Wong and Tse 2009). These difficulties change with scale, and achieving good uniformity during scale-up can often be very challenging. Summary A new method of conjugating antibodies to particles was compared to covalent conjugation in a lateral flow assay for human Chorionic Gonadotropin (hCG). Anteo Technologies will be launching a new Lateral Flow Coupling Kit at AACC 2016 that contains Anteo’s patented Activation Reagent with all necessary buffers for conjugating antibodies to a range of particles used for lateral flow applications. Here we demonstrate the performance advantages of using activated magnetic particles, only requiring the addition of antibody solution and blocking. Keeping all other variables constant, this kit was directly compared to covalently (EDAC chemistry) conjugated magnetic particles of the same type using visual detection as the readout. In this hCG study, this coupling kit achieves five times more sensitivity than the covalently conjugated magnetic particle based assay using half the amount of antibody. Comparison of Anteo’s Activated Magnetic Particles and Covalently Conjugated Magnetic Particles in a Lateral Flow Assay for Human Chorionic Gonadotropin (hCG) Find out more at anteotech.com
  • 3. Using 200 nm Magnetic Particles The use of 200 nm magnetic nanoparticles activated with Anteo’s reagent for lateral flow as an alternative to colloidal gold addresses a number of potential problematic situations encountered in the development of new lateral flow assays. For example, preparation of labelled conjugates is simplified using magnetic particles as they are easy to separate during conjugation, washing and blocking steps, with the additional benefit of inclusion of a magnetic label for performing magnetic immunoassays (MIA) (Shi et al 2015). Alternatively, when used in conjunction with sufficiently dark magnetic particles visual detection alone can produce results rivalling those of other coloured particle types. The kit contains Anteo’s Activation Reagent that can be used to activate particle sizes from 200 nm to 1 um that become immediately reactive to antibodies once Anteo’s reagent is applied. Conventionally, conjugation of antibodies to magnetic particles is commonly done by covalent chemistry using reagents such as 1-Ethyl-3-(3-dimethylaminopropyl) carbodiimide (EDC) and N-hydroxysulfosuccinimide (S-NHS). A reactive ester is formed on the magnetic particles, which reacts with amino groups on the protein. Chemically, the reaction is simple but fraught with multiple variables to consider when trying to achieve optimum results. Conditions that increase stability of the ester during synthesis make it less reactive to amino groups on the protein. Increasing ester reactivity leads to side reactions. While different approaches can mitigate many such problems, these methods require pre-activation of particles which cannot be stored for any length of time. The magnetic particles activated by Anteo’s Activation Reagent are claimed to be stable in storage and highly reactive to proteins with other advantages such as ease of use, less antibodies used, faster conjugation and reduced aggregation amongst its benefits. Using a hCG lateral flow assay previously developed using colloidal gold at Diagnostic Consulting Network (DCN), the Anteo coupling kits were compared to covalently conjugated magnetic particles using the same assay materials (membrane, pads, and antibodies) with the exception of the gold colloids. Find out more at anteotech.com
  • 4. Find out more at anteotech.com Method The hCG assay has previously been developed using colloidal gold at Diagnostic Consulting Network (DCN), Carlsbad, CA for internal use and demonstration purposes. The same assay materials (membrane, pads, and antibodies) were used with Anteo’s technology and covalently conjugated magnetic particles. All materials and tests were performed in DCN’s laboratories using magnetic particles conjugated with anti-hCG antibody using Anteo’s protocol and DCN’s standard covalent conjugation procedure. hCG concentrations from 0 to 250 mIU/ml were prepared in 1%BSA, 0.1%Tw- 20 1xPBS and in negative pooled human urine. Tests were performed using 100ul of sample and test strips with the covalently conjugated and Anteo’s conjugated particles dried onto a conjugate pad. Five replicates for each concentration were tested and results were recorded 15 minutes after the addition of the sample using DCN’s visual grading scale and an ESE Quant (Qiagen Lake Constance, Germany) reflectance reader. The amount of antibody conjugated to the different particles and the amount of particles used per test were optimised to produce the highest sensitivity. Anteo’s conjugate was tested with an antibody loading of 25ug of anti-hCG antibody per mg of particles at 0.05% solids. The covalently conjugated particle was tested with an antibody loading of 50ug of anti-hCG per mg particles at 0.05% solids. Results The results for the visual grades and the values recorded by the reflectance reader, including standard deviation and %CV are shown in Table 1 for each of the two tests. The limit of detection for the Anteo assay was ~10 mIU/ ml hCG in buffer and ~25 mIU/ml in urine for the visual and reader based results. The limit of detection for the covalently conjugated hCG assay was ~50 mIU/mL in buffer and ~100 mIU/mL in urine. As a reference the limit of the detection for the optimised colloidal gold assay is at ~2.5 mIU/mL in buffer and ~25 mIU/mL in urine. NOTE: A visual grade of 2 or higher at the test line is considered positive, while the cut-off for the reader is at 30mV.
  • 5. Find out more at anteotech.com Figure 1: Graph of the results from the Anteo and covalently conjugated magnetic particle hCG assays in buffer measured by a Qiagen reader. Table 1: Visual and Reader results of the Anteo’s technology and covalently conjugated magnetic particle based hCG assays in buffer. hCG Concentration (mIU/mL) in Buffer Anteo Mix&Go Magnetic Particle Covalently Conjugated Magnetic Particle Visual Grade Average TL (mv) Stdev %CV Visual Grade Average TL (mv) Stdev %CV 250 9.0 486.9 16.3 3% 5.4 163.5 25.8 16% 100 7.0 314.0 11.3 4% 4.0 108.0 12.3 11% 50 6.0 174.8 8.2 5% 3.0 62.8 6.0 10% 25 4.0 90.9 7.7 8% 2.6 59.5 12.8 21% 10 2.0 36.8 5.1 14% 2.2 44.7 6.2 14% 5 0.0 0.0 0.0 NA 1.8 34.3 11.4 33% 0 0.0 0.0 0.0 NA 1.0 6.5 12.4 189% 0.0 200.0 100.0 300.0 400.0 500.0 600.0 TestLineIntensity(mV) hCG Concentration in Buffer (mIU/mL) 0 100 200 300 Anteo Mix&Go Magnetic Particle Covalently Conjugated Magnetic Particle
  • 6. Find out more at anteotech.com hCG Concentration (mIU/mL) in Urine Anteo Mix&Go Magnetic Particle Covalently Conjugated Magnetic Particle Visual Grade Reader Average TL (mv) Stdev %CV Visual Grade Reader Average TL (mv) Stdev %CV 250 8.0 378.6 32.0 8% 4.2 86.9 35.2 40% 100 5.8 174.9 26.6 15% 3.2 43.7 29.4 67% 50 4.0 79.8 26.2 33% 1.6 9.0 20.1 224% 25 2.0 10.0 16.0 160% 0.6 7.2 16.0 224% 10 0.2 6.4 14.4 224% 0.2 0.0 0.0 NA 5 0.0 0.0 0.0 NA 0.2 7.8 17.4 224% 0 0.0 0.0 0.0 NA 0.0 0.0 0.0 NA Table 2: Visual and reader results of the Anteo and covalently conjugated magnetic particle hCG assays with urine. Figures 3 and 4: Pictures of lateral flow strips run with the Anteo and covalently conjugated magnetic particles with hCG diluted in buffer. Figure 3: Anteo magnetic particle test strips run with hCG diluted in buffer Figure 4: Covalently conjugated magnetic particle test strips run with hCG diluted in buffer. 250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL 250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL
  • 7. Find out more at anteotech.com Magnetic Particle Conjugation Evaluation hCG Spiked in Urine Dose Response Curve Visual Grades (Average of N=5; <25mIU/mL) VisualGrades(0-10) hCG Spiked in Urine (mIU/mL) Anteo Mix&Go Magnetic Particle Covalently Conjugated Magnetic Particle0.0 1.0 0.5 1.5 2.0 2.5 0 10 20 30 Figures 5 and 6: Graphs of the reader results from the Anteo and covalently conjugated magnetic particle hCG assays in urine. Magnetic Particle Conjugation Evaluation hCG Spiked in Urine Dose Response Curve Visual Grades (Average of N=5) 0.0 2.0 4.0 6.0 8.0 10.0 VisualGrades(0-10) hCG Spiked in Urine (mIU/mL) 0 100 200 300 Anteo Mix&Go Magnetic Particle Covalently Conjugated Magnetic Particle
  • 8. Conclusions The direct comparison of the two tests show: The Anteo technology based hCG assay is 5 times more sensitive than the covalently conjugated magnetic particle based test using the same critical reagents. The estimated limit of detection for the Anteo assay is ~10 mIU/ mL of hCG in buffer and ~25 mIU/mL of hCG in urine while the covalently conjugated hCG assay has limits of detection of ~50 and 100 mIU/ml of hCG in buffer and urine respectively. The hCG assay using the Anteo particles used half the amount of antibody to achieve better sensitivity than the covalently conjugated assay. The variability of the Anteo technology based hCG assay is lower than the covalently conjugated hCG assay. There is also less non-specific binding with the Anteo particles. The Anteo technology based magnetic particle hCG lateral flow assay produced a comparable limit of detection to the optimised colloidal gold reference assay of ~25 mIU/mL in urine. This confirms that particles conjugated with the Anteo Activation Reagent can rival traditionally used gold colloids in lateral flow applications and bring with it benefits associated with cost, handling and manufacturability improvements. Find out more at anteotech.com 250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL 250 mIU/mL 100 mIU/mL 50 mIU/mL 25 mIU/mL 10 mIU/mL 5 mIU/mL 0 mIU/mL Figures 7 and 8: Pictures of the Anteo and covalently conjugated magnetic particle hCG assays run with hCG diluted in urine. Figure 7: Anteo magnetic particle test strips run with hCG diluted with urine. Figure 8: Covalently conjugated magnetic particle test strips run with hCG diluted with urine.
  • 9. Anteo Products Lateral Flow Anteo’s Lateral Flow Coupling Kit makes conducting lateral flow tests with magnetic beads easier and more flexible. Magnetic separation removes the need to perform centrifugation and filtration concentration. Advantages • An alternative to EDC chemistry the Anteo Kit is easy to use • Less hazardous chemicals to work with, Anteo’s Activation Reagent is water based • Flexibility: you can bind any antibody subclass to your surface • Promotes monolayer binding which increases antibody functionality • Reduces aggregation • Low background noise The Anteo Lateral Flow Coupling Kit is launching at AACC 2016. For more information on this kit email sales@anteotech.com About Anteo Technologies Anteo Technologies is transforming the way scientists work by providing new, powerful tools developed with their patented nanoscale engineered Activation Reagent. This Reagent significantly advances traditional protein coupling methods, leading to increased assay sensitivity. Anteo’s Activation Reagent is designed by using cross- linking metal ions to enable gentle yet secure multi- point binding of biomolecules to synthetic materials. The Reagent also delivers solutions to the challenges of establishing highly functional biological interfaces between fragile biomolecules, and synthetic and often incompatible hydrophilic/hydrophobic materials such as silica, polystyrene, COC/COP, metals and ceramics. Water based, Anteo’s Activation Reagent is easy to use. Simply pipette the reagent onto your surface to activate. Hands on time for surface activation and coupling is 60 minutes. Find out more at anteotech.com Find out more at anteotech.com References Sharma S, Zapatero-Rodríguez J, Estrela P, O’Kennedy R. Point-of-Care Diagnostics in Low Resource Settings: Present Status and Future Role of Microfluidics. Biosensors (Basel). 2015 Aug 13;5(3):577-601. Shi L, Wu F, Wen Y, Zhao F, Xiang J, Ma L. A novel method to detect Listeria monocytogenes via superparamagnetic lateral flow immunoassay. Anal Bioanal Chem. 2015 Jan;407(2):529-35. Wong RC and Tse HY. Lateral flow immunoassays, Humana press, DOI 10.1007/978-1-59745-240-3, 2009