SlideShare a Scribd company logo
1 of 18
Lecture No. 08
Course title:
Atomic spectroscopy
Topic: Spectral Interferences, background absorption
Course instructor: Dr. Salma Amir
GFCW Peshawar
Spectral Interferences
Spectral interferences are those in which the measured light
absorption is erroneously high due to absorption by a
species other than the analyte element.
Absorption due to the presence of light-absorbing molecules
in the flame and light dimming due to the presence of small
particles in the flame are much more common spectral
interferences. Such phenomena are referred to as
background absorption.
The only cures for direct atomic spectral interference are
(1) to choose an alternate analytical wavelength
(2) to remove the interfering element from the sample
Background Absorption
 A common occurrence that results in spectral interference is absorption of
the HCL radiation by molecules or polyatomic species in the atomizer. This
is called “background absorption”; it occurs because not all of the molecules
from the sample matrix and the flame gases are completely atomized. This
type of interference is more commonplace at short wavelengths (<250 nm)
where many compounds absorb.
 Since atoms have extremely narrow absorption lines, there are few
problems involving interferences where one element absorbs at the
wavelength of another. Even when an absorbing wavelength of another
element falls within the spectral bandwidth used, no absorption can occur
unless the light source produces light at that wavelength, i.e., that element
is also present in the light source. However, undissociated molecular forms
of matrix materials may have broadband absorption spectra, and tiny solid
particles in the flame may scatter light over a wide wavelength region.
When this type of nonspecific absorption overlaps the atomic absorption
wavelength of the analyte, background absorption occurs.
Causes of Background Absorption
 Incomplete combustion of organic molecules in the atomizer can cause
serious background interference problems.
 If a flame atomizer is used, incomplete combustion of the solvent may
take place, particularly if the flame is too reducing (fuel rich). The
extent of the interference depends on flame conditions (reducing or
oxidizing), the flame temperature used, the solvent used, and the
sample aspiration rate.
 Background interference is much more severe when graphite furnace
atomizers are used because the pyrolysis step is limited to a maximum
temperature that does not volatilize the analyte. Consequently, many
matrix molecules are not thermally decomposed. They then volatilize
into the light path as the higher atomization temperature is reached
and absorb significant amounts of the source radiation.
Background absorption correction
 To compensate for this problem, the background absorption
must be measured and subtracted from the total measured
absorption to determine the true atomic absorption
component. Various Background correction techniques
employed are
 1. Two line background correction
 2. Continuum source background correction
 3. Zeeman background correction
 4. Smith–Hieftje Background Corrector
1. Two-Line Background Correction
An early manual method of measuring background and applying a correction
for it used the absorption of a nearby nonresonance line. The emission
spectrum from a hollow cathode is quite rich in emission lines and contains
the resonance lines of the element of interest plus many other nonresonance
emission lines from the element and the filler gas. The analyte atoms do not
absorb these nonresonance lines; however, the broad molecular background
absorbs them.
Two measurements are made.
 The absorbance at a nonresonance line close to the analyte resonance line
but far enough away that atomic absorption does not occur is measured.
 Then the absorbance at the resonance line of the analyte is measured.
The absorbance at the nonresonance
line is due only to background
absorption. The absorbance at the
resonance line is due to both atomic
and background absorption. The
difference in absorbance between
the resonance and nonresonance
lines is the net atomic absorption.
SUMMARY:
Resonance line= analyte + background
Non-resonance line= background
Net absorption due to analyte = Resonance - Non-resonance
2. Continuum source background
correction
 Continuum source background correction is a technique for automatically
measuring and compensating for any background component which might be
present in an atomic absorption measurement. This method incorporates a
continuum light source in a modified optical system
 In the continuum source background correction method, when an HCL source
is used, the absorption measured is the total of the atomic and background
absorptions. When a continuum lamp source is used, only the background
absorption is measured. The continuum lamp, which emits light over a range
or band of wavelengths, is placed into the spectrometer system as shown in
Figure 3.7. This setup allows radiation from both the HCL and the continuum
lamp to follow the same path to reach the detector. The detector observes
each source alternately in time, either through the use of mirrored choppers
or through pulsing of the lamp currents.
When the HCL lamp is in position as the source, the emission line from the HCL is
only about 0.002 nm wide; in other words, it fills only about 1% of the spectral
window as shown in the upper left-hand-side drawing in Figure 3.8. When the HCL
radiation passes through the flame, both the free atoms at the resonance
line and the broadband absorbing
molecules will absorb the line. This
results in significant attenuation of the
light reaching the detector, as shown in
the upper center and right-hand-side
drawing in Figure 3.8. When the
continuum source is in place, the
continuum lamp emission fills the entire
spectral window as shown in the lower
left-hand-side drawing in Figure 3.8.
Any absorption of the radiation from the
continuum lamp observed is broadband
background absorption, since it will be
absorbed over the entire 0.2 nm as seen
in the lower right-hand-side drawing in
Figure 3.8
 The absorbance of the continuum source is therefore an accurate
measure of background absorption. An advantage of this method is
that the background is measured at the same nominal wavelength as
the resonance line, resulting in more accurate correction.
3. Zeeman Background correction
 Atomic absorption lines occur at discrete wavelengths because the transition
that gives rise to the absorption is between two discrete energy levels.
However, when a vapor-phase atom is placed in a strong magnetic field, the
electronic energy levels split. This gives rise to several absorption lines for a
given transition in place of the single absorption line in the absence of a
magnetic field. This occurs in all atomic spectra and is called Zeeman splitting
or the Zeeman effect.
 In the simplest case, the Zeeman effect splits an absorption line into two
components. The first component is the π-component, at the same wavelength
as before (unshifted); the second is the σ-component, which undergoes both a
positive and negative shift in wavelength, resulting in two equally spaced lines
on either side of the original line. This splitting pattern is presented in Figure
6.27. The splitting results in lines that are separated by approximately 0.01 nm
or less depending on the field strength. The strength of the magnetic field used
is between 7 and 15 kG. Background absorption and scatter are usually not
affected by a magnetic field.
The π- and σ-components respond differently to polarized light. The π-
component absorbs light polarized in the direction parallel to the magnetic
field. The σ-components absorb only radiation polarized 90° to the applied
field. The combination of splitting and polarization differences can be used
to measured total absorbance (atomic plus background) and background
only, permitting the net atomic absorption to be determined.
A permanent magnet can be placed around the furnace to split the energy levels.
A rotating polarizer is used in front of the HCL or EDL. During that portion of
the cycle when the light is polarized parallel to the magnetic field, both atomic
and background absorptions occur. No atomic absorption occurs when the light
is polarized perpendicular
to the field, but background
absorption still occurs.
The difference between
the two is the net atomic
absorption. Such a system
is a DC Zeeman correction
system.
 Alternately, a fixed polarizer can be placed in front of the light source and an
electromagnet can be placed around the furnace. By making absorption
measurements with the magnetic field off (atomic plus background) and with
the magnetic field on (background only), the net atomic absorption signal can
be determined. This is a transverse AC Zeeman correction system. The AC
electromagnet can also be oriented so that the field is along the light path (a
longitudinal AC Zeeman system) rather than across the light path. No
polarizer is required in a longitudinal AC Zeeman system.
 Zeeman correction can also be achieved by having an alternating magnetic
field surround the hollow cathode, causing the emission line to be split and
then not split as the field is turned on and off. By tuning the amplifier to this
frequency, it is possible to discriminate between the split and unsplit
radiation. A major difficulty with the technique is that the magnetic field used
to generate Zeeman splitting also interacts with the ions in the hollow
cathode. This causes the emission from the hollow cathode to be erratic,
which in turn introduces imprecision into the measurement. Most commercial
instruments with Zeeman background correctors put the magnet around the
atomizer.
4. Smith–Hieftje Background Corrector
 It will be remembered that the HCL functions by the creation of excited
atoms that radiate at the desired resonance wavelengths. After radiating, the
atoms form a cloud of neutral atoms that, unless removed, will absorb
resonance radiation from other emitting atoms. If the HCL is run at a high
current, an abundance of free atoms form. These free atoms absorb at
precisely the resonance lines the hollow cathode is intended to emit; an
example is shown in Figure 6.28, with the free atoms absorbing exactly at λ
over an extremely narrow bandwidth (Figure 6.28b). The result is that the
line emitted from the HCL is as depicted in Figure 6.28c instead of the
desired emission line depicted in Figure 6.28a. The phenomenon of
absorption of the central portion of the emission line by free atoms in the
lamp is called “self-reversal.” Such absorption is not easily detectable,
because it is at the very center of the emitted resonance line and very
difficult to resolve.
It is, of course, exactly the radiation that is most easily absorbed by the
atoms of the sample. In practice, if the HCL is operated at too high a current,
the self-reversal decreases the sensitivity of the analysis by removing
absorbable light. It also shortens the life of the lamp significantly.
Figure 6.28 distortion of spectral line shape in an HCL due to self-reversal.
(a) shape of the spectral line
emitted by the HCL.
(b) shape of the spectral
energy band absorbed by
cool atoms inside the HCL.
(c) shape of the net signal
emerging from an HCL,
showing self-absorption
of the center of the
emission band.
 The Smith–Hieftje background corrector has taken advantage of this self-
reversal phenomenon by pulsing the lamp, alternating between high current
and low current. At low current, a normal resonance line is emitted and the
sample undergoes normal atomic absorption. When the HCL is pulsed to a
high current, the center of the emission line is self-absorbed, leaving only
the wings of the emitted resonance line; the emitted line would look similar
to Figure 6.28c. The atoms of the sample do not significantly absorb such a
line. The broad background, however, absorbs the wings of the line.
Consequently, the absorption of the wings is a direct measurement of the
background absorption at the wavelength of the atomic resonance
absorption. The Smith–Hieftje technique can be used to automatically
correct for background. In practice, the low current is run for a fairly short
period of time and the absorbance measured. The high current is run for a
very short, sharp burst, liberating intense emission and free atoms inside
the hollow cathode and the absorbance is again measured. The difference
between the measurements is the net atomic absorption. There is then a
delay time to disperse the free atoms within the lamp before the cycle is
started again.

More Related Content

What's hot

Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyAnuradha Verma
 
Raman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumarRaman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumarNITISH KUMAR
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy Zainab&Sons
 
Inductively Coupled Plasma-Optical Emission Spectroscopy
Inductively Coupled Plasma-Optical Emission Spectroscopy Inductively Coupled Plasma-Optical Emission Spectroscopy
Inductively Coupled Plasma-Optical Emission Spectroscopy karthi keyan
 
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptSachin G
 
ATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYRaju Sanghvi
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopyJyotiMhoprekar
 
Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...UMT Lahore
 
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.Dr. UMESH KUMAR SHARMA
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopySeenam Iftikhar
 
Auger Electron Spectroscopy
Auger Electron SpectroscopyAuger Electron Spectroscopy
Auger Electron SpectroscopyMANISHSAHU106
 
Lecture 06; atomization by Dr. Salma Amir
Lecture 06; atomization by Dr. Salma AmirLecture 06; atomization by Dr. Salma Amir
Lecture 06; atomization by Dr. Salma Amirsalmaamir2
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy AyeshaRasty
 
Chemical ionization
Chemical ionizationChemical ionization
Chemical ionizationAarif Khan
 
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYINDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYParimi Anuradha
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopymanishasoral
 
NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...
NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...
NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...AYESHA NAZEER
 

What's hot (20)

Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopy
 
Raman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumarRaman spectroscopy by nitish kumar
Raman spectroscopy by nitish kumar
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Inductively Coupled Plasma-Optical Emission Spectroscopy
Inductively Coupled Plasma-Optical Emission Spectroscopy Inductively Coupled Plasma-Optical Emission Spectroscopy
Inductively Coupled Plasma-Optical Emission Spectroscopy
 
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy ppt
 
Advantages of electrothremal atomizer over flame
Advantages of electrothremal atomizer over flameAdvantages of electrothremal atomizer over flame
Advantages of electrothremal atomizer over flame
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
ATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRY
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
 
Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...
 
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Auger Electron Spectroscopy
Auger Electron SpectroscopyAuger Electron Spectroscopy
Auger Electron Spectroscopy
 
Lecture 06; atomization by Dr. Salma Amir
Lecture 06; atomization by Dr. Salma AmirLecture 06; atomization by Dr. Salma Amir
Lecture 06; atomization by Dr. Salma Amir
 
Atomic spectroscopy
Atomic spectroscopy Atomic spectroscopy
Atomic spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Chemical ionization
Chemical ionizationChemical ionization
Chemical ionization
 
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYINDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...
NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...
NMR spectroscopy- Spin-lattice & spin-spin relaxation, signal splitting & sig...
 

Similar to Lecture 08; spectral interferences and background absorption by Dr. Salma Amir

Atomic Absorption spectroscopy.pptx
Atomic Absorption spectroscopy.pptxAtomic Absorption spectroscopy.pptx
Atomic Absorption spectroscopy.pptxNeetuSoni21
 
UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applicationsNAFIHHASSANMEHABANI
 
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptxPHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptxdbhansali57
 
Instru booklet (1)
Instru booklet (1)Instru booklet (1)
Instru booklet (1)Zainab&Sons
 
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...Dario Scotto
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible SpectroscopyAshish Roge
 
Atomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptAtomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptZORAIZ HAIDER
 
Atomic Absorption Techniques & Applications
Atomic Absorption  Techniques & ApplicationsAtomic Absorption  Techniques & Applications
Atomic Absorption Techniques & ApplicationsGamal Abdel Hamid
 
Basics of Infrared Spectroscopy : Theory, principles and applications
Basics of Infrared Spectroscopy : Theory, principles and applicationsBasics of Infrared Spectroscopy : Theory, principles and applications
Basics of Infrared Spectroscopy : Theory, principles and applicationsHemant Khandoliya
 
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAhnaf maznun
 
Uv-visible spectroscopy
Uv-visible spectroscopyUv-visible spectroscopy
Uv-visible spectroscopyMzgin Mohammed
 

Similar to Lecture 08; spectral interferences and background absorption by Dr. Salma Amir (20)

Atomic Absorption spectroscopy.pptx
Atomic Absorption spectroscopy.pptxAtomic Absorption spectroscopy.pptx
Atomic Absorption spectroscopy.pptx
 
UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applications
 
UV Visible Spectrometry
UV Visible SpectrometryUV Visible Spectrometry
UV Visible Spectrometry
 
APA seminar
APA seminarAPA seminar
APA seminar
 
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptxPHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
 
Instru booklet (1)
Instru booklet (1)Instru booklet (1)
Instru booklet (1)
 
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible Spectroscopy
 
AAS.pptx
AAS.pptxAAS.pptx
AAS.pptx
 
Atomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptAtomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.ppt
 
Infrared / IR spectroscopy
Infrared / IR spectroscopyInfrared / IR spectroscopy
Infrared / IR spectroscopy
 
AAS interference.pptx
AAS interference.pptxAAS interference.pptx
AAS interference.pptx
 
Radiology.pptx
Radiology.pptxRadiology.pptx
Radiology.pptx
 
Atomic Absorption Techniques & Applications
Atomic Absorption  Techniques & ApplicationsAtomic Absorption  Techniques & Applications
Atomic Absorption Techniques & Applications
 
Uv spectroscopy sybsc
Uv spectroscopy sybscUv spectroscopy sybsc
Uv spectroscopy sybsc
 
Instru booklet
Instru bookletInstru booklet
Instru booklet
 
Basics of Infrared Spectroscopy : Theory, principles and applications
Basics of Infrared Spectroscopy : Theory, principles and applicationsBasics of Infrared Spectroscopy : Theory, principles and applications
Basics of Infrared Spectroscopy : Theory, principles and applications
 
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptx
 
Uv-visible spectroscopy
Uv-visible spectroscopyUv-visible spectroscopy
Uv-visible spectroscopy
 
Prabhakar singh ii sem-paper v-colorimeter &amp; spectrophotometer
Prabhakar singh  ii sem-paper v-colorimeter &amp; spectrophotometerPrabhakar singh  ii sem-paper v-colorimeter &amp; spectrophotometer
Prabhakar singh ii sem-paper v-colorimeter &amp; spectrophotometer
 

More from salmaamir2

Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...salmaamir2
 
Lecture 04; spectral lines and broadening by Dr. Salma Amir
Lecture 04; spectral lines and broadening by Dr. Salma AmirLecture 04; spectral lines and broadening by Dr. Salma Amir
Lecture 04; spectral lines and broadening by Dr. Salma Amirsalmaamir2
 
Lecture 03; Boltzmann equation by Dr. Salma Amir
Lecture 03; Boltzmann equation by Dr. Salma AmirLecture 03; Boltzmann equation by Dr. Salma Amir
Lecture 03; Boltzmann equation by Dr. Salma Amirsalmaamir2
 
Lecture 02.; spectroscopic notations by Dr. Salma Amir
Lecture 02.; spectroscopic notations by Dr. Salma AmirLecture 02.; spectroscopic notations by Dr. Salma Amir
Lecture 02.; spectroscopic notations by Dr. Salma Amirsalmaamir2
 
Lecture 01; atomic spectroscopy by Dr. Salma Amir
Lecture 01; atomic spectroscopy by Dr. Salma AmirLecture 01; atomic spectroscopy by Dr. Salma Amir
Lecture 01; atomic spectroscopy by Dr. Salma Amirsalmaamir2
 
Transition metal polymers by Dr. Salma Amir
Transition metal polymers by Dr. Salma AmirTransition metal polymers by Dr. Salma Amir
Transition metal polymers by Dr. Salma Amirsalmaamir2
 
Polyphosphazenes... preparation and properties by Dr. Salma Amir
Polyphosphazenes... preparation and properties by Dr. Salma AmirPolyphosphazenes... preparation and properties by Dr. Salma Amir
Polyphosphazenes... preparation and properties by Dr. Salma Amirsalmaamir2
 
Polysiloxanes,... preparation and properties by Dr. Salma Amir
Polysiloxanes,... preparation and properties by Dr. Salma AmirPolysiloxanes,... preparation and properties by Dr. Salma Amir
Polysiloxanes,... preparation and properties by Dr. Salma Amirsalmaamir2
 
Polymerization process or synthesis by Dr. Salma Amir
Polymerization process or synthesis by Dr. Salma AmirPolymerization process or synthesis by Dr. Salma Amir
Polymerization process or synthesis by Dr. Salma Amirsalmaamir2
 
Classification of inorganic polymers by Dr. Salma Amir
Classification of inorganic polymers by Dr. Salma AmirClassification of inorganic polymers by Dr. Salma Amir
Classification of inorganic polymers by Dr. Salma Amirsalmaamir2
 
Introduction to inorganic polymers by Dr. Salma AMIR
Introduction to inorganic polymers by Dr. Salma AMIRIntroduction to inorganic polymers by Dr. Salma AMIR
Introduction to inorganic polymers by Dr. Salma AMIRsalmaamir2
 

More from salmaamir2 (11)

Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
 
Lecture 04; spectral lines and broadening by Dr. Salma Amir
Lecture 04; spectral lines and broadening by Dr. Salma AmirLecture 04; spectral lines and broadening by Dr. Salma Amir
Lecture 04; spectral lines and broadening by Dr. Salma Amir
 
Lecture 03; Boltzmann equation by Dr. Salma Amir
Lecture 03; Boltzmann equation by Dr. Salma AmirLecture 03; Boltzmann equation by Dr. Salma Amir
Lecture 03; Boltzmann equation by Dr. Salma Amir
 
Lecture 02.; spectroscopic notations by Dr. Salma Amir
Lecture 02.; spectroscopic notations by Dr. Salma AmirLecture 02.; spectroscopic notations by Dr. Salma Amir
Lecture 02.; spectroscopic notations by Dr. Salma Amir
 
Lecture 01; atomic spectroscopy by Dr. Salma Amir
Lecture 01; atomic spectroscopy by Dr. Salma AmirLecture 01; atomic spectroscopy by Dr. Salma Amir
Lecture 01; atomic spectroscopy by Dr. Salma Amir
 
Transition metal polymers by Dr. Salma Amir
Transition metal polymers by Dr. Salma AmirTransition metal polymers by Dr. Salma Amir
Transition metal polymers by Dr. Salma Amir
 
Polyphosphazenes... preparation and properties by Dr. Salma Amir
Polyphosphazenes... preparation and properties by Dr. Salma AmirPolyphosphazenes... preparation and properties by Dr. Salma Amir
Polyphosphazenes... preparation and properties by Dr. Salma Amir
 
Polysiloxanes,... preparation and properties by Dr. Salma Amir
Polysiloxanes,... preparation and properties by Dr. Salma AmirPolysiloxanes,... preparation and properties by Dr. Salma Amir
Polysiloxanes,... preparation and properties by Dr. Salma Amir
 
Polymerization process or synthesis by Dr. Salma Amir
Polymerization process or synthesis by Dr. Salma AmirPolymerization process or synthesis by Dr. Salma Amir
Polymerization process or synthesis by Dr. Salma Amir
 
Classification of inorganic polymers by Dr. Salma Amir
Classification of inorganic polymers by Dr. Salma AmirClassification of inorganic polymers by Dr. Salma Amir
Classification of inorganic polymers by Dr. Salma Amir
 
Introduction to inorganic polymers by Dr. Salma AMIR
Introduction to inorganic polymers by Dr. Salma AMIRIntroduction to inorganic polymers by Dr. Salma AMIR
Introduction to inorganic polymers by Dr. Salma AMIR
 

Recently uploaded

Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfagholdier
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...Sapna Thakur
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfchloefrazer622
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxVishalSingh1417
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room servicediscovermytutordmt
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Disha Kariya
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introductionMaksud Ahmed
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactPECB
 

Recently uploaded (20)

Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdf
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptx
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room service
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 

Lecture 08; spectral interferences and background absorption by Dr. Salma Amir

  • 1. Lecture No. 08 Course title: Atomic spectroscopy Topic: Spectral Interferences, background absorption Course instructor: Dr. Salma Amir GFCW Peshawar
  • 2. Spectral Interferences Spectral interferences are those in which the measured light absorption is erroneously high due to absorption by a species other than the analyte element. Absorption due to the presence of light-absorbing molecules in the flame and light dimming due to the presence of small particles in the flame are much more common spectral interferences. Such phenomena are referred to as background absorption. The only cures for direct atomic spectral interference are (1) to choose an alternate analytical wavelength (2) to remove the interfering element from the sample
  • 3. Background Absorption  A common occurrence that results in spectral interference is absorption of the HCL radiation by molecules or polyatomic species in the atomizer. This is called “background absorption”; it occurs because not all of the molecules from the sample matrix and the flame gases are completely atomized. This type of interference is more commonplace at short wavelengths (<250 nm) where many compounds absorb.  Since atoms have extremely narrow absorption lines, there are few problems involving interferences where one element absorbs at the wavelength of another. Even when an absorbing wavelength of another element falls within the spectral bandwidth used, no absorption can occur unless the light source produces light at that wavelength, i.e., that element is also present in the light source. However, undissociated molecular forms of matrix materials may have broadband absorption spectra, and tiny solid particles in the flame may scatter light over a wide wavelength region. When this type of nonspecific absorption overlaps the atomic absorption wavelength of the analyte, background absorption occurs.
  • 4. Causes of Background Absorption  Incomplete combustion of organic molecules in the atomizer can cause serious background interference problems.  If a flame atomizer is used, incomplete combustion of the solvent may take place, particularly if the flame is too reducing (fuel rich). The extent of the interference depends on flame conditions (reducing or oxidizing), the flame temperature used, the solvent used, and the sample aspiration rate.  Background interference is much more severe when graphite furnace atomizers are used because the pyrolysis step is limited to a maximum temperature that does not volatilize the analyte. Consequently, many matrix molecules are not thermally decomposed. They then volatilize into the light path as the higher atomization temperature is reached and absorb significant amounts of the source radiation.
  • 5. Background absorption correction  To compensate for this problem, the background absorption must be measured and subtracted from the total measured absorption to determine the true atomic absorption component. Various Background correction techniques employed are  1. Two line background correction  2. Continuum source background correction  3. Zeeman background correction  4. Smith–Hieftje Background Corrector
  • 6. 1. Two-Line Background Correction An early manual method of measuring background and applying a correction for it used the absorption of a nearby nonresonance line. The emission spectrum from a hollow cathode is quite rich in emission lines and contains the resonance lines of the element of interest plus many other nonresonance emission lines from the element and the filler gas. The analyte atoms do not absorb these nonresonance lines; however, the broad molecular background absorbs them. Two measurements are made.  The absorbance at a nonresonance line close to the analyte resonance line but far enough away that atomic absorption does not occur is measured.  Then the absorbance at the resonance line of the analyte is measured.
  • 7. The absorbance at the nonresonance line is due only to background absorption. The absorbance at the resonance line is due to both atomic and background absorption. The difference in absorbance between the resonance and nonresonance lines is the net atomic absorption. SUMMARY: Resonance line= analyte + background Non-resonance line= background Net absorption due to analyte = Resonance - Non-resonance
  • 8. 2. Continuum source background correction  Continuum source background correction is a technique for automatically measuring and compensating for any background component which might be present in an atomic absorption measurement. This method incorporates a continuum light source in a modified optical system  In the continuum source background correction method, when an HCL source is used, the absorption measured is the total of the atomic and background absorptions. When a continuum lamp source is used, only the background absorption is measured. The continuum lamp, which emits light over a range or band of wavelengths, is placed into the spectrometer system as shown in Figure 3.7. This setup allows radiation from both the HCL and the continuum lamp to follow the same path to reach the detector. The detector observes each source alternately in time, either through the use of mirrored choppers or through pulsing of the lamp currents.
  • 9.
  • 10. When the HCL lamp is in position as the source, the emission line from the HCL is only about 0.002 nm wide; in other words, it fills only about 1% of the spectral window as shown in the upper left-hand-side drawing in Figure 3.8. When the HCL radiation passes through the flame, both the free atoms at the resonance line and the broadband absorbing molecules will absorb the line. This results in significant attenuation of the light reaching the detector, as shown in the upper center and right-hand-side drawing in Figure 3.8. When the continuum source is in place, the continuum lamp emission fills the entire spectral window as shown in the lower left-hand-side drawing in Figure 3.8. Any absorption of the radiation from the continuum lamp observed is broadband background absorption, since it will be absorbed over the entire 0.2 nm as seen in the lower right-hand-side drawing in Figure 3.8
  • 11.  The absorbance of the continuum source is therefore an accurate measure of background absorption. An advantage of this method is that the background is measured at the same nominal wavelength as the resonance line, resulting in more accurate correction.
  • 12. 3. Zeeman Background correction  Atomic absorption lines occur at discrete wavelengths because the transition that gives rise to the absorption is between two discrete energy levels. However, when a vapor-phase atom is placed in a strong magnetic field, the electronic energy levels split. This gives rise to several absorption lines for a given transition in place of the single absorption line in the absence of a magnetic field. This occurs in all atomic spectra and is called Zeeman splitting or the Zeeman effect.  In the simplest case, the Zeeman effect splits an absorption line into two components. The first component is the π-component, at the same wavelength as before (unshifted); the second is the σ-component, which undergoes both a positive and negative shift in wavelength, resulting in two equally spaced lines on either side of the original line. This splitting pattern is presented in Figure 6.27. The splitting results in lines that are separated by approximately 0.01 nm or less depending on the field strength. The strength of the magnetic field used is between 7 and 15 kG. Background absorption and scatter are usually not affected by a magnetic field.
  • 13. The π- and σ-components respond differently to polarized light. The π- component absorbs light polarized in the direction parallel to the magnetic field. The σ-components absorb only radiation polarized 90° to the applied field. The combination of splitting and polarization differences can be used to measured total absorbance (atomic plus background) and background only, permitting the net atomic absorption to be determined.
  • 14. A permanent magnet can be placed around the furnace to split the energy levels. A rotating polarizer is used in front of the HCL or EDL. During that portion of the cycle when the light is polarized parallel to the magnetic field, both atomic and background absorptions occur. No atomic absorption occurs when the light is polarized perpendicular to the field, but background absorption still occurs. The difference between the two is the net atomic absorption. Such a system is a DC Zeeman correction system.
  • 15.  Alternately, a fixed polarizer can be placed in front of the light source and an electromagnet can be placed around the furnace. By making absorption measurements with the magnetic field off (atomic plus background) and with the magnetic field on (background only), the net atomic absorption signal can be determined. This is a transverse AC Zeeman correction system. The AC electromagnet can also be oriented so that the field is along the light path (a longitudinal AC Zeeman system) rather than across the light path. No polarizer is required in a longitudinal AC Zeeman system.  Zeeman correction can also be achieved by having an alternating magnetic field surround the hollow cathode, causing the emission line to be split and then not split as the field is turned on and off. By tuning the amplifier to this frequency, it is possible to discriminate between the split and unsplit radiation. A major difficulty with the technique is that the magnetic field used to generate Zeeman splitting also interacts with the ions in the hollow cathode. This causes the emission from the hollow cathode to be erratic, which in turn introduces imprecision into the measurement. Most commercial instruments with Zeeman background correctors put the magnet around the atomizer.
  • 16. 4. Smith–Hieftje Background Corrector  It will be remembered that the HCL functions by the creation of excited atoms that radiate at the desired resonance wavelengths. After radiating, the atoms form a cloud of neutral atoms that, unless removed, will absorb resonance radiation from other emitting atoms. If the HCL is run at a high current, an abundance of free atoms form. These free atoms absorb at precisely the resonance lines the hollow cathode is intended to emit; an example is shown in Figure 6.28, with the free atoms absorbing exactly at λ over an extremely narrow bandwidth (Figure 6.28b). The result is that the line emitted from the HCL is as depicted in Figure 6.28c instead of the desired emission line depicted in Figure 6.28a. The phenomenon of absorption of the central portion of the emission line by free atoms in the lamp is called “self-reversal.” Such absorption is not easily detectable, because it is at the very center of the emitted resonance line and very difficult to resolve.
  • 17. It is, of course, exactly the radiation that is most easily absorbed by the atoms of the sample. In practice, if the HCL is operated at too high a current, the self-reversal decreases the sensitivity of the analysis by removing absorbable light. It also shortens the life of the lamp significantly. Figure 6.28 distortion of spectral line shape in an HCL due to self-reversal. (a) shape of the spectral line emitted by the HCL. (b) shape of the spectral energy band absorbed by cool atoms inside the HCL. (c) shape of the net signal emerging from an HCL, showing self-absorption of the center of the emission band.
  • 18.  The Smith–Hieftje background corrector has taken advantage of this self- reversal phenomenon by pulsing the lamp, alternating between high current and low current. At low current, a normal resonance line is emitted and the sample undergoes normal atomic absorption. When the HCL is pulsed to a high current, the center of the emission line is self-absorbed, leaving only the wings of the emitted resonance line; the emitted line would look similar to Figure 6.28c. The atoms of the sample do not significantly absorb such a line. The broad background, however, absorbs the wings of the line. Consequently, the absorption of the wings is a direct measurement of the background absorption at the wavelength of the atomic resonance absorption. The Smith–Hieftje technique can be used to automatically correct for background. In practice, the low current is run for a fairly short period of time and the absorbance measured. The high current is run for a very short, sharp burst, liberating intense emission and free atoms inside the hollow cathode and the absorbance is again measured. The difference between the measurements is the net atomic absorption. There is then a delay time to disperse the free atoms within the lamp before the cycle is started again.