Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA
HEAVY CRUDE OIL BY INDUCTIVELY COUPLED PLASMA MASS
SPECTROMETRY
Henry Wuyke*, LlinaberFeo, Miguel Murillo
*PDVSA-INTEVEP

S.A. Urbanizacion Santa Rosa. Sector El Tambor. Los Teques. Estado Miranda. Apdo.
76343, Caracas 1070A, Venezuela

Centro de QuĂ­micaAnalĂ­tica, Facultad de Ciencias, Universidad Central de Venezuela, Apdo.47102,
Caracas 1070A, Venezuela.

ABSTRACT
The Laser-Extra Heavy Crude Oil interaction was studied using a nanosecond Nd:YAG 213 nm laser
coupled to an ICP-MS. For this study, ~2.7 g Extra Heavy Crude Oil was mixed with ~3 mL xylene and 0,3 g
organic multielemental solution containing 100 µg g-1 of Ag, Al, Ba, Cd, Cr, Cu, Mg, Mn, Mo, Ni, Pb, Sn, Ti, V, and
Zn, xylene was left to evaporate at 90 °C and multielemental samples were obtained and analyzed by LA-ICPMS. Laser parameters, such as, ablation gas flow, sample scanning speed, spot size, fluence and focus position
were optimized and RSD (n = 8) from 3% to 5% for the net signal of 63Cu, 95Mo, 60Ni and 66Zn and from 0.9%
to 3.3% for the ratios Ni/Mo and Cu/Zn were obtained. Elemental fractionation was observed with both Zn
and Cd when fluence was varied. Nevertheless, evidence is not enough to explain results as a consequence of
high volatilityor highionization potential (PI)of these elements. Ablationcratersobtained showsigns ofmore
pronounced thermal processesat higher fluences. Elemental fractionation was also observed in the ICP due to
mass load effect. Isotopic ratios

95Mo/112Cd, 63Cu/66Zn

y

27Al/24Mg,

which have different ionizationpotential

(1.89; 1.68 and 1.66 eV, respectively expressed as PIdenominator - PInumerator) increase when mass load is increase.
On the other hand,
plasma

60Ni/63Cu

ionization

ratio (0.09 eV difference) remains constant. These results indicate that the

temperature

decreases

when

mass

load

is

increased

affecting

thoseelementswithhighionization potential.
Keywords:Laser Ablation, ICP-MS, Extra Heavy Crude Oil, Elemental Fractionation.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

367
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

INTRODUCTION
The use of LA-ICP-MS to analyze crude oil maybe advantageoustoeliminatethe extensivesample
treatmentstepsoftraditional methods such as calcination or acid digestion. However, few works using LA-ICPMS for direct elemental analysis in crude oil have been reported [1-3]. It can be probably due to difficulties
found during ablation, such as splashing, of this type of liquid samples. LA-ICP-MS is more suitable forsolid
samples [4-7]. Therefore, it is a potential tool for direct analysis of semi-solid extra heavy crude oil with less
trouble inthe ablation process.
The ablation processis strongly dependent onthe propertiesofthe sample such as composition,
thermal diffusivity and absorptivity [4].Therefore, matrix matched calibration standards or reference
certificated materials are required to ensure accuracyof results. However, calibration standards are not
available for the majority of sample matrixes, which limitsthe applicationsofLA-ICP-MS [8,9]. Therefore, it is
important to know how the laser ablation of extra heavy crude oil is affected by eachexperimental parameter
and identify which of them lead rise to fractionation or non-stoichiometric sampling.
Pulse length isoneofthe most important parametersinvolvedinmechanisms by whichthe ablation
laser processoccurs. Lasers with pulse length in nanosecond (ns) and femtosecond (fs) are the
mostfrequentlyused in LA-ICP-MS applications. Generally, fs lasers produce better results because the time
interval of energy delivering is much smaller than the time interval needed for the thermal stabilization of the
sample (intheorder of picoseconds) [6,10]. Furthermore, photon intensitiesreach valuesof >1014 W cm-2,
thereby processesare predominantlynon-thermal, causingless collateral damage, less dependence onthe
matrix, smaller particlesand nonlaser-plasma interaction [6]. Nevertheless, even thoughthe latest
developmentsare aimed to take advantageoffslasers, todaynslasers are still very populardue primarily totheir
low cost.
Fluence also plays an important role in ablation mechanisms. It must be above a threshold value to
lead the ablation process [10]. When ns lasers are used, fluence should not be too high to avoid the generation
of a high electron density plasma, because it increases the laser-plasma interaction and therefore, reduces the
interaction between the laser and the sample [11]. When fs lasers are used, fractionation appears to be only
near the threshold value of the sample and reach an asymptotic value as the fluence increases [9,12-14].
Fractionationaccording to the fluencein the rangeof2 to 42J cm-2with afslaserin brass, aluminumand
glass samples was assessed by Bian et al. [12].In those conditions, the ratio Zn/Cu in the brass and aluminum
samples was independent of the fluence while for glass samples there was a strong dependency especially for
the most transparent glass sample (NIST 610). However, at higher fluence, the ratio Zn/Cu reached an
expected asymptotic value, considering the results of brass and aluminum measures. Similar results on the
same glass sample type were obtained by Koch et al. [13]. Fractionation was observed with ratios Zn/Cu and

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

368
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381
Mo/Ni which have different ionization potentials but not with the ratio U/Th which have similar ionization
potentials (PI: U: 6.2 eV, Th: 6.3 eV y Pb: 7.5 eV), which evidence of that the ionization potential affects the
ablation process. It has been mentioned that Coulomb interactions are responsible for this behavior [11].
However, ablationmechanismsarestillnotfully understoodto the date [15]. There are other factors such as the
volatility of the elements that influence the signals when thermal processes occurred, i.e.
morevolatileelements arevaporizedpreferentially,resulting in greatersignal and causingfractionation [12,16].
Fractionationhasalso beenstudiedas a function oftime, i.e. shot to shot. Garcia et al. [9] studied this
effect using afs laser in binarymetallic, semiconductorsand glasses samples. They obtainedthatthe
Cu/Znchangesfrom shot to shotreachedan asymptotic valueindependentof fluence,which correspondsto
astoichiometricsampling. Furthermore, thisasymptotic valueis reachedfaster ifhigh fluenceisapplied.
There are other parameters that also affect the ablation process such as ablation gas and laser
wavelength. Helium and argon are the most used gases for ablation environmental and sample
transportation. However, as argon has a lower ionization potential than helium, there is a greater likelihood
of plasma formation and shielding effect that prevent an efficient interaction between laser and sample, so
that helium is preferred [11]. Wälle et al. [17]using ns and fs lasers in glass, zircon and silicon samples,
observed a depression of the detection efficiency by a factor of 5 when used argon instead of helium as
ablation gas. Studies have shown that using shorter wavelengths, with both ns and fs lasers in glass samples
fractionation effects are reduced [13,18,19].
Particles generated in the ablation process can have different compositions in terms of size,
especially if a ns laser is used [6]. It creates fractionation if there is no transport efficiency, because large
particles (> 1 µm) tend to be lost by gravity or deposited on chamber ablation. While, small particles (<10
nm) are lost by diffusion or electrostatic effects [12].The ideal condition is that all particles are fully
decomposed when reaching the ICP in order to to avoid affecting the characteristics of plasma. Nevertheless,
it has been found that mass load that reaches the ICP can create a matrix effect dependent on the ionization
potential [13] or on the melting point of the elements [20].
Although LA-ICP-MS has been used for elemental analysis in crude oil [1,3] a systematic study that
assess the fractionation of these types of samples, as is usually done on metal alloys or glass samples, has not
been performed. The aim of this study is to evaluate the behavior of laser - extra heavy crude oil interaction
by LA-ICP-MS as a function of various experimental parameters using a ns Nd: YAG laser at 213nm.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

369
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

MATERIALS AND METHODS
Instrumentation:
Laser ablation was performed using a Nd:YAG UP-213 LA New Wave Research operated at 213 nm
with a pulsed duration of ~ 4 ns. An ICP-MS Agilent Technologies 7500ce that consist in a plasma source,
aoctopole reaction cell and a quadrupole mass analyzer was used. Laser ablation and ICP-MS were connected
using Tygon tubes (S-50-HL Class VI, ~ 3 mm i.d.) as shown in Fig. 1. In analysis, the aerosol from the ablation
cell is mixed with Ar using a "Y" connection, then goes to a cyclone chamber of ~ 4 cm diameter and finally to
the ICP-MS. The bypass and purge valves are used for analysis preconditioning.

Figure 1: Coupling laser ablation to ICP-MS

Reagents, stardards and samples:
Extra heavy crude oil samples (API° < 10) were provided by PDVSA-INTEVEP. Xylene (puriss. p.a.)
Sigma-Aldrich and an organic multielemental solution containing 100 µg g-1 of Ag, Al, Ba, Cd, Cr, Cu, Mg, Mn,
Mo, Ni, Pb, Sn, Ti, V and Zn, S-21 Conostan, USA were used. The tuning solution to perform the ICP-MS was
preparedfrom ARMTM Absolute Standards, INC.1000 mg/Kg of Li, Mg, Co, Y, Ce and Pb in 2% HNO3.

Sample preparation:
Approximately 2.7 g of oil was dissolved in a beaker with 2 mL xylene and 0.3 g of organic
multielemental solution, the mix was homogenized by magnetic stirring and brought to an oven at 90 ° C for 2
hours until evaporation of xylene. Finally, part of the oil was placed in a sample holder tray of 1 cm diameter
and 3 mm high and led to LA-ICP-MS. The resulting mixture contained a concentration of 10 µg g -1 in metals
from the organic multielemental solution.

General conditions of analysis:
In this work, several experiments were performed in order to study the laser - extra heavy crude oil
interaction by ICP-MS. Table 1 shows the general conditions of analysis. Isotopesof each elementwere chosen

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

370
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381
tominimizeisobaricandpolyatomicinterferences.
The ICP-MS parameters were those that resulted in good sensitivity for all elements of the tuning
solution, covering a wide tuning range of masses: 7Li,

24Mg, 59Co, 89Y, 140Ce

and

208Pb.

Once sample is

introduced into the ablation cell, the system is purged with He 1 L min-1 flow for 5 min.
ICP-MS
Plasma gas flow, L

Agilent 7500ce
min-1

15

Foward power, W

1500

Sample depth, mm

6.5

Integration timeper point, s

0.1

Integration timeper mass, s

0.3 (3 point per mass)

Laser ablation

New Wave Research UP-213 LA

Ablation gas flow (He), L min-1

0.75

Mixing gas flow (Ar), L min-1

0.5

Fluence, J cm-2

~ 34 (100 %)

Focus, mm

4 (above the sample)

Spot size, µm

100 (1000 µm to 4 mm of defocus)

Pulse frequency, Hz

20

Scanning speed, µm s-1 (linear)

20

Table 1:General conditions of analysis

RESULTS AND DISCUSSION
A typical ablation profile of Ni, Cu, Zn and Mo is shown in Fig. 2. It is observed that the signal of Ni is
much greater than the signals of Cu, Zn and Mo. Thisis becauseNi isoneofthe major elementsinthecrude oil.
However, there is good signal stability for the four elements at different concentration levels. Therefore, Cu
and / or Zn signals were used to shown the optimization of the laser parameters and were selected for their
differences in ionization potential (PI) and melting point (mp) (PI/mp: Cu 7.72 eV / 1083 °C, Zn 9.42 eV / 419
°C).

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

371
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

Figure 2: Ablation profile of 60Ni, 63Cu, 66Zn and 95Mo at general analysis conditions. Ablation time was 50 s
with 10 s of delay relative to the acquisition start
Optimization of the laser parameters was performed with the aim of achieving the greatest intensity
for

60Ni, 63Cu, 66Zn

and

95Mo

isotopes while maintaining good repeatability. In experiments with He-Ar as

ablation gas and mixing gas respectively, flow rates were optimized between 0.4 – 0.9 L min-1 and the best
results were obtained with 0.75 L min-1 He and 0.5 L min-1 Ar. In the Ar-Ar system, flow rates were optimized
between 0 to 1.2 L / min, and the best results were obtained with 1.0 L min -1Ar (ablation gas) and 0.2 L min1Ar

(mixing gas).
Both He and Ar as ablation gases were compared in terms of the transient signal produced. The

results are shown in Fig. 3. Figures 3a and 3c show that

63Cu

and 66Zn signals are greater when He is used as

ablation gas. Furthermore, a more pronounced memory effect is observed when Ar is used as ablation gas
(figures 3b and 3d).
The damage on crude oil surface is greater when Ar is used in the ablation cell (figure 4) at both
optimum analysis conditions (4 mm of defocus) and focus. These results are consistent with previous studies
[11,17], where it is argued that Ar, which posses an ionization potential (15.7 eV) lower than He (24.6 eV),
promotes the formation of plasma on the surface, which interacts with the sample and deform it, i.e. there
aresignificantthermal processes.
The memory effect produced when Ar is used can be explained by the expansion to large volume of
the ablated material into the ablation cell. It makes that the system takes longer to transport this volume. It
has been mentioned that this expansion is more pronounced when the ionization potential is decreased [11].
It agrees with the results obtained in this study.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

372
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381
He was chosen as ablation gas for the subsequent experiments because it produces a significantly
high signal for Cu, Zn, Ni and Mo (Student t test with α = 0.05) and an improved precision to the Cu/Zn and
Ni/Mo ratios (Fisher F test α = 0.05) when compared to the obtained with Ar as ablation gas. Results are
shown in table 2.

Figure 3: Signal versus time for two ablation gases, He and Ar: a and b) CPS of 63Cu, c and d) CPS of 66Zn. a and
c) approximately at 10 s, the laser was switched on 7s (140 pulses), b and d) zoom between times 30 and 90 s

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

373
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

Figure 4:Picturesof cratersat different ablation conditions. It is indicated in order: ablation gas, pulses
number and defocus level (mm): a) He, 1000 pulses, 4 mm, b) Ar, 1000 pulses, 4 mm, c) He, 400 pulses, 0 mm
(in focus), d) Ar, 100 pulses, 0 mm (in focus). The mark corresponds to 100 µm. He and Ar flow in the ablation
cell were 0.75 L min-1

Element o
ratio

Signal
Average
(CPS). He

SD

(CPS).

He

Signal
CV (%). He

Average
(CPS). Ar

SD

(CPS).

Ar

CV (%). Ar

Cu

8.1E5

0.3E5

3.7

4.6E5

0.3E5

6.5

Zn

4.6E5

0.2E5

4.3

2.0E5

0.1E5

5.0

Ni

4.2E6

0.1E6

2.4

2.4E6

0.2E6

8.3

Mo

6.1E5

0.2E5

3.3

3.5E5

0.3E5

8.6

Cu/Zn

1.754

0.058

3.3

2.15

0.12

5.6

Ni/Mo

6.990

0.061

0.9

6.99

0.22

3.1

Table 2: Media, standarddeviation (SD), coefficientofvariation (CV) and samplesize (n) for Cu, Zn, Ni and Mo
signals and Cu/Zn and Ni/Mo ratios using He or Ar as ablation gas
The effect of two cyclone chambers with different sizes (~ 4 cm and ~6 cm internal diameter) was
evaluated. Figure 5 shows that per run, the signal is more stable when a cyclone chamber is used, because it
seems to dampen the effect produced in the ablation cell. Among ~ 4 cm and ~ 6 cm chambers no great
differences were obtained. Therefore, ~ 4 cm chamber was taken as condition routine to make smaller the
volume of the transport region.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

374
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

Figure 5:63Cu signal versus time with two cyclone chambers and without them. Ablation time was 50 s with
10 s of delay with 10 s delay relative to the acquisition start
In most of the works reported on solid samples, it is preferred ablate with a scanning speed different
to zero, because at the same point, it begins to form an increasingly deep crater with each impact of the laser.
The efficiency of energy transfer from the laser to the sample is continuously decreased and fractionation is
observed. With a scanning speed different to zero, ablation is carried out over a greater area, which ensures a
stoichiometric sampling.

Figure 6:63Cu signal versus time for the four scanning speeds indicated. Four replicates are shown. Ablation
time was 50 s with 10 s of delay relative to the acquisition start

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

375
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381
Four linear scanning speeds 0, 20, 40 and 60 µm s -1 were tested in this work. Fig. 6 shows the
ablation profiles of the four scanning speeds with four repetitions each one. At 0 µm s -1 reproducible profiles
with a signal drop from the beginning to the end of the run between 10 and 15% are observed. However, Fig.
7 shows that there is no fractionation of

63Cu/66Zn

ratio at 0 µm s-1. Profiles at speeds 20 and 40 µm s-1 are

little less stable and reproducible than at 0 µm s-1, but without signal drop. The profile at 60 µm s-1 is more
random probably for the larger sample area covering, which is not on the same horizontal axis (the crude oil
sample on holder may be concave or convex and it is not easy to reproduce a flat surface). Besides from the
beginning to the end of the run, its vertical position has changed, varying the level of defocus and so the
sensitivity. Even though, the

63Cu/66Zn

ratio remained stable through time at four scanning speeds studied

(see Fig. 7). Finally, it canbe established thatthe most appropriate conditions fortheanalysisare
atsmallscanning speeds~ 20µm s-1.

Figure 7:63Cu/66Zn ratio versus time at four scanning speeds. Signals were recorded in the same experiment
of Fig. 6
The fractionation effect versus fluence was evaluated with 15 elements by varying of defucus level
which allows to work in a greater range instead of varying the energy to a fixed diameter. As can be seen in
Fig. 8, the greater signal was obtained at higher positive defocus levels. Although fluence becomes smaller, it
is still sufficient to ablate a larger area the crude oil sample, which explains the results.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

376
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

Figure 8:60Ni, 63Cu, 66Zn and 95Mo signals versus defocus level. Each point represents the average of three 30
s runs of signal acquisition
Figure 9 shows the relationship between the elements studied and

66Zn

(Zn was chosen for its high

ionization potential and high volatility) versus fluence. In general, fractionation was observed, i.e. M/ 66Zn
ratio increases with increasing fluence, except for Cd that appears to be stable, since it has very similar
characteristics to Zn. Zn and Cd that present this fractionation with respect to the other elements, have high
ionization potential and high volatility.
The observed behavior is hard to explain because ablation mechanisms are not yet fully known at the
date [15]. However, trends indicate thatwhentheablationis carried outbythermal processes, more
volatileelementsarepreferentiallyvolatilized [6]. Recent results[13] suggest thatelementswithlow ionization
potentialare preferentiallyableatedand coulomb interactions are responsible of that [11].
The results shown in Fig. 9 indicate that at lower fluences, Zn signal is favored compared to the other
elements (except for Cd), contrary effect to that observed in previous studies on glass samples [12,13]. The
results can not be explained by volatility if it is considered that there are greater thermal processes at high
fluencesor at leastgreatersurface damageby plasma-sampleinteraction (see Fig. 4a (less fluence) and 4c
(greater fluence, greatersurface damage)). As the Zn is more volatile, it should be favored at high fluence and
that does not happen. Similarly, the Mo that is an element with high melting and boilingpoint, presents
nofractionation, i.e. volatilitydoes not appear tobe responsible. The Al, that has relatively low ionization
potential, did not present fractionation. In conclusion, Zn and Cd were the only elements that showed
fractionation as function of fluence, but without sufficient evidence to explain it by volatility or ionization
potentials.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

377
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

Figure 9: Normalized M (isotopes indicated) / 66Zn ratios versus fluence (J cm-2). Each point represents three
2 min 30 s runs average of acquisition signal at a scanning speed of 20 µm
It can be thought that the results of Fig. 9 may be influenced by mass load effect, because there is
greater quantity of particles at low fluences (greater defocus and larger diameter) than at high fluences (see
Fig. 8).
The ICP may be affected, decreasing its ionization temperature as a result of a large mass load, which
can cause different ionization profiles for elements with different ionization potentials.
To evaluate this effect with crude oil sample, the analysis with different crater diameters while
remained fixed the fluence was performed. Fig. 10 shows

63Cu/66Zn,95Mo/112Cd, 27Al/24Mg

and

60Ni/63Cu

ratios which differ inionizationpotential (expressed as PIdenominator - PInumerator)1.68; 1.89; 1.66 y 0.09
eVrespectively. It was observed that ratios tend to increase when mass load increases. This effect was more
marked for the
60Ni/63Cu

95Mo/112Cd

ratio, which has the largest difference in ionization potential. On the contrary

ratio shows no significant change, since their PI not differ greatly.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

378
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381

Figure 10: Normalized 63Cu/66Zn, 95Mo/112Cd, 27Al/24Mg and 60Ni/63Cu ratios versus crater surface. Each
point represents three 2 min 30 s runs average of acquisition signal at a scanning speed of 20 µm
At higher mass load, the ICP diminishes its ionization temperature and those elements with high
ionization potential are more affected which explains the results of Fig. 10.
If mass load had an effect on the results of Fig. 9, they hada decreasing trend (because the greater
mass load is found in the low fluenceanalyse), confirming that the results are effects mainly produced by the
ablation processes at different fluences and not by the processes occurring in the ICP.

CONCLUSIONS
Fractionation was observed as function of fluence, i.e. Zn and Cd signals were affected in different
manner, respect to the other 13 elements studied, when fluence is increased. Elements with low ionization
potential as Al or low volatility as Mo, showed no fractionation when fluence is varied. Therefore, there was
insufficient evidenceto explainthe results in termsofthe volatility andtheionization potential. Craters pictures
are evidence that ablation mechanisms are changed at different fluence (see Fig. 4). They showed greater
surface damage at higher fluence. It can be probably due to plasma-sample interaction. Also, differences
among He and Ar ablation gases were observed. The best results, in terms of sensitivity and precision were
obtained when He was used as ablation gas. Fractionation was also observed in the ICP. At higher mass load
were disadvantaged those elements with high ionization potential due to the decreased of plasma ionization
temperature. Conditions of increased sensitivity to the extra heavy crude oil were determined in the LA-ICPMS system. It was found that they correspond mainly to a larger area of ablation (larger diameter and higher
defocus level), highlaser energy and high pulse frequency. Furthermore,an improvement inthe stability of
thesignal was observedwhen acyclonicchamberwas used inthe transport system, this may promote the use of

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

379
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381
various devices in the transport system to improve analysis precision of related samples with high organic
matter content. Under performed conditions, %RSD from 3 to 5% in the net signal of Cu, Zn, Ni and Mo and
from 0.9 to 3.3% in the Ni/Mo and Cu/Zn ratios were found, respectively. The results of this study show the
possibility of performing elemental determinations in extra heavy crude oil by ns-LA-ICP-MS. However,
further experiments are needed for the purpose of increasing the sensitivity of the analytes, improve
precision, reduce mass load effects and make the method robust to small matrix changes.

REFERENCES
1.

Ricard E, PĂ©cheyran C, Sanabria G, Prinzhofer A, Donard OFX, Direct analysis of trace elements in
crude oils by high-repetition-rate femtosecond laser ablation coupled to ICPMS detection, Anal
BioanalChem, 2011; 399:2153-2165.

2.

Vorapalawut N, Pohl P, Bouyssiere B, Shiowatana J, Lobinski R, Multielement analysis of petroleum
samples by laser ablation double focusing sector field inductively coupled plasma mass spectrometry
(LA-ICP MS), J Anal At Spectrom, 2011; 26:618-622.

3.

Heilmann J, Boulyga SF, Heumann KG, Development of an isotope dilution laser ablation ICP-MS
method for multi-element determination in crude and fuel oil samples, J Anal At Spectrom, 2009;
24:385-390.

4.

Herbert CG, Johnstone RAW, Mass Spectrometry Basics. CRC Press, New York, 2003.

5.

Koch J, Gunther D, Laser Ablation ICP-MS. Encyclopedia of Spectroscopy and Spectrometry, 2nd edn,
2009;1262-1269.

6.

Fernández B, Claverie F, Pécheyran C. Donard OFX, Direct analysis of solid samples by fs-LA-ICP-MS,
Trends in Anal. Chem, 2007; 26:951-966.

7.

Hergenröder R, Laser-generated aerosols in laser ablation for inductively coupled plasma
spectrometry,SpectrochimActa Part B, 2006; 61:284-300.

8.

Hill SJ, Inductively Coupled Plasma Spectrometry and its Applications. Blackwell Publishing, Oxford,
2007.

9.

Garcia CC, Lindner H, Von Bohlen A, Vadla C, Niemax K, Elemental fractionation and stoichiometric
sampling in femtosecond laser ablation, J Anal At Spectrom, 2008; 23:470-478.

10. Miller JC, Hanglund RF, Laser Ablation and Desorption, Academic Press, San Diego, 1998.
11. Garcia CC, Lindner H, Niemax K, Laser ablation inductively coupled plasma mass spectrometry—
current shortcomings, practical suggestions for improving performance, and experiments to guide
future development, J Anal At Spectrom, 2009; 24:14-26.
12. Bian Q, Garcia CC, Koch J, Niemax K, Non-matrix matched calibration of major and minor

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

380
Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381
concentrations of Zn and Cu in brass, aluminium and silicate glass using NIR femtosecond laser
ablation inductively coupled plasma mass spectrometry, J Anal At Spectrom, 2006; 21:187-191.
13. Koch J, Wälle M, Pisonero J, Günther D, Performance characteristics of ultra-violet femtosecond laser
ablation inductively coupled plasma mass spectrometry at ~265 and ~200 nm, J Anal At Spectrom,
2006; 21:932-940.
14. Bian Q, Koch J, Lindner H, Berndt H, Hergenröder R, Niemax K, Non-matrix matched calibration using
near-IR femtosecond laser ablation inductively coupled plasma optical emission spectrometry, J Anal
At Spectrom, 2005; 20:736-740.
15. Glaus R, Kaegi R, Krumeich F, GĂĽnther D, Phenomenological studies on structure and elemental
composition of nanosecond and femtosecond laser-generated aerosols with implications on laser
ablation inductively coupled plasma mass spectrometry,SpectrochimActa Part B, 2010; 65:812-822.
16. Kosler J, Wiedenbeck M, Wirth R, Hovorka J, Sylvester P, Míková J, Chemical and phase composition of
particles produced by laser ablation of silicate glass and zircon—implications for elemental
fractionation during ICP-MS analysis, J Anal At Spectrom, 2005; 20:402-409.
17. Wälle M, Koch J, Flamigni L, Heiroth S, Lippert T, Hartung W, Günther D, Detection efficiencies in nanoand femtosecond laser ablation inductively coupled plasma mass spectrometry,SpectrochimActa Part
B, 2009; 64:109-112.
18. Guillong M, Horn I, GĂĽnther D, A comparison of 266 nm, 213 nm and 193 nm produced from a single
solid state Nd:YAG laser for laser ablation ICP-MS,J Anal At Spectrom, 2003; 18:1224-1230.
19. Jochum KP, Stoll B, Herwig K, Willbold M, Validation of LA-ICP-MS trace element analysis of geological
glasses using a new solid-state 193 nm Nd:YAG laser and matrix-matched calibration, J Anal At
Spectrom, 2007; 22:112-121.
20. Kroslakova I, GĂĽnther D, Elemental fractionation in laser ablation-inductively coupled plasma-mass
spectrometry: evidence for mass load induced matrix effects in the ICP during ablation of a silicate
glass, J Anal AtSpectrom, 2007; 22:51-62.

IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013

381

More Related Content

What's hot

Lab 4 Photosynthesis
Lab 4 PhotosynthesisLab 4 Photosynthesis
Lab 4 Photosynthesisericavanetten
 
Investigation on thermodynamics in separation for ethylene glycol neopentyl g...
Investigation on thermodynamics in separation for ethylene glycol neopentyl g...Investigation on thermodynamics in separation for ethylene glycol neopentyl g...
Investigation on thermodynamics in separation for ethylene glycol neopentyl g...Josemar Pereira da Silva
 
On the coupling between Dissipartive Particle Dynamics and Computational Flui...
On the coupling between Dissipartive Particle Dynamics and Computational Flui...On the coupling between Dissipartive Particle Dynamics and Computational Flui...
On the coupling between Dissipartive Particle Dynamics and Computational Flui...Hermes Droghetti
 
Biosorption of hg (ii) from aqueous solutions
Biosorption of hg (ii) from aqueous solutionsBiosorption of hg (ii) from aqueous solutions
Biosorption of hg (ii) from aqueous solutionseSAT Publishing House
 
Principle and application of dsc,dta,ftir and x ray diffraction
Principle and application of dsc,dta,ftir and x ray diffractionPrinciple and application of dsc,dta,ftir and x ray diffraction
Principle and application of dsc,dta,ftir and x ray diffractionBhavesh Maktarpara
 
Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...
Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...
Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...UK Carbon Capture and Storage Research Centre
 
Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...
Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...
Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...Shimadzu Scientific Instruments
 
CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...
CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...
CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...UK Carbon Capture and Storage Research Centre
 
Study of the Electric Properties of Azo/Hydrazone Tautomeric Mixture of the ...
Study of the Electric Properties of Azo/Hydrazone Tautomeric  Mixture of the ...Study of the Electric Properties of Azo/Hydrazone Tautomeric  Mixture of the ...
Study of the Electric Properties of Azo/Hydrazone Tautomeric Mixture of the ...Scientific Review SR
 
Effect of salt magnesium chloride on vapour liquid equilibria of binary aze...
Effect of salt magnesium chloride on vapour   liquid equilibria of binary aze...Effect of salt magnesium chloride on vapour   liquid equilibria of binary aze...
Effect of salt magnesium chloride on vapour liquid equilibria of binary aze...eSAT Journals
 
ENVIRONMENTAL LAB REPORT - 6
ENVIRONMENTAL LAB REPORT - 6ENVIRONMENTAL LAB REPORT - 6
ENVIRONMENTAL LAB REPORT - 6Caleb Harmon
 
Effect of nanofluid on efficiency of centrifugal pump
Effect of nanofluid on efficiency of centrifugal pumpEffect of nanofluid on efficiency of centrifugal pump
Effect of nanofluid on efficiency of centrifugal pumpeSAT Journals
 
GHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIEGHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIEQian Xie
 
Environmental Engineering Lab Manual
Environmental Engineering Lab ManualEnvironmental Engineering Lab Manual
Environmental Engineering Lab ManualMalla Reddy University
 
33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...
33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...
33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...Cristian Randieri PhD
 
Review of magnetic refrigeration system as alternative to conventional refrig...
Review of magnetic refrigeration system as alternative to conventional refrig...Review of magnetic refrigeration system as alternative to conventional refrig...
Review of magnetic refrigeration system as alternative to conventional refrig...Naji Abdullah
 

What's hot (19)

Lab 4 Photosynthesis
Lab 4 PhotosynthesisLab 4 Photosynthesis
Lab 4 Photosynthesis
 
Ikard and Revil 2014 - JoH
Ikard and Revil 2014 - JoHIkard and Revil 2014 - JoH
Ikard and Revil 2014 - JoH
 
EESI
EESIEESI
EESI
 
Investigation on thermodynamics in separation for ethylene glycol neopentyl g...
Investigation on thermodynamics in separation for ethylene glycol neopentyl g...Investigation on thermodynamics in separation for ethylene glycol neopentyl g...
Investigation on thermodynamics in separation for ethylene glycol neopentyl g...
 
On the coupling between Dissipartive Particle Dynamics and Computational Flui...
On the coupling between Dissipartive Particle Dynamics and Computational Flui...On the coupling between Dissipartive Particle Dynamics and Computational Flui...
On the coupling between Dissipartive Particle Dynamics and Computational Flui...
 
Biosorption of hg (ii) from aqueous solutions
Biosorption of hg (ii) from aqueous solutionsBiosorption of hg (ii) from aqueous solutions
Biosorption of hg (ii) from aqueous solutions
 
Principle and application of dsc,dta,ftir and x ray diffraction
Principle and application of dsc,dta,ftir and x ray diffractionPrinciple and application of dsc,dta,ftir and x ray diffraction
Principle and application of dsc,dta,ftir and x ray diffraction
 
Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...
Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...
Impact of Equation of State on Simulating CO2 Pipeline Decompression, Solomon...
 
Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...
Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...
Fast, Sensitive, and Cost-effective Analysis of Trace Metals in Water by EPA ...
 
CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...
CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...
CO2QUEST - Fluid Properties and phase behaviour of CO2 with impurities - Geor...
 
Study of the Electric Properties of Azo/Hydrazone Tautomeric Mixture of the ...
Study of the Electric Properties of Azo/Hydrazone Tautomeric  Mixture of the ...Study of the Electric Properties of Azo/Hydrazone Tautomeric  Mixture of the ...
Study of the Electric Properties of Azo/Hydrazone Tautomeric Mixture of the ...
 
Effect of salt magnesium chloride on vapour liquid equilibria of binary aze...
Effect of salt magnesium chloride on vapour   liquid equilibria of binary aze...Effect of salt magnesium chloride on vapour   liquid equilibria of binary aze...
Effect of salt magnesium chloride on vapour liquid equilibria of binary aze...
 
ENVIRONMENTAL LAB REPORT - 6
ENVIRONMENTAL LAB REPORT - 6ENVIRONMENTAL LAB REPORT - 6
ENVIRONMENTAL LAB REPORT - 6
 
Effect of nanofluid on efficiency of centrifugal pump
Effect of nanofluid on efficiency of centrifugal pumpEffect of nanofluid on efficiency of centrifugal pump
Effect of nanofluid on efficiency of centrifugal pump
 
GHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIEGHGT 11-poster-QIAN XIE
GHGT 11-poster-QIAN XIE
 
Rtm assignment 2
Rtm assignment 2Rtm assignment 2
Rtm assignment 2
 
Environmental Engineering Lab Manual
Environmental Engineering Lab ManualEnvironmental Engineering Lab Manual
Environmental Engineering Lab Manual
 
33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...
33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...
33 Measurement of beam-recoil observables Ox, Oz and target asymmetry T for t...
 
Review of magnetic refrigeration system as alternative to conventional refrig...
Review of magnetic refrigeration system as alternative to conventional refrig...Review of magnetic refrigeration system as alternative to conventional refrig...
Review of magnetic refrigeration system as alternative to conventional refrig...
 

Viewers also liked

laser ablation vs radioiodine in toxic nodular goiters jcem
laser ablation vs radioiodine in toxic nodular goiters jcemlaser ablation vs radioiodine in toxic nodular goiters jcem
laser ablation vs radioiodine in toxic nodular goiters jcemGanesh Kumar
 
Master Presentation Part 1
Master Presentation Part 1Master Presentation Part 1
Master Presentation Part 1micoolcho
 
Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...
Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...
Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...Vein Global
 
Endovenous laser ablation
Endovenous laser ablationEndovenous laser ablation
Endovenous laser ablationdiliprajpal
 
Nanotechnology in treatment of cancer
Nanotechnology in treatment of cancerNanotechnology in treatment of cancer
Nanotechnology in treatment of cancerRAJASEKHARREDDY POLAM
 
Shareslide presentation
Shareslide presentationShareslide presentation
Shareslide presentationksross
 
State of the Word 2011
State of the Word 2011State of the Word 2011
State of the Word 2011photomatt
 

Viewers also liked (9)

laser ablation vs radioiodine in toxic nodular goiters jcem
laser ablation vs radioiodine in toxic nodular goiters jcemlaser ablation vs radioiodine in toxic nodular goiters jcem
laser ablation vs radioiodine in toxic nodular goiters jcem
 
LAT Portfolio
LAT PortfolioLAT Portfolio
LAT Portfolio
 
Nanotubes
NanotubesNanotubes
Nanotubes
 
Master Presentation Part 1
Master Presentation Part 1Master Presentation Part 1
Master Presentation Part 1
 
Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...
Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...
Review of Randomized Controlled Trials Comparing Endovenous Thermal and Chemi...
 
Endovenous laser ablation
Endovenous laser ablationEndovenous laser ablation
Endovenous laser ablation
 
Nanotechnology in treatment of cancer
Nanotechnology in treatment of cancerNanotechnology in treatment of cancer
Nanotechnology in treatment of cancer
 
Shareslide presentation
Shareslide presentationShareslide presentation
Shareslide presentation
 
State of the Word 2011
State of the Word 2011State of the Word 2011
State of the Word 2011
 

Similar to STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL BY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

Silver nanowires and nanoparticles from a millifluidic reactor-application to...
Silver nanowires and nanoparticles from a millifluidic reactor-application to...Silver nanowires and nanoparticles from a millifluidic reactor-application to...
Silver nanowires and nanoparticles from a millifluidic reactor-application to...Alex Tangy
 
11. malibs in argon (jo pcs)
11. malibs in argon (jo pcs)11. malibs in argon (jo pcs)
11. malibs in argon (jo pcs)Ali Khumaeni
 
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...PerkinElmer, Inc.
 
Effect of cold atmospheric pressure plasma needle on DNA
Effect of cold atmospheric pressure plasma needle on DNAEffect of cold atmospheric pressure plasma needle on DNA
Effect of cold atmospheric pressure plasma needle on DNAiosrjce
 
Carbohydrate Research 405 (2015) 55–65Contents lists availab.docx
Carbohydrate Research 405 (2015) 55–65Contents lists availab.docxCarbohydrate Research 405 (2015) 55–65Contents lists availab.docx
Carbohydrate Research 405 (2015) 55–65Contents lists availab.docxwendolynhalbert
 
Fluorescence Spectroscopy for Quantifying OH Radicals Produced by
Fluorescence Spectroscopy for Quantifying OH Radicals Produced byFluorescence Spectroscopy for Quantifying OH Radicals Produced by
Fluorescence Spectroscopy for Quantifying OH Radicals Produced byStephanie Hernandez
 
Master-Ag2S
Master-Ag2SMaster-Ag2S
Master-Ag2SRock Yang
 
An autoignition performance comparison of chemical kinetics models for n-heptane
An autoignition performance comparison of chemical kinetics models for n-heptaneAn autoignition performance comparison of chemical kinetics models for n-heptane
An autoignition performance comparison of chemical kinetics models for n-heptaneOregon State University
 
Gamma radiation induced transformational change in ir spectrum of ebha nemati...
Gamma radiation induced transformational change in ir spectrum of ebha nemati...Gamma radiation induced transformational change in ir spectrum of ebha nemati...
Gamma radiation induced transformational change in ir spectrum of ebha nemati...IAEME Publication
 
GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...
GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...
GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...IAEME Publication
 
IEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and AbstractIEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and Abstracttsysglobalsolutions
 
Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...
Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...
Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...IOSRJAC
 
approved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MR
approved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MRapproved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MR
approved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MRTim Rose
 
Atomicabsorptionspectroscopy aa sforelementalanalysisofplantsamples
Atomicabsorptionspectroscopy aa sforelementalanalysisofplantsamplesAtomicabsorptionspectroscopy aa sforelementalanalysisofplantsamples
Atomicabsorptionspectroscopy aa sforelementalanalysisofplantsamplesBoovizhikannan Thangabalan
 
EFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATION
EFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATIONEFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATION
EFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATIONAM Publications
 

Similar to STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL BY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY (20)

Dimitriu SR 15 Poster
Dimitriu SR 15 PosterDimitriu SR 15 Poster
Dimitriu SR 15 Poster
 
Silver nanowires and nanoparticles from a millifluidic reactor-application to...
Silver nanowires and nanoparticles from a millifluidic reactor-application to...Silver nanowires and nanoparticles from a millifluidic reactor-application to...
Silver nanowires and nanoparticles from a millifluidic reactor-application to...
 
Laser ablation depth
Laser ablation depthLaser ablation depth
Laser ablation depth
 
11. malibs in argon (jo pcs)
11. malibs in argon (jo pcs)11. malibs in argon (jo pcs)
11. malibs in argon (jo pcs)
 
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometall...
 
Effect of cold atmospheric pressure plasma needle on DNA
Effect of cold atmospheric pressure plasma needle on DNAEffect of cold atmospheric pressure plasma needle on DNA
Effect of cold atmospheric pressure plasma needle on DNA
 
Carbohydrate Research 405 (2015) 55–65Contents lists availab.docx
Carbohydrate Research 405 (2015) 55–65Contents lists availab.docxCarbohydrate Research 405 (2015) 55–65Contents lists availab.docx
Carbohydrate Research 405 (2015) 55–65Contents lists availab.docx
 
Fluorescence Spectroscopy for Quantifying OH Radicals Produced by
Fluorescence Spectroscopy for Quantifying OH Radicals Produced byFluorescence Spectroscopy for Quantifying OH Radicals Produced by
Fluorescence Spectroscopy for Quantifying OH Radicals Produced by
 
Master-Ag2S
Master-Ag2SMaster-Ag2S
Master-Ag2S
 
An autoignition performance comparison of chemical kinetics models for n-heptane
An autoignition performance comparison of chemical kinetics models for n-heptaneAn autoignition performance comparison of chemical kinetics models for n-heptane
An autoignition performance comparison of chemical kinetics models for n-heptane
 
Gamma radiation induced transformational change in ir spectrum of ebha nemati...
Gamma radiation induced transformational change in ir spectrum of ebha nemati...Gamma radiation induced transformational change in ir spectrum of ebha nemati...
Gamma radiation induced transformational change in ir spectrum of ebha nemati...
 
GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...
GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...
GAMMA RADIATION-INDUCED TRANSFORMATIONAL CHANGE IN IR SPECTRUM OF EBHA NEMATI...
 
IEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and AbstractIEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and Abstract
 
Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...
Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...
Preparation of Au-Ag composite Nanoparticles by Pulsed Laser Ablation in Wate...
 
approved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MR
approved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MRapproved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MR
approved final draft Tim Rose Poster Plasmonic Nano-particles for Energy MR
 
cme320_lab04
cme320_lab04cme320_lab04
cme320_lab04
 
Atomicabsorptionspectroscopy aa sforelementalanalysisofplantsamples
Atomicabsorptionspectroscopy aa sforelementalanalysisofplantsamplesAtomicabsorptionspectroscopy aa sforelementalanalysisofplantsamples
Atomicabsorptionspectroscopy aa sforelementalanalysisofplantsamples
 
Austin Biomolecules: open access
Austin Biomolecules: open accessAustin Biomolecules: open access
Austin Biomolecules: open access
 
Ijmet 10 01_157
Ijmet 10 01_157Ijmet 10 01_157
Ijmet 10 01_157
 
EFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATION
EFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATIONEFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATION
EFFECT OF TEMPERATURE & PRESSURE ON OXIDATIVE DESULFURIZATION
 

More from IJSIT Editor

CHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF STREPTOMYCES BACTERIS
CHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF  STREPTOMYCES BACTERISCHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF  STREPTOMYCES BACTERIS
CHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF STREPTOMYCES BACTERISIJSIT Editor
 
MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE ENZYMES, FAMILY 19 STR...
MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE  ENZYMES, FAMILY 19 STR...MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE  ENZYMES, FAMILY 19 STR...
MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE ENZYMES, FAMILY 19 STR...IJSIT Editor
 
THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN NITROGEN GREEN BEANS F...
THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN  NITROGEN GREEN BEANS F...THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN  NITROGEN GREEN BEANS F...
THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN NITROGEN GREEN BEANS F...IJSIT Editor
 
ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF EXTRACTS OF JATRO...
ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF  EXTRACTS OF JATRO...ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF  EXTRACTS OF JATRO...
ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF EXTRACTS OF JATRO...IJSIT Editor
 
BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...
BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...
BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...IJSIT Editor
 
THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS CISPLATININDUCED RENAL INSUFFIC...
THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS  CISPLATININDUCED RENAL INSUFFIC...THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS  CISPLATININDUCED RENAL INSUFFIC...
THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS CISPLATININDUCED RENAL INSUFFIC...IJSIT Editor
 
EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM MELEGUETA LEAVE ...
EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM  MELEGUETA LEAVE ...EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM  MELEGUETA LEAVE ...
EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM MELEGUETA LEAVE ...IJSIT Editor
 
THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING WASTE WATER EFFLUE...
THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING  WASTE WATER EFFLUE...THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING  WASTE WATER EFFLUE...
THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING WASTE WATER EFFLUE...IJSIT Editor
 
WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN SOUTHERN BORNO NIGERIA
WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN  SOUTHERN BORNO NIGERIA WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN  SOUTHERN BORNO NIGERIA
WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN SOUTHERN BORNO NIGERIA IJSIT Editor
 
DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE FRUIT TO CALCULATION ...
DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE  FRUIT TO CALCULATION ...DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE  FRUIT TO CALCULATION ...
DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE FRUIT TO CALCULATION ...IJSIT Editor
 
COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN LOUR. EXTRACTS AND ...
COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN  LOUR. EXTRACTS AND ...COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN  LOUR. EXTRACTS AND ...
COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN LOUR. EXTRACTS AND ...IJSIT Editor
 
DIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND COOLING
DIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND  COOLINGDIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND  COOLING
DIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND COOLINGIJSIT Editor
 
BIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE DIATOM CULTURE MEDIA
BIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE  DIATOM CULTURE MEDIABIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE  DIATOM CULTURE MEDIA
BIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE DIATOM CULTURE MEDIAIJSIT Editor
 
COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND TUMOR TISSUE, SARDINE...
COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND  TUMOR TISSUE, SARDINE...COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND  TUMOR TISSUE, SARDINE...
COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND TUMOR TISSUE, SARDINE...IJSIT Editor
 
ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS AGAINST PLANT PATHOG...
ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS  AGAINST PLANT PATHOG...ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS  AGAINST PLANT PATHOG...
ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS AGAINST PLANT PATHOG...IJSIT Editor
 
OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE YEARS SINGLE CE...
OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE  YEARS SINGLE CE...OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE  YEARS SINGLE CE...
OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE YEARS SINGLE CE...IJSIT Editor
 
ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES)
ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES) ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES)
ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES) IJSIT Editor
 
COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF LEACHATE FROM MUNICIPAL ...
COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF  LEACHATE FROM MUNICIPAL ...COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF  LEACHATE FROM MUNICIPAL ...
COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF LEACHATE FROM MUNICIPAL ...IJSIT Editor
 
PHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF RAKTADUSHTIHAR YOGA
PHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF  RAKTADUSHTIHAR YOGAPHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF  RAKTADUSHTIHAR YOGA
PHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF RAKTADUSHTIHAR YOGAIJSIT Editor
 
ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL INFLUENZA A, H1N1 IN MALA...
ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL  INFLUENZA A, H1N1 IN MALA...ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL  INFLUENZA A, H1N1 IN MALA...
ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL INFLUENZA A, H1N1 IN MALA...IJSIT Editor
 

More from IJSIT Editor (20)

CHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF STREPTOMYCES BACTERIS
CHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF  STREPTOMYCES BACTERISCHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF  STREPTOMYCES BACTERIS
CHITINASE AS THE MOST IMPORTANT SECONDARY METABOLITES OF STREPTOMYCES BACTERIS
 
MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE ENZYMES, FAMILY 19 STR...
MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE  ENZYMES, FAMILY 19 STR...MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE  ENZYMES, FAMILY 19 STR...
MOLECULAR ANALYSIS OF BACTERIAL GENE CODING CHITINASE ENZYMES, FAMILY 19 STR...
 
THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN NITROGEN GREEN BEANS F...
THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN  NITROGEN GREEN BEANS F...THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN  NITROGEN GREEN BEANS F...
THE EFFECTS OF HELPING BACTERIA (PSEUDOMONAS SPP.) IN NITROGEN GREEN BEANS F...
 
ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF EXTRACTS OF JATRO...
ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF  EXTRACTS OF JATRO...ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF  EXTRACTS OF JATRO...
ANTIMICROBIAL PROPERTY OF AQUEOUS AND PETROLEUM ETHER LEAF EXTRACTS OF JATRO...
 
BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...
BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...
BIO CHEMICAL EFFECT OF 1, 5-BIS (3, 5-DIMETHYLPYRAZOL-1-YL)-3- OXAPENTANE-DIA...
 
THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS CISPLATININDUCED RENAL INSUFFIC...
THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS  CISPLATININDUCED RENAL INSUFFIC...THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS  CISPLATININDUCED RENAL INSUFFIC...
THE EFFECT OF ALSTONEA BOONEI STEM BARK PLUS CISPLATININDUCED RENAL INSUFFIC...
 
EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM MELEGUETA LEAVE ...
EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM  MELEGUETA LEAVE ...EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM  MELEGUETA LEAVE ...
EFFECTS OFMYRISTICA FRAGRANS, MURRAYA KOENIGI AND AFRAMOMUM MELEGUETA LEAVE ...
 
THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING WASTE WATER EFFLUE...
THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING  WASTE WATER EFFLUE...THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING  WASTE WATER EFFLUE...
THE INFLUENCE OF SILICONE ANTIFOAM FROM LEATHER AND DYING WASTE WATER EFFLUE...
 
WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN SOUTHERN BORNO NIGERIA
WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN  SOUTHERN BORNO NIGERIA WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN  SOUTHERN BORNO NIGERIA
WATER INTAKE CHARACTERISTICS OF DIFFERENT SOIL TYPES IN SOUTHERN BORNO NIGERIA
 
DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE FRUIT TO CALCULATION ...
DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE  FRUIT TO CALCULATION ...DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE  FRUIT TO CALCULATION ...
DETERMINATION OF ENGINEERING PROPERTIES OF POMEGRANATE FRUIT TO CALCULATION ...
 
COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN LOUR. EXTRACTS AND ...
COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN  LOUR. EXTRACTS AND ...COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN  LOUR. EXTRACTS AND ...
COMPARSION OF ANTIOXIDANT POTENTIAL OF DIMOCARPUS LONGAN LOUR. EXTRACTS AND ...
 
DIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND COOLING
DIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND  COOLINGDIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND  COOLING
DIRECT EXPANSION GROUND SOURCE HEAT PUMPS FOR HEATING AND COOLING
 
BIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE DIATOM CULTURE MEDIA
BIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE  DIATOM CULTURE MEDIABIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE  DIATOM CULTURE MEDIA
BIOMINERALISED SILICA-NANOPARTICLES DETECTION FROM MARINE DIATOM CULTURE MEDIA
 
COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND TUMOR TISSUE, SARDINE...
COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND  TUMOR TISSUE, SARDINE...COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND  TUMOR TISSUE, SARDINE...
COMPARATIVE STUDIES ON NUTRITIONAL VALUE OF NORMAL AND TUMOR TISSUE, SARDINE...
 
ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS AGAINST PLANT PATHOG...
ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS  AGAINST PLANT PATHOG...ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS  AGAINST PLANT PATHOG...
ANTIFUNGAL ACTIVITY OF SELECTED MEDICINAL PLANT EXTRACS AGAINST PLANT PATHOG...
 
OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE YEARS SINGLE CE...
OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE  YEARS SINGLE CE...OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE  YEARS SINGLE CE...
OUTCOME OF TUNNELED CATHETERS IN HEMODIALYSIS PATIENTS: FIVE YEARS SINGLE CE...
 
ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES)
ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES) ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES)
ANTIBACTERIAL ACTIVITY OF Citrus limonON Acnevulgaris (PIMPLES)
 
COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF LEACHATE FROM MUNICIPAL ...
COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF  LEACHATE FROM MUNICIPAL ...COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF  LEACHATE FROM MUNICIPAL ...
COMPARATIVE STUDY ON HEAVY METAL CHARACTERISTICS OF LEACHATE FROM MUNICIPAL ...
 
PHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF RAKTADUSHTIHAR YOGA
PHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF  RAKTADUSHTIHAR YOGAPHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF  RAKTADUSHTIHAR YOGA
PHARMACOGNOSTICAL AND PHYTO–CHEMICAL EVALUATION OF RAKTADUSHTIHAR YOGA
 
ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL INFLUENZA A, H1N1 IN MALA...
ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL  INFLUENZA A, H1N1 IN MALA...ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL  INFLUENZA A, H1N1 IN MALA...
ESTIMATION OF THE REPRODUCTION NUMBER OF THE NOVEL INFLUENZA A, H1N1 IN MALA...
 

Recently uploaded

1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room servicediscovermytutordmt
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
 
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 đź’ž Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 đź’ž Full Nigh...Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 đź’ž Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 đź’ž Full Nigh...Pooja Nehwal
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...anjaliyadav012327
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformChameera Dedduwage
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Krashi Coaching
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAssociation for Project Management
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityGeoBlogs
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationnomboosow
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Mastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionMastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionSafetyChain Software
 

Recently uploaded (20)

1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room service
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
 
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 đź’ž Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 đź’ž Full Nigh...Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 đź’ž Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 đź’ž Full Nigh...
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy Reform
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
CĂłdigo Creativo y Arte de Software | Unidad 1
CĂłdigo Creativo y Arte de Software | Unidad 1CĂłdigo Creativo y Arte de Software | Unidad 1
CĂłdigo Creativo y Arte de Software | Unidad 1
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communication
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Mastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionMastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory Inspection
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 

STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL BY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

  • 1. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 STUDY OF NANOSECOND ND: YAG 213 nm LASER ABLATION OF EXTRA HEAVY CRUDE OIL BY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Henry Wuyke*, LlinaberFeo, Miguel Murillo *PDVSA-INTEVEP S.A. Urbanizacion Santa Rosa. Sector El Tambor. Los Teques. Estado Miranda. Apdo. 76343, Caracas 1070A, Venezuela Centro de QuĂ­micaAnalĂ­tica, Facultad de Ciencias, Universidad Central de Venezuela, Apdo.47102, Caracas 1070A, Venezuela. ABSTRACT The Laser-Extra Heavy Crude Oil interaction was studied using a nanosecond Nd:YAG 213 nm laser coupled to an ICP-MS. For this study, ~2.7 g Extra Heavy Crude Oil was mixed with ~3 mL xylene and 0,3 g organic multielemental solution containing 100 µg g-1 of Ag, Al, Ba, Cd, Cr, Cu, Mg, Mn, Mo, Ni, Pb, Sn, Ti, V, and Zn, xylene was left to evaporate at 90 °C and multielemental samples were obtained and analyzed by LA-ICPMS. Laser parameters, such as, ablation gas flow, sample scanning speed, spot size, fluence and focus position were optimized and RSD (n = 8) from 3% to 5% for the net signal of 63Cu, 95Mo, 60Ni and 66Zn and from 0.9% to 3.3% for the ratios Ni/Mo and Cu/Zn were obtained. Elemental fractionation was observed with both Zn and Cd when fluence was varied. Nevertheless, evidence is not enough to explain results as a consequence of high volatilityor highionization potential (PI)of these elements. Ablationcratersobtained showsigns ofmore pronounced thermal processesat higher fluences. Elemental fractionation was also observed in the ICP due to mass load effect. Isotopic ratios 95Mo/112Cd, 63Cu/66Zn y 27Al/24Mg, which have different ionizationpotential (1.89; 1.68 and 1.66 eV, respectively expressed as PIdenominator - PInumerator) increase when mass load is increase. On the other hand, plasma 60Ni/63Cu ionization ratio (0.09 eV difference) remains constant. These results indicate that the temperature decreases when mass load is increased affecting thoseelementswithhighionization potential. Keywords:Laser Ablation, ICP-MS, Extra Heavy Crude Oil, Elemental Fractionation. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 367
  • 2. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 INTRODUCTION The use of LA-ICP-MS to analyze crude oil maybe advantageoustoeliminatethe extensivesample treatmentstepsoftraditional methods such as calcination or acid digestion. However, few works using LA-ICPMS for direct elemental analysis in crude oil have been reported [1-3]. It can be probably due to difficulties found during ablation, such as splashing, of this type of liquid samples. LA-ICP-MS is more suitable forsolid samples [4-7]. Therefore, it is a potential tool for direct analysis of semi-solid extra heavy crude oil with less trouble inthe ablation process. The ablation processis strongly dependent onthe propertiesofthe sample such as composition, thermal diffusivity and absorptivity [4].Therefore, matrix matched calibration standards or reference certificated materials are required to ensure accuracyof results. However, calibration standards are not available for the majority of sample matrixes, which limitsthe applicationsofLA-ICP-MS [8,9]. Therefore, it is important to know how the laser ablation of extra heavy crude oil is affected by eachexperimental parameter and identify which of them lead rise to fractionation or non-stoichiometric sampling. Pulse length isoneofthe most important parametersinvolvedinmechanisms by whichthe ablation laser processoccurs. Lasers with pulse length in nanosecond (ns) and femtosecond (fs) are the mostfrequentlyused in LA-ICP-MS applications. Generally, fs lasers produce better results because the time interval of energy delivering is much smaller than the time interval needed for the thermal stabilization of the sample (intheorder of picoseconds) [6,10]. Furthermore, photon intensitiesreach valuesof >1014 W cm-2, thereby processesare predominantlynon-thermal, causingless collateral damage, less dependence onthe matrix, smaller particlesand nonlaser-plasma interaction [6]. Nevertheless, even thoughthe latest developmentsare aimed to take advantageoffslasers, todaynslasers are still very populardue primarily totheir low cost. Fluence also plays an important role in ablation mechanisms. It must be above a threshold value to lead the ablation process [10]. When ns lasers are used, fluence should not be too high to avoid the generation of a high electron density plasma, because it increases the laser-plasma interaction and therefore, reduces the interaction between the laser and the sample [11]. When fs lasers are used, fractionation appears to be only near the threshold value of the sample and reach an asymptotic value as the fluence increases [9,12-14]. Fractionationaccording to the fluencein the rangeof2 to 42J cm-2with afslaserin brass, aluminumand glass samples was assessed by Bian et al. [12].In those conditions, the ratio Zn/Cu in the brass and aluminum samples was independent of the fluence while for glass samples there was a strong dependency especially for the most transparent glass sample (NIST 610). However, at higher fluence, the ratio Zn/Cu reached an expected asymptotic value, considering the results of brass and aluminum measures. Similar results on the same glass sample type were obtained by Koch et al. [13]. Fractionation was observed with ratios Zn/Cu and IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 368
  • 3. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Mo/Ni which have different ionization potentials but not with the ratio U/Th which have similar ionization potentials (PI: U: 6.2 eV, Th: 6.3 eV y Pb: 7.5 eV), which evidence of that the ionization potential affects the ablation process. It has been mentioned that Coulomb interactions are responsible for this behavior [11]. However, ablationmechanismsarestillnotfully understoodto the date [15]. There are other factors such as the volatility of the elements that influence the signals when thermal processes occurred, i.e. morevolatileelements arevaporizedpreferentially,resulting in greatersignal and causingfractionation [12,16]. Fractionationhasalso beenstudiedas a function oftime, i.e. shot to shot. Garcia et al. [9] studied this effect using afs laser in binarymetallic, semiconductorsand glasses samples. They obtainedthatthe Cu/Znchangesfrom shot to shotreachedan asymptotic valueindependentof fluence,which correspondsto astoichiometricsampling. Furthermore, thisasymptotic valueis reachedfaster ifhigh fluenceisapplied. There are other parameters that also affect the ablation process such as ablation gas and laser wavelength. Helium and argon are the most used gases for ablation environmental and sample transportation. However, as argon has a lower ionization potential than helium, there is a greater likelihood of plasma formation and shielding effect that prevent an efficient interaction between laser and sample, so that helium is preferred [11]. Wälle et al. [17]using ns and fs lasers in glass, zircon and silicon samples, observed a depression of the detection efficiency by a factor of 5 when used argon instead of helium as ablation gas. Studies have shown that using shorter wavelengths, with both ns and fs lasers in glass samples fractionation effects are reduced [13,18,19]. Particles generated in the ablation process can have different compositions in terms of size, especially if a ns laser is used [6]. It creates fractionation if there is no transport efficiency, because large particles (> 1 µm) tend to be lost by gravity or deposited on chamber ablation. While, small particles (<10 nm) are lost by diffusion or electrostatic effects [12].The ideal condition is that all particles are fully decomposed when reaching the ICP in order to to avoid affecting the characteristics of plasma. Nevertheless, it has been found that mass load that reaches the ICP can create a matrix effect dependent on the ionization potential [13] or on the melting point of the elements [20]. Although LA-ICP-MS has been used for elemental analysis in crude oil [1,3] a systematic study that assess the fractionation of these types of samples, as is usually done on metal alloys or glass samples, has not been performed. The aim of this study is to evaluate the behavior of laser - extra heavy crude oil interaction by LA-ICP-MS as a function of various experimental parameters using a ns Nd: YAG laser at 213nm. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 369
  • 4. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 MATERIALS AND METHODS Instrumentation: Laser ablation was performed using a Nd:YAG UP-213 LA New Wave Research operated at 213 nm with a pulsed duration of ~ 4 ns. An ICP-MS Agilent Technologies 7500ce that consist in a plasma source, aoctopole reaction cell and a quadrupole mass analyzer was used. Laser ablation and ICP-MS were connected using Tygon tubes (S-50-HL Class VI, ~ 3 mm i.d.) as shown in Fig. 1. In analysis, the aerosol from the ablation cell is mixed with Ar using a "Y" connection, then goes to a cyclone chamber of ~ 4 cm diameter and finally to the ICP-MS. The bypass and purge valves are used for analysis preconditioning. Figure 1: Coupling laser ablation to ICP-MS Reagents, stardards and samples: Extra heavy crude oil samples (API° < 10) were provided by PDVSA-INTEVEP. Xylene (puriss. p.a.) Sigma-Aldrich and an organic multielemental solution containing 100 µg g-1 of Ag, Al, Ba, Cd, Cr, Cu, Mg, Mn, Mo, Ni, Pb, Sn, Ti, V and Zn, S-21 Conostan, USA were used. The tuning solution to perform the ICP-MS was preparedfrom ARMTM Absolute Standards, INC.1000 mg/Kg of Li, Mg, Co, Y, Ce and Pb in 2% HNO3. Sample preparation: Approximately 2.7 g of oil was dissolved in a beaker with 2 mL xylene and 0.3 g of organic multielemental solution, the mix was homogenized by magnetic stirring and brought to an oven at 90 ° C for 2 hours until evaporation of xylene. Finally, part of the oil was placed in a sample holder tray of 1 cm diameter and 3 mm high and led to LA-ICP-MS. The resulting mixture contained a concentration of 10 µg g -1 in metals from the organic multielemental solution. General conditions of analysis: In this work, several experiments were performed in order to study the laser - extra heavy crude oil interaction by ICP-MS. Table 1 shows the general conditions of analysis. Isotopesof each elementwere chosen IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 370
  • 5. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 tominimizeisobaricandpolyatomicinterferences. The ICP-MS parameters were those that resulted in good sensitivity for all elements of the tuning solution, covering a wide tuning range of masses: 7Li, 24Mg, 59Co, 89Y, 140Ce and 208Pb. Once sample is introduced into the ablation cell, the system is purged with He 1 L min-1 flow for 5 min. ICP-MS Plasma gas flow, L Agilent 7500ce min-1 15 Foward power, W 1500 Sample depth, mm 6.5 Integration timeper point, s 0.1 Integration timeper mass, s 0.3 (3 point per mass) Laser ablation New Wave Research UP-213 LA Ablation gas flow (He), L min-1 0.75 Mixing gas flow (Ar), L min-1 0.5 Fluence, J cm-2 ~ 34 (100 %) Focus, mm 4 (above the sample) Spot size, µm 100 (1000 µm to 4 mm of defocus) Pulse frequency, Hz 20 Scanning speed, µm s-1 (linear) 20 Table 1:General conditions of analysis RESULTS AND DISCUSSION A typical ablation profile of Ni, Cu, Zn and Mo is shown in Fig. 2. It is observed that the signal of Ni is much greater than the signals of Cu, Zn and Mo. Thisis becauseNi isoneofthe major elementsinthecrude oil. However, there is good signal stability for the four elements at different concentration levels. Therefore, Cu and / or Zn signals were used to shown the optimization of the laser parameters and were selected for their differences in ionization potential (PI) and melting point (mp) (PI/mp: Cu 7.72 eV / 1083 °C, Zn 9.42 eV / 419 °C). IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 371
  • 6. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Figure 2: Ablation profile of 60Ni, 63Cu, 66Zn and 95Mo at general analysis conditions. Ablation time was 50 s with 10 s of delay relative to the acquisition start Optimization of the laser parameters was performed with the aim of achieving the greatest intensity for 60Ni, 63Cu, 66Zn and 95Mo isotopes while maintaining good repeatability. In experiments with He-Ar as ablation gas and mixing gas respectively, flow rates were optimized between 0.4 – 0.9 L min-1 and the best results were obtained with 0.75 L min-1 He and 0.5 L min-1 Ar. In the Ar-Ar system, flow rates were optimized between 0 to 1.2 L / min, and the best results were obtained with 1.0 L min -1Ar (ablation gas) and 0.2 L min1Ar (mixing gas). Both He and Ar as ablation gases were compared in terms of the transient signal produced. The results are shown in Fig. 3. Figures 3a and 3c show that 63Cu and 66Zn signals are greater when He is used as ablation gas. Furthermore, a more pronounced memory effect is observed when Ar is used as ablation gas (figures 3b and 3d). The damage on crude oil surface is greater when Ar is used in the ablation cell (figure 4) at both optimum analysis conditions (4 mm of defocus) and focus. These results are consistent with previous studies [11,17], where it is argued that Ar, which posses an ionization potential (15.7 eV) lower than He (24.6 eV), promotes the formation of plasma on the surface, which interacts with the sample and deform it, i.e. there aresignificantthermal processes. The memory effect produced when Ar is used can be explained by the expansion to large volume of the ablated material into the ablation cell. It makes that the system takes longer to transport this volume. It has been mentioned that this expansion is more pronounced when the ionization potential is decreased [11]. It agrees with the results obtained in this study. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 372
  • 7. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 He was chosen as ablation gas for the subsequent experiments because it produces a significantly high signal for Cu, Zn, Ni and Mo (Student t test with α = 0.05) and an improved precision to the Cu/Zn and Ni/Mo ratios (Fisher F test α = 0.05) when compared to the obtained with Ar as ablation gas. Results are shown in table 2. Figure 3: Signal versus time for two ablation gases, He and Ar: a and b) CPS of 63Cu, c and d) CPS of 66Zn. a and c) approximately at 10 s, the laser was switched on 7s (140 pulses), b and d) zoom between times 30 and 90 s IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 373
  • 8. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Figure 4:Picturesof cratersat different ablation conditions. It is indicated in order: ablation gas, pulses number and defocus level (mm): a) He, 1000 pulses, 4 mm, b) Ar, 1000 pulses, 4 mm, c) He, 400 pulses, 0 mm (in focus), d) Ar, 100 pulses, 0 mm (in focus). The mark corresponds to 100 µm. He and Ar flow in the ablation cell were 0.75 L min-1 Element o ratio Signal Average (CPS). He SD (CPS). He Signal CV (%). He Average (CPS). Ar SD (CPS). Ar CV (%). Ar Cu 8.1E5 0.3E5 3.7 4.6E5 0.3E5 6.5 Zn 4.6E5 0.2E5 4.3 2.0E5 0.1E5 5.0 Ni 4.2E6 0.1E6 2.4 2.4E6 0.2E6 8.3 Mo 6.1E5 0.2E5 3.3 3.5E5 0.3E5 8.6 Cu/Zn 1.754 0.058 3.3 2.15 0.12 5.6 Ni/Mo 6.990 0.061 0.9 6.99 0.22 3.1 Table 2: Media, standarddeviation (SD), coefficientofvariation (CV) and samplesize (n) for Cu, Zn, Ni and Mo signals and Cu/Zn and Ni/Mo ratios using He or Ar as ablation gas The effect of two cyclone chambers with different sizes (~ 4 cm and ~6 cm internal diameter) was evaluated. Figure 5 shows that per run, the signal is more stable when a cyclone chamber is used, because it seems to dampen the effect produced in the ablation cell. Among ~ 4 cm and ~ 6 cm chambers no great differences were obtained. Therefore, ~ 4 cm chamber was taken as condition routine to make smaller the volume of the transport region. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 374
  • 9. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Figure 5:63Cu signal versus time with two cyclone chambers and without them. Ablation time was 50 s with 10 s of delay with 10 s delay relative to the acquisition start In most of the works reported on solid samples, it is preferred ablate with a scanning speed different to zero, because at the same point, it begins to form an increasingly deep crater with each impact of the laser. The efficiency of energy transfer from the laser to the sample is continuously decreased and fractionation is observed. With a scanning speed different to zero, ablation is carried out over a greater area, which ensures a stoichiometric sampling. Figure 6:63Cu signal versus time for the four scanning speeds indicated. Four replicates are shown. Ablation time was 50 s with 10 s of delay relative to the acquisition start IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 375
  • 10. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Four linear scanning speeds 0, 20, 40 and 60 µm s -1 were tested in this work. Fig. 6 shows the ablation profiles of the four scanning speeds with four repetitions each one. At 0 µm s -1 reproducible profiles with a signal drop from the beginning to the end of the run between 10 and 15% are observed. However, Fig. 7 shows that there is no fractionation of 63Cu/66Zn ratio at 0 µm s-1. Profiles at speeds 20 and 40 µm s-1 are little less stable and reproducible than at 0 µm s-1, but without signal drop. The profile at 60 µm s-1 is more random probably for the larger sample area covering, which is not on the same horizontal axis (the crude oil sample on holder may be concave or convex and it is not easy to reproduce a flat surface). Besides from the beginning to the end of the run, its vertical position has changed, varying the level of defocus and so the sensitivity. Even though, the 63Cu/66Zn ratio remained stable through time at four scanning speeds studied (see Fig. 7). Finally, it canbe established thatthe most appropriate conditions fortheanalysisare atsmallscanning speeds~ 20µm s-1. Figure 7:63Cu/66Zn ratio versus time at four scanning speeds. Signals were recorded in the same experiment of Fig. 6 The fractionation effect versus fluence was evaluated with 15 elements by varying of defucus level which allows to work in a greater range instead of varying the energy to a fixed diameter. As can be seen in Fig. 8, the greater signal was obtained at higher positive defocus levels. Although fluence becomes smaller, it is still sufficient to ablate a larger area the crude oil sample, which explains the results. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 376
  • 11. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Figure 8:60Ni, 63Cu, 66Zn and 95Mo signals versus defocus level. Each point represents the average of three 30 s runs of signal acquisition Figure 9 shows the relationship between the elements studied and 66Zn (Zn was chosen for its high ionization potential and high volatility) versus fluence. In general, fractionation was observed, i.e. M/ 66Zn ratio increases with increasing fluence, except for Cd that appears to be stable, since it has very similar characteristics to Zn. Zn and Cd that present this fractionation with respect to the other elements, have high ionization potential and high volatility. The observed behavior is hard to explain because ablation mechanisms are not yet fully known at the date [15]. However, trends indicate thatwhentheablationis carried outbythermal processes, more volatileelementsarepreferentiallyvolatilized [6]. Recent results[13] suggest thatelementswithlow ionization potentialare preferentiallyableatedand coulomb interactions are responsible of that [11]. The results shown in Fig. 9 indicate that at lower fluences, Zn signal is favored compared to the other elements (except for Cd), contrary effect to that observed in previous studies on glass samples [12,13]. The results can not be explained by volatility if it is considered that there are greater thermal processes at high fluencesor at leastgreatersurface damageby plasma-sampleinteraction (see Fig. 4a (less fluence) and 4c (greater fluence, greatersurface damage)). As the Zn is more volatile, it should be favored at high fluence and that does not happen. Similarly, the Mo that is an element with high melting and boilingpoint, presents nofractionation, i.e. volatilitydoes not appear tobe responsible. The Al, that has relatively low ionization potential, did not present fractionation. In conclusion, Zn and Cd were the only elements that showed fractionation as function of fluence, but without sufficient evidence to explain it by volatility or ionization potentials. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 377
  • 12. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Figure 9: Normalized M (isotopes indicated) / 66Zn ratios versus fluence (J cm-2). Each point represents three 2 min 30 s runs average of acquisition signal at a scanning speed of 20 µm It can be thought that the results of Fig. 9 may be influenced by mass load effect, because there is greater quantity of particles at low fluences (greater defocus and larger diameter) than at high fluences (see Fig. 8). The ICP may be affected, decreasing its ionization temperature as a result of a large mass load, which can cause different ionization profiles for elements with different ionization potentials. To evaluate this effect with crude oil sample, the analysis with different crater diameters while remained fixed the fluence was performed. Fig. 10 shows 63Cu/66Zn,95Mo/112Cd, 27Al/24Mg and 60Ni/63Cu ratios which differ inionizationpotential (expressed as PIdenominator - PInumerator)1.68; 1.89; 1.66 y 0.09 eVrespectively. It was observed that ratios tend to increase when mass load increases. This effect was more marked for the 60Ni/63Cu 95Mo/112Cd ratio, which has the largest difference in ionization potential. On the contrary ratio shows no significant change, since their PI not differ greatly. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 378
  • 13. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 Figure 10: Normalized 63Cu/66Zn, 95Mo/112Cd, 27Al/24Mg and 60Ni/63Cu ratios versus crater surface. Each point represents three 2 min 30 s runs average of acquisition signal at a scanning speed of 20 µm At higher mass load, the ICP diminishes its ionization temperature and those elements with high ionization potential are more affected which explains the results of Fig. 10. If mass load had an effect on the results of Fig. 9, they hada decreasing trend (because the greater mass load is found in the low fluenceanalyse), confirming that the results are effects mainly produced by the ablation processes at different fluences and not by the processes occurring in the ICP. CONCLUSIONS Fractionation was observed as function of fluence, i.e. Zn and Cd signals were affected in different manner, respect to the other 13 elements studied, when fluence is increased. Elements with low ionization potential as Al or low volatility as Mo, showed no fractionation when fluence is varied. Therefore, there was insufficient evidenceto explainthe results in termsofthe volatility andtheionization potential. Craters pictures are evidence that ablation mechanisms are changed at different fluence (see Fig. 4). They showed greater surface damage at higher fluence. It can be probably due to plasma-sample interaction. Also, differences among He and Ar ablation gases were observed. The best results, in terms of sensitivity and precision were obtained when He was used as ablation gas. Fractionation was also observed in the ICP. At higher mass load were disadvantaged those elements with high ionization potential due to the decreased of plasma ionization temperature. Conditions of increased sensitivity to the extra heavy crude oil were determined in the LA-ICPMS system. It was found that they correspond mainly to a larger area of ablation (larger diameter and higher defocus level), highlaser energy and high pulse frequency. Furthermore,an improvement inthe stability of thesignal was observedwhen acyclonicchamberwas used inthe transport system, this may promote the use of IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 379
  • 14. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 various devices in the transport system to improve analysis precision of related samples with high organic matter content. Under performed conditions, %RSD from 3 to 5% in the net signal of Cu, Zn, Ni and Mo and from 0.9 to 3.3% in the Ni/Mo and Cu/Zn ratios were found, respectively. The results of this study show the possibility of performing elemental determinations in extra heavy crude oil by ns-LA-ICP-MS. However, further experiments are needed for the purpose of increasing the sensitivity of the analytes, improve precision, reduce mass load effects and make the method robust to small matrix changes. REFERENCES 1. Ricard E, PĂ©cheyran C, Sanabria G, Prinzhofer A, Donard OFX, Direct analysis of trace elements in crude oils by high-repetition-rate femtosecond laser ablation coupled to ICPMS detection, Anal BioanalChem, 2011; 399:2153-2165. 2. Vorapalawut N, Pohl P, Bouyssiere B, Shiowatana J, Lobinski R, Multielement analysis of petroleum samples by laser ablation double focusing sector field inductively coupled plasma mass spectrometry (LA-ICP MS), J Anal At Spectrom, 2011; 26:618-622. 3. Heilmann J, Boulyga SF, Heumann KG, Development of an isotope dilution laser ablation ICP-MS method for multi-element determination in crude and fuel oil samples, J Anal At Spectrom, 2009; 24:385-390. 4. Herbert CG, Johnstone RAW, Mass Spectrometry Basics. CRC Press, New York, 2003. 5. Koch J, Gunther D, Laser Ablation ICP-MS. Encyclopedia of Spectroscopy and Spectrometry, 2nd edn, 2009;1262-1269. 6. Fernández B, Claverie F, PĂ©cheyran C. Donard OFX, Direct analysis of solid samples by fs-LA-ICP-MS, Trends in Anal. Chem, 2007; 26:951-966. 7. Hergenröder R, Laser-generated aerosols in laser ablation for inductively coupled plasma spectrometry,SpectrochimActa Part B, 2006; 61:284-300. 8. Hill SJ, Inductively Coupled Plasma Spectrometry and its Applications. Blackwell Publishing, Oxford, 2007. 9. Garcia CC, Lindner H, Von Bohlen A, Vadla C, Niemax K, Elemental fractionation and stoichiometric sampling in femtosecond laser ablation, J Anal At Spectrom, 2008; 23:470-478. 10. Miller JC, Hanglund RF, Laser Ablation and Desorption, Academic Press, San Diego, 1998. 11. Garcia CC, Lindner H, Niemax K, Laser ablation inductively coupled plasma mass spectrometry— current shortcomings, practical suggestions for improving performance, and experiments to guide future development, J Anal At Spectrom, 2009; 24:14-26. 12. Bian Q, Garcia CC, Koch J, Niemax K, Non-matrix matched calibration of major and minor IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 380
  • 15. Henry Wuyke -et al., IJSIT, 2013, 2(5),367-381 concentrations of Zn and Cu in brass, aluminium and silicate glass using NIR femtosecond laser ablation inductively coupled plasma mass spectrometry, J Anal At Spectrom, 2006; 21:187-191. 13. Koch J, Wälle M, Pisonero J, GĂĽnther D, Performance characteristics of ultra-violet femtosecond laser ablation inductively coupled plasma mass spectrometry at ~265 and ~200 nm, J Anal At Spectrom, 2006; 21:932-940. 14. Bian Q, Koch J, Lindner H, Berndt H, Hergenröder R, Niemax K, Non-matrix matched calibration using near-IR femtosecond laser ablation inductively coupled plasma optical emission spectrometry, J Anal At Spectrom, 2005; 20:736-740. 15. Glaus R, Kaegi R, Krumeich F, GĂĽnther D, Phenomenological studies on structure and elemental composition of nanosecond and femtosecond laser-generated aerosols with implications on laser ablation inductively coupled plasma mass spectrometry,SpectrochimActa Part B, 2010; 65:812-822. 16. Kosler J, Wiedenbeck M, Wirth R, Hovorka J, Sylvester P, MĂ­ková J, Chemical and phase composition of particles produced by laser ablation of silicate glass and zircon—implications for elemental fractionation during ICP-MS analysis, J Anal At Spectrom, 2005; 20:402-409. 17. Wälle M, Koch J, Flamigni L, Heiroth S, Lippert T, Hartung W, GĂĽnther D, Detection efficiencies in nanoand femtosecond laser ablation inductively coupled plasma mass spectrometry,SpectrochimActa Part B, 2009; 64:109-112. 18. Guillong M, Horn I, GĂĽnther D, A comparison of 266 nm, 213 nm and 193 nm produced from a single solid state Nd:YAG laser for laser ablation ICP-MS,J Anal At Spectrom, 2003; 18:1224-1230. 19. Jochum KP, Stoll B, Herwig K, Willbold M, Validation of LA-ICP-MS trace element analysis of geological glasses using a new solid-state 193 nm Nd:YAG laser and matrix-matched calibration, J Anal At Spectrom, 2007; 22:112-121. 20. Kroslakova I, GĂĽnther D, Elemental fractionation in laser ablation-inductively coupled plasma-mass spectrometry: evidence for mass load induced matrix effects in the ICP during ablation of a silicate glass, J Anal AtSpectrom, 2007; 22:51-62. IJSIT (www.ijsit.com), Volume 2, Issue 5, September-October 2013 381