HOW TO USE MEDICAL DEVICES SAFELY
VOL: 97, ISSUE: 43, PAGE NO: 36
Wendy Murray, PGDip (health visiting), BSc, RGN, RM, RHV, is nursing officer
Helen Glenister, PhD, MBA, BSc, RN, is nursing director of the Medical Devices Agency
Unfortunately, when using medical devices, things do sometimes go wrong - and the
outcome can be fatal. Last year, the Medical Devices Agency (MDA) received 7,249
adverse incident reports, of which 824 involved serious injury or fatality. The consequences
of these incidents can be devastating for patients and their families, as well as for the health
care professionals involved.
Unfortunately, when using medical devices, things do sometimes go wrong - and the
outcome can be fatal. Last year, the Medical Devices Agency (MDA) received 7,249
adverse incident reports, of which 824 involved serious injury or fatality. The consequences
of these incidents can be devastating for patients and their families, as well as for the health
care professionals involved.
Assessing devices
We all assume that in the event of an emergency, equipment will operate as we expect it to.
However, this is not always the case.
One incident reported to the MDA in March this year highlights the importance of checking
equipment regularly to ensure that it is working correctly. The report came from an
ambulance driver who had experienced problems with a suction pump unit during a
resuscitation procedure. Possibly because of the displacement of a reusable suction canister,
the unit failed to produce adequate suction. This may have been due to the canister being
moved or knocked.
This incident shows that after the initial assessment of any medical device (see last week's
article, 'Safety tips', October 18, p24), regular checks need to be carried out to ensure the
device is functioning correctly and to find out whether reassessment is required.
Nurses must ask themselves two important questions when assessing devices. First, is the
device doing what the manufacturer intended it to do, and, second, is it safe? Establishing
these two points is vital if adverse incidents are to be avoided (Box 1). There have been a
number of incidences where these simple questions could have avoided tragic incidents.
Knowing what to check
It is also important that nurses understand the instructions which accompany medical
devices. If they misinterpret the manufacturer's instructions, they may unwittingly cause
serious injury.
For example, the MDA recently received reports that the prolonged use of pulse oximeter
probes and the tape used to hold them in place had led to tissue necrosis in infants (MDA,
2001a).
The manufacturer's instructions stated that the site should be inspected every four hours.
The manufacturers had assumed that the inspection would result in a change of site.
However, nurses inspecting the site believed they were doing so simply to check that the
pulse oximeter was still in situ. To avoid confusion in future, revised instructions suggest
that the site should be changed at least every four hours.
Knowledge of the device being used
Health care professionals need to be familiar with the devices they are using and should not
attempt to operate any device they have not been fully trained to use.
Reports to the MDA have included incidences where over-infusion of intravenous drugs
has occurred because nurses did not follow the prescription, and administered them too
quickly. Investigations revealed that in some cases the device had been over-infusing for
some time without action being taken, despite written records clearly identifying over-
infusion at the time (Glenister, 2000).
There is ongoing confusion over two commonly used models of syringe pump: the Graseby
Ambulatory MS16 and MS16A (which are set at mm per hour) and the MS26 (which is set
at mm per day). There have been a number of cases where confusion has led to over-
infusion and caused the death of patients.
The range of medical devices available is enormous and it would be impossible for nurses
to be adequately trained to use of them all. The MDA has published comprehensive
guidance on rationalising a number of different devices (MDA, 1998a), and encourages
limiting the number of devices used. At least one hospital has opted to use only one type of
syringe pump, despite having both available.
This appears to be an ideal solution, although it may have cost implications. In the long
term, costs can be reduced through collective ward or unit purchasing power. Purchasers
should ensure staff training is included in any package.
Failing to take action
Another issue nurses must consider is what is being monitored by a device and why. There
is no point monitoring something if no action is going to be taken when things go wrong.
The question all nurses should ask themselves is: 'Are the results within recognised limits?'
If they are not, take action. Do not just continue to monitor the situation. There have been
incidences reported on labour wards where cardiotocograph recordings have indicated
problems in labour but action was taken too late, resulting in stillbirths (MDA, 1998b).
Using devices correctly
Patients' needs can change, sometimes gradually but often very quickly. It is important that
patients are reassessed regularly to ensure that their needs are still being met by the devices
being used. This provision should ideally be made at the outset of treatment in the patient's
care plan.
If there are problems with any device, steps should be taken immediately to rectify the
situation. Replacement equipment should be obtained if necessary, and faulty devices
returned for repair through the appropriate channels. There have been occasions where
devices removed from one incident were used again through failure to follow the correct
procedure.
It is not the correct procedure to put the device in the corner and wait for someone else to
sort it out. There is no excuse for the reoccurrence of an incident involving a device that is
already known to be faulty.
Sometimes devices that would appear to be harmless can in fact be lethal if used
incorrectly, especially where there is a change in circumstances.
There have been a number of incidents reported to the MDA where children have been
injured, even killed, through the inappropriate use of a medical device. In one incident
reported recently, a young child was trapped under an electric bed. When the bed was
lowered, the child's skull was crushed.
This is where education and continual reassessment play an important part.
Any device-related incidents should be reported as soon as possible to the MDA to allow
for any necessary follow-up. This also allows for trends to be identified and again, where
indicated, further action to be taken.
It is now possible to report these incidents quickly and easily via the MDA's website at:
www.medical-devices.gov.uk
Record-keeping
It is essential, as in any health care intervention, for accurate documentation to be
maintained (Box 2). Records should identify the device and provide evidence that it has
been maintained in good condition. This includes recording the device's identification
details, for example, manufacturer, model and serial number. Health care professionals
should know who they have to report incidents to.
Incident reports to the MDA highlight that the routine servicing of devices in some areas
does not take place. It is acknowledged that servicing can be difficult, especially in
community settings when devices are sometimes located in remote places. However, it is
the health care professional's responsibility to ensure that any equipment has been regularly
serviced in accordance with the manufacturer's recommendations before it is used.
With technology changing faster than ever before, it is increasingly difficult to keep track
of instructions on the use of medical devices. This has serious implications for users.
Ideally all instructions should be available at all times, if necessary by keeping them
securely attached to the device.
After use
Nurses' responsibility does not end when their patient is no longer using a device. It is their
individual responsibility to ensure that the correct procedures are followed when disposing
of the device.
There should be a standard method for the return, disposal or cleaning of all devices (Box
1). An example where this works effectively is the disposal of syringes in sharps boxes.
Where there is a need for decontamination, servicing or repair, procedures are a little more
involved. An assessment of the equipment should be conducted, and if there are any faults
or damage these should be dealt with. This should be done in accordance with the
manufacturer's instructions.
Reuse of single-use devices
The reuse of medical devices supplied for single use is a serious concern. Nurses are often
involved in this practice without having considered the full consequences of their actions.
The MDA has received calls from health care professionals concerned that they are
encouraged to reuse devices designed for single use only.
The MDA has published guidelines that clearly state that devices designated for single use
must not be reused under any circumstances (MDA, 2000). Nurses who do not adhere to
these guidelines are putting their patients and themselves at unnecessary risk.

More Related Content

What's hot

2015 Intelligent Health Association Award
2015 Intelligent Health Association Award2015 Intelligent Health Association Award
2015 Intelligent Health Association Award
Ashley Fontenot
 
Quality Improvement Analysis 2009
Quality Improvement Analysis 2009Quality Improvement Analysis 2009
Quality Improvement Analysis 2009
Melissa Nichols
 

What's hot (7)

2015 Intelligent Health Association Award
2015 Intelligent Health Association Award2015 Intelligent Health Association Award
2015 Intelligent Health Association Award
 
Speak Up for Safe Health Worker
Speak Up for Safe Health WorkerSpeak Up for Safe Health Worker
Speak Up for Safe Health Worker
 
Sess08 My Back Up Brain
Sess08 My Back Up BrainSess08 My Back Up Brain
Sess08 My Back Up Brain
 
The effect of comp prescrp
The effect of comp prescrpThe effect of comp prescrp
The effect of comp prescrp
 
Quality Improvement Analysis 2009
Quality Improvement Analysis 2009Quality Improvement Analysis 2009
Quality Improvement Analysis 2009
 
Using big data in hospital fm to enhance employee productivity and quality of...
Using big data in hospital fm to enhance employee productivity and quality of...Using big data in hospital fm to enhance employee productivity and quality of...
Using big data in hospital fm to enhance employee productivity and quality of...
 
Patient Safety Presentation
Patient Safety PresentationPatient Safety Presentation
Patient Safety Presentation
 

Similar to How to use medical devices safely

4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx
4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx
4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx
gilbertkpeters11344
 
Medicolegal aspects of anaesthesia and dilemmas to anaesthetist
Medicolegal aspects of anaesthesia and dilemmas to anaesthetistMedicolegal aspects of anaesthesia and dilemmas to anaesthetist
Medicolegal aspects of anaesthesia and dilemmas to anaesthetist
narasimha reddy
 
Just-Glance-Reducing-Alarm-Fatigue (1)
Just-Glance-Reducing-Alarm-Fatigue (1)Just-Glance-Reducing-Alarm-Fatigue (1)
Just-Glance-Reducing-Alarm-Fatigue (1)
Jessica Hathaway, RN
 
1-Evidence-based practice is what keeps the health care system u
1-Evidence-based practice is what keeps the health care system u1-Evidence-based practice is what keeps the health care system u
1-Evidence-based practice is what keeps the health care system u
sandibabcock
 

Similar to How to use medical devices safely (20)

Safety monitoring and reporting of adverse events of medical devices national...
Safety monitoring and reporting of adverse events of medical devices national...Safety monitoring and reporting of adverse events of medical devices national...
Safety monitoring and reporting of adverse events of medical devices national...
 
Equipment failure hbi-sep-2017
Equipment failure hbi-sep-2017Equipment failure hbi-sep-2017
Equipment failure hbi-sep-2017
 
4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx
4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx
4. Use α=.01, and n=100Determine the Chi-Square value, and come to.docx
 
Barcoding 2.0: Improving Patient Safety
Barcoding 2.0: Improving Patient SafetyBarcoding 2.0: Improving Patient Safety
Barcoding 2.0: Improving Patient Safety
 
Medical Devices-WPS Office.pptx
Medical Devices-WPS Office.pptxMedical Devices-WPS Office.pptx
Medical Devices-WPS Office.pptx
 
Presentation: Increasing post-market vigilance requirements for medical devices
Presentation: Increasing post-market vigilance requirements for medical devicesPresentation: Increasing post-market vigilance requirements for medical devices
Presentation: Increasing post-market vigilance requirements for medical devices
 
Medicolegal aspects of anaesthesia and dilemmas to anaesthetist
Medicolegal aspects of anaesthesia and dilemmas to anaesthetistMedicolegal aspects of anaesthesia and dilemmas to anaesthetist
Medicolegal aspects of anaesthesia and dilemmas to anaesthetist
 
Recent advances in safer surgery
Recent advances in safer surgeryRecent advances in safer surgery
Recent advances in safer surgery
 
FDA the Gold Standard for Medical Device Safety
FDA the Gold Standard for Medical Device SafetyFDA the Gold Standard for Medical Device Safety
FDA the Gold Standard for Medical Device Safety
 
Medical Device FDA Regulations and Classifications infographic
Medical Device FDA Regulations and Classifications infographicMedical Device FDA Regulations and Classifications infographic
Medical Device FDA Regulations and Classifications infographic
 
Introduction to medical equipments safety and testing
Introduction to medical equipments safety and testingIntroduction to medical equipments safety and testing
Introduction to medical equipments safety and testing
 
Reliability of Med-El Cochlear Implants in children. The Romania Experience.
Reliability of Med-El Cochlear Implants in children. The Romania Experience.Reliability of Med-El Cochlear Implants in children. The Romania Experience.
Reliability of Med-El Cochlear Implants in children. The Romania Experience.
 
Patient safety
Patient safetyPatient safety
Patient safety
 
Just-Glance-Reducing-Alarm-Fatigue (1)
Just-Glance-Reducing-Alarm-Fatigue (1)Just-Glance-Reducing-Alarm-Fatigue (1)
Just-Glance-Reducing-Alarm-Fatigue (1)
 
Ipsg patient safety
Ipsg  patient safetyIpsg  patient safety
Ipsg patient safety
 
1-Evidence-based practice is what keeps the health care system u
1-Evidence-based practice is what keeps the health care system u1-Evidence-based practice is what keeps the health care system u
1-Evidence-based practice is what keeps the health care system u
 
TGA presentation: Postmarket Monitoring
TGA presentation: Postmarket MonitoringTGA presentation: Postmarket Monitoring
TGA presentation: Postmarket Monitoring
 
Were You Aware?
Were You Aware?Were You Aware?
Were You Aware?
 
1660 - Scope Article
1660 - Scope Article1660 - Scope Article
1660 - Scope Article
 
Patient Safety (Final).ppt
Patient Safety (Final).pptPatient Safety (Final).ppt
Patient Safety (Final).ppt
 

More from guayacan87

Tendencias en radiologia imagenes de radiologia
Tendencias en radiologia imagenes de radiologiaTendencias en radiologia imagenes de radiologia
Tendencias en radiologia imagenes de radiologia
guayacan87
 
Linac detectores y aceleradores que curan
Linac detectores y aceleradores que curanLinac detectores y aceleradores que curan
Linac detectores y aceleradores que curan
guayacan87
 
Acelerador lineal principios
Acelerador lineal principiosAcelerador lineal principios
Acelerador lineal principios
guayacan87
 
Ecografia dopler principios y aplicaciones
Ecografia dopler principios y aplicacionesEcografia dopler principios y aplicaciones
Ecografia dopler principios y aplicaciones
guayacan87
 
Parametros para la presentacion del examen final de la materia de equipos med...
Parametros para la presentacion del examen final de la materia de equipos med...Parametros para la presentacion del examen final de la materia de equipos med...
Parametros para la presentacion del examen final de la materia de equipos med...
guayacan87
 
Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...
Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...
Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...
guayacan87
 
Tema 8-interpretacion-de-una-gasometria
Tema 8-interpretacion-de-una-gasometriaTema 8-interpretacion-de-una-gasometria
Tema 8-interpretacion-de-una-gasometria
guayacan87
 
19039 automated analyser lect_1__instrumentation lect 2nd yr mt_2nd semester
19039 automated analyser  lect_1__instrumentation lect 2nd yr mt_2nd semester19039 automated analyser  lect_1__instrumentation lect 2nd yr mt_2nd semester
19039 automated analyser lect_1__instrumentation lect 2nd yr mt_2nd semester
guayacan87
 

More from guayacan87 (20)

Tendencias en radiologia imagenes de radiologia
Tendencias en radiologia imagenes de radiologiaTendencias en radiologia imagenes de radiologia
Tendencias en radiologia imagenes de radiologia
 
Evolucion tendencias hospitales_-mayo2012
Evolucion tendencias hospitales_-mayo2012Evolucion tendencias hospitales_-mayo2012
Evolucion tendencias hospitales_-mayo2012
 
Linac detectores y aceleradores que curan
Linac detectores y aceleradores que curanLinac detectores y aceleradores que curan
Linac detectores y aceleradores que curan
 
Acelerador lineal principios
Acelerador lineal principiosAcelerador lineal principios
Acelerador lineal principios
 
Medicina nuclear diagnostica pet y spect
Medicina nuclear diagnostica pet y spectMedicina nuclear diagnostica pet y spect
Medicina nuclear diagnostica pet y spect
 
Ecografia y lenguaje ecografico
Ecografia y lenguaje ecograficoEcografia y lenguaje ecografico
Ecografia y lenguaje ecografico
 
Ecografia dopler principios y aplicaciones
Ecografia dopler principios y aplicacionesEcografia dopler principios y aplicaciones
Ecografia dopler principios y aplicaciones
 
Principios de ultrasonido i
Principios de ultrasonido iPrincipios de ultrasonido i
Principios de ultrasonido i
 
Parametros para la presentacion del examen final de la materia de equipos med...
Parametros para la presentacion del examen final de la materia de equipos med...Parametros para la presentacion del examen final de la materia de equipos med...
Parametros para la presentacion del examen final de la materia de equipos med...
 
Manual electroforesis 38
Manual electroforesis 38Manual electroforesis 38
Manual electroforesis 38
 
Tema6
Tema6Tema6
Tema6
 
Cromatografia de gases
Cromatografia de gasesCromatografia de gases
Cromatografia de gases
 
Xs 800i-español
Xs 800i-españolXs 800i-español
Xs 800i-español
 
Guia7
Guia7Guia7
Guia7
 
Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...
Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...
Tema 15-tad-de-medicina-y-cirugía-de-urgencias.-alteraciones-ácido-base.-gaso...
 
Tema 8-interpretacion-de-una-gasometria
Tema 8-interpretacion-de-una-gasometriaTema 8-interpretacion-de-una-gasometria
Tema 8-interpretacion-de-una-gasometria
 
Analizadores de-laboratorio
Analizadores de-laboratorioAnalizadores de-laboratorio
Analizadores de-laboratorio
 
19039 automated analyser lect_1__instrumentation lect 2nd yr mt_2nd semester
19039 automated analyser  lect_1__instrumentation lect 2nd yr mt_2nd semester19039 automated analyser  lect_1__instrumentation lect 2nd yr mt_2nd semester
19039 automated analyser lect_1__instrumentation lect 2nd yr mt_2nd semester
 
28 pag 17_20_nuevas_alternativ seca
28 pag 17_20_nuevas_alternativ seca28 pag 17_20_nuevas_alternativ seca
28 pag 17_20_nuevas_alternativ seca
 
1.conceptos teoricos
1.conceptos teoricos1.conceptos teoricos
1.conceptos teoricos
 

Recently uploaded

Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.
MateoGardella
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
heathfieldcps1
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
QucHHunhnh
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
PECB
 
Gardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch LetterGardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch Letter
MateoGardella
 

Recently uploaded (20)

INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SD
 
Class 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdfClass 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdf
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17  How to Extend Models Using Mixin ClassesMixin Classes in Odoo 17  How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
Gardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch LetterGardella_PRCampaignConclusion Pitch Letter
Gardella_PRCampaignConclusion Pitch Letter
 

How to use medical devices safely

  • 1. HOW TO USE MEDICAL DEVICES SAFELY VOL: 97, ISSUE: 43, PAGE NO: 36 Wendy Murray, PGDip (health visiting), BSc, RGN, RM, RHV, is nursing officer Helen Glenister, PhD, MBA, BSc, RN, is nursing director of the Medical Devices Agency Unfortunately, when using medical devices, things do sometimes go wrong - and the outcome can be fatal. Last year, the Medical Devices Agency (MDA) received 7,249 adverse incident reports, of which 824 involved serious injury or fatality. The consequences of these incidents can be devastating for patients and their families, as well as for the health care professionals involved. Unfortunately, when using medical devices, things do sometimes go wrong - and the outcome can be fatal. Last year, the Medical Devices Agency (MDA) received 7,249 adverse incident reports, of which 824 involved serious injury or fatality. The consequences of these incidents can be devastating for patients and their families, as well as for the health care professionals involved. Assessing devices We all assume that in the event of an emergency, equipment will operate as we expect it to. However, this is not always the case. One incident reported to the MDA in March this year highlights the importance of checking equipment regularly to ensure that it is working correctly. The report came from an ambulance driver who had experienced problems with a suction pump unit during a resuscitation procedure. Possibly because of the displacement of a reusable suction canister, the unit failed to produce adequate suction. This may have been due to the canister being moved or knocked.
  • 2. This incident shows that after the initial assessment of any medical device (see last week's article, 'Safety tips', October 18, p24), regular checks need to be carried out to ensure the device is functioning correctly and to find out whether reassessment is required. Nurses must ask themselves two important questions when assessing devices. First, is the device doing what the manufacturer intended it to do, and, second, is it safe? Establishing these two points is vital if adverse incidents are to be avoided (Box 1). There have been a number of incidences where these simple questions could have avoided tragic incidents. Knowing what to check It is also important that nurses understand the instructions which accompany medical devices. If they misinterpret the manufacturer's instructions, they may unwittingly cause serious injury. For example, the MDA recently received reports that the prolonged use of pulse oximeter probes and the tape used to hold them in place had led to tissue necrosis in infants (MDA, 2001a). The manufacturer's instructions stated that the site should be inspected every four hours. The manufacturers had assumed that the inspection would result in a change of site. However, nurses inspecting the site believed they were doing so simply to check that the pulse oximeter was still in situ. To avoid confusion in future, revised instructions suggest that the site should be changed at least every four hours. Knowledge of the device being used Health care professionals need to be familiar with the devices they are using and should not attempt to operate any device they have not been fully trained to use. Reports to the MDA have included incidences where over-infusion of intravenous drugs has occurred because nurses did not follow the prescription, and administered them too quickly. Investigations revealed that in some cases the device had been over-infusing for some time without action being taken, despite written records clearly identifying over- infusion at the time (Glenister, 2000).
  • 3. There is ongoing confusion over two commonly used models of syringe pump: the Graseby Ambulatory MS16 and MS16A (which are set at mm per hour) and the MS26 (which is set at mm per day). There have been a number of cases where confusion has led to over- infusion and caused the death of patients. The range of medical devices available is enormous and it would be impossible for nurses to be adequately trained to use of them all. The MDA has published comprehensive guidance on rationalising a number of different devices (MDA, 1998a), and encourages limiting the number of devices used. At least one hospital has opted to use only one type of syringe pump, despite having both available. This appears to be an ideal solution, although it may have cost implications. In the long term, costs can be reduced through collective ward or unit purchasing power. Purchasers should ensure staff training is included in any package. Failing to take action Another issue nurses must consider is what is being monitored by a device and why. There is no point monitoring something if no action is going to be taken when things go wrong. The question all nurses should ask themselves is: 'Are the results within recognised limits?' If they are not, take action. Do not just continue to monitor the situation. There have been incidences reported on labour wards where cardiotocograph recordings have indicated problems in labour but action was taken too late, resulting in stillbirths (MDA, 1998b). Using devices correctly Patients' needs can change, sometimes gradually but often very quickly. It is important that patients are reassessed regularly to ensure that their needs are still being met by the devices being used. This provision should ideally be made at the outset of treatment in the patient's care plan.
  • 4. If there are problems with any device, steps should be taken immediately to rectify the situation. Replacement equipment should be obtained if necessary, and faulty devices returned for repair through the appropriate channels. There have been occasions where devices removed from one incident were used again through failure to follow the correct procedure. It is not the correct procedure to put the device in the corner and wait for someone else to sort it out. There is no excuse for the reoccurrence of an incident involving a device that is already known to be faulty. Sometimes devices that would appear to be harmless can in fact be lethal if used incorrectly, especially where there is a change in circumstances. There have been a number of incidents reported to the MDA where children have been injured, even killed, through the inappropriate use of a medical device. In one incident reported recently, a young child was trapped under an electric bed. When the bed was lowered, the child's skull was crushed. This is where education and continual reassessment play an important part. Any device-related incidents should be reported as soon as possible to the MDA to allow for any necessary follow-up. This also allows for trends to be identified and again, where indicated, further action to be taken. It is now possible to report these incidents quickly and easily via the MDA's website at: www.medical-devices.gov.uk Record-keeping It is essential, as in any health care intervention, for accurate documentation to be maintained (Box 2). Records should identify the device and provide evidence that it has been maintained in good condition. This includes recording the device's identification
  • 5. details, for example, manufacturer, model and serial number. Health care professionals should know who they have to report incidents to. Incident reports to the MDA highlight that the routine servicing of devices in some areas does not take place. It is acknowledged that servicing can be difficult, especially in community settings when devices are sometimes located in remote places. However, it is the health care professional's responsibility to ensure that any equipment has been regularly serviced in accordance with the manufacturer's recommendations before it is used. With technology changing faster than ever before, it is increasingly difficult to keep track of instructions on the use of medical devices. This has serious implications for users. Ideally all instructions should be available at all times, if necessary by keeping them securely attached to the device. After use Nurses' responsibility does not end when their patient is no longer using a device. It is their individual responsibility to ensure that the correct procedures are followed when disposing of the device. There should be a standard method for the return, disposal or cleaning of all devices (Box 1). An example where this works effectively is the disposal of syringes in sharps boxes. Where there is a need for decontamination, servicing or repair, procedures are a little more involved. An assessment of the equipment should be conducted, and if there are any faults or damage these should be dealt with. This should be done in accordance with the manufacturer's instructions. Reuse of single-use devices The reuse of medical devices supplied for single use is a serious concern. Nurses are often involved in this practice without having considered the full consequences of their actions.
  • 6. The MDA has received calls from health care professionals concerned that they are encouraged to reuse devices designed for single use only. The MDA has published guidelines that clearly state that devices designated for single use must not be reused under any circumstances (MDA, 2000). Nurses who do not adhere to these guidelines are putting their patients and themselves at unnecessary risk.