AN ION SOURCE FOR THE HVEE 14C ISOTOPE RATIO MASS SPECTROMETER
FOR BIOMEDICALAPPLICATIONS
DIRKJ. W. MOUS, WIM FOKKER, REIN VANDENBROEK, RON KOOPMANS
High Voltage Engineering Europa B.V., PO Box 99, 3800 AB Amersfoort, The Netherlands
CHRISTOPHER BRONKRAMSEYI andR. E. M. HEDGESI
Oxford Radiocarbon Accelerator Unit, Oxford University, 6 Keble Road, Oxford OX13QJ
United Kingdom
ABSTRACT. During the past two decades, accelerator mass spectrometry (AMS) has allowed major developments in many
areas of geosciences and archaeology. In the near future, AMS should realize a similar potential in the field of biomedical
research, leading ultimately to clinical applications. For such applications, the required instrument differs significantly from
thatpresently used in the field of 14C dating. Whereas the needed accuracy and sensitivity is more than an order ofmagnitude
less demanding than that for present state-of-the-art14C instrumentation, the widespread acceptance of 14C AMS in biomed-
ical research will require AMS spectrometers that are small, simple to operate and capable of handling CO2samples. In order
to satisfy these demands, HVEE has developed a compact 14C AMS spectrometer dedicated to biomedical research. The
instrument consists of a compact accelerator with a footprint of 2.25 x 1.25 m and an ion source that features direct CO2
acceptance and optimal user friendliness. Having previously described the layout and design of the accelerator, we here dis-
cuss progress on the accelerator and present the design and first results of the CO2 ion source.
INTRODUCTION
In the past two decades, AMS has developed into a mature scientific tool for a variety of applica-
tions in the geosciences and archaeology. The usefulness ofAMS has been recognized in other dis-
ciplines as well, notably biomedical research (Davis et al. 1990; Davis 1994), as demonstrated by an
exponential increase in the number of contributions in this field at recent AMS conferences in Can-
berra-Sydney (Fifield et a1.1994) and Tucson (Jull, Beck and Burr 1997).
The required instrumentation for biomedical research differs significantly from that presently used
in the field of 14C dating. The use ofAMS instrumentation withinthe biomedical community is pres-
ently handicapped by the size of existing systems and their need for expensive support personnel;
widespread acceptance will require small AMS spectrometers that need little or no maintenance and
that are easy to operate. Size reduction of the instrument can be achieved by sacrificing sensitivity
and accuracy, since these specifications are at least one order of magnitude less demanding in bio-
medical research. Throughput requirements are an order of magnitude higher, however.
The present AMS technique relies on a laborious graphitization procedure that is unknown to the
biomedical community. The introduction of AMS instrumentation to this community would be
eased by the availability of ion sources that directly accept C02, enabling direct interfacing with a
CHN analyzer or a gas chromatograph.
HVEE has developed a compact AMS spectrometer dedicated to biomedical research that over-
comes the shortcomings of the presently available AMS instruments. With a footprint of2.25 x 1.25
m, it is easy to operate and capable of analyzing CO2 samples. Previously (Mows et al. 1997) we
reported on the design and layout of the instrument, whose accuracy and throughput have yet to be
fully characterized. Here we discuss progress on the system and present the development and first
results of a compact cesium-type sputter source that directly accepts CO2.
1These authors were exclusively involvedin the design and development of the CO2 ion source.
Proceedings ofthe 16th International 14C Conference, edited by W. G. Mookand J. van derPlicht
RADIOCARBON, Vol. 40, No. 1, 1998, P. 283-288 283
284 D. J. W. Mous et al.
THE CO2 ION SOURCE
The design of the CO2 ion source results from cooperation between HVEE and Oxford University.
Its design is based on that of the OXGIS source developed at Oxford (Bronk 1990), with improve-
ments seeking to minimize memory effects and optimize user friendliness. The operation character-
istics of the ion source are reflected in the following features:
acceptance ofboth graphite samples and CO2 samples admitted from ground potential;
an easily exchanged 50-sample carousel;
source embodiment at ground potential ensuring easy and safe operation;
almost complete absence ofradiation hazard as a result of a stainless-steel source embodiment;
optimal vacuum conditions for low memory effects;
easy, quick source maintenance via sideways disassembly of the source-head.
A schematic layout ofthe Cs sputter-type ion source is shown in Figure 1. Its principle of operation
is well understood and described extensively in the literature (e.g., Alton 1990). In this section we
concentrate on the technical aspects ofthe new source design.
»-360 l/s TURBINE PUMP
CAROUSEL
COOLED
TARGET HOLDER
ION BEAM
iR
H. CARRIER
C02
SOURCE HEAD FLANGE
Cs RESERVOIR
PRESSURE RELIEF HV INSULATION TUBE CAPILLARY SOURCE HEAD
Fig. 1. Schematic layout ofthe CO2 ion source
The stainless-steel source embodiment, including carousel, is at ground potential during operation,
avoiding the necessity ofa large isolation cage, thereby reducing the space requirements of the total
assembly. Isolation of the extraction and target voltage is achieved by insulators within the source
embodiment (Fig. 1). All high-voltage connections to the source, including the Cs reservoir, the CO2
feed tube and the cooling lines, are brought together on the source-head insulation flange, keeping
the space needed for external high-voltage isolation to a minimum. The source-head (Fig. 2) can be
pulled sideways from the source embodiment, making maintenance easy and quick.
A 360 L sturbomolecular pump is directly mounted on the source embodiment, just above the
source-head, providing optimal pumping speed in the source area. This fulfills an important condi-
Ion Sourcefor the HVEE Mass Spectrometer 285
I? Ifl M
P°l+tt 11 p
t`31
Fig. 2. Photograph ofthe source-head
tion for low memory in the case of CO2 targets, which is of paramount importance for biomedical
applications, where the 14C content of a sample can vary by many orders of magnitude.
The sample carousel can accommodate up to 50 samples and is isolated from the source by a 100-
mm diameter gate-valve, ensuring good vacuum conditions at the carousel at all times. A pneumat-
ically operated cylinder transports the sample under investigation into the ion source.
The source accepts graphite as well as CO2 samples. Solid graphite sample material is pressed in a
target holder in a similar manner as is done in most modern AMS laboratories. CO2 samples can be
admitted into the ion source from ground potential. A He carrier ofca. 1 sccm transports the CO2 via
an isolation tube and a stainless-steel capillary into the ion source (Fig. 1). Here, the CO2 is absorbed
on the surface ofa titanium target, where it is sputtered and ionized to form a negative carbon beam.
During gas admission, the vacuum level in the source is maintained at ca. 1 x 10-5 mbar.
OPERATION AND FIRST RESULTS
The source is operated at a target potential of6 kV and 19 kV extraction, giving a 25 keV ion energy.
Analyzed 12C currents from graphite samples were measured at between 50 and 100 .tA, with mod-
erate Cs supply into the ion source. No attempts have been made to push the source to its limits, but
currents well in excess of 100µA should be achievable. Figure 3 shows the analyzed 12C- current
for different graphite samples as a function of time. After target insertion, the 12C current rises
quickly, stabilizing within ca. 30 s. Furthermore, it can be seenfrom the figure that the sample trans-
fer time (for neighboring samples) is ca. 7 s. During sample exchange (t = -55 s in Fig. 3), a few,uA
of 12C-current is still present. This current probably results from the distorted ion optics ofthe sput-
tering Cs beam due to the absence ofthe targetin the source and does not reflect actual source back-
ground.
286
50
40
30
20
10
0
----------- --------------------- ---------- ----------
I I
^_--'___--L- _---1--_-------L'-'--'--'-1 __ -------- J
-
------- J-----
-----f----------f----------t-------------------- ------- -------
-___----__L--___---_-1--_---_--_-__-__-__-J __--_-_ - ------
sets;
--- I
;7
__--___--_r____--____T____----__T--__-_-_--1 ------- -1_-_-_
-
------L----------L'---------L'-_----------------- ------- -J-__
-
-------'r----------'r----------' -------' ------ ------ ----
0 10 20 30 40 50 60 70
Time (seconds)
Fig. 3. Analyzed 12C- ion source current as a function of time for two different graphite samples
The Cs spot size on the target was measured at ca. 0.8 mm in diameter, which corresponds reason-
ably well to the design value of 0.6 mm. The emittance ofthe source for a 50µA 12C beam was mea-
sured at ca. 6 mm mrad MeV- 5.
Exchanging of the carousel, which houses 50 targets, is easy; pump-down takesca. 5 min before the
next analysis can be started.
The source was briefly tested with CO2 samples, using a He carrier of ca. 1 sccm to transport the
CO2 gas from ground potential into the source (Fig. 1). Analyzed 12C currents of 5-10 uA were
achieved, which is ca. one order ofmagnitude less than currents from graphite samples. Further tests
in progress seek to optimize source output and to quantify the memory effects of the source in the
case of CO2 samples.
PROGRESS ON THE 14C AMS SPECTROMETER
Figure 4 shows a schematic diagramof the HVEE biomedical 14C AMS spectrometer. Negative ions
are extracted from the ion source and mass analyzed by a small 90° permanent magnet. At the exit
ofthe magnet, 12C ions are collected in a Faraday cup for normalization; ions with a mass of 14 pro-
ceed to pass a 90° electrostatic analyzer (ESA), the tandem accelerator with a 180° ESA in its ter-
minal and a magnetic deflector before they are collected in a particle detector. A detailed descrip-
tion, together with the principles of background reduction of the instrument, is given elsewhere
(Mous et aL 1997).
Since the early tests on the system, various improvements have been made to optimize the beam
optics. A remotely controlled power supply in the high-voltage terminal for the adjustment of the
D. J. W. Mous etal.
70
60
ESA and a gridded lens following the high-energy (HE) tube were incorporated.
Ion Sourcefor theHVEE Mass Spectrometer 287
INLET
180 ELECTROSTATIC
DEFLECTOR
GC OR CNN
HE ACCEL. TUBE
GRIDOED
LENS
1111111 n
-----------
Fig. 4. Schematic layout of the HVEE 14C AMS spectrometer
ANALYSIS
MAGNET IONIZATION
OR
SSB DETECTOR
To evaluate the ion optical quality ofthe accelerator, the 13C2+ current in the HE Faraday cuplocated
just in front of the detector was measured during a scan of the low energy (LE) steerer and the 180°
ESA in the terminal. Figure 5 presents the results. Both traces show a flat-top tuned transmission,
indicating that the beam is transported through the system without loss. Furthermore, the figure indi-
cates the usefulness of the LE aperture during tuning of the system.
T
Injector steerer voltage --* 1800 deflector voltage -*
Fig. 5. HE Faraday cup current as a function of injector steerer voltage and 180° electrostatic
analyzer voltage
288 D. J. W. Mous et at.
CONCLUSION
The cooperation between HVEE and Oxford University has resulted in a unique, compact ion source
that accepts both graphite samples and CO2 samples from ground potential. The source will be
applied on the HVEE 14C AMS spectrometer dedicated to biomedical research. Early tests have
showed that analyzed 12C- currents of 50-100 ,uA are easily extracted from graphite samples,
whereas 5-101uA is obtained from CO2 samples. ,
Beam transmission of the HVEE 14C AMS spectrometer is flat-top tuned, which guarantees that
accurate 14C analysis can be achieved.
REFERENCES
Alton, G. D. 1990 Recent advancementsin high intensity
heavy negative ion sources based on the sputter prin-
ciple. Nuclear Instruments and Methods in Physics
Research A287:139-149.
Bronk, C. R. and Hedges, R. E. M. 1990 A gaseous ion
source for routine AMS radiocarbon dating. In Yiou,
F. and Raisbeck, G. M., eds., Proceedings of the 5th
International Conference on Accelerator Mass Spec-
trometry. Nuclear Instruments and Methods in Phys-
ics Research B52: 322-326.
Davis, J. C. 1994 AMS beyond 2000. In Fifeld, L. K.,
Fink, D., Sic, S. H. and Tuniz, C., eds., Proceedings of
the 6th International Conference on Accelerator Mass
Spectrometry. Nuclear Instruments and Methods in
Physics Research B92:1-6.
Davis, J. C., Proctor, I. D., Southon, J. R., Caffee, M. W.,
Heikkinen, D. W., Roberts, M. L., Moore, T. L., Tur-
teltaub, K. W., Nelson, D. E., Loyd, D. H. and Vogel,
J. S. 1990 LLNIIUC AMS facility and research pro-
gram. In Yiou, F. and Raisbeck, G. M., eds., Proceed-
ings of the 5th International Conference on Accelera-
tor Mass Spectrometry. Nuclear Instruments and
Methods inPhysics Research B52: 269-272.
Fifield, L. K., Fink, D., Sic, S. H. andTniz, C., eds. 1994
Proceedings of the 6th International Conference on
Accelerator Mass Spectrometry. NuclearInstruments
and Methods in Physics B92.
Jull, A. J. T., Beck, J. W. and Burr, G. S., eds. 1997 Pro-
ceedings of the 7th International Conference on Ac-
celerator Mass Spectrometry. Nuclear Instruments
andMethods B123.
Mous, D. J. W., Purser, K. H., Fokker, W., van den Broek,
R. and Koopmans, R. B. 1997 A compact 14C Isotope
Ratio MassSpectrometer for Biomedical applications.
In Jull, A. J. T., Beck, J. W. and Burr, G.S., eds., Pro-
ceedings of the 7th International Conference on Ac-
celerator Mass Spectrometry. Nuclear Instruments
andMethods in PhysicsResearch B123:159-162.

More Related Content

What's hot

Haiyang-2015 Symposium Poster (9 22 modified)-Final
Haiyang-2015 Symposium Poster (9 22 modified)-FinalHaiyang-2015 Symposium Poster (9 22 modified)-Final
Haiyang-2015 Symposium Poster (9 22 modified)-Final
Haiyang Sheng
 
Importance of plasma conditions for vertically aligned carbon nanowalls synth...
Importance of plasma conditions for vertically aligned carbon nanowalls synth...Importance of plasma conditions for vertically aligned carbon nanowalls synth...
Importance of plasma conditions for vertically aligned carbon nanowalls synth...
Marcel Mesko
 
Poster_Electrocoagulation Final Version
Poster_Electrocoagulation Final VersionPoster_Electrocoagulation Final Version
Poster_Electrocoagulation Final Version
Jabari Lee, E.I.T
 
Walker Electrochemical Paper
Walker Electrochemical PaperWalker Electrochemical Paper
Walker Electrochemical Paper
Patrick Walker
 
A Novel Method of Synthesizing Graphene for Electronic Device Applications
A Novel Method of Synthesizing Graphene for Electronic Device ApplicationsA Novel Method of Synthesizing Graphene for Electronic Device Applications
A Novel Method of Synthesizing Graphene for Electronic Device Applications
Mariana Amorim Fraga
 
sci comm research-vivek
sci comm research-viveksci comm research-vivek
sci comm research-vivek
Vivek Saraswat
 

What's hot (15)

10.1088@1361 6595@ab123c
10.1088@1361 6595@ab123c10.1088@1361 6595@ab123c
10.1088@1361 6595@ab123c
 
Fall2015
Fall2015Fall2015
Fall2015
 
Haiyang-2015 Symposium Poster (9 22 modified)-Final
Haiyang-2015 Symposium Poster (9 22 modified)-FinalHaiyang-2015 Symposium Poster (9 22 modified)-Final
Haiyang-2015 Symposium Poster (9 22 modified)-Final
 
Importance of plasma conditions for vertically aligned carbon nanowalls synth...
Importance of plasma conditions for vertically aligned carbon nanowalls synth...Importance of plasma conditions for vertically aligned carbon nanowalls synth...
Importance of plasma conditions for vertically aligned carbon nanowalls synth...
 
Ce icp-ms
Ce icp-msCe icp-ms
Ce icp-ms
 
Sparc Ebit September 2008
Sparc Ebit September 2008Sparc Ebit September 2008
Sparc Ebit September 2008
 
Poster_Electrocoagulation Final Version
Poster_Electrocoagulation Final VersionPoster_Electrocoagulation Final Version
Poster_Electrocoagulation Final Version
 
Final REU Poster
Final REU PosterFinal REU Poster
Final REU Poster
 
C121326
C121326C121326
C121326
 
Photothermal Boiling in Aqueous Nanofluids
Photothermal Boiling in Aqueous NanofluidsPhotothermal Boiling in Aqueous Nanofluids
Photothermal Boiling in Aqueous Nanofluids
 
CAMS General Assembly Chabrillat
CAMS  General Assembly ChabrillatCAMS  General Assembly Chabrillat
CAMS General Assembly Chabrillat
 
Walker Electrochemical Paper
Walker Electrochemical PaperWalker Electrochemical Paper
Walker Electrochemical Paper
 
SDP poster
SDP posterSDP poster
SDP poster
 
A Novel Method of Synthesizing Graphene for Electronic Device Applications
A Novel Method of Synthesizing Graphene for Electronic Device ApplicationsA Novel Method of Synthesizing Graphene for Electronic Device Applications
A Novel Method of Synthesizing Graphene for Electronic Device Applications
 
sci comm research-vivek
sci comm research-viveksci comm research-vivek
sci comm research-vivek
 

Viewers also liked

Concurso de Leitura
Concurso de LeituraConcurso de Leitura
Concurso de Leitura
O Ciclista
 
HQ49assoportrP9P10v2
HQ49assoportrP9P10v2HQ49assoportrP9P10v2
HQ49assoportrP9P10v2
Iain Bitran
 
Flex3中文教程
Flex3中文教程Flex3中文教程
Flex3中文教程
yiditushe
 

Viewers also liked (17)

The Secret Garden
The Secret GardenThe Secret Garden
The Secret Garden
 
Multiple Contract Pharmacies in a Disproportionate Share Hospital System
Multiple Contract Pharmacies in a Disproportionate Share Hospital SystemMultiple Contract Pharmacies in a Disproportionate Share Hospital System
Multiple Contract Pharmacies in a Disproportionate Share Hospital System
 
Competency Based E Portfolios Slide Share Final March 26 2010
Competency Based E Portfolios Slide Share Final March 26 2010Competency Based E Portfolios Slide Share Final March 26 2010
Competency Based E Portfolios Slide Share Final March 26 2010
 
Concurso de Leitura
Concurso de LeituraConcurso de Leitura
Concurso de Leitura
 
Pasi Sahlberg, Finland
Pasi Sahlberg, FinlandPasi Sahlberg, Finland
Pasi Sahlberg, Finland
 
Instrucciones Originales
Instrucciones OriginalesInstrucciones Originales
Instrucciones Originales
 
edging
edgingedging
edging
 
Using FME to Build an Enterprise Geodatabase
Using FME to Build an Enterprise GeodatabaseUsing FME to Build an Enterprise Geodatabase
Using FME to Build an Enterprise Geodatabase
 
HQ49assoportrP9P10v2
HQ49assoportrP9P10v2HQ49assoportrP9P10v2
HQ49assoportrP9P10v2
 
Finance Bill 2010 Kalyan 97 2003
Finance  Bill 2010  Kalyan 97 2003Finance  Bill 2010  Kalyan 97 2003
Finance Bill 2010 Kalyan 97 2003
 
Restaurant Daily Press Choice Awards
Restaurant  Daily  Press  Choice  AwardsRestaurant  Daily  Press  Choice  Awards
Restaurant Daily Press Choice Awards
 
Agua
AguaAgua
Agua
 
Managing Technology on a Budget
Managing Technology on a BudgetManaging Technology on a Budget
Managing Technology on a Budget
 
Flex3中文教程
Flex3中文教程Flex3中文教程
Flex3中文教程
 
Linked In Intro Power Point
Linked In Intro Power PointLinked In Intro Power Point
Linked In Intro Power Point
 
القاعدة الثالثة من قواعد الميزان النبري كاملة
القاعدة الثالثة من قواعد الميزان النبري كاملةالقاعدة الثالثة من قواعد الميزان النبري كاملة
القاعدة الثالثة من قواعد الميزان النبري كاملة
 
cv only
cv onlycv only
cv only
 

Similar to publicatie C14_2014-2263-1-PB

4.2.cappelli 06(3)
4.2.cappelli 06(3)4.2.cappelli 06(3)
4.2.cappelli 06(3)
ELIMENG
 
4.2.cappelli 06(2)
4.2.cappelli 06(2)4.2.cappelli 06(2)
4.2.cappelli 06(2)
ELIMENG
 
4.2.cappelli 06(1)
4.2.cappelli 06(1)4.2.cappelli 06(1)
4.2.cappelli 06(1)
ELIMENG
 
4.2.cappelli 06
4.2.cappelli 064.2.cappelli 06
4.2.cappelli 06
ELIMENG
 
Optimal Generation of 254nm ultraviolet radiation
Optimal Generation of 254nm ultraviolet radiationOptimal Generation of 254nm ultraviolet radiation
Optimal Generation of 254nm ultraviolet radiation
IOSR Journals
 
Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...
Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...
Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...
IOSRJAP
 
Potential enhancement of thermoelectric energy conversion in cobaltite superl...
Potential enhancement of thermoelectric energy conversion in cobaltite superl...Potential enhancement of thermoelectric energy conversion in cobaltite superl...
Potential enhancement of thermoelectric energy conversion in cobaltite superl...
Anastasios Englezos
 
Wamg -graphene tubes_Nanoscale
Wamg -graphene tubes_NanoscaleWamg -graphene tubes_Nanoscale
Wamg -graphene tubes_Nanoscale
Haiyang Sheng
 

Similar to publicatie C14_2014-2263-1-PB (20)

Microscopy microanalysis microstructures_the european physical journal_applie...
Microscopy microanalysis microstructures_the european physical journal_applie...Microscopy microanalysis microstructures_the european physical journal_applie...
Microscopy microanalysis microstructures_the european physical journal_applie...
 
Perovskite Solar Cell
Perovskite Solar CellPerovskite Solar Cell
Perovskite Solar Cell
 
4.2.cappelli 06(3)
4.2.cappelli 06(3)4.2.cappelli 06(3)
4.2.cappelli 06(3)
 
4.2.cappelli 06(2)
4.2.cappelli 06(2)4.2.cappelli 06(2)
4.2.cappelli 06(2)
 
4.2.cappelli 06(1)
4.2.cappelli 06(1)4.2.cappelli 06(1)
4.2.cappelli 06(1)
 
4.2.cappelli 06
4.2.cappelli 064.2.cappelli 06
4.2.cappelli 06
 
Optimal Generation of 254nm ultraviolet radiation
Optimal Generation of 254nm ultraviolet radiationOptimal Generation of 254nm ultraviolet radiation
Optimal Generation of 254nm ultraviolet radiation
 
Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...
Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...
Low Cost Synthesis of Single Walled Carbon Nanotubes from Coal Tar Using Arc ...
 
Perovskite solar cells
Perovskite solar cellsPerovskite solar cells
Perovskite solar cells
 
Estimation of flow accelerated corrosion (fac) in feeder pipes using cf dd so...
Estimation of flow accelerated corrosion (fac) in feeder pipes using cf dd so...Estimation of flow accelerated corrosion (fac) in feeder pipes using cf dd so...
Estimation of flow accelerated corrosion (fac) in feeder pipes using cf dd so...
 
THZ_paper
THZ_paperTHZ_paper
THZ_paper
 
Production of CH4 and C2 hydrocarbons by axial and radial pulse H2/CO2 discha...
Production of CH4 and C2 hydrocarbons by axial and radial pulse H2/CO2 discha...Production of CH4 and C2 hydrocarbons by axial and radial pulse H2/CO2 discha...
Production of CH4 and C2 hydrocarbons by axial and radial pulse H2/CO2 discha...
 
j.ijrmhm.2020.105280.pdf
j.ijrmhm.2020.105280.pdfj.ijrmhm.2020.105280.pdf
j.ijrmhm.2020.105280.pdf
 
International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)
 
Potential enhancement of thermoelectric energy conversion in cobaltite superl...
Potential enhancement of thermoelectric energy conversion in cobaltite superl...Potential enhancement of thermoelectric energy conversion in cobaltite superl...
Potential enhancement of thermoelectric energy conversion in cobaltite superl...
 
Federico Capasso: Photonics as Design
Federico Capasso: Photonics as DesignFederico Capasso: Photonics as Design
Federico Capasso: Photonics as Design
 
APGC-MS, ASAP-MS, and MAI-MS: Expanding the Horizons of API-MS
APGC-MS, ASAP-MS, and MAI-MS: Expanding the Horizons of API-MSAPGC-MS, ASAP-MS, and MAI-MS: Expanding the Horizons of API-MS
APGC-MS, ASAP-MS, and MAI-MS: Expanding the Horizons of API-MS
 
20040086056
2004008605620040086056
20040086056
 
Modeling and Optimization of Cold Crucible Furnaces for Melting Metals
Modeling and Optimization of Cold Crucible Furnaces for Melting MetalsModeling and Optimization of Cold Crucible Furnaces for Melting Metals
Modeling and Optimization of Cold Crucible Furnaces for Melting Metals
 
Wamg -graphene tubes_Nanoscale
Wamg -graphene tubes_NanoscaleWamg -graphene tubes_Nanoscale
Wamg -graphene tubes_Nanoscale
 

publicatie C14_2014-2263-1-PB

  • 1. AN ION SOURCE FOR THE HVEE 14C ISOTOPE RATIO MASS SPECTROMETER FOR BIOMEDICALAPPLICATIONS DIRKJ. W. MOUS, WIM FOKKER, REIN VANDENBROEK, RON KOOPMANS High Voltage Engineering Europa B.V., PO Box 99, 3800 AB Amersfoort, The Netherlands CHRISTOPHER BRONKRAMSEYI andR. E. M. HEDGESI Oxford Radiocarbon Accelerator Unit, Oxford University, 6 Keble Road, Oxford OX13QJ United Kingdom ABSTRACT. During the past two decades, accelerator mass spectrometry (AMS) has allowed major developments in many areas of geosciences and archaeology. In the near future, AMS should realize a similar potential in the field of biomedical research, leading ultimately to clinical applications. For such applications, the required instrument differs significantly from thatpresently used in the field of 14C dating. Whereas the needed accuracy and sensitivity is more than an order ofmagnitude less demanding than that for present state-of-the-art14C instrumentation, the widespread acceptance of 14C AMS in biomed- ical research will require AMS spectrometers that are small, simple to operate and capable of handling CO2samples. In order to satisfy these demands, HVEE has developed a compact 14C AMS spectrometer dedicated to biomedical research. The instrument consists of a compact accelerator with a footprint of 2.25 x 1.25 m and an ion source that features direct CO2 acceptance and optimal user friendliness. Having previously described the layout and design of the accelerator, we here dis- cuss progress on the accelerator and present the design and first results of the CO2 ion source. INTRODUCTION In the past two decades, AMS has developed into a mature scientific tool for a variety of applica- tions in the geosciences and archaeology. The usefulness ofAMS has been recognized in other dis- ciplines as well, notably biomedical research (Davis et al. 1990; Davis 1994), as demonstrated by an exponential increase in the number of contributions in this field at recent AMS conferences in Can- berra-Sydney (Fifield et a1.1994) and Tucson (Jull, Beck and Burr 1997). The required instrumentation for biomedical research differs significantly from that presently used in the field of 14C dating. The use ofAMS instrumentation withinthe biomedical community is pres- ently handicapped by the size of existing systems and their need for expensive support personnel; widespread acceptance will require small AMS spectrometers that need little or no maintenance and that are easy to operate. Size reduction of the instrument can be achieved by sacrificing sensitivity and accuracy, since these specifications are at least one order of magnitude less demanding in bio- medical research. Throughput requirements are an order of magnitude higher, however. The present AMS technique relies on a laborious graphitization procedure that is unknown to the biomedical community. The introduction of AMS instrumentation to this community would be eased by the availability of ion sources that directly accept C02, enabling direct interfacing with a CHN analyzer or a gas chromatograph. HVEE has developed a compact AMS spectrometer dedicated to biomedical research that over- comes the shortcomings of the presently available AMS instruments. With a footprint of2.25 x 1.25 m, it is easy to operate and capable of analyzing CO2 samples. Previously (Mows et al. 1997) we reported on the design and layout of the instrument, whose accuracy and throughput have yet to be fully characterized. Here we discuss progress on the system and present the development and first results of a compact cesium-type sputter source that directly accepts CO2. 1These authors were exclusively involvedin the design and development of the CO2 ion source. Proceedings ofthe 16th International 14C Conference, edited by W. G. Mookand J. van derPlicht RADIOCARBON, Vol. 40, No. 1, 1998, P. 283-288 283
  • 2. 284 D. J. W. Mous et al. THE CO2 ION SOURCE The design of the CO2 ion source results from cooperation between HVEE and Oxford University. Its design is based on that of the OXGIS source developed at Oxford (Bronk 1990), with improve- ments seeking to minimize memory effects and optimize user friendliness. The operation character- istics of the ion source are reflected in the following features: acceptance ofboth graphite samples and CO2 samples admitted from ground potential; an easily exchanged 50-sample carousel; source embodiment at ground potential ensuring easy and safe operation; almost complete absence ofradiation hazard as a result of a stainless-steel source embodiment; optimal vacuum conditions for low memory effects; easy, quick source maintenance via sideways disassembly of the source-head. A schematic layout ofthe Cs sputter-type ion source is shown in Figure 1. Its principle of operation is well understood and described extensively in the literature (e.g., Alton 1990). In this section we concentrate on the technical aspects ofthe new source design. »-360 l/s TURBINE PUMP CAROUSEL COOLED TARGET HOLDER ION BEAM iR H. CARRIER C02 SOURCE HEAD FLANGE Cs RESERVOIR PRESSURE RELIEF HV INSULATION TUBE CAPILLARY SOURCE HEAD Fig. 1. Schematic layout ofthe CO2 ion source The stainless-steel source embodiment, including carousel, is at ground potential during operation, avoiding the necessity ofa large isolation cage, thereby reducing the space requirements of the total assembly. Isolation of the extraction and target voltage is achieved by insulators within the source embodiment (Fig. 1). All high-voltage connections to the source, including the Cs reservoir, the CO2 feed tube and the cooling lines, are brought together on the source-head insulation flange, keeping the space needed for external high-voltage isolation to a minimum. The source-head (Fig. 2) can be pulled sideways from the source embodiment, making maintenance easy and quick. A 360 L sturbomolecular pump is directly mounted on the source embodiment, just above the source-head, providing optimal pumping speed in the source area. This fulfills an important condi-
  • 3. Ion Sourcefor the HVEE Mass Spectrometer 285 I? Ifl M P°l+tt 11 p t`31 Fig. 2. Photograph ofthe source-head tion for low memory in the case of CO2 targets, which is of paramount importance for biomedical applications, where the 14C content of a sample can vary by many orders of magnitude. The sample carousel can accommodate up to 50 samples and is isolated from the source by a 100- mm diameter gate-valve, ensuring good vacuum conditions at the carousel at all times. A pneumat- ically operated cylinder transports the sample under investigation into the ion source. The source accepts graphite as well as CO2 samples. Solid graphite sample material is pressed in a target holder in a similar manner as is done in most modern AMS laboratories. CO2 samples can be admitted into the ion source from ground potential. A He carrier ofca. 1 sccm transports the CO2 via an isolation tube and a stainless-steel capillary into the ion source (Fig. 1). Here, the CO2 is absorbed on the surface ofa titanium target, where it is sputtered and ionized to form a negative carbon beam. During gas admission, the vacuum level in the source is maintained at ca. 1 x 10-5 mbar. OPERATION AND FIRST RESULTS The source is operated at a target potential of6 kV and 19 kV extraction, giving a 25 keV ion energy. Analyzed 12C currents from graphite samples were measured at between 50 and 100 .tA, with mod- erate Cs supply into the ion source. No attempts have been made to push the source to its limits, but currents well in excess of 100µA should be achievable. Figure 3 shows the analyzed 12C- current for different graphite samples as a function of time. After target insertion, the 12C current rises quickly, stabilizing within ca. 30 s. Furthermore, it can be seenfrom the figure that the sample trans- fer time (for neighboring samples) is ca. 7 s. During sample exchange (t = -55 s in Fig. 3), a few,uA of 12C-current is still present. This current probably results from the distorted ion optics ofthe sput- tering Cs beam due to the absence ofthe targetin the source and does not reflect actual source back- ground.
  • 4. 286 50 40 30 20 10 0 ----------- --------------------- ---------- ---------- I I ^_--'___--L- _---1--_-------L'-'--'--'-1 __ -------- J - ------- J----- -----f----------f----------t-------------------- ------- ------- -___----__L--___---_-1--_---_--_-__-__-__-J __--_-_ - ------ sets; --- I ;7 __--___--_r____--____T____----__T--__-_-_--1 ------- -1_-_-_ - ------L----------L'---------L'-_----------------- ------- -J-__ - -------'r----------'r----------' -------' ------ ------ ---- 0 10 20 30 40 50 60 70 Time (seconds) Fig. 3. Analyzed 12C- ion source current as a function of time for two different graphite samples The Cs spot size on the target was measured at ca. 0.8 mm in diameter, which corresponds reason- ably well to the design value of 0.6 mm. The emittance ofthe source for a 50µA 12C beam was mea- sured at ca. 6 mm mrad MeV- 5. Exchanging of the carousel, which houses 50 targets, is easy; pump-down takesca. 5 min before the next analysis can be started. The source was briefly tested with CO2 samples, using a He carrier of ca. 1 sccm to transport the CO2 gas from ground potential into the source (Fig. 1). Analyzed 12C currents of 5-10 uA were achieved, which is ca. one order ofmagnitude less than currents from graphite samples. Further tests in progress seek to optimize source output and to quantify the memory effects of the source in the case of CO2 samples. PROGRESS ON THE 14C AMS SPECTROMETER Figure 4 shows a schematic diagramof the HVEE biomedical 14C AMS spectrometer. Negative ions are extracted from the ion source and mass analyzed by a small 90° permanent magnet. At the exit ofthe magnet, 12C ions are collected in a Faraday cup for normalization; ions with a mass of 14 pro- ceed to pass a 90° electrostatic analyzer (ESA), the tandem accelerator with a 180° ESA in its ter- minal and a magnetic deflector before they are collected in a particle detector. A detailed descrip- tion, together with the principles of background reduction of the instrument, is given elsewhere (Mous et aL 1997). Since the early tests on the system, various improvements have been made to optimize the beam optics. A remotely controlled power supply in the high-voltage terminal for the adjustment of the D. J. W. Mous etal. 70 60 ESA and a gridded lens following the high-energy (HE) tube were incorporated.
  • 5. Ion Sourcefor theHVEE Mass Spectrometer 287 INLET 180 ELECTROSTATIC DEFLECTOR GC OR CNN HE ACCEL. TUBE GRIDOED LENS 1111111 n ----------- Fig. 4. Schematic layout of the HVEE 14C AMS spectrometer ANALYSIS MAGNET IONIZATION OR SSB DETECTOR To evaluate the ion optical quality ofthe accelerator, the 13C2+ current in the HE Faraday cuplocated just in front of the detector was measured during a scan of the low energy (LE) steerer and the 180° ESA in the terminal. Figure 5 presents the results. Both traces show a flat-top tuned transmission, indicating that the beam is transported through the system without loss. Furthermore, the figure indi- cates the usefulness of the LE aperture during tuning of the system. T Injector steerer voltage --* 1800 deflector voltage -* Fig. 5. HE Faraday cup current as a function of injector steerer voltage and 180° electrostatic analyzer voltage
  • 6. 288 D. J. W. Mous et at. CONCLUSION The cooperation between HVEE and Oxford University has resulted in a unique, compact ion source that accepts both graphite samples and CO2 samples from ground potential. The source will be applied on the HVEE 14C AMS spectrometer dedicated to biomedical research. Early tests have showed that analyzed 12C- currents of 50-100 ,uA are easily extracted from graphite samples, whereas 5-101uA is obtained from CO2 samples. , Beam transmission of the HVEE 14C AMS spectrometer is flat-top tuned, which guarantees that accurate 14C analysis can be achieved. REFERENCES Alton, G. D. 1990 Recent advancementsin high intensity heavy negative ion sources based on the sputter prin- ciple. Nuclear Instruments and Methods in Physics Research A287:139-149. Bronk, C. R. and Hedges, R. E. M. 1990 A gaseous ion source for routine AMS radiocarbon dating. In Yiou, F. and Raisbeck, G. M., eds., Proceedings of the 5th International Conference on Accelerator Mass Spec- trometry. Nuclear Instruments and Methods in Phys- ics Research B52: 322-326. Davis, J. C. 1994 AMS beyond 2000. In Fifeld, L. K., Fink, D., Sic, S. H. and Tuniz, C., eds., Proceedings of the 6th International Conference on Accelerator Mass Spectrometry. Nuclear Instruments and Methods in Physics Research B92:1-6. Davis, J. C., Proctor, I. D., Southon, J. R., Caffee, M. W., Heikkinen, D. W., Roberts, M. L., Moore, T. L., Tur- teltaub, K. W., Nelson, D. E., Loyd, D. H. and Vogel, J. S. 1990 LLNIIUC AMS facility and research pro- gram. In Yiou, F. and Raisbeck, G. M., eds., Proceed- ings of the 5th International Conference on Accelera- tor Mass Spectrometry. Nuclear Instruments and Methods inPhysics Research B52: 269-272. Fifield, L. K., Fink, D., Sic, S. H. andTniz, C., eds. 1994 Proceedings of the 6th International Conference on Accelerator Mass Spectrometry. NuclearInstruments and Methods in Physics B92. Jull, A. J. T., Beck, J. W. and Burr, G. S., eds. 1997 Pro- ceedings of the 7th International Conference on Ac- celerator Mass Spectrometry. Nuclear Instruments andMethods B123. Mous, D. J. W., Purser, K. H., Fokker, W., van den Broek, R. and Koopmans, R. B. 1997 A compact 14C Isotope Ratio MassSpectrometer for Biomedical applications. In Jull, A. J. T., Beck, J. W. and Burr, G.S., eds., Pro- ceedings of the 7th International Conference on Ac- celerator Mass Spectrometry. Nuclear Instruments andMethods in PhysicsResearch B123:159-162.