The effect of copper conversion plates on low-Z target image quality
David Parsonsa)
Department of Physics and Atmospheric Science, Dalhousie University, 5820 University Avenue, Halifax,
Nova Scotia B3H 1V7, Canada
James L. Robar
Department of Radiation Oncology, Dalhousie University, 5820 University Avenue, Halifax,
Nova Scotia B3H 1V7, Canada
(Received 8 May 2012; revised 11 July 2012; accepted for publication 12 July 2012; published 13
August 2012)
Purpose: Common electronic portal imaging devices (EPIDs) contain a 1.0 mm copper conversion
plate to increase detection efficiency of a therapeutic megavoltage spectrum. When used in imaging
with a photon beam generated with a low atomic number (Z) target, the conversion plate attenuates
a substantial proportion of photons in the diagnostic range, thereby reducing the achievable image
quality. In this work, we measure directly dependence on low-Z target image quality as a function of
copper plate thickness, for both planar imaging and cone beam computed tomography (CBCT).
Methods: Monte Carlo modeling was used to quantify changes to the diagnostic spectrum and de-
tector response for low-Z target beams generated with either 2.35 or 7.00 MeV electrons incident on
a carbon target. Planar contrast-to-noise ratio (CNR) and spatial resolution measurements were made
as a function of copper thickness. CNR measurements were made for CBCT imaging as a function of
dose both with and without the copper plate present in the EPID.
Results: The presence of copper in the EPID decreased the diagnostic photon population by up to
20% and suppressed the peak detector response (dose deposited in the scintillator) at 60 keV by
a factor of 6.4. Planar CNR was increased by a factor ranging from 1.4 to 4.0, depending on the
material imaged, with no copper present compared to a standard 1.0 mm thickness. Planar spatial
resolution was only slightly degraded with increasing copper thickness. Increases in CBCT image
CNR ranged from a factor of 1.3–2.1 with the copper plate removed.
Conclusions: It is possible to increase the proportion of photons in the diagnostic energy range
(25 keV–150 keV) reaching the phosphor screen by as much as 20% when removing the cop-
per conversion plate. This results in significant increases of planar and CBCT image CNR. Con-
sequently, we suggest that the copper conversion plate be removed from the EPID when used for
low-Z target planar or CBCT imaging. © 2012 American Association of Physicists in Medicine.
[http://dx.doi.org/10.1118/1.4742052]
Key words: low-Z target, EPID, planar imaging, CBCT
I. INTRODUCTION
Megavoltage planar or cone beam computed tomography
(MV CBCT) with a therapeutic 6 MV radiation beam has
been used to provide information for image guided radiation
therapy.1
However, MV imaging suffers from poor contrast-
to-noise (CNR) characteristics and involves a relatively high
imaging dose. While the introduction of a kilo-voltage
x-ray tube and detector orthogonal to the treatment beam
line2
has greatly improved available image quality, it has also
increased the associated quality assurance of matching imag-
ing and treatment isocenters, added additional equipment, in-
creased maintenance costs, and does not allow beam’s-eye-
view imaging. The replacement of the high atomic number (Z)
target with of a low-Z target3–13
has been shown to increase
the low energy component of the beam, thereby improving
the available image quality in MV-CBCT.7,10,14–16
While the
majority of the research regarding low-Z targets has focused
on target composition, incident electron energy and their re-
sulting effects on image quality for either planar or CBCT
imaging, relatively little has been reported with regard to op-
timizing the detector for a low-Z target energy spectrum. A
typical electronic portal imaging device (EPID) is composed
of a metal conversion plate (typically copper or steel), a de-
tection receptor (typically a scintillating phosphor), and pho-
todiode array. In MV imaging, photons interact with the metal
conversion plate to produce Compton recoil electrons, which
interact with the scintillating phosphor to produce visible light
that can be detected by the photodiodes.17
While a typical flat-
tened 6 MV beam contains less than 0.5% of photons in the
diagnostic energy range (e.g., 25–150 keV), this fraction in-
creases to as much as 50% for a low-Z target beam.11
While
this component of the spectrum is the most useful from the
point of view of image formation, they also have a high prob-
ability of absorption in the copper conversion plate.
Several groups have investigated the effect of detector
composition on image quality in combination with low-Z tar-
gets. Orton and Robar11
modeled 4 MeV and 6 MeV elec-
trons incident on aluminum and beryllium low-Z targets in a
high-energy linac, with an aS500 detector (Varian Medical,
Inc.), demonstrating that the removal of the copper conver-
sion plate can result in a contrast enhancement factor of 1.2, at
5362 Med. Phys. 39 (9), September 2012 © 2012 Am. Assoc. Phys. Med. 53620094-2405/2012/39(9)/5362/10/$30.00
5363 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5363
both 4 MeV and 6 MeV. Faddegon et al.7
qualitatively showed
that MV-CBCT image quality could be increased with the re-
moval of the copper conversion plate when using a carbon tar-
get with 4.2 MeV incident electrons. Roberts et al.9
demon-
strated that altering the detector receptor from Gb2S2O:Tb to
CsI:Tl crystal can significantly increase the detector response
from 4 MeV electrons incident on nickel and carbon target.
Compared to a 6 MV therapeutic beam, they showed a con-
trast enhancement factor ranging from 1.3 for a thick water
phantom (25.8 cm) to 4.62 for a thin water phantom (5.8 cm).
More recently, Roberts et al.15
showed a twofold increase in
contrast by reducing the incident electron energy to 1.4 MeV
in combination with a CsI:Tl crystalline scintillator detector
compared to a 6 MV therapeutic beam and a standard mega-
voltage EPID in the thick water phantom. However, with the
increase in image quality the associated cost of a crystalline
scintillator is greatly increased, as is the complexity of man-
ufacturing a crystal without defects. In lieu of a crystalline
scintillator, Breitbach et al.18
showed that the copper conver-
sion plate and Gb2S2O:Tb could be replaced with a segmented
scintillator composed of a Gb2S2O ceramic. Using the same
imaging beam line as Faddegon et al.,7
they showed that it
is possible to increase the CNR in CBCT images by a factor
of 1.64 compared to the unaltered EPID with no statistically
significant change in spatial resolution.
It is clear that by altering the detector composition, it is
possible to significantly affect image contrast and CNR. How-
ever, to date no one has quantitatively measured the effect
the copper conversion plate has on image CNR in standard
MV EPID detectors. Through measurement we investigate
(i) the effect that the thickness of the copper conversion plate
has on a 2.35 and 7.00 MV/carbon spectra and the associ-
ated detector response, and (ii) the resultant changes in image
quality, both qualitatively and quantitatively. This paper in-
cludes the following: In Sec. II, we present the methodology
on beam generation, the imaging system, Monte Carlo mod-
els, and imaging protocols. In Sec. III, we present results and
discussion, which include Monte Carlo simulations, planar,
and CBCT image quality analysis.
II. MATERIALS AND METHODS
II.A. Low-Z target beam production
The low-Z target beams were generated using a Varian
2100EX linear accelerator (Varian Medical, Inc., Palo
Alto, CA). Two separate carbon targets were investigated
(i) 7.6 mm thick (2.35 MV/carbon) and (ii) 22.8 mm thick
(7.00 MV/carbon). The thicknesses of the carbon targets were
calculated to coincide with the continuous slowing down
approximation range for 2.50 MeV and 7.00 MeV electrons.
While it has been shown that an increase in low-Z target
thickness slightly degrades resolution,12
a “full-thickness”
target alleviates the need for a polystyrene filter to absorb
transmitted electrons and increases the production of lower
energy photons.3
As described previously,10–12
when used
experimentally, the carousel is operated in manual mode
and the appropriate target is rotated into the beam line via
a rotary switch. For the 2.35 MV/carbon beam, the linac is
operated in electron mode and equipped with a customized
4 MeV program board (used for research purposes only). For
the 7.00 MV/carbon beam, a standard 6 MeV program board
is employed. To generate the 2.35 MeV beam, the bending
magnet was adjusted to tune the beam energy down from the
nominal operating point of 4.5 down to 2.35 MeV. To account
for the diminished beam current observed with the reduced
mean electron energy, the electron gun current, grid voltage,
solenoid, and buncher cavity steering set points were adjusted
to maximize electron current. The design and installation of
similar targets similar to those used in this study have been
published recently11,12
and will not be explained in full here.
Briefly, targets were mounted to the beam side of the carousel
and vertically offset 0.9 cm from the beryllium exit window
of the primary collimation vacuum. By keeping the target as
close as possible to the beryllium exit window, the amount
of electron scattering in air is minimized.11
It is possible to
decrease mean electron energy below 2.35 MeV, however,
this significantly increases acquisition times, degrades spatial
resolution, and has no noticeable advantage with respect to
image CNR.
II.B. Imaging system
An aS1000/IAS3 imaging system (Varian Medical Sys-
tems, Inc.) was used for all imaging. The system was setup as
a stand-alone configuration separate from the clinical imag-
ing system on the treatment unit. The aS1000 panel (IDU20)
has an active area of 30 × 40 cm2
and includes a 1.0 mm Cu
buildup plate, a Gd2O2S:Tb scintillating phosphor layer and a
1024 × 768 array of photodiodes switched by thin-film tran-
sistors deposited on a glass substrate. The number of monitor
units (MU) per exposure was varied to control the dose to the
phantoms, where the IAS3 allows centi-MU control of this
parameter. Prior to imaging, flood field, and dark field cali-
brations were performed where the former has the effect of
correcting for the forward peaked nature of the low-Z target,
flattening filter free beam.
II.C. Monte Carlo simulations
In order to study the effects of the copper conversion plate
on the 2.35 and 7.00 MV spectra, Monte Carlo simulations
were run using BEAMnrc.19
The linac and IDU20 were
modeled in accordance to the physical setup using validated
models provided by Orton and Robar.11
It should be noted
that these models were originally of an aS500, however
between the two detectors, the physical composition is
relatively consistent with the main difference being the
number of pixels within the active area of the detector. 2.0
× 107
incident electron histories were run for both 2.35 and
7.00 MeV beams. Linac collimation was set to match the
active area of the detector at a source to detector distance
(SDD) of 140 cm. Selective bremsstrahlung splitting was
used, and as recommended by Rogers et al.,20
an effective
splitting field size of 50 × 50 cm2
was used for source to sur-
face distance (SSD) of 140 cm with minimum and maximum
bremsstrahlung splitting values of 200 and 1000, respectively.
Medical Physics, Vol. 39, No. 9, September 2012
5364 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5364
FIG. 1. Diagram of the full imaging system. (a) A schematic drawing on the linac and detector. (b) A photo of the detector with the cover off with copper plates
and Cerrobend weights.
Global electron (ECUT) and photon (PCUT) cut-off energies
of 0.700 MeV and of 0.010 MeV, respectively, were used.
The resulting phase space was recorded at the exit side of the
copper conversion plate, above the phosphor. To study the
effect of the copper conversion plate thickness on the energy
spectra, the thickness was varied in 0.2 mm increments from
0 mm to 1.6 mm, corresponding with the physical setup of
experiments, as described below. To determine the spectral
distributions, BEAMdp (Ref. 21) was used to analyze the
phase space data. The resulting data were analyzed and
plotted using MATLAB (Mathworks, Natick, MA).
The IDU20 was also modeled in DOSXYZnrc (Ref. 22)
to determine the effect the thickness of the copper conver-
sion plate has on detector response. Similar to the methods
used by Roberts et al.,9,15
mono-energetic pencil beams of
varying energy from 0.010 MeV to 7.00 MeV were inci-
dent on the detector, and the detector response was quanti-
fied by scoring dose deposited in the phosphor screen per
incident particle. 5.0 × 107
incident photon histories were
used for all simulations, with PCUT and ECUT set to 0.010
MeV and 0.521 MeV, respectively. Detector response sim-
ulations were repeated for 0 mm, 1.0 mm, and 1.6 mm
copper thicknesses. The dose distributions were analyzed
with MATLAB and plotted with energy fluence for 2.35 and
7.00 MV/carbon beams.
II.D. Planar imaging
To observe changes in image quality with varying thick-
nesses of copper, the gantry was rotated to 0◦
(in IEC coordi-
nates) and the detector was oriented perpendicular to the cen-
tral axis at a SDD of 140 cm, as shown in Fig. 1. To decrease
the amount of backscatter radiation, Styrofoam blocks sup-
ported the detector from the floor. The detector cover, foam
spacing materials, and copper conversion plate were removed.
To vary the thickness of copper, 14 × 14 × 0.02 cm3
copper
sheets were stacked on the paper sheet covering the scintilla-
tor. This allowed for the thickness of copper to range from 0
mm to 1.6 mm in 0.2 mm increments. To compress the cop-
per sheets, a customized Cerrobend frame was manufactured
to apply pressure around the periphery of the stacked cop-
per sheets, and the weight of the frame was increased until it
was observed that all layers of copper were compressed firmly
against the panel. To quantify changes in image quality, a pla-
nar CNR and QC3 spatial resolution phantom23
were alter-
nately positioned on the central axis. The CNR phantom con-
sisted of various tissue substitutes surrounded by polystyrene
(electron density relative to water, ρe
w
= 1.08). The tissue
substitutes were 2.5 cm in diameter and 3.0 cm thick and sim-
ulated cortical bone 60% (ρe
w
= 1.69), cortical bone 30%
(ρe
w
= 1.28), inner bone (ρe
w
= 1.09), brain (ρe
w
= 1.05),
and lung (ρe
w
= 0.44). A relatively thin phantom was used as
it allows sensitivity in measuring changes of CNR with cop-
per thickness. The beam was collimated to 10 × 10 cm2
at
a SSD of 140 cm. An imaging dose of 0.02 cGy was deliv-
ered to the center of the CNR phantom to be representative of
the approximate dose per projection given in CBCT. A higher
dose of 0.19 cGy was delivered to center of the QC3 phan-
tom for spatial resolution measurements. A higher dose was
used as a consequence of high noise at 0.02 cGy resulting in
spurious results. CNR was calculated as
CNR =
Pmaterial − Ppolystyrene
σ2
material + σ2
polystyrene
, (1)
where Pmaterial is the average pixel value within the tis-
sue substitute, Ppolystyrene is the average pixel value in the
Medical Physics, Vol. 39, No. 9, September 2012
5365 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5365
surrounding polystyrene, σmaterial is the average noise within
the tissue substitute, and σpolystryene is the average noise within
the surrounding polystyrene. Error bars were found by calcu-
lating the mean and standard deviation of CNR measured in
multiple images. Spatial resolution was analyzed according to
the method outlined by Rajapakshe et al.,23
where the relative
modulation transfer function (RMTF) is defined as
RMTF =
M(f )
M(f1)
, (2)
where M(f) is the output modulation of the line pair in a re-
gion of interest and M(f1) is the output modulation for the
lowest frequency line pair region. As suggested by Droege
and Morin,24
the output modulation is obtained by using the
relationship between the signal amplitude and its variance de-
fined as
M2
(f ) = σ2
m (f ) − σ2
(f ), (3)
where σm
2
(f) is the measured total variance within a region of
interest, and σ2
(f) is the variance due to random noise within
the region of interest. The variance due to random noise as
calculated by Rajapakshe et al.23
is
σ2
(f ) =
σ2
sub
2
, (4)
where σ2
sub(f) represents the variance within the region of in-
terest of two subtracted images. Uncertainty in these measure-
ments was estimated by calculating the mean and standard
deviation of RMTF measured in multiple image sets.
Planar imaging of a sheep head was done to qualitatively
observe the effect the copper conversion plate has on planar
image quality at both low and high energies. The setup for
imaging of the head was as follows: The gantry and couch
were rotated to 90◦
. The imaging panel was placed in a stand
located on the couch with fine adjustments screws and lev-
eled to ensure orthogonality of the panel and the beam axis.
The head was centered on isocenter and a SDD of 130 cm was
used. All imaging involved an approximate dose of 0.26 cGy
delivered to the center of the head. Images were first taken
with the copper conversion plate within the detector. The cop-
per plate was then removed and replaced with a 2.1 mm sheet
of foam (to ensure adequate compression of the panel assem-
bly) and the detector was reassembled. Images were then ac-
quired without the copper conversion plate present within the
detector. Images were analyzed in MATLAB and were com-
pared with identical gray level window settings.
II.E. CBCT imaging
The CBCT images were acquired using the method de-
scribed previously by Robar et al.10
Briefly, this involved
arranging the beam and detector as described above for planar
imaging of the sheep head and rotating the object on a rotation
stage consisting of a microstepping motor (Intelligent Mo-
tion Systems, Inc., Marlborough, CT). The rotation stage was
aligned such that the axis was at the isocenter. Image acquisi-
tion control was accomplished using MATLAB with the Digital
I/O toolbox. All projection images were saved in DICOM for-
mat and subsequently read by the reconstruction software. Im-
age reconstruction was done using the Feldkamp Davis Kress
(FDK) algorithm25
implemented in MATLAB. The phantom
used for CBCT was cylindrical, 12.1 cm in diameter by 6.1 cm
in length, composed of polystyrene and contained the same
tissue substitutes as described above for planar imaging. Tis-
sue substitute materials were 2.5 cm in diameter and 7 cm
in length. CBCT images were acquired both with and with-
out the copper conversion plate present in the detector, with
central doses to the phantom ranging from 1.1 to 11.8 cGy.
Imaging was repeated for both the 2.35 MV/carbon and
7.00 MV/carbon beams. CNR was measured as described in
Eq. (1), with the exception that error bars were calculated by
measuring the mean and standard deviation of CNR in three
slices above and below the analyzed slice.
To observe qualitative changes, CBCT imaging of the
sheep head was done as well, for both beam energies, with
and without the copper plate. An approximate imaging dose
of 4.1 cGy was delivered to the center of the head, located
on isocenter. The resultant images were analyzed in MATLAB
and were compared with identical gray level window settings.
III. RESULTS AND DISCUSSION
III.A. Monte Carlo characterization
Figure 2(a) shows modeled fractional fluences at the exit
side of the copper conversion plate, illustrating the effect
of various copper thicknesses on the energy spectrum, over
the range from 25 keV to 150 keV, for both 2.35 and
7.00 MV/carbon beams. When increasing copper thickness
greater than 0 mm, the fluence below 100 keV is significantly
attenuated [Fig. 2(a)], and accordingly the percentage of di-
agnostic photons is reduced. It should be noted that for the
copper-free cases, the increase in fluence below 38 keV for the
0 mm copper spectra was due to the presence of barium
(k-edge at 37.45 keV) within the detector cover. For the
2.35 MV/carbon beam, the diagnostic photon population
(25 keV–150 keV) was decreased by 20% [Fig. 2(b)] with
the standard 1.0 mm copper conversion plate in place, com-
pared to 0 mm of copper. A corresponding 15% decrease
was observed for the 7.00 MV/carbon beam. Figure 3 illus-
trates the response of the IDU20 with various thicknesses of
copper as well as the photon spectra produced with the 2.35
and 7.00 MV/carbon beams. With no copper present in the
IDU20, detector response peaks at approximately 60 keV and
decreases rapidly with increasing photon energy. When cop-
per was placed within the IDU20, the response at 60 keV was
suppressed by factors of 6.4 and 8.5 for 1.0 mm and 1.6 mm
thicknesses, respectively. At 0.8 MeV, the response was ap-
proximately the same regardless of copper thickness, above
which the responses diverge with the copper responses in-
creasing with increasing energy. Figure 3 demonstrates that
the 2.35 MV/carbon beam contains a substantial fluence of
photons within the peak detector response region (approxi-
mately 30–120 keV). For the 7.00 MV/carbon beam peak flu-
ence occurs at approximately 400 keV, at which point the re-
sponse is locally minimized for the copper-less detector. In
Medical Physics, Vol. 39, No. 9, September 2012
5366 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5366
FIG. 2. (a) The diagnostic spectral distribution with 0 mm, 0.4 mm, 1.0 mm, and 1.6 mm of copper present in the detector for 2.35 MV/carbon and
7.00 MV/carbon beams. (b) The percentage of photons within the diagnostic energy domain (25–150 kV) at the phosphor surface as a function of copper
thickness.
summary, the results of Figs. 2 and 3 indicate that copper
removal is advantageous for either low-Z target beam with
regard to maximizing the diagnostic population reaching the
phosphor, and particularly for the 2.35 MV/carbon beam, al-
lows the energy spectrum to be well matched to the detector
response. The detector response curves shown here agree with
similar results given by Roberts et al.9
and Faddegon et al.7
for an Elekta iViewGT (Elekta AB, Stockholm, Sweden) and
FIG. 3. Response curves for 0 mm, 1.0 mm, and 1.6 mm thick copper layers within the IDU20 (left axis) and photon energy fluence produced with the
2.35 MV/carbon and 7.00 MV/carbon beams (right axis).
Medical Physics, Vol. 39, No. 9, September 2012
5367 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5367
FIG. 4. Planar image contrast-to-noise ratio as a function of copper thickness for both 2.35 MV/carbon and 7.00 MV/carbon beams, for cortical bone 60%
(top-left), cortical bone 30% (top-right), inner bone (middle-left), brain (middle-right), and lung (bottom-left).
a Perkin Elmer XRD 1640 AG9 panel (Perkin Elmer, Inc.,
Waltham, MA), respectively. Both of these detectors include
a 1.0 mm copper plate, a gadolinium oxysulphate scintillat-
ing layer, and an array of photodiodes.16,26
Both papers have
shown that a high-Z scintillating material strongly interacts
with the diagnostic portion of the energy spectrum. Roberts
et al.9,15
also show the detector response for an Eleckta XVI
panel (Elekta AB, Stockholm, Sweden) which differs substan-
tially in design, with the omission of a copper conversion
plate and replacement of the gadolinium oxysulphate scin-
tillating layer by columnar, thallium doped, cesium iodide
(CsI:Tl) crystal. The responses of this detector and the IDU20
FIG. 5. Relative modulation transfer function (RMTF) for 0.0 mm, 1.0 mm, and 1.6 mm thick copper layers in the IDU20 detector for 2.35 MV/carbon and
7.00 MV/carbon beams.
Medical Physics, Vol. 39, No. 9, September 2012
5368 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5368
FIG. 6. Planar imaging of a sheep head at 0.26 cGy (at isocenter) for both
0.0 mm (right) and 1.0 mm (left) of copper within the detector at both 2.35
(bottom) and 7.00 MV/carbon (top). For comparison, the images were nor-
malized to the pixel value at a point within all the images and with identical
gray level window settings.
without copper are similar in that they share a peak response
at approximately 50–70 keV and a local minimum between
approximately 500–600 keV, with the main difference in re-
sponses being in energies less than 40 keV.
III.B. Planar image quality
Figure 4 shows image CNR as a function of copper thick-
ness for various tissue equivalent materials. For most materi-
als, with the exception of lung, image CNR decreases quickly
with the addition of the first few 0.2 mm thick layers of cop-
per and approaches a relatively constant CNR value beyond
0.8 mm of copper. The rapid degradation in CNR with the
first few layers corresponds to the decrease in the diagnostic
fluence with copper shown in Fig. 2; both the fluence in the
diagnostic range and CNR show the largest decreases with
the addition of the first 0.4 mm of copper. With no copper
present, image CNR was increased by factors of 1.6, 2.3, 3.2,
1.4, and 2.3 for cortical bone 60%, cortical bone 30%, inner
bone, brain, and lung, respectively, compared to the standard
1.0 mm of copper. Similar gains in CNR were observed for
the 7.00 MV/carbon beam, with factor increases ranging from
1.6 to 4.0; however, the absolute values of CNR were typically
less than half of those observed at 2.35 MV. The results pre-
sented here confirm the modeled results of Orton and Robar,11
in that a gain in image contrast was observed with the removal
of the copper layer.
Figure 5 shows RMTF at various thicknesses of copper for
the 2.35 and 7.00 MV/carbon beams. At most frequencies,
there appears to be little variation with changing copper thick-
ness. At 0.25 lp/mm, there is a detectable but small change
in RMTF, with relative improvement of RMTF by 18% and
15% for the 2.35 and 7.00 MV/carbon beams, with removal of
the 1.0 mm copper layer. The frequency where RMTF equals
0.5, or f50, did not vary significantly with copper removal, for
either low-Z target beam. With the 1.0 mm copper layer, here
FIG. 7. Profiles of Fig. 6 with and without the copper conversion plate within the detector for 2.35 MV/carbon and 7.00 MV/carbon beams.
Medical Physics, Vol. 39, No. 9, September 2012
5369 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5369
FIG. 8. Cone beam computed tomography images at approximately 4.1 cGy
with copper plate (top) and without copper (bottom) in the detector, with the
7.00 MV/carbon and 2.35 MV/carbon beams. The tissue equivalent materials
imaged are (clockwise starting from the darkest) lung, cortical bone 60%,
cortical bone 30%, inner bone, and brain.
we see decrease of f50 0.03 lp/mm when reducing electron
energy from 7.00 to 2.35 MeV. In comparison, Connell and
Robar12
reported a change in f50 of approximately 0.02 and
0.03 lp/mm for beryllium and aluminum target beams, when
reducing the incident electron energy from 7.0 to 4.5 MeV.
Figure 6 shows planar sagittal sheep images both with
and without the copper conversion plate with the 2.35 and
7.00 MV/carbon beams. The images show increased image
contrast with the removal of the copper conversion plate, com-
pared to images with the copper conversion plate in place.
This is most noticeable within the squamous, nasal, and
mandible regions. These differences are highlighted in the
profiles shown in Fig. 7, in which the contrast is notably in-
creased with the removal of the copper conversion plate for
the 2.35 MV/carbon beam. Figure 7 also shows slight loss
of fine detail when using the 2.35 MV/carbon beam com-
pared to 7.00 MV/carbon. As noted by Connell and Robar,12
this is largely a result of electron beam broadening when de-
creasing incident electron energy. However, the changes are
relatively minor between images with and without the cop-
per conversion plate, which agrees with the results shown in
Fig. 6.
III.C. CBCT image quality
Figure 8 shows CBCT images of tissue equivalent materi-
als in polystyrene both with and without copper in the IDU20
for the 2.35 MV/carbon and 7.00 MV/carbon beams. The
CBCT images show greater contrast with the copper conver-
sion plate removed both at 7.00 and 2.35 MV/carbon, with
the greatest improvement seen for 2.35 MV/carbon. This is
FIG. 9. Contrast-to-noise ratios of tissue substitutes in CBCT images as a function of dose, comparing the 2.35 MV/carbon and 7.00 MV/carbon beams both
with and without copper present in the detector.
Medical Physics, Vol. 39, No. 9, September 2012
5370 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5370
FIG. 10. CBCT images of a sheep head at approximately 4.1 cGy (at isocen-
ter) with copper plate (top) and without copper (bottom) in the detector, with
the 7.00 MV/carbon and 2.35 MV/carbon beams.
evident also in the corresponding CNR curves as a function
of dose, shown in Fig. 9. At approximately 4.1 cGy and for
the 2.35 MV/carbon beam, for example, copper removal pro-
duced improvement of CNR by factors of 1.5, 1.9, 2.1, 1.4,
and 1.3 for cortical bone 60%, cortical bone 30%, inner bone,
brain, and lung, respectively. Similar factor increases are seen
for the 7.00 MV/carbon beam, but as for the planar imag-
ing results, CNR values are typically half of that observed
at 2.35 MV/carbon. The CNR improvement with copper re-
moval seen here is comparable in magnitude to the factor
1.64 gain in CNR observed by Breitbach et al.18
when using a
segmented ceramic Gb2S2O scintillator, compared to a stan-
dard 1.0 mm copper conversion plate with Gb2S2O scintillat-
ing phosphor. However, this group used a different beam-line7
within a Siemens linac at 4.2 MV.
Figure 10 shows CBCT images of the sheep head with and
without the copper conversion plate with the 2.35 MV/carbon
and 7.00 MV/carbon beams. These images highlight the effect
the presence of copper in the IDU20 has on image contrast-to-
noise characteristics. This can be seen in the bony structures
of the skull, and in the nasal cavity, where at 2.35 MV/carbon
are clearly visualized without copper, compared to the image
acquired with the copper layer. These trends are also present
in the 7.00 MV/carbon images; however, the CNR is clearly
degraded. These images taken here agree with the qualitative
findings reported by Faddegon et al.,7
in which there is an
overall improvement in bone and soft tissue visibility with
copper removal; however, in this work we show that the re-
duction of energy to 2.35 MeV offers an overall improve-
ment in image quality. While we believe that with a more
efficient detector such as a crystalline scintillator would be
preferable,9,15
these results demonstrate that substantial im-
provements in image quality may be realized with a standard
MV EPID, following removal of the copper conversion plate.
IV. CONCLUSIONS
In this work, we have shown that the removal of the copper
conversion plate within an IDU20 EPID can significantly im-
prove image contrast and CNR when used in conjunction with
low-Z target imaging beams. The presence of the copper con-
version plate caused a decrease in the diagnostic population
of 20% and 15% for 2.35 MV/carbon and 7.00 MV/carbon
beams, respectively. The peak response within the detector
without copper was reduced by a factor of 6.4 when 1.0 mm
of copper was present in the detector. Planar images acquired
with a copper-less detector showed a factor gain in CNR rang-
ing from 1.4 to 3.2 and 1.6 to 4.0, for 2.35 MV/carbon and
7.00 MV/carbon beams, respectively, compared to the stan-
dard 1.0 mm of copper. Similar gains in CNR were observed
in CBCT images, where CNR was improved by a factor rang-
ing from 1.3 to 2.1. Spatial resolution was not affected signif-
icantly by copper removal, for either beam. Qualitative imag-
ing of a sheep head showed that there was an increase in pla-
nar and CBCT image contrast with a copper-less detector.
ACKNOWLEDGMENTS
The authors are grateful for support provided by Varian
Medical Systems for the funding of this project and the Nat-
ural Sciences and Engineering Research Council of Canada
for their additional financial support. We would like to thank
Robert Moran for his vital electronic and technical support,
Ian Porter for fabrication of apparatus, Dr. Robin Kelly and
Dr. Mammo Yewondwossen for their knowledge and teach-
ing of techniques utilized in this project.
a)Author to whom correspondence should be addressed. Electronic mail:
david.parsons@dal.ca
1O. Morin, A. Gillis, J. Chen, M. Aubin, M. K. Bucci, M. Roach, and
J. Pouliot, “Megavoltage cone-beam CT: System description and clinical
applications,” Med. Dosim. 31, 51–61 (2006).
2D. A. Jaffray and J. H. Siewerdsen, “Cone-beam computed tomography
with a flat-panel imager: Initial performance characterization,” Med. Phys.
27, 1311–1323 (2000).
3D. M. Galbraith, “Low-energy imaging with high-energy bremsstrahlung
beams,” Med. Phys. 16, 734–746 (1989).
4D. W. Mah, D. M. Galbraith, and J. A. Rawlinson, “Low-energy imaging
with high-energy bremsstrahlung beams: Analysis and scatter reduction,”
Med. Phys. 16, 653–665 (1993).
5O. Z. Ostapiak, P. F. O’Brien, and B. A. Faddegon, “Megavoltage imaging
with low Z targets: Implementation and characterization of an investiga-
tional system,” Med. Phys. 25, 1910–1918 (1998).
6A. Tsechanski, A. F. Bielajew, S. Faermann, and Y. Krutman, “A thin target
approach for portal imaging in medical accelerators,” Phys. Med. Biol. 43,
2221–2236 (1998).
7B. A. Faddegon, V. Wu, J. Pouliot, B. Gangadharan, and A. Bani-Hashemi,
“Low dose megavoltage cone beam computed tomography with an unflat-
tened 4 MV beam from a carbon target,” Med. Phys. 35, 5777–5786 (2008).
8S. Flampouri, P. M. Evans, F. Verhaegen, A. E. Nahum, E. Spezi, and
M. Partridge, “Optimization of accelerator target and detector for portal
imaging using Monte Carlo simulation and experiment,” Phys. Med. Biol.
47, 3331–3349 (2002).
9D. A. Roberts, V. N. Hansen, A. C. Niven, M. G. Thompson, J. Seco, and
P. M. Evans, “A low Z linac and flat panel imager: Comparison with the
conventional imaging approach,” Phys. Med. Biol. 53, 6305–6319 (2008).
10J. L. Robar, T. Connell, W. Huang, and R. G. Kelly, “Megavoltage planar
and cone-beam imaging with low-Z targets: Dependence of image qual-
ity improvement on beam energy and patient separation,” Med. Phys. 36,
3955–3963 (2009).
Medical Physics, Vol. 39, No. 9, September 2012
5371 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5371
11E. J. Orton and J. L. Robar, “Megavoltage image contrast with low-atomic
number target materials and amorphous silicon electronic portal imagers,”
Phys. Med. Biol. 54, 1275–1289 (2009).
12T. Connell and J. L. Robar, “Low-Z target optimization for spatial res-
olution improvement in megavoltage imaging,” Med. Phys. 37, 124–131
(2010).
13D. Sawkey, M. Lu, O. Morin, M. Aubin, S. S. Yom, A. R. Gottschalk,
A. Bani-Hashemi, and B. A. Faddegon, “A diamond target for megavoltage
cone-beam CT,” Med. Phys. 37, 1246–1253 (2010).
14D. A. Roberts, V. N. Hansen, M. G. Thompson, G. Poludniowski, A. Niven,
J. Seco, and P. M. Evans, “Comparative study of a low-Z cone-beam com-
puted tomography system,” Phys. Med. Biol. 56, 4453–4464 (2011).
15D. A. Roberts, V. N. Hansen, M. G. Thompson, G. Poludniowski, A. Niven,
J. Seco, and P. M. Evans, “Kilovoltage energy imaging with a radiotherapy
linac with a continuously variable energy range,” Med. Phys. 39, 1218–26
(2012).
16B. A. Faddegon, M. Aubin, A. Bani-Hashemi, B. Gangadharan,
A. R. Gottschalk, O. Morin, D. Sawkey, V. Wu, and S. S. Yom, “Com-
parison of patient megavoltage cone beam CT images acquired with an un-
flattened beam from a carbon target and a flattened treatment beam,” Med.
Phys. 37, 1737–1741 (2010).
17D. A. Jaffray, J. J. Battista, A. Fenster, and P. Munro, “Monte Carlo studies
of x-ray energy absorption and quantum noise in megavoltage transmission
radiography,” Med. Phys. 22, 1077–88 (1995).
18E. K. Breitbach, J. S. Maltz, B. Gangadharan, A. Bani-Hashemi, C. M.
Anderson, S. K. Bhatia, J. Stiles, D. S. Edwards, and R. T. Flynn, “Image
quality improvement in megavoltage cone beam CT using an imaging beam
line and a sintered pixelated array system,” Med. Phys. 38, 5969–5979
(2011).
19D. W. Rogers, B. A. Faddegon, G. X. Ding, C. M. Ma, J. We, and
T. R. Mackie, “BEAM: A Monte Carlo code to simulate radiotherapy treat-
ment units,” Med. Phys. 22, 503–524 (1995).
20D. W. O. Rogers, B. Walters, and I. Kawrakow, “BEAMnrc users manual,”
NRCC Report No. PIRS-0509(A)revL (NRCC, Ottawa, 2011).
21C. M. Ma and D. W. O. Rogers, “BEAMdp users manual,” NRCC Report
No. PIRS-0509(C)revA (NRCC, Ottawa, 2009).
22I. Kawrakow, “Accurate condensed history Monte Carlo simulation of elec-
tron transport. I. EGSnrc, the new EGS4 version,” Med. Phys. 27, 485–498
(2000).
23R. Rajapakshe, K. Luchka, and S. Shalev, “A quality control test for elec-
tronic portal imaging devices,” Med. Phys. 23, 1237–1244 (1996).
24R. T. Droege and R. L. Morin, “A practical method to measure the MTF of
CT scanners,” Med. Phys. 9, 758–760 (1982).
25L. Feldkamp, L. Davis, and J. Kress, “Practical cone-beam algorithm,”
J. Opt. Soc. Am. A 1, 612–619 (1984).
26L. Parent, J. Seco, P. M. Evans, A. Fielding, and D. R. Dance, “Monte
Carlo modelling of a-Si EPID response: The effect of spectral variations
with field size and position,” Med. Phys. 33, 4527 (2006).
Medical Physics, Vol. 39, No. 9, September 2012

More Related Content

What's hot

Dye sensitized solar cells
Dye sensitized solar cellsDye sensitized solar cells
Dye sensitized solar cellssaromemarzadeh
 
Broadband high photoresponse from pure monolayer graphene photodetector
Broadband high photoresponse from pure monolayer graphene photodetectorBroadband high photoresponse from pure monolayer graphene photodetector
Broadband high photoresponse from pure monolayer graphene photodetectorCarlos Bella
 
Comparison of calcium carbonate and titania particles on improving color homo...
Comparison of calcium carbonate and titania particles on improving color homo...Comparison of calcium carbonate and titania particles on improving color homo...
Comparison of calcium carbonate and titania particles on improving color homo...TELKOMNIKA JOURNAL
 
Excellent luminous flux of WLEDs with flat dual-layer remote phosphor geometry
Excellent luminous flux of WLEDs with flat dual-layer remote phosphor geometryExcellent luminous flux of WLEDs with flat dual-layer remote phosphor geometry
Excellent luminous flux of WLEDs with flat dual-layer remote phosphor geometryTELKOMNIKA JOURNAL
 
1 s2.0-s0022369702003566-main
1 s2.0-s0022369702003566-main1 s2.0-s0022369702003566-main
1 s2.0-s0022369702003566-maintonamiry
 
APS march meeting 2012
APS march meeting 2012APS march meeting 2012
APS march meeting 2012Po-Chun Yeh
 
CEC Symposium Poster - Vineet J Nair
CEC Symposium Poster - Vineet J NairCEC Symposium Poster - Vineet J Nair
CEC Symposium Poster - Vineet J NairVineet Nair
 
A Front Surface Optimization Study for Photovoltaic Application
A Front Surface Optimization Study for Photovoltaic ApplicationA Front Surface Optimization Study for Photovoltaic Application
A Front Surface Optimization Study for Photovoltaic ApplicationTELKOMNIKA JOURNAL
 
Recent progress in non platinum counter electrode materials for dye sensitize...
Recent progress in non platinum counter electrode materials for dye sensitize...Recent progress in non platinum counter electrode materials for dye sensitize...
Recent progress in non platinum counter electrode materials for dye sensitize...Science Padayatchi
 
Preparation and study optical properties of (pva pvp-cr cl2) composites
Preparation and study optical properties of (pva pvp-cr cl2) compositesPreparation and study optical properties of (pva pvp-cr cl2) composites
Preparation and study optical properties of (pva pvp-cr cl2) compositesAlexander Decker
 
10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...
10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...
10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...Mahdi Robat Sarpoushi
 
The options in remote phosphor structure for better white LEDs color quality
The options in remote phosphor structure for better white LEDs color qualityThe options in remote phosphor structure for better white LEDs color quality
The options in remote phosphor structure for better white LEDs color qualityTELKOMNIKA JOURNAL
 
IRJET- Dual Band Cylindrical DRA with Carbon Nano Tube
IRJET- Dual Band Cylindrical DRA with Carbon Nano TubeIRJET- Dual Band Cylindrical DRA with Carbon Nano Tube
IRJET- Dual Band Cylindrical DRA with Carbon Nano TubeIRJET Journal
 
Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...
Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...
Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...Journal Papers
 
STUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZ
STUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZSTUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZ
STUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZjmicro
 
Carbon quantum dots_synthesis_book
Carbon quantum dots_synthesis_bookCarbon quantum dots_synthesis_book
Carbon quantum dots_synthesis_bookFilipa Correia
 

What's hot (20)

Dye sensitized solar cells
Dye sensitized solar cellsDye sensitized solar cells
Dye sensitized solar cells
 
Broadband high photoresponse from pure monolayer graphene photodetector
Broadband high photoresponse from pure monolayer graphene photodetectorBroadband high photoresponse from pure monolayer graphene photodetector
Broadband high photoresponse from pure monolayer graphene photodetector
 
1-s2.0-S1369800114000055-main
1-s2.0-S1369800114000055-main1-s2.0-S1369800114000055-main
1-s2.0-S1369800114000055-main
 
Comparison of calcium carbonate and titania particles on improving color homo...
Comparison of calcium carbonate and titania particles on improving color homo...Comparison of calcium carbonate and titania particles on improving color homo...
Comparison of calcium carbonate and titania particles on improving color homo...
 
Excellent luminous flux of WLEDs with flat dual-layer remote phosphor geometry
Excellent luminous flux of WLEDs with flat dual-layer remote phosphor geometryExcellent luminous flux of WLEDs with flat dual-layer remote phosphor geometry
Excellent luminous flux of WLEDs with flat dual-layer remote phosphor geometry
 
Carbon dots
Carbon dotsCarbon dots
Carbon dots
 
1 s2.0-s0022369702003566-main
1 s2.0-s0022369702003566-main1 s2.0-s0022369702003566-main
1 s2.0-s0022369702003566-main
 
Using Green Emitting Phosphor for Improving Lighting Performance of In-cup Pa...
Using Green Emitting Phosphor for Improving Lighting Performance of In-cup Pa...Using Green Emitting Phosphor for Improving Lighting Performance of In-cup Pa...
Using Green Emitting Phosphor for Improving Lighting Performance of In-cup Pa...
 
APS march meeting 2012
APS march meeting 2012APS march meeting 2012
APS march meeting 2012
 
CEC Symposium Poster - Vineet J Nair
CEC Symposium Poster - Vineet J NairCEC Symposium Poster - Vineet J Nair
CEC Symposium Poster - Vineet J Nair
 
A Front Surface Optimization Study for Photovoltaic Application
A Front Surface Optimization Study for Photovoltaic ApplicationA Front Surface Optimization Study for Photovoltaic Application
A Front Surface Optimization Study for Photovoltaic Application
 
Recent progress in non platinum counter electrode materials for dye sensitize...
Recent progress in non platinum counter electrode materials for dye sensitize...Recent progress in non platinum counter electrode materials for dye sensitize...
Recent progress in non platinum counter electrode materials for dye sensitize...
 
Preparation and study optical properties of (pva pvp-cr cl2) composites
Preparation and study optical properties of (pva pvp-cr cl2) compositesPreparation and study optical properties of (pva pvp-cr cl2) composites
Preparation and study optical properties of (pva pvp-cr cl2) composites
 
10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...
10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...
10.1016-j.mssp.2014.10.034-Graphene nanosheets as electrode materials for sup...
 
The options in remote phosphor structure for better white LEDs color quality
The options in remote phosphor structure for better white LEDs color qualityThe options in remote phosphor structure for better white LEDs color quality
The options in remote phosphor structure for better white LEDs color quality
 
IRJET- Dual Band Cylindrical DRA with Carbon Nano Tube
IRJET- Dual Band Cylindrical DRA with Carbon Nano TubeIRJET- Dual Band Cylindrical DRA with Carbon Nano Tube
IRJET- Dual Band Cylindrical DRA with Carbon Nano Tube
 
Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...
Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...
Electrical transportation mechanisms of molybdenum disulfide flakes graphene ...
 
STUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZ
STUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZSTUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZ
STUDY OF ABSORPTION IN CARBON NANOTUBE COMPOSITES FROM 1HZ TO 40GHZ
 
Carbon quantum dots_synthesis_book
Carbon quantum dots_synthesis_bookCarbon quantum dots_synthesis_book
Carbon quantum dots_synthesis_book
 
Ptf V8
Ptf V8Ptf V8
Ptf V8
 

Similar to Effect of copper plates on low-Z target image quality

Microscopy microanalysis microstructures_the european physical journal_applie...
Microscopy microanalysis microstructures_the european physical journal_applie...Microscopy microanalysis microstructures_the european physical journal_applie...
Microscopy microanalysis microstructures_the european physical journal_applie...Andrea Sentimenti
 
Semiconductor part-2
Semiconductor part-2Semiconductor part-2
Semiconductor part-2Santanu Paria
 
An analytical method with numerical results to be used in the design of optic...
An analytical method with numerical results to be used in the design of optic...An analytical method with numerical results to be used in the design of optic...
An analytical method with numerical results to be used in the design of optic...nooriasukmaningtyas
 
Plasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum Dots
Plasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum DotsPlasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum Dots
Plasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum Dotsijrap
 
Advances In Single-Charge Detectors And Their Applications
Advances In Single-Charge Detectors And Their ApplicationsAdvances In Single-Charge Detectors And Their Applications
Advances In Single-Charge Detectors And Their ApplicationsRichard Hogue
 
Transparent Conducting Oxides for Thin Film PV
Transparent Conducting Oxides for Thin Film PVTransparent Conducting Oxides for Thin Film PV
Transparent Conducting Oxides for Thin Film PVcdtpv
 
Transparent conducting oxides for thin film PV
Transparent conducting oxides for thin film PVTransparent conducting oxides for thin film PV
Transparent conducting oxides for thin film PVUniversity of Liverpool
 
Ultrafast response of metal-semiconductor hybrid.pptx
Ultrafast response of metal-semiconductor hybrid.pptxUltrafast response of metal-semiconductor hybrid.pptx
Ultrafast response of metal-semiconductor hybrid.pptxSabinaGurung12
 
10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...
10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...
10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...Mahdi Robat Sarpoushi
 
Influence of Ca[Mg 3 SiN 4 ]Ce 3+ phosphor’s concentration on optical proper...
Influence of Ca[Mg 3 SiN 4 ]Ce 3+  phosphor’s concentration on optical proper...Influence of Ca[Mg 3 SiN 4 ]Ce 3+  phosphor’s concentration on optical proper...
Influence of Ca[Mg 3 SiN 4 ]Ce 3+ phosphor’s concentration on optical proper...IJECEIAES
 
37-.ocn_.ppt Optimizing slideshow... photodiiode
37-.ocn_.ppt Optimizing slideshow... photodiiode37-.ocn_.ppt Optimizing slideshow... photodiiode
37-.ocn_.ppt Optimizing slideshow... photodiiodeNandanas11
 
Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...
Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...
Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...TELKOMNIKA JOURNAL
 
ECS Canada Spring Symposium 2018
ECS Canada Spring Symposium 2018ECS Canada Spring Symposium 2018
ECS Canada Spring Symposium 2018Taufique Z. Redhwan
 
Application of dual-layer phosphor geometries for enhancing the optical prope...
Application of dual-layer phosphor geometries for enhancing the optical prope...Application of dual-layer phosphor geometries for enhancing the optical prope...
Application of dual-layer phosphor geometries for enhancing the optical prope...TELKOMNIKA JOURNAL
 
Semiconductor Nanomaterials
Semiconductor NanomaterialsSemiconductor Nanomaterials
Semiconductor NanomaterialsSantanu Paria
 
Temperature Dependence of the Band-Edge Transitions of ZnCdBeSe
Temperature Dependence of the Band-Edge Transitions of ZnCdBeSeTemperature Dependence of the Band-Edge Transitions of ZnCdBeSe
Temperature Dependence of the Band-Edge Transitions of ZnCdBeSeOleg Maksimov
 

Similar to Effect of copper plates on low-Z target image quality (20)

Microscopy microanalysis microstructures_the european physical journal_applie...
Microscopy microanalysis microstructures_the european physical journal_applie...Microscopy microanalysis microstructures_the european physical journal_applie...
Microscopy microanalysis microstructures_the european physical journal_applie...
 
Semiconductor part-2
Semiconductor part-2Semiconductor part-2
Semiconductor part-2
 
An analytical method with numerical results to be used in the design of optic...
An analytical method with numerical results to be used in the design of optic...An analytical method with numerical results to be used in the design of optic...
An analytical method with numerical results to be used in the design of optic...
 
Al4102280285
Al4102280285Al4102280285
Al4102280285
 
Plasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum Dots
Plasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum DotsPlasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum Dots
Plasmon Enhanced Internal Quantum Efficiency of CdSe/ZnS Quantum Dots
 
Advances In Single-Charge Detectors And Their Applications
Advances In Single-Charge Detectors And Their ApplicationsAdvances In Single-Charge Detectors And Their Applications
Advances In Single-Charge Detectors And Their Applications
 
Transparent Conducting Oxides for Thin Film PV
Transparent Conducting Oxides for Thin Film PVTransparent Conducting Oxides for Thin Film PV
Transparent Conducting Oxides for Thin Film PV
 
Transparent conducting oxides for thin film PV
Transparent conducting oxides for thin film PVTransparent conducting oxides for thin film PV
Transparent conducting oxides for thin film PV
 
Ultrafast response of metal-semiconductor hybrid.pptx
Ultrafast response of metal-semiconductor hybrid.pptxUltrafast response of metal-semiconductor hybrid.pptx
Ultrafast response of metal-semiconductor hybrid.pptx
 
10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...
10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...
10.1016-j.synthmet.2014.12.031-Graphite nanosheets_nanoporous carbon black_ce...
 
Influence of Ca[Mg 3 SiN 4 ]Ce 3+ phosphor’s concentration on optical proper...
Influence of Ca[Mg 3 SiN 4 ]Ce 3+  phosphor’s concentration on optical proper...Influence of Ca[Mg 3 SiN 4 ]Ce 3+  phosphor’s concentration on optical proper...
Influence of Ca[Mg 3 SiN 4 ]Ce 3+ phosphor’s concentration on optical proper...
 
37-.ocn_.ppt Optimizing slideshow... photodiiode
37-.ocn_.ppt Optimizing slideshow... photodiiode37-.ocn_.ppt Optimizing slideshow... photodiiode
37-.ocn_.ppt Optimizing slideshow... photodiiode
 
Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...
Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...
Acquiring higher lumen efficacy and color rendering index with green NaYF4:Er...
 
Ce icp-ms
Ce icp-msCe icp-ms
Ce icp-ms
 
ECS Canada Spring Symposium 2018
ECS Canada Spring Symposium 2018ECS Canada Spring Symposium 2018
ECS Canada Spring Symposium 2018
 
Application of dual-layer phosphor geometries for enhancing the optical prope...
Application of dual-layer phosphor geometries for enhancing the optical prope...Application of dual-layer phosphor geometries for enhancing the optical prope...
Application of dual-layer phosphor geometries for enhancing the optical prope...
 
6-oc ceramic kilowatts
6-oc ceramic kilowatts6-oc ceramic kilowatts
6-oc ceramic kilowatts
 
Semiconductor Nanomaterials
Semiconductor NanomaterialsSemiconductor Nanomaterials
Semiconductor Nanomaterials
 
Optical Antenna
Optical AntennaOptical Antenna
Optical Antenna
 
Temperature Dependence of the Band-Edge Transitions of ZnCdBeSe
Temperature Dependence of the Band-Edge Transitions of ZnCdBeSeTemperature Dependence of the Band-Edge Transitions of ZnCdBeSe
Temperature Dependence of the Band-Edge Transitions of ZnCdBeSe
 

More from David Parsons

Stevens et al, Continuous monitoring of prostate position using stereoscopic ...
Stevens et al, Continuous monitoring of prostate position using stereoscopic ...Stevens et al, Continuous monitoring of prostate position using stereoscopic ...
Stevens et al, Continuous monitoring of prostate position using stereoscopic ...David Parsons
 
Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...
Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...
Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...David Parsons
 
Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...
Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...
Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...David Parsons
 
Parsons, David, MSc, PHYC, August 2012
Parsons, David, MSc, PHYC, August 2012Parsons, David, MSc, PHYC, August 2012
Parsons, David, MSc, PHYC, August 2012David Parsons
 
Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...
Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...
Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...David Parsons
 
Parsons and Robar, Volume of interest CBCT and tube current modulation for i...
 Parsons and Robar, Volume of interest CBCT and tube current modulation for i... Parsons and Robar, Volume of interest CBCT and tube current modulation for i...
Parsons and Robar, Volume of interest CBCT and tube current modulation for i...David Parsons
 

More from David Parsons (6)

Stevens et al, Continuous monitoring of prostate position using stereoscopic ...
Stevens et al, Continuous monitoring of prostate position using stereoscopic ...Stevens et al, Continuous monitoring of prostate position using stereoscopic ...
Stevens et al, Continuous monitoring of prostate position using stereoscopic ...
 
Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...
Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...
Berbeco et al, Low Z target switching to increase tumor endothelial cell dose...
 
Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...
Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...
Parsons et al, A Monte Carlo investigation of low-Z target image quality gene...
 
Parsons, David, MSc, PHYC, August 2012
Parsons, David, MSc, PHYC, August 2012Parsons, David, MSc, PHYC, August 2012
Parsons, David, MSc, PHYC, August 2012
 
Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...
Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...
Parsons and Robar, An investigation of kV CBCT image quality and dose reducti...
 
Parsons and Robar, Volume of interest CBCT and tube current modulation for i...
 Parsons and Robar, Volume of interest CBCT and tube current modulation for i... Parsons and Robar, Volume of interest CBCT and tube current modulation for i...
Parsons and Robar, Volume of interest CBCT and tube current modulation for i...
 

Effect of copper plates on low-Z target image quality

  • 1. The effect of copper conversion plates on low-Z target image quality David Parsonsa) Department of Physics and Atmospheric Science, Dalhousie University, 5820 University Avenue, Halifax, Nova Scotia B3H 1V7, Canada James L. Robar Department of Radiation Oncology, Dalhousie University, 5820 University Avenue, Halifax, Nova Scotia B3H 1V7, Canada (Received 8 May 2012; revised 11 July 2012; accepted for publication 12 July 2012; published 13 August 2012) Purpose: Common electronic portal imaging devices (EPIDs) contain a 1.0 mm copper conversion plate to increase detection efficiency of a therapeutic megavoltage spectrum. When used in imaging with a photon beam generated with a low atomic number (Z) target, the conversion plate attenuates a substantial proportion of photons in the diagnostic range, thereby reducing the achievable image quality. In this work, we measure directly dependence on low-Z target image quality as a function of copper plate thickness, for both planar imaging and cone beam computed tomography (CBCT). Methods: Monte Carlo modeling was used to quantify changes to the diagnostic spectrum and de- tector response for low-Z target beams generated with either 2.35 or 7.00 MeV electrons incident on a carbon target. Planar contrast-to-noise ratio (CNR) and spatial resolution measurements were made as a function of copper thickness. CNR measurements were made for CBCT imaging as a function of dose both with and without the copper plate present in the EPID. Results: The presence of copper in the EPID decreased the diagnostic photon population by up to 20% and suppressed the peak detector response (dose deposited in the scintillator) at 60 keV by a factor of 6.4. Planar CNR was increased by a factor ranging from 1.4 to 4.0, depending on the material imaged, with no copper present compared to a standard 1.0 mm thickness. Planar spatial resolution was only slightly degraded with increasing copper thickness. Increases in CBCT image CNR ranged from a factor of 1.3–2.1 with the copper plate removed. Conclusions: It is possible to increase the proportion of photons in the diagnostic energy range (25 keV–150 keV) reaching the phosphor screen by as much as 20% when removing the cop- per conversion plate. This results in significant increases of planar and CBCT image CNR. Con- sequently, we suggest that the copper conversion plate be removed from the EPID when used for low-Z target planar or CBCT imaging. © 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4742052] Key words: low-Z target, EPID, planar imaging, CBCT I. INTRODUCTION Megavoltage planar or cone beam computed tomography (MV CBCT) with a therapeutic 6 MV radiation beam has been used to provide information for image guided radiation therapy.1 However, MV imaging suffers from poor contrast- to-noise (CNR) characteristics and involves a relatively high imaging dose. While the introduction of a kilo-voltage x-ray tube and detector orthogonal to the treatment beam line2 has greatly improved available image quality, it has also increased the associated quality assurance of matching imag- ing and treatment isocenters, added additional equipment, in- creased maintenance costs, and does not allow beam’s-eye- view imaging. The replacement of the high atomic number (Z) target with of a low-Z target3–13 has been shown to increase the low energy component of the beam, thereby improving the available image quality in MV-CBCT.7,10,14–16 While the majority of the research regarding low-Z targets has focused on target composition, incident electron energy and their re- sulting effects on image quality for either planar or CBCT imaging, relatively little has been reported with regard to op- timizing the detector for a low-Z target energy spectrum. A typical electronic portal imaging device (EPID) is composed of a metal conversion plate (typically copper or steel), a de- tection receptor (typically a scintillating phosphor), and pho- todiode array. In MV imaging, photons interact with the metal conversion plate to produce Compton recoil electrons, which interact with the scintillating phosphor to produce visible light that can be detected by the photodiodes.17 While a typical flat- tened 6 MV beam contains less than 0.5% of photons in the diagnostic energy range (e.g., 25–150 keV), this fraction in- creases to as much as 50% for a low-Z target beam.11 While this component of the spectrum is the most useful from the point of view of image formation, they also have a high prob- ability of absorption in the copper conversion plate. Several groups have investigated the effect of detector composition on image quality in combination with low-Z tar- gets. Orton and Robar11 modeled 4 MeV and 6 MeV elec- trons incident on aluminum and beryllium low-Z targets in a high-energy linac, with an aS500 detector (Varian Medical, Inc.), demonstrating that the removal of the copper conver- sion plate can result in a contrast enhancement factor of 1.2, at 5362 Med. Phys. 39 (9), September 2012 © 2012 Am. Assoc. Phys. Med. 53620094-2405/2012/39(9)/5362/10/$30.00
  • 2. 5363 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5363 both 4 MeV and 6 MeV. Faddegon et al.7 qualitatively showed that MV-CBCT image quality could be increased with the re- moval of the copper conversion plate when using a carbon tar- get with 4.2 MeV incident electrons. Roberts et al.9 demon- strated that altering the detector receptor from Gb2S2O:Tb to CsI:Tl crystal can significantly increase the detector response from 4 MeV electrons incident on nickel and carbon target. Compared to a 6 MV therapeutic beam, they showed a con- trast enhancement factor ranging from 1.3 for a thick water phantom (25.8 cm) to 4.62 for a thin water phantom (5.8 cm). More recently, Roberts et al.15 showed a twofold increase in contrast by reducing the incident electron energy to 1.4 MeV in combination with a CsI:Tl crystalline scintillator detector compared to a 6 MV therapeutic beam and a standard mega- voltage EPID in the thick water phantom. However, with the increase in image quality the associated cost of a crystalline scintillator is greatly increased, as is the complexity of man- ufacturing a crystal without defects. In lieu of a crystalline scintillator, Breitbach et al.18 showed that the copper conver- sion plate and Gb2S2O:Tb could be replaced with a segmented scintillator composed of a Gb2S2O ceramic. Using the same imaging beam line as Faddegon et al.,7 they showed that it is possible to increase the CNR in CBCT images by a factor of 1.64 compared to the unaltered EPID with no statistically significant change in spatial resolution. It is clear that by altering the detector composition, it is possible to significantly affect image contrast and CNR. How- ever, to date no one has quantitatively measured the effect the copper conversion plate has on image CNR in standard MV EPID detectors. Through measurement we investigate (i) the effect that the thickness of the copper conversion plate has on a 2.35 and 7.00 MV/carbon spectra and the associ- ated detector response, and (ii) the resultant changes in image quality, both qualitatively and quantitatively. This paper in- cludes the following: In Sec. II, we present the methodology on beam generation, the imaging system, Monte Carlo mod- els, and imaging protocols. In Sec. III, we present results and discussion, which include Monte Carlo simulations, planar, and CBCT image quality analysis. II. MATERIALS AND METHODS II.A. Low-Z target beam production The low-Z target beams were generated using a Varian 2100EX linear accelerator (Varian Medical, Inc., Palo Alto, CA). Two separate carbon targets were investigated (i) 7.6 mm thick (2.35 MV/carbon) and (ii) 22.8 mm thick (7.00 MV/carbon). The thicknesses of the carbon targets were calculated to coincide with the continuous slowing down approximation range for 2.50 MeV and 7.00 MeV electrons. While it has been shown that an increase in low-Z target thickness slightly degrades resolution,12 a “full-thickness” target alleviates the need for a polystyrene filter to absorb transmitted electrons and increases the production of lower energy photons.3 As described previously,10–12 when used experimentally, the carousel is operated in manual mode and the appropriate target is rotated into the beam line via a rotary switch. For the 2.35 MV/carbon beam, the linac is operated in electron mode and equipped with a customized 4 MeV program board (used for research purposes only). For the 7.00 MV/carbon beam, a standard 6 MeV program board is employed. To generate the 2.35 MeV beam, the bending magnet was adjusted to tune the beam energy down from the nominal operating point of 4.5 down to 2.35 MeV. To account for the diminished beam current observed with the reduced mean electron energy, the electron gun current, grid voltage, solenoid, and buncher cavity steering set points were adjusted to maximize electron current. The design and installation of similar targets similar to those used in this study have been published recently11,12 and will not be explained in full here. Briefly, targets were mounted to the beam side of the carousel and vertically offset 0.9 cm from the beryllium exit window of the primary collimation vacuum. By keeping the target as close as possible to the beryllium exit window, the amount of electron scattering in air is minimized.11 It is possible to decrease mean electron energy below 2.35 MeV, however, this significantly increases acquisition times, degrades spatial resolution, and has no noticeable advantage with respect to image CNR. II.B. Imaging system An aS1000/IAS3 imaging system (Varian Medical Sys- tems, Inc.) was used for all imaging. The system was setup as a stand-alone configuration separate from the clinical imag- ing system on the treatment unit. The aS1000 panel (IDU20) has an active area of 30 × 40 cm2 and includes a 1.0 mm Cu buildup plate, a Gd2O2S:Tb scintillating phosphor layer and a 1024 × 768 array of photodiodes switched by thin-film tran- sistors deposited on a glass substrate. The number of monitor units (MU) per exposure was varied to control the dose to the phantoms, where the IAS3 allows centi-MU control of this parameter. Prior to imaging, flood field, and dark field cali- brations were performed where the former has the effect of correcting for the forward peaked nature of the low-Z target, flattening filter free beam. II.C. Monte Carlo simulations In order to study the effects of the copper conversion plate on the 2.35 and 7.00 MV spectra, Monte Carlo simulations were run using BEAMnrc.19 The linac and IDU20 were modeled in accordance to the physical setup using validated models provided by Orton and Robar.11 It should be noted that these models were originally of an aS500, however between the two detectors, the physical composition is relatively consistent with the main difference being the number of pixels within the active area of the detector. 2.0 × 107 incident electron histories were run for both 2.35 and 7.00 MeV beams. Linac collimation was set to match the active area of the detector at a source to detector distance (SDD) of 140 cm. Selective bremsstrahlung splitting was used, and as recommended by Rogers et al.,20 an effective splitting field size of 50 × 50 cm2 was used for source to sur- face distance (SSD) of 140 cm with minimum and maximum bremsstrahlung splitting values of 200 and 1000, respectively. Medical Physics, Vol. 39, No. 9, September 2012
  • 3. 5364 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5364 FIG. 1. Diagram of the full imaging system. (a) A schematic drawing on the linac and detector. (b) A photo of the detector with the cover off with copper plates and Cerrobend weights. Global electron (ECUT) and photon (PCUT) cut-off energies of 0.700 MeV and of 0.010 MeV, respectively, were used. The resulting phase space was recorded at the exit side of the copper conversion plate, above the phosphor. To study the effect of the copper conversion plate thickness on the energy spectra, the thickness was varied in 0.2 mm increments from 0 mm to 1.6 mm, corresponding with the physical setup of experiments, as described below. To determine the spectral distributions, BEAMdp (Ref. 21) was used to analyze the phase space data. The resulting data were analyzed and plotted using MATLAB (Mathworks, Natick, MA). The IDU20 was also modeled in DOSXYZnrc (Ref. 22) to determine the effect the thickness of the copper conver- sion plate has on detector response. Similar to the methods used by Roberts et al.,9,15 mono-energetic pencil beams of varying energy from 0.010 MeV to 7.00 MeV were inci- dent on the detector, and the detector response was quanti- fied by scoring dose deposited in the phosphor screen per incident particle. 5.0 × 107 incident photon histories were used for all simulations, with PCUT and ECUT set to 0.010 MeV and 0.521 MeV, respectively. Detector response sim- ulations were repeated for 0 mm, 1.0 mm, and 1.6 mm copper thicknesses. The dose distributions were analyzed with MATLAB and plotted with energy fluence for 2.35 and 7.00 MV/carbon beams. II.D. Planar imaging To observe changes in image quality with varying thick- nesses of copper, the gantry was rotated to 0◦ (in IEC coordi- nates) and the detector was oriented perpendicular to the cen- tral axis at a SDD of 140 cm, as shown in Fig. 1. To decrease the amount of backscatter radiation, Styrofoam blocks sup- ported the detector from the floor. The detector cover, foam spacing materials, and copper conversion plate were removed. To vary the thickness of copper, 14 × 14 × 0.02 cm3 copper sheets were stacked on the paper sheet covering the scintilla- tor. This allowed for the thickness of copper to range from 0 mm to 1.6 mm in 0.2 mm increments. To compress the cop- per sheets, a customized Cerrobend frame was manufactured to apply pressure around the periphery of the stacked cop- per sheets, and the weight of the frame was increased until it was observed that all layers of copper were compressed firmly against the panel. To quantify changes in image quality, a pla- nar CNR and QC3 spatial resolution phantom23 were alter- nately positioned on the central axis. The CNR phantom con- sisted of various tissue substitutes surrounded by polystyrene (electron density relative to water, ρe w = 1.08). The tissue substitutes were 2.5 cm in diameter and 3.0 cm thick and sim- ulated cortical bone 60% (ρe w = 1.69), cortical bone 30% (ρe w = 1.28), inner bone (ρe w = 1.09), brain (ρe w = 1.05), and lung (ρe w = 0.44). A relatively thin phantom was used as it allows sensitivity in measuring changes of CNR with cop- per thickness. The beam was collimated to 10 × 10 cm2 at a SSD of 140 cm. An imaging dose of 0.02 cGy was deliv- ered to the center of the CNR phantom to be representative of the approximate dose per projection given in CBCT. A higher dose of 0.19 cGy was delivered to center of the QC3 phan- tom for spatial resolution measurements. A higher dose was used as a consequence of high noise at 0.02 cGy resulting in spurious results. CNR was calculated as CNR = Pmaterial − Ppolystyrene σ2 material + σ2 polystyrene , (1) where Pmaterial is the average pixel value within the tis- sue substitute, Ppolystyrene is the average pixel value in the Medical Physics, Vol. 39, No. 9, September 2012
  • 4. 5365 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5365 surrounding polystyrene, σmaterial is the average noise within the tissue substitute, and σpolystryene is the average noise within the surrounding polystyrene. Error bars were found by calcu- lating the mean and standard deviation of CNR measured in multiple images. Spatial resolution was analyzed according to the method outlined by Rajapakshe et al.,23 where the relative modulation transfer function (RMTF) is defined as RMTF = M(f ) M(f1) , (2) where M(f) is the output modulation of the line pair in a re- gion of interest and M(f1) is the output modulation for the lowest frequency line pair region. As suggested by Droege and Morin,24 the output modulation is obtained by using the relationship between the signal amplitude and its variance de- fined as M2 (f ) = σ2 m (f ) − σ2 (f ), (3) where σm 2 (f) is the measured total variance within a region of interest, and σ2 (f) is the variance due to random noise within the region of interest. The variance due to random noise as calculated by Rajapakshe et al.23 is σ2 (f ) = σ2 sub 2 , (4) where σ2 sub(f) represents the variance within the region of in- terest of two subtracted images. Uncertainty in these measure- ments was estimated by calculating the mean and standard deviation of RMTF measured in multiple image sets. Planar imaging of a sheep head was done to qualitatively observe the effect the copper conversion plate has on planar image quality at both low and high energies. The setup for imaging of the head was as follows: The gantry and couch were rotated to 90◦ . The imaging panel was placed in a stand located on the couch with fine adjustments screws and lev- eled to ensure orthogonality of the panel and the beam axis. The head was centered on isocenter and a SDD of 130 cm was used. All imaging involved an approximate dose of 0.26 cGy delivered to the center of the head. Images were first taken with the copper conversion plate within the detector. The cop- per plate was then removed and replaced with a 2.1 mm sheet of foam (to ensure adequate compression of the panel assem- bly) and the detector was reassembled. Images were then ac- quired without the copper conversion plate present within the detector. Images were analyzed in MATLAB and were com- pared with identical gray level window settings. II.E. CBCT imaging The CBCT images were acquired using the method de- scribed previously by Robar et al.10 Briefly, this involved arranging the beam and detector as described above for planar imaging of the sheep head and rotating the object on a rotation stage consisting of a microstepping motor (Intelligent Mo- tion Systems, Inc., Marlborough, CT). The rotation stage was aligned such that the axis was at the isocenter. Image acquisi- tion control was accomplished using MATLAB with the Digital I/O toolbox. All projection images were saved in DICOM for- mat and subsequently read by the reconstruction software. Im- age reconstruction was done using the Feldkamp Davis Kress (FDK) algorithm25 implemented in MATLAB. The phantom used for CBCT was cylindrical, 12.1 cm in diameter by 6.1 cm in length, composed of polystyrene and contained the same tissue substitutes as described above for planar imaging. Tis- sue substitute materials were 2.5 cm in diameter and 7 cm in length. CBCT images were acquired both with and with- out the copper conversion plate present in the detector, with central doses to the phantom ranging from 1.1 to 11.8 cGy. Imaging was repeated for both the 2.35 MV/carbon and 7.00 MV/carbon beams. CNR was measured as described in Eq. (1), with the exception that error bars were calculated by measuring the mean and standard deviation of CNR in three slices above and below the analyzed slice. To observe qualitative changes, CBCT imaging of the sheep head was done as well, for both beam energies, with and without the copper plate. An approximate imaging dose of 4.1 cGy was delivered to the center of the head, located on isocenter. The resultant images were analyzed in MATLAB and were compared with identical gray level window settings. III. RESULTS AND DISCUSSION III.A. Monte Carlo characterization Figure 2(a) shows modeled fractional fluences at the exit side of the copper conversion plate, illustrating the effect of various copper thicknesses on the energy spectrum, over the range from 25 keV to 150 keV, for both 2.35 and 7.00 MV/carbon beams. When increasing copper thickness greater than 0 mm, the fluence below 100 keV is significantly attenuated [Fig. 2(a)], and accordingly the percentage of di- agnostic photons is reduced. It should be noted that for the copper-free cases, the increase in fluence below 38 keV for the 0 mm copper spectra was due to the presence of barium (k-edge at 37.45 keV) within the detector cover. For the 2.35 MV/carbon beam, the diagnostic photon population (25 keV–150 keV) was decreased by 20% [Fig. 2(b)] with the standard 1.0 mm copper conversion plate in place, com- pared to 0 mm of copper. A corresponding 15% decrease was observed for the 7.00 MV/carbon beam. Figure 3 illus- trates the response of the IDU20 with various thicknesses of copper as well as the photon spectra produced with the 2.35 and 7.00 MV/carbon beams. With no copper present in the IDU20, detector response peaks at approximately 60 keV and decreases rapidly with increasing photon energy. When cop- per was placed within the IDU20, the response at 60 keV was suppressed by factors of 6.4 and 8.5 for 1.0 mm and 1.6 mm thicknesses, respectively. At 0.8 MeV, the response was ap- proximately the same regardless of copper thickness, above which the responses diverge with the copper responses in- creasing with increasing energy. Figure 3 demonstrates that the 2.35 MV/carbon beam contains a substantial fluence of photons within the peak detector response region (approxi- mately 30–120 keV). For the 7.00 MV/carbon beam peak flu- ence occurs at approximately 400 keV, at which point the re- sponse is locally minimized for the copper-less detector. In Medical Physics, Vol. 39, No. 9, September 2012
  • 5. 5366 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5366 FIG. 2. (a) The diagnostic spectral distribution with 0 mm, 0.4 mm, 1.0 mm, and 1.6 mm of copper present in the detector for 2.35 MV/carbon and 7.00 MV/carbon beams. (b) The percentage of photons within the diagnostic energy domain (25–150 kV) at the phosphor surface as a function of copper thickness. summary, the results of Figs. 2 and 3 indicate that copper removal is advantageous for either low-Z target beam with regard to maximizing the diagnostic population reaching the phosphor, and particularly for the 2.35 MV/carbon beam, al- lows the energy spectrum to be well matched to the detector response. The detector response curves shown here agree with similar results given by Roberts et al.9 and Faddegon et al.7 for an Elekta iViewGT (Elekta AB, Stockholm, Sweden) and FIG. 3. Response curves for 0 mm, 1.0 mm, and 1.6 mm thick copper layers within the IDU20 (left axis) and photon energy fluence produced with the 2.35 MV/carbon and 7.00 MV/carbon beams (right axis). Medical Physics, Vol. 39, No. 9, September 2012
  • 6. 5367 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5367 FIG. 4. Planar image contrast-to-noise ratio as a function of copper thickness for both 2.35 MV/carbon and 7.00 MV/carbon beams, for cortical bone 60% (top-left), cortical bone 30% (top-right), inner bone (middle-left), brain (middle-right), and lung (bottom-left). a Perkin Elmer XRD 1640 AG9 panel (Perkin Elmer, Inc., Waltham, MA), respectively. Both of these detectors include a 1.0 mm copper plate, a gadolinium oxysulphate scintillat- ing layer, and an array of photodiodes.16,26 Both papers have shown that a high-Z scintillating material strongly interacts with the diagnostic portion of the energy spectrum. Roberts et al.9,15 also show the detector response for an Eleckta XVI panel (Elekta AB, Stockholm, Sweden) which differs substan- tially in design, with the omission of a copper conversion plate and replacement of the gadolinium oxysulphate scin- tillating layer by columnar, thallium doped, cesium iodide (CsI:Tl) crystal. The responses of this detector and the IDU20 FIG. 5. Relative modulation transfer function (RMTF) for 0.0 mm, 1.0 mm, and 1.6 mm thick copper layers in the IDU20 detector for 2.35 MV/carbon and 7.00 MV/carbon beams. Medical Physics, Vol. 39, No. 9, September 2012
  • 7. 5368 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5368 FIG. 6. Planar imaging of a sheep head at 0.26 cGy (at isocenter) for both 0.0 mm (right) and 1.0 mm (left) of copper within the detector at both 2.35 (bottom) and 7.00 MV/carbon (top). For comparison, the images were nor- malized to the pixel value at a point within all the images and with identical gray level window settings. without copper are similar in that they share a peak response at approximately 50–70 keV and a local minimum between approximately 500–600 keV, with the main difference in re- sponses being in energies less than 40 keV. III.B. Planar image quality Figure 4 shows image CNR as a function of copper thick- ness for various tissue equivalent materials. For most materi- als, with the exception of lung, image CNR decreases quickly with the addition of the first few 0.2 mm thick layers of cop- per and approaches a relatively constant CNR value beyond 0.8 mm of copper. The rapid degradation in CNR with the first few layers corresponds to the decrease in the diagnostic fluence with copper shown in Fig. 2; both the fluence in the diagnostic range and CNR show the largest decreases with the addition of the first 0.4 mm of copper. With no copper present, image CNR was increased by factors of 1.6, 2.3, 3.2, 1.4, and 2.3 for cortical bone 60%, cortical bone 30%, inner bone, brain, and lung, respectively, compared to the standard 1.0 mm of copper. Similar gains in CNR were observed for the 7.00 MV/carbon beam, with factor increases ranging from 1.6 to 4.0; however, the absolute values of CNR were typically less than half of those observed at 2.35 MV. The results pre- sented here confirm the modeled results of Orton and Robar,11 in that a gain in image contrast was observed with the removal of the copper layer. Figure 5 shows RMTF at various thicknesses of copper for the 2.35 and 7.00 MV/carbon beams. At most frequencies, there appears to be little variation with changing copper thick- ness. At 0.25 lp/mm, there is a detectable but small change in RMTF, with relative improvement of RMTF by 18% and 15% for the 2.35 and 7.00 MV/carbon beams, with removal of the 1.0 mm copper layer. The frequency where RMTF equals 0.5, or f50, did not vary significantly with copper removal, for either low-Z target beam. With the 1.0 mm copper layer, here FIG. 7. Profiles of Fig. 6 with and without the copper conversion plate within the detector for 2.35 MV/carbon and 7.00 MV/carbon beams. Medical Physics, Vol. 39, No. 9, September 2012
  • 8. 5369 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5369 FIG. 8. Cone beam computed tomography images at approximately 4.1 cGy with copper plate (top) and without copper (bottom) in the detector, with the 7.00 MV/carbon and 2.35 MV/carbon beams. The tissue equivalent materials imaged are (clockwise starting from the darkest) lung, cortical bone 60%, cortical bone 30%, inner bone, and brain. we see decrease of f50 0.03 lp/mm when reducing electron energy from 7.00 to 2.35 MeV. In comparison, Connell and Robar12 reported a change in f50 of approximately 0.02 and 0.03 lp/mm for beryllium and aluminum target beams, when reducing the incident electron energy from 7.0 to 4.5 MeV. Figure 6 shows planar sagittal sheep images both with and without the copper conversion plate with the 2.35 and 7.00 MV/carbon beams. The images show increased image contrast with the removal of the copper conversion plate, com- pared to images with the copper conversion plate in place. This is most noticeable within the squamous, nasal, and mandible regions. These differences are highlighted in the profiles shown in Fig. 7, in which the contrast is notably in- creased with the removal of the copper conversion plate for the 2.35 MV/carbon beam. Figure 7 also shows slight loss of fine detail when using the 2.35 MV/carbon beam com- pared to 7.00 MV/carbon. As noted by Connell and Robar,12 this is largely a result of electron beam broadening when de- creasing incident electron energy. However, the changes are relatively minor between images with and without the cop- per conversion plate, which agrees with the results shown in Fig. 6. III.C. CBCT image quality Figure 8 shows CBCT images of tissue equivalent materi- als in polystyrene both with and without copper in the IDU20 for the 2.35 MV/carbon and 7.00 MV/carbon beams. The CBCT images show greater contrast with the copper conver- sion plate removed both at 7.00 and 2.35 MV/carbon, with the greatest improvement seen for 2.35 MV/carbon. This is FIG. 9. Contrast-to-noise ratios of tissue substitutes in CBCT images as a function of dose, comparing the 2.35 MV/carbon and 7.00 MV/carbon beams both with and without copper present in the detector. Medical Physics, Vol. 39, No. 9, September 2012
  • 9. 5370 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5370 FIG. 10. CBCT images of a sheep head at approximately 4.1 cGy (at isocen- ter) with copper plate (top) and without copper (bottom) in the detector, with the 7.00 MV/carbon and 2.35 MV/carbon beams. evident also in the corresponding CNR curves as a function of dose, shown in Fig. 9. At approximately 4.1 cGy and for the 2.35 MV/carbon beam, for example, copper removal pro- duced improvement of CNR by factors of 1.5, 1.9, 2.1, 1.4, and 1.3 for cortical bone 60%, cortical bone 30%, inner bone, brain, and lung, respectively. Similar factor increases are seen for the 7.00 MV/carbon beam, but as for the planar imag- ing results, CNR values are typically half of that observed at 2.35 MV/carbon. The CNR improvement with copper re- moval seen here is comparable in magnitude to the factor 1.64 gain in CNR observed by Breitbach et al.18 when using a segmented ceramic Gb2S2O scintillator, compared to a stan- dard 1.0 mm copper conversion plate with Gb2S2O scintillat- ing phosphor. However, this group used a different beam-line7 within a Siemens linac at 4.2 MV. Figure 10 shows CBCT images of the sheep head with and without the copper conversion plate with the 2.35 MV/carbon and 7.00 MV/carbon beams. These images highlight the effect the presence of copper in the IDU20 has on image contrast-to- noise characteristics. This can be seen in the bony structures of the skull, and in the nasal cavity, where at 2.35 MV/carbon are clearly visualized without copper, compared to the image acquired with the copper layer. These trends are also present in the 7.00 MV/carbon images; however, the CNR is clearly degraded. These images taken here agree with the qualitative findings reported by Faddegon et al.,7 in which there is an overall improvement in bone and soft tissue visibility with copper removal; however, in this work we show that the re- duction of energy to 2.35 MeV offers an overall improve- ment in image quality. While we believe that with a more efficient detector such as a crystalline scintillator would be preferable,9,15 these results demonstrate that substantial im- provements in image quality may be realized with a standard MV EPID, following removal of the copper conversion plate. IV. CONCLUSIONS In this work, we have shown that the removal of the copper conversion plate within an IDU20 EPID can significantly im- prove image contrast and CNR when used in conjunction with low-Z target imaging beams. The presence of the copper con- version plate caused a decrease in the diagnostic population of 20% and 15% for 2.35 MV/carbon and 7.00 MV/carbon beams, respectively. The peak response within the detector without copper was reduced by a factor of 6.4 when 1.0 mm of copper was present in the detector. Planar images acquired with a copper-less detector showed a factor gain in CNR rang- ing from 1.4 to 3.2 and 1.6 to 4.0, for 2.35 MV/carbon and 7.00 MV/carbon beams, respectively, compared to the stan- dard 1.0 mm of copper. Similar gains in CNR were observed in CBCT images, where CNR was improved by a factor rang- ing from 1.3 to 2.1. Spatial resolution was not affected signif- icantly by copper removal, for either beam. Qualitative imag- ing of a sheep head showed that there was an increase in pla- nar and CBCT image contrast with a copper-less detector. ACKNOWLEDGMENTS The authors are grateful for support provided by Varian Medical Systems for the funding of this project and the Nat- ural Sciences and Engineering Research Council of Canada for their additional financial support. We would like to thank Robert Moran for his vital electronic and technical support, Ian Porter for fabrication of apparatus, Dr. Robin Kelly and Dr. Mammo Yewondwossen for their knowledge and teach- ing of techniques utilized in this project. a)Author to whom correspondence should be addressed. Electronic mail: david.parsons@dal.ca 1O. Morin, A. Gillis, J. Chen, M. Aubin, M. K. Bucci, M. Roach, and J. Pouliot, “Megavoltage cone-beam CT: System description and clinical applications,” Med. Dosim. 31, 51–61 (2006). 2D. A. Jaffray and J. H. Siewerdsen, “Cone-beam computed tomography with a flat-panel imager: Initial performance characterization,” Med. Phys. 27, 1311–1323 (2000). 3D. M. Galbraith, “Low-energy imaging with high-energy bremsstrahlung beams,” Med. Phys. 16, 734–746 (1989). 4D. W. Mah, D. M. Galbraith, and J. A. Rawlinson, “Low-energy imaging with high-energy bremsstrahlung beams: Analysis and scatter reduction,” Med. Phys. 16, 653–665 (1993). 5O. Z. Ostapiak, P. F. O’Brien, and B. A. Faddegon, “Megavoltage imaging with low Z targets: Implementation and characterization of an investiga- tional system,” Med. Phys. 25, 1910–1918 (1998). 6A. Tsechanski, A. F. Bielajew, S. Faermann, and Y. Krutman, “A thin target approach for portal imaging in medical accelerators,” Phys. Med. Biol. 43, 2221–2236 (1998). 7B. A. Faddegon, V. Wu, J. Pouliot, B. Gangadharan, and A. Bani-Hashemi, “Low dose megavoltage cone beam computed tomography with an unflat- tened 4 MV beam from a carbon target,” Med. Phys. 35, 5777–5786 (2008). 8S. Flampouri, P. M. Evans, F. Verhaegen, A. E. Nahum, E. Spezi, and M. Partridge, “Optimization of accelerator target and detector for portal imaging using Monte Carlo simulation and experiment,” Phys. Med. Biol. 47, 3331–3349 (2002). 9D. A. Roberts, V. N. Hansen, A. C. Niven, M. G. Thompson, J. Seco, and P. M. Evans, “A low Z linac and flat panel imager: Comparison with the conventional imaging approach,” Phys. Med. Biol. 53, 6305–6319 (2008). 10J. L. Robar, T. Connell, W. Huang, and R. G. Kelly, “Megavoltage planar and cone-beam imaging with low-Z targets: Dependence of image qual- ity improvement on beam energy and patient separation,” Med. Phys. 36, 3955–3963 (2009). Medical Physics, Vol. 39, No. 9, September 2012
  • 10. 5371 D. Parsons and J. L. Robar: Effect copper conversion plates low-Z target image quality 5371 11E. J. Orton and J. L. Robar, “Megavoltage image contrast with low-atomic number target materials and amorphous silicon electronic portal imagers,” Phys. Med. Biol. 54, 1275–1289 (2009). 12T. Connell and J. L. Robar, “Low-Z target optimization for spatial res- olution improvement in megavoltage imaging,” Med. Phys. 37, 124–131 (2010). 13D. Sawkey, M. Lu, O. Morin, M. Aubin, S. S. Yom, A. R. Gottschalk, A. Bani-Hashemi, and B. A. Faddegon, “A diamond target for megavoltage cone-beam CT,” Med. Phys. 37, 1246–1253 (2010). 14D. A. Roberts, V. N. Hansen, M. G. Thompson, G. Poludniowski, A. Niven, J. Seco, and P. M. Evans, “Comparative study of a low-Z cone-beam com- puted tomography system,” Phys. Med. Biol. 56, 4453–4464 (2011). 15D. A. Roberts, V. N. Hansen, M. G. Thompson, G. Poludniowski, A. Niven, J. Seco, and P. M. Evans, “Kilovoltage energy imaging with a radiotherapy linac with a continuously variable energy range,” Med. Phys. 39, 1218–26 (2012). 16B. A. Faddegon, M. Aubin, A. Bani-Hashemi, B. Gangadharan, A. R. Gottschalk, O. Morin, D. Sawkey, V. Wu, and S. S. Yom, “Com- parison of patient megavoltage cone beam CT images acquired with an un- flattened beam from a carbon target and a flattened treatment beam,” Med. Phys. 37, 1737–1741 (2010). 17D. A. Jaffray, J. J. Battista, A. Fenster, and P. Munro, “Monte Carlo studies of x-ray energy absorption and quantum noise in megavoltage transmission radiography,” Med. Phys. 22, 1077–88 (1995). 18E. K. Breitbach, J. S. Maltz, B. Gangadharan, A. Bani-Hashemi, C. M. Anderson, S. K. Bhatia, J. Stiles, D. S. Edwards, and R. T. Flynn, “Image quality improvement in megavoltage cone beam CT using an imaging beam line and a sintered pixelated array system,” Med. Phys. 38, 5969–5979 (2011). 19D. W. Rogers, B. A. Faddegon, G. X. Ding, C. M. Ma, J. We, and T. R. Mackie, “BEAM: A Monte Carlo code to simulate radiotherapy treat- ment units,” Med. Phys. 22, 503–524 (1995). 20D. W. O. Rogers, B. Walters, and I. Kawrakow, “BEAMnrc users manual,” NRCC Report No. PIRS-0509(A)revL (NRCC, Ottawa, 2011). 21C. M. Ma and D. W. O. Rogers, “BEAMdp users manual,” NRCC Report No. PIRS-0509(C)revA (NRCC, Ottawa, 2009). 22I. Kawrakow, “Accurate condensed history Monte Carlo simulation of elec- tron transport. I. EGSnrc, the new EGS4 version,” Med. Phys. 27, 485–498 (2000). 23R. Rajapakshe, K. Luchka, and S. Shalev, “A quality control test for elec- tronic portal imaging devices,” Med. Phys. 23, 1237–1244 (1996). 24R. T. Droege and R. L. Morin, “A practical method to measure the MTF of CT scanners,” Med. Phys. 9, 758–760 (1982). 25L. Feldkamp, L. Davis, and J. Kress, “Practical cone-beam algorithm,” J. Opt. Soc. Am. A 1, 612–619 (1984). 26L. Parent, J. Seco, P. M. Evans, A. Fielding, and D. R. Dance, “Monte Carlo modelling of a-Si EPID response: The effect of spectral variations with field size and position,” Med. Phys. 33, 4527 (2006). Medical Physics, Vol. 39, No. 9, September 2012