CAUTION: There are live circuits inside the controller even when the power switch is in the
OFF position! Servicing and troubleshooting the controller must be performed by Qualified
Service Personnel.
WM1, WIND, W400, W100, W600, Intuition-6: ORP Sensor
Troubleshooting Guide (Electrode and Housing)
Table of Contents
Section 1: Sensor (Electrode) Responds Slowly............................................................................ 2
Section 2: Sensor Reading is Stuck on one Value .......................................................................... 3
Section 3: Sensor Reads Low or High Versus a Calibrated Handheld Meter; Cannot Calibrate.... 4
Section 4: Sensor Error................................................................................................................... 5
Section 5: ORP Reading is Unstable............................................................................................... 7
2
CAUTION: There are live circuits inside the controller even when the power switch is in the
OFF position! Servicing and troubleshooting the controller must be performed by Qualified
Service Personnel.
Section 1: Sensor (Electrode) Responds Slowly
Step 1:
The pH or ORP electrodes require periodic cleaning and calibration. These electrodes are like batteries
and their voltage outputs will change with time even if they are not being used.
 The method of cleaning the electrode will depend upon the coating, as well as the materials of
construction of the electrode. Do not use a solvent that will attack the electrode
 Oily coatings should be removed with a mild detergent or isopropyl alcohol.
 Hard scales such as calcium carbonate can usually be removed with a dilute hydrochloric acid
solution.
 Soft coatings can be removed using a soft cloth or soft toothbrush.
 A two point calibration should always be performed after cleaning the electrode.
Step 2:
If cleaning does not speed up the electrode's response
 Replace the electrode.
3
CAUTION: There are live circuits inside the controller even when the power switch is in the
OFF position! Servicing and troubleshooting the controller must be performed by Qualified
Service Personnel.
Section 2: Sensor Reading is Stuck on one Value
Step 1:
 If the sensor reading is stuck on one number
 If it is stuck near 0 mV then there is most likely an electrical short.
 If it is stuck at a high mV reading (~1200), then the pre-amp circuit is most likely open.
 Broken glass on the electrode tip could cause an open circuit which sends the signal drifting
to the end, or lower, or all the way to Sensor Error.
Note: If water leaks into the preamp, you may get a mV reading, but the reading will not hold accuracy.
Step 2:
Check the sensor (electrode) for an electrical short.
 Power down the controller.
 Remove/unthread the electrode from the housing.
 Power the controller up.
 If reading drifts then the electrode has a short. Replace the electrode. If the reading is still
stuck on one number, proceed to Step 3.
Step 3:
Check the preamp (housing) for an electrical short.
 Power down the controller.
 Disconnect the preamp from the controller.
 Power the controller up.
 If reading drifts then the preamp has a short.
Replace the preamp. If the reading is stuck on
one number, proceed to Section 4.
4
CAUTION: There are live circuits inside the controller even when the power switch is in the
OFF position! Servicing and troubleshooting the controller must be performed by Qualified
Service Personnel.
Section 3: Sensor Reads Low or High Versus a Calibrated Handheld
Meter; Cannot Calibrate
Step 1:
Perform a sensor calibration in accordance with the
controller manual.
 If the sensor cannot be calibrated, first thing to try is
clean the sensor.
 If the sensor cannot be calibrated after cleaning,
proceed to Step 2.
Step 2:
Go to the Sensor mV menu. Resetting the calibration settings.
 Go into the Sensor menu, go to Sensor Type, change it from ORP to pH, hit ENTER to save this
change. Go back into the Sensor Type menu, change it from pH back to ORP, hit ENTER to save
this change.
 Try to calibrate the sensor in accordance with the controller manual. If you still cannot
calibrate, proceed to Step 3.
Step 3:
Go to the Sensor mV menu. Resetting to factory defaults.
 Proceed as follows depending on which controller model you have.
 For a WM1 or WIND controller, go into the Sensor Input menu, and press the reset defaults
button. Proceed to Section 4.
 For a WPH400 controller reset the controller to factory defaults as follows.
 Caution: Resetting defaults erases any user defined settings programmed into the
controller.
 Power down the controller, press and hold the ENTER and EXIT keys, power up the
controller keeping the ENTER and EXIT keys pressed in until the controller boots up.
 When the display comes up, you will be asked if you want to reset defaults, release the
ENTER and EXIT keys and then press ENTER for yes. Re-check the mV and pH values. If the
pH and mV values do not correlate, then either the preamp or the electrode is faulty.
Proceed to Section 4.
5
CAUTION: There are live circuits inside the controller even when the power switch is in the
OFF position! Servicing and troubleshooting the controller must be performed by Qualified
Service Personnel.
+5V
T-
T+
G
R
N
/
W
H
T
W
H
I
T
E
/
G
R
N
SHIELD
-5V
IN+ IN-
O
R
N
/W
H
T
WH
ITE
/OR
N
B
L
U
/W
H
T
W
H
I
T
E
/
B
L
U
WEL Electrode
Diffamp Housing
(102581, 102606,
102582,102607,
102758, 102759) pH/ORP electrode
cartridge
TO GRND
Dashed lines indicate
optional temperature
compensation
Section 4: Sensor Error
Step 1:
Perform a self-test for this sensor input.
 If the self-test passes, then the problem is not inside
the controller. The preamp or electrode is faulty.
Continue to Step 2.
 If the self-test fails, disconnect sensor from the
controller and repeat the self-test and if it still fails,
send the controller back for repair. Contact your
distributor or Walchem Customer Service with the
controller model number and serial number to initiate the Return Authorization process.
Step 2:
On the ORP sensor input terminals inside the controller, using a multimeter:
 Measure the VDC across IN- and +5V, should read +5 VDC +/- 5%
 Measure the VDC across IN- and –5 VDC, should read –4.6 VDC +/- 5%.
 If these measurements are out of spec with and without the preamp connected, the controller is
faulty. Contact your distributor or Walchem Customer Service with the controller model number
and serial number to initiate the Return Authorization process.
 If these measurements are in spec without the preamp connected, then replace the preamp. If
these measurements are in spec with and without the preamp, then continue to Step 3.
 Make sure the preamp is wired
correctly into the controller, as
shown.
 Be cautious about junction box
connections. The sensor cable
shielding should only be
grounded on one end (inside the
controller) otherwise there will
be a ground loop.
6
CAUTION: There are live circuits inside the controller even when the power switch is in the
OFF position! Servicing and troubleshooting the controller must be performed by Qualified
Service Personnel.
Step 3:
Check the sensor cable, sensor extension cable (if installed).
 Make sure you are using shielded sensor cable, 3 twisted pair wire. Use Walchem p/n 102535
or Belden 9680.
 Make sure the sensor cable is not run near, or in the same conduit as AC voltage wire (All wires
carry voltage. We are concerned with adjacent wires that are carrying voltage above 24
VAC). If you suspect the sensor cable may be the problem, then take the existing sensor (with
the 5 or 20 feet of cable), and wire it directly into the controller by-passing the sensor
extension cable. If the sensor reading gets corrected, then the sensor extension cable, or
where it was routed, was the problem.
 If wiring the sensor cable directly into the controller does not solve the issue, then replace the
preamp.
7
CAUTION: There are live circuits inside the controller even when the power switch is in the
OFF position! Servicing and troubleshooting the controller must be performed by Qualified
Service Personnel.
Section 5: ORP Reading is Unstable
Step 1:
The ORP reading is unstable when the ORP sensor is placed in
the system water.
 Remove the ORP sensor from the system water and
place it in a beaker of water. Does the reading stabilize?
If yes, then proceed to Step 2. If no, proceed to Step 3.
Step 2:
The sensor is still in the beaker, and the reading is stable.
 Obtain a length of wire (the ends of the wire must be striped) long enough to place one end in the
water in the beaker, and the other end in the system water.
 With this length of wire connecting the beaker water to the system water, does the reading stay
stable?
 If the reading does not stay stable, this can indicate a ground loop. A ground loop could cause
readings to be stuck at a high or low value. Or there is some electrical interference in the system
water, possibly from a mixer or recirculation pump(s).
Step 3:
The sensor is still in the beaker, and the reading is still unstable.
 Power down the controller, disconnect the ORP sensor shield wire from the Earth ground stud
inside the controller.
 Does the reading stabilize?
 If the reading stabilizes, then there is possibly some electrical noise coming in through the
controller VAC mains.
 If the reading does not stay stable, then connect the controller to an alternate supply voltage
circuit. This may necessitate temporarily running an extension cord for this test. If the reading
still does not stay stable, please contact Walchem for further instructions.

More Related Content

Similar to ORP Sensor Troubleshooting Guide

Evespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in EnglishEvespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in EnglishDomotica daVinci
 
SenseTek Fireray 3000 installers handbook
SenseTek Fireray 3000 installers handbookSenseTek Fireray 3000 installers handbook
SenseTek Fireray 3000 installers handbookHans Bronkhorst
 
HELE / TLEelectronics
HELE / TLEelectronicsHELE / TLEelectronics
HELE / TLEelectronicsdjpprkut
 
Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
 Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20 Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20Tenmars Việt Nam
 
RAC-050 Engineer Training V1.0e-202202.ppt
RAC-050 Engineer Training V1.0e-202202.pptRAC-050 Engineer Training V1.0e-202202.ppt
RAC-050 Engineer Training V1.0e-202202.pptArnoldoFelix3
 
Basic electrical measuring instruments
Basic electrical measuring instrumentsBasic electrical measuring instruments
Basic electrical measuring instrumentsPramod A
 
Basicelectricalmeasuringinstruments 160111044802
Basicelectricalmeasuringinstruments 160111044802Basicelectricalmeasuringinstruments 160111044802
Basicelectricalmeasuringinstruments 160111044802shivmittal6
 
PLC input and output devices
PLC input and output devices PLC input and output devices
PLC input and output devices Ameen San
 
Diagra de instalacion de alarma prestige
Diagra de instalacion de alarma prestigeDiagra de instalacion de alarma prestige
Diagra de instalacion de alarma prestigemartinezd1
 
Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745
Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745
Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745Tenmars Việt Nam
 
Everspring sp816 user manual
Everspring sp816 user manualEverspring sp816 user manual
Everspring sp816 user manualDomotica daVinci
 
PLC INPUT&OUTPUT DEVICES.pptx
PLC INPUT&OUTPUT DEVICES.pptxPLC INPUT&OUTPUT DEVICES.pptx
PLC INPUT&OUTPUT DEVICES.pptxJannaRicaBeason
 
IRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET Journal
 
Emotional Stress Indicator and Digital Thermometer-Project-8thsem
Emotional Stress Indicator and Digital Thermometer-Project-8thsemEmotional Stress Indicator and Digital Thermometer-Project-8thsem
Emotional Stress Indicator and Digital Thermometer-Project-8thsemkaushikbandopadhyay
 

Similar to ORP Sensor Troubleshooting Guide (20)

Evespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in EnglishEvespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in English
 
SenseTek Fireray 3000 installers handbook
SenseTek Fireray 3000 installers handbookSenseTek Fireray 3000 installers handbook
SenseTek Fireray 3000 installers handbook
 
HELE / TLEelectronics
HELE / TLEelectronicsHELE / TLEelectronics
HELE / TLEelectronics
 
Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
 Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20 Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
 
RAC-050 Engineer Training V1.0e-202202.ppt
RAC-050 Engineer Training V1.0e-202202.pptRAC-050 Engineer Training V1.0e-202202.ppt
RAC-050 Engineer Training V1.0e-202202.ppt
 
Basic electrical measuring instruments
Basic electrical measuring instrumentsBasic electrical measuring instruments
Basic electrical measuring instruments
 
Basicelectricalmeasuringinstruments 160111044802
Basicelectricalmeasuringinstruments 160111044802Basicelectricalmeasuringinstruments 160111044802
Basicelectricalmeasuringinstruments 160111044802
 
PLC input and output devices
PLC input and output devices PLC input and output devices
PLC input and output devices
 
Diagra de instalacion de alarma prestige
Diagra de instalacion de alarma prestigeDiagra de instalacion de alarma prestige
Diagra de instalacion de alarma prestige
 
Analog Multimeter.pdf
Analog Multimeter.pdfAnalog Multimeter.pdf
Analog Multimeter.pdf
 
Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745
Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745
Hướng dẫn sử dụng bút thử điện không tiếp xúc Testo 745
 
huong dan su dung Testo 745
huong dan su dung Testo 745huong dan su dung Testo 745
huong dan su dung Testo 745
 
Dc dc converter
Dc dc converterDc dc converter
Dc dc converter
 
Everspring sp816 user manual
Everspring sp816 user manualEverspring sp816 user manual
Everspring sp816 user manual
 
PLC INPUT&OUTPUT DEVICES.pptx
PLC INPUT&OUTPUT DEVICES.pptxPLC INPUT&OUTPUT DEVICES.pptx
PLC INPUT&OUTPUT DEVICES.pptx
 
Ks 307 dct user's manual of PIR detector
Ks 307 dct user's manual of PIR detectorKs 307 dct user's manual of PIR detector
Ks 307 dct user's manual of PIR detector
 
IRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level Indicator
 
Quizbank quiz buzzer
Quizbank quiz buzzer Quizbank quiz buzzer
Quizbank quiz buzzer
 
Manual pop 004308
Manual pop 004308Manual pop 004308
Manual pop 004308
 
Emotional Stress Indicator and Digital Thermometer-Project-8thsem
Emotional Stress Indicator and Digital Thermometer-Project-8thsemEmotional Stress Indicator and Digital Thermometer-Project-8thsem
Emotional Stress Indicator and Digital Thermometer-Project-8thsem
 

More from Walchem, Iwaki America Inc.

Understanding Chemical Dosing Pumps Applications, Types, and Considerations.pdf
Understanding Chemical Dosing Pumps Applications, Types, and Considerations.pdfUnderstanding Chemical Dosing Pumps Applications, Types, and Considerations.pdf
Understanding Chemical Dosing Pumps Applications, Types, and Considerations.pdfWalchem, Iwaki America Inc.
 
Intuition-9™ Series Water Treatment Controllers Brochure
Intuition-9™ Series Water Treatment Controllers BrochureIntuition-9™ Series Water Treatment Controllers Brochure
Intuition-9™ Series Water Treatment Controllers BrochureWalchem, Iwaki America Inc.
 
WECT400 Series Cooling Tower Conductivity Controller.pdf
WECT400 Series Cooling Tower Conductivity Controller.pdfWECT400 Series Cooling Tower Conductivity Controller.pdf
WECT400 Series Cooling Tower Conductivity Controller.pdfWalchem, Iwaki America Inc.
 
W100 CT-BL Series Cooling Tower and Boiler Controller Manual
W100 CT-BL Series Cooling Tower and Boiler Controller ManualW100 CT-BL Series Cooling Tower and Boiler Controller Manual
W100 CT-BL Series Cooling Tower and Boiler Controller ManualWalchem, Iwaki America Inc.
 

More from Walchem, Iwaki America Inc. (16)

Understanding Chemical Dosing Pumps Applications, Types, and Considerations.pdf
Understanding Chemical Dosing Pumps Applications, Types, and Considerations.pdfUnderstanding Chemical Dosing Pumps Applications, Types, and Considerations.pdf
Understanding Chemical Dosing Pumps Applications, Types, and Considerations.pdf
 
Web Master® WIND Modbus Instruction Manual
Web Master® WIND Modbus Instruction ManualWeb Master® WIND Modbus Instruction Manual
Web Master® WIND Modbus Instruction Manual
 
Intuition-9™ Series Water Treatment Controllers Brochure
Intuition-9™ Series Water Treatment Controllers BrochureIntuition-9™ Series Water Treatment Controllers Brochure
Intuition-9™ Series Water Treatment Controllers Brochure
 
EWN-R Electromagnetic Metering Pump
EWN-R Electromagnetic Metering PumpEWN-R Electromagnetic Metering Pump
EWN-R Electromagnetic Metering Pump
 
Intuition 6 Instruction Manual
Intuition 6 Instruction ManualIntuition 6 Instruction Manual
Intuition 6 Instruction Manual
 
EZ Series Electronic Metering Pump
EZ Series Electronic Metering PumpEZ Series Electronic Metering Pump
EZ Series Electronic Metering Pump
 
Intuition 6 Water Treatment Controller Manual
Intuition 6 Water Treatment Controller ManualIntuition 6 Water Treatment Controller Manual
Intuition 6 Water Treatment Controller Manual
 
WECT400 Series Cooling Tower Conductivity Controller.pdf
WECT400 Series Cooling Tower Conductivity Controller.pdfWECT400 Series Cooling Tower Conductivity Controller.pdf
WECT400 Series Cooling Tower Conductivity Controller.pdf
 
EHE Series Electronic Metering Pump
EHE Series Electronic Metering PumpEHE Series Electronic Metering Pump
EHE Series Electronic Metering Pump
 
WebMaster ONE Water Treatment Controller Manual
WebMaster ONE Water Treatment Controller ManualWebMaster ONE Water Treatment Controller Manual
WebMaster ONE Water Treatment Controller Manual
 
W100 CT-BL Series Cooling Tower and Boiler Controller Manual
W100 CT-BL Series Cooling Tower and Boiler Controller ManualW100 CT-BL Series Cooling Tower and Boiler Controller Manual
W100 CT-BL Series Cooling Tower and Boiler Controller Manual
 
Contacting Conductivity Sensors Brochure.pdf
Contacting Conductivity Sensors Brochure.pdfContacting Conductivity Sensors Brochure.pdf
Contacting Conductivity Sensors Brochure.pdf
 
Intuition 6 Water Treatment Controllers
Intuition 6 Water Treatment ControllersIntuition 6 Water Treatment Controllers
Intuition 6 Water Treatment Controllers
 
EWN-Y Series Metering Pump
EWN-Y Series Metering PumpEWN-Y Series Metering Pump
EWN-Y Series Metering Pump
 
W100W Series Controller Manual.pdf
W100W Series Controller Manual.pdfW100W Series Controller Manual.pdf
W100W Series Controller Manual.pdf
 
W600 Series Controller Manual
W600 Series Controller ManualW600 Series Controller Manual
W600 Series Controller Manual
 

Recently uploaded

BEST ✨ Call Girls In Indirapuram Ghaziabad ✔️ 9871031762 ✔️ Escorts Service...
BEST ✨ Call Girls In  Indirapuram Ghaziabad  ✔️ 9871031762 ✔️ Escorts Service...BEST ✨ Call Girls In  Indirapuram Ghaziabad  ✔️ 9871031762 ✔️ Escorts Service...
BEST ✨ Call Girls In Indirapuram Ghaziabad ✔️ 9871031762 ✔️ Escorts Service...noida100girls
 
Mondelez State of Snacking and Future Trends 2023
Mondelez State of Snacking and Future Trends 2023Mondelez State of Snacking and Future Trends 2023
Mondelez State of Snacking and Future Trends 2023Neil Kimberley
 
Vip Dewas Call Girls #9907093804 Contact Number Escorts Service Dewas
Vip Dewas Call Girls #9907093804 Contact Number Escorts Service DewasVip Dewas Call Girls #9907093804 Contact Number Escorts Service Dewas
Vip Dewas Call Girls #9907093804 Contact Number Escorts Service Dewasmakika9823
 
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
Pharma Works Profile of Karan Communications
Pharma Works Profile of Karan CommunicationsPharma Works Profile of Karan Communications
Pharma Works Profile of Karan Communicationskarancommunications
 
Catalogue ONG NUOC PPR DE NHAT .pdf
Catalogue ONG NUOC PPR DE NHAT      .pdfCatalogue ONG NUOC PPR DE NHAT      .pdf
Catalogue ONG NUOC PPR DE NHAT .pdfOrient Homes
 
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine ServiceCall Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine Serviceritikaroy0888
 
RE Capital's Visionary Leadership under Newman Leech
RE Capital's Visionary Leadership under Newman LeechRE Capital's Visionary Leadership under Newman Leech
RE Capital's Visionary Leadership under Newman LeechNewman George Leech
 
The Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case studyThe Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case studyEthan lee
 
A DAY IN THE LIFE OF A SALESMAN / WOMAN
A DAY IN THE LIFE OF A  SALESMAN / WOMANA DAY IN THE LIFE OF A  SALESMAN / WOMAN
A DAY IN THE LIFE OF A SALESMAN / WOMANIlamathiKannappan
 
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...anilsa9823
 
Eni 2024 1Q Results - 24.04.24 business.
Eni 2024 1Q Results - 24.04.24 business.Eni 2024 1Q Results - 24.04.24 business.
Eni 2024 1Q Results - 24.04.24 business.Eni
 
Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...Roland Driesen
 
Insurers' journeys to build a mastery in the IoT usage
Insurers' journeys to build a mastery in the IoT usageInsurers' journeys to build a mastery in the IoT usage
Insurers' journeys to build a mastery in the IoT usageMatteo Carbone
 
Catalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdf
Catalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdfCatalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdf
Catalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdfOrient Homes
 
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779Delhi Call girls
 
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRLMONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRLSeo
 
Regression analysis: Simple Linear Regression Multiple Linear Regression
Regression analysis:  Simple Linear Regression Multiple Linear RegressionRegression analysis:  Simple Linear Regression Multiple Linear Regression
Regression analysis: Simple Linear Regression Multiple Linear RegressionRavindra Nath Shukla
 

Recently uploaded (20)

BEST ✨ Call Girls In Indirapuram Ghaziabad ✔️ 9871031762 ✔️ Escorts Service...
BEST ✨ Call Girls In  Indirapuram Ghaziabad  ✔️ 9871031762 ✔️ Escorts Service...BEST ✨ Call Girls In  Indirapuram Ghaziabad  ✔️ 9871031762 ✔️ Escorts Service...
BEST ✨ Call Girls In Indirapuram Ghaziabad ✔️ 9871031762 ✔️ Escorts Service...
 
Mondelez State of Snacking and Future Trends 2023
Mondelez State of Snacking and Future Trends 2023Mondelez State of Snacking and Future Trends 2023
Mondelez State of Snacking and Future Trends 2023
 
Vip Dewas Call Girls #9907093804 Contact Number Escorts Service Dewas
Vip Dewas Call Girls #9907093804 Contact Number Escorts Service DewasVip Dewas Call Girls #9907093804 Contact Number Escorts Service Dewas
Vip Dewas Call Girls #9907093804 Contact Number Escorts Service Dewas
 
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
 
Pharma Works Profile of Karan Communications
Pharma Works Profile of Karan CommunicationsPharma Works Profile of Karan Communications
Pharma Works Profile of Karan Communications
 
Catalogue ONG NUOC PPR DE NHAT .pdf
Catalogue ONG NUOC PPR DE NHAT      .pdfCatalogue ONG NUOC PPR DE NHAT      .pdf
Catalogue ONG NUOC PPR DE NHAT .pdf
 
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine ServiceCall Girls In Panjim North Goa 9971646499 Genuine Service
Call Girls In Panjim North Goa 9971646499 Genuine Service
 
RE Capital's Visionary Leadership under Newman Leech
RE Capital's Visionary Leadership under Newman LeechRE Capital's Visionary Leadership under Newman Leech
RE Capital's Visionary Leadership under Newman Leech
 
The Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case studyThe Coffee Bean & Tea Leaf(CBTL), Business strategy case study
The Coffee Bean & Tea Leaf(CBTL), Business strategy case study
 
A DAY IN THE LIFE OF A SALESMAN / WOMAN
A DAY IN THE LIFE OF A  SALESMAN / WOMANA DAY IN THE LIFE OF A  SALESMAN / WOMAN
A DAY IN THE LIFE OF A SALESMAN / WOMAN
 
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
 
Eni 2024 1Q Results - 24.04.24 business.
Eni 2024 1Q Results - 24.04.24 business.Eni 2024 1Q Results - 24.04.24 business.
Eni 2024 1Q Results - 24.04.24 business.
 
Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...
 
KestrelPro Flyer Japan IT Week 2024 (English)
KestrelPro Flyer Japan IT Week 2024 (English)KestrelPro Flyer Japan IT Week 2024 (English)
KestrelPro Flyer Japan IT Week 2024 (English)
 
Insurers' journeys to build a mastery in the IoT usage
Insurers' journeys to build a mastery in the IoT usageInsurers' journeys to build a mastery in the IoT usage
Insurers' journeys to build a mastery in the IoT usage
 
Catalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdf
Catalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdfCatalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdf
Catalogue ONG NƯỚC uPVC - HDPE DE NHAT.pdf
 
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
Best VIP Call Girls Noida Sector 40 Call Me: 8448380779
 
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRLMONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
MONA 98765-12871 CALL GIRLS IN LUDHIANA LUDHIANA CALL GIRL
 
Regression analysis: Simple Linear Regression Multiple Linear Regression
Regression analysis:  Simple Linear Regression Multiple Linear RegressionRegression analysis:  Simple Linear Regression Multiple Linear Regression
Regression analysis: Simple Linear Regression Multiple Linear Regression
 

ORP Sensor Troubleshooting Guide

  • 1. 1 CAUTION: There are live circuits inside the controller even when the power switch is in the OFF position! Servicing and troubleshooting the controller must be performed by Qualified Service Personnel. WM1, WIND, W400, W100, W600, Intuition-6: ORP Sensor Troubleshooting Guide (Electrode and Housing) Table of Contents Section 1: Sensor (Electrode) Responds Slowly............................................................................ 2 Section 2: Sensor Reading is Stuck on one Value .......................................................................... 3 Section 3: Sensor Reads Low or High Versus a Calibrated Handheld Meter; Cannot Calibrate.... 4 Section 4: Sensor Error................................................................................................................... 5 Section 5: ORP Reading is Unstable............................................................................................... 7
  • 2. 2 CAUTION: There are live circuits inside the controller even when the power switch is in the OFF position! Servicing and troubleshooting the controller must be performed by Qualified Service Personnel. Section 1: Sensor (Electrode) Responds Slowly Step 1: The pH or ORP electrodes require periodic cleaning and calibration. These electrodes are like batteries and their voltage outputs will change with time even if they are not being used.  The method of cleaning the electrode will depend upon the coating, as well as the materials of construction of the electrode. Do not use a solvent that will attack the electrode  Oily coatings should be removed with a mild detergent or isopropyl alcohol.  Hard scales such as calcium carbonate can usually be removed with a dilute hydrochloric acid solution.  Soft coatings can be removed using a soft cloth or soft toothbrush.  A two point calibration should always be performed after cleaning the electrode. Step 2: If cleaning does not speed up the electrode's response  Replace the electrode.
  • 3. 3 CAUTION: There are live circuits inside the controller even when the power switch is in the OFF position! Servicing and troubleshooting the controller must be performed by Qualified Service Personnel. Section 2: Sensor Reading is Stuck on one Value Step 1:  If the sensor reading is stuck on one number  If it is stuck near 0 mV then there is most likely an electrical short.  If it is stuck at a high mV reading (~1200), then the pre-amp circuit is most likely open.  Broken glass on the electrode tip could cause an open circuit which sends the signal drifting to the end, or lower, or all the way to Sensor Error. Note: If water leaks into the preamp, you may get a mV reading, but the reading will not hold accuracy. Step 2: Check the sensor (electrode) for an electrical short.  Power down the controller.  Remove/unthread the electrode from the housing.  Power the controller up.  If reading drifts then the electrode has a short. Replace the electrode. If the reading is still stuck on one number, proceed to Step 3. Step 3: Check the preamp (housing) for an electrical short.  Power down the controller.  Disconnect the preamp from the controller.  Power the controller up.  If reading drifts then the preamp has a short. Replace the preamp. If the reading is stuck on one number, proceed to Section 4.
  • 4. 4 CAUTION: There are live circuits inside the controller even when the power switch is in the OFF position! Servicing and troubleshooting the controller must be performed by Qualified Service Personnel. Section 3: Sensor Reads Low or High Versus a Calibrated Handheld Meter; Cannot Calibrate Step 1: Perform a sensor calibration in accordance with the controller manual.  If the sensor cannot be calibrated, first thing to try is clean the sensor.  If the sensor cannot be calibrated after cleaning, proceed to Step 2. Step 2: Go to the Sensor mV menu. Resetting the calibration settings.  Go into the Sensor menu, go to Sensor Type, change it from ORP to pH, hit ENTER to save this change. Go back into the Sensor Type menu, change it from pH back to ORP, hit ENTER to save this change.  Try to calibrate the sensor in accordance with the controller manual. If you still cannot calibrate, proceed to Step 3. Step 3: Go to the Sensor mV menu. Resetting to factory defaults.  Proceed as follows depending on which controller model you have.  For a WM1 or WIND controller, go into the Sensor Input menu, and press the reset defaults button. Proceed to Section 4.  For a WPH400 controller reset the controller to factory defaults as follows.  Caution: Resetting defaults erases any user defined settings programmed into the controller.  Power down the controller, press and hold the ENTER and EXIT keys, power up the controller keeping the ENTER and EXIT keys pressed in until the controller boots up.  When the display comes up, you will be asked if you want to reset defaults, release the ENTER and EXIT keys and then press ENTER for yes. Re-check the mV and pH values. If the pH and mV values do not correlate, then either the preamp or the electrode is faulty. Proceed to Section 4.
  • 5. 5 CAUTION: There are live circuits inside the controller even when the power switch is in the OFF position! Servicing and troubleshooting the controller must be performed by Qualified Service Personnel. +5V T- T+ G R N / W H T W H I T E / G R N SHIELD -5V IN+ IN- O R N /W H T WH ITE /OR N B L U /W H T W H I T E / B L U WEL Electrode Diffamp Housing (102581, 102606, 102582,102607, 102758, 102759) pH/ORP electrode cartridge TO GRND Dashed lines indicate optional temperature compensation Section 4: Sensor Error Step 1: Perform a self-test for this sensor input.  If the self-test passes, then the problem is not inside the controller. The preamp or electrode is faulty. Continue to Step 2.  If the self-test fails, disconnect sensor from the controller and repeat the self-test and if it still fails, send the controller back for repair. Contact your distributor or Walchem Customer Service with the controller model number and serial number to initiate the Return Authorization process. Step 2: On the ORP sensor input terminals inside the controller, using a multimeter:  Measure the VDC across IN- and +5V, should read +5 VDC +/- 5%  Measure the VDC across IN- and –5 VDC, should read –4.6 VDC +/- 5%.  If these measurements are out of spec with and without the preamp connected, the controller is faulty. Contact your distributor or Walchem Customer Service with the controller model number and serial number to initiate the Return Authorization process.  If these measurements are in spec without the preamp connected, then replace the preamp. If these measurements are in spec with and without the preamp, then continue to Step 3.  Make sure the preamp is wired correctly into the controller, as shown.  Be cautious about junction box connections. The sensor cable shielding should only be grounded on one end (inside the controller) otherwise there will be a ground loop.
  • 6. 6 CAUTION: There are live circuits inside the controller even when the power switch is in the OFF position! Servicing and troubleshooting the controller must be performed by Qualified Service Personnel. Step 3: Check the sensor cable, sensor extension cable (if installed).  Make sure you are using shielded sensor cable, 3 twisted pair wire. Use Walchem p/n 102535 or Belden 9680.  Make sure the sensor cable is not run near, or in the same conduit as AC voltage wire (All wires carry voltage. We are concerned with adjacent wires that are carrying voltage above 24 VAC). If you suspect the sensor cable may be the problem, then take the existing sensor (with the 5 or 20 feet of cable), and wire it directly into the controller by-passing the sensor extension cable. If the sensor reading gets corrected, then the sensor extension cable, or where it was routed, was the problem.  If wiring the sensor cable directly into the controller does not solve the issue, then replace the preamp.
  • 7. 7 CAUTION: There are live circuits inside the controller even when the power switch is in the OFF position! Servicing and troubleshooting the controller must be performed by Qualified Service Personnel. Section 5: ORP Reading is Unstable Step 1: The ORP reading is unstable when the ORP sensor is placed in the system water.  Remove the ORP sensor from the system water and place it in a beaker of water. Does the reading stabilize? If yes, then proceed to Step 2. If no, proceed to Step 3. Step 2: The sensor is still in the beaker, and the reading is stable.  Obtain a length of wire (the ends of the wire must be striped) long enough to place one end in the water in the beaker, and the other end in the system water.  With this length of wire connecting the beaker water to the system water, does the reading stay stable?  If the reading does not stay stable, this can indicate a ground loop. A ground loop could cause readings to be stuck at a high or low value. Or there is some electrical interference in the system water, possibly from a mixer or recirculation pump(s). Step 3: The sensor is still in the beaker, and the reading is still unstable.  Power down the controller, disconnect the ORP sensor shield wire from the Earth ground stud inside the controller.  Does the reading stabilize?  If the reading stabilizes, then there is possibly some electrical noise coming in through the controller VAC mains.  If the reading does not stay stable, then connect the controller to an alternate supply voltage circuit. This may necessitate temporarily running an extension cord for this test. If the reading still does not stay stable, please contact Walchem for further instructions.