SlideShare a Scribd company logo
1 of 20
Implant Insertion and
Removal
Materials for Insertion
• Sterile surgical drapes
• Sterile gloves
• Antiseptic solution
• Surgical marker
• Local anesthetic
• Needles and syringe
• Sterile gauze
• Adhesive bandage
• Pressure bandage
The following equipment is needed for insertion:
Patient Insertion Position
View applicator from the side and not from
above the arm.
Non-dominant arm flexed, hand underneath
head.
This position deflects ulnar nerve away from
insertion site.
Insertion
 Step 1. Position and mark
Insertion
 Step 2. Clean and anesthetize (with epi recommended)
 Step 3. Open blister pack and remove cap - should see implant by looking into
tip of needle (DO NOT touch slider)
 Step 4. Stretch skin toward elbow
If the purple slider is
released prematurely,
restart the procedure
with a new applicator
Insertion
 Step 5. Viewing from the side puncture the skin – bevel in only
 Step 6. Lower the applicator, lift the skin, slide needle in fully
Readjust as needed
Insert bevel
on an angle
Lower angle
& lift skin
Slide in – all the way
Insertion
 Step 7. Without moving applicator, move the slider fully back until it stops.
 Step 8. Bandage, verify position and have pt. palpate
 Step 9. Apply small bandage then pressure bandage over it
Insertion
If applicator is moved during insertion or if purple
slider is not moved fully back, implant may
protrude from insertion site
If implant is protruding,
remove and perform new
procedure at same insertion
site using a new applicator
Do not push
protruding implant
back into incision
Post-insertion Steps
 May remove:
 Pressure bandage in 24 hours
 Small adhesive bandage over insertion site after 3 to 5 days
 Complete Patient Alert Card and give to woman to keep
 Complete adhesive label and affix to woman’s medical record (and/or record
batch number if electronic medical record)
If Implant is Not Palpable After Insertion
 May not have been inserted or inserted too deeply:
 Check applicator: Needle should be fully retracted and
only purple tip of the obturator should be visible
 Use other methods to confirm the presence of implant.
X-ray, CT scan, Ultrasound, MRI
Radiopaque Implant X-ray Localization
Mansour D et al. Contraception 2010;82:243–9. Adapted with permission.
If Imaging Methods Fail to Locate the
Implant
 Until presence of implant verified non-hormonal contraceptive method must
be used
 Deeply-placed implants removed ASAP to avoid distant migration
Serum
etonogestrel
assay can be
requested to
confirm
presence of
etonogestrel
subdermal
implant
Contact Merck
Canada for
further
guidance
Most Frequent Clinician-reported Events
at Insertion (N=7364)
*
*Number of insertions; †26 (60%) were reported by a single clinician
CI, confidence interval
Reed S et al. Contraception 2019;100:31-6.
13
Insertion event Number* % incidence 95% CI
Incorrect insertion 93 1.3 1.0–1.6
Deep insertion 65† 0.9 0.7–1.1
Partial insertion 27 0.4 0.2–0.5
Non-insertion 1 0.01 0.0–0.08
Insertion-related challenge
Difficulty removing protection
cap
93 1.3 1.0–1.5
Difficulty sliding needle to full
length in skin
30 0.4 0.3–0.6
Difficulty handling the
device/visualization
17 0.2 0.1–0.4
Real-world Evidence: Nexplanon Observational Risk Assessment (NORA) Study
• Sterile surgical drapes
• Sterile gloves
• Antiseptic solution
• Surgical marker
• Local anesthetic
• Needles and syringe
• Sterile scalpel
• Forceps (straight and curved
mosquito)
• Sterile adhesive wound closure
• Sterile gauze
• Pressure bandage
Removal
The following equipment is needed for the implant removal:
Removal
 Step 1. Locate the implant and push down, mark the distal end
 Step 2. Clean, anesthetize (inject under implant!)
 Step 3. Push down, make parallel incision
Removal
 Step 4. Implant should pop through incision, grasp the implant with forceps,
remove and confirm length
 Step 5. Dissect if needed
 Step 6. Insert new Nexplanon in same incision prn
 Step 7. Sterile adhesive wound closure and pressure bandage
After Removing the Implant
 Remove:
 Pressure bandage after 24 hours
 Sterile adhesive wound closure after 3 to 5 days
 Restart contraception immediately
Possible Factors Complicating Removal
Implant
not
inserted
correctly
Implant
inserted
too
deeply
Implant
not
palpable
Implant
encased
in fibrous
tissue
Implant
has
migrated
Removal of a Non-Palpable Implant
 Localization and removal of non-palpable implants is recommended ASAP
 Exploratory surgery is strongly discouraged
 Once localized, implant should be removed by HCP experienced in removing
deeply placed implants, consider ultrasound guidance
 If not found consider radiographic examination of chest
Insertion Video
Link to:
http://www.nexplanonvideos.com/home_lander_subs.html?country=EN|Nexplan
on

More Related Content

Similar to Implant Insertion and Removal Procedures

WOUND CARE for Public health professionals .ppt
WOUND CARE for Public health professionals .pptWOUND CARE for Public health professionals .ppt
WOUND CARE for Public health professionals .pptAme Mehadi
 
INTRADERMAL INJECTION
INTRADERMAL INJECTIONINTRADERMAL INJECTION
INTRADERMAL INJECTIONNeethuskaria
 
Subcutaneous injection.pptx
Subcutaneous injection.pptxSubcutaneous injection.pptx
Subcutaneous injection.pptxJenishia M
 
Lab 150308094954-conversion-gate01
Lab 150308094954-conversion-gate01Lab 150308094954-conversion-gate01
Lab 150308094954-conversion-gate01MESHALALKHAMMASH
 
Lab 150308094954-conversion-gate01 (1)
Lab 150308094954-conversion-gate01 (1)Lab 150308094954-conversion-gate01 (1)
Lab 150308094954-conversion-gate01 (1)MESHALALKHAMMASH
 
Suture care
Suture careSuture care
Suture carearunmtin
 
005 Suctioning.ppt
005 Suctioning.ppt005 Suctioning.ppt
005 Suctioning.pptsavitri49
 
Proc. 2 pharma. skills sensitivity test
Proc. 2 pharma. skills sensitivity testProc. 2 pharma. skills sensitivity test
Proc. 2 pharma. skills sensitivity testjhonee balmeo
 
Fundamental of Nursing 7. : Parenteral Medication Administration
Fundamental of Nursing 7. : Parenteral Medication AdministrationFundamental of Nursing 7. : Parenteral Medication Administration
Fundamental of Nursing 7. : Parenteral Medication AdministrationParya J. Ahmad
 
Subcutaneous Injections.pptx Nursing students
Subcutaneous Injections.pptx Nursing studentsSubcutaneous Injections.pptx Nursing students
Subcutaneous Injections.pptx Nursing studentsAnjuAnnMani1
 
Principles of Anesthesia & suture.pptx
Principles of Anesthesia & suture.pptxPrinciples of Anesthesia & suture.pptx
Principles of Anesthesia & suture.pptxGEDIONZERIHUN1
 
surgicalwounddressing-201022180123 (1).pdf
surgicalwounddressing-201022180123 (1).pdfsurgicalwounddressing-201022180123 (1).pdf
surgicalwounddressing-201022180123 (1).pdfBob Taguba
 
Young margot administrating intramuscular injections
Young margot administrating intramuscular injectionsYoung margot administrating intramuscular injections
Young margot administrating intramuscular injectionsmargot125
 
Intramuscular Injection (Im)
Intramuscular Injection (Im)Intramuscular Injection (Im)
Intramuscular Injection (Im)chuckiecalsado
 
Tracheostomy site care procedure
Tracheostomy site care procedureTracheostomy site care procedure
Tracheostomy site care procedurechrissie argana
 
Introduction to suturing
Introduction to suturingIntroduction to suturing
Introduction to suturing92763
 

Similar to Implant Insertion and Removal Procedures (20)

WOUND CARE for Public health professionals .ppt
WOUND CARE for Public health professionals .pptWOUND CARE for Public health professionals .ppt
WOUND CARE for Public health professionals .ppt
 
INTRADERMAL INJECTION
INTRADERMAL INJECTIONINTRADERMAL INJECTION
INTRADERMAL INJECTION
 
Fine Needle Aspiration cytology(FNAC)
Fine Needle Aspiration cytology(FNAC)Fine Needle Aspiration cytology(FNAC)
Fine Needle Aspiration cytology(FNAC)
 
Subcutaneous injection.pptx
Subcutaneous injection.pptxSubcutaneous injection.pptx
Subcutaneous injection.pptx
 
Lab 150308094954-conversion-gate01
Lab 150308094954-conversion-gate01Lab 150308094954-conversion-gate01
Lab 150308094954-conversion-gate01
 
Lab 150308094954-conversion-gate01 (1)
Lab 150308094954-conversion-gate01 (1)Lab 150308094954-conversion-gate01 (1)
Lab 150308094954-conversion-gate01 (1)
 
Suture care
Suture careSuture care
Suture care
 
005 Suctioning.ppt
005 Suctioning.ppt005 Suctioning.ppt
005 Suctioning.ppt
 
Proc. 2 pharma. skills sensitivity test
Proc. 2 pharma. skills sensitivity testProc. 2 pharma. skills sensitivity test
Proc. 2 pharma. skills sensitivity test
 
Fundamental of Nursing 7. : Parenteral Medication Administration
Fundamental of Nursing 7. : Parenteral Medication AdministrationFundamental of Nursing 7. : Parenteral Medication Administration
Fundamental of Nursing 7. : Parenteral Medication Administration
 
Subcutaneous Injections.pptx Nursing students
Subcutaneous Injections.pptx Nursing studentsSubcutaneous Injections.pptx Nursing students
Subcutaneous Injections.pptx Nursing students
 
Principles of Anesthesia & suture.pptx
Principles of Anesthesia & suture.pptxPrinciples of Anesthesia & suture.pptx
Principles of Anesthesia & suture.pptx
 
Surgical wound dressing
Surgical wound dressingSurgical wound dressing
Surgical wound dressing
 
surgicalwounddressing-201022180123 (1).pdf
surgicalwounddressing-201022180123 (1).pdfsurgicalwounddressing-201022180123 (1).pdf
surgicalwounddressing-201022180123 (1).pdf
 
Young margot administrating intramuscular injections
Young margot administrating intramuscular injectionsYoung margot administrating intramuscular injections
Young margot administrating intramuscular injections
 
Intramuscular Injection (Im)
Intramuscular Injection (Im)Intramuscular Injection (Im)
Intramuscular Injection (Im)
 
Tracheostomy site care procedure
Tracheostomy site care procedureTracheostomy site care procedure
Tracheostomy site care procedure
 
part ii.pptx
part ii.pptxpart ii.pptx
part ii.pptx
 
Introduction to suturing
Introduction to suturingIntroduction to suturing
Introduction to suturing
 
Session 4 a
Session 4 aSession 4 a
Session 4 a
 

Recently uploaded

Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...VICTOR MAESTRE RAMIREZ
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort servicejennyeacort
 
Introduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptxIntroduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptxvipinkmenon1
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSKurinjimalarL3
 
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...VICTOR MAESTRE RAMIREZ
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZTE
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVRajaP95
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝soniya singh
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineeringmalavadedarshan25
 
chaitra-1.pptx fake news detection using machine learning
chaitra-1.pptx  fake news detection using machine learningchaitra-1.pptx  fake news detection using machine learning
chaitra-1.pptx fake news detection using machine learningmisbanausheenparvam
 
microprocessor 8085 and its interfacing
microprocessor 8085  and its interfacingmicroprocessor 8085  and its interfacing
microprocessor 8085 and its interfacingjaychoudhary37
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girlsssuser7cb4ff
 
Heart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxHeart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxPoojaBan
 
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...srsj9000
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 

Recently uploaded (20)

Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
 
Introduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptxIntroduction to Microprocesso programming and interfacing.pptx
Introduction to Microprocesso programming and interfacing.pptx
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
 
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
VICTOR MAESTRE RAMIREZ - Planetary Defender on NASA's Double Asteroid Redirec...
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
 
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCRCall Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineering
 
chaitra-1.pptx fake news detection using machine learning
chaitra-1.pptx  fake news detection using machine learningchaitra-1.pptx  fake news detection using machine learning
chaitra-1.pptx fake news detection using machine learning
 
microprocessor 8085 and its interfacing
microprocessor 8085  and its interfacingmicroprocessor 8085  and its interfacing
microprocessor 8085 and its interfacing
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girls
 
Heart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptxHeart Disease Prediction using machine learning.pptx
Heart Disease Prediction using machine learning.pptx
 
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 

Implant Insertion and Removal Procedures

  • 2. Materials for Insertion • Sterile surgical drapes • Sterile gloves • Antiseptic solution • Surgical marker • Local anesthetic • Needles and syringe • Sterile gauze • Adhesive bandage • Pressure bandage The following equipment is needed for insertion:
  • 3. Patient Insertion Position View applicator from the side and not from above the arm. Non-dominant arm flexed, hand underneath head. This position deflects ulnar nerve away from insertion site.
  • 4. Insertion  Step 1. Position and mark
  • 5. Insertion  Step 2. Clean and anesthetize (with epi recommended)  Step 3. Open blister pack and remove cap - should see implant by looking into tip of needle (DO NOT touch slider)  Step 4. Stretch skin toward elbow If the purple slider is released prematurely, restart the procedure with a new applicator
  • 6. Insertion  Step 5. Viewing from the side puncture the skin – bevel in only  Step 6. Lower the applicator, lift the skin, slide needle in fully Readjust as needed Insert bevel on an angle Lower angle & lift skin Slide in – all the way
  • 7. Insertion  Step 7. Without moving applicator, move the slider fully back until it stops.  Step 8. Bandage, verify position and have pt. palpate  Step 9. Apply small bandage then pressure bandage over it
  • 8. Insertion If applicator is moved during insertion or if purple slider is not moved fully back, implant may protrude from insertion site If implant is protruding, remove and perform new procedure at same insertion site using a new applicator Do not push protruding implant back into incision
  • 9. Post-insertion Steps  May remove:  Pressure bandage in 24 hours  Small adhesive bandage over insertion site after 3 to 5 days  Complete Patient Alert Card and give to woman to keep  Complete adhesive label and affix to woman’s medical record (and/or record batch number if electronic medical record)
  • 10. If Implant is Not Palpable After Insertion  May not have been inserted or inserted too deeply:  Check applicator: Needle should be fully retracted and only purple tip of the obturator should be visible  Use other methods to confirm the presence of implant. X-ray, CT scan, Ultrasound, MRI
  • 11. Radiopaque Implant X-ray Localization Mansour D et al. Contraception 2010;82:243–9. Adapted with permission.
  • 12. If Imaging Methods Fail to Locate the Implant  Until presence of implant verified non-hormonal contraceptive method must be used  Deeply-placed implants removed ASAP to avoid distant migration Serum etonogestrel assay can be requested to confirm presence of etonogestrel subdermal implant Contact Merck Canada for further guidance
  • 13. Most Frequent Clinician-reported Events at Insertion (N=7364) * *Number of insertions; †26 (60%) were reported by a single clinician CI, confidence interval Reed S et al. Contraception 2019;100:31-6. 13 Insertion event Number* % incidence 95% CI Incorrect insertion 93 1.3 1.0–1.6 Deep insertion 65† 0.9 0.7–1.1 Partial insertion 27 0.4 0.2–0.5 Non-insertion 1 0.01 0.0–0.08 Insertion-related challenge Difficulty removing protection cap 93 1.3 1.0–1.5 Difficulty sliding needle to full length in skin 30 0.4 0.3–0.6 Difficulty handling the device/visualization 17 0.2 0.1–0.4 Real-world Evidence: Nexplanon Observational Risk Assessment (NORA) Study
  • 14. • Sterile surgical drapes • Sterile gloves • Antiseptic solution • Surgical marker • Local anesthetic • Needles and syringe • Sterile scalpel • Forceps (straight and curved mosquito) • Sterile adhesive wound closure • Sterile gauze • Pressure bandage Removal The following equipment is needed for the implant removal:
  • 15. Removal  Step 1. Locate the implant and push down, mark the distal end  Step 2. Clean, anesthetize (inject under implant!)  Step 3. Push down, make parallel incision
  • 16. Removal  Step 4. Implant should pop through incision, grasp the implant with forceps, remove and confirm length  Step 5. Dissect if needed  Step 6. Insert new Nexplanon in same incision prn  Step 7. Sterile adhesive wound closure and pressure bandage
  • 17. After Removing the Implant  Remove:  Pressure bandage after 24 hours  Sterile adhesive wound closure after 3 to 5 days  Restart contraception immediately
  • 18. Possible Factors Complicating Removal Implant not inserted correctly Implant inserted too deeply Implant not palpable Implant encased in fibrous tissue Implant has migrated
  • 19. Removal of a Non-Palpable Implant  Localization and removal of non-palpable implants is recommended ASAP  Exploratory surgery is strongly discouraged  Once localized, implant should be removed by HCP experienced in removing deeply placed implants, consider ultrasound guidance  If not found consider radiographic examination of chest

Editor's Notes

  1. Explain that etonogestrel subdermal implant should be inserted under aseptic conditions. Display slide and review the materials needed for insertion of etonogestrel subdermal implant. An examination table for the woman to lie on Sterile surgical drapes Sterile gloves Antiseptic solution Surgical marker Local anesthetic Needles and syringe Sterile gauze Adhesive bandage Pressure bandage
  2. Position Woman Prior to Insertion To help make sure the implant is inserted just under the skin, the healthcare providers should be positioned to see the advancement of the needle by viewing the applicator from the side and not from above the arm. From the side view, the insertion site and the movement of the needle just under the skin can be clearly visualized. Have the woman lie on her back on the examination table with her non-dominant arm flexed at the elbow and externally rotated so that her hand is underneath her head (or as close as possible). Please note that for illustrative purposes, Figure depicts the left inner arm.
  3. Display slide with images of insertion site: Emphasize: Before inserting the etonogestrel subdermal implant (etonogestrel implant) implant, identify the insertion site, which is at the inner side of the non-dominant upper arm. The insertion site is overlying the triceps muscle about 8-10 cm (3-4 inches) from the medial epicondyle of the humerus and 3-5 cm (1.25-2 inches) posterior to the sulcus (groove) between the biceps and triceps muscles. This location is intended to avoid the large blood vessels and nerves lying within and surrounding the sulcus. If it is not possible to insert the implant in this location (eg, in women with thin arms), it should be inserted as far posterior from the sulcus as possible. Have the woman lie on her back on the examination table with her non-dominant arm flexed at the elbow and externally rotated so that her hand is underneath her head (or as close as possible) Display slide on marking insertion site: Emphasize: Mark the insertion site with a surgical marker. Make two marks: First, mark the spot where the etonogestrel subdermal implant will be inserted Second, mark a spot at 5 centimeters (2 inches) proximal (toward the shoulder) to the first mark as a guiding mark After marking the arm, confirm the site is in the correct location on the inner side of the arm
  4. Mention backups that you can get from Merck rep– they will give 2 Clean the skin from the insertion site to the guiding mark with an antiseptic solution Anesthetize the insertion area (for example, with anesthetic spray or by injecting 2 mL of 1% lidocaine just under the skin along the planned insertion tunnel) Explain steps of insertion procedure: Remove the sterile preloaded disposable etonogestrel subdermal implant applicator carrying the implant from its blister The applicator should not be used if sterility is in question Hold the applicator just above the needle at the textured surface area and remove the transparent protection cap by sliding it horizontally in the direction of the arrow away from the needle If the cap does not come off easily, the applicator should not be used You should see the white colored implant by looking into the tip of the needle Emphasize: Do not touch the purple slider until you have fully inserted the needle subdermally, as doing so will retract the needle and prematurely release the implant from the applicator If the purple slider is released prematurely, restart the procedure with a new applicator Emphasize: To help make sure the implant is inserted just under the skin, you should position yourself to see the advancement of the needle by viewing the applicator from the side and not from above the arm. From the side view, you can clearly see the insertion site and the movement of the needle just under the skin Describe next steps: With your free hand, stretch the skin around the insertion site towards the elbow
  5. Emphasize that the needle should be inserted until the bevel (slanted opening of the tip) is just under the skin (and no further). If you insert the needle past the bevel, withdraw it until only the bevel is beneath the skin. Describe next steps: Lower the applicator to a nearly horizontal position. To facilitate subdermal placement, lift the skin with the needle while sliding the needle to its full length. You may feel slight resistance but do not exert excessive force. Emphasize that if the needle is not inserted to its full length, the implant will not be inserted properly. If the needle tip emerges from the skin before needle insertion is complete, the needle should be pulled back and be readjusted to subdermal position before completing the insertion procedure. Emphasize that the needle should be inserted until the bevel (slanted opening of the tip) is just under the skin (and no further). If you insert the needle past the bevel, withdraw it until only the bevel is beneath the skin. Describe next steps: Lower the applicator to a nearly horizontal position. To facilitate subdermal placement, lift the skin with the needle while sliding the needle to its full length. You may feel slight resistance but do not exert excessive force. Emphasize that if the needle is not inserted to its full length, the implant will not be inserted properly. If the needle tip emerges from the skin before needle insertion is complete, the needle should be pulled back and be readjusted to subdermal position before completing the insertion procedure.
  6. Display slide which describes the final steps of the insertion process. Emphasize: Keep the applicator in the same position with the needle inserted to its full length. If needed, you may use your free hand to stabilize the applicator. Describe next steps: Unlock the purple slider by pushing it slightly down Move the slider fully back until it stops Emphasize: Do not move the applicator while moving the purple slider The implant is now in its final subdermal position, and the needle is locked inside the body of the applicator The applicator can now be removed Verify the Presence of the Implant Apply a small adhesive bandage over the insertion site. Apply sterile gauze with a pressure bandage to minimize bruising Emphasize: Always verify the presence of the implant in the woman’s arm immediately after insertion by palpation. By palpating both ends of the implant, you should be able to confirm the presence of the 4-cm rod.
  7. Emphasize that if the applicator is not kept in the same position during this procedure or if the purple slider is not moved fully back until it stops, the implant will not be inserted properly and may protrude from the insertion site If the implant is protruding from the insertion site, remove the implant and perform a new procedure at the same insertion site using a new applicator Do not push the protruding implant back into the incision
  8. Review each of the post-insertion steps as described on slide. Tell participants to refer to the Prescribing Information if they have any questions about counseling the woman regarding follow-up care. You may wish to remind participants to counsel women to leave the pressure bandage in place for the recommended 24-hour period as its primary purpose is to minimize bruising at the insertion site.
  9. Display slide Emphasize that if you cannot feel the implant or are in doubt of its presence, the implant may not have been inserted or it may have been inserted deeply Review next steps: Check the applicator. The needle should be fully retracted and only the purple tip of the obturator should be visible Use other methods to confirm the presence of the implant Review suitable methods for localization
  10. Display slide and tell participants that it shows examples of implants localized via the use of X-ray. Make sure participants are able to identify the location of the implant on each of the 3 examples. Make the following points: The implant in the examples is radiopaque, that’s why it is visible with X-ray. The way the implant looks on the X-ray may vary a bit depending on the angle of the X-ray. In the third example, the implant is visible even though the angle of projection makes it appear as though it is projected on the bone.
  11. Review the information regarding the etonogestrel (ENG) assay technique by displaying and discussing slide. Emphasize the purpose and use of an ENG assay is to confirm the presence of the implant but not its location. Remind participants that they should contact their local Merck office in any situation in which an ENG assay may be required. They will assist you with detailed prescription of the procedure.
  12. In the NORA study, 85 HCPs reported a total of 208 insertion-related events involving 189 patients during 7364 insertions, including 93 incorrect insertions. One HCP reported 26 of these (all deep insertions). There were 65 deep insertions, defined as non-palpable implants following insertion, implants within muscle or adjacent to the deep fascia overlying the muscle, and injury to a nerve or blood vessel during insertion. Deep insertions did not result in serious injuries. Over 60% were due to the single HCP with 26 incorrect insertions. The non-insertion was initially unrecognized. There was no difference in incorrect insertion rates based on HCP prior experience (<5 insertions vs ≥5 insertions) or physician vs non-physician. The most common insertion-related challenge was difficulty in removing the protection cap. Reference Reed S, Do Minh T, Lange JA, Koro C, Fox M, Heinemann K. Real world data on Nexplanon® procedure-related events: final results from the Nexplanon Observational Risk Assessment study (NORA). Contraception 2019;100:31-36.
  13. Display slide and review the materials needed for removal of etonogestrel subdermal implant An examination table for the woman to lie on Sterile surgical drapes Sterile gloves Antiseptic solution Surgical marker Local anesthetic Needles and syringe Sterile scalpel Forceps (straight and curved mosquito) Skin closure Sterile gauze Pressure bandage
  14. Review:. Have the woman lie on her back on the table. The arm should be positioned with the elbow flexed and the hand underneath the head (or as close as possible). Locate the implant by palpation. Push down the end of the implant closest to the shoulder to stabilize it; a bulge should appear indicating the tip of the implant that is closest to the elbow. Emphasize: If the tip does not pop up, removal of the implant may be more challenging and should be performed by providers experienced with removing deeper implants. Contact the local Merck office for further information. Exact location of implant in the arm should be verified by palpation If not use other methods to localize and consider referral if deep or migrated Mark the distal end (end closest to the elbow), for example, with a surgical marker. Review: Clean the site with an antiseptic solution. Anesthetize the site, for example, with 0.5 to 1 mL 1% lidocaine, where the incision will be made (near the tip of etonogestrel subdermal implant that is closest to the elbow). Emphasize: Be sure to inject the local anesthetic under the implant to keep the implant close to the skin surface. Injection of local anesthetic over the implant may make removal more difficult. Review: Push down the end of the implant closest to the shoulder to stabilize it throughout the procedure. Starting over the tip of the implant closest to the elbow, make a longitudinal (parallel to the implant) incision of 2 mm towards the elbow. Take care not to cut the tip of the implant. NOTE: If desired, you may wish to ask participants to follow along by using the dummy arm they were given earlier and using the process described to identify the distal end of an implant they inserted earlier. Make sure participants understand where to make the longitudinal incision and approximately how deep the incision should be.
  15. Be sure to tell participants what to do if they encounter adherent tissue that is interfering with visualization or removal of the implant. The tip of the implant should pop out of the incision. If it does not, gently push the implant towards the incision until the tip is visible. If needed, gently remove adherent tissue from the tip of the implant using blunt dissection. If the implant tip is not exposed following blunt dissection, make an incision into the tissue sheath and then remove the implant with the forceps If the tip of the implant does not become visible in the incision, insert forceps (preferably curved mosquito forceps, with the tips pointed up) superficially into the incision Gently grasp the implant and then flip the forceps over into your other hand With a second pair of forceps carefully dissect the tissue around the implant and grasp the implant The implant can then be removed. If the implant cannot be grasped, stop the procedure and refer the woman to a healthcare provider experienced with complex removals or call 1-877-888-4231. Figure 18 Confirm that the entire implant, which is 4 cm long, has been removed by measuring its length. There have been reports of broken implants while in the patient’s arm. In some cases, difficult removal of the broken implant has been reported. If a partial implant (less than 4 cm) is removed, the remaining piece should be removed by following the instructions in section 2.3. If the woman would like to continue using NEXPLANON, a new implant may be inserted immediately after the old implant is removed using the same incision as long as the site is in the correct location Apply adhesive then pressure bandage with sterile gauze to minimize bruising.
  16. Display slide and review final post-removal steps. Emphasize: Bandaging and timing of removal The woman should restart contraception immediately after removal of the implant if continued contraceptive protection is desired.
  17. Display slide and introduce section on topic of Factors Complicating Removal Review slide
  18. Display slide on Removal of Non-Palpable implants Emphasize that exact location of the implant must be determined prior to a removal attempt; exploratory surgery is strongly discouraged. Instruct that once localized, the implant should be removed by a healthcare provider experienced in removing deeply placed implants and familiar with the anatomy of the arm. U/S guidance should be considered.
  19. Use insertion video prn