MRI ARTIFACTS
Introduction
• An Artifact is something observed in a scientific
investigation that is not naturally present but occurs as
a result of the investigative procedure.
• It is a structure not normally present, but visible as a
result of malfunction in the hardware or software of the
device, or a consequence of environmental influences
Types of MRI image
artifacts
MR hardware and room shielding
• Zipper artifact
• Herringbone artifact
• Zebrastripes
• Moire fringes
 MR software
 Slice-overlap artifact (cross talkartifact)
 Cross excitation
 Patient and physiologic motion
 -Phase-encoded motion artifact
 -Entry slice phenomenon
 Tissue heterogeneity and foreign bodies
 -Magic angleeffect
 -Magnetic susceptibility artifact
 -Chemical shift artifact
Fourier transform
-Gibbs artifact /truncation artifact
--Aliasing/wrap around artifact
• Cause: This type of artifact results from the
contamination of the MR scanner environment
(shielded MR suite) by spurious Radio Frequency
signals.
Zipper
artifact
 A Faraday cage is a
conductive enclosure
used to shield a
space from
electromagnetic
interference.
 In MR imaging, this
type of structure
provides
radiofrequency
shielding to the
scanning room to
minimize occurrence
of imaging artifacts.
• The contamination can result dueto
 1)BreechintheRadiofrequencycagei.e. RF entering the shielded
room from outside
 2) Illumination devices & other medical equipment (e.g pulse
oximeter )
 Appearance- A dense broken line of alternatively dark
and bright pixels running across the image perpendicular
to the frequency encoding direction.
 Solution:
 1) Scan door should be closed
 2) Breech in RF cage should be located and repaired
 3) Equipments in the scan room should be serviced
Herringbone
artifact
• Herringbone artifact (also known as ascrisscross
artifact or corduroyartifact)
• Due to interferences which occur while filing‐in of
the k‐space
• These interferences are usually RF spikes which
lead to points in k‐space with high intensity
• Causes ;
• Fluctuating power supply
• RF pulse discripencies
Appearance- It appears as obliquely oriented stripes seen throughout the image. Artifact is
scattered all over the image in a single slice or multiple slices .
Moire
fringes
• Moire fringes are an interference pattern most commonly seen when
doing gradient echo images with the body coil.
• Because of lack of perfect homogeneity of the main magnetic field
from one side of the body to the other
, aliasing of one side of the
body to the other results in superimposition of signals of different
phases that alternatively add and cancel.
• This causes the banding appearance and is similar to the effect of
looking though two screenwindows.
Wavy
appearance
at the edge
of the
images
 Solutions:
 1) Use of spin echo sequences
 2) Use of anti Aliasing remedies
 3) Plug on the receiver coil should be firmly put in the
socket
Slice-overlap
artifact
• Loss of signal in an image due to a multi‐slice, multi‐angle acquisition
or imperfect sliceprofile
• The mechanism behind this artifact is spinsaturation
Image depicts Two sets of spine images, at different angles, covering L4‐5 and L5‐S1 respectively. The
set acquired later(red) will have image spins that have already been excited and that are
saturated,leading to signal loss
If slices at different angles cross, then spins that have been previously
excited (saturated) can not be excited again, thus leading to a signal
drop in the crossing region
Solution
Cross excitation
artifact
• There is loss of signal within a slice due to pre-excitation from RF
pulse meant for an adjacent slice
• Appearance- ThisArtefact causes a reduction in SNR in adjacent
slices in a Slice stack.
.
 Cause : Ideally the profile of a slice
should be square or rectangular
when viewed from the edge, but in
practice a radio frequency excitation
pulse is not able to achieve this.
 The adjacent slices receive energy
from the RF excitation pulse of their
neighbors .
 When the tissues in next slice receive
their own RF pulse they are unable to
produce a signal as they are already
saturated
Solution:-
• Cross excitation can be reduced by insuring that
there is at least at 30% gap between the slices
(Gap= 30% of the Slice thickness )
This reduces the likelihood of RF pulse exciting
adjacent slice.
• Interleaving between slices i.e.
FIrst acquiring the odd number slices(1,3,5,7..)
Then acquiring the even number slices( 2,4,6,8..)
Effect of cross-talk on image contrast. On left is
a SE image with 50% gap showing expected spin-
density contrast. On right the same sequence with 0%
gap demonstrating impairedcontrast.
On the coronal scout, the lumen of the gallbladder is sharply divided into low signal
medially and high signal laterally with a "level" appearance suggesting layering
sludge. However, the axial T2W shows no sludge in the gallbladder. This finding is
attributed to cross-excitation artifact.
Phase-encoded motion
artifact
• Phase-encoded motion artifact occurs as a result
of tissue/fluid moving during the scan. It manifests
as ghosting in the direction of phase-encoding.
• These artifacts may be seen from arterial pulsations,
swallowing, breathing, peristalsis, and physical
movement of a patient. When projected over
anatomy it can mimic pathology, and needs to be
recognized.
• Motion that is random such as the patient moving
produces a smear in the phase direction. Periodic
motion, such as respiratory or cardiac/vascular
pulsation, produces discrete, well-defined ghosts.
This is a typical example of
phase-encoded artifact secondary
to aortic pulsation. The ghost
images of the aorta are:
discrete, well defined, lining with
the aorta along the short axis of
the abdomen (AP axis) at regular
intervals and extending beyond
the field of view.
These features help recognize
them and not to confuse them
with any pathology.
Blood flow in the sagittal sinus and carotids causes Periodic ghosting which gives the
impression of an enhancing lesion in the occipital lobe as well as anterior to the right
temporal horn. Wide windowing demonstrates the ghosting extending beyond the confines
of the patient's anatomy.
Apparent enhancement in the
occipital lobe and right temporal
lobe in a patient with
malignancy.
windowed slice demonstrates
the phase mismapping artefact from
the superior sagittal sinus.
 Swallowing artifact has
produced horizontal
band of noise along
phase-encoded
direction. Where it
passes over vertebral
body and spinal cord,
there are local intensity
and contrast changes,
rendering these portions
of scan suspect.
Solutions
 cardiac/respiratory gating
 spatial presaturation bands placed over moving tissues (e.g.
over the anterior neck in sagittal cervical spines)
 spatial presaturation bands placed outside the FOV, especially
before the entry or after the exit slice for reducing ghosting
from vascular flow
 scanning prone to reduce abdominal excursion
 increasing the number of signal averages
 shorten the scan time when motion is from patient movement
Cardiac Gating
 This involves the acquistion of a series of MR images of a
single anatomic section during a specific point in the
cardiac cycle ,which is being monitored with ECG.
 Data within the k-space is linked with specific time point
in cardiac cycle.
 The acquired datasets are then sorted according to time
stamp to produce sequential images allowing a dynamic
evaluation of myocardial function.
Respiratory gating
 respiration related motion is monitored, using bellows or
breathing belt, and image acquisition is timed to take
place at end expiration, when there is little or no motion.
however this is infrequently used in clinical imaging.
-Breath holding method most often used
-Sequences with short acquisition time may be used, or if
breath holding possible for limited time, exam may be
divided into several brief acquisitions.
 Saturation pulses
Saturation pulses involve the application of RF energy to
suppress the MR signal from moving tissues outside the
imaged volume to reduce or eliminate motion artifacts.
Saturation pulse
-Most commonly used in abdominal imaging at the interface of lung and diaphragm
-Application of small, one dimensional spatial encoding gradient in a plane perpendicular to
diaphragm.
- Echo measured at this location allows correction of imaging dataset and ensures that only the
imaging data acquired when diaphragm is at its peak (end expiration), is used in image
reconstruction.
Navigator Echo
Entry slice
phenomenon
Entry slice phenomenon occurs when unsaturated spins in
blood first enter into a slice .
Unsaturated spins = protons without transverse magnetisation
When a large number of such protons encounter a RF pulse for
the first time(i.e. at the entry slice),they undergo transverse
magnetisation and emit a strong signal.
- It is characterized by the bright signal in a blood vessel
(artery or vein) at the first slice that the vessel enters.
-
 This artefact can be confused with thrombosis
Solutions-
 -The characteristic location and if necessary, the use of
gradient echo flow techniques can be used to
differentiate entry slice artefacts from occlusions.
 -Spatial saturation bands place before the first slice and
after the last can also be used to eliminate this artefact.
Magic angle
artefact
• The magic angle effect is important in the clinical MR
imaging of certain tissues that are highly structured
and are oriented obliquely to the main magnetic field
(especially tendons, cartilage, and peripheral nerves).
• In tendons, the water molecules are aligned along the tendon sheath
and have strong dipolar interactions. Due to these dipolar
interactions, protons in the collagen rapidly dissipate their energy
after encountering a RF pulse i.e. tendons have a short intrinsic
T2 value (1-10 ms).
The signal of tendons is uniformly low on all conventional MR
sequences due to this rapid T2 decay.
 however, when the tendons are aligned at an angle of 55degree to the
main magnetic field , these dipolar interactions go to zero, leading to
an increase in T2 relaxation time and consequent sudden increase in
signal.
Common sites:
 proximal part of the Posterior cruciate ligament
 infrapatellar tendon at the tibial insertion
 peroneal tendons as they hook around the lateral
malleolus
 Supraspinatus tendon
 High signal within a tendon can mimic tendinopathy so it
is important that radiologists are aware of the magic
angle phenomena.
 Look for it in the outer portions of the rotator cuff,
supraspinatus tendon, the distil patellar tendon, tendons
of the ankle, extensor and flexor policis tendons of the
wrist or any part made of collagen that lies in a curve
Susceptibility
artifact
• Due to the presence of substances ( e.g.ferromagnetic materials) that
causes variation in the local magneticfield
• Results in increased magnetic field inhomogeneity ,leading to spin
dephasing and consequent signal drop
• Becomes worse for long echo times and GRE sequences( no RF 180
pulse)
• Less visible in fast spin echosequence
• Worse for higher fieldstrength
Susceptibility artefact due
to hip prosthesis
Susceptibility artefact
due to dental fillings
Susceptibility artefact
due to dental fillings
Remedies
• Use spin-echo sequences
A spin echo sequence uses a 180-degree RF
pulse .
The main purpose of this 180-degree pulse is
to correct local field inhomogeneities.
A gradient echo pulse sequence does not
utilize such a “refocusing” pulse.
As such, local field inhomogeneities are
much more readily apparent when gradient
echo sequences are utilized
Chemical Shift
-Chemical Shift is the difference between precessional
frequencies of water and fat.(3.5ppm)
-Artefact is the displacement of signal intensity due to this
difference
AT 1.5 T W>F by 220Hz
At 3T W>F by 440Hz
WATER FAT
 Water molecules have low
electron shielding and their
nucleus is more susceptible to
external magnetic field.
 As a result water molecules
precess at a higher frequency
at the same magnetic field.
 Fat molecules have higher
electron shielding and their
nucleus is comparitively less
susceptible to external
magnetic field.
 Fat molecules precess at a
lower frequency at the same
mag field
 water and lipid
protons coexist at
the interface, the
signal emitted by
the lipid protons will
have a lower
frequency than that
of the water
protons,leading to
mismapping of fat.
CHEMICAL SHIFT TYPE 1: MISMAPPING ARTIFACT
-Most commonly seen around water containg structures
sorrounded bt Fat(Liver,Kidneys,Orbits)
-Artifact appears as a dark band on one side of the
water/fat interface and a bright band on the other side
 a black band is seen on the left side of each kidney and a
white band on the right.
 T2 weighted image demonstrates
chemical shift artefact around the
lipoma.
 Note the signal loss anteriorly
(black arrows) and hyperintensity
posteriorly (white arrows).
 This is the result of inaccurate
spatial encoding, resulting in the
lipoma being incorrectly placed
posterior to its actual location.
LIPOMA
 axial T1 image also
demonstrates a chemical
shift artifact at the edges
between the mass and
the orbital fat.
ORBITAL HEMANGIOMA
 Chemical Shift: 2nd Kind
The chemical shift artifact of the second kind is sometimes
known as the India Ink or black line artifact because it
looks like organs and muscle bundles have been outlined
with a black pen. The line does not correspond to a true
anatomic structure, but results from destructive
interference of signals at the boundary pixels that contain
both water and fat. For this reason it is sometimes often
called the phase cancellation artifact. Examples in the
spine, thigh, and abdomen.
spine
Abdomen
Thigh
Gibbs artifact /
truncationartifact
• It refers to a series of bright or dark lines, parallel & next to
borders of abrupt intensity change
Areas of abrupt intensity change such as
-the CSF-Spinal Cord Interface.
-the Skull-Brain Interface.
 Bright and dark bands are
seen parallel and adjacent
to the outer convexity of
calvaria
Skull-Brain Interface
. Figure illustrates a syrinx-like artifact
(red arrow) within the center of the cord, a
classic truncation artifact that appears as a
result of alternating dark and light signal
bands overlying the spinal cord.
This artifact can be easily mistaken for
hydromyelia (dilatation of the central canal,
more commonly referred to as a “syrinx”),
Truncation artifacts are reduced by
the use of a smaller field of view
(FOV) or a larger matrix size.
Aliasing / wrap
aroundartifact
• It occurs when the field of view (FOV) is smaller than
the body part being imaged. The part of the body that
lies beyond the edge of the FOV is projected onto the
other side of the image
• Caused by under‐sampling in the phase encoding direction
-phase shifts of between 0° and 360°
encompass the field-of-view.
The subject's left flank in the drawing
extends outside the field-of-view and
encompasses phase shifts from 361°
to 450°.
Since all meaningful frequencies
have been defined over the range of
0° to 360°, a phase shift of 361° will
be assigned to the spatial position of
1°, and a shift of 450° will be
assigned to 450°-360° = 90°.
Hence ,the left side of the patient's
body will therefore be "wrapped
around" and spatially mismapped to
the opposite (right) side of the image.
Solution
-Enlarging the field of view (FOV)
-Using pre-saturation bands on areas outside the FOV
-Anti-aliasing software
-Use a surface coil to reduce the signal outside of the
area of interest
 PARTIAL VOLUME ARTIFACT
-Partial volume artefact occurs if slice thickness >
thickness of the tissue of interest.
-Occurs when multiple tissue types are encompassed
within a single voxel
-If a small structure is entirely contained within the slice
thickness along with other tissue of differing signal
intensities, the resulting signal displayed on the image is a
combination of these two intensities.
This reduces contrast of the small structure.
Slice thickness 10mm
Unable to depict finer
structures due to partial
volume artefact
Slice thickness 3mm
Depicts cranial nerves
THANKS

More Related Content

What's hot

Ct quality control
Ct quality controlCt quality control
Ct quality controlSAJIDEJAZ1
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopySelf
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifactsshajitha khan
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexVivek Elangovan
 
Information regarding e lora for AERB registration
Information regarding e lora for AERB registrationInformation regarding e lora for AERB registration
Information regarding e lora for AERB registrationDr Jitu Lal Meena
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdfmitians
 
CT Image Reconstruction- Avinesh Shrestha
CT Image Reconstruction- Avinesh ShresthaCT Image Reconstruction- Avinesh Shrestha
CT Image Reconstruction- Avinesh ShresthaAvinesh Shrestha
 
6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptxRohit Bansal
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its careSaruGosain
 
4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.pptRohit Bansal
 
General Consideration of all Radiology (imaging) Modalities
General Consideration of all Radiology (imaging) ModalitiesGeneral Consideration of all Radiology (imaging) Modalities
General Consideration of all Radiology (imaging) ModalitiesMuhammad Arif Afridi
 
CR, DR and recent advances
CR, DR and recent advancesCR, DR and recent advances
CR, DR and recent advancesVishal Sankpal
 
Radiological anatomy of the Carotid arteries
Radiological anatomy of the Carotid arteriesRadiological anatomy of the Carotid arteries
Radiological anatomy of the Carotid arteriesMohamed M.A. Zaitoun
 
Image Reconstruction in Computed Tomography
Image Reconstruction in Computed TomographyImage Reconstruction in Computed Tomography
Image Reconstruction in Computed TomographyAnjan Dangal
 

What's hot (20)

MRI artifacts
MRI artifactsMRI artifacts
MRI artifacts
 
Ct quality control
Ct quality controlCt quality control
Ct quality control
 
MR Elastography
MR ElastographyMR Elastography
MR Elastography
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopy
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifacts
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Information regarding e lora for AERB registration
Information regarding e lora for AERB registrationInformation regarding e lora for AERB registration
Information regarding e lora for AERB registration
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdf
 
MRI SAFETY.pptx
MRI SAFETY.pptxMRI SAFETY.pptx
MRI SAFETY.pptx
 
CT Image Reconstruction- Avinesh Shrestha
CT Image Reconstruction- Avinesh ShresthaCT Image Reconstruction- Avinesh Shrestha
CT Image Reconstruction- Avinesh Shrestha
 
6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx
 
Mri safety
Mri safetyMri safety
Mri safety
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
 
CT Coronary angiography
CT Coronary angiography CT Coronary angiography
CT Coronary angiography
 
240 spatial encoding
240 spatial encoding240 spatial encoding
240 spatial encoding
 
4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt
 
General Consideration of all Radiology (imaging) Modalities
General Consideration of all Radiology (imaging) ModalitiesGeneral Consideration of all Radiology (imaging) Modalities
General Consideration of all Radiology (imaging) Modalities
 
CR, DR and recent advances
CR, DR and recent advancesCR, DR and recent advances
CR, DR and recent advances
 
Radiological anatomy of the Carotid arteries
Radiological anatomy of the Carotid arteriesRadiological anatomy of the Carotid arteries
Radiological anatomy of the Carotid arteries
 
Image Reconstruction in Computed Tomography
Image Reconstruction in Computed TomographyImage Reconstruction in Computed Tomography
Image Reconstruction in Computed Tomography
 

Similar to artifactsusgctmriradiologypk-190830131406.pptx

ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxKeerthan39
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxAnayaAnaya14
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundArjun Reddy
 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet Ranjan
 
MRI Artifact Radiology
MRI Artifact RadiologyMRI Artifact Radiology
MRI Artifact RadiologyHenock Negasi
 
Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer Ganesan Yogananthem
 
Sonography
SonographySonography
SonographyFLI
 
BASICS OF EUS PHYSICS.pptx
BASICS OF EUS PHYSICS.pptxBASICS OF EUS PHYSICS.pptx
BASICS OF EUS PHYSICS.pptxPriyaranjanDas22
 
Ophthalmic ultrasonography
Ophthalmic ultrasonographyOphthalmic ultrasonography
Ophthalmic ultrasonographyRiyad Banayot
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkDr pradeep Kumar
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound ArtifactAmer Sohail
 

Similar to artifactsusgctmriradiologypk-190830131406.pptx (20)

MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
Artifacts
ArtifactsArtifacts
Artifacts
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptx
 
mri artefacts.pptx
mri artefacts.pptxmri artefacts.pptx
mri artefacts.pptx
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasound
 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to it
 
MRI Artifact Radiology
MRI Artifact RadiologyMRI Artifact Radiology
MRI Artifact Radiology
 
ULTRA.pptx
ULTRA.pptxULTRA.pptx
ULTRA.pptx
 
Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer Ultrasound assignment 3 mark question and answer
Ultrasound assignment 3 mark question and answer
 
Physics
PhysicsPhysics
Physics
 
MRI Artifacts
MRI ArtifactsMRI Artifacts
MRI Artifacts
 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
 
Sonography
SonographySonography
Sonography
 
Clinical electro physio assessment
Clinical electro physio assessmentClinical electro physio assessment
Clinical electro physio assessment
 
BASICS OF EUS PHYSICS.pptx
BASICS OF EUS PHYSICS.pptxBASICS OF EUS PHYSICS.pptx
BASICS OF EUS PHYSICS.pptx
 
Ophthalmic ultrasonography
Ophthalmic ultrasonographyOphthalmic ultrasonography
Ophthalmic ultrasonography
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pk
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound Artifact
 

More from Shubham661884

PHEOCHROMOCYTOMA.pptx
PHEOCHROMOCYTOMA.pptxPHEOCHROMOCYTOMA.pptx
PHEOCHROMOCYTOMA.pptxShubham661884
 
Contrast imaging phases and contrast media used in CT.pptx
Contrast imaging phases and contrast media used in CT.pptxContrast imaging phases and contrast media used in CT.pptx
Contrast imaging phases and contrast media used in CT.pptxShubham661884
 
cystic lung diseases (1).pptx
cystic lung diseases (1).pptxcystic lung diseases (1).pptx
cystic lung diseases (1).pptxShubham661884
 
Prune belly syndrome.pptx
Prune belly syndrome.pptxPrune belly syndrome.pptx
Prune belly syndrome.pptxShubham661884
 
Tree-In-Bud Pattern.pptx
Tree-In-Bud Pattern.pptxTree-In-Bud Pattern.pptx
Tree-In-Bud Pattern.pptxShubham661884
 
pulmonary sarcoidosis
pulmonary sarcoidosispulmonary sarcoidosis
pulmonary sarcoidosisShubham661884
 
CTMR OF HEAD & NECK.pptx
CTMR OF HEAD & NECK.pptxCTMR OF HEAD & NECK.pptx
CTMR OF HEAD & NECK.pptxShubham661884
 
PRESENTATION BOWEL.pptx
PRESENTATION BOWEL.pptxPRESENTATION BOWEL.pptx
PRESENTATION BOWEL.pptxShubham661884
 
sbo-171027200226.pptx
sbo-171027200226.pptxsbo-171027200226.pptx
sbo-171027200226.pptxShubham661884
 
Radiology day final.pptx
Radiology day final.pptxRadiology day final.pptx
Radiology day final.pptxShubham661884
 
oral hygiene & eye care.pptx
oral hygiene & eye care.pptxoral hygiene & eye care.pptx
oral hygiene & eye care.pptxShubham661884
 
radiologyofdemyelinatingdiseasesmbs-170105191815.pptx
radiologyofdemyelinatingdiseasesmbs-170105191815.pptxradiologyofdemyelinatingdiseasesmbs-170105191815.pptx
radiologyofdemyelinatingdiseasesmbs-170105191815.pptxShubham661884
 
Presentation MSpptx.pptx
Presentation MSpptx.pptxPresentation MSpptx.pptx
Presentation MSpptx.pptxShubham661884
 
Hysterosalpingography.pptx
Hysterosalpingography.pptxHysterosalpingography.pptx
Hysterosalpingography.pptxShubham661884
 

More from Shubham661884 (20)

coronary class.pptx
coronary class.pptxcoronary class.pptx
coronary class.pptx
 
SARCOIDOSIS.pptx
SARCOIDOSIS.pptxSARCOIDOSIS.pptx
SARCOIDOSIS.pptx
 
PHEOCHROMOCYTOMA.pptx
PHEOCHROMOCYTOMA.pptxPHEOCHROMOCYTOMA.pptx
PHEOCHROMOCYTOMA.pptx
 
Contrast imaging phases and contrast media used in CT.pptx
Contrast imaging phases and contrast media used in CT.pptxContrast imaging phases and contrast media used in CT.pptx
Contrast imaging phases and contrast media used in CT.pptx
 
cystic lung diseases (1).pptx
cystic lung diseases (1).pptxcystic lung diseases (1).pptx
cystic lung diseases (1).pptx
 
Prune belly syndrome.pptx
Prune belly syndrome.pptxPrune belly syndrome.pptx
Prune belly syndrome.pptx
 
Tree-In-Bud Pattern.pptx
Tree-In-Bud Pattern.pptxTree-In-Bud Pattern.pptx
Tree-In-Bud Pattern.pptx
 
pulmonary sarcoidosis
pulmonary sarcoidosispulmonary sarcoidosis
pulmonary sarcoidosis
 
CTMR OF HEAD & NECK.pptx
CTMR OF HEAD & NECK.pptxCTMR OF HEAD & NECK.pptx
CTMR OF HEAD & NECK.pptx
 
Pancreatitis F.pptx
Pancreatitis F.pptxPancreatitis F.pptx
Pancreatitis F.pptx
 
PRESENTATION BOWEL.pptx
PRESENTATION BOWEL.pptxPRESENTATION BOWEL.pptx
PRESENTATION BOWEL.pptx
 
ABDOMINAL TB.pptx
ABDOMINAL TB.pptxABDOMINAL TB.pptx
ABDOMINAL TB.pptx
 
sbo-171027200226.pptx
sbo-171027200226.pptxsbo-171027200226.pptx
sbo-171027200226.pptx
 
Radiology day final.pptx
Radiology day final.pptxRadiology day final.pptx
Radiology day final.pptx
 
oral hygiene & eye care.pptx
oral hygiene & eye care.pptxoral hygiene & eye care.pptx
oral hygiene & eye care.pptx
 
radiologyofdemyelinatingdiseasesmbs-170105191815.pptx
radiologyofdemyelinatingdiseasesmbs-170105191815.pptxradiologyofdemyelinatingdiseasesmbs-170105191815.pptx
radiologyofdemyelinatingdiseasesmbs-170105191815.pptx
 
Presentation MSpptx.pptx
Presentation MSpptx.pptxPresentation MSpptx.pptx
Presentation MSpptx.pptx
 
neurocut S (1).pptx
neurocut S (1).pptxneurocut S (1).pptx
neurocut S (1).pptx
 
Hysterosalpingography.pptx
Hysterosalpingography.pptxHysterosalpingography.pptx
Hysterosalpingography.pptx
 
ppt_ch02.ppt
ppt_ch02.pptppt_ch02.ppt
ppt_ch02.ppt
 

Recently uploaded

Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...indiancallgirl4rent
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 

Recently uploaded (20)

Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 

artifactsusgctmriradiologypk-190830131406.pptx

  • 2. Introduction • An Artifact is something observed in a scientific investigation that is not naturally present but occurs as a result of the investigative procedure. • It is a structure not normally present, but visible as a result of malfunction in the hardware or software of the device, or a consequence of environmental influences
  • 3. Types of MRI image artifacts MR hardware and room shielding • Zipper artifact • Herringbone artifact • Zebrastripes • Moire fringes
  • 4.  MR software  Slice-overlap artifact (cross talkartifact)  Cross excitation
  • 5.  Patient and physiologic motion  -Phase-encoded motion artifact  -Entry slice phenomenon
  • 6.  Tissue heterogeneity and foreign bodies  -Magic angleeffect  -Magnetic susceptibility artifact  -Chemical shift artifact
  • 7. Fourier transform -Gibbs artifact /truncation artifact --Aliasing/wrap around artifact
  • 8. • Cause: This type of artifact results from the contamination of the MR scanner environment (shielded MR suite) by spurious Radio Frequency signals. Zipper artifact
  • 9.  A Faraday cage is a conductive enclosure used to shield a space from electromagnetic interference.  In MR imaging, this type of structure provides radiofrequency shielding to the scanning room to minimize occurrence of imaging artifacts.
  • 10. • The contamination can result dueto  1)BreechintheRadiofrequencycagei.e. RF entering the shielded room from outside  2) Illumination devices & other medical equipment (e.g pulse oximeter )
  • 11.  Appearance- A dense broken line of alternatively dark and bright pixels running across the image perpendicular to the frequency encoding direction.
  • 12.
  • 13.  Solution:  1) Scan door should be closed  2) Breech in RF cage should be located and repaired  3) Equipments in the scan room should be serviced
  • 14. Herringbone artifact • Herringbone artifact (also known as ascrisscross artifact or corduroyartifact) • Due to interferences which occur while filing‐in of the k‐space • These interferences are usually RF spikes which lead to points in k‐space with high intensity • Causes ; • Fluctuating power supply • RF pulse discripencies
  • 15. Appearance- It appears as obliquely oriented stripes seen throughout the image. Artifact is scattered all over the image in a single slice or multiple slices .
  • 16. Moire fringes • Moire fringes are an interference pattern most commonly seen when doing gradient echo images with the body coil. • Because of lack of perfect homogeneity of the main magnetic field from one side of the body to the other , aliasing of one side of the body to the other results in superimposition of signals of different phases that alternatively add and cancel. • This causes the banding appearance and is similar to the effect of looking though two screenwindows.
  • 18.  Solutions:  1) Use of spin echo sequences  2) Use of anti Aliasing remedies  3) Plug on the receiver coil should be firmly put in the socket
  • 19. Slice-overlap artifact • Loss of signal in an image due to a multi‐slice, multi‐angle acquisition or imperfect sliceprofile • The mechanism behind this artifact is spinsaturation
  • 20. Image depicts Two sets of spine images, at different angles, covering L4‐5 and L5‐S1 respectively. The set acquired later(red) will have image spins that have already been excited and that are saturated,leading to signal loss If slices at different angles cross, then spins that have been previously excited (saturated) can not be excited again, thus leading to a signal drop in the crossing region
  • 21.
  • 23. Cross excitation artifact • There is loss of signal within a slice due to pre-excitation from RF pulse meant for an adjacent slice • Appearance- ThisArtefact causes a reduction in SNR in adjacent slices in a Slice stack. .
  • 24.  Cause : Ideally the profile of a slice should be square or rectangular when viewed from the edge, but in practice a radio frequency excitation pulse is not able to achieve this.  The adjacent slices receive energy from the RF excitation pulse of their neighbors .  When the tissues in next slice receive their own RF pulse they are unable to produce a signal as they are already saturated
  • 25. Solution:- • Cross excitation can be reduced by insuring that there is at least at 30% gap between the slices (Gap= 30% of the Slice thickness ) This reduces the likelihood of RF pulse exciting adjacent slice. • Interleaving between slices i.e. FIrst acquiring the odd number slices(1,3,5,7..) Then acquiring the even number slices( 2,4,6,8..)
  • 26. Effect of cross-talk on image contrast. On left is a SE image with 50% gap showing expected spin- density contrast. On right the same sequence with 0% gap demonstrating impairedcontrast.
  • 27. On the coronal scout, the lumen of the gallbladder is sharply divided into low signal medially and high signal laterally with a "level" appearance suggesting layering sludge. However, the axial T2W shows no sludge in the gallbladder. This finding is attributed to cross-excitation artifact.
  • 28. Phase-encoded motion artifact • Phase-encoded motion artifact occurs as a result of tissue/fluid moving during the scan. It manifests as ghosting in the direction of phase-encoding. • These artifacts may be seen from arterial pulsations, swallowing, breathing, peristalsis, and physical movement of a patient. When projected over anatomy it can mimic pathology, and needs to be recognized. • Motion that is random such as the patient moving produces a smear in the phase direction. Periodic motion, such as respiratory or cardiac/vascular pulsation, produces discrete, well-defined ghosts.
  • 29. This is a typical example of phase-encoded artifact secondary to aortic pulsation. The ghost images of the aorta are: discrete, well defined, lining with the aorta along the short axis of the abdomen (AP axis) at regular intervals and extending beyond the field of view. These features help recognize them and not to confuse them with any pathology.
  • 30. Blood flow in the sagittal sinus and carotids causes Periodic ghosting which gives the impression of an enhancing lesion in the occipital lobe as well as anterior to the right temporal horn. Wide windowing demonstrates the ghosting extending beyond the confines of the patient's anatomy. Apparent enhancement in the occipital lobe and right temporal lobe in a patient with malignancy. windowed slice demonstrates the phase mismapping artefact from the superior sagittal sinus.
  • 31.  Swallowing artifact has produced horizontal band of noise along phase-encoded direction. Where it passes over vertebral body and spinal cord, there are local intensity and contrast changes, rendering these portions of scan suspect.
  • 32. Solutions  cardiac/respiratory gating  spatial presaturation bands placed over moving tissues (e.g. over the anterior neck in sagittal cervical spines)  spatial presaturation bands placed outside the FOV, especially before the entry or after the exit slice for reducing ghosting from vascular flow  scanning prone to reduce abdominal excursion  increasing the number of signal averages  shorten the scan time when motion is from patient movement
  • 33. Cardiac Gating  This involves the acquistion of a series of MR images of a single anatomic section during a specific point in the cardiac cycle ,which is being monitored with ECG.  Data within the k-space is linked with specific time point in cardiac cycle.  The acquired datasets are then sorted according to time stamp to produce sequential images allowing a dynamic evaluation of myocardial function.
  • 34. Respiratory gating  respiration related motion is monitored, using bellows or breathing belt, and image acquisition is timed to take place at end expiration, when there is little or no motion. however this is infrequently used in clinical imaging. -Breath holding method most often used -Sequences with short acquisition time may be used, or if breath holding possible for limited time, exam may be divided into several brief acquisitions.
  • 35.  Saturation pulses Saturation pulses involve the application of RF energy to suppress the MR signal from moving tissues outside the imaged volume to reduce or eliminate motion artifacts.
  • 37. -Most commonly used in abdominal imaging at the interface of lung and diaphragm -Application of small, one dimensional spatial encoding gradient in a plane perpendicular to diaphragm. - Echo measured at this location allows correction of imaging dataset and ensures that only the imaging data acquired when diaphragm is at its peak (end expiration), is used in image reconstruction. Navigator Echo
  • 38. Entry slice phenomenon Entry slice phenomenon occurs when unsaturated spins in blood first enter into a slice . Unsaturated spins = protons without transverse magnetisation When a large number of such protons encounter a RF pulse for the first time(i.e. at the entry slice),they undergo transverse magnetisation and emit a strong signal. - It is characterized by the bright signal in a blood vessel (artery or vein) at the first slice that the vessel enters. -
  • 39.
  • 40.
  • 41.  This artefact can be confused with thrombosis Solutions-  -The characteristic location and if necessary, the use of gradient echo flow techniques can be used to differentiate entry slice artefacts from occlusions.  -Spatial saturation bands place before the first slice and after the last can also be used to eliminate this artefact.
  • 42. Magic angle artefact • The magic angle effect is important in the clinical MR imaging of certain tissues that are highly structured and are oriented obliquely to the main magnetic field (especially tendons, cartilage, and peripheral nerves). • In tendons, the water molecules are aligned along the tendon sheath and have strong dipolar interactions. Due to these dipolar interactions, protons in the collagen rapidly dissipate their energy after encountering a RF pulse i.e. tendons have a short intrinsic T2 value (1-10 ms). The signal of tendons is uniformly low on all conventional MR sequences due to this rapid T2 decay.
  • 43.
  • 44.
  • 45.  however, when the tendons are aligned at an angle of 55degree to the main magnetic field , these dipolar interactions go to zero, leading to an increase in T2 relaxation time and consequent sudden increase in signal.
  • 46.
  • 47.
  • 48.
  • 49.
  • 50.
  • 51. Common sites:  proximal part of the Posterior cruciate ligament  infrapatellar tendon at the tibial insertion  peroneal tendons as they hook around the lateral malleolus  Supraspinatus tendon
  • 52.  High signal within a tendon can mimic tendinopathy so it is important that radiologists are aware of the magic angle phenomena.  Look for it in the outer portions of the rotator cuff, supraspinatus tendon, the distil patellar tendon, tendons of the ankle, extensor and flexor policis tendons of the wrist or any part made of collagen that lies in a curve
  • 53. Susceptibility artifact • Due to the presence of substances ( e.g.ferromagnetic materials) that causes variation in the local magneticfield • Results in increased magnetic field inhomogeneity ,leading to spin dephasing and consequent signal drop • Becomes worse for long echo times and GRE sequences( no RF 180 pulse) • Less visible in fast spin echosequence • Worse for higher fieldstrength
  • 57. Remedies • Use spin-echo sequences A spin echo sequence uses a 180-degree RF pulse . The main purpose of this 180-degree pulse is to correct local field inhomogeneities. A gradient echo pulse sequence does not utilize such a “refocusing” pulse. As such, local field inhomogeneities are much more readily apparent when gradient echo sequences are utilized
  • 58. Chemical Shift -Chemical Shift is the difference between precessional frequencies of water and fat.(3.5ppm) -Artefact is the displacement of signal intensity due to this difference
  • 59. AT 1.5 T W>F by 220Hz At 3T W>F by 440Hz WATER FAT  Water molecules have low electron shielding and their nucleus is more susceptible to external magnetic field.  As a result water molecules precess at a higher frequency at the same magnetic field.  Fat molecules have higher electron shielding and their nucleus is comparitively less susceptible to external magnetic field.  Fat molecules precess at a lower frequency at the same mag field
  • 60.  water and lipid protons coexist at the interface, the signal emitted by the lipid protons will have a lower frequency than that of the water protons,leading to mismapping of fat.
  • 61. CHEMICAL SHIFT TYPE 1: MISMAPPING ARTIFACT -Most commonly seen around water containg structures sorrounded bt Fat(Liver,Kidneys,Orbits) -Artifact appears as a dark band on one side of the water/fat interface and a bright band on the other side
  • 62.  a black band is seen on the left side of each kidney and a white band on the right.
  • 63.  T2 weighted image demonstrates chemical shift artefact around the lipoma.  Note the signal loss anteriorly (black arrows) and hyperintensity posteriorly (white arrows).  This is the result of inaccurate spatial encoding, resulting in the lipoma being incorrectly placed posterior to its actual location. LIPOMA
  • 64.  axial T1 image also demonstrates a chemical shift artifact at the edges between the mass and the orbital fat. ORBITAL HEMANGIOMA
  • 65.  Chemical Shift: 2nd Kind The chemical shift artifact of the second kind is sometimes known as the India Ink or black line artifact because it looks like organs and muscle bundles have been outlined with a black pen. The line does not correspond to a true anatomic structure, but results from destructive interference of signals at the boundary pixels that contain both water and fat. For this reason it is sometimes often called the phase cancellation artifact. Examples in the spine, thigh, and abdomen.
  • 67. Gibbs artifact / truncationartifact • It refers to a series of bright or dark lines, parallel & next to borders of abrupt intensity change Areas of abrupt intensity change such as -the CSF-Spinal Cord Interface. -the Skull-Brain Interface.
  • 68.  Bright and dark bands are seen parallel and adjacent to the outer convexity of calvaria
  • 70. . Figure illustrates a syrinx-like artifact (red arrow) within the center of the cord, a classic truncation artifact that appears as a result of alternating dark and light signal bands overlying the spinal cord. This artifact can be easily mistaken for hydromyelia (dilatation of the central canal, more commonly referred to as a “syrinx”),
  • 71. Truncation artifacts are reduced by the use of a smaller field of view (FOV) or a larger matrix size.
  • 72. Aliasing / wrap aroundartifact • It occurs when the field of view (FOV) is smaller than the body part being imaged. The part of the body that lies beyond the edge of the FOV is projected onto the other side of the image • Caused by under‐sampling in the phase encoding direction
  • 73. -phase shifts of between 0° and 360° encompass the field-of-view. The subject's left flank in the drawing extends outside the field-of-view and encompasses phase shifts from 361° to 450°. Since all meaningful frequencies have been defined over the range of 0° to 360°, a phase shift of 361° will be assigned to the spatial position of 1°, and a shift of 450° will be assigned to 450°-360° = 90°. Hence ,the left side of the patient's body will therefore be "wrapped around" and spatially mismapped to the opposite (right) side of the image.
  • 74.
  • 75. Solution -Enlarging the field of view (FOV) -Using pre-saturation bands on areas outside the FOV -Anti-aliasing software -Use a surface coil to reduce the signal outside of the area of interest
  • 76.  PARTIAL VOLUME ARTIFACT -Partial volume artefact occurs if slice thickness > thickness of the tissue of interest. -Occurs when multiple tissue types are encompassed within a single voxel -If a small structure is entirely contained within the slice thickness along with other tissue of differing signal intensities, the resulting signal displayed on the image is a combination of these two intensities. This reduces contrast of the small structure.
  • 77. Slice thickness 10mm Unable to depict finer structures due to partial volume artefact Slice thickness 3mm Depicts cranial nerves