SlideShare a Scribd company logo
1 of 30
ULTRASOUND
ARTIFACT
By:Muhammad Amer Sohail
Roll no: 19
DVM 6th morning
Session : 2012-2017
DEFINITION
 An ultrasound artifact is a structure in an
image which does not directly similar with
actual tissue being scanned.
 Artifact assumes different forms including :
• Structures in the image that are not
actually present
• Objects that should be represented but
are missing from the image.
• Structures which are misregistered on the
image.
COMMON ARTIFACTS
1. REVERBERATION
2. ACOUSTIC SHADOWING
3. ACOUSTIC ENHANCEMENT
4. EDGE SHADOWING
5. BEAM WIDTH ARTIFACT
6. SLICE THICKNESS ARTIFACT
7. SIDE LOBE ARTIFACT
8. MIROR IMAGE
9. DOUBLE IMAGE
10. EQUIPMENT-GENERATED ARTIFACT
11. REFRACTION ARTIFACT
REVERBERATION
• This is the production of false echoes due to repeated
reflections between two interfaces with a high acoustic
impedance mismatch.
• The echo from the interface is received by the transducer
and displayed on the image.
• Some of the energy in the returned echo is reflected at the
transducer face, and return to the reflecting interface as if it
was a weak transmitted pulse, returning as a second echo.
• As the time taken for the second echo to arrive is twice that
taken by the depth.
• This sequence of reflection and transmission can occur
many times, with the third echo taking three times as
long to return to the transducer and being displayed at
three times the depth, and so on.
• The reverberation echoes will be equally spaced
because the time for each additional echo is multiple of
the time of return of the first echo.
• These reverberation echoes will be strong because of
the high acoustic mismatch.
This artifact will be seen at the skin-transducer interface
and behind bowel gas.
Rectification:
• Increase the amount of gel used.
• Use a stand-off pad.
• Reduce the gain.
• Move the position of the transducer.
ACOUSTIC SHADOWING
• This appears as an area of low amplitude echoes
behind an area of strongly attenuating tissue.
• It is caused by severe attenuation of the beam at an
interface, resulting in very little sound being transmitted
beyond.
• The attenuation can be due to either absorption or
reflection of the sound waves, or a combination of the
two.
• Acoustic shadowing will occur at interfaces with large
acoustic mismatch such as:
Soft tissue and gas
Soft tissue and bone or calculus
ACOUSTIC ENHANCEMENT
• This artifact appears as a localized area of increased
echo amplitude behind an area of low attenuation.
• On a scan it will appears as an area of increased
brightness, and can commonly be seen distal to fluid-
filled structures such as the urinary bladder, GB or a
cyst.
• The artifact arises due to the application of the time-gain
compensation(TGC) to areas of low attenuating
structures such as fluid.
• It is caused by the low level of attenuation of the beam
as it passes through fluid relative to the greater
attenuation of the beam in the adjacent more solid
tissue.
• This artifact can often be an useful diagnostic aid,
particularly when scanning a soft-tissue mass or cyst
containing low level echoes.
• These echoes may often cause the structure to
disappear in the image as it blend into the surrounding
echo pattern.
EDGE SHADOWING
• A combination of refraction and reflection occurring at
the edges of rounded structures will result in edge
shadowing artifact.
• It arises due to refraction of the beam caused by both
the curvature of the rounded edges and difference in
speed of two materials.
• When the ultrasound beam reaches the rounded edge
of a structure, reflection will occur, with an angle of
incidence equal to the angle of reflection.
• The outer part of the beam will be totally reflected, but
the reminder of the beam passes through the rounded
structure and is refracted.
• This combination of reflection and refraction of the
beam at the edges of a rounded structure results in a
thin strip of tissue behind the edge not being
insonated and causes a shadow.
BEAM WIDTH ARTIFACT
• This artifact can be demonstrated by scanning a point
reflector in a phantom, where the display will clearly
portray this as a line.
• During routine scanning, the artifact can be seen
when spurious echoes are displayed in an echo-free
area.
• Correct positioning of the focal zone will help to
reduce this artifact.
• The focal zone is controlled by electronically
narrowing the beam
SLICE THICKNESS ARTIFACT
• These occurs due to the thickness of the beam, and
are similar to beam width artifacts.
• These artifacts will typically be seen in transverse
views of the urinary bladder when structures adjacent
to the slice through the bladder being scanned will be
incorporated into the image.
• These echoes are then displayed as if they were
arising from within the bladder.
• Although the appearance of this artifact is similar to the
beam width artifact, the differentiating factors is that the
reflector causing the slice thickness artifact will not be
seen on the display.
 This artifact is a result of inherent characteristics of the
transducer, and apart from trying a different transducer,
cannot be eliminated
SIDE LOBE ARTIFACT
• The energy within the ultrasound beam exists as
several side lobes radiating at a number of angles
from a central lobe.
• Echoes are generated by these side lobes in addition
to the main lobe, but all the returning echoes are
assumed by the transducer to have arisen from the
central axis of the main lobe.
• Side lobe echoes will therefore be misregistered in the
display.
• This artifact can often be seen in area such as the
urinary bladder and may also arise within a cyst.
MIRROR IMAGE ARTIFACT
• These artifacts results in a mirror image of a structure
occurring in an ultrasound display.
• They arise due to specular reflection of the beam at a
large smooth interface.
• An area close to a specular reflector will be imaged
twice, once by the original ultrasound beam and once
by the beam after it has reflected off the specular
reflector.
• Mirror image artifacts are most commonly seen where
there is a large acoustic mismatch, such as a fluid-air
interface.
• Typically this artifact can occur during the scanning
of a full bladder, when air in the rectum behind the
bladder act as specular reflector and mirror image of
the bladder is displayed posteriorly.
• It will then have the appearance of a large cyst
behind the bladder.
• It can also be seen when scanning the liver, and the
diaphragm act as a specular reflector.
DOUBLE IMAGE ARTIFACT
• This image is caused by refraction of the beam and
may occur in areas such as the rectus abdominis
muscle on the anterior abdominal wall.
• In the transverse plane the edges of the muscle act as
a lens and the ultrasound beam to be refracted and
this causes the single structure to be interrogated by
two separate refracted beams.
• Two sets of echoes will therefore be returned and
these will cause display of two structures in the image.
This results in, for example, two images of the transverse
aorta side by side in the abdomen.
EQUIPMENT-GENERATED ARTIFACT
• Incorrect use of the equipment controls can lead to
artifact appearing.
• Misuse of controls such as the gain or TGC can result
in echoes being recorded as too bright or too dark.
• Care must be taken when setting these controls, to
ensure an even brightness throughout the image.
• If the dynamic range control is incorrectly set, this can
lead to an image which has too much contrast, and
result in the loss of subtle echo information.
• Gain must be in medium level.
• Blurring of a moving image can occur if the frame rate
is too low or if the persistence is too high.
• It is important to ensure that the frame rate is capable
of recording a moving structure at the speed.
• Use of multiple focal zones can give rise to a
prominent banding effect within the image.
REFRACTION ARTIFCT
• The refraction is the change of the sound direction
on passing from one medium to another.
• In ultrasound, refraction is due to sound velocity
mismatches combines with oblique angles of
incidence, most commonly with convex scanheads.
• When the ultrasound wave crosses at an oblique
angle the interface of two materials, through which
the waves propagate at different velocities,
refraction occurs, caused by bending of the wave
beam.
• Refraction artifact cause spatial distortion and loss
of resolution in the image.
Ultrasound Artifact

More Related Content

What's hot

Prostate MRI anatomy from UNIVERSITY OF MICHIGAN
Prostate MRI anatomy from UNIVERSITY OF MICHIGANProstate MRI anatomy from UNIVERSITY OF MICHIGAN
Prostate MRI anatomy from UNIVERSITY OF MICHIGANKanhu Charan
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifactsansaripv
 
magnetic resonance angiography
magnetic resonance angiographymagnetic resonance angiography
magnetic resonance angiographyqavi786
 
Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)Ankit Mishra
 
radiological anatomy of retroperitoneum powerpoint
radiological anatomy of  retroperitoneum powerpointradiological anatomy of  retroperitoneum powerpoint
radiological anatomy of retroperitoneum powerpointDactarAdhikari
 
Radiological procedure of retrograde urethrography(rgu) and micturating
Radiological procedure of retrograde urethrography(rgu) and micturatingRadiological procedure of retrograde urethrography(rgu) and micturating
Radiological procedure of retrograde urethrography(rgu) and micturatingSanzzuTimilsina
 
CT Physics
CT PhysicsCT Physics
CT PhysicsRMLIMS
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictorsvishwanath0908
 
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni WadhwaniUltrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni WadhwaniChandni Wadhwani
 
Presentation1.pptx, radiological imaging of prostatic diseases
Presentation1.pptx, radiological imaging of prostatic diseasesPresentation1.pptx, radiological imaging of prostatic diseases
Presentation1.pptx, radiological imaging of prostatic diseasesAbdellah Nazeer
 
TRUS in Evaluation of Male Infertility
TRUS in Evaluation of Male InfertilityTRUS in Evaluation of Male Infertility
TRUS in Evaluation of Male InfertilitySiewhong Ho
 
Barium meal follow through
Barium meal follow throughBarium meal follow through
Barium meal follow throughShiva Prakash
 
Learn Barium Meal & Follow Through
Learn Barium Meal & Follow ThroughLearn Barium Meal & Follow Through
Learn Barium Meal & Follow ThroughDr.Santosh Atreya
 
ULTRASOUND PHYSICS
ULTRASOUND PHYSICSULTRASOUND PHYSICS
ULTRASOUND PHYSICSNavni Garg
 
Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.Abdellah Nazeer
 
Radiological procedure questions and answers2 converted
Radiological procedure questions and answers2 convertedRadiological procedure questions and answers2 converted
Radiological procedure questions and answers2 convertedGanesan Yogananthem
 
Anterograde/Retrograde urethrography (RGU/MCU)
Anterograde/Retrograde urethrography (RGU/MCU)Anterograde/Retrograde urethrography (RGU/MCU)
Anterograde/Retrograde urethrography (RGU/MCU)Shubham Singhal
 
KEYS OF RADIOLOGY SPOTTERS GIT
KEYS OF RADIOLOGY SPOTTERS GITKEYS OF RADIOLOGY SPOTTERS GIT
KEYS OF RADIOLOGY SPOTTERS GITAnish Choudhary
 

What's hot (20)

Prostate MRI anatomy from UNIVERSITY OF MICHIGAN
Prostate MRI anatomy from UNIVERSITY OF MICHIGANProstate MRI anatomy from UNIVERSITY OF MICHIGAN
Prostate MRI anatomy from UNIVERSITY OF MICHIGAN
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
magnetic resonance angiography
magnetic resonance angiographymagnetic resonance angiography
magnetic resonance angiography
 
Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)
 
radiological anatomy of retroperitoneum powerpoint
radiological anatomy of  retroperitoneum powerpointradiological anatomy of  retroperitoneum powerpoint
radiological anatomy of retroperitoneum powerpoint
 
Radiological procedure of retrograde urethrography(rgu) and micturating
Radiological procedure of retrograde urethrography(rgu) and micturatingRadiological procedure of retrograde urethrography(rgu) and micturating
Radiological procedure of retrograde urethrography(rgu) and micturating
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
 
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni WadhwaniUltrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
 
Presentation1.pptx, radiological imaging of prostatic diseases
Presentation1.pptx, radiological imaging of prostatic diseasesPresentation1.pptx, radiological imaging of prostatic diseases
Presentation1.pptx, radiological imaging of prostatic diseases
 
TRUS in Evaluation of Male Infertility
TRUS in Evaluation of Male InfertilityTRUS in Evaluation of Male Infertility
TRUS in Evaluation of Male Infertility
 
Barium meal follow through
Barium meal follow throughBarium meal follow through
Barium meal follow through
 
Learn Barium Meal & Follow Through
Learn Barium Meal & Follow ThroughLearn Barium Meal & Follow Through
Learn Barium Meal & Follow Through
 
ULTRASOUND PHYSICS
ULTRASOUND PHYSICSULTRASOUND PHYSICS
ULTRASOUND PHYSICS
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CT
 
Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.Presentation1.pptx, ultrasound examination of the orbit.
Presentation1.pptx, ultrasound examination of the orbit.
 
Radiological procedure questions and answers2 converted
Radiological procedure questions and answers2 convertedRadiological procedure questions and answers2 converted
Radiological procedure questions and answers2 converted
 
Anterograde/Retrograde urethrography (RGU/MCU)
Anterograde/Retrograde urethrography (RGU/MCU)Anterograde/Retrograde urethrography (RGU/MCU)
Anterograde/Retrograde urethrography (RGU/MCU)
 
KEYS OF RADIOLOGY SPOTTERS GIT
KEYS OF RADIOLOGY SPOTTERS GITKEYS OF RADIOLOGY SPOTTERS GIT
KEYS OF RADIOLOGY SPOTTERS GIT
 

Similar to Ultrasound Artifact

ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxAnayaAnaya14
 
Ultrasound of eye - B scan
Ultrasound of eye - B scan Ultrasound of eye - B scan
Ultrasound of eye - B scan Shruti Laddha
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imagingmtuzer
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects Nusirat Adedewe
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT PAWAN JARWAL
 
Artifacts in echocardiography
Artifacts  in echocardiographyArtifacts  in echocardiography
Artifacts in echocardiographyDilzith Katepogu
 
artefacts in ultrasound imaging: physics .pptx
artefacts in ultrasound imaging: physics .pptxartefacts in ultrasound imaging: physics .pptx
artefacts in ultrasound imaging: physics .pptxWangjuSumnyan
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducersRaghu Kishore Galla
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxShubham661884
 
BScan and Ascan in ophthalmology and eye field
BScan and Ascan  in ophthalmology and eye fieldBScan and Ascan  in ophthalmology and eye field
BScan and Ascan in ophthalmology and eye fieldAsif469093
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptxdypradio
 

Similar to Ultrasound Artifact (20)

ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptx
 
Ultrasound of eye - B scan
Ultrasound of eye - B scan Ultrasound of eye - B scan
Ultrasound of eye - B scan
 
Us artefacts
Us artefactsUs artefacts
Us artefacts
 
Ct artiact
Ct artiactCt artiact
Ct artiact
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imaging
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects
 
Usg
UsgUsg
Usg
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT
 
ULTRA.pptx
ULTRA.pptxULTRA.pptx
ULTRA.pptx
 
Artifacts in echocardiography
Artifacts  in echocardiographyArtifacts  in echocardiography
Artifacts in echocardiography
 
usg artifacts.pptx
usg artifacts.pptxusg artifacts.pptx
usg artifacts.pptx
 
artefacts in ultrasound imaging: physics .pptx
artefacts in ultrasound imaging: physics .pptxartefacts in ultrasound imaging: physics .pptx
artefacts in ultrasound imaging: physics .pptx
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducers
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
BScan and Ascan in ophthalmology and eye field
BScan and Ascan  in ophthalmology and eye fieldBScan and Ascan  in ophthalmology and eye field
BScan and Ascan in ophthalmology and eye field
 
Us physics 3
Us physics 3Us physics 3
Us physics 3
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
 
Usg artifacts
Usg artifactsUsg artifacts
Usg artifacts
 

Recently uploaded

Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfUjwalaBharambe
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxRaymartEstabillo3
 
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxiammrhaywood
 
HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...
HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...
HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...Nguyen Thanh Tu Collection
 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxpboyjonauth
 
Hierarchy of management that covers different levels of management
Hierarchy of management that covers different levels of managementHierarchy of management that covers different levels of management
Hierarchy of management that covers different levels of managementmkooblal
 
AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.arsicmarija21
 
ACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfSpandanaRallapalli
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentInMediaRes1
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfMr Bounab Samir
 
Introduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher EducationIntroduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher Educationpboyjonauth
 
Keynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-designKeynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-designMIPLM
 
Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxOH TEIK BIN
 
Romantic Opera MUSIC FOR GRADE NINE pptx
Romantic Opera MUSIC FOR GRADE NINE pptxRomantic Opera MUSIC FOR GRADE NINE pptx
Romantic Opera MUSIC FOR GRADE NINE pptxsqpmdrvczh
 
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxGrade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxChelloAnnAsuncion2
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Celine George
 

Recently uploaded (20)

Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
 
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdfTataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
 
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdfFraming an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
Framing an Appropriate Research Question 6b9b26d93da94caf993c038d9efcdedb.pdf
 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
 
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptxECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
ECONOMIC CONTEXT - PAPER 1 Q3: NEWSPAPERS.pptx
 
HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...
HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...
HỌC TỐT TIẾNG ANH 11 THEO CHƯƠNG TRÌNH GLOBAL SUCCESS ĐÁP ÁN CHI TIẾT - CẢ NĂ...
 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptx
 
9953330565 Low Rate Call Girls In Rohini Delhi NCR
9953330565 Low Rate Call Girls In Rohini  Delhi NCR9953330565 Low Rate Call Girls In Rohini  Delhi NCR
9953330565 Low Rate Call Girls In Rohini Delhi NCR
 
Hierarchy of management that covers different levels of management
Hierarchy of management that covers different levels of managementHierarchy of management that covers different levels of management
Hierarchy of management that covers different levels of management
 
AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.AmericanHighSchoolsprezentacijaoskolama.
AmericanHighSchoolsprezentacijaoskolama.
 
ACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdf
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media Component
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
 
Introduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher EducationIntroduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher Education
 
OS-operating systems- ch04 (Threads) ...
OS-operating systems- ch04 (Threads) ...OS-operating systems- ch04 (Threads) ...
OS-operating systems- ch04 (Threads) ...
 
Keynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-designKeynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-design
 
Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptx
 
Romantic Opera MUSIC FOR GRADE NINE pptx
Romantic Opera MUSIC FOR GRADE NINE pptxRomantic Opera MUSIC FOR GRADE NINE pptx
Romantic Opera MUSIC FOR GRADE NINE pptx
 
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxGrade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17
 

Ultrasound Artifact

  • 1. ULTRASOUND ARTIFACT By:Muhammad Amer Sohail Roll no: 19 DVM 6th morning Session : 2012-2017
  • 2. DEFINITION  An ultrasound artifact is a structure in an image which does not directly similar with actual tissue being scanned.  Artifact assumes different forms including : • Structures in the image that are not actually present • Objects that should be represented but are missing from the image. • Structures which are misregistered on the image.
  • 3. COMMON ARTIFACTS 1. REVERBERATION 2. ACOUSTIC SHADOWING 3. ACOUSTIC ENHANCEMENT 4. EDGE SHADOWING 5. BEAM WIDTH ARTIFACT 6. SLICE THICKNESS ARTIFACT 7. SIDE LOBE ARTIFACT 8. MIROR IMAGE 9. DOUBLE IMAGE 10. EQUIPMENT-GENERATED ARTIFACT 11. REFRACTION ARTIFACT
  • 4. REVERBERATION • This is the production of false echoes due to repeated reflections between two interfaces with a high acoustic impedance mismatch. • The echo from the interface is received by the transducer and displayed on the image. • Some of the energy in the returned echo is reflected at the transducer face, and return to the reflecting interface as if it was a weak transmitted pulse, returning as a second echo. • As the time taken for the second echo to arrive is twice that taken by the depth.
  • 5. • This sequence of reflection and transmission can occur many times, with the third echo taking three times as long to return to the transducer and being displayed at three times the depth, and so on. • The reverberation echoes will be equally spaced because the time for each additional echo is multiple of the time of return of the first echo. • These reverberation echoes will be strong because of the high acoustic mismatch.
  • 6. This artifact will be seen at the skin-transducer interface and behind bowel gas. Rectification: • Increase the amount of gel used. • Use a stand-off pad. • Reduce the gain. • Move the position of the transducer.
  • 7.
  • 8. ACOUSTIC SHADOWING • This appears as an area of low amplitude echoes behind an area of strongly attenuating tissue. • It is caused by severe attenuation of the beam at an interface, resulting in very little sound being transmitted beyond. • The attenuation can be due to either absorption or reflection of the sound waves, or a combination of the two. • Acoustic shadowing will occur at interfaces with large acoustic mismatch such as: Soft tissue and gas Soft tissue and bone or calculus
  • 9.
  • 10.
  • 11. ACOUSTIC ENHANCEMENT • This artifact appears as a localized area of increased echo amplitude behind an area of low attenuation. • On a scan it will appears as an area of increased brightness, and can commonly be seen distal to fluid- filled structures such as the urinary bladder, GB or a cyst. • The artifact arises due to the application of the time-gain compensation(TGC) to areas of low attenuating structures such as fluid. • It is caused by the low level of attenuation of the beam as it passes through fluid relative to the greater attenuation of the beam in the adjacent more solid tissue.
  • 12. • This artifact can often be an useful diagnostic aid, particularly when scanning a soft-tissue mass or cyst containing low level echoes. • These echoes may often cause the structure to disappear in the image as it blend into the surrounding echo pattern.
  • 13. EDGE SHADOWING • A combination of refraction and reflection occurring at the edges of rounded structures will result in edge shadowing artifact. • It arises due to refraction of the beam caused by both the curvature of the rounded edges and difference in speed of two materials. • When the ultrasound beam reaches the rounded edge of a structure, reflection will occur, with an angle of incidence equal to the angle of reflection. • The outer part of the beam will be totally reflected, but the reminder of the beam passes through the rounded structure and is refracted.
  • 14. • This combination of reflection and refraction of the beam at the edges of a rounded structure results in a thin strip of tissue behind the edge not being insonated and causes a shadow.
  • 15. BEAM WIDTH ARTIFACT • This artifact can be demonstrated by scanning a point reflector in a phantom, where the display will clearly portray this as a line. • During routine scanning, the artifact can be seen when spurious echoes are displayed in an echo-free area. • Correct positioning of the focal zone will help to reduce this artifact. • The focal zone is controlled by electronically narrowing the beam
  • 16.
  • 17. SLICE THICKNESS ARTIFACT • These occurs due to the thickness of the beam, and are similar to beam width artifacts. • These artifacts will typically be seen in transverse views of the urinary bladder when structures adjacent to the slice through the bladder being scanned will be incorporated into the image. • These echoes are then displayed as if they were arising from within the bladder. • Although the appearance of this artifact is similar to the beam width artifact, the differentiating factors is that the reflector causing the slice thickness artifact will not be seen on the display.
  • 18.  This artifact is a result of inherent characteristics of the transducer, and apart from trying a different transducer, cannot be eliminated
  • 19. SIDE LOBE ARTIFACT • The energy within the ultrasound beam exists as several side lobes radiating at a number of angles from a central lobe. • Echoes are generated by these side lobes in addition to the main lobe, but all the returning echoes are assumed by the transducer to have arisen from the central axis of the main lobe. • Side lobe echoes will therefore be misregistered in the display. • This artifact can often be seen in area such as the urinary bladder and may also arise within a cyst.
  • 20.
  • 21. MIRROR IMAGE ARTIFACT • These artifacts results in a mirror image of a structure occurring in an ultrasound display. • They arise due to specular reflection of the beam at a large smooth interface. • An area close to a specular reflector will be imaged twice, once by the original ultrasound beam and once by the beam after it has reflected off the specular reflector. • Mirror image artifacts are most commonly seen where there is a large acoustic mismatch, such as a fluid-air interface.
  • 22. • Typically this artifact can occur during the scanning of a full bladder, when air in the rectum behind the bladder act as specular reflector and mirror image of the bladder is displayed posteriorly. • It will then have the appearance of a large cyst behind the bladder. • It can also be seen when scanning the liver, and the diaphragm act as a specular reflector.
  • 23.
  • 24. DOUBLE IMAGE ARTIFACT • This image is caused by refraction of the beam and may occur in areas such as the rectus abdominis muscle on the anterior abdominal wall. • In the transverse plane the edges of the muscle act as a lens and the ultrasound beam to be refracted and this causes the single structure to be interrogated by two separate refracted beams. • Two sets of echoes will therefore be returned and these will cause display of two structures in the image.
  • 25. This results in, for example, two images of the transverse aorta side by side in the abdomen.
  • 26. EQUIPMENT-GENERATED ARTIFACT • Incorrect use of the equipment controls can lead to artifact appearing. • Misuse of controls such as the gain or TGC can result in echoes being recorded as too bright or too dark. • Care must be taken when setting these controls, to ensure an even brightness throughout the image. • If the dynamic range control is incorrectly set, this can lead to an image which has too much contrast, and result in the loss of subtle echo information.
  • 27. • Gain must be in medium level. • Blurring of a moving image can occur if the frame rate is too low or if the persistence is too high. • It is important to ensure that the frame rate is capable of recording a moving structure at the speed. • Use of multiple focal zones can give rise to a prominent banding effect within the image.
  • 28.
  • 29. REFRACTION ARTIFCT • The refraction is the change of the sound direction on passing from one medium to another. • In ultrasound, refraction is due to sound velocity mismatches combines with oblique angles of incidence, most commonly with convex scanheads. • When the ultrasound wave crosses at an oblique angle the interface of two materials, through which the waves propagate at different velocities, refraction occurs, caused by bending of the wave beam. • Refraction artifact cause spatial distortion and loss of resolution in the image.