Separation and Quantitation
of Oxysterol Compounds
Using LC-MS/MS
Shimadzu Scientific Instruments, Columbia, MD, USA
Summary
A mixture of 16 oxysterols and cholesterol related compounds were separated using a Shimadzu
Nexera UHPLC and analyzed by a Shimadzu LCMS-8060 system. Detection and quantitation
limits were obtained using APCI, ESI, and Dual Ion (DUIS) sources.
Background
There is high demand for oxysterol quantitation due to their correlation with neurodegenerative
diseases. The ratios of various oxysterols in biological fluids are used by researchers to study
disease states.
An LCMS oxysterol quantitation method was developed using a Shimadzu LCMS-8060.
Detection and quantitation limits were determined using multiple reaction monitoring (MRM)
mode for each analyte.
Method
Oxysterol standards were obtained in methanol and diluted
with 10:90 H2O:MeOH. Standards included the compounds
indicated in Table 1.
An MRM method was determined and optimized using
LabSolutions 5.82 on a Shimadzu triple quadrupole mass
spectrometer (LCMS-8060). A Shim-pack XR-ODS III
column was used for the separation. Injection volume was
2 µL with autosampler temperature being set at 10°C. Flow
rate was initially at 0.4 mL/min and adjusted to 0.7 mL/min
toward the end of the run to speed the elution of highly
retained compounds.
Compound name Abbreviation
24 (S)-hydroxycholesterol 24HC
(D7)22-hydroscycholesterol 22HC(d7)
25-hydroxycholesterol 25HC
27-hydroxycholesterol 27HC
(D7)7α-hydroxycholesterol 7αHC(d7)
7α-hydroxycholesterol 7αHC
7β-hydroxycholesterol 7βHC
7-Ketocholesterol 7KC
(D7) 7-Ketocholesterol 7KC(d7)
7α-hydroxycholestenone 7αHCn
(D3) Vitamin D3 VitD3(d3)
Zymosterol Zymo
Desmosterol Desmo
7α,27 dihydroxycholestenone 7α,27diHC,3one
Cholesterol CH
7dehydrocholesterol 7DHC
Table 1. Oxysterol Related Compounds
Experimental
MRM transitions were developed and optimized on the triple quadrupole mass spectrometer
using DUIS, APCI, and ESI sources. In cases where two precursor ions showed a strong signal,
MRM development and optimization were done on both precursor ions. This may provide an
option for better sensitivity in different matrices for future applications.
The DUIS method established nebulizing, heating, and drying gas flows at 3, 17, and 3 L/min.
Interface, desolvation line, and heat block temperatures were set at 300°C, 250°C, and 400°C,
respectively.
With the APCI source, temperatures for the interface, desolvation line, and heat block were set
at 350°C, 200°C, and 200°C, respectively. Drying gas flow rate was 5 L/min.
For ESI, nebulizing, heating, and drying gas flows were set at 3, 10, and 10 L/min. Interface,
desolvation line, and heat block temperatures were 300°C, 250°C, and 400°C.
Experimental
Compound
Name
Retention
time (min)
DUIS APCI ESI
Target MRM
Reference
MRM
Target MRM
Reference
MRM
Target MRM
Reference
MRM
24HC 4.0 385.45>367.30 385.45>324.30 367.35>91.05 367.35>104.85 385.10>367.25 385.10>109.00
24HC 4.0 367.30>281.10 367.30>104.90 385.40>367.30 385.40>104.90 367.45>95.15 367.45>105.25
22HC(d7) 4.2 374.40>91.10 374.40>255.25 374.40>104.80 374.40>133.15 374.20>104.90 374.20>132.95
25HC 4.1 385.30>367.40 385.30>324.25 367.35>95.00 367.35>135.10 367.35>105.15 367.35>147.15
25HC 4.1 367.35>81.15 367.35>105.10 385.35>367.30 385.35>133.10 385.20>367.30 385.20>324.45
27HC 4.5 385.05>67.15 385.05>93.10 385.10>135.10 385.10>149.10 385.15>95.05 385.15>324.35
7αHC(d7) 7.6 374.35>159.15 374.35>91.15 374.40>145.30 374.40>159.25 374.20>158.95 374.20>144.80
7αHC 7.7 367.35>117.10 367.35>66.90 367.45>144.95 367.45>95.00 367.15>145.40 367.15>159.35
7αHC 7.7 385.30>367.40 385.30>367.40 385.10>367.25 385.10>159.10 385.45>367.35 385.45>159.05
7βHC 7.9 385.30>367.30 385.30>324.35 367.15>159.10 367.15>145.30 367.10>95.15 367.10>158.95
7βHC 7.9 367.35>81.10 367.35>104.85 385.10>367.30 385.10>159.00 385.10>367.15 385.10>158.95
7KC 8.3 401.15>95.10 401.15>383.30 401.10>80.95 401.10>383.35 401.45>95.30 401.45>383.30
7KC(d7) 8.1 408.30>390.15 408.30>95.15 408.40>81.15 408.40>95.25 408.40>390.25 408.40>95.05
7αHCn 6.5 401.25>383.30 401.25>97.10 401.35>383.45 401.35>97.10 401.10>382.95 401.10>96.90
VitD3(d3) 3.0 386.25>232.20 386.25>368.45 386.40>368.35 386.40>92.90 386.40>368.30 386.40>95.20
Zymo 12.8 367.30>95.00 367.30>80.90 367.15>95.20 367.15>109.20 367.15>95.10 367.15>80.95
Desmo 13.3 367.25>95.05 367.25>95.05 367.35>135.25 367.35>104.90 367.15>95.00 367.15>104.95
7α,27diHC,3one 2.2 417.25>399.10 417.25>381.30 417.30>399.25 417.30>381.45 417.35>399.30 417.35>97.25
CH 17.4 369.30>161.10 369.30>94.90 369.40>95.20 369.40>146.95 369.40>147.25 369.40>135.25
7DHC 14.5 367.25>95.05 367.25>158.90 367.30>145.25 367.30>159.25 367.15>145.10 367.15>159.20
Table 2. MRM
transitions of oxysterol
related compounds
Multiple MRM
transitions were
determined from
each precursor ion,
the two most
intense are shown.
Results
Chromatographic separation was accomplished using a Shimadzu Nexera UHPLC system
with a Shim-pack XR-ODS III column. Challenging separations of 24HC and 25HC as well as
7αHC and 7βHC were achieved to allow individual quantitation of these analytes. Maximum
operating pressure for a typical run was about 11,000 psi.
Figure 1. Chromatogram of the 16 oxysterol related compounds.
0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 min
0
2500000
5000000
7a,27diHC,3one
VitD3(d3)
7aHCn
7βHC
7aHC
7aHC(d7)
27HC24HC CH
Intensity
Results
4.0 5.0 6.0 7.0 8.0 min
24HC
27HC
7βHC
7aHC
Figure 2. Chromatogram showing separation of key components.
Detection limits determined using ESI, APCI, and DUIS sources are shown below. Limits of
detection (LOD) and limits of quantitation (LOQ) were determined by a signal to noise ratio
of 3:1 and 10:1, respectively.
Results
LOD LOQ LOD LOQ LOD LOQ
24 (S)-hydroxycholesterol 20 30 30 50 20 50
(D7) 22-hydroscycholesterol 10 20 30 50 10 20
25-hydroxycholesterol 20 30 50 100 20 30
27-hydroxycholesterol 20 50 50 50 20 100
(D7) 7a-hydroxycholesterol 20 20 10 10 10 10
7a-hydroxycholesterol 4 10 10 10 10 20
7b-hydroxycholesterol 10 10 10 30 10 20
7-Ketocholesterol 1 4 4 10 1 4
(D7) 7-Ketocholesterol 0.5 2 2 4 0.5 2
7a-hydroxycholestenone 0.5 4 4 4 1 4
(D3) Vitamin D3 30 300 50 50 20 30
Zymosterol 100 >2000 30 300 300 300
Desmosterol 300 >2000 30 50 100 300
7a,27 dihydroxycholestenone 0.25 1 2 4 0.5 2
Cholesterol 100 1000 100 500 100 500
7dehydrocholesterol 100 300 100 300 100 500
ESI APCI DUIS
Table 3. LODs and LOQs determined by LCMS-8060 with ESI, APCI, and DUIS sources. Values shown are
picograms on-column. Green highlighting indicates lower or lowest limits. Yellow highlighting indicates those limits
which are somewhat higher than the lowest limits. Red highlighting indicates the highest limits.
Results
Although ESI was able to obtain low detection limits for most oxysterol related compounds,
it had difficulty reaching practical quantitation limits for zymosterol and desmosterol, even in
neat standards. APCI allowed better quantitation for these two compounds; however, the
LODs and LOQs for many other compounds were sacrificed.
DUIS offers the benefit of both APCI and ESI by allowing quantitation of zymosterol and
desmosterol without significantly sacrificing the LODs of other compounds. Therefore, DUIS
is the optimal source for analyzing this oxysterol related mixture. The dual ion DUIS source
is able to quantify all sixteen analytes in a single run instead of two runs using ESI and
APCI sources separately.
In the case of (D3) Vitamin D3, detection and quantitation was even lower using DUIS vs.
APCI and ESI. This is likely the result of probe position optimization, which is possible with
DUIS (and ESI) but not with APCI.
Results
MRM transitions for sixteen oxysterol related compounds were identified and optimized
using the LCMS-8060. A UHPLC column was used to obtain separation. 24HC, 25HC,
7αHC, and 7βHC were chromatographically separated in a manner sufficient for individual
quantitation. The total run time for this method was 21 minutes.
Limits of detection and limits of quantitation for oxysterol related compounds were
determined using APCI, ESI, and DUIS sources. Although LOD and LOQ vary per
compound, the LCMS-8060 was able to reach detection limits in the low picogram (mass
on-column) range for oxysterol related compounds.
DUIS was shown to be the optimal source for this oxysterol related compound mixture
analysis. It demonstrates the advantages of both ESI and APCI and allows all sixteen
compounds to be analyzed in a single run.
Conclusion
A fast and sensitive method using a triple quadrupole mass spectrometer was developed
to assist future applications for oxysterol related compound quantitation.
Need More Info?
Thank you for viewing this presentation. Should you have any
questions or require additional information about our research,
products, or services, please visit our Web site at:
www.ssi.shimadzu.com
Follow us on Twitter @shimadzussi

More Related Content

What's hot

Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Shimadzu Scientific Instruments
 
High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...
High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...
High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...Shimadzu Scientific Instruments
 
Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...
Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...
Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...Shimadzu Scientific Instruments
 
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer Shimadzu Scientific Instruments
 
Development and Validation of the HPLC Method for the Analysis of Ametridio...
	 Development and Validation of the HPLC Method for the Analysis of Ametridio...	 Development and Validation of the HPLC Method for the Analysis of Ametridio...
Development and Validation of the HPLC Method for the Analysis of Ametridio...iosrphr_editor
 
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...PerkinElmer, Inc.
 
Comparative investigation of food supplements containing ascorbic acid Danka ...
Comparative investigation of food supplements containing ascorbic acid Danka ...Comparative investigation of food supplements containing ascorbic acid Danka ...
Comparative investigation of food supplements containing ascorbic acid Danka ...Georgi Daskalov
 
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...pharmaindexing
 
SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014Kenneth Wong
 
Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...BRNSS Publication Hub
 
Case Study: Nylon Syringe Filter E&L
Case Study: Nylon Syringe Filter E&LCase Study: Nylon Syringe Filter E&L
Case Study: Nylon Syringe Filter E&LJordi Labs
 
Vijendra
VijendraVijendra
Vijendravij1989
 
Revisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMSRevisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMSShimadzu Scientific Instruments
 
Charged aerosol detector in HPLC.ppt
Charged aerosol detector in HPLC.pptCharged aerosol detector in HPLC.ppt
Charged aerosol detector in HPLC.pptsas3333
 
Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...ijtsrd
 

What's hot (20)

Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
 
High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...
High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...
High-sensitivity analysis of pesticides in dried hops cones and hops pellets ...
 
Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...
Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...
Fully Automated Online Sample Extraction and Analysis of Mycotoxins in Foods ...
 
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
 
Development and Validation of the HPLC Method for the Analysis of Ametridio...
	 Development and Validation of the HPLC Method for the Analysis of Ametridio...	 Development and Validation of the HPLC Method for the Analysis of Ametridio...
Development and Validation of the HPLC Method for the Analysis of Ametridio...
 
B03708012
B03708012B03708012
B03708012
 
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinEl...
 
A360105
A360105A360105
A360105
 
Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water
 
Obreshkova
ObreshkovaObreshkova
Obreshkova
 
Comparative investigation of food supplements containing ascorbic acid Danka ...
Comparative investigation of food supplements containing ascorbic acid Danka ...Comparative investigation of food supplements containing ascorbic acid Danka ...
Comparative investigation of food supplements containing ascorbic acid Danka ...
 
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
 
SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014SGSLSS_Migration_WONG_Poster_2014
SGSLSS_Migration_WONG_Poster_2014
 
Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...
 
Case Study: Nylon Syringe Filter E&L
Case Study: Nylon Syringe Filter E&LCase Study: Nylon Syringe Filter E&L
Case Study: Nylon Syringe Filter E&L
 
Chromatography: Concentration of Human Hormones in Drinking Water Using a N...
Chromatography:   Concentration of Human Hormones in Drinking Water Using a N...Chromatography:   Concentration of Human Hormones in Drinking Water Using a N...
Chromatography: Concentration of Human Hormones in Drinking Water Using a N...
 
Vijendra
VijendraVijendra
Vijendra
 
Revisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMSRevisions to EPA Method 624 for Analysis of VOCs by GCMS
Revisions to EPA Method 624 for Analysis of VOCs by GCMS
 
Charged aerosol detector in HPLC.ppt
Charged aerosol detector in HPLC.pptCharged aerosol detector in HPLC.ppt
Charged aerosol detector in HPLC.ppt
 
Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...Multiple Method Development and Validation for Simultaneous Estimation of Chl...
Multiple Method Development and Validation for Simultaneous Estimation of Chl...
 

Similar to LC-MS/MS Separation and Quantitation of Oxysterol Compounds

consensus superiority of the pharmacophore based alignment, over maximum comm...
consensus superiority of the pharmacophore based alignment, over maximum comm...consensus superiority of the pharmacophore based alignment, over maximum comm...
consensus superiority of the pharmacophore based alignment, over maximum comm...Deepak Rohilla
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...PerkinElmer, Inc.
 
ASMS 2016 Lipid and metabolite distribution in Tissue_Final
ASMS 2016 Lipid and metabolite distribution in Tissue_FinalASMS 2016 Lipid and metabolite distribution in Tissue_Final
ASMS 2016 Lipid and metabolite distribution in Tissue_FinalJoe Kennedy
 
Ultrafast Analysis of Anti-Cancer Drugs
Ultrafast Analysis of Anti-Cancer DrugsUltrafast Analysis of Anti-Cancer Drugs
Ultrafast Analysis of Anti-Cancer DrugsMichael Roberts
 
Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...
Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...
Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...Annex Publishers
 
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...PerkinElmer, Inc.
 
Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...
Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...
Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...TELKOMNIKA JOURNAL
 
Swift and efficient_sono-hydrolysis_of_n
Swift and efficient_sono-hydrolysis_of_nSwift and efficient_sono-hydrolysis_of_n
Swift and efficient_sono-hydrolysis_of_nKarlitox Saoj
 
A brief review of exosome & its detection
A brief review of exosome & its detectionA brief review of exosome & its detection
A brief review of exosome & its detectionPeng-Wen Liu
 
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesRapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesShimadzu Scientific Instruments
 
Extractables and Leachables Analysis of Common Household Food Storage Product...
Extractables and Leachables Analysis of Common Household Food Storage Product...Extractables and Leachables Analysis of Common Household Food Storage Product...
Extractables and Leachables Analysis of Common Household Food Storage Product...Shimadzu Scientific Instruments
 
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...PerkinElmer, Inc.
 
Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...
Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...
Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...Covance
 
LCMS PHYTOCHEMICAL ANALYSIS
LCMS PHYTOCHEMICAL ANALYSISLCMS PHYTOCHEMICAL ANALYSIS
LCMS PHYTOCHEMICAL ANALYSISBushraYasin4
 
Poster-mAbChem-ppt file
Poster-mAbChem-ppt filePoster-mAbChem-ppt file
Poster-mAbChem-ppt fileRongliang Lou
 

Similar to LC-MS/MS Separation and Quantitation of Oxysterol Compounds (20)

consensus superiority of the pharmacophore based alignment, over maximum comm...
consensus superiority of the pharmacophore based alignment, over maximum comm...consensus superiority of the pharmacophore based alignment, over maximum comm...
consensus superiority of the pharmacophore based alignment, over maximum comm...
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
 
An 0903-0042 en
An 0903-0042 enAn 0903-0042 en
An 0903-0042 en
 
An 0903-0042 en
An 0903-0042 enAn 0903-0042 en
An 0903-0042 en
 
ASMS 2016 Lipid and metabolite distribution in Tissue_Final
ASMS 2016 Lipid and metabolite distribution in Tissue_FinalASMS 2016 Lipid and metabolite distribution in Tissue_Final
ASMS 2016 Lipid and metabolite distribution in Tissue_Final
 
Ultrafast Analysis of Anti-Cancer Drugs
Ultrafast Analysis of Anti-Cancer DrugsUltrafast Analysis of Anti-Cancer Drugs
Ultrafast Analysis of Anti-Cancer Drugs
 
Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...
Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...
Identification of-designer-drugs-using-gas-chromatography-high-resolution-mas...
 
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
 
ASMSPoster_2015
ASMSPoster_2015ASMSPoster_2015
ASMSPoster_2015
 
Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...
Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...
Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS ...
 
Swift and efficient_sono-hydrolysis_of_n
Swift and efficient_sono-hydrolysis_of_nSwift and efficient_sono-hydrolysis_of_n
Swift and efficient_sono-hydrolysis_of_n
 
A brief review of exosome & its detection
A brief review of exosome & its detectionA brief review of exosome & its detection
A brief review of exosome & its detection
 
Acamprosate Analytical
Acamprosate AnalyticalAcamprosate Analytical
Acamprosate Analytical
 
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesRapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
 
Extractables and Leachables Analysis of Common Household Food Storage Product...
Extractables and Leachables Analysis of Common Household Food Storage Product...Extractables and Leachables Analysis of Common Household Food Storage Product...
Extractables and Leachables Analysis of Common Household Food Storage Product...
 
c4ra02698e
c4ra02698ec4ra02698e
c4ra02698e
 
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
 
Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...
Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...
Quantitative Analysis of Oligonucleotides in Human Muscle Tissue Using Liquid...
 
LCMS PHYTOCHEMICAL ANALYSIS
LCMS PHYTOCHEMICAL ANALYSISLCMS PHYTOCHEMICAL ANALYSIS
LCMS PHYTOCHEMICAL ANALYSIS
 
Poster-mAbChem-ppt file
Poster-mAbChem-ppt filePoster-mAbChem-ppt file
Poster-mAbChem-ppt file
 

More from Shimadzu Scientific Instruments

A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...
A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...
A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...Shimadzu Scientific Instruments
 
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsApplication of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsShimadzu Scientific Instruments
 
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...Shimadzu Scientific Instruments
 
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...Shimadzu Scientific Instruments
 
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR MicroscopyIllicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR MicroscopyShimadzu Scientific Instruments
 
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Shimadzu Scientific Instruments
 
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR MicroscopeExploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR MicroscopeShimadzu Scientific Instruments
 
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...Shimadzu Scientific Instruments
 
Microplastics Detection and Characterization using FTIR Microscopy
Microplastics Detection and Characterization using FTIR MicroscopyMicroplastics Detection and Characterization using FTIR Microscopy
Microplastics Detection and Characterization using FTIR MicroscopyShimadzu Scientific Instruments
 
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...Shimadzu Scientific Instruments
 
The PFAS conundrum: Mass spectrometry solutions for addressing it.
The PFAS conundrum: Mass spectrometry solutions for addressing it.The PFAS conundrum: Mass spectrometry solutions for addressing it.
The PFAS conundrum: Mass spectrometry solutions for addressing it.Shimadzu Scientific Instruments
 
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...Shimadzu Scientific Instruments
 
Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...
Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...
Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...Shimadzu Scientific Instruments
 
The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...
The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...
The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...Shimadzu Scientific Instruments
 
Characterization of residual gunpowder recovered from articles of clothing ut...
Characterization of residual gunpowder recovered from articles of clothing ut...Characterization of residual gunpowder recovered from articles of clothing ut...
Characterization of residual gunpowder recovered from articles of clothing ut...Shimadzu Scientific Instruments
 
Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...
Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...
Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...Shimadzu Scientific Instruments
 

More from Shimadzu Scientific Instruments (17)

A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...
A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...
A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...
 
Determination of Heavy Metals in Baby Foods by ICP-MS
Determination of Heavy Metals in Baby Foods by ICP-MSDetermination of Heavy Metals in Baby Foods by ICP-MS
Determination of Heavy Metals in Baby Foods by ICP-MS
 
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsApplication of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
 
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...
 
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...
 
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR MicroscopyIllicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
 
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
 
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR MicroscopeExploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
 
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...
 
Microplastics Detection and Characterization using FTIR Microscopy
Microplastics Detection and Characterization using FTIR MicroscopyMicroplastics Detection and Characterization using FTIR Microscopy
Microplastics Detection and Characterization using FTIR Microscopy
 
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
 
The PFAS conundrum: Mass spectrometry solutions for addressing it.
The PFAS conundrum: Mass spectrometry solutions for addressing it.The PFAS conundrum: Mass spectrometry solutions for addressing it.
The PFAS conundrum: Mass spectrometry solutions for addressing it.
 
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...
 
Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...
Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...
Testing the Tensile Properties of Rigid and Semi-rigid Plastics (ASTM D638 an...
 
The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...
The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...
The Determination of CBD and General Cannabinoid Content in Hemp Oils Using H...
 
Characterization of residual gunpowder recovered from articles of clothing ut...
Characterization of residual gunpowder recovered from articles of clothing ut...Characterization of residual gunpowder recovered from articles of clothing ut...
Characterization of residual gunpowder recovered from articles of clothing ut...
 
Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...
Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...
Cannabis Analyzer for Potency Provides Effortless Determination of Cannabinoi...
 

Recently uploaded

BUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdf
BUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdfBUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdf
BUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdfWildaNurAmalia2
 
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...lizamodels9
 
Pests of safflower_Binomics_Identification_Dr.UPR.pdf
Pests of safflower_Binomics_Identification_Dr.UPR.pdfPests of safflower_Binomics_Identification_Dr.UPR.pdf
Pests of safflower_Binomics_Identification_Dr.UPR.pdfPirithiRaju
 
Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologycaarthichand2003
 
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxGenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxBerniceCayabyab1
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentationtahreemzahra82
 
《Queensland毕业文凭-昆士兰大学毕业证成绩单》
《Queensland毕业文凭-昆士兰大学毕业证成绩单》《Queensland毕业文凭-昆士兰大学毕业证成绩单》
《Queensland毕业文凭-昆士兰大学毕业证成绩单》rnrncn29
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.PraveenaKalaiselvan1
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPirithiRaju
 
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)Columbia Weather Systems
 
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)riyaescorts54
 
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxLIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxmalonesandreagweneth
 
Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024AyushiRastogi48
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPirithiRaju
 
Transposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.pptTransposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.pptArshadWarsi13
 
Citronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyayCitronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyayupadhyaymani499
 
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...Universidade Federal de Sergipe - UFS
 
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...Universidade Federal de Sergipe - UFS
 
Microphone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptxMicrophone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptxpriyankatabhane
 

Recently uploaded (20)

BUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdf
BUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdfBUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdf
BUMI DAN ANTARIKSA PROJEK IPAS SMK KELAS X.pdf
 
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
Best Call Girls In Sector 29 Gurgaon❤️8860477959 EscorTs Service In 24/7 Delh...
 
Pests of safflower_Binomics_Identification_Dr.UPR.pdf
Pests of safflower_Binomics_Identification_Dr.UPR.pdfPests of safflower_Binomics_Identification_Dr.UPR.pdf
Pests of safflower_Binomics_Identification_Dr.UPR.pdf
 
Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technology
 
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxGenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentation
 
《Queensland毕业文凭-昆士兰大学毕业证成绩单》
《Queensland毕业文凭-昆士兰大学毕业证成绩单》《Queensland毕业文凭-昆士兰大学毕业证成绩单》
《Queensland毕业文凭-昆士兰大学毕业证成绩单》
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
 
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
 
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
 
Volatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -IVolatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -I
 
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxLIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
 
Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
 
Transposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.pptTransposable elements in prokaryotes.ppt
Transposable elements in prokaryotes.ppt
 
Citronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyayCitronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyay
 
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
 
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
REVISTA DE BIOLOGIA E CIÊNCIAS DA TERRA ISSN 1519-5228 - Artigo_Bioterra_V24_...
 
Microphone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptxMicrophone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptx
 

LC-MS/MS Separation and Quantitation of Oxysterol Compounds

  • 1. Separation and Quantitation of Oxysterol Compounds Using LC-MS/MS Shimadzu Scientific Instruments, Columbia, MD, USA
  • 2. Summary A mixture of 16 oxysterols and cholesterol related compounds were separated using a Shimadzu Nexera UHPLC and analyzed by a Shimadzu LCMS-8060 system. Detection and quantitation limits were obtained using APCI, ESI, and Dual Ion (DUIS) sources.
  • 3. Background There is high demand for oxysterol quantitation due to their correlation with neurodegenerative diseases. The ratios of various oxysterols in biological fluids are used by researchers to study disease states. An LCMS oxysterol quantitation method was developed using a Shimadzu LCMS-8060. Detection and quantitation limits were determined using multiple reaction monitoring (MRM) mode for each analyte.
  • 4. Method Oxysterol standards were obtained in methanol and diluted with 10:90 H2O:MeOH. Standards included the compounds indicated in Table 1. An MRM method was determined and optimized using LabSolutions 5.82 on a Shimadzu triple quadrupole mass spectrometer (LCMS-8060). A Shim-pack XR-ODS III column was used for the separation. Injection volume was 2 µL with autosampler temperature being set at 10°C. Flow rate was initially at 0.4 mL/min and adjusted to 0.7 mL/min toward the end of the run to speed the elution of highly retained compounds. Compound name Abbreviation 24 (S)-hydroxycholesterol 24HC (D7)22-hydroscycholesterol 22HC(d7) 25-hydroxycholesterol 25HC 27-hydroxycholesterol 27HC (D7)7α-hydroxycholesterol 7αHC(d7) 7α-hydroxycholesterol 7αHC 7β-hydroxycholesterol 7βHC 7-Ketocholesterol 7KC (D7) 7-Ketocholesterol 7KC(d7) 7α-hydroxycholestenone 7αHCn (D3) Vitamin D3 VitD3(d3) Zymosterol Zymo Desmosterol Desmo 7α,27 dihydroxycholestenone 7α,27diHC,3one Cholesterol CH 7dehydrocholesterol 7DHC Table 1. Oxysterol Related Compounds
  • 5. Experimental MRM transitions were developed and optimized on the triple quadrupole mass spectrometer using DUIS, APCI, and ESI sources. In cases where two precursor ions showed a strong signal, MRM development and optimization were done on both precursor ions. This may provide an option for better sensitivity in different matrices for future applications. The DUIS method established nebulizing, heating, and drying gas flows at 3, 17, and 3 L/min. Interface, desolvation line, and heat block temperatures were set at 300°C, 250°C, and 400°C, respectively. With the APCI source, temperatures for the interface, desolvation line, and heat block were set at 350°C, 200°C, and 200°C, respectively. Drying gas flow rate was 5 L/min. For ESI, nebulizing, heating, and drying gas flows were set at 3, 10, and 10 L/min. Interface, desolvation line, and heat block temperatures were 300°C, 250°C, and 400°C.
  • 6. Experimental Compound Name Retention time (min) DUIS APCI ESI Target MRM Reference MRM Target MRM Reference MRM Target MRM Reference MRM 24HC 4.0 385.45>367.30 385.45>324.30 367.35>91.05 367.35>104.85 385.10>367.25 385.10>109.00 24HC 4.0 367.30>281.10 367.30>104.90 385.40>367.30 385.40>104.90 367.45>95.15 367.45>105.25 22HC(d7) 4.2 374.40>91.10 374.40>255.25 374.40>104.80 374.40>133.15 374.20>104.90 374.20>132.95 25HC 4.1 385.30>367.40 385.30>324.25 367.35>95.00 367.35>135.10 367.35>105.15 367.35>147.15 25HC 4.1 367.35>81.15 367.35>105.10 385.35>367.30 385.35>133.10 385.20>367.30 385.20>324.45 27HC 4.5 385.05>67.15 385.05>93.10 385.10>135.10 385.10>149.10 385.15>95.05 385.15>324.35 7αHC(d7) 7.6 374.35>159.15 374.35>91.15 374.40>145.30 374.40>159.25 374.20>158.95 374.20>144.80 7αHC 7.7 367.35>117.10 367.35>66.90 367.45>144.95 367.45>95.00 367.15>145.40 367.15>159.35 7αHC 7.7 385.30>367.40 385.30>367.40 385.10>367.25 385.10>159.10 385.45>367.35 385.45>159.05 7βHC 7.9 385.30>367.30 385.30>324.35 367.15>159.10 367.15>145.30 367.10>95.15 367.10>158.95 7βHC 7.9 367.35>81.10 367.35>104.85 385.10>367.30 385.10>159.00 385.10>367.15 385.10>158.95 7KC 8.3 401.15>95.10 401.15>383.30 401.10>80.95 401.10>383.35 401.45>95.30 401.45>383.30 7KC(d7) 8.1 408.30>390.15 408.30>95.15 408.40>81.15 408.40>95.25 408.40>390.25 408.40>95.05 7αHCn 6.5 401.25>383.30 401.25>97.10 401.35>383.45 401.35>97.10 401.10>382.95 401.10>96.90 VitD3(d3) 3.0 386.25>232.20 386.25>368.45 386.40>368.35 386.40>92.90 386.40>368.30 386.40>95.20 Zymo 12.8 367.30>95.00 367.30>80.90 367.15>95.20 367.15>109.20 367.15>95.10 367.15>80.95 Desmo 13.3 367.25>95.05 367.25>95.05 367.35>135.25 367.35>104.90 367.15>95.00 367.15>104.95 7α,27diHC,3one 2.2 417.25>399.10 417.25>381.30 417.30>399.25 417.30>381.45 417.35>399.30 417.35>97.25 CH 17.4 369.30>161.10 369.30>94.90 369.40>95.20 369.40>146.95 369.40>147.25 369.40>135.25 7DHC 14.5 367.25>95.05 367.25>158.90 367.30>145.25 367.30>159.25 367.15>145.10 367.15>159.20 Table 2. MRM transitions of oxysterol related compounds Multiple MRM transitions were determined from each precursor ion, the two most intense are shown.
  • 7. Results Chromatographic separation was accomplished using a Shimadzu Nexera UHPLC system with a Shim-pack XR-ODS III column. Challenging separations of 24HC and 25HC as well as 7αHC and 7βHC were achieved to allow individual quantitation of these analytes. Maximum operating pressure for a typical run was about 11,000 psi. Figure 1. Chromatogram of the 16 oxysterol related compounds. 0.0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 min 0 2500000 5000000 7a,27diHC,3one VitD3(d3) 7aHCn 7βHC 7aHC 7aHC(d7) 27HC24HC CH Intensity
  • 8. Results 4.0 5.0 6.0 7.0 8.0 min 24HC 27HC 7βHC 7aHC Figure 2. Chromatogram showing separation of key components.
  • 9. Detection limits determined using ESI, APCI, and DUIS sources are shown below. Limits of detection (LOD) and limits of quantitation (LOQ) were determined by a signal to noise ratio of 3:1 and 10:1, respectively. Results LOD LOQ LOD LOQ LOD LOQ 24 (S)-hydroxycholesterol 20 30 30 50 20 50 (D7) 22-hydroscycholesterol 10 20 30 50 10 20 25-hydroxycholesterol 20 30 50 100 20 30 27-hydroxycholesterol 20 50 50 50 20 100 (D7) 7a-hydroxycholesterol 20 20 10 10 10 10 7a-hydroxycholesterol 4 10 10 10 10 20 7b-hydroxycholesterol 10 10 10 30 10 20 7-Ketocholesterol 1 4 4 10 1 4 (D7) 7-Ketocholesterol 0.5 2 2 4 0.5 2 7a-hydroxycholestenone 0.5 4 4 4 1 4 (D3) Vitamin D3 30 300 50 50 20 30 Zymosterol 100 >2000 30 300 300 300 Desmosterol 300 >2000 30 50 100 300 7a,27 dihydroxycholestenone 0.25 1 2 4 0.5 2 Cholesterol 100 1000 100 500 100 500 7dehydrocholesterol 100 300 100 300 100 500 ESI APCI DUIS Table 3. LODs and LOQs determined by LCMS-8060 with ESI, APCI, and DUIS sources. Values shown are picograms on-column. Green highlighting indicates lower or lowest limits. Yellow highlighting indicates those limits which are somewhat higher than the lowest limits. Red highlighting indicates the highest limits.
  • 10. Results Although ESI was able to obtain low detection limits for most oxysterol related compounds, it had difficulty reaching practical quantitation limits for zymosterol and desmosterol, even in neat standards. APCI allowed better quantitation for these two compounds; however, the LODs and LOQs for many other compounds were sacrificed. DUIS offers the benefit of both APCI and ESI by allowing quantitation of zymosterol and desmosterol without significantly sacrificing the LODs of other compounds. Therefore, DUIS is the optimal source for analyzing this oxysterol related mixture. The dual ion DUIS source is able to quantify all sixteen analytes in a single run instead of two runs using ESI and APCI sources separately. In the case of (D3) Vitamin D3, detection and quantitation was even lower using DUIS vs. APCI and ESI. This is likely the result of probe position optimization, which is possible with DUIS (and ESI) but not with APCI.
  • 11. Results MRM transitions for sixteen oxysterol related compounds were identified and optimized using the LCMS-8060. A UHPLC column was used to obtain separation. 24HC, 25HC, 7αHC, and 7βHC were chromatographically separated in a manner sufficient for individual quantitation. The total run time for this method was 21 minutes. Limits of detection and limits of quantitation for oxysterol related compounds were determined using APCI, ESI, and DUIS sources. Although LOD and LOQ vary per compound, the LCMS-8060 was able to reach detection limits in the low picogram (mass on-column) range for oxysterol related compounds. DUIS was shown to be the optimal source for this oxysterol related compound mixture analysis. It demonstrates the advantages of both ESI and APCI and allows all sixteen compounds to be analyzed in a single run.
  • 12. Conclusion A fast and sensitive method using a triple quadrupole mass spectrometer was developed to assist future applications for oxysterol related compound quantitation.
  • 13. Need More Info? Thank you for viewing this presentation. Should you have any questions or require additional information about our research, products, or services, please visit our Web site at: www.ssi.shimadzu.com Follow us on Twitter @shimadzussi