Chapter 11
Atomic Mass
Spectrometry
Atomic Mass Spectrometry
Atomic mass spectrometry offers A number of
advantages over the atomic optical spectrometric
methods that we have thus far considered, including
(1)detection limits that are, for many elements, as
great as three orders of magnitude better than
optical methods; (2) remarkably Simple . spectra
that are usually unique and often easily
interpretable (3) the ability to measure atomic
isotopic ratios. Disadvantages include (1)
instrument costs that are two to three times that of
optical atomic instruments, (2) instrument drift that
can be as high as 5% to 10% per hour (3) certain
types of interference effects that are discussed later.
11A
Atomic Mass
Spectrometry
SOME GENERAL FEATURES OF
ATOMIC MASS SPECTROMETRY
An atomic mass spectrometric analysis 1 involves the
following steps: (1) atomization, (2) conversion of a
substantial fraction of the atoms formed in step I to
a stream of ions (usually singly charged positive
ions), (3) separating the ions formed in step 2 on the
basis of their mass-to-charge ratio (m/z), where m is
the mass number of the ion in atomic mass units and
z is the number of fundamental charges that it bears,
and (4) counting the number of ions of each type or
measuring the ion current produced when the ions
formed from the sample strike a suitable transducer.
11A-1
Atomic Masses
in Mass Spectrometry
Atomic Masses
in Mass Spectrometry
Atomic and molecular masses are generally
expressed in atomic mass units (amu), or
daltons (Da).
Uses of Mass Spec
 forms ions, usually positive, study
charge/mass ratio
 very characteristic fragmentation
pattern in charge/mass ratio
 data easier to interpret than IR and/or
NMR
 provides accurate MW of sample
 used to determine isotopic abundances
 Nearly all elements in the periodic table can
be determined by mass spectrometry
 More selective and sensitive than optical
instruments
 Simple spectra
 Isotope ratios
 Much more expensive instrumentation
11A-2
Mass-to-Charge Ratio
Mass-to-Charge Ratio
The mass-to-charge ratio of an ion is the
unitless ratio of its mass number to the
number of fundamental charges z on the
ion. Thus: For 12C1H4
+
m/z = 16.0313/1 =16.0313
For 13C1H4
2+
m/z = 17.0346/2 = 8.5173
11A-3
Types ofAtomic Mass
Spectrometry
Types ofAtomic mass spectrometers
11B
mass spectrometer
mass spectrometer
A mass spectrometer is an instrument that produces
ions and separates them according to their mass-to-
charge ratios, m/z. Most of the ions we will discuss
are singly charged so that the ratio is simply equal
to the mass number of the ion. Several types of mass
spectrometers are currently available from
instrument manufacturers. In this chapter, we
describe the three types that are used in atomic mass
spectrometry: the quadrupole mass spectrometer, the
time-af-flight mass spectrometer, and the double-
focusing mass spectrometer.
Components of Mass Spec
Inlet
System
Ion
Source
Mass
Analyzer
Detector
Sample
Signal
Processor
Readout
10-5 - 10-8 torr
Vacuum
System
Wavelength selector
Atomic mass spectrometer
AMS
MMS
11 B-1
Transducers for Mass
Spectrometry
Transducers for Mass Spectrometry
Several types of transducers are
commercially available for mass
spectrometers. The electron multiplier is
the transducer of choice for most routine
experiments.
electron multiplier
electron multiplier
Figure 11-2a is a schematic of a discrete-
dynode electron multiplier designed for
collecting and converting positive ions into
an electrical signal. This device is very much
like the photomultiplier transducer for
ultraviolet-visible radiation, with each
dynode held at a successively higher voltage.
When energetic ions or electrons strike the
Cu-Be surfaces of the cathode and the
dynodes, bursts of electrons are emitted.
The Faraday Cup
The Faraday Cup
Figure 11-3 is a schematic of a Faraday cup collector. The
transducer is aligned so that ions exiting the analyzer strike the
collector electrode. This electrode is
surrounded by a cage that prevents the escape of reflected ions
and ejected secondary electrons. The collector
electrode is inclined with respect to the path of the entering ions
so that particles striking or leaving the electrode are reflected
from the entrance to the
cup. The collector electrode and cage are connected to ground
through a large resistor. The charge of the positive
ions striking the plate is neutralized by a flow of electrons from
ground through the resistor.
FIGURE 11-3 Faraday cup detector. The
voltage on the ion suppressor plates is
adjusted to minimize differential response
as a function of mass.
Detectors: Electron multipliers and Faraday Cup
11 B-2
Quadrupole MassAnalyzers
Quadrupole MassAnalyzers
The most common type of mass spectrometer
used in atomic mass spectroscopy is the
quadrupole mass analyzer shown in Figure
11-6. This instrument is more compact, less
expensive, and more rugged than most other
types of mass spectrometers. It also has the
advantage of high scan rates so that an entire
mass spectrum can be obtained in less than
100 ms.
Mass analyzers
1. Quadrupole mass analyzer
2. Time-of-flight mass analyzers
3. Double Focus Mass analyzers
4. Single Focus Mass analyzers
5. Ion Trap Analyzer (FT Mass spectrometry)
ICPMS
Molecular Mass spectrometers
Mass Analyzer (Quadrupole)
Skoog et al., 1999, Instrumental Analysis
Two pairs of rods:
Attach + and - sides of a
variable dc source
Apply variable radio-
frequency ac potentials to
each pair of rods.
Ions are accelerated into
the space between the rods by
a small potential (5-10V)
Ions having a limited range
of m/z value reach the
transducer.
Ion Trajectories in a quadrupole
Ion Trajectories in a quadrupole
To understand the filtering capability of a quadrupole,
we need to consider the effect of the dc and ac voltages
on the trajectory of ions as they pass through the channel
between the rods. Let us first focus on the pair of
positive rods, which are shown in Figure 11-7 as lying
In the xz plane. In the absence of a dc voltage, ions in
the channel will tend to converge in the center of the
channel during the positive half of the ac cycle and will
tend to diverge during the negative half. This behavior is
illustrated at points A and B in the figure. If during the
negative half cycle an ion strikes the rod, the positive
charge will be neutralized, and the resulting molecule
will be carried away. Whether a positive ion strikes the
rod depends on the rate of movement of the ion along
the z-axis, its mass-to-charge ratio. And the frequency
and magnitude of the ac signal.
Ion trajectories in a quadrupole
 A pair of positive rods (as lying in the xz plane).
In the absence of a dc potential:
 Positive half of the ac cycle: Converge (ion in the channel
will tend to converge in the center of the channel during the
positive half of the ac cycle).
 Negative half of the ac cycle: Diverge (ions will tend to
diverge during the negative half).
Skoog et al., 1999, Instrumental Analysis
Whether or not a positive ion strikes
the rod will depend upon the rate of
movement of ion along the z axis, its
m/z, and the frequency and magnitude
of the ac signal.
~
AC
F = E*ze (E: electric field intensity)
F = am (a: acceleration, m: mass)
Pass high-M
Pass low-M
Like a band-pass filter
Scanning with a quadrupole Filter
Scanning with a quadrupole Filter
To scan a mass spectrum with a quadrupole
instrument, the ac voltage V and the dc
voltage U are increased simultaneously from
zero to some maximum value while their
ratio is maintained at slightly less than 6. The
changes in voltage during a typical scan are
shown in Figure 11-9. The two diverging
straight lines show the variation in the two dc
voltages as a function of time.
Scanning with a quadrupole Filter
FIGURE 11-9 Voltage relationships during a
mass scan with a quadrupole analyzer.
11 B-3
Time of FlightAnalyzers
Time of FlightAnalyzers
In time-of-flight (TOF) instruments, positive ions
are produced periodically by bombardment of the
sample with brief pulses of electrons, secondary
ions, or laser generated photons. These pulses
typically have a frequency of 10 to 50 kHz and a
lifetime of 0.25 ms. The ions produced in this
way are then accelerated by an electric field
pulse of 103 to 10' V that has the same frequency
as, but lags behind, the ionization pulse. The
accelerated particles pass into a field-free drift
tube about a meter long (Figure 11-10).
Time of FlightAnalyzers
 non-magnetic separation
 detector - electron multiplier tube
 instantaneous display of results
Example: (a) calculate the kinetic energy that a singly charged ion
(z=1) Will Acquired if it is accelerated through a potential of 103 V in
an electron-Impact source. (b) Does the kinetic energy is depend
upon its mass? (c) Does the velocity of the ion depend upon its mass?
11 B-4
Double-FocusingAnalyzers
Double-FocusingAnalyzers
As shown in Figure 11-11, a double-focusing
mass spectrometer contains two devices
for focusing a beam of ions: an
electrostatic analyzer and a magnetic
sector analyzer.
Mattacuh-Herzog type double-
focusing mass spectrometer
11C
INDUCTIVELY COUPLED
PLASMA
MASS SPECTROMETRY
INDUCTIVELYCOUPLED PLASMA
MASS SPECTROMETRY
Since the early 1980, ICPMS has grown to
be one of the most important techniques
for elemental analysis because of its low
detection limits for most elements, its high
degree of selectivity, and its reasonably
good precision and accuracy. In these
applications an ICP torch serves as an
atomizer and ionizer.
Instruments for
ICPMS
11 C-1
Instruments for ICPMS
Figure 11-12 shows schematically the
components of a commercial ICPMS
systemI2 A critical part of the instrument
is the interface that couples the ICP torch,
which operates at atmospheric pressure
with the mass spectrometer that requires a
pressure of less than 10-4 torr.
ICP-MS: a handy tool!
Typical mass spectrum
11C-2
Atomic Mass Spectra
and Interferences
Atomic Mass Spectra
and Interferences
One of the advantages of using mass
spectrometric detection with ICPs as
opposed to optical detection is that mass
spectra are usually much simpler and
easier to interpret than corresponding
optical spectra, This property is especially
true for those elements such as the rare
earths that may exhibit thousands of
emission lines.
Spectral Interferences
Spectral Interferences?
 Refractory oxide
As a result of incomplete dissociation of the sample
matrix or from recombination in the plasma tail
MO+, MO2+, MO3+
 Doubly charged ions
Spectral Interferences?
 Isobaric overlap
Due to two elements that have isotopes having
substantially the same mass
40Ar+ and 40Ca+
 Poly atomic
Due to interactions between species in the plasma
and species in matrix or atmosphere
56Fe and 40Ar16O
44Ca and 12C16O16O.
Isobaric interferences?
Matrix Effects
Matrix Effects
In ICPMS, matrix effects become
noticeable at concomitant concentrations
of greater than about 500 to 1000 mg/mL.
Usually these effects cause a reduction in
the analyte signal, although under certain
experimental conditions signal
enhancement is observed.
Applications of
ICPMS
11C-3
Applications of ICPMS
ICPMS can be used for qualitative,
semiquantitative, and quantitative
determination of one or more elements in
samples of matter.
Qualitative and Semiquantitative
Applications
Qualitative and Semiquantitative
Applications
FIGURE11-16 ICPMS spectrum for the rare earth elements. Solutions
contain 1 mg/mL of each element. (From K. E. Jarvis, J. Anal. Atom.
Spectrum., 1989, 4, 563. With permission.)
Detection Limits
Detection Limits
One of the main attractions of ICPMS lies
with the lower detection limits attainable
with mass spectrometric detection than
with optical detection. These limits in
many cases equal and sometimes exceed
those that can be realized by
electrothermal atomic absorption
methods.
Detection Limits
FIGURE11-17 Detection limits for selected elements by ICPMS (black bars)
compared with those for ICP-OES (blue bars) and ETMS (gray bars), plotted
on a logarithmic scale in concentrations of ppb (or mg/L).
Because ETMS detection limits are inherently in mass units (pg), they have
been converted to concentration by assuming a 20-JLLsample. (From M. Selby
and G. M. Hieftje,Amer. Lab., 1987 (8),20. With permission.)
quantitative Analyses
quantitative Analyses
The most widely used quantitative method
of ICPMSuses a set of calibration
standards for preparing a calibrationcurve.
Simple aqueous standards are usually
adequate if the unknown solutions are
sufjiciently dilute - less than 2000 g/mL of
total dissolved solid.
quantitative Analyses
How good this is?
Linear calibrations over 4 orders of magnitude Multi-elemental analysis of a standard
Isotope Ratio
Measurements
Isotope Ratio Measurements
The measurement of isotope ratios is of
considerable importance in several fields
of science and medicine. For example,
archeologists and geologists use such data
to establish the age of artifacts and various
types of deposits. Chemists and clinicians
use isotopically enriched materials as
tracers in various types of studies. The
outcome of these studies is based on
isotope ratio measurements.
SPARK SOURCE
MASS
SPECTROMETY
11 D
SPARK SOURCE MASS SPECTROMETRY
In SSMS. the atomic constituents of a sample are
converted by a high-voltage (-30 kv), radio-frequency
spark to gaseous ions for mass analysis. The spark is
housed in a vacuum chamber located immediately
adjacent to the mass analyzer. The chamber is
equipped with a separate high-speed pumping system
that quickly reduces the internal pressure to about 10-8
torr after sample changes. Often, the sample serves as
one or both electrodes. Alternatively. it is mixed with
graphite and loaded into a cup-shape electrode. The
gaseous positive ions formed in the spark plasma arc
drawn into the analyzer by a de voltage.
spectra
11 D-1
spectra
Like ICP mass spectra, spark source mass
spectra are much simpler than atomic
emission spectra, consisting of one major
peak for each isotope of an element as
well as a few weaker lines corresponding
to multiply charged ions and ionized oxide
and hydroxide species. The presence of
these additional ions creates the potential
for interference just as in ICPMS.
Qualitative
Applications
11 D-2
QualitativeApplications
SSMS is a powerful tool for qualitative and
semiquantitaative analysis. All elements in the
periodic table from 7Li through 238Ucan be identified
in a single excitation. By varying data-acquisition
parameters, it is possible to determine order of
magnitude concentrations for major constituents of a
sample as well as for constituents in the parts-per-
billion concentration range. Interpretation of spectra
does require skill and experience, however, because
of the presence of multiply charged species,
polymeric species, and molecular ions.
Quantitative Applications
A radio-frequency spark is not a very
reproducible source over short periods. As
a consequence. it is necessary to integrate
the output signals from a spark for periods
that range from several seconds to
hundreds of seconds if good quantitative
data are to be obtained .The detection
system must be capable of electronic
signal integration,
GLOW DISCHARGE
MASS
SPECTROMETY
11 E
GLOW DISCHARGE MASS SPECTROMETRY
As shown in Section 8C-2, a glow-discharge
source is used as an atomization device for various
types of atomic spectroscopy.In addition to
atomizing samples, it also produces a cloud of
positive analyte ions from solid samples. This
device consists of a simple two electrode closed
system containing argon at a pressure of 0.1 to
10torr. A voltage of 5 to 15 kv from a pulsed dc
power supply is applied between the electrodes,
causing the formation of positive argon ions,
which are then accelerated toward the cathode.
Methods for
elemental surface
analysis
11 F
Methods for elemental surface analysis
Two atomic mass spectrometric methods
are often used to determine the elemental
composition of solid surfaces: secondary-
ion mass spectrometry and laser
microprobe mass spectrometry.
Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International
University
Updated on 9/13/2006 Chapter 3 ICPMS
Isotope Dilution
 Isotope dilution is a super internal standard
addition method on the basis of isotope
ratios.
 Add a known amount (spike) of a stable
enriched isotope of the element considered,
which has at least two stable isotopes 1 and
2, to the sample
 Measure the isotope ratio of isotopes 1 and 2
in the Spike, the unspiked sample and finally
the spiked sample.
 The concentration of the element of interest
can then be deducted from these isotopic
ratios and from the amount of spike added.
Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International
University
Updated on 9/13/2006 Chapter 3 ICPMS
 Advantages:
 Simplified chemical and physical separation
procedures
 Elimination (reduction) of matrix effects
 Elimination of the effect of instrumental drift
Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International
University
Updated on 9/13/2006 Chapter 3 ICPMS
 Theory
In principle, any element with at least two
isotopes that can be measured is suitable for
determination by isotope dilution.The two
selected are designed 1 and 2.
Three solutions will be used:
Sample (s) Standard (t) Spiked
sample (m)
Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International
University
Updated on 9/13/2006 Chapter 3 ICPMS
 1ns is the number of moles of isotope 1 in the
sample.
 2ns is the number of moles of isotope 2 in the
sample.
 1nt is the number of moles of isotope 1 in the
standard.
 2nt is the number of moles of isotope 2 in the
standard.
 Rs is the ratio of isotope 1 to isotope 2 in the
sample solution.
 Rt is the ratio of isotope 1 to isotope 2 in the
standard.
 Rm is the ratio of isotope 1 to isotope 2 in the
spiked sample.
MassAnalyzer
Double-Focusing Analyzers
 higher resolution, need higher amplification
 2 magnets or 1 magnet & 1 electrostatic field

More Related Content

What's hot

instrumentation of mass spectrometry
instrumentation of mass spectrometryinstrumentation of mass spectrometry
instrumentation of mass spectrometryManali Parab
 
Auger Electron Spectroscopy
Auger Electron SpectroscopyAuger Electron Spectroscopy
Auger Electron Spectroscopyhasanjamal13
 
X ray powder diffraction
X ray powder diffractionX ray powder diffraction
X ray powder diffractionchiranjibi68
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopymanishasoral
 
Principle and working of Nmr spectroscopy
Principle and working of Nmr spectroscopyPrinciple and working of Nmr spectroscopy
Principle and working of Nmr spectroscopyArpitSuralkar
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyUlivin Al Farisi
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyChetan Sharma
 
Instrumentation presentation - Auger Electron Spectroscopy (AES)
Instrumentation presentation - Auger Electron Spectroscopy (AES)Instrumentation presentation - Auger Electron Spectroscopy (AES)
Instrumentation presentation - Auger Electron Spectroscopy (AES)Amirah Basir
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy AyeshaRasty
 
NMR, principle and instrumentation by kk sahu sir
NMR, principle and instrumentation by kk sahu sirNMR, principle and instrumentation by kk sahu sir
NMR, principle and instrumentation by kk sahu sirKAUSHAL SAHU
 
MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...
MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...
MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...Sachin Kale
 
Photo Electron Spectroscopy
Photo Electron SpectroscopyPhoto Electron Spectroscopy
Photo Electron SpectroscopyRadha Mini
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyAnshul2593
 

What's hot (20)

instrumentation of mass spectrometry
instrumentation of mass spectrometryinstrumentation of mass spectrometry
instrumentation of mass spectrometry
 
Atomic absorption spectrophotometer
Atomic absorption spectrophotometerAtomic absorption spectrophotometer
Atomic absorption spectrophotometer
 
Auger Electron Spectroscopy
Auger Electron SpectroscopyAuger Electron Spectroscopy
Auger Electron Spectroscopy
 
X ray powder diffraction
X ray powder diffractionX ray powder diffraction
X ray powder diffraction
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic absorption spectroscopy.pptx
Atomic absorption spectroscopy.pptxAtomic absorption spectroscopy.pptx
Atomic absorption spectroscopy.pptx
 
Principle and working of Nmr spectroscopy
Principle and working of Nmr spectroscopyPrinciple and working of Nmr spectroscopy
Principle and working of Nmr spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic absorption spectrometry (aas)
Atomic absorption spectrometry (aas)Atomic absorption spectrometry (aas)
Atomic absorption spectrometry (aas)
 
XRD
XRD XRD
XRD
 
Instrumentation presentation - Auger Electron Spectroscopy (AES)
Instrumentation presentation - Auger Electron Spectroscopy (AES)Instrumentation presentation - Auger Electron Spectroscopy (AES)
Instrumentation presentation - Auger Electron Spectroscopy (AES)
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Flame photometer
Flame photometerFlame photometer
Flame photometer
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
NMR, principle and instrumentation by kk sahu sir
NMR, principle and instrumentation by kk sahu sirNMR, principle and instrumentation by kk sahu sir
NMR, principle and instrumentation by kk sahu sir
 
MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...
MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...
MASS SPECTROSCOPY ( Molecular ion, Base peak, Isotopic abundance, Metastable ...
 
Photo Electron Spectroscopy
Photo Electron SpectroscopyPhoto Electron Spectroscopy
Photo Electron Spectroscopy
 
NUCLEAR QUADRUPOLE RESONANCE SPECTROSCOPY
NUCLEAR QUADRUPOLE RESONANCE SPECTROSCOPY NUCLEAR QUADRUPOLE RESONANCE SPECTROSCOPY
NUCLEAR QUADRUPOLE RESONANCE SPECTROSCOPY
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 

Similar to Chapter-11-Atomic-Mass-Spectrometry (1).ppt

mass spectroscopy presentation.pptx
mass spectroscopy presentation.pptxmass spectroscopy presentation.pptx
mass spectroscopy presentation.pptxAsefakushkiwala
 
4R6Notes3_GasFilled_Detectors.pdf
4R6Notes3_GasFilled_Detectors.pdf4R6Notes3_GasFilled_Detectors.pdf
4R6Notes3_GasFilled_Detectors.pdfAnjaliVcew
 
mass spectro pres.pptx
mass spectro pres.pptxmass spectro pres.pptx
mass spectro pres.pptxbreenaawan
 
Chapter 4 electronic
Chapter 4 electronicChapter 4 electronic
Chapter 4 electronicAsrar Humaira
 
Instrumentation of Mass Spectrometry
 Instrumentation of Mass Spectrometry Instrumentation of Mass Spectrometry
Instrumentation of Mass SpectrometrySowmya B U Jain
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopylavanya0238
 
Introduction of spectroscopy
Introduction of spectroscopyIntroduction of spectroscopy
Introduction of spectroscopyZainab&Sons
 
Mass spectroscopy by Shantanu Nimbalkar
Mass spectroscopy by Shantanu NimbalkarMass spectroscopy by Shantanu Nimbalkar
Mass spectroscopy by Shantanu NimbalkarShantanu Nimbalkar
 
Mass spectroscopy.pptx
Mass spectroscopy.pptxMass spectroscopy.pptx
Mass spectroscopy.pptxshanes8
 
Quadrupole and Time of Flight Mass analysers.
 Quadrupole and Time of Flight Mass analysers. Quadrupole and Time of Flight Mass analysers.
Quadrupole and Time of Flight Mass analysers.Gagangowda58
 
Analyser of Quadrupole and time of flight.pptx
   Analyser of Quadrupole and time of flight.pptx   Analyser of Quadrupole and time of flight.pptx
Analyser of Quadrupole and time of flight.pptxnishenandansuryawans
 
Mass spectroscopy -> by Mohit kumar
Mass spectroscopy -> by Mohit kumarMass spectroscopy -> by Mohit kumar
Mass spectroscopy -> by Mohit kumarMohit Kumar
 
Mass Spectrometry
Mass SpectrometryMass Spectrometry
Mass SpectrometryPujanSasmal
 
Chapter2-Structure of Atom 2023.pptx
Chapter2-Structure of Atom 2023.pptxChapter2-Structure of Atom 2023.pptx
Chapter2-Structure of Atom 2023.pptxValarmathiRajendran1
 
Mass principle FOR PG PHARMACEUTICAL ANALYSIS
Mass principle FOR PG PHARMACEUTICAL ANALYSISMass principle FOR PG PHARMACEUTICAL ANALYSIS
Mass principle FOR PG PHARMACEUTICAL ANALYSISprakash64742
 

Similar to Chapter-11-Atomic-Mass-Spectrometry (1).ppt (20)

mass spectrometery
mass spectrometerymass spectrometery
mass spectrometery
 
Mass Analyser my ppt.pptx
Mass Analyser my ppt.pptxMass Analyser my ppt.pptx
Mass Analyser my ppt.pptx
 
mass spectroscopy presentation.pptx
mass spectroscopy presentation.pptxmass spectroscopy presentation.pptx
mass spectroscopy presentation.pptx
 
4R6Notes3_GasFilled_Detectors.pdf
4R6Notes3_GasFilled_Detectors.pdf4R6Notes3_GasFilled_Detectors.pdf
4R6Notes3_GasFilled_Detectors.pdf
 
mass spectro pres.pptx
mass spectro pres.pptxmass spectro pres.pptx
mass spectro pres.pptx
 
Chapter 4 electronic
Chapter 4 electronicChapter 4 electronic
Chapter 4 electronic
 
Instrumentation of Mass Spectrometry
 Instrumentation of Mass Spectrometry Instrumentation of Mass Spectrometry
Instrumentation of Mass Spectrometry
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Introduction of spectroscopy
Introduction of spectroscopyIntroduction of spectroscopy
Introduction of spectroscopy
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Mass spectroscopy by Shantanu Nimbalkar
Mass spectroscopy by Shantanu NimbalkarMass spectroscopy by Shantanu Nimbalkar
Mass spectroscopy by Shantanu Nimbalkar
 
mass spectrOSCOPY
 mass spectrOSCOPY mass spectrOSCOPY
mass spectrOSCOPY
 
Mass spectroscopy.pptx
Mass spectroscopy.pptxMass spectroscopy.pptx
Mass spectroscopy.pptx
 
Quadrupole and Time of Flight Mass analysers.
 Quadrupole and Time of Flight Mass analysers. Quadrupole and Time of Flight Mass analysers.
Quadrupole and Time of Flight Mass analysers.
 
Analyser of Quadrupole and time of flight.pptx
   Analyser of Quadrupole and time of flight.pptx   Analyser of Quadrupole and time of flight.pptx
Analyser of Quadrupole and time of flight.pptx
 
Mass spectroscopy -> by Mohit kumar
Mass spectroscopy -> by Mohit kumarMass spectroscopy -> by Mohit kumar
Mass spectroscopy -> by Mohit kumar
 
Mass Spectrometry
Mass SpectrometryMass Spectrometry
Mass Spectrometry
 
Chapter2-Structure of Atom 2023.pptx
Chapter2-Structure of Atom 2023.pptxChapter2-Structure of Atom 2023.pptx
Chapter2-Structure of Atom 2023.pptx
 
Mass principle FOR PG PHARMACEUTICAL ANALYSIS
Mass principle FOR PG PHARMACEUTICAL ANALYSISMass principle FOR PG PHARMACEUTICAL ANALYSIS
Mass principle FOR PG PHARMACEUTICAL ANALYSIS
 
Ap unit 1
Ap unit 1Ap unit 1
Ap unit 1
 

Recently uploaded

Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfagholdier
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphThiyagu K
 
Role Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptxRole Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptxNikitaBankoti2
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfJayanti Pande
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.pptRamjanShidvankar
 
Making and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfMaking and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfChris Hunter
 
Python Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docxPython Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docxRamakrishna Reddy Bijjam
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfPoh-Sun Goh
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...Nguyen Thanh Tu Collection
 
Energy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural Resources
Energy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural ResourcesEnergy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural Resources
Energy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural ResourcesShubhangi Sonawane
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.christianmathematics
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxDenish Jangid
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxVishalSingh1417
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxVishalSingh1417
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdfQucHHunhnh
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...christianmathematics
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 

Recently uploaded (20)

Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
Asian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptxAsian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptx
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
Role Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptxRole Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptx
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
Making and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfMaking and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdf
 
Python Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docxPython Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docx
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
Energy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural Resources
Energy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural ResourcesEnergy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural Resources
Energy Resources. ( B. Pharmacy, 1st Year, Sem-II) Natural Resources
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 

Chapter-11-Atomic-Mass-Spectrometry (1).ppt

  • 2. Atomic Mass Spectrometry Atomic mass spectrometry offers A number of advantages over the atomic optical spectrometric methods that we have thus far considered, including (1)detection limits that are, for many elements, as great as three orders of magnitude better than optical methods; (2) remarkably Simple . spectra that are usually unique and often easily interpretable (3) the ability to measure atomic isotopic ratios. Disadvantages include (1) instrument costs that are two to three times that of optical atomic instruments, (2) instrument drift that can be as high as 5% to 10% per hour (3) certain types of interference effects that are discussed later.
  • 4. SOME GENERAL FEATURES OF ATOMIC MASS SPECTROMETRY An atomic mass spectrometric analysis 1 involves the following steps: (1) atomization, (2) conversion of a substantial fraction of the atoms formed in step I to a stream of ions (usually singly charged positive ions), (3) separating the ions formed in step 2 on the basis of their mass-to-charge ratio (m/z), where m is the mass number of the ion in atomic mass units and z is the number of fundamental charges that it bears, and (4) counting the number of ions of each type or measuring the ion current produced when the ions formed from the sample strike a suitable transducer.
  • 6. Atomic Masses in Mass Spectrometry Atomic and molecular masses are generally expressed in atomic mass units (amu), or daltons (Da).
  • 7. Uses of Mass Spec  forms ions, usually positive, study charge/mass ratio  very characteristic fragmentation pattern in charge/mass ratio  data easier to interpret than IR and/or NMR  provides accurate MW of sample  used to determine isotopic abundances
  • 8.  Nearly all elements in the periodic table can be determined by mass spectrometry  More selective and sensitive than optical instruments  Simple spectra  Isotope ratios  Much more expensive instrumentation
  • 10. Mass-to-Charge Ratio The mass-to-charge ratio of an ion is the unitless ratio of its mass number to the number of fundamental charges z on the ion. Thus: For 12C1H4 + m/z = 16.0313/1 =16.0313 For 13C1H4 2+ m/z = 17.0346/2 = 8.5173
  • 12. Types ofAtomic mass spectrometers
  • 14. mass spectrometer A mass spectrometer is an instrument that produces ions and separates them according to their mass-to- charge ratios, m/z. Most of the ions we will discuss are singly charged so that the ratio is simply equal to the mass number of the ion. Several types of mass spectrometers are currently available from instrument manufacturers. In this chapter, we describe the three types that are used in atomic mass spectrometry: the quadrupole mass spectrometer, the time-af-flight mass spectrometer, and the double- focusing mass spectrometer.
  • 15. Components of Mass Spec Inlet System Ion Source Mass Analyzer Detector Sample Signal Processor Readout 10-5 - 10-8 torr Vacuum System Wavelength selector
  • 17. 11 B-1 Transducers for Mass Spectrometry
  • 18. Transducers for Mass Spectrometry Several types of transducers are commercially available for mass spectrometers. The electron multiplier is the transducer of choice for most routine experiments.
  • 20. electron multiplier Figure 11-2a is a schematic of a discrete- dynode electron multiplier designed for collecting and converting positive ions into an electrical signal. This device is very much like the photomultiplier transducer for ultraviolet-visible radiation, with each dynode held at a successively higher voltage. When energetic ions or electrons strike the Cu-Be surfaces of the cathode and the dynodes, bursts of electrons are emitted.
  • 22. The Faraday Cup Figure 11-3 is a schematic of a Faraday cup collector. The transducer is aligned so that ions exiting the analyzer strike the collector electrode. This electrode is surrounded by a cage that prevents the escape of reflected ions and ejected secondary electrons. The collector electrode is inclined with respect to the path of the entering ions so that particles striking or leaving the electrode are reflected from the entrance to the cup. The collector electrode and cage are connected to ground through a large resistor. The charge of the positive ions striking the plate is neutralized by a flow of electrons from ground through the resistor. FIGURE 11-3 Faraday cup detector. The voltage on the ion suppressor plates is adjusted to minimize differential response as a function of mass.
  • 25. Quadrupole MassAnalyzers The most common type of mass spectrometer used in atomic mass spectroscopy is the quadrupole mass analyzer shown in Figure 11-6. This instrument is more compact, less expensive, and more rugged than most other types of mass spectrometers. It also has the advantage of high scan rates so that an entire mass spectrum can be obtained in less than 100 ms.
  • 26. Mass analyzers 1. Quadrupole mass analyzer 2. Time-of-flight mass analyzers 3. Double Focus Mass analyzers 4. Single Focus Mass analyzers 5. Ion Trap Analyzer (FT Mass spectrometry) ICPMS Molecular Mass spectrometers
  • 27. Mass Analyzer (Quadrupole) Skoog et al., 1999, Instrumental Analysis Two pairs of rods: Attach + and - sides of a variable dc source Apply variable radio- frequency ac potentials to each pair of rods. Ions are accelerated into the space between the rods by a small potential (5-10V) Ions having a limited range of m/z value reach the transducer.
  • 28. Ion Trajectories in a quadrupole
  • 29. Ion Trajectories in a quadrupole To understand the filtering capability of a quadrupole, we need to consider the effect of the dc and ac voltages on the trajectory of ions as they pass through the channel between the rods. Let us first focus on the pair of positive rods, which are shown in Figure 11-7 as lying In the xz plane. In the absence of a dc voltage, ions in the channel will tend to converge in the center of the channel during the positive half of the ac cycle and will tend to diverge during the negative half. This behavior is illustrated at points A and B in the figure. If during the negative half cycle an ion strikes the rod, the positive charge will be neutralized, and the resulting molecule will be carried away. Whether a positive ion strikes the rod depends on the rate of movement of the ion along the z-axis, its mass-to-charge ratio. And the frequency and magnitude of the ac signal.
  • 30. Ion trajectories in a quadrupole  A pair of positive rods (as lying in the xz plane). In the absence of a dc potential:  Positive half of the ac cycle: Converge (ion in the channel will tend to converge in the center of the channel during the positive half of the ac cycle).  Negative half of the ac cycle: Diverge (ions will tend to diverge during the negative half). Skoog et al., 1999, Instrumental Analysis Whether or not a positive ion strikes the rod will depend upon the rate of movement of ion along the z axis, its m/z, and the frequency and magnitude of the ac signal.
  • 31.
  • 32. ~ AC F = E*ze (E: electric field intensity) F = am (a: acceleration, m: mass) Pass high-M Pass low-M Like a band-pass filter
  • 33. Scanning with a quadrupole Filter
  • 34. Scanning with a quadrupole Filter To scan a mass spectrum with a quadrupole instrument, the ac voltage V and the dc voltage U are increased simultaneously from zero to some maximum value while their ratio is maintained at slightly less than 6. The changes in voltage during a typical scan are shown in Figure 11-9. The two diverging straight lines show the variation in the two dc voltages as a function of time.
  • 35. Scanning with a quadrupole Filter FIGURE 11-9 Voltage relationships during a mass scan with a quadrupole analyzer.
  • 36. 11 B-3 Time of FlightAnalyzers
  • 37. Time of FlightAnalyzers In time-of-flight (TOF) instruments, positive ions are produced periodically by bombardment of the sample with brief pulses of electrons, secondary ions, or laser generated photons. These pulses typically have a frequency of 10 to 50 kHz and a lifetime of 0.25 ms. The ions produced in this way are then accelerated by an electric field pulse of 103 to 10' V that has the same frequency as, but lags behind, the ionization pulse. The accelerated particles pass into a field-free drift tube about a meter long (Figure 11-10).
  • 38. Time of FlightAnalyzers  non-magnetic separation  detector - electron multiplier tube  instantaneous display of results Example: (a) calculate the kinetic energy that a singly charged ion (z=1) Will Acquired if it is accelerated through a potential of 103 V in an electron-Impact source. (b) Does the kinetic energy is depend upon its mass? (c) Does the velocity of the ion depend upon its mass?
  • 40. Double-FocusingAnalyzers As shown in Figure 11-11, a double-focusing mass spectrometer contains two devices for focusing a beam of ions: an electrostatic analyzer and a magnetic sector analyzer.
  • 43. INDUCTIVELYCOUPLED PLASMA MASS SPECTROMETRY Since the early 1980, ICPMS has grown to be one of the most important techniques for elemental analysis because of its low detection limits for most elements, its high degree of selectivity, and its reasonably good precision and accuracy. In these applications an ICP torch serves as an atomizer and ionizer.
  • 45. Instruments for ICPMS Figure 11-12 shows schematically the components of a commercial ICPMS systemI2 A critical part of the instrument is the interface that couples the ICP torch, which operates at atmospheric pressure with the mass spectrometer that requires a pressure of less than 10-4 torr.
  • 49. Atomic Mass Spectra and Interferences One of the advantages of using mass spectrometric detection with ICPs as opposed to optical detection is that mass spectra are usually much simpler and easier to interpret than corresponding optical spectra, This property is especially true for those elements such as the rare earths that may exhibit thousands of emission lines.
  • 50.
  • 52. Spectral Interferences?  Refractory oxide As a result of incomplete dissociation of the sample matrix or from recombination in the plasma tail MO+, MO2+, MO3+  Doubly charged ions
  • 53. Spectral Interferences?  Isobaric overlap Due to two elements that have isotopes having substantially the same mass 40Ar+ and 40Ca+  Poly atomic Due to interactions between species in the plasma and species in matrix or atmosphere 56Fe and 40Ar16O 44Ca and 12C16O16O.
  • 56. Matrix Effects In ICPMS, matrix effects become noticeable at concomitant concentrations of greater than about 500 to 1000 mg/mL. Usually these effects cause a reduction in the analyte signal, although under certain experimental conditions signal enhancement is observed.
  • 58. Applications of ICPMS ICPMS can be used for qualitative, semiquantitative, and quantitative determination of one or more elements in samples of matter.
  • 60. Qualitative and Semiquantitative Applications FIGURE11-16 ICPMS spectrum for the rare earth elements. Solutions contain 1 mg/mL of each element. (From K. E. Jarvis, J. Anal. Atom. Spectrum., 1989, 4, 563. With permission.)
  • 62. Detection Limits One of the main attractions of ICPMS lies with the lower detection limits attainable with mass spectrometric detection than with optical detection. These limits in many cases equal and sometimes exceed those that can be realized by electrothermal atomic absorption methods.
  • 63. Detection Limits FIGURE11-17 Detection limits for selected elements by ICPMS (black bars) compared with those for ICP-OES (blue bars) and ETMS (gray bars), plotted on a logarithmic scale in concentrations of ppb (or mg/L). Because ETMS detection limits are inherently in mass units (pg), they have been converted to concentration by assuming a 20-JLLsample. (From M. Selby and G. M. Hieftje,Amer. Lab., 1987 (8),20. With permission.)
  • 65. quantitative Analyses The most widely used quantitative method of ICPMSuses a set of calibration standards for preparing a calibrationcurve. Simple aqueous standards are usually adequate if the unknown solutions are sufjiciently dilute - less than 2000 g/mL of total dissolved solid.
  • 67. How good this is? Linear calibrations over 4 orders of magnitude Multi-elemental analysis of a standard
  • 69. Isotope Ratio Measurements The measurement of isotope ratios is of considerable importance in several fields of science and medicine. For example, archeologists and geologists use such data to establish the age of artifacts and various types of deposits. Chemists and clinicians use isotopically enriched materials as tracers in various types of studies. The outcome of these studies is based on isotope ratio measurements.
  • 71. SPARK SOURCE MASS SPECTROMETRY In SSMS. the atomic constituents of a sample are converted by a high-voltage (-30 kv), radio-frequency spark to gaseous ions for mass analysis. The spark is housed in a vacuum chamber located immediately adjacent to the mass analyzer. The chamber is equipped with a separate high-speed pumping system that quickly reduces the internal pressure to about 10-8 torr after sample changes. Often, the sample serves as one or both electrodes. Alternatively. it is mixed with graphite and loaded into a cup-shape electrode. The gaseous positive ions formed in the spark plasma arc drawn into the analyzer by a de voltage.
  • 73. spectra Like ICP mass spectra, spark source mass spectra are much simpler than atomic emission spectra, consisting of one major peak for each isotope of an element as well as a few weaker lines corresponding to multiply charged ions and ionized oxide and hydroxide species. The presence of these additional ions creates the potential for interference just as in ICPMS.
  • 75. QualitativeApplications SSMS is a powerful tool for qualitative and semiquantitaative analysis. All elements in the periodic table from 7Li through 238Ucan be identified in a single excitation. By varying data-acquisition parameters, it is possible to determine order of magnitude concentrations for major constituents of a sample as well as for constituents in the parts-per- billion concentration range. Interpretation of spectra does require skill and experience, however, because of the presence of multiply charged species, polymeric species, and molecular ions.
  • 76. Quantitative Applications A radio-frequency spark is not a very reproducible source over short periods. As a consequence. it is necessary to integrate the output signals from a spark for periods that range from several seconds to hundreds of seconds if good quantitative data are to be obtained .The detection system must be capable of electronic signal integration,
  • 78. GLOW DISCHARGE MASS SPECTROMETRY As shown in Section 8C-2, a glow-discharge source is used as an atomization device for various types of atomic spectroscopy.In addition to atomizing samples, it also produces a cloud of positive analyte ions from solid samples. This device consists of a simple two electrode closed system containing argon at a pressure of 0.1 to 10torr. A voltage of 5 to 15 kv from a pulsed dc power supply is applied between the electrodes, causing the formation of positive argon ions, which are then accelerated toward the cathode.
  • 80. Methods for elemental surface analysis Two atomic mass spectrometric methods are often used to determine the elemental composition of solid surfaces: secondary- ion mass spectrometry and laser microprobe mass spectrometry.
  • 81. Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International University Updated on 9/13/2006 Chapter 3 ICPMS Isotope Dilution  Isotope dilution is a super internal standard addition method on the basis of isotope ratios.  Add a known amount (spike) of a stable enriched isotope of the element considered, which has at least two stable isotopes 1 and 2, to the sample  Measure the isotope ratio of isotopes 1 and 2 in the Spike, the unspiked sample and finally the spiked sample.  The concentration of the element of interest can then be deducted from these isotopic ratios and from the amount of spike added.
  • 82. Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International University Updated on 9/13/2006 Chapter 3 ICPMS  Advantages:  Simplified chemical and physical separation procedures  Elimination (reduction) of matrix effects  Elimination of the effect of instrumental drift
  • 83. Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International University Updated on 9/13/2006 Chapter 3 ICPMS  Theory In principle, any element with at least two isotopes that can be measured is suitable for determination by isotope dilution.The two selected are designed 1 and 2. Three solutions will be used: Sample (s) Standard (t) Spiked sample (m)
  • 84. Advanced Analytical Chemistry – CHM 6157 ® Y. CAI Florida International University Updated on 9/13/2006 Chapter 3 ICPMS  1ns is the number of moles of isotope 1 in the sample.  2ns is the number of moles of isotope 2 in the sample.  1nt is the number of moles of isotope 1 in the standard.  2nt is the number of moles of isotope 2 in the standard.  Rs is the ratio of isotope 1 to isotope 2 in the sample solution.  Rt is the ratio of isotope 1 to isotope 2 in the standard.  Rm is the ratio of isotope 1 to isotope 2 in the spiked sample.
  • 85. MassAnalyzer Double-Focusing Analyzers  higher resolution, need higher amplification  2 magnets or 1 magnet & 1 electrostatic field