SlideShare a Scribd company logo
1 of 9
Download to read offline
Some Critical Facets in Air Pollution Monitoring
Dr. Rajendra Prasad
Managing Director
Envirotech Instruments Pvt. Ltd.
A-271, Okhla Industrial Area, Phase -1, New Delhi - 110020 (email: envirotech@vsnl.com)
1.0 BACKGROUND
The prime objective of air quality monitoring is to find out prevailing
concentration of air pollutants with accuracy of 1 Dg/m3
. This accuracy is
depend upon several factors. Man and machine are the two basic facets
equally responsible for reliable data generation which can meet aforesaid
accuracy expectations.
Qualification, training and experience of man power can improve quality of
generated data. However if monitoring personnel are motivated absolute
reliability in collected data can ensured without further efforts even in existing
setup. Similarly the monitoring instruments have to be accurate, precise and
regularly calibrated without which quality data generation is out of question.
CPCB have already specified following accuracy limits for critical parameters
recorded by a monitoring system used for AAQM.
Flow rate accurate upto + 5 %
Time accurate upto ± 1 min.
Above parameters required to be critically examined in monitoring
instruments and must be ensured within specified accuracy limits for
maintaining desired accuracy of measurements.
1.1 PRESENT SCENARIO
1. About Man power :
The manpower deployed for regular air quality monitoring either in NAAQM
or in any other monitoring program most of time are not adequately qualified
and experienced. Formal training has never been given to these staff. Most
60
critical part is that the people engaged in these monitorings are not motivated
for the assigned jobs.
Besides this there are several other serious limitations with man power like
they are overloaded with work due to less staff, have no clear cut instructions
about monitoring for acceptance and rejection of sample, have no freedom to
put there problems and view point with senior and experienced officers. At
last there salaries are neither attractive not revised since long hence
whenever there are opportunities people are escaping.
About the Machines :
Monitoring instruments used for assessment of air pollution are prone to drift
and may show variation with time in measured parameters. Hence it is a
false expectation that their calibrations are stable over long period of time.
As per standard requirements calibration need to be done.
• When instrument is manufactured.
• When instrument is transported to a new location and in different climatic
condition.
• If instrument is operated after interruption of several months.
• After repairs or overhaul of key components.
• After observing significant drift either in zero or span value or in recorded
value.
The monitoring instruments used in regular air pollution monitoring either
under NAAQM or in other project are not calibrated as per above
requirements. Even when it is known that regular calibration of monitoring
instrument is a basic need for reliable data generation.
• In general life of a HVS/RDS is about 5 years. But instruments are not
retired even after crossing age of 10 years and more with old machine
(not regularly calibrated with age of 10 years and above) reliability of data
are expected similar as expected with new calibrated machines.
61
Table below show the age distribution of installed HVS at various NAAQM
stations.
S.No. Age (years) No. of Machine
1. 10 and above 32
2. 7 - 1 0 75
3. 5 - 7 28
4. Less than 5 06
Total 141
(Source Status Report on Servicing of HVS Installed at Various NAAQM Station in India - 1995)
Over all it can be said that conditions of machines used in air quality
monitoring are critical and are not good for reliable data generation. In such
situation when conditions of man and machine are weak reliable data can not be
generated as per requirements.
1.2 Difficulties of Calibration :
Users face following difficulties while calibration of instruments.
• Calibration facilities are not available at number of places in the country.
Further all required set-ups related to the calibration of air pollution
monitoring are not available with one agency.
• Reference set up at calibrating agencies are not portable enough hence
calibration at site is not possible. Thus monitoring system need to be taken
to calibration agencies for calibration. It is not easy to transport HVS/RDS
from distant places. Besides this cost of transportation is also significant
where possibilities of damages are always there.
• Time taken for calibration is several weeks since spare machines are not
available hence calibration is postponed due to this limitation.
• Cost of calibration is very high most of the time more than the cost of
instrument to be calibrated. Some calibration costs are given below.
62
S.No, Name of the item
Cost of item
calibrator (Rs.)
Cost of calibration
(Rs.)
1.
Portable calibrator for flow
rate calibration
15,000 28,000
2. Thermometer (6 points) 100 1,200
3. Dry gas meter 5,000 25,000
4. Pyrometer + thermocouple 4,000 10,000
5. Pitot Tube 1,500 3,000
6. Stop Watch 1,000 2,500
7. Weights (5 Nos.) 1,000 3,000
• Calibration of supplied instrument is done as per understanding of the
scientist undertaking the calibration. Number of situations have been faced
when calibrations are not done as per requirements and significant errors
are observed in calibration data.
• Once instrument is calibrated user do not try to look into the calibration data
and graph. Further values are not corrected for observed errors indicated in
calibration. Calibration graph and reports are only used to show that this
instrument is calibrated.
• There is no clarity about frequency of calibration. For one type of instrument
one organization says recalibration is due after one year while other says it
is due after 3 years. It appears that frequency of calibration is decided by
calibrating agency arbitrarily.
1.3 SUGGESTED METHODOLOGY:
(A) For field calibration.
Keeping above difficulties in mind Envirotech has setup Roots meter, a
primary standard device for finding out possitive displacement of air. This
unit is traceable to NISST and used for calibration of portable Top Loading
calibrator and also monitoring instrument. The portable calibrator is useful in
the field for calibration of machines at site of their operation. How calibrator
is calibrated? How sampler is calibrated using calibrated calibrator?
Procedure has been discussed in the methodology of calibration given in
Annexure -1.
63
Since calibration is required only for those parameters which are taking part
in the calculation. In HVS/RDS following units need to be calibrated:
for flow - orifice - manometer and rotameter
for time - time totalizer
All these units are possible to calibrate in the field using calibrated portable
Top Loading Calibrator, Dry Gas Meter and Stop Watch.
(B) For Calibration at Our Works.
Machines which can be brought at our works are calibrated using Roots
meter for flow rate at various pressure drop equal to the clean filter and dust
loaded filter. Rotameter calibration is carried out using soap bubble meter or
using dry gas meter depending upon the capacity of Rotameter.
For establishing cut off of the cyclone unit of RDS, collected dust on filter
paper and cyclone cup is analysed for particle size distribution on lesser
based particle size analyser. Rough estimate of cut-off of the cyclone can
also be done using manual method with help of projection microscope.
1.4 OTHER CRITICAL FACTS OF AIR POLLUTION MONITORING
RESULTING UNRELIABLE DATA GENERATION
1. Filter handling is not done as per recommended procedure.
2. Filter are not checked for pin holes. Experiences of this author say that in
every box of 100 sheets 4-5 filters are found damaged with pin holes need to
be rejected and should not be used in sampling.
3. At some places it has been observed that filters are dryed at 105°C for few
hours prior to weighing instead of putting them in desiccator with active
absorbent at room temp (17-27°C) and at low humidity (0-50 % R.H.) as per
standard procedure notified by BIS/CPCB.
4. Cleaning of filter gaskets, top cover, inlet pipe, cyclone assembly is not done
at periodic intervals.
64
5. Manometer tube water is not replaced at specified intervals and there are
situation, when tap water is found filled in the tube responsible for erroneous
recording of flow readings.
6. Initial zero level in manometer is not maintained accurately.
7. Carbon accumulated in blower is not cleaned at periodic interval causes
break down and variation of rpm of armatures. Resulting flow fluctuations.
8. Manual of instrument most of time neither available with operator nor read by
operators. Thus precautions suggested by manufacturer are never followed
resulting errors in sampling and break down of the machine due to
mishandling.
9. In case of gaseous attachment ice is not kept in ice box, flow setting is done
casually, Cleaning of manifold is never done where fine particulates escaping
from impinger solution found to be deposited on the body of needle valve
ultimately responsible for reduction of flow rate because pressure drop at
needle valve is changed during sampling of gaseous pollutants.
10. Impingers are not properly greased and leakages are common in impingers
in present practices. This ultimately responsible for escaping a part of gases
without absorption in the reagents.
11. It has been seen that impingers used for sampling of gaseous pollutants do
not meet BIS/CPCB published specifications. Due to change in dimensions
concentration of gaseous pollutants absorbed in the reagent influenced
significantly.
12. Cleaning of impinger tip with pore size of 1mm is not done after each
sampling. At tip deposition of salts and particulates while sampling causes
reduction of flow rate due to change in pressure drop at the tip.
13. In case of Respirable Dust Sampler collected dust in cyclone cup must be
removed with every filter change otherwise if accumulated dust in cyclone
cup is collected beyond the capacity it shall re-entrained with air stream and
may reach to filter paper and effect collection of Respirable Dust.
14. It has been observed that flow reading are taken immediate after switching
on the blower. Which results recording of inaccurate flow rate because
65
steady flow is only achieved after 4-5 minute of operation. This time is
needed to set the motor brushes.
15. Availability and quality of power is the most critical parameter influencing
monitoring duration and thus measured concentrations also. Frequency of
power (50 cycles/sec) become critical about electrical components like Time
Totalizer where readings are known to be influenced significantly if frequency
of power is changed.
1.5 RECOMMENDATIONS :
« Man power operating instruments need to be qualified, trained and motivated.
• Periodic check of monitoring methodology must be done by an experienced
team through field evaluation on each site where actual sample collection
analysis and calculation of results must be supervised and operators are advised
. Such evaluation must be done once in year for all locations.
• Regular calibration of monitoring instruments must be done as per requirements
of calibration.
• Calibration requirements and frequency of calibration must be made a part of the
standard as method of sampling and analysis is mentioned in air quality
standard.
e On going practices which do not match with standard procedure need to be
stopped so that generation of reliable data can be ensured.
a CPCB should set-up calibration centres at their various zonal laboratories where
calibration of HVS/RDS and other air pollution monitoring instruments is done for
various users in the area on payment basis. While calibration freedom need to
be given to the user to witness the method of calibration and obtained data.
66
Annexure -1
METHODOLOGY OF CALIBRATION OF HVS/RDS
• Flow metric devices used in HVS/RDS are orifice and manometer which indicate
flow in m3
/min.
• Flow measurements are based on observed differential pressure. Thus for flow
calibration, procedure has to be performed against a known air flow.
• Top loading calibrators where a calibration orifice unit with resistance plates
(load plates) are used for calibration of flow rate. Calibration orifice resistance
plates and manometer is shown in figure 1.0.
• Calibration unit is consist of a modular pipe with static pressure tap located at
one end (closed to calibration orifice). Farther end away from static tap is
flanged. In this portion resistance plates are accommodated while calibration.
Near end of the static pressure tap is fitted with a metal plate with an air inlet
which has a fix dia hole. This hole is the orifice on which the calibration
procedures are based. Five resistance plates are used to create resistance of
filter with varying particulate loading. The resistance plates have 5,7,10, 13 & 18
holes. The plate with 18 holes represents a clean filter while other plates
represents a filter with increasingly heavy dust loading.
• Series of steady flows are drawn through the orifice unit for each resistance
plate. The air flow is recorded along with corresponding pressure differential
from the manometer attached to the pressure tap of the orifice unit. Placement
of orifice unit before the primary standard (Roots meter) reduces the inlet
pressure to this meter below atmospheric pressure. To compensate for this a
second manometer is attached to an inlet pressure tap of the primary standard.
After recording barometric pressure, the true volume of air drawn though primary
roots meter is calculated.
P b a r o m e t r i c —
P m e a s u r e d
V t r u e ~ [ -I * ^ m e a s u r e d
P b a r o m e t r i c
Where,
P b a r o m e t r i c
=
Barometeic Pressure at place of calibration mmHg.
P m e a s u r e d
=
Measured Pressure mmHg
Vmeasured = Value measured
Vtme =
True flow rate at ambient conditions
True flow rate can be calculated by
VT rue(m3
)
Q (m3
/rain) =
Time of flow (min)
A plot of orifice manometer pressure reading versus true floW' rate is generated
which become calibration curve for calibrator which is used to calibrate HVS/RDS.
This Top Loading Calibrated Calibrator is fitted at the inlet of sampler with 18 hole
resistance plate in place properly so that there is no leakages. Calibrator pressure
tap is connected to one end of manometer and other end left open to atmosphere.
67
Sampler is switched on and allowed to run for 5 minute for steady flow (this time is
needed to set the motor brushes). Reading indicated by the sampler and pressure
reading of the manometer is recorded. This is done for different resistance plates.
Find out true flow rate reading against recorded pressures in manometer. A plot of
sampler flow rate versus true flow rate reading is plotted and best fit curve not more
than one point of inflection is used as calibration curve of HVS/RDS.
When using the Orifice Calibration unit to calibrate a HVS/RDS correction must be
made to indicated flow if actual atmospheric temperature and pressure vary from
calibration conditions. Corrected flow can be calculated as.
Qcorrcct - QlJncorrcct [(T2 Pl)/(Tj P2)]
Ti
T2
Pi
P2
Corrected flow rate
Uncorrected flow rate reading from orifice manometer
calibration curve.
Absolute temp at orifice calibration condition.
Absolute temp when calibrating the sampler.
Barometric pressure at orifice calibration conditions mmHg.
Barometric pressures when calibrating the sampler
monitoring mmHg
68

More Related Content

What's hot

Qualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERQualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERDr.K.Venkateswara raju
 
Bridging the Gap in Flow Measurement
Bridging the Gap in Flow MeasurementBridging the Gap in Flow Measurement
Bridging the Gap in Flow MeasurementBailey LeRoux
 
Workshop On Risk Assesment by Palash Ch Das
Workshop On Risk Assesment by Palash Ch DasWorkshop On Risk Assesment by Palash Ch Das
Workshop On Risk Assesment by Palash Ch DasPalash Das
 
a1-cbiss Gas Detectors Brochure
a1-cbiss Gas Detectors Brochurea1-cbiss Gas Detectors Brochure
a1-cbiss Gas Detectors Brochurea1-cbiss
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIRMehulJain143
 
Thariq_Ahamed.M(Nov 15)
Thariq_Ahamed.M(Nov 15)Thariq_Ahamed.M(Nov 15)
Thariq_Ahamed.M(Nov 15)Thariq Ahamed
 
Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Atchaya Thalapathy
 
Calibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsCalibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsISF COLLEGE OF PHARMACY MOGA
 
Gas detect. brochure 10 12
Gas detect. brochure 10 12Gas detect. brochure 10 12
Gas detect. brochure 10 12a1-cbiss
 
Calibration of Instruments
Calibration of InstrumentsCalibration of Instruments
Calibration of Instrumentskaroline Enoch
 
Gas detection brochure 60
Gas detection brochure 60Gas detection brochure 60
Gas detection brochure 60a1-cbiss
 
Accab certificate & scope
Accab certificate & scopeAccab certificate & scope
Accab certificate & scopeMaxpromotion
 
Water Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check ProceduresWater Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check ProceduresSeaBirdScientific
 
Qualification of gc equipment by manoj kumar
Qualification of gc equipment by manoj kumarQualification of gc equipment by manoj kumar
Qualification of gc equipment by manoj kumarhimaja donthula
 
Instrumentation For Process Control 09
Instrumentation For Process Control 09Instrumentation For Process Control 09
Instrumentation For Process Control 09GreggACunn
 
C-0131 Mechanical Calibration
C-0131 Mechanical CalibrationC-0131 Mechanical Calibration
C-0131 Mechanical CalibrationMaxpromotion
 
AIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth K
AIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth KAIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth K
AIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth KKartik Vora
 

What's hot (20)

Qualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERQualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETER
 
Bridging the Gap in Flow Measurement
Bridging the Gap in Flow MeasurementBridging the Gap in Flow Measurement
Bridging the Gap in Flow Measurement
 
Workshop On Risk Assesment by Palash Ch Das
Workshop On Risk Assesment by Palash Ch DasWorkshop On Risk Assesment by Palash Ch Das
Workshop On Risk Assesment by Palash Ch Das
 
a1-cbiss Gas Detectors Brochure
a1-cbiss Gas Detectors Brochurea1-cbiss Gas Detectors Brochure
a1-cbiss Gas Detectors Brochure
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIR
 
Thariq_Ahamed.M(Nov 15)
Thariq_Ahamed.M(Nov 15)Thariq_Ahamed.M(Nov 15)
Thariq_Ahamed.M(Nov 15)
 
Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography
 
Calibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsCalibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical Instruments
 
Gas detect. brochure 10 12
Gas detect. brochure 10 12Gas detect. brochure 10 12
Gas detect. brochure 10 12
 
Calibration of Instruments
Calibration of InstrumentsCalibration of Instruments
Calibration of Instruments
 
Gas detection brochure 60
Gas detection brochure 60Gas detection brochure 60
Gas detection brochure 60
 
CALIBRATION INSTRUMENTS
CALIBRATION INSTRUMENTSCALIBRATION INSTRUMENTS
CALIBRATION INSTRUMENTS
 
Accab certificate & scope
Accab certificate & scopeAccab certificate & scope
Accab certificate & scope
 
Water Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check ProceduresWater Quality Monitor - Sensor Check Procedures
Water Quality Monitor - Sensor Check Procedures
 
Qualification
QualificationQualification
Qualification
 
Qualification of gc equipment by manoj kumar
Qualification of gc equipment by manoj kumarQualification of gc equipment by manoj kumar
Qualification of gc equipment by manoj kumar
 
Instrumentation For Process Control 09
Instrumentation For Process Control 09Instrumentation For Process Control 09
Instrumentation For Process Control 09
 
Concepts of Process Instrument Calibration
Concepts of Process Instrument CalibrationConcepts of Process Instrument Calibration
Concepts of Process Instrument Calibration
 
C-0131 Mechanical Calibration
C-0131 Mechanical CalibrationC-0131 Mechanical Calibration
C-0131 Mechanical Calibration
 
AIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth K
AIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth KAIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth K
AIHce 2015 PBZ monitoing and wipe sampling for lead BY Prashanth K
 

Viewers also liked

Monitoring of inorganic ions
Monitoring of inorganic ionsMonitoring of inorganic ions
Monitoring of inorganic ionsECRD2015
 
Preventive measures for occupational
Preventive measures for occupationalPreventive measures for occupational
Preventive measures for occupationalECRD2015
 
2015 tgcsa best bagger manual
2015 tgcsa best bagger manual2015 tgcsa best bagger manual
2015 tgcsa best bagger manualAnurag Sinha
 
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...taxonbytes
 

Viewers also liked (7)

Dieu Hien - Bai 17
Dieu Hien - Bai 17Dieu Hien - Bai 17
Dieu Hien - Bai 17
 
Monitoring of inorganic ions
Monitoring of inorganic ionsMonitoring of inorganic ions
Monitoring of inorganic ions
 
Preventive measures for occupational
Preventive measures for occupationalPreventive measures for occupational
Preventive measures for occupational
 
Autoservicio
AutoservicioAutoservicio
Autoservicio
 
2015 tgcsa best bagger manual
2015 tgcsa best bagger manual2015 tgcsa best bagger manual
2015 tgcsa best bagger manual
 
BIM Portfolio
BIM PortfolioBIM Portfolio
BIM Portfolio
 
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...
 

Similar to Some critical facets in air pollution

calibrationintervalspresentation.pdf
calibrationintervalspresentation.pdfcalibrationintervalspresentation.pdf
calibrationintervalspresentation.pdfAbdulSalamSagir1
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollutionECRD2015
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollutionECRD IN
 
Calipers presentation mukesh
Calipers presentation mukeshCalipers presentation mukesh
Calipers presentation mukeshmukeshjeswani
 
Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology...
 Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology... Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology...
Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology...hydrologyproject0
 
Untitled.fr10
Untitled.fr10Untitled.fr10
Untitled.fr10ECRD IN
 
Implementation of condition based maintenance in manufacturing industry
Implementation of condition based maintenance in manufacturing industryImplementation of condition based maintenance in manufacturing industry
Implementation of condition based maintenance in manufacturing industryAbhilash sharan
 
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Mudasir Gulzar
 
Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviii
 Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviii Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviii
Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviiihydrologyproject001
 
Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...
Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...
Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...Kartik Vora
 
CONDITION BASED MAINTAINACE BASICS presentation
CONDITION BASED MAINTAINACE BASICS presentation CONDITION BASED MAINTAINACE BASICS presentation
CONDITION BASED MAINTAINACE BASICS presentation RAJKUMAR Rswagmare
 
In-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab PracticeIn-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab PracticeLiza Tallon
 
Particulate Containment Validation By Prashant kondragunta, ISS
Particulate Containment Validation By Prashant kondragunta, ISSParticulate Containment Validation By Prashant kondragunta, ISS
Particulate Containment Validation By Prashant kondragunta, ISSInternational Safety Systems
 
PDM IMPLIMENTATION
PDM IMPLIMENTATIONPDM IMPLIMENTATION
PDM IMPLIMENTATIONUttam Misra
 
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTIONPHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTIONPrincy Agarwal
 
CONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICS
CONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICSCONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICS
CONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICSijmnct
 

Similar to Some critical facets in air pollution (20)

calibrationintervalspresentation.pdf
calibrationintervalspresentation.pdfcalibrationintervalspresentation.pdf
calibrationintervalspresentation.pdf
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollution
 
Calibration of air pollution
Calibration of air pollutionCalibration of air pollution
Calibration of air pollution
 
Calipers presentation mukesh
Calipers presentation mukeshCalipers presentation mukesh
Calipers presentation mukesh
 
Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology...
 Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology... Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology...
Download-manuals-surface water-manual-sw-volume3fieldmanualhydro-meteorology...
 
Calibration
CalibrationCalibration
Calibration
 
Preventive maintenance
Preventive maintenancePreventive maintenance
Preventive maintenance
 
Untitled.fr10
Untitled.fr10Untitled.fr10
Untitled.fr10
 
Implementation of condition based maintenance in manufacturing industry
Implementation of condition based maintenance in manufacturing industryImplementation of condition based maintenance in manufacturing industry
Implementation of condition based maintenance in manufacturing industry
 
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
 
Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviii
 Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviii Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviii
Download-manuals-surface water-manual-sw-volume4fieldmanualhydrometrypartviii
 
All you Need to Know About Sampling and Analysis - OGSI
All you Need to Know About Sampling and Analysis - OGSIAll you Need to Know About Sampling and Analysis - OGSI
All you Need to Know About Sampling and Analysis - OGSI
 
Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...
Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...
Particulate Containment Validation: An Effective Tool in Reducing Pharmaceuti...
 
CONDITION BASED MAINTAINACE BASICS presentation
CONDITION BASED MAINTAINACE BASICS presentation CONDITION BASED MAINTAINACE BASICS presentation
CONDITION BASED MAINTAINACE BASICS presentation
 
In-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab PracticeIn-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab Practice
 
Particulate Containment Validation By Prashant kondragunta, ISS
Particulate Containment Validation By Prashant kondragunta, ISSParticulate Containment Validation By Prashant kondragunta, ISS
Particulate Containment Validation By Prashant kondragunta, ISS
 
PDM IMPLIMENTATION
PDM IMPLIMENTATIONPDM IMPLIMENTATION
PDM IMPLIMENTATION
 
Condition Based Maintenance
Condition Based MaintenanceCondition Based Maintenance
Condition Based Maintenance
 
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTIONPHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
 
CONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICS
CONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICSCONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICS
CONDITION-BASED MAINTENANCE USING SENSOR ARRAYS AND TELEMATICS
 

More from ECRD2015

Pollution control in small and medium scale industry
Pollution control in small and medium scale industryPollution control in small and medium scale industry
Pollution control in small and medium scale industryECRD2015
 
Indoor air quality
Indoor air qualityIndoor air quality
Indoor air qualityECRD2015
 
Air pollution control
Air pollution controlAir pollution control
Air pollution controlECRD2015
 
Vehicular Pollution Management
Vehicular Pollution ManagementVehicular Pollution Management
Vehicular Pollution ManagementECRD2015
 
Strategy of control of urban air pollution
Strategy of control of urban air pollutionStrategy of control of urban air pollution
Strategy of control of urban air pollutionECRD2015
 
Spectrophotometer meter
Spectrophotometer meterSpectrophotometer meter
Spectrophotometer meterECRD2015
 
Shop floor monitoring
Shop floor monitoringShop floor monitoring
Shop floor monitoringECRD2015
 
Quality assurance and quality control
Quality assurance and quality controlQuality assurance and quality control
Quality assurance and quality controlECRD2015
 
Pm10 measmnt
Pm10 measmntPm10 measmnt
Pm10 measmntECRD2015
 
Nitrogen oxide monitoring
Nitrogen oxide monitoringNitrogen oxide monitoring
Nitrogen oxide monitoringECRD2015
 
Minutes of brain storming session on air pollution load carrying capacity of ...
Minutes of brain storming session on air pollution load carrying capacity of ...Minutes of brain storming session on air pollution load carrying capacity of ...
Minutes of brain storming session on air pollution load carrying capacity of ...ECRD2015
 
Measurement of respirable suspended pm10
Measurement of respirable suspended pm10Measurement of respirable suspended pm10
Measurement of respirable suspended pm10ECRD2015
 
Measurement of hydrogen ion concentration (p h)
Measurement of hydrogen ion concentration (p h)Measurement of hydrogen ion concentration (p h)
Measurement of hydrogen ion concentration (p h)ECRD2015
 
Maintenance
MaintenanceMaintenance
MaintenanceECRD2015
 
Maintenance of high volume
Maintenance of high volumeMaintenance of high volume
Maintenance of high volumeECRD2015
 
Industrial emission control techniques
Industrial emission control techniquesIndustrial emission control techniques
Industrial emission control techniquesECRD2015
 
Indophenol method
Indophenol methodIndophenol method
Indophenol methodECRD2015
 
Industrial ventilation
Industrial ventilationIndustrial ventilation
Industrial ventilationECRD2015
 

More from ECRD2015 (20)

Pollution control in small and medium scale industry
Pollution control in small and medium scale industryPollution control in small and medium scale industry
Pollution control in small and medium scale industry
 
Indoor air quality
Indoor air qualityIndoor air quality
Indoor air quality
 
Air pollution control
Air pollution controlAir pollution control
Air pollution control
 
Vehicular Pollution Management
Vehicular Pollution ManagementVehicular Pollution Management
Vehicular Pollution Management
 
Strategy of control of urban air pollution
Strategy of control of urban air pollutionStrategy of control of urban air pollution
Strategy of control of urban air pollution
 
Spectrophotometer meter
Spectrophotometer meterSpectrophotometer meter
Spectrophotometer meter
 
Slm_1
Slm_1Slm_1
Slm_1
 
Shop floor monitoring
Shop floor monitoringShop floor monitoring
Shop floor monitoring
 
Quality assurance and quality control
Quality assurance and quality controlQuality assurance and quality control
Quality assurance and quality control
 
Pm10 measmnt
Pm10 measmntPm10 measmnt
Pm10 measmnt
 
Ozone
OzoneOzone
Ozone
 
Nitrogen oxide monitoring
Nitrogen oxide monitoringNitrogen oxide monitoring
Nitrogen oxide monitoring
 
Minutes of brain storming session on air pollution load carrying capacity of ...
Minutes of brain storming session on air pollution load carrying capacity of ...Minutes of brain storming session on air pollution load carrying capacity of ...
Minutes of brain storming session on air pollution load carrying capacity of ...
 
Measurement of respirable suspended pm10
Measurement of respirable suspended pm10Measurement of respirable suspended pm10
Measurement of respirable suspended pm10
 
Measurement of hydrogen ion concentration (p h)
Measurement of hydrogen ion concentration (p h)Measurement of hydrogen ion concentration (p h)
Measurement of hydrogen ion concentration (p h)
 
Maintenance
MaintenanceMaintenance
Maintenance
 
Maintenance of high volume
Maintenance of high volumeMaintenance of high volume
Maintenance of high volume
 
Industrial emission control techniques
Industrial emission control techniquesIndustrial emission control techniques
Industrial emission control techniques
 
Indophenol method
Indophenol methodIndophenol method
Indophenol method
 
Industrial ventilation
Industrial ventilationIndustrial ventilation
Industrial ventilation
 

Recently uploaded

Horizon Net Zero Dawn – keynote slides by Ben Abraham
Horizon Net Zero Dawn – keynote slides by Ben AbrahamHorizon Net Zero Dawn – keynote slides by Ben Abraham
Horizon Net Zero Dawn – keynote slides by Ben Abrahamssuserbb03ff
 
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...Cluster TWEED
 
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Kalighat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Roomdivyansh0kumar0
 
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girlsMumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girlsPooja Nehwal
 
VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130
VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130
VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130Suhani Kapoor
 
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur EscortsCall Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Spiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsSpiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsprasan26
 
Freegle User Survey as visual display - BH
Freegle User Survey as visual display - BHFreegle User Survey as visual display - BH
Freegle User Survey as visual display - BHbill846304
 
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service NashikRussian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashikranjana rawat
 
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service MumbaiCall Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service MumbaiCall girls in Ahmedabad High profile
 
VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...Suhani Kapoor
 
Sustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and ChallengesSustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and ChallengesDr. Salem Baidas
 

Recently uploaded (20)

Gandhi Nagar (Delhi) 9953330565 Escorts, Call Girls Services
Gandhi Nagar (Delhi) 9953330565 Escorts, Call Girls ServicesGandhi Nagar (Delhi) 9953330565 Escorts, Call Girls Services
Gandhi Nagar (Delhi) 9953330565 Escorts, Call Girls Services
 
Horizon Net Zero Dawn – keynote slides by Ben Abraham
Horizon Net Zero Dawn – keynote slides by Ben AbrahamHorizon Net Zero Dawn – keynote slides by Ben Abraham
Horizon Net Zero Dawn – keynote slides by Ben Abraham
 
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
(ANAYA) Call Girls Hadapsar ( 7001035870 ) HI-Fi Pune Escorts Service
 
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
webinaire-green-mirror-episode-2-Smart contracts and virtual purchase agreeme...
 
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Kalighat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Kalighat 👉 8250192130 Available With Room
 
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girlsMumbai Call Girls, 💞  Prity 9892124323, Navi Mumbai Call girls
Mumbai Call Girls, 💞 Prity 9892124323, Navi Mumbai Call girls
 
Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...
Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...
Call Girls In { Delhi } South Extension Whatsup 9873940964 Enjoy Unlimited Pl...
 
Escort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCR
Escort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCREscort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCR
Escort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCR
 
VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130
VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130
VIP Call Girls Service Tolichowki Hyderabad Call +91-8250192130
 
Sustainable Packaging
Sustainable PackagingSustainable Packaging
Sustainable Packaging
 
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur EscortsCall Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
 
Spiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnidsSpiders by Slidesgo - an introduction to arachnids
Spiders by Slidesgo - an introduction to arachnids
 
Freegle User Survey as visual display - BH
Freegle User Survey as visual display - BHFreegle User Survey as visual display - BH
Freegle User Survey as visual display - BH
 
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service NashikRussian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
Russian Call Girls Nashik Anjali 7001305949 Independent Escort Service Nashik
 
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service MumbaiCall Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
Call Girls Mumbai Gayatri 8617697112 Independent Escort Service Mumbai
 
E Waste Management
E Waste ManagementE Waste Management
E Waste Management
 
VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
VIP Call Girls Moti Ganpur ( Hyderabad ) Phone 8250192130 | ₹5k To 25k With R...
 
FULL ENJOY Call Girls In kashmiri gate (Delhi) Call Us 9953056974
FULL ENJOY Call Girls In  kashmiri gate (Delhi) Call Us 9953056974FULL ENJOY Call Girls In  kashmiri gate (Delhi) Call Us 9953056974
FULL ENJOY Call Girls In kashmiri gate (Delhi) Call Us 9953056974
 
Call Girls In R.K. Puram 9953056974 Escorts ServiCe In Delhi Ncr
Call Girls In R.K. Puram 9953056974 Escorts ServiCe In Delhi NcrCall Girls In R.K. Puram 9953056974 Escorts ServiCe In Delhi Ncr
Call Girls In R.K. Puram 9953056974 Escorts ServiCe In Delhi Ncr
 
Sustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and ChallengesSustainable Clothing Strategies and Challenges
Sustainable Clothing Strategies and Challenges
 

Some critical facets in air pollution

  • 1. Some Critical Facets in Air Pollution Monitoring Dr. Rajendra Prasad Managing Director Envirotech Instruments Pvt. Ltd. A-271, Okhla Industrial Area, Phase -1, New Delhi - 110020 (email: envirotech@vsnl.com) 1.0 BACKGROUND The prime objective of air quality monitoring is to find out prevailing concentration of air pollutants with accuracy of 1 Dg/m3 . This accuracy is depend upon several factors. Man and machine are the two basic facets equally responsible for reliable data generation which can meet aforesaid accuracy expectations. Qualification, training and experience of man power can improve quality of generated data. However if monitoring personnel are motivated absolute reliability in collected data can ensured without further efforts even in existing setup. Similarly the monitoring instruments have to be accurate, precise and regularly calibrated without which quality data generation is out of question. CPCB have already specified following accuracy limits for critical parameters recorded by a monitoring system used for AAQM. Flow rate accurate upto + 5 % Time accurate upto ± 1 min. Above parameters required to be critically examined in monitoring instruments and must be ensured within specified accuracy limits for maintaining desired accuracy of measurements. 1.1 PRESENT SCENARIO 1. About Man power : The manpower deployed for regular air quality monitoring either in NAAQM or in any other monitoring program most of time are not adequately qualified and experienced. Formal training has never been given to these staff. Most 60
  • 2. critical part is that the people engaged in these monitorings are not motivated for the assigned jobs. Besides this there are several other serious limitations with man power like they are overloaded with work due to less staff, have no clear cut instructions about monitoring for acceptance and rejection of sample, have no freedom to put there problems and view point with senior and experienced officers. At last there salaries are neither attractive not revised since long hence whenever there are opportunities people are escaping. About the Machines : Monitoring instruments used for assessment of air pollution are prone to drift and may show variation with time in measured parameters. Hence it is a false expectation that their calibrations are stable over long period of time. As per standard requirements calibration need to be done. • When instrument is manufactured. • When instrument is transported to a new location and in different climatic condition. • If instrument is operated after interruption of several months. • After repairs or overhaul of key components. • After observing significant drift either in zero or span value or in recorded value. The monitoring instruments used in regular air pollution monitoring either under NAAQM or in other project are not calibrated as per above requirements. Even when it is known that regular calibration of monitoring instrument is a basic need for reliable data generation. • In general life of a HVS/RDS is about 5 years. But instruments are not retired even after crossing age of 10 years and more with old machine (not regularly calibrated with age of 10 years and above) reliability of data are expected similar as expected with new calibrated machines. 61
  • 3. Table below show the age distribution of installed HVS at various NAAQM stations. S.No. Age (years) No. of Machine 1. 10 and above 32 2. 7 - 1 0 75 3. 5 - 7 28 4. Less than 5 06 Total 141 (Source Status Report on Servicing of HVS Installed at Various NAAQM Station in India - 1995) Over all it can be said that conditions of machines used in air quality monitoring are critical and are not good for reliable data generation. In such situation when conditions of man and machine are weak reliable data can not be generated as per requirements. 1.2 Difficulties of Calibration : Users face following difficulties while calibration of instruments. • Calibration facilities are not available at number of places in the country. Further all required set-ups related to the calibration of air pollution monitoring are not available with one agency. • Reference set up at calibrating agencies are not portable enough hence calibration at site is not possible. Thus monitoring system need to be taken to calibration agencies for calibration. It is not easy to transport HVS/RDS from distant places. Besides this cost of transportation is also significant where possibilities of damages are always there. • Time taken for calibration is several weeks since spare machines are not available hence calibration is postponed due to this limitation. • Cost of calibration is very high most of the time more than the cost of instrument to be calibrated. Some calibration costs are given below. 62
  • 4. S.No, Name of the item Cost of item calibrator (Rs.) Cost of calibration (Rs.) 1. Portable calibrator for flow rate calibration 15,000 28,000 2. Thermometer (6 points) 100 1,200 3. Dry gas meter 5,000 25,000 4. Pyrometer + thermocouple 4,000 10,000 5. Pitot Tube 1,500 3,000 6. Stop Watch 1,000 2,500 7. Weights (5 Nos.) 1,000 3,000 • Calibration of supplied instrument is done as per understanding of the scientist undertaking the calibration. Number of situations have been faced when calibrations are not done as per requirements and significant errors are observed in calibration data. • Once instrument is calibrated user do not try to look into the calibration data and graph. Further values are not corrected for observed errors indicated in calibration. Calibration graph and reports are only used to show that this instrument is calibrated. • There is no clarity about frequency of calibration. For one type of instrument one organization says recalibration is due after one year while other says it is due after 3 years. It appears that frequency of calibration is decided by calibrating agency arbitrarily. 1.3 SUGGESTED METHODOLOGY: (A) For field calibration. Keeping above difficulties in mind Envirotech has setup Roots meter, a primary standard device for finding out possitive displacement of air. This unit is traceable to NISST and used for calibration of portable Top Loading calibrator and also monitoring instrument. The portable calibrator is useful in the field for calibration of machines at site of their operation. How calibrator is calibrated? How sampler is calibrated using calibrated calibrator? Procedure has been discussed in the methodology of calibration given in Annexure -1. 63
  • 5. Since calibration is required only for those parameters which are taking part in the calculation. In HVS/RDS following units need to be calibrated: for flow - orifice - manometer and rotameter for time - time totalizer All these units are possible to calibrate in the field using calibrated portable Top Loading Calibrator, Dry Gas Meter and Stop Watch. (B) For Calibration at Our Works. Machines which can be brought at our works are calibrated using Roots meter for flow rate at various pressure drop equal to the clean filter and dust loaded filter. Rotameter calibration is carried out using soap bubble meter or using dry gas meter depending upon the capacity of Rotameter. For establishing cut off of the cyclone unit of RDS, collected dust on filter paper and cyclone cup is analysed for particle size distribution on lesser based particle size analyser. Rough estimate of cut-off of the cyclone can also be done using manual method with help of projection microscope. 1.4 OTHER CRITICAL FACTS OF AIR POLLUTION MONITORING RESULTING UNRELIABLE DATA GENERATION 1. Filter handling is not done as per recommended procedure. 2. Filter are not checked for pin holes. Experiences of this author say that in every box of 100 sheets 4-5 filters are found damaged with pin holes need to be rejected and should not be used in sampling. 3. At some places it has been observed that filters are dryed at 105°C for few hours prior to weighing instead of putting them in desiccator with active absorbent at room temp (17-27°C) and at low humidity (0-50 % R.H.) as per standard procedure notified by BIS/CPCB. 4. Cleaning of filter gaskets, top cover, inlet pipe, cyclone assembly is not done at periodic intervals. 64
  • 6. 5. Manometer tube water is not replaced at specified intervals and there are situation, when tap water is found filled in the tube responsible for erroneous recording of flow readings. 6. Initial zero level in manometer is not maintained accurately. 7. Carbon accumulated in blower is not cleaned at periodic interval causes break down and variation of rpm of armatures. Resulting flow fluctuations. 8. Manual of instrument most of time neither available with operator nor read by operators. Thus precautions suggested by manufacturer are never followed resulting errors in sampling and break down of the machine due to mishandling. 9. In case of gaseous attachment ice is not kept in ice box, flow setting is done casually, Cleaning of manifold is never done where fine particulates escaping from impinger solution found to be deposited on the body of needle valve ultimately responsible for reduction of flow rate because pressure drop at needle valve is changed during sampling of gaseous pollutants. 10. Impingers are not properly greased and leakages are common in impingers in present practices. This ultimately responsible for escaping a part of gases without absorption in the reagents. 11. It has been seen that impingers used for sampling of gaseous pollutants do not meet BIS/CPCB published specifications. Due to change in dimensions concentration of gaseous pollutants absorbed in the reagent influenced significantly. 12. Cleaning of impinger tip with pore size of 1mm is not done after each sampling. At tip deposition of salts and particulates while sampling causes reduction of flow rate due to change in pressure drop at the tip. 13. In case of Respirable Dust Sampler collected dust in cyclone cup must be removed with every filter change otherwise if accumulated dust in cyclone cup is collected beyond the capacity it shall re-entrained with air stream and may reach to filter paper and effect collection of Respirable Dust. 14. It has been observed that flow reading are taken immediate after switching on the blower. Which results recording of inaccurate flow rate because 65
  • 7. steady flow is only achieved after 4-5 minute of operation. This time is needed to set the motor brushes. 15. Availability and quality of power is the most critical parameter influencing monitoring duration and thus measured concentrations also. Frequency of power (50 cycles/sec) become critical about electrical components like Time Totalizer where readings are known to be influenced significantly if frequency of power is changed. 1.5 RECOMMENDATIONS : « Man power operating instruments need to be qualified, trained and motivated. • Periodic check of monitoring methodology must be done by an experienced team through field evaluation on each site where actual sample collection analysis and calculation of results must be supervised and operators are advised . Such evaluation must be done once in year for all locations. • Regular calibration of monitoring instruments must be done as per requirements of calibration. • Calibration requirements and frequency of calibration must be made a part of the standard as method of sampling and analysis is mentioned in air quality standard. e On going practices which do not match with standard procedure need to be stopped so that generation of reliable data can be ensured. a CPCB should set-up calibration centres at their various zonal laboratories where calibration of HVS/RDS and other air pollution monitoring instruments is done for various users in the area on payment basis. While calibration freedom need to be given to the user to witness the method of calibration and obtained data. 66
  • 8. Annexure -1 METHODOLOGY OF CALIBRATION OF HVS/RDS • Flow metric devices used in HVS/RDS are orifice and manometer which indicate flow in m3 /min. • Flow measurements are based on observed differential pressure. Thus for flow calibration, procedure has to be performed against a known air flow. • Top loading calibrators where a calibration orifice unit with resistance plates (load plates) are used for calibration of flow rate. Calibration orifice resistance plates and manometer is shown in figure 1.0. • Calibration unit is consist of a modular pipe with static pressure tap located at one end (closed to calibration orifice). Farther end away from static tap is flanged. In this portion resistance plates are accommodated while calibration. Near end of the static pressure tap is fitted with a metal plate with an air inlet which has a fix dia hole. This hole is the orifice on which the calibration procedures are based. Five resistance plates are used to create resistance of filter with varying particulate loading. The resistance plates have 5,7,10, 13 & 18 holes. The plate with 18 holes represents a clean filter while other plates represents a filter with increasingly heavy dust loading. • Series of steady flows are drawn through the orifice unit for each resistance plate. The air flow is recorded along with corresponding pressure differential from the manometer attached to the pressure tap of the orifice unit. Placement of orifice unit before the primary standard (Roots meter) reduces the inlet pressure to this meter below atmospheric pressure. To compensate for this a second manometer is attached to an inlet pressure tap of the primary standard. After recording barometric pressure, the true volume of air drawn though primary roots meter is calculated. P b a r o m e t r i c — P m e a s u r e d V t r u e ~ [ -I * ^ m e a s u r e d P b a r o m e t r i c Where, P b a r o m e t r i c = Barometeic Pressure at place of calibration mmHg. P m e a s u r e d = Measured Pressure mmHg Vmeasured = Value measured Vtme = True flow rate at ambient conditions True flow rate can be calculated by VT rue(m3 ) Q (m3 /rain) = Time of flow (min) A plot of orifice manometer pressure reading versus true floW' rate is generated which become calibration curve for calibrator which is used to calibrate HVS/RDS. This Top Loading Calibrated Calibrator is fitted at the inlet of sampler with 18 hole resistance plate in place properly so that there is no leakages. Calibrator pressure tap is connected to one end of manometer and other end left open to atmosphere. 67
  • 9. Sampler is switched on and allowed to run for 5 minute for steady flow (this time is needed to set the motor brushes). Reading indicated by the sampler and pressure reading of the manometer is recorded. This is done for different resistance plates. Find out true flow rate reading against recorded pressures in manometer. A plot of sampler flow rate versus true flow rate reading is plotted and best fit curve not more than one point of inflection is used as calibration curve of HVS/RDS. When using the Orifice Calibration unit to calibrate a HVS/RDS correction must be made to indicated flow if actual atmospheric temperature and pressure vary from calibration conditions. Corrected flow can be calculated as. Qcorrcct - QlJncorrcct [(T2 Pl)/(Tj P2)] Ti T2 Pi P2 Corrected flow rate Uncorrected flow rate reading from orifice manometer calibration curve. Absolute temp at orifice calibration condition. Absolute temp when calibrating the sampler. Barometric pressure at orifice calibration conditions mmHg. Barometric pressures when calibrating the sampler monitoring mmHg 68