White Paper | CARESTREAM Tube & Line Visualization Software,
CARESTREAM Pneumothorax Visualization Software
Carestream’s companion-view processing for
improved visualization of tubes, lines, and
pneumothoraces in digital, portable chest radiography
Introduction
Carestream has implemented companion views in its digital
radiography systems. A companion view is designed to
complement the standard processed radiographic image
delivered from the digital radiography capture modality to
PACS, to provide an additional rendering tailored for the visual
interpretation needed for a specific diagnostic or clinical
purpose. Two companion views are available in Carestream
products for chest radiography: one for the optimal
visualization of tubes and lines in chest radiographs
(CARESTREAM Tube & Line Visualization Software), and the
other for enhancing the conspicuity of a pneumothorax
(CARESTREAM Pneumothorax Visualization Software).
Companion Views
Carestream’s digital radiography image processing software is
designed to automatically render images with an appearance
suitable for a radiologist to assess a patient’s condition with
respect to the spectrum of potential abnormalities. A single
chest image typically comprises a wide range of exposure
(gray) levels that exceed the number of brightness levels
available on a PACS workstation monitor. Therefore, an
appropriate overall rendering may not necessarily display the
full grayscale fidelity for any particular indication. In other
words, there are simply more gray levels in the raw, captured
image than what the PACS monitor can display in a single,
static presentation. This effect exists for radiologists who
typically read images on high-resolution displays in rooms with
reduced ambient lighting—but the effect is exacerbated for
ICU clinicians who may need to interpret portable chest
radiographs at the point of care, on lower-resolution monitors
under bright light, which is common in the ICU setting.
For instance, a portable chest radiograph may be properly
rendered and displayed with excellent overall diagnostic
quality, yet it may still be challenging to localize tube and line
tips in underpenetrated regions, such as in the mediastinum
and subdiaphragm. In order to preserve global contrast and
brightness for the overall image, gray levels may be quantized
in the underpenetrated regions, which will cause some degree
of detail contrast loss in the mediastinum (Figure 1 – image
on left).
To address this issue, Carestream has developed a unique and
robust image-processing algorithm that specifically highlights
tube and catheter features 1
(CARESTREAM Tube & Line
Visualization Software). To generate the companion view, the
image is first decomposed into eight spatial frequency bands.
These bands are individually manipulated using non-linear
functions designed to achieve an equalized contrast
enhancement of image details, while at the same time
minimizing noise. The manipulated frequency bands are then
reconstructed and the output image is passed through a
grayscale look-up table for display. This method boosts the
spatial frequencies required to enhance the conspicuity of
tubes and lines while it suppresses the spatial frequencies
containing noise. By enhancing the tube and line features
while simultaneously suppressing noise, this method also
works robustly for images captured at low exposure levels and
for images of large patients containing excessive X-ray scatter.
While the companion view presentation may not be suitable
for general diagnostic interpretation, the image features
necessary for interpretation of the tube and line structures are
significantly enhanced (Figure 1 – image on right).
White Paper | CARESTREAM Tube & Line Visualization Software,
CARESTREAM Pneumothorax Visualization Software
2
Figure 1 – Left: a portable chest X-ray image processed using default automatic processing. Right: the same image processed using
the Tube & Line Visualization Software. The two images are delivered together to PACS.
For some abnormalities, the visual characteristics of the key
diagnostic features may be masked by, or confounded with,
normal anatomical structures. This is a very different effect
from the insufficient grayscale resolution issue described
above. This effect can be more readily thought of as visual
interference between the features corresponding to the
abnormality and the normal anatomical structures.
For instance, consider a pneumothorax, defined as the
presence of air in the pleural space. A pneumothorax is
visualized by a delineation of the lung edge, which is displaced
from the chest wall by the air contained between the visceral
and parietal pleura. Depending on positioning, there will be no
visible lung markings in this region. The delineation of the lung
edge, and the absence of lung markings, may be difficult to
perceive and be potentially masked by the radiographic
presentation of the scapula or ribs. The conspicuity of the
pneumothorax features may be further dampened by
insufficient grayscale resolution of the PACS display.
In an approach analogous to that used in tube- and line-
enhanced visualization, Carestream has developed a
companion view for pneumothorax that is specifically designed
to enhance the appearance of the associated features
(CARESTREAM Pneumothorax Visualization Software). The
algorithm differs from the tube- and line-enhancement
technique. Instead of focusing on boosting the spatial
frequencies associated with the diagnostic features of interest,
the pneumothorax algorithm uses a texture filter that
enhances the appearance differences between the
characteristic markings for the regions inside and outside the
lung. On page 3, Figure 2 (image on left) shows an example of
a large pneumothorax (the lung edge is indicated by arrows).
The image on the right in Figure 2 shows an insert that
illustrates the effect of the enhancement resulting from the
application of the texture filter.
White Paper | CARESTREAM Tube & Line Visualization Software,
CARESTREAM Pneumothorax Visualization Software
3
Figure 2 – Left: a portable chest X-ray image illustrating a large pneumothorax. Right: an insert showing the texture differences
resulting from processing the image using the pneumothorax-enhancement filter.
The images shown in Figure 3 compare a portable chest X-ray image presented using the default processing (on the left) against
the same image processed with the pneumothorax-enhancement filter (on the right).
Figure 3 – Left: a portable chest X-ray image of a patient with a pneumothorax. Right: the companion view processed using the
pneumothorax texture-enhancement filter. The lung edge is indicated by arrows.
White Paper | CARESTREAM Tube & Line Visualization Software,
CARESTREAM Pneumothorax Visualization Software
www.carestream.com
© Carestream Health, Inc., 2012. CARESTREAM is a trademark of
Carestream Health. CAT 200 0012 7/12
Companion views improve interpretation
confidence and radiologist reading efficiency
PACS workstations provide window/level adjustment
capabilities to help address the generic shortcomings of
insufficient luminance dynamic range on PACS monitors, and
for enhancing the contrast of subtle features. However,
interactive window and leveling is time-consuming and
degrades radiologist reading efficiency. The enhanced
visualization delivered to PACS as a companion view, together
with the default processed image, makes it possible for
radiologists and clinicians to rapidly toggle between the
standard visualization optimized for general interpretation, and
the customized visualization that is optimized for a specific
diagnostic task. Because the companion view can be toggled
with the default processed image, radiologists can more easily
make their interpretations with minimal workstation
interaction.
A study was conducted with both radiologists and pulmonary
ICU physicians to assess the utility of the tube and line
companion view in terms of interpretation and reading
efficiency 2
. The companion view method was evaluated by
two reader groups, radiologists and pulmonary ICU physicians.
Each reader group evaluated the images under viewing
conditions consistent with their typical respective reading
environment. The companion view method was shown to
provide improved interpretation confidence and improved
diagnostic efficiency for radiologists (30% less reading time),
and more confident interpretation for pulmonary ICU
physicians in the same amount of time, when used in lieu of
window and level adjustments at the PACS workstation.
Summary
Carestream now offers companion-view processing that is
designed to enhance the visualization of tubes, lines, and
pneumothoraces in chest images captured using Carestream
CR and DR modalities. These companion views complement
the standard default rendering delivered from the capture
modality to PACS with a second image that is processed to
enhance the features for a particular diagnostic or clinical
indication. The Tube & Line Visualization Software has been
shown to improve interpretation confidence for radiologists
and pulmonary ICU physicians, and improve reading efficiency
for radiologists.
References
1. Couwenhoven ME, Senn RA, Foos DH. Enhancement method that provides direct and independent control of fundamental attributes of
image quality for radiographic imagery. SPIE Medical Imaging Proceedings; 5367: 474-481, 2004.
2. Foos DH, Yankelevitz DF, Wang X, Berlin D, Zappetti D, Cham M, Sanders A, Novak Parker K, Henschke CI. Improved visualization of tubes
and lines in portable intensive care unit radiographs: a study comparing a new approach to the standard approach. Clinical Imaging
Volume 35, Issue 5, September–October 2011, Pages 346–352.

More Related Content

What's hot

Carestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound ImagingCarestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound ImagingCarestream
 
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAEDose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAEhaijaypee_dan
 
Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedyYashawant Yadav
 
CT Image reconstruction
CT Image reconstructionCT Image reconstruction
CT Image reconstructionSantosh Ojha
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scanMohd Aiman Azmardi
 
Automatic detection of lung cancer in ct images
Automatic detection of lung cancer in ct imagesAutomatic detection of lung cancer in ct images
Automatic detection of lung cancer in ct imageseSAT Publishing House
 
Basics of ct lecture 1
Basics of ct  lecture 1Basics of ct  lecture 1
Basics of ct lecture 1Gamal Mahdaly
 
CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstructionGunjan Patel
 
4 computed tomography Dr. Muhammad Bin Zulfiqar
4 computed tomography Dr. Muhammad Bin Zulfiqar4 computed tomography Dr. Muhammad Bin Zulfiqar
4 computed tomography Dr. Muhammad Bin ZulfiqarDr. Muhammad Bin Zulfiqar
 
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisAn Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisCSCJournals
 
Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...
Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...
Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...IDES Editor
 
Segmentation of Lung Lobes and Nodules in CT Images
Segmentation of Lung Lobes and Nodules in CT ImagesSegmentation of Lung Lobes and Nodules in CT Images
Segmentation of Lung Lobes and Nodules in CT Imagessipij
 

What's hot (20)

Ct i patient_whitepaper
Ct i patient_whitepaperCt i patient_whitepaper
Ct i patient_whitepaper
 
Carestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound ImagingCarestream Whitepaper Advanced Ultrasound Imaging
Carestream Whitepaper Advanced Ultrasound Imaging
 
Session 6 image quality ct and session 8 dose ct
Session 6  image quality ct and  session 8 dose ctSession 6  image quality ct and  session 8 dose ct
Session 6 image quality ct and session 8 dose ct
 
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAEDose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
Dose reduction in MDCT . Daniel J.P , Khorfakhan hospital . UAE
 
Radiotherapy Treatment Simulation
Radiotherapy Treatment SimulationRadiotherapy Treatment Simulation
Radiotherapy Treatment Simulation
 
Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedy
 
CT Image reconstruction
CT Image reconstructionCT Image reconstruction
CT Image reconstruction
 
MR reconstruction 101
MR reconstruction 101MR reconstruction 101
MR reconstruction 101
 
Medphysics Planning
Medphysics PlanningMedphysics Planning
Medphysics Planning
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scan
 
Automatic detection of lung cancer in ct images
Automatic detection of lung cancer in ct imagesAutomatic detection of lung cancer in ct images
Automatic detection of lung cancer in ct images
 
Digital X-ray Problems..
Digital X-ray Problems..Digital X-ray Problems..
Digital X-ray Problems..
 
Basics of ct lecture 1
Basics of ct  lecture 1Basics of ct  lecture 1
Basics of ct lecture 1
 
Dose reduction
Dose reductionDose reduction
Dose reduction
 
CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstruction
 
Virtopsy
VirtopsyVirtopsy
Virtopsy
 
4 computed tomography Dr. Muhammad Bin Zulfiqar
4 computed tomography Dr. Muhammad Bin Zulfiqar4 computed tomography Dr. Muhammad Bin Zulfiqar
4 computed tomography Dr. Muhammad Bin Zulfiqar
 
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component AnalysisAn Ultrasound Image Despeckling Approach Based on Principle Component Analysis
An Ultrasound Image Despeckling Approach Based on Principle Component Analysis
 
Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...
Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...
Lung Nodule Segmentation in CT Images using Rotation Invariant Local Binary P...
 
Segmentation of Lung Lobes and Nodules in CT Images
Segmentation of Lung Lobes and Nodules in CT ImagesSegmentation of Lung Lobes and Nodules in CT Images
Segmentation of Lung Lobes and Nodules in CT Images
 

Similar to Tube and Line and Pneumothorax Visualization Software

DIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus SoftwareDIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus SoftwareCarestream
 
Diagnostic imaging./ cosmetic dentistry training
Diagnostic imaging./ cosmetic dentistry trainingDiagnostic imaging./ cosmetic dentistry training
Diagnostic imaging./ cosmetic dentistry trainingIndian dental academy
 
Diagnostic imaging / dental implant courses by Indian dental academy
Diagnostic imaging / dental implant courses by Indian dental academy Diagnostic imaging / dental implant courses by Indian dental academy
Diagnostic imaging / dental implant courses by Indian dental academy Indian dental academy
 
Diagnostic imaging / dental crown & bridge courses
Diagnostic imaging / dental crown & bridge coursesDiagnostic imaging / dental crown & bridge courses
Diagnostic imaging / dental crown & bridge coursesIndian dental academy
 
IRJET - A Review on Segmentation of Chest Radiographs
IRJET - A Review on Segmentation of Chest RadiographsIRJET - A Review on Segmentation of Chest Radiographs
IRJET - A Review on Segmentation of Chest RadiographsIRJET Journal
 
Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...
Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...
Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...Indian dental academy
 
Tube and Grid Alignment System
Tube and Grid Alignment System Tube and Grid Alignment System
Tube and Grid Alignment System Carestream
 
An Enhanced ILD Diagnosis Method using DWT
An Enhanced ILD Diagnosis Method using DWTAn Enhanced ILD Diagnosis Method using DWT
An Enhanced ILD Diagnosis Method using DWTIOSR Journals
 
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...acijjournal
 
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...acijjournal
 
A novel CAD system to automatically detect cancerous lung nodules using wav...
  A novel CAD system to automatically detect cancerous lung nodules using wav...  A novel CAD system to automatically detect cancerous lung nodules using wav...
A novel CAD system to automatically detect cancerous lung nodules using wav...IJECEIAES
 
Efficient lung air volume estimation using
Efficient lung air volume estimation usingEfficient lung air volume estimation using
Efficient lung air volume estimation usingijcsa
 
Atlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgeryAtlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgerySpringer
 
Atlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgeryAtlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgerySpringer
 
Diagnostic imaging/ oral surgery courses  
Diagnostic imaging/ oral surgery courses  Diagnostic imaging/ oral surgery courses  
Diagnostic imaging/ oral surgery courses  Indian dental academy
 
The International Journal of Engineering and Science (The IJES)
The International Journal of Engineering and Science (The IJES)The International Journal of Engineering and Science (The IJES)
The International Journal of Engineering and Science (The IJES)theijes
 
Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...
Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...
Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...Associate Professor in VSB Coimbatore
 

Similar to Tube and Line and Pneumothorax Visualization Software (20)

DIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus SoftwareDIRECTVIEW EVP Plus Software
DIRECTVIEW EVP Plus Software
 
Dental Diagnostic imaging
Dental Diagnostic imagingDental Diagnostic imaging
Dental Diagnostic imaging
 
Diagnostic imaging./ cosmetic dentistry training
Diagnostic imaging./ cosmetic dentistry trainingDiagnostic imaging./ cosmetic dentistry training
Diagnostic imaging./ cosmetic dentistry training
 
Diagnostic imaging / dental implant courses by Indian dental academy
Diagnostic imaging / dental implant courses by Indian dental academy Diagnostic imaging / dental implant courses by Indian dental academy
Diagnostic imaging / dental implant courses by Indian dental academy
 
Diagnostic imaging / dental crown & bridge courses
Diagnostic imaging / dental crown & bridge coursesDiagnostic imaging / dental crown & bridge courses
Diagnostic imaging / dental crown & bridge courses
 
IRJET - A Review on Segmentation of Chest Radiographs
IRJET - A Review on Segmentation of Chest RadiographsIRJET - A Review on Segmentation of Chest Radiographs
IRJET - A Review on Segmentation of Chest Radiographs
 
Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...
Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...
Diagnostic imaging /certified fixed orthodontic courses by Indian dental acad...
 
Tube and Grid Alignment System
Tube and Grid Alignment System Tube and Grid Alignment System
Tube and Grid Alignment System
 
CT dose reduction
CT dose reductionCT dose reduction
CT dose reduction
 
An Enhanced ILD Diagnosis Method using DWT
An Enhanced ILD Diagnosis Method using DWTAn Enhanced ILD Diagnosis Method using DWT
An Enhanced ILD Diagnosis Method using DWT
 
2011BMES Abstract radDE5DE
2011BMES Abstract radDE5DE2011BMES Abstract radDE5DE
2011BMES Abstract radDE5DE
 
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
 
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
AUTOMATED SEGMENTATION OF FLUORESCENT AND FUNDS IMAGES BASED ON RETINAL BLOOD...
 
A novel CAD system to automatically detect cancerous lung nodules using wav...
  A novel CAD system to automatically detect cancerous lung nodules using wav...  A novel CAD system to automatically detect cancerous lung nodules using wav...
A novel CAD system to automatically detect cancerous lung nodules using wav...
 
Efficient lung air volume estimation using
Efficient lung air volume estimation usingEfficient lung air volume estimation using
Efficient lung air volume estimation using
 
Atlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgeryAtlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgery
 
Atlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgeryAtlas of advanced endoaortic surgery
Atlas of advanced endoaortic surgery
 
Diagnostic imaging/ oral surgery courses  
Diagnostic imaging/ oral surgery courses  Diagnostic imaging/ oral surgery courses  
Diagnostic imaging/ oral surgery courses  
 
The International Journal of Engineering and Science (The IJES)
The International Journal of Engineering and Science (The IJES)The International Journal of Engineering and Science (The IJES)
The International Journal of Engineering and Science (The IJES)
 
Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...
Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...
Automatic Segmentation of the Lumen of the Carotid Artery in Ultrasound B-mod...
 

More from Carestream

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Carestream
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsCarestream
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-rayCarestream
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentCarestream
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityCarestream
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Carestream
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Carestream
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesCarestream
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seramCarestream
 
Cloud Security
Cloud Security Cloud Security
Cloud Security Carestream
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Carestream
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareCarestream
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundCarestream
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Carestream
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Carestream
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Carestream
 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16Carestream
 
Casa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital ItalyCasa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital ItalyCarestream
 
The Power of Together
The Power of TogetherThe Power of Together
The Power of TogetherCarestream
 
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging StrategyWhitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging StrategyCarestream
 

More from Carestream (20)

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-ray
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray Equipment
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image Quality
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seram
 
Cloud Security
Cloud Security Cloud Security
Cloud Security
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
 
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA163 Reasons to Visit Carestream at RSNA2016 #RNSA16
3 Reasons to Visit Carestream at RSNA2016 #RNSA16
 
Casa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital ItalyCasa di Cura Sileno Hospital Italy
Casa di Cura Sileno Hospital Italy
 
The Power of Together
The Power of TogetherThe Power of Together
The Power of Together
 
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging StrategyWhitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
Whitepaper: Leveraging the Cloud to Enhance an Enterprise Imaging Strategy
 

Recently uploaded

Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Dipal Arora
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan 087776558899
 
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...soniyagrag336
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableSteve Davis
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...TanyaAhuja34
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...rajnisinghkjn
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxSwetaba Besh
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...dishamehta3332
 
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service AvailableLucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Availablesoniyagrag336
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Janvi Singh
 
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...dilbirsingh0889
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxSwetaba Besh
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...Rashmi Entertainment
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationMedicoseAcademics
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsMedicoseAcademics
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...amritaverma53
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...call girls hydrabad
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryJyoti singh
 

Recently uploaded (20)

Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Wayanad Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
 
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service AvailableLucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
Lucknow Call Girls Just Call 👉👉8630512678 Top Class Call Girl Service Available
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 

Tube and Line and Pneumothorax Visualization Software

  • 1. White Paper | CARESTREAM Tube & Line Visualization Software, CARESTREAM Pneumothorax Visualization Software Carestream’s companion-view processing for improved visualization of tubes, lines, and pneumothoraces in digital, portable chest radiography Introduction Carestream has implemented companion views in its digital radiography systems. A companion view is designed to complement the standard processed radiographic image delivered from the digital radiography capture modality to PACS, to provide an additional rendering tailored for the visual interpretation needed for a specific diagnostic or clinical purpose. Two companion views are available in Carestream products for chest radiography: one for the optimal visualization of tubes and lines in chest radiographs (CARESTREAM Tube & Line Visualization Software), and the other for enhancing the conspicuity of a pneumothorax (CARESTREAM Pneumothorax Visualization Software). Companion Views Carestream’s digital radiography image processing software is designed to automatically render images with an appearance suitable for a radiologist to assess a patient’s condition with respect to the spectrum of potential abnormalities. A single chest image typically comprises a wide range of exposure (gray) levels that exceed the number of brightness levels available on a PACS workstation monitor. Therefore, an appropriate overall rendering may not necessarily display the full grayscale fidelity for any particular indication. In other words, there are simply more gray levels in the raw, captured image than what the PACS monitor can display in a single, static presentation. This effect exists for radiologists who typically read images on high-resolution displays in rooms with reduced ambient lighting—but the effect is exacerbated for ICU clinicians who may need to interpret portable chest radiographs at the point of care, on lower-resolution monitors under bright light, which is common in the ICU setting. For instance, a portable chest radiograph may be properly rendered and displayed with excellent overall diagnostic quality, yet it may still be challenging to localize tube and line tips in underpenetrated regions, such as in the mediastinum and subdiaphragm. In order to preserve global contrast and brightness for the overall image, gray levels may be quantized in the underpenetrated regions, which will cause some degree of detail contrast loss in the mediastinum (Figure 1 – image on left). To address this issue, Carestream has developed a unique and robust image-processing algorithm that specifically highlights tube and catheter features 1 (CARESTREAM Tube & Line Visualization Software). To generate the companion view, the image is first decomposed into eight spatial frequency bands. These bands are individually manipulated using non-linear functions designed to achieve an equalized contrast enhancement of image details, while at the same time minimizing noise. The manipulated frequency bands are then reconstructed and the output image is passed through a grayscale look-up table for display. This method boosts the spatial frequencies required to enhance the conspicuity of tubes and lines while it suppresses the spatial frequencies containing noise. By enhancing the tube and line features while simultaneously suppressing noise, this method also works robustly for images captured at low exposure levels and for images of large patients containing excessive X-ray scatter. While the companion view presentation may not be suitable for general diagnostic interpretation, the image features necessary for interpretation of the tube and line structures are significantly enhanced (Figure 1 – image on right).
  • 2. White Paper | CARESTREAM Tube & Line Visualization Software, CARESTREAM Pneumothorax Visualization Software 2 Figure 1 – Left: a portable chest X-ray image processed using default automatic processing. Right: the same image processed using the Tube & Line Visualization Software. The two images are delivered together to PACS. For some abnormalities, the visual characteristics of the key diagnostic features may be masked by, or confounded with, normal anatomical structures. This is a very different effect from the insufficient grayscale resolution issue described above. This effect can be more readily thought of as visual interference between the features corresponding to the abnormality and the normal anatomical structures. For instance, consider a pneumothorax, defined as the presence of air in the pleural space. A pneumothorax is visualized by a delineation of the lung edge, which is displaced from the chest wall by the air contained between the visceral and parietal pleura. Depending on positioning, there will be no visible lung markings in this region. The delineation of the lung edge, and the absence of lung markings, may be difficult to perceive and be potentially masked by the radiographic presentation of the scapula or ribs. The conspicuity of the pneumothorax features may be further dampened by insufficient grayscale resolution of the PACS display. In an approach analogous to that used in tube- and line- enhanced visualization, Carestream has developed a companion view for pneumothorax that is specifically designed to enhance the appearance of the associated features (CARESTREAM Pneumothorax Visualization Software). The algorithm differs from the tube- and line-enhancement technique. Instead of focusing on boosting the spatial frequencies associated with the diagnostic features of interest, the pneumothorax algorithm uses a texture filter that enhances the appearance differences between the characteristic markings for the regions inside and outside the lung. On page 3, Figure 2 (image on left) shows an example of a large pneumothorax (the lung edge is indicated by arrows). The image on the right in Figure 2 shows an insert that illustrates the effect of the enhancement resulting from the application of the texture filter.
  • 3. White Paper | CARESTREAM Tube & Line Visualization Software, CARESTREAM Pneumothorax Visualization Software 3 Figure 2 – Left: a portable chest X-ray image illustrating a large pneumothorax. Right: an insert showing the texture differences resulting from processing the image using the pneumothorax-enhancement filter. The images shown in Figure 3 compare a portable chest X-ray image presented using the default processing (on the left) against the same image processed with the pneumothorax-enhancement filter (on the right). Figure 3 – Left: a portable chest X-ray image of a patient with a pneumothorax. Right: the companion view processed using the pneumothorax texture-enhancement filter. The lung edge is indicated by arrows.
  • 4. White Paper | CARESTREAM Tube & Line Visualization Software, CARESTREAM Pneumothorax Visualization Software www.carestream.com © Carestream Health, Inc., 2012. CARESTREAM is a trademark of Carestream Health. CAT 200 0012 7/12 Companion views improve interpretation confidence and radiologist reading efficiency PACS workstations provide window/level adjustment capabilities to help address the generic shortcomings of insufficient luminance dynamic range on PACS monitors, and for enhancing the contrast of subtle features. However, interactive window and leveling is time-consuming and degrades radiologist reading efficiency. The enhanced visualization delivered to PACS as a companion view, together with the default processed image, makes it possible for radiologists and clinicians to rapidly toggle between the standard visualization optimized for general interpretation, and the customized visualization that is optimized for a specific diagnostic task. Because the companion view can be toggled with the default processed image, radiologists can more easily make their interpretations with minimal workstation interaction. A study was conducted with both radiologists and pulmonary ICU physicians to assess the utility of the tube and line companion view in terms of interpretation and reading efficiency 2 . The companion view method was evaluated by two reader groups, radiologists and pulmonary ICU physicians. Each reader group evaluated the images under viewing conditions consistent with their typical respective reading environment. The companion view method was shown to provide improved interpretation confidence and improved diagnostic efficiency for radiologists (30% less reading time), and more confident interpretation for pulmonary ICU physicians in the same amount of time, when used in lieu of window and level adjustments at the PACS workstation. Summary Carestream now offers companion-view processing that is designed to enhance the visualization of tubes, lines, and pneumothoraces in chest images captured using Carestream CR and DR modalities. These companion views complement the standard default rendering delivered from the capture modality to PACS with a second image that is processed to enhance the features for a particular diagnostic or clinical indication. The Tube & Line Visualization Software has been shown to improve interpretation confidence for radiologists and pulmonary ICU physicians, and improve reading efficiency for radiologists. References 1. Couwenhoven ME, Senn RA, Foos DH. Enhancement method that provides direct and independent control of fundamental attributes of image quality for radiographic imagery. SPIE Medical Imaging Proceedings; 5367: 474-481, 2004. 2. Foos DH, Yankelevitz DF, Wang X, Berlin D, Zappetti D, Cham M, Sanders A, Novak Parker K, Henschke CI. Improved visualization of tubes and lines in portable intensive care unit radiographs: a study comparing a new approach to the standard approach. Clinical Imaging Volume 35, Issue 5, September–October 2011, Pages 346–352.