2. Main points
• Specifications
• Device presentation
• Main Instrument
• Installation
• Settings
• Maintanance
• Troubleshooting
• Spare parts
3. Specifications
General
• 216 x 205 x 75 mm ; ~1 kg
• 17-32°C ; 80% Relative Humidity
• 100-240 VAC and 50/60 Hz; 45W; 150VA
• High capacity
• Calibrations -up to 6 calibration levels
• QC - 12 different control names
11. Installation
• Table - 220 mm (W) x 210 mm (D)
• 100 mm back space – ventilation, cables
• Temperature 17-32°C, humidity max 80%
• avoid direct sunlight, dust . . . .
Put the supply tightly to instrumet
Plug the adaptor to wall socket
Instal paper roll and close lid
12. Settings
User level
• Remove all QC data
• Remove all results
• Set to factory data – ECL will
preserve stored quantities
Service level
• Usage all of user methods
• Edit device properties
• Test peripheries
14. Settings
USB
• Test USB communication
• Mass storage test – r/w to USB
• USB keyboard test – w witk kb
• USB B test – virtual COM port
PRINTER
• Test motor and printer
• Printer temt – temperature of
head
17. Maintanance
Daily Beginning of day
• Check any damage, electrical and LIS connections are secure
• Swipe dust and any other liquids from surface wet and dry towel
• turn ON and wait for temperature 37+/-0.5°C
End of day
• Turn OFF
• Remove all cuvets and reagent tubes
• Clean surface
Note: For decontamination of the surface use a 1+1 solution of alcohol and water.
Note: Do not use strong bleach that could demage cover plastics and other.
18. Troubleshooting
• No back light
• Blank LCD / faulty display
• Touch screen error
• Printer error
• Time and data lost
• RFID error
• LED / Detector error
• Power error
19. No back light
Measeure voltage 15 – 22V DC
• OK – replace LCD
• NOT OK – replace main board
20. Blank LCD / faulty display
Check connection
• NOT OK – replace LCD
23. Time and data lost
Data are not stored
• Replace CR2032 battery
24. RFID error
Check the tag in menu
• Check connection
• NOT OK – replace RFID panel
25. LED / Detector error
Check If LEDS and detector are OK
• Check visually
• Check by controls
• NOT OK - replace LED
• Measure and calibrate
temperature
Incubator test
Measurement test