Training material of SUNBRIGHT brand.
#8-Parameter 15-inch Patient monitor model SUN-603.
Explains a brief feature of functions that embedded in SUN-603, with each of 8-parameters guide to use, possible modes, and the accessories provided.
Training material that suitable for healthcare professional & biomedical engineering personnel.
2. INTRODUCTION
- 8 parameter patient monitor
- Used for adult, neonate and pediatric
- Mobile (2.8 kg)
- Powered with 50 Hz, 220 V
- 12.1 inch with TFT monitor
- Displaying real-time data & waveform
- Inserted with 12 V battery
Altica Sdn. Bhd.
3. FRONT VIEW
12 inch SCREEN
-for display
NOT TOUCH
SCREEN!
KNOB
-to navigate for use
ON/OFF BUTTON
-as a switch
USE THUMB FOR
PROLONGED LIFESPAN
FUNCTIONAL KEY
-for user operation Altica Sdn. Bhd.
4. BACK VIEW
NET CSSI PORT
-to connect with wifi router
ISOPOTENTIAL TERMINAL
-for electrical safety test
USB PORT
-to update software
DON’T PUT ANY CHARGING CABLE
HERE!
POWER CORD
-for charging the unit
Altica Sdn. Bhd.
8. ASSEMBLING
1. Set the trolley to
desired height
3. Push main unit
until halfway
2. Disconnect all
screw from plating
Altica Sdn. Bhd.
9. ASSEMBLING
4. Pull (not twist!) stopper
while pushing main unit
5. After stopper “clicked”,
reconnect all screw to plating
Altica Sdn. Bhd.
10. FEATURE
SILENCE
-to silent alarm until
press again
PAUSE
-to silent alarm
(2 minutes)
FREEZE
-to hold the interface
(press again to
continue)
NIBP
-to inflate NIBP cuff
(press again to stop)
MENU
-to open config
PRINT
-to activate thermal
printing
(optional)
Altica Sdn. Bhd.
11. FEATURE - part 1
To inflate NIBP cuff
(press again to stop)
To start thermal
printing
(optional)
To set all parameter setting to factory setting
Data log will not deleted
All customized setting must be set back
Altica Sdn. Bhd.
12. FEATURE - part 2
Control interval of trend
Record data for print
(only with THERMAL
PRINTER)
Altica Sdn. Bhd.
13. FEATURE - part 3
Control parameter of trend
Move axis in second
Move axis in minutes
Altica Sdn. Bhd.
14. FEATURE - part 4 Bed number
(1-200)
Sex
-female
-male
Type of patient
-adult
-paediatric
-neonates
To set pacemaker
-off
(patient without
pacemaker)
-on
(patient with
pacemaker)
To put new patient setting
Will lose previous data log to avoid
misinformation of patient! Altica Sdn. Bhd.
15. FEATURE - part 5
FREEZE
-to hold the interface
(press again to continue)
Knob loudness
-low
-medium
-high
Pulse rate loudness
(higher value, louder)
ECG beat loudness
(higher value, louder)
Altica Sdn. Bhd.
16. FEATURE - part 6
Alert value
above max
Alert value
below min
Can setup alarm for all
parameter
EXCEPT IBP & CO2
(go to respective parameter to set
alarm limit)
Altica Sdn. Bhd.
17. FEATURE - part 7
Choose specific parameter
alarm record
Set duration for
recent alarm
Click on yellow bar to view trend graph
when alarm triggered
Altica Sdn. Bhd.
18. FEATURE - part 8
Set drug name
Cannot add, only the one
set by manufacturer
Set weight of patient
1. Set drug name
2. Set weight
3. Wait for value to
auto-calculate
For new drug name/weight, close the
calculator first and open it back
Altica Sdn. Bhd.
19. FEATURE - part 9
Set baseline for patient PQRS wave
ONLY WHEN THE PATIENT ALREADY
STABLE!
To view recent arrhythmia alarm
ONLY ABLE TO CLICK IF TERE IS
ALARM TRIGGER
Set type of arrhythmia
-Asystole
-Vfib/Vtac
-R on T
-VT > 2
-Couplet
-PVC
-Bigeminy
-Trigeminy
-Tachy
-Brady
-PNC
-PNP
-Missed
beats
Altica Sdn. Bhd.
21. FEATURE - part 11
To set main unit into
standby mode
Use for fast charging/
Not in use mode
STANDBY MODE STILL CONSUME BATTERY!
Altica Sdn. Bhd.
22. FEATURE - part 12
MUTE
-to silent alarm until
press again
PAUSE
-to silent alarm
(2 minutes)
HOME
-to refresh the screen
Altica Sdn. Bhd.
23. FEATURE - part 13
DRUG CALCULATOR OPEN/EXIT DEMO
Key: 101
To set all parameter
setting to factory
setting
Altica Sdn. Bhd.
24. FEATURE- part 14 Change
screen
style
To set main
unit clock
MUST BE SET
WHEN FIRST
TIME USE!
Knob loudness
-low
-medium
-high
Screen brightness
(higher value,
brighter)
Record data for print
(only with THERMAL
PRINTER) Altica Sdn. Bhd.
25. FEATURE - part 15 To check software version
To detect is the port are detected by
software? YES = /
Altica Sdn. Bhd.
26. PARAMETER - ECG
ECG monitoring produces a
continuous waveform of the
patient's cardiac electrical
activity to enable an
accurate assessment of his
current physiological state.
Provide:
5-lead set x2
ECG-electrode set x2
Altica Sdn. Bhd.
27. PARAMETER - ECG
To observe heart rate from:
-ECG
-SPO2
-AUTO (alternate between SPO2
& ECG)
To observe parameter:
-Lead I, II, III
-aVR, aVL, aVF
-V1, V2, V3, V4, V5, V6
Filter mode:
-monitoring
-diagnosis
-surgery Altica Sdn. Bhd.
28. PARAMETER - ECG
To set size of graph
(higher value, larger size)
x1, x2, x0.25, x0.5
To choose type of lead
-5 lead
-3 lead
ENSURE THE LEAD TYPE SAME AS
THE NUMBER OF ELECTRODE !
For further option
To set speed of graph
(higher value, faster)
12.5, 25.0, 50.0
To set back all ECG setting to factory setting
Altica Sdn. Bhd.
29. PARAMETER - ECG
To set pacemaker
-off
(patient without
pacemaker)
-on
(patient with
pacemaker)
To set display
-normal
-ECG-specialized
screen
To control beat loudness
(higher, louder)
Altica Sdn. Bhd.
30. PARAMETER - SPO2
SpO2 Plethysmogram
measurement is employed to
determine the oxygen
saturation of hemoglobin in
the arterial blood.
Provide:
5 PIN Nellcor Adult
SPO2 probe x2
Altica Sdn. Bhd.
31. PARAMETER - SPO2 Loudness of pulse rate
(higher,louder)
To set speed of graph
(higher value, faster)
12.5, 25.0, 50.0
Control sensitivity of SPO2 sensor
(higher, faster reading)
high , med, low
To set back all SPO2 setting to factory setting
Altica Sdn. Bhd.
32. PARAMETER - RESP
The monitor measures the thoracic impedance between two ECG
electrodes on the patient's chest. Changes in the impedance due to
thoracic movement produce the Resp waveform on the monitor screen.
Altica Sdn. Bhd.
33. PARAMETER - RESP
To refer which
electrode for
respiratory rate
-LL
-LA
To set size of graph
(higher value, larger size)
1, 2, 0.25, 0.5
To set speed of graph
(higher value, faster)
12.5, 25.0, 50.0
ENSURE THE ECG PORT ARE CONNECTED!
To set back all RESP setting to factory setting
Altica Sdn. Bhd.
35. PARAMETER - TEMP
To change the unit
of temperature
parameter
-Celcius
-Fahrenheit
To set back all TEMP setting
to factory setting
Altica Sdn. Bhd.
36. PARAMETER - IBP
IBP monitoring measures direct
blood pressure of one selected
blood vessel through two
channels, and displays two BP
waveforms measures direct
blood pressure.
Provide:
IBP cable x2
Pressure transducer set x1 Altica Sdn. Bhd.
37. PARAMETER - IBP
To change output
-systole, mean, diastole
-mean ONLY
To set speed of graph
(higher value, faster)
12.5, 25.0, 50.0
To change unit
-mmHg
-kPa
To observe parameter:
-ART
-PA
-CVP
-RAP
-LAP
-ICP
-P1
-P2
To set back all IBP setting to factory setting
For further option
Altica Sdn. Bhd.
38. PARAMETER - IBP
1. Connect IBP
cable to
main unit
3. Ensure transducer
is OFF to patient &
the pressure is fully
flushed
2. Connect IBP
cable with
transducer
Step for IBP zeroing (part 1)
Altica Sdn. Bhd.
39. PARAMETER - IBP
Step for IBP zeroing (part 2)
4. Press IBP > other setup > IBP zero
5. Select Channel that inserted
6. Wait until successful zero
Altica Sdn. Bhd.
40. PARAMETER - CO2
CO2 monitoring measures Patients’
Gas Circuit CO2 pressure, show one
channel CO2 waveform and
EtCO2,FiCO2,AwRR.
Provide:
Sidestream EtCO2 module
x1
Water filter x1
L connector x1
Adult nasal cannula x1 Altica Sdn. Bhd.
41. PARAMETER - CO2
To change unit
-mmHg
-kPa
For further option
To set speed of graph
(higher value, faster)
12.5, 25.0, 50.0
Altica Sdn. Bhd.
42. PARAMETER - CO2
1. Connect
EtCO2
module to
main unit
3. Connect the filter
with cannula
2. Push and twist
the filter under
EtCO2 module
Step for CO2 zeroing (part 1)
Altica Sdn. Bhd.
43. PARAMETER - CO2
5. Press ZERO and
change gas type into
“ROOM AIR”
4. Press CO2 > OTHER SETUP
and set the work mode to
“MEASUREMENT”
6. Press CO2 ZERO and
wait for the graph to
return normal
Step for CO2 zeroing (part 2)
Altica Sdn. Bhd.
45. PARAMETER - NIBP
To change unit
-mmHg
-kPa
To set auto pump
-manual
-1, 2, 3, 4, 5, 10, 15,
30, 60, 90, 120, 180,
240, 480 minutes
To set max
pressure
discharge in cuff
To refresh NIBP
setting
To obtain as
many reading as
possible in a
interval
To produce
pressure for leak
test
To set back all NIBP setting to factory setting
Altica Sdn. Bhd.
46. Muhammad Ali Aizat
Certified biomedical technician
Altica Sdn. Bhd.