Call Girls Kakinada Just Call 9907093804 Top Class Call Girl Service Available
Creative Patient Monitor
1. CREATIVE UP-7000 PATIENT MONITOR
Gradian Health Systems
Simulation-Based Product Training
Version 2.0
2. Agenda
2
UAM Simulation-based Training I Lecture Content | Overview of Gradian
I. Creative Patient Monitor
a. Product Description
b. Accessory Connectors
c. Display Features
d. Back Panel
e. Alarm Status
3. Creative Patient Monitor: Product Description
3
UAM Simulation-based Training I Lecture Content | Creative Patient Monitor
The Creative Monitor supports:
• Electrocardiography (ECG)
• Temperature measurement
• Non-invasive blood pressure measurement
• Pulse oximetry (SpO2)
Monitoring Specs
• 5-channel waveform display: 3 x ECG, RESP,
SpO2
• 7 ECG waveforms displayed simultaneously
• Arrhythmia analysis and ST segment analysis
• Adult / Pediatric / Neonatal modes
Other Features
• 24-hour ECG waveform storage and recall
• Visual and audible alarms
• Runs on internal battery for up to 4 hours
• Protection against interference from defibrillator,
electrosurgical unit
• Invasive blood
pressure monitoring
• Print capabilities
THE CREATIVE PATIENT
MONITOR DOES NOT
PROVIDE
5. Creative Patient Monitor: Display Features
5
UAM Simulation-based Training I Lecture Content | Creative Patient Monitor
Battery indicator
Power switch
Alarm indicator
Navigation knob
Display mode
NIBP start / stop
AC indicator
ECG lead selector
Alarm silence
Waveform freeze
Display
6. Creative Patient Monitor: Back Panel
6
UAM Simulation-based Training I Lecture Content | Creative Patient Monitor
Fuses
AC power connector
Grounding post
7. Creative Patient Monitor: Alarm Status
7
UAM Simulation-based Training I Lecture Content | Creative Patient Monitor
Indicator Color Alarm Priority Alarm Event
Red Flashing High Exceeds limits, pulse stop, suffocation
Orange Flashing Medium Leads and probe off, critical arrhythmias
Orange Solid Low Other arrhythmia phenomenon
Green Solid None N/A
Note: Should the user decide not to
connect the patient to ECG leads,
simply ensure that the ECG leads
are disconnected from the patient
monitor prior to turning it on. This will
prevent the monitor from presenting
the ECG Leads Disconnect Alarm.