Basic Approach to Dissolution Method Development – Challenges and Regulatory Issues
1. Dr. Harshal Ashok Pawar
Assistant Professor & HOD (Quality Assurance),
Dr. L. H. Hiranandani College of Pharmacy, Ulhasnagar
Email : harshal.dlhhcop@gmail.com , hapkmk@rediffmail.com
2. CONTENTS
Theoretical aspect
Instrumentation
Pre-requisite
Selection of Dissolution medium
Optimization of dissolution parameters Optimization of dissolution parameters
Specifications /Acceptance criteria
Comparison of Dissolution Profiling
Biovaivers
4. DISSOLUTION
In the body, a pharmaceutical
active ingredient must be “in
solution” before it can be
absorbed by the blood andabsorbed by the blood and
ultimately carried to the
receptor site to render a
therapeutic effect (in vivo).
Dissolution is the process
by which that active
ingredient enters into
solvent to yield a solution.
5. DISSOLUTION SOLUBILITY
Dissolution rate is defined as
the amount of solid substance
that goes into solution per
unit time under standard
Absolute solubility is defined
as the maximum amount of
solute dissolved in a given
solvent under standardunit time under standard
conditions of temperature,
pH, solvent composition and
constant solid surface area.
solvent under standard
conditions of temperature ,
pressure and pH.
It is a dynamic process. It is a static process.
6. MECHAMISM OF DISSOLUTION
Initial mechanical lag
Wetting of dosage
form
Penetration of
dissolution medium
Disintegration Disintegration
Deaggregation
Dissolution
Occlusion of some
particles
7. Dissolution process of solid dosage Forms :
DISINTEGRATION DISSOLUTION
DISSOLUTION ABSORPTION
IN-VIVO
TABLETS OR
CAPSULES
GRANULES OR
AGGREGATES
DRUG IN
SOLUTION
(IN-VITRO OR IN-
DRUG IN
BLOOD,OTHER
FLUIDS,AND
IN-VIVO
DISAGGREGATION
DISSOLUTION
AGGREGATES
FINE PARTICLES
(IN-VITRO OR IN-
VIVO)
FLUIDS,AND
TISSUES
8. Dissolution Testing
Dissolution is one the three primary tests used to
release a finished drug product:
Assay – determines the overall potency of the batch and ensures
the accuracy of the finished drug product.
Dose Uniformity – determines the consistency among the Dose Uniformity – determines the consistency among the
individual dosage units and ensures the precision of the
manufacturing process.
Dissolution – ensures that the performance of the finished drug
product is consistent with the release rates of the API as determined
in bioavailability studies during the clinical trials.
9. Which product to be tested?
USP <1088>:
“No product, including suspensions and chewable tablets,
should be developed without dissolution or drug release
characterization where a solid phase exists.”
and
“Dissolution testing is required for all solid oral Pharmacopeial“Dissolution testing is required for all solid oral Pharmacopeial
dosage forms in which absorption of the drug is necessary
for the product to exert the desired therapeutic effect.
Exceptions are for tablets meeting a requirement for
completeness of solution or for rapid (10 to 15 minutes)
disintegration for soluble or radiolabeled drugs.”
10. NEED FOR DISSOLUTION TESTING
Development and Optimization of dosage forms.
Batch to batch drug release uniformity.
Ensures quality, safety, efficacy and stability of the product.
Evaluation of IVIV Correlation / bioavailability .
To support waiver for bioequivalence requirement.
For assessing Pre and Post approval changes in manufacturing
process, or formulation
11. Dissolution and Drug Development
Drug
Clinical trial
(IVIV)
Manufacturing
(Quality
control)
Discovery ( Little to
no dissolution)
Characterization
(Intrinsic
dissolution)
Drug
development
(DMD)
control)
13. I.P U.S.P B.P E.P
TYPE 1 Paddle
apparatus
Basket
apparatus
Basket
apparatus
Basket
apparatus
TYPE 2 Basket
apparatus
Paddle
apparatus
Paddle
apparatus
Paddle
apparatus
TYPE 3 Reciprocating
cylinder
Flow through
cell
Flow through
cell
TYPE 4 Flow through
cell
TYPE 5 Paddle over
disk
TYPE 6 Rotating
cylinder
TYPE 7 Reciprocating
disk
14. APPARATUS-1(ROTATING BASKET)
DESIGN:
Vessel: -Made of borosilicate glass.
-Semi hemispherical bottom
-Capacity 1000ml
Shaft : -Stainless steel type 316
-Rotates smoothly without
significance wobble(100 rpm)
-Speed regulator
Water bath:-Maintained at 37±0.5ºC
USE: Tablets, capsules, delayed release
suppositories, floating dosage forms.
16. Advantages
Full pH change during the test
Can be easily automated
which is important for routine
investigations.
Disadvantages
Basket screen is clogged with
gummy particles.
Hydrodynamic dead zone
under the basket
Degassing is particularly
important
Mesh gets corroded by HCl solution.
17. APPARATUS-2 (PADDLE)
DESIGN:
Vessel: -Same as basket apparatus
Shaft: -The blade passes through the shaft so that the bottom of the
blade fuses with bottom of the shaft.
Stirring elements: -Made of tefflon
For laboratory purposeFor laboratory purpose
-Stainless steel 316
Water-bath: -Maintains at 37±0.5°C
Sinkers : -Platinum wire used to prevent
tablet/capsule from floating
18. Advantages
Easy to use
Robust
Can be easily automated which is important for
routine investigations
Disadvantages Disadvantages
pH/media change is often difficult
Sinkers for floating dosage forms
20. APPARATUS-3(RECIPROCATING CYLINDER)
Tester was designed to test the dissolution rates of extended release
products or any dosage form requiring release profiling at multiple pH
levels.
DESIGN:
Vessel: -Set of cylindrical flat bottom glass vessels
-Set of reciprocating cylinders
-stainless steel fittings(type 316) and
screens made of nonsorbing orscreens made of nonsorbing or
non-reactive materials.
Agitation type: -Reciprocating at 5-35 rpm
Volume of dissolution medium:-200-250ml
Water bath:- Maintain at 37±0.5°C
USE: Chewable Tablets, beads, controlled and
Extended release formulations, Soft gel cap,
Non Disintegrating type products
21. METHOD(Reciprocating cylinder):
Place the stated volume of dissolution medium in each vessel of the
apparatus, assemble the apparatus, equilibrate the dissolution
medium to 37±0.5 and remove the thermometer
Place one dosage form unit in each of the cylinders taking care to
exclude the air bubbles from the surface of each dosage unit and
immediately operate the apparatus as specified in the monograph.
During the upward and downward stroke, the reciprocating cylinder During the upward and downward stroke, the reciprocating cylinder
moves through a total distance of 9.9 to 10.1cm.
Within the time interval specified, raise the cylinders and withdraw a
portion of the solution under test from a zone midway between the
surface of the dissolution medium and bottom of each vessel.
22. Advantages
Easy to change the pH
pH-profiles
Hydrodynamics can be
directly influenced by
varying the dip rate
Disadvantages
Small volume (max. 250 ml)
Little experience
Limited data
23. APPARATUS-4 (FLOW THROUGH CELL)
DESIGN:
Reservoir : -For dissolution medium
Pump : -Forces dissolution medium through cell
-Holding a sample
-Flow rate 10-100ml/min
-Laminar flow is maintained
-Peristaltic/centrifugal pumps are not recommended
Water bath:- Maintain at 37±0.5°C
USE:
Modified-release dosage forms and immediate-release dosage forms
Soft gelatin capsules, beaded products, suppositories, or injectable-depot dosage forms
Suspension-type extended-release dosage forms for oral or parenteral use, or ocular
application.
24. METHOD(Flow through cell):
The flow through cell is transparent & inert mounted vertically
with filters.
Standard cell diameters are 12 & 22.6 mm.
The bottom cone usually filled with glass beads of 1 mm
diameter.
Tablet holder used for positioning special dosage form e.g. inlay
tablets.
Place the glass beads into the cell as specified in the monograph. Place the glass beads into the cell as specified in the monograph.
Place one dosage unit on top of the beads or on a wire carrier.
Assemble the filter head and fix the parts together by means of a
suitable clamping device.
Introduce by the pump of the dissolution medium warmed to
37±0.5 through the bottom of the cell to obtain the flow rate
specified and measured with an accuracy of 5%.
Collect the eluate by fractions at each of the times stated.
25. Advantages
Easy to change media pH
pH-profile possible
Sink conditions maintained.
Different modes
a) open system
b) closed system
Disadvantages
Deaeration necessary
High volumes of media
Labor intensive
26. Cell types:
Tablets 12 mm Tablets 22.6 mm Powders / Granules Implants Suppositories /
Soft gelatincapsules
28. APPARATUS-5(PADDLE-OVER-DISK)
DESIGN:
Vessel
Shaft
Stirring elements- rotating speed 25-50 rpm
Sample holder:-disk assembly that hold a product in such a way Sample holder:-disk assembly that hold a product in such a way
that release surface is parallel with paddle
-Paddle is directly attached over disk assembly
-Samples are drawn between surface off the medium
and top of the paddle blade
Volume:900ml
Temperature:32°C (Similar to skin)
29. USE: Transdermal patches, ointments, floaters , emulsions.
Modification: Disk design and volume
Advantages:
Easy to handle
Sink conditions are maintained.
Membrane effect is minimum.
i.e. drug is placed on a disc at the bottom.
Disadvantages: Disadvantages:
Disk assembly restricts the patch size
17 mesh is standard(others available)
Accommodates patches up to 90mm.
30. METHOD(Paddle over disk)
This method is used for testing the release of drugs from
transdermal products.
The apparatus consists of a sample holder or disc assembly that
holds the product.
The entire preparation is placed in a dissolution flask filled with
specified medium maintained at 32ºC.specified medium maintained at 32ºC.
The paddle is placed directly over the disc assembly.
The disk assembly holds the system flat and is positioned such that
release surface is placed parallel with the bottom of the paddle
blade. Vessel is covered to minimize evaporation during test.
Samples are drawn midway between the surface of dissolution
medium and the top of the paddle blade at specified times.
31. APPARATUS-6(ROTATING CYLINDER)
DESIGN:
Vessel:- In place of basket, cylinder is used.
Shaft :-Stainless steel 316
Sample :- Mounted to cuprophan (inner porous cellulosic material)
an entire system adheres to cylinder.an entire system adheres to cylinder.
- Dosage unit is placed in cylinder and release from side out.
Water-bath: maintained at 32±0.5°C
USE:
Transdermal patches cannot be cut into small size.
Solid dosage forms, pH profile , small volumes
32. METHOD( Rotating cylinder):
Use the assembly from apparatus 1 except to replace the basket and
shaft with a stainless steel cylinder stirring element.
The temperature is maintained at 32±0.5°C.
The dosage unit is placed on the cylinder with side out .
The dosage unit is placed to the exterior of the cylinder such that
long axis of the system fits around the circumference of the cylinderlong axis of the system fits around the circumference of the cylinder
and removes trapped air bubbles.
Place the cylinder in the apparatus and immediately rotate at the rate
specified in the individual monograph.
Samples are drawn midway between the surface of the dissolution
medium and the top of the rotating cylinder for analysis.
33. Carefully apply the
adhesive coated side of the
system to the exterior of
the cylinder with the long
axis of the system fittingaxis of the system fitting
around the circumference
of the cylinder.
34. Rotating cylinder:
Advantages: -Equipment (apparatus 1)available with the
manufacturers can be used with modification as apparatus 6.
Disadvantages:-Large volume of medium is required.
-Drug gets diluted & causes difficulties in analysis
-Difficult to clean the cylinder.
35. APPARATUS-7(RECIPROCATING-DISK)
DESIGN:
Vessel:-Flat bottomed cylindrical vessel
-Volume of dissolution medium
Shaft :
Sample : -Placed on disk shaped holders
Agitation :-Reciprocation
shaft
disk
dissolution medium
constant temp
water bath
Agitation :-Reciprocation
-Reciprocating frequency 30 cycle/sec
Water-bath:-Maintain at 32±0.5°C
USE:
Non-disintegrating, oral modified-release dosage forms, stents, and
implants
Transdermal dosage forms.
36. METHOD(Reciprocating disk):
The assembly consists of a set of volumetrically calibrated solution
containers made of glass or suitable inert material, a motor , a drive
assembly used to reciprocate the system vertically.
The samples are placed on the disk shaped holders using cuprophan
supports
The test is carried out at 32°C.
The reciprocating frequency is 30cycles/min. The reciprocating frequency is 30cycles/min.
Advantages:-Convenient method for selecting the volume of the
medium. Modifications can been made to accommodate
300 mL vessels
-sink conditions can be maintained.
Disadvantages: -Investment is high because the design is totally
different from standard equipment already available in industry.
37.
38. UNOFFICIAL METHODS
1.ROTATING/STATIC DISK METHOD
Developed by late Eino nelson and described by Levy and Sahli.
In this method ,the drug is compressed in a non-disintegrating
disc without excipients.
The disc is mounted in a holder so that only one face of the disc is
exposed to the dissolution medium.exposed to the dissolution medium.
The holder and disc are immersed in medium and held in a
fixed position as in static disc method and rotated at a given
speed in rotating disc method.
Samples are collected at predetermined times.
Surface area of the drug through which dissolution
occurs is kept constant –intrinsic dissolution rate.
39. 2.BEAKER METHOD:
Reported by Levy and Hayes(1960).
Dissolution medium, 250ml of 0.1N HCl at 37°C placed
in a 400ml beaker.
Agitation by three blade polyethylene stirrer, 5cm diameter and
rotates at 60 rpm.
Stirrer immersed to a depth of 2.7 cm in medium and in the center.
Tablets are placed in a beaker and test was carried out.Tablets are placed in a beaker and test was carried out.
Samples are removed and assayed for the content.
3.FLASK STIRRER METHOD
Developed by Poole(1969).It includes RBF and a stirring element
similar to that of beaker method.
RBF used to avoid the formation of moulds of particles in different
positions on the flat bottom of a beaker.
40. 4.PERISTALSIS METHOD:
To stimulate hydrodynamic condition of GIT tract in an in-vitro
dissolution device.
It consists of rigid plastic cylindrical tubing fitted with septum and
rubber stopper at both ends.
Dissolution chamber consists of a space between septum and lower
stopper.
Dissolution medium is pumped with peristaltic action through Dissolution medium is pumped with peristaltic action through
the dosage form.
5.ROTATING BOTTLE METHOD:
It consists of rotating rack to hold sample drug products in bottles
and they are capped tightly & rotated in 37°C temperature bath.
Sample are decanted through a 40 mesh screen and residue are
assayed.
41. 6.DIALYSIS METHOD:
Cell consist of 32mm inflated membrane.
Plugged at the lower end by tight fitting cylindrical perspex box.
Upper end of the tube held by thin perspex ring inserted into the
tube and secured by an elastic band.
The cell suspended , from the arm of the tablet disintegration
apparatus and containing the dosage form in 150ml of distilled
water at 37°C.water at 37°C.
The cell is raised or lowered 30times a min, into 150ml of distilled
water at same temperature.
Agitation by slight flexing and stretching of the dialysis membrane
as it enters and leaves the bath. Rotated at 60rpm.
.
42. 7.DIFFUSION CELL
Static or flow through diffusion cells are used to characterize in-
vitro drug release and drug permeation kinetics from a topical drug
product e.g.: Ointment, cream or transdermal drug product.
The Franz diffusion cell is static diffusion system used to
characterize drug permeation through skin model.
The skin is mounted on the Franz diffusion cell and the drug
product is placed on the skin surface.product is placed on the skin surface.
The drug permeates across the skin into a receptor fluid
compartment that may be sampled at various times.
This system is used for selection of appropriate formulation that
has optimum drug delivery.
46. Instrument should be calibrated /Operational Checks ( document
each time of use):
•Basket/shaft examination
•Paddle examination
•Vessel examination
•VesselTemperature
•Vibration
Analytical balance must be calibrated Analytical balance must be calibrated
pH meter should be calibrated
Pipettes and other glassware's
Volumetric accuracy : ± 1%
Deareation of dissolution medium.
Lighting should be sufficient to perform visual observations.
47. Calibration of Dissolution Apparatus
Why ?
• To confirm suitability of the equipment and proper operation of
the apparatus
How ?How ?
• Mechanical calibration (verification of physical parameters)
• Chemical calibration (Apparatus SuitabilityTest – USP)
When ?
• Before using new test equipment
• After relocation or major maintenance
• At regular intervals (Every 6 months)
48. Current Harmonized Physical
Parameters and Tolerances
USP Mechanical Calibration Parameters include:
Basket/ShaftWobble (No significant wobble)
Vessel/Shaft Centering (2 mm from centerline)
Height check/Basket or Paddle Depth as measured at basket
bottom or Paddle bottom (25 + 2 mm)bottom or Paddle bottom (25 + 2 mm)
No significant vibration
Rotational speed (+ 4%)
VesselTemperature (37.0 + 0.5 C)
BasketWobble (bottom rim) (+ 1mm)
49. USP Performance Verification Test
(PVT)
Official since 1978, USP Calibration with Prednisone and Salicylic
Acid has been the means of qualifying the dissolution apparatus.
Initially, the primary purpose was to indicate environmental
effects on the apparatus and vibration since most other parameters
could be controlled by mechanical measurements
The original test was called “Calibration” which was not a true
indication of the test being performed, later changed to
“PerformanceVerification” (PVT)
The PVT, has been responsible for detecting problems associated
with dissolution apparatus that are found to be within mechanical
tolerances
50. Calibrator Tablets
1970’s : USP CalibratorTablets Introduced
Disintegrating – 50 mg Prednisone (Upjohn)
Non Disintegrating – 300 mg Salicylic Acid (Hoffman
LaRoche)
1997 : 50 mg Prednisone replaced with 10 mg Prednisone
manufactured at University of Marylandmanufactured at University of Maryland
2004 : USP begins search for replacement for 10 mg
Prednisone tablet
USP: Both Calibrators on a given apparatus (i.e. 4 calibration
tests if instrument is used for paddle and basket methods)
JP, BP and EP: No calibrator tablets
51. Allowable Variations
A basket with gold coating 2.5 μm thick (0.0001 inch) is
an allowable variation of the standard 40-mesh basket.
Some changes can be made to the compendial apparatus; for
example, a basket mesh size other than the typical 40-mesh
basket (e.g., 10-, 20-, or 80-mesh)basket (e.g., 10-, 20-, or 80-mesh)
Larger vessels accommodating up to two and four liters are
now allowable variations in the USP. Such vessels are
advantageous for poorly soluble drugs.
For example, a small-volume apparatus with mini paddles
and baskets may be considered for low-dosage strength
products.
52. De-areation
Air bubbles can act as a barrier to the dissolution process
if present on the dosage unit or basket mesh and can
adversely affect the reliability of the test results.
Furthermore, bubbles can cause particles to cling to the
apparatus and vessel walls.apparatus and vessel walls.
Bubbles on the dosage unit may increase buoyancy, leading
to an increase in the dissolution rate, or may decrease the
available surface area, leading to a decrease in the dissolution
rate.
Poorly soluble drugs are most sensitive to interference
from air bubbles, therefore, deaeration may be needed.
53. Degassing as per USP
Prepare dissolution media and
properly deaerate.
USP Method:USP Method:
Heat media to 41°C, vacuum
filter through 0.45μm filter,
continue to pull vacuum for 5
additional minutes
54. Common Degassing Methods
Acceptable Methods
USP Vacuum Filtration Method
(default unless another approach is
validated)
Helium Sparging*
Automated Degassing*
Unacceptable Methods
Nitrogen Sparging
Sonication
Automated Degassing*
Superheating*
Not Degassing At All*
*when validated against USP method
59. Development of a Discriminating
Method
The procedure should be capable of distinguishing significant
changes in composition or manufacturing process that might
be expected to affect in vivo performance.
Factors to consider:
Qualitative and quantitative excipient changes Qualitative and quantitative excipient changes
Manufacturing parameters:
Lubrication
Blend time
Compression force
Drying parameters
60. Steps for dissolution method
development
Literature survey : Drug and drug product knowledge
Selection ofApparatus
Selection of dissolution medium
Optimization of dissolution parameters Optimization of dissolution parameters
Development of suitable analytical method for estimation of
content
Validation of analytical method
61. Knowledge of Drug and Drug
Product
Characteristics of the API e.g.,particle size, crystal form,
bulk density, solubility
Product composition e.g., drug loading /dose, and the
identity, type, and levels of excipients
Manufacturing process e.g., compression forces, Manufacturing process e.g., compression forces,
equipment
Effects of stability storage conditions e.g., temperature,
humidity
Incompatibility of the drug with certain buffers or salts
62. Dissolution Tester Choice
Paddles and Baskets tend to be the choice for most solid
oral dosage forms.
If pH changes, greater/smaller volumes, or different
agitation is needed then Apparatus 3 and 4 are often
considered after exhausting Paddle and Basket testing
ForTransdermals Apparatus 5-7 are the primary
choices.
For semisolid dosage forms, the generally used
apparatus include the vertical diffusion cell,apparatus include the vertical diffusion cell,
immersion cell, and flow-through cell apparatus with
the insert for topical dosage forms
A rotating bottle or dialysis tubes may have utility
for microspheres and implants; peak vessels for
eliminating coning; and modified flow-through cells for
special dosage forms including powders and stents.
63. Agitation Rate
Should be sufficient to allow for
media to interact with dosage form
Decreasing or increasing the
apparatus rotation speed may be
justified if to achieve an in-vitro–in-
vivo correlation (IVIVC)
For better discrimination
Too much agitation can result in
non-discriminatory profiles
•Baskets – 50-100 RPM
•Paddles – 25-100 RPM
64. USP.APPARATUS DESCRIPTION ROT.SPEED DOSAGE FORM
TYPE 1 Basket apparatus 50-100 rpm IDR,DR,ER
TYPE 2 Paddle apparatus 50-75rpm
25-50rpm
IDR,DR,ER
Suspension
TYPE 3 Reciprocating
cylinder
5-30 dips/min IDR,ER
TYPE 4 Flow through cell 2-50ml/min ER,Poorly soluble
API
TYPE 5 Paddle over disk 25-50 rpm TRANSDERMAL
TYPE 6 Rotating cylinder N/A TRANSDERMAL
TYPE 7 Reciprocating
holder
30 rpm ER
65. The Ideal Dissolution Media
•Meets sink conditions
•Simple preparation
•Drug is Stable in media 24
hrs+
•Uses as little extras as
possiblepossible
–Surfactants
–Alcohol
•Biologically relevant for
site of dissolution in vivo
–IR typically in acid
–DR typically in acid, then
neutral
–MR typically in neutral
solution
66. Determination of Intrinsic Dissolution
Intrinsic dissolution is defined as
the dissolution of a pure drug
substance from a specified constant
surface area.
A special punch and die is used to
compress pure drug substances into
a disk or tablet.a disk or tablet.
The disk is placed into a special
holder that allows only one flat
surface to come in contact with a
dissolution media at any time
during the test.
67. Application of ID
Intrinsic Dissolution
Intrinsic dissolution data is generally used in drug
screening but can provide helpful solubility information
for method development.
Rates lower than 0.1 mg/min/cm2 generally mean that Rates lower than 0.1 mg/min/cm2 generally mean that
bioavailability will be determined by the dissolution
rate.
Rates higher than 1 mg/min/cm2 mean that
bioavailability will most likely be determined by the
drug permeability.
68. pH Dependent Solubility
Solubility screen in multiple media (pH 1.2-7.5) should be
done to determine optimal solubility.
If needed, use as little surfactant as necessary.
Evaluate multiple surfactants (pay attention to grades and
vendors)
Investigations of the stability of the drug substance should Investigations of the stability of the drug substance should
be carried out, when needed, in the selected dissolution
medium with excipients present, at 37°.
Effect of pH on solubility and stability need to be evaluated.
69. Sink condition and Solubility
Sink condition :Volume of medium at least three times that required in
order to form a saturated solution of drug.
Solubility of the drug substance is usually evaluated by determining the
saturation concentration of the drug in different media at 37° using the shake-
flask solubility method (equilibrium solubility).Alternative methods for
solubility determination may also be used.
In the absence of sink conditions, investigate methods to enhance
solubility, e.g. use of a surfactant. If a surfactant is used, its concentration
should be properly justified (e.g. typically <2% Sodium Lauryl Sulfate (SLS)).
In certain cases, it may be necessary to evaluate the solubility of the drug at
room temperature.
The pH of the clear supernatant should be checked to determine
whether the pH changes during the solubility test.
70. Rules of Thumb for Media Limits
Surfactants below 1% tend to be accepted with appropriate checks that lower
limits aren’t acceptable
>1% require greater scrutiny, other surfactants usually
>1.5% tends to be very difficult to handle with automation
Alcohol is generally a last resort – unless doing a dose dumping study
specifically.specifically.
Stay within pH 1.1 – pH 7.5 if at all possible
The use of surfactants needs to be justified by data that show low
solubility in the aqueous media. The chosen concentration of surfactant
also needs to be justified by providing dissolution profiles in media
containing the surfactant at concentrations higher and lower than the chosen
concentration.
72. Use of enzymes
The use of enzymes in the dissolution medium is permitted,
in accordance with general chapter Dissolution 711 ,when
dissolution failures occur as a result of cross-linking
with gelatin capsules or gelatin-coated products.
A discussion of the phenomenon of cross-linking and method
development using enzymes can be found in proposed
general information chapter Capsules–Dissolution Testing and
Related Quality Attributes 1094 .
73. Antifoaming Agents and
antioxidants
The hydrodynamics are influenced by the cylinder's
reciprocating motion and the resulting movement of the
sample in the medium. The reciprocating motion of the
cylinder and screen may cause foaming if the medium
contains surfactants. Addition of an anti-foaming agent suchcontains surfactants. Addition of an anti-foaming agent such
as simethicone or n-octanol may be useful for avoiding
foaming from surfactants.
In some cases, antioxidants such as ascorbic acid may be
used in the dissolution medium to stabilize the drug.
74. Media Cautions
Be careful with water
–No buffering capacity
–Quality can differ b/w sites
–Quality can differ b/w DI systems, filters, etc.
Check pH before and after run to ensure buffering capacity is Check pH before and after run to ensure buffering capacity is
acceptable
Beware of methods needing tight pH limits
Do not use SLS with Potassium Phosphate Buffers –
Sodium Phosphate only
75. Bio-relevant Dissolution Medium
Bio-relevant media is the media that represent the
conditions same as that of the in-vivo condition.
The fed and fasted state may have significant effects on the
absorption or solubility of a compound.
Composition of media that simulate the fed and fasted Composition of media that simulate the fed and fasted
condition is necessary to establish in-vivo in-vitro
correlations.
This media reflect changes in the pH, bile concentration and
osmolarity after meal intake and therefore have a different
composition than that of typical compendial media.
76. Why the Need for Better
Biorelevant Information
Better IVIVC
Fewer ClinicalTrials
Better Predictive tools
Shorter Development
CyclesCycles
77. Choice of bio-relevant media is usually
based upon:
A mechanistic approach that considers the absorption site, if
known.
Whether the rate-limiting step to absorption is the dissolution
or permeability of the compound.
Fed and fasted states may have significant effects on theFed and fasted states may have significant effects on the
absorption or solubility of a compound.
These media are primarily used to establish in vitro-in vivo
correlation during formulation development and to assess potential
food effects; they are not always intended for Quality Control
purposes.
78. Selection of Dissolution medium
Primary requirement for selection of dissolution media is
that, it should be able to reflect in vivo situations when it is
used to establish an IVIVC.
For Class I and III drugs, use of simple aqueous media such as
SGF without enzymes or SIF without enzymes isSGF without enzymes or SIF without enzymes is
recommended.
For Class II and III drugs, use of biorelevant media for
dissolution testing is recommended. They are: 1) SGF plus
surfactant 2) Milk with 3.5 % fat to stimulate fed state
condition 3)FaSSIF is used for poorly soluble drugs.
79. Temperature of Dissolution Medium
The standard temperature- 37±0.5 °C for oral dosage form .
Slightly increased temperatures such as 38±0.5 °C have been
recommended for dosages forms such as suppositories.
Lower temperatures such as 32±0.5 °C are utilized for Lower temperatures such as 32±0.5 °C are utilized for
topical dosage forms such as trans-dermal patches and topical
ointments.
Media temperature readings must be taken at least
twice during the dissolution test, at the start and
end of a test.
80. Appropriate Test Duration
Time points should be selected to adequately characterize the
ascending and plateau phases of the dissolution curve.
Typically:
Immediate release, 85% in <15 minutes, one time point
Immediate release, 15, 20, 30, 45, 60 minutes
For some products, including suspensions, useful information mayFor some products, including suspensions, useful information may
be obtained from earlier points, e.g., 5–10 min.
Delayed release,Acid 1-2 hours, Buffer +30 minutes
Extended release, minimum of three points
– Initial (1-2 hours) to show potential dose dumping
– Intermediate point to define similar in vivo profile
– Final point to show that essentially complete release (>80%) of the
drug is achieved.
81. When to add dosage form?
Tablet introduction may be
performed manually or
automatically
Dosage forms may be
introduced simultaneously orintroduced simultaneously or
sequentially but they must be
introduced into non-
rotating media.
Evaporation cover shown with
Dosage Delivery Module
(DDM)
82. How to handle sample?
Always handle dosage units
with gloves (not cotton),
forceps or tweezers which
will not scratch or damage
the surface of the dosagethe surface of the dosage
unit.
Examine the six dosage
units.
Do not use chipped,
cracked or capped tablets
83. Weighing of the sample?
Option: Record the dosage
unit weights?
Weight is for
information and
investigation purposes
only.
Dosage units are to be
chosen at random and may
not be selected or
discarded based on weight.
84. Sinkers (USP Apparatus 2)
In addition to sinking, floating dosage forms, sinkers may
assist in keeping a dosage form from sticking to the vessel
inappropriately as in the case with some film coated tablets.
Sinkers Specifications must be adequately described in
the method to eliminate hydrodynamic variation associated
with different sinker devices.
Sinkers may be fabricated by wrapping around a cork
bore.
Should not be too tight – will restrict any disintegrating
release mechanism
Too loosely, the dosage form may escape soon after the test
begins.
88. Examples of typical observations
Uneven distribution of particles throughout the
vessel
1. Particles cling to the sides of the vessel,
2. coning or mounding directly under the apparatus, e.g.,
below the basket or paddle,below the basket or paddle,
3. Particles float at the surface of the medium,
4. Film-coated tablets stick to the vessel
89. Examples of typical observations
Air bubbles on the inside of the vessel or on the
apparatus or dosage unit
1. Air bubbles on the inside of the vessel or on the
apparatus or dosage unit.
2. Sheen on the apparatus is also a sign of air bubbles.2. Sheen on the apparatus is also a sign of air bubbles.
This observation would typically be made when
assessing the need to deaerate the medium
90. Examples of typical observations
Dancing or spinning of the dosage unit, or the dosage
unit being hit by the paddle.
Adhesion of particles to the paddle or the inside of the
basket, which may be observed upon removal of the
stirring device at the end of the run.stirring device at the end of the run.
Pellicles or analogous formations, such as transparent
sacs or rubbery, swollen masses surrounding the capsule
contents.
91. Examples of typical observations
Observation of the disintegration rate (e.g.,
percentage reduction in size of the dosage unit within a
certain time frame).
Complex disintegration of the coating of
modified or enteric-coated products.modified or enteric-coated products.
Whether the dosage form lands in the vessel center
or off-center, and if off-center, whether it sticks there.
Time required for the complete dissolution of the
capsule shell or for tablet disintegration.
92. Usual Remedies
Visual observations
Changing any of the following
factors:
1) Apparatus type,
2) Speed of agitation,
3) Level of deaeration,3) Level of deaeration,
4) Sinker type, or
5) composition of the medium.
For dosage forms that exhibit
coning (mounding) under the
paddle at 50 rpm, the coning can be
reduced by increasing the paddle
speed to 75 rpm
93. How To Sample Properly?
Filtration must occur at USP location
and at appropriate time
±2 % from time point or 15
minutes (whichever less)
Halfway between top of paddle Halfway between top of paddle
or basket and media
No closer than 1cm to vessel
wall
Recommend not sampling close
to shaft due to poor hydrodynamics
94. Acceptable Method Requirements
Low variability (<20% at initial time point,
<10% at later points)
Complete Release (85%+ or Asymptote)
Proper understanding of dissolution release
Discrimination between batches -Challenged
with other formulations
Robust/Rugged/Reproducible results
96. Specification Parameters -
Dissolution
Dissolution is considered product-specific.
The method and limits should be appropriate for the proposed
product.
It is useful to have the parameters (medium, apparatus, speed) in
specs.specs.
Dissolution specs at release and shelf-life should be
identical.
Surfactant use should be exceptional and appropriate.—not
exceed 2% normally
97. Dissolution Acceptance Criteria
Q –Value :
Define as a percentage of drug content
dissolved in a given time period.
It is commonly used in the USP for immediate release and It is commonly used in the USP for immediate release and
delayed release dosage forms.
The quantity of Q is the amount of dissolved active
ingredient specified in the individual monograph
expressed as a percentage of the labeled content.
98. Q - Value
For highly soluble and rapidly dissolving drug products (BCS
classes 1 and 3), a single-point dissolution test specification of NLT 85%
(Q=80%) in 60 minutes or less is sufficient as a routine quality control test for
batch-to-batch uniformity.
For slowly dissolving or poorly water soluble drugs (BCS class 2), a
two-point dissolution specification, one at 15 minutes to include a
dissolution range (a dissolution window) and the other at a later point (30, 45,dissolution range (a dissolution window) and the other at a later point (30, 45,
or 60 minutes) to ensure 85% dissolution, is recommended to characterize the
quality of the product.
For products containing water insoluble APIs, it is recommended to
have a two tire dissolution limit. For example Artemether dissolution:
NLT 40% in 1 hour and NLT 60% at the 3rd hour.
99. Immediate Release Forms – Acceptance
Table USP <711>
STAGE No. of Dosage
units tested
Acceptance criteria
S1 6 Each unit is ≥ Q + 5%.
S2 6 Average of 12 units (S1 + S2) is ≥Q and no
unit is < Q – 15%.unit is < Q – 15%.
S3 12(6+6+12=24) Average of 24 units (S1 + S2 + S3) is ≥ Q,
not more than 2 units are < Q – 15% and
no single unit is less than Q – 25%.
9
9
If a sample fails either Stage S1 or S2, proceed to the next stage and test
the number of units indicated.
100. Delayed Release Forms – Method A
USP <711>
USP Delayed Release Dosage Forms – MethodA:
Media is 750 mL of 0.1N HCl
Samples are removed for analysis after 2 hours ± 2%
Within 5 minutes of withdrawing the acid stage sample
aliquots, add 250 mL of 0.20-M tribasic sodium phosphate,
•
aliquots, add 250 mL of 0.20-M tribasic sodium phosphate,
adjusting to pH + 0.05 at 37 °C while stirring at the
specified rate.
Dissolution continues for 0.75 h (or per monograph) or less.
Sample aliquots are then analyzed with Q being the
total % dissolved for both acid and buffer stages.
101. Delayed Release Forms – Acid Stage
Acceptance Table USP <711>
STAGE No. of Dosage
units tested
Acceptance criteria
A1 6 No individual value exceeds 10%
dissolved.
A2 6 Average of 12 units (A1 + A2) is no more
than 10% dissolved and no single unit isthan 10% dissolved and no single unit is
greater than 25% dissolved.
A3 12(6+6+12=24) Average of 24 units (A1 +A2 +A3) is not
more than 10% dissolved and no single
unit is more than 25% dissolved.
1
0
1
If a sample fails either Stage A1 or A2, proceed to the next stage and test
the number of units indicated.
102. Delayed Release Forms - Buffer Stage
Acceptance Table USP <711>
STAGE No. of Dosage
units tested
Acceptance criteria
B1 6 Each unit is ≥ Q + 5%.
B2 6 Average of 12 units (B1 + B2) is ≥ Q and no unit is < Q
– 15%.– 15%.
B3 12(6+6+12=24) Average of 24 units (B1 + B2 + B3) is ≥ Q, not more
than 2 units are < Q – 15% and no single unit is less than
Q – 25%.
1
0
2
If a sample fails either Stage B1 or B2, proceed to the next stage and test
the number of units indicated.
103. Delayed Release Forms – Method B
USP <711>
USP Delayed Release Dosage Forms – Method B:
Media is 1000 mL of 0.1N HCl
Samples are removed for analysis after 2 hours ―
USP Delayed Release Dosage Forms – Method B:
Drain the acid from the original vessel and replace with
•
Drain the acid from the original vessel and replace with
1000mL pH6.8 phosphate buffer pre equilibrated at 37°C
or switch the dosage form to a second vessel containing the
phosphate buffer
104. Disssolution specification – Case
study -1
Ethambutol hydrochloride 400mg tablets.The applicant
claimed USP standard for the product.
Bioequivalent of the product is accepted as per BCS class 3
based biowavier
The applicant set the dissolution specification limits as below: The applicant set the dissolution specification limits as below:
NLT 80% (Q) in 45min at release
NLT 75% (Q) in 45min at shelf life, which is in line with the
requirement of USP monograph.
Is it acceptable?What limits should be applied?
105. Disssolution specification – Case
study -1
Answer: not acceptable
The dissolution limits at release and shelf life should be the
same.
A limit of NLT 80% (Q) in 15min should be set for A limit of NLT 80% (Q) in 15min should be set for
both release and shelf life as for the BCS class 3
biowaiver.
107. Why Profiling / Comparision?
It reflects its release pattern under the selected condition sets. i.e. either
sustained release or immediate release of the formulated formulas.
For optimizing the dosage formula by comparing the dissolution profiles
of various formulas of the same drug.of various formulas of the same drug.
Dissolution profile comparison between pre change and post change products
for SUPAC (scale up post approval change ) related changes or with different
strengths, helps to assure the similarity in the product performance and green
signals to bioequivalence.
Modify formulation as per Brand Leader.
108. Requirement for Comparative dissolution testing
Two or more products or batches containing the same API
are compared
The strength of products / batches may or may not be the
same (depending on purpose of test)
The dissolution conditions are similar, e.g.
Apparatus, medium, volume, rotation speed & temp.• Apparatus, medium, volume, rotation speed & temp.
• Minimize possible experimental differences in
conditions
Samples are taken at the same time points and the data
(dissolution profiles) compared
109. How to Compare?
Aim: To show similar in vitro dissolution under physiologically
relevant experimental pH conditions.
Advisable to investigate more than one batch of test and reference
products; must include bioequivalence batches.
Investigate within pH 1-6.8 (normally pH 1.2, 4.5 and 6.8) and QC media
(if different).
Water may be used as an additional medium, especially when the APIWater may be used as an additional medium, especially when the API
is unstable in buffered media to the extent that data is unusable.
Additional investigations may be required at pH values in which the
drug has minimum solubility.
Use12 units to enable statistical evaluation
For each condition, present comparative dissolution profiles(mean values
vs.time) together with statistics (max, min, mean, RSD; f2 similarity factor if
calculated; individual values)
110. METHODSTO COMPARE DISSOLUTION PROFILEMETHODSTO COMPARE DISSOLUTION PROFILE
Graphical methodGraphical method Statistical
Analysis
Statistical
Analysis
Model Dependent
method
Model Dependent
method
Model Independent
Method
Model Independent
Method
t-Testt-Test ANOVAANOVA
ZeroZero FirstFirst Hixson-Hixson- HiguchiHiguchi WeibullWeibull KorsemeyarKorsemeyar Baker-Baker-Zero
order
Zero
order
First
order
First
order
Hixson-
crowell
law
Hixson-
crowell
law
Higuchi
model
Higuchi
model
Weibull
model
Weibull
model
Korsemeyar
and peppas
model
Korsemeyar
and peppas
model
Baker-
Lonsdale
model
Baker-
Lonsdale
model
RatioTest
Procedure
RatioTest
Procedure
PairWise
Procedure
PairWise
Procedure
Multivariate
Confidence Region
Procedure
Multivariate
Confidence Region
Procedure
Index of RescignoIndex of Rescigno
111. Paired Wise Procedure
DIFFERENCE FACTOR (f1) & SIMILARITY FACTOR (f2)
The difference factor (f1) as defined by FDA calculates the % difference
between 2 curves at each time point and is a measurement of the relative
error between 2 curves.
n
TtRt
f1 = × 100
where, n = number of time points
Rt = % dissolved at time t of reference product (pre change)
Tt = % dissolved at time t of test product (post change)
n
t
t
Rt
TtRt
1
1
112. The similarity factor (f2) as defined by FDA is logarithmic reciprocal
square root transformation of sum of squared error and is a measurement
of the similarity in the percentage (%) dissolution between the two curves
f2 = 50 ×
100
1
log )(
1
1
5.0
n
r
TtRtwt
n 1rn
113. Guidance for Industry
A specific procedure to determine difference and similarity factors is as
follows:
1. Determine the dissolution profile of two products (12 units each) of the test
(postchange) and reference (prechange) products.
2. Using the mean dissolution values from both curves at each time interval,
calculate the difference factor (f1 ) and similarity factor (f2) using the above
equations.equations.
3. For curves to be considered similar, f1 values should be close to 0, and f2 values
should be close to 100. Generally, f1 values up to 15 (0-15) and f2 values greater
than 50 (50-100) ensure equivalence of the two curves and thus, of the
performance of the test (postchange) and reference (prechange) products.
4.This model independent method is most suitable for dissolution profile
comparison when three to four or more dissolution time points are
available.
114. The following recommendations should also be considered:
The dissolution measurements of the test and reference batches should be
made under exactly the same conditions.
The dissolution time points for both the profiles should be the same (e.g., 15,
30, 45, 60 minutes).
The reference batch used should be the most recently manufactured
prechange product.
A minimum of three points required for comparison of profile.
Only one measurement should be considered after 85% dissolution of both the Only one measurement should be considered after 85% dissolution of both the
products.
To allow use of mean data, the percent coefficient of variation at the earlier
time points (e.g., 10 minutes) should not be more than 20%, and at other time
points should not be more than 10%.
The mean dissolution values for R can be derived either from
(1) last prechange (reference) batch or
(2) last two or more consecutively manufactured prechange
batches.
115. Example
Determination of similarity of profiles
Example 1-B
% API dissolved
Time
(min)
Tablet D
(Ref)
Tablet E
(Test)
10 55 57
15 72 78
Example 1-A
% API dissolved
Time
(min)
Tablet A
(Ref)
Tablet B
(Test)
10 87 94
15 96 99 15 72 78
20 85 91
30 97 100
45 102 100
60 103 101
f2 required? Yes
f2 (n = 3 ?) 64 (similar)
15 96 99
20 99 99
30 100 99
45 101 99
60 101 99
f2 required? No, ≥ 85% in 15 min
f2 (n = N/A ?) profiles similar
116. Example
Determination of similarity of profiles (cont.)
Example 1-D
% API dissolved
Time
(min)
Tablet A
(Ref)
Tablet Y
(Test)
10 87 55
15 96 72
Example 1-C
% API dissolved
Time
(min)
Tablet X
(Ref)
Tablet Y
(Test)
10 29 34
15 38 41 15 96 72
20 99 85
30 100 97
45 101 102
60 101 103
f2 required? Yes
f2 (n = 3 ?) 31 (not similar)
15 38 41
20 47 50
30 63 64
45 80 79
60 95 91
f2 required? Yes
f2 (n = 6 ?) 74 (similar)
117. Comparison to reference medicinal
product
Immediate ReleaseTablets/Capsules
> 85% dissolved within15 minutes : test and
reference similar without any further
calculationcalculation
≤ 85 % dissolved within15minutes: calculate f2
similarity factor
Modified Release Preparations
Calculate f2 similarity factor
118. Comparison to reference medicinal
product
Prolonged Release Preparations
Minimum of 3 time points, but may be prudent to do more
Particularly important if desired release profile not uniform Particularly important if desired release profile not uniform
(e . g. immediate release outer coat and prolonged release core)
If only3 time points: expected to mirror final specification time
points (20-30%(dose dumping),50%(defines profile),>80%)
More time points =↑confidence of bioequivalence
e.g. for once daily preparations(1,2,4,8,12,16,20 & 24 hours)
119. Alternative methods to dissolution testing
In ICH Q6A permits use of disintegration testing as a surrogate
for conventional Compendial dissolution tests, provided
highly soluble drug substances
intrinsic rate of solubilization is rapid intrinsic rate of solubilization is rapid
overall drug release rate is dominated by cohesive properties
of the formulation
120. DT in place of dissolution?
ICH Q6A decision trees #7 can be used to assess the proposed dissolution criteria,
however:
Highly soluble throughout physiological pH range.
Solubility at 37°C ± 0.5°C, dose + solubility < 250 ml, pH 1.2 - 6.8.
For considering /accepting DT in place of dissolution: all the considerationsFor considering /accepting DT in place of dissolution: all the considerations
should be carefully assessed: highly soluble and very rapidly dissolving, plus
significant supporting development data – including
when DT is more discriminating or
has a demonstrated relationship to dissolution, robustness of the
formulation/manufacturing process have been demonstrated wrt DT, etc.
Dissolution may not be necessary or proposed as a
skip test.
121. Example
APIs with good solubility at gastric pH levels may be
granted BCS Class I and III classification i.e. may be
characterized by disintegration testing alone.
In liquid filled capsules, drug dissolved in solubilisation aids
offering a true mechanism for drug release is likely to be the
rupture of the capsule use disintegration as a surrogate for
the QC dissolution test
122. Bio vaiver (In vitro equivalence testing)
The term biowaiver is applied to a regulatory drug approval
process where the efficacy and safety part of the dossier
(application) is approved based on evidence of equivalence other
than through in vivo equivalence testing.
A biowaiver can be applied only for products which meet
requirements on pharmaceutical similarity, as well as similarity inrequirements on pharmaceutical similarity, as well as similarity in
comparative dissolution tests.
A BCS-based biowaiver has become an important and
cost-saving tool in approval of generic drugs.
The bio-relevance of the BCS properties and the in vitro release
are best expressed through a correlation between in vitro and in
vivo data.
123. BCS Classification
Class 1 Class 2 Class 3 Class 4
Highly Poorly Highly PoorlyHighly
Soluble
Poorly
Soluble
Highly
Soluble
Poorly
Soluble
Highly
Permeable
Highly
Permeable
Poorly
Permeable
Poorly
Permeable
124. ‘High solubility
The definition of ‘high solubility’ refers to the highest dose
strength of an immediate release product, which has to be
soluble in 250 ml or less of aqueous media over the pH
range of 1 – 7.5,. Solubility measurements should berange of 1 – 7.5,. Solubility measurements should be
performed at 37 °C using a stability indicating, validated
method.
125. High permeability
The classification regarding high permeability refers to the
extent of absorption in humans, i.e. high permeability is
concluded if the extent of absorption in humans
reaches at least 90 % of an orally administered dose.reaches at least 90 % of an orally administered dose.
This conclusion may be based on either pharmacokinetic
studies in humans (e.g. mass balance, or absolute
bioavailability studies) or intestinal permeability methods like
e.g. in vivo intestinal perfusion studies in humans or validated in
vitro permeation studies across a monolayer of cultured epithelial
cells.
126. Requirement
Comparative in vitro dissolution investigations should ensure that no less
than 85 % of the labeled amount is dissolved within 30 min in each
of the required media: 0.1 N HCl, pH 4.5 and 6.8 buffers. Regarding
experimental requirements, reference is made to the US Pharmacopoeia and
the US-FDA guidance for industry on DissolutionTesting of Immediate Release Solid
Oral Dosage Forms (August 1997) 3.Oral Dosage Forms (August 1997) 3.
Resulting profiles should be compared using the similarity factor (f2),
unless 85% or more of the labeled amount dissolves within 15 min
from both products.
The latter case would allow the conclusion that the investigated
products are similar without requiring any further statistical
calculations.
127. Restrictions
BCS based waivers are not applicable for the initial in vivo
bioavailability characterization for NDAs.
Other restrictions of application include:
1) narrow therapeutic index (NTI) drug; and1) narrow therapeutic index (NTI) drug; and
2) drug products intended to be absorbed in the oral cavity.
3) Similarly, prodrugs and excipients require special consideration.
In case of prodrugs, whether to measure the prodrug or the drug
for permeability determination will depend on where the
conversion occurs.
4
128. Comparison of FDA, EU and WHO
guidance on BCS based biowaiver.
Parameters US-FDA EU WHO
Allowed classes 1 1 and 3 1, 2 (weak acids), and
3
High solubility
Highest strength completely dissolved in 250mL of aqueous media at 37 ˚C ± 1 ˚C.Highest strength completely dissolved in 250mL of aqueous media at 37 ˚C ± 1 ˚C.
pH range pH 1-7.5 pH 1-6.8 pH 1.2-6.8
High permeability >90% absolute BA or
mass balance study
>85% absolute BA or mass balance study
129. Comparison of FDA, EU and WHO
guidance on BCS based biowaiver.
Parameters FDA EU WHO
Rapid dissolution
Media (studies
should be
conducted at 37 ±
900 mL or less
aqueous
media (0.1N HCl
900 mL or less
aqueous
media (pH 1.0-1.2
900 mL or less
aqueous
media (pH 1.2conducted at 37 ±
1 C)
media (0.1N HCl
or
SGF; pH 4.5
buffer; and
pH 6.8 buffer or
SIF)
media (pH 1.0-1.2
buffer, usually
0.1N
HCl or SGF; pH
4.5
buffer; and pH 6.8
buffer or SIF)
media (pH 1.2
HCl solution; pH
4.5
acetate buffer;
and pH
6.8 phosphate
buffer)
Apparatus (APP) USP APP I - 100
rpm
USP APP II - 50
rpm
Paddle APP - 50
rpm
Basket APP - 100
rpm
Paddle APP - 75
rpm
Basket APP - 100
rpm
130. Comparison of FDA, EU and WHO
guidance on BCS based biowaiver.
Parameters FDA EU WHO
Rapid dissolution
Criteria >85% in 30 min
in 3 media
Class 1: >85% in
30 min in 3 media
Class 1: >85% in 30 min
in 3 media (Rapid)in 3 media 30 min in 3 media
(Rapid)
Class 3: >85% in 15
min in 3 media (Very
Rapid); or, >85% in
30 min and similar
dissolution profile to
RLD (Similarly Rapid)
in 3 media (Rapid)
Class 2: >85% in 30 min
in pH 6.8 medium and
similar dissolution
profile in 3 media
Class 3: >85% in 15 min
in 3 media (Very Rapid)
Restrictions Narrow therapeutic drugs
Oral products intended to be absorbed in the oral cavity
Modified release drug products
131. CONCLUSION:
In vitro dissolution is the best available tool today which can at
least quantitatively assure about the biological availability of drug.
Great Scope for New Method Development / Pharmacopoeial
monographs
Systematic and scientific Approach is needed. Systematic and scientific Approach is needed.
Academic standards / Research ??????
Basic infrastructure / facilities / sophisticated instrument /
positive efforts / attitude and honesty is the need of time to
improve the quality of research.
132. Guidelines Referred
BCS Guidance (Waiver of InVivo Bioavailability and
Bioequivalence Studies for Immediate-Release Solid Oral Dosage
Forms Based on a Biopharmaceutics Classification System”);
August 2000
IR Dissolution Guidance (DissolutionTesting of Immediate
Release Solid Oral Dosage Forms);August 1997
IVIVC Guidance (Extended Release Oral Dosage Forms: IVIVC Guidance (Extended Release Oral Dosage Forms:
Development, Evaluation, andApplication of InVitro/InVivo
Correlations); September 1997
General BA/BE Guidance (Bioavailability and Bioequivalence
Studies for OrallyAdministered Drug Products - General
Considerations); 2003
WHO, US FDA and EMEA guidelines
ICH guidelines