WHAT IS pH?
pH is a measure of the acidity or alkalinity of a water solution. The
acidity or alkalinity of a water solution is determined by the relative
number of hydrogen ions (H+) or hydroxyl ions (OH-) present. Acidic
solutions have a higher relative number of hydrogen ions, while alkaline
(also called basic) solutions have a higher relative number of hydroxyl
ions.
Acids are substances which either dissociate (split apart) to release
hydrogen ions or react with water to form hydrogen ions.
Bases are substances that dissociate to release hydroxyl ions or
react with water to form hydroxyl ions.
In practice, pH is often assumed to be the negative logarithm of
the hydrogen ion concentration:
pH = - log10 [H+]
At 25°C, a neutral solution has a pH of 7.0, while solutions with pH < 7
are acidic and solutions with pH > 7 are alkaline. The normal overall pH
range is 0 to 14 pH, although solutions containing non-water solvents
can have pH values outside this range.
The cell consists of a measuring electrode, a reference electrode, a
temperature sensing element, and the liquid being measured. The
voltage of the cell is directly proportional to the pH of the liquid. The pH
meter measures the voltage and uses a temperature-dependent factor to
convert the voltage to pH. Because the cell has high internal resistance,
the pH meter must have a very high input impedance.
THE MEASUREMENT OF pH
The pH electrode uses a specially formulated, pH sensitive glass in
contact with the solution, which develops a potential (voltage)
proportional to the pH of the solution.
A glass pH electrode consists of an inert glass tube with a pH sensitive
glass tip, either hemispherical (bulb) or flat in shape, blown onto it. The
tip contains a fill solution with a known pH,
The reference electrode is designed to maintain Constant potential at any
given temperature Serves to complete the pH measuring circuit within the
solution It provides a known reference potential for the pH electrode.
The difference in the potentials of the pH and reference electrodes provides
a millivolt signal proportional to pH..
Most pH sensors are designed to produce a 0 mV signal at 7.0 pH, with a
(theoretically ideal) slope (sensitivity) of -59.16 mV / pH at 25°C
The millivolt signals produced by the pH and reference electrodes are
temperature dependent.
However, the pH and reference electrode combination exhibits an
isopotential point, which is a pH and millivolt potential at which the potential
is constant with temperature changes. The isopotential point is most
frequently designed to be at 7.0 pH and 0 mV. Using the isopotential point
with a theoretical knowledge of electrode behavior makes it possible to
compensate (correct) the pH measurement at any temperature to a
reference temperature (usually 25°C),
Compensation (correct) the pH measurement
Compensation (correct) the pH measurement means to remove the
temperature dependence in the intercept and to shift the origin of the axes to pH 7.
The result is plotted in the Figure. Two lines appear on the graph. One line shows
how cell voltage changes with pH at 25°C, and the other line shows the relationship
at 50°C.
The lines, which are commonly called isotherms, intersect at the point (pH
7, 0 mV). An entire family of curves, each having a slope determined by the
temperature and all passing through the point (pH 7, 0 mV) can be drawn on the
graph.
Shows why temperature is important in making pH measurements. When
temperature changes, the slope of the isotherm changes. Therefore, a given cell
voltage corresponds to a different pH value, depending on the temperature.
Reference Poisoning
The mechanism of reference poisoning is a conversion of the
reference from a silver-silver chloride based electrode to an
electrode based on a different silver compound.
These ions include bromide, iodide, and sulfide ions. When
these ions enter the fill solution, they form insoluble precipitates
with the silver ions in the fill solution.
But there is no initial effect on the potential of the reference,
because the silver ions lost to precipitation are replenished by
silver ions dissolving off the silver chloride coating of the silver
wire. It is not until the silver chloride coating is completely lost
that a large change in the potential so the reference electrode
must be replaced.
To counter this effect, multiple junction reference electrodes are
used, which consist of two or more liquid junctions and fill
solutions to slow the progress of the poisoning ions. Gelling of
the reference fill solution is also used to prevent the transport of
poisoning ions by convection.
Liquid Junction Potential
Potassium chloride is chosen for the fill solution not only for its ability to solubilize
silver ion, but also because it is equitransferent. Equitransferent means that the
positive potassium ion and the negative chloride ion diffuse through a water
solution at nearly the same rate.
When diffusing through the liquid junction of a reference electrode, both the
positive and negative ions tend to move together and maintain a net zero charge
at all points within the liquid junction.
However, not all solutions are equitransferent, and when the process solution is
not, a liquid junction potential can result. When a positive ion diffuses through the
liquid junction faster than a negative ion, or vice versa, a charge imbalance will
result. This gives rise to an opposing potential, which is liquid junction potential.
This potential gets added to the potentials of the pH and reference electrodes
and causes an offset to the pH measurement of typically a few tenths of a pH.
pH IN AQUEOUS SOLUTIONS
Practical pH Measurement Range
Although the range of pH measurements is defined to be 0 to 14 pH, solutions
with pH values near the extremes of this range are often better measured
using conductivity.
Tx: Ph202 YOKOGAWA
Sensor: PH8EHP
Place :DEMI WATER
The Model PH8EHP pH Sensor is used to configure a pure water pH
transmission
system.
•Measurement principle: Glass electrode method
•Sample temperature range : 0 to 508C
•KCl solution consumption: Maximum 3 ml/day
•Sample pressure : Under atmospheric pressure
•Sample flow rate : 30 to 600 ml/min
Note : The maximum flow rate is restricted by the measurement
solution conductivity.
Process application:why put at this phase in the process?
XMT-P
CALIBRATING TEMPERATURE
XMT-P
CALIBRATING pH
For pH sensors, two-point buffer calibration
is standard. Both automatic calibration and
manual calibration are available
Auto calibration avoids common pitfalls and reduces errors. Its use is
recommended. In auto calibration the Solu Comp Xmt calculates the actual
pH of the buffer from the nominal value entered by the user and does not
accept calibration data until readings are stable
Manual calibration the user enters buffer values and judges when readings
are stable. The pH reading can also be standardized, that is, forced to
match the reading from a referee instrument. Finally, if the user knows the
electrode slope (at 25°C), he can enter it directly.
XMT-P Auto CALIBRATING pH
XMT-P Auto CALIBRATING pH
XMT-P Auto CALIBRATING pH
XMT-P Manual CALIBRATING pH
XMT-P Manual CALIBRATING pH
XMT-P ENTERING A KNOWN SLOPE VALUE CALIBRATING pH
If the electrode slope is known from other measurements, it can be entered
directly into the transmitter. The slope must be entered as the slope at 25°C. To
calculate the slope at 25°C from the slope at temperature t°C,
use the equation:
XMT-P TROUBLESHOOTING
Broken pH Glass and pH Glass Z High
These messages mean that the pH sensor glass impedance is outside the
programmed limits. To read the impedance go to the main display and press until
Glass Imp appears in the display. The default lower limit is 10 MΩ. The default upper
limit is 1000 MΩ. Low glass impedance means the glass membrane — the sensing
element in a pH sensor — is cracked or broken. High glass impedance means the
membrane is aging and nearing the end of its useful life. High impedance can also
mean the pH sensor is not completely submerged in the process liquid.
1. Check the sensor wiring, including connections in a junction box.
2. Verify that the sensor is completely submerged in the process liquid.
3. Check the sensor response in buffers. If the sensor can be calibrated, it is in
satisfactory condition. To disable the fault message, reprogram the glass impedance
limits
XMT-P TROUBLESHOOTING
PH202G TROUBLESHOOTING
PH202G TROUBLESHOOTING
pH Analzyer

More Related Content

What's hot

What's hot (20)

Coulometry.pptx presentation assignment copy
Coulometry.pptx presentation assignment   copyCoulometry.pptx presentation assignment   copy
Coulometry.pptx presentation assignment copy
 
pH and pH meter
pH and pH meterpH and pH meter
pH and pH meter
 
Full automation
Full automationFull automation
Full automation
 
Polarography
PolarographyPolarography
Polarography
 
pH measurement
pH measurementpH measurement
pH measurement
 
Ion selective electrodes
Ion selective electrodesIon selective electrodes
Ion selective electrodes
 
Laboratory reagent water
Laboratory reagent waterLaboratory reagent water
Laboratory reagent water
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
pH & its measurement
pH & its measurementpH & its measurement
pH & its measurement
 
HPTLC.pptx
HPTLC.pptxHPTLC.pptx
HPTLC.pptx
 
Analytical balance
Analytical balanceAnalytical balance
Analytical balance
 
pH electrodes
pH electrodespH electrodes
pH electrodes
 
pH Meter
pH MeterpH Meter
pH Meter
 
Ion selective electrodes(ise)
Ion selective electrodes(ise)Ion selective electrodes(ise)
Ion selective electrodes(ise)
 
Advances in Ion Selective Electrodes(ISE)
Advances in Ion Selective Electrodes(ISE) Advances in Ion Selective Electrodes(ISE)
Advances in Ion Selective Electrodes(ISE)
 
LC NMR HYPHENATED TECHNIQUES
LC NMR HYPHENATED TECHNIQUESLC NMR HYPHENATED TECHNIQUES
LC NMR HYPHENATED TECHNIQUES
 
Electrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseElectrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, ise
 
Ion specific electrodes
Ion specific electrodesIon specific electrodes
Ion specific electrodes
 
Automation in chemical analysis-need,objectives,instrumentation and process c...
Automation in chemical analysis-need,objectives,instrumentation and process c...Automation in chemical analysis-need,objectives,instrumentation and process c...
Automation in chemical analysis-need,objectives,instrumentation and process c...
 
HPLC Sample Preparation and Instrumentatio
HPLC Sample Preparation and InstrumentatioHPLC Sample Preparation and Instrumentatio
HPLC Sample Preparation and Instrumentatio
 

Viewers also liked

Steam water analysis system
Steam water analysis systemSteam water analysis system
Steam water analysis systempriyank.modi
 
D 548 – 97 r02 ;rdu0 oa _
D 548 – 97 r02  ;rdu0 oa _D 548 – 97 r02  ;rdu0 oa _
D 548 – 97 r02 ;rdu0 oa _hamfrost
 
Pre and post treatment of water
Pre and post treatment of waterPre and post treatment of water
Pre and post treatment of waterPrudhvi raj
 
Von Willebrand Disease: A Review
Von Willebrand Disease: A ReviewVon Willebrand Disease: A Review
Von Willebrand Disease: A ReviewBhargav Kiran
 
Examples of educational boards
Examples of educational boardsExamples of educational boards
Examples of educational boardsFabrizio Cantini
 
Unigenetics_Product Catalogue_2017
Unigenetics_Product Catalogue_2017Unigenetics_Product Catalogue_2017
Unigenetics_Product Catalogue_2017AJAY JAIN
 
Niezwykle ptaki1 birds of paradise humming-birds1
Niezwykle ptaki1 birds of paradise humming-birds1Niezwykle ptaki1 birds of paradise humming-birds1
Niezwykle ptaki1 birds of paradise humming-birds1Renée Bukay
 
Waste water treatment
Waste water treatmentWaste water treatment
Waste water treatmentNathiyadevi K
 
Waste water treatment
Waste water treatment Waste water treatment
Waste water treatment Ghassan Hadi
 
Nationalgeographicbest2013 2014
Nationalgeographicbest2013 2014Nationalgeographicbest2013 2014
Nationalgeographicbest2013 2014Renée Bukay
 
Alkalinity in Water
Alkalinity in WaterAlkalinity in Water
Alkalinity in WaterFirasMfarrej
 
Parameters of waste water analysis
Parameters of waste water analysisParameters of waste water analysis
Parameters of waste water analysisStudent
 
Water treatment , a simple presentation....
Water treatment , a simple presentation....Water treatment , a simple presentation....
Water treatment , a simple presentation....Thoshina Thomas
 

Viewers also liked (20)

Steam water analysis system
Steam water analysis systemSteam water analysis system
Steam water analysis system
 
P h and ph meter
P h and ph meterP h and ph meter
P h and ph meter
 
D 548 – 97 r02 ;rdu0 oa _
D 548 – 97 r02  ;rdu0 oa _D 548 – 97 r02  ;rdu0 oa _
D 548 – 97 r02 ;rdu0 oa _
 
Acidity
AcidityAcidity
Acidity
 
Shaik Saleem
Shaik SaleemShaik Saleem
Shaik Saleem
 
Oil in water analyzer
Oil in water analyzerOil in water analyzer
Oil in water analyzer
 
ECOLOGY
ECOLOGYECOLOGY
ECOLOGY
 
Pre and post treatment of water
Pre and post treatment of waterPre and post treatment of water
Pre and post treatment of water
 
Von Willebrand Disease: A Review
Von Willebrand Disease: A ReviewVon Willebrand Disease: A Review
Von Willebrand Disease: A Review
 
Examples of educational boards
Examples of educational boardsExamples of educational boards
Examples of educational boards
 
Unigenetics_Product Catalogue_2017
Unigenetics_Product Catalogue_2017Unigenetics_Product Catalogue_2017
Unigenetics_Product Catalogue_2017
 
Acidity in Water
Acidity in WaterAcidity in Water
Acidity in Water
 
Niezwykle ptaki1 birds of paradise humming-birds1
Niezwykle ptaki1 birds of paradise humming-birds1Niezwykle ptaki1 birds of paradise humming-birds1
Niezwykle ptaki1 birds of paradise humming-birds1
 
Waste water treatment
Waste water treatmentWaste water treatment
Waste water treatment
 
Waste water treatment
Waste water treatment Waste water treatment
Waste water treatment
 
Nationalgeographicbest2013 2014
Nationalgeographicbest2013 2014Nationalgeographicbest2013 2014
Nationalgeographicbest2013 2014
 
Orp analyzer
Orp analyzerOrp analyzer
Orp analyzer
 
Alkalinity in Water
Alkalinity in WaterAlkalinity in Water
Alkalinity in Water
 
Parameters of waste water analysis
Parameters of waste water analysisParameters of waste water analysis
Parameters of waste water analysis
 
Water treatment , a simple presentation....
Water treatment , a simple presentation....Water treatment , a simple presentation....
Water treatment , a simple presentation....
 

Similar to pH Analzyer

Introduction to-ph
Introduction to-phIntroduction to-ph
Introduction to-phOswaldoRuiz
 
Measurement of p h
Measurement of p hMeasurement of p h
Measurement of p hPraveen Garg
 
Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhimegha bagri
 
Share_Preparation_of_buffers_and_estimation_of_pH.pptx
Share_Preparation_of_buffers_and_estimation_of_pH.pptxShare_Preparation_of_buffers_and_estimation_of_pH.pptx
Share_Preparation_of_buffers_and_estimation_of_pH.pptxRiyaKaushal21ZO042
 
Analytical Measurements: Troubleshooting, Maintenance and the Future
Analytical Measurements:  Troubleshooting, Maintenance and the FutureAnalytical Measurements:  Troubleshooting, Maintenance and the Future
Analytical Measurements: Troubleshooting, Maintenance and the FutureISA Boston Section
 
working and application of pH meter
working and application of pH meterworking and application of pH meter
working and application of pH meterkhushirahulkar
 
PH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwash
PH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwashPH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwash
PH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwashDr. Karrar Alwash
 
pH determination (electrometric method)
pH determination (electrometric method)pH determination (electrometric method)
pH determination (electrometric method)JahanviTankaria
 
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdf
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdfDrilling Fluid Engineering-PH Lab-Drilling Lab#9.pdf
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdfSARWAR SALAM
 
Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.
Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.
Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.Acmas Technologies Pvt. Ltd.
 

Similar to pH Analzyer (20)

Introduction to-ph
Introduction to-phIntroduction to-ph
Introduction to-ph
 
Measurement of p h
Measurement of p hMeasurement of p h
Measurement of p h
 
Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhi
 
pH - MEASUREMENT .ppt
pH - MEASUREMENT .pptpH - MEASUREMENT .ppt
pH - MEASUREMENT .ppt
 
Ph electrodes
Ph electrodesPh electrodes
Ph electrodes
 
pH meter
pH meterpH meter
pH meter
 
Share_Preparation_of_buffers_and_estimation_of_pH.pptx
Share_Preparation_of_buffers_and_estimation_of_pH.pptxShare_Preparation_of_buffers_and_estimation_of_pH.pptx
Share_Preparation_of_buffers_and_estimation_of_pH.pptx
 
pH Measurement
pH MeasurementpH Measurement
pH Measurement
 
Ph meter.pptx
Ph meter.pptxPh meter.pptx
Ph meter.pptx
 
Analytical Measurements: Troubleshooting, Maintenance and the Future
Analytical Measurements:  Troubleshooting, Maintenance and the FutureAnalytical Measurements:  Troubleshooting, Maintenance and the Future
Analytical Measurements: Troubleshooting, Maintenance and the Future
 
Basics of pH
Basics of pHBasics of pH
Basics of pH
 
18 pH Meter.pptx
18 pH Meter.pptx18 pH Meter.pptx
18 pH Meter.pptx
 
Table top microprocessor p h meters
Table top microprocessor p h metersTable top microprocessor p h meters
Table top microprocessor p h meters
 
pH meter.pdf
pH meter.pdfpH meter.pdf
pH meter.pdf
 
pH METER
pH METERpH METER
pH METER
 
working and application of pH meter
working and application of pH meterworking and application of pH meter
working and application of pH meter
 
PH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwash
PH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwashPH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwash
PH meter of buffer solution الدكتور كرار رأفت علوش < Dr. karrar raafat alwash
 
pH determination (electrometric method)
pH determination (electrometric method)pH determination (electrometric method)
pH determination (electrometric method)
 
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdf
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdfDrilling Fluid Engineering-PH Lab-Drilling Lab#9.pdf
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdf
 
Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.
Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.
Digital pH Meters Table Model by ACMAS Technologies Pvt Ltd.
 

Recently uploaded

Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...srsj9000
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSSIVASHANKAR N
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidNikhilNagaraju
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...ranjana rawat
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxpurnimasatapathy1234
 

Recently uploaded (20)

Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
Gfe Mayur Vihar Call Girls Service WhatsApp -> 9999965857 Available 24x7 ^ De...
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfid
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptx
 

pH Analzyer

  • 1.
  • 2. WHAT IS pH? pH is a measure of the acidity or alkalinity of a water solution. The acidity or alkalinity of a water solution is determined by the relative number of hydrogen ions (H+) or hydroxyl ions (OH-) present. Acidic solutions have a higher relative number of hydrogen ions, while alkaline (also called basic) solutions have a higher relative number of hydroxyl ions. Acids are substances which either dissociate (split apart) to release hydrogen ions or react with water to form hydrogen ions. Bases are substances that dissociate to release hydroxyl ions or react with water to form hydroxyl ions.
  • 3. In practice, pH is often assumed to be the negative logarithm of the hydrogen ion concentration: pH = - log10 [H+]
  • 4. At 25°C, a neutral solution has a pH of 7.0, while solutions with pH < 7 are acidic and solutions with pH > 7 are alkaline. The normal overall pH range is 0 to 14 pH, although solutions containing non-water solvents can have pH values outside this range.
  • 5. The cell consists of a measuring electrode, a reference electrode, a temperature sensing element, and the liquid being measured. The voltage of the cell is directly proportional to the pH of the liquid. The pH meter measures the voltage and uses a temperature-dependent factor to convert the voltage to pH. Because the cell has high internal resistance, the pH meter must have a very high input impedance. THE MEASUREMENT OF pH
  • 6. The pH electrode uses a specially formulated, pH sensitive glass in contact with the solution, which develops a potential (voltage) proportional to the pH of the solution. A glass pH electrode consists of an inert glass tube with a pH sensitive glass tip, either hemispherical (bulb) or flat in shape, blown onto it. The tip contains a fill solution with a known pH,
  • 7. The reference electrode is designed to maintain Constant potential at any given temperature Serves to complete the pH measuring circuit within the solution It provides a known reference potential for the pH electrode. The difference in the potentials of the pH and reference electrodes provides a millivolt signal proportional to pH..
  • 8. Most pH sensors are designed to produce a 0 mV signal at 7.0 pH, with a (theoretically ideal) slope (sensitivity) of -59.16 mV / pH at 25°C
  • 9. The millivolt signals produced by the pH and reference electrodes are temperature dependent. However, the pH and reference electrode combination exhibits an isopotential point, which is a pH and millivolt potential at which the potential is constant with temperature changes. The isopotential point is most frequently designed to be at 7.0 pH and 0 mV. Using the isopotential point with a theoretical knowledge of electrode behavior makes it possible to compensate (correct) the pH measurement at any temperature to a reference temperature (usually 25°C),
  • 10. Compensation (correct) the pH measurement Compensation (correct) the pH measurement means to remove the temperature dependence in the intercept and to shift the origin of the axes to pH 7. The result is plotted in the Figure. Two lines appear on the graph. One line shows how cell voltage changes with pH at 25°C, and the other line shows the relationship at 50°C. The lines, which are commonly called isotherms, intersect at the point (pH 7, 0 mV). An entire family of curves, each having a slope determined by the temperature and all passing through the point (pH 7, 0 mV) can be drawn on the graph. Shows why temperature is important in making pH measurements. When temperature changes, the slope of the isotherm changes. Therefore, a given cell voltage corresponds to a different pH value, depending on the temperature.
  • 11. Reference Poisoning The mechanism of reference poisoning is a conversion of the reference from a silver-silver chloride based electrode to an electrode based on a different silver compound. These ions include bromide, iodide, and sulfide ions. When these ions enter the fill solution, they form insoluble precipitates with the silver ions in the fill solution. But there is no initial effect on the potential of the reference, because the silver ions lost to precipitation are replenished by silver ions dissolving off the silver chloride coating of the silver wire. It is not until the silver chloride coating is completely lost that a large change in the potential so the reference electrode must be replaced. To counter this effect, multiple junction reference electrodes are used, which consist of two or more liquid junctions and fill solutions to slow the progress of the poisoning ions. Gelling of the reference fill solution is also used to prevent the transport of poisoning ions by convection.
  • 12. Liquid Junction Potential Potassium chloride is chosen for the fill solution not only for its ability to solubilize silver ion, but also because it is equitransferent. Equitransferent means that the positive potassium ion and the negative chloride ion diffuse through a water solution at nearly the same rate. When diffusing through the liquid junction of a reference electrode, both the positive and negative ions tend to move together and maintain a net zero charge at all points within the liquid junction. However, not all solutions are equitransferent, and when the process solution is not, a liquid junction potential can result. When a positive ion diffuses through the liquid junction faster than a negative ion, or vice versa, a charge imbalance will result. This gives rise to an opposing potential, which is liquid junction potential. This potential gets added to the potentials of the pH and reference electrodes and causes an offset to the pH measurement of typically a few tenths of a pH.
  • 13. pH IN AQUEOUS SOLUTIONS Practical pH Measurement Range Although the range of pH measurements is defined to be 0 to 14 pH, solutions with pH values near the extremes of this range are often better measured using conductivity.
  • 14. Tx: Ph202 YOKOGAWA Sensor: PH8EHP Place :DEMI WATER
  • 15. The Model PH8EHP pH Sensor is used to configure a pure water pH transmission system. •Measurement principle: Glass electrode method •Sample temperature range : 0 to 508C •KCl solution consumption: Maximum 3 ml/day •Sample pressure : Under atmospheric pressure •Sample flow rate : 30 to 600 ml/min Note : The maximum flow rate is restricted by the measurement solution conductivity. Process application:why put at this phase in the process?
  • 17. XMT-P CALIBRATING pH For pH sensors, two-point buffer calibration is standard. Both automatic calibration and manual calibration are available Auto calibration avoids common pitfalls and reduces errors. Its use is recommended. In auto calibration the Solu Comp Xmt calculates the actual pH of the buffer from the nominal value entered by the user and does not accept calibration data until readings are stable Manual calibration the user enters buffer values and judges when readings are stable. The pH reading can also be standardized, that is, forced to match the reading from a referee instrument. Finally, if the user knows the electrode slope (at 25°C), he can enter it directly.
  • 23. XMT-P ENTERING A KNOWN SLOPE VALUE CALIBRATING pH If the electrode slope is known from other measurements, it can be entered directly into the transmitter. The slope must be entered as the slope at 25°C. To calculate the slope at 25°C from the slope at temperature t°C, use the equation:
  • 25. Broken pH Glass and pH Glass Z High These messages mean that the pH sensor glass impedance is outside the programmed limits. To read the impedance go to the main display and press until Glass Imp appears in the display. The default lower limit is 10 MΩ. The default upper limit is 1000 MΩ. Low glass impedance means the glass membrane — the sensing element in a pH sensor — is cracked or broken. High glass impedance means the membrane is aging and nearing the end of its useful life. High impedance can also mean the pH sensor is not completely submerged in the process liquid. 1. Check the sensor wiring, including connections in a junction box. 2. Verify that the sensor is completely submerged in the process liquid. 3. Check the sensor response in buffers. If the sensor can be calibrated, it is in satisfactory condition. To disable the fault message, reprogram the glass impedance limits XMT-P TROUBLESHOOTING