What is a PH meter?
Theory of operation
Electrode
Operation modes
What is a PH meter?
 The pH meter is a laboratory equipment which used to

measure acidity or alkalinity of a solution.
 pH = -log [H+]





pH = 4.0 acidic solution
pH = 7.0 neutral solution
pH = 10.0 basic solution
0< pH <14

 The pH meter measures the concentration of hydrogen ions

[H+] using an ion-sensitive electrode.
Theory of operation
Operation is similar to galvanize cell,
Electrode or pH probe
Different operating modes
a) Standby mode -In this position the electrodes are protected from
electrical currents. It is the position used for maintaining the
equipment while stored.

b) pH mode- In this position the equipment can take pH
measurements after performing the required calibration procedures

c) Millivolt mode (mV).In this position the equipment is capable
of performing millivoltage readings.

d) ATC mode- The Automatic Temperature Control mode is used
when the pH is measured in solutions for which the temperature
varies. This function requires the use of a special probe. Not all pH
meters have this control.
Calibration
a) One point calibration- This is carried out for normal

working conditions and for normal use. It uses one
known pH reference solution.
b) Two point calibration - This is done prior to
performing very precise measurements. It uses two
known pH reference solutions.
 Necessary calibrating steps are described in each
operating manuals and it should be carried out
according to manufacturer instructions.
Calibration
One point calibration

Two point calibration

 Adjust the temperature selector

 Adjust the temperature selector to the






to the environmental
temperature.
Rinse the electrode with distilled
water in an empty beaker.
Dry the electrode
Place electrodes in the
calibration solution. (pH=7.0)
Adjust the meter to read the pH
of the calibration solution using
the button marked Cal










environmental temperature.
Rinse the electrode with distilled water in
an empty beaker.
Dry the electrode
Place electrodes in the first calibration
solution. (pH=4.0)
Adjust the meter to read the pH of the
calibration solution using the button
marked Cal 1
Rinse the electrode with distilled water
Dry the electrode
Place electrodes in the first calibration
solution. (pH=10.0)
Adjust the meter to read the pH of the
calibration solution using the button
marked Cal 2
Measuring the pH of a solution
a)
b)
c)
d)
e)

Rinse the electrode with distilled water and dry it.
Place the electrode in the solution of unknown pH.
Turn the functions selector from the Standby
position to the pH position.
Read the pH of the solution on the meter’s scale or
the screen.
Turn the functions selector again to the Standby
position.
General Maintenance
pH Electrode : Every four
 Replace Electrolyte solution (KCl)
month the electrode considering affected
 Clean

pHand evaluate its:general physical
Analyzer’s Every six month

a)



b)

c)
d)

contaminants
General cleaning- Soak the pH electrode
in a 0.1 M HCl solution or 0.1 M HNO3,
for 20 minutes. Rinse with water.
Removal of deposits and bacteria- Soak
the pH electrode in a diluted domestic
bleach solution, for 10 minutes. Rinse
abundantly with water
Cleaning oil and grease- Rinse the pH
electrode with a mild detergent or with
methyl alcohol. Rinse with water.
Cleaning of protein deposits- Soak the
pH electrode in 1% pepsin and 0.1 M HCl
for 5 minutes. Rinse with water.

 Examine the exterior of the equipment









condition.
Test the connection cable and its
system of connections.
Examine the equipment controls.
Verify that these are in good condition
and activated without difficulty.
Verify the state of the electrode
carrying arm.
Check the batteries (if applicable);
change them if necessary.
Test its function by measuring the pH
of a known solution
Inspect the ground connection and
check for escaping current
Basic fault Identification
 1.0 The pH meter shows

unstable readings.
 2.0 The electrode’s response
is slow.
1.0 The pH meter shows unstable readings.
There are air bubbles in
the electrode.

The electrode is dirty.

The electrode is not
immersed.

 Soak the electrode to eliminate the

bubbles.

 Clean the electrode and recalibrate.

 Verify that the sample covers the tip of the

electrode perfectly.
The electrode is broken.
2.0 The electrode’s response is slow.

The electrode is dirty or
greasy.

 Clean the electrode and recalibrate.
3.0 The screen shows a calibration or error message.

There is a calibration
error.

 Recalibrate the pH meter.

The calibration of the
 Verify the buffer values used
buffer value is erroneous.

The electrode is dirty.

 Clean and calibrate the electrode.
End of pH Meter

More Related Content

What's hot (20)

Colorimetry
ColorimetryColorimetry
Colorimetry
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Analytical balance
Analytical balanceAnalytical balance
Analytical balance
 
Ion selective electrodes(ise)
Ion selective electrodes(ise)Ion selective electrodes(ise)
Ion selective electrodes(ise)
 
Partition chromatography 3
Partition chromatography 3Partition chromatography 3
Partition chromatography 3
 
water bath instrument
 water bath instrument water bath instrument
water bath instrument
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Calibration of glassware ppt
Calibration of glassware pptCalibration of glassware ppt
Calibration of glassware ppt
 
pH meter design and working principle
pH meter design and working principlepH meter design and working principle
pH meter design and working principle
 
pH Measurement
pH MeasurementpH Measurement
pH Measurement
 
pH & its measurement
pH & its measurementpH & its measurement
pH & its measurement
 
centrifuge principle and application
centrifuge principle and applicationcentrifuge principle and application
centrifuge principle and application
 
pH Meter
pH MeterpH Meter
pH Meter
 
Calibration of glassware
Calibration of glasswareCalibration of glassware
Calibration of glassware
 
Gas liquid chromatography
Gas liquid chromatographyGas liquid chromatography
Gas liquid chromatography
 
Chromatography and its types
Chromatography and its typesChromatography and its types
Chromatography and its types
 
Balances
BalancesBalances
Balances
 
Colorimeter
ColorimeterColorimeter
Colorimeter
 
Magnetic Stirrer a brief guide
Magnetic Stirrer a brief guideMagnetic Stirrer a brief guide
Magnetic Stirrer a brief guide
 
HPLC
HPLCHPLC
HPLC
 

Similar to PH meter

Analytical Measurements: Troubleshooting, Maintenance and the Future
Analytical Measurements:  Troubleshooting, Maintenance and the FutureAnalytical Measurements:  Troubleshooting, Maintenance and the Future
Analytical Measurements: Troubleshooting, Maintenance and the FutureISA Boston Section
 
Calibration of pH meter
Calibration of pH meterCalibration of pH meter
Calibration of pH meterJanine Samelo
 
Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhimegha bagri
 
Ph and ph meter and application in foods
Ph and ph meter and application in foodsPh and ph meter and application in foods
Ph and ph meter and application in foodsNirenjana V
 
Ph meter ppt.pptx
Ph meter ppt.pptxPh meter ppt.pptx
Ph meter ppt.pptxNeha757208
 
Analytical-Instrumentation.pptx
Analytical-Instrumentation.pptxAnalytical-Instrumentation.pptx
Analytical-Instrumentation.pptxSymonFeolino1
 

Similar to PH meter (20)

Analytical Measurements: Troubleshooting, Maintenance and the Future
Analytical Measurements:  Troubleshooting, Maintenance and the FutureAnalytical Measurements:  Troubleshooting, Maintenance and the Future
Analytical Measurements: Troubleshooting, Maintenance and the Future
 
Calibration of pH meter
Calibration of pH meterCalibration of pH meter
Calibration of pH meter
 
Presentation1
Presentation1Presentation1
Presentation1
 
pH meter.pdf
pH meter.pdfpH meter.pdf
pH meter.pdf
 
pH - MEASUREMENT .ppt
pH - MEASUREMENT .pptpH - MEASUREMENT .ppt
pH - MEASUREMENT .ppt
 
Industrial Analytical Instruments
Industrial Analytical Instruments Industrial Analytical Instruments
Industrial Analytical Instruments
 
Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhi
 
PH METER.pptx
PH METER.pptxPH METER.pptx
PH METER.pptx
 
70th training program of nitub shahin
70th training program of nitub shahin70th training program of nitub shahin
70th training program of nitub shahin
 
Ph and ph meter and application in foods
Ph and ph meter and application in foodsPh and ph meter and application in foods
Ph and ph meter and application in foods
 
Ph meter ppt.pptx
Ph meter ppt.pptxPh meter ppt.pptx
Ph meter ppt.pptx
 
18 pH Meter.pptx
18 pH Meter.pptx18 pH Meter.pptx
18 pH Meter.pptx
 
Analytical-Instrumentation.pptx
Analytical-Instrumentation.pptxAnalytical-Instrumentation.pptx
Analytical-Instrumentation.pptx
 
CALIBRATION OF pH METER
CALIBRATION OF pH METER CALIBRATION OF pH METER
CALIBRATION OF pH METER
 
Exp 1 (A)
Exp 1 (A)Exp 1 (A)
Exp 1 (A)
 
Experiment5 chem210
Experiment5 chem210Experiment5 chem210
Experiment5 chem210
 
Pharmaceutical validation
Pharmaceutical  validationPharmaceutical  validation
Pharmaceutical validation
 
Potentio lab report
Potentio lab reportPotentio lab report
Potentio lab report
 
pH Analzyer
pH AnalzyerpH Analzyer
pH Analzyer
 
Basics of pH
Basics of pHBasics of pH
Basics of pH
 

Recently uploaded

Dev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebDev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebUiPathCommunity
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupFlorian Wilhelm
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
costume and set research powerpoint presentation
costume and set research powerpoint presentationcostume and set research powerpoint presentation
costume and set research powerpoint presentationphoebematthew05
 
My Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationMy Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationRidwan Fadjar
 
Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Scott Keck-Warren
 
Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!Commit University
 
SAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxSAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxNavinnSomaal
 
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticsKotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticscarlostorres15106
 
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...shyamraj55
 
Install Stable Diffusion in windows machine
Install Stable Diffusion in windows machineInstall Stable Diffusion in windows machine
Install Stable Diffusion in windows machinePadma Pradeep
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 3652toLead Limited
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfAlex Barbosa Coqueiro
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek SchlawackFwdays
 
Developer Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLDeveloper Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLScyllaDB
 
APIForce Zurich 5 April Automation LPDG
APIForce Zurich 5 April  Automation LPDGAPIForce Zurich 5 April  Automation LPDG
APIForce Zurich 5 April Automation LPDGMarianaLemus7
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationSafe Software
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsMemoori
 
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Mark Simos
 

Recently uploaded (20)

Dev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebDev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio Web
 
Streamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project SetupStreamlining Python Development: A Guide to a Modern Project Setup
Streamlining Python Development: A Guide to a Modern Project Setup
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
costume and set research powerpoint presentation
costume and set research powerpoint presentationcostume and set research powerpoint presentation
costume and set research powerpoint presentation
 
My Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationMy Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 Presentation
 
Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024Advanced Test Driven-Development @ php[tek] 2024
Advanced Test Driven-Development @ php[tek] 2024
 
Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!Nell’iperspazio con Rocket: il Framework Web di Rust!
Nell’iperspazio con Rocket: il Framework Web di Rust!
 
SAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxSAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptx
 
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmaticsKotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
Kotlin Multiplatform & Compose Multiplatform - Starter kit for pragmatics
 
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
 
Install Stable Diffusion in windows machine
Install Stable Diffusion in windows machineInstall Stable Diffusion in windows machine
Install Stable Diffusion in windows machine
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
 
Unraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdfUnraveling Multimodality with Large Language Models.pdf
Unraveling Multimodality with Large Language Models.pdf
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
 
Developer Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLDeveloper Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQL
 
APIForce Zurich 5 April Automation LPDG
APIForce Zurich 5 April  Automation LPDGAPIForce Zurich 5 April  Automation LPDG
APIForce Zurich 5 April Automation LPDG
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial Buildings
 
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
Tampa BSides - Chef's Tour of Microsoft Security Adoption Framework (SAF)
 

PH meter

  • 1. What is a PH meter? Theory of operation Electrode Operation modes
  • 2. What is a PH meter?  The pH meter is a laboratory equipment which used to measure acidity or alkalinity of a solution.  pH = -log [H+]    pH = 4.0 acidic solution pH = 7.0 neutral solution pH = 10.0 basic solution 0< pH <14  The pH meter measures the concentration of hydrogen ions [H+] using an ion-sensitive electrode.
  • 3. Theory of operation Operation is similar to galvanize cell,
  • 5. Different operating modes a) Standby mode -In this position the electrodes are protected from electrical currents. It is the position used for maintaining the equipment while stored. b) pH mode- In this position the equipment can take pH measurements after performing the required calibration procedures c) Millivolt mode (mV).In this position the equipment is capable of performing millivoltage readings. d) ATC mode- The Automatic Temperature Control mode is used when the pH is measured in solutions for which the temperature varies. This function requires the use of a special probe. Not all pH meters have this control.
  • 6. Calibration a) One point calibration- This is carried out for normal working conditions and for normal use. It uses one known pH reference solution. b) Two point calibration - This is done prior to performing very precise measurements. It uses two known pH reference solutions.  Necessary calibrating steps are described in each operating manuals and it should be carried out according to manufacturer instructions.
  • 7. Calibration One point calibration Two point calibration  Adjust the temperature selector  Adjust the temperature selector to the     to the environmental temperature. Rinse the electrode with distilled water in an empty beaker. Dry the electrode Place electrodes in the calibration solution. (pH=7.0) Adjust the meter to read the pH of the calibration solution using the button marked Cal         environmental temperature. Rinse the electrode with distilled water in an empty beaker. Dry the electrode Place electrodes in the first calibration solution. (pH=4.0) Adjust the meter to read the pH of the calibration solution using the button marked Cal 1 Rinse the electrode with distilled water Dry the electrode Place electrodes in the first calibration solution. (pH=10.0) Adjust the meter to read the pH of the calibration solution using the button marked Cal 2
  • 8. Measuring the pH of a solution a) b) c) d) e) Rinse the electrode with distilled water and dry it. Place the electrode in the solution of unknown pH. Turn the functions selector from the Standby position to the pH position. Read the pH of the solution on the meter’s scale or the screen. Turn the functions selector again to the Standby position.
  • 9. General Maintenance pH Electrode : Every four  Replace Electrolyte solution (KCl) month the electrode considering affected  Clean pHand evaluate its:general physical Analyzer’s Every six month a)  b) c) d) contaminants General cleaning- Soak the pH electrode in a 0.1 M HCl solution or 0.1 M HNO3, for 20 minutes. Rinse with water. Removal of deposits and bacteria- Soak the pH electrode in a diluted domestic bleach solution, for 10 minutes. Rinse abundantly with water Cleaning oil and grease- Rinse the pH electrode with a mild detergent or with methyl alcohol. Rinse with water. Cleaning of protein deposits- Soak the pH electrode in 1% pepsin and 0.1 M HCl for 5 minutes. Rinse with water.  Examine the exterior of the equipment      condition. Test the connection cable and its system of connections. Examine the equipment controls. Verify that these are in good condition and activated without difficulty. Verify the state of the electrode carrying arm. Check the batteries (if applicable); change them if necessary. Test its function by measuring the pH of a known solution Inspect the ground connection and check for escaping current
  • 10. Basic fault Identification  1.0 The pH meter shows unstable readings.  2.0 The electrode’s response is slow.
  • 11. 1.0 The pH meter shows unstable readings. There are air bubbles in the electrode. The electrode is dirty. The electrode is not immersed.  Soak the electrode to eliminate the bubbles.  Clean the electrode and recalibrate.  Verify that the sample covers the tip of the electrode perfectly. The electrode is broken.
  • 12. 2.0 The electrode’s response is slow. The electrode is dirty or greasy.  Clean the electrode and recalibrate.
  • 13. 3.0 The screen shows a calibration or error message. There is a calibration error.  Recalibrate the pH meter. The calibration of the  Verify the buffer values used buffer value is erroneous. The electrode is dirty.  Clean and calibrate the electrode.