Intraoperative neurophysiological monitoring (IONM) is the use of electrophysiological methods such as electroencephalography (EEG), electromyography (EMG), and evoked potentials to monitor the functional integrity of certain neural structures during surgery. The purpose this lecture is to Introduce you to the neurophysiological signals in intra operative neurophysiological signals.
The AIM of IONM is to reduce the risk to the patient of iatrogenic damage to the nervous system, and/or to provide functional guidance to the surgeon and anesthesiologist.
3. AnuragTewariMD
Anurag Tewari MD
AMPLIFICATION
• The amplifier not only determines the morphology and polarity of its output
signal, it also amplifies the signal so that it can be
• FILTERED
• DIGITIZED
• DISPLAYED
• The amplifier in an IONM machine must be able to handle amplification of a
wide variety of neurologic signals
Anurag Tewari MD
3
Medsurant Health Academy
4. AnuragTewariMD
Anurag Tewari MD
AMPLIFICATION
• EP signals are in the order of 0.1 – 5 μV
• EEG signals are in the 10 to 100 μV range and
• EMG signals are in the millivolt range
Amplifier has to be
• SENSITIVE to record very tiny signals
• VERSATILE to amplify a wide range of signal types
Anurag Tewari MD
4
Medsurant Health Academy
5. AnuragTewariMD
Anurag Tewari MD
GAIN & SENSITIVITY
• Gain describes how much our equipment amplifies our signal before we see it
on the screen.
• This is different than display gain and is NOT assigned any units.
• Power Gain
• Voltage Gain
•P=V2/R
Anurag Tewari MD
5
Medsurant Health Academy
6. AnuragTewariMD
Anurag Tewari MD
GAIN & SENSITIVITY
• Sensitivity: Sensitivity or display gain on EP machines is a measure of
• INPUT VOLTAGE PER DIVISION on the screen
• The display screen is divided into equal intervals called divisions
Sensitivity = Input voltage / vertical deflection (in divisions or mm)
Anurag Tewari MD
6
Medsurant Health Academy
7. AnuragTewariMD
Anurag Tewari MDDECIBEL-POWER GAIN
AMPLIFIER
• Amplifier is an electronic device for increasing the amplitude of electrical signals
Anurag Tewari MD
7
Medsurant Health Academy
8. AnuragTewariMD
Anurag Tewari MDDECIBEL-POWER GAIN
AMPLIFIER
• The power gain (Ap) of an Amplifier is the logarithmic ratio between
• Input Power (Pi) and
• output power (Po)
Anurag Tewari MD
8
Medsurant Health Academy
11. AnuragTewariMD
Anurag Tewari MDDECIBEL-POWER GAIN
AMPLIFIER
Decibel – Attenuation: A reduction of one hundred times is expressed as -20dB
Anurag Tewari MD
11
Medsurant Health Academy
12. AnuragTewariMD
Anurag Tewari MD
OPERATIONAL AMPLIFIER
•The operational amplifier (op-amp) is the basic building
block of the amplification system in IONM machine
•It consists of
• two input poles and
• an output pole
Anurag Tewari MD
12
OUTPUT
INPUT
INPUT
Medsurant Health Academy
13. AnuragTewariMD
Anurag Tewari MD
OPERATIONAL AMPLIFIER
• Is a DIFFERENTIAL AMPLIFIER,
• means that its output depends on the nature of
both of its inputs
• The input marked (–), is the inverting pole,
also known as
• grid 1 (G1)
• the negative input, or the
• active input
• The other pole, (+), is the noninverting pole,
also known as
• grid 2 (G2)
• the positive input, or
• the reference input
Anurag Tewari MD
13
INVERTING POLE
NON-INVERTING POLE
Medsurant Health Academy
14. AnuragTewariMD
Anurag Tewari MD
OPERATIONAL AMPLIFIER
• The output signal consists of the difference between the input signals at the two
poles
• Signals introduced at the inverting pole are inverted at the output
• Signals introduced at the noninverting pole have the same polarity at the output
Anurag Tewari MD
14
NO output signalSame signal introduced at both poles
Medsurant Health Academy
15. AnuragTewariMD
Anurag Tewari MD
OPERATIONAL AMPLIFIER
• If an upgoing signal is placed at the inverting pole and no signal at the
noninverting pole
• If an upgoing signal is introduced at the noninverting pole and no signal at the
inverting pole, the resulting output will be an upgoing signal
Anurag Tewari MD
15
Output will be a DOWN-going signal
Output will be a UP-going signal
Medsurant Health Academy
17. AnuragTewariMD
Anurag Tewari MD
OPERATIONAL AMPLIFIER
• When NEAR-FIELD potentials are recorded
• The electrode over the generator site is
placed in the inverting pole of the op-amp
• The input to the noninverting pole is placed
at a site away from the generator, where no
activity will be recorded
Anurag Tewari MD
17
Medsurant Health Academy
18. AnuragTewariMD
Anurag Tewari MD
OPERATIONAL AMPLIFIER
• Electrically negative signals introduced into the op-amp are displayed as an
upgoing signal at the output
• This is why the normal convention for most IONM modalities is “NEGATIVE UP”
• One exception to this is the BAEP
Anurag Tewari MD
18
Medsurant Health Academy
19. AnuragTewariMD
Anurag Tewari MD
OPERATIONAL AMPLIFIER
To be classified as an operational amplifier, the circuit must have three
important characteristics
•Very HIGH gain
•Very HIGH input impedance
•Very LOW output impedance
Anurag Tewari MD
19
Medsurant Health Academy
20. AnuragTewariMD
Anurag Tewari MD
COMMON MODE-REJECTION
• Common mode rejection: elimination of signals that are common to both
inputs of a differential amplifier
• The benefit is that the far-field distant signals such as noise artifacts will be
rejected if they are common to both recording electrodes
• The closer the electrodes are to each other, the more alike their input signals
will be and the more far-field rejection you get
The danger of this is if your recording electrodes are too close together for the
signal you are trying to measure, you will reject much of the signal
Anurag Tewari MD
20
Medsurant Health Academy
21. AnuragTewariMD
Anurag Tewari MD
COMMON MODE-REJECTION
• DISCRIMINATION is the ability of an amplifier to pass differential-
mode signals and to reject common-mode signals
• It is measured as a common mode rejection ratio or CMRR
• Higher the common mode-rejection ratio, the more efficient the
amplification system is at discerning the differences between signals
introduced at the input poles
Anurag Tewari MD
21
Medsurant Health Academy
22. AnuragTewariMD
Anurag Tewari MD
COMMON MODE-REJECTION
• In an ideal op-amp, when the same signal is introduced at both input poles,
there should be absolutely no signal at the output pole
• In practice however, the op-amp’s ability to do this is imperfect
• There is always a stray signal that is seen at the output when a common signal
is introduced at the input
Anurag Tewari MD
22
Medsurant Health Academy
23. AnuragTewariMD
Anurag Tewari MD
COMMON MODE-REJECTION
• The ratio is as follows:
For every increase in 20dB the CMRR increases 10X
• Guidelines require EP amplifiers to have a minimum CMRR = 10,000:1 or 80 dB
Anurag Tewari MD
23
Medsurant Health Academy
24. AnuragTewariMD
Anurag Tewari MD
COMMON MODE-REJECTION
•Manufacturers offer amplifiers with common-mode rejection
•The common-mode rejection ratio given by manufacturers
refers to an ideal situation that is rarely attainable
•Thus, the practical obtainable common-mode rejection ratio is
lower than that given in the specifications for any amplifier
Anurag Tewari MD
24
Medsurant Health Academy
25. AnuragTewariMD
Anurag Tewari MD
INPUT IMPEDANCE
• Impedance as measure ACROSS THE INPUTS of a differential amplifier
• It is NOT the same as electrode to skin impedance
• We want impedance to be very high so that no current flows through the
recording electrodes, influencing the potentials being measured.
• High input impedance is important, as it ensures that most of the recorded
signal voltage will be across the inputs of the amplifier
• The ACNS guidelines state that neuro-diagnostic equipment’s amplification
system should have an input impedance of at least 100 Mega ohms
• Most of the commercially available IONM machines meet this specification
Anurag Tewari MD
25
Medsurant Health Academy
26. AnuragTewariMD
Anurag Tewari MD
HOW DOES INPUT IMPEDANCE WORK?
• The impedance of the patient and input impedance of the amplifier are in series
• Since resistors in series act as a voltage dividers, the voltage recorded across the
patient and the voltage recorded across the amplifier are proportional to their
respective impedances
Anurag Tewari MD
26
Medsurant Health Academy
27. AnuragTewariMD
Anurag Tewari MD
HOW DOES INPUT IMPEDANCE WORK?
• Assume that the impedance of the electrodes on the patient is 1 Kohm and the
input impedance across the amplifier is 100 MEGA Ohm
• That means that the portion of the total signal voltage across the amplifier is
100,000 times greater than that across the patient
Anurag Tewari MD
27
Medsurant Health Academy
28. AnuragTewariMD
Anurag Tewari MD
HOW DOES INPUT IMPEDANCE WORK?
• For all intents and purposes, all of the signal voltage is across the amplifier; only
a minuscule amount (1 x 10–5) is across the patient electrode
• Therefore the output signal is a true representation of the input signal
Anurag Tewari MD
28
Medsurant Health Academy
29. AnuragTewariMD
Anurag Tewari MD
HOW DOES INPUT IMPEDANCE WORK?
• Voltage Divider Rule:
• The voltage recorded across the patient and across the amplifier is proportional
to their impedance
• Vamp is 100,000 greater than Vpatient
Anurag Tewari MD
29
Medsurant Health Academy