SlideShare a Scribd company logo
• 7½ digits resolution 
• 100 pV, 100 nΩ sensitivity 
• 1.3 nVrms, 8 nVpp noise 
performance 
• Built-in low noise 2 channel 
scanner 
• Direct SPRT, RTD, Thermistor, and 
Thermocouple measurements 
Nanovolt Performance 
at a Microvolt Price 
The Agilent Technologies 34420A 
nanoVolt/micro-Ohm meter is a 
high-sensitivity multimeter optimized for 
performing low-level measurements. It 
combines low-noise voltage measurements 
with resistance and temperature functions, 
setting a new standard in low-level flexibil-ity 
and performance. 
Take the Uncertainty Out of Your 
Low-Level Measurements 
Low-noise input amplifiers and a highly 
tuned input protection scheme bring read-ing 
noise down to 8 nVpp. Combine this 
with 71/2 digits of resolution, selectable 
analog and digital filtering, 2 ppm basic 
24-hour dcV accuracy, and a shielded, cop-per 
pin connector and you’ve got accurate, 
repeatable measurements you can 
count on. 
Agilent 34420A 
NanoVolt/Micro-Ohm Meter 
Data Sheet 
Two Input Channels 
An integral two-channel programmable scan-ner 
simplifies voltage comparisons. Built-in 
ratio and difference functions enable 
auto mated two channel measurements 
without the need for an external nanoVolt 
scanner. Both channels share the same 
low noise specifications to ensure accurate 
comparisons. 
Built-In Resistance and 
Temperature 
The 34420A combines its low-noise nano- 
Volt input circuits with a high-stability 
current source to provide precise low-level 
resistance measurements – no more has-sling 
with the cost and complexity of an 
external current source. Three resistance 
modes are included: 
• Standard 
• Low-power 
• Voltage-limited for 
dry-circuit testing 
Offset compensation is also provided to 
minimize thermal EMFs and associated 
errors. 
SPRT Measurements 
Built-in ITS-90 conversion routines accept 
the calibration coefficients from your 
SPRT probe for direct temperature mea-surement 
and conversion. Thermocouples, 
thermistors, and RTDs are also supported. 
Unequaled Versatility 
The 34420A gives you the versatility to 
tackle your most challenging tasks, both 
on the benchtop and in your automated 
system. Standard features include RS-232 
and GPIB interfaces, SCPI and Keithley 
181 program ming language, 1024-reading 
memory, scaling and statistics, and a chart 
recorder analog output. 
Agilent IntuiLink:Easy Data 
Access 
The included Agilent IntuiLink 
software allows your captured data 
to be put to work easily, using PC applica-tions 
such as Microsoft Excel® or Word® 
to analyze, interpret, display, print, and doc-ument 
the data you get from the 34420A. 
You can specify the meter setup and take 
a single reading or log data to the Excel 
spreadsheet in specified time intervals. To 
find out more about IntuiLink visit 
www.agilent.com/find/intuilink 
Quality You Can Count On 
The 34420A gives you the quality and 
reliability you expect from Agilent 
Technologies. From the product’s proven 
>150,000 hour Mean Time Between Failure, 
to its standard 1-year warranty, Agilent 
stands behind you to bring a new level 
of confidence to your low-level measure-ments.
Specifications 
Accuracy Specifications ±(% of reading + % of range) 1 
Function Range2 Test Current 24 Hour 90 Day 1 Year Temperature Maximum 
2 
23 °C ± 1 °C 23 °C ± 5 °C 23 °C ± 5 °C Coefficient per Lead 
0 °C—18 °C Resistance 
28 °C—55 °C 
dc Voltage 1.0000000 mV3 0.0025 + .0020 0.0040 + .0020 0.0050 + .0020 0.0004 + .0001 
10.000000 mV3 0.0025 + .0002 0.0040 + .0002 0.0050 + .0003 0.0004 + .0001 
100.00000 mV 0.0015 + .0003 0.0030 + .0004 0.0040 + .0004 0.0004 + .00006 
1.0000000 V 0.0010 + .0003 0.0025 + .0004 0.0035 + .0004 0.0004 + .00004 
10.000000 V 0.0002 + .0001 0.0020 + .0004 0.0030 + .0004 0.0001 + .00002 
100.00000 V4 0.0010 + .0004 0.0025 + .0005 0.0035 + .0005 0.0004 + .00005 
Resistance 5 1.0000000 Ω 10 mA 0.0015 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω 
10.000000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 Ω 
100.00000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 10 Ω 
1.0000000 KΩ 1 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 100 Ω 
10.000000 KΩ 100 μA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 KΩ 
100.00000 KΩ 10 μA 0.0015 + .0003 0.0040 + .0004 0.0060 + .0004 0.0005 + .00002 1 KΩ 
1.0000000 MΩ 5 μA 0.0020 + .0003 0.0050 + .0004 0.0070 + .0004 0.0006 + .00003 1 KΩ 
Low Power 1.0000000 Ω 10 mA 0.0015 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω 
Resistance 5 10.000000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 Ω 
100.00000 Ω 1 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 10 Ω 
1.0000000 KΩ 100 μA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 100 Ω 
10.000000 KΩ 10 μA 0.0015 + .0004 0.0040 + .0004 0.0060 + .0004 0.0005 + .00001 1 KΩ 
100.00000 KΩ 5 μA 0.0015 + .0012 0.0040 + .0015 0.0060 + .0015 0.0005 + .00003 1 KΩ 
1.0000000 MΩ 5 μA 0.0020 + .0003 0.0050 + .0004 0.0070 + .0004 0.0006 + .00003 1 KΩ 
Voltage Limited 10.000000 Ω 1 mA 0.0020 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω 
Resistance 5, 6 100.00000 Ω 100 μA 0.0025 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 5 Ω 
Channel 1 / Channel 2 (dcV Ratio) Ratio Error in % = Channel 1 accuracy in % + Channel 2 accuracy in % 
Channel 1-Channel 2 (dcV Difference) Difference Error = Channel 1 (% of reading + % of range) + Channel 2 (% of reading + % of range) 
Temperature (resolution = 0.001 °C) 
SPRT 7 SPRT Probe Accuracy + 0.003°C 
RTD RTD Probe Accuracy + 0.05°C 
Thermistor Thermistor Probe Accuracy + 0.1°C 
Thermocouple 8 Thermocouple Probe Accuracy + 0.2°C 
DC Voltage Noise 9 
Observation Period 
Range 2-Minute 2-Minute 24-Hour 
RMS Peak-Peak Peak-Peak 
Noise Noise Noise 
1 mV 1.3 nVrms 8 nVpp 12 nVpp 
10 mV 1.5 nVrms 10 nVpp 14 nVpp 
100 mV 10 nVrms 65 nVpp 80 nVpp 
1 V 100 nVrms 650 nVpp 800 nVpp 
10 V 450 nVrms 3 μVpp 3.7 μVpp 
100 V 11 μVrms 75 μVpp 90 μVpp 
DC Voltage Noise vs Source Resistance 10 
Source Analog Digital 
Resistance Noise Filter Filter 
0 Ω 1.3 nVrms Off Med 
100 Ω 1.7 nVrms Off Med 
1k Ω 4 nVrms Off Med 
10k Ω 13 nVrms Off Med 
100k Ω 41 nVrms On Med 
1M Ω 90 nVrms On Slow 
1 Specifications are for Channel 1 or Channel 2, after 2-hour warm-up, resolution 
at 7.5 digits (100 NPLC), with FILTERS off. RESISTANCE specifications are for 
4-wire Ohms or 2-wire ohms using Null. Without Null, add 0.2 Ohms additional 
error in 2-wire Ohms function. For Analog Filter ON, add 0.002% of reading. 
2 20% overrange on all ranges except 5% on Voltage Limited Resistance. 
3 After using Math Null. If Null is not used add 100 nanoVolts. 
4 Channel 1 only. 
5 Channel 1 only. Resistance measurements, for NPLC <1, add 160 μΩ rms 
noise. 
6 Voltage limit can be set to 20 mV (default), 100 mV, or 500 mV. Measured 
resistance plus Channel 1 HI and LO lead resistance is limited to 10.5 Ω on the 
10 Ω range and 105 Ω on the 100 Ω range. 
7 For 25 Ω SPRT with triple-point of water check within the last 4 hours. 
Without the triple-point of water check, add 0.013°C for 24 hours, add 0.035°C 
for 
90 day, and add 0.055°C for 1 year specifications. 
8 For fixed reference junction. Add 0.3ûC for external reference junction, add 
2.0ûC for internal reference junction. 
9 After a 2-hour warm-up, ± 1ûC, 6.5 digits (10 PLC) with Analog Filter Off and 
Digital Filter Medium (50 reading average). 
2-minute rms and 24-hour noise typical. For measurements using 0.02 or 0.2 
NPLC, add 800 nV rms noise. 
10 Typical noise behavior for Ch 1 or Ch 2, after 2 hour warm-up, 6.5 digits (10 
PLC), 2 minute observation period on 1 mV range. For peak-to-peak noise, mul-tiply 
rms noise by 6.
3 
Measurement Characteristics 
DC Voltage 
Measurement Method: 
Continuously integrating multi-slope III 
A-D Converter 
A-D Linearity: 
0.00008% of reading + 0.00005% of range 
Input Resistance: 
100V (Ch1 only): 10 MΩ +- 1% 
1mV through 10V: > 10 GΩ, in parallel with < 3.6 nF 
Input Bias Current: <50 pA at 25 °C 
Injected Current: <50 nA pp at 50 or 60 Hz 
Input Protection: 
150 V peak any input terminal 
to Channel 1 LO, continuous 
Channel-to-channel switching error (typical): 
3 nV 
Channel Isolation: 
Isolation between input channels >1010 Ω 
Earth Isolation: 
350 V peak any input terminal to earth. 
Impedance from any input terminal to earth 
is >10 GΩ and <400 pF 
Maximum Voltage: 
Channel 1 LO to Channel 2 LO, 150V peak 
Resistance 
Measurement Method: 
Selectable 4-wire or 2-wire ohms. Current 
Source referenced to Channel 1 LO input 
Offset Compensation: 
Used on all ranges except 100 kΩ and 1 MΩ. 
Can be turned off if desired 
Protection: 150 V peak 
Open Circuit Voltage: 
For Resistance and Low Power Resistance 
<14 V. 20 mV, 100 mV, 500 mV selectable clamp 
Temperature 
SPRT: 
ITS-90 calibrated temperature with the range 
of -190°C to +660°C 
Thermocouple: 
ITS-90 conversions of Type B, E, J, K, N, R, S, T 
Thermistor: 5 kΩ 
RTD: Type α =.00385 and α =.00392. R0 from 
4.9 Ω to 2.1 kΩ. ITS -90 (IEC-751) Callendar 
Van Dusen conversion. 
Measurement Noise Rejection 60 (50) Hz1 
dc CMRR: 140 dB 
ac CMRR: 70 dB 
Integration Time Normal Mode 
Rejection2 
200 plc/3.335 s (4 s) 110 dB3 
100 plc/1.675 s (2 s) 105 dB3 
20 plc/334 ms (400 ms) 100 dB3 
10 plc/167 ms (200 ms) 95 dB3 
2 plc/33.3 ms (40 ms) 90 dB 
1 plc/16.7 ms (20 ms) 60 dB 
<1 plc 0 
Operating Characteristics4 
Function Digits Integration Time Readings/s 5 
dcV 71/2 200 plc .15 (.125) 
Thermocouple 71/2 100 plc .3 (.25) 
61/2 20 plc 1.5 (1.25) 
61/2 10 plc 3 (2.5) 
51/2 1 plc 25 (20.8) 
51/2 0.2 plc 100 (100) 
41/2 0.02 plc 250 (250) 
Resistance 71/2 200 plc .075 (.062) 
dcV1/DCV2 71/2 100 plc .15 (.125) 
dcV 1-2 61/2 20 plc .75 (.625) 
RTD 61/2 10 plc 1.5 (1.25) 
Thermistor 51/2 1 plc 12.5 (10.4) 
0.2 plc 50 (50) 
41/2 
0.02 plc 125 (125) 
System Speeds6 
Configuration Rates: 26/s to 50/s 
Autorange Rate (Volts): >30/s 
ASCII reading to RS-232: 55/s 
ASCII reading to GPIB: 250/s 
Max. Internal Trigger Rate: 250/s 
Max. Ext. Trig. Rate to Memory: 250/s 
Triggering and Memory 
Reading HOLD Sensitivity: 
10%, 1%, 0.1%, or 0.01% of range 
Samples/Trigger: 1 to 50,000 
Trigger Delay: 0 to 3600 s; 10 μs step size 
External Trigger Delay: <1 ms 
External Trigger Jitter: <500 μs 
Memory: 1024 readings 
Math Functions 
NULL (Channel 1 dcV, Channel 2 dcV, 
Difference, Resistance, Temperature) 
STATS (Min, Max, Average, Peak-Peak, 
Standard Deviation, Number of readings) 
SCALE (Allows linear scaling as y = mx+b) 
CHART NULL (Establishes zero for rear panel 
output) 
Filter (Analog or Digital or Both) 
Analog: 
Low pass 2 pole @ 13Hz, available for dcV 
on 
1 mV, 10 mV, 100 mV range 
Digital: 
Moving average filter, 10 (fast), 50 (medium), 
or 100 (slow) reading averages. 
Chart Out (Analog Out) 
Maximum output: ± 3V 
Resolution: 16 bits 
Accuracy: ± 0.1% of output + 1 mV 
Output Resistance: 1 kΩ ± 5% 
Update rate: once per reading 
Span and Offset: Adjustable 
Standard Programming Languages 
SCPI (IEEE 488.2), Keithley 181 
Accessories Included 
4 ft low thermal cable with copper spade lugs, 
Kelvin clip set, 4-wire shorting plug, user’s man-ual, 
service manual, test report, contact cleaner, 
and power cord. 
General Specifications 
Front Panel Connection: 
Shielded, low thermal, 99% copper contacts. 
Power Supply: 
100V/120V/220V(230V)/240V +- 10%. 
Power Line Frequency: 
45 Hz to 66 Hz and 360 Hz to 440 Hz. 
Automatically sensed at power-on. 
Power Consumption: 
25VA peak (10W average). 
Operating Environment: 
Full accuracy for 0 °C to 55 °C. Full accuracy 
to 80% R.H. up to 30 °C. 
Storage Environment: 
-40 °C to 75 °C. 
Size: 254.4 mm W x 374.0 mm L x 103.6 mm H 
(10.02” W x 14.72” L x 4.08” H) 
Weight: 3 kg (6.5 lbs). 
Safety: 
Designed to CSA, UL-1244, IEC-1010. 
RFI and ESD: CISPR 11. 
1 For 1 kΩ unbalanced in LO lead. 
2 For power line frequency ± - 0.1%, Filters OFF. For 
Digital Filter slow add 20 db, for medium or fast add 
10 db for NPLC ³ 1. 
3 For power line frequency ± - 1%, use 80 db, 
for ± - 3% use 60 db. 
4 Speeds are for delay 0, Display OFF, Filters OFF, 
Offset Compensation OFF. 
5 Reading speeds for 60 Hz or (50 Hz), 100 mV 
through 100 V ranges. 1 mV range 30/s MAX, 10 
mV range 170/s MAX, thermocouple 120/s MAX. 
6 Speeds are for NPLC 0.02, Delay 0, Display OFF, 
Chart Out OFF.
Agilent Email Updates 
www.agilent.com/find/emailupdates 
Get the latest information on the 
products and applications you select. 
www.lxistandard.org 
LAN eXtensions for Instruments puts 
the power of Ethernet and the Web 
inside your test systems. Agilent is a 
founding member of the LXI consor-tium. 
Agilent Channel Partners 
www.agilent.com/find/channelpartners 
Get the best of both worlds: Agilent’s 
measurement expertise and product 
breadth, combined with channel 
partner convenience. 
www.agilent.com 
For more information on Agilent Tech-nologies’ 
products, applications or services, 
please contact your local Agilent office. The 
complete list is available at: 
www.agilent.com/find/contactus 
Americas 
Canada (877) 894 4414 
Brazil (11) 4197 3500 
Mexico 01800 5064 800 
United States (800) 829 4444 
Asia Pacific 
Australia 1 800 629 485 
China 800 810 0189 
Hong Kong 800 938 693 
India 1 800 112 929 
Japan 0120 (421) 345 
Korea 080 769 0800 
Malaysia 1 800 888 848 
Singapore 1 800 375 8100 
Taiwan 0800 047 866 
Other AP Countries (65) 375 8100 
Europe & Middle East 
Belgium 32 (0) 2 404 93 40 
Denmark 45 70 13 15 15 
Finland 358 (0) 10 855 2100 
France 0825 010 700* 
*0.125 €/minute 
Germany 49 (0) 7031 464 6333 
Ireland 1890 924 204 
Israel 972-3-9288-504/544 
Italy 39 02 92 60 8484 
Netherlands 31 (0) 20 547 2111 
Spain 34 (91) 631 3300 
Sweden 0200-88 22 55 
United Kingdom 44 (0) 131 452 0200 
For other unlisted countries: 
www.agilent.com/find/contactus 
Revised: June 8, 2011 
Product specifications and descriptions 
in this document subject to change 
without notice. 
© Agilent Technologies, Inc. 2011 
Published in USA, October 11, 2011 
5968-0161EN 
Agilent Advantage Services is committed 
to your success throughout your equip-ment’s 
lifetime. To keep you competitive, 
we continually invest in tools and 
processes that speed up calibration and 
repair and reduce your cost of ownership. 
You can also use Infoline Web Services 
to manage equipment and services more 
effectively. By sharing our measurement 
and service expertise, we help you create 
the products that change our world. 
www.agilent.com/find/advantageservices 
www.agilent.com/quality 
Ordering Information 
Includes low thermal input cable 
(34102A), low thermal shorting plug 
(34103A), Kelvin clip set (11062A), cali-bration 
certificate, power cord. 
Also includes CD with: IntuiLink software, 
IVI and VXI PnP drivers, user's guide, 
service guide, and data sheet. 
Options 
34420A-ABA English localization 
34420A-ABD German localization: 
translated operating manual 
34420A-ABF French localization: 
translated operating manual 
34420A-ABJ Japanese localization: 
translated operating manual 
34420A-A6J ANSI Z540 compliant 
calibration 
Accessories 
34102A Low-thermal input cable (four-conductor) 
with copper spade lugs 
34103A Low-thermal shorting plug 
34104A Low-thermal input connector 
34131A Transit Case 
34161A Accessory pouch 
34190A Rackmount Kit: designed for use 
with only one instrument, mounted 
on either the left or the right side 
of the rack. 
34191A 2U Dual Flange Kit: secures the 
instrument to the front of the rack. 
This kit can be used with the 
34194A Dual Lock Link Kit to mount 
two half-width, 2U height instruments 
side-by-side. 
34194A Dual Lock Link Kit: recommended 
for side-by-side combinations and 
includes links for instruments of differ-ent 
depths. This kit can be used 
with the 34191A 2U Dual Flange Kit 
to mount two half-width, 2U height 
instruments side-by-side.

More Related Content

What's hot

K48 Controller
K48 ControllerK48 Controller
K48 Controller
Joseph
 
Aml catalogue 13
Aml catalogue 13Aml catalogue 13
Aml catalogue 13slideroma
 
Catalog Relay Idec - Beeteco.com
Catalog Relay Idec - Beeteco.comCatalog Relay Idec - Beeteco.com
Catalog Relay Idec - Beeteco.com
Beeteco
 
Original Relay G5RL 1A 12V New Omron 
Original Relay G5RL 1A 12V New Omron Original Relay G5RL 1A 12V New Omron 
Original Relay G5RL 1A 12V New Omron 
AUTHELECTRONIC
 
Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...
Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...
Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...
PT. Siwali Swantika
 
Yehai catalogue -tracy chow
Yehai catalogue -tracy chowYehai catalogue -tracy chow
Yehai catalogue -tracy chowslideroma
 
Catalog relay idec 2019
Catalog relay idec 2019Catalog relay idec 2019
Catalog relay idec 2019
CTY TNHH HẠO PHƯƠNG
 
Catalog timer idec mới nhất 2019
Catalog timer idec mới nhất 2019Catalog timer idec mới nhất 2019
Catalog timer idec mới nhất 2019
CTY TNHH HẠO PHƯƠNG
 
VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...
VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...
VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...Ricky Henderson
 
Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
180625_W600 PH WCT DIS Wall Mount
180625_W600 PH WCT DIS Wall Mount180625_W600 PH WCT DIS Wall Mount
180625_W600 PH WCT DIS Wall MountEn Campos
 
Original High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 New
Original High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 NewOriginal High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 New
Original High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 New
AUTHELECTRONIC
 
Catalog bộ nguồn omron S8 FS-C Beeteco.com
Catalog bộ nguồn omron S8 FS-C Beeteco.comCatalog bộ nguồn omron S8 FS-C Beeteco.com
Catalog bộ nguồn omron S8 FS-C Beeteco.com
Beeteco
 
Cataloge schneider contactor ct-dienhathe.vn
Cataloge schneider contactor ct-dienhathe.vnCataloge schneider contactor ct-dienhathe.vn
Cataloge schneider contactor ct-dienhathe.vn
Dien Ha The
 

What's hot (17)

K48 Controller
K48 ControllerK48 Controller
K48 Controller
 
Aml catalogue 13
Aml catalogue 13Aml catalogue 13
Aml catalogue 13
 
PTC31 - PTP31
PTC31 - PTP31PTC31 - PTP31
PTC31 - PTP31
 
Ad524 e
Ad524 eAd524 e
Ad524 e
 
Catalog Relay Idec - Beeteco.com
Catalog Relay Idec - Beeteco.comCatalog Relay Idec - Beeteco.com
Catalog Relay Idec - Beeteco.com
 
Original Relay G5RL 1A 12V New Omron 
Original Relay G5RL 1A 12V New Omron Original Relay G5RL 1A 12V New Omron 
Original Relay G5RL 1A 12V New Omron 
 
Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...
Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...
Datasheet Fluke Automated AC Measurement Standard. Hubungi PT. Siwali Swantik...
 
Yehai catalogue -tracy chow
Yehai catalogue -tracy chowYehai catalogue -tracy chow
Yehai catalogue -tracy chow
 
Catalog relay idec 2019
Catalog relay idec 2019Catalog relay idec 2019
Catalog relay idec 2019
 
Catalog timer idec mới nhất 2019
Catalog timer idec mới nhất 2019Catalog timer idec mới nhất 2019
Catalog timer idec mới nhất 2019
 
VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...
VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...
VAISALA MHT410 HYDROGEN+MOISTURE-IN-OIL TRANSMITTER NA PRODUCT ORDER GUIDE AN...
 
Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 1750. Hubungi PT. Siwali Swantika 021-45850618
 
180625_W600 PH WCT DIS Wall Mount
180625_W600 PH WCT DIS Wall Mount180625_W600 PH WCT DIS Wall Mount
180625_W600 PH WCT DIS Wall Mount
 
Original High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 New
Original High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 NewOriginal High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 New
Original High Voltage Isolation IC ACS710T KLA-12CB 710T SOP-16 New
 
Catalog bộ nguồn omron S8 FS-C Beeteco.com
Catalog bộ nguồn omron S8 FS-C Beeteco.comCatalog bộ nguồn omron S8 FS-C Beeteco.com
Catalog bộ nguồn omron S8 FS-C Beeteco.com
 
PTP35
PTP35PTP35
PTP35
 
Cataloge schneider contactor ct-dienhathe.vn
Cataloge schneider contactor ct-dienhathe.vnCataloge schneider contactor ct-dienhathe.vn
Cataloge schneider contactor ct-dienhathe.vn
 

Viewers also liked

Programacion Estructurada Unidad 3 Practica 5
Programacion Estructurada Unidad 3 Practica 5Programacion Estructurada Unidad 3 Practica 5
Programacion Estructurada Unidad 3 Practica 5
Instituto Tecnológico de Tijuana
 
evaluacion
evaluacionevaluacion
evaluacion
yohananq
 
Universal design
Universal designUniversal design
Universal designbarlowmp
 
Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543
Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543
Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543
Estojos
 
Big DiP US 2014 - Sponsorship Introduction
Big DiP US 2014 - Sponsorship IntroductionBig DiP US 2014 - Sponsorship Introduction
Big DiP US 2014 - Sponsorship IntroductionCurtis Dingley
 
Ppt(은평청년 잠비아에 가다)2
Ppt(은평청년 잠비아에 가다)2Ppt(은평청년 잠비아에 가다)2
Ppt(은평청년 잠비아에 가다)2SunMi Ban
 
Effects of Wind Direction on VOC Concentrations in Southeast Kansas
Effects of Wind Direction on VOC  Concentrations in Southeast KansasEffects of Wind Direction on VOC  Concentrations in Southeast Kansas
Effects of Wind Direction on VOC Concentrations in Southeast Kansas
Sergio A. Guerra
 
Pieces of Wales
Pieces of WalesPieces of Wales
Pieces of Wales
Swansea-Jack
 
Evaluation question 3
Evaluation question 3Evaluation question 3
Evaluation question 3SeyiDaramola
 
Sample
SampleSample
Sample
ijuanmiguel
 
DOM E-5064 09102015
DOM E-5064 09102015DOM E-5064 09102015
DOM E-5064 09102015
samir_bhowmik
 
Marketing director cover letter
Marketing director cover letterMarketing director cover letter
Marketing director cover letterclairerobin425
 
Pendekatan saintifik ilmiah
Pendekatan saintifik ilmiahPendekatan saintifik ilmiah
Pendekatan saintifik ilmiah
Joe Zidane
 
Mi calicatura calcada
Mi calicatura calcadaMi calicatura calcada
Mi calicatura calcadajuancho805
 
Levinas and Education - Chapter 9
Levinas and Education - Chapter 9 Levinas and Education - Chapter 9
Levinas and Education - Chapter 9
Grecia Falzon
 
CPP PRIME2 Next-Generation-Downwash_Model
CPP PRIME2 Next-Generation-Downwash_ModelCPP PRIME2 Next-Generation-Downwash_Model
CPP PRIME2 Next-Generation-Downwash_Model
Sergio A. Guerra
 

Viewers also liked (18)

Programacion Estructurada Unidad 3 Practica 5
Programacion Estructurada Unidad 3 Practica 5Programacion Estructurada Unidad 3 Practica 5
Programacion Estructurada Unidad 3 Practica 5
 
evaluacion
evaluacionevaluacion
evaluacion
 
Universal design
Universal designUniversal design
Universal design
 
Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543
Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543
Necessaire Masculina Feminina Infantil Esportiva Personalizadas (11)98950-3543
 
Big DiP US 2014 - Sponsorship Introduction
Big DiP US 2014 - Sponsorship IntroductionBig DiP US 2014 - Sponsorship Introduction
Big DiP US 2014 - Sponsorship Introduction
 
Ppt(은평청년 잠비아에 가다)2
Ppt(은평청년 잠비아에 가다)2Ppt(은평청년 잠비아에 가다)2
Ppt(은평청년 잠비아에 가다)2
 
Effects of Wind Direction on VOC Concentrations in Southeast Kansas
Effects of Wind Direction on VOC  Concentrations in Southeast KansasEffects of Wind Direction on VOC  Concentrations in Southeast Kansas
Effects of Wind Direction on VOC Concentrations in Southeast Kansas
 
Pieces of Wales
Pieces of WalesPieces of Wales
Pieces of Wales
 
Evaluation question 3
Evaluation question 3Evaluation question 3
Evaluation question 3
 
Sample
SampleSample
Sample
 
Calibration
CalibrationCalibration
Calibration
 
DOM E-5064 09102015
DOM E-5064 09102015DOM E-5064 09102015
DOM E-5064 09102015
 
Marketing director cover letter
Marketing director cover letterMarketing director cover letter
Marketing director cover letter
 
Pendekatan saintifik ilmiah
Pendekatan saintifik ilmiahPendekatan saintifik ilmiah
Pendekatan saintifik ilmiah
 
Mi calicatura calcada
Mi calicatura calcadaMi calicatura calcada
Mi calicatura calcada
 
Levinas and Education - Chapter 9
Levinas and Education - Chapter 9 Levinas and Education - Chapter 9
Levinas and Education - Chapter 9
 
TURNOS DE RECOLECCIÓN DE BASURA EN ASUNCIÓN.
TURNOS DE RECOLECCIÓN DE BASURA EN ASUNCIÓN.TURNOS DE RECOLECCIÓN DE BASURA EN ASUNCIÓN.
TURNOS DE RECOLECCIÓN DE BASURA EN ASUNCIÓN.
 
CPP PRIME2 Next-Generation-Downwash_Model
CPP PRIME2 Next-Generation-Downwash_ModelCPP PRIME2 Next-Generation-Downwash_Model
CPP PRIME2 Next-Generation-Downwash_Model
 

Similar to Katalog agilent-34420 a-micro-ohm-meter-tridinamika

Katalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamikaKatalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamikaKatalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamika
PT. Tridinamika Jaya Instrument
 
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTHData Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
PT. Siwali Swantika
 
Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
Hioki series battery_tester_denkei
Hioki series battery_tester_denkeiHioki series battery_tester_denkei
Hioki series battery_tester_denkei
NIHON DENKEI SINGAPORE
 
Insulation Test Hioki 3455
Insulation Test Hioki 3455Insulation Test Hioki 3455
Insulation Test Hioki 3455Nanan Ginanjar
 
Hioki High End digital multimeter DT4281
Hioki High End digital multimeter DT4281Hioki High End digital multimeter DT4281
Hioki High End digital multimeter DT4281
NIHON DENKEI SINGAPORE
 
Fluke 725 Multifunction Calibrator
Fluke 725 Multifunction  CalibratorFluke 725 Multifunction  Calibrator
Fluke 725 Multifunction Calibrator
Dalian Zero Instrument Technology Co., Ltd China
 
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
PT. Siwali Swantika
 
Wago jumpflex catalog
Wago jumpflex catalogWago jumpflex catalog
Wago jumpflex catalog
Electromate
 
Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3AData Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
PT. Siwali Swantika
 
HIOKI Digital Multimeter DT4281 catalogue
HIOKI Digital Multimeter DT4281 catalogueHIOKI Digital Multimeter DT4281 catalogue
HIOKI Digital Multimeter DT4281 catalogue
NIHON DENKEI SINGAPORE
 
W100P WCT PH Panel type
W100P WCT PH Panel typeW100P WCT PH Panel type
W100P WCT PH Panel typeEn Campos
 

Similar to Katalog agilent-34420 a-micro-ohm-meter-tridinamika (20)

Katalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamikaKatalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamika
 
Katalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamikaKatalog agilent-digital-multimeter-L4411 a-system-tridinamika
Katalog agilent-digital-multimeter-L4411 a-system-tridinamika
 
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTHData Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
 
Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 52102A. Hubungi PT. Siwali Swantika 021-45850618
 
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
 
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
 
Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 115. Hubungi PT. Siwali Swantika 021-45850618
 
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
 
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
 
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
 
Hioki series battery_tester_denkei
Hioki series battery_tester_denkeiHioki series battery_tester_denkei
Hioki series battery_tester_denkei
 
Insulation Test Hioki 3455
Insulation Test Hioki 3455Insulation Test Hioki 3455
Insulation Test Hioki 3455
 
Hioki High End digital multimeter DT4281
Hioki High End digital multimeter DT4281Hioki High End digital multimeter DT4281
Hioki High End digital multimeter DT4281
 
Fluke 725 Multifunction Calibrator
Fluke 725 Multifunction  CalibratorFluke 725 Multifunction  Calibrator
Fluke 725 Multifunction Calibrator
 
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62
 
Wago jumpflex catalog
Wago jumpflex catalogWago jumpflex catalog
Wago jumpflex catalog
 
Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 117. Hubungi PT. Siwali Swantika 021-45850618
 
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3AData Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
 
HIOKI Digital Multimeter DT4281 catalogue
HIOKI Digital Multimeter DT4281 catalogueHIOKI Digital Multimeter DT4281 catalogue
HIOKI Digital Multimeter DT4281 catalogue
 
W100P WCT PH Panel type
W100P WCT PH Panel typeW100P WCT PH Panel type
W100P WCT PH Panel type
 

More from PT. Tridinamika Jaya Instrument

Company profile Tridinamika 2017
Company profile Tridinamika 2017Company profile Tridinamika 2017
Company profile Tridinamika 2017
PT. Tridinamika Jaya Instrument
 
Katalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamikaKatalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamikaKatalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamikaKatalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
PT. Tridinamika Jaya Instrument
 
Katalog agilent-dc-power-analyzer-n6705 b-tridinamika
Katalog agilent-dc-power-analyzer-n6705 b-tridinamikaKatalog agilent-dc-power-analyzer-n6705 b-tridinamika
Katalog agilent-dc-power-analyzer-n6705 b-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamikaKatalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog kanomax ih-solutions-tridinamika
Katalog kanomax ih-solutions-tridinamikaKatalog kanomax ih-solutions-tridinamika
Katalog kanomax ih-solutions-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamika
Katalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamikaKatalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamika
Katalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamika
Katalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamikaKatalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamika
Katalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog trotec-cm-device-moisture-measurement
Katalog trotec-cm-device-moisture-measurementKatalog trotec-cm-device-moisture-measurement
Katalog trotec-cm-device-moisture-measurement
PT. Tridinamika Jaya Instrument
 
Katalog trotec-thermal-anemometer-tridinamika
Katalog trotec-thermal-anemometer-tridinamikaKatalog trotec-thermal-anemometer-tridinamika
Katalog trotec-thermal-anemometer-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog trotec-special-infrared-thermal-imaging-camera-tridinamika
Katalog trotec-special-infrared-thermal-imaging-camera-tridinamikaKatalog trotec-special-infrared-thermal-imaging-camera-tridinamika
Katalog trotec-special-infrared-thermal-imaging-camera-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...
Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...
Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...
PT. Tridinamika Jaya Instrument
 
Katalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamika
Katalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamikaKatalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamika
Katalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamika
PT. Tridinamika Jaya Instrument
 

More from PT. Tridinamika Jaya Instrument (20)

Company profile Tridinamika 2017
Company profile Tridinamika 2017Company profile Tridinamika 2017
Company profile Tridinamika 2017
 
Katalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamikaKatalog-Katronic-ultrasonic-flow-measurement-tridinamika
Katalog-Katronic-ultrasonic-flow-measurement-tridinamika
 
Katalog-Katronic technologies brochure 2014
Katalog-Katronic technologies brochure 2014Katalog-Katronic technologies brochure 2014
Katalog-Katronic technologies brochure 2014
 
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamikaKatalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
Katalog agilent-3458 a-digital-multimeter- 8½-digit-tridinamika
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
 
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamikaKatalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
Katalog agilent-digital-multimeter-3458 a-digital-multimeter-tridinamika
 
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamikaKatalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
Katalog agilent-n6731 b-dc-power-module-05v- 010a-50w-tridinamika
 
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
Katalog agilent-dc-power-analyzer-n6715 b-base-model-custom-configured-tridin...
 
Katalog agilent-dc-power-analyzer-n6705 b-tridinamika
Katalog agilent-dc-power-analyzer-n6705 b-tridinamikaKatalog agilent-dc-power-analyzer-n6705 b-tridinamika
Katalog agilent-dc-power-analyzer-n6705 b-tridinamika
 
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs1000-series-tridinamika
 
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamikaKatalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
Katalog siglent-handheld-digital-oscilloscope-shs800-series-tridinamika
 
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamikaKatalog siglent-digital-oscilloscope-sds1000cml-tridinamika
Katalog siglent-digital-oscilloscope-sds1000cml-tridinamika
 
Katalog kanomax ih-solutions-tridinamika
Katalog kanomax ih-solutions-tridinamikaKatalog kanomax ih-solutions-tridinamika
Katalog kanomax ih-solutions-tridinamika
 
Katalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamika
Katalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamikaKatalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamika
Katalog trotec-xl63-videoprobe-system-videoendoscopy-tridinamika
 
Katalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamika
Katalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamikaKatalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamika
Katalog trotec-xlvu-xlgo-videoprobe-system-visual-inspection-tridinamika
 
Katalog trotec-cm-device-moisture-measurement
Katalog trotec-cm-device-moisture-measurementKatalog trotec-cm-device-moisture-measurement
Katalog trotec-cm-device-moisture-measurement
 
Katalog trotec-thermal-anemometer-tridinamika
Katalog trotec-thermal-anemometer-tridinamikaKatalog trotec-thermal-anemometer-tridinamika
Katalog trotec-thermal-anemometer-tridinamika
 
Katalog trotec-special-infrared-thermal-imaging-camera-tridinamika
Katalog trotec-special-infrared-thermal-imaging-camera-tridinamikaKatalog trotec-special-infrared-thermal-imaging-camera-tridinamika
Katalog trotec-special-infrared-thermal-imaging-camera-tridinamika
 
Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...
Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...
Katalog trotec-infrared-camera-surveillance-security-camera-icsm-series-tridi...
 
Katalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamika
Katalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamikaKatalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamika
Katalog trotec-infrared-camera-thermal-imager-thermography-ic-series-tridinamika
 

Recently uploaded

State of ICS and IoT Cyber Threat Landscape Report 2024 preview
State of ICS and IoT Cyber Threat Landscape Report 2024 previewState of ICS and IoT Cyber Threat Landscape Report 2024 preview
State of ICS and IoT Cyber Threat Landscape Report 2024 preview
Prayukth K V
 
UiPath Test Automation using UiPath Test Suite series, part 4
UiPath Test Automation using UiPath Test Suite series, part 4UiPath Test Automation using UiPath Test Suite series, part 4
UiPath Test Automation using UiPath Test Suite series, part 4
DianaGray10
 
Key Trends Shaping the Future of Infrastructure.pdf
Key Trends Shaping the Future of Infrastructure.pdfKey Trends Shaping the Future of Infrastructure.pdf
Key Trends Shaping the Future of Infrastructure.pdf
Cheryl Hung
 
JMeter webinar - integration with InfluxDB and Grafana
JMeter webinar - integration with InfluxDB and GrafanaJMeter webinar - integration with InfluxDB and Grafana
JMeter webinar - integration with InfluxDB and Grafana
RTTS
 
Bits & Pixels using AI for Good.........
Bits & Pixels using AI for Good.........Bits & Pixels using AI for Good.........
Bits & Pixels using AI for Good.........
Alison B. Lowndes
 
Search and Society: Reimagining Information Access for Radical Futures
Search and Society: Reimagining Information Access for Radical FuturesSearch and Society: Reimagining Information Access for Radical Futures
Search and Society: Reimagining Information Access for Radical Futures
Bhaskar Mitra
 
De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...
De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...
De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...
Product School
 
UiPath Test Automation using UiPath Test Suite series, part 3
UiPath Test Automation using UiPath Test Suite series, part 3UiPath Test Automation using UiPath Test Suite series, part 3
UiPath Test Automation using UiPath Test Suite series, part 3
DianaGray10
 
The Art of the Pitch: WordPress Relationships and Sales
The Art of the Pitch: WordPress Relationships and SalesThe Art of the Pitch: WordPress Relationships and Sales
The Art of the Pitch: WordPress Relationships and Sales
Laura Byrne
 
To Graph or Not to Graph Knowledge Graph Architectures and LLMs
To Graph or Not to Graph Knowledge Graph Architectures and LLMsTo Graph or Not to Graph Knowledge Graph Architectures and LLMs
To Graph or Not to Graph Knowledge Graph Architectures and LLMs
Paul Groth
 
FIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdf
FIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdfFIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdf
FIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdf
FIDO Alliance
 
Neuro-symbolic is not enough, we need neuro-*semantic*
Neuro-symbolic is not enough, we need neuro-*semantic*Neuro-symbolic is not enough, we need neuro-*semantic*
Neuro-symbolic is not enough, we need neuro-*semantic*
Frank van Harmelen
 
DevOps and Testing slides at DASA Connect
DevOps and Testing slides at DASA ConnectDevOps and Testing slides at DASA Connect
DevOps and Testing slides at DASA Connect
Kari Kakkonen
 
Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...
Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...
Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...
UiPathCommunity
 
Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...
Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...
Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...
Jeffrey Haguewood
 
Smart TV Buyer Insights Survey 2024 by 91mobiles.pdf
Smart TV Buyer Insights Survey 2024 by 91mobiles.pdfSmart TV Buyer Insights Survey 2024 by 91mobiles.pdf
Smart TV Buyer Insights Survey 2024 by 91mobiles.pdf
91mobiles
 
ODC, Data Fabric and Architecture User Group
ODC, Data Fabric and Architecture User GroupODC, Data Fabric and Architecture User Group
ODC, Data Fabric and Architecture User Group
CatarinaPereira64715
 
Mission to Decommission: Importance of Decommissioning Products to Increase E...
Mission to Decommission: Importance of Decommissioning Products to Increase E...Mission to Decommission: Importance of Decommissioning Products to Increase E...
Mission to Decommission: Importance of Decommissioning Products to Increase E...
Product School
 
How world-class product teams are winning in the AI era by CEO and Founder, P...
How world-class product teams are winning in the AI era by CEO and Founder, P...How world-class product teams are winning in the AI era by CEO and Founder, P...
How world-class product teams are winning in the AI era by CEO and Founder, P...
Product School
 
Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...
Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...
Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...
Ramesh Iyer
 

Recently uploaded (20)

State of ICS and IoT Cyber Threat Landscape Report 2024 preview
State of ICS and IoT Cyber Threat Landscape Report 2024 previewState of ICS and IoT Cyber Threat Landscape Report 2024 preview
State of ICS and IoT Cyber Threat Landscape Report 2024 preview
 
UiPath Test Automation using UiPath Test Suite series, part 4
UiPath Test Automation using UiPath Test Suite series, part 4UiPath Test Automation using UiPath Test Suite series, part 4
UiPath Test Automation using UiPath Test Suite series, part 4
 
Key Trends Shaping the Future of Infrastructure.pdf
Key Trends Shaping the Future of Infrastructure.pdfKey Trends Shaping the Future of Infrastructure.pdf
Key Trends Shaping the Future of Infrastructure.pdf
 
JMeter webinar - integration with InfluxDB and Grafana
JMeter webinar - integration with InfluxDB and GrafanaJMeter webinar - integration with InfluxDB and Grafana
JMeter webinar - integration with InfluxDB and Grafana
 
Bits & Pixels using AI for Good.........
Bits & Pixels using AI for Good.........Bits & Pixels using AI for Good.........
Bits & Pixels using AI for Good.........
 
Search and Society: Reimagining Information Access for Radical Futures
Search and Society: Reimagining Information Access for Radical FuturesSearch and Society: Reimagining Information Access for Radical Futures
Search and Society: Reimagining Information Access for Radical Futures
 
De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...
De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...
De-mystifying Zero to One: Design Informed Techniques for Greenfield Innovati...
 
UiPath Test Automation using UiPath Test Suite series, part 3
UiPath Test Automation using UiPath Test Suite series, part 3UiPath Test Automation using UiPath Test Suite series, part 3
UiPath Test Automation using UiPath Test Suite series, part 3
 
The Art of the Pitch: WordPress Relationships and Sales
The Art of the Pitch: WordPress Relationships and SalesThe Art of the Pitch: WordPress Relationships and Sales
The Art of the Pitch: WordPress Relationships and Sales
 
To Graph or Not to Graph Knowledge Graph Architectures and LLMs
To Graph or Not to Graph Knowledge Graph Architectures and LLMsTo Graph or Not to Graph Knowledge Graph Architectures and LLMs
To Graph or Not to Graph Knowledge Graph Architectures and LLMs
 
FIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdf
FIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdfFIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdf
FIDO Alliance Osaka Seminar: The WebAuthn API and Discoverable Credentials.pdf
 
Neuro-symbolic is not enough, we need neuro-*semantic*
Neuro-symbolic is not enough, we need neuro-*semantic*Neuro-symbolic is not enough, we need neuro-*semantic*
Neuro-symbolic is not enough, we need neuro-*semantic*
 
DevOps and Testing slides at DASA Connect
DevOps and Testing slides at DASA ConnectDevOps and Testing slides at DASA Connect
DevOps and Testing slides at DASA Connect
 
Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...
Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...
Dev Dives: Train smarter, not harder – active learning and UiPath LLMs for do...
 
Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...
Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...
Slack (or Teams) Automation for Bonterra Impact Management (fka Social Soluti...
 
Smart TV Buyer Insights Survey 2024 by 91mobiles.pdf
Smart TV Buyer Insights Survey 2024 by 91mobiles.pdfSmart TV Buyer Insights Survey 2024 by 91mobiles.pdf
Smart TV Buyer Insights Survey 2024 by 91mobiles.pdf
 
ODC, Data Fabric and Architecture User Group
ODC, Data Fabric and Architecture User GroupODC, Data Fabric and Architecture User Group
ODC, Data Fabric and Architecture User Group
 
Mission to Decommission: Importance of Decommissioning Products to Increase E...
Mission to Decommission: Importance of Decommissioning Products to Increase E...Mission to Decommission: Importance of Decommissioning Products to Increase E...
Mission to Decommission: Importance of Decommissioning Products to Increase E...
 
How world-class product teams are winning in the AI era by CEO and Founder, P...
How world-class product teams are winning in the AI era by CEO and Founder, P...How world-class product teams are winning in the AI era by CEO and Founder, P...
How world-class product teams are winning in the AI era by CEO and Founder, P...
 
Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...
Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...
Builder.ai Founder Sachin Dev Duggal's Strategic Approach to Create an Innova...
 

Katalog agilent-34420 a-micro-ohm-meter-tridinamika

  • 1. • 7½ digits resolution • 100 pV, 100 nΩ sensitivity • 1.3 nVrms, 8 nVpp noise performance • Built-in low noise 2 channel scanner • Direct SPRT, RTD, Thermistor, and Thermocouple measurements Nanovolt Performance at a Microvolt Price The Agilent Technologies 34420A nanoVolt/micro-Ohm meter is a high-sensitivity multimeter optimized for performing low-level measurements. It combines low-noise voltage measurements with resistance and temperature functions, setting a new standard in low-level flexibil-ity and performance. Take the Uncertainty Out of Your Low-Level Measurements Low-noise input amplifiers and a highly tuned input protection scheme bring read-ing noise down to 8 nVpp. Combine this with 71/2 digits of resolution, selectable analog and digital filtering, 2 ppm basic 24-hour dcV accuracy, and a shielded, cop-per pin connector and you’ve got accurate, repeatable measurements you can count on. Agilent 34420A NanoVolt/Micro-Ohm Meter Data Sheet Two Input Channels An integral two-channel programmable scan-ner simplifies voltage comparisons. Built-in ratio and difference functions enable auto mated two channel measurements without the need for an external nanoVolt scanner. Both channels share the same low noise specifications to ensure accurate comparisons. Built-In Resistance and Temperature The 34420A combines its low-noise nano- Volt input circuits with a high-stability current source to provide precise low-level resistance measurements – no more has-sling with the cost and complexity of an external current source. Three resistance modes are included: • Standard • Low-power • Voltage-limited for dry-circuit testing Offset compensation is also provided to minimize thermal EMFs and associated errors. SPRT Measurements Built-in ITS-90 conversion routines accept the calibration coefficients from your SPRT probe for direct temperature mea-surement and conversion. Thermocouples, thermistors, and RTDs are also supported. Unequaled Versatility The 34420A gives you the versatility to tackle your most challenging tasks, both on the benchtop and in your automated system. Standard features include RS-232 and GPIB interfaces, SCPI and Keithley 181 program ming language, 1024-reading memory, scaling and statistics, and a chart recorder analog output. Agilent IntuiLink:Easy Data Access The included Agilent IntuiLink software allows your captured data to be put to work easily, using PC applica-tions such as Microsoft Excel® or Word® to analyze, interpret, display, print, and doc-ument the data you get from the 34420A. You can specify the meter setup and take a single reading or log data to the Excel spreadsheet in specified time intervals. To find out more about IntuiLink visit www.agilent.com/find/intuilink Quality You Can Count On The 34420A gives you the quality and reliability you expect from Agilent Technologies. From the product’s proven >150,000 hour Mean Time Between Failure, to its standard 1-year warranty, Agilent stands behind you to bring a new level of confidence to your low-level measure-ments.
  • 2. Specifications Accuracy Specifications ±(% of reading + % of range) 1 Function Range2 Test Current 24 Hour 90 Day 1 Year Temperature Maximum 2 23 °C ± 1 °C 23 °C ± 5 °C 23 °C ± 5 °C Coefficient per Lead 0 °C—18 °C Resistance 28 °C—55 °C dc Voltage 1.0000000 mV3 0.0025 + .0020 0.0040 + .0020 0.0050 + .0020 0.0004 + .0001 10.000000 mV3 0.0025 + .0002 0.0040 + .0002 0.0050 + .0003 0.0004 + .0001 100.00000 mV 0.0015 + .0003 0.0030 + .0004 0.0040 + .0004 0.0004 + .00006 1.0000000 V 0.0010 + .0003 0.0025 + .0004 0.0035 + .0004 0.0004 + .00004 10.000000 V 0.0002 + .0001 0.0020 + .0004 0.0030 + .0004 0.0001 + .00002 100.00000 V4 0.0010 + .0004 0.0025 + .0005 0.0035 + .0005 0.0004 + .00005 Resistance 5 1.0000000 Ω 10 mA 0.0015 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω 10.000000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 Ω 100.00000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 10 Ω 1.0000000 KΩ 1 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 100 Ω 10.000000 KΩ 100 μA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 KΩ 100.00000 KΩ 10 μA 0.0015 + .0003 0.0040 + .0004 0.0060 + .0004 0.0005 + .00002 1 KΩ 1.0000000 MΩ 5 μA 0.0020 + .0003 0.0050 + .0004 0.0070 + .0004 0.0006 + .00003 1 KΩ Low Power 1.0000000 Ω 10 mA 0.0015 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω Resistance 5 10.000000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 Ω 100.00000 Ω 1 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 10 Ω 1.0000000 KΩ 100 μA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 100 Ω 10.000000 KΩ 10 μA 0.0015 + .0004 0.0040 + .0004 0.0060 + .0004 0.0005 + .00001 1 KΩ 100.00000 KΩ 5 μA 0.0015 + .0012 0.0040 + .0015 0.0060 + .0015 0.0005 + .00003 1 KΩ 1.0000000 MΩ 5 μA 0.0020 + .0003 0.0050 + .0004 0.0070 + .0004 0.0006 + .00003 1 KΩ Voltage Limited 10.000000 Ω 1 mA 0.0020 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω Resistance 5, 6 100.00000 Ω 100 μA 0.0025 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 5 Ω Channel 1 / Channel 2 (dcV Ratio) Ratio Error in % = Channel 1 accuracy in % + Channel 2 accuracy in % Channel 1-Channel 2 (dcV Difference) Difference Error = Channel 1 (% of reading + % of range) + Channel 2 (% of reading + % of range) Temperature (resolution = 0.001 °C) SPRT 7 SPRT Probe Accuracy + 0.003°C RTD RTD Probe Accuracy + 0.05°C Thermistor Thermistor Probe Accuracy + 0.1°C Thermocouple 8 Thermocouple Probe Accuracy + 0.2°C DC Voltage Noise 9 Observation Period Range 2-Minute 2-Minute 24-Hour RMS Peak-Peak Peak-Peak Noise Noise Noise 1 mV 1.3 nVrms 8 nVpp 12 nVpp 10 mV 1.5 nVrms 10 nVpp 14 nVpp 100 mV 10 nVrms 65 nVpp 80 nVpp 1 V 100 nVrms 650 nVpp 800 nVpp 10 V 450 nVrms 3 μVpp 3.7 μVpp 100 V 11 μVrms 75 μVpp 90 μVpp DC Voltage Noise vs Source Resistance 10 Source Analog Digital Resistance Noise Filter Filter 0 Ω 1.3 nVrms Off Med 100 Ω 1.7 nVrms Off Med 1k Ω 4 nVrms Off Med 10k Ω 13 nVrms Off Med 100k Ω 41 nVrms On Med 1M Ω 90 nVrms On Slow 1 Specifications are for Channel 1 or Channel 2, after 2-hour warm-up, resolution at 7.5 digits (100 NPLC), with FILTERS off. RESISTANCE specifications are for 4-wire Ohms or 2-wire ohms using Null. Without Null, add 0.2 Ohms additional error in 2-wire Ohms function. For Analog Filter ON, add 0.002% of reading. 2 20% overrange on all ranges except 5% on Voltage Limited Resistance. 3 After using Math Null. If Null is not used add 100 nanoVolts. 4 Channel 1 only. 5 Channel 1 only. Resistance measurements, for NPLC <1, add 160 μΩ rms noise. 6 Voltage limit can be set to 20 mV (default), 100 mV, or 500 mV. Measured resistance plus Channel 1 HI and LO lead resistance is limited to 10.5 Ω on the 10 Ω range and 105 Ω on the 100 Ω range. 7 For 25 Ω SPRT with triple-point of water check within the last 4 hours. Without the triple-point of water check, add 0.013°C for 24 hours, add 0.035°C for 90 day, and add 0.055°C for 1 year specifications. 8 For fixed reference junction. Add 0.3ûC for external reference junction, add 2.0ûC for internal reference junction. 9 After a 2-hour warm-up, ± 1ûC, 6.5 digits (10 PLC) with Analog Filter Off and Digital Filter Medium (50 reading average). 2-minute rms and 24-hour noise typical. For measurements using 0.02 or 0.2 NPLC, add 800 nV rms noise. 10 Typical noise behavior for Ch 1 or Ch 2, after 2 hour warm-up, 6.5 digits (10 PLC), 2 minute observation period on 1 mV range. For peak-to-peak noise, mul-tiply rms noise by 6.
  • 3. 3 Measurement Characteristics DC Voltage Measurement Method: Continuously integrating multi-slope III A-D Converter A-D Linearity: 0.00008% of reading + 0.00005% of range Input Resistance: 100V (Ch1 only): 10 MΩ +- 1% 1mV through 10V: > 10 GΩ, in parallel with < 3.6 nF Input Bias Current: <50 pA at 25 °C Injected Current: <50 nA pp at 50 or 60 Hz Input Protection: 150 V peak any input terminal to Channel 1 LO, continuous Channel-to-channel switching error (typical): 3 nV Channel Isolation: Isolation between input channels >1010 Ω Earth Isolation: 350 V peak any input terminal to earth. Impedance from any input terminal to earth is >10 GΩ and <400 pF Maximum Voltage: Channel 1 LO to Channel 2 LO, 150V peak Resistance Measurement Method: Selectable 4-wire or 2-wire ohms. Current Source referenced to Channel 1 LO input Offset Compensation: Used on all ranges except 100 kΩ and 1 MΩ. Can be turned off if desired Protection: 150 V peak Open Circuit Voltage: For Resistance and Low Power Resistance <14 V. 20 mV, 100 mV, 500 mV selectable clamp Temperature SPRT: ITS-90 calibrated temperature with the range of -190°C to +660°C Thermocouple: ITS-90 conversions of Type B, E, J, K, N, R, S, T Thermistor: 5 kΩ RTD: Type α =.00385 and α =.00392. R0 from 4.9 Ω to 2.1 kΩ. ITS -90 (IEC-751) Callendar Van Dusen conversion. Measurement Noise Rejection 60 (50) Hz1 dc CMRR: 140 dB ac CMRR: 70 dB Integration Time Normal Mode Rejection2 200 plc/3.335 s (4 s) 110 dB3 100 plc/1.675 s (2 s) 105 dB3 20 plc/334 ms (400 ms) 100 dB3 10 plc/167 ms (200 ms) 95 dB3 2 plc/33.3 ms (40 ms) 90 dB 1 plc/16.7 ms (20 ms) 60 dB <1 plc 0 Operating Characteristics4 Function Digits Integration Time Readings/s 5 dcV 71/2 200 plc .15 (.125) Thermocouple 71/2 100 plc .3 (.25) 61/2 20 plc 1.5 (1.25) 61/2 10 plc 3 (2.5) 51/2 1 plc 25 (20.8) 51/2 0.2 plc 100 (100) 41/2 0.02 plc 250 (250) Resistance 71/2 200 plc .075 (.062) dcV1/DCV2 71/2 100 plc .15 (.125) dcV 1-2 61/2 20 plc .75 (.625) RTD 61/2 10 plc 1.5 (1.25) Thermistor 51/2 1 plc 12.5 (10.4) 0.2 plc 50 (50) 41/2 0.02 plc 125 (125) System Speeds6 Configuration Rates: 26/s to 50/s Autorange Rate (Volts): >30/s ASCII reading to RS-232: 55/s ASCII reading to GPIB: 250/s Max. Internal Trigger Rate: 250/s Max. Ext. Trig. Rate to Memory: 250/s Triggering and Memory Reading HOLD Sensitivity: 10%, 1%, 0.1%, or 0.01% of range Samples/Trigger: 1 to 50,000 Trigger Delay: 0 to 3600 s; 10 μs step size External Trigger Delay: <1 ms External Trigger Jitter: <500 μs Memory: 1024 readings Math Functions NULL (Channel 1 dcV, Channel 2 dcV, Difference, Resistance, Temperature) STATS (Min, Max, Average, Peak-Peak, Standard Deviation, Number of readings) SCALE (Allows linear scaling as y = mx+b) CHART NULL (Establishes zero for rear panel output) Filter (Analog or Digital or Both) Analog: Low pass 2 pole @ 13Hz, available for dcV on 1 mV, 10 mV, 100 mV range Digital: Moving average filter, 10 (fast), 50 (medium), or 100 (slow) reading averages. Chart Out (Analog Out) Maximum output: ± 3V Resolution: 16 bits Accuracy: ± 0.1% of output + 1 mV Output Resistance: 1 kΩ ± 5% Update rate: once per reading Span and Offset: Adjustable Standard Programming Languages SCPI (IEEE 488.2), Keithley 181 Accessories Included 4 ft low thermal cable with copper spade lugs, Kelvin clip set, 4-wire shorting plug, user’s man-ual, service manual, test report, contact cleaner, and power cord. General Specifications Front Panel Connection: Shielded, low thermal, 99% copper contacts. Power Supply: 100V/120V/220V(230V)/240V +- 10%. Power Line Frequency: 45 Hz to 66 Hz and 360 Hz to 440 Hz. Automatically sensed at power-on. Power Consumption: 25VA peak (10W average). Operating Environment: Full accuracy for 0 °C to 55 °C. Full accuracy to 80% R.H. up to 30 °C. Storage Environment: -40 °C to 75 °C. Size: 254.4 mm W x 374.0 mm L x 103.6 mm H (10.02” W x 14.72” L x 4.08” H) Weight: 3 kg (6.5 lbs). Safety: Designed to CSA, UL-1244, IEC-1010. RFI and ESD: CISPR 11. 1 For 1 kΩ unbalanced in LO lead. 2 For power line frequency ± - 0.1%, Filters OFF. For Digital Filter slow add 20 db, for medium or fast add 10 db for NPLC ³ 1. 3 For power line frequency ± - 1%, use 80 db, for ± - 3% use 60 db. 4 Speeds are for delay 0, Display OFF, Filters OFF, Offset Compensation OFF. 5 Reading speeds for 60 Hz or (50 Hz), 100 mV through 100 V ranges. 1 mV range 30/s MAX, 10 mV range 170/s MAX, thermocouple 120/s MAX. 6 Speeds are for NPLC 0.02, Delay 0, Display OFF, Chart Out OFF.
  • 4. Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consor-tium. Agilent Channel Partners www.agilent.com/find/channelpartners Get the best of both worlds: Agilent’s measurement expertise and product breadth, combined with channel partner convenience. www.agilent.com For more information on Agilent Tech-nologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Americas Canada (877) 894 4414 Brazil (11) 4197 3500 Mexico 01800 5064 800 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 375 8100 Europe & Middle East Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 €/minute Germany 49 (0) 7031 464 6333 Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 United Kingdom 44 (0) 131 452 0200 For other unlisted countries: www.agilent.com/find/contactus Revised: June 8, 2011 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 2011 Published in USA, October 11, 2011 5968-0161EN Agilent Advantage Services is committed to your success throughout your equip-ment’s lifetime. To keep you competitive, we continually invest in tools and processes that speed up calibration and repair and reduce your cost of ownership. You can also use Infoline Web Services to manage equipment and services more effectively. By sharing our measurement and service expertise, we help you create the products that change our world. www.agilent.com/find/advantageservices www.agilent.com/quality Ordering Information Includes low thermal input cable (34102A), low thermal shorting plug (34103A), Kelvin clip set (11062A), cali-bration certificate, power cord. Also includes CD with: IntuiLink software, IVI and VXI PnP drivers, user's guide, service guide, and data sheet. Options 34420A-ABA English localization 34420A-ABD German localization: translated operating manual 34420A-ABF French localization: translated operating manual 34420A-ABJ Japanese localization: translated operating manual 34420A-A6J ANSI Z540 compliant calibration Accessories 34102A Low-thermal input cable (four-conductor) with copper spade lugs 34103A Low-thermal shorting plug 34104A Low-thermal input connector 34131A Transit Case 34161A Accessory pouch 34190A Rackmount Kit: designed for use with only one instrument, mounted on either the left or the right side of the rack. 34191A 2U Dual Flange Kit: secures the instrument to the front of the rack. This kit can be used with the 34194A Dual Lock Link Kit to mount two half-width, 2U height instruments side-by-side. 34194A Dual Lock Link Kit: recommended for side-by-side combinations and includes links for instruments of differ-ent depths. This kit can be used with the 34191A 2U Dual Flange Kit to mount two half-width, 2U height instruments side-by-side.