SlideShare a Scribd company logo
3-447-012-03
6/1.19
GMC-I Messtechnik GmbH
METRALINE DM 61/62
Analog-Digital Multimeter
Features
Automatic Blocking Sockets (ABS) *
Automatic blocking sockets prevent incorrect connection of
measurement cables and inadvertent selection of the wrong
measured quantity. This significantly reduces danger to the user,
the instrument and the system under test, and eliminates it
entirely in many cases.
Automatic / Manual Measuring Range Selection
Measured quantities are selected with the rotary switch. The
measuring range is automatically matched to measured values.
The measuring range can be selected manually as well with the
help of the AUTO/MAN key.
Display of Negative Values at the Analog Scale
Negative values are also displayed at the analog scale for zero-
frequency quantities, allowing for observation of measured
quantity fluctuation around the zero-point.
Storage of Measured Values
By pressing the HOLD/MIN/MAX key, the currently displayed
measurement value can be „frozen“ in the display.
The minimum and maximum values which were present at the
input of the measuring instrument after activation of the MIN/MAX
mode can be selectively "retained" with the MIN/ MAX function.
The most important application is the determination of the
minimum or maximum value during long-term observation of
measurement quantities. MIN/MAX has no effect on the analog
display; it continues to display the current measurement value.
Continuity Test
Allows for the detection of short-circuits and interrupted
conductors. In addition to displaying test results, an acoustic
signal can also be generated if desired.
Power Saving Circuit
The device is switched off automatically if the measured value
remains unchanged for a period of approximately 15 minutes, and
if none of the controls are activated during this time. Automatic
shutdown can be deactivated.
Protective Cover for Harsh Conditions
The instrument is protected against damage in the event of
impacts or dropping by means of a soft rubber cover with tilt
stand. The rubber material also assures that the instrument does
not wander if it is set up on a vibrating surface.
Duty Cycle Measurement – Square-Wave Signals
This function makes it possible to test circuits and transmission
cables by measuring the frequency and the duty cycle of pulses.
Voluntary Manufacturer’s Warranty
36 months for material and workmanship
* Patented (patent no. EP 1801 598, US 7,439,725)
• Voltage: DC / AC 100 μV ... 1000 V
• Current: DC / AC: 10 μA ... 660.0 mA (DM 61) / 10.00A (DM 62)
• Clip function 1000:1 for current transformers (DM 61 only)
• Resistance: 100 mΩ... 60.00 MΩ
• Capacitance: 1 pF ... 40.00 mF (DM 62 only)
• Frequency: 10.00 Hz ... 10.00 MHz (DM 62 only)
• Diode / Continuity
• Duty cycle (%) measurement (DM 62 only)
• Temperature TC with K-type: –50 ... 1300 °C
• TRMS bandwidth: 2 kHz (DM 62 only)
• Hold / Peak / Min-Max / Relative (Zero)
• Auto / Manual ranging
• Dual digital display with analog scale and backlight
• ABS Automatic Blocking Sockets
• UL Certification
• 3 year warranty
2 GMC-I Messtechnik GmbH
METRALINE DM 61/62
Analog-Digital Multimeter
Characteristic Values
1)
At 0 °C ... + 40 °C
2)
At input > 3.5 Vrms, typical 5 Vp-p, square wave, bipolar inputs
3) For < 10 kHz at 5 Vp-p, square wave, bipolar inputs
4)
Without sensor
5)
Display with current transformers 1000 : 1
6) Limited by 10 A fuse
Influencing Quantities and Influence Error
1)
With temperature: Error data apply per 10 K change in temperature. With frequen-
cy: Error data apply to a display from 300 digits onwards.
2)
With unknown waveform (crest factor CF > 2), measure with manual range selection
3)
With the exception of sinusoidal waveform.
4) After the „ “ symbol is displayed.
Display
Liquid crystal display (58 mm x 31.4 mm) with analog indication
and digital display and with display of the unit of measured
quantity, function and various special functions.
Analog
Indication LCD scale with bar graph
Scale length 55 mm
Scaling 65 scale divisions during all the
measurement
Polarity indication With automatic reversal
Overrange indication By triangle
Sampling rate 28 times/s
Meas.
Func-
tion
Measuring
Range
DM61 DM62
(TRMS)
Resolu-
tion
Input
Impedance
Digital dis-
play Inherent
deviation at
reference
condition
+(...%rdg
+...digits)
Overload
capacity 1)
Over-
load
values
Over-
load
dura-
tion
V(DC)
660.0 mV • • 100 μV >100 MΩ //
<40pF
0.7 + 5
1000V
DC
AC
eff/rms
Sine
wave
Cont.
6.600 V • • 1 mV 11 MΩ //<40pF 0.4 + 5
66.00 V • • 10 mV 10 MΩ // <40pF 0.4 + 5
660.0 V • • 100 mV 10 MΩ // <40pF 0.4 + 5
1000 V • • 1 V 10 MΩ // <40pF 0.4 + 5
V(AC)
660.0 mV • • 100 μV >100 MΩ //
<40pF 1.2 + 5
6.600 V • • 1 mV 11 MΩ //<40pF
1.0 + 3
66.00 V • • 10 mV 10 MΩ // <40pF
660.0 V • • 100 mV 10 MΩ // <40pF
1000 V • • 1 V 10 MΩ // <40pF
A(DC)
Voltage Drop
66.00 mA • • 10 μA 66.00 mV 0.8 + 5
0.7 A Cont.
660.0 mA • • 100 μA 66.00 mV 0.8 + 5
10.00 A 6)
— • 10 mA 10.00 mV 1.5 + 5 — —
A(AC)
66.00 mA • • 10 μA 66.00 mV 0.8 + 5
0.7 A Cont.
660.0 mA • • 100 μA 66.00 mV 0.8 + 5
10.00 A 6)
— • 10 mA 10.00 mV 1.5 + 5 — —
(AC) 5)
66.00 A • — 10 mA 66.00 mV 0.8 + 5
0.7 A Cont.
660.0 A • — 100 mA 66.00 mV 0.8 + 5
Ω
No load Voltage
660.0 Ω • • 100 mΩ –3.3 V 0.8 + 5
1000 V
DC
AC
eff/rms
Sine
wave
max.
10 s
6.600 kΩ • • 1 Ω –1.08 V 0.8 + 5
66.00 kΩ • • 10 Ω –1.08 V 0.8 + 5
660.0 kΩ • • 100 Ω –1.08 V 0.8 + 5
6.600 MΩ • • 1 kΩ –1.08 V 1.0 + 5
66.00 MΩ • • 10 kΩ –1.08 V 2.0 + 5
660.0 Ω • • 100 mΩ –3.3 V 0.8 + 5
1000V
DC
AC
eff/rms
Sine
wave
max.
10 s
DIODE 2.000 V • • 1 mV 3.3 V 2.0 + 10
F
6.600 nF — • 1 pF
—
3.0 + 40
66.00 nF — • 10 pF 2.0 + 10
660.0 nF — • 100 pF 2.0 + 10
6.600 μF — • 1 nF 2.0 + 10
66.00 μF — • 10 nF 2.0 + 10
660.0 μF — • 100 nF 5.0 + 10
6.600 mF — • 1 μF 5.0 + 10
40.00 mF — • 10 μF 5.0 + 10
Hz
f min
66.00 Hz — • 0.01 Hz
10 Hz 0.2 + 2 2)
660.0 Hz — • 0.1 Hz
6.600 kHz — • 1 Hz
66.00 kHz — • 10 Hz
660.0 kHz — • 100 Hz
6.600 MHz — • 1 kHz
10.00 MHz — • 10 kHz
%
1.0 ... 98.90% — • 0.01 %
0.9% (% min)
10 Hz...1 kHz
±5 Digit 3)
1...10 kHz;
±5 Digit/kHz
°C/°F
0 ... 1300 °C • • 1 °C — 2.0 + 3 4)
–50 ... 0 °C • • 1 °C — 2.0 ±10 4)
Influencing
Quantity Range of Influence
Measured Quantity/
Measuring Range
Influence Error1)
±(... % of rdg. + ... digits)
Temperature
0 °C ... +21 °C
and
+25 °C ... +40 °C
V DC, V AC
1 x Intrinsic uncertainty/K
A DC, A AC
Ω
Diode
F, Hz, %, °C
Measured
Quantity
Frequency
20 Hz ... < 50 Hz
660 mV~
1.0 + 3
> 50 Hz ... 200 Hz 5.0 + 3
20 Hz ... < 50 Hz
6.6 ... 1000 V~
1.0 + 3
> 50 Hz ... 2 kHz 5.0 + 7
> 50 Hz ... 200 Hz
A~
1.0 + 3
20 Hz ... < 2 kHz 5.0 + 3
Crest
Factor CF
1 ... 1.4
V~3)
, A~3) ±1% of rdg
1.4 ... 5 2) ±5% of rdg
Battery
Voltage
4) ... < 2.49
V
> 2.49 V ... 3 V
V DC 5 Digit
V~, A DC 10 Digit
A AC 6 Digit
660 Ω 4 Digit
6.600 kΩ ...
66.00 MΩ
3 Digit
nF, F, mF, Hz, % 5 Digit
Relative
Humidity
75%
3 days
Meter off
V~, V DC
A~, A DC
Ω
F
Hz
°C
%
1 x intrinsic uncertainty
Influencing
Quantity
Range of Influence
Measuring
Range
Attenuation
Common Mode
Interference
Voltage
Noise quantity max. 1000 V
V > 100 dB
V > 100 dB
Noise quantity max. 1000 V
50 Hz, 60 Hz sinusoidal
V > 100 dB
V > 50 dB
Normal Mode
Interference
Voltage
Noise quantity: V ,
value of the measuring range at a time
max. 1000 V , 50 Hz, 60 Hz sinusoidal
660 mV, 6.6 V,
660 V, 1000 V
DC
> 43 dB
66 V DC > 35 dB
Noise quantity max. 1000 V V > 45 dB
GMC-I Messtechnik GmbH 3
METRALINE DM 61/62
Analog-Digital Multimeter
Digital
Height of Main
Display numerals 7 segment numerals: 12 mm
Height of Sub
Display numerals 7 segment numerals: 7 mm
Number of counts 4 digit: 6600 steps
Overrange display „OL“ is shown
Polarity display „–“ sign is shown, When positive pole
connected to „ “
Sampling rate 2.8 times/s
Power supply
Battery 2 AA size batteries alkaline manganese
cells as per IEC LR6.
Service life for METRALINE DM 61:
600 hrs. for V DC, A DC
300 hrs. for V AC, A AC
for METRALINE DM 62:
400 hrs. for V DC, A DC
200 hrs. for V AC, A AC
Battery test Automatic display of „ “ symbol when
battery voltage falls below following value:
approx. 2.4 V.
Electromagnetic compatibility (EMC)
Emission EN 61326: 2013 Class B
Immunity IEC 61000-4-2:
8 kV atmosphere discharge
4 kV contact discharge
IEC 61000-4-3: 3 V/m
Short-term measured value deviation may occur during electromagnetic interference
thus reducing the specified operating quality.
Safety: IEC 61010-1-2010
Measuring category 600 V CAT III, 300 V CAT IV
The maximum voltage of 1000 V may only
be used with CAT II.
High Voltage Test 6.7 kV (IEC 61010-1-2010)
Fuses
Fuse for up to 660 mA ranges
FF (UR) 1.6 A/1000 V AC/DC; 6.3 mm X 32 mm; rating 10 kA
with 1000 VAC/DC and ohmic load; in conjunction with power
diodes, protects all current measuring ranges up to 660 mA.
Fuse for up to 10 A ranges (METRALINE DM 62)
FF (UR)10 A/1000 V AC/DC; 10 mm x 38 mm; rating 30 kA with
1000 VAC/DC and ohmic load; protects the 10 A ranges up to
1000 V AC/DC.
Defective fuses are not displayed.
Response Time (after manual range selection)
Reference conditions
Ambient temperature 23 °C+ 2 K
Relative humidity 45% ... 55 % RH
Frequency of
measured quantity 50 or 60 Hz ±2%
Waveform of the
measured quantity sinusoidal
Battery voltage 3 V ±0.1 V
Environmental conditions
Functional
temperature range 0 °C ... +50 °C
Storage
temperature range –25 °C ... +70 °C (without batteries)
Relative humidity 45 ... 75 %
Altitude up to 2000 m
Mechanical configuration
Protection
for the meter IP50
Pollution degree 2
Connection sockets IP20 according to
EN 60529 / DIN VDE 0470-1
Dimensions with holster: 86 mm x 188 mm x 53 mm
without holster: 79 mm x 174 mm x 38 mm
Weight 480 g approx., including battery and holster
Applicable Regulations and Standards
Measured Quantity/
Measuring Range
Response Time Transient response for
step function of the
measured quantity
Analog
Display
Digital Display
V , V , °C 0.1 s 1 s from 0 to 80%
of the upper range limit
A , A 0.1 s 1 s
from 0 to 50%
of the upper range limit
660 Ω ... 6.6 MΩ 0.1 s 1 s
66 MΩ 0.2 s 2 s
0.1 s 1 s
from 0 to 80%
of the upper range limit
6.6 nF ... 66 μF 0.7 s max. 1 s
660 μF ... 6.6 mF 1.4 s max. 3 s
66 mF 7.0 s max. 15 s
660 Hz, 6.6 kHz 2.0 s max. 2 s
66 kHz, 660 kHz, 1 MHz 0.5 s max. 1 s
% (≥ 10 Hz) 0.7 s max. 2.5 s
IEC 61010-1/EN 61010-1/
VDE 0411-1
Safety requirements for electrical equipment for
measurement, control and laboratory use
EN 60529
VDE 0470, Part 1
Test instruments and test procedures
Protection provided by enclosures (IP code)
DIN EN 61326-2-1
VDE 0843-02-2-1
Electrical equipment for measurement, control and labo-
ratory use – EMC requirements –
Part 2-1: Particular requirements for sensitive test and
measurement equipment
DIN EN 60529
DIN VDE 0470 Part 1
Test Instruments and test procedures
– Degree of protection provided by enclosures (IP code)
Edited in Germany • Subject to change without notice • A pdf version is available on the Internet
Phone +49 911 8602-111
Fax +49 911 8602-777
E-Mail info@gossenmetrawatt.com
www.gossenmetrawatt.com
GMC-I Messtechnik GmbH
Südwestpark 15
90449 Nürnberg • Germany
METRALINE DM 61/62
Analog-Digital Multimeter
Standard Equipment
1 Multimeter
1 Rubber holster with carrying strap
1 Cable set
1 Battery set
1 Operating instructions
1 Test report
Order Information
For additional information on accessories, please refer to
• our „Measuring Instruments and Testers“ catalogue
• our website www.gossenmetrawatt.com
Description Type Article Number
Clipping multimeter, clip factor 1:1000
for current measurement with optional
clamp WZ1001 as accessory METRALINE DM 61 M194A
TRMS Multimeter METRALINE DM 62 M197A
Accessories
AC clamp 1000:1 WZ1001 Z194A

More Related Content

What's hot

Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100
Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100
Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100
Tenmars Việt Nam
 
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRAData Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
PT. Siwali Swantika
 
Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10
Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10
Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10
PT. Siwali Swantika
 
SOLAR WATER HEATER AND LED LANTERN By Vtech Electrocon
SOLAR WATER HEATER AND LED LANTERN By Vtech ElectroconSOLAR WATER HEATER AND LED LANTERN By Vtech Electrocon
SOLAR WATER HEATER AND LED LANTERN By Vtech Electrocon
IndiaMART InterMESH Limited
 
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Ref 615
Ref 615Ref 615
Neway inverter range
Neway inverter rangeNeway inverter range
Neway inverter range
NEWAY POWER
 
Katalog agilent-34420 a-micro-ohm-meter-tridinamika
Katalog agilent-34420 a-micro-ohm-meter-tridinamikaKatalog agilent-34420 a-micro-ohm-meter-tridinamika
Katalog agilent-34420 a-micro-ohm-meter-tridinamika
PT. Tridinamika Jaya Instrument
 
Hioki 9651 clamp_on_sensor_datasheet 
Hioki 9651 clamp_on_sensor_datasheet Hioki 9651 clamp_on_sensor_datasheet 
Hioki 9651 clamp_on_sensor_datasheet 
Angus Sankaran
 
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
PT. Tridinamika Jaya Instrument
 
189 specification- www.ttbvs.com
189 specification- www.ttbvs.com189 specification- www.ttbvs.com
189 specification- www.ttbvs.com
TTBVS
 
constant voltage dali dimmable driver
constant voltage dali dimmable driverconstant voltage dali dimmable driver
constant voltage dali dimmable driver
Shanghai Euchips Industrial Co.,Ltd.
 
Hioki 3286 20 Clamp On Power HiTester in Bangladesh
Hioki 3286 20 Clamp On Power HiTester in BangladeshHioki 3286 20 Clamp On Power HiTester in Bangladesh
Hioki 3286 20 Clamp On Power HiTester in Bangladesh
ROnex International
 
Hioki series battery_tester_denkei
Hioki series battery_tester_denkeiHioki series battery_tester_denkei
Hioki series battery_tester_denkei
NIHON DENKEI SINGAPORE
 
04 specification- www.ttbvs.com
04 specification- www.ttbvs.com04 specification- www.ttbvs.com
04 specification- www.ttbvs.com
TTBVS
 

What's hot (18)

Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100
Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100
Hướng dẫn sử dụng máy đo tốc độ vòng quay Tenmars TM-4100
 
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRAData Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
 
Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10
Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10
Data Teknis Gossen Metrawatt Insulation Tester : METRISO PRIME 10
 
SOLAR WATER HEATER AND LED LANTERN By Vtech Electrocon
SOLAR WATER HEATER AND LED LANTERN By Vtech ElectroconSOLAR WATER HEATER AND LED LANTERN By Vtech Electrocon
SOLAR WATER HEATER AND LED LANTERN By Vtech Electrocon
 
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
 
Ref 615
Ref 615Ref 615
Ref 615
 
Neway inverter range
Neway inverter rangeNeway inverter range
Neway inverter range
 
Katalog agilent-34420 a-micro-ohm-meter-tridinamika
Katalog agilent-34420 a-micro-ohm-meter-tridinamikaKatalog agilent-34420 a-micro-ohm-meter-tridinamika
Katalog agilent-34420 a-micro-ohm-meter-tridinamika
 
Hioki 9651 clamp_on_sensor_datasheet 
Hioki 9651 clamp_on_sensor_datasheet Hioki 9651 clamp_on_sensor_datasheet 
Hioki 9651 clamp_on_sensor_datasheet 
 
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1272a-series-handheld-digital-multimeter-tridinamika
 
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1271a-series-handheld-digital-multimeter-tridinamika
 
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamikaKatalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
Katalog agilent-u1270-series-handheld-digital-multimeter-tridinamika
 
189 specification- www.ttbvs.com
189 specification- www.ttbvs.com189 specification- www.ttbvs.com
189 specification- www.ttbvs.com
 
constant voltage dali dimmable driver
constant voltage dali dimmable driverconstant voltage dali dimmable driver
constant voltage dali dimmable driver
 
Hioki 3286 20 Clamp On Power HiTester in Bangladesh
Hioki 3286 20 Clamp On Power HiTester in BangladeshHioki 3286 20 Clamp On Power HiTester in Bangladesh
Hioki 3286 20 Clamp On Power HiTester in Bangladesh
 
Hioki series battery_tester_denkei
Hioki series battery_tester_denkeiHioki series battery_tester_denkei
Hioki series battery_tester_denkei
 
04 specification- www.ttbvs.com
04 specification- www.ttbvs.com04 specification- www.ttbvs.com
04 specification- www.ttbvs.com
 
Part 1 - Weighing & Weights 2015-16
Part 1 - Weighing & Weights 2015-16Part 1 - Weighing & Weights 2015-16
Part 1 - Weighing & Weights 2015-16
 

Similar to Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62

Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTHData Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
PT. Siwali Swantika
 
Katalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamikaKatalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamika
PT. Tridinamika Jaya Instrument
 
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3AData Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
PT. Siwali Swantika
 
Hioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in BangladeshHioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in Bangladesh
ROnex International
 
Hioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in BangladeshHioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in Bangladesh
Ronexmeterbd
 
Kingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme Technology
Kingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme TechnologyKingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme Technology
Kingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme Technology
Angus Sankaran
 
Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
PT. Siwali Swantika
 
Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Hioki 3286-20 Single Phase Clamp On Power Hi-Tester in Bangladesh
Hioki 3286-20 Single Phase Clamp On Power Hi-Tester in BangladeshHioki 3286-20 Single Phase Clamp On Power Hi-Tester in Bangladesh
Hioki 3286-20 Single Phase Clamp On Power Hi-Tester in Bangladesh
Ronexmeterbd
 
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 
Magnetic flow meter for industrial process measurement
Magnetic flow meter for industrial process measurementMagnetic flow meter for industrial process measurement
Magnetic flow meter for industrial process measurement
CTi Controltech
 
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COILData Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
PT. Siwali Swantika
 
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUS
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUSData Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUS
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUS
PT. Siwali Swantika
 
Htc digital multimeter dm 112 r
Htc digital multimeter dm 112 rHtc digital multimeter dm 112 r
Htc digital multimeter dm 112 rMonarch Controls
 
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Beeteco
 
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
CTY TNHH HẠO PHƯƠNG
 
Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...
Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...
Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...
Angus Sankaran
 
Brochure FMI - Productos comerciales para espacio
Brochure FMI - Productos comerciales para espacioBrochure FMI - Productos comerciales para espacio
Brochure FMI - Productos comerciales para espacio
Marketing Donalba
 
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
PT. Siwali Swantika
 

Similar to Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62 (20)

Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTHData Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
 
Katalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamikaKatalog agilent-digital-multimeter-34401 a-tridinamika
Katalog agilent-digital-multimeter-34401 a-tridinamika
 
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3AData Teknis Gossen Metrawatt Multimeter : METRAport 3A
Data Teknis Gossen Metrawatt Multimeter : METRAport 3A
 
Hioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in BangladeshHioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in Bangladesh
 
Hioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in BangladeshHioki Electrical Measuring Instruments in Bangladesh
Hioki Electrical Measuring Instruments in Bangladesh
 
Kingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme Technology
Kingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme TechnologyKingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme Technology
Kingsine K3066i Protection Relay Testing - Datasheet Manual - Supreme Technology
 
Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9240. Hubungi PT. Siwali Swantika 021-45850618
 
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
 
Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MD9230. Hubungi PT. Siwali Swantika 021-45850618
 
Hioki 3286-20 Single Phase Clamp On Power Hi-Tester in Bangladesh
Hioki 3286-20 Single Phase Clamp On Power Hi-Tester in BangladeshHioki 3286-20 Single Phase Clamp On Power Hi-Tester in Bangladesh
Hioki 3286-20 Single Phase Clamp On Power Hi-Tester in Bangladesh
 
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 113. Hubungi PT. Siwali Swantika 021-45850618
 
Magnetic flow meter for industrial process measurement
Magnetic flow meter for industrial process measurementMagnetic flow meter for industrial process measurement
Magnetic flow meter for industrial process measurement
 
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COILData Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
 
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUS
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUSData Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUS
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRIME PLUS
 
Htc digital multimeter dm 112 r
Htc digital multimeter dm 112 rHtc digital multimeter dm 112 r
Htc digital multimeter dm 112 r
 
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
 
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
Bộ điều khiển nhiệt độ (Temperature Controller) Fuji Electric PXF9
 
Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...
Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...
Omicron CMC 256plus High Precision Relay Test Set and Universal Calibrator Da...
 
Brochure FMI - Productos comerciales para espacio
Brochure FMI - Productos comerciales para espacioBrochure FMI - Productos comerciales para espacio
Brochure FMI - Productos comerciales para espacio
 
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Fluke 5080A. Hubungi PT. Siwali Swantika 021-45850618
 

More from PT. Siwali Swantika

NetAlly LinkRunner 10G
NetAlly LinkRunner 10GNetAlly LinkRunner 10G
NetAlly LinkRunner 10G
PT. Siwali Swantika
 
NetAlly Etherscope nXG
NetAlly Etherscope nXGNetAlly Etherscope nXG
NetAlly Etherscope nXG
PT. Siwali Swantika
 
Fluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdfFluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdf
PT. Siwali Swantika
 
Kalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasiKalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasi
PT. Siwali Swantika
 
Kalibrasi Transmitter
Kalibrasi TransmitterKalibrasi Transmitter
Kalibrasi Transmitter
PT. Siwali Swantika
 
Kategori Pengukuran Kelistrikan
Kategori Pengukuran KelistrikanKategori Pengukuran Kelistrikan
Kategori Pengukuran Kelistrikan
PT. Siwali Swantika
 
5 Langkah Kalibrasi RTD
5 Langkah Kalibrasi RTD5 Langkah Kalibrasi RTD
5 Langkah Kalibrasi RTD
PT. Siwali Swantika
 
Keuntungan AnyWare Cloud.pdf
Keuntungan AnyWare Cloud.pdfKeuntungan AnyWare Cloud.pdf
Keuntungan AnyWare Cloud.pdf
PT. Siwali Swantika
 
3 Metode Kalibrasi Sensor Suhu
3 Metode Kalibrasi Sensor Suhu3 Metode Kalibrasi Sensor Suhu
3 Metode Kalibrasi Sensor Suhu
PT. Siwali Swantika
 
10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdf10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdf
PT. Siwali Swantika
 
Pengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdfPengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdf
PT. Siwali Swantika
 
Fluke Process Calibration Tools Catalog
Fluke Process Calibration Tools CatalogFluke Process Calibration Tools Catalog
Fluke Process Calibration Tools Catalog
PT. Siwali Swantika
 
Fluke Temperature Calibration Catalog
Fluke Temperature Calibration CatalogFluke Temperature Calibration Catalog
Fluke Temperature Calibration Catalog
PT. Siwali Swantika
 
Vitrek Catalog
Vitrek CatalogVitrek Catalog
Vitrek Catalog
PT. Siwali Swantika
 
Rotronic Water Activity Series
Rotronic Water Activity SeriesRotronic Water Activity Series
Rotronic Water Activity Series
PT. Siwali Swantika
 
Rotronic Catalog 2022
Rotronic Catalog 2022Rotronic Catalog 2022
Rotronic Catalog 2022
PT. Siwali Swantika
 
Rotronic RMS Catalog
Rotronic RMS CatalogRotronic RMS Catalog
Rotronic RMS Catalog
PT. Siwali Swantika
 
Fluke Catalog 2022
Fluke Catalog 2022Fluke Catalog 2022
Fluke Catalog 2022
PT. Siwali Swantika
 
Metrel Catalog 2022
Metrel Catalog 2022Metrel Catalog 2022
Metrel Catalog 2022
PT. Siwali Swantika
 
NetAlly Product Family Brochure
NetAlly Product Family BrochureNetAlly Product Family Brochure
NetAlly Product Family Brochure
PT. Siwali Swantika
 

More from PT. Siwali Swantika (20)

NetAlly LinkRunner 10G
NetAlly LinkRunner 10GNetAlly LinkRunner 10G
NetAlly LinkRunner 10G
 
NetAlly Etherscope nXG
NetAlly Etherscope nXGNetAlly Etherscope nXG
NetAlly Etherscope nXG
 
Fluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdfFluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdf
 
Kalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasiKalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasi
 
Kalibrasi Transmitter
Kalibrasi TransmitterKalibrasi Transmitter
Kalibrasi Transmitter
 
Kategori Pengukuran Kelistrikan
Kategori Pengukuran KelistrikanKategori Pengukuran Kelistrikan
Kategori Pengukuran Kelistrikan
 
5 Langkah Kalibrasi RTD
5 Langkah Kalibrasi RTD5 Langkah Kalibrasi RTD
5 Langkah Kalibrasi RTD
 
Keuntungan AnyWare Cloud.pdf
Keuntungan AnyWare Cloud.pdfKeuntungan AnyWare Cloud.pdf
Keuntungan AnyWare Cloud.pdf
 
3 Metode Kalibrasi Sensor Suhu
3 Metode Kalibrasi Sensor Suhu3 Metode Kalibrasi Sensor Suhu
3 Metode Kalibrasi Sensor Suhu
 
10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdf10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdf
 
Pengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdfPengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdf
 
Fluke Process Calibration Tools Catalog
Fluke Process Calibration Tools CatalogFluke Process Calibration Tools Catalog
Fluke Process Calibration Tools Catalog
 
Fluke Temperature Calibration Catalog
Fluke Temperature Calibration CatalogFluke Temperature Calibration Catalog
Fluke Temperature Calibration Catalog
 
Vitrek Catalog
Vitrek CatalogVitrek Catalog
Vitrek Catalog
 
Rotronic Water Activity Series
Rotronic Water Activity SeriesRotronic Water Activity Series
Rotronic Water Activity Series
 
Rotronic Catalog 2022
Rotronic Catalog 2022Rotronic Catalog 2022
Rotronic Catalog 2022
 
Rotronic RMS Catalog
Rotronic RMS CatalogRotronic RMS Catalog
Rotronic RMS Catalog
 
Fluke Catalog 2022
Fluke Catalog 2022Fluke Catalog 2022
Fluke Catalog 2022
 
Metrel Catalog 2022
Metrel Catalog 2022Metrel Catalog 2022
Metrel Catalog 2022
 
NetAlly Product Family Brochure
NetAlly Product Family BrochureNetAlly Product Family Brochure
NetAlly Product Family Brochure
 

Recently uploaded

HYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generationHYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generation
Robbie Edward Sayers
 
Water Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdfWater Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation & Control
 
RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...
RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...
RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...
thanhdowork
 
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
MdTanvirMahtab2
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
gerogepatton
 
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&BDesign and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Sreedhar Chowdam
 
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
ydteq
 
MCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdfMCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdf
Osamah Alsalih
 
6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)
ClaraZara1
 
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdfGoverning Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
WENKENLI1
 
Fundamentals of Electric Drives and its applications.pptx
Fundamentals of Electric Drives and its applications.pptxFundamentals of Electric Drives and its applications.pptx
Fundamentals of Electric Drives and its applications.pptx
manasideore6
 
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdfAKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
SamSarthak3
 
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdfHybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
fxintegritypublishin
 
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Dr.Costas Sachpazis
 
Final project report on grocery store management system..pdf
Final project report on grocery store management system..pdfFinal project report on grocery store management system..pdf
Final project report on grocery store management system..pdf
Kamal Acharya
 
Railway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdfRailway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdf
TeeVichai
 
block diagram and signal flow graph representation
block diagram and signal flow graph representationblock diagram and signal flow graph representation
block diagram and signal flow graph representation
Divya Somashekar
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
AJAYKUMARPUND1
 
14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application
SyedAbiiAzazi1
 
space technology lecture notes on satellite
space technology lecture notes on satellitespace technology lecture notes on satellite
space technology lecture notes on satellite
ongomchris
 

Recently uploaded (20)

HYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generationHYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generation
 
Water Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdfWater Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdf
 
RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...
RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...
RAT: Retrieval Augmented Thoughts Elicit Context-Aware Reasoning in Long-Hori...
 
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
Industrial Training at Shahjalal Fertilizer Company Limited (SFCL)
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
 
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&BDesign and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
 
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
 
MCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdfMCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdf
 
6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)6th International Conference on Machine Learning & Applications (CMLA 2024)
6th International Conference on Machine Learning & Applications (CMLA 2024)
 
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdfGoverning Equations for Fundamental Aerodynamics_Anderson2010.pdf
Governing Equations for Fundamental Aerodynamics_Anderson2010.pdf
 
Fundamentals of Electric Drives and its applications.pptx
Fundamentals of Electric Drives and its applications.pptxFundamentals of Electric Drives and its applications.pptx
Fundamentals of Electric Drives and its applications.pptx
 
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdfAKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
 
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdfHybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
 
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
 
Final project report on grocery store management system..pdf
Final project report on grocery store management system..pdfFinal project report on grocery store management system..pdf
Final project report on grocery store management system..pdf
 
Railway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdfRailway Signalling Principles Edition 3.pdf
Railway Signalling Principles Edition 3.pdf
 
block diagram and signal flow graph representation
block diagram and signal flow graph representationblock diagram and signal flow graph representation
block diagram and signal flow graph representation
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
 
14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application
 
space technology lecture notes on satellite
space technology lecture notes on satellitespace technology lecture notes on satellite
space technology lecture notes on satellite
 

Data Teknis Gossen Metrawatt Multimeter : METRALINE DM61, METRALINE DM62

  • 1. 3-447-012-03 6/1.19 GMC-I Messtechnik GmbH METRALINE DM 61/62 Analog-Digital Multimeter Features Automatic Blocking Sockets (ABS) * Automatic blocking sockets prevent incorrect connection of measurement cables and inadvertent selection of the wrong measured quantity. This significantly reduces danger to the user, the instrument and the system under test, and eliminates it entirely in many cases. Automatic / Manual Measuring Range Selection Measured quantities are selected with the rotary switch. The measuring range is automatically matched to measured values. The measuring range can be selected manually as well with the help of the AUTO/MAN key. Display of Negative Values at the Analog Scale Negative values are also displayed at the analog scale for zero- frequency quantities, allowing for observation of measured quantity fluctuation around the zero-point. Storage of Measured Values By pressing the HOLD/MIN/MAX key, the currently displayed measurement value can be „frozen“ in the display. The minimum and maximum values which were present at the input of the measuring instrument after activation of the MIN/MAX mode can be selectively "retained" with the MIN/ MAX function. The most important application is the determination of the minimum or maximum value during long-term observation of measurement quantities. MIN/MAX has no effect on the analog display; it continues to display the current measurement value. Continuity Test Allows for the detection of short-circuits and interrupted conductors. In addition to displaying test results, an acoustic signal can also be generated if desired. Power Saving Circuit The device is switched off automatically if the measured value remains unchanged for a period of approximately 15 minutes, and if none of the controls are activated during this time. Automatic shutdown can be deactivated. Protective Cover for Harsh Conditions The instrument is protected against damage in the event of impacts or dropping by means of a soft rubber cover with tilt stand. The rubber material also assures that the instrument does not wander if it is set up on a vibrating surface. Duty Cycle Measurement – Square-Wave Signals This function makes it possible to test circuits and transmission cables by measuring the frequency and the duty cycle of pulses. Voluntary Manufacturer’s Warranty 36 months for material and workmanship * Patented (patent no. EP 1801 598, US 7,439,725) • Voltage: DC / AC 100 μV ... 1000 V • Current: DC / AC: 10 μA ... 660.0 mA (DM 61) / 10.00A (DM 62) • Clip function 1000:1 for current transformers (DM 61 only) • Resistance: 100 mΩ... 60.00 MΩ • Capacitance: 1 pF ... 40.00 mF (DM 62 only) • Frequency: 10.00 Hz ... 10.00 MHz (DM 62 only) • Diode / Continuity • Duty cycle (%) measurement (DM 62 only) • Temperature TC with K-type: –50 ... 1300 °C • TRMS bandwidth: 2 kHz (DM 62 only) • Hold / Peak / Min-Max / Relative (Zero) • Auto / Manual ranging • Dual digital display with analog scale and backlight • ABS Automatic Blocking Sockets • UL Certification • 3 year warranty
  • 2. 2 GMC-I Messtechnik GmbH METRALINE DM 61/62 Analog-Digital Multimeter Characteristic Values 1) At 0 °C ... + 40 °C 2) At input > 3.5 Vrms, typical 5 Vp-p, square wave, bipolar inputs 3) For < 10 kHz at 5 Vp-p, square wave, bipolar inputs 4) Without sensor 5) Display with current transformers 1000 : 1 6) Limited by 10 A fuse Influencing Quantities and Influence Error 1) With temperature: Error data apply per 10 K change in temperature. With frequen- cy: Error data apply to a display from 300 digits onwards. 2) With unknown waveform (crest factor CF > 2), measure with manual range selection 3) With the exception of sinusoidal waveform. 4) After the „ “ symbol is displayed. Display Liquid crystal display (58 mm x 31.4 mm) with analog indication and digital display and with display of the unit of measured quantity, function and various special functions. Analog Indication LCD scale with bar graph Scale length 55 mm Scaling 65 scale divisions during all the measurement Polarity indication With automatic reversal Overrange indication By triangle Sampling rate 28 times/s Meas. Func- tion Measuring Range DM61 DM62 (TRMS) Resolu- tion Input Impedance Digital dis- play Inherent deviation at reference condition +(...%rdg +...digits) Overload capacity 1) Over- load values Over- load dura- tion V(DC) 660.0 mV • • 100 μV >100 MΩ // <40pF 0.7 + 5 1000V DC AC eff/rms Sine wave Cont. 6.600 V • • 1 mV 11 MΩ //<40pF 0.4 + 5 66.00 V • • 10 mV 10 MΩ // <40pF 0.4 + 5 660.0 V • • 100 mV 10 MΩ // <40pF 0.4 + 5 1000 V • • 1 V 10 MΩ // <40pF 0.4 + 5 V(AC) 660.0 mV • • 100 μV >100 MΩ // <40pF 1.2 + 5 6.600 V • • 1 mV 11 MΩ //<40pF 1.0 + 3 66.00 V • • 10 mV 10 MΩ // <40pF 660.0 V • • 100 mV 10 MΩ // <40pF 1000 V • • 1 V 10 MΩ // <40pF A(DC) Voltage Drop 66.00 mA • • 10 μA 66.00 mV 0.8 + 5 0.7 A Cont. 660.0 mA • • 100 μA 66.00 mV 0.8 + 5 10.00 A 6) — • 10 mA 10.00 mV 1.5 + 5 — — A(AC) 66.00 mA • • 10 μA 66.00 mV 0.8 + 5 0.7 A Cont. 660.0 mA • • 100 μA 66.00 mV 0.8 + 5 10.00 A 6) — • 10 mA 10.00 mV 1.5 + 5 — — (AC) 5) 66.00 A • — 10 mA 66.00 mV 0.8 + 5 0.7 A Cont. 660.0 A • — 100 mA 66.00 mV 0.8 + 5 Ω No load Voltage 660.0 Ω • • 100 mΩ –3.3 V 0.8 + 5 1000 V DC AC eff/rms Sine wave max. 10 s 6.600 kΩ • • 1 Ω –1.08 V 0.8 + 5 66.00 kΩ • • 10 Ω –1.08 V 0.8 + 5 660.0 kΩ • • 100 Ω –1.08 V 0.8 + 5 6.600 MΩ • • 1 kΩ –1.08 V 1.0 + 5 66.00 MΩ • • 10 kΩ –1.08 V 2.0 + 5 660.0 Ω • • 100 mΩ –3.3 V 0.8 + 5 1000V DC AC eff/rms Sine wave max. 10 s DIODE 2.000 V • • 1 mV 3.3 V 2.0 + 10 F 6.600 nF — • 1 pF — 3.0 + 40 66.00 nF — • 10 pF 2.0 + 10 660.0 nF — • 100 pF 2.0 + 10 6.600 μF — • 1 nF 2.0 + 10 66.00 μF — • 10 nF 2.0 + 10 660.0 μF — • 100 nF 5.0 + 10 6.600 mF — • 1 μF 5.0 + 10 40.00 mF — • 10 μF 5.0 + 10 Hz f min 66.00 Hz — • 0.01 Hz 10 Hz 0.2 + 2 2) 660.0 Hz — • 0.1 Hz 6.600 kHz — • 1 Hz 66.00 kHz — • 10 Hz 660.0 kHz — • 100 Hz 6.600 MHz — • 1 kHz 10.00 MHz — • 10 kHz % 1.0 ... 98.90% — • 0.01 % 0.9% (% min) 10 Hz...1 kHz ±5 Digit 3) 1...10 kHz; ±5 Digit/kHz °C/°F 0 ... 1300 °C • • 1 °C — 2.0 + 3 4) –50 ... 0 °C • • 1 °C — 2.0 ±10 4) Influencing Quantity Range of Influence Measured Quantity/ Measuring Range Influence Error1) ±(... % of rdg. + ... digits) Temperature 0 °C ... +21 °C and +25 °C ... +40 °C V DC, V AC 1 x Intrinsic uncertainty/K A DC, A AC Ω Diode F, Hz, %, °C Measured Quantity Frequency 20 Hz ... < 50 Hz 660 mV~ 1.0 + 3 > 50 Hz ... 200 Hz 5.0 + 3 20 Hz ... < 50 Hz 6.6 ... 1000 V~ 1.0 + 3 > 50 Hz ... 2 kHz 5.0 + 7 > 50 Hz ... 200 Hz A~ 1.0 + 3 20 Hz ... < 2 kHz 5.0 + 3 Crest Factor CF 1 ... 1.4 V~3) , A~3) ±1% of rdg 1.4 ... 5 2) ±5% of rdg Battery Voltage 4) ... < 2.49 V > 2.49 V ... 3 V V DC 5 Digit V~, A DC 10 Digit A AC 6 Digit 660 Ω 4 Digit 6.600 kΩ ... 66.00 MΩ 3 Digit nF, F, mF, Hz, % 5 Digit Relative Humidity 75% 3 days Meter off V~, V DC A~, A DC Ω F Hz °C % 1 x intrinsic uncertainty Influencing Quantity Range of Influence Measuring Range Attenuation Common Mode Interference Voltage Noise quantity max. 1000 V V > 100 dB V > 100 dB Noise quantity max. 1000 V 50 Hz, 60 Hz sinusoidal V > 100 dB V > 50 dB Normal Mode Interference Voltage Noise quantity: V , value of the measuring range at a time max. 1000 V , 50 Hz, 60 Hz sinusoidal 660 mV, 6.6 V, 660 V, 1000 V DC > 43 dB 66 V DC > 35 dB Noise quantity max. 1000 V V > 45 dB
  • 3. GMC-I Messtechnik GmbH 3 METRALINE DM 61/62 Analog-Digital Multimeter Digital Height of Main Display numerals 7 segment numerals: 12 mm Height of Sub Display numerals 7 segment numerals: 7 mm Number of counts 4 digit: 6600 steps Overrange display „OL“ is shown Polarity display „–“ sign is shown, When positive pole connected to „ “ Sampling rate 2.8 times/s Power supply Battery 2 AA size batteries alkaline manganese cells as per IEC LR6. Service life for METRALINE DM 61: 600 hrs. for V DC, A DC 300 hrs. for V AC, A AC for METRALINE DM 62: 400 hrs. for V DC, A DC 200 hrs. for V AC, A AC Battery test Automatic display of „ “ symbol when battery voltage falls below following value: approx. 2.4 V. Electromagnetic compatibility (EMC) Emission EN 61326: 2013 Class B Immunity IEC 61000-4-2: 8 kV atmosphere discharge 4 kV contact discharge IEC 61000-4-3: 3 V/m Short-term measured value deviation may occur during electromagnetic interference thus reducing the specified operating quality. Safety: IEC 61010-1-2010 Measuring category 600 V CAT III, 300 V CAT IV The maximum voltage of 1000 V may only be used with CAT II. High Voltage Test 6.7 kV (IEC 61010-1-2010) Fuses Fuse for up to 660 mA ranges FF (UR) 1.6 A/1000 V AC/DC; 6.3 mm X 32 mm; rating 10 kA with 1000 VAC/DC and ohmic load; in conjunction with power diodes, protects all current measuring ranges up to 660 mA. Fuse for up to 10 A ranges (METRALINE DM 62) FF (UR)10 A/1000 V AC/DC; 10 mm x 38 mm; rating 30 kA with 1000 VAC/DC and ohmic load; protects the 10 A ranges up to 1000 V AC/DC. Defective fuses are not displayed. Response Time (after manual range selection) Reference conditions Ambient temperature 23 °C+ 2 K Relative humidity 45% ... 55 % RH Frequency of measured quantity 50 or 60 Hz ±2% Waveform of the measured quantity sinusoidal Battery voltage 3 V ±0.1 V Environmental conditions Functional temperature range 0 °C ... +50 °C Storage temperature range –25 °C ... +70 °C (without batteries) Relative humidity 45 ... 75 % Altitude up to 2000 m Mechanical configuration Protection for the meter IP50 Pollution degree 2 Connection sockets IP20 according to EN 60529 / DIN VDE 0470-1 Dimensions with holster: 86 mm x 188 mm x 53 mm without holster: 79 mm x 174 mm x 38 mm Weight 480 g approx., including battery and holster Applicable Regulations and Standards Measured Quantity/ Measuring Range Response Time Transient response for step function of the measured quantity Analog Display Digital Display V , V , °C 0.1 s 1 s from 0 to 80% of the upper range limit A , A 0.1 s 1 s from 0 to 50% of the upper range limit 660 Ω ... 6.6 MΩ 0.1 s 1 s 66 MΩ 0.2 s 2 s 0.1 s 1 s from 0 to 80% of the upper range limit 6.6 nF ... 66 μF 0.7 s max. 1 s 660 μF ... 6.6 mF 1.4 s max. 3 s 66 mF 7.0 s max. 15 s 660 Hz, 6.6 kHz 2.0 s max. 2 s 66 kHz, 660 kHz, 1 MHz 0.5 s max. 1 s % (≥ 10 Hz) 0.7 s max. 2.5 s IEC 61010-1/EN 61010-1/ VDE 0411-1 Safety requirements for electrical equipment for measurement, control and laboratory use EN 60529 VDE 0470, Part 1 Test instruments and test procedures Protection provided by enclosures (IP code) DIN EN 61326-2-1 VDE 0843-02-2-1 Electrical equipment for measurement, control and labo- ratory use – EMC requirements – Part 2-1: Particular requirements for sensitive test and measurement equipment DIN EN 60529 DIN VDE 0470 Part 1 Test Instruments and test procedures – Degree of protection provided by enclosures (IP code)
  • 4. Edited in Germany • Subject to change without notice • A pdf version is available on the Internet Phone +49 911 8602-111 Fax +49 911 8602-777 E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.com GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg • Germany METRALINE DM 61/62 Analog-Digital Multimeter Standard Equipment 1 Multimeter 1 Rubber holster with carrying strap 1 Cable set 1 Battery set 1 Operating instructions 1 Test report Order Information For additional information on accessories, please refer to • our „Measuring Instruments and Testers“ catalogue • our website www.gossenmetrawatt.com Description Type Article Number Clipping multimeter, clip factor 1:1000 for current measurement with optional clamp WZ1001 as accessory METRALINE DM 61 M194A TRMS Multimeter METRALINE DM 62 M197A Accessories AC clamp 1000:1 WZ1001 Z194A