DIGITAL CLAMP METER
MODEL : 27-AUTO/27T-AUTO
INSTRUCTION MANUAL
®
1
4
Temperature (27T-AUTO only)
Range : -200
C to 7500
C
-40
F to 14000
F
Accuracy : ± (3% rdg + 5 dgts)
Resolution : 10
C / 10
F
Sensor : K type thermocouple
2. OPERATION
Before taking any measurements, read the safety information section.
Always examine the instrument for damage, contamination (excessive
dirt, grease, etc.) and defects. Examine the test leads for cracked or
frayed insulation. If any abnormal conditions exist do not attempt to
make any measurements.
AUTO POWER OFF (APO)
The meter will be switch off if no range switch or key is used for approx
15 minutes.
FUNCTION BUTTON
The Button is used for select / range and 0
C/0
F (27T-AUTO only)
HOLD BUTTON
Press Hold button to toggle in and out of Hold mode, In the Hold mode,
the “ H ” annunciator is displayed.
MAX BUTTON
Press MAX button to toggle in and out of MAX mode (holding the highest
absolute reading) In the MAX mode, the “ MAX ” annunciator is displayed.
This function is available for DCV, ACV & Current ranges.
2.1 Current Measurements
1. Set the Range switch to the highest 400A range.
2. Press the trigger to open transformer jaws, clamp onto one
conductor only and release trigger. Jaws should be completely
closed. Read the current directly on the display. It is
recommended that the conductor be placed at th e center of
the closed jaws for maximum accuracy (Fig. - 1).
1. SPECIFICATIONS
1.1 General Specification
Display n 3½ digit liquid crystal display (LCD) with a
maximum reading of 2000 counts.
Polarity n Automatic, positive implied, negative polarity
indication.
Overrange n (OL) or (-OL) is displayed.
Zero n Automatic.
Low battery n “ ” is displayed when the battery
indication voltage drops below the operating level.
Measurement n 3 times per second, nominal.
rate
Operating n 0°C to 50°C at < 70% relative humidity.
Environment
Storage n -20°C to 60°C, 0 to 80% R.H. with battery
Temperature removed from meter.
Accuracy n Stated accuracy at 270
C ± 50
C,
<75% relative humidity.
Power n Two 1.5V ‘AAA’ Size Battery
Battery life n 200 hours typical
Dimensions n 185 x 65 x 28mm (approx.)
Weight n 170gms including battery (approx.)
Accessories n Pair of test leads x 1, Instruction manual x 1,
1.5V batterys (installed) x 2,
Carring Case x 1, (K type thermocouple upto
2600
C for 27T-AUTO only) x 1
Maximum Jaw n 25mm
Opening
2
1.2 Electrical Specification
Accuracies are ± (% reading + number of digits) at 27 ± 5°C and
humidity of less than 75% RH.
AC CURRENT (Auto Ranging) 50-60Hz
DC VOLTAGE (Auto Ranging)
Range Accuracy Overload Protection
600V ±(0.8% rdg + 3 dgt) 600V DC / AC rms
Range Accuracy Overload Protection
600V ±(1.2% rdg + 3 dgt) 600V DC / AC rms
RESISTANCE (Auto Ranging)
Range Accuracy Overload Protection
20MVVVVV ±(1.2% rdg + 3 dgt) 250V DC / AC rms
AC VOLTAGE (Auto Ranging) 50-60Hz
Continuity Check
Threshold Level : 40V Approx.
Response Time : 1m Sec. Approx.
Open Circuit Voltage : 0.4V Approx.
Indication : ‘ ’ is displayed on LCD and buzzer sounds
at continuity.
Diode Test
Measurement Current : 1.0 ± 0.6 mA Approx.
Open Circuit Voltage : 1.6V Approx.
3
Range Accuracy Overload Protection
2A ±(3% rdg + 5 dgt)
20A ±(2% rdg + 3 dgt) 400A AC Max.
200A
±(2% rdg + 5 dgt)
for 1 minute
400A
27-27T : 28-09-2012
6
3. Select Diode range by using Function Button.
4. Turn off power to the circuit under test.
5. Touch probes to the diode. A forward-voltage drop is about
0.6V (typical for a silicon diode).
6. If the digital display reads overrange “ OL ”, reverse the lead
connections. The placement of the test leads when the forward
reading is displayed indicates the orientation of the diode. The
red lead is positive and the black lead is negative. If overrange
“ OL ” is displayed with both lead connections, the junction is
open. If a low reading (less than 1000) is obtained with both
lead connections, the junction is shorted internally or (if
junction is measured in a circuit) the junction is shunted by a
resistance less than 1KV. In the letter case the junction must
be disconnected from the circuit in order to verify its operartion.
2.5 Continuity Measurement
1. Connect red test lead to the “VV” jack and black test lead
to the “COM” jack.
2. Set Range switch to the / range.
3. Select Continuity range by using Function Button.
In the continuty test, the beeper sounds cuntinuously, if the
resistance is less than 40V.
2.6 NCV Check (ACV only)
Indicates presence of voltage in an electrical circuit or equipment
without touching them.
1. Set Range switch to the NCV Range.
2. The NCV indicator flashing every 1-2 sec.
3. When the clamp jaw near to the object under test it detected voltage.
The NCV LED is quickly flashing.
l Detection against inwall outlet is possible.
l In NCV range meter will be auto power off within 3 minutes,
if no signals is obtained
2.7 Temperature Measurment (27T-AUTO only)
1. Set range switch to 0
C possition.
2. Using function button select 0
C or 0
F.
3. Connect the thermocouple “+, -” at VV and COM input terminals.
4. Touch the end of temperature probe to the area or surface of the
object whose temperature is to be measured.
8
®
Certificate of Calibration
We hereby certify that this product has been calibrated and found
to be in accordance with the applicable SPECIFICATIONS and
STANDARDS.
Accuracies of the standard equipment used in this calibration
are traceable to the National Standards.
SR. NO. :
CHECKED BY :
DATE :
MODEL NO. :
MECO METERS PVT. LTD.
Plot No. EL-60, MIDC Electronic Zone, TTC Industrial Area,
Mahape, Navi Mumbai - 400710 (INDIA)
Tel : 0091-22-27673311-16, 27673300 (Board)
Fax : 0091-22-27673310, 27673330
E-mail : sales@mecoinst.com
Web : www.mecoinst.com
3. When the reading is lower than 20A, set the Range switch to
the next lower range position.
2.2 Voltage Measurements (AC or DC)
1. Connect the red test lead to the “VV” jack and the black test lead
to the “COM” jack.
2. Set the Range switch to the desired Voltage type (AC or DC)
3. Connect the test leads to the device or circuit being measured.
4. For DC, a (-) sign is displayed for negative polarity; positive
polarity is implied.
2.3 Resistance Measurements
1. Connect red test lead to the “VV” jack and black test lead
to the “COM” jack.
2. Set Range switch to the V range.
3. If the resistance being measured is connected to a circuit,
turn off power to the circuit being tested and discharge all
capacitors.
Fig. 1
7 3
4. Connect test leads across the resistance being measured.
When measuring high resistance, be sure not to contact
adjacent points even if insulated because some insulators
have a relatively low insulation resistance, causing the measured
resistance to be lower than the actual resistance.
5. Read resistance value on digital display. If a high resistance
value is shunted by a large value of capacitance allow display to
stabilize.
2.4 Diode Test
1. Connect the red test lead to the “VV” jack and black test lead
to the “ COM ” jack.
2. Set the Range switch to the “ / ”.5
7
3. MAINTENANCE
WARNING : Remove test leads before changing battery or servicing.
Battery Replacement
Power is supplied by two 1.5V ‘AAA’ size battery or Equivalent.
The “ ” appears on the LCD display when replacement is needed, To
replace the battery, remove the screw from the battery cover and lift off
the battery cover. Remove the batterys & repleced with new batterys.
4. SAFETY INFORMATION
The following safety information must be observed to ensure maximum
personal safety during the operation of this meter:
1. Do not use the meter if the meter or test leads look damaged,
or if you suspect that the meter is not operating properly.
2. This Clamp Meter is designed to take current measurements
on circuits with a maximum voltage difference of 500VAC
between any conductor and ground potential. Using the
instrument for current measurements on circuits above this
voltage may cause electric shock, instrument damage or
damage to the equipment under test.
Before measuring current make certain the test leads are
removed from the instrument.
3. The instrument is protected for overload upto 600 VAC for 1
minute. Do not take current readings on circuits where the
maximum current potential is not known.
Do not exceed the maximum currents that this instrument is
designed to measure.
4. Turn off power to the circuit under test before cutting,
unsoldering, or breaking the circuit. Small amounts of current
can be dangerous.
5. Use caution when working above 60V DC or 30V AC rms.
Such voltages pose a shock hazard.
6. When using the probes, keep your fingers behind the finger
guards on the probes.
7. Measuring voltage which exceeds the limits of the clamp meter
may damage the meter and expose the operator to a shock
hazard. Always recognize the meter voltage limits as stated
on the front of the meter.
Important : To avoid heat damage to the meter keep it away from
sources of very high temperature. The life of the temperature probe is
also reduced. When subjected to very high temperature.
225AUTO : 28-09-2012
AUTORANGING
DIGITAL CLAMP METER
MODEL : 225AUTO
INSTRUCTION MANUAL
®
1. SPECIFICATIONS
1.1 General Specification
Display n 3¾ digit liquid crystal display (LCD) with a
maximum reading of 3999.
Polarity n Automatic, positive implied, negative polarity
indication.
Overrange n (OL) or (-OL) is displayed.
Zero n Automatic.
Low battery n “ ” is displayed when the battery
indication voltage drops below the operating level.
Measurement n 3 times per second, nominal.
rate
Operating n 0°C to 50°C at < 70% relative humidity.
Environment
Storage n -20°C to 60°C, 0 to 80% R.H. with battery
Temperature removed from meter.
Accuracy n Stated accuracy at 270
C ± 50
C,
<75% relative humidity.
Power n Two 1.5V ‘AAA’ Size Battery
Battery life n 200 hours typical
Dimensions n 186 x 58 x 30mm (approx.)
Weight n 165gms including battery (approx.)
Accessories n Pair of test leads x 1, instruction manual x 1,
1.5V batterys (installed) x 2, Carring Case x 1
Maximum Jaw n 32mm
Opening
1.2 Electrical Specification
Accuracies are ± (% reading + number of digits) at 27 ± 5°C and
humidity of less than 75% RH.
AC CURRENT (50-60Hz)
DC VOLTAGE (Auto Ranging)
Range Accuracy Overload Protection
40A
±(1.5% rdg + 5dgt)
400A AC Max.
400A for 1 minute
Range Accuracy Overload Protection
600V ±(0.8% rdg + 1dgt)
AC voltage ranges 600V AC
Resistance ranges 250V AC
Range Accuracy Overload Protection
600V ±(1.0% rdg + 3dgt)
AC voltage ranges 600V AC
Resistance ranges 250V AC
RESISTANCE (Auto Ranging)
Range Accuracy Overload Protection
40MVVVVV ±(1.0% rdg + 2dgt)
AC voltage ranges 600V AC
Resistance ranges 250V AC
AC VOLTAGE (Auto Ranging) 50-60Hz
Continuity Check
Threshold Level : 50V Approx.
Response Time : 1m Sec. Approx.
Open Circuit Voltage: 0.4V Approx.
Indication : ‘ ’ is displayed on LCD and buzzer sounds
at continuity.
Diode Test
Measurement Current : 1.0 ± 0.6 mA Approx.
Open Circuit Voltage : 0.4V Approx.
2. OPERATION
Before taking any measurements, read the safety information section.
Always examine the instrument for damage, contamination (excessive
dirt, grease, etc.) and defects. Examine the test leads for cracked or
frayed insulation. If any abnormal conditions exist do not attempt to
make any measurements.
SELECT BUTTON
The Button is used for select V / / range.
DATA HOLD BUTTON
Press Data Hold button to toggle in and out of Data Hold mode, In the
Data Hold mode, the “ HOLD ” annunciator is displayed.
3 4
21
225AUTO : 28-09-2012
3. When the reading is lower than 40A, set the Function / Range
switch to the next lower range position.
2.2 Voltage Measurements
1. Connect the red test lead to the “VV” jack and the black test lead
to the “COM” jack.
2. Set the Function/Range switch to the desired Voltage type
(AC or DC)
3. Connect the test leads to the device or circuit being measured.
4. For DC, a (-) sign is displayed for negative polarity; positive
polarity is implied.
2.3 Resistance Measurements
1. Connect red test lead to the “VV” jack and black test lead
to the “COM” jack.
2. Set Function / Range switch to the V / / range.
3. Select resistance range by using Select Button.
4. If the resistance being measured is connected to a circuit,
turn off power to the circuit being tested and discharge all
capacitors.
2.1 Current Measurements
1. Set the Function / Range switch to the highest 400A range.
2. Press the trigger to open transformer jaws, clamp onto one
conductor only and release trigger. Jaws should be completely
closed. Read the current directly on the display. It is
recommended that the conductor be placed at th e center of
the closed jaws for maximum accuracy (Fig. - 1).
Fig. 1
7 3
5. Connect test leads across the resistance being measured.
When measuring high resistance, be sure not to contact
adjacent points even if insulated because some insulators
have a relatively low insulation resistance, causing the measured
resistance to be lower than the actual resistance.
6. Read resistance value on digital display. If a high resistance
value is shunted by a large value of capacitance allow display to
stabilize.
2.4 Diode Test
1. Connect the red test lead to the “VV” jack and black test lead
to the “ COM ” jack.
2. Set the Function / Range switch to the “ V / / ”.
3. Select Diode range by using Select Button.
4. Turn off power to the circuit under test.
5. Touch probes to the diode. A forward-voltage drop is about
0.6V (typical for a silicon diode).
6. If the digital display reads overrange “ OL ”, reverse the lead
connections. The placement of the test leads when the forward
reading is displayed indicates the orientation of the diode. The
red lead is positive and the black lead is negative. If overrange
“ OL ” is displayed with both lead connections, the junction is
open. If a low reading (less than 1000) is obtained with both
lead connections, the junction is shorted internally or (if
junction is measured in a circuit) the junction is shunted by a
resistance less than 1KV. In the letter case the junction must
be disconnected from the circuit in order to verify its operartion.
2.5 Continuity Measurement
1. Connect red test lead to the “VV” jack and black test lead
to the “COM” jack.
2. Set Function / Range switch to the V / / range.
3. Select Continuity range by using Select Button.
In the continuty test, the beeper sounds cuntinuously, if the resistance
is less than 50V.
3. MAINTENANCE
WARNING
Remove test leads before changing battery or servicing.
Battery Replacement
Power is supplied by two 1.5V ‘AAA’ size battery or Equivalent.
The “ ” appears on the LCD display when replacement is needed,
To replace the battery, remove the screw from the battery cover and
lift off the battery case. Remove the batterys & repleced with new
batterys.
4. SAFETY INFORMATION
The following safety information must be observed to ensure maximum
personal safety during the operation of this meter:
1. Do not use the meter if the meter or test leads look damaged,
or if you suspect that the meter is not operating properly.
2. This Clamp Meter is designed to take current measurements
on circuits with a maximum voltage difference of 500VAC
between any conductor and ground potential. Using the
instrument for current measurements on circuits above this
voltage may cause electric shock, instrument damage or
damage to the equipment under test.
Before measuring current make certain the test leads are
removed from the instrument.
3. The instrument is protected for overload upto 600 VAC for 1
minute. Do not take current readings on circuits where the
maximum current potential is not known.
Do not exceed the maximum currents that this instrument is
designed to measure.
4. Turn off power to the circuit under test before cutting,
unsoldering, or breaking the circuit. Small amounts of current
can be dangerous.
5. Use caution when working above 60V DC or 30V AC rms.
Such voltages pose a shock hazard.
6. When using the probes, keep your fingers behind the finger
guards on the probes.
7. Measuring voltage which exceeds the limits of the clamp meter
may damage the meter and expose the operator to a shock
hazard. Always recognize the meter voltage limits as stated
on the front of the meter.7
8
65
CERTIFICATE OF CALIBRATION
We hereby certify that this product has been
calibrated and found to be in accordance with
the applicable SPECIFICATIONS and MECO
STANDARDS.
Accuracies of the standard equipment used in
this calibration are traceable to the National
Standards.
MECO METERS PVT. LTD.
Plot No. EL-60, MIDC Electronic Zone, TTC Industrial Area,
Mahape, Navi Mumbai - 400710 (INDIA)
Tel : 0091-22-27673311-16, 27673300 (Board)
Fax : 0091-22-27673310, 27673330
E-mail : sales@mecoinst.com Web : www.mecoinst.com
SR. NO : ________________________________
CHECKED BY : ________________________________
DATE : ________________________________
MODEL NO : ________________________________
®

More Related Content

What's hot

Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...
Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...
Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...
Thorne & Derrick International
 
Aml catalogue 13
Aml catalogue 13Aml catalogue 13
Aml catalogue 13
slideroma
 
Yehai catalogue -tracy chow
Yehai catalogue -tracy chowYehai catalogue -tracy chow
Yehai catalogue -tracy chow
slideroma
 
W100P WCT PH Panel type
W100P WCT PH Panel typeW100P WCT PH Panel type
W100P WCT PH Panel type
En Campos
 

What's hot (20)

Kaise sk 20 22-30 manual
Kaise sk 20 22-30 manualKaise sk 20 22-30 manual
Kaise sk 20 22-30 manual
 
2 n6509g
2 n6509g2 n6509g
2 n6509g
 
Fe 712 instruction_sheet_1
Fe 712 instruction_sheet_1Fe 712 instruction_sheet_1
Fe 712 instruction_sheet_1
 
Martindale Electric Metrohm Flash Testers & Insulation Testers - MET7A501 Saf...
Martindale Electric Metrohm Flash Testers & Insulation Testers - MET7A501 Saf...Martindale Electric Metrohm Flash Testers & Insulation Testers - MET7A501 Saf...
Martindale Electric Metrohm Flash Testers & Insulation Testers - MET7A501 Saf...
 
Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...
Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...
Martindale Electric Metrohm METE3640 4kV Flash Tester - Martindale Electric M...
 
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTHData Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
Data Teknis Gossen Metrawatt Earth Ground Clamp : METRACLIP EARTH
 
Aml catalogue 13
Aml catalogue 13Aml catalogue 13
Aml catalogue 13
 
Yehai catalogue -tracy chow
Yehai catalogue -tracy chowYehai catalogue -tracy chow
Yehai catalogue -tracy chow
 
Imegger, mit 4002, cat iv insulation tester 1000 v, op 54
Imegger, mit 4002, cat iv insulation tester 1000 v, op 54Imegger, mit 4002, cat iv insulation tester 1000 v, op 54
Imegger, mit 4002, cat iv insulation tester 1000 v, op 54
 
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRO
Data Teknis Gossen Metrawatt Insulation Tester METRISO PROData Teknis Gossen Metrawatt Insulation Tester METRISO PRO
Data Teknis Gossen Metrawatt Insulation Tester METRISO PRO
 
Triac 2 n6073a
Triac 2 n6073aTriac 2 n6073a
Triac 2 n6073a
 
6 n137 datasheet_en_20170808 (1)
6 n137 datasheet_en_20170808 (1)6 n137 datasheet_en_20170808 (1)
6 n137 datasheet_en_20170808 (1)
 
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COILData Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
Data Teknis Gossen Metrawatt Multimeter : METRAHIT ISO & METRAHIT COIL
 
Seaward Catalog High Voltage Test Tools
Seaward Catalog High Voltage Test ToolsSeaward Catalog High Voltage Test Tools
Seaward Catalog High Voltage Test Tools
 
Rotork YTC YT-2501 Smart Positioner Manual | YTC INDIA
Rotork YTC YT-2501 Smart Positioner Manual | YTC INDIARotork YTC YT-2501 Smart Positioner Manual | YTC INDIA
Rotork YTC YT-2501 Smart Positioner Manual | YTC INDIA
 
Data Teknis Gossen Metrawatt TRMS Multimeter : METRAHIT 2+
Data Teknis Gossen Metrawatt TRMS Multimeter : METRAHIT 2+Data Teknis Gossen Metrawatt TRMS Multimeter : METRAHIT 2+
Data Teknis Gossen Metrawatt TRMS Multimeter : METRAHIT 2+
 
Data Teknis Gossen Metrawatt Multimeter : METRAHIT 2 PLUS
Data Teknis Gossen Metrawatt Multimeter : METRAHIT 2 PLUSData Teknis Gossen Metrawatt Multimeter : METRAHIT 2 PLUS
Data Teknis Gossen Metrawatt Multimeter : METRAHIT 2 PLUS
 
Datasheet Metrel MI 3250. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3250. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MI 3250. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3250. Hubungi PT. Siwali Swantika 021-45850618
 
W100P WCT PH Panel type
W100P WCT PH Panel typeW100P WCT PH Panel type
W100P WCT PH Panel type
 
Av02 3563 en-ds_acpl-c87x_04mar2013-2
Av02 3563 en-ds_acpl-c87x_04mar2013-2Av02 3563 en-ds_acpl-c87x_04mar2013-2
Av02 3563 en-ds_acpl-c87x_04mar2013-2
 

Viewers also liked

Viewers also liked (12)

Creating your DRF program
Creating your DRF programCreating your DRF program
Creating your DRF program
 
Mapa conceptual
Mapa conceptualMapa conceptual
Mapa conceptual
 
Banco Guipuzcoano San Sebastian 2005
Banco Guipuzcoano San Sebastian 2005Banco Guipuzcoano San Sebastian 2005
Banco Guipuzcoano San Sebastian 2005
 
TOP DIGITAL MULTIMETER
TOP DIGITAL MULTIMETERTOP DIGITAL MULTIMETER
TOP DIGITAL MULTIMETER
 
Form and survey builder training
Form and survey builder training Form and survey builder training
Form and survey builder training
 
(School.antoree.com) Đề ôn thi Đại học môn Anh cô Mai Phương
(School.antoree.com) Đề ôn thi Đại học môn Anh cô Mai Phương(School.antoree.com) Đề ôn thi Đại học môn Anh cô Mai Phương
(School.antoree.com) Đề ôn thi Đại học môn Anh cô Mai Phương
 
Recursos educativos abiertos
Recursos educativos abiertosRecursos educativos abiertos
Recursos educativos abiertos
 
Clamp-On, Precision, Split Core Current Transformer
Clamp-On, Precision, Split Core Current TransformerClamp-On, Precision, Split Core Current Transformer
Clamp-On, Precision, Split Core Current Transformer
 
Cómo analizar y traducir oraciones
Cómo analizar y traducir oracionesCómo analizar y traducir oraciones
Cómo analizar y traducir oraciones
 
Origenes de roma
Origenes de romaOrigenes de roma
Origenes de roma
 
Ladrones en el foro
Ladrones en el foroLadrones en el foro
Ladrones en el foro
 
Textos repaso sintaxis 28 9-2016
Textos repaso  sintaxis 28 9-2016Textos repaso  sintaxis 28 9-2016
Textos repaso sintaxis 28 9-2016
 

Similar to Digital clampmeters

HELE / TLEelectronics
HELE / TLEelectronicsHELE / TLEelectronics
HELE / TLEelectronics
djpprkut
 
lem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdf
lem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdflem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdf
lem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdf
NestorMartin8
 

Similar to Digital clampmeters (20)

dt830d-multimeter-datasheet.pdf
dt830d-multimeter-datasheet.pdfdt830d-multimeter-datasheet.pdf
dt830d-multimeter-datasheet.pdf
 
Hướng dẫn sử dụng Testo 755-1
Hướng dẫn sử dụng Testo 755-1Hướng dẫn sử dụng Testo 755-1
Hướng dẫn sử dụng Testo 755-1
 
Megger-BM802-User-Manual.pdf
Megger-BM802-User-Manual.pdfMegger-BM802-User-Manual.pdf
Megger-BM802-User-Manual.pdf
 
HELE / TLEelectronics
HELE / TLEelectronicsHELE / TLEelectronics
HELE / TLEelectronics
 
Battery tester manual
Battery tester manualBattery tester manual
Battery tester manual
 
How to use a multimeter
How to use a multimeterHow to use a multimeter
How to use a multimeter
 
Hướng dẫn sử dụng Testo 750-1
Hướng dẫn sử dụng Testo 750-1Hướng dẫn sử dụng Testo 750-1
Hướng dẫn sử dụng Testo 750-1
 
Hướng dẫn sử dụngTesto 750-1
 Hướng dẫn sử dụngTesto 750-1 Hướng dẫn sử dụngTesto 750-1
Hướng dẫn sử dụngTesto 750-1
 
GM328 Tester condensadores, diodos, transistores
GM328 Tester condensadores, diodos, transistoresGM328 Tester condensadores, diodos, transistores
GM328 Tester condensadores, diodos, transistores
 
Advanced motion controls 25a20
Advanced motion controls 25a20Advanced motion controls 25a20
Advanced motion controls 25a20
 
Zero100E insertion type electromagnetic flow meter catalog
Zero100E insertion type electromagnetic flow meter catalogZero100E insertion type electromagnetic flow meter catalog
Zero100E insertion type electromagnetic flow meter catalog
 
Brymen BM157 BM357 User's Manual
Brymen BM157 BM357 User's ManualBrymen BM157 BM357 User's Manual
Brymen BM157 BM357 User's Manual
 
Manual de uso del megohmetro METRISO 2000
Manual de uso del megohmetro METRISO 2000Manual de uso del megohmetro METRISO 2000
Manual de uso del megohmetro METRISO 2000
 
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
Data Teknis Gossen Metrawatt Insulation Tester METRISO INTRO-METRISO BASE-MET...
 
Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413
Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413
Data Teknis Gossen Metrawatt Ohmmeter : METRAOHM 413
 
K39 Controller
K39 ControllerK39 Controller
K39 Controller
 
lem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdf
lem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdflem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdf
lem_procon-plus_400p_750p_1000p_1250p_1500p_1800p_2200p_adjustments_sch.pdf
 
Datasheet profitest mbase_mtech (1)
Datasheet profitest mbase_mtech (1)Datasheet profitest mbase_mtech (1)
Datasheet profitest mbase_mtech (1)
 
Advanced motion controls 50a20
Advanced motion controls 50a20Advanced motion controls 50a20
Advanced motion controls 50a20
 
K49 Controller
K49 ControllerK49 Controller
K49 Controller
 

More from mecoinst

More from mecoinst (16)

Infrared Thermometers-Mecoinst
Infrared Thermometers-MecoinstInfrared Thermometers-Mecoinst
Infrared Thermometers-Mecoinst
 
TESTING & MEASURING INSTRUMENTS by Meco
TESTING & MEASURING INSTRUMENTS by Meco TESTING & MEASURING INSTRUMENTS by Meco
TESTING & MEASURING INSTRUMENTS by Meco
 
THERMO HYGROMETER by Meco
THERMO HYGROMETER by Meco THERMO HYGROMETER by Meco
THERMO HYGROMETER by Meco
 
Digital Multimeter by Mecoinst
Digital Multimeter by Mecoinst 	Digital Multimeter by Mecoinst
Digital Multimeter by Mecoinst
 
Calibrating equipment
Calibrating equipmentCalibrating equipment
Calibrating equipment
 
Calibrating equipment
Calibrating equipmentCalibrating equipment
Calibrating equipment
 
Meters
MetersMeters
Meters
 
Testing & measuring instruments and environmental testing instruments
Testing & measuring instruments and environmental testing instrumentsTesting & measuring instruments and environmental testing instruments
Testing & measuring instruments and environmental testing instruments
 
Leakage current tester
Leakage current testerLeakage current tester
Leakage current tester
 
Digital clampmeters
Digital clampmetersDigital clampmeters
Digital clampmeters
 
Digital multimeter
Digital multimeterDigital multimeter
Digital multimeter
 
4 DIGIT PROGRAMMABLE V & A METER
4 DIGIT PROGRAMMABLE V & A METER4 DIGIT PROGRAMMABLE V & A METER
4 DIGIT PROGRAMMABLE V & A METER
 
Power line transducer introduction-spec-selection-
Power line transducer introduction-spec-selection-Power line transducer introduction-spec-selection-
Power line transducer introduction-spec-selection-
 
Analog Panel & Switchboard Meters
Analog Panel & Switchboard MetersAnalog Panel & Switchboard Meters
Analog Panel & Switchboard Meters
 
ANALOG PANEL AND SWITCHBOARD METERS
ANALOG PANEL AND SWITCHBOARD METERS ANALOG PANEL AND SWITCHBOARD METERS
ANALOG PANEL AND SWITCHBOARD METERS
 
Multifunction Meters- Powerguard
Multifunction Meters- PowerguardMultifunction Meters- Powerguard
Multifunction Meters- Powerguard
 

Recently uploaded

Memorandum Of Association Constitution of Company.ppt
Memorandum Of Association Constitution of Company.pptMemorandum Of Association Constitution of Company.ppt
Memorandum Of Association Constitution of Company.ppt
seri bangash
 
FINAL PRESENTATION.pptx12143241324134134
FINAL PRESENTATION.pptx12143241324134134FINAL PRESENTATION.pptx12143241324134134
FINAL PRESENTATION.pptx12143241324134134
LR1709MUSIC
 
chapter 10 - excise tax of transfer and business taxation
chapter 10 - excise tax of transfer and business taxationchapter 10 - excise tax of transfer and business taxation
chapter 10 - excise tax of transfer and business taxation
AUDIJEAngelo
 
NewBase 24 May 2024 Energy News issue - 1727 by Khaled Al Awadi_compresse...
NewBase   24 May  2024  Energy News issue - 1727 by Khaled Al Awadi_compresse...NewBase   24 May  2024  Energy News issue - 1727 by Khaled Al Awadi_compresse...
NewBase 24 May 2024 Energy News issue - 1727 by Khaled Al Awadi_compresse...
Khaled Al Awadi
 

Recently uploaded (20)

Luxury Artificial Plants Dubai | Plants in KSA, UAE | Shajara
Luxury Artificial Plants Dubai | Plants in KSA, UAE | ShajaraLuxury Artificial Plants Dubai | Plants in KSA, UAE | Shajara
Luxury Artificial Plants Dubai | Plants in KSA, UAE | Shajara
 
Easy Way to Download and Set Up Gen TDS Software on Your Computer
Easy Way to Download and Set Up Gen TDS Software on Your ComputerEasy Way to Download and Set Up Gen TDS Software on Your Computer
Easy Way to Download and Set Up Gen TDS Software on Your Computer
 
State of D2C in India: A Logistics Update
State of D2C in India: A Logistics UpdateState of D2C in India: A Logistics Update
State of D2C in India: A Logistics Update
 
Falcon Invoice Discounting Setup for Small Businesses
Falcon Invoice Discounting Setup for Small BusinessesFalcon Invoice Discounting Setup for Small Businesses
Falcon Invoice Discounting Setup for Small Businesses
 
USA classified ads posting – best classified sites in usa.pdf
USA classified ads posting – best classified sites in usa.pdfUSA classified ads posting – best classified sites in usa.pdf
USA classified ads posting – best classified sites in usa.pdf
 
Evolution and Growth of Supply chain.pdf
Evolution and Growth of Supply chain.pdfEvolution and Growth of Supply chain.pdf
Evolution and Growth of Supply chain.pdf
 
Memorandum Of Association Constitution of Company.ppt
Memorandum Of Association Constitution of Company.pptMemorandum Of Association Constitution of Company.ppt
Memorandum Of Association Constitution of Company.ppt
 
Hyundai capital 2024 1quarter Earnings release
Hyundai capital 2024 1quarter Earnings releaseHyundai capital 2024 1quarter Earnings release
Hyundai capital 2024 1quarter Earnings release
 
India’s Recommended Women Surgeons to Watch in 2024.pdf
India’s Recommended Women Surgeons to Watch in 2024.pdfIndia’s Recommended Women Surgeons to Watch in 2024.pdf
India’s Recommended Women Surgeons to Watch in 2024.pdf
 
Using Generative AI for Content Marketing
Using Generative AI for Content MarketingUsing Generative AI for Content Marketing
Using Generative AI for Content Marketing
 
FINAL PRESENTATION.pptx12143241324134134
FINAL PRESENTATION.pptx12143241324134134FINAL PRESENTATION.pptx12143241324134134
FINAL PRESENTATION.pptx12143241324134134
 
chapter 10 - excise tax of transfer and business taxation
chapter 10 - excise tax of transfer and business taxationchapter 10 - excise tax of transfer and business taxation
chapter 10 - excise tax of transfer and business taxation
 
Unlock Your TikTok Potential: Free TikTok Likes with InstBlast
Unlock Your TikTok Potential: Free TikTok Likes with InstBlastUnlock Your TikTok Potential: Free TikTok Likes with InstBlast
Unlock Your TikTok Potential: Free TikTok Likes with InstBlast
 
Special Purpose Vehicle (Purpose, Formation & examples)
Special Purpose Vehicle (Purpose, Formation & examples)Special Purpose Vehicle (Purpose, Formation & examples)
Special Purpose Vehicle (Purpose, Formation & examples)
 
The Inspiring Personality To Watch In 2024.pdf
The Inspiring Personality To Watch In 2024.pdfThe Inspiring Personality To Watch In 2024.pdf
The Inspiring Personality To Watch In 2024.pdf
 
5 Things You Need To Know Before Hiring a Videographer
5 Things You Need To Know Before Hiring a Videographer5 Things You Need To Know Before Hiring a Videographer
5 Things You Need To Know Before Hiring a Videographer
 
NewBase 24 May 2024 Energy News issue - 1727 by Khaled Al Awadi_compresse...
NewBase   24 May  2024  Energy News issue - 1727 by Khaled Al Awadi_compresse...NewBase   24 May  2024  Energy News issue - 1727 by Khaled Al Awadi_compresse...
NewBase 24 May 2024 Energy News issue - 1727 by Khaled Al Awadi_compresse...
 
Global Interconnection Group Joint Venture[960] (1).pdf
Global Interconnection Group Joint Venture[960] (1).pdfGlobal Interconnection Group Joint Venture[960] (1).pdf
Global Interconnection Group Joint Venture[960] (1).pdf
 
8 Questions B2B Commercial Teams Can Ask To Help Product Discovery
8 Questions B2B Commercial Teams Can Ask To Help Product Discovery8 Questions B2B Commercial Teams Can Ask To Help Product Discovery
8 Questions B2B Commercial Teams Can Ask To Help Product Discovery
 
Byrd & Chen’s Canadian Tax Principles 2023-2024 Edition 1st edition Volumes I...
Byrd & Chen’s Canadian Tax Principles 2023-2024 Edition 1st edition Volumes I...Byrd & Chen’s Canadian Tax Principles 2023-2024 Edition 1st edition Volumes I...
Byrd & Chen’s Canadian Tax Principles 2023-2024 Edition 1st edition Volumes I...
 

Digital clampmeters

  • 1. 27-27T : 28-09-2012 DIGITAL CLAMP METER MODEL : 27-AUTO/27T-AUTO INSTRUCTION MANUAL ® 1 4 Temperature (27T-AUTO only) Range : -200 C to 7500 C -40 F to 14000 F Accuracy : ± (3% rdg + 5 dgts) Resolution : 10 C / 10 F Sensor : K type thermocouple 2. OPERATION Before taking any measurements, read the safety information section. Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements. AUTO POWER OFF (APO) The meter will be switch off if no range switch or key is used for approx 15 minutes. FUNCTION BUTTON The Button is used for select / range and 0 C/0 F (27T-AUTO only) HOLD BUTTON Press Hold button to toggle in and out of Hold mode, In the Hold mode, the “ H ” annunciator is displayed. MAX BUTTON Press MAX button to toggle in and out of MAX mode (holding the highest absolute reading) In the MAX mode, the “ MAX ” annunciator is displayed. This function is available for DCV, ACV & Current ranges. 2.1 Current Measurements 1. Set the Range switch to the highest 400A range. 2. Press the trigger to open transformer jaws, clamp onto one conductor only and release trigger. Jaws should be completely closed. Read the current directly on the display. It is recommended that the conductor be placed at th e center of the closed jaws for maximum accuracy (Fig. - 1). 1. SPECIFICATIONS 1.1 General Specification Display n 3½ digit liquid crystal display (LCD) with a maximum reading of 2000 counts. Polarity n Automatic, positive implied, negative polarity indication. Overrange n (OL) or (-OL) is displayed. Zero n Automatic. Low battery n “ ” is displayed when the battery indication voltage drops below the operating level. Measurement n 3 times per second, nominal. rate Operating n 0°C to 50°C at < 70% relative humidity. Environment Storage n -20°C to 60°C, 0 to 80% R.H. with battery Temperature removed from meter. Accuracy n Stated accuracy at 270 C ± 50 C, <75% relative humidity. Power n Two 1.5V ‘AAA’ Size Battery Battery life n 200 hours typical Dimensions n 185 x 65 x 28mm (approx.) Weight n 170gms including battery (approx.) Accessories n Pair of test leads x 1, Instruction manual x 1, 1.5V batterys (installed) x 2, Carring Case x 1, (K type thermocouple upto 2600 C for 27T-AUTO only) x 1 Maximum Jaw n 25mm Opening 2 1.2 Electrical Specification Accuracies are ± (% reading + number of digits) at 27 ± 5°C and humidity of less than 75% RH. AC CURRENT (Auto Ranging) 50-60Hz DC VOLTAGE (Auto Ranging) Range Accuracy Overload Protection 600V ±(0.8% rdg + 3 dgt) 600V DC / AC rms Range Accuracy Overload Protection 600V ±(1.2% rdg + 3 dgt) 600V DC / AC rms RESISTANCE (Auto Ranging) Range Accuracy Overload Protection 20MVVVVV ±(1.2% rdg + 3 dgt) 250V DC / AC rms AC VOLTAGE (Auto Ranging) 50-60Hz Continuity Check Threshold Level : 40V Approx. Response Time : 1m Sec. Approx. Open Circuit Voltage : 0.4V Approx. Indication : ‘ ’ is displayed on LCD and buzzer sounds at continuity. Diode Test Measurement Current : 1.0 ± 0.6 mA Approx. Open Circuit Voltage : 1.6V Approx. 3 Range Accuracy Overload Protection 2A ±(3% rdg + 5 dgt) 20A ±(2% rdg + 3 dgt) 400A AC Max. 200A ±(2% rdg + 5 dgt) for 1 minute 400A
  • 2. 27-27T : 28-09-2012 6 3. Select Diode range by using Function Button. 4. Turn off power to the circuit under test. 5. Touch probes to the diode. A forward-voltage drop is about 0.6V (typical for a silicon diode). 6. If the digital display reads overrange “ OL ”, reverse the lead connections. The placement of the test leads when the forward reading is displayed indicates the orientation of the diode. The red lead is positive and the black lead is negative. If overrange “ OL ” is displayed with both lead connections, the junction is open. If a low reading (less than 1000) is obtained with both lead connections, the junction is shorted internally or (if junction is measured in a circuit) the junction is shunted by a resistance less than 1KV. In the letter case the junction must be disconnected from the circuit in order to verify its operartion. 2.5 Continuity Measurement 1. Connect red test lead to the “VV” jack and black test lead to the “COM” jack. 2. Set Range switch to the / range. 3. Select Continuity range by using Function Button. In the continuty test, the beeper sounds cuntinuously, if the resistance is less than 40V. 2.6 NCV Check (ACV only) Indicates presence of voltage in an electrical circuit or equipment without touching them. 1. Set Range switch to the NCV Range. 2. The NCV indicator flashing every 1-2 sec. 3. When the clamp jaw near to the object under test it detected voltage. The NCV LED is quickly flashing. l Detection against inwall outlet is possible. l In NCV range meter will be auto power off within 3 minutes, if no signals is obtained 2.7 Temperature Measurment (27T-AUTO only) 1. Set range switch to 0 C possition. 2. Using function button select 0 C or 0 F. 3. Connect the thermocouple “+, -” at VV and COM input terminals. 4. Touch the end of temperature probe to the area or surface of the object whose temperature is to be measured. 8 ® Certificate of Calibration We hereby certify that this product has been calibrated and found to be in accordance with the applicable SPECIFICATIONS and STANDARDS. Accuracies of the standard equipment used in this calibration are traceable to the National Standards. SR. NO. : CHECKED BY : DATE : MODEL NO. : MECO METERS PVT. LTD. Plot No. EL-60, MIDC Electronic Zone, TTC Industrial Area, Mahape, Navi Mumbai - 400710 (INDIA) Tel : 0091-22-27673311-16, 27673300 (Board) Fax : 0091-22-27673310, 27673330 E-mail : sales@mecoinst.com Web : www.mecoinst.com 3. When the reading is lower than 20A, set the Range switch to the next lower range position. 2.2 Voltage Measurements (AC or DC) 1. Connect the red test lead to the “VV” jack and the black test lead to the “COM” jack. 2. Set the Range switch to the desired Voltage type (AC or DC) 3. Connect the test leads to the device or circuit being measured. 4. For DC, a (-) sign is displayed for negative polarity; positive polarity is implied. 2.3 Resistance Measurements 1. Connect red test lead to the “VV” jack and black test lead to the “COM” jack. 2. Set Range switch to the V range. 3. If the resistance being measured is connected to a circuit, turn off power to the circuit being tested and discharge all capacitors. Fig. 1 7 3 4. Connect test leads across the resistance being measured. When measuring high resistance, be sure not to contact adjacent points even if insulated because some insulators have a relatively low insulation resistance, causing the measured resistance to be lower than the actual resistance. 5. Read resistance value on digital display. If a high resistance value is shunted by a large value of capacitance allow display to stabilize. 2.4 Diode Test 1. Connect the red test lead to the “VV” jack and black test lead to the “ COM ” jack. 2. Set the Range switch to the “ / ”.5 7 3. MAINTENANCE WARNING : Remove test leads before changing battery or servicing. Battery Replacement Power is supplied by two 1.5V ‘AAA’ size battery or Equivalent. The “ ” appears on the LCD display when replacement is needed, To replace the battery, remove the screw from the battery cover and lift off the battery cover. Remove the batterys & repleced with new batterys. 4. SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: 1. Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly. 2. This Clamp Meter is designed to take current measurements on circuits with a maximum voltage difference of 500VAC between any conductor and ground potential. Using the instrument for current measurements on circuits above this voltage may cause electric shock, instrument damage or damage to the equipment under test. Before measuring current make certain the test leads are removed from the instrument. 3. The instrument is protected for overload upto 600 VAC for 1 minute. Do not take current readings on circuits where the maximum current potential is not known. Do not exceed the maximum currents that this instrument is designed to measure. 4. Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit. Small amounts of current can be dangerous. 5. Use caution when working above 60V DC or 30V AC rms. Such voltages pose a shock hazard. 6. When using the probes, keep your fingers behind the finger guards on the probes. 7. Measuring voltage which exceeds the limits of the clamp meter may damage the meter and expose the operator to a shock hazard. Always recognize the meter voltage limits as stated on the front of the meter. Important : To avoid heat damage to the meter keep it away from sources of very high temperature. The life of the temperature probe is also reduced. When subjected to very high temperature.
  • 3. 225AUTO : 28-09-2012 AUTORANGING DIGITAL CLAMP METER MODEL : 225AUTO INSTRUCTION MANUAL ® 1. SPECIFICATIONS 1.1 General Specification Display n 3¾ digit liquid crystal display (LCD) with a maximum reading of 3999. Polarity n Automatic, positive implied, negative polarity indication. Overrange n (OL) or (-OL) is displayed. Zero n Automatic. Low battery n “ ” is displayed when the battery indication voltage drops below the operating level. Measurement n 3 times per second, nominal. rate Operating n 0°C to 50°C at < 70% relative humidity. Environment Storage n -20°C to 60°C, 0 to 80% R.H. with battery Temperature removed from meter. Accuracy n Stated accuracy at 270 C ± 50 C, <75% relative humidity. Power n Two 1.5V ‘AAA’ Size Battery Battery life n 200 hours typical Dimensions n 186 x 58 x 30mm (approx.) Weight n 165gms including battery (approx.) Accessories n Pair of test leads x 1, instruction manual x 1, 1.5V batterys (installed) x 2, Carring Case x 1 Maximum Jaw n 32mm Opening 1.2 Electrical Specification Accuracies are ± (% reading + number of digits) at 27 ± 5°C and humidity of less than 75% RH. AC CURRENT (50-60Hz) DC VOLTAGE (Auto Ranging) Range Accuracy Overload Protection 40A ±(1.5% rdg + 5dgt) 400A AC Max. 400A for 1 minute Range Accuracy Overload Protection 600V ±(0.8% rdg + 1dgt) AC voltage ranges 600V AC Resistance ranges 250V AC Range Accuracy Overload Protection 600V ±(1.0% rdg + 3dgt) AC voltage ranges 600V AC Resistance ranges 250V AC RESISTANCE (Auto Ranging) Range Accuracy Overload Protection 40MVVVVV ±(1.0% rdg + 2dgt) AC voltage ranges 600V AC Resistance ranges 250V AC AC VOLTAGE (Auto Ranging) 50-60Hz Continuity Check Threshold Level : 50V Approx. Response Time : 1m Sec. Approx. Open Circuit Voltage: 0.4V Approx. Indication : ‘ ’ is displayed on LCD and buzzer sounds at continuity. Diode Test Measurement Current : 1.0 ± 0.6 mA Approx. Open Circuit Voltage : 0.4V Approx. 2. OPERATION Before taking any measurements, read the safety information section. Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements. SELECT BUTTON The Button is used for select V / / range. DATA HOLD BUTTON Press Data Hold button to toggle in and out of Data Hold mode, In the Data Hold mode, the “ HOLD ” annunciator is displayed. 3 4 21
  • 4. 225AUTO : 28-09-2012 3. When the reading is lower than 40A, set the Function / Range switch to the next lower range position. 2.2 Voltage Measurements 1. Connect the red test lead to the “VV” jack and the black test lead to the “COM” jack. 2. Set the Function/Range switch to the desired Voltage type (AC or DC) 3. Connect the test leads to the device or circuit being measured. 4. For DC, a (-) sign is displayed for negative polarity; positive polarity is implied. 2.3 Resistance Measurements 1. Connect red test lead to the “VV” jack and black test lead to the “COM” jack. 2. Set Function / Range switch to the V / / range. 3. Select resistance range by using Select Button. 4. If the resistance being measured is connected to a circuit, turn off power to the circuit being tested and discharge all capacitors. 2.1 Current Measurements 1. Set the Function / Range switch to the highest 400A range. 2. Press the trigger to open transformer jaws, clamp onto one conductor only and release trigger. Jaws should be completely closed. Read the current directly on the display. It is recommended that the conductor be placed at th e center of the closed jaws for maximum accuracy (Fig. - 1). Fig. 1 7 3 5. Connect test leads across the resistance being measured. When measuring high resistance, be sure not to contact adjacent points even if insulated because some insulators have a relatively low insulation resistance, causing the measured resistance to be lower than the actual resistance. 6. Read resistance value on digital display. If a high resistance value is shunted by a large value of capacitance allow display to stabilize. 2.4 Diode Test 1. Connect the red test lead to the “VV” jack and black test lead to the “ COM ” jack. 2. Set the Function / Range switch to the “ V / / ”. 3. Select Diode range by using Select Button. 4. Turn off power to the circuit under test. 5. Touch probes to the diode. A forward-voltage drop is about 0.6V (typical for a silicon diode). 6. If the digital display reads overrange “ OL ”, reverse the lead connections. The placement of the test leads when the forward reading is displayed indicates the orientation of the diode. The red lead is positive and the black lead is negative. If overrange “ OL ” is displayed with both lead connections, the junction is open. If a low reading (less than 1000) is obtained with both lead connections, the junction is shorted internally or (if junction is measured in a circuit) the junction is shunted by a resistance less than 1KV. In the letter case the junction must be disconnected from the circuit in order to verify its operartion. 2.5 Continuity Measurement 1. Connect red test lead to the “VV” jack and black test lead to the “COM” jack. 2. Set Function / Range switch to the V / / range. 3. Select Continuity range by using Select Button. In the continuty test, the beeper sounds cuntinuously, if the resistance is less than 50V. 3. MAINTENANCE WARNING Remove test leads before changing battery or servicing. Battery Replacement Power is supplied by two 1.5V ‘AAA’ size battery or Equivalent. The “ ” appears on the LCD display when replacement is needed, To replace the battery, remove the screw from the battery cover and lift off the battery case. Remove the batterys & repleced with new batterys. 4. SAFETY INFORMATION The following safety information must be observed to ensure maximum personal safety during the operation of this meter: 1. Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly. 2. This Clamp Meter is designed to take current measurements on circuits with a maximum voltage difference of 500VAC between any conductor and ground potential. Using the instrument for current measurements on circuits above this voltage may cause electric shock, instrument damage or damage to the equipment under test. Before measuring current make certain the test leads are removed from the instrument. 3. The instrument is protected for overload upto 600 VAC for 1 minute. Do not take current readings on circuits where the maximum current potential is not known. Do not exceed the maximum currents that this instrument is designed to measure. 4. Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit. Small amounts of current can be dangerous. 5. Use caution when working above 60V DC or 30V AC rms. Such voltages pose a shock hazard. 6. When using the probes, keep your fingers behind the finger guards on the probes. 7. Measuring voltage which exceeds the limits of the clamp meter may damage the meter and expose the operator to a shock hazard. Always recognize the meter voltage limits as stated on the front of the meter.7 8 65 CERTIFICATE OF CALIBRATION We hereby certify that this product has been calibrated and found to be in accordance with the applicable SPECIFICATIONS and MECO STANDARDS. Accuracies of the standard equipment used in this calibration are traceable to the National Standards. MECO METERS PVT. LTD. Plot No. EL-60, MIDC Electronic Zone, TTC Industrial Area, Mahape, Navi Mumbai - 400710 (INDIA) Tel : 0091-22-27673311-16, 27673300 (Board) Fax : 0091-22-27673310, 27673330 E-mail : sales@mecoinst.com Web : www.mecoinst.com SR. NO : ________________________________ CHECKED BY : ________________________________ DATE : ________________________________ MODEL NO : ________________________________ ®