ī‚§ Artifact is used to describe any part of an image that
does not accurately represent the anatomic structures
present within the subject being evaluated.
ī‚§ In ultrasonography (US), artifacts may cause structures
to appear in an image that are not present
anatomically or a structure that is present anatomically
may be missing from the image. US artifacts may also
show structures as present but incorrect in location,
size, or brightness.
ī‚§ US is prone to numerous imaging artifacts, and these
are commonly encountered in clinical practice.
Artifacts have the potential to interfere with image
interpretation.
ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
ī‚§ Shadowing
ī‚§ Increased Through
Transmission
ī‚§ Mirror Image Artifact
ī‚§ Comet-Tail Artifact
ī‚§ Refraction
ī‚§ Anisotropy
ī‚§ Reverberation Artifact
ī‚§ Reverberation Artifact
ī‚§ Ring-Down Artifact
ī‚§ Dirty Shadowing Artifact
ī‚§ Side Lobe & Grating Lobe
Artifacts
ī‚§ Partial Volume Averaging
ī‚§ Loss or Distortion of
Doppler
Information
ī‚§ Artifactual Vascular Flow
ī‚§ Tissue Vibration Artifact
ī‚§ Aliasing and Velocity
Scale Errors
ī‚§ Spectral Broadening
ī‚§ Blooming Artifact
ī‚§ Twinkle Artifact
PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
SHADOWING
Shadowing results when there is a marked reduction in the
intensity of the ultrasound deep to a strong reflector, attenuator,
or refractor. Clean dark shadows will be seen behind calcified
objects when the focal zone is at or just below the structure
Gallstone With Shadowing Shadowing From Hernia Repair Mesh
Shadowing of Intrauterine Device Coned down view of distal neonatal spine shows shadowing from vertebral bodies that
interrupt the linear appearance of the nerve roots
Shadowing From Ovarian Fibroma. This is a slightly less
“clean” shadow since the fibroma is not as dense as a calcified stone.
Edge shadowing artifact behind the posterior aspect of the fetal skull.
The beam bends at curved surface and loses intensity, producing a
shadow
PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Increased Through Transmission
The hepatic parenchyma distal to the cysts is falsely displayed as increased in echogenicity (arrows)
secondary to increased through-transmission artifact. The cyst attenuates 7 dB less than the normal tissue, and time gain curve correction for
normal tissue results in over amplification of the signals deep to the cyst, producing increased through transmission of these tissues
Increased through transmission occurs when an
object (such as a cyst) attenuates the sound waves
less than the surrounding tissues.
Nebothian cyst
ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
ī‚§ Shadowing
ī‚§ Increased Through
Transmission
ī‚§ Mirror Image Artifact
ī‚§ Comet-Tail Artifact
ī‚§ Refraction
ī‚§ Anisotropy
ī‚§ Reverberation Artifact
ī‚§ Reverberation Artifact
ī‚§ Ring-Down Artifact
ī‚§ Dirty Shadowing Artifact
ī‚§ Side Lobe & Grating Lobe
Artifacts
ī‚§ Partial Volume Averaging
ī‚§ Loss or Distortion of
Doppler
Information
ī‚§ Artifactual Vascular Flow
ī‚§ Tissue Vibration Artifact
ī‚§ Aliasing and Velocity
Scale Errors
ī‚§ Spectral Broadening
ī‚§ Blooming Artifact
ī‚§ Twinkle Artifact
PATH OF SOUND-RELATED ARTIFACTS
Mirror Image Artifact
Mirror image artifact in sonography is seen when there is
a highly reflective surface in the path of the primary
beam. (e.g. diaphragm)
The primary beam reflects from such a surface (e.g diaphragm ) but instead of directly being received by
the transducer, it encounters another structure (e.g. a nodular lesion) in its path and is reflected back to the highly
reflective surface (e.g. diaphragm). It then again reflects back towards the transducer.
The ultrasound machine makes a false assumption that the returning echo has been reflected once and hence the
delayed echoes are judged as if being returned from a deeper structure, thus giving a mirror artifact on the other
side of the reflective surface.
It is a friendly artifact that allows the sonographer to exclude pleural effusion by the reflection of the liver image
through the diaphragm.
Examples:
ī‚§ reflection of a liver lesion into the thorax (the commonest example)
ī‚§ reflection of abdominal ascites mimicking pleural effusion
ī‚§ duplication of gestational sac (either ghost twin or heterotopic pregnancy) 3
ī‚§ duplication of the uterus
To avoid this artifact, change the position and angle of scanning to change the angel of insonation of the primary
ultrasound beam.
Mirror Image Artifact. Soft tissue–gas interfaces, such as the diaphragm, are excellent
reflectors of the sound beam owing to the large difference in the acoustic impedance
between the two materials. Therefore hepatic structures are frequently seen mirrored
beyond the diaphragm onto the lungs. These two figures show liver parenchyma, hepatic
veins in part A, and a hepatic hemangioma in part B all inversely projected onto lung.
PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Comet-Tail Artifact
The comet tail artifact seen when small calcific / crystalline /
highly reflective objects are interrogated and is believed to
be a special form of reverberation artifact.
ī‚§ Small renal or ureteric calculi
ī‚§ Small common bile duct stones
ī‚§ Adenomyomatosis of the gallbladder
ī‚§ Pancreatic calcifications of chronic pancreatitis
ī‚§ Testicular microlithiasis (sometimes)
ī‚§ Thyroid colloid nodules
ī‚§ Identification of foreign bodies
Surgical clips
Catheter, Needle tips
Debris/glass/metal
Gallbladder adenomyomatosis
Comet-Tail Artifact Behind Von Meyenburg Complex
Thyroid colloid cysts
PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
REFRACTION
A change in velocity of the ultrasound beam as it
travels through two adjacent tissues with different
density and elastic properties may produce a
refraction artifact such as from muscle to fat.
Second gestational sac due to refraction artifact “ghosting”
Refraction artifact (arrowheads) is caused by the interfaces between the ruptured tendon edges and hemorrhage.
PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Anisotrophy
Anisotropy is an artifact caused by structures that are
composed of bundles of highly reflective fibers
running parallel to each other, such as tendons and
ligaments
The insertion of supraspinatus tendon
artifactually appears hypoechoic due to anisotropy.
“Peritrigonal blush” region of the white matte
(ellipse) appears relatively echogenic
PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Reverberation Artifact
Reverberation artifact occurs when a strong
reflector runs perpendicular to the direction of the
beam (i.e., parallel to the probe surface), usually
close to the skin surface.
The sound waves may then get partially “trapped” between the reflector and skin, reverberating back and forth and
causing the appearance of multiple regular lines when some of the waves eventually.
Reverberation artifacts arise when the ultrasound signal reflects repeatedly between highly reflective interfaces near
the transducer, resulting in delayed echo return to the transducer. This appears in the image as a series of regularly
spaced echoes at increasing depth reach the probe. Often the reflectors are the skin and subcutaneous
fascia.Reverberation artifact is angle dependent, so moving the transducer slightly will cause a decrease in or
elimination of the artifact.
Reverberation can be helpful during biopsies, to see a needle. If it is distracting, try changing the angle of insonation,
using a different window, or decreasing the gain.
The highly reflective and mirror like surface of the pacemaker (arrow) results in repeated bouncing of sound wave among the probe
surface, pacemaker and skin.
ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
ī‚§ Shadowing
ī‚§ Increased Through
Transmission
ī‚§ Mirror Image Artifact
ī‚§ Comet-Tail Artifact
ī‚§ Refraction
ī‚§ Anisotropy
ī‚§ Reverberation Artifact
ī‚§ Reverberation Artifact
ī‚§ Ring-Down Artifact
ī‚§ Dirty Shadowing Artifact
ī‚§ Side Lobe & Grating Lobe
Artifacts
ī‚§ Partial Volume Averaging
ī‚§ Loss or Distortion of
Doppler
Information
ī‚§ Artifactual Vascular Flow
ī‚§ Tissue Vibration Artifact
ī‚§ Aliasing and Velocity
Scale Errors
ī‚§ Spectral Broadening
ī‚§ Blooming Artifact
ī‚§ Twinkle Artifact
GAS-RELATED ARTIFACTS The physical pressure of the probe on the soft
tissues, such as the abdominal wall, flattens their
shape in the near field, making them perpendicular
to the surface of the probe and the incident
sound beam. Because gases conform to the shape of
their containers, they too will form a linear and
perpendicular interface to the incident sound beam,
especially when in the form of a large bubble. The
highly reflective gas interface will act like a
mirror,resulting in reverberation artifacts.
Reverberation Artifact
Small amount of gas within the
bladder is recognizable by the
reverberation artifact
GAS-RELATED ARTIFACTS
Ring-Down Artifact
ī‚§ Ring-down artifact has been thought to be a variant of
comet tailartifact. This assumption was based on the
often similar appearance
of the two artifacts.
ī‚§ In ring-down artifact, the transmitted ultrasound
energy causes resonant vibrations within fluid trapped
between a tetrahedron of air bubbles. These vibrations
create a continuous sound wave that is transmitted
back to the transducer .
ī‚§ This phenomenon is displayed as a line or series of
parallel bands extending posterior to a gas collection.
Despite the similar sonographic appearance, these two
artifacts have separate mechanisms
Ring-Down Artifact- Pneumobilia causing ring-down artifact in the biliary tree (A) and gallbladder (B).
A B
Ring-down artifact-peripheral ducts of the liver(C) Iatrogenic air in the bladder introduced at cystoscopy
appears as a nondependent bright region (D)
C D
GAS-RELATED ARTIFACTS
Dirty Shadowing Artifact
Dirty shadowing results from a combination of
reflection,reverberation, and ring-down artifacts arising
from multiple variably sized bubbles in foam.
Dirty Shadowing Posterior to Emphysematous Cholecystitis
Dirty Shadowing in the Kidney Due to Emphysematous Pyelonephritis
ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
ī‚§ Shadowing
ī‚§ Increased Through
Transmission
ī‚§ Mirror Image Artifact
ī‚§ Comet-Tail Artifact
ī‚§ Refraction
ī‚§ Anisotropy
ī‚§ Reverberation Artifact
ī‚§ Reverberation Artifact
ī‚§ Ring-Down Artifact
ī‚§ Dirty Shadowing Artifact
ī‚§ Side Lobe & Grating Lobe
Artifacts
ī‚§ Partial Volume Averaging
ī‚§ Loss or Distortion of
Doppler
Information
ī‚§ Artifactual Vascular Flow
ī‚§ Tissue Vibration Artifact
ī‚§ Aliasing and Velocity
Scale Errors
ī‚§ Spectral Broadening
ī‚§ Blooming Artifact
ī‚§ Twinkle Artifact
BEAM PROFILE–RELATED ARTIFACTS
Side Lobe & Grating Lobe Artifacts
Side lobe artifacts occur where side lobes reflect sound from
a strong reflector that is outside of the central beam, and
where the echoes are displayed as if they originated from
within the central beam. Side lobe beams are low-intensity
beams that surround the central beam
Side lobe artifacts are echogenic, linear or curvilinear
artifacts. Strong reflectors include bowel gas adjacent to the
gallbladder or urinary bladder.
Pseudosludge
Echoes seen in the nondependent portion of
the bladder. Note also the shadowing behind
the bladder stone
BEAM PROFILE–RELATED ARTIFACTS
Partial Volume Averaging
The ultrasound beam not only has a complex shape in the imaging
plane but also has a real profile in the third dimension, called
the “elevation plane” or “Z plane.” The ultrasound image appears
as a flat two-dimensional image, but the brightness of each pixel
is representative of the average sum of all the echoes received
within the thickness of the beam in the elevation plane. This
results in echoes being projected in structures that should be
anechoic.
Partial Volume Averaging. The bladder wall is indistinct
in region of this image. It is unclear if there are masses or if this is due
to grating lobe artifact or partial volume artifact or both. Moving the
patient or transducer to avoid bowel gas and putting the focal zone in
the region in question should aid in evaluation. For the area closer to
the patient’s anterior abdominal wall, a higher-frequency transducer
might be helpful
ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
ī‚§ Shadowing
ī‚§ Increased Through
Transmission
ī‚§ Mirror Image Artifact
ī‚§ Comet-Tail Artifact
ī‚§ Refraction
ī‚§ Anisotropy
ī‚§ Reverberation Artifact
ī‚§ Reverberation Artifact
ī‚§ Ring-Down Artifact
ī‚§ Dirty Shadowing Artifact
ī‚§ Side Lobe & Grating Lobe
Artifacts
ī‚§ Partial Volume Averaging
ī‚§ Loss or Distortion of
Doppler
Information
ī‚§ Artifactual Vascular Flow
ī‚§ Tissue Vibration Artifact
ī‚§ Aliasing and Velocity
Scale Errors
ī‚§ Spectral Broadening
ī‚§ Blooming Artifact
ī‚§ Twinkle Artifact
DOPPLER IMAGING ARTIFACTS
Loss or Distortion of Doppler
Information
The appearance of color Doppler signal is affected by several
variables including color-write priority, gray-scale gain, and
pulse repetition frequency. If the color gain is too low, flow
might not be visualized. If the gain is too high, artifactual flow
may be seen in adjacent soft tissues, and thrombus within the
vessel might be missed.
Incorrect gain, wall-filter, and velocity scale can all lead to
loss or distortion of Doppler signal.
Frequency, the rule of thumb is to use a Doppler angle less than
60 degrees (but not 0). Wall filters eliminate low-frequency
noise, but a high setting can lead to loss of signal.
In general, wall filters should be kept at the lowest practical
level, typically in the range of 50 to 100 Hz.
A (PRF = 700 Hz) B (PRF = 4500 Hz)
Artifactual Lack of Flow. Color Doppler image of a carotid artery and jugular vein. In (A), the pulse repetition frequency
(PRF) is 700 Hz and there is aliasing in the carotid artery, but slow flow in the jugular vein is seen. In (B), the PRF is 4500 Hz,
eliminating aliasing in the artery but also suppressing the display of the low Doppler frequencies in the internal jugular vein.
Artifactual Lack of Flow. (A) Color Doppler image shows flow in the portal vein. (B) When a color image is taken to assess
the common duct, no flow is visible in the portal vein. Note the difference in overall gain in image B.
A B
DOPPLER IMAGING ARTIFACTS
Artifactual Vascular Flow
The Doppler effect (shift) is not specific to vascular flow and
occurs with movement of any reflector toward or away from the
ultrasound beam. Fluid or solid tissue motion can mimic
vascular flow. Transmitted pulsations, especially close to the
heart or major arteries, can therefore result in artifactual
appearance of flow within thrombosed veins or avascular areas.
Artifactual vascular flow can also be visualized if the color gain
is too high or the color-write priority is too high.
Color Doppler jets in the bladder use color motion
to indicate that there is flow from each ureteral
orifice
Color flow Doppler signal in pleural fluid with debris. (A) Gray-scale sonogram shows left pleural fluid
with much debris. (B) Color flow Doppler signal
A B
Color Streaking Artifact. (A) This complicated cyst contains floating punctate echoes that move posteriorly
while being scanned,creating a scintillating appearance on gray-scale sonography. (B) Power Doppler
ultrasound pushes the echoes posteriorly faster and with more energy than does the gray-scale beam. The
echoes move fast enough that color persistence creates the appearance of “color streaking” artifact
Flash Artifact From Pulsation. In this image of the left portal vein, note how
transmitted pulsation from the heart causes artifactual flow in the
adjacent liver owing to motion.
DOPPLER IMAGING ARTIFACTS
Tissue Vibration Artifact
The color bruit or tissue vibration artifact is a type of color
Doppler ultrasound artifact which results in color signal
overflowing to the perivascular tissues most often caused by
stenosis, AV fistulas, or shunts. Thus, this artifact is useful by
pinpointing areas of potentially pathological blood flow.
Tissue Vibration Artifact From Common Femoral Artery
to Common Femoral Vein Arteriovenous Fistula. Turbulent
flow through the fistula may affect surrounding tissues,
causing a tissue vibration artifact, which may be the first
clue that an arteriovenous fistula (AVF) is present. Common
femoral artery to common femoral vein
AVF. (A)
Color Doppler shows common femoral artery to common
femoral vein AVF. Note the adjacent tissue vibration artifact
(arrowheads).
Turbulent flow causes vibration of the vessel wall and
perivascular soft tissues, resulting in a detectable doppler
shift which typically results in a mixture of red and blue
speckles .
The color bruit is a beneficial artifact that helps with
localizing areas of critical stenosis or turbulent flow due to
other reasons
Tissue Vibration Artifact. Postbiopsy arteriovenous fistula. Color Doppler image shows markedly increased lower-pole
blood flow (arrows) with adjacent soft tissue color artifact related to tissue vibration (arrowheads)
DOPPLER IMAGING ARTIFACTS
Aliasing and Velocity Scale Errors
Aliasing is an artifact due to undersampling of Doppler
signal, resulting in incorrect estimation of flow velocity
To measure the Doppler frequency shift appropriately by
pulsed Doppler, at least two samples from the cycle are
required (Nyquist limit). If the pulse repetition frequency is
less than twice the maximum
frequency shift produced by movement of the target, aliasing
results. The Doppler signal wraps around either the spectrum
(spectral Doppler) or the color scale (color Doppler). The most
common methods for correcting for aliasing are to shift the
baseline down or up, or increase the pulse repetition
frequency (or velocity scale). A lower-frequency transducer
can also be used to help in correcting for aliasing. As
angulation approaches 90 degrees, directional ambiguity can
occur, suggesting bidirectional flow. Errors in Doppler angle
correction can also lead to misleading assessments of flow
velocity.
Tissue Vibration Artifact. Spectral waveform at site of
superior mesenteric artery stenosis shows very high systolic
velocity, and diastolic velocities of greater than 500 cm/sec.
Color portion of the image shows color bruit consisting of
color outside the vessel near the stenosis site. Color bruit is
believed to be caused by tissue vibration aliasing prevents
precise calculation of the maximum systolic velocity
A B
Aliasing (A) appears in the spectral display as a “wraparound” of the higher frequencies to display below
the baseline. (B) In color Doppler display, aliasing results in a wraparound of the frequency color map
from one flow direction to the opposite of unsaturated color.
Aliasing in Superior Mesenteric Artery (SMA) Due to
High-Velocity Flow
DOPPLER IMAGING ARTIFACTS
Spectral Broadening
Spectral broadening occurs when there are multiple
different velocities of flow within a vessel. This can be a sign
of stenosis.
However, artifactual spectral broadening can occur owing to
improper positioning of the sample volume near the vessel
wall, use of an excessively large sample volume, or excessive
system gain.
Spectral Broadening. The range of velocities
detected at a given time in the pulse cycle is
reflected in the Doppler spectrum
as spectral broadening. (A) Normal spectrum.
Spectral broadening may arise from turbulent
flow in association with vessel stenosis.
(B and C)
Artifactual spectral broadening may be
produced by improper positioning
of the sample volume near the vessel wall, use
of an excessively large sample volume (B), or
excessive system gain (C)
Spectral Broadening. True spectral broadening due to ECA
stenosis.
DOPPLER IMAGING ARTIFACTS
Blooming Artifact
Blooming artifact occurs when color reaches beyond the
vessel wall, making the vessels look larger than expected. It
is gain dependent in that lowering the gain will decrease
blooming. It is also wall filter and color-overwrite
dependent, in that increasing the wall filter will decrease the
artifact
Two images of the same carotid artery. In (A) the gain and wall filter are appropriate to show the flow centrallyin the vessel. In (B) the
gain and wall filter have been changed (in a manner beyond what would be used in clinical imaging) to demonstrate how these can
make the vessel “bloom” and appear larger
DOPPLER IMAGING ARTIFACTS
Twinkle Artifact
Twinkle artifact occurs as a focus of alternating
colors on Doppler signal behind a reflective object
(such as a calculus), which gives the appearance of
turbulent blood flow 2. It appears with or without an
associated color comet tail artifact.
The underlying mechanism of this artifact is thought
to be a result of inherent noise within the ultrasound
scanner, specifically phase (a.k.a. clock) jitter within
the Doppler electronics
Twinkling artifact is more sensitive for detection of
small stones (e.g. urolithiasis,cholelithiasis ) than is
acoustic shadowing.
Ureterovesical junction stone
ULTRASOUND - ARTIFACTS.pptx
ULTRASOUND - ARTIFACTS.pptx

More Related Content

What's hot

Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifactsdonishajohnson
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging MAMTA PANDA
 
Ultrasound Physics
Ultrasound PhysicsUltrasound Physics
Ultrasound Physicsu.surgery
 
Ultrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsUltrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsshariq ahmad shah
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CTSudil Paudyal
 
ultrasound physics
ultrasound physicsultrasound physics
ultrasound physicsRakesh Ca
 
Ultrasound Transducer Types
Ultrasound Transducer TypesUltrasound Transducer Types
Ultrasound Transducer TypesMette Lyng
 
Ultrasound physics
Ultrasound physicsUltrasound physics
Ultrasound physicsAbhilasha Singh
 
Color Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsColor Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsSyed Yousaf Gilani
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundArjun Reddy
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct Muntaser S.Ahmad
 
Principle of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersPrinciple of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersDev Lakhera
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxZubairUlIslam5
 
Principles of Doppler ultrasound
Principles of Doppler ultrasoundPrinciples of Doppler ultrasound
Principles of Doppler ultrasoundSamir Haffar
 
Ct instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationCt instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationSUJAN KARKI
 
Ultrasound & properties
Ultrasound &  propertiesUltrasound &  properties
Ultrasound & propertiesMahesh Kumar
 

What's hot (20)

Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging
 
Usg artifacts
Usg artifactsUsg artifacts
Usg artifacts
 
Ultrasound Physics
Ultrasound PhysicsUltrasound Physics
Ultrasound Physics
 
Ultrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsUltrasound instrumentation practical applications
Ultrasound instrumentation practical applications
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CT
 
ultrasound physics
ultrasound physicsultrasound physics
ultrasound physics
 
Ultrasound Transducer Types
Ultrasound Transducer TypesUltrasound Transducer Types
Ultrasound Transducer Types
 
Ultrasound physics
Ultrasound physicsUltrasound physics
Ultrasound physics
 
Doppler physics by Dr Nani
Doppler physics by Dr NaniDoppler physics by Dr Nani
Doppler physics by Dr Nani
 
Basics Physics of ultrasound
Basics Physics of ultrasoundBasics Physics of ultrasound
Basics Physics of ultrasound
 
Color Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsColor Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physics
 
Usg artifacts
Usg artifactsUsg artifacts
Usg artifacts
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasound
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 
Principle of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersPrinciple of usg imaging, construction of transducers
Principle of usg imaging, construction of transducers
 
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
 
Principles of Doppler ultrasound
Principles of Doppler ultrasoundPrinciples of Doppler ultrasound
Principles of Doppler ultrasound
 
Ct instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationCt instrumentation and types of detector configuration
Ct instrumentation and types of detector configuration
 
Ultrasound & properties
Ultrasound &  propertiesUltrasound &  properties
Ultrasound & properties
 

Similar to ULTRASOUND - ARTIFACTS.pptx

ultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdfultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdfGenesisGeronaga
 
Probe construction, types and usg artefacts
Probe construction, types and usg artefactsProbe construction, types and usg artefacts
Probe construction, types and usg artefactsVishakha Mittal
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxAnayaAnaya14
 
Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pkDr pradeep Kumar
 
Reverberation Artifacts in ultrasound imaging
Reverberation Artifacts in ultrasound imagingReverberation Artifacts in ultrasound imaging
Reverberation Artifacts in ultrasound imagingFranciscoBotelho19
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxKeerthan39
 
M-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.ppt
M-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.pptM-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.ppt
M-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.pptAliHussain4701
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptxShubham661884
 
MRI Artefacts
MRI ArtefactsMRI Artefacts
MRI ArtefactsAnjan Dangal
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound ArtifactAmer Sohail
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects Nusirat Adedewe
 
usg artifacts.pptx
usg artifacts.pptxusg artifacts.pptx
usg artifacts.pptxrohanjohnjacob
 
Ophthalmic ultrasonography
Ophthalmic ultrasonographyOphthalmic ultrasonography
Ophthalmic ultrasonographyRiyad Banayot
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxShubham661884
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifactsshajitha khan
 

Similar to ULTRASOUND - ARTIFACTS.pptx (20)

ultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdfultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdf
 
Probe construction, types and usg artefacts
Probe construction, types and usg artefactsProbe construction, types and usg artefacts
Probe construction, types and usg artefacts
 
Us artefacts
Us artefactsUs artefacts
Us artefacts
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptx
 
Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pk
 
Reverberation Artifacts in ultrasound imaging
Reverberation Artifacts in ultrasound imagingReverberation Artifacts in ultrasound imaging
Reverberation Artifacts in ultrasound imaging
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
M-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.ppt
M-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.pptM-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.ppt
M-2 2018-ag-7688 Diagnostic ultrasonography in domestic animal reproduction.ppt
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
MRI Artefacts
MRI ArtefactsMRI Artefacts
MRI Artefacts
 
Anatomic landmarks
Anatomic landmarksAnatomic landmarks
Anatomic landmarks
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound Artifact
 
Anatomic Landmarks.pptx
Anatomic Landmarks.pptxAnatomic Landmarks.pptx
Anatomic Landmarks.pptx
 
Sillu thesis
Sillu thesisSillu thesis
Sillu thesis
 
Classical signs in oral and maxillofacial radiology.pptx
Classical signs in oral and maxillofacial radiology.pptxClassical signs in oral and maxillofacial radiology.pptx
Classical signs in oral and maxillofacial radiology.pptx
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects
 
usg artifacts.pptx
usg artifacts.pptxusg artifacts.pptx
usg artifacts.pptx
 
Ophthalmic ultrasonography
Ophthalmic ultrasonographyOphthalmic ultrasonography
Ophthalmic ultrasonography
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifacts
 

Recently uploaded

Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000aliya bhat
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...narwatsonia7
 

Recently uploaded (20)

Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Russian Call Girls in Delhi Tanvi ➡ī¸ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡ī¸ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡ī¸ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡ī¸ 9711199012 💋📞 Independent Escort Service...
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Majestic 📞 9907093804 High Profile Service 100% Safe
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
 

ULTRASOUND - ARTIFACTS.pptx

  • 1. DR. DHARMENDER JUNIOR RESIDENT DEPTT. OF RADIODIAGNOSIS PGIMS ROHTAK
  • 2. ī‚§ Artifact is used to describe any part of an image that does not accurately represent the anatomic structures present within the subject being evaluated. ī‚§ In ultrasonography (US), artifacts may cause structures to appear in an image that are not present anatomically or a structure that is present anatomically may be missing from the image. US artifacts may also show structures as present but incorrect in location, size, or brightness. ī‚§ US is prone to numerous imaging artifacts, and these are commonly encountered in clinical practice. Artifacts have the potential to interfere with image interpretation.
  • 3. ARTIFACT PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS PATH OF SOUND- RELATED GAS-RELATED ARTIFACTS BEAM PROFILE– RELATED ARTIFACTS DOPPLER IMAGING ARTIFACTS ī‚§ Shadowing ī‚§ Increased Through Transmission ī‚§ Mirror Image Artifact ī‚§ Comet-Tail Artifact ī‚§ Refraction ī‚§ Anisotropy ī‚§ Reverberation Artifact ī‚§ Reverberation Artifact ī‚§ Ring-Down Artifact ī‚§ Dirty Shadowing Artifact ī‚§ Side Lobe & Grating Lobe Artifacts ī‚§ Partial Volume Averaging ī‚§ Loss or Distortion of Doppler Information ī‚§ Artifactual Vascular Flow ī‚§ Tissue Vibration Artifact ī‚§ Aliasing and Velocity Scale Errors ī‚§ Spectral Broadening ī‚§ Blooming Artifact ī‚§ Twinkle Artifact
  • 4. PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS SHADOWING Shadowing results when there is a marked reduction in the intensity of the ultrasound deep to a strong reflector, attenuator, or refractor. Clean dark shadows will be seen behind calcified objects when the focal zone is at or just below the structure Gallstone With Shadowing Shadowing From Hernia Repair Mesh
  • 5. Shadowing of Intrauterine Device Coned down view of distal neonatal spine shows shadowing from vertebral bodies that interrupt the linear appearance of the nerve roots Shadowing From Ovarian Fibroma. This is a slightly less “clean” shadow since the fibroma is not as dense as a calcified stone. Edge shadowing artifact behind the posterior aspect of the fetal skull. The beam bends at curved surface and loses intensity, producing a shadow
  • 6. PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS Increased Through Transmission The hepatic parenchyma distal to the cysts is falsely displayed as increased in echogenicity (arrows) secondary to increased through-transmission artifact. The cyst attenuates 7 dB less than the normal tissue, and time gain curve correction for normal tissue results in over amplification of the signals deep to the cyst, producing increased through transmission of these tissues Increased through transmission occurs when an object (such as a cyst) attenuates the sound waves less than the surrounding tissues. Nebothian cyst
  • 7. ARTIFACT PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS PATH OF SOUND- RELATED GAS-RELATED ARTIFACTS BEAM PROFILE– RELATED ARTIFACTS DOPPLER IMAGING ARTIFACTS ī‚§ Shadowing ī‚§ Increased Through Transmission ī‚§ Mirror Image Artifact ī‚§ Comet-Tail Artifact ī‚§ Refraction ī‚§ Anisotropy ī‚§ Reverberation Artifact ī‚§ Reverberation Artifact ī‚§ Ring-Down Artifact ī‚§ Dirty Shadowing Artifact ī‚§ Side Lobe & Grating Lobe Artifacts ī‚§ Partial Volume Averaging ī‚§ Loss or Distortion of Doppler Information ī‚§ Artifactual Vascular Flow ī‚§ Tissue Vibration Artifact ī‚§ Aliasing and Velocity Scale Errors ī‚§ Spectral Broadening ī‚§ Blooming Artifact ī‚§ Twinkle Artifact
  • 8. PATH OF SOUND-RELATED ARTIFACTS Mirror Image Artifact Mirror image artifact in sonography is seen when there is a highly reflective surface in the path of the primary beam. (e.g. diaphragm) The primary beam reflects from such a surface (e.g diaphragm ) but instead of directly being received by the transducer, it encounters another structure (e.g. a nodular lesion) in its path and is reflected back to the highly reflective surface (e.g. diaphragm). It then again reflects back towards the transducer. The ultrasound machine makes a false assumption that the returning echo has been reflected once and hence the delayed echoes are judged as if being returned from a deeper structure, thus giving a mirror artifact on the other side of the reflective surface. It is a friendly artifact that allows the sonographer to exclude pleural effusion by the reflection of the liver image through the diaphragm. Examples: ī‚§ reflection of a liver lesion into the thorax (the commonest example) ī‚§ reflection of abdominal ascites mimicking pleural effusion ī‚§ duplication of gestational sac (either ghost twin or heterotopic pregnancy) 3 ī‚§ duplication of the uterus To avoid this artifact, change the position and angle of scanning to change the angel of insonation of the primary ultrasound beam.
  • 9. Mirror Image Artifact. Soft tissue–gas interfaces, such as the diaphragm, are excellent reflectors of the sound beam owing to the large difference in the acoustic impedance between the two materials. Therefore hepatic structures are frequently seen mirrored beyond the diaphragm onto the lungs. These two figures show liver parenchyma, hepatic veins in part A, and a hepatic hemangioma in part B all inversely projected onto lung.
  • 10. PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS Comet-Tail Artifact The comet tail artifact seen when small calcific / crystalline / highly reflective objects are interrogated and is believed to be a special form of reverberation artifact. ī‚§ Small renal or ureteric calculi ī‚§ Small common bile duct stones ī‚§ Adenomyomatosis of the gallbladder ī‚§ Pancreatic calcifications of chronic pancreatitis ī‚§ Testicular microlithiasis (sometimes) ī‚§ Thyroid colloid nodules ī‚§ Identification of foreign bodies Surgical clips Catheter, Needle tips Debris/glass/metal Gallbladder adenomyomatosis
  • 11. Comet-Tail Artifact Behind Von Meyenburg Complex Thyroid colloid cysts
  • 12. PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS REFRACTION A change in velocity of the ultrasound beam as it travels through two adjacent tissues with different density and elastic properties may produce a refraction artifact such as from muscle to fat.
  • 13. Second gestational sac due to refraction artifact “ghosting” Refraction artifact (arrowheads) is caused by the interfaces between the ruptured tendon edges and hemorrhage.
  • 14. PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS Anisotrophy Anisotropy is an artifact caused by structures that are composed of bundles of highly reflective fibers running parallel to each other, such as tendons and ligaments The insertion of supraspinatus tendon artifactually appears hypoechoic due to anisotropy. “Peritrigonal blush” region of the white matte (ellipse) appears relatively echogenic
  • 15. PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS Reverberation Artifact Reverberation artifact occurs when a strong reflector runs perpendicular to the direction of the beam (i.e., parallel to the probe surface), usually close to the skin surface. The sound waves may then get partially “trapped” between the reflector and skin, reverberating back and forth and causing the appearance of multiple regular lines when some of the waves eventually. Reverberation artifacts arise when the ultrasound signal reflects repeatedly between highly reflective interfaces near the transducer, resulting in delayed echo return to the transducer. This appears in the image as a series of regularly spaced echoes at increasing depth reach the probe. Often the reflectors are the skin and subcutaneous fascia.Reverberation artifact is angle dependent, so moving the transducer slightly will cause a decrease in or elimination of the artifact. Reverberation can be helpful during biopsies, to see a needle. If it is distracting, try changing the angle of insonation, using a different window, or decreasing the gain. The highly reflective and mirror like surface of the pacemaker (arrow) results in repeated bouncing of sound wave among the probe surface, pacemaker and skin.
  • 16. ARTIFACT PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS PATH OF SOUND- RELATED GAS-RELATED ARTIFACTS BEAM PROFILE– RELATED ARTIFACTS DOPPLER IMAGING ARTIFACTS ī‚§ Shadowing ī‚§ Increased Through Transmission ī‚§ Mirror Image Artifact ī‚§ Comet-Tail Artifact ī‚§ Refraction ī‚§ Anisotropy ī‚§ Reverberation Artifact ī‚§ Reverberation Artifact ī‚§ Ring-Down Artifact ī‚§ Dirty Shadowing Artifact ī‚§ Side Lobe & Grating Lobe Artifacts ī‚§ Partial Volume Averaging ī‚§ Loss or Distortion of Doppler Information ī‚§ Artifactual Vascular Flow ī‚§ Tissue Vibration Artifact ī‚§ Aliasing and Velocity Scale Errors ī‚§ Spectral Broadening ī‚§ Blooming Artifact ī‚§ Twinkle Artifact
  • 17. GAS-RELATED ARTIFACTS The physical pressure of the probe on the soft tissues, such as the abdominal wall, flattens their shape in the near field, making them perpendicular to the surface of the probe and the incident sound beam. Because gases conform to the shape of their containers, they too will form a linear and perpendicular interface to the incident sound beam, especially when in the form of a large bubble. The highly reflective gas interface will act like a mirror,resulting in reverberation artifacts. Reverberation Artifact Small amount of gas within the bladder is recognizable by the reverberation artifact
  • 18. GAS-RELATED ARTIFACTS Ring-Down Artifact ī‚§ Ring-down artifact has been thought to be a variant of comet tailartifact. This assumption was based on the often similar appearance of the two artifacts. ī‚§ In ring-down artifact, the transmitted ultrasound energy causes resonant vibrations within fluid trapped between a tetrahedron of air bubbles. These vibrations create a continuous sound wave that is transmitted back to the transducer . ī‚§ This phenomenon is displayed as a line or series of parallel bands extending posterior to a gas collection. Despite the similar sonographic appearance, these two artifacts have separate mechanisms
  • 19. Ring-Down Artifact- Pneumobilia causing ring-down artifact in the biliary tree (A) and gallbladder (B). A B Ring-down artifact-peripheral ducts of the liver(C) Iatrogenic air in the bladder introduced at cystoscopy appears as a nondependent bright region (D) C D
  • 20. GAS-RELATED ARTIFACTS Dirty Shadowing Artifact Dirty shadowing results from a combination of reflection,reverberation, and ring-down artifacts arising from multiple variably sized bubbles in foam.
  • 21. Dirty Shadowing Posterior to Emphysematous Cholecystitis Dirty Shadowing in the Kidney Due to Emphysematous Pyelonephritis
  • 22. ARTIFACT PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS PATH OF SOUND- RELATED GAS-RELATED ARTIFACTS BEAM PROFILE– RELATED ARTIFACTS DOPPLER IMAGING ARTIFACTS ī‚§ Shadowing ī‚§ Increased Through Transmission ī‚§ Mirror Image Artifact ī‚§ Comet-Tail Artifact ī‚§ Refraction ī‚§ Anisotropy ī‚§ Reverberation Artifact ī‚§ Reverberation Artifact ī‚§ Ring-Down Artifact ī‚§ Dirty Shadowing Artifact ī‚§ Side Lobe & Grating Lobe Artifacts ī‚§ Partial Volume Averaging ī‚§ Loss or Distortion of Doppler Information ī‚§ Artifactual Vascular Flow ī‚§ Tissue Vibration Artifact ī‚§ Aliasing and Velocity Scale Errors ī‚§ Spectral Broadening ī‚§ Blooming Artifact ī‚§ Twinkle Artifact
  • 23. BEAM PROFILE–RELATED ARTIFACTS Side Lobe & Grating Lobe Artifacts Side lobe artifacts occur where side lobes reflect sound from a strong reflector that is outside of the central beam, and where the echoes are displayed as if they originated from within the central beam. Side lobe beams are low-intensity beams that surround the central beam Side lobe artifacts are echogenic, linear or curvilinear artifacts. Strong reflectors include bowel gas adjacent to the gallbladder or urinary bladder.
  • 24. Pseudosludge Echoes seen in the nondependent portion of the bladder. Note also the shadowing behind the bladder stone
  • 25. BEAM PROFILE–RELATED ARTIFACTS Partial Volume Averaging The ultrasound beam not only has a complex shape in the imaging plane but also has a real profile in the third dimension, called the “elevation plane” or “Z plane.” The ultrasound image appears as a flat two-dimensional image, but the brightness of each pixel is representative of the average sum of all the echoes received within the thickness of the beam in the elevation plane. This results in echoes being projected in structures that should be anechoic. Partial Volume Averaging. The bladder wall is indistinct in region of this image. It is unclear if there are masses or if this is due to grating lobe artifact or partial volume artifact or both. Moving the patient or transducer to avoid bowel gas and putting the focal zone in the region in question should aid in evaluation. For the area closer to the patient’s anterior abdominal wall, a higher-frequency transducer might be helpful
  • 26. ARTIFACT PROPAGATION VELOCITY ARTIFACT ATTENUATION ERRORS PATH OF SOUND- RELATED GAS-RELATED ARTIFACTS BEAM PROFILE– RELATED ARTIFACTS DOPPLER IMAGING ARTIFACTS ī‚§ Shadowing ī‚§ Increased Through Transmission ī‚§ Mirror Image Artifact ī‚§ Comet-Tail Artifact ī‚§ Refraction ī‚§ Anisotropy ī‚§ Reverberation Artifact ī‚§ Reverberation Artifact ī‚§ Ring-Down Artifact ī‚§ Dirty Shadowing Artifact ī‚§ Side Lobe & Grating Lobe Artifacts ī‚§ Partial Volume Averaging ī‚§ Loss or Distortion of Doppler Information ī‚§ Artifactual Vascular Flow ī‚§ Tissue Vibration Artifact ī‚§ Aliasing and Velocity Scale Errors ī‚§ Spectral Broadening ī‚§ Blooming Artifact ī‚§ Twinkle Artifact
  • 27. DOPPLER IMAGING ARTIFACTS Loss or Distortion of Doppler Information The appearance of color Doppler signal is affected by several variables including color-write priority, gray-scale gain, and pulse repetition frequency. If the color gain is too low, flow might not be visualized. If the gain is too high, artifactual flow may be seen in adjacent soft tissues, and thrombus within the vessel might be missed. Incorrect gain, wall-filter, and velocity scale can all lead to loss or distortion of Doppler signal. Frequency, the rule of thumb is to use a Doppler angle less than 60 degrees (but not 0). Wall filters eliminate low-frequency noise, but a high setting can lead to loss of signal. In general, wall filters should be kept at the lowest practical level, typically in the range of 50 to 100 Hz.
  • 28. A (PRF = 700 Hz) B (PRF = 4500 Hz) Artifactual Lack of Flow. Color Doppler image of a carotid artery and jugular vein. In (A), the pulse repetition frequency (PRF) is 700 Hz and there is aliasing in the carotid artery, but slow flow in the jugular vein is seen. In (B), the PRF is 4500 Hz, eliminating aliasing in the artery but also suppressing the display of the low Doppler frequencies in the internal jugular vein.
  • 29. Artifactual Lack of Flow. (A) Color Doppler image shows flow in the portal vein. (B) When a color image is taken to assess the common duct, no flow is visible in the portal vein. Note the difference in overall gain in image B. A B
  • 30. DOPPLER IMAGING ARTIFACTS Artifactual Vascular Flow The Doppler effect (shift) is not specific to vascular flow and occurs with movement of any reflector toward or away from the ultrasound beam. Fluid or solid tissue motion can mimic vascular flow. Transmitted pulsations, especially close to the heart or major arteries, can therefore result in artifactual appearance of flow within thrombosed veins or avascular areas. Artifactual vascular flow can also be visualized if the color gain is too high or the color-write priority is too high. Color Doppler jets in the bladder use color motion to indicate that there is flow from each ureteral orifice
  • 31. Color flow Doppler signal in pleural fluid with debris. (A) Gray-scale sonogram shows left pleural fluid with much debris. (B) Color flow Doppler signal A B
  • 32. Color Streaking Artifact. (A) This complicated cyst contains floating punctate echoes that move posteriorly while being scanned,creating a scintillating appearance on gray-scale sonography. (B) Power Doppler ultrasound pushes the echoes posteriorly faster and with more energy than does the gray-scale beam. The echoes move fast enough that color persistence creates the appearance of “color streaking” artifact
  • 33. Flash Artifact From Pulsation. In this image of the left portal vein, note how transmitted pulsation from the heart causes artifactual flow in the adjacent liver owing to motion.
  • 34. DOPPLER IMAGING ARTIFACTS Tissue Vibration Artifact The color bruit or tissue vibration artifact is a type of color Doppler ultrasound artifact which results in color signal overflowing to the perivascular tissues most often caused by stenosis, AV fistulas, or shunts. Thus, this artifact is useful by pinpointing areas of potentially pathological blood flow. Tissue Vibration Artifact From Common Femoral Artery to Common Femoral Vein Arteriovenous Fistula. Turbulent flow through the fistula may affect surrounding tissues, causing a tissue vibration artifact, which may be the first clue that an arteriovenous fistula (AVF) is present. Common femoral artery to common femoral vein AVF. (A) Color Doppler shows common femoral artery to common femoral vein AVF. Note the adjacent tissue vibration artifact (arrowheads). Turbulent flow causes vibration of the vessel wall and perivascular soft tissues, resulting in a detectable doppler shift which typically results in a mixture of red and blue speckles . The color bruit is a beneficial artifact that helps with localizing areas of critical stenosis or turbulent flow due to other reasons
  • 35. Tissue Vibration Artifact. Postbiopsy arteriovenous fistula. Color Doppler image shows markedly increased lower-pole blood flow (arrows) with adjacent soft tissue color artifact related to tissue vibration (arrowheads)
  • 36. DOPPLER IMAGING ARTIFACTS Aliasing and Velocity Scale Errors Aliasing is an artifact due to undersampling of Doppler signal, resulting in incorrect estimation of flow velocity To measure the Doppler frequency shift appropriately by pulsed Doppler, at least two samples from the cycle are required (Nyquist limit). If the pulse repetition frequency is less than twice the maximum frequency shift produced by movement of the target, aliasing results. The Doppler signal wraps around either the spectrum (spectral Doppler) or the color scale (color Doppler). The most common methods for correcting for aliasing are to shift the baseline down or up, or increase the pulse repetition frequency (or velocity scale). A lower-frequency transducer can also be used to help in correcting for aliasing. As angulation approaches 90 degrees, directional ambiguity can occur, suggesting bidirectional flow. Errors in Doppler angle correction can also lead to misleading assessments of flow velocity. Tissue Vibration Artifact. Spectral waveform at site of superior mesenteric artery stenosis shows very high systolic velocity, and diastolic velocities of greater than 500 cm/sec. Color portion of the image shows color bruit consisting of color outside the vessel near the stenosis site. Color bruit is believed to be caused by tissue vibration aliasing prevents precise calculation of the maximum systolic velocity
  • 37. A B Aliasing (A) appears in the spectral display as a “wraparound” of the higher frequencies to display below the baseline. (B) In color Doppler display, aliasing results in a wraparound of the frequency color map from one flow direction to the opposite of unsaturated color. Aliasing in Superior Mesenteric Artery (SMA) Due to High-Velocity Flow
  • 38. DOPPLER IMAGING ARTIFACTS Spectral Broadening Spectral broadening occurs when there are multiple different velocities of flow within a vessel. This can be a sign of stenosis. However, artifactual spectral broadening can occur owing to improper positioning of the sample volume near the vessel wall, use of an excessively large sample volume, or excessive system gain. Spectral Broadening. The range of velocities detected at a given time in the pulse cycle is reflected in the Doppler spectrum as spectral broadening. (A) Normal spectrum. Spectral broadening may arise from turbulent flow in association with vessel stenosis. (B and C) Artifactual spectral broadening may be produced by improper positioning of the sample volume near the vessel wall, use of an excessively large sample volume (B), or excessive system gain (C)
  • 39. Spectral Broadening. True spectral broadening due to ECA stenosis.
  • 40. DOPPLER IMAGING ARTIFACTS Blooming Artifact Blooming artifact occurs when color reaches beyond the vessel wall, making the vessels look larger than expected. It is gain dependent in that lowering the gain will decrease blooming. It is also wall filter and color-overwrite dependent, in that increasing the wall filter will decrease the artifact Two images of the same carotid artery. In (A) the gain and wall filter are appropriate to show the flow centrallyin the vessel. In (B) the gain and wall filter have been changed (in a manner beyond what would be used in clinical imaging) to demonstrate how these can make the vessel “bloom” and appear larger
  • 41. DOPPLER IMAGING ARTIFACTS Twinkle Artifact Twinkle artifact occurs as a focus of alternating colors on Doppler signal behind a reflective object (such as a calculus), which gives the appearance of turbulent blood flow 2. It appears with or without an associated color comet tail artifact. The underlying mechanism of this artifact is thought to be a result of inherent noise within the ultrasound scanner, specifically phase (a.k.a. clock) jitter within the Doppler electronics Twinkling artifact is more sensitive for detection of small stones (e.g. urolithiasis,cholelithiasis ) than is acoustic shadowing.