X-RAY TUBE & PRODUCTION
OF X-RAYS.
Presented by- Tajdar Amber
Radiology 1st Year Resident
 Major Components Of X-ray Tube
X-RAY TUBE DESIGN
Cathode- Negatively charged
Anode- Positively charged
Glass envelope- Supports the anode and
cathode structure
X-rays are produced by energy conversion when a fast moving
stream of electrons from the cathode is suddenly decelerated in the
target anode of an x-ray tube.
Cathode Assembly consist of
 Filament
 Focusing cup
Filament
 Tungsten wire
 Embedded in a focusing cup which is negatively charged
 Powered by 10-15volts at 4-6 amperes
 Thermionic emission
 Source of electrons
 About 0.2mm in diameter, coiled form a vertical spiral about
0.2cm in diameter and 1cm or less in length
Current
 Filament current- Flow of electrons through the filament in order to raise its
temperature and release electrons.
 X-Ray Tube current- the number of electrons flowing per second from the
filament to the target.(measured in milliamperes)
Advantages of Tungsten filament
o High melting point(3370°C)
o Good thermionic emitter
o Rugged and can be twirled into thin wire
o Less tendency to vaporize
o Stability and strength
o Long life expectancy
o Trace amount of thorium with tungsten is found to be
efficient for electron emission and prolong the life of
filament
Thermionic emission
When current flows through the filament, the heating cause
the release of electrons from the surface of the tungsten
metal wire. The electrons are liberated at a rate
proportional to the filament current.
Therefore, Electrons emitted resulting from the absorption
of thermal energy is known as thermionic emission.
The electron cloud surrounding the filament, produced by
thermionic emission has been termed “The Edison Effect”.
Space charge effect
Electrons emitted from the tungsten filament form a cloud in
vicinity of the filament. This collection of these negatively
charged electrons is called the space charge.
The tendency of space charge is to limit emission of further
electrons from the filament is known as Space charge effect.
Equilibrium state
when a filament is heated to its emission temperature, a
state of equilibrium is quickly reached
 The number of electrons returning to the filament is equal to the
number of electrons being emitted
 Hence, space charge will remain constant depending on the filament
temperature.
X-Ray tubes may be supplied with single or, more commonly, a double filament.
Single filament
Double filament
 Broad focus, larger filament
Higher current, lower resolution, larger area of exposure
 Fine focus, smaller filament
lower current, higher resolution, small area of exposure
Triple focus(three filament)- used for Angiography
Single/Double/Triple Filament Modern X-Ray tube.
Focusing Cup
Focusing cup is concave
Made of Molybdenum nickel alloy
High melting point
Poor thermionic emitter
Converge the electrons to the focal spot on
target.
Supports the filament
Anode
Positively charged
Attracts the electrons when the exposure button is pressed
Made of
Tungsten in all radiographic tubes including digital mammography
 Higher X-Ray production efficiency due to higher atomic number
 Improved heat absorption capacity and rapid dissipation of the heat away from the target area
 High melting point
 Wider dynamic range of digital detectors relaxing the requirement of X-Ray beam
Molybdenum and Rhodium anode targets are used in conventional
mammography
 Characteristic X-rays near 20 keV, desired for breast imaging
Two types- Stationary or Rotating
Stationary anode
 Tungsten(Anode target) is fixed within a large mass of
copper block
 Low output machines
 Eg Dental, Mobile radiography
Disadvantage(or problem with stationary anode)
In spite of tungsten good thermal characteristics, it can’t withstand the heat of
repeated exposure.
Copper base is used to increase the total thermal capacity of the anode and speed
its rate of cooling but copper and tungsten have different coefficients of expansions,
hence, the tungsten target would tend to peel away from the copper anode
Rotating anode X-Ray tube
The purpose of the rotating anode is to spread the heat produced during an
exposure over a large area of the anode, and made it capable of
withstanding the heat generated by large exposures
Tungsten anode target is rotating
High end machine which requires producing high intensity X-rays in short
time
Eg- general purpose X-Ray machines such as DR, Angiography.
Rotating anode assembly consist of
o Anode disk
o Anode stem
o Rotator assembly
o High tension connection from
positive end of the tube circuit
Anode disk
 Tungsten(90%) alloy with rhenium(10%) to avoid cracking and has more resistant to
surface roughening and has a higher thermal capacity compare to pure tungsten’
 Tungsten layered with Molybdenum/Graphite; easier to rotate
(Molybdenum and graphite is lighter and has lesser specific gravity which help to
decrease disc inertia)
 Electron beam interacts with larger area on rotating target and hence heat is not
concentrated at one point; good thermal dissipation
 High speed rotation 3400rpm – 10000rpm
• The rotor is made of copper which is actually part of
a motor that is made to rotate by the device called
Stator
• The stator is a set of electric coils that produces a
very strong magnetic field outside of the glass
envelope that in turn “pulls along” the copper rotor
• The stator causes the rotor to rotate very fast just
like any electric motor. The stator is also called an
Induction motor
• The bearings are very sensitive and if the X-RAY tube
heats up abnormally high, they can be damaged and
can cause the rotor to rotate at less than 3200rpm. If
this occurs, the target surface can be damaged or
“pitted”.
Line Focus Principle
The focal spot is the area of the tungsten target(anode) that is
bombarded by electrons from the cathode.
Most of the energy of electron is converted into heat, with less
than 1% being converted into X-Rays.
therefore, we need a large focal spot to allow greater heat
loading, and the conflicting need for a small focal area to
produce good radiographic detail, were resolved in 1918 with
the development of Line Focus Principle.
Line Focus Principal
o Effective focal spot smaller than the actual focal spot
o Angling the target 5°-20°
o Decrease in target angle or anode angle also decreases the effective
focal spot
Actual focal
spot
Anode Heel Effect
The intensity of the X-Ray beam is greater towards the
cathode side than near the anode
Most of the energy is absorbed at the anode heel
This effect depends on the anode angle
o Steeper the angle more is the effect
o Source to detector distance(SID) should be increased to minimize the
effect
A B
Image A- Patient positioned with thinner region towards the cathode
Image B- Patient positioned with thicker region towards the cathode.
X-Ray photons that are not directed at the focal spot area.
It has no diagnostic value
Low intensity beam
Increases patient radiation dose
Reduces image contrast
Use of lead collimators(exit port of the X-Ray tube) can reduce
the off focus radiation.
Off Focus Radiation
Glass Envelope
Evacuated tube made of borosilicate (Pyrex) to withstand high heat
generated
Encompasses anode and cathode
Good electrical insulator
Long term usage leads to tube arcing (tungsten vaporizes and
forms a thin coating on the inner surface of the glass)
Improvement using metal/ceramic which maintains constant
electrical potential between the electrons and envelope thus
resulting in longer life
Tube Housing
Metal case made of aluminium alloy
Lead lined inside layer protects and supports
the glass tube insert
Oil bath- provide electrical and thermal
insulation
Tube housing provides efficient radiation
barrier, shielding for high voltage
Leakage radiation from tube housing
Limit- not exceeding 114mR/hr (1mGy/hr) at
1meter distance
Cooling Mechanism Of X-Ray Tube
Heat generated in the X-Ray tube is dissipated by radiation from the anode
Radiation- transfer of heat by emission of IR
Conduction- transfer of heat through solid parts of anode
Convection- transfer of heat through oil bath
Tube rating charts
 The maximum temperature that can be safely accepted by an X-
Ray tube is a function of the heat energy produced during the
exposure
o Tungsten can be safely raised to 3000 °C above this level
considerable vaporization of the tungsten target
occurs
 The Heat Unit (HU) is defined as the product of
current(mA) and kVp and time(sec) for single-phase power
supplies
 The safe limit within which an X-Ray tube can be operated
for a single exposure can be easily determined by the
Tube Rating Chart supplied with all X-Ray tube
 Every machine has its own unique chart
 The highest technique factors that are possible without
damaging the target.
Anode Heating and Cooling Curves
 This chart help to determine the length of time the
tube must be allowed to cool before additional
exposure can be made
Eg- if a rapid filming procedure required 20exposures,
the total anode heating will be 100,000 J. It could be
determined from the chart that it would require
approximately 6min for the tube to cool from 100,000 to
10,000 J
 Thermal capacity of anode and its heat dissipation
characteristics.
X-ray Tube & Production.pptx

More Related Content

What's hot

X ray attenuation and scatter radiation
X ray attenuation and scatter radiationX ray attenuation and scatter radiation
X ray attenuation and scatter radiationNISHANT RAJ
 
Interactions of X rays with matter
Interactions of X rays with matterInteractions of X rays with matter
Interactions of X rays with matterSadwala Jr
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiationNitish Rajput
 
Line focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectLine focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectJoaquinLuisRinehart1
 
Production of x rays
Production of x raysProduction of x rays
Production of x raysManish Sandhu
 
interaction of xray with matter
interaction of xray with matterinteraction of xray with matter
interaction of xray with matterdypradio
 
physical interaction of x ray with matter
physical interaction of x ray with matter physical interaction of x ray with matter
physical interaction of x ray with matter charusmita chaudhary
 
Image characteristics of x ray film
Image characteristics of x ray filmImage characteristics of x ray film
Image characteristics of x ray filmYashawant Yadav
 
Conventional X-RAY Tube
 Conventional X-RAY Tube Conventional X-RAY Tube
Conventional X-RAY TubeVibhaJoshi12
 
Basic of x ray production
Basic of x ray productionBasic of x ray production
Basic of x ray productionKhalis Karim
 
The x ray tube
The x ray tubeThe x ray tube
The x ray tubeairwave12
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionNitish Virmani
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...DrKajalLimbad
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Sarin Vidya
 

What's hot (20)

X-Ray Tube
X-Ray TubeX-Ray Tube
X-Ray Tube
 
X ray attenuation and scatter radiation
X ray attenuation and scatter radiationX ray attenuation and scatter radiation
X ray attenuation and scatter radiation
 
X ray generators
X ray generatorsX ray generators
X ray generators
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Interactions of X rays with matter
Interactions of X rays with matterInteractions of X rays with matter
Interactions of X rays with matter
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiation
 
Line focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectLine focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effect
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
interaction of xray with matter
interaction of xray with matterinteraction of xray with matter
interaction of xray with matter
 
physical interaction of x ray with matter
physical interaction of x ray with matter physical interaction of x ray with matter
physical interaction of x ray with matter
 
Image characteristics of x ray film
Image characteristics of x ray filmImage characteristics of x ray film
Image characteristics of x ray film
 
X-ray Production
X-ray ProductionX-ray Production
X-ray Production
 
Conventional X-RAY Tube
 Conventional X-RAY Tube Conventional X-RAY Tube
Conventional X-RAY Tube
 
Basic of x ray production
Basic of x ray productionBasic of x ray production
Basic of x ray production
 
The x ray tube
The x ray tubeThe x ray tube
The x ray tube
 
Grids
GridsGrids
Grids
 
Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolution
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968
 

Similar to X-ray Tube & Production.pptx

Production of x rays chapter -1
Production  of  x   rays  chapter -1Production  of  x   rays  chapter -1
Production of x rays chapter -1sadhamhussain52
 
Xray tubes .pptx
Xray tubes .pptxXray tubes .pptx
Xray tubes .pptxrose25eva
 
Basic principle of x ray
Basic principle of x rayBasic principle of x ray
Basic principle of x rayprahlad maurya
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdfgametheory686sharech
 
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptxPRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptxManojMandal65
 
X ray tube, cassette and screens
X ray tube, cassette and screensX ray tube, cassette and screens
X ray tube, cassette and screensTarun Goyal
 
Production of x rays
Production of x raysProduction of x rays
Production of x rayspraveen kumar
 
Introduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptxIntroduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptxbabaaijaz01
 
X ray tube
X ray tubeX ray tube
X ray tubeRad Tech
 
Production of x rays
Production of x rays Production of x rays
Production of x rays DrMeghaBahal
 
X ray tube and production of xrays
X ray tube and production of xraysX ray tube and production of xrays
X ray tube and production of xraysIndermeet Mangat
 
medical equipment lec 2 x-rays
medical equipment lec 2 x-raysmedical equipment lec 2 x-rays
medical equipment lec 2 x-rayscairo university
 
Ch.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.pptCh.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.pptAbubakar Hammadama
 

Similar to X-ray Tube & Production.pptx (20)

Neelam
NeelamNeelam
Neelam
 
X-RAY TUBE.pptx
X-RAY TUBE.pptxX-RAY TUBE.pptx
X-RAY TUBE.pptx
 
Production of x rays chapter -1
Production  of  x   rays  chapter -1Production  of  x   rays  chapter -1
Production of x rays chapter -1
 
Xray tubes .pptx
Xray tubes .pptxXray tubes .pptx
Xray tubes .pptx
 
X-ray ppt.pptx
X-ray ppt.pptxX-ray ppt.pptx
X-ray ppt.pptx
 
Basic principle of x ray
Basic principle of x rayBasic principle of x ray
Basic principle of x ray
 
Study of Xray machines and functions.pdf
Study of Xray machines and functions.pdfStudy of Xray machines and functions.pdf
Study of Xray machines and functions.pdf
 
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptxPRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
PRODUCTION AND PROPERTIES OF X.pptx BY MANOJ MANDAL(1).pptx
 
X ray tube, cassette and screens
X ray tube, cassette and screensX ray tube, cassette and screens
X ray tube, cassette and screens
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
Introduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptxIntroduction-to-Rotating-Anode-Tube.pptx
Introduction-to-Rotating-Anode-Tube.pptx
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Production of x rays
Production of x rays Production of x rays
Production of x rays
 
X ray tube and production of xrays
X ray tube and production of xraysX ray tube and production of xrays
X ray tube and production of xrays
 
medical equipment lec 2 x-rays
medical equipment lec 2 x-raysmedical equipment lec 2 x-rays
medical equipment lec 2 x-rays
 
X RAY MACHINE.pptx
X RAY MACHINE.pptxX RAY MACHINE.pptx
X RAY MACHINE.pptx
 
Ch.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.pptCh.2-Anatomy of x-ray machine.ppt
Ch.2-Anatomy of x-ray machine.ppt
 
L 2 ct physics
L 2  ct physics L 2  ct physics
L 2 ct physics
 
PHYSICS2.PPT
PHYSICS2.PPTPHYSICS2.PPT
PHYSICS2.PPT
 
PHYSICS2.PPT
PHYSICS2.PPTPHYSICS2.PPT
PHYSICS2.PPT
 

Recently uploaded

(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...indiancallgirl4rent
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls ServiceMiss joya
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...Neha Kaur
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call girls in Ahmedabad High profile
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 

Recently uploaded (20)

(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
Call Girls Service Navi Mumbai Samaira 8617697112 Independent Escort Service ...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 

X-ray Tube & Production.pptx

  • 1. X-RAY TUBE & PRODUCTION OF X-RAYS. Presented by- Tajdar Amber Radiology 1st Year Resident
  • 2.  Major Components Of X-ray Tube X-RAY TUBE DESIGN Cathode- Negatively charged Anode- Positively charged Glass envelope- Supports the anode and cathode structure X-rays are produced by energy conversion when a fast moving stream of electrons from the cathode is suddenly decelerated in the target anode of an x-ray tube.
  • 3. Cathode Assembly consist of  Filament  Focusing cup Filament  Tungsten wire  Embedded in a focusing cup which is negatively charged  Powered by 10-15volts at 4-6 amperes  Thermionic emission  Source of electrons  About 0.2mm in diameter, coiled form a vertical spiral about 0.2cm in diameter and 1cm or less in length Current  Filament current- Flow of electrons through the filament in order to raise its temperature and release electrons.  X-Ray Tube current- the number of electrons flowing per second from the filament to the target.(measured in milliamperes)
  • 4. Advantages of Tungsten filament o High melting point(3370°C) o Good thermionic emitter o Rugged and can be twirled into thin wire o Less tendency to vaporize o Stability and strength o Long life expectancy o Trace amount of thorium with tungsten is found to be efficient for electron emission and prolong the life of filament
  • 5. Thermionic emission When current flows through the filament, the heating cause the release of electrons from the surface of the tungsten metal wire. The electrons are liberated at a rate proportional to the filament current. Therefore, Electrons emitted resulting from the absorption of thermal energy is known as thermionic emission. The electron cloud surrounding the filament, produced by thermionic emission has been termed “The Edison Effect”.
  • 6. Space charge effect Electrons emitted from the tungsten filament form a cloud in vicinity of the filament. This collection of these negatively charged electrons is called the space charge. The tendency of space charge is to limit emission of further electrons from the filament is known as Space charge effect. Equilibrium state when a filament is heated to its emission temperature, a state of equilibrium is quickly reached  The number of electrons returning to the filament is equal to the number of electrons being emitted  Hence, space charge will remain constant depending on the filament temperature.
  • 7. X-Ray tubes may be supplied with single or, more commonly, a double filament. Single filament Double filament  Broad focus, larger filament Higher current, lower resolution, larger area of exposure  Fine focus, smaller filament lower current, higher resolution, small area of exposure Triple focus(three filament)- used for Angiography Single/Double/Triple Filament Modern X-Ray tube.
  • 8. Focusing Cup Focusing cup is concave Made of Molybdenum nickel alloy High melting point Poor thermionic emitter Converge the electrons to the focal spot on target. Supports the filament
  • 9. Anode Positively charged Attracts the electrons when the exposure button is pressed Made of Tungsten in all radiographic tubes including digital mammography  Higher X-Ray production efficiency due to higher atomic number  Improved heat absorption capacity and rapid dissipation of the heat away from the target area  High melting point  Wider dynamic range of digital detectors relaxing the requirement of X-Ray beam Molybdenum and Rhodium anode targets are used in conventional mammography  Characteristic X-rays near 20 keV, desired for breast imaging
  • 10. Two types- Stationary or Rotating Stationary anode  Tungsten(Anode target) is fixed within a large mass of copper block  Low output machines  Eg Dental, Mobile radiography Disadvantage(or problem with stationary anode) In spite of tungsten good thermal characteristics, it can’t withstand the heat of repeated exposure. Copper base is used to increase the total thermal capacity of the anode and speed its rate of cooling but copper and tungsten have different coefficients of expansions, hence, the tungsten target would tend to peel away from the copper anode
  • 11. Rotating anode X-Ray tube The purpose of the rotating anode is to spread the heat produced during an exposure over a large area of the anode, and made it capable of withstanding the heat generated by large exposures Tungsten anode target is rotating High end machine which requires producing high intensity X-rays in short time Eg- general purpose X-Ray machines such as DR, Angiography.
  • 12. Rotating anode assembly consist of o Anode disk o Anode stem o Rotator assembly o High tension connection from positive end of the tube circuit Anode disk  Tungsten(90%) alloy with rhenium(10%) to avoid cracking and has more resistant to surface roughening and has a higher thermal capacity compare to pure tungsten’  Tungsten layered with Molybdenum/Graphite; easier to rotate (Molybdenum and graphite is lighter and has lesser specific gravity which help to decrease disc inertia)  Electron beam interacts with larger area on rotating target and hence heat is not concentrated at one point; good thermal dissipation  High speed rotation 3400rpm – 10000rpm
  • 13. • The rotor is made of copper which is actually part of a motor that is made to rotate by the device called Stator • The stator is a set of electric coils that produces a very strong magnetic field outside of the glass envelope that in turn “pulls along” the copper rotor • The stator causes the rotor to rotate very fast just like any electric motor. The stator is also called an Induction motor • The bearings are very sensitive and if the X-RAY tube heats up abnormally high, they can be damaged and can cause the rotor to rotate at less than 3200rpm. If this occurs, the target surface can be damaged or “pitted”.
  • 14. Line Focus Principle The focal spot is the area of the tungsten target(anode) that is bombarded by electrons from the cathode. Most of the energy of electron is converted into heat, with less than 1% being converted into X-Rays. therefore, we need a large focal spot to allow greater heat loading, and the conflicting need for a small focal area to produce good radiographic detail, were resolved in 1918 with the development of Line Focus Principle. Line Focus Principal o Effective focal spot smaller than the actual focal spot o Angling the target 5°-20° o Decrease in target angle or anode angle also decreases the effective focal spot Actual focal spot
  • 15. Anode Heel Effect The intensity of the X-Ray beam is greater towards the cathode side than near the anode Most of the energy is absorbed at the anode heel This effect depends on the anode angle o Steeper the angle more is the effect o Source to detector distance(SID) should be increased to minimize the effect A B Image A- Patient positioned with thinner region towards the cathode Image B- Patient positioned with thicker region towards the cathode.
  • 16. X-Ray photons that are not directed at the focal spot area. It has no diagnostic value Low intensity beam Increases patient radiation dose Reduces image contrast Use of lead collimators(exit port of the X-Ray tube) can reduce the off focus radiation. Off Focus Radiation
  • 17. Glass Envelope Evacuated tube made of borosilicate (Pyrex) to withstand high heat generated Encompasses anode and cathode Good electrical insulator Long term usage leads to tube arcing (tungsten vaporizes and forms a thin coating on the inner surface of the glass) Improvement using metal/ceramic which maintains constant electrical potential between the electrons and envelope thus resulting in longer life
  • 18. Tube Housing Metal case made of aluminium alloy Lead lined inside layer protects and supports the glass tube insert Oil bath- provide electrical and thermal insulation Tube housing provides efficient radiation barrier, shielding for high voltage Leakage radiation from tube housing Limit- not exceeding 114mR/hr (1mGy/hr) at 1meter distance
  • 19. Cooling Mechanism Of X-Ray Tube Heat generated in the X-Ray tube is dissipated by radiation from the anode Radiation- transfer of heat by emission of IR Conduction- transfer of heat through solid parts of anode Convection- transfer of heat through oil bath
  • 20. Tube rating charts  The maximum temperature that can be safely accepted by an X- Ray tube is a function of the heat energy produced during the exposure o Tungsten can be safely raised to 3000 °C above this level considerable vaporization of the tungsten target occurs  The Heat Unit (HU) is defined as the product of current(mA) and kVp and time(sec) for single-phase power supplies  The safe limit within which an X-Ray tube can be operated for a single exposure can be easily determined by the Tube Rating Chart supplied with all X-Ray tube  Every machine has its own unique chart  The highest technique factors that are possible without damaging the target.
  • 21. Anode Heating and Cooling Curves  This chart help to determine the length of time the tube must be allowed to cool before additional exposure can be made Eg- if a rapid filming procedure required 20exposures, the total anode heating will be 100,000 J. It could be determined from the chart that it would require approximately 6min for the tube to cool from 100,000 to 10,000 J  Thermal capacity of anode and its heat dissipation characteristics.