Oxygen Demand, Chemical with chloride removal 10067


Oxygen Demand, Chemical                                                                                          DOC316.53.01101

Manganese III Reactor Digestion Method
                                                                                                                   Method 10067
(with optional chloride removal)1
30 to 1000 mg/L COD Mn
Scope and Application: For water and wastewater
1   U.S. Patent 5,556,787


          Test preparation



How to use instrument-specific information
                     The Instrument-specific information table displays requirements that may vary between
                     instruments. To use this table, select an instrument then read across to find the corresponding
                     information required to perform this test.

                                        Table 1 Instrument-specific information
    Instrument                                                                                    Light shield

    DR 3900                                                                                         LZV849

    DR 3800, DR 2800, DR 2700                                                                       LZV646



Before starting the test:

    DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before performing this test.

    To determine if the sample contains chloride, use Quantab® Titrator Strips for low range chloride.

    If the sample COD is expected to exceed 1000 mg/L, dilute the sample as described in the Multiplication factors table.

    Homogenize the sample for even distribution of solids and better accuracy and reliability.

    Run one blank with each lot of reagent. Run all samples and blanks with the same lot of vials. The lot number appears on the
    container label.

    The stability of the reagent blank allows for reuse. Verify the reagent blank quality by measuring the absorbance of the blank
    vs. a clean COD vial filled with deionized water. The absorbance range should be about 1.41–1.47.

    If the sample boils during the digestion, the vial is not properly sealed. Test results will be invalid.

    Spilled reagent will affect test accuracy and is hazardous. Do not run tests with spilled vials.

    The maximum range of the VPD gauge is 40 inches of water; it will not indicate the full vacuum level obtained. Full vacuum is
    20–25 inches of mercury; this can be measured at the vacuum pump with a gauge calibrated for inches of mercury.




                                                                                                        Oxygen Demand, Chemical
                                                                                                                    Page 1 of 10
Oxygen Demand, Chemical



Collect the following items:

 Description                                                                                         Quantity

 Blender                                                                                                1

 DRB200 Reactor                                                                                         1

 Forceps, extra fine                                                                                    1

 Light Shield or Adapter (see Instrument-specific information)                                          1

 Manganese III COD Reagent Vials, 20–1000 mg/L COD                                                      1

 Pipet, TenSette® (0.1 to 1.0 mL and 1.0 to 10.0 mL)                                                  1 each

 Pipet Tips for TenSette Pipet                                                                       2 of each

 Sulfuric Acid, concentrated, ACS                                                                      1 mL

 Test Tube Rack                                                                                         1

 Vacuum pretreatment device                                                                             1

 Vacuum Pump                                                                                            1

 Vial, glass, for sample and acid                                                                       2

 Water, deionized                                                                                     varies

 See Consumables and replacement items for reorder information.


Acidified sample preparation




1. Turn on the DRB200            2. Homogenize 100 mL            3. Blank Preparation:       4. Prepared Sample:
Reactor and heat to              of sample for 30 seconds        Pipet 9.0 mL of deionized   Pipet 9.0 mL of
150 °C or set to COD             in a blender.                   water into an empty glass   homogenized sample into
program.                         If suspended solids are         mixing cell.                another empty glass
                                 present, continue to mix                                    mixing cell
                                 the sample while pipetting.




Oxygen Demand, Chemical
Page 2 of 10
Oxygen Demand, Chemical


Acidified sample preparation (continued)




5. Using an automatic               6. Cap the cells tightly                            Proceed to Vacuum
dispenser or TenSette®              and invert several times.                           pretreatment.
Pipet, add 1.0 mL of                The solution will become
concentrated sulfuric acid          hot. Cool to room
to both the sample and the          temperature before
blank.                              proceeding.
                                    Acidified samples are
                                    stable for several months
                                    when refrigerated at 4 °C.


Calculate the multiplication factor
                  All dilutions require that the ratio of sample to sulfuric acid remain at 9:1. For other dilutions that
                  are not listed in the Multiplication factors table, add the sample volume and the deionized water
                  and divide by the sample volume to obtain the multiplication factor.
                  Note: Mixing concentrated sulfuric acid and water is not additive. Adding 1.0 mL of concentrated sulfuric acid
                      to 9.0 mL of sample does not result in a final volume of 10.0 mL. This factor is built into the
                      calibration curve.


                                                     Table 2 Multiplication factors
        Sample (mL)                       Deionized Water (mL)                                 Range (mg/L COD)            Multiplication Factor
            6.0                                           3.0                                           30–1500                     1.5
            3.0                                           6.0                                           60–3000                     3
            1.0                                           8.0                                          180–9000                     9
            0.5                                           8.5                                         360–18,000                    18


                  For best results, use 0.5 mL or more of sample for diluting. If sample values exceed 18,000 mg/L
                  COD, use a separate sample dilution before performing the sample chloride removal procedure.
                  Example: Dilute the sample to a range of 90–4500 mg/L COD.
                      Sample Volume (2.0 mL) + Deionized water (7.0 mL) = Total Volume (9.0 mL)
                                                  Total Volume                            9.0 mL
                      Multiplication Factor = ----------------------------------------- = ----------------- = 4.5
                                                                                      -                   -
                                              Sample Volume                               2.0 mL
                      Standard test range is 50 to 1000 mg/L COD.
                      Example test range = 4.5(50) to 4.5(1000) = 225 to 4500 mg/L COD




                                                                                                                        Oxygen Demand, Chemical
                                                                                                                                    Page 3 of 10
Oxygen Demand, Chemical


Vacuum pretreatment




1. Attach the Vacuum          2. Label each Mn III           3. Place the VPD top on       4. Turn on the vacuum
Pretreatment Device           COD vial and remove the        the base. Insert a fresh      pump and adjust the
(VPD) to a vacuum pump        cap. Insert the vials in the   Chloride Removal              vacuum regulator valve on
(not an aspirator-type        numbered holes in the          Cartridge (CRC) directly      top of the VPD until the
vacuum) that can create a     VPD base.                      above each Mn III COD         internal gauge reads 20
vacuum of 20–25 inches of                                    Reagent Vial. Plug any        inches of water.
mercury.                                                     open holes in the VPD top
                                                             using the stoppers
                                                             provided.




5. Pipet 0.60 mL of           6. Close the vacuum            7. Open the VPD               Proceed to Sample
acidified sample (see         regulator valve completely     regulator valve to release    preparation and
Acidified sample              to achieve full vacuum.        the vacuum. Turn the          measurement.
preparation) into the CRC.    After one minute of full       pump off. Remove the
Pipet 0.60 mL of acidified    vacuum, slide the VPD          VPD top and set it beside
blank into another CRC. It    back and forth several         the base.
should take 30–45             times to dislodge any          Dispose of the used
seconds to draw the liquid    drops clinging to the          Chloride Removal
through the CRC into each     cartridge.                     Cartridge. Do not reuse it.
vial.
If the sample does not flow
through the Chloride
Removal Cartridge (CRC),
increase the vacuum until
flow starts, then reduce
the vacuum down to 20
inches of water. Proceed
as usual.




Oxygen Demand, Chemical
Page 4 of 10
Oxygen Demand, Chemical


Sample preparation and measurement




1. Use forceps to remove       2. Insert each filter in the   3. Remove the Mn III          4. Invert the vials several
the filter from the top of     corresponding Mn III COD       COD vial from the vacuum      times to mix.
each CRC.                      Vial. (Use numbers on the      chamber and replace the
If the sample does not         VPD as a guide.)               original cap. Screw the
contain suspended solids,      To avoid cross-                cap on tightly.
it is not necessary to         contamination between
transfer the filter to the     samples, clean forcep tips
digestion vial.                between samples by
                               wiping with a clean towel
                               or rinsing with deionized
                               water.




5. Insert the vials in the     6. Insert the vials in a       7. Cool the vials to room     8. Remove the vials from
DRB200 Reactor at              cooling rack for two           temperature in a cool         the water and wipe with a
150 °C. Close the              minutes.                       water bath or with running    clean, dry paper towel.
protective cover. Digest for   If the solution develops a     tap water for several
one hour.                      colorless upper layer and      minutes.
To oxidize resistant           a purple lower layer, invert
organics, samples can be       the vial several times to
digested for up to four        mix and proceed to the
hours. Digest the blank for    next step.
the same time period as
the samples.




                                                                                           Oxygen Demand, Chemical
                                                                                                       Page 5 of 10
Oxygen Demand, Chemical


Sample preparation and measurement (continued)



     Stored Programs

      432 COD Mn III
                                                                          Zero
           Start




9. Select the test.             10. Insert the blank into      11. ZERO the instrument.       12. Make sure the filter
Insert an adapter or a light    the 16-mm cell holder.         The display will show:         disc is not suspended in
shield if required (see                                                                       the middle of the vial; it
                                                                   0 mg/L COD Mn              can interfere with the
Instrument-specific
information).                                                                                 instrument reading.
                                                                                              The disc must be more
                                                                                              than 20 mm (0.8") or less
                                                                                              than 10 mm (0.4"), from
                                                                                              the bottom of the vial.
                                                                                              Move it by gently swirling
                                                                                              or by lightly tapping the
                                                                                              vial on the table top.




                                           Read




13. Wipe the prepared           14. READ the results in
sample and insert it into       mg/L COD Mn.
the .



Interferences
                   Inorganic materials may also be oxidized by trivalent manganese and constitute a positive
                   interference when present in significant amounts. Chloride is the most common interference and is
                   removed by sample pretreatment with the Chloride Removal Cartridge. If chloride is known to be
                   absent or present in insignificant levels, the pretreatment can be omitted. A simple way to
                   determine if chloride will affect test results is to run routine samples with and without the chloride
                   removal, then compare results. Other inorganic interferences (i.e., nitrite, ferrous iron, sulfide) are
                   not usually present in significant amounts. If necessary, these interferences can be corrected after
                   determining their concentrations with separate methods and adjusting the final COD test results
                   accordingly.
                   Ammonia nitrogen is known to interfere in the presence of chloride; it does not interfere if chloride
                   is absent.




Oxygen Demand, Chemical
Page 6 of 10
Oxygen Demand, Chemical


Sample collection, preservation and storage
               •   Collect samples in clean glass bottles.
               •   Use plastic bottles only if they are known to be free of organic contamination.
               •   Test biologically active samples as soon as possible.
               •   Homogenize samples containing solids to assure representative samples.
               •   Samples treated with concentrated sulfuric acid to a pH of less than 2 (about 2 mL per liter)
                   and refrigerated at 4 °C may be stored up to 28 days.
               •   Correct test results for volume additions.

Accuracy check
               Standard solution method
               Note: Refer to the instrument user manual for specific software navigation instructions.

               Required for accuracy check:
               •   COD standard solution, 800 mg/L

               1. Use 0.60 mL of the 800-mg/L solution in place of the sample.

               2. To adjust the calibration curve using the reading obtained with the standard solution, navigate
                  to Standard Adjust in the software: OPTIONS>MORE>STANDARD ADJUST.

               3. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate
                  concentration is used, enter the concentration and adjust the curve to that value.

Method performance

                                                            Precision—95% Confidence         Sensitivity—∆Concentration
 Program      Instrument               Standard
                                                               Limits of Distribution              per 0.010 ∆Abs
   432         DR 5000              600 mg/L COD                 576–624 mg/L COD                     8 mg/L COD

Estimated detection limit (EDL)
              The EDL for program 432 is 4 mg/L COD. The EDL is the calculated lowest average concentration
              in a deionized water matrix that is different from zero with a 99% level of confidence.

Summary of method
               Chemical Oxygen Demand (COD) is defined as “... a measure of the oxygen equivalent of the
               organic matter content of a sample that is susceptible to oxidation by a strong chemical oxidant”
               (APHA Standard Methods, 19th ed., 1995). Trivalent manganese is a strong, non-carcinogenic
               chemical oxidant that changes quantitatively from purple to colorless when it reacts with organic
               matter. It typically oxidizes about 80% of the organic compounds. Studies have shown that the
               reactions are highly reproducible and test results correlate closely to Biochemical Oxygen
               Demand (BOD) values and hexavalent chromium COD tests. None of the oxygen demand tests
               provide 100% oxidation of all organic compounds.
               A calibration is provided which is based on the oxidation of Potassium Acid Phthalate (KHP). A
               different response may be seen in analyzing various wastewaters. The KHP calibration is
               adequate for most applications. The highest degree of accuracy is obtained when test results are
               correlated to a standard reference method such as BOD or one of the chromium COD methods.
               Special waste streams or classes will require a separate calibration to obtain a direct
               mg/L COD reading or to generate a correction factor for the precalibrated KHP response. The



                                                                                            Oxygen Demand, Chemical
                                                                                                        Page 7 of 10
Oxygen Demand, Chemical

                   sample digestion time can be extended up to four hours for samples that are difficult to oxidize.
                   Test results are measured at 510 nm.

Consumables and replacement items

Required reagents
Description                                                           Quantity/Test       Unit       Catalog number
Manganese III COD Reagent Vials, 20–1000 mg/L COD                           1            25/pkg          2623425
Chloride Removal Cartridge (CRC)                                            1            25/pkg          2661825
Sulfuric Acid, concentrated, ACS                                           1 mL           2.5 L           97909
Water, deionized                                                          varies           4L             27256



Required apparatus
Description                                                           Quantity/Test        Unit       Catalog number
Blender, 120 VAC                                                            1             each           2616100
Cap, with inert Teflon liner, for mixing bottle                           varies          12/pkg         2401812
DRB200 Reactor, 110 V, 15 x 16 mm                                           1             each       LTV082.53.40001
DRB200 Reactor, 220 V, 15 x 16 mm                                           1             each       LTV082.52.40001
Forceps, extra fine point                                                   1             each           2669600
Pipet, TenSette®, 1.0 to 10.0 mL                                            1             each           1970010
Pipet Tips for TenSette Pipet 19700-10                                      2             50/pkg         2199796
Pipet, TenSette, 0.1 to 1.0 mL                                              1             each           1970001
Pipet Tips for TenSette Pipet 19700-01                                      2             50/pkg         2185696
Test Tube Rack                                                              1             each           1864100
Vacuum Pretreatment Device (VPD)                                            1             each           4900000
Vacuum Pump, 1.2 CFM 115V                                                   1             each           2824800
Vial, glass, for sample plus acid                                           2             each           2427700



Recommended standards
Description                                                                                Unit      Catalog number
COD Standard Solution, 800-mg/L COD                                                      200 mL          2672629
Oxygen Demand Standard for BOD, COD, TOC, 10-mL ampules                                  16/pkg          2833510
Potassium Acid Phthalate, ACS                                                             500 g           31534
Wastewater Standard, Influent Inorganics, for NH3–N, NO3–N, PO4, COD, SO4, TOC           500 mL          2833149



Optional reagents and apparatus
Description                                                                                Unit       Catalog number
Dispenser, automatic, 1.0–5.0 mL                                                          each           2563137
Titrator Strips, Quantab®, for low range chloride                                        40 tests        2744940
Finger cots                                                                               2/pkg          1464702
Flask, volumetric class A, 1 L                                                            each           1457453



Oxygen Demand, Chemical
Page 8 of 10
Oxygen Demand, Chemical

Optional reagents and apparatus
Description                                         Unit     Catalog number
pH Paper, 0–14 pH range                           100/pkg       2601300
Balance, analytical, 80 g capacity, 115 VAC        each         2936701
Weighing Paper, 76x76 mm                          500/pkg       1473800
Oven, Laboratory 120 VAC/60Hz                      each         1428900
Oven, Laboratory 240 VAC/50Hz                      each         1428902
COD Standard Solution, 300-mg/L                   200 mL        1218629
COD Standard Solution, 300-mg/L                   500 mL        1218649
COD Standard Solution, 1000-mg/L                  200 mL        2253929
Standard Methods book, most recent edition         each         2270800
Pipet Tips for TenSette Pipet 19700-10            250/pkg       2199725
Pipet Tips for TenSette Pipet 19700-01            1000/pkg      2185628




                                                  Oxygen Demand, Chemical
                                                              Page 9 of 10
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:                  HACH COMPANY
                  In the U.S.A. – Call toll-free 800-227-4224                                WORLD HEADQUARTERS
                  Outside the U.S.A. – Contact the HACH office or distributor serving you.   Telephone: (970) 669-3050
                  On the Worldwide Web – www.hach.com; E-mail – techhelp@hach.com            FAX: (970) 669-2932

© Hach Company, 2007, 2010. All rights reserved. Printed in the U.S.A.                                     Edition 6

More Related Content

What's hot

Analysis Of Atrazine Using DLLME and HS-LPME
Analysis Of Atrazine Using DLLME and HS-LPMEAnalysis Of Atrazine Using DLLME and HS-LPME
Analysis Of Atrazine Using DLLME and HS-LPMEEclecticElegance
 
Steroid radioimmunoassay
Steroid radioimmunoassaySteroid radioimmunoassay
Steroid radioimmunoassayTeuku Salman
 
Developing a discriminating dissolution
Developing a discriminating dissolutionDeveloping a discriminating dissolution
Developing a discriminating dissolutionV.Maheswaran ,
 
Ib biology lab_report_sample
Ib biology lab_report_sampleIb biology lab_report_sample
Ib biology lab_report_samplenpopova
 
Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...SriramNagarajan18
 
Research Technique Presentation
Research Technique PresentationResearch Technique Presentation
Research Technique PresentationEclecticElegance
 
Reveal Q+ for Fumonisin
Reveal Q+ for FumonisinReveal Q+ for Fumonisin
Reveal Q+ for Fumonisinleprecoilin386
 
D 548 – 97 ;rdu0 oc05nw _
D 548 – 97  ;rdu0 oc05nw _D 548 – 97  ;rdu0 oc05nw _
D 548 – 97 ;rdu0 oc05nw _hamfrost
 
Introduction to qu e chers
Introduction to qu e chersIntroduction to qu e chers
Introduction to qu e chersPradeep Kumar
 
Determine the content of vitamin C
Determine the content of vitamin CDetermine the content of vitamin C
Determine the content of vitamin COng Chee Kiong
 
Diffusion lab
Diffusion labDiffusion lab
Diffusion lab14771
 

What's hot (19)

water analysis sop
water analysis sopwater analysis sop
water analysis sop
 
Analysis Of Atrazine Using DLLME and HS-LPME
Analysis Of Atrazine Using DLLME and HS-LPMEAnalysis Of Atrazine Using DLLME and HS-LPME
Analysis Of Atrazine Using DLLME and HS-LPME
 
Lab Report 3
Lab Report 3Lab Report 3
Lab Report 3
 
Steroid radioimmunoassay
Steroid radioimmunoassaySteroid radioimmunoassay
Steroid radioimmunoassay
 
Developing a discriminating dissolution
Developing a discriminating dissolutionDeveloping a discriminating dissolution
Developing a discriminating dissolution
 
Ib biology lab_report_sample
Ib biology lab_report_sampleIb biology lab_report_sample
Ib biology lab_report_sample
 
Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...Stability indicating method development and validation for the estimation of ...
Stability indicating method development and validation for the estimation of ...
 
Sop water analysis-2
Sop water analysis-2Sop water analysis-2
Sop water analysis-2
 
googa
googagooga
googa
 
buffer preparation
buffer preparationbuffer preparation
buffer preparation
 
adfb[oadf
adfb[oadfadfb[oadf
adfb[oadf
 
Research Technique Presentation
Research Technique PresentationResearch Technique Presentation
Research Technique Presentation
 
Reveal Q+ for Fumonisin
Reveal Q+ for FumonisinReveal Q+ for Fumonisin
Reveal Q+ for Fumonisin
 
D 548 – 97 ;rdu0 oc05nw _
D 548 – 97  ;rdu0 oc05nw _D 548 – 97  ;rdu0 oc05nw _
D 548 – 97 ;rdu0 oc05nw _
 
Dissolution
DissolutionDissolution
Dissolution
 
Introduction to qu e chers
Introduction to qu e chersIntroduction to qu e chers
Introduction to qu e chers
 
Determine the content of vitamin C
Determine the content of vitamin CDetermine the content of vitamin C
Determine the content of vitamin C
 
Diffusion lab
Diffusion labDiffusion lab
Diffusion lab
 
Don
DonDon
Don
 

Similar to Biolife

Chemical Oxygen Demand
Chemical Oxygen DemandChemical Oxygen Demand
Chemical Oxygen DemandMuhd Effandi
 
Calibration of analytical instruments
Calibration of analytical instrumentsCalibration of analytical instruments
Calibration of analytical instrumentsVigneshVicky470
 
Miscellaneous test .pdf
Miscellaneous test .pdfMiscellaneous test .pdf
Miscellaneous test .pdfUVAS
 
Water test methods
Water test methodsWater test methods
Water test methodsSoumen Basu
 
Reducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC AnalysesReducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC AnalysesShimadzu Scientific Instruments
 
Lab1 spectrophotometry
Lab1 spectrophotometryLab1 spectrophotometry
Lab1 spectrophotometryCody Taylor
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxAsifSHAIKH50040
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxAsif Shaikh
 
Indophenol Method
Indophenol MethodIndophenol Method
Indophenol MethodECRD IN
 
Indophenol method
Indophenol methodIndophenol method
Indophenol methodECRD2015
 
Water Analysis using LAMBDA Spectrophotometer
Water Analysis using LAMBDA SpectrophotometerWater Analysis using LAMBDA Spectrophotometer
Water Analysis using LAMBDA SpectrophotometerPerkinElmer, Inc.
 
BENFIELD LIQUOR:Determination of Diethanolamine Using an Auto Titrator
BENFIELD LIQUOR:Determination of Diethanolamine Using an Auto TitratorBENFIELD LIQUOR:Determination of Diethanolamine Using an Auto Titrator
BENFIELD LIQUOR:Determination of Diethanolamine Using an Auto TitratorGerard B. Hawkins
 
Water analysis and testing
Water analysis and testingWater analysis and testing
Water analysis and testingNEHA MISHRA
 
all sops combined pdf.pdf
all sops combined pdf.pdfall sops combined pdf.pdf
all sops combined pdf.pdfAsifSHAIKH50040
 
all sops combined pdf.pdf
all sops combined pdf.pdfall sops combined pdf.pdf
all sops combined pdf.pdfAsif Shaikh
 
Men_ct25_0202.pdf SWAN manual swiss control
Men_ct25_0202.pdf SWAN manual swiss controlMen_ct25_0202.pdf SWAN manual swiss control
Men_ct25_0202.pdf SWAN manual swiss controlApplySofttech
 

Similar to Biolife (20)

Chemical Oxygen Demand
Chemical Oxygen DemandChemical Oxygen Demand
Chemical Oxygen Demand
 
Calibration of analytical instruments
Calibration of analytical instrumentsCalibration of analytical instruments
Calibration of analytical instruments
 
Miscellaneous test .pdf
Miscellaneous test .pdfMiscellaneous test .pdf
Miscellaneous test .pdf
 
Water test methods
Water test methodsWater test methods
Water test methods
 
Tester and rapid kit for analysis catalog
Tester and rapid kit for analysis catalogTester and rapid kit for analysis catalog
Tester and rapid kit for analysis catalog
 
Reducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC AnalysesReducing the Effects of Sample Solvent on UHPLC Analyses
Reducing the Effects of Sample Solvent on UHPLC Analyses
 
Micro chemlab
Micro chemlabMicro chemlab
Micro chemlab
 
Micro chem
Micro chemMicro chem
Micro chem
 
Lab1 spectrophotometry
Lab1 spectrophotometryLab1 spectrophotometry
Lab1 spectrophotometry
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptx
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptx
 
Colorimeter
ColorimeterColorimeter
Colorimeter
 
Indophenol Method
Indophenol MethodIndophenol Method
Indophenol Method
 
Indophenol method
Indophenol methodIndophenol method
Indophenol method
 
Water Analysis using LAMBDA Spectrophotometer
Water Analysis using LAMBDA SpectrophotometerWater Analysis using LAMBDA Spectrophotometer
Water Analysis using LAMBDA Spectrophotometer
 
BENFIELD LIQUOR:Determination of Diethanolamine Using an Auto Titrator
BENFIELD LIQUOR:Determination of Diethanolamine Using an Auto TitratorBENFIELD LIQUOR:Determination of Diethanolamine Using an Auto Titrator
BENFIELD LIQUOR:Determination of Diethanolamine Using an Auto Titrator
 
Water analysis and testing
Water analysis and testingWater analysis and testing
Water analysis and testing
 
all sops combined pdf.pdf
all sops combined pdf.pdfall sops combined pdf.pdf
all sops combined pdf.pdf
 
all sops combined pdf.pdf
all sops combined pdf.pdfall sops combined pdf.pdf
all sops combined pdf.pdf
 
Men_ct25_0202.pdf SWAN manual swiss control
Men_ct25_0202.pdf SWAN manual swiss controlMen_ct25_0202.pdf SWAN manual swiss control
Men_ct25_0202.pdf SWAN manual swiss control
 

Recently uploaded

How to convert PDF to text with Nanonets
How to convert PDF to text with NanonetsHow to convert PDF to text with Nanonets
How to convert PDF to text with Nanonetsnaman860154
 
GenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationGenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationMichael W. Hawkins
 
[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdfhans926745
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsMark Billinghurst
 
Slack Application Development 101 Slides
Slack Application Development 101 SlidesSlack Application Development 101 Slides
Slack Application Development 101 Slidespraypatel2
 
The Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptxThe Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptxMalak Abu Hammad
 
How to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected WorkerHow to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected WorkerThousandEyes
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitecturePixlogix Infotech
 
Factors to Consider When Choosing Accounts Payable Services Providers.pptx
Factors to Consider When Choosing Accounts Payable Services Providers.pptxFactors to Consider When Choosing Accounts Payable Services Providers.pptx
Factors to Consider When Choosing Accounts Payable Services Providers.pptxKatpro Technologies
 
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersEnhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersThousandEyes
 
SIEMENS: RAPUNZEL – A Tale About Knowledge Graph
SIEMENS: RAPUNZEL – A Tale About Knowledge GraphSIEMENS: RAPUNZEL – A Tale About Knowledge Graph
SIEMENS: RAPUNZEL – A Tale About Knowledge GraphNeo4j
 
Pigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions
 
Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)Allon Mureinik
 
08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking MenDelhi Call girls
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Patryk Bandurski
 
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Alan Dix
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 3652toLead Limited
 
How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?XfilesPro
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptxHampshireHUG
 
IAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI SolutionsIAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI SolutionsEnterprise Knowledge
 

Recently uploaded (20)

How to convert PDF to text with Nanonets
How to convert PDF to text with NanonetsHow to convert PDF to text with Nanonets
How to convert PDF to text with Nanonets
 
GenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day PresentationGenCyber Cyber Security Day Presentation
GenCyber Cyber Security Day Presentation
 
[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf[2024]Digital Global Overview Report 2024 Meltwater.pdf
[2024]Digital Global Overview Report 2024 Meltwater.pdf
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR Systems
 
Slack Application Development 101 Slides
Slack Application Development 101 SlidesSlack Application Development 101 Slides
Slack Application Development 101 Slides
 
The Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptxThe Codex of Business Writing Software for Real-World Solutions 2.pptx
The Codex of Business Writing Software for Real-World Solutions 2.pptx
 
How to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected WorkerHow to Troubleshoot Apps for the Modern Connected Worker
How to Troubleshoot Apps for the Modern Connected Worker
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC Architecture
 
Factors to Consider When Choosing Accounts Payable Services Providers.pptx
Factors to Consider When Choosing Accounts Payable Services Providers.pptxFactors to Consider When Choosing Accounts Payable Services Providers.pptx
Factors to Consider When Choosing Accounts Payable Services Providers.pptx
 
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersEnhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
 
SIEMENS: RAPUNZEL – A Tale About Knowledge Graph
SIEMENS: RAPUNZEL – A Tale About Knowledge GraphSIEMENS: RAPUNZEL – A Tale About Knowledge Graph
SIEMENS: RAPUNZEL – A Tale About Knowledge Graph
 
Pigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping ElbowsPigging Solutions Piggable Sweeping Elbows
Pigging Solutions Piggable Sweeping Elbows
 
Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)
 
08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men08448380779 Call Girls In Civil Lines Women Seeking Men
08448380779 Call Girls In Civil Lines Women Seeking Men
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
 
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
 
How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?How to Remove Document Management Hurdles with X-Docs?
How to Remove Document Management Hurdles with X-Docs?
 
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
04-2024-HHUG-Sales-and-Marketing-Alignment.pptx
 
IAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI SolutionsIAC 2024 - IA Fast Track to Search Focused AI Solutions
IAC 2024 - IA Fast Track to Search Focused AI Solutions
 

Biolife

  • 1. Oxygen Demand, Chemical with chloride removal 10067 Oxygen Demand, Chemical DOC316.53.01101 Manganese III Reactor Digestion Method Method 10067 (with optional chloride removal)1 30 to 1000 mg/L COD Mn Scope and Application: For water and wastewater 1 U.S. Patent 5,556,787 Test preparation How to use instrument-specific information The Instrument-specific information table displays requirements that may vary between instruments. To use this table, select an instrument then read across to find the corresponding information required to perform this test. Table 1 Instrument-specific information Instrument Light shield DR 3900 LZV849 DR 3800, DR 2800, DR 2700 LZV646 Before starting the test: DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before performing this test. To determine if the sample contains chloride, use Quantab® Titrator Strips for low range chloride. If the sample COD is expected to exceed 1000 mg/L, dilute the sample as described in the Multiplication factors table. Homogenize the sample for even distribution of solids and better accuracy and reliability. Run one blank with each lot of reagent. Run all samples and blanks with the same lot of vials. The lot number appears on the container label. The stability of the reagent blank allows for reuse. Verify the reagent blank quality by measuring the absorbance of the blank vs. a clean COD vial filled with deionized water. The absorbance range should be about 1.41–1.47. If the sample boils during the digestion, the vial is not properly sealed. Test results will be invalid. Spilled reagent will affect test accuracy and is hazardous. Do not run tests with spilled vials. The maximum range of the VPD gauge is 40 inches of water; it will not indicate the full vacuum level obtained. Full vacuum is 20–25 inches of mercury; this can be measured at the vacuum pump with a gauge calibrated for inches of mercury. Oxygen Demand, Chemical Page 1 of 10
  • 2. Oxygen Demand, Chemical Collect the following items: Description Quantity Blender 1 DRB200 Reactor 1 Forceps, extra fine 1 Light Shield or Adapter (see Instrument-specific information) 1 Manganese III COD Reagent Vials, 20–1000 mg/L COD 1 Pipet, TenSette® (0.1 to 1.0 mL and 1.0 to 10.0 mL) 1 each Pipet Tips for TenSette Pipet 2 of each Sulfuric Acid, concentrated, ACS 1 mL Test Tube Rack 1 Vacuum pretreatment device 1 Vacuum Pump 1 Vial, glass, for sample and acid 2 Water, deionized varies See Consumables and replacement items for reorder information. Acidified sample preparation 1. Turn on the DRB200 2. Homogenize 100 mL 3. Blank Preparation: 4. Prepared Sample: Reactor and heat to of sample for 30 seconds Pipet 9.0 mL of deionized Pipet 9.0 mL of 150 °C or set to COD in a blender. water into an empty glass homogenized sample into program. If suspended solids are mixing cell. another empty glass present, continue to mix mixing cell the sample while pipetting. Oxygen Demand, Chemical Page 2 of 10
  • 3. Oxygen Demand, Chemical Acidified sample preparation (continued) 5. Using an automatic 6. Cap the cells tightly Proceed to Vacuum dispenser or TenSette® and invert several times. pretreatment. Pipet, add 1.0 mL of The solution will become concentrated sulfuric acid hot. Cool to room to both the sample and the temperature before blank. proceeding. Acidified samples are stable for several months when refrigerated at 4 °C. Calculate the multiplication factor All dilutions require that the ratio of sample to sulfuric acid remain at 9:1. For other dilutions that are not listed in the Multiplication factors table, add the sample volume and the deionized water and divide by the sample volume to obtain the multiplication factor. Note: Mixing concentrated sulfuric acid and water is not additive. Adding 1.0 mL of concentrated sulfuric acid to 9.0 mL of sample does not result in a final volume of 10.0 mL. This factor is built into the calibration curve. Table 2 Multiplication factors Sample (mL) Deionized Water (mL) Range (mg/L COD) Multiplication Factor 6.0 3.0 30–1500 1.5 3.0 6.0 60–3000 3 1.0 8.0 180–9000 9 0.5 8.5 360–18,000 18 For best results, use 0.5 mL or more of sample for diluting. If sample values exceed 18,000 mg/L COD, use a separate sample dilution before performing the sample chloride removal procedure. Example: Dilute the sample to a range of 90–4500 mg/L COD. Sample Volume (2.0 mL) + Deionized water (7.0 mL) = Total Volume (9.0 mL) Total Volume 9.0 mL Multiplication Factor = ----------------------------------------- = ----------------- = 4.5 - - Sample Volume 2.0 mL Standard test range is 50 to 1000 mg/L COD. Example test range = 4.5(50) to 4.5(1000) = 225 to 4500 mg/L COD Oxygen Demand, Chemical Page 3 of 10
  • 4. Oxygen Demand, Chemical Vacuum pretreatment 1. Attach the Vacuum 2. Label each Mn III 3. Place the VPD top on 4. Turn on the vacuum Pretreatment Device COD vial and remove the the base. Insert a fresh pump and adjust the (VPD) to a vacuum pump cap. Insert the vials in the Chloride Removal vacuum regulator valve on (not an aspirator-type numbered holes in the Cartridge (CRC) directly top of the VPD until the vacuum) that can create a VPD base. above each Mn III COD internal gauge reads 20 vacuum of 20–25 inches of Reagent Vial. Plug any inches of water. mercury. open holes in the VPD top using the stoppers provided. 5. Pipet 0.60 mL of 6. Close the vacuum 7. Open the VPD Proceed to Sample acidified sample (see regulator valve completely regulator valve to release preparation and Acidified sample to achieve full vacuum. the vacuum. Turn the measurement. preparation) into the CRC. After one minute of full pump off. Remove the Pipet 0.60 mL of acidified vacuum, slide the VPD VPD top and set it beside blank into another CRC. It back and forth several the base. should take 30–45 times to dislodge any Dispose of the used seconds to draw the liquid drops clinging to the Chloride Removal through the CRC into each cartridge. Cartridge. Do not reuse it. vial. If the sample does not flow through the Chloride Removal Cartridge (CRC), increase the vacuum until flow starts, then reduce the vacuum down to 20 inches of water. Proceed as usual. Oxygen Demand, Chemical Page 4 of 10
  • 5. Oxygen Demand, Chemical Sample preparation and measurement 1. Use forceps to remove 2. Insert each filter in the 3. Remove the Mn III 4. Invert the vials several the filter from the top of corresponding Mn III COD COD vial from the vacuum times to mix. each CRC. Vial. (Use numbers on the chamber and replace the If the sample does not VPD as a guide.) original cap. Screw the contain suspended solids, To avoid cross- cap on tightly. it is not necessary to contamination between transfer the filter to the samples, clean forcep tips digestion vial. between samples by wiping with a clean towel or rinsing with deionized water. 5. Insert the vials in the 6. Insert the vials in a 7. Cool the vials to room 8. Remove the vials from DRB200 Reactor at cooling rack for two temperature in a cool the water and wipe with a 150 °C. Close the minutes. water bath or with running clean, dry paper towel. protective cover. Digest for If the solution develops a tap water for several one hour. colorless upper layer and minutes. To oxidize resistant a purple lower layer, invert organics, samples can be the vial several times to digested for up to four mix and proceed to the hours. Digest the blank for next step. the same time period as the samples. Oxygen Demand, Chemical Page 5 of 10
  • 6. Oxygen Demand, Chemical Sample preparation and measurement (continued) Stored Programs 432 COD Mn III Zero Start 9. Select the test. 10. Insert the blank into 11. ZERO the instrument. 12. Make sure the filter Insert an adapter or a light the 16-mm cell holder. The display will show: disc is not suspended in shield if required (see the middle of the vial; it 0 mg/L COD Mn can interfere with the Instrument-specific information). instrument reading. The disc must be more than 20 mm (0.8") or less than 10 mm (0.4"), from the bottom of the vial. Move it by gently swirling or by lightly tapping the vial on the table top. Read 13. Wipe the prepared 14. READ the results in sample and insert it into mg/L COD Mn. the . Interferences Inorganic materials may also be oxidized by trivalent manganese and constitute a positive interference when present in significant amounts. Chloride is the most common interference and is removed by sample pretreatment with the Chloride Removal Cartridge. If chloride is known to be absent or present in insignificant levels, the pretreatment can be omitted. A simple way to determine if chloride will affect test results is to run routine samples with and without the chloride removal, then compare results. Other inorganic interferences (i.e., nitrite, ferrous iron, sulfide) are not usually present in significant amounts. If necessary, these interferences can be corrected after determining their concentrations with separate methods and adjusting the final COD test results accordingly. Ammonia nitrogen is known to interfere in the presence of chloride; it does not interfere if chloride is absent. Oxygen Demand, Chemical Page 6 of 10
  • 7. Oxygen Demand, Chemical Sample collection, preservation and storage • Collect samples in clean glass bottles. • Use plastic bottles only if they are known to be free of organic contamination. • Test biologically active samples as soon as possible. • Homogenize samples containing solids to assure representative samples. • Samples treated with concentrated sulfuric acid to a pH of less than 2 (about 2 mL per liter) and refrigerated at 4 °C may be stored up to 28 days. • Correct test results for volume additions. Accuracy check Standard solution method Note: Refer to the instrument user manual for specific software navigation instructions. Required for accuracy check: • COD standard solution, 800 mg/L 1. Use 0.60 mL of the 800-mg/L solution in place of the sample. 2. To adjust the calibration curve using the reading obtained with the standard solution, navigate to Standard Adjust in the software: OPTIONS>MORE>STANDARD ADJUST. 3. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate concentration is used, enter the concentration and adjust the curve to that value. Method performance Precision—95% Confidence Sensitivity—∆Concentration Program Instrument Standard Limits of Distribution per 0.010 ∆Abs 432 DR 5000 600 mg/L COD 576–624 mg/L COD 8 mg/L COD Estimated detection limit (EDL) The EDL for program 432 is 4 mg/L COD. The EDL is the calculated lowest average concentration in a deionized water matrix that is different from zero with a 99% level of confidence. Summary of method Chemical Oxygen Demand (COD) is defined as “... a measure of the oxygen equivalent of the organic matter content of a sample that is susceptible to oxidation by a strong chemical oxidant” (APHA Standard Methods, 19th ed., 1995). Trivalent manganese is a strong, non-carcinogenic chemical oxidant that changes quantitatively from purple to colorless when it reacts with organic matter. It typically oxidizes about 80% of the organic compounds. Studies have shown that the reactions are highly reproducible and test results correlate closely to Biochemical Oxygen Demand (BOD) values and hexavalent chromium COD tests. None of the oxygen demand tests provide 100% oxidation of all organic compounds. A calibration is provided which is based on the oxidation of Potassium Acid Phthalate (KHP). A different response may be seen in analyzing various wastewaters. The KHP calibration is adequate for most applications. The highest degree of accuracy is obtained when test results are correlated to a standard reference method such as BOD or one of the chromium COD methods. Special waste streams or classes will require a separate calibration to obtain a direct mg/L COD reading or to generate a correction factor for the precalibrated KHP response. The Oxygen Demand, Chemical Page 7 of 10
  • 8. Oxygen Demand, Chemical sample digestion time can be extended up to four hours for samples that are difficult to oxidize. Test results are measured at 510 nm. Consumables and replacement items Required reagents Description Quantity/Test Unit Catalog number Manganese III COD Reagent Vials, 20–1000 mg/L COD 1 25/pkg 2623425 Chloride Removal Cartridge (CRC) 1 25/pkg 2661825 Sulfuric Acid, concentrated, ACS 1 mL 2.5 L 97909 Water, deionized varies 4L 27256 Required apparatus Description Quantity/Test Unit Catalog number Blender, 120 VAC 1 each 2616100 Cap, with inert Teflon liner, for mixing bottle varies 12/pkg 2401812 DRB200 Reactor, 110 V, 15 x 16 mm 1 each LTV082.53.40001 DRB200 Reactor, 220 V, 15 x 16 mm 1 each LTV082.52.40001 Forceps, extra fine point 1 each 2669600 Pipet, TenSette®, 1.0 to 10.0 mL 1 each 1970010 Pipet Tips for TenSette Pipet 19700-10 2 50/pkg 2199796 Pipet, TenSette, 0.1 to 1.0 mL 1 each 1970001 Pipet Tips for TenSette Pipet 19700-01 2 50/pkg 2185696 Test Tube Rack 1 each 1864100 Vacuum Pretreatment Device (VPD) 1 each 4900000 Vacuum Pump, 1.2 CFM 115V 1 each 2824800 Vial, glass, for sample plus acid 2 each 2427700 Recommended standards Description Unit Catalog number COD Standard Solution, 800-mg/L COD 200 mL 2672629 Oxygen Demand Standard for BOD, COD, TOC, 10-mL ampules 16/pkg 2833510 Potassium Acid Phthalate, ACS 500 g 31534 Wastewater Standard, Influent Inorganics, for NH3–N, NO3–N, PO4, COD, SO4, TOC 500 mL 2833149 Optional reagents and apparatus Description Unit Catalog number Dispenser, automatic, 1.0–5.0 mL each 2563137 Titrator Strips, Quantab®, for low range chloride 40 tests 2744940 Finger cots 2/pkg 1464702 Flask, volumetric class A, 1 L each 1457453 Oxygen Demand, Chemical Page 8 of 10
  • 9. Oxygen Demand, Chemical Optional reagents and apparatus Description Unit Catalog number pH Paper, 0–14 pH range 100/pkg 2601300 Balance, analytical, 80 g capacity, 115 VAC each 2936701 Weighing Paper, 76x76 mm 500/pkg 1473800 Oven, Laboratory 120 VAC/60Hz each 1428900 Oven, Laboratory 240 VAC/50Hz each 1428902 COD Standard Solution, 300-mg/L 200 mL 1218629 COD Standard Solution, 300-mg/L 500 mL 1218649 COD Standard Solution, 1000-mg/L 200 mL 2253929 Standard Methods book, most recent edition each 2270800 Pipet Tips for TenSette Pipet 19700-10 250/pkg 2199725 Pipet Tips for TenSette Pipet 19700-01 1000/pkg 2185628 Oxygen Demand, Chemical Page 9 of 10
  • 10. FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: HACH COMPANY In the U.S.A. – Call toll-free 800-227-4224 WORLD HEADQUARTERS Outside the U.S.A. – Contact the HACH office or distributor serving you. Telephone: (970) 669-3050 On the Worldwide Web – www.hach.com; E-mail – techhelp@hach.com FAX: (970) 669-2932 © Hach Company, 2007, 2010. All rights reserved. Printed in the U.S.A. Edition 6