Ludlum 14C Survey Meter
Troubleshooting
No power or meter does not reach BAT TEST mark
1. Check batteries. Replace if weak.
2. Check polarity. (There are marks inside the battery lid indicating which end of
the batteries should be facing up.)
3. Check battery contacts. Clean with rough sandpaper or use an engraver to
clean the tips.
4. Remove the meter can and check for loose or broken wires.
Nonlinear Readings
1. Check the high voltage (HV) using Ludlum model 500 Pulser or high
impedance multimeter.
2. Check for noise in the detector cable by disconnecting detector, placing meter
on lowest range setting, and wiggling the cable while observing the meter face
dial for significant changes in readings.
3. Check for “sticky” meter movement. Does the reading change when you tap
the meter? Does the needle stick at any spot?
4. Check the meter zero. Turn the power off or press the RESET button. The
meter needle should return to 0.
Meter goes full-scale or “Pegs out”
1. Replace the detector cable to determine whether the cable has failed, causing
excessive noise.
2. Check the HV and, if possible, the input threshold for proper setting.
3. Ensure the meter’s can is properly attached. The speaker should be on the left
side. If the can is on backwards, interference between the speaker and the
input preamplifier may cause noise.
No Response to radiation
1. Substitute a “known well” detector and/or cable.
2. Has the correct operating voltage been set? Refer to the calibration certificate
or detector instruction manual. If the meter uses multiple detectors, confirm
the high voltage is matched with the detector being used.
(cont’d)
No Audio
1. Ensure that the AUD ON/OFF switch is set to ON.
2. Remove the meter housing and check the connection between the circuit board
and the speaker. Plug in the two-pin connector if it has been disconnected.
Troubleshooting G-M Detectors
1. If the tube has a thin mica window, check for window breakage. If damage is
evident, the tube must be replaced. (See “Pancake Tube Replacement”.)
2. Check the HV. For most G-M tubes it will be 900 Vdc, or 460-550 Vdc for
“peanut” tubes.
3. If the input sensitivity is too low, the user could see some double-pulsing.
4. Wires to the tube may be broken or the crimped connector could have a loose
wire.
Troubleshooting Scintillators
1. Alpha or Alpha/Beta scintillators are prone to light leaks. They can be tested in
a dark room or with a bright light. If a light leak is determined, changing the
Mylar window assembly will usually fix the problem.
NOTE: Be sure to use a replacement window made to the same thickness and
with the same number of layers as the original.
2. Verify that the HV and input sensitivity are correct. Alpha and gamma
scintillators typically operate from 10-35 mV. High voltage varies with the
photomultiplier tubes (PMT) from 600 Vdc to 1400 Vdc.
3. On a gamma scintillator, visually inspect the crystal for breakage or humidity
leakage. Water inside the crystal will turn it yellow and will gradually degrade
performance.
4. Check the PMT to see if the photocathode still exists. If the end of the PMT is
clear (not brownish), there is a loss of vacuum, which renders the PMT
useless.

More Related Content

Viewers also liked

KF Investment Advisory Report Nov-2012
KF Investment Advisory Report Nov-2012KF Investment Advisory Report Nov-2012
KF Investment Advisory Report Nov-2012
Hitendra Gupta
 

Viewers also liked (18)

2016 Long Island Beauty Ball Deck: Bohlson Restaurant Group
2016 Long Island Beauty Ball Deck: Bohlson Restaurant Group2016 Long Island Beauty Ball Deck: Bohlson Restaurant Group
2016 Long Island Beauty Ball Deck: Bohlson Restaurant Group
 
James D kuhn - 10 Lessons to Become a Successful Entrepreneur
James D kuhn - 10 Lessons to Become a Successful EntrepreneurJames D kuhn - 10 Lessons to Become a Successful Entrepreneur
James D kuhn - 10 Lessons to Become a Successful Entrepreneur
 
Ken caryl-ranch-directory-proof-binder-01
Ken caryl-ranch-directory-proof-binder-01Ken caryl-ranch-directory-proof-binder-01
Ken caryl-ranch-directory-proof-binder-01
 
2016 Long Island Beauty Ball Deck: Smith Haven Mall
2016 Long Island Beauty Ball Deck: Smith Haven Mall2016 Long Island Beauty Ball Deck: Smith Haven Mall
2016 Long Island Beauty Ball Deck: Smith Haven Mall
 
Panduan Membeli Projektor
Panduan Membeli ProjektorPanduan Membeli Projektor
Panduan Membeli Projektor
 
KF Investment Advisory Report Nov-2012
KF Investment Advisory Report Nov-2012KF Investment Advisory Report Nov-2012
KF Investment Advisory Report Nov-2012
 
12 Years a Slave Presentation cf
12 Years a Slave Presentation cf12 Years a Slave Presentation cf
12 Years a Slave Presentation cf
 
Psychology of influence
Psychology of influencePsychology of influence
Psychology of influence
 
Νοσηλεία και Έρευνα - Τεύχος 33
Νοσηλεία και Έρευνα - Τεύχος 33Νοσηλεία και Έρευνα - Τεύχος 33
Νοσηλεία και Έρευνα - Τεύχος 33
 
Teyxos 30
Teyxos 30Teyxos 30
Teyxos 30
 
Νοσηλεία και Έρευνα - Τεύχος 27
Νοσηλεία και Έρευνα - Τεύχος 27Νοσηλεία και Έρευνα - Τεύχος 27
Νοσηλεία και Έρευνα - Τεύχος 27
 
Νοσηλεία και Έρευνα - Τεύχος 28
Νοσηλεία και Έρευνα - Τεύχος 28Νοσηλεία και Έρευνα - Τεύχος 28
Νοσηλεία και Έρευνα - Τεύχος 28
 
Νοσηλεία και Έρευνα - Τεύχος 34
Νοσηλεία και Έρευνα - Τεύχος 34Νοσηλεία και Έρευνα - Τεύχος 34
Νοσηλεία και Έρευνα - Τεύχος 34
 
Surgical Approach To Open Fractures by Dr Sarah Murniati
Surgical Approach To Open Fractures by Dr Sarah MurniatiSurgical Approach To Open Fractures by Dr Sarah Murniati
Surgical Approach To Open Fractures by Dr Sarah Murniati
 
2016 Long Island Beauty Ball Deck: Revitalash
2016 Long Island Beauty Ball Deck: Revitalash2016 Long Island Beauty Ball Deck: Revitalash
2016 Long Island Beauty Ball Deck: Revitalash
 
Cloud based cad modeling
Cloud based cad modelingCloud based cad modeling
Cloud based cad modeling
 
25 Quotes to Inspire Success in Your Marketing Efforts
25 Quotes to Inspire Success in Your Marketing Efforts25 Quotes to Inspire Success in Your Marketing Efforts
25 Quotes to Inspire Success in Your Marketing Efforts
 
Hyderabad Real Estate Report H1 2016 Presentation
Hyderabad Real Estate Report H1 2016 PresentationHyderabad Real Estate Report H1 2016 Presentation
Hyderabad Real Estate Report H1 2016 Presentation
 

Similar to troubleshooting

HELE / TLEelectronics
HELE / TLEelectronicsHELE / TLEelectronics
HELE / TLEelectronics
djpprkut
 
Chapter 57
Chapter 57Chapter 57
Chapter 57
mcfalltj
 
Smoke Alarm using Photodiode
Smoke Alarm using PhotodiodeSmoke Alarm using Photodiode
Smoke Alarm using Photodiode
Chintan Patel
 
Fit extra electrical sockets
Fit extra electrical socketsFit extra electrical sockets
Fit extra electrical sockets
Zulaiha Amaria
 

Similar to troubleshooting (20)

Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
 Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20 Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
Hướng dẫn sử dụng thiết bị kiểm tra dây cáp mạng TG - 20
 
ORP Sensor Troubleshooting Guide
ORP Sensor Troubleshooting GuideORP Sensor Troubleshooting Guide
ORP Sensor Troubleshooting Guide
 
Extech EX330 Mini Multimeter User Manual
Extech EX330 Mini Multimeter User ManualExtech EX330 Mini Multimeter User Manual
Extech EX330 Mini Multimeter User Manual
 
3. Introduction of Electrical protective devices (1).pptx
3. Introduction of Electrical protective devices (1).pptx3. Introduction of Electrical protective devices (1).pptx
3. Introduction of Electrical protective devices (1).pptx
 
Kaise sk 20 22-30 manual
Kaise sk 20 22-30 manualKaise sk 20 22-30 manual
Kaise sk 20 22-30 manual
 
Wire Harness Test Simple
Wire Harness Test SimpleWire Harness Test Simple
Wire Harness Test Simple
 
Hướng dẫn sử dụng bộ kiểm tra mạng Extech 40180
Hướng dẫn sử dụng bộ kiểm tra mạng Extech 40180Hướng dẫn sử dụng bộ kiểm tra mạng Extech 40180
Hướng dẫn sử dụng bộ kiểm tra mạng Extech 40180
 
Digital clampmeters
Digital clampmetersDigital clampmeters
Digital clampmeters
 
How to use a multimeter
How to use a multimeterHow to use a multimeter
How to use a multimeter
 
Etekcity MSR-R500 Digital Multimeter User Manual
Etekcity MSR-R500 Digital Multimeter User ManualEtekcity MSR-R500 Digital Multimeter User Manual
Etekcity MSR-R500 Digital Multimeter User Manual
 
HELE / TLEelectronics
HELE / TLEelectronicsHELE / TLEelectronics
HELE / TLEelectronics
 
Quizbank quiz buzzer
Quizbank quiz buzzer Quizbank quiz buzzer
Quizbank quiz buzzer
 
Electrical testing
Electrical testingElectrical testing
Electrical testing
 
Chapter 57
Chapter 57Chapter 57
Chapter 57
 
KL1034 Lect2 Electrical wiring.pdf
KL1034 Lect2 Electrical wiring.pdfKL1034 Lect2 Electrical wiring.pdf
KL1034 Lect2 Electrical wiring.pdf
 
Hướng dẫn sử dụng Ampe kìm với 100A với đo nhiệt độ hồng ngoại và Bluetooth E...
Hướng dẫn sử dụng Ampe kìm với 100A với đo nhiệt độ hồng ngoại và Bluetooth E...Hướng dẫn sử dụng Ampe kìm với 100A với đo nhiệt độ hồng ngoại và Bluetooth E...
Hướng dẫn sử dụng Ampe kìm với 100A với đo nhiệt độ hồng ngoại và Bluetooth E...
 
Smoke Alarm using Photodiode
Smoke Alarm using PhotodiodeSmoke Alarm using Photodiode
Smoke Alarm using Photodiode
 
Test Switch Operation and Accessories
Test Switch Operation and AccessoriesTest Switch Operation and Accessories
Test Switch Operation and Accessories
 
Fit extra electrical sockets
Fit extra electrical socketsFit extra electrical sockets
Fit extra electrical sockets
 
Theories of cellphones including the types and model
Theories of cellphones  including the types and modelTheories of cellphones  including the types and model
Theories of cellphones including the types and model
 

troubleshooting

  • 1. Ludlum 14C Survey Meter Troubleshooting No power or meter does not reach BAT TEST mark 1. Check batteries. Replace if weak. 2. Check polarity. (There are marks inside the battery lid indicating which end of the batteries should be facing up.) 3. Check battery contacts. Clean with rough sandpaper or use an engraver to clean the tips. 4. Remove the meter can and check for loose or broken wires. Nonlinear Readings 1. Check the high voltage (HV) using Ludlum model 500 Pulser or high impedance multimeter. 2. Check for noise in the detector cable by disconnecting detector, placing meter on lowest range setting, and wiggling the cable while observing the meter face dial for significant changes in readings. 3. Check for “sticky” meter movement. Does the reading change when you tap the meter? Does the needle stick at any spot? 4. Check the meter zero. Turn the power off or press the RESET button. The meter needle should return to 0. Meter goes full-scale or “Pegs out” 1. Replace the detector cable to determine whether the cable has failed, causing excessive noise. 2. Check the HV and, if possible, the input threshold for proper setting. 3. Ensure the meter’s can is properly attached. The speaker should be on the left side. If the can is on backwards, interference between the speaker and the input preamplifier may cause noise. No Response to radiation 1. Substitute a “known well” detector and/or cable. 2. Has the correct operating voltage been set? Refer to the calibration certificate or detector instruction manual. If the meter uses multiple detectors, confirm the high voltage is matched with the detector being used. (cont’d)
  • 2. No Audio 1. Ensure that the AUD ON/OFF switch is set to ON. 2. Remove the meter housing and check the connection between the circuit board and the speaker. Plug in the two-pin connector if it has been disconnected. Troubleshooting G-M Detectors 1. If the tube has a thin mica window, check for window breakage. If damage is evident, the tube must be replaced. (See “Pancake Tube Replacement”.) 2. Check the HV. For most G-M tubes it will be 900 Vdc, or 460-550 Vdc for “peanut” tubes. 3. If the input sensitivity is too low, the user could see some double-pulsing. 4. Wires to the tube may be broken or the crimped connector could have a loose wire. Troubleshooting Scintillators 1. Alpha or Alpha/Beta scintillators are prone to light leaks. They can be tested in a dark room or with a bright light. If a light leak is determined, changing the Mylar window assembly will usually fix the problem. NOTE: Be sure to use a replacement window made to the same thickness and with the same number of layers as the original. 2. Verify that the HV and input sensitivity are correct. Alpha and gamma scintillators typically operate from 10-35 mV. High voltage varies with the photomultiplier tubes (PMT) from 600 Vdc to 1400 Vdc. 3. On a gamma scintillator, visually inspect the crystal for breakage or humidity leakage. Water inside the crystal will turn it yellow and will gradually degrade performance. 4. Check the PMT to see if the photocathode still exists. If the end of the PMT is clear (not brownish), there is a loss of vacuum, which renders the PMT useless.