SlideShare a Scribd company logo
1 of 34
ENVIRONMENTAL POLLUTION
MONITORING DEVICES
BY
N.P.DHARANI
ASSISTANT PROFESSOR
CONTENTS
• INTRODUCTION
• CO MONITOR
• NOX ANALYZER
• SULPHUR DIOXIDE MEASUREMENT
• HYDRO CARBONS ESTIMATION
INTRODUCTION
• Used to analyze the amount different gases present
in a mixture.
• 02,CO,CO2,NO2 can be analyzed from air.
• Gas analyzers used in industries are:
• In power plants oxygen and CO can be measured in
flue gas.
• In ammonia synthesis, presence of CO,
CO2,H2S,Water vapour will poison the synthesis
catalyst. So, these are analyzed and prevention are
taken to avoid the damage.
Instrumental Techniques and
measurements range
FROM R.S.KHANDPUR
CO MONITOR
• Non dispersive infrared analyzer
• Gas Chromatography
Non dispersive infrared analyzer
• It depends on characteristic energy of
absorption of CO molecule at wavelength of
4.6µ.
• It also absorbs other gases like CO2, H2O,SO2 and
NO2.
• Infrared radiation is passed through a sample
cell containing CO and a reference cell
containing a fixed quantity of nitrogen , CO and
water vapour.
Infrared Gas Analyzer
Construction:
• Two identical IR sources
emits beam of radiation
that are pulsed by a motor
driven chopper.
• The source of IR radiation I
shot wire spiral.
• For industrial analyzers,
the chopping frequency is
2-10Hz whereas for
medical application, it is 2-
50Hz.
FROM R.S.KHANDPUR
Block diagram of measuring system for
carbon monoxide
FROM R.S.KHANDPUR
• Reference cell contains a fixed quantity of the gases
whereas air sample is made to flow at about 150ml/min
though the sample cell.
• CO analyzer with IR absorption would give greater
sensitivity with larger cell path length.
• Instruments with 1mm cell length would measure from 1
to 50 ppm.
• Some latest instruments are capable of measuring 1 to
25 ppm even with cell path length of 10 cm.
• Response time of such instruments varies from 1 to 5
mins.
• Calibration is done by passing a known ppm
concentration of CO in nitrogen in the sample cell.
• The zero is set by using pure nitrogen as a sample.
Advantages
• Effect of flow rate is small.
• Response speed is high.
• High sensitivity measurements are possible.
• Effect of interfering components is small.
• Equipments are easy to maintain.
Gas Chromatography
• Gas chromatography (GC) is a common type of chromatography used
in analytical chemistry for separating and analyzing compounds that
can be vaporized without decomposition
• Air sample is passed though a stripper column, the heavy
hydrocarbons are retained, CO and methane are passed into
chromatographic column and then to catalytic reducing chamber.
• The methane pass through the reducing chamber unaffected, while CO
is reduced to methane.
• By using hydrogen flame ionization detector, both methane peaks can
be detected.
• The first peak is due to methane and second peak is due to CO.
• Accuracy is ±2%
• Peak heights of CO and methane give sensitivity of about 50ppb.
NOX ANALYZER
• COLORIMETRY
• CHEMILUMINESCENCE
• USE OF CO LASER
• LASER OPTO-ACOUSTIC SPETROSCOPY
COLORIMETRY
• This method can be used for the determination of nitrogen dioxide by
using Saltzman method.
• This method is based on a reaction in which a pink colored dye
complex is formed when air containing NO2 is passed in an absorbing
solution consisting of sulphanilic acid and diamine dissolved in the
acetic medium.
• Sensitivity is in ppm range.
• Nitric oxide may be determined by passing the sample through an acid
permanganate bubbler which oxidizes nitric oxide to nitrogen dioxide.
• Used for monitoring in stack sewages in concentration range of 5 ppm
to several thousand ppm, consists in passing sample into an evacuated
flask containing a solution of H2O2 in sulphuric acid.
• The oxides of nitrogen are converted into nitric acid and nitrate ions
react with phenoldisupluric acid to produce a yellow color, which is
measured calorimetrically.
CHEMILUMINESCENCE
• The phenomenon of emission of radiation from
a chemically excited species is known as
chemiluminescence. It results due to the
formation of new chemical bonds .The atoms in
the excited states posses higher energy levels
than the ground state & while returning to the
ground state they emit radiations.
• This phenomenon is useful for measurement of
air pollutants, particularly NO, NO2
CONT.,
• The following reaction uses for the measurement of
chemiluminescence emission.
• Since NO2 reacts slowly with ozone and the reaction
which produces NO3 is not done by chemiluminescence,
so it is necessary to reduce NO2 to NO before admission
into reactor.
NO2 NO +1/2 O2
• Light emission is measured with photmultiplier with high
gain and low dark current can have low levels of
radiation can be measured.
• The response of instruments based on this method is
linear from 1ppm to 1000ppm of NO.
• Used extremely in automative exhaust gases.
MEASUREMENT OF NITROGEN OXIDE
USING CO LASERS
FROM R.S.KHANDPUR
• This method detects nitric oxide in 0.25ppm.
• The apparatus consists of a CO laser, which emits
radiation that is absorbed by NO in the mixture, the
amount of absorption is proportional to the
concentration of NO present.
• The wavelength match between laser and NO is made
exact and hence the absorption is enhanced by placing
NO in a magnetic field.
• The CO laser used is a dc excited continuous working
laser, which operates at a single wavelength of 5.307µm.
• A diffraction grating is used at one end of the cavity as a
line selector.
• The laser yields 5-30mW of single line power.
• Absorption cell is made of pyrex (low thermal expansion
borosilicate glass) of diameter 15mm and 90cm length.
• Pressure applied is 10^-6- 10^-5 torr.
• To produce modulating audio frequency magnetic field
along the axis of the cell, insulated wire is closely wound
around the outside of the cell over about half its length.
• The coil is excited with a current of 1A in the frequency
range of 5-150KHz.
• This produces a varying magnetic field of about 50Guass P-
P intensity.
• The DC magnetic field is produced by a solenoid which
produce a field upto 2.5KGauss.
• The detector is a liquid nitrogen cooled Ge-Au element.
• The signal is amplied in a lock in amplifier(extracts a signal
with a known carrier wave from an extremely noise
environment) before given to recorder.
• The signal amplitude varies linearly with the concentration
of NO in the sample.
LASER OPTO-ACOUSTIC SPETROSCOPY
• Opto acoustic detectors combined with thermal IR sources has
been widely used in gas detection.
• This technique is used to measure trace amounts of nitrogen oxides
in the stratosphere.
• Opto acoustic detector involves the absorption of an AM beam of IR
by a gas which results in the generation of sound.
• The energy absorbed by the gas molecules from IR beam excites the
molecules to the rotational-vibrational energy levels above the
ground state.
• The main path for the delay of these excited states is collisional de-
excitation which results in the transfer of absorbed energy into heat
and raises the gas temperature.
• The temperature rise causes a corresponding pressure rise in the
gas.
Block Diagram of laser opto acoustic
detector for nitric oxides
FROM R.S.KHANDPUR
• When the beam intensity is modulated, the gas
temperature and pressure changes accordingly.
• Thus periodic pressure vibrations in the ags result in the
generation of sound.
• The arrangement uses CO2 laser, tuned by rotating a
diffraction grating at ine end of the laser cavity.
• It is tunable to 64 different emission lines in the range
from 927 to 1085cm-1
• The laser beam is focused at chopping wheel and then
focused into the detector.
• The chopping frequency is selected to optimize the SNR.
• The microphone is available and acoustic signal is
amplified in a preamplifier and displayed.
• The detector is highly sensitive and sample amount of
20pg may be detected.
SULPHUR DIOXIDE MEASUREMENT
• Colorimetry
• Conductivitimetry
• Gas chromatography
• Coulometry
• Flame photometric detector
• UV fluorescence method
Colorimetry
• A known volume of air is passed through an aqueous
solution,which contains reagents that absorb so2 and
produce a colored substance.
• The amount of colored substance is proportional to so2.
• The air to be analysed is passed through an aqueous
solution of 0.1M sodium tetracholromercurate.
• The so2 reacts with mercuric salt to form a
sulphatomercuric compound.
• The solution is further treated with acid bleached para-
rasoniline and formaldehyde, which forms a red purple
colour.
• The amount of color is determined photometrically.
• So2 can also be determined by means of
decolorisation of an aqueous starch iodine
solution which normally has deep blue color.
• When so2 is passed through the solution
containing starch iodine solution,the iodine
reduced by sulphur dioxide which in turn
results in decrease of intensity of starch iodine
color.
Conductivitimetry
• This technique is based
on measuring the change
in conductivity of a
solution, when a sample
of air containing so2 is
bubbled through it.
• C- cell
• P-perspex cap
• J-jet
• E-electrodes
FROM R.S.KHANDPUR
coulometry
• The ability of sulphur dioxide to reduce iodine can be
employed to monitor so2 by using iodine coulometry in an
aqueous solution.
• The mass of iodine reacted per unit time during any given
interval of time would indicate concentration of so2 in air.
• The coulometric arrangement requires two electrodes
made of platinum which acts as anode and cathode.
• These electrodes maintain a trace of iodine in equilibrium
with potassium bromide to maintain conductance.
• The shift in anode-cathode potential is sensed by a third
reference electrode, when so2 from sample air is bubbled
through the solution in a detector cell.
Flame photometric detector
• In Flame photometers, sample air is
introduced into hydrogen rich air flame by
using a narrow band interference filter shields
the PMT detectors, but 394nm emission
energy of flame excited sulphur atoms.
UV fluorescence method
• This method is based on the principle that so2
molecules absorb UV light and became excited
at a particular wavelength.
• These molecules then decay to a lower energy
sate emitting UV light at a different
wavelength.
• The intensity of fluorescence is proportional
to the so2 concentration.
Block diagram of fluorescence so2
analyser
Hydrocarbons estimation
• Flame ionization detector
• Gas chromatography
Flame ionization detector
• Organic compounds easily pyrolyse when
introduced into an air-hydrogen flame.
• The pyrolysis produces ions that can be
collected by having a cylindrical grid
surrounding the flame.
• The detector response would be in proportion
to the number of carbon atoms in the chain.
Schematic diagram of FID
FROM R.S.KHANDPUR
Working principle
• The sample gas containing hydrocarbons controlled at
a constant flow rate is mixed with hydrogen for fuel.
• The mixed gas is burned at the end of a very fine
nozzle.
• Two electrodes are placed on either side of the flame
and an appropriate electric field is applied to them.
• Ionic current will flow between the collector electrode
and the other electrodes.
• The current is then amplified and displayed on a meter.
Gas chromatography
• Gas chromatography has been applied for the
detection and measurement of hydrocarbons,
CO,CO2 in air.
• They are generally detected by FID since they
are not sensitive to CO and CO2.
• These are first converted into methane by
hydrogenation and then measured.

More Related Content

What's hot

Monitoring CO -NDIR Method
Monitoring CO -NDIR MethodMonitoring CO -NDIR Method
Monitoring CO -NDIR MethodECRD IN
 
Important analytical terms in industry like pH, ORP, SCR, Gas Chromatography
Important analytical terms in industry like pH, ORP, SCR, Gas ChromatographyImportant analytical terms in industry like pH, ORP, SCR, Gas Chromatography
Important analytical terms in industry like pH, ORP, SCR, Gas ChromatographyHamzaNabeel1
 
Applications of ICPAES
Applications of ICPAESApplications of ICPAES
Applications of ICPAESPradeepa J
 
TURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTION
TURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTIONTURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTION
TURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTIONHaydar Mohammad Salim
 
Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...UMT Lahore
 
inductively coupled mass spectrometer
inductively coupled mass spectrometerinductively coupled mass spectrometer
inductively coupled mass spectrometerMANU MOHAN
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)megr1412
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopyMahendra G S
 
Atomic Absorption Spectroscopy
Atomic Absorption SpectroscopyAtomic Absorption Spectroscopy
Atomic Absorption SpectroscopySHANMUGAPRIYA I.G.
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy AyeshaRasty
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentationArgha Sen
 
Carbon monoxide
Carbon monoxideCarbon monoxide
Carbon monoxideChakri Cr
 
Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyAnuradha Verma
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyShivaji Burungale
 
A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...Adnan Sohail
 
inductively coupled plasma ICP techniques & applications
inductively coupled plasma ICP techniques &  applicationsinductively coupled plasma ICP techniques &  applications
inductively coupled plasma ICP techniques & applicationsGamal Abdel Hamid
 
A New Method for the Analysis of ppb
A New Method for the Analysis of ppb A New Method for the Analysis of ppb
A New Method for the Analysis of ppb Jennifer Maclachlan
 

What's hot (20)

Monitoring CO -NDIR Method
Monitoring CO -NDIR MethodMonitoring CO -NDIR Method
Monitoring CO -NDIR Method
 
Important analytical terms in industry like pH, ORP, SCR, Gas Chromatography
Important analytical terms in industry like pH, ORP, SCR, Gas ChromatographyImportant analytical terms in industry like pH, ORP, SCR, Gas Chromatography
Important analytical terms in industry like pH, ORP, SCR, Gas Chromatography
 
Applications of ICPAES
Applications of ICPAESApplications of ICPAES
Applications of ICPAES
 
TURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTION
TURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTIONTURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTION
TURBIDIMETRIC DETERMINATION OF SULPHATE ION (SO42- ) IN THE UNKNOWN SOLUTION
 
Icpms
IcpmsIcpms
Icpms
 
Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...Deference between atomic absorption spectrometry and atomic emission spectrom...
Deference between atomic absorption spectrometry and atomic emission spectrom...
 
inductively coupled mass spectrometer
inductively coupled mass spectrometerinductively coupled mass spectrometer
inductively coupled mass spectrometer
 
Presentation1
Presentation1Presentation1
Presentation1
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
Atomic Absorption Spectroscopy
Atomic Absorption SpectroscopyAtomic Absorption Spectroscopy
Atomic Absorption Spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentation
 
Carbon monoxide
Carbon monoxideCarbon monoxide
Carbon monoxide
 
ICP-MS
ICP-MSICP-MS
ICP-MS
 
Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...A presentation ,analytical methods-for-determination-of-metals-in-environment...
A presentation ,analytical methods-for-determination-of-metals-in-environment...
 
inductively coupled plasma ICP techniques & applications
inductively coupled plasma ICP techniques &  applicationsinductively coupled plasma ICP techniques &  applications
inductively coupled plasma ICP techniques & applications
 
A New Method for the Analysis of ppb
A New Method for the Analysis of ppb A New Method for the Analysis of ppb
A New Method for the Analysis of ppb
 

Similar to Environmental pollution monitoring devices

Nitrogen oxides and carbon oxides determination
Nitrogen oxides and carbon oxides determinationNitrogen oxides and carbon oxides determination
Nitrogen oxides and carbon oxides determinationIshfaq Ahmad
 
Detectors used in gas chromatography
Detectors used in gas chromatographyDetectors used in gas chromatography
Detectors used in gas chromatographyVrushali Tambe
 
Monitoring of gaseous pollutants
Monitoring of gaseous pollutantsMonitoring of gaseous pollutants
Monitoring of gaseous pollutantsGARRE RAVI KUMAR
 
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.Dr. Ravi Sankar
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYnadeem akhter
 
Types of Industrial Process Analyzers
Types of Industrial Process AnalyzersTypes of Industrial Process Analyzers
Types of Industrial Process AnalyzersAttiya Rehman
 
Instrumentation of IR spectroscopy
Instrumentation of IR spectroscopyInstrumentation of IR spectroscopy
Instrumentation of IR spectroscopyTalha Liaqat
 
Infra red spectroscopy
Infra red spectroscopyInfra red spectroscopy
Infra red spectroscopySiddu K M
 
atomic absorption spectroscopy.pdf
atomic absorption spectroscopy.pdfatomic absorption spectroscopy.pdf
atomic absorption spectroscopy.pdfssusera34ec2
 
M.Sc. Part I 1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsx
M.Sc. Part I  1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsxM.Sc. Part I  1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsx
M.Sc. Part I 1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsxSIMARANJAISWAL41M3S
 
Atomic absorption spectrcopy
Atomic absorption spectrcopyAtomic absorption spectrcopy
Atomic absorption spectrcopySidhu Taran
 

Similar to Environmental pollution monitoring devices (20)

Nitrogen oxides and carbon oxides determination
Nitrogen oxides and carbon oxides determinationNitrogen oxides and carbon oxides determination
Nitrogen oxides and carbon oxides determination
 
Infrared Spectroscopy
Infrared SpectroscopyInfrared Spectroscopy
Infrared Spectroscopy
 
Detectors used in gas chromatography
Detectors used in gas chromatographyDetectors used in gas chromatography
Detectors used in gas chromatography
 
AAS.pptx
AAS.pptxAAS.pptx
AAS.pptx
 
Monitoring of gaseous pollutants
Monitoring of gaseous pollutantsMonitoring of gaseous pollutants
Monitoring of gaseous pollutants
 
IR Spectroscopy
IR SpectroscopyIR Spectroscopy
IR Spectroscopy
 
Unit v
Unit vUnit v
Unit v
 
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
Types of Industrial Process Analyzers
Types of Industrial Process AnalyzersTypes of Industrial Process Analyzers
Types of Industrial Process Analyzers
 
Instrumentation of IR spectroscopy
Instrumentation of IR spectroscopyInstrumentation of IR spectroscopy
Instrumentation of IR spectroscopy
 
Ir spectroscopy
Ir spectroscopyIr spectroscopy
Ir spectroscopy
 
Infra red spectroscopy
Infra red spectroscopyInfra red spectroscopy
Infra red spectroscopy
 
atomic absorption spectroscopy.pdf
atomic absorption spectroscopy.pdfatomic absorption spectroscopy.pdf
atomic absorption spectroscopy.pdf
 
25601.pptx
25601.pptx25601.pptx
25601.pptx
 
ICP - ICP/MS comparison
ICP - ICP/MS comparisonICP - ICP/MS comparison
ICP - ICP/MS comparison
 
M.Sc. Part I 1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsx
M.Sc. Part I  1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsxM.Sc. Part I  1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsx
M.Sc. Part I 1.3 INFRARED SPECTROPHOTOMETRY PPT (1).ppsx
 
CEMS_Sunil.pptx
CEMS_Sunil.pptxCEMS_Sunil.pptx
CEMS_Sunil.pptx
 
Atomic absorption spectrcopy
Atomic absorption spectrcopyAtomic absorption spectrcopy
Atomic absorption spectrcopy
 
Detectors in GC
Detectors in GCDetectors in GC
Detectors in GC
 

Recently uploaded

Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AIabhishek36461
 
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2RajaP95
 
microprocessor 8085 and its interfacing
microprocessor 8085  and its interfacingmicroprocessor 8085  and its interfacing
microprocessor 8085 and its interfacingjaychoudhary37
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidNikhilNagaraju
 
Current Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLCurrent Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLDeelipZope
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girlsssuser7cb4ff
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxDeepakSakkari2
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerAnamika Sarkar
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxpurnimasatapathy1234
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
chaitra-1.pptx fake news detection using machine learning
chaitra-1.pptx  fake news detection using machine learningchaitra-1.pptx  fake news detection using machine learning
chaitra-1.pptx fake news detection using machine learningmisbanausheenparvam
 
CCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdf
CCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdfCCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdf
CCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdfAsst.prof M.Gokilavani
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZTE
 

Recently uploaded (20)

Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AI
 
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
microprocessor 8085 and its interfacing
microprocessor 8085  and its interfacingmicroprocessor 8085  and its interfacing
microprocessor 8085 and its interfacing
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
main PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfidmain PPT.pptx of girls hostel security using rfid
main PPT.pptx of girls hostel security using rfid
 
Current Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLCurrent Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCL
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girls
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptx
 
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptx
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 
young call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Service
young call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Serviceyoung call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Service
young call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Service
 
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Isha Call 7001035870 Meet With Nagpur Escorts
 
chaitra-1.pptx fake news detection using machine learning
chaitra-1.pptx  fake news detection using machine learningchaitra-1.pptx  fake news detection using machine learning
chaitra-1.pptx fake news detection using machine learning
 
CCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdf
CCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdfCCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdf
CCS355 Neural Network & Deep Learning UNIT III notes and Question bank .pdf
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
 

Environmental pollution monitoring devices

  • 2. CONTENTS • INTRODUCTION • CO MONITOR • NOX ANALYZER • SULPHUR DIOXIDE MEASUREMENT • HYDRO CARBONS ESTIMATION
  • 3. INTRODUCTION • Used to analyze the amount different gases present in a mixture. • 02,CO,CO2,NO2 can be analyzed from air. • Gas analyzers used in industries are: • In power plants oxygen and CO can be measured in flue gas. • In ammonia synthesis, presence of CO, CO2,H2S,Water vapour will poison the synthesis catalyst. So, these are analyzed and prevention are taken to avoid the damage.
  • 5. CO MONITOR • Non dispersive infrared analyzer • Gas Chromatography
  • 6. Non dispersive infrared analyzer • It depends on characteristic energy of absorption of CO molecule at wavelength of 4.6µ. • It also absorbs other gases like CO2, H2O,SO2 and NO2. • Infrared radiation is passed through a sample cell containing CO and a reference cell containing a fixed quantity of nitrogen , CO and water vapour.
  • 7. Infrared Gas Analyzer Construction: • Two identical IR sources emits beam of radiation that are pulsed by a motor driven chopper. • The source of IR radiation I shot wire spiral. • For industrial analyzers, the chopping frequency is 2-10Hz whereas for medical application, it is 2- 50Hz. FROM R.S.KHANDPUR
  • 8. Block diagram of measuring system for carbon monoxide FROM R.S.KHANDPUR
  • 9. • Reference cell contains a fixed quantity of the gases whereas air sample is made to flow at about 150ml/min though the sample cell. • CO analyzer with IR absorption would give greater sensitivity with larger cell path length. • Instruments with 1mm cell length would measure from 1 to 50 ppm. • Some latest instruments are capable of measuring 1 to 25 ppm even with cell path length of 10 cm. • Response time of such instruments varies from 1 to 5 mins. • Calibration is done by passing a known ppm concentration of CO in nitrogen in the sample cell. • The zero is set by using pure nitrogen as a sample.
  • 10. Advantages • Effect of flow rate is small. • Response speed is high. • High sensitivity measurements are possible. • Effect of interfering components is small. • Equipments are easy to maintain.
  • 11. Gas Chromatography • Gas chromatography (GC) is a common type of chromatography used in analytical chemistry for separating and analyzing compounds that can be vaporized without decomposition • Air sample is passed though a stripper column, the heavy hydrocarbons are retained, CO and methane are passed into chromatographic column and then to catalytic reducing chamber. • The methane pass through the reducing chamber unaffected, while CO is reduced to methane. • By using hydrogen flame ionization detector, both methane peaks can be detected. • The first peak is due to methane and second peak is due to CO. • Accuracy is ±2% • Peak heights of CO and methane give sensitivity of about 50ppb.
  • 12. NOX ANALYZER • COLORIMETRY • CHEMILUMINESCENCE • USE OF CO LASER • LASER OPTO-ACOUSTIC SPETROSCOPY
  • 13. COLORIMETRY • This method can be used for the determination of nitrogen dioxide by using Saltzman method. • This method is based on a reaction in which a pink colored dye complex is formed when air containing NO2 is passed in an absorbing solution consisting of sulphanilic acid and diamine dissolved in the acetic medium. • Sensitivity is in ppm range. • Nitric oxide may be determined by passing the sample through an acid permanganate bubbler which oxidizes nitric oxide to nitrogen dioxide. • Used for monitoring in stack sewages in concentration range of 5 ppm to several thousand ppm, consists in passing sample into an evacuated flask containing a solution of H2O2 in sulphuric acid. • The oxides of nitrogen are converted into nitric acid and nitrate ions react with phenoldisupluric acid to produce a yellow color, which is measured calorimetrically.
  • 14. CHEMILUMINESCENCE • The phenomenon of emission of radiation from a chemically excited species is known as chemiluminescence. It results due to the formation of new chemical bonds .The atoms in the excited states posses higher energy levels than the ground state & while returning to the ground state they emit radiations. • This phenomenon is useful for measurement of air pollutants, particularly NO, NO2
  • 15. CONT., • The following reaction uses for the measurement of chemiluminescence emission. • Since NO2 reacts slowly with ozone and the reaction which produces NO3 is not done by chemiluminescence, so it is necessary to reduce NO2 to NO before admission into reactor. NO2 NO +1/2 O2 • Light emission is measured with photmultiplier with high gain and low dark current can have low levels of radiation can be measured. • The response of instruments based on this method is linear from 1ppm to 1000ppm of NO. • Used extremely in automative exhaust gases.
  • 16. MEASUREMENT OF NITROGEN OXIDE USING CO LASERS FROM R.S.KHANDPUR
  • 17. • This method detects nitric oxide in 0.25ppm. • The apparatus consists of a CO laser, which emits radiation that is absorbed by NO in the mixture, the amount of absorption is proportional to the concentration of NO present. • The wavelength match between laser and NO is made exact and hence the absorption is enhanced by placing NO in a magnetic field. • The CO laser used is a dc excited continuous working laser, which operates at a single wavelength of 5.307µm. • A diffraction grating is used at one end of the cavity as a line selector. • The laser yields 5-30mW of single line power. • Absorption cell is made of pyrex (low thermal expansion borosilicate glass) of diameter 15mm and 90cm length.
  • 18. • Pressure applied is 10^-6- 10^-5 torr. • To produce modulating audio frequency magnetic field along the axis of the cell, insulated wire is closely wound around the outside of the cell over about half its length. • The coil is excited with a current of 1A in the frequency range of 5-150KHz. • This produces a varying magnetic field of about 50Guass P- P intensity. • The DC magnetic field is produced by a solenoid which produce a field upto 2.5KGauss. • The detector is a liquid nitrogen cooled Ge-Au element. • The signal is amplied in a lock in amplifier(extracts a signal with a known carrier wave from an extremely noise environment) before given to recorder. • The signal amplitude varies linearly with the concentration of NO in the sample.
  • 19. LASER OPTO-ACOUSTIC SPETROSCOPY • Opto acoustic detectors combined with thermal IR sources has been widely used in gas detection. • This technique is used to measure trace amounts of nitrogen oxides in the stratosphere. • Opto acoustic detector involves the absorption of an AM beam of IR by a gas which results in the generation of sound. • The energy absorbed by the gas molecules from IR beam excites the molecules to the rotational-vibrational energy levels above the ground state. • The main path for the delay of these excited states is collisional de- excitation which results in the transfer of absorbed energy into heat and raises the gas temperature. • The temperature rise causes a corresponding pressure rise in the gas.
  • 20. Block Diagram of laser opto acoustic detector for nitric oxides FROM R.S.KHANDPUR
  • 21. • When the beam intensity is modulated, the gas temperature and pressure changes accordingly. • Thus periodic pressure vibrations in the ags result in the generation of sound. • The arrangement uses CO2 laser, tuned by rotating a diffraction grating at ine end of the laser cavity. • It is tunable to 64 different emission lines in the range from 927 to 1085cm-1 • The laser beam is focused at chopping wheel and then focused into the detector. • The chopping frequency is selected to optimize the SNR. • The microphone is available and acoustic signal is amplified in a preamplifier and displayed. • The detector is highly sensitive and sample amount of 20pg may be detected.
  • 22. SULPHUR DIOXIDE MEASUREMENT • Colorimetry • Conductivitimetry • Gas chromatography • Coulometry • Flame photometric detector • UV fluorescence method
  • 23. Colorimetry • A known volume of air is passed through an aqueous solution,which contains reagents that absorb so2 and produce a colored substance. • The amount of colored substance is proportional to so2. • The air to be analysed is passed through an aqueous solution of 0.1M sodium tetracholromercurate. • The so2 reacts with mercuric salt to form a sulphatomercuric compound. • The solution is further treated with acid bleached para- rasoniline and formaldehyde, which forms a red purple colour. • The amount of color is determined photometrically.
  • 24. • So2 can also be determined by means of decolorisation of an aqueous starch iodine solution which normally has deep blue color. • When so2 is passed through the solution containing starch iodine solution,the iodine reduced by sulphur dioxide which in turn results in decrease of intensity of starch iodine color.
  • 25. Conductivitimetry • This technique is based on measuring the change in conductivity of a solution, when a sample of air containing so2 is bubbled through it. • C- cell • P-perspex cap • J-jet • E-electrodes FROM R.S.KHANDPUR
  • 26. coulometry • The ability of sulphur dioxide to reduce iodine can be employed to monitor so2 by using iodine coulometry in an aqueous solution. • The mass of iodine reacted per unit time during any given interval of time would indicate concentration of so2 in air. • The coulometric arrangement requires two electrodes made of platinum which acts as anode and cathode. • These electrodes maintain a trace of iodine in equilibrium with potassium bromide to maintain conductance. • The shift in anode-cathode potential is sensed by a third reference electrode, when so2 from sample air is bubbled through the solution in a detector cell.
  • 27. Flame photometric detector • In Flame photometers, sample air is introduced into hydrogen rich air flame by using a narrow band interference filter shields the PMT detectors, but 394nm emission energy of flame excited sulphur atoms.
  • 28. UV fluorescence method • This method is based on the principle that so2 molecules absorb UV light and became excited at a particular wavelength. • These molecules then decay to a lower energy sate emitting UV light at a different wavelength. • The intensity of fluorescence is proportional to the so2 concentration.
  • 29. Block diagram of fluorescence so2 analyser
  • 30. Hydrocarbons estimation • Flame ionization detector • Gas chromatography
  • 31. Flame ionization detector • Organic compounds easily pyrolyse when introduced into an air-hydrogen flame. • The pyrolysis produces ions that can be collected by having a cylindrical grid surrounding the flame. • The detector response would be in proportion to the number of carbon atoms in the chain.
  • 32. Schematic diagram of FID FROM R.S.KHANDPUR
  • 33. Working principle • The sample gas containing hydrocarbons controlled at a constant flow rate is mixed with hydrogen for fuel. • The mixed gas is burned at the end of a very fine nozzle. • Two electrodes are placed on either side of the flame and an appropriate electric field is applied to them. • Ionic current will flow between the collector electrode and the other electrodes. • The current is then amplified and displayed on a meter.
  • 34. Gas chromatography • Gas chromatography has been applied for the detection and measurement of hydrocarbons, CO,CO2 in air. • They are generally detected by FID since they are not sensitive to CO and CO2. • These are first converted into methane by hydrogenation and then measured.