CT ARTIFACTS
Definition – Artifacts
are defined as
anything that appears
on the image that is
not on the
object being scanned
CLASSIFICATION
1.Physics based artifacts – Result from the physical
processes involved in the data acquisition.
Examples are photon starvation and beam
hardening.
2.Patient based artifacts- Caused by patient
movement or presence of metallic objects in the
patient body.
Example of these type of artifacts are metallic
artifacts and motion artifacts
3.Scan based artifacts –Results from imperfection in
scanners functioning.
Examples are ring artifacts
A).BEAM
HARDENING
ARTIFACTS
• Caused by a polychromatic beam .
• This beam hardening artifact also produces
cupping artifact. The center of the object is
usually the thickest and therefore, the
beam will be harder in the center than at
the periphery of the image and is assigned
lower Hounsfield units.
• This can be corrected by beam correction
algorithms.
• Filtering can also reduce beam hardening
by creating a monochromatic x-ray beam.
• Prepatient filter – absorbs soft x-rays.
• Bow filter – prehardens the x-ray beam.
B).METALLIC
ARTIFACTS.
Highly dense objects in
and around patient can
cause significant
streaking.
Appearance; Streaking
on the image.
• Remove metal objects.
• Decrease section width.
• Increase kVp for better penetrability.
• Adjust window level settings.
• Reorient data acquisition (e.g. gantry tilt)
• Metallic Artifact Reduction (MAR) software.
MAR uses interpolation to substitute over
range values in attenuation profiles.
Solutions:
C).PHOTON
STARVATION
• Parts of individual projections can be very
noisy due to insufficient photons passing
through the widest part of the patient.
• When these views are reconsructed,the
noise is magnified, resulting in streaks.
• Example is scan across the shoulders.
Smaller signals reach the detector, therefore
large proportion of noise.
• Solution :Adaptive filtering- the regions in
which the attenuation exceeds
a specified level are smoothed before
undergoing back projection .
• Milliampere (mA) modulation –higher mAs is
used for the more attenuation profile to
reduce the effect of photon starvation.
D).PARTIAL VOLUME ARTIFACTS.
• Partial volume effect occurs when two objects of different densities are occupying the same
plane.
• This may cause the object of interest to not fill the plane resulting in an underestimated
CT number.
• Partial volume artifacts occur when an off center object is occupying part of the scan plane.
• The inconsistences between views causes streaks.
• Solution : Use thinner slices.
E). RING ARTIFACTS.
• Ring artifacts occur due to malfunctioning of one or more of the
detectors in the array.
• Due to failure of a particular detector or miscalibration, incorrect
data in every projection will appear as a ring in the image.
• Solution : recalibrate the detectors.
• If rings persist even after recalibration ,call service.
F).CONE BEAM ARTIFACTS.
• Only in multi detector CT systems.
• As the section being scanned increases per rotation, a wider collimation is
used.
• Because of this the x-ray beam becomes cone shaped instead of fan shaped
and area imaged by each detector as it rotates around the patient is now a
volume instead of a flat plane .
• It appears as lines in windmill formation particularly at the edges of the
image.
• Solution :Use pitch selection recommended by the manufacturer. Most
modern scanner cone beam reconstruction algorithms correct this artifact.
G). TUBE ARCING
TUBE ARCING IS CAUSED BY A
TEMPORARY SHORT CIRCUIT IN THE
X-RAY TUBE CAUSING MOMENTARY
LOSS OF X-RAY OUTPUT.
IT IS SEEN ON THE IMAGE AS NEAR-
PARALLEL AND EQUIDISTANT STREAK
PATTERN ON TRANS AXIAL CT IMAGES
AND AS A HORIZONTAL HYPODENSE
BAND ON THE CORONAL AND SAGITTAL
CT IMAGES.
SOLUTION : CALL SERVICE.
H).STEP
ARTIFACTS
• Stair step artifacts occur around edges
of structures in three dimensional (3D)
and multi planar reconstruction (MPR)
images when wide collimations and
non-overlapping reconstruction
intervals are used.
• They are found in straight structures
which are oriented obliquely with
respect to movement of the table.
• Solution : Reduce collimation.
• Use overlapping reconstruction in
helical imaging.
• Thinner acquisition slices.
I).MOTION
ARTIFACTS.
• Motion artifacts are caused by the
voluntary or involuntary movement of the
patient during data acquisition.
• It appears as shading, streaking and
blurring.
• Solution : use shortest scan time as
possible.
• Immobilization aids.
• Consider sedation.
• Beta blockers for cardiac protocols.
• Special software correction which reduces
weighting of views obtained towards the
beginning and end of the scan.
J).OUT-OF-FIELD ARTIFACTS.
• These occur when a portion of the patient is outside the field of view (FOV).
• It is seen on the image as streaks.
• Common problem areas include shoulders, abdomen and lower pelvis.
• Solution : adjust patient position (L/R).
• Expand the FOV if possible.
K).SPIRAL INTERPOLATION ARTIFACTS.
• These are caused by images that are created from views that are not in the same plane.
• This results in inaccuracies in CT number.
• Appearance : streaks
• Solution : Lower pitch.
L).ALIASING ARTIFACTS
• Also known as under sampling.
• Refers to the error in the accuracy component of the ADC during image
digitization.
• When sampling, the brightness of each pixel in the image is measured and
creates an analog signal.
• The more the samples, the more accurate the representation of signal will
be,hence,if a lack of sampling has occurred the computer will process an
inaccurate image resulting in an aliasing artifact.
• Appearance : Fine lines.
• Solution :reduce pitch
• Acquiring the largest possible number of projections per rotation.(on some
scanners this can be achieved by using slower rotation speed)
M).EDGE GRADIENT EFFECT
• The edge gradient effect results in shading arising from irregularly
shaped objects that have different densities.
• Occurs when barium and air lie adjacent to each other in the
stomach .
• Appearance : straight lines radiating from high-contrast areas.
• Solution :Use thinner slices, Use low or neutral HU value oral contrast
in place Barium.
THANK YOU !!!!!!

More Related Content

What's hot

Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Ali Ayaz
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
parthajyotidas11
 
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
Prem Murti
 

What's hot (20)

CT artifact
CT artifact CT artifact
CT artifact
 
Helical and Multislice CT
Helical and Multislice CTHelical and Multislice CT
Helical and Multislice CT
 
Principles of mdct
Principles of mdctPrinciples of mdct
Principles of mdct
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomography
 
CT numbers, window width and window level
CT numbers, window width and window levelCT numbers, window width and window level
CT numbers, window width and window level
 
Ct instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationCt instrumentation and types of detector configuration
Ct instrumentation and types of detector configuration
 
Computed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationComputed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector Configuration
 
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
 
CT SCAN - 3 Marks - QUESTION AND ANSWERS
CT SCAN - 3 Marks - QUESTION AND ANSWERSCT SCAN - 3 Marks - QUESTION AND ANSWERS
CT SCAN - 3 Marks - QUESTION AND ANSWERS
 
Radiographic grids
Radiographic gridsRadiographic grids
Radiographic grids
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
soft tissue radiography
soft tissue radiographysoft tissue radiography
soft tissue radiography
 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
 
Basics in ct
Basics in ctBasics in ct
Basics in ct
 
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...tomography tomography - Presentation Transcript     1. PRINCIPLE OFTOMOGRAPHY...
tomography tomography - Presentation Transcript 1. PRINCIPLE OFTOMOGRAPHY...
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Advances in ct technology
Advances in ct technologyAdvances in ct technology
Advances in ct technology
 
Gradient echo pulse sequence swastik
Gradient echo pulse sequence  swastikGradient echo pulse sequence  swastik
Gradient echo pulse sequence swastik
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 

Similar to Ct image artifacts

ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
Shubham661884
 
5.2. lithography 3,4,5 final,2013
5.2. lithography 3,4,5 final,20135.2. lithography 3,4,5 final,2013
5.2. lithography 3,4,5 final,2013
Bhargav Veepuri
 
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptxLecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
faiz3334
 

Similar to Ct image artifacts (20)

Ct artiact
Ct artiactCt artiact
Ct artiact
 
Artifacts
ArtifactsArtifacts
Artifacts
 
Cone beam ct
Cone beam ctCone beam ct
Cone beam ct
 
Cbct
CbctCbct
Cbct
 
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSDIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
 
Physics_of_CT , CT machine and it’s parts, ct generations
Physics_of_CT , CT machine and it’s parts, ct generationsPhysics_of_CT , CT machine and it’s parts, ct generations
Physics_of_CT , CT machine and it’s parts, ct generations
 
Cbct
CbctCbct
Cbct
 
ct image quality
ct image qualityct image quality
ct image quality
 
COMPUTED TOMOGRAPHY upload.pptx
COMPUTED TOMOGRAPHY upload.pptxCOMPUTED TOMOGRAPHY upload.pptx
COMPUTED TOMOGRAPHY upload.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
Back projection geometry in cbct
Back projection geometry in cbctBack projection geometry in cbct
Back projection geometry in cbct
 
color doppler
color dopplercolor doppler
color doppler
 
Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedy
 
5.2. lithography 3,4,5 final,2013
5.2. lithography 3,4,5 final,20135.2. lithography 3,4,5 final,2013
5.2. lithography 3,4,5 final,2013
 
Computed tomography - Physics and Clinical
Computed tomography - Physics and ClinicalComputed tomography - Physics and Clinical
Computed tomography - Physics and Clinical
 
Ct ppt 2
Ct ppt 2Ct ppt 2
Ct ppt 2
 
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxLCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
 
Chapter 1 basic principles of ct
Chapter 1 basic principles of ctChapter 1 basic principles of ct
Chapter 1 basic principles of ct
 
Digital image processing
Digital image processingDigital image processing
Digital image processing
 
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptxLecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
 

Recently uploaded

❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
daljeetkaur2026
 
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
dilpreetentertainmen
 
❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...
❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...
❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...
chandigarhentertainm
 
👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...
👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...
👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...
minkseocompany
 
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North CarolinaTIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
Mebane Rash
 

Recently uploaded (19)

The Events of Cardiac Cycle - Wigger's Diagram
The Events of Cardiac Cycle - Wigger's DiagramThe Events of Cardiac Cycle - Wigger's Diagram
The Events of Cardiac Cycle - Wigger's Diagram
 
❤️ Call Girls service In Panchkula☎️9815457724☎️ Call Girl service in Panchku...
❤️ Call Girls service In Panchkula☎️9815457724☎️ Call Girl service in Panchku...❤️ Call Girls service In Panchkula☎️9815457724☎️ Call Girl service in Panchku...
❤️ Call Girls service In Panchkula☎️9815457724☎️ Call Girl service in Panchku...
 
2024 PCP #IMPerative Updates in Rheumatology
2024 PCP #IMPerative Updates in Rheumatology2024 PCP #IMPerative Updates in Rheumatology
2024 PCP #IMPerative Updates in Rheumatology
 
💞 Safe And Secure Call Girls Coimbatore 🧿 9332606886 🧿 High Class Call Girl S...
💞 Safe And Secure Call Girls Coimbatore 🧿 9332606886 🧿 High Class Call Girl S...💞 Safe And Secure Call Girls Coimbatore 🧿 9332606886 🧿 High Class Call Girl S...
💞 Safe And Secure Call Girls Coimbatore 🧿 9332606886 🧿 High Class Call Girl S...
 
💸Cash Payment No Advance Call Girls Pune 🧿 9332606886 🧿 High Class Call Girl ...
💸Cash Payment No Advance Call Girls Pune 🧿 9332606886 🧿 High Class Call Girl ...💸Cash Payment No Advance Call Girls Pune 🧿 9332606886 🧿 High Class Call Girl ...
💸Cash Payment No Advance Call Girls Pune 🧿 9332606886 🧿 High Class Call Girl ...
 
❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...❤️ Zirakpur Call Girl Service  ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
❤️ Zirakpur Call Girl Service ☎️9878799926☎️ Call Girl service in Zirakpur ☎...
 
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
 
👉Bangalore Call Girl Service👉📞 6378878445 👉📞 Just📲 Call Manisha Call Girls Se...
👉Bangalore Call Girl Service👉📞 6378878445 👉📞 Just📲 Call Manisha Call Girls Se...👉Bangalore Call Girl Service👉📞 6378878445 👉📞 Just📲 Call Manisha Call Girls Se...
👉Bangalore Call Girl Service👉📞 6378878445 👉📞 Just📲 Call Manisha Call Girls Se...
 
❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...
❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...
❤️ Escorts Service in Bangalore ☎️81279-924O8☎️ Call Girl service in Bangalor...
 
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
💞 Safe And Secure Call Girls Mysore 🧿 9332606886 🧿 High Class Call Girl Servi...
 
Making change happen: learning from "positive deviancts"
Making change happen: learning from "positive deviancts"Making change happen: learning from "positive deviancts"
Making change happen: learning from "positive deviancts"
 
👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...
👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...
👉Indore Call Girl Service👉📞 7718850664 👉📞 Just📲 Call Anuj Call Girls In Indor...
 
💸Cash Payment No Advance Call Girls Nagpur 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Nagpur 🧿 9332606886 🧿 High Class Call Gir...💸Cash Payment No Advance Call Girls Nagpur 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Nagpur 🧿 9332606886 🧿 High Class Call Gir...
 
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North CarolinaTIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
 
💸Cash Payment No Advance Call Girls Kanpur 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Kanpur 🧿 9332606886 🧿 High Class Call Gir...💸Cash Payment No Advance Call Girls Kanpur 🧿 9332606886 🧿 High Class Call Gir...
💸Cash Payment No Advance Call Girls Kanpur 🧿 9332606886 🧿 High Class Call Gir...
 
💸Cash Payment No Advance Call Girls Hyderabad 🧿 9332606886 🧿 High Class Call ...
💸Cash Payment No Advance Call Girls Hyderabad 🧿 9332606886 🧿 High Class Call ...💸Cash Payment No Advance Call Girls Hyderabad 🧿 9332606886 🧿 High Class Call ...
💸Cash Payment No Advance Call Girls Hyderabad 🧿 9332606886 🧿 High Class Call ...
 
💞 Safe And Secure Call Girls gaya 🧿 9332606886 🧿 High Class Call Girl Service...
💞 Safe And Secure Call Girls gaya 🧿 9332606886 🧿 High Class Call Girl Service...💞 Safe And Secure Call Girls gaya 🧿 9332606886 🧿 High Class Call Girl Service...
💞 Safe And Secure Call Girls gaya 🧿 9332606886 🧿 High Class Call Girl Service...
 
Call Girls In Indore 💯Call Us 🔝 9987056364 🔝 💃 Independent Escort Service Ind...
Call Girls In Indore 💯Call Us 🔝 9987056364 🔝 💃 Independent Escort Service Ind...Call Girls In Indore 💯Call Us 🔝 9987056364 🔝 💃 Independent Escort Service Ind...
Call Girls In Indore 💯Call Us 🔝 9987056364 🔝 💃 Independent Escort Service Ind...
 
❤️Amritsar Escort Service☎️98151-129OO☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escort Service☎️98151-129OO☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escort Service☎️98151-129OO☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escort Service☎️98151-129OO☎️ Call Girl service in Amritsar☎️ Amri...
 

Ct image artifacts

  • 2. Definition – Artifacts are defined as anything that appears on the image that is not on the object being scanned CLASSIFICATION 1.Physics based artifacts – Result from the physical processes involved in the data acquisition. Examples are photon starvation and beam hardening. 2.Patient based artifacts- Caused by patient movement or presence of metallic objects in the patient body. Example of these type of artifacts are metallic artifacts and motion artifacts 3.Scan based artifacts –Results from imperfection in scanners functioning. Examples are ring artifacts
  • 3. A).BEAM HARDENING ARTIFACTS • Caused by a polychromatic beam . • This beam hardening artifact also produces cupping artifact. The center of the object is usually the thickest and therefore, the beam will be harder in the center than at the periphery of the image and is assigned lower Hounsfield units. • This can be corrected by beam correction algorithms. • Filtering can also reduce beam hardening by creating a monochromatic x-ray beam. • Prepatient filter – absorbs soft x-rays. • Bow filter – prehardens the x-ray beam.
  • 4. B).METALLIC ARTIFACTS. Highly dense objects in and around patient can cause significant streaking. Appearance; Streaking on the image. • Remove metal objects. • Decrease section width. • Increase kVp for better penetrability. • Adjust window level settings. • Reorient data acquisition (e.g. gantry tilt) • Metallic Artifact Reduction (MAR) software. MAR uses interpolation to substitute over range values in attenuation profiles. Solutions:
  • 5. C).PHOTON STARVATION • Parts of individual projections can be very noisy due to insufficient photons passing through the widest part of the patient. • When these views are reconsructed,the noise is magnified, resulting in streaks. • Example is scan across the shoulders. Smaller signals reach the detector, therefore large proportion of noise. • Solution :Adaptive filtering- the regions in which the attenuation exceeds a specified level are smoothed before undergoing back projection . • Milliampere (mA) modulation –higher mAs is used for the more attenuation profile to reduce the effect of photon starvation.
  • 6. D).PARTIAL VOLUME ARTIFACTS. • Partial volume effect occurs when two objects of different densities are occupying the same plane. • This may cause the object of interest to not fill the plane resulting in an underestimated CT number. • Partial volume artifacts occur when an off center object is occupying part of the scan plane. • The inconsistences between views causes streaks. • Solution : Use thinner slices.
  • 7. E). RING ARTIFACTS. • Ring artifacts occur due to malfunctioning of one or more of the detectors in the array. • Due to failure of a particular detector or miscalibration, incorrect data in every projection will appear as a ring in the image. • Solution : recalibrate the detectors. • If rings persist even after recalibration ,call service.
  • 8. F).CONE BEAM ARTIFACTS. • Only in multi detector CT systems. • As the section being scanned increases per rotation, a wider collimation is used. • Because of this the x-ray beam becomes cone shaped instead of fan shaped and area imaged by each detector as it rotates around the patient is now a volume instead of a flat plane . • It appears as lines in windmill formation particularly at the edges of the image. • Solution :Use pitch selection recommended by the manufacturer. Most modern scanner cone beam reconstruction algorithms correct this artifact.
  • 9. G). TUBE ARCING TUBE ARCING IS CAUSED BY A TEMPORARY SHORT CIRCUIT IN THE X-RAY TUBE CAUSING MOMENTARY LOSS OF X-RAY OUTPUT. IT IS SEEN ON THE IMAGE AS NEAR- PARALLEL AND EQUIDISTANT STREAK PATTERN ON TRANS AXIAL CT IMAGES AND AS A HORIZONTAL HYPODENSE BAND ON THE CORONAL AND SAGITTAL CT IMAGES. SOLUTION : CALL SERVICE.
  • 10. H).STEP ARTIFACTS • Stair step artifacts occur around edges of structures in three dimensional (3D) and multi planar reconstruction (MPR) images when wide collimations and non-overlapping reconstruction intervals are used. • They are found in straight structures which are oriented obliquely with respect to movement of the table. • Solution : Reduce collimation. • Use overlapping reconstruction in helical imaging. • Thinner acquisition slices.
  • 11. I).MOTION ARTIFACTS. • Motion artifacts are caused by the voluntary or involuntary movement of the patient during data acquisition. • It appears as shading, streaking and blurring. • Solution : use shortest scan time as possible. • Immobilization aids. • Consider sedation. • Beta blockers for cardiac protocols. • Special software correction which reduces weighting of views obtained towards the beginning and end of the scan.
  • 12. J).OUT-OF-FIELD ARTIFACTS. • These occur when a portion of the patient is outside the field of view (FOV). • It is seen on the image as streaks. • Common problem areas include shoulders, abdomen and lower pelvis. • Solution : adjust patient position (L/R). • Expand the FOV if possible. K).SPIRAL INTERPOLATION ARTIFACTS. • These are caused by images that are created from views that are not in the same plane. • This results in inaccuracies in CT number. • Appearance : streaks • Solution : Lower pitch.
  • 13. L).ALIASING ARTIFACTS • Also known as under sampling. • Refers to the error in the accuracy component of the ADC during image digitization. • When sampling, the brightness of each pixel in the image is measured and creates an analog signal. • The more the samples, the more accurate the representation of signal will be,hence,if a lack of sampling has occurred the computer will process an inaccurate image resulting in an aliasing artifact. • Appearance : Fine lines. • Solution :reduce pitch • Acquiring the largest possible number of projections per rotation.(on some scanners this can be achieved by using slower rotation speed)
  • 14. M).EDGE GRADIENT EFFECT • The edge gradient effect results in shading arising from irregularly shaped objects that have different densities. • Occurs when barium and air lie adjacent to each other in the stomach . • Appearance : straight lines radiating from high-contrast areas. • Solution :Use thinner slices, Use low or neutral HU value oral contrast in place Barium.