sensors
Article
Development and Test of a Portable ECG Device with Dry
Capacitive Electrodes and Driven Right Leg Circuit
Alessandro Zompanti 1,* , Anna Sabatini 1 , Simone Grasso 2 , Giorgio Pennazza 1 , Giuseppe Ferri 3 ,
Gianluca Barile 3 , Massimo Chello 4 , Mario Lusini 4 and Marco Santonico 2


Citation: Zompanti, A.; Sabatini, A.;
Grasso, S.; Pennazza, G.; Ferri, G.;
Barile, G.; Chello, M.; Lusini, M.;
Santonico, M. Development and Test
of a Portable ECG Device with Dry
Capacitive Electrodes and Driven
Right Leg Circuit. Sensors 2021, 21,
2777. https://doi.org/10.3390/
s21082777
Academic Editor: Giancarlo Fortino
Received: 24 March 2021
Accepted: 13 April 2021
Published: 15 April 2021
Publisher’s Note: MDPI stays neutral
with regard to jurisdictional claims in
published maps and institutional affil-
iations.
Copyright: © 2021 by the authors.
Licensee MDPI, Basel, Switzerland.
This article is an open access article
distributed under the terms and
conditions of the Creative Commons
Attribution (CC BY) license (https://
creativecommons.org/licenses/by/
4.0/).
1 Unit of Electronics for Sensor Systems, Department of Engineering, Campus Bio-Medico University of Rome,
00128 Rome, Italy; a.sabatini@unicampus.it (A.S.); g.pennazza@unicampus.it (G.P.)
2 Unit of Electronics for Sensor Systems, Department of Science and Technology for Humans and the
Environment, Campus Bio-Medico University of Rome, 00128 Rome, Italy; s.grasso@unicampus.it (S.G.);
m.santonico@unicampus.it (M.S.)
3 Department of Industrial and Information Engineering and Economics, University of L’Aquila,
67100 L’Aquila, Italy; giuseppe.ferri@univaq.it (G.F.); gianluca.barile@univaq.it (G.B.)
4 Unit of Cardiology, Department of Medicine, Campus Bio-Medico University of Rome, 00128 Rome, Italy;
m.chello@unicampus.it (M.C.); m.lusini@unicampus.it (M.L.)
* Correspondence: a.zompanti@unicampus.it
Abstract: The use of wearable sensors for health monitoring is rapidly growing. Over the past
decade, wearable technology has gained much attention from the tech industry for commercial
reasons and the interest of researchers and clinicians for reasons related to its potential benefit on
patients’ health. Wearable devices use advanced and specialized sensors able to monitor not only
activity parameters, such as heart rate or step count, but also physiological parameters, such as
heart electrical activity or blood pressure. Electrocardiogram (ECG) monitoring is becoming one of
the most attractive health-related features of modern smartwatches, and, because cardiovascular
disease (CVD) is one of the leading causes of death globally, the use of a smartwatch to monitor
patients could greatly impact the disease outcomes on health care systems. Commercial wearable
devices are able to record just single-lead ECG using a couple of metallic contact dry electrodes. This
kind of measurement can be used only for arrhythmia diagnosis. For the diagnosis of other cardiac
disorders, additional ECG leads are required. In this study, we characterized an electronic interface
to be used with multiple contactless capacitive electrodes in order to develop a wearable ECG device
able to perform several lead measurements. We verified the ability of the electronic interface to
amplify differential biopotentials and to reject common-mode signals produced by electromagnetic
interference (EMI). We developed a portable device based on the studied electronic interface that
represents a prototype system for further developments. We evaluated the performances of the
developed device. The signal-to-noise ratio of the output signal is favorable, and all the features
needed for a clinical evaluation (P waves, QRS complexes and T waves) are clearly readable.
Keywords: wearable sensors and devices; wearable monitoring systems; wearable healthcare elec-
trocardiography; wireless ECG monitoring systems; electronics for biosignals and biopotentials;
electronic interfaces and embedded sensor systems for biomedical application; dry electrodes; capaci-
tive coupling electrodes
1. Introduction
The use of point-of-care (POC) systems and wearable sensors for health monitoring is
rapidly growing. They are used both for clinical and for remote monitoring, improving
the possibility to collect health data about patients continuously and to prevent health
deterioration or clinical events [1].
POC and wearable sensors are able to empower patients, providing them with the
right tools to monitor themselves autonomously. Patients are able to cooperate with medical
Sensors 2021, 21, 2777. https://doi.org/10.3390/s21082777 https://www.mdpi.com/journal/sensors
Sensors 2021, 21, 2777 2 of 15
teams and contribute to their own health [2]. Moreover, remote monitoring devices prevent
the hospitalization of patients, with a consequent cost reduction for the healthcare system.
Several data can be gathered using wearable sensors in order to monitor and record
real-time information about patients’ health. Sensors measure physiological signs such as
an electrocardiogram (ECG), an electromyogram (EMG), heart rate (HR), body temperature,
electrodermal activity (EDA), arterial oxygen saturation (SpO2), blood pressure (BP) and
respiration rate (RR) [3].
There is currently a growing interest in using wearable devices to monitor cardiovas-
cular diseases [4]. According to recent statistics, cardiovascular disease (CVD) is one of the
leading causes of death globally. In Europe as a whole, it is responsible for over 3.9 million
deaths a year, or 45% of all deaths [5].
Over the past decade, wearable technology has gained the interest of researchers and
clinicians due to the benefits that could be associated with the long-term monitoring of
CVD patients, not necessarily at a hospital. Moreover, wearable devices have attracted
much attention from the tech industry. Even if the primary use of commercial wearable
devices is to monitor activity parameters (heart rate, step count, etc.), the sensors used
are now so advanced that it is possible to also monitor physiological parameters directly
related to CVD, such as heart electrical activity or blood pressure.
Among the list of features forming the wearable technology paradigm (which can be
found in many review papers or textbooks devoted to this subject [6–9]), it is here important
to discuss those in this paper. A wearable device, especially for medical use, should be
comfortable for the patient and adequate to follow movements. If possible, it should be
effectively wearable, meaning that it could be possible to wear it with clothes. It should
have electrical/electronic characteristics that make it the least disturbed by interferences.
Several commercial wearable devices, such as Apple Watch (Apple Inc., Cupertino, CA,
USA), Fitbit Sense (Fitbit Inc., San Francisco, CA, USA) or Withings ScanWatch (Withings,
Issy-les-Moulineaux, France), are able to record a single-lead ECG using metallic dry-
contact electrodes, in contrast to adhesive wet electrodes [10]. Typically, one electrode is
placed on the front of the watch and can be touched with a finger, and another electrode is
placed on the back, in contact with the wrist. This one-lead ECG can improve arrhythmia
diagnosis; however, for the diagnosis of other cardiac disorders, such as myocardial
infarction or ischemia, additional ECG leads are required [11].
Traditional 12-lead ECG is able to record cardiac potentials on the skin, with an am-
plitude of a few mV and frequencies in the range of 0.01–150 Hz. The measurement has
been performed for a limited time with the patient at rest, laying on an examination table.
Several cables and standard wet Ag/AgCl electrodes are used. This type of electrode
requires proper preparation of the skin [12]. Moreover, adhesive wet electrodes are not
suitable for long-term monitoring purposes because they can cause irritation and lose their
conductivity over time due to drying out [13]. As a consequence, dry-contact electrodes
represent an alternative to wet contact electrodes [14], in particular for long-term monitor-
ing, where they are more comfortable for the patient [15,16]. In recent years, noncontact
electrodes have been used in several studies [17]. In particular, capacitive electrodes, cou-
pled with the patient’s body, have been investigated [18–24]. A smart way to use more
leads, marginally compromising patient comfort, could be by embedding the electrodes
into clothes. Dry capacitive electrodes are highly suitable for this kind of application. In
order to use capacitive electrodes, new electronic interfaces should be developed to match
the application requirements. The two main requirements are electronic interface high
impedance and EMI rejection.
In this study, the possibility of recording multiple-lead ECG using dry capacitive elec-
trodes with a custom-developed portable device was investigated to explore the feasibility
of a wearable ECG device with electrodes embedded into clothes. The performances of
an electronic interface proposed by Chi et al. [25] and others [26,27] were evaluated by
characterizing it with simulations and electrical tests in order to deeper investigate its
performances. Moreover, the studied electronic interface was implemented into a custom
Sensors 2021, 21, 2777 3 of 15
realized portable device, and in order to verify the effectiveness of the system, a comparison
between signals obtained using the developed device, with dry capacitive electrodes, and
a commercial heart rate front-end IC, with standard wet electrodes, was performed.
The novelty of this study is the deeper understanding of the studied electronic inter-
face behavior, with respect to published results. This goal was achieved by characterizing
the electronic interface in detail using simulations and electrical tests. An opportune
experimental setup was designed to perform the measurements with experimental condi-
tions as close as possible to real applications; that is, an “ECG T-shirt” with an embedded
capacitive electrode.
According to the obtained results, a specific circuit optimization method was proposed.
Moreover, these results could be exploited by numerous real applications, such as drivers’
safety [21,28–33], remote monitoring of the elderly [34–37], sleep monitoring [20,38–42],
human identification [43–46] and emotion recognition [47–50].
2. Materials and Methods
Biopotential signals, such as those generated by the electrical activity of the heart,
are usually measured with a differential approach using voltage amplifiers with at least
two basic requirements: input impedance, which should be as high as possible in order
to match the electrode impedance, and the current flowing between the body and the
amplifiers, which should be as low as possible for safety reasons.
Biopotential measurements, such as ECG, are very sensitive to electromagnetic inter-
ference (EMI). One of the most experienced EMI is caused by the coupling between the
patient’s body and the power lines, resulting in an applied common-mode voltage (Vcm)
on the electrodes.
By performing differential biopotential measurement and improving the common-
mode rejection (CMR) of the system, it is possible to contain this phenomenon. Differential
measurement can be obtained by applying the biopotentials to the input stage of a classic
instrumentation amplifier (INA) topology.
EMI generates an Icm flowing through the patient’s body from the sinusoidal power-
line source (120 V/60 Hz or 230/50 Hz) through coupling capacitances to the ground. In
order to perform an effective ECG measurement, the patient’s body has to be grounded
with the ECG device. This grounding connection contributes to the rise in the Vcm primarily
because of its resistance. As a consequence, it is important to reduce the impedance of the
grounding electrode or to compensate for the EMI-generated current. This result can be
achieved using a driven right leg (DRL) electrode [51]. It is commonly applied to the right
leg, but it can be applied elsewhere on the patient’s body.
The DRL electrode senses the common-mode voltage on the body, giving feedback to
the patient’s body. This negative feedback reduces Vcm to µV values [52].
This study investigates the potentialities of a suitable circuit topology, originally
proposed by Chi et al. [25], to enhance its applicability via a deeper characterization, which
should allow exploiting this strategy for innovative devices.
The proposed electronic system uses two sensing capacitive electrodes and a DRL
capacitive electrode, as shown in Figure 1.
The capacitive electrodes can be placed on the patient’s clothes (e.g., a cotton shirt)
and not be in contact with the skin. Therefore, from now on, we will refer to the system as
a contactless ECG device.
Each electrode is an input of the buffering stage of a differential amplifier based
on the topology of a 3-input amplifier. The DRL is made of an inverting opamp and a
buffering stage.
The electrodes are active and use amplifiers in order to achieve a very high input
impedance, as shown in Figure 2. A high-pass filtering stage, formed by R1 and C1, is used
to reduce the signal artifacts.
Sensors 2021, 21, 2777 4 of 15
Sensors 2021, 21, x FOR PEER REVIEW 4 of 16
Figure 1. Schematic of the electronic interface. Each electrode is connected to the buffering stage of an
instrumentation amplifier (INA) configuration. The buffered and filtered output voltages will be
acquired by two ADC (Analog To Digital Converter) channels with respect to the ground. The com-
mon-mode voltage is connected to the driven right leg (DRL) electrode and is rejected.
Each electrode is an input of the buffering stage of a differential amplifier based on
the topology of a 3-input amplifier. The DRL is made of an inverting opamp and a buffer-
ing stage.
The electrodes are active and use amplifiers in order to achieve a very high input
impedance, as shown in Figure 2. A high-pass filtering stage, formed by R1 and C1, is used
to reduce the signal artifacts.
Figure 2. Schematic of the electrode interface. A double buffering stage is implemented in order to
achieve a very high input impedance.
The CMR and the frequency response of the electronic interface of the contactless
ECG were characterized by simulating the circuit in MULTISIM (National Instruments).
Figure 1. Schematic of the electronic interface. Each electrode is connected to the buffering stage of
an instrumentation amplifier (INA) configuration. The buffered and filtered output voltages will
be acquired by two ADC (Analog To Digital Converter) channels with respect to the ground. The
common-mode voltage is connected to the driven right leg (DRL) electrode and is rejected.
Figure 1. Schematic of the electronic interface. Each electrode is connected to the buffering stage of an
instrumentation amplifier (INA) configuration. The buffered and filtered output voltages will be
acquired by two ADC (Analog To Digital Converter) channels with respect to the ground. The com-
mon-mode voltage is connected to the driven right leg (DRL) electrode and is rejected.
Each electrode is an input of the buffering stage of a differential amplifier based on
the topology of a 3-input amplifier. The DRL is made of an inverting opamp and a buffer-
ing stage.
The electrodes are active and use amplifiers in order to achieve a very high input
impedance, as shown in Figure 2. A high-pass filtering stage, formed by R1 and C1, is used
to reduce the signal artifacts.
Figure 2. Schematic of the electrode interface. A double buffering stage is implemented in order to
achieve a very high input impedance.
The CMR and the frequency response of the electronic interface of the contactless
ECG were characterized by simulating the circuit in MULTISIM (National Instruments).
Figure 2. Schematic of the electrode interface. A double buffering stage is implemented in order to
achieve a very high input impedance.
The CMR and the frequency response of the electronic interface of the contactless
ECG were characterized by simulating the circuit in MULTISIM (National Instruments).
The circuit was implemented on a PCB and tested. According to the obtained results,
optimization of the circuit is proposed.
Moreover, we developed a portable device based on the presented electronic layout.
The output signals of the electronic interface are acquired using the 12-bit SAR ADCs of
the ESP32 microcontroller (ESPRESSIF SYSTEMS, Shanghai) with a sampling frequency of
250 Hz. The ESP32 microcontroller is equipped with an internal Bluetooth module; thus,
it is capable of transmitting digitized signals through a wireless connection to a PC or
smartphone. The device is battery operated.
A schematic of the developed device is reported in Figure 3.
Sensors 2021, 21, 2777 5 of 15
the ESP32 microcontroller (ESPRESSIF SYSTEMS, Shanghai) with a samplin
of 250 Hz. The ESP32 microcontroller is equipped with an internal Blueto
thus, it is capable of transmitting digitized signals through a wireless connec
or smartphone. The device is battery operated.
A schematic of the developed device is reported in Figure 3.
Figure 3. Block diagram of the developed contactless electrocardiogram (ECG) device.
Each electrode is connected to the electronic interface using a 4-pole aud
armature plate of the electrodes was implemented using directly the bottom
PCBs, as shown in Figure 4.
Figure 4. PCBs of the capacitive electrodes. The armature plate is placed in the bottom
board under an insulating layer. It is made of copper and has a radius of 3 cm.
To evaluate the effectiveness of the contactless ECG device, we performe
ison between signals obtained using this device and those obtained using a
fully integrated single-lead ECG front end by Analog Devices, using standa
trodes. The AD8232 is an integrated signal conditioning block for portable E
and other biopotential measurement applications. It implements a filter and a
amplification stage. To improve the common-mode rejection of the IC, it inc
ditional amplifier for the right leg drive [53].
We used the AD8232 in the “Cardiac Monitor Configuration,” as sugg
IC’s application note [53]. This configuration is designed to perform ECG m
with minimal distortion. The AD8232 is configured with a 0.5 Hz two-pole hig
followed by a 40 Hz two-pole low-pass filter, and the total system gain is ab
Figure 3. Block diagram of the developed contactless electrocardiogram (ECG) device.
Each electrode is connected to the electronic interface using a 4-pole audio cable. The
armature plate of the electrodes was implemented using directly the bottom layer of the
PCBs, as shown in Figure 4.
or smartphone. The device is battery operated.
A schematic of the developed device is reported in Figure 3.
Figure 3. Block diagram of the developed contactless electrocardiogram (ECG) device.
Each electrode is connected to the electronic interface using a 4-pole audio
armature plate of the electrodes was implemented using directly the bottom l
PCBs, as shown in Figure 4.
Figure 4. PCBs of the capacitive electrodes. The armature plate is placed in the bottom l
board under an insulating layer. It is made of copper and has a radius of 3 cm.
To evaluate the effectiveness of the contactless ECG device, we performed
ison between signals obtained using this device and those obtained using an
fully integrated single-lead ECG front end by Analog Devices, using standard
trodes. The AD8232 is an integrated signal conditioning block for portable EC
and other biopotential measurement applications. It implements a filter and a d
amplification stage. To improve the common-mode rejection of the IC, it inclu
ditional amplifier for the right leg drive [53].
We used the AD8232 in the “Cardiac Monitor Configuration,” as sugges
IC’s application note [53]. This configuration is designed to perform ECG mea
with minimal distortion. The AD8232 is configured with a 0.5 Hz two-pole high
followed by a 40 Hz two-pole low-pass filter, and the total system gain is abou
Figure 4. PCBs of the capacitive electrodes. The armature plate is placed in the bottom layer of the
board under an insulating layer. It is made of copper and has a radius of 3 cm.
To evaluate the effectiveness of the contactless ECG device, we performed a compari-
son between signals obtained using this device and those obtained using an AD8232, a fully
integrated single-lead ECG front end by Analog Devices, using standard wet electrodes.
The AD8232 is an integrated signal conditioning block for portable ECG devices and other
biopotential measurement applications. It implements a filter and a differential amplifi-
cation stage. To improve the common-mode rejection of the IC, it includes an additional
amplifier for the right leg drive [53].
We used the AD8232 in the “Cardiac Monitor Configuration,” as suggested by the
IC’s application note [53]. This configuration is designed to perform ECG measurements
with minimal distortion. The AD8232 is configured with a 0.5 Hz two-pole high-pass filter
followed by a 40 Hz two-pole low-pass filter, and the total system gain is about 60 dB. A
DRL is used to improve common-mode rejection; thus, the electronic approach is the same
as the developed contactless ECG device. The frequency response of the system is reported
in Figure 5.
Sensors 2021, 21, 2777 6 of 15
Sensors 2021, 21, x FOR PEER REVIEW 6 of 16
DRL is used to improve common-mode rejection; thus, the electronic approach is the same
as the developed contactless ECG device. The frequency response of the system is re-
ported in Figure 5.
Figure 5. Frequency response of the AD8232 implemented in the “Cardiac Monitor Configuration.”
The bode plot shows a band-pass behavior in the range of frequencies between 0.5 and 40 Hz and a
gain of about 60 dB (1000 V/V) (the bode plot is extracted from the AD8232 datasheet [53]).
The output signals from the AD8232 are acquired using the ADCs of the ESP32 mi-
crocontroller with a sampling frequency of 250 Hz. The ESP32 microcontroller is equipped
with an internal Bluetooth module; thus, it is capable of transmitting digitized signals
through a wireless connection to a PC or smartphone. The device is battery operated.
The scheme of the developed device is reported in Figure 6.
Figure 6. Block diagram of the developed ECG device based on the AD8232 IC. The device uses
standard wet adhesive electrodes.
In order to avoid interference between cables and electrodes of the two devices, ECG
measurements were performed sequentially, placing the electrodes in the same position
with the same experimental condition. The position of the electrodes implements Lead I
of Einthoven’s triangle.
Several measurements were performed on a single healthy subject at rest. This is be-
cause our primary goal in this study was to evaluate the feasibility and effectiveness of
Figure 5. Frequency response of the AD8232 implemented in the “Cardiac Monitor Configuration.”
The bode plot shows a band-pass behavior in the range of frequencies between 0.5 and 40 Hz and a
gain of about 60 dB (1000 V/V) (the bode plot is extracted from the AD8232 datasheet [53]).
The output signals from the AD8232 are acquired using the ADCs of the ESP32 micro-
controller with a sampling frequency of 250 Hz. The ESP32 microcontroller is equipped
with an internal Bluetooth module; thus, it is capable of transmitting digitized signals
through a wireless connection to a PC or smartphone. The device is battery operated.
The scheme of the developed device is reported in Figure 6.
Sensors 2021, 21, x FOR PEER REVIEW 6 of 16
DRL is used to improve common-mode rejection; thus, the electronic approach is the same
as the developed contactless ECG device. The frequency response of the system is re-
ported in Figure 5.
Figure 5. Frequency response of the AD8232 implemented in the “Cardiac Monitor Configuration.”
The bode plot shows a band-pass behavior in the range of frequencies between 0.5 and 40 Hz and a
gain of about 60 dB (1000 V/V) (the bode plot is extracted from the AD8232 datasheet [53]).
The output signals from the AD8232 are acquired using the ADCs of the ESP32 mi-
crocontroller with a sampling frequency of 250 Hz. The ESP32 microcontroller is equipped
with an internal Bluetooth module; thus, it is capable of transmitting digitized signals
through a wireless connection to a PC or smartphone. The device is battery operated.
The scheme of the developed device is reported in Figure 6.
Figure 6. Block diagram of the developed ECG device based on the AD8232 IC. The device uses
standard wet adhesive electrodes.
In order to avoid interference between cables and electrodes of the two devices, ECG
measurements were performed sequentially, placing the electrodes in the same position
with the same experimental condition. The position of the electrodes implements Lead I
of Einthoven’s triangle.
Several measurements were performed on a single healthy subject at rest. This is be-
cause our primary goal in this study was to evaluate the feasibility and effectiveness of
Figure 6. Block diagram of the developed ECG device based on the AD8232 IC. The device uses
standard wet adhesive electrodes.
In order to avoid interference between cables and electrodes of the two devices, ECG
measurements were performed sequentially, placing the electrodes in the same position
with the same experimental condition. The position of the electrodes implements Lead I of
Einthoven’s triangle.
Several measurements were performed on a single healthy subject at rest. This is
because our primary goal in this study was to evaluate the feasibility and effectiveness of
the studied electronic interface; even one subject was sufficient. Clearly, further studies on
a larger population will be necessary in order to clinically test the developed device.
The electrodes were placed as shown in Figure 7.
Sensors 2021, 21, 2777 7 of 15
Sensors 2021, 21, x FOR PEER REVIEW 7 of 16
the studied electronic interface; even one subject was sufficient. Clearly, further studies
on a larger population will be necessary in order to clinically test the developed device.
The electrodes were placed as shown in Figure 7.
Figure 7. Position of the 3 electrodes on the body of the patient.
3. Results
The electronic interface of the contactless ECG device was characterized. The fre-
quency response of the electrodes was investigated.
The electrical scheme of the electrode is reported in Figure 2. The simulation, re-
ported in Figure 8, shows a first-order high-pass characteristic with a cut-off frequency of
about 0.7 Hz and no gain.
Figure 8. Frequency response of the active capacitive electrodes.
It is useful to introduce a high-pass filtering stage to reduce the signal artifacts caused
by electrodes or body movements, such as thoracic expansion during breathing.
Figure 7. Position of the 3 electrodes on the body of the patient.
3. Results
The electronic interface of the contactless ECG device was characterized. The fre-
quency response of the electrodes was investigated.
The electrical scheme of the electrode is reported in Figure 2. The simulation, reported
in Figure 8, shows a first-order high-pass characteristic with a cut-off frequency of about
0.7 Hz and no gain.
Figure 7. Position of the 3 electrodes on the body of the patient.
3. Results
The electronic interface of the contactless ECG device was characterized. The fre-
quency response of the electrodes was investigated.
The electrical scheme of the electrode is reported in Figure 2. The simulation, re-
ported in Figure 8, shows a first-order high-pass characteristic with a cut-off frequency of
about 0.7 Hz and no gain.
Figure 8. Frequency response of the active capacitive electrodes.
It is useful to introduce a high-pass filtering stage to reduce the signal artifacts caused
by electrodes or body movements, such as thoracic expansion during breathing.
Figure 8. Frequency response of the active capacitive electrodes.
It is useful to introduce a high-pass filtering stage to reduce the signal artifacts caused
by electrodes or body movements, such as thoracic expansion during breathing.
The frequency response of the differential opamp stage was investigated. The electrical
scheme is reported in Figure 9.
Sensors 2021, 21, 2777 8 of 15
Sensors 2021, 21, x FOR PEER REVIEW 8 of 16
The frequency response of the differential opamp stage was investigated. The elec-
trical scheme is reported in Figure 9.
Figure 9. Schematic of the differential stage. In this stage, the input signal is amplified and filtered.
The simulation shows a second-order low-pass characteristic of the differential am-
plifier with a gain of 60 dB and a bandwidth of about 100 Hz (see Figure 10).
Figure 10. Frequency response of the differential stage.
The system frequency response of the electrodes coupled with the differential stage
was investigated. The complete circuit is shown in Figure 11.
Figure 9. Schematic of the differential stage. In this stage, the input signal is amplified and filtered.
The simulation shows a second-order low-pass characteristic of the differential ampli-
fier with a gain of 60 dB and a bandwidth of about 100 Hz (see Figure 10).
The frequency response of the differential opamp stage was investigated. The elec-
trical scheme is reported in Figure 9.
Figure 9. Schematic of the differential stage. In this stage, the input signal is amplified and filtered.
The simulation shows a second-order low-pass characteristic of the differential am-
plifier with a gain of 60 dB and a bandwidth of about 100 Hz (see Figure 10).
Figure 10. Frequency response of the differential stage.
The system frequency response of the electrodes coupled with the differential stage
was investigated. The complete circuit is shown in Figure 11.
Figure 10. Frequency response of the differential stage.
The system frequency response of the electrodes coupled with the differential stage
was investigated. The complete circuit is shown in Figure 11.
Sensors 2021, 21, x FOR PEER REVIEW 9 of 16
Figure 11. Schematic of the capacitive active electrodes coupled with the differential stage.
As expected, simulation results (see Figure 12) show a band-pass characteristic with
a gain of 40 dB.
Figure 11. Schematic of the capacitive active electrodes coupled with the differential stage.
Sensors 2021, 21, 2777 9 of 15
As expected, simulation results (see Figure 12) show a band-pass characteristic with a
gain of 40 dB.
Figure 11. Schematic of the capacitive active electrodes coupled with the differential stage.
As expected, simulation results (see Figure 12) show a band-pass characteristic with
a gain of 40 dB.
Figure 12. Frequency response of the capacitive active electrodes coupled with the differential
stage. The bode plot shows a band-pass behavior in the range of frequencies between 0.7 and 100
Hz and a gain of about 60 dB.
The voltage output of each channel with respect to the ground will exhibit a common-
mode voltage gain equal to 1.0. Obviously, by performing a differential measurement be-
tween the output voltage of Channel 1 and the output voltage of Channel 2, the Vcm will
theoretically disappear (in a real scenario, it is a small voltage). Using a DRL circuit, it is
possible to further increase the CMR of the system. Different gains of the DRL stage were
simulated to study its feedback effect on Vcm. The schematic is shown in Figure 13.
Figure 12. Frequency response of the capacitive active electrodes coupled with the differential stage. The bode plot shows a
band-pass behavior in the range of frequencies between 0.7 and 100 Hz and a gain of about 60 dB.
The voltage output of each channel with respect to the ground will exhibit a common-
mode voltage gain equal to 1.0. Obviously, by performing a differential measurement
between the output voltage of Channel 1 and the output voltage of Channel 2, the Vcm will
theoretically disappear (in a real scenario, it is a small voltage). Using a DRL circuit, it is
possible to further increase the CMR of the system. Different gains of the DRL stage were
simulated to study its feedback effect on Vcm. The schematic is shown in Figure 13.
Sensors 2021, 21, x FOR PEER REVIEW 10 of 16
Figure 13. Schematic of the complete system: capacitive electrodes, differential stage and DRL electrode. The electromag-
netic interference from the power line was modeled as a sine voltage source coupled to the electrodes through a 2 pF
capacitor. Moreover, the patient’s body was also coupled to the ground through a 200 pF capacitance [51].
The simulation results are shown in Figure 14.
Figure 13. Schematic of the complete system: capacitive electrodes, differential stage and DRL electrode. The electromagnetic
interference from the power line was modeled as a sine voltage source coupled to the electrodes through a 2 pF capacitor.
Moreover, the patient’s body was also coupled to the ground through a 200 pF capacitance [51].
Sensors 2021, 21, 2777 10 of 15
The simulation results are shown in Figure 14.
Figure 13. Schematic of the complete system: capacitive electrodes, differential stage and DRL electrode. The electromag-
netic interference from the power line was modeled as a sine voltage source coupled to the electrodes through a 2 pF
capacitor. Moreover, the patient’s body was also coupled to the ground through a 200 pF capacitance [51].
The simulation results are shown in Figure 14.
Figure 14. Plot showing the amplitude of the common-mode voltage Vcm versus the input re-
sistance R7 (see Figure 15) of the DRL opamp. As expected, by lowering the gain of the DRL, the
amplitude of the Vcm increased.
The implemented circuit shown in Figure 11 was tested to verify its frequency re-
sponse using a Lecroy Wave Station 3082 Waveform Generator as input and recording the
output signals using a Lecroy HDO6054-MS Oscilloscope. The experimental setup at the
block scheme level is shown in Figure 15.
Figure 14. Plot showing the amplitude of the common-mode voltage Vcm versus the input resistance
R7 (see Figure 15) of the DRL opamp. As expected, by lowering the gain of the DRL, the amplitude
of the Vcm increased.
Sensors 2021, 21, x FOR PEER REVIEW 11 of 16
Figure 15. Experimental setup at block scheme level used to characterize the frequency response of
the capacitive electrodes coupled with the differential stage.
The frequency response of the circuit is shown in Figure 16.
The output signals were acquired using the internal 12-bit SAR ADCs of the ESP32
microcontroller and transmitted using Bluetooth to a PC. The sampling frequency was 250
Hz.
About 20 s of recorded data obtained using the contactless ECG device are shown in
Figure 17.
In Figure 17, a sinusoidal pattern is clearly visible. It is caused by the movement of
the rib cage during breathing. Even if this pattern is a sort of signal artifact, it could be
used to determine the respiratory rate during an ECG measurement. In any case, this kind
of low-frequency artifact can be filtered both in the analog and in the digital domain by
applying a high-pass filter. In the original study, a compression vest was used to keep the
electrodes in fixed positions and to mitigate this kind of artifact, but we think that this
approach is not very comfortable for patients. In contrast, we used an elastic band in order
to apply the same compression that a cotton T-shirt would apply. One of our aims in this
study is to explore the feasibility of the embedding of the electrodes into clothes. In fact, a
smart way to use several electrodes for a long period of time, marginally compromising
patient comfort, could be, for example, by embedding the electrodes directly into a T-shirt.
Figure 15. Experimental setup at block scheme level used to characterize the frequency response of
the capacitive electrodes coupled with the differential stage.
The implemented circuit shown in Figure 11 was tested to verify its frequency response
using a Lecroy Wave Station 3082 Waveform Generator as input and recording the output
signals using a Lecroy HDO6054-MS Oscilloscope. The experimental setup at the block
scheme level is shown in Figure 15.
The frequency response of the circuit is shown in Figure 16.
Sensors 2021, 21, x FOR PEER REVIEW 11 of 16
Figure 15. Experimental setup at block scheme level used to characterize the frequency response of
the capacitive electrodes coupled with the differential stage.
The frequency response of the circuit is shown in Figure 16.
The output signals were acquired using the internal 12-bit SAR ADCs of the ESP32
microcontroller and transmitted using Bluetooth to a PC. The sampling frequency was 250
Hz.
About 20 s of recorded data obtained using the contactless ECG device are shown in
Figure 17.
In Figure 17, a sinusoidal pattern is clearly visible. It is caused by the movement of
the rib cage during breathing. Even if this pattern is a sort of signal artifact, it could be
used to determine the respiratory rate during an ECG measurement. In any case, this kind
of low-frequency artifact can be filtered both in the analog and in the digital domain by
applying a high-pass filter. In the original study, a compression vest was used to keep the
electrodes in fixed positions and to mitigate this kind of artifact, but we think that this
approach is not very comfortable for patients. In contrast, we used an elastic band in order
to apply the same compression that a cotton T-shirt would apply. One of our aims in this
study is to explore the feasibility of the embedding of the electrodes into clothes. In fact, a
smart way to use several electrodes for a long period of time, marginally compromising
patient comfort, could be, for example, by embedding the electrodes directly into a T-shirt.
Figure 16. Frequency response of the capacitive active electrodes coupled with the differential
stage. The band-pass behavior shown by the simulation was confirmed by the electrical tests. Dots
represent experimental data.
Figure 16. Frequency response of the capacitive active electrodes coupled with the differential stage.
The band-pass behavior shown by the simulation was confirmed by the electrical tests. Dots represent
experimental data.
Sensors 2021, 21, 2777 11 of 15
The output signals were acquired using the internal 12-bit SAR ADCs of the ESP32
microcontroller and transmitted using Bluetooth to a PC. The sampling frequency was
250 Hz.
About 20 s of recorded data obtained using the contactless ECG device are shown in
Figure 17.
Sensors 2021, 21, x FOR PEER REVIEW 12 of 16
Figure 17. Output signal obtained using the contactless ECG device.
About 20 s of recorded data obtained using the AD8232 device are shown in Figure
18.
In Figure 18, a slightly higher noise floor can be observed. No artifacts are shown.
In Figure 19, signals recorded from both the contactless device and the AD8232-based
device are reported. In order to better visualize the difference between the signals, several
PQRST complexes, extracted from the raw output signals, are plotted.
ECG complexes, extracted from the two devices, are quite comparable but not iden-
tical. The timings of intervals and segments of the complexes are the same, but the relative
amplitudes of the R peak and the T wave are different. Moreover, S peaks are not visible
in the AD8232 complexes.
Figure 18. Output signal obtained using the AD8232 device.
0 2 4 6 8 10 12 14 16 18
Time [s]
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1
Amplitude
[V]
0 2.5 5 7.5 10 12.5 15 17.5
Time [s]
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1
Amplitude
[V]
Figure 17. Output signal obtained using the contactless ECG device.
In Figure 17, a sinusoidal pattern is clearly visible. It is caused by the movement of
the rib cage during breathing. Even if this pattern is a sort of signal artifact, it could be
used to determine the respiratory rate during an ECG measurement. In any case, this kind
of low-frequency artifact can be filtered both in the analog and in the digital domain by
applying a high-pass filter. In the original study, a compression vest was used to keep the
electrodes in fixed positions and to mitigate this kind of artifact, but we think that this
approach is not very comfortable for patients. In contrast, we used an elastic band in order
to apply the same compression that a cotton T-shirt would apply. One of our aims in this
study is to explore the feasibility of the embedding of the electrodes into clothes. In fact, a
smart way to use several electrodes for a long period of time, marginally compromising
patient comfort, could be, for example, by embedding the electrodes directly into a T-shirt.
About 20 s of recorded data obtained using the AD8232 device are shown in Figure 18.
Sensors 2021, 21, x FOR PEER REVIEW 12 of 16
Figure 17. Output signal obtained using the contactless ECG device.
About 20 s of recorded data obtained using the AD8232 device are shown in Figure
18.
In Figure 18, a slightly higher noise floor can be observed. No artifacts are shown.
In Figure 19, signals recorded from both the contactless device and the AD8232-based
device are reported. In order to better visualize the difference between the signals, several
PQRST complexes, extracted from the raw output signals, are plotted.
ECG complexes, extracted from the two devices, are quite comparable but not iden-
tical. The timings of intervals and segments of the complexes are the same, but the relative
amplitudes of the R peak and the T wave are different. Moreover, S peaks are not visible
in the AD8232 complexes.
Figure 18. Output signal obtained using the AD8232 device.
0 2 4 6 8 10 12 14 16 18
Time [s]
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1
Amplitude
[V]
0 2.5 5 7.5 10 12.5 15 17.5
Time [s]
-1
-0.8
-0.6
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1
Amplitude
[V]
Figure 18. Output signal obtained using the AD8232 device.
In Figure 18, a slightly higher noise floor can be observed. No artifacts are shown.
In Figure 19, signals recorded from both the contactless device and the AD8232-based
device are reported. In order to better visualize the difference between the signals, several
PQRST complexes, extracted from the raw output signals, are plotted.
Sensors 2021, 21, 2777 12 of 15
Sensors 2021, 21, x FOR PEER REVIEW 13 of 16
Figure 19. Comparison between the contactless device and AD8232 device signals. About 60 PQRST
complexes were extracted from the raw output signals and overlapped in a single plot for each de-
vice.
4. Discussion and Conclusions
We simulated, implemented and tested an electronic interface for ECG devices. The
results obtained from simulations show the ability of the electronic interface to amplify
differential biopotentials and to reject common-mode signals produced by EMI in an ef-
fective way. The results obtained from the electrical testing on the implemented PCB con-
firm these capabilities. Further adjustments can be applied in order to mostly optimize the
response of the system at low frequencies.
The studied interface can be used with a capacitive electrode that can be placed on
the clothes of the patient. This contactless ECG device is more comfortable for long-term
monitoring instead of standard ECG devices using adhesive wet electrodes. The circuit
topology allows the implementation of the electronic interface with very low costs; in fact,
unlike many other electronic interfaces for ECG devices, it does not use expensive Instru-
mentation amplifiers. Moreover, the topology is highly modular, allowing the use of a
flexible number of electrodes.
To evaluate the effectiveness of the contactless ECG device, we performed ECG
measurements on a healthy subject. The biopotentials were acquired using both the stud-
ied electronic interface and a commercial fully integrated single-lead ECG front end by an
analog device. Both electronic interfaces use an approach based on differential amplifica-
tion, filtering and DRL electrodes in order to obtain a high signal-to-noise ratio. The first
electronic interface makes use of capacitive electrodes, and the second interface uses
standard wet Ag/AgCl electrodes.
The obtained measurements are quite comparable in terms of the amplitude and
shape of the recorded signal. The ECG traces, recorded using a contactless device, present
a satisfying signal-to-noise ratio. Moreover, the acquired signals present all the features
needed for a clinical evaluation: P waves, QRS complex and T wave. Contactless ECG
traces showed a sinusoidal pattern caused by the movement of the rib cage during breath-
ing. This artifact could be used to determine the respiratory rate of the patient, obtaining
one more useful piece of information during the ECG measurement. The characterization
of the electronic interface behavior in the frequency domain highlights how optimization
of the filtering stages could decrease this kind of artifact.
Moreover, capacitive electrodes are sensitive to nearby charge displacements. As was
expected, during the tests, we experienced signal artifacts of the output signal caused by
body movements, in particular arm movements and, in a minor way, movements of the
head and shoulders. We think that before testing the system on a person in motion (e.g.,
0 50 100 150 200 250
Sample Number
0
0.2
0.4
0.6
0.8
1
1.2
Amplitude
[V]
Contactless
0 50 100 150 200 250
Sample Number
0
0.2
0.4
0.6
0.8
1
1.2
Amplitude
[V]
AD8232
Figure 19. Comparison between the contactless device and AD8232 device signals. About 60 PQRST
complexes were extracted from the raw output signals and overlapped in a single plot for each device.
ECG complexes, extracted from the two devices, are quite comparable but not identical.
The timings of intervals and segments of the complexes are the same, but the relative
amplitudes of the R peak and the T wave are different. Moreover, S peaks are not visible in
the AD8232 complexes.
4. Discussion and Conclusions
We simulated, implemented and tested an electronic interface for ECG devices. The
results obtained from simulations show the ability of the electronic interface to amplify
differential biopotentials and to reject common-mode signals produced by EMI in an
effective way. The results obtained from the electrical testing on the implemented PCB
confirm these capabilities. Further adjustments can be applied in order to mostly optimize
the response of the system at low frequencies.
The studied interface can be used with a capacitive electrode that can be placed on
the clothes of the patient. This contactless ECG device is more comfortable for long-term
monitoring instead of standard ECG devices using adhesive wet electrodes. The circuit
topology allows the implementation of the electronic interface with very low costs; in
fact, unlike many other electronic interfaces for ECG devices, it does not use expensive
Instrumentation amplifiers. Moreover, the topology is highly modular, allowing the use of
a flexible number of electrodes.
To evaluate the effectiveness of the contactless ECG device, we performed ECG
measurements on a healthy subject. The biopotentials were acquired using both the
studied electronic interface and a commercial fully integrated single-lead ECG front end
by an analog device. Both electronic interfaces use an approach based on differential
amplification, filtering and DRL electrodes in order to obtain a high signal-to-noise ratio.
The first electronic interface makes use of capacitive electrodes, and the second interface
uses standard wet Ag/AgCl electrodes.
The obtained measurements are quite comparable in terms of the amplitude and
shape of the recorded signal. The ECG traces, recorded using a contactless device, present
a satisfying signal-to-noise ratio. Moreover, the acquired signals present all the features
needed for a clinical evaluation: P waves, QRS complex and T wave. Contactless ECG traces
showed a sinusoidal pattern caused by the movement of the rib cage during breathing.
This artifact could be used to determine the respiratory rate of the patient, obtaining one
more useful piece of information during the ECG measurement. The characterization of
the electronic interface behavior in the frequency domain highlights how optimization of
the filtering stages could decrease this kind of artifact.
Moreover, capacitive electrodes are sensitive to nearby charge displacements. As was
expected, during the tests, we experienced signal artifacts of the output signal caused by
Sensors 2021, 21, 2777 13 of 15
body movements, in particular arm movements and, in a minor way, movements of the
head and shoulders. We think that before testing the system on a person in motion (e.g.,
walking, running), further investigations are needed to understand how to reduce signal
interferences generated by movements of the upper body.
It is important to highlight that although we tested the contactless ECG device imple-
menting only two electrodes and a DRL grounding electrode in a single-lead configuration,
it can also be used in a multiple-lead configuration with more than two electrodes.
In further studies, we would like to implement and test the contactless ECG device
with at least three leads in order to record standard and augmented limb leads. With more
electrodes, it would be possible to record a complete 12-lead ECG.
Thus, the use of a greater number of electrodes allows the recording of information
content greater than those obtainable with wearable devices (e.g., smartwatches) currently
available on the market. The implementation of more leads and the recording of more
informative signals will allow going beyond mere arrhythmia diagnosis. In any case,
further studies will be needed to evaluate the possibility of using ECG recorded using a
multilead contactless ECG device to provide a diagnosis of other cardiac disorders.
Obviously, it will be less comfortable to measure an ECG using several electrodes
placed on the body instead of a smartwatch, but huge improvements can be performed
with regard to the implementation of the electrodes. As the capacitive electrodes do not
require direct contact with the patient’s skin, they could be embedded inside the clothes;
for example, in the fabric of a T-shirt using flexible electronics. Therefore, in further studies,
we would also like to optimize the implementation of the electrodes in order to make the
contactless ECG device more comfortable and wearable.
Author Contributions: Conceptualization, A.Z. and M.S.; methodology, A.Z. and M.S.; software, A.S.
and S.G.; validation, A.Z., G.P., M.S. and M.C.; formal analysis, A.Z., S.G., A.S. and G.B.; investigation,
G.P., M.C. and M.L.; resources, M.S.; data curation, A.Z., A.S. and S.G.; writing—original draft
preparation, A.Z. and M.S.; writing—review and editing, G.P., M.C. and G.F.; supervision, G.P., M.S.,
G.F. and M.C. All authors have read and agreed to the published version of the manuscript.
Funding: This research was funded by the Italian Ministry of Research and University in the frame
of the European FLAG-ERA funding scheme. Name of the project: CONVERGENCE: “Frictionless
Energy Efficient Convergent Wearables for Healthcare and Lifestyle Applications.”
Institutional Review Board Statement: UCBM Ethical Committee approved the trial on 5 March,
2018: Protocol Number 28/18OSS ComEtCBM.
Informed Consent Statement: Informed consent was obtained from all subjects involved in the study.
Data Availability Statement: Data are available upon reasonable request.
Conflicts of Interest: The authors declare no conflict of interest.
References
1. Kristoffersson, A.; Lindén, M. A systematic review on the use of wearable body sensors for health monitoring: A qualitative
synthesis. Sensors 2020, 20, 1502. [CrossRef] [PubMed]
2. Ding, E.Y.; Ensom, E.; Hafer, N.; Buchholz, B.; Picard, M.A.; Dunlap, D.; McManus, D.D. Point-of-care technologies in heart, lung,
blood and sleep disorders from the Center for Advancing Point-of-Care Technologies. Curr. Opin. Biomed. Eng. 2019, 11, 58–67.
[CrossRef] [PubMed]
3. Pantelopoulos, A.; Bourbakis, N.G. A survey on wearable sensor-based systems for health monitoring and prognosis. IEEE Trans.
Syst. Man. Cybern. Part. C (Appl. Rev.) 2019, 40, 1–12. [CrossRef]
4. Mizuno, A.; Changolkar, S.; Patel, M.S. Wearable Devices to Monitor and Reduce the Risk of Cardiovascular Disease: Evidence
and Opportunities. Annu. Rev. Med. 2020, 72, 459–471. [CrossRef] [PubMed]
5. Wilkins, E.; Wilson, L.; Wickramasinghe, K.; Bhatnagar, P.; Leal, J.; Luengo-Fernandez, R.; Townsend, N. European Cardiovascular
Disease Statistics; European Heart Network: Brussels, Belgium, 2017. Available online: http://www.ehnheart.org/images/CVD-
statistics-report-August-2017.pdf (accessed on 1 March 2021).
6. Patel, S.; Park, H.; Bonato, P.; Chan, L.; Rodgers, M. A review of wearable sensors and systems with application in rehabilitation.
J. Neuroeng. Rehabil. 2012, 9, 1–17. [CrossRef]
Sensors 2021, 21, 2777 14 of 15
7. Iqbal, M.H.; Aydin, A.; Brunckhorst, O.; Dasgupta, P.; Ahmed, K. A review of wearable technology in medicine. J. R. Soc. Med.
2016, 109, 372–380. [CrossRef] [PubMed]
8. McCann, J.; Bryson, D. Smart Clothes and Wearable Technology; Woodhead Publishing: Cambridge, UK, 2009; ISBN 9781845693572.
9. Tong, R. Wearable Technology in Medicine and Health Care; Academic Press: New York, NY, USA, 2018; ISBN 9780128118108.
10. Rashkovska, A.; Depolli, M.; Tomašić, I.; Avbelj, V.; Trobec, R. Medical-grade ECG sensor for long-term monitoring. Sensors 2020,
20, 1695. [CrossRef]
11. Samol, A.; Bischof, K.; Luani, B.; Pascut, D.; Wiemer, M.; Kaese, S. Single-lead ECG recordings including Einthoven and Wilson
leads by a smartwatch: A new era of patient directed early ECG differential diagnosis of cardiac diseases? Sensors 2019, 19, 4377.
[CrossRef]
12. Miller, H.A.; Harrison, D.C. Biomedical Electrode Technology: Theory and Practice; Academic Press: New York, NY, USA, 1974; ISBN
9780124968509.
13. Searle, A.; Kirkup, L. A direct comparison of wet, dry and insulating bioelectric recording electrodes. Physiol. Meas. 2000, 21.
[CrossRef]
14. Chi, Y.M.; Jung, T.P.; Cauwenberghs, G. Dry-Contact and Noncontact Biopotential Electrodes: Methodological Review. IEEE Rev.
Biomed. Eng. 2010, 3, 106–119. [CrossRef]
15. Burke, M.J.; Gleeson, D.T. A micropower dry-electrode ECG preamplifier. IEEE Trans. Biomed. Eng. 2000, 47, 155–162. [CrossRef]
[PubMed]
16. Lin, C.T.; Liao, L.D.; Liu, Y.H.; Wang, I.J.; Lin, B.S.; Chang, J.Y. Novel dry polymer foam electrodes for long-term EEG measurement.
IEEE Trans. Biomed. Eng. 2011, 58, 1200–1207. [CrossRef] [PubMed]
17. Teichmann, D.; Foussier, J.; Jia, J.; Leonhardt, S.; Walter, M. Noncontact monitoring of cardiorespiratory activity by electromagnetic
coupling. IEEE Trans. Biomed. Eng. 2013, 60, 2142–2152. [CrossRef] [PubMed]
18. Sullivan, T.J.; Deiss, S.R.; Cauwenberghs, G. A low-noise, non-contact EEG/ECG sensor. In Proceedings of the Biomedical Circuits
and Systems Conference, Montreal, QC, Canada, 27–30 November 2007; pp. 154–157.
19. Ueno, A.; Akabane, Y.; Kato, T.; Hoshino, H.; Kataoka, S.; Ishiyama, Y. Capacitive sensing of electrocardiographic potential
through cloth from the dorsal surface of the body in a supine position: A preliminary study. IEEE Trans. Biomed. Eng. 2007, 54,
759–766. [CrossRef] [PubMed]
20. Lim, Y.G.; Kim, K.K.; Park, K.S. ECG recording on a bed during sleep without direct skin-contact. IEEE Trans. Biomed. Eng. 2007,
54, 718–725. [CrossRef] [PubMed]
21. Wartzek, T.; Eilebrecht, B.; Lem, J.; Lindner, H.J.; Leonhardt, S.; Walter, M. ECG on the road: Robust and unobtrusive estimation
of heart rate contact. IEEE Trans. Biomed. Eng. 2011, 58, 3112–3120. [CrossRef]
22. Gao, S.; Huang, D.Y.; Fang, W.C. A non-contact ECG sensing system with motion artifact equalization. In Proceedings of the IEEE
Biomedical Circuits and Systems Conference (BioCAS 2016), Shanghai, China, 17–19 October 2016; pp. 244–247.
23. Parente, F.R.; Santonico, M.; Zompanti, A.; Benassai, M.; Ferri, G.; D’Amico, A.; Pennazza, G. An electronic system for the
contactless reading of ECG signals. Sensors 2017, 17, 2474. [CrossRef]
24. Lessard-Tremblay, M.; Weeks, J.; Morelli, L.; Cowan, G.; Gagnon, G.; Zednik, R.J. Contactless capacitive electrocardiography
using hybrid flexible printed electrodes. Sensors 2020, 20, 5156. [CrossRef] [PubMed]
25. Chi, Y.M.; Ng, P.; Kang, E.; Kang, J.; Fang, J.; Cauwenberghs, G. Wireless non-contact cardiac and neural monitoring. In
Proceedings of the Wireless Health, San Diego, CA, USA, 5−7 October 2010; pp. 15–23.
26. Van Rijn, A.M.; Peper, A.; Grimbergen, C.A. High-quality recording of bioelectric events. Med. Biol. Eng. Comput. 1990, 28,
389–397. [CrossRef]
27. Lim, Y.G.; Chung, G.S.; Park, K.S. Capacitive driven-right-leg grounding in indirect-contact ECG measurement. In Proceedings
of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC 2010), Buenos Aires,
Argentina, 31 August–4 September 2010; pp. 1250–1253. [CrossRef]
28. Singh, R.K.; Sarkar, A.; Anoop, C.S. A health monitoring system using multiple non-contact ECG sensors for automotive drivers.
In Proceedings of the IEEE International Instrumentation and Measurement Technology Conference, Taipei, Taiwan, 23–26 May
2016; pp. 1–6. [CrossRef]
29. Gromer, M.; Salb, D.; Walzer, T.; Madrid, N.M.; Seepold, R. ECG sensor for detection of driver’s drowsiness. Procedia Comput. Sci.
2019, 159, 1938–1946. [CrossRef]
30. Leonhardt, S.; Aleksandrowicz, A. Non-contact ECG monitoring for automotive application. In Proceedings of the IEEE 5th
International Summer School and Symposium on Medical Devices and Biosensors, Hong Kong, China, 1–3 June 2008; pp. 183–185.
[CrossRef]
31. Leicht, L.; Skobel, E.; Knackstedt, C.; Mathissen, M.; Sitter, A.; Wartzek, T.; Teichmann, D. Capacitive ECG monitoring in cardiac
patients during simulated driving. IEEE Trans. Biomed. Eng. 2018, 66, 749–758. [CrossRef]
32. Choi, Y.J.; Lee, J.Y.; Kong, S.H. Driver ECG Measuring System With a Conductive Fabric-Based Dry Electrode. IEEE Access 2018, 6,
415–427. [CrossRef]
33. Walter, M.; Eilebrecht, B.; Wartzek, T.; Leonhardt, S. The smart car seat: Personalized monitoring of vital signs in automotive
applications. Pers. Ubiquitous Comput. 2011, 15, 707–715. [CrossRef]
Sensors 2021, 21, 2777 15 of 15
34. Wang, Y.H.; Chung, C.G.; Lin, C.C.; Lin, C.M. The study of the electrocardiography monitoring for the elderly based on smart
clothes. In Proceedings of the IEEE 2018 Eighth International Conference on Information Science and Technology (ICIST), Istanbul,
Turkey, 16–18 March 2018; pp. 478–482. [CrossRef]
35. Lee, H.J.; Lee, S.H.; Ha, K.S.; Jang, H.C.; Chung, W.Y.; Kim, J.Y.; Chang, Y.S.; Yoo, D.H. Ubiquitous healthcare service using Zigbee
and mobile phone for elderly patients. Int. J. Med. Inform. 2009, 78, 193–198. [CrossRef]
36. Fan, X.; Zhao, Y.; Wang, H.; Tsui, K.L. Forecasting one-day-forward wellness conditions for community-dwelling elderly with
single lead short electrocardiogram signals. BMC Med. Inform. Decis. Mak. 2019, 19, 285. [CrossRef] [PubMed]
37. Chung, W.Y.; Bhardwaj, S.; Punvar, A.; Lee, D.S.; Myllylae, R. A fusion health monitoring using ECG and accelerometer sensors
for elderly persons at home. In Proceedings of the 2007 29th annual international conference of the IEEE Engineering in Medicine
and Biology Society, Lyon, France, 22–26 August 2007; pp. 3818–3821. [CrossRef]
38. Bsoul, M.; Minn, H.; Tamil, L. Apnea MedAssist: Real-time sleep apnea monitor using single-lead ECG. IEEE Trans. Inf. Technol.
Biomed. 2010, 15, 416–427. [CrossRef]
39. Kesper, K.; Canisius, S.; Penzel, T.; Ploch, T.; Cassel, W. ECG signal analysis for the assessment of sleep-disordered breathing and
sleep pattern. Med. Biol. Eng. Comput. 2012, 50, 135–144. [CrossRef] [PubMed]
40. Varon, C.; Caicedo, A.; Testelmans, D.; Buyse, B.; Van Huffel, S. A novel algorithm for the automatic detection of sleep apnea
from single-lead ECG. IEEE Trans. Biomed. Eng. 2015, 62, 2269–2278. [CrossRef]
41. Almazaydeh, L.; Elleithy, K.; Faezipour, M. Detection of obstructive sleep apnea through ECG signal features. In Proceedings
of the 2012 IEEE International Conference on Electro/Information Technology, Indianapolis, IN, USA, 6–8 May 2012; pp. 1–6.
[CrossRef]
42. Yilmaz, B.; Asyali, M.H.; Arikan, E.; Yetkin, S.; Özgen, F. Sleep stage and obstructive apneaic epoch classification using single-lead
ECG. Biomed. Eng. Online 2010, 9, 1–14. [CrossRef]
43. Biel, L.; Pettersson, O.; Philipson, L.; Wide, P. ECG analysis: A new approach in human identification. IEEE Trans. Instrum. Meas.
2001, 50, 808–812. [CrossRef]
44. Singh, Y.N.; Gupta, P. ECG to individual identification. In Proceeding of the 2008 IEEE Second International Conference on
Biometrics: Theory, Applications and Systems, Washington, DC, USA, 29 September–1 October 2008; pp. 1–8. [CrossRef]
45. Saechia, S.; Koseeyaporn, J.; Wardkein, P. Human identification system based ECG signal. In Proceedings of the TENCON
2005–2005 IEEE Region 10 Conference, Melbourne, VIC, Australia, 21–24 November 2005; pp. 1–4. [CrossRef]
46. Ye, C.; Coimbra, M.T.; Kumar, B.V. Investigation of human identification using two-lead electrocardiogram (ECG) signals.
In Proceedings of the 2010 Fourth IEEE International Conference on Biometrics: Theory, Applications and Systems (BTAS),
Washington, DC, USA, 27–29 September 2010; pp. 1–8. [CrossRef]
47. Covello, R.; Fortino, G.; Gravina, R.; Aguilar, A.; Breslin, J.G. Novel method and real-time system for detecting the Cardiac
Defense Response based on the ECG. In Proceedings of the 2013 IEEE International Symposium on Medical Measurements and
Applications (MeMeA), Gatineau, QC, Canada, 4–5 May 2013; pp. 53–57. [CrossRef]
48. Gravina, R.; Fortino, G. Automatic methods for the detection of accelerative cardiac defense response. IEEE Trans. Affect. Comput.
2016, 7, 286–298. [CrossRef]
49. Fortino, G.; Gravina, R. Real-time automatic detection of accelerative cardiac defense response. In Proceedings of the 9th
International Conference on Body Area Networks, London, UK, 29 September–1 October 2014; pp. 159–164. [CrossRef]
50. Ahamed, M.A.; Ahad, M.A.U.; Sohag, M.H.A.; Ahmad, M. Development of low cost wireless biosignal acquisition system for
ECG EMG and EOG. In Proceedings of the IEEE 2015 2nd International Conference on Electrical Information and Communication
Technologies (EICT), Khulna, Bangladesh, 10–12 December 2015; pp. 195–199. [CrossRef]
51. Winter, B.B.; Webster, J.G. Driven-right-leg circuit design. IEEE Trans. Biomed. Eng. 1983, 1, 62–66. [CrossRef] [PubMed]
52. Clark, J.W., Jr.; Neuman, M.R.; Olson, W.H.; Peura, R.A.; Primiano, F.P., Jr.; Siedband, M.P.; Webster, J.G.; Wheeler, L.A.
Contributing Authors. In Medical Instrumentation: Application and Design, 4th ed.; Webster, J.G., Ed.; John Wiley  Sons: New York,
NY, USA, 2010; ISBN 13-978-0471-57600-3.
53. Analog Devices, Inc. D8232 IC Datasheet. Available online: https://www.analog.com/media/en/technical-documentation/
data-sheets/ad8232.pdf (accessed on 1 March 2021).

More Related Content

What's hot

International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) ijceronline
 
A Wireless ECG Plaster for Real-Time Cardiac Health Monitoring in Body Senso...
A Wireless ECG Plaster for Real-Time Cardiac  Health Monitoring in Body Senso...A Wireless ECG Plaster for Real-Time Cardiac  Health Monitoring in Body Senso...
A Wireless ECG Plaster for Real-Time Cardiac Health Monitoring in Body Senso...ecgpapers
 
Low cost solar ecg with bluetooth transmitter
Low cost solar ecg with bluetooth transmitterLow cost solar ecg with bluetooth transmitter
Low cost solar ecg with bluetooth transmittersudhakar5472
 
Real time heart monitoring system
Real time heart monitoring systemReal time heart monitoring system
Real time heart monitoring systemShashank Kapoor
 
Final Documentation - Full
Final Documentation - FullFinal Documentation - Full
Final Documentation - FullEmeel Gayed
 
Acquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsAcquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsIJERA Editor
 
Biotelemetry
BiotelemetryBiotelemetry
BiotelemetrySamuely
 
A Real Time Electrocardiogram (ECG) Device for Cardiac Patients
A Real Time Electrocardiogram (ECG) Device for Cardiac PatientsA Real Time Electrocardiogram (ECG) Device for Cardiac Patients
A Real Time Electrocardiogram (ECG) Device for Cardiac PatientsIJERD Editor
 
wireless ECG system based on android
wireless ECG system based on androidwireless ECG system based on android
wireless ECG system based on androidGhadeerAlsandi
 
Bachelorthesis.compressed
Bachelorthesis.compressedBachelorthesis.compressed
Bachelorthesis.compressedDhara Shah
 
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...ijtsrd
 
The development of a wireless LCP-based intracranial pressure sensor for trau...
The development of a wireless LCP-based intracranial pressure sensor for trau...The development of a wireless LCP-based intracranial pressure sensor for trau...
The development of a wireless LCP-based intracranial pressure sensor for trau...IJECEIAES
 
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...TELKOMNIKA JOURNAL
 

What's hot (19)

International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER)
 
Medical applications-ecg
Medical applications-ecgMedical applications-ecg
Medical applications-ecg
 
Ecg
EcgEcg
Ecg
 
A Wireless ECG Plaster for Real-Time Cardiac Health Monitoring in Body Senso...
A Wireless ECG Plaster for Real-Time Cardiac  Health Monitoring in Body Senso...A Wireless ECG Plaster for Real-Time Cardiac  Health Monitoring in Body Senso...
A Wireless ECG Plaster for Real-Time Cardiac Health Monitoring in Body Senso...
 
Low cost solar ecg with bluetooth transmitter
Low cost solar ecg with bluetooth transmitterLow cost solar ecg with bluetooth transmitter
Low cost solar ecg with bluetooth transmitter
 
Real time heart monitoring system
Real time heart monitoring systemReal time heart monitoring system
Real time heart monitoring system
 
Final Documentation - Full
Final Documentation - FullFinal Documentation - Full
Final Documentation - Full
 
Acquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart AilmentsAcquiring Ecg Signals And Analysing For Different Heart Ailments
Acquiring Ecg Signals And Analysing For Different Heart Ailments
 
ECG Report_2014
ECG Report_2014ECG Report_2014
ECG Report_2014
 
Biotelemetry
BiotelemetryBiotelemetry
Biotelemetry
 
Eg24842846
Eg24842846Eg24842846
Eg24842846
 
A Real Time Electrocardiogram (ECG) Device for Cardiac Patients
A Real Time Electrocardiogram (ECG) Device for Cardiac PatientsA Real Time Electrocardiogram (ECG) Device for Cardiac Patients
A Real Time Electrocardiogram (ECG) Device for Cardiac Patients
 
Seminar
SeminarSeminar
Seminar
 
wireless ECG system based on android
wireless ECG system based on androidwireless ECG system based on android
wireless ECG system based on android
 
Bachelorthesis.compressed
Bachelorthesis.compressedBachelorthesis.compressed
Bachelorthesis.compressed
 
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
Real Time Signal Quality Aware Internet of Things IOT Framework for FPGA Base...
 
Health Monitoring
Health MonitoringHealth Monitoring
Health Monitoring
 
The development of a wireless LCP-based intracranial pressure sensor for trau...
The development of a wireless LCP-based intracranial pressure sensor for trau...The development of a wireless LCP-based intracranial pressure sensor for trau...
The development of a wireless LCP-based intracranial pressure sensor for trau...
 
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
A low-cost electro-cardiograph machine equipped with sensitivity and paper sp...
 

Similar to Sensors 21-02777

Electrocardiogram signal processing algorithm on microcontroller using wavele...
Electrocardiogram signal processing algorithm on microcontroller using wavele...Electrocardiogram signal processing algorithm on microcontroller using wavele...
Electrocardiogram signal processing algorithm on microcontroller using wavele...IJECEIAES
 
A Low Power Wearable Physiological Parameter Monitoring System
A Low Power Wearable Physiological Parameter Monitoring SystemA Low Power Wearable Physiological Parameter Monitoring System
A Low Power Wearable Physiological Parameter Monitoring Systemijsrd.com
 
Portable Real Time Cardiac Activity Monitoring System
Portable  Real Time Cardiac Activity Monitoring SystemPortable  Real Time Cardiac Activity Monitoring System
Portable Real Time Cardiac Activity Monitoring SystemIRJET Journal
 
Low-cost central monitor based personal computer with electrocardiogram and h...
Low-cost central monitor based personal computer with electrocardiogram and h...Low-cost central monitor based personal computer with electrocardiogram and h...
Low-cost central monitor based personal computer with electrocardiogram and h...TELKOMNIKA JOURNAL
 
Wireless Telemetry System
Wireless Telemetry SystemWireless Telemetry System
Wireless Telemetry Systemijsrd.com
 
Body Adapted Wearable Bio Sensor
Body Adapted Wearable Bio SensorBody Adapted Wearable Bio Sensor
Body Adapted Wearable Bio Sensorkavindrakrishna
 
IRJET- Design and Implementation of Health Monitoring System
IRJET-  	  Design and Implementation of Health Monitoring SystemIRJET-  	  Design and Implementation of Health Monitoring System
IRJET- Design and Implementation of Health Monitoring SystemIRJET Journal
 
WBSN based safe lifestyle: a case study of heartrate monitoring system
WBSN based safe lifestyle: a case study of heartrate monitoring system WBSN based safe lifestyle: a case study of heartrate monitoring system
WBSN based safe lifestyle: a case study of heartrate monitoring system IJECEIAES
 
Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...eSAT Journals
 
Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...eSAT Publishing House
 
Design and implementation of portable electrocardiogram recorder with field ...
Design and implementation of portable electrocardiogram  recorder with field ...Design and implementation of portable electrocardiogram  recorder with field ...
Design and implementation of portable electrocardiogram recorder with field ...IJECEIAES
 
Wireless Body Area Networks for Healthcare: A Survey
Wireless Body Area Networks for Healthcare: A SurveyWireless Body Area Networks for Healthcare: A Survey
Wireless Body Area Networks for Healthcare: A Surveyijasuc
 
Real time ECG Monitoring: A Review
Real time ECG Monitoring: A ReviewReal time ECG Monitoring: A Review
Real time ECG Monitoring: A Reviewijtsrd
 
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...IRJET Journal
 
Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...journalBEEI
 
A general framework for improving electrocardiography monitoring system with ...
A general framework for improving electrocardiography monitoring system with ...A general framework for improving electrocardiography monitoring system with ...
A general framework for improving electrocardiography monitoring system with ...journalBEEI
 
Technical seminar.pptx is based on the Wearable BIO Sensors
Technical seminar.pptx is based on the Wearable BIO SensorsTechnical seminar.pptx is based on the Wearable BIO Sensors
Technical seminar.pptx is based on the Wearable BIO Sensors4HG20EC020MouneshGow
 

Similar to Sensors 21-02777 (20)

Electrocardiogram signal processing algorithm on microcontroller using wavele...
Electrocardiogram signal processing algorithm on microcontroller using wavele...Electrocardiogram signal processing algorithm on microcontroller using wavele...
Electrocardiogram signal processing algorithm on microcontroller using wavele...
 
ECG Report_2014
ECG Report_2014ECG Report_2014
ECG Report_2014
 
A Low Power Wearable Physiological Parameter Monitoring System
A Low Power Wearable Physiological Parameter Monitoring SystemA Low Power Wearable Physiological Parameter Monitoring System
A Low Power Wearable Physiological Parameter Monitoring System
 
Portable Real Time Cardiac Activity Monitoring System
Portable  Real Time Cardiac Activity Monitoring SystemPortable  Real Time Cardiac Activity Monitoring System
Portable Real Time Cardiac Activity Monitoring System
 
Electrocardiograph: A Portable Bedside Monitor
Electrocardiograph: A Portable Bedside MonitorElectrocardiograph: A Portable Bedside Monitor
Electrocardiograph: A Portable Bedside Monitor
 
Low-cost central monitor based personal computer with electrocardiogram and h...
Low-cost central monitor based personal computer with electrocardiogram and h...Low-cost central monitor based personal computer with electrocardiogram and h...
Low-cost central monitor based personal computer with electrocardiogram and h...
 
Wireless Telemetry System
Wireless Telemetry SystemWireless Telemetry System
Wireless Telemetry System
 
Body Adapted Wearable Bio Sensor
Body Adapted Wearable Bio SensorBody Adapted Wearable Bio Sensor
Body Adapted Wearable Bio Sensor
 
IRJET- Design and Implementation of Health Monitoring System
IRJET-  	  Design and Implementation of Health Monitoring SystemIRJET-  	  Design and Implementation of Health Monitoring System
IRJET- Design and Implementation of Health Monitoring System
 
WBSN based safe lifestyle: a case study of heartrate monitoring system
WBSN based safe lifestyle: a case study of heartrate monitoring system WBSN based safe lifestyle: a case study of heartrate monitoring system
WBSN based safe lifestyle: a case study of heartrate monitoring system
 
Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...
 
Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...Implementation of ecg signal acquisition and transmission through bluetooth t...
Implementation of ecg signal acquisition and transmission through bluetooth t...
 
Design and implementation of portable electrocardiogram recorder with field ...
Design and implementation of portable electrocardiogram  recorder with field ...Design and implementation of portable electrocardiogram  recorder with field ...
Design and implementation of portable electrocardiogram recorder with field ...
 
Wireless Body Area Networks for Healthcare: A Survey
Wireless Body Area Networks for Healthcare: A SurveyWireless Body Area Networks for Healthcare: A Survey
Wireless Body Area Networks for Healthcare: A Survey
 
research paper
research paper research paper
research paper
 
Real time ECG Monitoring: A Review
Real time ECG Monitoring: A ReviewReal time ECG Monitoring: A Review
Real time ECG Monitoring: A Review
 
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
 
Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...Implementation of electronic stethoscope for online remote monitoring with mo...
Implementation of electronic stethoscope for online remote monitoring with mo...
 
A general framework for improving electrocardiography monitoring system with ...
A general framework for improving electrocardiography monitoring system with ...A general framework for improving electrocardiography monitoring system with ...
A general framework for improving electrocardiography monitoring system with ...
 
Technical seminar.pptx is based on the Wearable BIO Sensors
Technical seminar.pptx is based on the Wearable BIO SensorsTechnical seminar.pptx is based on the Wearable BIO Sensors
Technical seminar.pptx is based on the Wearable BIO Sensors
 

More from ArhamSheikh1

Portable out‐of‐hospital electrocardiography a review of current technologie...
Portable out‐of‐hospital electrocardiography  a review of current technologie...Portable out‐of‐hospital electrocardiography  a review of current technologie...
Portable out‐of‐hospital electrocardiography a review of current technologie...ArhamSheikh1
 
Physical activity and risk of cardiovascular disease—a
Physical activity and risk of cardiovascular disease—aPhysical activity and risk of cardiovascular disease—a
Physical activity and risk of cardiovascular disease—aArhamSheikh1
 
Heart diseases overview
Heart diseases overviewHeart diseases overview
Heart diseases overviewArhamSheikh1
 
Heart disease causes prevention and current
Heart disease causes prevention and currentHeart disease causes prevention and current
Heart disease causes prevention and currentArhamSheikh1
 
Development and test of a portable ecg device with dry capacitive electrodes ...
Development and test of a portable ecg device with dry capacitive electrodes ...Development and test of a portable ecg device with dry capacitive electrodes ...
Development and test of a portable ecg device with dry capacitive electrodes ...ArhamSheikh1
 
Cardiovascular diseases traditional_and_non-tradit
Cardiovascular diseases traditional_and_non-traditCardiovascular diseases traditional_and_non-tradit
Cardiovascular diseases traditional_and_non-traditArhamSheikh1
 
1472 6874-4-s1-s15
1472 6874-4-s1-s151472 6874-4-s1-s15
1472 6874-4-s1-s15ArhamSheikh1
 

More from ArhamSheikh1 (7)

Portable out‐of‐hospital electrocardiography a review of current technologie...
Portable out‐of‐hospital electrocardiography  a review of current technologie...Portable out‐of‐hospital electrocardiography  a review of current technologie...
Portable out‐of‐hospital electrocardiography a review of current technologie...
 
Physical activity and risk of cardiovascular disease—a
Physical activity and risk of cardiovascular disease—aPhysical activity and risk of cardiovascular disease—a
Physical activity and risk of cardiovascular disease—a
 
Heart diseases overview
Heart diseases overviewHeart diseases overview
Heart diseases overview
 
Heart disease causes prevention and current
Heart disease causes prevention and currentHeart disease causes prevention and current
Heart disease causes prevention and current
 
Development and test of a portable ecg device with dry capacitive electrodes ...
Development and test of a portable ecg device with dry capacitive electrodes ...Development and test of a portable ecg device with dry capacitive electrodes ...
Development and test of a portable ecg device with dry capacitive electrodes ...
 
Cardiovascular diseases traditional_and_non-tradit
Cardiovascular diseases traditional_and_non-traditCardiovascular diseases traditional_and_non-tradit
Cardiovascular diseases traditional_and_non-tradit
 
1472 6874-4-s1-s15
1472 6874-4-s1-s151472 6874-4-s1-s15
1472 6874-4-s1-s15
 

Recently uploaded

Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Nehru place Escorts
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...Miss joya
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
 
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Nehru place Escorts
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls ServiceMiss joya
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls ServiceMiss joya
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiNehru place Escorts
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Miss joya
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 

Recently uploaded (20)

Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
Russian Call Girls Chennai Madhuri 9907093804 Independent Call Girls Service ...
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
VIP Call Girls Pune Vani 9907093804 Short 1500 Night 6000 Best call girls Ser...
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
 
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
 
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Hadapsar ( Pune) Girls Service
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls ServiceCALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune)  Girls Service
CALL ON ➥9907093804 🔝 Call Girls Baramati ( Pune) Girls Service
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 

Sensors 21-02777

  • 1. sensors Article Development and Test of a Portable ECG Device with Dry Capacitive Electrodes and Driven Right Leg Circuit Alessandro Zompanti 1,* , Anna Sabatini 1 , Simone Grasso 2 , Giorgio Pennazza 1 , Giuseppe Ferri 3 , Gianluca Barile 3 , Massimo Chello 4 , Mario Lusini 4 and Marco Santonico 2 Citation: Zompanti, A.; Sabatini, A.; Grasso, S.; Pennazza, G.; Ferri, G.; Barile, G.; Chello, M.; Lusini, M.; Santonico, M. Development and Test of a Portable ECG Device with Dry Capacitive Electrodes and Driven Right Leg Circuit. Sensors 2021, 21, 2777. https://doi.org/10.3390/ s21082777 Academic Editor: Giancarlo Fortino Received: 24 March 2021 Accepted: 13 April 2021 Published: 15 April 2021 Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affil- iations. Copyright: © 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https:// creativecommons.org/licenses/by/ 4.0/). 1 Unit of Electronics for Sensor Systems, Department of Engineering, Campus Bio-Medico University of Rome, 00128 Rome, Italy; a.sabatini@unicampus.it (A.S.); g.pennazza@unicampus.it (G.P.) 2 Unit of Electronics for Sensor Systems, Department of Science and Technology for Humans and the Environment, Campus Bio-Medico University of Rome, 00128 Rome, Italy; s.grasso@unicampus.it (S.G.); m.santonico@unicampus.it (M.S.) 3 Department of Industrial and Information Engineering and Economics, University of L’Aquila, 67100 L’Aquila, Italy; giuseppe.ferri@univaq.it (G.F.); gianluca.barile@univaq.it (G.B.) 4 Unit of Cardiology, Department of Medicine, Campus Bio-Medico University of Rome, 00128 Rome, Italy; m.chello@unicampus.it (M.C.); m.lusini@unicampus.it (M.L.) * Correspondence: a.zompanti@unicampus.it Abstract: The use of wearable sensors for health monitoring is rapidly growing. Over the past decade, wearable technology has gained much attention from the tech industry for commercial reasons and the interest of researchers and clinicians for reasons related to its potential benefit on patients’ health. Wearable devices use advanced and specialized sensors able to monitor not only activity parameters, such as heart rate or step count, but also physiological parameters, such as heart electrical activity or blood pressure. Electrocardiogram (ECG) monitoring is becoming one of the most attractive health-related features of modern smartwatches, and, because cardiovascular disease (CVD) is one of the leading causes of death globally, the use of a smartwatch to monitor patients could greatly impact the disease outcomes on health care systems. Commercial wearable devices are able to record just single-lead ECG using a couple of metallic contact dry electrodes. This kind of measurement can be used only for arrhythmia diagnosis. For the diagnosis of other cardiac disorders, additional ECG leads are required. In this study, we characterized an electronic interface to be used with multiple contactless capacitive electrodes in order to develop a wearable ECG device able to perform several lead measurements. We verified the ability of the electronic interface to amplify differential biopotentials and to reject common-mode signals produced by electromagnetic interference (EMI). We developed a portable device based on the studied electronic interface that represents a prototype system for further developments. We evaluated the performances of the developed device. The signal-to-noise ratio of the output signal is favorable, and all the features needed for a clinical evaluation (P waves, QRS complexes and T waves) are clearly readable. Keywords: wearable sensors and devices; wearable monitoring systems; wearable healthcare elec- trocardiography; wireless ECG monitoring systems; electronics for biosignals and biopotentials; electronic interfaces and embedded sensor systems for biomedical application; dry electrodes; capaci- tive coupling electrodes 1. Introduction The use of point-of-care (POC) systems and wearable sensors for health monitoring is rapidly growing. They are used both for clinical and for remote monitoring, improving the possibility to collect health data about patients continuously and to prevent health deterioration or clinical events [1]. POC and wearable sensors are able to empower patients, providing them with the right tools to monitor themselves autonomously. Patients are able to cooperate with medical Sensors 2021, 21, 2777. https://doi.org/10.3390/s21082777 https://www.mdpi.com/journal/sensors
  • 2. Sensors 2021, 21, 2777 2 of 15 teams and contribute to their own health [2]. Moreover, remote monitoring devices prevent the hospitalization of patients, with a consequent cost reduction for the healthcare system. Several data can be gathered using wearable sensors in order to monitor and record real-time information about patients’ health. Sensors measure physiological signs such as an electrocardiogram (ECG), an electromyogram (EMG), heart rate (HR), body temperature, electrodermal activity (EDA), arterial oxygen saturation (SpO2), blood pressure (BP) and respiration rate (RR) [3]. There is currently a growing interest in using wearable devices to monitor cardiovas- cular diseases [4]. According to recent statistics, cardiovascular disease (CVD) is one of the leading causes of death globally. In Europe as a whole, it is responsible for over 3.9 million deaths a year, or 45% of all deaths [5]. Over the past decade, wearable technology has gained the interest of researchers and clinicians due to the benefits that could be associated with the long-term monitoring of CVD patients, not necessarily at a hospital. Moreover, wearable devices have attracted much attention from the tech industry. Even if the primary use of commercial wearable devices is to monitor activity parameters (heart rate, step count, etc.), the sensors used are now so advanced that it is possible to also monitor physiological parameters directly related to CVD, such as heart electrical activity or blood pressure. Among the list of features forming the wearable technology paradigm (which can be found in many review papers or textbooks devoted to this subject [6–9]), it is here important to discuss those in this paper. A wearable device, especially for medical use, should be comfortable for the patient and adequate to follow movements. If possible, it should be effectively wearable, meaning that it could be possible to wear it with clothes. It should have electrical/electronic characteristics that make it the least disturbed by interferences. Several commercial wearable devices, such as Apple Watch (Apple Inc., Cupertino, CA, USA), Fitbit Sense (Fitbit Inc., San Francisco, CA, USA) or Withings ScanWatch (Withings, Issy-les-Moulineaux, France), are able to record a single-lead ECG using metallic dry- contact electrodes, in contrast to adhesive wet electrodes [10]. Typically, one electrode is placed on the front of the watch and can be touched with a finger, and another electrode is placed on the back, in contact with the wrist. This one-lead ECG can improve arrhythmia diagnosis; however, for the diagnosis of other cardiac disorders, such as myocardial infarction or ischemia, additional ECG leads are required [11]. Traditional 12-lead ECG is able to record cardiac potentials on the skin, with an am- plitude of a few mV and frequencies in the range of 0.01–150 Hz. The measurement has been performed for a limited time with the patient at rest, laying on an examination table. Several cables and standard wet Ag/AgCl electrodes are used. This type of electrode requires proper preparation of the skin [12]. Moreover, adhesive wet electrodes are not suitable for long-term monitoring purposes because they can cause irritation and lose their conductivity over time due to drying out [13]. As a consequence, dry-contact electrodes represent an alternative to wet contact electrodes [14], in particular for long-term monitor- ing, where they are more comfortable for the patient [15,16]. In recent years, noncontact electrodes have been used in several studies [17]. In particular, capacitive electrodes, cou- pled with the patient’s body, have been investigated [18–24]. A smart way to use more leads, marginally compromising patient comfort, could be by embedding the electrodes into clothes. Dry capacitive electrodes are highly suitable for this kind of application. In order to use capacitive electrodes, new electronic interfaces should be developed to match the application requirements. The two main requirements are electronic interface high impedance and EMI rejection. In this study, the possibility of recording multiple-lead ECG using dry capacitive elec- trodes with a custom-developed portable device was investigated to explore the feasibility of a wearable ECG device with electrodes embedded into clothes. The performances of an electronic interface proposed by Chi et al. [25] and others [26,27] were evaluated by characterizing it with simulations and electrical tests in order to deeper investigate its performances. Moreover, the studied electronic interface was implemented into a custom
  • 3. Sensors 2021, 21, 2777 3 of 15 realized portable device, and in order to verify the effectiveness of the system, a comparison between signals obtained using the developed device, with dry capacitive electrodes, and a commercial heart rate front-end IC, with standard wet electrodes, was performed. The novelty of this study is the deeper understanding of the studied electronic inter- face behavior, with respect to published results. This goal was achieved by characterizing the electronic interface in detail using simulations and electrical tests. An opportune experimental setup was designed to perform the measurements with experimental condi- tions as close as possible to real applications; that is, an “ECG T-shirt” with an embedded capacitive electrode. According to the obtained results, a specific circuit optimization method was proposed. Moreover, these results could be exploited by numerous real applications, such as drivers’ safety [21,28–33], remote monitoring of the elderly [34–37], sleep monitoring [20,38–42], human identification [43–46] and emotion recognition [47–50]. 2. Materials and Methods Biopotential signals, such as those generated by the electrical activity of the heart, are usually measured with a differential approach using voltage amplifiers with at least two basic requirements: input impedance, which should be as high as possible in order to match the electrode impedance, and the current flowing between the body and the amplifiers, which should be as low as possible for safety reasons. Biopotential measurements, such as ECG, are very sensitive to electromagnetic inter- ference (EMI). One of the most experienced EMI is caused by the coupling between the patient’s body and the power lines, resulting in an applied common-mode voltage (Vcm) on the electrodes. By performing differential biopotential measurement and improving the common- mode rejection (CMR) of the system, it is possible to contain this phenomenon. Differential measurement can be obtained by applying the biopotentials to the input stage of a classic instrumentation amplifier (INA) topology. EMI generates an Icm flowing through the patient’s body from the sinusoidal power- line source (120 V/60 Hz or 230/50 Hz) through coupling capacitances to the ground. In order to perform an effective ECG measurement, the patient’s body has to be grounded with the ECG device. This grounding connection contributes to the rise in the Vcm primarily because of its resistance. As a consequence, it is important to reduce the impedance of the grounding electrode or to compensate for the EMI-generated current. This result can be achieved using a driven right leg (DRL) electrode [51]. It is commonly applied to the right leg, but it can be applied elsewhere on the patient’s body. The DRL electrode senses the common-mode voltage on the body, giving feedback to the patient’s body. This negative feedback reduces Vcm to µV values [52]. This study investigates the potentialities of a suitable circuit topology, originally proposed by Chi et al. [25], to enhance its applicability via a deeper characterization, which should allow exploiting this strategy for innovative devices. The proposed electronic system uses two sensing capacitive electrodes and a DRL capacitive electrode, as shown in Figure 1. The capacitive electrodes can be placed on the patient’s clothes (e.g., a cotton shirt) and not be in contact with the skin. Therefore, from now on, we will refer to the system as a contactless ECG device. Each electrode is an input of the buffering stage of a differential amplifier based on the topology of a 3-input amplifier. The DRL is made of an inverting opamp and a buffering stage. The electrodes are active and use amplifiers in order to achieve a very high input impedance, as shown in Figure 2. A high-pass filtering stage, formed by R1 and C1, is used to reduce the signal artifacts.
  • 4. Sensors 2021, 21, 2777 4 of 15 Sensors 2021, 21, x FOR PEER REVIEW 4 of 16 Figure 1. Schematic of the electronic interface. Each electrode is connected to the buffering stage of an instrumentation amplifier (INA) configuration. The buffered and filtered output voltages will be acquired by two ADC (Analog To Digital Converter) channels with respect to the ground. The com- mon-mode voltage is connected to the driven right leg (DRL) electrode and is rejected. Each electrode is an input of the buffering stage of a differential amplifier based on the topology of a 3-input amplifier. The DRL is made of an inverting opamp and a buffer- ing stage. The electrodes are active and use amplifiers in order to achieve a very high input impedance, as shown in Figure 2. A high-pass filtering stage, formed by R1 and C1, is used to reduce the signal artifacts. Figure 2. Schematic of the electrode interface. A double buffering stage is implemented in order to achieve a very high input impedance. The CMR and the frequency response of the electronic interface of the contactless ECG were characterized by simulating the circuit in MULTISIM (National Instruments). Figure 1. Schematic of the electronic interface. Each electrode is connected to the buffering stage of an instrumentation amplifier (INA) configuration. The buffered and filtered output voltages will be acquired by two ADC (Analog To Digital Converter) channels with respect to the ground. The common-mode voltage is connected to the driven right leg (DRL) electrode and is rejected. Figure 1. Schematic of the electronic interface. Each electrode is connected to the buffering stage of an instrumentation amplifier (INA) configuration. The buffered and filtered output voltages will be acquired by two ADC (Analog To Digital Converter) channels with respect to the ground. The com- mon-mode voltage is connected to the driven right leg (DRL) electrode and is rejected. Each electrode is an input of the buffering stage of a differential amplifier based on the topology of a 3-input amplifier. The DRL is made of an inverting opamp and a buffer- ing stage. The electrodes are active and use amplifiers in order to achieve a very high input impedance, as shown in Figure 2. A high-pass filtering stage, formed by R1 and C1, is used to reduce the signal artifacts. Figure 2. Schematic of the electrode interface. A double buffering stage is implemented in order to achieve a very high input impedance. The CMR and the frequency response of the electronic interface of the contactless ECG were characterized by simulating the circuit in MULTISIM (National Instruments). Figure 2. Schematic of the electrode interface. A double buffering stage is implemented in order to achieve a very high input impedance. The CMR and the frequency response of the electronic interface of the contactless ECG were characterized by simulating the circuit in MULTISIM (National Instruments). The circuit was implemented on a PCB and tested. According to the obtained results, optimization of the circuit is proposed. Moreover, we developed a portable device based on the presented electronic layout. The output signals of the electronic interface are acquired using the 12-bit SAR ADCs of the ESP32 microcontroller (ESPRESSIF SYSTEMS, Shanghai) with a sampling frequency of 250 Hz. The ESP32 microcontroller is equipped with an internal Bluetooth module; thus, it is capable of transmitting digitized signals through a wireless connection to a PC or smartphone. The device is battery operated. A schematic of the developed device is reported in Figure 3.
  • 5. Sensors 2021, 21, 2777 5 of 15 the ESP32 microcontroller (ESPRESSIF SYSTEMS, Shanghai) with a samplin of 250 Hz. The ESP32 microcontroller is equipped with an internal Blueto thus, it is capable of transmitting digitized signals through a wireless connec or smartphone. The device is battery operated. A schematic of the developed device is reported in Figure 3. Figure 3. Block diagram of the developed contactless electrocardiogram (ECG) device. Each electrode is connected to the electronic interface using a 4-pole aud armature plate of the electrodes was implemented using directly the bottom PCBs, as shown in Figure 4. Figure 4. PCBs of the capacitive electrodes. The armature plate is placed in the bottom board under an insulating layer. It is made of copper and has a radius of 3 cm. To evaluate the effectiveness of the contactless ECG device, we performe ison between signals obtained using this device and those obtained using a fully integrated single-lead ECG front end by Analog Devices, using standa trodes. The AD8232 is an integrated signal conditioning block for portable E and other biopotential measurement applications. It implements a filter and a amplification stage. To improve the common-mode rejection of the IC, it inc ditional amplifier for the right leg drive [53]. We used the AD8232 in the “Cardiac Monitor Configuration,” as sugg IC’s application note [53]. This configuration is designed to perform ECG m with minimal distortion. The AD8232 is configured with a 0.5 Hz two-pole hig followed by a 40 Hz two-pole low-pass filter, and the total system gain is ab Figure 3. Block diagram of the developed contactless electrocardiogram (ECG) device. Each electrode is connected to the electronic interface using a 4-pole audio cable. The armature plate of the electrodes was implemented using directly the bottom layer of the PCBs, as shown in Figure 4. or smartphone. The device is battery operated. A schematic of the developed device is reported in Figure 3. Figure 3. Block diagram of the developed contactless electrocardiogram (ECG) device. Each electrode is connected to the electronic interface using a 4-pole audio armature plate of the electrodes was implemented using directly the bottom l PCBs, as shown in Figure 4. Figure 4. PCBs of the capacitive electrodes. The armature plate is placed in the bottom l board under an insulating layer. It is made of copper and has a radius of 3 cm. To evaluate the effectiveness of the contactless ECG device, we performed ison between signals obtained using this device and those obtained using an fully integrated single-lead ECG front end by Analog Devices, using standard trodes. The AD8232 is an integrated signal conditioning block for portable EC and other biopotential measurement applications. It implements a filter and a d amplification stage. To improve the common-mode rejection of the IC, it inclu ditional amplifier for the right leg drive [53]. We used the AD8232 in the “Cardiac Monitor Configuration,” as sugges IC’s application note [53]. This configuration is designed to perform ECG mea with minimal distortion. The AD8232 is configured with a 0.5 Hz two-pole high followed by a 40 Hz two-pole low-pass filter, and the total system gain is abou Figure 4. PCBs of the capacitive electrodes. The armature plate is placed in the bottom layer of the board under an insulating layer. It is made of copper and has a radius of 3 cm. To evaluate the effectiveness of the contactless ECG device, we performed a compari- son between signals obtained using this device and those obtained using an AD8232, a fully integrated single-lead ECG front end by Analog Devices, using standard wet electrodes. The AD8232 is an integrated signal conditioning block for portable ECG devices and other biopotential measurement applications. It implements a filter and a differential amplifi- cation stage. To improve the common-mode rejection of the IC, it includes an additional amplifier for the right leg drive [53]. We used the AD8232 in the “Cardiac Monitor Configuration,” as suggested by the IC’s application note [53]. This configuration is designed to perform ECG measurements with minimal distortion. The AD8232 is configured with a 0.5 Hz two-pole high-pass filter followed by a 40 Hz two-pole low-pass filter, and the total system gain is about 60 dB. A DRL is used to improve common-mode rejection; thus, the electronic approach is the same as the developed contactless ECG device. The frequency response of the system is reported in Figure 5.
  • 6. Sensors 2021, 21, 2777 6 of 15 Sensors 2021, 21, x FOR PEER REVIEW 6 of 16 DRL is used to improve common-mode rejection; thus, the electronic approach is the same as the developed contactless ECG device. The frequency response of the system is re- ported in Figure 5. Figure 5. Frequency response of the AD8232 implemented in the “Cardiac Monitor Configuration.” The bode plot shows a band-pass behavior in the range of frequencies between 0.5 and 40 Hz and a gain of about 60 dB (1000 V/V) (the bode plot is extracted from the AD8232 datasheet [53]). The output signals from the AD8232 are acquired using the ADCs of the ESP32 mi- crocontroller with a sampling frequency of 250 Hz. The ESP32 microcontroller is equipped with an internal Bluetooth module; thus, it is capable of transmitting digitized signals through a wireless connection to a PC or smartphone. The device is battery operated. The scheme of the developed device is reported in Figure 6. Figure 6. Block diagram of the developed ECG device based on the AD8232 IC. The device uses standard wet adhesive electrodes. In order to avoid interference between cables and electrodes of the two devices, ECG measurements were performed sequentially, placing the electrodes in the same position with the same experimental condition. The position of the electrodes implements Lead I of Einthoven’s triangle. Several measurements were performed on a single healthy subject at rest. This is be- cause our primary goal in this study was to evaluate the feasibility and effectiveness of Figure 5. Frequency response of the AD8232 implemented in the “Cardiac Monitor Configuration.” The bode plot shows a band-pass behavior in the range of frequencies between 0.5 and 40 Hz and a gain of about 60 dB (1000 V/V) (the bode plot is extracted from the AD8232 datasheet [53]). The output signals from the AD8232 are acquired using the ADCs of the ESP32 micro- controller with a sampling frequency of 250 Hz. The ESP32 microcontroller is equipped with an internal Bluetooth module; thus, it is capable of transmitting digitized signals through a wireless connection to a PC or smartphone. The device is battery operated. The scheme of the developed device is reported in Figure 6. Sensors 2021, 21, x FOR PEER REVIEW 6 of 16 DRL is used to improve common-mode rejection; thus, the electronic approach is the same as the developed contactless ECG device. The frequency response of the system is re- ported in Figure 5. Figure 5. Frequency response of the AD8232 implemented in the “Cardiac Monitor Configuration.” The bode plot shows a band-pass behavior in the range of frequencies between 0.5 and 40 Hz and a gain of about 60 dB (1000 V/V) (the bode plot is extracted from the AD8232 datasheet [53]). The output signals from the AD8232 are acquired using the ADCs of the ESP32 mi- crocontroller with a sampling frequency of 250 Hz. The ESP32 microcontroller is equipped with an internal Bluetooth module; thus, it is capable of transmitting digitized signals through a wireless connection to a PC or smartphone. The device is battery operated. The scheme of the developed device is reported in Figure 6. Figure 6. Block diagram of the developed ECG device based on the AD8232 IC. The device uses standard wet adhesive electrodes. In order to avoid interference between cables and electrodes of the two devices, ECG measurements were performed sequentially, placing the electrodes in the same position with the same experimental condition. The position of the electrodes implements Lead I of Einthoven’s triangle. Several measurements were performed on a single healthy subject at rest. This is be- cause our primary goal in this study was to evaluate the feasibility and effectiveness of Figure 6. Block diagram of the developed ECG device based on the AD8232 IC. The device uses standard wet adhesive electrodes. In order to avoid interference between cables and electrodes of the two devices, ECG measurements were performed sequentially, placing the electrodes in the same position with the same experimental condition. The position of the electrodes implements Lead I of Einthoven’s triangle. Several measurements were performed on a single healthy subject at rest. This is because our primary goal in this study was to evaluate the feasibility and effectiveness of the studied electronic interface; even one subject was sufficient. Clearly, further studies on a larger population will be necessary in order to clinically test the developed device. The electrodes were placed as shown in Figure 7.
  • 7. Sensors 2021, 21, 2777 7 of 15 Sensors 2021, 21, x FOR PEER REVIEW 7 of 16 the studied electronic interface; even one subject was sufficient. Clearly, further studies on a larger population will be necessary in order to clinically test the developed device. The electrodes were placed as shown in Figure 7. Figure 7. Position of the 3 electrodes on the body of the patient. 3. Results The electronic interface of the contactless ECG device was characterized. The fre- quency response of the electrodes was investigated. The electrical scheme of the electrode is reported in Figure 2. The simulation, re- ported in Figure 8, shows a first-order high-pass characteristic with a cut-off frequency of about 0.7 Hz and no gain. Figure 8. Frequency response of the active capacitive electrodes. It is useful to introduce a high-pass filtering stage to reduce the signal artifacts caused by electrodes or body movements, such as thoracic expansion during breathing. Figure 7. Position of the 3 electrodes on the body of the patient. 3. Results The electronic interface of the contactless ECG device was characterized. The fre- quency response of the electrodes was investigated. The electrical scheme of the electrode is reported in Figure 2. The simulation, reported in Figure 8, shows a first-order high-pass characteristic with a cut-off frequency of about 0.7 Hz and no gain. Figure 7. Position of the 3 electrodes on the body of the patient. 3. Results The electronic interface of the contactless ECG device was characterized. The fre- quency response of the electrodes was investigated. The electrical scheme of the electrode is reported in Figure 2. The simulation, re- ported in Figure 8, shows a first-order high-pass characteristic with a cut-off frequency of about 0.7 Hz and no gain. Figure 8. Frequency response of the active capacitive electrodes. It is useful to introduce a high-pass filtering stage to reduce the signal artifacts caused by electrodes or body movements, such as thoracic expansion during breathing. Figure 8. Frequency response of the active capacitive electrodes. It is useful to introduce a high-pass filtering stage to reduce the signal artifacts caused by electrodes or body movements, such as thoracic expansion during breathing. The frequency response of the differential opamp stage was investigated. The electrical scheme is reported in Figure 9.
  • 8. Sensors 2021, 21, 2777 8 of 15 Sensors 2021, 21, x FOR PEER REVIEW 8 of 16 The frequency response of the differential opamp stage was investigated. The elec- trical scheme is reported in Figure 9. Figure 9. Schematic of the differential stage. In this stage, the input signal is amplified and filtered. The simulation shows a second-order low-pass characteristic of the differential am- plifier with a gain of 60 dB and a bandwidth of about 100 Hz (see Figure 10). Figure 10. Frequency response of the differential stage. The system frequency response of the electrodes coupled with the differential stage was investigated. The complete circuit is shown in Figure 11. Figure 9. Schematic of the differential stage. In this stage, the input signal is amplified and filtered. The simulation shows a second-order low-pass characteristic of the differential ampli- fier with a gain of 60 dB and a bandwidth of about 100 Hz (see Figure 10). The frequency response of the differential opamp stage was investigated. The elec- trical scheme is reported in Figure 9. Figure 9. Schematic of the differential stage. In this stage, the input signal is amplified and filtered. The simulation shows a second-order low-pass characteristic of the differential am- plifier with a gain of 60 dB and a bandwidth of about 100 Hz (see Figure 10). Figure 10. Frequency response of the differential stage. The system frequency response of the electrodes coupled with the differential stage was investigated. The complete circuit is shown in Figure 11. Figure 10. Frequency response of the differential stage. The system frequency response of the electrodes coupled with the differential stage was investigated. The complete circuit is shown in Figure 11. Sensors 2021, 21, x FOR PEER REVIEW 9 of 16 Figure 11. Schematic of the capacitive active electrodes coupled with the differential stage. As expected, simulation results (see Figure 12) show a band-pass characteristic with a gain of 40 dB. Figure 11. Schematic of the capacitive active electrodes coupled with the differential stage.
  • 9. Sensors 2021, 21, 2777 9 of 15 As expected, simulation results (see Figure 12) show a band-pass characteristic with a gain of 40 dB. Figure 11. Schematic of the capacitive active electrodes coupled with the differential stage. As expected, simulation results (see Figure 12) show a band-pass characteristic with a gain of 40 dB. Figure 12. Frequency response of the capacitive active electrodes coupled with the differential stage. The bode plot shows a band-pass behavior in the range of frequencies between 0.7 and 100 Hz and a gain of about 60 dB. The voltage output of each channel with respect to the ground will exhibit a common- mode voltage gain equal to 1.0. Obviously, by performing a differential measurement be- tween the output voltage of Channel 1 and the output voltage of Channel 2, the Vcm will theoretically disappear (in a real scenario, it is a small voltage). Using a DRL circuit, it is possible to further increase the CMR of the system. Different gains of the DRL stage were simulated to study its feedback effect on Vcm. The schematic is shown in Figure 13. Figure 12. Frequency response of the capacitive active electrodes coupled with the differential stage. The bode plot shows a band-pass behavior in the range of frequencies between 0.7 and 100 Hz and a gain of about 60 dB. The voltage output of each channel with respect to the ground will exhibit a common- mode voltage gain equal to 1.0. Obviously, by performing a differential measurement between the output voltage of Channel 1 and the output voltage of Channel 2, the Vcm will theoretically disappear (in a real scenario, it is a small voltage). Using a DRL circuit, it is possible to further increase the CMR of the system. Different gains of the DRL stage were simulated to study its feedback effect on Vcm. The schematic is shown in Figure 13. Sensors 2021, 21, x FOR PEER REVIEW 10 of 16 Figure 13. Schematic of the complete system: capacitive electrodes, differential stage and DRL electrode. The electromag- netic interference from the power line was modeled as a sine voltage source coupled to the electrodes through a 2 pF capacitor. Moreover, the patient’s body was also coupled to the ground through a 200 pF capacitance [51]. The simulation results are shown in Figure 14. Figure 13. Schematic of the complete system: capacitive electrodes, differential stage and DRL electrode. The electromagnetic interference from the power line was modeled as a sine voltage source coupled to the electrodes through a 2 pF capacitor. Moreover, the patient’s body was also coupled to the ground through a 200 pF capacitance [51].
  • 10. Sensors 2021, 21, 2777 10 of 15 The simulation results are shown in Figure 14. Figure 13. Schematic of the complete system: capacitive electrodes, differential stage and DRL electrode. The electromag- netic interference from the power line was modeled as a sine voltage source coupled to the electrodes through a 2 pF capacitor. Moreover, the patient’s body was also coupled to the ground through a 200 pF capacitance [51]. The simulation results are shown in Figure 14. Figure 14. Plot showing the amplitude of the common-mode voltage Vcm versus the input re- sistance R7 (see Figure 15) of the DRL opamp. As expected, by lowering the gain of the DRL, the amplitude of the Vcm increased. The implemented circuit shown in Figure 11 was tested to verify its frequency re- sponse using a Lecroy Wave Station 3082 Waveform Generator as input and recording the output signals using a Lecroy HDO6054-MS Oscilloscope. The experimental setup at the block scheme level is shown in Figure 15. Figure 14. Plot showing the amplitude of the common-mode voltage Vcm versus the input resistance R7 (see Figure 15) of the DRL opamp. As expected, by lowering the gain of the DRL, the amplitude of the Vcm increased. Sensors 2021, 21, x FOR PEER REVIEW 11 of 16 Figure 15. Experimental setup at block scheme level used to characterize the frequency response of the capacitive electrodes coupled with the differential stage. The frequency response of the circuit is shown in Figure 16. The output signals were acquired using the internal 12-bit SAR ADCs of the ESP32 microcontroller and transmitted using Bluetooth to a PC. The sampling frequency was 250 Hz. About 20 s of recorded data obtained using the contactless ECG device are shown in Figure 17. In Figure 17, a sinusoidal pattern is clearly visible. It is caused by the movement of the rib cage during breathing. Even if this pattern is a sort of signal artifact, it could be used to determine the respiratory rate during an ECG measurement. In any case, this kind of low-frequency artifact can be filtered both in the analog and in the digital domain by applying a high-pass filter. In the original study, a compression vest was used to keep the electrodes in fixed positions and to mitigate this kind of artifact, but we think that this approach is not very comfortable for patients. In contrast, we used an elastic band in order to apply the same compression that a cotton T-shirt would apply. One of our aims in this study is to explore the feasibility of the embedding of the electrodes into clothes. In fact, a smart way to use several electrodes for a long period of time, marginally compromising patient comfort, could be, for example, by embedding the electrodes directly into a T-shirt. Figure 15. Experimental setup at block scheme level used to characterize the frequency response of the capacitive electrodes coupled with the differential stage. The implemented circuit shown in Figure 11 was tested to verify its frequency response using a Lecroy Wave Station 3082 Waveform Generator as input and recording the output signals using a Lecroy HDO6054-MS Oscilloscope. The experimental setup at the block scheme level is shown in Figure 15. The frequency response of the circuit is shown in Figure 16. Sensors 2021, 21, x FOR PEER REVIEW 11 of 16 Figure 15. Experimental setup at block scheme level used to characterize the frequency response of the capacitive electrodes coupled with the differential stage. The frequency response of the circuit is shown in Figure 16. The output signals were acquired using the internal 12-bit SAR ADCs of the ESP32 microcontroller and transmitted using Bluetooth to a PC. The sampling frequency was 250 Hz. About 20 s of recorded data obtained using the contactless ECG device are shown in Figure 17. In Figure 17, a sinusoidal pattern is clearly visible. It is caused by the movement of the rib cage during breathing. Even if this pattern is a sort of signal artifact, it could be used to determine the respiratory rate during an ECG measurement. In any case, this kind of low-frequency artifact can be filtered both in the analog and in the digital domain by applying a high-pass filter. In the original study, a compression vest was used to keep the electrodes in fixed positions and to mitigate this kind of artifact, but we think that this approach is not very comfortable for patients. In contrast, we used an elastic band in order to apply the same compression that a cotton T-shirt would apply. One of our aims in this study is to explore the feasibility of the embedding of the electrodes into clothes. In fact, a smart way to use several electrodes for a long period of time, marginally compromising patient comfort, could be, for example, by embedding the electrodes directly into a T-shirt. Figure 16. Frequency response of the capacitive active electrodes coupled with the differential stage. The band-pass behavior shown by the simulation was confirmed by the electrical tests. Dots represent experimental data. Figure 16. Frequency response of the capacitive active electrodes coupled with the differential stage. The band-pass behavior shown by the simulation was confirmed by the electrical tests. Dots represent experimental data.
  • 11. Sensors 2021, 21, 2777 11 of 15 The output signals were acquired using the internal 12-bit SAR ADCs of the ESP32 microcontroller and transmitted using Bluetooth to a PC. The sampling frequency was 250 Hz. About 20 s of recorded data obtained using the contactless ECG device are shown in Figure 17. Sensors 2021, 21, x FOR PEER REVIEW 12 of 16 Figure 17. Output signal obtained using the contactless ECG device. About 20 s of recorded data obtained using the AD8232 device are shown in Figure 18. In Figure 18, a slightly higher noise floor can be observed. No artifacts are shown. In Figure 19, signals recorded from both the contactless device and the AD8232-based device are reported. In order to better visualize the difference between the signals, several PQRST complexes, extracted from the raw output signals, are plotted. ECG complexes, extracted from the two devices, are quite comparable but not iden- tical. The timings of intervals and segments of the complexes are the same, but the relative amplitudes of the R peak and the T wave are different. Moreover, S peaks are not visible in the AD8232 complexes. Figure 18. Output signal obtained using the AD8232 device. 0 2 4 6 8 10 12 14 16 18 Time [s] -1 -0.8 -0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8 1 Amplitude [V] 0 2.5 5 7.5 10 12.5 15 17.5 Time [s] -1 -0.8 -0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8 1 Amplitude [V] Figure 17. Output signal obtained using the contactless ECG device. In Figure 17, a sinusoidal pattern is clearly visible. It is caused by the movement of the rib cage during breathing. Even if this pattern is a sort of signal artifact, it could be used to determine the respiratory rate during an ECG measurement. In any case, this kind of low-frequency artifact can be filtered both in the analog and in the digital domain by applying a high-pass filter. In the original study, a compression vest was used to keep the electrodes in fixed positions and to mitigate this kind of artifact, but we think that this approach is not very comfortable for patients. In contrast, we used an elastic band in order to apply the same compression that a cotton T-shirt would apply. One of our aims in this study is to explore the feasibility of the embedding of the electrodes into clothes. In fact, a smart way to use several electrodes for a long period of time, marginally compromising patient comfort, could be, for example, by embedding the electrodes directly into a T-shirt. About 20 s of recorded data obtained using the AD8232 device are shown in Figure 18. Sensors 2021, 21, x FOR PEER REVIEW 12 of 16 Figure 17. Output signal obtained using the contactless ECG device. About 20 s of recorded data obtained using the AD8232 device are shown in Figure 18. In Figure 18, a slightly higher noise floor can be observed. No artifacts are shown. In Figure 19, signals recorded from both the contactless device and the AD8232-based device are reported. In order to better visualize the difference between the signals, several PQRST complexes, extracted from the raw output signals, are plotted. ECG complexes, extracted from the two devices, are quite comparable but not iden- tical. The timings of intervals and segments of the complexes are the same, but the relative amplitudes of the R peak and the T wave are different. Moreover, S peaks are not visible in the AD8232 complexes. Figure 18. Output signal obtained using the AD8232 device. 0 2 4 6 8 10 12 14 16 18 Time [s] -1 -0.8 -0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8 1 Amplitude [V] 0 2.5 5 7.5 10 12.5 15 17.5 Time [s] -1 -0.8 -0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8 1 Amplitude [V] Figure 18. Output signal obtained using the AD8232 device. In Figure 18, a slightly higher noise floor can be observed. No artifacts are shown. In Figure 19, signals recorded from both the contactless device and the AD8232-based device are reported. In order to better visualize the difference between the signals, several PQRST complexes, extracted from the raw output signals, are plotted.
  • 12. Sensors 2021, 21, 2777 12 of 15 Sensors 2021, 21, x FOR PEER REVIEW 13 of 16 Figure 19. Comparison between the contactless device and AD8232 device signals. About 60 PQRST complexes were extracted from the raw output signals and overlapped in a single plot for each de- vice. 4. Discussion and Conclusions We simulated, implemented and tested an electronic interface for ECG devices. The results obtained from simulations show the ability of the electronic interface to amplify differential biopotentials and to reject common-mode signals produced by EMI in an ef- fective way. The results obtained from the electrical testing on the implemented PCB con- firm these capabilities. Further adjustments can be applied in order to mostly optimize the response of the system at low frequencies. The studied interface can be used with a capacitive electrode that can be placed on the clothes of the patient. This contactless ECG device is more comfortable for long-term monitoring instead of standard ECG devices using adhesive wet electrodes. The circuit topology allows the implementation of the electronic interface with very low costs; in fact, unlike many other electronic interfaces for ECG devices, it does not use expensive Instru- mentation amplifiers. Moreover, the topology is highly modular, allowing the use of a flexible number of electrodes. To evaluate the effectiveness of the contactless ECG device, we performed ECG measurements on a healthy subject. The biopotentials were acquired using both the stud- ied electronic interface and a commercial fully integrated single-lead ECG front end by an analog device. Both electronic interfaces use an approach based on differential amplifica- tion, filtering and DRL electrodes in order to obtain a high signal-to-noise ratio. The first electronic interface makes use of capacitive electrodes, and the second interface uses standard wet Ag/AgCl electrodes. The obtained measurements are quite comparable in terms of the amplitude and shape of the recorded signal. The ECG traces, recorded using a contactless device, present a satisfying signal-to-noise ratio. Moreover, the acquired signals present all the features needed for a clinical evaluation: P waves, QRS complex and T wave. Contactless ECG traces showed a sinusoidal pattern caused by the movement of the rib cage during breath- ing. This artifact could be used to determine the respiratory rate of the patient, obtaining one more useful piece of information during the ECG measurement. The characterization of the electronic interface behavior in the frequency domain highlights how optimization of the filtering stages could decrease this kind of artifact. Moreover, capacitive electrodes are sensitive to nearby charge displacements. As was expected, during the tests, we experienced signal artifacts of the output signal caused by body movements, in particular arm movements and, in a minor way, movements of the head and shoulders. We think that before testing the system on a person in motion (e.g., 0 50 100 150 200 250 Sample Number 0 0.2 0.4 0.6 0.8 1 1.2 Amplitude [V] Contactless 0 50 100 150 200 250 Sample Number 0 0.2 0.4 0.6 0.8 1 1.2 Amplitude [V] AD8232 Figure 19. Comparison between the contactless device and AD8232 device signals. About 60 PQRST complexes were extracted from the raw output signals and overlapped in a single plot for each device. ECG complexes, extracted from the two devices, are quite comparable but not identical. The timings of intervals and segments of the complexes are the same, but the relative amplitudes of the R peak and the T wave are different. Moreover, S peaks are not visible in the AD8232 complexes. 4. Discussion and Conclusions We simulated, implemented and tested an electronic interface for ECG devices. The results obtained from simulations show the ability of the electronic interface to amplify differential biopotentials and to reject common-mode signals produced by EMI in an effective way. The results obtained from the electrical testing on the implemented PCB confirm these capabilities. Further adjustments can be applied in order to mostly optimize the response of the system at low frequencies. The studied interface can be used with a capacitive electrode that can be placed on the clothes of the patient. This contactless ECG device is more comfortable for long-term monitoring instead of standard ECG devices using adhesive wet electrodes. The circuit topology allows the implementation of the electronic interface with very low costs; in fact, unlike many other electronic interfaces for ECG devices, it does not use expensive Instrumentation amplifiers. Moreover, the topology is highly modular, allowing the use of a flexible number of electrodes. To evaluate the effectiveness of the contactless ECG device, we performed ECG measurements on a healthy subject. The biopotentials were acquired using both the studied electronic interface and a commercial fully integrated single-lead ECG front end by an analog device. Both electronic interfaces use an approach based on differential amplification, filtering and DRL electrodes in order to obtain a high signal-to-noise ratio. The first electronic interface makes use of capacitive electrodes, and the second interface uses standard wet Ag/AgCl electrodes. The obtained measurements are quite comparable in terms of the amplitude and shape of the recorded signal. The ECG traces, recorded using a contactless device, present a satisfying signal-to-noise ratio. Moreover, the acquired signals present all the features needed for a clinical evaluation: P waves, QRS complex and T wave. Contactless ECG traces showed a sinusoidal pattern caused by the movement of the rib cage during breathing. This artifact could be used to determine the respiratory rate of the patient, obtaining one more useful piece of information during the ECG measurement. The characterization of the electronic interface behavior in the frequency domain highlights how optimization of the filtering stages could decrease this kind of artifact. Moreover, capacitive electrodes are sensitive to nearby charge displacements. As was expected, during the tests, we experienced signal artifacts of the output signal caused by
  • 13. Sensors 2021, 21, 2777 13 of 15 body movements, in particular arm movements and, in a minor way, movements of the head and shoulders. We think that before testing the system on a person in motion (e.g., walking, running), further investigations are needed to understand how to reduce signal interferences generated by movements of the upper body. It is important to highlight that although we tested the contactless ECG device imple- menting only two electrodes and a DRL grounding electrode in a single-lead configuration, it can also be used in a multiple-lead configuration with more than two electrodes. In further studies, we would like to implement and test the contactless ECG device with at least three leads in order to record standard and augmented limb leads. With more electrodes, it would be possible to record a complete 12-lead ECG. Thus, the use of a greater number of electrodes allows the recording of information content greater than those obtainable with wearable devices (e.g., smartwatches) currently available on the market. The implementation of more leads and the recording of more informative signals will allow going beyond mere arrhythmia diagnosis. In any case, further studies will be needed to evaluate the possibility of using ECG recorded using a multilead contactless ECG device to provide a diagnosis of other cardiac disorders. Obviously, it will be less comfortable to measure an ECG using several electrodes placed on the body instead of a smartwatch, but huge improvements can be performed with regard to the implementation of the electrodes. As the capacitive electrodes do not require direct contact with the patient’s skin, they could be embedded inside the clothes; for example, in the fabric of a T-shirt using flexible electronics. Therefore, in further studies, we would also like to optimize the implementation of the electrodes in order to make the contactless ECG device more comfortable and wearable. Author Contributions: Conceptualization, A.Z. and M.S.; methodology, A.Z. and M.S.; software, A.S. and S.G.; validation, A.Z., G.P., M.S. and M.C.; formal analysis, A.Z., S.G., A.S. and G.B.; investigation, G.P., M.C. and M.L.; resources, M.S.; data curation, A.Z., A.S. and S.G.; writing—original draft preparation, A.Z. and M.S.; writing—review and editing, G.P., M.C. and G.F.; supervision, G.P., M.S., G.F. and M.C. All authors have read and agreed to the published version of the manuscript. Funding: This research was funded by the Italian Ministry of Research and University in the frame of the European FLAG-ERA funding scheme. Name of the project: CONVERGENCE: “Frictionless Energy Efficient Convergent Wearables for Healthcare and Lifestyle Applications.” Institutional Review Board Statement: UCBM Ethical Committee approved the trial on 5 March, 2018: Protocol Number 28/18OSS ComEtCBM. Informed Consent Statement: Informed consent was obtained from all subjects involved in the study. Data Availability Statement: Data are available upon reasonable request. Conflicts of Interest: The authors declare no conflict of interest. References 1. Kristoffersson, A.; Lindén, M. A systematic review on the use of wearable body sensors for health monitoring: A qualitative synthesis. Sensors 2020, 20, 1502. [CrossRef] [PubMed] 2. Ding, E.Y.; Ensom, E.; Hafer, N.; Buchholz, B.; Picard, M.A.; Dunlap, D.; McManus, D.D. Point-of-care technologies in heart, lung, blood and sleep disorders from the Center for Advancing Point-of-Care Technologies. Curr. Opin. Biomed. Eng. 2019, 11, 58–67. [CrossRef] [PubMed] 3. Pantelopoulos, A.; Bourbakis, N.G. A survey on wearable sensor-based systems for health monitoring and prognosis. IEEE Trans. Syst. Man. Cybern. Part. C (Appl. Rev.) 2019, 40, 1–12. [CrossRef] 4. Mizuno, A.; Changolkar, S.; Patel, M.S. Wearable Devices to Monitor and Reduce the Risk of Cardiovascular Disease: Evidence and Opportunities. Annu. Rev. Med. 2020, 72, 459–471. [CrossRef] [PubMed] 5. Wilkins, E.; Wilson, L.; Wickramasinghe, K.; Bhatnagar, P.; Leal, J.; Luengo-Fernandez, R.; Townsend, N. European Cardiovascular Disease Statistics; European Heart Network: Brussels, Belgium, 2017. Available online: http://www.ehnheart.org/images/CVD- statistics-report-August-2017.pdf (accessed on 1 March 2021). 6. Patel, S.; Park, H.; Bonato, P.; Chan, L.; Rodgers, M. A review of wearable sensors and systems with application in rehabilitation. J. Neuroeng. Rehabil. 2012, 9, 1–17. [CrossRef]
  • 14. Sensors 2021, 21, 2777 14 of 15 7. Iqbal, M.H.; Aydin, A.; Brunckhorst, O.; Dasgupta, P.; Ahmed, K. A review of wearable technology in medicine. J. R. Soc. Med. 2016, 109, 372–380. [CrossRef] [PubMed] 8. McCann, J.; Bryson, D. Smart Clothes and Wearable Technology; Woodhead Publishing: Cambridge, UK, 2009; ISBN 9781845693572. 9. Tong, R. Wearable Technology in Medicine and Health Care; Academic Press: New York, NY, USA, 2018; ISBN 9780128118108. 10. Rashkovska, A.; Depolli, M.; Tomašić, I.; Avbelj, V.; Trobec, R. Medical-grade ECG sensor for long-term monitoring. Sensors 2020, 20, 1695. [CrossRef] 11. Samol, A.; Bischof, K.; Luani, B.; Pascut, D.; Wiemer, M.; Kaese, S. Single-lead ECG recordings including Einthoven and Wilson leads by a smartwatch: A new era of patient directed early ECG differential diagnosis of cardiac diseases? Sensors 2019, 19, 4377. [CrossRef] 12. Miller, H.A.; Harrison, D.C. Biomedical Electrode Technology: Theory and Practice; Academic Press: New York, NY, USA, 1974; ISBN 9780124968509. 13. Searle, A.; Kirkup, L. A direct comparison of wet, dry and insulating bioelectric recording electrodes. Physiol. Meas. 2000, 21. [CrossRef] 14. Chi, Y.M.; Jung, T.P.; Cauwenberghs, G. Dry-Contact and Noncontact Biopotential Electrodes: Methodological Review. IEEE Rev. Biomed. Eng. 2010, 3, 106–119. [CrossRef] 15. Burke, M.J.; Gleeson, D.T. A micropower dry-electrode ECG preamplifier. IEEE Trans. Biomed. Eng. 2000, 47, 155–162. [CrossRef] [PubMed] 16. Lin, C.T.; Liao, L.D.; Liu, Y.H.; Wang, I.J.; Lin, B.S.; Chang, J.Y. Novel dry polymer foam electrodes for long-term EEG measurement. IEEE Trans. Biomed. Eng. 2011, 58, 1200–1207. [CrossRef] [PubMed] 17. Teichmann, D.; Foussier, J.; Jia, J.; Leonhardt, S.; Walter, M. Noncontact monitoring of cardiorespiratory activity by electromagnetic coupling. IEEE Trans. Biomed. Eng. 2013, 60, 2142–2152. [CrossRef] [PubMed] 18. Sullivan, T.J.; Deiss, S.R.; Cauwenberghs, G. A low-noise, non-contact EEG/ECG sensor. In Proceedings of the Biomedical Circuits and Systems Conference, Montreal, QC, Canada, 27–30 November 2007; pp. 154–157. 19. Ueno, A.; Akabane, Y.; Kato, T.; Hoshino, H.; Kataoka, S.; Ishiyama, Y. Capacitive sensing of electrocardiographic potential through cloth from the dorsal surface of the body in a supine position: A preliminary study. IEEE Trans. Biomed. Eng. 2007, 54, 759–766. [CrossRef] [PubMed] 20. Lim, Y.G.; Kim, K.K.; Park, K.S. ECG recording on a bed during sleep without direct skin-contact. IEEE Trans. Biomed. Eng. 2007, 54, 718–725. [CrossRef] [PubMed] 21. Wartzek, T.; Eilebrecht, B.; Lem, J.; Lindner, H.J.; Leonhardt, S.; Walter, M. ECG on the road: Robust and unobtrusive estimation of heart rate contact. IEEE Trans. Biomed. Eng. 2011, 58, 3112–3120. [CrossRef] 22. Gao, S.; Huang, D.Y.; Fang, W.C. A non-contact ECG sensing system with motion artifact equalization. In Proceedings of the IEEE Biomedical Circuits and Systems Conference (BioCAS 2016), Shanghai, China, 17–19 October 2016; pp. 244–247. 23. Parente, F.R.; Santonico, M.; Zompanti, A.; Benassai, M.; Ferri, G.; D’Amico, A.; Pennazza, G. An electronic system for the contactless reading of ECG signals. Sensors 2017, 17, 2474. [CrossRef] 24. Lessard-Tremblay, M.; Weeks, J.; Morelli, L.; Cowan, G.; Gagnon, G.; Zednik, R.J. Contactless capacitive electrocardiography using hybrid flexible printed electrodes. Sensors 2020, 20, 5156. [CrossRef] [PubMed] 25. Chi, Y.M.; Ng, P.; Kang, E.; Kang, J.; Fang, J.; Cauwenberghs, G. Wireless non-contact cardiac and neural monitoring. In Proceedings of the Wireless Health, San Diego, CA, USA, 5−7 October 2010; pp. 15–23. 26. Van Rijn, A.M.; Peper, A.; Grimbergen, C.A. High-quality recording of bioelectric events. Med. Biol. Eng. Comput. 1990, 28, 389–397. [CrossRef] 27. Lim, Y.G.; Chung, G.S.; Park, K.S. Capacitive driven-right-leg grounding in indirect-contact ECG measurement. In Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC 2010), Buenos Aires, Argentina, 31 August–4 September 2010; pp. 1250–1253. [CrossRef] 28. Singh, R.K.; Sarkar, A.; Anoop, C.S. A health monitoring system using multiple non-contact ECG sensors for automotive drivers. In Proceedings of the IEEE International Instrumentation and Measurement Technology Conference, Taipei, Taiwan, 23–26 May 2016; pp. 1–6. [CrossRef] 29. Gromer, M.; Salb, D.; Walzer, T.; Madrid, N.M.; Seepold, R. ECG sensor for detection of driver’s drowsiness. Procedia Comput. Sci. 2019, 159, 1938–1946. [CrossRef] 30. Leonhardt, S.; Aleksandrowicz, A. Non-contact ECG monitoring for automotive application. In Proceedings of the IEEE 5th International Summer School and Symposium on Medical Devices and Biosensors, Hong Kong, China, 1–3 June 2008; pp. 183–185. [CrossRef] 31. Leicht, L.; Skobel, E.; Knackstedt, C.; Mathissen, M.; Sitter, A.; Wartzek, T.; Teichmann, D. Capacitive ECG monitoring in cardiac patients during simulated driving. IEEE Trans. Biomed. Eng. 2018, 66, 749–758. [CrossRef] 32. Choi, Y.J.; Lee, J.Y.; Kong, S.H. Driver ECG Measuring System With a Conductive Fabric-Based Dry Electrode. IEEE Access 2018, 6, 415–427. [CrossRef] 33. Walter, M.; Eilebrecht, B.; Wartzek, T.; Leonhardt, S. The smart car seat: Personalized monitoring of vital signs in automotive applications. Pers. Ubiquitous Comput. 2011, 15, 707–715. [CrossRef]
  • 15. Sensors 2021, 21, 2777 15 of 15 34. Wang, Y.H.; Chung, C.G.; Lin, C.C.; Lin, C.M. The study of the electrocardiography monitoring for the elderly based on smart clothes. In Proceedings of the IEEE 2018 Eighth International Conference on Information Science and Technology (ICIST), Istanbul, Turkey, 16–18 March 2018; pp. 478–482. [CrossRef] 35. Lee, H.J.; Lee, S.H.; Ha, K.S.; Jang, H.C.; Chung, W.Y.; Kim, J.Y.; Chang, Y.S.; Yoo, D.H. Ubiquitous healthcare service using Zigbee and mobile phone for elderly patients. Int. J. Med. Inform. 2009, 78, 193–198. [CrossRef] 36. Fan, X.; Zhao, Y.; Wang, H.; Tsui, K.L. Forecasting one-day-forward wellness conditions for community-dwelling elderly with single lead short electrocardiogram signals. BMC Med. Inform. Decis. Mak. 2019, 19, 285. [CrossRef] [PubMed] 37. Chung, W.Y.; Bhardwaj, S.; Punvar, A.; Lee, D.S.; Myllylae, R. A fusion health monitoring using ECG and accelerometer sensors for elderly persons at home. In Proceedings of the 2007 29th annual international conference of the IEEE Engineering in Medicine and Biology Society, Lyon, France, 22–26 August 2007; pp. 3818–3821. [CrossRef] 38. Bsoul, M.; Minn, H.; Tamil, L. Apnea MedAssist: Real-time sleep apnea monitor using single-lead ECG. IEEE Trans. Inf. Technol. Biomed. 2010, 15, 416–427. [CrossRef] 39. Kesper, K.; Canisius, S.; Penzel, T.; Ploch, T.; Cassel, W. ECG signal analysis for the assessment of sleep-disordered breathing and sleep pattern. Med. Biol. Eng. Comput. 2012, 50, 135–144. [CrossRef] [PubMed] 40. Varon, C.; Caicedo, A.; Testelmans, D.; Buyse, B.; Van Huffel, S. A novel algorithm for the automatic detection of sleep apnea from single-lead ECG. IEEE Trans. Biomed. Eng. 2015, 62, 2269–2278. [CrossRef] 41. Almazaydeh, L.; Elleithy, K.; Faezipour, M. Detection of obstructive sleep apnea through ECG signal features. In Proceedings of the 2012 IEEE International Conference on Electro/Information Technology, Indianapolis, IN, USA, 6–8 May 2012; pp. 1–6. [CrossRef] 42. Yilmaz, B.; Asyali, M.H.; Arikan, E.; Yetkin, S.; Özgen, F. Sleep stage and obstructive apneaic epoch classification using single-lead ECG. Biomed. Eng. Online 2010, 9, 1–14. [CrossRef] 43. Biel, L.; Pettersson, O.; Philipson, L.; Wide, P. ECG analysis: A new approach in human identification. IEEE Trans. Instrum. Meas. 2001, 50, 808–812. [CrossRef] 44. Singh, Y.N.; Gupta, P. ECG to individual identification. In Proceeding of the 2008 IEEE Second International Conference on Biometrics: Theory, Applications and Systems, Washington, DC, USA, 29 September–1 October 2008; pp. 1–8. [CrossRef] 45. Saechia, S.; Koseeyaporn, J.; Wardkein, P. Human identification system based ECG signal. In Proceedings of the TENCON 2005–2005 IEEE Region 10 Conference, Melbourne, VIC, Australia, 21–24 November 2005; pp. 1–4. [CrossRef] 46. Ye, C.; Coimbra, M.T.; Kumar, B.V. Investigation of human identification using two-lead electrocardiogram (ECG) signals. In Proceedings of the 2010 Fourth IEEE International Conference on Biometrics: Theory, Applications and Systems (BTAS), Washington, DC, USA, 27–29 September 2010; pp. 1–8. [CrossRef] 47. Covello, R.; Fortino, G.; Gravina, R.; Aguilar, A.; Breslin, J.G. Novel method and real-time system for detecting the Cardiac Defense Response based on the ECG. In Proceedings of the 2013 IEEE International Symposium on Medical Measurements and Applications (MeMeA), Gatineau, QC, Canada, 4–5 May 2013; pp. 53–57. [CrossRef] 48. Gravina, R.; Fortino, G. Automatic methods for the detection of accelerative cardiac defense response. IEEE Trans. Affect. Comput. 2016, 7, 286–298. [CrossRef] 49. Fortino, G.; Gravina, R. Real-time automatic detection of accelerative cardiac defense response. In Proceedings of the 9th International Conference on Body Area Networks, London, UK, 29 September–1 October 2014; pp. 159–164. [CrossRef] 50. Ahamed, M.A.; Ahad, M.A.U.; Sohag, M.H.A.; Ahmad, M. Development of low cost wireless biosignal acquisition system for ECG EMG and EOG. In Proceedings of the IEEE 2015 2nd International Conference on Electrical Information and Communication Technologies (EICT), Khulna, Bangladesh, 10–12 December 2015; pp. 195–199. [CrossRef] 51. Winter, B.B.; Webster, J.G. Driven-right-leg circuit design. IEEE Trans. Biomed. Eng. 1983, 1, 62–66. [CrossRef] [PubMed] 52. Clark, J.W., Jr.; Neuman, M.R.; Olson, W.H.; Peura, R.A.; Primiano, F.P., Jr.; Siedband, M.P.; Webster, J.G.; Wheeler, L.A. Contributing Authors. In Medical Instrumentation: Application and Design, 4th ed.; Webster, J.G., Ed.; John Wiley Sons: New York, NY, USA, 2010; ISBN 13-978-0471-57600-3. 53. Analog Devices, Inc. D8232 IC Datasheet. Available online: https://www.analog.com/media/en/technical-documentation/ data-sheets/ad8232.pdf (accessed on 1 March 2021).