Z100E Series V1.0
Z100E
Z100ES
Z100ES
Z100EL
Z100EL
Z100ES Z100EL
Compact magnetic flow meter
Remote magnetic flow meter
Language Chinese,English
Defaultsetup Defaultinto factory setup
Save defaultsetup Savefactory default setup
LCDdisplay contract setup
LCDbacklight function Open /close
Backlight automaticclose 1.5.10.30.60minutes for option
Deletelogger Deleterecords
setup forward flowvolume setup
setup reverse flow volume setup
Clear sum key Deletevolume flow
Display password Displaylevel 1,2,3password
Password setup setup level1,2,3password
Communication Modbus Address
Bit rate
Parity bitsetup
Output setup Pulseoutputtype Pulse, frequency
Pulsepolarity Positive/negative
Scaledpulse unit L,setup tablebetween 0.001L~10000.000L
Pulsewidth Automatic/manual
Frequency output range 0~10000HZ
Digital output Flow direction,upper/lowflow limit alarm
Currentoutputadjustment setup
Diagnosis test 4-20mAoutputtest Simulatedcurrent output
Flow velocity test Simulate full pipe situation (output
frequency,pulse and current)
Pulseoutputtest Simulatepulseoutput
Frequency,outputtest Simulatefrequency output
Logger inquiry
Stop and startlogger Record recent 50times of stop and starting
flow
Monthly volumeflowrecord Recordtheflow ineach month
Daily volumeflow record Recordtheflow ineach day
First levelmenu Second levelmenu Third level menu
Parametersetup Measurement pipesize 3mm-3000mm
Damping time 0-99seconds
Flow rateunit L/h, L/m, L/s,m3/h,m3/m,m3/s, t/h, t/m,t/s,
kg/h, kg/m, kg/s
Decimalpoints Automatic; manual (up to 3 decimal points can
be setup)
Volumeflowunit 0.001m3,0.01m3,0.1m3,1m3,0.001L,0.01L,0.1L,
1L,1T,1KG
Flow range setup
Excitation frequency 1/2, 1/4, 1/8, 1/16, 1/32 power frequency
Excitation current 0,20%,40%,55%,70%,80%,90%,100% (250mA)
Liquiddensity setup
Functionsetup Flow direction Forward and reverse
Reverse flow measure Enable/disable
Reverse flow output Enable/disable
Flow cutoff Enable/disable
Low flowcutoff setup
Stabilizeflow function Enable/disable
Stabilizeflow coefficient setup
Alarmallowance Enable/disable
Systemexciting alarm Enable/disable
Empty pipealarm Enable/disable
Empty pipealarm trigger setup thevalue totrig empty pipealarm
Empty pipemode 0 , 1
Upperflow limitalarm Enable/disable
Upperflow alarm value setup
Low flowlimit alarm Enable/disable
Low flowalarm value setup
Batterpower alarm Enable/disable
Batterpower display Display lithiumbattery percent
Language Chinese,English
Defaultsetup Defaultinto factory setup
Save defaultsetup Savefactory default setup
LCDdisplay contract setup
LCDbacklight function Open /close
Backlight automaticclose 1.5.10.30.60minutes for option
Deletelogger Deleterecords
setup forward flowvolume setup
setup reverse flow volume setup
Clear sum key Deletevolume flow
Display password Displaylevel 1,2,3password
Password setup setup level1,2,3password
Communication Modbus Address
Bit rate
Parity bitsetup
Output setup Pulseoutputtype Pulse, frequency
Pulsepolarity Positive/negative
Scaledpulse unit L,setup tablebetween 0.001L~10000.000L
Pulsewidth Automatic/manual
Frequency output range 0~10000HZ
Digital output Flow direction,upper/lowflow limit alarm
Currentoutputadjustment setup
Diagnosis test 4-20mAoutputtest Simulatedcurrent output
Flow velocity test Simulate full pipe situation (output
frequency,pulse and current)
Pulseoutputtest Simulatepulseoutput
Frequency,outputtest Simulatefrequency output
Logger inquiry
Stop and startlogger Record recent 50times of stop and starting
flow
Monthly volumeflowrecord Recordtheflow ineach month
Daily volumeflow record Recordtheflow ineach day
First levelmenu Second levelmenu Third level menu
Parametersetup Measurement pipesize 3mm-3000mm
Damping time 0-99seconds
Flow rateunit L/h, L/m, L/s,m3/h,m3/m,m3/s, t/h, t/m,t/s,
kg/h, kg/m, kg/s
Decimalpoints Automatic; manual (up to 3 decimal points can
be setup)
Volumeflowunit 0.001m3,0.01m3,0.1m3,1m3,0.001L,0.01L,0.1L,
1L,1T,1KG
Flow range setup
Excitation frequency 1/2, 1/4, 1/8, 1/16, 1/32 power frequency
Excitation current 0,20%,40%,55%,70%,80%,90%,100% (250mA)
Liquiddensity setup
Functionsetup Flow direction Forward and reverse
Reverse flow measure Enable/disable
Reverse flow output Enable/disable
Flow cutoff Enable/disable
Low flowcutoff setup
Stabilizeflow function Enable/disable
Stabilizeflow coefficient setup
Alarmallowance Enable/disable
Systemexciting alarm Enable/disable
Empty pipealarm Enable/disable
Empty pipealarm trigger setup thevalue totrig empty pipealarm
Empty pipemode 0 , 1
Upperflow limitalarm Enable/disable
Upperflow alarm value setup
Low flowlimit alarm Enable/disable
Low flowalarm value setup
Batterpower alarm Enable/disable
Batterpower display Display lithiumbattery percent
System datesetup setup anddisplay date
System timesetup setup anddisplay time
Sensor manufacturedate setup anddisplay manufacturedate
Sensor SN. NO. setup anddisplay product SN.NO.
Meter manufacturedate setup and display transmitter manufacture
date
Meter SN.NO. setup anddisplay transmitter SN.NO.
Calibration date setup anddisplay previous calibrationdate
Maintenance date setup and display previous maintenance
date
Default Change intofactory default setup
Save defaultsetup Savefactory default setup
Calibration setup
Zero pointcorrection setup
Sensor coefficient setup
Sensor coefficient calculation setup
Transmitteruniformity setup setup
Linearity correction setup Enable/disable
Linearity correction point setup
Linearity correction value setup
Subsection correction setup
Flow correction unitM/S, M3/H
setupflow correction points 1
setupflow correction coefficient 1
setupflow correction points 2
setupflow correction coefficient 2
setupflow correction points 3
setupflow correction coefficient 3
setupflow correction points 4
setupflow correction coefficient 4
setupflow correction points 5
setupflow correction coefficient 5
Subsection correction setup Enable/disable
Sensor Signal
SIG1 Signal 1
SGND Signal Ground
SIG2 Signal 2
EXT+ Exciting Current+
EXT- Exciting Current-
Analog Current Output
Ivee 24V DC Power Supply for 4-20mA Output
IOUT+ Analog Current Output
ICOM- Analog Current Output Ground
Pulse/frequencyOutput POUT+ Frequency (Pulse) Output+
PCOM- Frequency (Pulse) Ground
Digital Signal Output DOUT+ Digital Signal Output+
DOUT- Digital Signal Ground
RS485 (Function Optional) TRX+ Communication RS485+
TRX- Communication RS485-
System datesetup setup anddisplay date
System timesetup setup anddisplay time
Sensor manufacturedate setup anddisplay manufacturedate
Sensor SN. NO. setup anddisplay product SN.NO.
Meter manufacturedate setup and display transmitter manufacture
date
Meter SN.NO. setup anddisplay transmitter SN.NO.
Calibration date setup anddisplay previous calibrationdate
Maintenance date setup and display previous maintenance
date
Default Change intofactory default setup
Save defaultsetup Savefactory default setup
Calibration setup
Zero pointcorrection setup
Sensor coefficient setup
Sensor coefficient calculation setup
Transmitteruniformity setup setup
Linearity correction setup Enable/disable
Linearity correction point setup
Linearity correction value setup
Subsection correction setup
Flow correction unitM/S, M3/H
setupflow correction points 1
setupflow correction coefficient 1
setupflow correction points 2
setupflow correction coefficient 2
setupflow correction points 3
setupflow correction coefficient 3
setupflow correction points 4
setupflow correction coefficient 4
setupflow correction points 5
setupflow correction coefficient 5
Subsection correction setup Enable/disable
Sensor Signal
SIG1 Signal 1
SGND Signal Ground
SIG2 Signal 2
EXT+ Exciting Current+
EXT- Exciting Current-
Analog Current Output
Ivee 24V DC Power Supply for 4-20mA Output
IOUT+ Analog Current Output
ICOM- Analog Current Output Ground
Pulse/frequencyOutput POUT+ Frequency (Pulse) Output+
PCOM- Frequency (Pulse) Ground
Digital Signal Output DOUT+ Digital Signal Output+
DOUT- Digital Signal Ground
RS485 (Function Optional) TRX+ Communication RS485+
TRX- Communication RS485-
Sensor Signal
SIG1 Signal 1
SGND Signal Ground
SIG2 Signal 2
EXT+ Exciting Current+
EXT- Exciting Current-
Analog Current Output
Ivee 24V DC Power Supply for 4-20mA Output
IOUT+ Analog Current Output
ICOM- Analog Current Output Ground
Pulse/frequency Output POUT+ Frequency (Pulse) Output+
PCOM- Frequency (Pulse) Ground
Digital Signal Output DOUT+ Digital Signal Output+
DOUT- Digital Signal Ground
RS485 (Function Optional) TRX+ Communication RS485+
TRX- Communication RS485-
Sensor Signal
SIG1 Signal 1
SGND Signal Ground
SIG2 Signal 2
EXT+ Exciting Current+
EXT- Exciting Current-
Analog Current Output
Ivee 24V DC Power Supply for 4-20mA Output
IOUT+ Analog Current Output
ICOM- Analog Current Output Ground
Pulse/frequency Output POUT+ Frequency (Pulse) Output+
PCOM- Frequency (Pulse) Ground
Digital Signal Output DOUT+ Digital Signal Output+
DOUT- Digital Signal Ground
RS485 (Function Optional) TRX+ Communication RS485+
TRX- Communication RS485-
Z100E series eng version magnetic flow meter datasheet
Z100E series eng version magnetic flow meter datasheet

Z100E series eng version magnetic flow meter datasheet

  • 1.
  • 4.
  • 6.
    Compact magnetic flowmeter Remote magnetic flow meter
  • 12.
    Language Chinese,English Defaultsetup Defaultintofactory setup Save defaultsetup Savefactory default setup LCDdisplay contract setup LCDbacklight function Open /close Backlight automaticclose 1.5.10.30.60minutes for option Deletelogger Deleterecords setup forward flowvolume setup setup reverse flow volume setup Clear sum key Deletevolume flow Display password Displaylevel 1,2,3password Password setup setup level1,2,3password Communication Modbus Address Bit rate Parity bitsetup Output setup Pulseoutputtype Pulse, frequency Pulsepolarity Positive/negative Scaledpulse unit L,setup tablebetween 0.001L~10000.000L Pulsewidth Automatic/manual Frequency output range 0~10000HZ Digital output Flow direction,upper/lowflow limit alarm Currentoutputadjustment setup Diagnosis test 4-20mAoutputtest Simulatedcurrent output Flow velocity test Simulate full pipe situation (output frequency,pulse and current) Pulseoutputtest Simulatepulseoutput Frequency,outputtest Simulatefrequency output Logger inquiry Stop and startlogger Record recent 50times of stop and starting flow Monthly volumeflowrecord Recordtheflow ineach month Daily volumeflow record Recordtheflow ineach day First levelmenu Second levelmenu Third level menu Parametersetup Measurement pipesize 3mm-3000mm Damping time 0-99seconds Flow rateunit L/h, L/m, L/s,m3/h,m3/m,m3/s, t/h, t/m,t/s, kg/h, kg/m, kg/s Decimalpoints Automatic; manual (up to 3 decimal points can be setup) Volumeflowunit 0.001m3,0.01m3,0.1m3,1m3,0.001L,0.01L,0.1L, 1L,1T,1KG Flow range setup Excitation frequency 1/2, 1/4, 1/8, 1/16, 1/32 power frequency Excitation current 0,20%,40%,55%,70%,80%,90%,100% (250mA) Liquiddensity setup Functionsetup Flow direction Forward and reverse Reverse flow measure Enable/disable Reverse flow output Enable/disable Flow cutoff Enable/disable Low flowcutoff setup Stabilizeflow function Enable/disable Stabilizeflow coefficient setup Alarmallowance Enable/disable Systemexciting alarm Enable/disable Empty pipealarm Enable/disable Empty pipealarm trigger setup thevalue totrig empty pipealarm Empty pipemode 0 , 1 Upperflow limitalarm Enable/disable Upperflow alarm value setup Low flowlimit alarm Enable/disable Low flowalarm value setup Batterpower alarm Enable/disable Batterpower display Display lithiumbattery percent
  • 13.
    Language Chinese,English Defaultsetup Defaultintofactory setup Save defaultsetup Savefactory default setup LCDdisplay contract setup LCDbacklight function Open /close Backlight automaticclose 1.5.10.30.60minutes for option Deletelogger Deleterecords setup forward flowvolume setup setup reverse flow volume setup Clear sum key Deletevolume flow Display password Displaylevel 1,2,3password Password setup setup level1,2,3password Communication Modbus Address Bit rate Parity bitsetup Output setup Pulseoutputtype Pulse, frequency Pulsepolarity Positive/negative Scaledpulse unit L,setup tablebetween 0.001L~10000.000L Pulsewidth Automatic/manual Frequency output range 0~10000HZ Digital output Flow direction,upper/lowflow limit alarm Currentoutputadjustment setup Diagnosis test 4-20mAoutputtest Simulatedcurrent output Flow velocity test Simulate full pipe situation (output frequency,pulse and current) Pulseoutputtest Simulatepulseoutput Frequency,outputtest Simulatefrequency output Logger inquiry Stop and startlogger Record recent 50times of stop and starting flow Monthly volumeflowrecord Recordtheflow ineach month Daily volumeflow record Recordtheflow ineach day First levelmenu Second levelmenu Third level menu Parametersetup Measurement pipesize 3mm-3000mm Damping time 0-99seconds Flow rateunit L/h, L/m, L/s,m3/h,m3/m,m3/s, t/h, t/m,t/s, kg/h, kg/m, kg/s Decimalpoints Automatic; manual (up to 3 decimal points can be setup) Volumeflowunit 0.001m3,0.01m3,0.1m3,1m3,0.001L,0.01L,0.1L, 1L,1T,1KG Flow range setup Excitation frequency 1/2, 1/4, 1/8, 1/16, 1/32 power frequency Excitation current 0,20%,40%,55%,70%,80%,90%,100% (250mA) Liquiddensity setup Functionsetup Flow direction Forward and reverse Reverse flow measure Enable/disable Reverse flow output Enable/disable Flow cutoff Enable/disable Low flowcutoff setup Stabilizeflow function Enable/disable Stabilizeflow coefficient setup Alarmallowance Enable/disable Systemexciting alarm Enable/disable Empty pipealarm Enable/disable Empty pipealarm trigger setup thevalue totrig empty pipealarm Empty pipemode 0 , 1 Upperflow limitalarm Enable/disable Upperflow alarm value setup Low flowlimit alarm Enable/disable Low flowalarm value setup Batterpower alarm Enable/disable Batterpower display Display lithiumbattery percent
  • 14.
    System datesetup setupanddisplay date System timesetup setup anddisplay time Sensor manufacturedate setup anddisplay manufacturedate Sensor SN. NO. setup anddisplay product SN.NO. Meter manufacturedate setup and display transmitter manufacture date Meter SN.NO. setup anddisplay transmitter SN.NO. Calibration date setup anddisplay previous calibrationdate Maintenance date setup and display previous maintenance date Default Change intofactory default setup Save defaultsetup Savefactory default setup Calibration setup Zero pointcorrection setup Sensor coefficient setup Sensor coefficient calculation setup Transmitteruniformity setup setup Linearity correction setup Enable/disable Linearity correction point setup Linearity correction value setup Subsection correction setup Flow correction unitM/S, M3/H setupflow correction points 1 setupflow correction coefficient 1 setupflow correction points 2 setupflow correction coefficient 2 setupflow correction points 3 setupflow correction coefficient 3 setupflow correction points 4 setupflow correction coefficient 4 setupflow correction points 5 setupflow correction coefficient 5 Subsection correction setup Enable/disable Sensor Signal SIG1 Signal 1 SGND Signal Ground SIG2 Signal 2 EXT+ Exciting Current+ EXT- Exciting Current- Analog Current Output Ivee 24V DC Power Supply for 4-20mA Output IOUT+ Analog Current Output ICOM- Analog Current Output Ground Pulse/frequencyOutput POUT+ Frequency (Pulse) Output+ PCOM- Frequency (Pulse) Ground Digital Signal Output DOUT+ Digital Signal Output+ DOUT- Digital Signal Ground RS485 (Function Optional) TRX+ Communication RS485+ TRX- Communication RS485-
  • 15.
    System datesetup setupanddisplay date System timesetup setup anddisplay time Sensor manufacturedate setup anddisplay manufacturedate Sensor SN. NO. setup anddisplay product SN.NO. Meter manufacturedate setup and display transmitter manufacture date Meter SN.NO. setup anddisplay transmitter SN.NO. Calibration date setup anddisplay previous calibrationdate Maintenance date setup and display previous maintenance date Default Change intofactory default setup Save defaultsetup Savefactory default setup Calibration setup Zero pointcorrection setup Sensor coefficient setup Sensor coefficient calculation setup Transmitteruniformity setup setup Linearity correction setup Enable/disable Linearity correction point setup Linearity correction value setup Subsection correction setup Flow correction unitM/S, M3/H setupflow correction points 1 setupflow correction coefficient 1 setupflow correction points 2 setupflow correction coefficient 2 setupflow correction points 3 setupflow correction coefficient 3 setupflow correction points 4 setupflow correction coefficient 4 setupflow correction points 5 setupflow correction coefficient 5 Subsection correction setup Enable/disable Sensor Signal SIG1 Signal 1 SGND Signal Ground SIG2 Signal 2 EXT+ Exciting Current+ EXT- Exciting Current- Analog Current Output Ivee 24V DC Power Supply for 4-20mA Output IOUT+ Analog Current Output ICOM- Analog Current Output Ground Pulse/frequencyOutput POUT+ Frequency (Pulse) Output+ PCOM- Frequency (Pulse) Ground Digital Signal Output DOUT+ Digital Signal Output+ DOUT- Digital Signal Ground RS485 (Function Optional) TRX+ Communication RS485+ TRX- Communication RS485-
  • 16.
    Sensor Signal SIG1 Signal1 SGND Signal Ground SIG2 Signal 2 EXT+ Exciting Current+ EXT- Exciting Current- Analog Current Output Ivee 24V DC Power Supply for 4-20mA Output IOUT+ Analog Current Output ICOM- Analog Current Output Ground Pulse/frequency Output POUT+ Frequency (Pulse) Output+ PCOM- Frequency (Pulse) Ground Digital Signal Output DOUT+ Digital Signal Output+ DOUT- Digital Signal Ground RS485 (Function Optional) TRX+ Communication RS485+ TRX- Communication RS485-
  • 17.
    Sensor Signal SIG1 Signal1 SGND Signal Ground SIG2 Signal 2 EXT+ Exciting Current+ EXT- Exciting Current- Analog Current Output Ivee 24V DC Power Supply for 4-20mA Output IOUT+ Analog Current Output ICOM- Analog Current Output Ground Pulse/frequency Output POUT+ Frequency (Pulse) Output+ PCOM- Frequency (Pulse) Ground Digital Signal Output DOUT+ Digital Signal Output+ DOUT- Digital Signal Ground RS485 (Function Optional) TRX+ Communication RS485+ TRX- Communication RS485-