This document discusses recent developments in microcatheter technology, including new microcatheters from Asahi such as the Corsair Pro XS. It describes design features of various microcatheters like tapered shafts, braided coils, and tip configurations. New devices aim to improve trackability, torque response, and crossability. While microcatheter technology facilitates endovascular techniques, the document notes that further coordinated evolution is still needed between microcatheters and guidewires.
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Microcatheter Developments and Technologies
1. Prof Georgios Sianos, MD, PhD, FESC
Aristotle University
AHEPA University Hospital
Thessaloniki, Greece
A’ Cardiology Department
AHEPA University Hospital
Aristotle University
of Thessaloniki
MICROCATHETERS
NEW
DEVELOPMENTS
6. OVERVIEW
OTW catheters with dual-layer coil + braid shaft design
for superior flexibility & torque for delivery over 0.014"
wires
6
Turnpike
superior tracking
in tortuous vessels
Turnpike Spiral
rotational assistance,
trackability
Turnpike Gold
metallic tip for
enhanced advancement
Turnpike LP
low profile & flexible
for extreme tortuosity
15. CENTERCROSS AND MULTICROSS
MICRO 14
Jalal S et al, Novel CenterCross and MultiCross devices for the treatment of infrainguinal chronic
total occlusions: initial single-centre experience. EuroIntervention. 2016
16. HISTORY OF ASAHI CORONARY MICROCATHETERS
ASAHI Corsair Pro
• 2017
ASAHI Caravel
• 2016
ASAHI Corsair
• 2011
ASAHI Corsair Pro XS
• 2020
19
17. 0.88~0.95mm
(2.6~2.9 Fr)
Tip portion (A) Proximal portion (C)Distal portion (B)
A B C
Thinner new SHINKA-Shaft – Cross section
0.38 mm
(0.015 inch)
0.38 ~0.48mm
(0.015~0.019 inch)
0.48mm
(0.018 inch)
0.44 mm
(1.3 Fr)
0.71~0.88mm
(2.1~2.6 Fr)
14 wires
braiding
20
18. ASAHI CORSAIR PRO XS - KEY FEATURES
Product Tip Entry profile Distal shaft OD Proximal shaft OD
ASAHI Corsair Pro 0.42mm (1.3Fr) 0.87mm (2.6 Fr) 0.93mm (2.8 Fr)
ASAHI Caravel 0.48mm (1.4Fr) 0.62mm (1.9 Fr) 0.85mm (2.6 Fr)
ASAHI Corsair Pro XS 0.44mm (1.3Fr) 0.71mm (2.1 Fr) 0.95mm (2.9 Fr)
Proximal shaft OD
O.D. 0.95 mm (2.9 Fr)
New SHINKA-Shaft
New SHINKA-Shaft
Innovative rotational shaft design increases
device performance in retrograde approach
Distal shaft OD
O.D. 0.71 mm (2.1 Fr)
21
19. Improved distal tip
• Lower profile and thin tip portion
• Shorter welding between the tip and shaft
• Distal portion with smooth taper
ASAHI Corsair Pro XS
Product Tip Length Tip OD
ASAHI Corsair Pro XS 6.5 mm 0.44mm (1.3 Fr)
High crossability within complex lesions due to the new design of the tip portion.
22
21. Torqueable
Wire support
Pushability and crossability
Torqueable
Flexibility and trackability
Low profile
ASAHI Microcatheters
24
Non torqueable
Low profile
Flexibility and trackability
32. GUIDEZILLA GUIDE EXTENSION
1x1 Braid provides extra back-up support without
over-straightening the vessel. Soft, flexible, and
atraumatic tip designed to minimize the risk of
complications to the vessel
33. • Coaxial alignment & backup support
• Distal device delivery
• Selective delivery of contrast
Available sizes:
• 5F, 5.5F, 6F, 7F & 8F
• 6F XL (40cm guide extension)
“Child” guide extension delivered through
“mother” guide with rapid exchange construction
Half-pipe to align devices through collar transition
37. Proximal shaft change
Current : Round shaft diameter 0.43 (mm)
New : Oval shaft 0.41 mm x 0.50 mm
Distal flex zone
Low density coiling at the distal zone
Transition improvement
Current : Some variation in the steepness of transition
between proximal and distal shaft
New : Completely perfect gradual controlled transition
Proximal end stop color change
Current : White proximal end stop
New : Yellow proximal end stop
IV Generation
39. MICROCATHETER CLASSIFICATION
A B C D E
"Big" "Small" Angulated Dual Lumen Plaque modification
Vammou et al. Expert Rev Med Devices. 2019 Apr;16(4):267-273
CORSAIR PRO XS
NX3
40. CONCLUSION
There is a great evolution in microcatheter
technology both for antegrade and retrograde
approach greatly facilitating the efficacy and
safety of these techniques.
Still microcatheter evolution lacks guidewire
evolution and needs further improvement.
41. CONCLUSION
There is a great evolution in microcatheter
technology both for antegrade and retrograde
approach greatly facilitating the efficacy and
safety of these techniques.
Still microcatheter evolution lacks guidewire
evolution and needs further improvement.