LABORATORY DESIGN, SAFETY AND MANAGEMENT
1
 Designing laboratory – equipment
 Heating devices
 Ovens
 Hot plates
 Heating mantles
 Oil, salt, or sand baths
 Hot air baths and tube furnaces
 Heat guns
 Air conditioning systems
 Types of air conditioning
 Ventilation
 Laboratory fume hoods
 References
CONTENTS:
2
 The process of constructing a new laboratory revolves around four general
activities: planning, design, construction, and moving.
 Specific activities vary within each phase of the process, and each stage.
Scope
 The primary objective in laboratory design is to provide a safe environment
for laboratory personnel to conduct their work.
 Therefore, all health and safety hazards must be identified and carefully
evaluated so that protective measures can be incorporated into the design.
Introduction
3
 The most common type of electrical equipment
 Used to supply the heat needed to effect a reaction or separation.
 The use of steam-heated devices rather than electrically heated devices is
generally preferred whenever temperatures of 100 °C or less are required.
 Because they don’t present shock or spark risks, they can be left unattended with
assurance that their temperature will never exceed 100 °C.
General precautions:
 Enclose the actual heating element in any heating device in a glass,
ceramic, or insulated metal case to prevent laboratory personnel from
accidentally touching the wire carrying the electric current. This
construction minimizes the risk of electric shock.
 Resistance devices used to heat oil baths should not contain bare wires.
 If any heating device becomes so worn or damaged that its heating element
is exposed, either discard the device or repair it before it is used again.
 Be aware that dry and concentrated residues can ignite when overheated in
stills, ovens, dryers, and other heating devices.
Heating Devices
4
 Electrically heated ovens to remove water or other solvents from
chemical samples and to dry laboratory glassware.
 Never use laboratory ovens to prepare food for human consumption.
 Purchase or construct laboratory ovens with their heating elements and their
temperature controls physically separated from their interior atmospheres.
 Small household ovens and similar heating devices should not be used in
laboratories.
 Vacuum drying ovens, discharge volatilized
substances inside the oven to form explosive
mixtures with the air.
 This hazard can be reduced by connecting
the oven vent directly to an exhaust system.
Ovens
5
 Hot plates solutions are to be heated to 100 °C or higher and the
inherently safer steam baths cannot be used as the source of heat.
 Use only hot plates that have completely enclosed heating elements in
laboratories.
 Although almost all laboratory hot plates currently sold meet this criterion,
many older ones pose an electrical spark hazard arising from either the on/off
switch located on the hot plate, the bimetallic thermostat used to regulate the
temperature, or both.
 Normally, these two spark sources are located in the lower part of the hot
plate in a region where any heavier-than-air and possibly flammable vapors
evolving from a boiling liquid on the hot plate would tend to accumulate.
Hot Plates
6
 Heating mantles heat round-bottom flasks, reaction kettles, and related
reaction vessels.
 These mantles enclose a heating element in layers of fiber glass cloth.
 As long as the fiberglass coating is not worn or broken and no water or other
chemicals are spilled into the mantle heating mantles pose minimal shock
hazard.
 Always use heating mantles with a variable autotransformer to control the
input voltage.
 Never plug them directly into a 110-V line. Higher voltages will cause a
mantle to overheat, melting the fiberglass insulation and exposing the bare
heating element.
Heating Mantles
7
 Electrically heated oil baths small or irregularly shaped vessels or to
maintain a constant temperature.
 For ≤ 200 °C, a saturated paraffin oil is often used;
 For ≥ 300 °C, a silicone oil should be used.
 Care must be taken with hot oil baths not to generate smoke or have the oil
burst into flames from overheating.
 When using oil, salt, or sand baths, take care not to spill water and other
volatile substances into the baths. Such an accident can splatter hot material
over a wide area and cause serious injuries.
Oil, Salt,or SandBaths
8
 Always monitor an oil bath by using a thermometer or other thermal sensing
device to ensure that its temperature does not exceed the flash point of the oil
being used.
 For the same reason, fit oil baths (left unattended ) with thermal-sensing
devices that turn off the electric power if the bath overheats.
 Molten salt baths, like hot oil baths, offer the advantages of good heat transfer,
commonly have a higher operating range (e.g., 200 to 425 °C), and may have a
high thermal stability (e.g., 540 °C).
 Care must be taken to keep salt baths dry, because they are hygroscopic, a
property that can cause hazardous popping and splattering if the absorbed water
vaporizes during heating.
9
 Hot air baths:-
useful heating devices,
electrically heated,
frequently used to heat small or irregularly shaped vessels.
constructed from metal, ceramic, or, less desirably, glass vessels
low heat capacity
• Tube furnaces are often used for high-temperature reactions under reduced
pressure. The proper choice of glassware or metal tubes and joints is required,
and the procedures should conform to safe practice with electrical equipment
and evacuated apparatus.
Hot Air Baths and Tube Furnaces
10
• Laboratory heat guns are constructed with a motor-driven fan that blows air
over an electrically heated filament.
• They are frequently used to dry glassware or to heat the upper parts of a
distillation apparatus during distillation of high-boiling point materials.
• Heat guns almost always pose a serious spark hazard.
• Never use them near open containers of flammable liquids, in environments
where appreciable concentrations of flammable vapors may be present, or in
laboratory chemical hoods used to remove flammable vapors.
HeatGuns
11
 Air-conditioning is a process of controlling the air temperature, relative
humidity, ventilation, air movement and air cleanliness of a given space in
order to provide the occupants with a comfortable indoor temperature.
 One of the costliest items in a laboratory construction project is the air
conditioning system.
 Labs need more air than almost any other type of facility and supplying that
air can be expensive.
 Two Major Types of Systems:
1.Constant Volume (CV)
2.Variable Air Volume (VAV)
Air conditioning systems
12
Constant Volume System
 It exhausts a constant amount (volume) of air through the lab fume hoods while
at the same time supplying a constant volume of air into the laboratory.
 The amount of air exhausted and supplied is constant.
Advantages:
 The constant volume system is simple (to design, install and maintain).
 The start-up and testing, installation, and maintenance can be performed by
conventional experienced HVAC contractors.
 Since all hoods will be in operation at all times and that they will be exhausting
at a constant rate it is likely that the equipment specified for a CV system is of
sufficient size to provide enough air for a safe, comfortable lab.
 The initial cost of a CV system is typically lower than a VAV system requires
more controls and more sophisticated equipment they tend to be more expensive.
13
Variable Air Volume Systems
 Variable Air Volume (VAV) systems vary the amount of air being exhausted and
supplied to a lab based upon the usage of the fume hoods.
 Systems utilize a sensing device built into the side wall of the hood to measure
the air volume drawn into the hood.
 As the volume increases the sensor activates and increases the exhaust. A
decrease in the air volume results in a reduction of the exhaust.
Advantages:
 Cost savings resulting from reduced energy consumption. Since the system
exhausts less conditioned air it requires less conditioned air with resulting lower
energy costs.
 Reduction in size of the HVAC equipment. distinct advantage if space is at a
premium. A subsequent reduction in equipment cost is also achieved.
14
When considering and selecting an air conditioning system, the designer must
evaluate the following factors:
 Building location, surrounding environment and external climate.
 Uses and functional requirements of the building
 Client’s budget, investment policy and expected quality of service
 System qualities– e.g. aesthetics, life, reliability and maintainability. However
for the present work unitary air conditioning system is selected.
 Energy consumption – for both economic and environment reasons
Air Conditioning System Selection
15
 It is required to promote and maintain laboratory safety and protection to life
and property.
 Items such as fume containment, worker safety, proper cleanliness through
pressure relationships, filtration, air changes per hour (ACH), point of fume
capture, temperature, and Relative humidity requirements are elements
necessary to design the ventilation system depending on the laboratory type.
 When working in a required clean environment such as when mixing
medications (pharmacy) or working with infectious bacteria (research), airflow
direction and air cleanliness are requirements.
Ventilation
16
 Laboratory fume hoods are a type of specialized local exhaust that is
designed to isolate and remove gases, vapours, and particles from laboratory
sources and limit exposure to the user.
 The location of the fume hood with respect to open windows, doorways, and
personnel traffic directly influences the containment ability of the hood.
 Fume hoods should not be located adjacent to a single means of access to an
exit.
 A fire hazard or chemical release incident, both of which may start in a fume
hood, can block an exit rendering it impassable.
 A fire or explosion in a fume hood located adjacent to a path of egress could
trap someone in the lab.
 Hoods should be located more than 10 feet from any door or doorway.
Laboratory Fume Hoods
17
 Architectural features and their recommended distance from fume hoods.
 Drying ovens or drying cabinets shall not be installed underneath fume hoods.
 The heat from drying ovens increases the risk of volatilization of chemicals
inside the fume hood, particularly chemical solvents and may increase the risk
of explosions or fires.
 In addition, heat transfer to the fume hood may affect the functionality of the
fume hood
Architectural Feature Minimum Distance from
Hood (feet)
Pedestrian Walkways 4
Opposite Laboratory Bench 5
Opposite Laboratory Fume Hood 10
Adjoining Laboratory Fume
Hood
3
Opposing Wall 7
Adjoining Wall 1
Doorway 10
19
• Laboratory Design Guidelines-University of North Carolina, USA
 Designing and planning of laboratories (2009)
 Laboratory Design and Construction Guidelines (2010)-University of South
Carolina, USA
 International Journal of Application or Innovation in Engineering &
Management (IJAIEM), Volume 2, Issue 9, September 2013 Page 287
 Forensic Laboratories: Handbook for Facility Planning, Design,
Construction, and Moving
References
20

More Related Content

What's hot

Basic design and organisation of diagnostic laboratory
Basic design and organisation of diagnostic laboratoryBasic design and organisation of diagnostic laboratory
Basic design and organisation of diagnostic laboratoryRohit Hari
 
Laboratory hazards, safety and contamination
Laboratory hazards, safety and contaminationLaboratory hazards, safety and contamination
Laboratory hazards, safety and contaminationNadjo Deen
 
Introduction to haematology laboratory
Introduction to haematology laboratoryIntroduction to haematology laboratory
Introduction to haematology laboratoryHussein Al-tameemi
 
Sample collection, Preservation and its Estimation
Sample collection, Preservation and its EstimationSample collection, Preservation and its Estimation
Sample collection, Preservation and its EstimationMD Abdul Haleem
 
Microbiology laboratory wastes disposal
Microbiology  laboratory  wastes disposalMicrobiology  laboratory  wastes disposal
Microbiology laboratory wastes disposalMicrobiology
 
Safe Use and Storage of Chemicals and Reagents
Safe Use and Storage of Chemicals and ReagentsSafe Use and Storage of Chemicals and Reagents
Safe Use and Storage of Chemicals and ReagentsTapeshwar Yadav
 
Responsibilities of Clinical Laboratory Scientist and Technicians
Responsibilities of Clinical Laboratory Scientist and TechniciansResponsibilities of Clinical Laboratory Scientist and Technicians
Responsibilities of Clinical Laboratory Scientist and TechniciansBest care Lab
 
Medical Laboratory Sciences
Medical Laboratory SciencesMedical Laboratory Sciences
Medical Laboratory SciencesTapeshwar Yadav
 
Automation in the Clinical Lab
Automation in the Clinical LabAutomation in the Clinical Lab
Automation in the Clinical LabAli Raza Ph.D
 
Laboratory accreditation
Laboratory accreditationLaboratory accreditation
Laboratory accreditationGift Sam
 
Clinical laboratory total quality management (TQM) system
Clinical laboratory total quality management (TQM) systemClinical laboratory total quality management (TQM) system
Clinical laboratory total quality management (TQM) systemTapeshwar Yadav
 
Laboratory management and safety good
Laboratory management and safety goodLaboratory management and safety good
Laboratory management and safety goodChala Dandessa
 
Laboratory Information Management System
Laboratory Information Management SystemLaboratory Information Management System
Laboratory Information Management SystemDr. Rajesh Bendre
 

What's hot (20)

Basic design and organisation of diagnostic laboratory
Basic design and organisation of diagnostic laboratoryBasic design and organisation of diagnostic laboratory
Basic design and organisation of diagnostic laboratory
 
Laboratory safety
Laboratory safetyLaboratory safety
Laboratory safety
 
Laboratory hazards, safety and contamination
Laboratory hazards, safety and contaminationLaboratory hazards, safety and contamination
Laboratory hazards, safety and contamination
 
Introduction to haematology laboratory
Introduction to haematology laboratoryIntroduction to haematology laboratory
Introduction to haematology laboratory
 
Laboratory safety
Laboratory safetyLaboratory safety
Laboratory safety
 
Sample collection, Preservation and its Estimation
Sample collection, Preservation and its EstimationSample collection, Preservation and its Estimation
Sample collection, Preservation and its Estimation
 
Biosafety
Biosafety Biosafety
Biosafety
 
Microbiology lab QC
Microbiology lab QCMicrobiology lab QC
Microbiology lab QC
 
Laboratory management
Laboratory managementLaboratory management
Laboratory management
 
Laboratory Centrifuge
Laboratory CentrifugeLaboratory Centrifuge
Laboratory Centrifuge
 
Microbiology laboratory wastes disposal
Microbiology  laboratory  wastes disposalMicrobiology  laboratory  wastes disposal
Microbiology laboratory wastes disposal
 
Safe Use and Storage of Chemicals and Reagents
Safe Use and Storage of Chemicals and ReagentsSafe Use and Storage of Chemicals and Reagents
Safe Use and Storage of Chemicals and Reagents
 
lab safety
 lab safety  lab safety
lab safety
 
Responsibilities of Clinical Laboratory Scientist and Technicians
Responsibilities of Clinical Laboratory Scientist and TechniciansResponsibilities of Clinical Laboratory Scientist and Technicians
Responsibilities of Clinical Laboratory Scientist and Technicians
 
Medical Laboratory Sciences
Medical Laboratory SciencesMedical Laboratory Sciences
Medical Laboratory Sciences
 
Automation in the Clinical Lab
Automation in the Clinical LabAutomation in the Clinical Lab
Automation in the Clinical Lab
 
Laboratory accreditation
Laboratory accreditationLaboratory accreditation
Laboratory accreditation
 
Clinical laboratory total quality management (TQM) system
Clinical laboratory total quality management (TQM) systemClinical laboratory total quality management (TQM) system
Clinical laboratory total quality management (TQM) system
 
Laboratory management and safety good
Laboratory management and safety goodLaboratory management and safety good
Laboratory management and safety good
 
Laboratory Information Management System
Laboratory Information Management SystemLaboratory Information Management System
Laboratory Information Management System
 

Similar to Laboratory design, safety and management

P (6) Autoclave and ovens.pdf
P (6) Autoclave and ovens.pdfP (6) Autoclave and ovens.pdf
P (6) Autoclave and ovens.pdfSaad49687
 
Airflow Industrial Process Ovens Drum Heating Ovens - Brochure
Airflow Industrial Process Ovens Drum Heating Ovens - BrochureAirflow Industrial Process Ovens Drum Heating Ovens - Brochure
Airflow Industrial Process Ovens Drum Heating Ovens - BrochureThorne & Derrick UK
 
Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...
Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...
Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...yogeshysumthane
 
abhi presentation autoclave.pptx
abhi presentation autoclave.pptxabhi presentation autoclave.pptx
abhi presentation autoclave.pptxabhijeetpadhi001
 
Retorting Technology - Operating Principle, Components of Retort, Types of Re...
Retorting Technology - Operating Principle, Components of Retort, Types of Re...Retorting Technology - Operating Principle, Components of Retort, Types of Re...
Retorting Technology - Operating Principle, Components of Retort, Types of Re...PMG Engineering Private Limited
 
Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...
Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...
Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...Thorne & Derrick UK
 
Autoclave the hit sterilisation
Autoclave the hit sterilisationAutoclave the hit sterilisation
Autoclave the hit sterilisationrenuka10496
 
MECHANICAL VENTILATION commercial kitchen.pptx
MECHANICAL VENTILATION commercial kitchen.pptxMECHANICAL VENTILATION commercial kitchen.pptx
MECHANICAL VENTILATION commercial kitchen.pptxFizaNaaz8
 
Tray drying-chamber
Tray drying-chamberTray drying-chamber
Tray drying-chamberacmasindia
 
4068 plant tissue culture lab design
4068 plant tissue culture lab design4068 plant tissue culture lab design
4068 plant tissue culture lab designSheena Prem
 

Similar to Laboratory design, safety and management (20)

P (6) Autoclave and ovens.pdf
P (6) Autoclave and ovens.pdfP (6) Autoclave and ovens.pdf
P (6) Autoclave and ovens.pdf
 
Presentation of industrial pharmacy 2
Presentation of industrial pharmacy 2Presentation of industrial pharmacy 2
Presentation of industrial pharmacy 2
 
Airflow Industrial Process Ovens Drum Heating Ovens - Brochure
Airflow Industrial Process Ovens Drum Heating Ovens - BrochureAirflow Industrial Process Ovens Drum Heating Ovens - Brochure
Airflow Industrial Process Ovens Drum Heating Ovens - Brochure
 
Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...
Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...
Dr. Yogesh Sumthane Assistant Professor College of Forestry Banda University ...
 
abhi presentation autoclave.pptx
abhi presentation autoclave.pptxabhi presentation autoclave.pptx
abhi presentation autoclave.pptx
 
MET 214 Module 6
MET 214 Module 6MET 214 Module 6
MET 214 Module 6
 
Retorting Technology - Operating Principle, Components of Retort, Types of Re...
Retorting Technology - Operating Principle, Components of Retort, Types of Re...Retorting Technology - Operating Principle, Components of Retort, Types of Re...
Retorting Technology - Operating Principle, Components of Retort, Types of Re...
 
Industrial oven
Industrial ovenIndustrial oven
Industrial oven
 
Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...
Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...
Rwc, Reliance Water Controls Caremix H3 - Deck Mounted 2 Hole TMV3 Thermostat...
 
amzzz.equipments.pdf
amzzz.equipments.pdfamzzz.equipments.pdf
amzzz.equipments.pdf
 
PMG Newsletter (Volume 2. Issue 14)- DIGITAL.pdf
PMG Newsletter (Volume 2. Issue 14)- DIGITAL.pdfPMG Newsletter (Volume 2. Issue 14)- DIGITAL.pdf
PMG Newsletter (Volume 2. Issue 14)- DIGITAL.pdf
 
Agnee E-catalouge Oven.
Agnee E-catalouge Oven.Agnee E-catalouge Oven.
Agnee E-catalouge Oven.
 
Agnee e catalouge oven.
Agnee e catalouge oven.Agnee e catalouge oven.
Agnee e catalouge oven.
 
Presentation of industrial pharmacy 2
Presentation  of industrial pharmacy 2Presentation  of industrial pharmacy 2
Presentation of industrial pharmacy 2
 
Autoclave the hit sterilisation
Autoclave the hit sterilisationAutoclave the hit sterilisation
Autoclave the hit sterilisation
 
electric boiler user boiler.pdf
electric boiler user boiler.pdfelectric boiler user boiler.pdf
electric boiler user boiler.pdf
 
MECHANICAL VENTILATION commercial kitchen.pptx
MECHANICAL VENTILATION commercial kitchen.pptxMECHANICAL VENTILATION commercial kitchen.pptx
MECHANICAL VENTILATION commercial kitchen.pptx
 
DW172
DW172DW172
DW172
 
Tray drying-chamber
Tray drying-chamberTray drying-chamber
Tray drying-chamber
 
4068 plant tissue culture lab design
4068 plant tissue culture lab design4068 plant tissue culture lab design
4068 plant tissue culture lab design
 

Recently uploaded

9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...
9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...
9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...Pooja Nehwal
 
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...Call Girls in Nagpur High Profile
 
Thane Escorts, (Pooja 09892124323), Thane Call Girls
Thane Escorts, (Pooja 09892124323), Thane Call GirlsThane Escorts, (Pooja 09892124323), Thane Call Girls
Thane Escorts, (Pooja 09892124323), Thane Call GirlsPooja Nehwal
 
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)kojalkojal131
 
VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...
VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...
VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...Call Girls in Nagpur High Profile
 
(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...ranjana rawat
 
Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...
Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...
Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...ranjana rawat
 
Call Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service NashikLow Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,Pooja Nehwal
 
Russian Call Girls Kolkata Chhaya 🤌 8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Chhaya 🤌  8250192130 🚀 Vip Call Girls KolkataRussian Call Girls Kolkata Chhaya 🤌  8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Chhaya 🤌 8250192130 🚀 Vip Call Girls Kolkataanamikaraghav4
 
High Profile Call Girls In Andheri 7738631006 Call girls in mumbai Mumbai ...
High Profile Call Girls In Andheri 7738631006 Call girls in mumbai  Mumbai ...High Profile Call Girls In Andheri 7738631006 Call girls in mumbai  Mumbai ...
High Profile Call Girls In Andheri 7738631006 Call girls in mumbai Mumbai ...Pooja Nehwal
 
Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...
Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...
Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...Pooja Nehwal
 
定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一
定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一
定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一zul5vf0pq
 
WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service - Bandra F...
WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service -  Bandra F...WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service -  Bandra F...
WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service - Bandra F...Pooja Nehwal
 
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Gaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service GayaGaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service Gayasrsj9000
 
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样qaffana
 
Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...
Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...
Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...nagunakhan
 

Recently uploaded (20)

9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...
9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...
9892124323 Pooja Nehwal Call Girls Services Call Girls service in Santacruz A...
 
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...
 
Thane Escorts, (Pooja 09892124323), Thane Call Girls
Thane Escorts, (Pooja 09892124323), Thane Call GirlsThane Escorts, (Pooja 09892124323), Thane Call Girls
Thane Escorts, (Pooja 09892124323), Thane Call Girls
 
🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate
🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate
🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate
 
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
 
VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...
VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...
VVIP Pune Call Girls Balaji Nagar (7001035870) Pune Escorts Nearby with Compl...
 
(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(ANIKA) Wanwadi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
 
Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...
Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...
Book Paid Lohegaon Call Girls Pune 8250192130Low Budget Full Independent High...
 
Call Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Bhavna Call 7001035870 Meet With Nagpur Escorts
 
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service NashikLow Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
 
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
Call Girls In Andheri East Call 9892124323 Book Hot And Sexy Girls,
 
Russian Call Girls Kolkata Chhaya 🤌 8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Chhaya 🤌  8250192130 🚀 Vip Call Girls KolkataRussian Call Girls Kolkata Chhaya 🤌  8250192130 🚀 Vip Call Girls Kolkata
Russian Call Girls Kolkata Chhaya 🤌 8250192130 🚀 Vip Call Girls Kolkata
 
High Profile Call Girls In Andheri 7738631006 Call girls in mumbai Mumbai ...
High Profile Call Girls In Andheri 7738631006 Call girls in mumbai  Mumbai ...High Profile Call Girls In Andheri 7738631006 Call girls in mumbai  Mumbai ...
High Profile Call Girls In Andheri 7738631006 Call girls in mumbai Mumbai ...
 
Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...
Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...
Call Girls in Thane 9892124323, Vashi cAll girls Serivces Juhu Escorts, powai...
 
定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一
定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一
定制加拿大滑铁卢大学毕业证(Waterloo毕业证书)成绩单(文凭)原版一比一
 
WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service - Bandra F...
WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service -  Bandra F...WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service -  Bandra F...
WhatsApp 9892124323 ✓Call Girls In Khar ( Mumbai ) secure service - Bandra F...
 
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
(ZARA) Call Girls Jejuri ( 7001035870 ) HI-Fi Pune Escorts Service
 
Gaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service GayaGaya Call Girls #9907093804 Contact Number Escorts Service Gaya
Gaya Call Girls #9907093804 Contact Number Escorts Service Gaya
 
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
 
Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...
Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...
Russian Escorts in lucknow 💗 9719455033 💥 Lovely Lasses: Radiant Beauties Shi...
 

Laboratory design, safety and management

  • 1. P.SAIKUMAR 1008-16-884-005 M.PHARMACY PHARMACEUTICAL CHEMISTRY UNIVERSITY COLLEGE OF TECHNOLOGY OSMANIA UNIVERSITY LABORATORY DESIGN, SAFETY AND MANAGEMENT 1
  • 2.  Designing laboratory – equipment  Heating devices  Ovens  Hot plates  Heating mantles  Oil, salt, or sand baths  Hot air baths and tube furnaces  Heat guns  Air conditioning systems  Types of air conditioning  Ventilation  Laboratory fume hoods  References CONTENTS: 2
  • 3.  The process of constructing a new laboratory revolves around four general activities: planning, design, construction, and moving.  Specific activities vary within each phase of the process, and each stage. Scope  The primary objective in laboratory design is to provide a safe environment for laboratory personnel to conduct their work.  Therefore, all health and safety hazards must be identified and carefully evaluated so that protective measures can be incorporated into the design. Introduction 3
  • 4.  The most common type of electrical equipment  Used to supply the heat needed to effect a reaction or separation.  The use of steam-heated devices rather than electrically heated devices is generally preferred whenever temperatures of 100 °C or less are required.  Because they don’t present shock or spark risks, they can be left unattended with assurance that their temperature will never exceed 100 °C. General precautions:  Enclose the actual heating element in any heating device in a glass, ceramic, or insulated metal case to prevent laboratory personnel from accidentally touching the wire carrying the electric current. This construction minimizes the risk of electric shock.  Resistance devices used to heat oil baths should not contain bare wires.  If any heating device becomes so worn or damaged that its heating element is exposed, either discard the device or repair it before it is used again.  Be aware that dry and concentrated residues can ignite when overheated in stills, ovens, dryers, and other heating devices. Heating Devices 4
  • 5.  Electrically heated ovens to remove water or other solvents from chemical samples and to dry laboratory glassware.  Never use laboratory ovens to prepare food for human consumption.  Purchase or construct laboratory ovens with their heating elements and their temperature controls physically separated from their interior atmospheres.  Small household ovens and similar heating devices should not be used in laboratories.  Vacuum drying ovens, discharge volatilized substances inside the oven to form explosive mixtures with the air.  This hazard can be reduced by connecting the oven vent directly to an exhaust system. Ovens 5
  • 6.  Hot plates solutions are to be heated to 100 °C or higher and the inherently safer steam baths cannot be used as the source of heat.  Use only hot plates that have completely enclosed heating elements in laboratories.  Although almost all laboratory hot plates currently sold meet this criterion, many older ones pose an electrical spark hazard arising from either the on/off switch located on the hot plate, the bimetallic thermostat used to regulate the temperature, or both.  Normally, these two spark sources are located in the lower part of the hot plate in a region where any heavier-than-air and possibly flammable vapors evolving from a boiling liquid on the hot plate would tend to accumulate. Hot Plates 6
  • 7.  Heating mantles heat round-bottom flasks, reaction kettles, and related reaction vessels.  These mantles enclose a heating element in layers of fiber glass cloth.  As long as the fiberglass coating is not worn or broken and no water or other chemicals are spilled into the mantle heating mantles pose minimal shock hazard.  Always use heating mantles with a variable autotransformer to control the input voltage.  Never plug them directly into a 110-V line. Higher voltages will cause a mantle to overheat, melting the fiberglass insulation and exposing the bare heating element. Heating Mantles 7
  • 8.  Electrically heated oil baths small or irregularly shaped vessels or to maintain a constant temperature.  For ≤ 200 °C, a saturated paraffin oil is often used;  For ≥ 300 °C, a silicone oil should be used.  Care must be taken with hot oil baths not to generate smoke or have the oil burst into flames from overheating.  When using oil, salt, or sand baths, take care not to spill water and other volatile substances into the baths. Such an accident can splatter hot material over a wide area and cause serious injuries. Oil, Salt,or SandBaths 8
  • 9.  Always monitor an oil bath by using a thermometer or other thermal sensing device to ensure that its temperature does not exceed the flash point of the oil being used.  For the same reason, fit oil baths (left unattended ) with thermal-sensing devices that turn off the electric power if the bath overheats.  Molten salt baths, like hot oil baths, offer the advantages of good heat transfer, commonly have a higher operating range (e.g., 200 to 425 °C), and may have a high thermal stability (e.g., 540 °C).  Care must be taken to keep salt baths dry, because they are hygroscopic, a property that can cause hazardous popping and splattering if the absorbed water vaporizes during heating. 9
  • 10.  Hot air baths:- useful heating devices, electrically heated, frequently used to heat small or irregularly shaped vessels. constructed from metal, ceramic, or, less desirably, glass vessels low heat capacity • Tube furnaces are often used for high-temperature reactions under reduced pressure. The proper choice of glassware or metal tubes and joints is required, and the procedures should conform to safe practice with electrical equipment and evacuated apparatus. Hot Air Baths and Tube Furnaces 10
  • 11. • Laboratory heat guns are constructed with a motor-driven fan that blows air over an electrically heated filament. • They are frequently used to dry glassware or to heat the upper parts of a distillation apparatus during distillation of high-boiling point materials. • Heat guns almost always pose a serious spark hazard. • Never use them near open containers of flammable liquids, in environments where appreciable concentrations of flammable vapors may be present, or in laboratory chemical hoods used to remove flammable vapors. HeatGuns 11
  • 12.  Air-conditioning is a process of controlling the air temperature, relative humidity, ventilation, air movement and air cleanliness of a given space in order to provide the occupants with a comfortable indoor temperature.  One of the costliest items in a laboratory construction project is the air conditioning system.  Labs need more air than almost any other type of facility and supplying that air can be expensive.  Two Major Types of Systems: 1.Constant Volume (CV) 2.Variable Air Volume (VAV) Air conditioning systems 12
  • 13. Constant Volume System  It exhausts a constant amount (volume) of air through the lab fume hoods while at the same time supplying a constant volume of air into the laboratory.  The amount of air exhausted and supplied is constant. Advantages:  The constant volume system is simple (to design, install and maintain).  The start-up and testing, installation, and maintenance can be performed by conventional experienced HVAC contractors.  Since all hoods will be in operation at all times and that they will be exhausting at a constant rate it is likely that the equipment specified for a CV system is of sufficient size to provide enough air for a safe, comfortable lab.  The initial cost of a CV system is typically lower than a VAV system requires more controls and more sophisticated equipment they tend to be more expensive. 13
  • 14. Variable Air Volume Systems  Variable Air Volume (VAV) systems vary the amount of air being exhausted and supplied to a lab based upon the usage of the fume hoods.  Systems utilize a sensing device built into the side wall of the hood to measure the air volume drawn into the hood.  As the volume increases the sensor activates and increases the exhaust. A decrease in the air volume results in a reduction of the exhaust. Advantages:  Cost savings resulting from reduced energy consumption. Since the system exhausts less conditioned air it requires less conditioned air with resulting lower energy costs.  Reduction in size of the HVAC equipment. distinct advantage if space is at a premium. A subsequent reduction in equipment cost is also achieved. 14
  • 15. When considering and selecting an air conditioning system, the designer must evaluate the following factors:  Building location, surrounding environment and external climate.  Uses and functional requirements of the building  Client’s budget, investment policy and expected quality of service  System qualities– e.g. aesthetics, life, reliability and maintainability. However for the present work unitary air conditioning system is selected.  Energy consumption – for both economic and environment reasons Air Conditioning System Selection 15
  • 16.  It is required to promote and maintain laboratory safety and protection to life and property.  Items such as fume containment, worker safety, proper cleanliness through pressure relationships, filtration, air changes per hour (ACH), point of fume capture, temperature, and Relative humidity requirements are elements necessary to design the ventilation system depending on the laboratory type.  When working in a required clean environment such as when mixing medications (pharmacy) or working with infectious bacteria (research), airflow direction and air cleanliness are requirements. Ventilation 16
  • 17.  Laboratory fume hoods are a type of specialized local exhaust that is designed to isolate and remove gases, vapours, and particles from laboratory sources and limit exposure to the user.  The location of the fume hood with respect to open windows, doorways, and personnel traffic directly influences the containment ability of the hood.  Fume hoods should not be located adjacent to a single means of access to an exit.  A fire hazard or chemical release incident, both of which may start in a fume hood, can block an exit rendering it impassable.  A fire or explosion in a fume hood located adjacent to a path of egress could trap someone in the lab.  Hoods should be located more than 10 feet from any door or doorway. Laboratory Fume Hoods 17
  • 18.  Architectural features and their recommended distance from fume hoods.  Drying ovens or drying cabinets shall not be installed underneath fume hoods.  The heat from drying ovens increases the risk of volatilization of chemicals inside the fume hood, particularly chemical solvents and may increase the risk of explosions or fires.  In addition, heat transfer to the fume hood may affect the functionality of the fume hood Architectural Feature Minimum Distance from Hood (feet) Pedestrian Walkways 4 Opposite Laboratory Bench 5 Opposite Laboratory Fume Hood 10 Adjoining Laboratory Fume Hood 3 Opposing Wall 7 Adjoining Wall 1 Doorway 10 19
  • 19. • Laboratory Design Guidelines-University of North Carolina, USA  Designing and planning of laboratories (2009)  Laboratory Design and Construction Guidelines (2010)-University of South Carolina, USA  International Journal of Application or Innovation in Engineering & Management (IJAIEM), Volume 2, Issue 9, September 2013 Page 287  Forensic Laboratories: Handbook for Facility Planning, Design, Construction, and Moving References 20