SlideShare a Scribd company logo
1 of 41
Process validation of Sterilization 
& Water Process Systems
STERILIZATION VALIDATION 
• Sterile products have several unique properties such as 
1. Free from micro organisms 
2. Free from pyrogens 
3. Free from particulates 
4. High standards of purity and quality 
References: [1,2,3, & 4]
Methods of sterilization of products 
HEAT:- 
1. Moist heat-auto clave 
2. Dry heat-hot air oven 
GAS:- 
1. ethylene oxide 
2. Peracetic acid 
3. Hydrogen peroxide(vapor phase) 
4. Chlorine di oxide 
References: [5& 6]
Radiation 
1. Gamma rays 
2. Beta rays 
3. Ultraviolet 
4. Microwave 
References: [5& 6]
Validation of steam sterilization cycles 
Qualification and calibration 
1. Mechanically checking ,upgrading, and qualifying the 
sterilizer unit
Selection and calibration of thermocouples 
• Cu constantan wires coated with teflon are a popular choice as 
thermocouple monitors 
• Accuracy of thermocouples should be ±0.5°C. Temperature accuracy is 
especially important in steam sterilization validation. 
• Thermocouple accuracy is determined using NATIONAL BUREAU OF 
STANDARDS (NBS) traceable constant temperature calibration 
instruments. 
• Thermocouples should be calibrated before and after validation 
experiment at 2 temperatures i.e. 0C & 125 C . 
• New thermocouple-recording devices are capable of automatically 
correcting temperature 
• Any thermocouple that senses a temperature of more than 0.5 C away 
from the calibration temperature bath should be discarded 
• Temperature recorders should be capable of printing temperature data in 
0.1°C increments 
References: [8 & 9]
Selection & calibration of BI 
• The organism most resistant to steam heat is the 
bacterial spore B. stearothermophilus. This bacterial 
spore is commonly used BI’s in validating steam 
sterilization cycles. 
• Spore trips or spore suspensions are used in the 
validation studies. the no. of mo’s per ml of suspension 
must be as accurately known as D value. 
• Precautions should be taken to use proper storage 
conditions for B. stearothermophilus BIs .storing in the 
freezer provides a more stable resistance profile for the 
shelf life of the indicator. 
References: [7 ]
Heat distribution studies 
It include 2 phases 
1. Heat distribution in any empty autoclave chamber. 
2. Heat distribution in a loaded autoclave chamber. 
a. 10-20 thermocouples should be used/cycle. thermocouples 
should be secured inside the chamber. 
b. The trips where the wires are soldered should not make contact 
with the autoclave interior walls or any metal surface. 
c. 1 end of thermocouple should remain in an ice bath and high 
temperature oil bath during each cycle for reference when the 
temp monitoring equipement has the capability for electronically 
compensating each temp measurement against an internal 
reference. 
d. Heat distribution studies following the initial study may employ 
fewer thermocouples as the cool spot in the chamber & in the 
load is identified. 
e. The difference in temp b/n the coolest spot $ the mean chamber 
temperature should not be greater than  2.5C. 
References: [8 ]
Heat penetration studies 
• This is the most critical component of the entire validation 
process 
• Thermocouples will be placed both inside and outside the 
container at the coolspot location(s) in the steam exhaust 
line and in constant temperature baths outside the 
chamber 
• The sterilization cycle design must be based on the heating 
charecteristics of the load and containers located in the 
slowest heating zone of the load. 
• The variation in the rate of heating of the slowest heating 
zone must be known, so this variation must be determined 
under fully loaded conditions 
• The effect of load to load variation on the time-temperature 
profile must also be determined. 
• Then the statistically worst case conditions should be used 
in the final sterilization process design
• The final step in steam sterilization process the 
establishment of a monitoring program to ensure 
that the validated cycle remains essentially 
unchanged in the future. 
• Cycle monitoring usually involves the use of 
thermocouples to measure heat penetration at the 
cool spot location. 
• Any changes in the load size, load configuration, or 
container charecteristics (volume, geometry etc) 
must be accompanied by repeat validation studies to 
prove that the cool spot location has not changed 
References: [10 ]
Validation of dry-heat sterilization 
cycles 
1. Batch oven validation 
• Air balance determination in an empty oven 
data are obtained on the flow rates of both 
intake and exhaust air. air should be balanced 
so that positive pressure is exerted to the 
non sterile side when the door is opend and 
air velocity across and up and down the 
opening of the door is ±50 FPM of the 
average velocity
• Heat distribution of an empty chamber 
thermocouples should be situated according to a 
specific predetermined pattern. Repeatability of 
temp attainment and identification of cold spot 
can be achieved if the temp range is ±15°C at all 
monitored locations. 
• Heat penetration studies. These studies should be 
designed to determine the location of slowest 
heating point within a commodity at various 
locations of test load in sterilizer. 
• Mechanical repeatability. during all these studies 
mechanical repeatability in terms of air velocity, 
temp consistency, reliability and sensitivity of all 
the oven and instrumental controls must be 
verified. 
References: [11 ]
2.Tunnel sterilizer validation 
 Air balance determination 
• Proper air balance is more critical to a tunnel sterile 
process than a batch oven process .since the items being 
sterilized are exposed to a different air systems(eg:-heating 
zone $ cooling zone).in the absence of a critical balance of 
air dynamics, either the items will not be cooled sufficiently 
once they exit the tunnel or they will be cooled too quickly. 
causing the glass to shatter and contaminate the entire 
tunnel area with particles. 
• The major problem in validating tunnel sterilizers is the 
control of particules. not only are items exposed to great 
extreams in temp, but also the conveyer belt is a natural 
source of particulates because metal is moving against 
metal. 
• Air must be particulate-free as it enters the tunnel area; 
therefore, all high efficiency particulate air(HEPA)filters in 
the tunnel must be tested and certified prior to validation 
studies.
Heat distribution studies 
• Thermocouples used in tunnel sterilizer validation 
must be sufficiently durable to withstand the 
extremely high( ≥ 300 c)temperatures in the 
heating zone area of the tunnel heat-distribution 
studies should determine where the cold spots 
are located as a function of the width of the belt 
$ height of the tunnel chamber. trays or tracks of 
ampules are vials should run through the tunnel 
• Bottle-mapping studies may also be conducted 
during this phase. the purpose of these studies is 
to determine possible locations inside the 
container that are most difficult to heat. 
References: [12 ]
Heat penetration studies 
• For testing of the tunnel sterilization, heat-penetration 
studies must be completed in order to identify the coolest 
container in the entire load. Results of heat-distribution 
studies should aid in the predicting where the coolest 
location with in the load should be. Thermocouples should 
be diposited at or near the coolest point inside the 
container from bottle-mapping studies. 
• The containers inner surface should be in contact with the 
thermocouple tip because the objective is to sterilize the 
inner walls of the container, as well as the inner space. 
• Every loading should be done using 10-20 thermocouples 
distributed through out the load. 
References: [ 9 ]
 Mechanical repeatability 
• Air velocity, air particulates, temp consistency 
and reliability of all the tunnel controls(heat 
zone temperatures, belt speed, and blower 
functions)must be proved during the physical 
validation studies. 
References: [ 9 ]
3.Biological process validation of dry heat 
sterilization cycles 
• mo’s known to be most resistant to dry heat must be used to 
prove the ability of dry heat cycle to destroy them at the 
coolest location in load. the dry heat process is claimed to 
produce both sterile and pyrogen-free commodities, validation 
studies must be done using both mo’s $ microbial endotoxins. 
• Biological validation of dry heat cycles should be based on the 
destruction of endotoxin rather than on the destruction of 
mo’s because of the enormous dry heat resistance of 
endotoxin compared to mo’s. 
• With both mo’s $ endotoxin challenges, the cool spot 
identified in heat distribution $ the heat penetration studies 
will be the logical location to run the microbial challenge tests. 
containers inoculated with the microbial cells or endotoxin will 
be situated adjacent to identical containers into which 
thermocouples are secured to monitor temp. References: [ 9 ]
Step by step sequence in the microbial validation of a dry 
heat process for sterilizing and depyrogenating large volume 
glass containers by wegel $ akers et al 
1. Place spore carrier in approximately 12 glass bottles located at the 
coolest area of the oven. bottles adjucent to the inoculated bottles 
should contain thermocouples for the monitoring purposes . 
2. Run a complete cycle using the desired loading pattern for future 
dry heat overkill cycles. 
3. After the cycle, aseptically transfer the spore strip to vessels of 
culture meedia. if spore suspensions were used, aseptically transfer 
the inoculated bottles to a laminar air flow work station $ add 
culture media to the bottles. use approximate possitive $ negative 
controls 
4. Determine the no. of survivors by plate counting or fraction 
negative methods. 
References: [11]
Validation of ethylene oxide sterilization cycles 
Eto has been a sterilant for over 50 years. 
• 5 variables critical to the Eto process. they are 
1. Eto concentration 
2. Relative humidity 
3. Temperature 
4. Time 
5. Pressure/vaccume. 
temp is the easiest variable to measure $ monitor, 
therefore temp is used as the indicator of the worst-case location 
within the loaded Eto strilizer. Once the worst case location is 
identified, the validation studies are conducted with the goal of 
inactivating a known conc of indicator mo’s in the worst-case 
location using a specific loading pattern with a specific Eto cycles. 
References: [15]
Procedure for the Eto cycle validation 
1. Use a laboratory sized Eto sterilizer during early phases of the 
validation process as long as the sterilizer is equipped with 
devices allowing variability in vaccume ,relative humidity, temp, 
gas pressure, timing,$ rate of gassing the chamber. 
2. Verify the calibration of all instrumentation involved in 
monitoring the Eto cycle. 
3. Perform an extensive temp distribution study using an empty 
sterilizer. 
4. Do a series of repetitive runs for each sterilization cycle in an 
empty vessel in order to verify the accuracy and reliability of the 
sterilizer contorls and monitoring equipment. 
5. Do a series of repetitive heat distribution and heat penetration 
runs using a loaded Eto sterilizer.
7.Test should be conducted on the final packaged product. 
8.Institute a documented monitoring system primary relying on 
bio-logical indicators,with lesser reliance on end-product 
sterility testing. 
References: [15]
Validation of radiation sterilization process 
• The major objective in validating a radiation sterilization 
process regardless of whether the mode of radiation is 
cobalt-60,cesium-137 or electron beam. 
• The radiation sterilization cycles are validated based upon 
the achievement of sterility ,many factors must be 
considered in the utilization and approval of the radiation 
sterilization process. such factors include 
 The physical appearance of the container system and its 
contents, 
 Stability of the active ingradient, if present, and 
 Safety of the irradiated material. 
References: [16]
VALIDATION OF WATER SUPPLY 
SYSTEMS
OBJECTIVE 
To understand: 
1. The need for water quality manual 
2. reason for usage of pharmaceutical water supply systems. 
3. The technical requirements for water supply systems. 
4. Different types of water supply systems. 
5. Validation requirements. 
6. Qualification & inspection requirement 
References: [17] 
24
INTRODUCTION 
 High-quality water is essential for the manufacturing of 
pharmaceuticals. Water is the most commonly used raw 
material in pharmaceutical manufacturing. 
 water is directly or indirectly used in the 
pharmaceutical manufacturing such as a major 
component in injectable products and in cleaning of 
manufacturing equipment. 
 It is one of the raw material that is usually processed by 
the pharmaceutical manufacturer prior to use because it 
cannot be supplied by the vendor. 
Water is thus an important raw material in GMP and in 
validating the manufacturing process. 
References: [17] 25
INTRODUCTION 
Quality of water should be specific for product quality. 
Water contains, 
• Organic and inorganic impurities 
• Microbial contamination 
• Endotoxin 
• Particulate contamination 
Low quality of water can lead to 
 product degradation 
 product contamination 
 loss of product and profit 
26 
References: [17]
TYPES OF WATER 
 Different grades of Water for Pharmaceutical Purposes-each 
type has its on characteristic for all parameters. 
Potable water 
Purified water 
Water for injection(WFI) 
Sterile water for injection, inhalation, irrigation 
Sterile bacteriostatic water for injection 
References: [18] 
27
References: [18]
DIFFERENT TECHNIQUES USED FOR 
WATER TREATMENT 
– De-chlorination (Sodium Bisulphite, Carbon Filter) 
– Filtration 
– Ultra Filtration 
– Softening 
– Demineralization 
– Reverse Osmosis 
– UV Treatment 
– Deionization 
– Ozonization 
References: [18] 
29
DIFFERENT EQUIPMENTS AND 
COMPONENTS FOR WATER SYSTEM 
• Piping 
• Valves 
• Pumps 
• Pressure gauges 
• Heat exchangers 
• Distillation unit 
• Filters 
• Deionizers 
• Sensors 
• Auxiliary equipment 
References: [19]
WATER STORAGE AND 
DISTRIBUTION – CONSIDERATIONS 
– Materials of Construction (Chemical and Heat 
Compatibility) 
• Stainless Steel (316 or 316L) 
• Teflon, Silicone, Viton (gaskets, diaphragms) 
– Minimize Dead Legs (<= 2 pipe diameters) 
– Smooth Surfaces (Mechanical Polish , Electropolish) 
– Clean joints (sanitary Tri®Clamp, automatic orbital 
welding) 
– Passivate interior surfaces to form barrier between water 
and free iron (0.5 to 1% alkali at 160ºF for 30 minutes 
followed by 1% Phosphoric Acid or Nitric Acid at 150ºF 
to 180º F for 10 minutes.) 
References: [19] 31
Conti…. 
• Design of the following should be appropriate to prevent 
recontamination after treatment- 
–Vent filter 
–Sanitary overflow 
–Tank UV light 
–Conical Bottom 
–Steam sterilization 
• Combination of on-line (TOC, Conductivity meter etc.) and 
off-line monitoring (lab testing by proper sampling) to ensure 
compliance with water specification 
32 
References: [19]
VALIDATION CONCEPT 
 To prove the performance of processes or systems under 
all conditions expected to be encountered during future 
operations. 
 To prove the performance, one must demonstrate 
(document) that the processes or systems consistently 
produce the specified quantity and quality of water when 
operated and maintained according to specific written 
operating and maintenance procedures. 
 validation involves proving- 
1. Engineering design 
2.Operating procedures and acceptable ranges for control 
parameters 
3. Maintenance procedures to accomplish it 
References: [17] 
33
Conti.. 
• the system must be carefully, 
-designed 
-installed 
-tested during processing, after construction, and 
under all operating conditions. 
• Variations in daily, weekly and annual system usage 
patterns must be validated. 
References: [17] 
34
STEPS OF VALIDATION 
• Establishing standards for quality attributes 
• Defining system and subsystem 
• Designing equipment, control, & monitoring 
technologies 
• Establishing standards for operating parameters 
• Developing an IQ stage & OQ stage 
• Establishing alert and action levels 
• Developing a prospective PQ stage 
• Completing protocols and documenting each steps 
References: [17] 
35
Conti… 
36
DESIGN QUALIFICATION OF WATER SYSTEM 
Based on the URS, supplier designs the equipment. 
• This is 1st step in the qualification of new water supply systems. 
• Define process schematically by use of PFD and P&IDs. 
• It is documented the design of the system & will include : 
-Functional Specification.(Storage, purification, etc) 
-Technical/Performance specification for 
equipment.(requirements of water volume and flow, define 
pumps and pipe sizes ) 
-Detailed layout of the system. 
Design must be in compliance with GMPs and other regulatory 
requirements. 
References: [18] 
37
REFERENCES 
1. Mascoli , C,C should end product sterility testing continue, 
med dev Diag ind 3: 8-9(1981) 
2. Bowman. F.W.the sterility testing of pharmaceuticals. j. 
pharm sci 58:1301-1308(1969) 
3. Ernst R.R., West, K.L.Doyle, J.E.problemareas in sterility 
testing .Bull parenter drug assoc 23:29-39(1969) 
4. Akers ,M.J. In : parenteral quality control :sterility , 
pyrogen,particulate matter, and package integrity testing, 2nd 
addition . Newyork: Marcel Dekker ,pp.1-4 (1994) 
5. Brewer , J.H. In: G.L. Redish , ed. Anticeptics, disinifectants, 
fungicides, and sterilization , 2nd ed. Philadelphia: Lea $ 
Febiger,pp,160-161(1957) 
6. Food $ drug administration. Guidelines on general principles 
of process validation. Rockville , MD:FDA(1984)
7. Reich, R.R.,Whibourne, J.E.,Mcdaniel, A.W.effect of storage 
conditions on the performance of B.steaothermophilus 
biological indicators. 
8.validation of steam sterilization cycles .technical monograph 
no.1 parenteral drug association (1978) . 
9.Tsuji,K.harrison,S.J.dry-heat destroction of lipopolysaccharide : 
dry heat destruction kinetics. 
10.pflug .I.J. holcomb .R.J principles of thermal destruction of 
mo’s. 
11.validation of dry heat sterilization used for sterilization and 
depyrogenation, technical no.3parenteral drug 
administration(1981) 
12.simmons,P.L.validation of dry heat sterilizer.pharm eng 
38(may- july 1981).
13.simmons,P.L.Hot air and continuous sterilization.akers,M.J.,Ketron 
,K.,Thompson. B.F.value requirements for the destruction of endotoxin 
in the validation of dry heat sterilization cycles. 
14.Avis,K.E.jewell,ludwig,J.D.avis,K.E.validation of a heating cellfor 
precisely controlled studies on the thermal destruction of endotoxin in 
glass. 
15.Robertson,J.H.,Townsend.M.W,Allen;valenti,simmons,P.L.ETO 
sterilization,caputo,R.A.Rohn. 
16.ISO11137:1995.sterilization of health care products –requirements for 
validation $ routine control –radiation sterilization . 
17. ICH.GMP guide for active pharmaceutical ingredients.Q7A(march 
15,2000). 
18.general information:water for pharmaceutical 
purposes.U.S.pharmacopeia.vol.25.rockville,MD:U.S.pharmacopeial 
convention,pp.2261-2270(2002). 
19.Pharmaceutical engineering guide vol.4:water $ steam guide. tampa, 
FL:ISPE, (1997).
41 
THANK 
YOU

More Related Content

What's hot

Validation of dry heat sterilization
Validation of dry heat sterilizationValidation of dry heat sterilization
Validation of dry heat sterilizationDeepak Shanbhag
 
1 2 sterilisation-validationqualification
1 2 sterilisation-validationqualification1 2 sterilisation-validationqualification
1 2 sterilisation-validationqualificationAnAnD SiNgH PaTeL
 
6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATION
6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATION6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATION
6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATIONaccessamg
 
Stopper Siliconization
Stopper SiliconizationStopper Siliconization
Stopper Siliconizationaaron_mertens
 
EQUIPMENT VALIDATION : HOT AIR OVEN
EQUIPMENT VALIDATION : HOT AIR OVENEQUIPMENT VALIDATION : HOT AIR OVEN
EQUIPMENT VALIDATION : HOT AIR OVENSagar Savale
 
Autoclave validatin
Autoclave validatinAutoclave validatin
Autoclave validatinSagar Savale
 
PROCESS VALIDATION OF INJECTABLES
PROCESS VALIDATION OF INJECTABLESPROCESS VALIDATION OF INJECTABLES
PROCESS VALIDATION OF INJECTABLESKhushboo Pasbola
 
One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...Palash Das
 
Bowie Dick Tests - Failure Modes and Specification
Bowie Dick Tests - Failure Modes and SpecificationBowie Dick Tests - Failure Modes and Specification
Bowie Dick Tests - Failure Modes and SpecificationChris Maughan
 
Validation of dry_powder_mixer_ .
Validation of dry_powder_mixer_ .Validation of dry_powder_mixer_ .
Validation of dry_powder_mixer_ .Aditi Roy
 
377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITY377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITYGOVIND YuvRaju
 
Environmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingEnvironmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingJerry Fahrni
 
Pkging of sterile products
Pkging of sterile productsPkging of sterile products
Pkging of sterile productsDr Qureshi
 

What's hot (20)

Validation of dry heat sterilization
Validation of dry heat sterilizationValidation of dry heat sterilization
Validation of dry heat sterilization
 
1 2 sterilisation-validationqualification
1 2 sterilisation-validationqualification1 2 sterilisation-validationqualification
1 2 sterilisation-validationqualification
 
Validation of sterelization_equipment
Validation of sterelization_equipmentValidation of sterelization_equipment
Validation of sterelization_equipment
 
6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATION
6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATION6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATION
6 CLASSES OF CHEMICAL INDICATORS FOR STEAM STERILIZATION
 
Stopper Siliconization
Stopper SiliconizationStopper Siliconization
Stopper Siliconization
 
EQUIPMENT VALIDATION : HOT AIR OVEN
EQUIPMENT VALIDATION : HOT AIR OVENEQUIPMENT VALIDATION : HOT AIR OVEN
EQUIPMENT VALIDATION : HOT AIR OVEN
 
Tunnel thermographs
Tunnel thermographsTunnel thermographs
Tunnel thermographs
 
Autoclave validatin
Autoclave validatinAutoclave validatin
Autoclave validatin
 
Stopper Processing
Stopper ProcessingStopper Processing
Stopper Processing
 
Filter validation
Filter validationFilter validation
Filter validation
 
PROCESS VALIDATION OF INJECTABLES
PROCESS VALIDATION OF INJECTABLESPROCESS VALIDATION OF INJECTABLES
PROCESS VALIDATION OF INJECTABLES
 
One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...
 
Bowie Dick Tests - Failure Modes and Specification
Bowie Dick Tests - Failure Modes and SpecificationBowie Dick Tests - Failure Modes and Specification
Bowie Dick Tests - Failure Modes and Specification
 
Environmental monitoring
Environmental monitoringEnvironmental monitoring
Environmental monitoring
 
ENVIRONMENTAL MONITORING
ENVIRONMENTAL MONITORINGENVIRONMENTAL MONITORING
ENVIRONMENTAL MONITORING
 
Validation of dry_powder_mixer_ .
Validation of dry_powder_mixer_ .Validation of dry_powder_mixer_ .
Validation of dry_powder_mixer_ .
 
377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITY377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITY
 
Environmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingEnvironmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testing
 
Pkging of sterile products
Pkging of sterile productsPkging of sterile products
Pkging of sterile products
 
Depyrogenation by dry heat
Depyrogenation by dry heatDepyrogenation by dry heat
Depyrogenation by dry heat
 

Similar to Presentation of industrial pharmacy 2

Qualification of Autoclave - Dipankar.pptx
Qualification of Autoclave - Dipankar.pptxQualification of Autoclave - Dipankar.pptx
Qualification of Autoclave - Dipankar.pptxGNIPST
 
Thermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree DiwakarThermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree DiwakarDurgashree Diwakar
 
Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...Talal Albudayri
 
Thermogravimetric analysis.pptx
Thermogravimetric analysis.pptxThermogravimetric analysis.pptx
Thermogravimetric analysis.pptxkaroline Enoch
 
Final Testing Report TES
Final Testing Report TESFinal Testing Report TES
Final Testing Report TESZachery Pope
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17PerkinElmer, Inc.
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
An Innovative Approach for Humidity Control by Using Deliquescent Materials i...
An Innovative Approach for Humidity Control by Using Deliquescent Materials i...An Innovative Approach for Humidity Control by Using Deliquescent Materials i...
An Innovative Approach for Humidity Control by Using Deliquescent Materials i...IJMER
 
Qualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptxQualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptxDhruvi50
 
Laboratory design, safety and management
Laboratory design, safety and managementLaboratory design, safety and management
Laboratory design, safety and managementponnala saikumar
 
equipmentsvalidation.pdf
equipmentsvalidation.pdfequipmentsvalidation.pdf
equipmentsvalidation.pdfpshashank0409
 

Similar to Presentation of industrial pharmacy 2 (20)

Qualification of Autoclave - Dipankar.pptx
Qualification of Autoclave - Dipankar.pptxQualification of Autoclave - Dipankar.pptx
Qualification of Autoclave - Dipankar.pptx
 
Thermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree DiwakarThermal techniques- DSC- Durgashree Diwakar
Thermal techniques- DSC- Durgashree Diwakar
 
Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...
 
Thermogravimetric analysis.pptx
Thermogravimetric analysis.pptxThermogravimetric analysis.pptx
Thermogravimetric analysis.pptx
 
Final Testing Report TES
Final Testing Report TESFinal Testing Report TES
Final Testing Report TES
 
4_2018_04_07!06_25_24_AM (1).pptx
4_2018_04_07!06_25_24_AM (1).pptx4_2018_04_07!06_25_24_AM (1).pptx
4_2018_04_07!06_25_24_AM (1).pptx
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
MET 214 Module 6
MET 214 Module 6MET 214 Module 6
MET 214 Module 6
 
Validation of sterilization methods by nilesh sharma
Validation of sterilization methods by nilesh sharmaValidation of sterilization methods by nilesh sharma
Validation of sterilization methods by nilesh sharma
 
An Innovative Approach for Humidity Control by Using Deliquescent Materials i...
An Innovative Approach for Humidity Control by Using Deliquescent Materials i...An Innovative Approach for Humidity Control by Using Deliquescent Materials i...
An Innovative Approach for Humidity Control by Using Deliquescent Materials i...
 
Qualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptxQualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptx
 
Laboratory design, safety and management
Laboratory design, safety and managementLaboratory design, safety and management
Laboratory design, safety and management
 
Mpat presentation
Mpat presentationMpat presentation
Mpat presentation
 
Autoclaves
AutoclavesAutoclaves
Autoclaves
 
Validation ppt
Validation pptValidation ppt
Validation ppt
 
final adsorbtion.pptx
final adsorbtion.pptxfinal adsorbtion.pptx
final adsorbtion.pptx
 
Sterilization
SterilizationSterilization
Sterilization
 
equipmentsvalidation.pdf
equipmentsvalidation.pdfequipmentsvalidation.pdf
equipmentsvalidation.pdf
 

More from Malla Reddy College of Pharmacy (20)

Rna secondary structure prediction
Rna secondary structure predictionRna secondary structure prediction
Rna secondary structure prediction
 
Proteomics
ProteomicsProteomics
Proteomics
 
Proteins basics
Proteins basicsProteins basics
Proteins basics
 
Protein structure classification
Protein structure classificationProtein structure classification
Protein structure classification
 
Protein identication characterization
Protein identication characterizationProtein identication characterization
Protein identication characterization
 
Protein modeling
Protein modelingProtein modeling
Protein modeling
 
Primerdesign
PrimerdesignPrimerdesign
Primerdesign
 
Phylogenetic studies
Phylogenetic studiesPhylogenetic studies
Phylogenetic studies
 
Multiple sequence alignment
Multiple sequence alignmentMultiple sequence alignment
Multiple sequence alignment
 
Homology modeling tools
Homology modeling toolsHomology modeling tools
Homology modeling tools
 
Homology modeling
Homology modelingHomology modeling
Homology modeling
 
Genome assembly
Genome assemblyGenome assembly
Genome assembly
 
Genome analysis2
Genome analysis2Genome analysis2
Genome analysis2
 
Genome analysis
Genome analysisGenome analysis
Genome analysis
 
Fasta
FastaFasta
Fasta
 
Drug design intro
Drug design introDrug design intro
Drug design intro
 
Drug design
Drug designDrug design
Drug design
 
Data retrieval
Data retrievalData retrieval
Data retrieval
 
Blast
BlastBlast
Blast
 
Biological databases
Biological databasesBiological databases
Biological databases
 

Presentation of industrial pharmacy 2

  • 1. Process validation of Sterilization & Water Process Systems
  • 2. STERILIZATION VALIDATION • Sterile products have several unique properties such as 1. Free from micro organisms 2. Free from pyrogens 3. Free from particulates 4. High standards of purity and quality References: [1,2,3, & 4]
  • 3. Methods of sterilization of products HEAT:- 1. Moist heat-auto clave 2. Dry heat-hot air oven GAS:- 1. ethylene oxide 2. Peracetic acid 3. Hydrogen peroxide(vapor phase) 4. Chlorine di oxide References: [5& 6]
  • 4. Radiation 1. Gamma rays 2. Beta rays 3. Ultraviolet 4. Microwave References: [5& 6]
  • 5. Validation of steam sterilization cycles Qualification and calibration 1. Mechanically checking ,upgrading, and qualifying the sterilizer unit
  • 6. Selection and calibration of thermocouples • Cu constantan wires coated with teflon are a popular choice as thermocouple monitors • Accuracy of thermocouples should be ±0.5°C. Temperature accuracy is especially important in steam sterilization validation. • Thermocouple accuracy is determined using NATIONAL BUREAU OF STANDARDS (NBS) traceable constant temperature calibration instruments. • Thermocouples should be calibrated before and after validation experiment at 2 temperatures i.e. 0C & 125 C . • New thermocouple-recording devices are capable of automatically correcting temperature • Any thermocouple that senses a temperature of more than 0.5 C away from the calibration temperature bath should be discarded • Temperature recorders should be capable of printing temperature data in 0.1°C increments References: [8 & 9]
  • 7. Selection & calibration of BI • The organism most resistant to steam heat is the bacterial spore B. stearothermophilus. This bacterial spore is commonly used BI’s in validating steam sterilization cycles. • Spore trips or spore suspensions are used in the validation studies. the no. of mo’s per ml of suspension must be as accurately known as D value. • Precautions should be taken to use proper storage conditions for B. stearothermophilus BIs .storing in the freezer provides a more stable resistance profile for the shelf life of the indicator. References: [7 ]
  • 8. Heat distribution studies It include 2 phases 1. Heat distribution in any empty autoclave chamber. 2. Heat distribution in a loaded autoclave chamber. a. 10-20 thermocouples should be used/cycle. thermocouples should be secured inside the chamber. b. The trips where the wires are soldered should not make contact with the autoclave interior walls or any metal surface. c. 1 end of thermocouple should remain in an ice bath and high temperature oil bath during each cycle for reference when the temp monitoring equipement has the capability for electronically compensating each temp measurement against an internal reference. d. Heat distribution studies following the initial study may employ fewer thermocouples as the cool spot in the chamber & in the load is identified. e. The difference in temp b/n the coolest spot $ the mean chamber temperature should not be greater than  2.5C. References: [8 ]
  • 9. Heat penetration studies • This is the most critical component of the entire validation process • Thermocouples will be placed both inside and outside the container at the coolspot location(s) in the steam exhaust line and in constant temperature baths outside the chamber • The sterilization cycle design must be based on the heating charecteristics of the load and containers located in the slowest heating zone of the load. • The variation in the rate of heating of the slowest heating zone must be known, so this variation must be determined under fully loaded conditions • The effect of load to load variation on the time-temperature profile must also be determined. • Then the statistically worst case conditions should be used in the final sterilization process design
  • 10. • The final step in steam sterilization process the establishment of a monitoring program to ensure that the validated cycle remains essentially unchanged in the future. • Cycle monitoring usually involves the use of thermocouples to measure heat penetration at the cool spot location. • Any changes in the load size, load configuration, or container charecteristics (volume, geometry etc) must be accompanied by repeat validation studies to prove that the cool spot location has not changed References: [10 ]
  • 11. Validation of dry-heat sterilization cycles 1. Batch oven validation • Air balance determination in an empty oven data are obtained on the flow rates of both intake and exhaust air. air should be balanced so that positive pressure is exerted to the non sterile side when the door is opend and air velocity across and up and down the opening of the door is ±50 FPM of the average velocity
  • 12. • Heat distribution of an empty chamber thermocouples should be situated according to a specific predetermined pattern. Repeatability of temp attainment and identification of cold spot can be achieved if the temp range is ±15°C at all monitored locations. • Heat penetration studies. These studies should be designed to determine the location of slowest heating point within a commodity at various locations of test load in sterilizer. • Mechanical repeatability. during all these studies mechanical repeatability in terms of air velocity, temp consistency, reliability and sensitivity of all the oven and instrumental controls must be verified. References: [11 ]
  • 13. 2.Tunnel sterilizer validation  Air balance determination • Proper air balance is more critical to a tunnel sterile process than a batch oven process .since the items being sterilized are exposed to a different air systems(eg:-heating zone $ cooling zone).in the absence of a critical balance of air dynamics, either the items will not be cooled sufficiently once they exit the tunnel or they will be cooled too quickly. causing the glass to shatter and contaminate the entire tunnel area with particles. • The major problem in validating tunnel sterilizers is the control of particules. not only are items exposed to great extreams in temp, but also the conveyer belt is a natural source of particulates because metal is moving against metal. • Air must be particulate-free as it enters the tunnel area; therefore, all high efficiency particulate air(HEPA)filters in the tunnel must be tested and certified prior to validation studies.
  • 14. Heat distribution studies • Thermocouples used in tunnel sterilizer validation must be sufficiently durable to withstand the extremely high( ≥ 300 c)temperatures in the heating zone area of the tunnel heat-distribution studies should determine where the cold spots are located as a function of the width of the belt $ height of the tunnel chamber. trays or tracks of ampules are vials should run through the tunnel • Bottle-mapping studies may also be conducted during this phase. the purpose of these studies is to determine possible locations inside the container that are most difficult to heat. References: [12 ]
  • 15. Heat penetration studies • For testing of the tunnel sterilization, heat-penetration studies must be completed in order to identify the coolest container in the entire load. Results of heat-distribution studies should aid in the predicting where the coolest location with in the load should be. Thermocouples should be diposited at or near the coolest point inside the container from bottle-mapping studies. • The containers inner surface should be in contact with the thermocouple tip because the objective is to sterilize the inner walls of the container, as well as the inner space. • Every loading should be done using 10-20 thermocouples distributed through out the load. References: [ 9 ]
  • 16.  Mechanical repeatability • Air velocity, air particulates, temp consistency and reliability of all the tunnel controls(heat zone temperatures, belt speed, and blower functions)must be proved during the physical validation studies. References: [ 9 ]
  • 17. 3.Biological process validation of dry heat sterilization cycles • mo’s known to be most resistant to dry heat must be used to prove the ability of dry heat cycle to destroy them at the coolest location in load. the dry heat process is claimed to produce both sterile and pyrogen-free commodities, validation studies must be done using both mo’s $ microbial endotoxins. • Biological validation of dry heat cycles should be based on the destruction of endotoxin rather than on the destruction of mo’s because of the enormous dry heat resistance of endotoxin compared to mo’s. • With both mo’s $ endotoxin challenges, the cool spot identified in heat distribution $ the heat penetration studies will be the logical location to run the microbial challenge tests. containers inoculated with the microbial cells or endotoxin will be situated adjacent to identical containers into which thermocouples are secured to monitor temp. References: [ 9 ]
  • 18. Step by step sequence in the microbial validation of a dry heat process for sterilizing and depyrogenating large volume glass containers by wegel $ akers et al 1. Place spore carrier in approximately 12 glass bottles located at the coolest area of the oven. bottles adjucent to the inoculated bottles should contain thermocouples for the monitoring purposes . 2. Run a complete cycle using the desired loading pattern for future dry heat overkill cycles. 3. After the cycle, aseptically transfer the spore strip to vessels of culture meedia. if spore suspensions were used, aseptically transfer the inoculated bottles to a laminar air flow work station $ add culture media to the bottles. use approximate possitive $ negative controls 4. Determine the no. of survivors by plate counting or fraction negative methods. References: [11]
  • 19. Validation of ethylene oxide sterilization cycles Eto has been a sterilant for over 50 years. • 5 variables critical to the Eto process. they are 1. Eto concentration 2. Relative humidity 3. Temperature 4. Time 5. Pressure/vaccume. temp is the easiest variable to measure $ monitor, therefore temp is used as the indicator of the worst-case location within the loaded Eto strilizer. Once the worst case location is identified, the validation studies are conducted with the goal of inactivating a known conc of indicator mo’s in the worst-case location using a specific loading pattern with a specific Eto cycles. References: [15]
  • 20. Procedure for the Eto cycle validation 1. Use a laboratory sized Eto sterilizer during early phases of the validation process as long as the sterilizer is equipped with devices allowing variability in vaccume ,relative humidity, temp, gas pressure, timing,$ rate of gassing the chamber. 2. Verify the calibration of all instrumentation involved in monitoring the Eto cycle. 3. Perform an extensive temp distribution study using an empty sterilizer. 4. Do a series of repetitive runs for each sterilization cycle in an empty vessel in order to verify the accuracy and reliability of the sterilizer contorls and monitoring equipment. 5. Do a series of repetitive heat distribution and heat penetration runs using a loaded Eto sterilizer.
  • 21. 7.Test should be conducted on the final packaged product. 8.Institute a documented monitoring system primary relying on bio-logical indicators,with lesser reliance on end-product sterility testing. References: [15]
  • 22. Validation of radiation sterilization process • The major objective in validating a radiation sterilization process regardless of whether the mode of radiation is cobalt-60,cesium-137 or electron beam. • The radiation sterilization cycles are validated based upon the achievement of sterility ,many factors must be considered in the utilization and approval of the radiation sterilization process. such factors include  The physical appearance of the container system and its contents,  Stability of the active ingradient, if present, and  Safety of the irradiated material. References: [16]
  • 23. VALIDATION OF WATER SUPPLY SYSTEMS
  • 24. OBJECTIVE To understand: 1. The need for water quality manual 2. reason for usage of pharmaceutical water supply systems. 3. The technical requirements for water supply systems. 4. Different types of water supply systems. 5. Validation requirements. 6. Qualification & inspection requirement References: [17] 24
  • 25. INTRODUCTION  High-quality water is essential for the manufacturing of pharmaceuticals. Water is the most commonly used raw material in pharmaceutical manufacturing.  water is directly or indirectly used in the pharmaceutical manufacturing such as a major component in injectable products and in cleaning of manufacturing equipment.  It is one of the raw material that is usually processed by the pharmaceutical manufacturer prior to use because it cannot be supplied by the vendor. Water is thus an important raw material in GMP and in validating the manufacturing process. References: [17] 25
  • 26. INTRODUCTION Quality of water should be specific for product quality. Water contains, • Organic and inorganic impurities • Microbial contamination • Endotoxin • Particulate contamination Low quality of water can lead to  product degradation  product contamination  loss of product and profit 26 References: [17]
  • 27. TYPES OF WATER  Different grades of Water for Pharmaceutical Purposes-each type has its on characteristic for all parameters. Potable water Purified water Water for injection(WFI) Sterile water for injection, inhalation, irrigation Sterile bacteriostatic water for injection References: [18] 27
  • 29. DIFFERENT TECHNIQUES USED FOR WATER TREATMENT – De-chlorination (Sodium Bisulphite, Carbon Filter) – Filtration – Ultra Filtration – Softening – Demineralization – Reverse Osmosis – UV Treatment – Deionization – Ozonization References: [18] 29
  • 30. DIFFERENT EQUIPMENTS AND COMPONENTS FOR WATER SYSTEM • Piping • Valves • Pumps • Pressure gauges • Heat exchangers • Distillation unit • Filters • Deionizers • Sensors • Auxiliary equipment References: [19]
  • 31. WATER STORAGE AND DISTRIBUTION – CONSIDERATIONS – Materials of Construction (Chemical and Heat Compatibility) • Stainless Steel (316 or 316L) • Teflon, Silicone, Viton (gaskets, diaphragms) – Minimize Dead Legs (<= 2 pipe diameters) – Smooth Surfaces (Mechanical Polish , Electropolish) – Clean joints (sanitary Tri®Clamp, automatic orbital welding) – Passivate interior surfaces to form barrier between water and free iron (0.5 to 1% alkali at 160ºF for 30 minutes followed by 1% Phosphoric Acid or Nitric Acid at 150ºF to 180º F for 10 minutes.) References: [19] 31
  • 32. Conti…. • Design of the following should be appropriate to prevent recontamination after treatment- –Vent filter –Sanitary overflow –Tank UV light –Conical Bottom –Steam sterilization • Combination of on-line (TOC, Conductivity meter etc.) and off-line monitoring (lab testing by proper sampling) to ensure compliance with water specification 32 References: [19]
  • 33. VALIDATION CONCEPT  To prove the performance of processes or systems under all conditions expected to be encountered during future operations.  To prove the performance, one must demonstrate (document) that the processes or systems consistently produce the specified quantity and quality of water when operated and maintained according to specific written operating and maintenance procedures.  validation involves proving- 1. Engineering design 2.Operating procedures and acceptable ranges for control parameters 3. Maintenance procedures to accomplish it References: [17] 33
  • 34. Conti.. • the system must be carefully, -designed -installed -tested during processing, after construction, and under all operating conditions. • Variations in daily, weekly and annual system usage patterns must be validated. References: [17] 34
  • 35. STEPS OF VALIDATION • Establishing standards for quality attributes • Defining system and subsystem • Designing equipment, control, & monitoring technologies • Establishing standards for operating parameters • Developing an IQ stage & OQ stage • Establishing alert and action levels • Developing a prospective PQ stage • Completing protocols and documenting each steps References: [17] 35
  • 37. DESIGN QUALIFICATION OF WATER SYSTEM Based on the URS, supplier designs the equipment. • This is 1st step in the qualification of new water supply systems. • Define process schematically by use of PFD and P&IDs. • It is documented the design of the system & will include : -Functional Specification.(Storage, purification, etc) -Technical/Performance specification for equipment.(requirements of water volume and flow, define pumps and pipe sizes ) -Detailed layout of the system. Design must be in compliance with GMPs and other regulatory requirements. References: [18] 37
  • 38. REFERENCES 1. Mascoli , C,C should end product sterility testing continue, med dev Diag ind 3: 8-9(1981) 2. Bowman. F.W.the sterility testing of pharmaceuticals. j. pharm sci 58:1301-1308(1969) 3. Ernst R.R., West, K.L.Doyle, J.E.problemareas in sterility testing .Bull parenter drug assoc 23:29-39(1969) 4. Akers ,M.J. In : parenteral quality control :sterility , pyrogen,particulate matter, and package integrity testing, 2nd addition . Newyork: Marcel Dekker ,pp.1-4 (1994) 5. Brewer , J.H. In: G.L. Redish , ed. Anticeptics, disinifectants, fungicides, and sterilization , 2nd ed. Philadelphia: Lea $ Febiger,pp,160-161(1957) 6. Food $ drug administration. Guidelines on general principles of process validation. Rockville , MD:FDA(1984)
  • 39. 7. Reich, R.R.,Whibourne, J.E.,Mcdaniel, A.W.effect of storage conditions on the performance of B.steaothermophilus biological indicators. 8.validation of steam sterilization cycles .technical monograph no.1 parenteral drug association (1978) . 9.Tsuji,K.harrison,S.J.dry-heat destroction of lipopolysaccharide : dry heat destruction kinetics. 10.pflug .I.J. holcomb .R.J principles of thermal destruction of mo’s. 11.validation of dry heat sterilization used for sterilization and depyrogenation, technical no.3parenteral drug administration(1981) 12.simmons,P.L.validation of dry heat sterilizer.pharm eng 38(may- july 1981).
  • 40. 13.simmons,P.L.Hot air and continuous sterilization.akers,M.J.,Ketron ,K.,Thompson. B.F.value requirements for the destruction of endotoxin in the validation of dry heat sterilization cycles. 14.Avis,K.E.jewell,ludwig,J.D.avis,K.E.validation of a heating cellfor precisely controlled studies on the thermal destruction of endotoxin in glass. 15.Robertson,J.H.,Townsend.M.W,Allen;valenti,simmons,P.L.ETO sterilization,caputo,R.A.Rohn. 16.ISO11137:1995.sterilization of health care products –requirements for validation $ routine control –radiation sterilization . 17. ICH.GMP guide for active pharmaceutical ingredients.Q7A(march 15,2000). 18.general information:water for pharmaceutical purposes.U.S.pharmacopeia.vol.25.rockville,MD:U.S.pharmacopeial convention,pp.2261-2270(2002). 19.Pharmaceutical engineering guide vol.4:water $ steam guide. tampa, FL:ISPE, (1997).