Hemodynamic principles basics
Pressure wave: Complex periodic fluctuation in force per unit area
Fundamental frequency: number of times the pressure wave cycles in 1 second
Harmonic: multiple of fundamental frequency Essential physiologic information is contained within the first 10 harmonics
At a HR of 120, the fundamental frequency is 2Hz, and 10th harmonic is 20Hz.
A pressure response system with a frequency response range that is flat to atleast 20 Hz will be required.
Damped Natural frequency should be 3 times as fast as the 10thHarmonic of the pressure measured ie 60 Hz.
Fidelity of the recording drops with increasing heart rate.HR. End pressure artifact
Flowing blood has kinetic energy which when suddenly halted in part coverts into pressure and records a pressure which is artifactually elevated
Seen with end hole catheters
This added pressure may range from 2-10 mm Hg
Catheter impact artifact
Pressure due to impact on adjacent structure– valve, papillary muscles, moderator band.
Common with the pigtail catheter in the LV, where the MV hits the catheter as they open in early diastole Normally RA diastolic pressure is equal to RV diastolic pressure except in early diastole which drives the rapid filling from atria to ventricles( E wave on ECHO)
An elevated early RA-RV pressure gradient may be seen in patients with decompensated right heart failure who characteristically have elevated RA pressure that pushes blood in RV
Furthermore due to the loss of RV compliance , the diastolic pressure rises rapidly to a high plateau level Wedge Pressure
Pressure obtained when an end-hole catheter is positioned in a “designated” blood vessel with its open end-hole facing a capillary bed, with no connecting vessels conducting flow into or away from the “designated” blood vessel between the catheter’s tip and the capillary bed.
True wedge pressure can be measured only in the absence of flow, allowing pressure to equilibrate across the capillary be
3. +
A complex periodic fluctuation in force per unit area
A pressure wave is the cyclical force generated by
cardiac muscle contraction
Its amplitude and duration are influenced by various
mechanical and physiological parameters
1. Force of the contracting chamber
2. Surrounding structures - contiguous chambers of the heart
pericardium, lungs, vasculature
3. Physiological variables - heart rate, respiratory cycle
Pressure Wave
4. +
Pressure wave: Complex periodic fluctuation in force per
unit area
Fundamental frequency: number of times the pressure
wave cycles in 1 second
Harmonic: multiple of fundamental frequency
Pressure Measurement
Terminology
5. +
Essential physiologic information is contained within the
first 10 harmonics
At a HR of 120, the fundamental frequency is 2Hz, and
10th harmonic is 20Hz.
A pressure response system with a frequency response
range that is flat to atleast 20 Hz will be required.
Damped Natural frequency should be 3 times as fast as
the 10thHarmonic of the pressure measured ie 60 Hz.
Fidelity of the recording drops with increasing heart
rate.HR.
Pressure Measurement
Fourier analysis
7. +
Sensitivity
Ratio of amplitude of record signals to input
signals(The more the rigid membrane the lower the
sensitivity)
Frequency resposnse
Ratio of output amplitude to input amplitude over a range of
frequencies of the input pressure
To measure pressure accurately , frequency response must be
constant over a given range of frequencies.
Pressure Measurement Devices
Terminology
8. +
NATURAL FREQUENCY AND
DAMPING
Natural frequency
Frequency at which fluid oscillates in a catheter when it is
tapped.
At Natural frequency, ratio of output/input amplitude is
maximal.
The higher the natural frequency of the system, the more
accurate the pressure measurement.
DAMPING
It is a means of dissipating the energy of the oscillations of
sensing membrane
Optimal damping dissipates the energy gradually maintaining
frequency response curve nearly flat
10. +
Baim DS and Grossman W. Cardiac Catheterization, Angiography, and Intervention. 8h Edition. Baltimore: Williams and
Wilkins, 2014
Pressure Measurement
optimal damping
Natural
frequency
catheter
radius
𝑓𝑙𝑢𝑖𝑑 𝑑𝑒𝑛𝑠𝑖𝑡𝑦 𝑥 𝑐𝑎𝑡ℎ𝑒𝑡𝑒𝑟 𝑐𝑜𝑚𝑝𝑙𝑖𝑎𝑛𝑐𝑒
= x
SHORTER catheter
LARGER catheter lumen
LIGHTER fluid
HIGHER natural frequency
Catheter
length
√
1
1
11. +
Transforming pressure wave to electrical
signals
Strain gauge pressure transducer Strain gauge connection of Wheatstone
bridge
12. +
Balancing and Zeroing
Balancing a transducer
Add variable resistance into the circuit so that at an arbitrary baseline
pressure the output can be reduced to zero
Zeroing
Zero reference level at mid chest
14. + Fluid Filled Catheter
Fluid-filled catheter attached to a pressure transducer
Pressure wave is transmitted by the fluid column within
the catheter
Data should be collected ,with the patient in steady state
and before introduction of radiographic contrast.
15. +
Deterioration of frequency response
Introduction of air permits damping and reduces natural frequency by
serving as added compliance.
When natural frequency of pressure system falls, high frequency
components of the pressure waveform (intraventricular pressure rise
and fall) may set the system into oscillation, producing “pressure
overshoots” ( early systole & diastole of ventricular pressure curve).
Flushing-restores the frequency response of system.
Artifacts
16. +
Movement artifact (WHIP Artifact)
May produce superimposed waves of ±10 mm Hg
Motion of tip of the catheter within the measured chamber
→ Enhance the fluid oscillations of the transducer system
Render systolic and to a lesser extent diastolic pressures unreliable.
No way to fix it internally.
Stabilize externally.
Artifacts
17. +
End pressure artifact
Flowing blood has kinetic energy which when suddenly halted in part
coverts into pressure and records a pressure which is artifactually elevated
Seen with end hole catheters
This added pressure may range from 2-10 mm Hg
Catheter impact artifact
Pressure due to impact on adjacent structure– valve, papillary muscles,
moderator band.
Common with the pigtail catheter in the LV, where the MV hits the catheter
as they open in early diastole
18. +
Systolic Pressure amplification in the periphery
Reflected wave (tidal wave)
Peak SBP in radial, brachial, femoral > peak SBP in central Aorta by
up to 20 mmHg
Largely as a consequence of reflected wave from Aortic bifurcation,
arterial branching, small peripheral vessels.
Use of a double lumen catheter(double lumen pigtail) allows
measurement of LV and central aorta measurements simultaneously ,
thus avoiding this problem
19. + Micromanometer –Tipped Catheters
Fluid filled system- distortion of wave forms- artifacts, amplification of
systolic pressure in periphery, damping or augmentation of frequency
response system.
For precise, undistorted ,high fidelity pressure recordings
Micromamometer chips at the end of catheters
Interposing fluid column is eliminated
Have higher natural frequencies and more optimal damping
characteristics
a. To assess pressure waveform contours in a tachycardia
situation, rate of ventricular pressure rise(dp/dt) etc
b. Limitation- additional cost, fragility , time needed for properly
calibrating and using the system
22. +
Hemodynamic Parameters
Reference Values
Davidson CJ, et al. Cardiac Catheterization. In: Heart Disease: A Textbook of Cardiovascular Medicine,
Edited by E. Braunwald
Average Range Average Range
a wave
v wave
mean
Right ventricle
peak systolic
end diastolic
Pulmonary artery
peak systolic
Right atrium
end diastolic
mean
6
5
25
9
15
25
4
3
2 - 7
2 - 7
15 - 30
4-12
9-19
15-30
1 - 7
1 - 5
mean
Left atrium
a wave
v wave
mean
Left ventricle
peak systolic
end diastolic
PCWP
Central aorta
peak systolic
9
12
8
130
8
10
4 - 12
6 - 21
2 - 12
90 - 140
5 - 12
4 - 16
130 90 - 140
70 60 - 90
end diastolic
mean 85 70 -105
23. +
Right Heart Catheterization
Right Atrial Pressure
• “a” wave
– Atrial systole
• “c” wave
– Protrusion of TV into RA
• “a” wave
– Atrial systole
• “c” wave
– Protrusion of TV into RA
• “x” descent
– Relaxation of RA
– Downward pulling of tricuspid
annulus by RV contraction
• “v” wave
– RV contraction
– Height related to atrial compliance & amount of blood return
– Smaller than a wave
• “a” wave
– Atrial systole
• “c” wave
– Protrusion of TV into RA
• “x” descent
– Relaxation of RA
– Downward pulling of tricuspid
annulus by RV contraction
• “v” wave
– RV contraction
– Height --- atrial compliance & amount of blood return
• “y” descent
– opening and RA emptying into RV
• “a” wave
– Atrial systole
• “c” wave
– Protrusion of TV into RA
• “x” descent
– Relaxation of RA
– Downward pulling of tricuspid
annulus by RV contraction
25. +
Kern MJ. Right Heart Catheterization. CATHSAP II CD-ROM. Bethesda, American College of Cardiology, 2001.
Right Heart Catheterization
Inspiratory Effect on Right Atrial Pressure
Normal physiology
Inhalation: Intrathoracic pressure falls RA pressure falls
Exhalation: Intrathoracic pressure increases RA pressure
increases
Pressures are best measured at end expiration at which the
pressure is maximum
Exp
26. +
Elevated a wave
Tricuspid stenosis,
Decreased RV compliance ; PS, PAH
Cannon a wave
A-V asynchrony(3rd degree AVB, V-paced)
Absent a wave
Atrial fibrillation
Elevated v wave
SEVERE TR
RV failure
Reduced atrial compliance(restrictive myopathy)
Equal a and v waves
ASD
Abnormalities in RA tracing
27. +
Prominent x descent
– Tamponade
– CP/RCMP
Prominent y descent
– TR
– Constrictive pericarditis
– Restrictive myopathy
Abnormalities in RA tracing
28. +
Right Ventricular Pressure
Systole
Isovolumetric contraction
From TV closure to PV opening
Ejection
From PV opening to PV closure
Diastole
Isovolumetric relaxation
From PV closure to TV opening
Filling
From TV opening to TV closure
Early Rapid Phase
Slow Phase
Atrial Contraction (“a”wave”)
29. +
Normally RA diastolic pressure is equal to RV diastolic
pressure except in early diastole which drives the rapid filling
from atria to ventricles( E wave on ECHO)
An elevated early RA-RV pressure gradient may be seen in
patients with decompensated right heart failure who
characteristically have elevated RA pressure that pushes blood
in RV
Furthermore due to the loss of RV compliance , the diastolic
pressure rises rapidly to a high plateau level
30. +
HENCE RV FAILIURE IS CHARACTERISED BY EARLY
DIASTOLIC DIP AND PLATEAUED HIGH DIASTOLIC
PRESSURE
36. + How to differentiate atrial,
ventricular and arterial tracings
Atrial Presuure Ventricular
Pressure
Arterial Pressure
No of dominanat
waves for every
QRS
2 (A and V) 1 1
Timing of pressure
peak in relation to
ECG
V wave peaks at or
after the end of T
Peaks during ST/T Peaks during ST/T
Shape in diastole Upsloping or
horizontal
Upsloping Downsloping
Presence of A wave Yes Yes No
Dicrotic Notch No No Yes
37. +
Pulmonary Capillary Wedge
Pressure
“a” wave- occurs 240 msec after p wave
on ecg
– Atrial systole
“c” wave
– Protrusion of MV into LA
“x” descent
– Relaxation of LA
– Downward pulling of mitral annulus by LV
contraction
“v” wave
– LV contraction
– Height related to atrial compliance & amount
of blood return – Higher than ‘a’ wave
“y” descent
– MV opening and LA emptying into LV
38. +
Baim DS and Grossman W. Cardiac Catheterization, Angiography, and Intervention. 8th Edition. Baltimore: Williams and
Wilkins, 2014
Physiologic characteristics
Wedge Pressure
Wedge Pressure
Pressure obtained when an end-hole catheter is positioned in
a “designated” blood vessel with its open end-hole facing a
capillary bed, with no connecting vessels conducting flow into
or away from the “designated” blood vessel between the
catheter’s tip and the capillary bed.
True wedge pressure can be measured only in the absence
of flow, allowing pressure to equilibrate across the capillary
bed
39. +
CHARACTERISTICS OF HIGH
QUALITY PCWP
Presence of well defined a and v waves
Appropriate fluoroscopic confirmation of the catheter tip in
distal pulmonary artery and no apparent motion of the catheter
with balloon inflated
Oxygen saturation obtained from the PCWP position greater
than 90%
40. +
Baim DS and Grossman W. Cardiac Catheterization, Angiography, and Intervention. 5th Edition. Baltimore: Williams and
Wilkins, 1996.
Right Heart Catheterization
Left Atrial and PCW Pressure
PCW tracing “approximates” actual LA tracing but is
slightly delayed since pressure wave is transmitted
retrograde through pulmonary veins
PCWP is delayed approximately 50-150 milli sec in
comparison to LA pressure
41. +
PCWP has a smoother contour with less deep X and Y
descents than LA pressure, as the pressure gets damped while
being transmitted from LA through pulmonary capillaries
42. +
1. LARGE V WAVE:
- Acute severe/decompensated MR (not in chronic
compensated MR)
- VSD
- MS
- AF( as only one wave)
Abnormalities of PCWP Tracing
48. +
Right vs Left Ventricular Pressure
Diastolic amplitude similar between RV and LV tracings
Systolic amplitude higher for LV than RV
Duration of systole, isovolumetric contraction, and
isovolumetric relaxation are longer for LV compared to RV
Duration of ejection is shorter for LV than RV.
49. +
Arterial Pressure Monitoring
Aortic waveform varies along length of
the aorta
Systolic wave increases in amplitude
while diastolic wave decreases
Dicrotic notch less apparent in peripheral
tracing
Mean aortic pressure is somewhat
constant
54. +
Spike and dome pattern
Hypertrophic obstructive cardiomyopathy
Abnormalities in Central Aortic
Tracing
55. +
Abnormalities in Central Aortic
Tracing
Dicrotic Pulse
Low stroke volume
Severe LV failure
Hypovolemia
56. +
Topics
① Introduction
② Normal Pressure Tracings
③ Estimation of Cardiac Output
④ Estimation of Valve Area
⑤ Estimation of shunt
57. +Techniques for determination of cardiac output
Fick Oxygen technique
Thermodilution technique
Indicator dilution technique
Angiographic technique
58. +
Technique described by Adolph Fick in 1870
The total uptake or release of any substance by an organ is the product of
blood flow to the organ and arterio-venous oxygen difference of the
substance
If no intracardiac shunt, ie PBF=SBF
Pulmonary blood flow = oxygen consumption/ arteriovenous oxygen
content difference across the lungs
FICKS OXYGEN TECHNIQUE
59. +
AV O2 difference is defined as arterial oxygen content – mixed
venous oxygen content
CO = O2 consumption (ml/min)
-------------------------------------------------------------
arterial O2 content - mixed venous (PA) O2
content
60. +
At low cardiac output states ,greater extraction of oxygen is
present from the tissues and MVO2 is low, resulting in high
AVO2 difference and vice versa
MVO2 alone is a crude estimation of cardiac output states, with
low saturation indicating a low output state and vice versa
The AVO2 content difference is accurately measured by
simultaneously obtaining systemic arterial and mixed venous
saturations.
Ideally arterial sample to be taken from pulmonary veins and
venous from pulmonary artery
61. +
To determine oxygen content , the AVO2 difference is multiplied
by Hb content and oxygen carrying capacity(1.36ml O2/g of
Hb). This is then multiplied by 10 to convert the vaule from g/dl
to g/L
Oxygen Consumption: At a steady state, oxygen consumption
is the rate at which oxygen is taken by the blood from the
lungs.
62. +Methods of measuring oxygen consumption
• Douglas bag method
• Polarographic method
• Paramagnetic method
Assumed Fick method – 125ml/min/m2 for average
indivisuals and 110ml/min/m2 for elderly (Lafarge &
Meittinem)
64. +SOURCES OF ERROR
Use of assumed rather than directly measured oxygen
consumption
Assumes prevalance of steady state
Improper collection of the mixed venous sample
O2 con- error-6%. A-V O2 diff -error -5%
The total error Fick CO– about 10 %
65. +
Merely a specific application of fick’s general principle
“An indicator mixed into a unit volume of constantly flowing blood can
be used to identify that volume of blood in time, provided the
indicator remains in the system between injection and
measurement and mixes completely in the blood”
Indocyanine green is the indicator usually used
Continous infusion & single injection method
It is injected as a bolus into the pulmonary artery and samples
are taken from the peripheral systemic artery
Severe allergic reactions can occur
Indicator dilution methods
66. Amount of dye added = 5 mg
Average dye concentration = 2 mg/L
Therefore the volume that diluted the dye =
5mg/2mg per L = 2.5 L
Time it took to go past = 0.5 min
ie flow rate = 2.5 L /0.5 min = 5 L/min
Concentration
(g/L)
Time (min)
0 0.5
average conc (X) = 2 mg/L
time of passage (t) = 0.5 min
~
67. Flow rate = mass of dye (Q g)
____________________
average dye conc (X g/L) x time of passage (t min)
68. +
THERMODILUTION TECHNIQUE
Fegler described the thermodilution method in 1954.
This method is the easiest to perform and the most widely used
. The thermodilution technique is a variation of the indicator-
dilution technique, using blood temperature as the indicator.
Saline at a known temperature is injected into the right atrium
from the proximal port of a Swan–Ganz catheter.
The saline mixes with blood and lowers its temperature.
.
69. +
The temperature of blood is measured in the pulmonary artery
by a thermistor mounted on the distal tip of the catheter which
is a variable resistor in which the resistance is proportional to
the temperature
As the resistance changes, a change in voltage occurs.
The measured change in voltage over time generates a
temperature curve that is related to the cardiac output.
Similar to the indicator-dilution method, cardiac output is
inversely related to the area under the time–temperature curve.
70. +
If the area under the curve is small, this means the temperature
equilibrates rapidly with the ambient body temperature,
indicating a high cardiac output. Conversely, if the area under
the curve is large, it takes longer for the blood temperature to
reach ambient body temperature, implying low cardiac output.
The thermodilution method is preferable to the indicator-dilution
method because right-sided injection and right-sided sampling
of the cold saline yields a curve that is less subject to
recirculation-induced distortion than right-sided injection and
left-sided sampling of indocyanine dye.
71. +
normal curve
sharp upstroke
smooth curve
mildly prolonged
downslope until baseline
Thermodilution
72. +
area under the curve is inversely
proportional to the flow rate in the
pulmonary artery which equals the
cardiac output in absence of
intracardiac shunt
73. + Advantages
Easy widely available
No withdrawal of blood
No arterial puncture
Inert and inexpensive indicator
No recirculation –analysis simple
Less measurement variability and
correlation with Fick.
Sources of error
Unreliable- TR,PR, INTRA
CARDIAC SHUNTS
Inaccurate in low flow- low output
states (overestimation upto 35%)
74. +
ANGIOGRAPHIC CO
Calculated by tracing end diastolic and end systolic images
SV= EDV-ESV
CO= ( EDV-ESV ) × HR
AF- not accurate
But preferred over fick to measure co in calculation of valve areas in
combined stenotic and regurgitant lesions
Erroneous in RWMA or structurally abnormal ventricles
75. +
METHOD MOST RELIABLE LEAST RELIABLE
FICK METHOD LOW CO HIGH CO
THERMODILUTION HIGH CO PR, TR
INTRACARDIAC
SHUNTS
ANGIORAPHIC Mixed stenotic and
regurgitant lesions
EXTENSIVE RWMA
DILATED VENTRICLE