Nuc chemsummerschool 072106-v2
1. 1
Physics & Instrumentation in
Positron Emission Tomography
Paul Vaska, Ph.D.
Center for Translational Neuroscience
Brookhaven National Laboratory
July 21, 2006
7. 7
ββ++
DecayDecay
+
+
+
+
+
+
+
+
+
Neutron-deficient isotopes can decay by emittingNeutron-deficient isotopes can decay by emitting
positronspositrons
+
anti-neutrinoanti-neutrino
positronpositron Net effect: oneNet effect: one
protonproton
replaced byreplaced by
• neutronneutron
• anti-neutrinoanti-neutrino
• positronpositron
8. 8
Positron annihilation
Annihilation gives
• 2x 511 keV gamma rays
• 180 degrees apart
• Line of response
Positron range & gamma
noncollinearity
Scanner is just a photon
counter!
• Counts gamma-ray pairs vs.
single gammas
• Time window ~ 1 ns
511 keV
511 keV
e+
e-
9. 9
Raw Data & Image Reconstruction
0° projection
0°
180°
90°
90°projection
image
reconstruction
“sinogram”
10. 10
Important Detector PropertiesImportant Detector Properties
- Spatial resolution
- Directly controls spatial resolution
in reconstructed image
- Currently ~ 1 - 5 mm
- Depth-of-interaction?
- Reduces “parallax”
11. 11
Important Detector PropertiesImportant Detector Properties
- Detection efficiency (aka sensitivity, stopping power)
- Reduces noise from counting statistics
- Currently > ~ 30% (singles)
55M Events1M Events
12. 12
Important Detector PropertiesImportant Detector Properties
Random (accidental) coincidence
- Time resolution
- Affects acceptance of random
coincidences
- Currently ~ 1 - 10 ns
- Time-of-flight (TOF)?
- c = ~ 1 ft/ns
- Need << 1 ns resolution
13. 13
Important Detector PropertiesImportant Detector Properties
Scatter and Attenuation
511 keV
- Energy resolution
- Scattered gammas change
direction AND lose energy
- Affects acceptance of scattered
coincidences
- Currently ~ 20%
- Deadtime
- Handle MHz count rates!
511 keV 400 keV
14. 14
Scintillation
Crystal
PMT
Pre-Amplifier
+ Electronics
Gamma photon
converts to optical
photons
(proportional to
gamma energy, typ.
1000’s)
photons are
collected at the end
of the crystal
light is converted to an electrical signal
& amplified
Front-end electronics
condition the signal for
further processing
Prototypical PET Detector
Gamma Ray
Optical reflector
23. 23
RatCAP: Rat Conscious Animal PET
Eliminate anesthesia in preclinical neuroscience
using PET in order to:
• Remove confounding effects of anesthetic on
neurochemistry
• Enable stimulation in animal PET
• Enable correlations of behavior and neuro-PET
24. 24
Architecture
Detector blocks x12
• LSO 2.2 x 2.2 x 5 mm in 4 x 8 array
• 1:1 coupling to APD
• ASIC - single all digital output
Timestamp & Signal Processing Module
• Programmable real-time logic (FPGA)
• 1 ns bins (debugging, now 10 ns)
Data acquisition
• PCI card in standard PC
• Up to 70 MB/s = ~10 Mcps singles
• Offline software for coincidences, corrections, recon, …
TSPM
TDC
PCI
card
ASIC
opticaldifferential
RatCAP