Describes the basic of applanation tonometry, the factors affecting it and also how to perform the ideal tonometry. The slide are borrowed but it gives complete idea of mastering Applanation tonometry.
If the original owner of the slides has an objection i shall take down the ppt with due apologies.
3. FACTORS INFLUENCING IOP
ďźRate of aqueous formation
ďźResistance of aqueous outflow (
drainage)
ďźIncreased episcleral venous pressure
ďźDilatation of pupil
ďźHeredity
ďźAge
ďźSex
ďźDiurnal variation
ďźPostural variation
ďźBlood pressure
ďźOsmotic pressure of blood
4. ⢠The concept was introduced by Goldmann is 1954.
⢠It is based on IMBERT FICKS LAW.
⢠It states that the pressure inside an ideal sphere (P) is equal to force (F) necessary
to flatten its surface divided by the area of the flattening (A).
⢠P can be determined if Force F is fixed or Area A is fixed
P=F/A
5. ⢠The ideal sphere is dry, thin-walled and flexible.
⢠The cornea is not ideal sphere.
Two extra forces acting on cornea -
ďźCapillary attraction of tear meniscus (T), tends to pull tonometer towards cornea
ďźCorneal rigidity (C) resists flattening
Thus, Modified Imbert Ficks Law
F = PA , becomes
F + T = PA + C , or
P =( F + T - C) / A
6. ⢠These two forces cancel each other
when flattened area has diameter of
3.06 mm.
⢠Applanation tonometry displaces
only about 0.5 microlitre of aqueous
humor, which raises IOP by about
3%. Because the volume displaced is
so small, ocular rigidity, or the
âstretchabilityâ of the globe, has little
effect on the pressure readings.
7.
8.
9.
10.
11. PROCEDURE
⢠Patient is asked not to drink alcoholic beverages or large amounts of fluid (e.g., 500 ml or more)
for 2 hours before the test, as the former will lower IOP and the latter may raise it.
⢠Patient is told the purpose of the test and is reassured that the measurement is not painful.
⢠The patient is instructed to relax, maintain position, and hold the eyes open wide.
⢠One drop of a topical anesthetic, such as 0.5% proparacaine, is placed in each eye, and the tip of a
moistened fluorescein strip is touched to the tear layer on the inner surface of each lower lid.
⢠Tonometer tip is cleaned with a sterilizing solution, and the tip and prism are set in correct
position on the slit lamp.
⢠Tension knob is set at 1g. If the knob is set at 0, the prism head may vibrate when it touches the
eye and damage the corneal epithelium.
12. ⢠The 1 g position is used before each measurement. As a rule, it is more accurate to measure IOP
by increasing rather than decreasing the force of applanation.
⢠0 graduation mark of the prism is set at the white line on the prism holder.
⢠Cobalt blue filter is used with the slit beam opened maximally. The angle between the
illumination and the microscope should be
⢠approximately 60°.
⢠The room illumination is reduced.
13. ⢠Heights of the slit lamp, chair, and chin rest are adjusted until the patient is comfortable and in
the correct position for the measurement.
⢠Palpebral fissure is a little wider if the patient looks up. However, the gaze should be no more
than 15° above the horizontal to prevent an elevation of IOP that is especially marked in the
presence of restrictive neuromuscular disease such as dysthyroid ophthalmopathy.
⢠Operator sits opposite the patient, the assembly is advanced towards the patient with the tester
observing from the side until the limbal zone has a bluish hue.
⢠If the tonometer tip touches the lids, the fluorescein rings will thicken, which may cause an
overestimation of IOP.
⢠Clinician observes the applanation through the biprism at low power. A monocular view is
obtained of the central applanated zone and the surrounding fluorescein-stained tear film.
14. ⢠Using the control stick, the observer raises, lowers, and centers the assembly until two equal
semicircles are seen in the center of the field of view.
⢠If the two semicircles are not equal in size, IOP is overestimated. The clinician turns the tension knob
in both directions to ensure that the instrument is in good position.
⢠If the semicircles cannot be made âtoo small,â the instrument is too far forward. If the semicircles
cannot be made âtoo large,â the instrument is too far from the eye.
⢠Fluorescein rings should be approximately 0.25â0.3mm in thickness â or about one-tenth the diameter
of the flattened area. If the rings are too narrow, the patient should blink two or three times to replenish
the fluorescein.
⢠If the fluorescein rings are too narrow, IOP is underestimated. The tension knob is rotated until the
inner borders of the fluorescein rings touch each other at the midpoint of their pulsations.
⢠Intraocular pressure is measured in the right eye until three successive readings are within 1 mmHg.
Intraocular pressure is then measured in the left eye.
⢠Reading obtained in grams is multiplied by 10 to give the IOP in millimeters of mercury.
15. ⢠The fluorescein rings should be pproximately
0.25â0.3 mm in thickness â or about one-
tenth the diameter of the flattened area.
⢠The fluorescent semicircles are viewed
through the biprism and the force against the
cornea is adjusted until the inner edges
overlap.
16. FALSELY LOW IOP FALSELY HIGH IOP
Too little fluorescein Too much fluorescein
Thin cornea Thick cornea
Corneal edema Steep cornea
With the rule astigmatism - 1mm Hg per
4D
Against the rule astigmatism - 1 m Hg
per 3D
Repeated tonometry Wider meniscus, Widening the lid fissure
Excessively, Elevating the eyes more than
150
17. ⢠For every 3D increase in corneal curvature, IOP raises about 1 mm Hg as more fluid is
displaced under steeper corneas causing increase in ocular rigidity.
⢠More than 6 D astigmatism produces an elliptical area on applanation that gives erroneous IOP.
4D with-the rule astigmatism underestimate IOP and 4D against-the rule astigmatism
overestimate IOP.
⢠Goldmann applanation tonometer was designed to give accurate readings when the
CCT was 550 Îźm.
⢠The deviation of CCT from 550 Οm yields a change in applanation readings of 0.7
mm Hg per 10 Îźm.
18.
19.
20.
21.
22.
23.
24.
25.
26. ⢠Biprism should be rinsed and dried immediately after use.
⢠Between uses, the prism head should be soaked in a solution such as diluted bleach or 3%
hydrogen peroxide.
⢠70% ethanol and 70% isopropanol are effective as sterilizing solutions but were shown in one
study to cause mild damage to the tonometer tip after one month of immersion.
⢠Other methods of sterilization include: 10 min of rinsing in running tap water, wash with soap
and water, cover the tip with a disposable film, and exposure to UV light.
⢠It is possible to transfer bacteria, viruses, and other infectious agents with the tonometer head,
including such potentially serious infections as epidemic keratoconjunctivitis, hepatitis B, Jacob-
Kreutzfeld and, theoretically, acquired immunodeficiency syndrome
27. ⢠Care must be taken to be sure any sterilizing solution has been completely rinsed off the
tonometer tip, as some of these solutions may be toxic to the corneal epithelium, especially after
LASIK or other corneal procedures.
⢠If the tonometer tip is not mechanically wiped after each use, epithelial cells may stick to the tip
with the small but serious risk of transmitting Jacob-Kreutzfeld virus.
⢠Disposable tonometer tips may be an acceptable alternative to soaking in, and wiping with,
antiseptic solutions.