SlideShare a Scribd company logo
1 of 29
BASICS…
FACTORS INFLUENCING IOP
Rate of aqueous formation
Resistance of aqueous outflow (
drainage)
Increased episcleral venous pressure
Dilatation of pupil
Heredity
Age
Sex
Diurnal variation
Postural variation
Blood pressure
Osmotic pressure of blood
• The concept was introduced by Goldmann is 1954.
• It is based on IMBERT FICKS LAW.
• It states that the pressure inside an ideal sphere (P) is equal to force (F) necessary
to flatten its surface divided by the area of the flattening (A).
• P can be determined if Force F is fixed or Area A is fixed
P=F/A
• The ideal sphere is dry, thin-walled and flexible.
• The cornea is not ideal sphere.
Two extra forces acting on cornea -
Capillary attraction of tear meniscus (T), tends to pull tonometer towards cornea
Corneal rigidity (C) resists flattening
Thus, Modified Imbert Ficks Law
F = PA , becomes
F + T = PA + C , or
P =( F + T - C) / A
• These two forces cancel each other
when flattened area has diameter of
3.06 mm.
• Applanation tonometry displaces
only about 0.5 microlitre of aqueous
humor, which raises IOP by about
3%. Because the volume displaced is
so small, ocular rigidity, or the
‘stretchability’ of the globe, has little
effect on the pressure readings.
PROCEDURE
• Patient is asked not to drink alcoholic beverages or large amounts of fluid (e.g., 500 ml or more)
for 2 hours before the test, as the former will lower IOP and the latter may raise it.
• Patient is told the purpose of the test and is reassured that the measurement is not painful.
• The patient is instructed to relax, maintain position, and hold the eyes open wide.
• One drop of a topical anesthetic, such as 0.5% proparacaine, is placed in each eye, and the tip of a
moistened fluorescein strip is touched to the tear layer on the inner surface of each lower lid.
• Tonometer tip is cleaned with a sterilizing solution, and the tip and prism are set in correct
position on the slit lamp.
• Tension knob is set at 1g. If the knob is set at 0, the prism head may vibrate when it touches the
eye and damage the corneal epithelium.
• The 1 g position is used before each measurement. As a rule, it is more accurate to measure IOP
by increasing rather than decreasing the force of applanation.
• 0 graduation mark of the prism is set at the white line on the prism holder.
• Cobalt blue filter is used with the slit beam opened maximally. The angle between the
illumination and the microscope should be
• approximately 60°.
• The room illumination is reduced.
• Heights of the slit lamp, chair, and chin rest are adjusted until the patient is comfortable and in
the correct position for the measurement.
• Palpebral fissure is a little wider if the patient looks up. However, the gaze should be no more
than 15° above the horizontal to prevent an elevation of IOP that is especially marked in the
presence of restrictive neuromuscular disease such as dysthyroid ophthalmopathy.
• Operator sits opposite the patient, the assembly is advanced towards the patient with the tester
observing from the side until the limbal zone has a bluish hue.
• If the tonometer tip touches the lids, the fluorescein rings will thicken, which may cause an
overestimation of IOP.
• Clinician observes the applanation through the biprism at low power. A monocular view is
obtained of the central applanated zone and the surrounding fluorescein-stained tear film.
• Using the control stick, the observer raises, lowers, and centers the assembly until two equal
semicircles are seen in the center of the field of view.
• If the two semicircles are not equal in size, IOP is overestimated. The clinician turns the tension knob
in both directions to ensure that the instrument is in good position.
• If the semicircles cannot be made ‘too small,’ the instrument is too far forward. If the semicircles
cannot be made ‘too large,’ the instrument is too far from the eye.
• Fluorescein rings should be approximately 0.25–0.3mm in thickness – or about one-tenth the diameter
of the flattened area. If the rings are too narrow, the patient should blink two or three times to replenish
the fluorescein.
• If the fluorescein rings are too narrow, IOP is underestimated. The tension knob is rotated until the
inner borders of the fluorescein rings touch each other at the midpoint of their pulsations.
• Intraocular pressure is measured in the right eye until three successive readings are within 1 mmHg.
Intraocular pressure is then measured in the left eye.
• Reading obtained in grams is multiplied by 10 to give the IOP in millimeters of mercury.
• The fluorescein rings should be pproximately
0.25–0.3 mm in thickness – or about one-
tenth the diameter of the flattened area.
• The fluorescent semicircles are viewed
through the biprism and the force against the
cornea is adjusted until the inner edges
overlap.
FALSELY LOW IOP FALSELY HIGH IOP
Too little fluorescein Too much fluorescein
Thin cornea Thick cornea
Corneal edema Steep cornea
With the rule astigmatism - 1mm Hg per
4D
Against the rule astigmatism - 1 m Hg
per 3D
Repeated tonometry Wider meniscus, Widening the lid fissure
Excessively, Elevating the eyes more than
150
• For every 3D increase in corneal curvature, IOP raises about 1 mm Hg as more fluid is
displaced under steeper corneas causing increase in ocular rigidity.
• More than 6 D astigmatism produces an elliptical area on applanation that gives erroneous IOP.
4D with-the rule astigmatism underestimate IOP and 4D against-the rule astigmatism
overestimate IOP.
• Goldmann applanation tonometer was designed to give accurate readings when the
CCT was 550 Îźm.
• The deviation of CCT from 550 μm yields a change in applanation readings of 0.7
mm Hg per 10 Îźm.
• Biprism should be rinsed and dried immediately after use.
• Between uses, the prism head should be soaked in a solution such as diluted bleach or 3%
hydrogen peroxide.
• 70% ethanol and 70% isopropanol are effective as sterilizing solutions but were shown in one
study to cause mild damage to the tonometer tip after one month of immersion.
• Other methods of sterilization include: 10 min of rinsing in running tap water, wash with soap
and water, cover the tip with a disposable film, and exposure to UV light.
• It is possible to transfer bacteria, viruses, and other infectious agents with the tonometer head,
including such potentially serious infections as epidemic keratoconjunctivitis, hepatitis B, Jacob-
Kreutzfeld and, theoretically, acquired immunodeficiency syndrome
• Care must be taken to be sure any sterilizing solution has been completely rinsed off the
tonometer tip, as some of these solutions may be toxic to the corneal epithelium, especially after
LASIK or other corneal procedures.
• If the tonometer tip is not mechanically wiped after each use, epithelial cells may stick to the tip
with the small but serious risk of transmitting Jacob-Kreutzfeld virus.
• Disposable tonometer tips may be an acceptable alternative to soaking in, and wiping with,
antiseptic solutions.
PERKIN’S TONOMETER
Applanation tonometer

More Related Content

What's hot

Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.Dr Samarth Mishra
 
Amsler's grid test
Amsler's grid testAmsler's grid test
Amsler's grid testRASHAD MUHAMMED
 
Soft toric Contact Lens
Soft toric Contact LensSoft toric Contact Lens
Soft toric Contact LensManish Dahal
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopyHira Dahal
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondenceRajeshwori
 
Slit lamp (methods of illumination)
Slit lamp (methods of illumination)Slit lamp (methods of illumination)
Slit lamp (methods of illumination)maclester manahan
 
Pediatric contact lens
Pediatric contact lensPediatric contact lens
Pediatric contact lensNoor Munirah Aab
 
Measurements of the optical constants of the eye
Measurements  of  the optical constants of the eyeMeasurements  of  the optical constants of the eye
Measurements of the optical constants of the eyeDIVYAMURUGESAN8
 
Clinical examination of squint
Clinical examination of squintClinical examination of squint
Clinical examination of squintReshma Peter
 
Maddox rod and double maddox rod
Maddox rod and double maddox rodMaddox rod and double maddox rod
Maddox rod and double maddox rodAnuMusyakhwo7
 
corneal Pachymetry
 corneal Pachymetry corneal Pachymetry
corneal PachymetryKavita Kumari
 
Fundus Fluoroscein Angiography
Fundus Fluoroscein AngiographyFundus Fluoroscein Angiography
Fundus Fluoroscein AngiographyRashmi Ranjan
 
Non contact tonometer ppt
Non contact tonometer pptNon contact tonometer ppt
Non contact tonometer pptsatish kumar
 

What's hot (20)

Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.
 
Keratometer
KeratometerKeratometer
Keratometer
 
Amsler's grid test
Amsler's grid testAmsler's grid test
Amsler's grid test
 
Soft toric Contact Lens
Soft toric Contact LensSoft toric Contact Lens
Soft toric Contact Lens
 
Biometry
BiometryBiometry
Biometry
 
Jackson cross cylinder
Jackson cross cylinderJackson cross cylinder
Jackson cross cylinder
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Keratometry
KeratometryKeratometry
Keratometry
 
B SCAN
B SCAN B SCAN
B SCAN
 
Pediatric refraction
Pediatric       refractionPediatric       refraction
Pediatric refraction
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondence
 
Slit lamp (methods of illumination)
Slit lamp (methods of illumination)Slit lamp (methods of illumination)
Slit lamp (methods of illumination)
 
Pediatric contact lens
Pediatric contact lensPediatric contact lens
Pediatric contact lens
 
Measurements of the optical constants of the eye
Measurements  of  the optical constants of the eyeMeasurements  of  the optical constants of the eye
Measurements of the optical constants of the eye
 
Clinical examination of squint
Clinical examination of squintClinical examination of squint
Clinical examination of squint
 
B scan
B scanB scan
B scan
 
Maddox rod and double maddox rod
Maddox rod and double maddox rodMaddox rod and double maddox rod
Maddox rod and double maddox rod
 
corneal Pachymetry
 corneal Pachymetry corneal Pachymetry
corneal Pachymetry
 
Fundus Fluoroscein Angiography
Fundus Fluoroscein AngiographyFundus Fluoroscein Angiography
Fundus Fluoroscein Angiography
 
Non contact tonometer ppt
Non contact tonometer pptNon contact tonometer ppt
Non contact tonometer ppt
 

Similar to Applanation tonometer

Similar to Applanation tonometer (20)

Tonometry.pptx
Tonometry.pptxTonometry.pptx
Tonometry.pptx
 
Goldmanntonometry.ppt
Goldmanntonometry.pptGoldmanntonometry.ppt
Goldmanntonometry.ppt
 
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTXTONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
 
Tonometry
TonometryTonometry
Tonometry
 
Tonometer
TonometerTonometer
Tonometer
 
Tonometry
TonometryTonometry
Tonometry
 
INTRAOCULAR PRESSURE .pdf
INTRAOCULAR PRESSURE                  .pdfINTRAOCULAR PRESSURE                  .pdf
INTRAOCULAR PRESSURE .pdf
 
INTRAOCULAR PRESSURE.pdf
INTRAOCULAR                    PRESSURE.pdfINTRAOCULAR                    PRESSURE.pdf
INTRAOCULAR PRESSURE.pdf
 
Goldman Applanation Tonometer
Goldman Applanation TonometerGoldman Applanation Tonometer
Goldman Applanation Tonometer
 
08 GAT
08 GAT08 GAT
08 GAT
 
Tonometry
TonometryTonometry
Tonometry
 
Tonometry by dr. mamta meena
Tonometry by dr. mamta meenaTonometry by dr. mamta meena
Tonometry by dr. mamta meena
 
INTRAOCULAR PRESSURE &TONOMETRY Dr. ZAW MIN HTET (OPHTHALMOLOGIST)
INTRAOCULAR PRESSURE &TONOMETRY Dr. ZAW MIN HTET (OPHTHALMOLOGIST)INTRAOCULAR PRESSURE &TONOMETRY Dr. ZAW MIN HTET (OPHTHALMOLOGIST)
INTRAOCULAR PRESSURE &TONOMETRY Dr. ZAW MIN HTET (OPHTHALMOLOGIST)
 
tonometer.pptx
tonometer.pptxtonometer.pptx
tonometer.pptx
 
Gonioscopy
Gonioscopy Gonioscopy
Gonioscopy
 
Tonometer ppt
Tonometer pptTonometer ppt
Tonometer ppt
 
TONOMETRS---11.pptx
TONOMETRS---11.pptxTONOMETRS---11.pptx
TONOMETRS---11.pptx
 
Means of IOP measurement
Means of IOP measurementMeans of IOP measurement
Means of IOP measurement
 
Tonometry and tonography
Tonometry and tonographyTonometry and tonography
Tonometry and tonography
 
Tonometry Procedure
Tonometry ProcedureTonometry Procedure
Tonometry Procedure
 

More from drindeevarmishra

Evaluation of squint - The Basics
Evaluation of squint - The BasicsEvaluation of squint - The Basics
Evaluation of squint - The Basicsdrindeevarmishra
 
Lasers in glaucoma - A Brief review
Lasers in glaucoma - A Brief reviewLasers in glaucoma - A Brief review
Lasers in glaucoma - A Brief reviewdrindeevarmishra
 
Neuroprotection in glaucoma - A brief review
Neuroprotection in glaucoma - A brief reviewNeuroprotection in glaucoma - A brief review
Neuroprotection in glaucoma - A brief reviewdrindeevarmishra
 
Managing bleb failure - A brief guide
Managing bleb failure - A brief guideManaging bleb failure - A brief guide
Managing bleb failure - A brief guidedrindeevarmishra
 
Automated-perimetry - Brief review and basics
Automated-perimetry - Brief review and basicsAutomated-perimetry - Brief review and basics
Automated-perimetry - Brief review and basicsdrindeevarmishra
 
Attack on doctors - how to prevent it
Attack on doctors - how to prevent it Attack on doctors - how to prevent it
Attack on doctors - how to prevent it drindeevarmishra
 
Occult Posterior capsular rent
Occult Posterior capsular rentOccult Posterior capsular rent
Occult Posterior capsular rentdrindeevarmishra
 
Intermittent exotropia
Intermittent exotropiaIntermittent exotropia
Intermittent exotropiadrindeevarmishra
 
REview of Phacoemulsification
REview of PhacoemulsificationREview of Phacoemulsification
REview of Phacoemulsificationdrindeevarmishra
 
Anatomy of eye and its clinical relevance in Squint
Anatomy of eye and its clinical relevance in SquintAnatomy of eye and its clinical relevance in Squint
Anatomy of eye and its clinical relevance in Squintdrindeevarmishra
 

More from drindeevarmishra (10)

Evaluation of squint - The Basics
Evaluation of squint - The BasicsEvaluation of squint - The Basics
Evaluation of squint - The Basics
 
Lasers in glaucoma - A Brief review
Lasers in glaucoma - A Brief reviewLasers in glaucoma - A Brief review
Lasers in glaucoma - A Brief review
 
Neuroprotection in glaucoma - A brief review
Neuroprotection in glaucoma - A brief reviewNeuroprotection in glaucoma - A brief review
Neuroprotection in glaucoma - A brief review
 
Managing bleb failure - A brief guide
Managing bleb failure - A brief guideManaging bleb failure - A brief guide
Managing bleb failure - A brief guide
 
Automated-perimetry - Brief review and basics
Automated-perimetry - Brief review and basicsAutomated-perimetry - Brief review and basics
Automated-perimetry - Brief review and basics
 
Attack on doctors - how to prevent it
Attack on doctors - how to prevent it Attack on doctors - how to prevent it
Attack on doctors - how to prevent it
 
Occult Posterior capsular rent
Occult Posterior capsular rentOccult Posterior capsular rent
Occult Posterior capsular rent
 
Intermittent exotropia
Intermittent exotropiaIntermittent exotropia
Intermittent exotropia
 
REview of Phacoemulsification
REview of PhacoemulsificationREview of Phacoemulsification
REview of Phacoemulsification
 
Anatomy of eye and its clinical relevance in Squint
Anatomy of eye and its clinical relevance in SquintAnatomy of eye and its clinical relevance in Squint
Anatomy of eye and its clinical relevance in Squint
 

Recently uploaded

Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowNehru place Escorts
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Call Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service IndoreCall Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service IndoreRiya Pathan
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipurparulsinha
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 

Recently uploaded (20)

Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Call Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service IndoreCall Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service Indore
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 

Applanation tonometer

  • 1.
  • 3. FACTORS INFLUENCING IOP Rate of aqueous formation Resistance of aqueous outflow ( drainage) Increased episcleral venous pressure Dilatation of pupil Heredity Age Sex Diurnal variation Postural variation Blood pressure Osmotic pressure of blood
  • 4. • The concept was introduced by Goldmann is 1954. • It is based on IMBERT FICKS LAW. • It states that the pressure inside an ideal sphere (P) is equal to force (F) necessary to flatten its surface divided by the area of the flattening (A). • P can be determined if Force F is fixed or Area A is fixed P=F/A
  • 5. • The ideal sphere is dry, thin-walled and flexible. • The cornea is not ideal sphere. Two extra forces acting on cornea - Capillary attraction of tear meniscus (T), tends to pull tonometer towards cornea Corneal rigidity (C) resists flattening Thus, Modified Imbert Ficks Law F = PA , becomes F + T = PA + C , or P =( F + T - C) / A
  • 6. • These two forces cancel each other when flattened area has diameter of 3.06 mm. • Applanation tonometry displaces only about 0.5 microlitre of aqueous humor, which raises IOP by about 3%. Because the volume displaced is so small, ocular rigidity, or the ‘stretchability’ of the globe, has little effect on the pressure readings.
  • 7.
  • 8.
  • 9.
  • 10.
  • 11. PROCEDURE • Patient is asked not to drink alcoholic beverages or large amounts of fluid (e.g., 500 ml or more) for 2 hours before the test, as the former will lower IOP and the latter may raise it. • Patient is told the purpose of the test and is reassured that the measurement is not painful. • The patient is instructed to relax, maintain position, and hold the eyes open wide. • One drop of a topical anesthetic, such as 0.5% proparacaine, is placed in each eye, and the tip of a moistened fluorescein strip is touched to the tear layer on the inner surface of each lower lid. • Tonometer tip is cleaned with a sterilizing solution, and the tip and prism are set in correct position on the slit lamp. • Tension knob is set at 1g. If the knob is set at 0, the prism head may vibrate when it touches the eye and damage the corneal epithelium.
  • 12. • The 1 g position is used before each measurement. As a rule, it is more accurate to measure IOP by increasing rather than decreasing the force of applanation. • 0 graduation mark of the prism is set at the white line on the prism holder. • Cobalt blue filter is used with the slit beam opened maximally. The angle between the illumination and the microscope should be • approximately 60°. • The room illumination is reduced.
  • 13. • Heights of the slit lamp, chair, and chin rest are adjusted until the patient is comfortable and in the correct position for the measurement. • Palpebral fissure is a little wider if the patient looks up. However, the gaze should be no more than 15° above the horizontal to prevent an elevation of IOP that is especially marked in the presence of restrictive neuromuscular disease such as dysthyroid ophthalmopathy. • Operator sits opposite the patient, the assembly is advanced towards the patient with the tester observing from the side until the limbal zone has a bluish hue. • If the tonometer tip touches the lids, the fluorescein rings will thicken, which may cause an overestimation of IOP. • Clinician observes the applanation through the biprism at low power. A monocular view is obtained of the central applanated zone and the surrounding fluorescein-stained tear film.
  • 14. • Using the control stick, the observer raises, lowers, and centers the assembly until two equal semicircles are seen in the center of the field of view. • If the two semicircles are not equal in size, IOP is overestimated. The clinician turns the tension knob in both directions to ensure that the instrument is in good position. • If the semicircles cannot be made ‘too small,’ the instrument is too far forward. If the semicircles cannot be made ‘too large,’ the instrument is too far from the eye. • Fluorescein rings should be approximately 0.25–0.3mm in thickness – or about one-tenth the diameter of the flattened area. If the rings are too narrow, the patient should blink two or three times to replenish the fluorescein. • If the fluorescein rings are too narrow, IOP is underestimated. The tension knob is rotated until the inner borders of the fluorescein rings touch each other at the midpoint of their pulsations. • Intraocular pressure is measured in the right eye until three successive readings are within 1 mmHg. Intraocular pressure is then measured in the left eye. • Reading obtained in grams is multiplied by 10 to give the IOP in millimeters of mercury.
  • 15. • The fluorescein rings should be pproximately 0.25–0.3 mm in thickness – or about one- tenth the diameter of the flattened area. • The fluorescent semicircles are viewed through the biprism and the force against the cornea is adjusted until the inner edges overlap.
  • 16. FALSELY LOW IOP FALSELY HIGH IOP Too little fluorescein Too much fluorescein Thin cornea Thick cornea Corneal edema Steep cornea With the rule astigmatism - 1mm Hg per 4D Against the rule astigmatism - 1 m Hg per 3D Repeated tonometry Wider meniscus, Widening the lid fissure Excessively, Elevating the eyes more than 150
  • 17. • For every 3D increase in corneal curvature, IOP raises about 1 mm Hg as more fluid is displaced under steeper corneas causing increase in ocular rigidity. • More than 6 D astigmatism produces an elliptical area on applanation that gives erroneous IOP. 4D with-the rule astigmatism underestimate IOP and 4D against-the rule astigmatism overestimate IOP. • Goldmann applanation tonometer was designed to give accurate readings when the CCT was 550 Îźm. • The deviation of CCT from 550 Îźm yields a change in applanation readings of 0.7 mm Hg per 10 Îźm.
  • 18.
  • 19.
  • 20.
  • 21.
  • 22.
  • 23.
  • 24.
  • 25.
  • 26. • Biprism should be rinsed and dried immediately after use. • Between uses, the prism head should be soaked in a solution such as diluted bleach or 3% hydrogen peroxide. • 70% ethanol and 70% isopropanol are effective as sterilizing solutions but were shown in one study to cause mild damage to the tonometer tip after one month of immersion. • Other methods of sterilization include: 10 min of rinsing in running tap water, wash with soap and water, cover the tip with a disposable film, and exposure to UV light. • It is possible to transfer bacteria, viruses, and other infectious agents with the tonometer head, including such potentially serious infections as epidemic keratoconjunctivitis, hepatitis B, Jacob- Kreutzfeld and, theoretically, acquired immunodeficiency syndrome
  • 27. • Care must be taken to be sure any sterilizing solution has been completely rinsed off the tonometer tip, as some of these solutions may be toxic to the corneal epithelium, especially after LASIK or other corneal procedures. • If the tonometer tip is not mechanically wiped after each use, epithelial cells may stick to the tip with the small but serious risk of transmitting Jacob-Kreutzfeld virus. • Disposable tonometer tips may be an acceptable alternative to soaking in, and wiping with, antiseptic solutions.